by Automated Cell Counters

Size: px
Start display at page:

Download "by Automated Cell Counters"

Transcription

1 Cytometry (Communications in Clinical Cytometry) 22:26-34 (1995) Variability in Absolute Lymphocyte Counts Obtained by Automated Cell Counters E. Simson and W. Groner Department of Pathology, Long Island Jewish Medical Center, the Long Island campus for the Albert Einstein College of Medicine, New Hyde Park, New York (E.S.); Center for Laboratory Technology, Inc., Great Neck, New York (W.G.) There is increasing interest in the absolute lymphocyte count. This is partly driven by the need to obtain absolute values for lymphocyte subsets such as absolute CD4+ counts in human immunodeficiency virus (HIV)-infected persons. The absolute total lymphocyte count is usually determined in the routine hematology laboratory on a separate sample from the same patient specimen and then combined with percentage results from flow cytometry to obtain the absolute value of the lymphocyte subsets. We have studied analytic variability in the absolute lymphocyte determination and compared it to the variability of the total white blood count (WBC). In a series of 524 specimens, four different automated methods were compared to each other and to the traditional eye count differential. The automated methods were four widely used automated cell counters (Technicon H*l, TOA NE8, Coulter STKS, and Abbott CD3). The results indicate that analytic variability in the absolute lymphocyte counts, due, primarily, to method variability, is significant and is larger than the variability typically observed on interlaboratory trials of relative CD4 counts. These method biases cannot easily be reduced by calibration, since the cell classification algorithms are built-in features of the various cell counters. Analytic variability of the absolute lymphocyte counts was found to be 12.4% compared with analytic variability of only 4.9% for total WBC counts on the same samples. Our data suggest that more precise results would be obtained if flow cytometry results expressed each phenotype as a fraction of the leukocytes as well as total lymphocytes. Conversion to absolute values could then be accomplished through determination of the total WBC in the routine hematology laboratory Wiley-Liss, Inc. Key terms: Flow cytometry, hematology methods, lymphocyte counts, HIV, CD4+ lymphocytes The determination of the absolute value of CD4+ lymphocytes has been shown to be useful in monitoring persons with human immunodeficiency virus (HIV) infection (12). The U.S. Public Health Service has recommended monitoring CD4+ T lymphocytes every 3-6 months in all HIV-infected persons (5), and the level of + CD4 T lymphocytes has been included as a criterion in the classification system for HIV infection (2). It is therefore of great interest to consider the potential for and magnitude of error in this determination. The measurement of CD4+ T-cell levels presently involves the combination of at least two (sometimes three) independent laboratory determinations. The first, which is the counting of the relative number (percentage) of CD4 lymphocytes by flow cytometry, has received considerable attention. Stand ards for performance have been issued recently by the Centers for Disease Control (CDC), the National Committee for Clinical hbordtory Standards (NCCLS), and the National Institute for Allergy and Infectious Disease (NIAID; 6,19,4) in an effort to minimize the impact of technical and analytic factors (such as staining procedures and alignment and calibration of flow cytometers) on intralaboratory and interlaboratory variability in the measurement of CD4 cell number. Each of these documents also refers to the need for additional accurate hematology measurements to convert the relative values into the required absolute value (cells/ volume). The documents stipulate that the hematology counts be performed on fresh specimens drawn at the same time, and they suggest that automated counters be used (except for samples rejected or flagged by the automated counter) for the determination of the absolute lymphocyte count to reduce the imprecision resulting from the relatively low numbers of cells typically classified by manual microscopy. The NIAID guidelines state that, when manual counting is required, a minimum of Received for publication February 1, 1994; accepted June 14, Address reprint requests to Elkin Simson, Associate Chairman, Clinical Pathology, Long Island Jewish Medical Center, New Hyde Park, NY Wiley-Liss, Inc.

2 VARIABILITY IN ABSOLUTE LYMPHOCYTE COUNTS 27 4 cells must be classified, and all atypical lymphocytes and large unstained cells (LUCS) are to be included in the lymphocytes. Studies of the imprecision of the determination of absolute lymphocytes have also been reported ( 16,17). These studies demonstrate that the total variability in the computed CD4 counts can be substantially affected by specimen handling and counting imprecision. A recent comment has also indicated that significant bias may result from the use of different automated cell counters (24) and that flow cytometers may be better in determining the percentage of lymphocytes ( 13). In this study, an attempt was made to characterize the analytic variability of absolute lymphocyte counts by considering the variation across different automated cell counting instruments in addition to within-system imprecision. Results were obtained on the same samples with each of four different multiparameter cell counters. The percentage of total lymphocytes from each system was compared with the reference for manual (eye count) differential, per NCCLS H2A (22), to ascertain the performance of lymphocyte counting for each system. The results were also compared with each other for total white blood cell count (WBC) and for absolute lymphocyte count. Protocols were established to isolate the effects of specimen handling, method bias, and imprecision. The results were then analyzed to demonstrate the variability in absolute CD4 counts that can be expected as a result of the variability of the hematology measurements. MATERIALS AND METHODS Specimen Collection and Handling Samples of blood from 697 patients, including 12 normal individuals, were run on each of the four instruments over a period of 2 days. Patient specimens were part of the daily workload arriving in the hematology laboratory of the Department of Pathology at Long Island Jewish Medical Center, an 87-bed tertiary-care hospital complex. Specimens were drawn in tripotassium EDTA and were analyzed within 18 h from phlebotomy. Specimens were analyzed within 6 h from their receipt in the laboratory. Specimen handling errors due to aliquoting were reduced by running the identical samples sequentially on each analyzer. The analyzer sequence was varied in a random manner. Most samples were run in duplicate on each analyzer, but approximately 7 were only run singly on each analyzer due to insufficient sample volume for duplicate analysis. Reference Differential Wedge smears were made from each sample and analyzed by eye microscopy. A 4-cell eye differential was obtained from two independent 2-cell differentials as described in NCCLS H2A (22). Instruments Four instruments were used for the study: 1) Technicon H* 1 System (Miles Diagnostics, Tarrytown, NY), us- ing software version 1.3; 2) Coulter STKS (Coulter, Hialeah, FL), using software version 1D and ld.l; 3) Sysmex NE8 (TOA Instruments, Kobe, Japan), using software version 27; and 4) CD3 (Abbott Diagnostics, Chicago, IL), using software version Revision L. Each of the instruments was installed by the manufacturer and run according to the manufacturer s instructions. The instruments were calibrated independently according to the manufacturers specifications. Quality control materials recommended by each manufacturer were run throughout the study to verify constancy of calibration. Any sample on which any of the analyzers did not produce results for WBC, percentage lymphocytes, or absolute lymphocytes was eliminated. This left 524 specimens for comparative analysis. In this sample set, the WBC ranged from 2 X 1 Aiter to 6 X 1 Aiter. The absolute lymphocyte counts ranged kom.15 X lo /liter to 9 X lo /liter, and the proportional lymphocytes from 1% of WBC to 75% of WBC. Accuracy The accuracy of each analyzer in determining lymphocyte percentage was obtained by comparing the result from each specimen with the result from the 4-cell reference manual differential. When duplicate analyzer results were available, only the first result was used for comparison. Imprecision For each analyzer, duplicate imprecision was determined for both absolute lymphocytes and total WBC by analyzing the paired readings from each of the analyzers. For comparison, the duplicate imprecision of 2-cell manual counts multiplied by WBC was calculated as per NCCLS H2A. In addition, the duplicates for each method were compared to each other by regression analysis to determine the residual scatter around the regression line contributed by method imprecision. Analytic Variability To determine the analytic variability in the determinations, the first result for both absolute lymphocytes and total WBC from each analyzer on each specimen was used. As a measure of the analytic variability across the four analyzers, a coefficient of variation was calculated for each specimen. Specimen Aging A series of four specimens was drawn, and aliquots of each specimen were stored both at room temperature and refrigerated (-4 C) over a period of 24 h. These samples were then analyzed repeatedly by each of the four instruments over a period of 24 hours. RESULTS Accuracy Figure l a4 shows graphic summaries for each system compared to the reference method for percentage lym-

3 28 SIMSON AND GRONER phocytes. Table 1 summarizes the population and regression statistics for the accuracy study. The results show relatively good agreement between each method and the reference method, with the bulk of the results falling within the bounds of imprecision for the reference 4- cell manual. Imprecision Table 2 summarizes the results of the duplicate imprecision studies for absolute lymphocytes and for total WBC obtained from each instrument. Duplicates of the 2- cell microscopic eye count for lymphocytes are also included in Table 2 for comparison. The results in Table 2 demonstrate that the larger numbers of cells typically counted with the automated instruments have the expected effect of substantially reducing the duplicate imprecision as compared with the microscopic differential, even when 2 cells are classified by eye rather than the routine 1-cell differential. The imprecision for lymphocyte counts for the analyzers ranged from a standard deviation (S.D.) of.7 X 1 cells/liter to.18 X 1 ceils/liter yielding coefficients of variation (C.V.) from 3.95% to 1.45%. There were greater differences among systems for WBC counts where the S.D. varied from.14 to.47 X IO cells/liter. However, due to the larger denominator, C.V.s were smaller, ranging from 1.71% to 5.13%. The mean values for absolute lymphocytes and WBC varied somewhat among the analyzers as a result of the differences in samples included in the precision study. As expected, the standard error on regression analysis (Sy.x) was similar to the S.D. Analytic Variability In order to focus on the total analytic variability, the data were subjected to a one-way analysis of variance. The coefficient of variation (C.V.) across the four measurement systems for each of the 524 specimens was reviewed; one sample, whose C.V. was greater than 5% for WBC and greater than 1% for absolute lymphocyte count, was considered an outlier and was excluded from further analysis of variability, leaving 523 samples. The results were then analyzed in a manner analogous to the determination of comparability between two methods ( 15), with the C.V.s replacing the difference between methods. Figure 2 shows the coefficients of variation across the analyzers for absolute lymphocytes and for total WBC plotted vs. the level of each parameter. In order to test for any relationship between the variability and the level (the analogy for proportional bias), the data were subjected to regression analysis. Figure 2a demonstrates the results obtained for WBC. For a few samples, the C.V. was slightly greater than 2%, but, for most, it was less than 1%. The regression line was horizontal, indicating no relationship between the magnitude of the variability and the magnitude of the WBC. The mean value of the C.V. for the WBC count was 4.9%. Figure 2b shows the C.V. across the analyzers for absolute lymphocytes vs. the mean absolute lymphocyte count. The C.V. for several samples was as high as 5%, with many samples greater than 1%. The mean C.V. was 12.4%. When tested by regression variability analysis, variability also increased significantly as absolute lymphocyte count decreased. However, this increase could be the result of increased imprecision, because fewer cells are classified as the absolute lymphocytes count decreases. Figure 3 compares the distribution of the coefficient of variation for WBC with absolute lymphocytes. The variability for WBC is generally much less than that for absolute lymphocyte counts, with 99% of the WBC C.V. less than the average value for absolute lymphocyte counts ( 12.4% ). Specimen Handling Figures 4 and 5 show results obtained from specimens stored at room temperature for 1 hours from each method for absolute lymphocytes and total WBC respectively. This is a greater period than that stipulated in current standards for determining CD4+ counts. The dashed lines for each analyzer are derived from the duplicate imprecision 2 2 S.D. from the starting value. Thus, since each data point is a single determination, only results exceeding this boundary would indicate a shift from the initial value with greater than 95% confidence. It can be seen from the figures that no significant drifts (defined as two consecutive results exceeding the boundary) occurred within the 1 h postphlebotomy study period for either absolute lymphocytes or total WBC. Similar results were obtained from the samples stored at refrigerated temperatures. DISCUSSION The results demonstrate that, as was previously noted (25), increased precision in lymphocyte counting is achieved with the use of automated cell counters. Furthermore, the results demonstrate that there is generally good agreement between each analyzer and the traditional microscopic eye count, as has also been previously reported for individual analyzer comparisons (3,8,9,1 I, 23,26,27). However, these comparisons are typically limited by the imprecision of the microscopic eye count and, therefore, are not generally useful in detecting small method variances that may be significant. It was the intent of this study to examine the overall comparability of the various methods for producing absolute lymphocyte counts to be used to complement immunophenotyping performed on a flow cytometer. Thus, in our analysis, we attempt to characterize the total analytic variability in lymphocyte counts that may be encountered in, for instance, tracking an HIV-positive individual by obtaining CD4 and lymphocyte counts every 6 months, not necessarily from the same analyzer. We assumed only that the hematology results were obtained in a manner consistent with the current guidelines for immunophenotyping, i.e., that the determination was performed with a properly calibrated and quality controlled automated cell counter on a fresh blood sample.

4 VARIABILITY IN ABSOLUTE LYMPHOCYTE COUNTS 29 f 2. r P E, -1 D - C95%-LINE - C95%+LINE CLYMP4 ae a Manual 4 cell Lymphocyte% 1 I I b o Manual 4 cell Lymphocyte % N95%-LINE - N95%+LINE NLYMPA Manual 4 cell Lymphocyte% 1 7 f 2. c D 5 m Y I-m S95%-LINE - S95%+LINE SLYMP% d o Manual 4 cell Lymphocyte% FIG. 1. Comparison to reference method. Scatterplot and regression analysis for relative lymphocytes (%I vs. the 4 cell manual microscopic reference method. A single reading for each specimen is taken for comparison. Where duplicate readings were available, the first of the pair is taken. The oval boundary represents the 95% confidence limits for the reference method. a: CD3 vs. reference. b: H*l vs. reference. c: NE8 vs. reference. d: STKS vs. reference.

5 3 SIMSON AND GRONER Table 1 Summary of Accuracy Data: Percentage Lymphocytes Sample Population Regression Statistics Analyzer Mean Rangea Slope Intercept Corr. Coeff. H* CD NE STKS aentire range of specimens tested. The results of this study indicate that substantially greater analytic variability can be encountered in the absolute lymphocyte determination than in the determination of total WBC. The additional variability of absolute lymphocyte counts is due, in part, to increased imprecision but, to a larger degree, to both scatter and biases between the cell counters. These biases are most probably a result of differences in both the technologies and the cell classification algorithms used in the cell counter design. Thus, the differences are essentially built into the cell counter and cannot be adjusted out by calibration. These instrument algorithms can result in a bias for lymphocytes or granulocytes, even between two instruments from the same manufacturer (7). Although no results of interlaboratory trials across instruments are available for the determination of absolute lymphocytes, our data suggest that the analytic variability is greater than 1% or significantly greater than the C.V. reported in interlaboratory studies of relative CD4 counts (24). The data in this study show the analytic variability of total WBC to be approximately 5%. This is approximately one-third of the analytic variability of absolute lymphocytes. This result for WBC is quite consistent with results of interlaboratory trials for this determination ( 1). Further, each of the automated instruments contains a means for adjusting out the bias in total WBC by means of calibration. The absolute number of a lymphocyte subset is currently obtained from the product of two independent measurements; a relative measurement (Rel) of the percentage of lymphocytes isolated by one or more monoclonal antibodies, and the determination of the absolute lymphocyte counts (ABS lymph). It is likely that advances in technology will be made to overcome this situation for the special case of CD4+ counting. This may occur through better flow cytometric methods (2 1 ) or through the use of an alternate technology in which a direct determination by sensitive immunoassay is made of the concentration of CD4' (18). However, it is unlikely that the need to combine results from quantitative cell counters with flow cytometers will be totally eliminated for some time to come. Therefore, it remains of interest to consider how this combination can be made with minimal variability. The theory of the propagation of errors states that the coefficient of variation for a simple product of two independent determinations is given by the root sum square of the coefficients of variation for each measurement ( 1 ). Thus, the coefficient of variation for the absolute cell count can be expressed as: CVZ(ABS) = CV2(REL) + CV2(ABS LYMPH) (1) Interlaboratory trials (2) have shown the C.V. for the relative determination of lymphocyte subsets to be typically less than 1% and, for CD4, on the order of 5%. As an example, assume that 5% of lymphocytes are CD4+ at an absolute count of 1. X 1" cells/liter. The value from Figure 2b for the variability of absolute lymphocytes at 1. X 1" cellsfliter is 15%. Combining it with a C.V. for CD4 (Rel) of 5%, one obtains an estimate of analytic variation of absolute CD4+ counts on the order of 16%, or +.8 X lo9 cells/liter. This analytic variability will increase as the absolute lymphocyte count decreases as a result of both counting imprecision and method variability. Following the same theory, the coefficient of variation for absolute lymphocytes could alternatively be given in terms of the C.V. of the lymphocyte subset expressed as a percentage of total leukocytes (Rel* ) together with the C.V. of the total WBC (TOT WBC) as: CVZ(ABS) = CV2(REL*) + CV2(TOT WBC) (2) In this case, the relative count obtained on the flow cytometer is the fraction obtained by dividing the cells isolated by the monoclonal antibodies by total gated leukocytes. The data we have collected show that both of the components of the C.V. (imprecision and analytic variability) are much less for the total WBC than for absolute lymphocytes. Furthermore, the total WBC is a more robust parameter, is less subject to abnormal flagging, can be calibrated independently, and is regularly assessed in interlaboratory trials. Thus, the component of variability contributed by the hematology result should be reduced by using the total WBC. In addition, we have conducted a brief trial with five normal samples to confirm the implications of equations 1 and 2. Each of five normal specimens was analyzed for CD4+ lymphocytes 3 times on a flow cytometer (Becton Dickinson Facscan) with the total lymphocyte population isolated with CD45 and gated. The fraction of + CD4 lymphocytes was calculated. The samples were then reanalyzed 3 times with only the total leukocyte population gated (using forward vs. side scatter), and the CD4+ lymphocytes were calculated as a fraction of the total leukocytes. Isolation of + CD4 cells was by dual fluorescence (CD4/CD3) in each case, and Becton Dickinson reagents were used throughout the study. A sample of each specimen was also analyzed on a hematology cell counter (Miles Diagnostics H* 1) ten times. Results for the concentration of absolute + CD4 lymphocytes were then obtained by each of the methods described above, and the C.V. for the two methods was calculated. Concentrations of CD4+ were similar by both methods, and replication showed that the imprecision of the flow cytometry results were similar whether the

6 VARIABILITY IN ABSOLUTE LYMPHOCYTE COUNTS 31 Absolute lymphocyte count Table 2 Duplicate Imprecision" Total WBC Method N Mean Value S.D. C.V. (Yo) Sv.xb Mean Value S.D. C.V. (Yo) SV.Xb CD H*l NE STKS Manualc N /A N /A NIA - "All values and S.D. in 1O911iter. bsy.x, root mean square of residuals when a linear regression analysis is performed on the duplicate results. It is a measure of the scatter of observation s about the line of best fit. 'Two hundred cell count multiplied by the WBC for each sample to obtain an absolute count > 8 4 3o 2 1 a WBC MEAN Ln - El2 m 1 = a o WBC CV.% Y = -2.9X Mean CV = 12.4 Mean CV = 12.4% t bo I Abs. LYMP MEAN FIG. 2. Analytic variability. Each data point represents the coefficient of variation (C.V.) across four methods of 523 single specimens. The results of a regression analysis between the C.V. and level is also indicated. a: C.V. across methods for total WBC vs. total WBC x lo9 cellsil. b: C.V. across methods for absolute lymphocytes vs. absolute lymphocytes x 1Og/liter. CD4+ lymphocytes were expressed as a fraction of lymphocytes or as a fraction of leukocytes, as shown in Table 3. The average C.V. for CD4+ concentration, as determined using the leukocyte fraction (method 2), was 5.% bo abs LYMP. CV.% FIG. 3. Distribution of variability. The frequency of occurrence of the C.V. across methods for 523 specimens is plotted vs. the C.V. a: Frequency distribution for total WBC. b: Frequency distribution for absolute lymphocytes. compared with 5.3% for CD4 concentration using the lymphocyte fraction (method 1). Thus, we propose that the total variability of absolute lymphocyte subset counts may be reduced significantly if the bridge between the results obtained with the flow cytometer and those obtained in hematology is made through the total WBC. This would simnlifv the gating nrocedure on the flow

7 32 SIMSON AND GRONER lo T STKS T O I I CD T lo N E lo T H"l - 1 A Hours After Draw 8 1 FIG. 4. Specimen stabilityitotal WBC. Each of four specimens was aliquoted at phlebotomy to each of four analyzers. Samples were stored at room temperature and analyzed at the tlmes indicated. Each point represents a single sampling.

8 VARIABILITY IN ABSOLUTE LYMPHOCYTE COUNTS 33 2 J CD t Hours After Draw 8 1 FIG. 5. Specimen stabilityiabsolute lymphocytes. Each of four specimens was aliquoted at phlebotomy to each of four analyzers. Samples were stored at room temperature and analyzed at the times indicated. Each point represents a single sampling.

9 34 SlMSON AND GRONER Table 3 Imprecision Comparison on Normal Specimens: Flow Cytometry Gating of Lymphocytes (Method I) vs. flow Cvtometric Gating of WBC (Method 2) Measurement Average C.V. (%) CD4 divided by lymphocyte 4.1 Lymphocytes cellsiliter x lo9 3.1 Concentration of CD4 5.3 Method 1 (a x b) CD4 divided by WBC 4.7 WBC cells/liter x lo9 1.8 Concentration of CD4 5. Method 2 (d x e) cytometer and eliminate the need for reagents used specifically to isolate lymphocytes. This hypothesis will have to be confirmed by more extensive studies of the equivalence of results including studies of abnormal specimens and interlaboratory trials. These studies are being pursued in our laboratory and will be reported in due course LITERATURE CITED Beers Y: Introduction to the Theory of Errors. Addison-Wesley, Reading. MA, Brando B, Sommaruga E: Nationwide quality control trial on lymphocyte immunophenotyping and flow cytometer performance in Italy. Cytometry 14:294-36, Buttarello M. Gedotti M, Lorenz C, Toffalori E, Ceschini N, Valentini A, Rizzotti P Evaluation of four automated hematology analyzers. Am J Clin Pathol , Calvelli T, Denny TN, Paxton H, Gelman R, Kagan J: Guideline for flow cytometric immunophenotyping: A report from the National Institute of Allergy and Infectious Disease, Division of AIDS. Cytometry 14:72-715, CDC recommendations for prophylaxis against Pneumocysfis curinii pneumonia for adults and adolescents infected with HIV. MMWR RR4, Vol 41, Centers for Disease Control: Guidelines for the performance of CD4+ T cell determination in persons with HIV infection. MMWR RR8 4l:l-17, Cohen AJ, Peerschke E, Stegbigel RT: A comparison of the Coulter STKS, Coulter S + IV and manual analysis of white blood cell differential counts in a human immunodeficiency virus infected population. Am J Clin Pathol 1:611, Cornbleet J, Myrick D, Judkins S, Levy R: Evaluation of the Cell-Dyn 3 Differential. Am J Clin Pathol 98:63, Cornbleet JR, Myrick D, Levy R: Evaluation of the Coulter STKS five part differential. Am J Clin Pathol , Data recap CAP, Skokie, IL, 1981, p Devreese K, DeLogi E, Francart C, Heyndrickx B, Phillipe J, LeRoux- Roels J: Evaluation of the automated hematology analyser Sysmex NE8. Eur J Clin Chem Clin Biochem , Fahey JL, Taylor JMG, Detels R, Hofmann B, Melmed R, Nishanian P. Giorgi JV: Prognostic value of cellular and serologic markers in infection with human immunodeficiency virus type I. N En@ J Med 322: , Gave HB, Henry K Measuring percent lymphocytes by flow cytometry to calculate absolute lymphocytes subset counts for HIV specimens. Cytometry 13: , Homburger HA, McCarthy R, Deodhar S: Assessment of interlaboratory variability in analytic cytology. Arch Pathol Lab Med I13: , International Committee for Standardization in Hematology: Protocol for evaluation of automated blood cell analyzers. Clin Lab Haematol , Koepke JA, Jones MS: Lymphocyte counting in HIV positive individuals. Sysmex J Int 2:71-74, Koepke JA, Landay AL: Precision and accuracy of absolute lymphocyte counts. Clin Immunol Immunopathol 52:19-27, Landay A: Putting alternative CD4 T cell enumeration methodologies in perspective. Symposium at the IX International Conference on AIDS, June, National Committee for Clinical Laboratory Standards: Clinical applications of flow cytometry: Quality assurance and immunophenotyping of peripheral blood lymphocytes. Tentative Guideline H42T. NCCLS, Villanova, PA, Revised classification system for HIV infection and expanded AIDS surveillance case definition for adolescents and adults. MMWR RR8, Vol 41, Prince HE, Lesar W: Simultaneous determination of absolute total lymphocyte and CD4+ lymphocytes in peripheral blood by flow cytometry. Am J Clin Pathol 92:26-29, National Committee for Clinical Laboratory Standards: Reference leukocyte differential count (proportional) and evaluation of instrumental methods. Approved Standard. NCCLS document H2A, Villanova, PA, Robertson EP, Lai NW, Wei DC: Analysis of leukocyte analysis on the Coulter STKS. Clin Lab Hematol 14:53-68, Robinson G, Robinson G, Morgan L, Evans M, McDermott S, Pereira S, Wainsbrough-Jones M, Griffin G: Effect of type of haematology analyser on CD4 count. Lancet 2:485, Rumke CL: The statistically expected variability in differential leukocyte counting. In: Differential leukocyte counting. Koepke JA (ed). Coll Am Pathol, Skokie, IL, pp 39-45, Van Lieuwen L, Eggels PH, Bullen JA: A short evaluation of a new hematological cell counter. Eur J Clin Chem Clin Biochem 29:15-11, Wen2 B, Burns E: The H'1 hematology analyzer: Its performance characteristics and value in the diagnosis of infectious disease. Am J Pathol Lab Med 111: , 1987.

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer Hematopathology / EVALUATION OF THE ABBOTT CELL-DYN 4 HEMATOLOGY ANALYZER Evaluation of the Abbott CELL-DYN 4 Hematology Analyzer Ernesto Grimaldi, MD, and Francesco Scopacasa, PhD Key Words: Abbott CD

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

Original Article Time and temperature stability of T-cell subsets evaluated by a dual-platform method

Original Article Time and temperature stability of T-cell subsets evaluated by a dual-platform method Am J Blood Res 2012;2(2):128-135 www.ajblood.us /ISSN: 2160-1992/AJBR1204003 Original Article Time and temperature stability of T-cell subsets evaluated by a dual-platform method Horatiu Olteanu 1, Bernard

More information

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Thomas J. Spira, M.D. International Laboratory Branch Global AIDS Program Centers for Disease

More information

Evaluation of a Method for Counting Absolute Numbers of Cells with a Flow Cytometer

Evaluation of a Method for Counting Absolute Numbers of Cells with a Flow Cytometer CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, May 1997, p. 309 313 Vol. 4, No. 3 1071-412X/97/$04.00 0 Copyright 1997, American Society for Microbiology Evaluation of a Method for Counting Absolute Numbers

More information

Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation

Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation Laboratory Hematology 8:85-90 2002 Carden Jennings Publishing Co.. Ltd. Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation T. SCHISANO, L. VAN HOVE Abbott

More information

PointCare NOW TM Technical Note

PointCare NOW TM Technical Note PointCare NOW TM Technical Note Introduction to our Technology And Summary of Evaluation and Performance data 1 of 19 Introduction The PointCare NOW is a portable, and very fast, diagnostic instrument

More information

Method Comparison Report Semi-Annual 1/5/2018

Method Comparison Report Semi-Annual 1/5/2018 Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community

More information

Evaluation of a Single-Platform Technology for Lymphocyte Immunophenotyping

Evaluation of a Single-Platform Technology for Lymphocyte Immunophenotyping CLINICAL AND VACCINE IMMUNOLOGY, Oct. 2007, p. 1342 1348 Vol. 14, No. 10 1556-6811/07/$08.00 0 doi:10.1128/cvi.00168-07 Copyright 2007, American Society for Microbiology. All Rights Reserved. Evaluation

More information

8.2 Principles of Quantitative Hematologic Determinations (1)

8.2 Principles of Quantitative Hematologic Determinations (1) 3 8. FIELD CENTER HEMATOLOGY SERVICES 8.1 Clinical Significance Quantitation of the formed elements of the blood (erythrocytes -RBCs, leukocytes - WBCs, and platelets) is important in the ARIC study primarily

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: BD FACSCount Instrument System with FACSCount Control Kit and BD FACSCount Reagent Kit (Absolute CD4+, CD8+, and CD3+ Counts) Number:

More information

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100 Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100 R. HERKLOTZ and A. R. HUBER Zentrum für Labormedizin, Kantonsspital Aarau, 5001 Aarau, Switzerland. The automated leukocyte(wbc)

More information

CAROL BRIGGS, DONNA GRANT, SAMUEL J. MACHIN INTRODUCTION. Department of Haematology, University College London Hospital, London, United Kingdom

CAROL BRIGGS, DONNA GRANT, SAMUEL J. MACHIN INTRODUCTION. Department of Haematology, University College London Hospital, London, United Kingdom ISLH Laboratory Hematology 7:75-80 2001 Carden Jennings Publishing Co., Ltd. Official Publication Comparison of the Automated Reticulocyte Counts and Immature Reticulocyte Fraction Measurements Obtained

More information

CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, May 2000, p Vol. 7, No. 3

CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, May 2000, p Vol. 7, No. 3 CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, May 2000, p. 344 351 Vol. 7, No. 3 1071-412X/00/$04.00 0 Copyright 2000, American Society for Microbiology. All Rights Reserved. Multisite Comparison of CD4

More information

Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population

Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population Clinical immunology Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population EL BIETA GÓRSKA, URSZULA DEMKOW, ROMAN PIÑKOWSKI, BARBARA JAKUBCZAK,

More information

Technical Bulletin No. 157

Technical Bulletin No. 157 CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 157 January 11, 2017 Flow Cytometry Lymphocyte Subset Analysis Testing Platform Change Change Effective Date: February 6, 2017 Methods

More information

How can Blood Stabilizers be used for Validation of New Instruments? UK NEQAS for Leucocyte Immunophenotyping

How can Blood Stabilizers be used for Validation of New Instruments? UK NEQAS for Leucocyte Immunophenotyping How can Blood Stabilizers be used for Validation of New Instruments? Quality Pathology is essential for about 60% of all diagnoses. Thus, quality of pathology services is critical for good care and management.

More information

Detection of important abnormalities of the differential

Detection of important abnormalities of the differential J Clin Pathol 1989;42:772-776 Detection of important abnormalities of the differential count using the Coulter STKR blood counter D F BARNARD, S A BARNARD, A B CARTER, S J MACHN, K G PATTERSON, A YARDUMAN

More information

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa The Role of the FBC in clinical decision making History Examination Investigations Decision 70% FBC Laboratory

More information

Performance evaluation of Celltac G: a new automated hematology analyzer

Performance evaluation of Celltac G: a new automated hematology analyzer Original Article Performance evaluation of Celltac G: a new automated hematology analyzer Masaaki Sugiyama 1, Tsukasa Kobayashi 1, Yuki Jisyage 2, Shigeko Yamamoto 2, Yutaka Nagai 2,3 and Hiroshi Kondo

More information

Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential

Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential H e m a t o p a t h o l o g y / REFRIGERATION IMPROVES CBC COUNT STABILITY Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential Brent L. Wood, MD, PhD,

More information

Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. ABSTRACT

Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. ABSTRACT OriginalArticle Comparison of Leukocyte Differential Count by Beckman Coulter Unicel DxH800, Beckman Coulter LH780 and Sysmex XE-5000 Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. Department

More information

Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature

Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature Gene L. Gulati, PhD; Lawrence J. Hyland, MD; William Kocher, MD;

More information

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes A. M. CENCI *1, M. MACONI *2, and B. CASOLARI *1 *1 Laboratory of

More information

Application Note. Productivity and Efficiency of 6-Color BD Multitest and BD Trucount Technologies for Enumeration of Lymphocyte Subsets

Application Note. Productivity and Efficiency of 6-Color BD Multitest and BD Trucount Technologies for Enumeration of Lymphocyte Subsets Sept 27 Productivity and Efficiency of 6-Color BD Multitest and BD Trucount Technologies for Enumeration of Lymphocyte Subsets Phillip Ruiz, MD, PhD*, Michael J. Borowitz, MD, PhD**, Henok Tilahun***,

More information

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation.

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Sysmex America White Paper One Nelson C. White Parkway,

More information

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including Week 17 and 21 Comparing two assays and Measurement of Uncertainty 2.4.1.4. Explain tools used to compare the performance of two assays, including 2.4.1.4.1. Linear regression 2.4.1.4.2. Bland-Altman plots

More information

BACKGROUND. Absolute neutrophil counts (ANCs) and absolute phagocyte counts

BACKGROUND. Absolute neutrophil counts (ANCs) and absolute phagocyte counts 2681 Utility of Automated Counting to Determine Absolute Neutrophil Counts and Absolute Phagocyte Counts for Pediatric Cancer Treatment Protocols Nobuko Hijiya, M.D. 1,2 Mihaela Onciu, M.D. 3 Scott C.

More information

Peripheral Blood Monocyte Counting: Towards a New Reference Method

Peripheral Blood Monocyte Counting: Towards a New Reference Method Hübl et al.: Peripheral blood monocyte counting 839 Eur J Clin Chem Clin Biochem 1995; 33:839-845 1995 Walter de Gruyter & Co. Berlin New York Peripheral Blood Monocyte Counting: Towards a New Reference

More information

LYDIA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE, M.D., SANDRA WINGRELD, MT(ASCP), AND EDWARD KEYSER, MT(ASCP)

LYDIA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE, M.D., SANDRA WINGRELD, MT(ASCP), AND EDWARD KEYSER, MT(ASCP) Laboratory Evaluation of Differential White Blood Cell Count nformation from the S-Plus V and H-1 in Patient Populations Requiring Rapid "Turnaround" Time LYDA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE,

More information

Comparison of Methodological Data Measurement Limits in CD4 1 T Lymphocyte Flow Cytometric Enumeration and Their Clinical Impact on HIV Management

Comparison of Methodological Data Measurement Limits in CD4 1 T Lymphocyte Flow Cytometric Enumeration and Their Clinical Impact on HIV Management Cytometry Part B (Clinical Cytometry) 84B:248 254 (2013) Original Article Comparison of Methodological Data Measurement Limits in CD4 1 T Lymphocyte Flow Cytometric Enumeration and Their Clinical Impact

More information

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S 111 Riv Med Lab - JLM, Vol. 4, N. 2, 2003 First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S H. Vandeputte a, D. Darre

More information

Customer Information Literature List White Blood Cells

Customer Information Literature List White Blood Cells Customer Information Literature List White Blood Cells Date: September 2015 Subject: Literature List White Blood Cells Issued by: Scientific Customer Services Number: 150901 Note: Whether references are

More information

Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities

Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities JOANNE CORNBLEET, M.D. AND SALLY KESSINGER, MT(ASCP)SH The authors evaluated the Coulter 15 S-Plus

More information

Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis

Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis Hematopathology / LEFT SHIFT MODEL PREDICTS INFECTION Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis Michael J. Wile, MD, 1 Louis D. Homer, MD, PhD, 2 Stede Gaehler,

More information

The measurements of pleural fluid WBC count

The measurements of pleural fluid WBC count Variations in Pleural Fluid WBC Count and Differential Counts With Different Sample Containers and Different Methods* Barrett D. Conner, MD; Y. C. Gary Lee, MD; Paul Branca, MD, FCCP; Jeffrey T. Rogers,

More information

White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation

White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation ISLH Laboratory Hematology 7:255 266 2001 Carden Jennings Publishing Co., Ltd. Official Publication White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation BERNARD

More information

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags.

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags. WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags www.sysmex.com/us Introduction For many decades, physicians have relied on

More information

Changes of hematological references depends on storage period and temperature conditions in rats and dogs

Changes of hematological references depends on storage period and temperature conditions in rats and dogs ISSN 1738-6055 (Print) ISSN 2233-7660 (Online) Lab Anim Res 2016: 32(4), 241-248 https://doi.org/10.5625/lar.2016.32.4.241 Changes of hematological references depends on storage period and temperature

More information

Establishing a Pure Lymphocyte Gate for Subset Analysis by Flow Cytometry

Establishing a Pure Lymphocyte Gate for Subset Analysis by Flow Cytometry Cytornetry (Communications in Clinical Cytometry) 26:172-177 (1996) Establishing a Pure Lymphocyte Gate for Subset nalysis by Flow Cytometry Joseph M. Homtinovich, Sara D. Sparks, and Karen P. Mann Clinical

More information

Sue Jung Kim, MD, MS; Yoonjung Kim, MD; Saeam Shin, MD; Jaewoo Song, MD, MS; Jong Rak Choi, MD, PhD

Sue Jung Kim, MD, MS; Yoonjung Kim, MD; Saeam Shin, MD; Jaewoo Song, MD, MS; Jong Rak Choi, MD, PhD Comparison Study of the Rates of Manual Peripheral Blood Smear Review From 3 Automated Hematology Analyzers, Unicel DxH 800, ADVIA 2120i, and XE 2100, Using International Consensus Group Guidelines Sue

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital

Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital HEMATOPATHOLOGY Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital JOAN-LLUIS VIVES-CORRONS, MD, ISABEL BESSON, PHD, JOSEP MARIA JOU, MD, AND GABRIELA GUTIERREZ, MB The

More information

The Virtues and Pitfalls of Implementing a New Test

The Virtues and Pitfalls of Implementing a New Test The Virtues and Pitfalls of Implementing a New Test James H. Nichols, Ph.D., DABCC, FACB Professor of Clinical Pathology, Microbiology and Immunology Associate Medical Director for Clinical Operations

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

Evaluation of Monoclonal Antibody Reagents from Three Different Manufacturers Using Flow Cytometric Two-Color Immunophenotyping

Evaluation of Monoclonal Antibody Reagents from Three Different Manufacturers Using Flow Cytometric Two-Color Immunophenotyping ASIAN PACIFIC JOURNAL OF ALLERGYAND IMMUNOLOGY (1998) 16: 185-192 Evaluation of Monoclonal Antibody Reagents from Three Different Manufacturers Using Flow Cytometric Two-Color Immunophenotyping 1 2 2 1

More information

Analytic Performance of the iq200 Automated Urine Microscopy Analyzer and Comparison With Manual Counts Using Fuchs-Rosenthal Cell Chambers

Analytic Performance of the iq200 Automated Urine Microscopy Analyzer and Comparison With Manual Counts Using Fuchs-Rosenthal Cell Chambers Clinical Chemistry / PERFORMANCE OF THE IQ2 URINALYSIS SYSTEM Analytic Performance of the iq2 Automated Urine Microscopy Analyzer and Comparison With Manual Counts Using Fuchs-Rosenthal Cell Chambers David

More information

Performance Characteristics of Eight Estradiol Immunoassays

Performance Characteristics of Eight Estradiol Immunoassays Clinical Chemistry / EVALUATION OF EIGHT ESTRADIOL IMMUNOASSAYS Performance Characteristics of Eight Estradiol Immunoassays David T. Yang, MD, 1 William E. Owen, MT(ASCP), 2 Carol S. Ramsay, 3 Hui Xie,

More information

Comparison between Manual and Automated Methods for Determination of Canine and Feline Hematocrit and Hemoglobin Concentration

Comparison between Manual and Automated Methods for Determination of Canine and Feline Hematocrit and Hemoglobin Concentration Kasetsart J. (Nat. Sci.) 4 : 655-659 (8) Comparison between and d Methods for Determination of Canine and Feline Hematocrit and Hemoglobin Concentration Kreangsak Prihirunkit *, Chalermpol Lekcharoensuk

More information

Chapter 11: Advanced Remedial Measures. Weighted Least Squares (WLS)

Chapter 11: Advanced Remedial Measures. Weighted Least Squares (WLS) Chapter : Advanced Remedial Measures Weighted Least Squares (WLS) When the error variance appears nonconstant, a transformation (of Y and/or X) is a quick remedy. But it may not solve the problem, or it

More information

Do Your Flow Cytometric LDTs. Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine

Do Your Flow Cytometric LDTs. Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine Do Your Flow Cytometric LDTs Conform to the ICSH ICCS Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine How should LDTs be validated? Accuracy Specificity Sensitivity

More information

Farouk AL Quadan lecture 3 Implentation of

Farouk AL Quadan lecture 3 Implentation of How to implement a program? Establish written policies and procedures Assign responsibility for monitoring and reviewing Train staff Obtain control materials Collect data Set target values (mean, SD) Establish

More information

ACS GUIDELINE FOR LYMPHOCYTE SUBSET IMMUNOPHENOTYPING. Second Edition Australasian Cytometry Society Guideline Document

ACS GUIDELINE FOR LYMPHOCYTE SUBSET IMMUNOPHENOTYPING. Second Edition Australasian Cytometry Society Guideline Document ACS GUIDELINE FOR LYMPHOCYTE SUBSET IMMUNOPHENOTYPING Second Edition 2017 Australasian Cytometry Society Guideline Document Paper-based publications This work is copyright. You may reproduce the whole

More information

Simple Linear Regression the model, estimation and testing

Simple Linear Regression the model, estimation and testing Simple Linear Regression the model, estimation and testing Lecture No. 05 Example 1 A production manager has compared the dexterity test scores of five assembly-line employees with their hourly productivity.

More information

Individual Lab Report Ci-Trol Nov,2016. Abnormal Fibrinogen (mg/dl) Abnormal Fbg Control - Lot# LFC Your Lab

Individual Lab Report Ci-Trol Nov,2016. Abnormal Fibrinogen (mg/dl) Abnormal Fbg Control - Lot# LFC Your Lab Individual Lab Report Ci-Trol Nov,2016 ST VINCENT MEDICAL CENTER LABORATORY(LAB# 7300 ) 2131 WEST THIRD STREET LOS ANGELES CA USA 90057 Abnormal Fibrinogen (mg/dl) Abnormal Fbg Control - Lot# LFC SYSMEX

More information

Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer

Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer Science Gene Gulati, PhD, Sean Cote, DO, Eric Behling, MD, William Kocher, MD

More information

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Polymer Technology Systems, Inc. CardioChek PA Comparison Study Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory

More information

METHOD VALIDATION: WHY, HOW AND WHEN?

METHOD VALIDATION: WHY, HOW AND WHEN? METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test

More information

Human Progenitor Cell Enumeration by Flow Cytometry Practical Considerations

Human Progenitor Cell Enumeration by Flow Cytometry Practical Considerations Human Progenitor Cell Enumeration by Flow Cytometry Practical Considerations Bruce Briggs, M.S., S.I. (A.S.C.P.) OSU Medical Center Clinical Flow Cytometry Objectives Conceptual overview: How the Flow

More information

3. Blood Cell Histograms:

3. Blood Cell Histograms: LECTURE MODULE 6c: ELECTRONIC CELL COUNTING PART III 3. Blood Cell Histograms: a. The Coulter cell counters today provides size distributions of the cellular content: 1) volume given in µm 3 or fl vs relative

More information

Eileen Crimmins 1 Jianwei Hu 2 Peifeng Hu 3 Wei Huang 2 Jungki Kim 1 Yuhui Shi 2 John Strauss 1 Lu Zhang 1 Xiaohui Zhao 2 Yaohui Zhao 2

Eileen Crimmins 1 Jianwei Hu 2 Peifeng Hu 3 Wei Huang 2 Jungki Kim 1 Yuhui Shi 2 John Strauss 1 Lu Zhang 1 Xiaohui Zhao 2 Yaohui Zhao 2 CHARLS Pilot: Blood-based Biomarker Documentation Eileen Crimmins 1 Jianwei Hu 2 Peifeng Hu 3 Wei Huang 2 Jungki Kim 1 Yuhui Shi 2 John Strauss 1 Lu Zhang 1 Xiaohui Zhao 2 Yaohui Zhao 2 June 2011 1 University

More information

Application Information Bulletin: Navios Multi-center Study

Application Information Bulletin: Navios Multi-center Study Application Information Bulletin: Navios Multi-center Study Demonstration of the analytical performance of the Navios Flow Cytometer in a multi-center study. Diana Careaga 1, Robert Magari 1, Karen Lo

More information

Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples

Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples Hematopathology / HEMOGLOBIN-BASED BLOOD SUBSTITUTES Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples Jolanta Kunicka, PhD, 1 Michael Malin, PhD, 1 David Zelmanovic,

More information

CRITERIA FOR USE. A GRAPHICAL EXPLANATION OF BI-VARIATE (2 VARIABLE) REGRESSION ANALYSISSys

CRITERIA FOR USE. A GRAPHICAL EXPLANATION OF BI-VARIATE (2 VARIABLE) REGRESSION ANALYSISSys Multiple Regression Analysis 1 CRITERIA FOR USE Multiple regression analysis is used to test the effects of n independent (predictor) variables on a single dependent (criterion) variable. Regression tests

More information

WELCOME! Lecture 11 Thommy Perlinger

WELCOME! Lecture 11 Thommy Perlinger Quantitative Methods II WELCOME! Lecture 11 Thommy Perlinger Regression based on violated assumptions If any of the assumptions are violated, potential inaccuracies may be present in the estimated regression

More information

ACCESS hstni SCIENTIFIC LITERATURE

ACCESS hstni SCIENTIFIC LITERATURE ACCESS hstni SCIENTIFIC LITERATURE 2017 2018 Table of contents Performance Evaluation of Access hstni A critical evaluation of the Beckman Coulter Access hstni: Analytical performance, reference interval

More information

Detection and Classification of Acute Leukemia by the Coulter STKS Hematology Analyzer

Detection and Classification of Acute Leukemia by the Coulter STKS Hematology Analyzer HEMATOPATHOLOGY Detection and Classification of Acute Leukemia by the Coulter STKS Hematology Analyzer JAMES D. HOYER, MD, 1 CATHERINE P. FISHER, MD, 1 VICKI M. SOPPA, BS, 1 KAY L. LANTIS, SH, MT(ASCP),

More information

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%) cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care

More information

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803) Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial

More information

Flow cytometric CD4 enumeration of four different HIV-Infected blood samples at the cost of one monoclonal antibody reagent

Flow cytometric CD4 enumeration of four different HIV-Infected blood samples at the cost of one monoclonal antibody reagent Original article Flow cytometric CD4 enumeration of four different HIV-Infected blood samples at the cost of one monoclonal antibody reagent Egarit Noulsri 1, Surada Lerdwana 1, Kasama Sukapirom 1, Charin

More information

bivariate analysis: The statistical analysis of the relationship between two variables.

bivariate analysis: The statistical analysis of the relationship between two variables. bivariate analysis: The statistical analysis of the relationship between two variables. cell frequency: The number of cases in a cell of a cross-tabulation (contingency table). chi-square (χ 2 ) test for

More information

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Alert or critical values represent those assay results that require prompt, rapid clinical attention to avert significant study-participant

More information

Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions

Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions Clinica Chimica Acta 300 (2000) 13 21 www.elsevier.com/ locate/ clinchim Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions L.V. Rao *, A.O.

More information

Evaluation of Automated Reticulocyte Counts and Their Reliability in the Presence of Howell-Jolly Bodies

Evaluation of Automated Reticulocyte Counts and Their Reliability in the Presence of Howell-Jolly Bodies HEMATOPATHOLOGY Evaluation of Automated Reticulocyte Counts and Their Reliability in the Presence of Howell-Jolly Bodies KAREN G. LOFSNESS, MS, MONICA L. KOHNKE, BS, AND NANCY A. GEIER, BS An automated

More information

Unit 1 Exploring and Understanding Data

Unit 1 Exploring and Understanding Data Unit 1 Exploring and Understanding Data Area Principle Bar Chart Boxplot Conditional Distribution Dotplot Empirical Rule Five Number Summary Frequency Distribution Frequency Polygon Histogram Interquartile

More information

Insulin (Porcine/Canine) ELISA

Insulin (Porcine/Canine) ELISA Insulin (Porcine/Canine) ELISA For the quantitative measurement of insulin in Porcine/Canine serum and plasma (EDTA) For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSPO-E01

More information

Quantitation of T-lymphocyte Subsets Using the CELL-DYN Sapphire Automated Haematology Analyser

Quantitation of T-lymphocyte Subsets Using the CELL-DYN Sapphire Automated Haematology Analyser ORIGINAL ARTICLE http://dx.doi.org/10.15263/jlmqa.2014.36.1.39 Quantitation of T-lymphocyte Subsets Using the CELL-DYN Sapphire Automated Haematology Analyser Jimyung Kim 1, Seung Ryu 2, Jinsook Lim 1,

More information

PanLeucogating Akin Abayomi

PanLeucogating Akin Abayomi PanLeucogating Akin Abayomi Cost effective CD4 counting at a central laboratory. The Barbados experience. CCAS Surinam 2008 Appreciation Frank Mandy George Janossy Phil McCoy John Codrington Clive Landis

More information

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2013

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2013 EXPERTISE, SERVICE PROVISION AND CUSTOMER RELATIONS QUALITY OF MEDICAL LABORATORIES CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS EXTERNAL QUALITY ASSESSMENT IN CLINICAL BIOLOGY DEFINITIVE ANNUAL REPORT

More information

STATISTICS & PROBABILITY

STATISTICS & PROBABILITY STATISTICS & PROBABILITY LAWRENCE HIGH SCHOOL STATISTICS & PROBABILITY CURRICULUM MAP 2015-2016 Quarter 1 Unit 1 Collecting Data and Drawing Conclusions Unit 2 Summarizing Data Quarter 2 Unit 3 Randomness

More information

Bovine Insulin ELISA

Bovine Insulin ELISA Bovine Insulin ELISA For quantitative determination of insulin in bovine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSBO-E01 Size: 96 wells Version:

More information

WDHS Curriculum Map Probability and Statistics. What is Statistics and how does it relate to you?

WDHS Curriculum Map Probability and Statistics. What is Statistics and how does it relate to you? WDHS Curriculum Map Probability and Statistics Time Interval/ Unit 1: Introduction to Statistics 1.1-1.3 2 weeks S-IC-1: Understand statistics as a process for making inferences about population parameters

More information

Introduction: Table/Figure Descriptions:

Introduction: Table/Figure Descriptions: Introduction: We have completed the analysis of your HIV RNA Validation Study. The validation plan was designed to verify the installation of an unmodified FDA-approved HIV RNA assay into your laboratory.

More information

QUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES

QUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES AACC24 QUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES Graham Jones Department of Chemical Pathology, St Vincent s Hospital,

More information

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2014

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2014 EXPERTISE, SERVICE PROVISION AND CUSTOMER RELATIONS QUALITY OF MEDICAL LABORATORIES CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS EXTERNAL QUALITY ASSESSMENT IN CLINICAL BIOLOGY DEFINITIVE ANNUAL REPORT

More information

UK NEQAS Low Level Leucocyte Enumeration Programme. Mr Matthew Fletcher Advanced Biomedical Scientist UK NEQAS for Leucocyte Immunophenotyping

UK NEQAS Low Level Leucocyte Enumeration Programme. Mr Matthew Fletcher Advanced Biomedical Scientist UK NEQAS for Leucocyte Immunophenotyping UK NEQAS Low Level Leucocyte Enumeration Programme Mr Matthew Fletcher Advanced Biomedical Scientist UK NEQAS for Leucocyte Immunophenotyping Content Results of the UKNEQASLI flow cytometry survey Analysis

More information

Flow Cytometric Analysis of T -Lym phocytes Subsets in Adult Thais

Flow Cytometric Analysis of T -Lym phocytes Subsets in Adult Thais j ASAN PACFC JOURNAL OF ALLERGY AND MMUNOLOGY (1997): 15: 141-146 Flow Cytometric Analysis of T -Lym phocytes Subsets in Adult Thais Vicham Vithayasai, Thira Sirisanthana 1, Choompone Sakonwasun and Chontavat

More information

Porcine/Canine Insulin ELISA

Porcine/Canine Insulin ELISA Porcine/Canine Insulin ELISA For the quantitative determination of insulin in porcine or canine serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research

More information

Evaluation of the QBC Star centrifugal three-part differential haematology system

Evaluation of the QBC Star centrifugal three-part differential haematology system ORIGINAL ARTICLE 67 Evaluation of the QBC Star centrifugal three-part differential haematology system O. ERHABOR, G. RICHARDSON, I. MOHAMMED, C. THORNTON, J. BARK, M. HURST, D. HAMER and P. KINSELLA Blood

More information

Sysmex Journal International Vol.26 No.1 (2016)

Sysmex Journal International Vol.26 No.1 (2016) Sysmex Journal International Vol.26 No.1 (216) Performance Evaluation of the XN-55 Automated Hematology Analyzer Body Fluid Mode Considerations for Operational Conditions for Cell Counting with Cerebrospinal

More information

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Rat Insulin ELISA For the quantitative determination of insulin in rat serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: Size: 80-INSRT-E01, E10 96 wells, 10

More information

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,

More information

Patricia C. Fallest-Strobl, 1 Elin Olafsdottir, 2 Donald A. Wiebe, 1 and James O. Westgard 1 * Lipoproteins

Patricia C. Fallest-Strobl, 1 Elin Olafsdottir, 2 Donald A. Wiebe, 1 and James O. Westgard 1 * Lipoproteins Clinical Chemistry 43:11 2164 2168 (1997) Lipids and Lipoproteins Comparison of NCEP performance specifications for triglycerides, HDL-, and LDL-cholesterol with operating specifications based on NCEP

More information

BIOLOGICAL VARIABILITY: THE CIRME CONTRIBUTION. F. Braga Centre for Metrological Traceability in Laboratory Medicine (CIRME)

BIOLOGICAL VARIABILITY: THE CIRME CONTRIBUTION. F. Braga Centre for Metrological Traceability in Laboratory Medicine (CIRME) BIOLOGICAL VARIABILITY: THE CIRME CONTRIBUTION F. Braga Centre for Metrological Traceability in Laboratory Medicine (CIRME) TOTAL VARIATION (CV (CV T ) T ) PREANALYTICAL VARIATION ANALYTICAL VARIATION

More information

MMI 409 Spring 2009 Final Examination Gordon Bleil. 1. Is there a difference in depression as a function of group and drug?

MMI 409 Spring 2009 Final Examination Gordon Bleil. 1. Is there a difference in depression as a function of group and drug? MMI 409 Spring 2009 Final Examination Gordon Bleil Table of Contents Research Scenario and General Assumptions Questions for Dataset (Questions are hyperlinked to detailed answers) 1. Is there a difference

More information

Presentation by Dr. Robert Keller 1998 Conference on the Laboratory Science of HIV

Presentation by Dr. Robert Keller 1998 Conference on the Laboratory Science of HIV Presentation by Dr. Robert Keller 1998 Conference on the Laboratory Science of HIV ABSTRACT It has been previously demonstrated that HIV infection is associated with reduced levels of intracellular GSH.

More information

2.75: 84% 2.5: 80% 2.25: 78% 2: 74% 1.75: 70% 1.5: 66% 1.25: 64% 1.0: 60% 0.5: 50% 0.25: 25% 0: 0%

2.75: 84% 2.5: 80% 2.25: 78% 2: 74% 1.75: 70% 1.5: 66% 1.25: 64% 1.0: 60% 0.5: 50% 0.25: 25% 0: 0% Capstone Test (will consist of FOUR quizzes and the FINAL test grade will be an average of the four quizzes). Capstone #1: Review of Chapters 1-3 Capstone #2: Review of Chapter 4 Capstone #3: Review of

More information

BD Multitest IMK Kit

BD Multitest IMK Kit BD Multitest IMK Kit 50 Tests Catalog No. 340503 50 Tests with BD Trucount Tubes Catalog No. 340504 IVD 2016 BD. BD, the BD Logo and all other trademarks are property of Becton, Dickinson and Company.

More information

Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan

Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan 729488DSTXXX10.1177/1932296817729488Journal of Diabetes Science and TechnologyWestgard et al research-article2017 Original Article Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan Journal

More information