Dave van Woensel 08 June Installation effects in ultrasonic gas flow meters Calibration results of a Z-configuration measurement skid
|
|
- Cory Underwood
- 5 years ago
- Views:
Transcription
1 Dave van Woensel 08 June 2017 Installation effects in ultrasonic gas flow meters Calibration results of a Z-configuration measurement skid
2 1. Installation Effects 2. Skid Calibration 3. Results Ultrasonic flowmeter for Custody Transfer Agenda
3 Customer request Will the meter measure as accurate as calibrated once assembled on the skid? 3
4 Uncertainty Installation effects 2,00% 0,50% 1,50% 0,40% (Initial) Base 15 bar 1,00% 0,30% 0,20% 0,10% 0,00% -0,10% -0,20% -0,30% -0,40% Base 15 bar FC Base 10 bar Diameter step +3% 0,50% Diameter step -3% 0,00% Elbow 10D 0deg Elbow 10D 90deg 0% 20% 40% 60% 80% 100% -0,50% OOP 10D 0deg OOP 10D 90deg OOP Exp. 10D 0deg -1,00% Base 15 bar REVERSE -1,50% Renewed base 15 bar Elbow 10D 0deg -2,00% -0,50% Renewed base 15 bar repeat flow rate (Qmax = m3/h) 4
5 Swirl 5
6 Swirl and flow conditioning Computational Fluid Dynamics (CFD) 6
7 Remaining swirl after flow conditioner Most swirl has been removed 7
8 Swirl causing meter to overread 8
9 Swirl causing meter to underread 9
10 Swirl compensation in plane 10
11 Multiple paths for flow profile recognition 11
12 How standards describe installation effects Calibrate meter under ideal conditions Redo calibration with upstream disturbance, manufacturer to advise required upstream piping to comply to standard and class Allowed deviation depends on standard and accuracy class Qmin Qt Qmax 12
13 Allowed deviations due to installation effects OIML R137 class 1 OIML R137 class 0.5 AGA 9 ISO ± 0.33% (1/3th of class) ± 0.17% (1/3th of class) Perturbation testing is done during one-off type test Uncertainty due to installation effect comes on top of regular calibration result (installation effects are not seen during a regular calibration) OIML certificates are issued by an independent 3 rd party (e.g. NMi, PTB). AGA9 and ISO compliance is not checked by 3 rd party. ± 0.3% ± 0.3% 13
14 Example of perturbation tests ISO and OIML R137-1 requirement shown below 80D 80D 0 D 0D SB Re/Ex DBooP Dboop/Ex DBooP/Ex/HMP 14
15 1. Installation Effects 2. Skid Calibration 3. Results Ultrasonic flowmeter for Custody Transfer Agenda
16 Customer concern Will the meter measure as accurate as calibrated once assembled on the skid? The only way to find out is to calibrate the entire skid 16
17 Flow measurement skid required by customer 17
18 Flow measurement skid Normal operation Verification 18
19 Flow measurement skid Maintenance for duty meter Extra capacity 19
20 Force Technology small calibration loop Flowrates 8-10,000 m3/h Calibration pressure 0-50 barg Lab uncertainty
21 Skid calibration (2x 8 /1500#) 21
22 Skid calibrated at FORCE (Denmark) 22
23 Skid calibration (2x 8 /1500#) Meter B Meter A Calibration pressure 50 bar Flow rates loose meter calibration 62 m3/h (0,8 m/s) 2500 m3/h (31,3 m/s) Flow rates skid calibration 62 m3/h (0,8 m/s) 1875 m3/h (23,4 m/s) 23
24 1. Installation Effects 2. Skid Calibration 3. Results Ultrasonic flowmeter for Custody Transfer Agenda
25 Calibration of loose meters 5D 3D 25
26 Flow conditioner and upstream piping KROHNE Flowcon3 (modified Spearmann plate), position locked by bracket Inlet pipe permanently affixed to flow meter This is not normal procedure but specifically done for this project 26
27 Flow profiles at different flow rates Meter A, loose meter calibration 27
28 Flow profiles at different flow rates Meter A, loose meter calibration 28
29 Flow profiles at different flow rates Meter B, loose meter calibration 29
30 Modified flow conditioner to minimize noise production Flow conditioner of meter B produced ultrasonic noise at flowrates 2200 m3/h. Flow conditioner modified to reduce noise production. After modification less noise and only starting at 2450 m3/h. Modifications caused flow profile of meter B to be slightly asymmetric (combination with upstream conditions). Noise production by flow conditioner occasionally happens and is a typical case of bad luck. 30
31 As left calibration results meter A and B Baseline after linearization 31
32 Calibration of meter A in skid B A Inlet conditions: - 3x 90 bend, 1x T-section, 1x 90 bend (all swirl generators) - 2 fully open ball valves, 6D straight - Flow conditioner, 5D straight (used in original meter calibration) 32
33 Flow profile meter A Loose meter vs meter in skid at 1500 m3/h 33
34 Calibration results meter A Loose meter vs meter in skid AGA 9/ ISO perturbation limits OIML R137 class 0.5 perturbation limits 34
35 Calibration of meter B in skid B A Inlet conditions meter B: - 3x 90 bend, 1x T-section, 1x 90 bend (all swirl generators) - 2 fully open ball valves, 6D straight - Flow conditioner, 5D straight (used in original meter calibration) 35
36 Flow profile meter B Loose meter vs meter in skid at 1500 m3/h 36
37 Calibration results meter B Loose meter vs meter in skid AGA 9/ ISO perturbation limits OIML R137 class 0.5 perturbation limits 37
38 Calibration of meter A and B in skid Meter B Meter A Inlet conditions meter A: - 3x 90 bend, 1x T-section, 1x 90 bend (all swirl generators) - 2 fully open ball valves, 6D straight - Flow conditioner, 5D straight (used in original meter calibration) 38
39 Flow profile meter A at 1500 m3/h Loose meter vs meter in skid in Z-configuration 39
40 Calibration results meter A Skid in parallel vs Z AGA 9/ ISO perturbation limits OIML R137 class 0.5 perturbation limits 40
41 Calibration of meter A and B in skid B A Inlet conditions meter B: - 3x 90 bend, 1x T-section, 1x 90 bend (all swirl generators) - 2 fully open ball valves, 6D straight, flow conditioner, 15D straight - 1x T-section, 2x 90 bend, 1x T-section, 1D straight - Flow conditioner, 5D straight (used in original meter calibration) 41
42 Flow profile meter B at 1500 m3/h Loose meter vs meter in skid in Z-configuration 42
43 Calibration results meter B Skid in parallel vs Z AGA 9/ ISO perturbation limits OIML R137 class 0.5 perturbation limits 43
44 Summary and conclusions Due to overall length restrictions meter B in Z-mode only had 1D straight inlet (settling chamber) upstream of the flow conditioner, less then the required 2D. Despite the 1D settling chamber and significant perturbations all shift where well within AGA9 and ISO perturbation limits of ± 0.3%. All results, except the highest flow rate of meter B in Z-mode, were within the OIML R137 class 0.5 perturbation limits of ± 0.17%. The modulations based on the log files afterwards indicate a 1.4D settling chamber was required. So it is expected that with a 2D settling chamber also meter B in Z-mode would have been within the severe limits. Different flow profiles do not automatically mean bad results, if you can recognize them. 44
45 Summary and conclusions It is possible to calibrate a complete skid but even for a small skid is it a lot of work. Especially if the recalibration periods are short in the laws of the country. Z-configuration skids are perfectly possible with ultrasonic flowmeters. Calibrating result of meters remains valid when meters are placed in a skid, this means that there is no need to (re-)calibrate entire skids. Loose meters including upstream conditions seem sufficient. It would be interesting to do the same test without flow conditioners to see if results would be similar. But we would need a new customer with a special request! 45 To be continued
46 Thank you for your attention!
1. Introduction. 2. API 5.7 Testing Protocol for Differential Pressure Flow Measurement Devices
Testing the Wafer V-Cone Flowmeters in accordance with API 5.7 Testing Protocol for Differential Pressure Flow Measurement Devices in the CEESI Colorado Test Facility Dr R.J.W.Peters Flow Measurement Technology
More informationORIFICE FLOW METERS WITH CHAMBERS MEASURING RINGS
Bulletin E79 rev0 02/03/03 ORIFICE FLOW METERS WITH CHAMBERS MEASURING RINGS POP-U-S SERIES FEATURES UNI PN6 flanges: iron Chamber measuring rings: iron Disc: AISI30 Max. operating temperature: 00 C Max.
More informationEXPERIMENTAL INVESTIGATION OF STROUHAL S NUMBER LINEARITY FOR DUAL TRAPEZOIDAL BLUFF BODY VORTEX FLOWMETER
ISSN: 244-787 EXPERIMENTAL INVESTIGATION OF STROUHAL S NUMBER LINEARITY FOR DUAL TRAPEZOIDAL BLUFF BODY VORTEX FLOWMETER ATUL S. BACHAL Mechanical Engineering Department Government College of Engineering
More informationIndustrial Flow Measurement 3rd Edition. David W. Spitzer
Industrial Flow Measurement 3rd Edition David W. Spitzer Contents Preface WHY MEASURE FLOW? xi ABOUT THE BOOK xv ACKNOWLEDGMENTS xvii Chapter 1 INTRODUCTION 1 Objectives, 1 Prerequisites and Audience,
More informationINFLUENCE OF AFTERBODY SHAPE ANGLE OF TRAPEZOIDAL BLUFF BODY ON MEASURED SIGNAL PARAMETERS
27 P a g e INFLUENCE OF AFTERBODY SHAPE ANGLE OF TRAPEZOIDAL BLUFF BODY ON MEASURED SIGNAL PARAMETERS ATUL BACHAL Mechanical Department, Government College of Engineering Aurnagbad, MH, India, atul.bachal@in.endress.com
More informationTraining and qualification inspectors for advanced GWT inspections. David Alleyne 7MENDT, Bahrain 14 September 2015
Training and qualification inspectors for advanced GWT inspections David Alleyne 7MENDT, Bahrain 14 September 2015 1 Presentation Outline Biography History of GWT and inspector training GULT training scheme
More informationAutomated ultrasonic testing of submerged-arc welded (SAW) pipes using phased-arrays
18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa Automated ultrasonic testing of submerged-arc welded (SAW) pipes using phased-arrays 1 Jörg Ininger, 1 Ulrich Semmler,
More informationPerformance Evaluation of 200 and 400 mm PARTI MAG II, Bailey Fischer & Porter Electromagnetic Flowmeters, Installed in Partially Filled Pipes
Performance Evaluation of 200 and 400 mm PARTI MAG II, Bailey Fischer & Porter Electromagnetic Flowmeters, Installed in Partially Filled Pipes A Report for NMSPU Department of Trade and Industry 151 Buckingham
More informationUltrasonic Flow Meter Best Practices for commissioning and Balancing
AABC Commissioning Group AIA Provider Number 50111116 Ultrasonic Flow Meter Best Practices for commissioning and Balancing Course Number: CXENERGY1605 Brent Baird Instruments Direct April 12, 2016 Credit(s)
More informationwww.wefoundyourflowmeter.com For more information call 1.866.462.6839 / 905.764.7736 Ohio Medical has found a flowmeter that won t be moved or lost. Introducing the new Tube Style Integrated Flowmeter
More informationSolutions at GE Inspection Technologies
Solutions at GE Inspection Technologies Business Overview Paul A. Meyer Our Company A global leader in technology-driven inspection solutions that deliver productivity, quality, and safety to our customers
More informationBROEN UNIFLEX FITTINGS FOR NON-BURNING 2.0 GASES / VACUUM
BROEN UNIFLEX FITTINGS FOR NON-BURNING 2.0 GASES / ACUUM LAB TABLE OF CONTENTS Content Page Table of Content 2 Product Overview 3 BROEN UniFlex TM BASIC 4-5 Fittings for Non-burnig 2.0 Gases/acuum Turret
More informationDescription. Modell-Nr Airless ProfessionalLine. According to ISO 1167 ASTM D 1598 ASTM D Simple and safe operation. Reliable test results
Modell-Nr. 1720 Airless ProfessionalLine According to ISO 1167 ASTM D 1598 ASTM D 1599 Description The internal pressure creep test is a test procedure for determining the strength of thermoplastic pipes.
More informationInstallation Guide for ultrasonic transducers
Installation Guide for ultrasonic transducers 1 1. Welcome at systec Thank you very much for using deltawave. This document provides you with all necessary information to install the transducers accurately
More informationA comparative study of the flow rate characteristics of flangeembedded averaging Pitot tube (APT) flow meters for various crosssection
A comparative study of the flow rate characteristics of flangeembedded averaging Pitot tube (AP) flow meters for various crosssection shapes Choong oon Lee Department of Mechanical and Automotive Engineering,
More informationOIML R 122 Annex C RECOMMENDATION. Edition 1999 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL RECOMMENDATION OIML R 122 Annex C Edition 1999 (E) Equipment for speech audiometry Annex C Test report format Appareils pour l audiométrie vocale Annexe C Format du rapport d essai OIML R
More information4-20 ma Pressure Transducer
4-20 ma Pressure Transducer 56410099 Copyright 2017 Durham Geo Slope Indicator All Rights Reserved. This equipment should be installed, maintained, and operated by technically qualified personnel. Any
More informationEI10C - Pre-Instructional Survey
EI10C - Pre-Instructional Survey Name: Date: 1. Describe the following as they apply to flow measurement: Reynolds Number - High Accuracy - Applicable Range - Ideal Flowmeter - Cavitation - 2. Which is
More informationMONTANA FLUME. User s Manual. Montana (Short Parshall) Flume User s Manual
MONTANA FLUME User s Manual i Montana (Short Parshall) Flume User s Manual TABLE OF CONTENTS Introduction to the Montana Flume... 1 Function... 1 Design... 1 Standards... 2 Accuracy... 3 Dimensions...
More informationChoose smart valves for balancing Flow
Choose smart valves for balancing Flow For easy balancing of the flow NC Flow balancing valves NC Flow is designed for balancing heating and cooling systems. Incorrectly balanced flow is a common cause
More information8 Performing Medical Procedures
8 The Terason usmart3200t Ultrasound System can aid in performing medical procedures such as biopsies. Depending on whether you purchased the additional equipment required for these procedures, you may
More informationFluid Flow Calibration Facilities in India- A Case Study
flotek.g 2017- Innovative Solutions in Flow Measurement and Control - Oil, Water and Gas August 28-30, 2017, FCRI, Palakkad, Kerala, India Fluid Flow Calibration Facilities in India- A Case Study Battal
More informationTru-Fit 10k Fittings
Tru-Fit 10k Fittings TruFit 10K Pipe and Adapter Fittings are designed and manufactured to provide a reliable connection in high pressure systems running to 10,000 psi. Manufactured from fully-traceable
More informationMulti-Channel Film Dosimetry and Gamma Map Verification
Multi-Channel Film Dosimetry and Gamma Map Verification Micke A International Specialty Products Ashland proprietary technology, patents pending Single Channel Film Dosimetry Calibration Curve X=R R ave
More informationInspection of High Temperature Pipe-work using Guided Waves Mark J. EVANS 1, Simon BUTLER 2
18th World Conference on Non-destructive Testing, 16-20 April 2012, Durban, South Africa Inspection of High Temperature Pipe-work using Guided Waves Mark J. EVANS 1, Simon BUTLER 2 1 Guided Ultrasonics
More informationDental Extraoral X-ray Systems
Dental Extraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL EXTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability. A registrant s x-ray system used for dental
More informationAdvanced in vitro exposure systems VITROCELL VC 10 S-TYPE SMOKING ROBOT
Advanced in vitro exposure systems VITROCELL VC 10 S-TYPE SMOKING ROBOT Platform concept conventional cigarettes e-cigarettes fast cleaning VITROCELL SMOKING ROBOT VC 10 S-TYPE Features all special requirements
More informationPerformance of phased array and conventional ultrasonic probes on the new ISO reference block
Performance of phased array and conventional ultrasonic probes on the new ISO 19675 reference block C. Udell, D. Chai 1 and F. Gattiker Proceq S.A., Ringstrasse 2, Schwerzenbach, Switzerland. More info
More informationManual Ultrasonic Inspection of Thin Metal Welds
11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Manual Ultrasonic Inspection of Thin Metal Welds More Info at Open Access Database www.ndt.net/?id=16364
More informationQuickSplitTM Flow Splitters
QuickTM Flow ters between 2 or more detectors such as Mass Spectroscopy and evaporative light scattering, UV, or IR. This is especially important when one of the detectors is destructive, such as Mass
More informationST SERIES. design, development, production made in Germany
ST SERIES design, development, production made in Germany ST 01 - Crunch Bench ST 02 - Push-up Bars ST 03 - Pull-up Bars ST 04 - Back Bench ST 05 - Balance Board ST 06 - Stretching Bars SE ST 01 - Crunch
More informationCERTIFICATE OF COMPETENCE IN WORKPLACE NOISE RISK ASSESSMENT
CERTIFICATE OF COMPETENCE IN WORKPLACE NOISE RISK ASSESSMENT Entry Requirements The Institute of Acoustics is committed to an open access policy. The main principle is to admit all who will benefit from
More informationUltrasonic Measurement of Aircraft Strut Hydraulic Fluid Level
02WAC-19 Ultrasonic Measurement of Aircraft Strut Hydraulic Fluid Level Sidney G. Allison NASA Langley Research Center ABSTRACT An ultrasonic method is presented for non-intrusively measuring hydraulic
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 60601-2-37 2001 AMENDMENT 1 2004-08 Amendment 1 Medical electrical equipment Part 2-37: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 MASTER GAGE & TOOL CO. 112 Maplewood Street Danville, VA 24540 Sean Cobb Phone: 434 836 4243 CALIBRATION Valid To: June 30, 2018 Certificate
More informationDräger Alcotest 5510 Alcohol Screening Device
Dräger Alcotest 5510 Alcohol Screening Device The Dräger Alcotest 5510 handheld screener provides the quality, accuracy, and reliability that our customers have come to expect from our breath alcohol detection
More informationQingtian Instrument Kaifeng Qingtianweiye Flow Instrument Co.,Ltd
Qingtian Instrument Kaifeng Qingtianweiye Flow Instrument Co.,Ltd Section One Company Profile 2 Kaifeng Qingtianweiye Flow instrument Co., Ltd was founded in 2005,located in China flow instrument production
More informationPortable Equine Nebuliser System. User Manual
Portable Equine Nebuliser System User Manual Table of Contents INTENDED USE... 3 SAFETY INFORMATION... 3 TECHNICAL SPECIFICATION... 4 INSTRUCTIONS FOR USE... 6 MAINTENANCE... 12 TROUBLESHOOTING... 13 WARRANTY...
More information10 Years Experience in Industrial Phased Array Testing of Rolled Bars
18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa 10 Years Experience in Industrial Phased Array Testing of Rolled Bars Josef MAIER 1 and Gerhard Ferstl 1, 1 Böhler
More informationAIR PREPARATION UNITS 3/4" NPT & 1" NPT SERIES
PNEUMATIC GROUP AIR PREPARATION UNITS 3/" NPT & " NPT SERIES 6 0, 0,6 0, 0,8 8 0,0 0 0 bar 6 0, 0,6 0, 0,8 8 0,0 0 0 bar 83 6 0, 0,6 0, 0,8 8 0,0 0 0 bar 6 0, 0,6 0, 0,8 0,0 0 0 bar 8 Port size Recommended
More informationA new geometric and mechanical verification device for medical LINACs
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 A new geometric and mechanical verification device for medical LINACs Keith T. Welsh,* Robert A. Wlodarczyk, and L. E. Reinstein
More informationHearing Conservation Services Specifications
Hearing Conservation Services Specifications FACILITIES AND EQUIPMENT The following equipment is used in assessing hearing, middle ear function and noise levels: - Maico MA40 audiometer - Maico MA41 audiometer
More informationFlowmeters and. Specialty Products
Flowmeters and Specialty Products Chrome Flowmeter Durable, Accurate, Precise, Dependable 1MFA9005 4MFA1001 1MFA2001 1MFA1101 6MFA1001 1MFA3001 1MFA3501 1MFA1001 Specifications Model Gas Scale Increments
More informationLAB5 EDITION 2 NOVEMBER 2012
2007 LAB5 EDITION 2 NOVEMBER 2012 Reporting Calibration Results JH CONTENTS SECTION PAGE 1 Introduction 2 2 UKAS calibration accreditation symbol 3 3 General guidance for the presentation of calibration
More informationGambro Renal Products Technical Assistance Services
Gambro Renal Products Technical Assistance Services http://tech.usa-gambro.com 800/525-2623 303/232-6800 Technical Update PH,TR,11060 12/2008 TROUBLESHOOTING Equipment Type: Phoenix Dialysis System Subject:
More informationQuickSplit TM Flow Splitters
Quick TM Flow ters between 2 or more detecrs such as Mass Spectroscopy and evaporative light scattering, UV, or IR. This is especially important when one of the detecrs is destructive, such as Mass Spec
More informationGE Measurement & Control. Model C-ET. Ultrasonic Flow Transducer System. Installation Guide
GE Measurement & Control Model C-ET Ultrasonic Flow Transducer System Installation Guide 916-083 Rev. B January 2013 Model C-ET Ultrasonic Flow Transducer System Installation Guide 916-083 Rev. B January
More informationBM-1500 POWER TRAINER.
BM-1500 POWER TRAINER www.body--sculpture.com INDEX IMPORTANT SAFETY INFORMATION. 1 EXPLODED VIEW... 2 PARTS LIST... 3 ASSEMBLY INSTRUCTIONS.... 4 EXERCISE SESSION.... 5 EXERCISE INSTRCUTION..... 9 EXERCISE
More informationFlume Installation Instructions
Flume Installation Instructions To ensure a successful installation, it is important to keep the following points in mind: 1) There should be no bend, dips, elbows, or flow junction immediately upstream
More informationLUNG FUNCTION MEASUREMENT INSTRUMENT
LUNG FUNCTION MEASUREMENT INSTRUMENT THE WHISTLER IS THE FIRST HANDHELD, WIRELESS DEVICE IN THE WORLD FOR PERFORMING PULMONARY FUNCTION TESTS WITH BOTH BABIES AND PRESCHOOL CHILDREN 2 3 THE WHISTLER MAKES
More informationAPI. Defined Procedure. for. Ultrasonic Thickness Measurement API-UT-21
API Defined Procedure for Ultrasonic Thickness Measurement API-UT-21 This Procedure Defines the Recommended Ultrasonic Methods and Techniques for Thickness Measurements Page 1 1.0 PURPOSE 1.1 This procedure
More informationCase Study. Complex Components Demand Innovative Ultrasound Solution
Complex Components Demand Innovative Ultrasound Solution Critical components in the aerospace and defense industries, such as engine housings, wings and control surfaces, undergo rigorous testing to ensure
More informationTHE RIGAKU JOURNAL VOL. 23 / 2006, A1-A10 CONSIDERATIONS REGARDING THE ALIGNMENT OF DIFFRACTOMETERS FOR RESIDUAL STRESS ANALYSIS
THE RIGAKU JOURNAL VOL. 23 / 2006, A1-A10 CONSIDERATIONS REGARDING THE ALIGNMENT OF DIFFRACTOMETERS FOR RESIDUAL STRESS ANALYSIS THOMAS R. WATKINS, O. BURL CAVIN, CAMDEN R. HUBBARD, BETH MATLOCK, AND ROGER
More informationIn-Line Accessories. In-Line Pressure Gauges. Pressure Gauges. In-Line Accessories
Pressure Gauges Pressure gauges for monitoring system pressure Hoses and test points for sampling oil and determining ISO cleanliness levels Flanges to connect components Valves for system control In-Line
More informationENGR 83 Final Project
ENGR 83 Final Project Affordable Flowmeter for Measurement of Low- Flowrate Suspended- Particulate Flow Julian Leland ENGR 83, 2011 Swarthmore College Introduction Measurement of fluid flow is an important
More informationThe design and development of a force comparator standard machine to provide national traceability in force measurement to industry.
The design and development of a force comparator standard machine to provide national traceability in force measurement to industry. Speaker/ Author: S. Dlamini Co-author: C. Gouws National Metrology Institute
More informationw. D. Jolly, F. A. Bruton, and C. Fedor
ULTRASONIC TRANSDUCER CHARACTERIZATION STATION w. D. Jolly, F. A. Bruton, and C. Fedor Southwest Research Institute 6220 Culebra Road San Antonio, Texas 78284 INTRODUCTION The portable ultrasonic transducer
More information3000 Series Articulator System
3000 Series Articulator System Models 3040 and 3140 Instruction Manual Whip Mix Corporation 361 Farmington Ave. P.O. Box 17183 Louisville, KY 40217-0183 USA 502-637-1451 800-626-5651 Fax 502-634-4512 www.whipmix.com
More informationModern Hadron Therapy Gantry Developments. 16th Jan 2014
Modern Hadron Therapy Gantry Developments 16th Jan 2014 Protect, Enhance, and Save Lives - 2 - IBA ProteusONE THE TREND: MORE COMPACT, CHEAPER Access to PT is key ProteusPLUS ProteusONE 18000 ft² / 1672
More informationGRUNDFOS DATA BOOKLET. BM, BMhp. Booster Modules 50/60 Hz
GRUNDFOS DATA BOOKLET BM, BMhp Booster Modules 5/6 z BM, BMhp Table of contents 1. BM applications 4 Pumped liquids 4 2. BM product introduction 4 Standard pumps 4 Construction 4 Motor 4 3. BM, BMhp performance
More informationSignet 2850 Conductivity/Resistivity Sensor Electronics and Integral Systems
Signet 2850 Conductivity/Resistivity Sensor Electronics and Integral Systems Features Certificate of calibration supplied with all sensors. Universal Mount Custom cell constant programmed into the electronics.
More informationComparability and quality of experimental data under different quality systems. S. Caroli Istituto Superiore di Sanità Rome
Comparability and quality of experimental data under different quality systems S. Caroli Istituto Superiore di Sanità Rome Programme of this presentation Part I. Background information Part II. Key aspects
More informationKiwa Inspecta. Kiwa Inspecta. Trust Quality Progress
Pressure equipment Third party agency In-Service Inspection Condition controls Crane and Lifting Risk assesments Elevator acts independently with integrity and full respect for the customers needs. As
More informationShcherbakov O.N., Petrov A.E., Polevoy A.G., Annenkov A.S. ULTES LLC, Moscow, Russia
METHODS AND MEANS OF ULTRASONIC INSPECTION (USI) OF WELDED JOINTS WITH ACOUSTIC CONTACT LEVEL TRACKING AND EVALUATION OF THE NATURE AND ACTUAL CHARACTERISTICS OF FLAWS Shcherbakov O.N., Petrov A.E., Polevoy
More informationOrganisation Internationale de Métrologie Légale
Organisation Internationale de Métrologie Légale OIML INTERNATIONAL RECOMMENDATION Equipment for speech audiometry Appareils pour l audiométrie vocale OIML R 122 Edition 1996 (E) CONTENTS Foreword... 3
More informationAgilent 1220 Infinity LC System Performance Specifications
IET International Equipment Trading, Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 1220 Infinity LC System Performance
More informationSureFlow. Oxygen Flow Station PROVIDER TECHNICAL MANUAL
SureFlow Oxygen Flow Station PROVIDER TECHNICAL MANUAL Table of Contents Definition of Product Symbols... 3 General Information... 3 Warning and Caution Statements...3 Introduction to the SureFlow Oxygen
More informationTable 1: Samples used in experimental program. Block ID/thickness [ mm ]
CONTRIBUTION TO DETECTION AND SIZING LINEAR DEFECTS BY CONVENTIONAL AND PHASED ARRAY ULTRASONIC TECHNIQUES P. Ciorau Ontario Power Generation-Inspection Services Division-NDE Systems Department, Pickering,
More informationMost of the Conductivity/Resistivity electrodes used with the 2850 can be installed into standard 3 /4 in. (NPT or ISO 7-R 3 /4) fittings.
InLine Installation Most of the Conductivity/Resistivity electrodes used with the 850 can be installed into standard /4 in. (NPT or ISO 7R /4) fittings. The 8505X models are designed for integral systems,
More informationCopyrighted Material. Do not duplicate.
THAT IS THE EASY PART Copyrighted Material. Do not duplicate. JOHN TSCHIDA - UNIVERSITY OF ST. THOMAS HITTING Our Goal in Hitting: Consistently Achieve Maximum Force with Minimal Effort!! 1) Force: Bat
More informationRUBBER WATERBLAST HOSES
up to 15, PSI (1 Bar) RUBBER WATERBLAST HOSES These high pressure hose assemblies combine rugged construction, dependability and safe performance. Designed specifically for waterblast applications Lightweight
More informationDevelopment and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.
Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Gonzalo Walwyn Salas, Stefan Gutiérrez Lores Center for Radiation Protection and Hygiene,
More informationProficiency Testings (PTs) and Interlaboratory Comparisons (ILCs) for the Department of Calibration Laboratories
Title Reference Proficiency Testings (PTs) and Interlaboratory Comparisons (ILCs) for the Department of Calibration Laboratories RT-36 Revision 00 Date Preparation Approval Authorization Application date
More informationSignet 2850 Conductivity/Resistivity Sensor Electronics and Integral Systems with PVDF Sensor
Signet 2850 Conductivity/Resistivity Sensor Electronics and Integral Systems with PVDF Sensor Features Test certificate supplied with all sensors Universal Mount Custom cell constant programmed into the
More informationActive Imaging TM transducers: Active Imaging 3-IN-1 Active Imaging SideScan
Active Imaging TM transducers: Active Imaging 3-IN-1 Active Imaging SideScan Installation manual ENGLISH www.simrad-yachting.com www.lowrance.com 1 Preface Warning: It is your sole responsibility to install
More informationSTART & RUN A MARIJUANA DISPENSARY OR POT SHOP: WHEREVER IT IS LEGAL! (START & RUN A BUSINESS SERIES) BY JAY CURRIE
START & RUN A MARIJUANA DISPENSARY OR POT SHOP: WHEREVER IT IS LEGAL! (START & RUN A BUSINESS SERIES) BY JAY CURRIE DOWNLOAD EBOOK : START & RUN A MARIJUANA DISPENSARY OR POT SHOP: SERIES) BY JAY CURRIE
More informationOctober 1999 ACCURACY REQUIREMENTS FOR NET QUANTITY DECLARATIONS CONSUMER PACKAGING AND LABELLING ACT AND REGULATIONS
October 1999 ACCURACY REQUIREMENTS FOR NET QUANTITY DECLARATIONS CONSUMER PACKAGING AND LABELLING ACT AND REGULATIONS TABLE OF CONTENTS SUBJECT PAGE Accuracy Requirements for Net Quantity Declarations
More informationMethods of validation of occupational noise exposure measurement with multi aspect personal sound exposure meter
Proceedings of ACOUSTICS 2016 Methods of validation of occupational noise exposure measurement with multi aspect personal sound exposure meter Marek Podgórski Svantek Sp. z o.o., Warsaw, Poland ABSTRACT
More informationStainless Steel Trough Urinal Stainless Steel Urinals UWHA
Stainless Steel Trough Urinal Stainless Steel Urinals UWHA Product Code UWHA Material Manufactured from 1.6mm thick stainless steel with a satin finish. Length The urinal is available in five standard
More informationHearing Conservation Program
Last Reviewed Date: 3/07/2018 Last Revised Date: 7/27/2017 Effective Date: 6/27/1994 Applies To: Employees, Faculty, Students, Others For More Information contact: EHS, Coordinator at 860-486-3613 or valerie.brangan@uconn.edu
More informationUser Manual. RaySafe i2 dose viewer
User Manual RaySafe i2 dose viewer 2012.03 Unfors RaySafe 5001048-A All rights are reserved. Reproduction or transmission in whole or in part, in any form or by any means, electronic, mechanical or otherwise,
More informationRaySafe i3 INSTALLATION & SERVICE MANUAL
RaySafe i3 INSTALLATION & SERVICE MANUAL 2017.06 Unfors RaySafe 5001104-1.1 All rights are reserved. Reproduction or transmission in whole or in part, in any form or by any means, electronic, mechanical
More informationCertificate of Competence in Workplace Noise Risk Assessment (CCWPNRA):
Certificate of Competence in Workplace Noise Risk Assessment (CCWPNRA): The course follows the prescribed Institute of Acoustics (IOA) syllabus and it additionally addresses some of the differences in
More informationGeneral Single Ion Calibration. Pete 14-May-09
General Single Ion Calibration Pete 14-May-09 Purpose of SI calibration Measure the instrument response of a single ion. Necessary for understanding of error in instrument (counting statistics) Calculation
More informationImplementation Guide for the DoseControl Dosimetry System
1.0 SCOPE The GEX DoseControl System is intended for use as a single dosimetry system solution, specifically developed to satisfy the generic dosimetry needs of radiation sterilization application users
More informationRoot Cause Analysis of vibrations in a Roots Blower system
Root Cause Analysis of vibrations in a Roots Blower system Leonard van Lier, Pieter van Beek Christof Fischer, Stefan J. Uibel Problem description A roots blower system is suffering from excessive vibrations
More informationHearing Conservation Program
Hearing Conservation Program Revised: September 2017 2009-2017. University of New Hampshire Office of Environmental Health and Safety. All rights reserved. The most current version of this document can
More informationyour bioassay is in good hands: transfer from a CRO perspective MIKE MERGES March 4, 2013
your bioassay is in good hands: transfer from a CRO perspective MIKE MERGES March 4, 2013 Contact Information Michael Merges, M.S. Director Analytical Development Solutions Development & Clinical Services
More informationOccupational Noise. Contents. OHSS: Guidance Occupational Noise
OHSS: Guidance 105.1 Occupational Noise Occupational Noise Contents Scope... 2 Introduction to Noise... 2 Effects of Hearing Loss... 2 Managing Noise Risks... 2 Estimating Noise Exposure... 4 Noise Risk
More informationGRIMM. European Approval PM10 & PM2.5. situation & legislation for. Aerosol Technik
GRIMM Aerosol Technik European Approval situation & legislation for PM10 & PM2.5 AGENDA Legislation Guidance of euqivalence EDM180 - PM10 EDM180 - PM2.5 Certifications PM2.5 start with 1.1.2009 For PM10
More informationWhat is the Difference Between the New BAM-1020 PM2.5 FEM and Older BAM-1020 Monitors?
Met One Technical Bulletin Revised: April 21, 2009 By: Dennis Hart Met One Instruments, Inc. 1600 N.W. Washington Blvd. Grants Pass, OR 97526 (541) 471-7111 www.metone.com What is the Difference Between
More informationHigh-end solution STEP-BY-STEP USER GUIDE
High-end solution STEP-BY-STEP USER GUIDE System Contents REAL TIME Restoration Guided Conical drill Multi-Unit abutment, Conical connection Guided kit for conical connection implant procedure C1 & V3
More informationLowering of the Alcohol Limits for Drivers in Ireland and the Analytical Impact
Lowering of the Alcohol Limits for Drivers in Ireland and the Analytical Impact Helen Kearns, Pauline Leavy, David Reynolds, Denis Cusack. Medical Bureau of Road Safety, University College Dublin. Abstract
More informationUnion Bonnet Valves - VU6 Series
BlowOur Proof Back Seating Hardened Stem Thread Mountable Panel nut Metal Seal Bonnet Union Bonnet Valves VU6 Series Features Pressure rating up to 6000psig (13bar) @38 C (100 F) Temperature rating from
More informationMore about inferential control models
PETROCONTROL Advanced Control and Optimization More about inferential control models By Y. Zak Friedman, PhD Principal Consultant 34 East 30 th Street, New York, NY 10016 212-481-6195 Fax: 212-447-8756
More informationFundamentals of Pressure Transducers
print close Fundamentals of Pressure Transducers (Unpublished) Hydraulics & Pneumatics Alan Hitchcox Tue, 2013 09 17 (All day) Pressure is a key factor in almost every fluid power circuit. Pressure transducers
More informationCleveland State University Department of Electrical and Computer Engineering Control Systems Laboratory. Experiment #3
Cleveland State University Department of Electrical and Computer Engineering Control Systems Laboratory Experiment #3 Closed Loop Steady State Error and Transient Performance INTRODUCTION The two primary
More informationThere are many varieties of particulate pollutants in our homes and offices. Ones from outside include industrial
Model: IAQ Indoor PM2.5 -PM2.5 Particulate matter (PM) is a particle pollution, which is produced in a great number of ways that can be classified into either mechanical or chemical processes. Traditionally,
More informationBS EN ISO 14971: Risk Analysis and Risk Management Report
BLACKHORSE ROAD, LETCHWORTH, HERTS, SG6 1HB, ENGLAND Tel: + 44 (0) 1462 686221 www.stoddard.co.uk Fax: + 44 (0) 1462 480711 BS EN ISO 14971: Risk Analysis and Risk Management Report Product: Gates and
More information