Evaluation of Automated Reticulocyte Counts and Their Reliability in the Presence of Howell-Jolly Bodies

Size: px
Start display at page:

Download "Evaluation of Automated Reticulocyte Counts and Their Reliability in the Presence of Howell-Jolly Bodies"

Transcription

1 HEMATOPATHOLOGY Evaluation of Automated Reticulocyte Counts and Their Reliability in the Presence of Howell-Jolly Bodies KAREN G. LOFSNESS, MS, MONICA L. KOHNKE, BS, AND NANCY A. GEIER, BS An automated reticulocyte procedure using a flow cytometer and the fluorescent dye thiazole orange was evaluated for clinical use. The mean reticulocyte count on 118 hematologically healthy adults was 1.56% (standard deviation SD] 0.54), with virtually no difference in percentage between women and men. The mean absolute values were 68.4 X 1O'/L (SD 24.6) and 75.7 X 10»/L (SD 27.2), respectively. When compared with the standard microscopic technique, the automated method showed excellent correlation (r = 0.98) and greatly improved precision (coefficient of variation CV] 4.1%) over the manual method (CV 22.8%). Preanalytic storage of blood samples at 4 C for up to 48 hours did not significantly affect results, nor did varying the incubation time of diluted samples from */i to 2 hours. In a group of patients with appreciable numbers of Howell-Jolly bodies, automated reticulocyte counts were spuriously elevated. The difference between the manual and automated counts on these patients approximated the percentage of Howell-Jolly bodies observed on their Wright-Giemsa stained blood smears. (Key words: Automated counting; Flow cytometry; Howell-Jolly bodies; Reticulocytes) Am J Clin Pathol 1994;101: Reticulocytes are immature red blood cells (RBC) that have recently left the bone marrow and still contain intracellular RNA. The effectiveness of erythropoiesis can be measured by staining blood with a supravital dye and determining the number of RBCs that contain this residual RNA. Reticulocyte counts provide clinically useful information, not only in the diagnosis and classification of anemias, but also in monitoring therapeutic response. The most widely used method for counting reticulocytes is a manual microscopic procedure. 1 Although it is inexpensive and relatively simple to perform, this method is labor intensive and imprecise. Individual differences in the preparation and staining of the sample, distributional variation on the smear, the limited number of cells actually counted, and interobserver subjectivity in identifying reticulocytes all contribute to the poor reproducibility of the manual procedure. Recently, automated reticulocyte methods, using flow cytometry and fluorescent dyes that bind RNA, have shown improved precision and cost effectiveness. 2 " 12 However, most of the currently used fluorescent dyes bind polynucleotides nonspecifically, and thus stain DNA as well as RNA.' 3 Nucleated cells, because of their high DNA content, have fluorescent activity that is many times greater than that of reticulocytes and can usually be excluded or "gated out" from the RBC population. Small lymphocytes, such as those seen in chronic lymphocytic leukemia, occasionally cause interference. Although the rela- From the Department of Laboratory Medicine and Pathology, University of Minnesota, Minneapolis. Minnesota. Manuscript received October 8, 1992; revision accepted February 1, Address correspondence to Ms. Lofsness: Box 198 UMHC, University of Minnesota, Minneapolis, MN tive concentration of nucleated RBCs and erythrocytes containing malarial parasites or Howell-Jolly bodies is usually very low, these inclusions also fluoresce and have the potential to spuriously elevate an automated reticulocyte count. The objectives of this study were to evaluate the Coulter Epics Profile II flow cytometer (Coulter Electronics, Hialeah, FL) for reticulocyte counting in a routine hematology laboratory, to use this instrument to establish reference values, and to study the effect of significant numbers of Howell-Jolly bodies on automated reticulocyte counts. MATERIALS AND METHODS Automated Reticulocyte Counting Automated reticulocyte counts were performed on a Coulter Epics Profile II flow cytometer, using a modification of the method of Lee and coworkers. 13 A 1:10,000 working solution of thiazole orange was prepared fresh daily by diluting 5 jul stock solution (1 mg/ml thiazole orange in methanol) in 50 ml Isoton II (Coulter Electronics, Hialeah, FL); the solution was then stored in a brown bottle. For each blood specimen, two dilutions (labeled "blank" and "thiazole orange") were prepared in 10 X 75 mm capped polypropylene tubes. Each blank tube contained 5 nl blood diluted in 2.0 ml Isoton II, and each thiazole orange tube contained 5 nl blood diluted in 2.0 ml working thiazole orange solution. Both tubes were mixed well and incubated in the dark at room temperature for more than 30, but less than 120, minutes before analysis. The tubes were remixed two to three times during this incubation period, and again immediately before counting. The following instrument settings were used for reticulocyte counting. The laser was set at 15 m W. The green fluorescence 85

2 86 HEMATOPATHOLOGY of 50,000 RBC (measured with 525 nm band pass and 550 nm short pass dichroic filters) was collected on a 256 channel 4 generation log analyzer. Forward scatter and log side scatter gates were used to encompass the RBC in a bitmap. The high voltage and gain were set so that 99.9% of thefluorescenceof an unstained sample fell between channels 0 and 50. The sample flow rate was set at 50 ^L/min ( cells/sec). The instrument start-up procedure recommended by the manufacturer was performed daily, and any necessary adjustments were made before the analysis of the blood samples. For each determination, the blank sample was analyzed first, and results were checked to ensure that the background count was approximately 50 events, the reported percentage was 0.1%, and the histogram did not appear abnormal. The range bar cursor was adjusted if necessary, and the thiazole orange sample was analyzed using the same settings as the blank (Figure 1). To correct for the background count, 0.1% was subtracted from the reticulocyte percentage obtained on the thiazole orange sample. Manual Reticulocyte Counting Equal amounts of blood and new methylene blue were combined and incubated at room temperature for 5-15 minutes. The dilutions were remixed, and three wedge smears were prepared from each sample. The number of reticulocytes per 500 erythrocytes was determined microscopically by each of two technologists on separate slides (1000 cells total). If the number of reticulocytes did not agree withinfivecells, a third technologist counted 500 cells on the third slide, and results were averaged. A reticulocyte was defined as an RBC containing at least two granules of reticulum. Specimen Collection and Method Evaluation Venous blood samples were collected in Na 2 EDTA (1.5 mg/ ml), and reticulocyte counts were determined as described. Complete blood counts (CBC) were performed by either a Coulter S-Plus IV or STKR (Coulter Electronics, Hialeah, FL), and blood smears were prepared and stained by a Wright- Giemsa method. Results were analyzed statistically using Statview II software (Abacus Concepts, Berkeley, CA). Healthy Subjects Reference values were determined by performing automated reticulocyte counts on blood samples from 118 healthy nonhospitalized subjects, aged 16 to 80 years. All 67 women and 51 men were judged to be hematologically healthy on the basis of their automated CBC. Absolute reticulocyte counts were calculated by multiplying the relative count (in percent) by the RBC count determined on the same blood sample. Patient Samples For correlation studies, 215 random blood samples from hospitalized patients, on whom reticulocyte counts had been ordered, were analyzed by both the manual and automated methods. Reproducibility was evaluated by performing three replicate analyses on 10 patient samples by both the manual and automated methods. To determine the effect of preanalytic specimen storage time on automated reticulocyte results, 15 patient blood samples were first analyzed within 4 hours of B MR MAX CODNT LFL1 PBRCBHT HEAR SD I gpcv MIR MAI 1 O.ÍOTTÍÍX LFL1 COURT PEBCEHT 1527 rr HEAR TUT SO ÜTÍ7T \ BPCJ FIG. 1. Histograms of blank (A) and thiazole orange (B) dilutions. Cell number is shown on the vertical axis (count), andfluorescentintensity, plotted on a four-generation log scale, is shown on the horizontal axis (LFL1). The range bar cursor [1] is set so that the event count on the blank is around 50, and the percentage of cells showingfluorescenceis 0.1%. The thiazole orange sample is analyzed using the same settings, and 0.1 % is subtracted from the result. This patient's reticulocyte count was 4.9%. collection, then stored at 4 C and reanalyzed at 24 and 48 hours. The stability of the fluorescence in the blood-working thiazole orange dilution was measured by comparing the results on 10 patient samples after 'A, 1, and 2 hours incubation time. Forty-one blood samples from 19 patients, on which Howell- Jolly bodies had been observed microscopically, were used to study the potential interference of these inclusions on automated reticulocyte counts. The automated and manual reticulocyte percentages on these patients were obtained by the methods described above. The percentage of Howell-Jolly bodies was determined independently on a Wright-Giemsa stained AJ.C.P.-January 1994

3 TABLE 1. AUTOMATED RETICULOCYTE REFERENCE VALUES LOFSNESS, KOHNKE, AND GEIER 87 Automated Reticulocyte Counts Mean ±2SD Mean (XIO'/L) ±2SD 118 adults (ages yr) women men s 5 blood smear of the same sample by having two technologists count the number of RBCs containing Howell-Jolly bodies per 500 cells (1000 cells total) in the manner of a manual reticulocyte count. RESULTS Reference Values and Correlation Studies Reticulocyte results for 118 hematologically healthy adults are shown in Table 1. The mean reticulocyte percentage was 1.56%, with a standard deviation (SD) of 0.54%. The average absolute reticulocyte count was 71.5 X 10 9 /L (SD 25.9). For the 67 women, the means were 1.56% (SD 0.55) and 68.4 X 10 9 /L (SD 24.6), with observed ranges of 0.4% to 3.4% and 17 to 136 X 10 9 /L, respectively. The 51 men had means of 1.55% (SD 0.54) and 75.7 X 10 9 /L (SD 27.2), and observed ranges of 0.6% to 2.9% and 30 to 144 X 10 9 /L. Figure 2 shows the frequency distribution of reticulocyte percentages. Because this distribution was not Gaussian, the 95% reference interval was also determined by percentile rank. When the lowest 2.5% and highest 2.5% of the values were excluded, the range was 0.6% to 2.7%. Figure 3 illustrates the frequency distribution of absolute reticulocyte values for these 118 subjects. The correlation between manual and automated reticulocyte results on 215 patient samples with manual percentages ranging from 0.1% to 28.2% is shown as Figure 4. By linear regression analysis, the coefficient of correlation (r) was The coefficient of determination (i 2 ) was 0.95, and y = 0.92x Relic absolute « FIG. 3. Frequency distribution of absolute reticulocyte values (X 10 9 /L) by flow cytometer on 118 healthy adults (67 women, 51 men). Reproducibility and Stability The 10 patient samples used to evaluate reproducibility had automated counts ranging from 1.2% to 9.3%. For the automated method, the average coefficient of variation (CV) was 4.1%, with a range of 0.9% to 7.7%. By the manual method, the same samples showed a mean CV of 22.8% (range 11.1% to 36.1%). The 15 random bloods used to study the effect of preanalytic specimen storage time had initial automated reticulocyte values ranging from 1.0% to 10.3%, with a mean of 3.3%. After 24 hours of storage at 4 C, the results ranged from 0.8% to 10.3%, with a mean of 3.4%. Only one 24-hour sample varied as much as 0.5% from the original value. After 48 hours at 4 C, the mean was still 3.4% and the range was 0.9% to 10.3%. One 48-hour sample was 0.7% higher than the initial result; the rest were all within ± 0.5%. These differences were not statistically significant by paired /-tests. The 10 samples that were analyzed after l h, 1, and 2 hours incubation in the working thiazole orange solution had initial automated reticulocyte percentages of 1.1% to 6.8%, with a mean of 2.4%. After 1 hour of incubation, they ranged from 1.0% to 6.5% (mean 2.4%), and after 2 hours the range was 1.0% to 6.4% (mean 2.4%). One 2-hour dilution was 0.4% lower than the initial value; the rest were all within ± 0.2%. These differences were not statistically significant by paired i-tests. y Six *.17, r > 48, r-aqu«i*d m M (n 21S) ,w,..,VK.i...v..., Relic % FIG. 2. Frequency distribution of reticulocyte percentages byflowcytometer on 118 healthy adults. FIG. 4. Flow cytometer versus manual method, reticulocyte percentages. Scattergram and regression coefficients. Vol. 101 No. I

4 88 HEMATOPATHOLOO T TABLE 2. EFFECT OF HOWELL-JOLLY BODIES Patient No n HJ (%) Manual reticulocyte (%) HJ + manual reticulocyte (%) Automated reticulocyte (%) l OS "> ? ? I HJ = Howell-Jolly bodies. Effect of Howell-Jolly Bodies Forty-one samples in which Howell-Jolly bodies were easily found on the Wright-Giemsa smear were selected for reticulocyte analysis by both the manual and automated methods. These samples had manual reticulocyte counts ranging from 0.2% to 19.0%, and automated counts from 0.7% to 24.3%. In every case, the automated percentage was higher than the manual. The percentage of Howell-Jolly bodies was determined microscopically and ranged from 0.2 to 8.8%. When this percentage was added to the manual reticulocyte count, the result approximated the automated count in most cases. Table 2 shows data from the initial samples of the 19 patients with increased Howell-Jolly bodies. The flow cytometer histogram patterns on these samples with Howell-Jolly bodies did not appear to show any unusual characteristics (Fig. 5). When the thiazole orange dilutions from one of the Howell-Jolly samples was examined by fluorescent microscopy, the Howell-Jolly bodies appeared as bright green discrete intracellular dots, whereas the reticulocytes were diffusely fluorescent. DISCUSSION The clinical utility of the reticulocyte count is compromised by the fact that, in most laboratories, it is still determined microscopically. Although other routine hematologic counts are NIK COOHT LFL1 PERCENT HEAR SD \ HPCV FIG. 5. Histogram of thiazole orange dilution on a patient with an automated reticulocyte count of 5.4%, a manual count of 2.2%, and a Howell-Jolly body count of 2.5%. now performed by complex instruments that give highly accurate and reproducible results, the standard method for counting reticulocytes has not changed for many years and remains imprecise and unreliable. 14 ' 15 The recent development of practical techniques usingflowcytometry and fluorescent dyes will allow many clinical laboratories to automate reticulocyte counting and bring it to an acceptable level of reliability. On the basis of an internal study performed in 1966, the reference range for reticulocytes in our laboratory was considered to be 0.5% to 3.3% for women and 0.5% to 2.1% for men. As a result of the present study, the range has been adjusted to 0.5% to 2.7% for both sexes. Although we found virtually no difference between the reticulocyte percentages for women and men, reports in the literature vary on this point. Earlier studies by both manual 16 and automated 4 methods describe significantly higher reticulocyte percentages in women than men. However, other authors report no appreciable difference by flow cytometry, 2 ' 3 ' 7 and one study of 265 adults actually found a slightly lower mean percentage in women (0.92%) than in men (1.07%)." Because method comparison for reticulocyte counting necessitates using the admittedly imprecise manual technique as the reference by which to judge a new automated system, it is somewhat surprising to find a high degree of correlation between the two methods, especially in the higher clinical ranges. Our correlation and reproducibility data were similar to those reported by others, 4-6 ' 7 ' 10 ' 12 and showed that the automated method agreed closely with the manual, and was far more precise. The greatly improved precision obtained withflowcytometry reflects not only the much larger number of cells counted (~ 50,000 versus 1000 with the microscope), but also the elimination of subjectivity on the part of the observer. In a preliminary report from the CAP reticulocyte project, Savage and coworkers 15 recommend that the r value for an automated reticulocyte system should be greater than 0.95, with a y intercept of 0.25 or less when correlated with manual reticulocyte results. They also state that the system CV should be 15% or less. The automated system evaluated in this study, with r equal to 0.98, a y intercept of 0.17, and a CV of 4.1%, easily meets these criteria. The stability of this system was not affected by specimen storage time, either before analysis or following dilution in thiazole orange. Although some studies of sample storage reported a progressive fall in reticulocytes after blood was stored more than 24 hours at 4 C, 4-9 we found no significant differences up to 48 hours. Because refrigerated blood may be held as long as 2 days without appreciable changes in reticulocyte values, routine samples can be batched and run every other day, with specimens requiring immediate attention being performed by the manual method. Selected samples can also be stored at 4 C and reanalyzed with the next batch as a quality control check of system reproducibility. Because the thiazole A.J.C.P.-January 1994

5 LOFSNESS, KOHNKE, AND GEIER 89 Automated Reticulocyte Counts LFL1 HIN HAX COUNT PERCENT HEAD SD % HPCV T 27T TTJOT X5IT FIG. 6. Histogram of thiazole orange dilution on patient with chronic lymphocytic leukemia. The second peak caused by the WBC (200 X 10 9 /L) is evident. The initial range bar cursor setting [1] included some of the WBC, and gave a reticulocyte value of 2.6%. Range bar cursor setting [2] was extended to include all of the WBC, whereas setting [3] excluded the WBC peak and gave afinalresult of 2.2%. orange incubation timing is somewhatflexible,individual samples do not have to be analyzed in any specific order. Samples are easily added or repeated, and batches may be even interrupted for a reasonable period of time. Thiazole orange and the otherfluorescentdyes used in automated reticulocyte counting bind to DNA as well as RNA, thus nuclear material can potentially cause interference. Most leukocytes are so large and highly fluorescent that they are easily excluded from the RBC population being analyzed. Occasionally, it is difficult to "gate out" the small lymphocytes seen in chronic lymphocytic leukemia. However, the second peak they produce is evident on the histogram, and adjustment of the bitmap and the upper range bar cursor will usually provide acceptable results (Fig. 6). Large numbers of nucleated RBC would probably require similar adjustments. Erythrocytes containing smaller inclusions of DNA, such as malarial parasites or Howell-Jolly bodies, are more similar to reticulocytes in both size and fluorescent activity. When present in appreciable numbers, these RBC inclusions have the potential to elevate automated reticulocyte counts incorrectly. A recently published method for quantitating malarial parasites in in vitro culture systems actually uses flow cytometry with thiazole orange as the binding dye. 17 Some studies evaluating automated reticulocyte counting mention the possibility of spurious results in samples containing Howell-Jolly bodies. 78 However, others state that either Howell-Jolly bodies do not interfere, 2 ' 10 '" or their concentration is generally so low that they have no significant effect. 61 ' 8 Our investigation of the interference caused by Howell-Jolly bodies was prompted by a patient sample that had a manual reticulocyte count of 1.8% and an automated count of 6.0%. Howell-Jolly bodies were noted on the new methylene blue preparation, and examination of the patient's Wright-Giemsa stained blood smear showed several Howell-Jolly bodies in everyfield.by counting the number of Howell-Jolly bodies per 1000 RBC in the manner of a manual reticulocyte count, 4.1% Howell-Jolly bodies were found on this blood smear. Because the total number of microscopically determined reticulocytes and Howell-Jolly bodies so closely approximated the automated reticulocyte count, it was decided to select samples with noticeable numbers of Howell-Jolly bodies deliberately to see if this pattern persisted. Although this method of determining the percentage of Howell-Jolly bodies is admittedly as imprecise as the manual reticulocyte method itself, in every case the automated count was demonstrated to be higher than the manual, and the numeric difference between the two was somewhat close to the Howell-Jolly percentage. Howell-Jolly bodies are nuclear remnants that may be seen in erythrocytes following splenectomy. They may also be seen in hemolytic anemia, megaloblastic anemia, and hyposplenic states. 19 The question might well be raised as to the prevalence of samples containing appreciable numbers of Howell-Jolly bodies in a patient population. The samples in this study were from 19 patients who were seen in our 550-bed tertiary care hospital over a span of 4 weeks. Because such samples are not rare we recommend screening Wright-Giemsa blood smears on all samples submitted for reticulocyte counts, and performing manual counts on those that have more than one Howell-Jolly body per oil immersion (X 1000) field. Automated reticulocyte counting, which has consistently proven superior to the manual technique, should eventually replace the manual technique as the method of choice. In our experience, agreement between the two methods has been good, except for the occasional sample containing Howell-Jolly bodies or otherfluorescinginterferences. Errors associated with these factors can be minimized by carefully monitoring histograms and screening Wright-Giemsa smears for RBC inclusions. Although automated reticulocyte counting may not be feasible for smaller laboratories with low test volume, it should be a practical alternative for many clinical laboratories, particularly those that have already acquired aflowcytometer to perform immunologic marker studies. REFERENCES 1. National Committee for Clinical Laboratory Standards. Method for reticulocyte counting: Proposed standard. Villanova, Pa: NCCLS, 1985, Publication H16-P. 2. Batjer JD, Lin LI, Fritsma GA. Stability and reproducibility of reticulocyte analysis by flow cytometry. Hematology monograph. McGaw Park, 111: Baxter Diagnostics, 1991, pp Bowen D, Bentley N, Hoy T, Cavill I. Comparison of a modified thiazole orange technique with a fully automated analyser for reticulocyte counting. J din Pathol 1991;44: Carter JM, McSweeney PA, Wakem PJ, Nemet AM: Counting reticulocytes by flow cytometry: use of thiazole orange. Clin Lab Haemal 1989; 11: Corash L, Rheinschmidt M, Lieu S, Meers P, Brew E. Enumeration of reticulocytes usingfluorescence-activatedflowcytometry. Pathol Immunopathol Res 1988;7: Davis BH, Bigelow NC. Flow cytometric reticulocyte quantification provides clinically useful reticulocyte maturity index. Arch Pathol Lab Med 1989; 113: Ferguson DJ, Lee S, Gordon PA. Evaluation of reticulocyte counts by flow cytometry in a routine laboratory. Am J Hematol 1990;33: Vol. 101-No. 1

6 90 HEMATOPATHOLOGY 8. Hackney JR, Cembrowski GS, Prystowsky MB, Kant JA. Automated reticulocyte counting by image analysis andflowcytometry. LabMed 1989;20: Kojima K, Niri M, Setoguchi K, Tsuda I, Tatsumi N. An automated optoelectric reticulocyte counter. Am J Clin Palhol 1989:92: Nobes PR, Carter AB. Reticulocyte counting usingflowcytometry. J Clin Pathol 1990;43: Tatsumi N, Tsuda I, Kojima K, Niri M, Setoguchi K. An automated reticulocyte counting method: Preliminary observations. MedLabSci 1989;46: Tichelli A, Gratwohl A, Driessen A, et al. Evaluation of the Sysmex R-1000: An automated reticulocyte analyzer. Am J Clin Pathol 1990;93: Lee LG, Chen C, Chu LA. Thiazole orange: A new dye for reticulocyte analysis. Cylometry 1986; 7: Peebles DA, Hochberg A, Clarke TD. Analysis of manual reticulocyte counting. Am J Clin Pathol 1981; 76: Savage RA, Skoog DP, Rabinovitch A. Analytic inaccuracy and imprecision in reticulocyte counting: A preliminary report from the College of American Pathologists reticulocyte project. Blood Cells 1985:11: Deiss A, Kurth D. Circulating reticulocytes in normal adults as determined by the new methylene blue method. Am J Clin Pathol 1970:53: Makler MT, Lee LG, Recktenwald D. Thiazole orange: A new dye for Plasmodium species analysis. Cytometry 1987; 8: Davis BH, Bigelow NC. Clinicalflowcytometric reticulocyte analysis. Pathobiology 1990; 58: Bull BS, Breton-Gorius J, Beutler E. Morphology of the erythron. In: Williams WJ, Beutler E, Erslev A, Lichtman MA, eds. Hematology. 4th ed. New York, NY: McGraw-Hill, 1990, p A.J.C.P.-January 1994

Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation

Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation Laboratory Hematology 8:85-90 2002 Carden Jennings Publishing Co.. Ltd. Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation T. SCHISANO, L. VAN HOVE Abbott

More information

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer Hematopathology / EVALUATION OF THE ABBOTT CELL-DYN 4 HEMATOLOGY ANALYZER Evaluation of the Abbott CELL-DYN 4 Hematology Analyzer Ernesto Grimaldi, MD, and Francesco Scopacasa, PhD Key Words: Abbott CD

More information

CAROL BRIGGS, DONNA GRANT, SAMUEL J. MACHIN INTRODUCTION. Department of Haematology, University College London Hospital, London, United Kingdom

CAROL BRIGGS, DONNA GRANT, SAMUEL J. MACHIN INTRODUCTION. Department of Haematology, University College London Hospital, London, United Kingdom ISLH Laboratory Hematology 7:75-80 2001 Carden Jennings Publishing Co., Ltd. Official Publication Comparison of the Automated Reticulocyte Counts and Immature Reticulocyte Fraction Measurements Obtained

More information

Role of immature reticulocyte fraction in evaluation of aplastic anemia in cases of pancytopenia

Role of immature reticulocyte fraction in evaluation of aplastic anemia in cases of pancytopenia Original article Role of immature reticulocyte fraction in evaluation of aplastic anemia in cases of pancytopenia Dr. R. Sindhu, S.K. Behera, D.P. Mishra Department of Pathology, MKCG medical college,

More information

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa The Role of the FBC in clinical decision making History Examination Investigations Decision 70% FBC Laboratory

More information

History. n 1865 First description of reticulocytes by Erb, who. n 1881 Using supravital staining, Ehrlich demonstrated

History. n 1865 First description of reticulocytes by Erb, who. n 1881 Using supravital staining, Ehrlich demonstrated SYSMEX EDUCATIONAL ENHANCEMENT AND DEVELOPMENT JANUARY 2016 SEED HAEMATOLOGY The importance of reticulocyte detection Production of reticulocytes All blood cells emanate from a stem cell. Under pronounced

More information

8.2 Principles of Quantitative Hematologic Determinations (1)

8.2 Principles of Quantitative Hematologic Determinations (1) 3 8. FIELD CENTER HEMATOLOGY SERVICES 8.1 Clinical Significance Quantitation of the formed elements of the blood (erythrocytes -RBCs, leukocytes - WBCs, and platelets) is important in the ARIC study primarily

More information

Performance evaluation of Celltac G: a new automated hematology analyzer

Performance evaluation of Celltac G: a new automated hematology analyzer Original Article Performance evaluation of Celltac G: a new automated hematology analyzer Masaaki Sugiyama 1, Tsukasa Kobayashi 1, Yuki Jisyage 2, Shigeko Yamamoto 2, Yutaka Nagai 2,3 and Hiroshi Kondo

More information

Sysmex Journal International Vol.26 No.1 (2016)

Sysmex Journal International Vol.26 No.1 (2016) Sysmex Journal International Vol.26 No.1 (216) Performance Evaluation of the XN-55 Automated Hematology Analyzer Body Fluid Mode Considerations for Operational Conditions for Cell Counting with Cerebrospinal

More information

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Thomas J. Spira, M.D. International Laboratory Branch Global AIDS Program Centers for Disease

More information

2008 CAP TODAY Q & A

2008 CAP TODAY Q & A 2008 CAP TODAY Q & A Q. How often should we document hematology competencies to ensure consistency of morphologic observations for manual differentials and fluids? Should we do this every six months or

More information

EDUCATIONAL COMMENTARY BLOOD CELL IDENTIFICATION

EDUCATIONAL COMMENTARY BLOOD CELL IDENTIFICATION EDUCATIONAL COMMENTARY BLOOD CELL IDENTIFICATION Educational commentary is provided through our affiliation with the American Society for Clinical Pathology (ASCP). To obtain FREE CME/CMLE credits click

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

CELL-DYN 3700 Strength in Technology, Proven Reliability

CELL-DYN 3700 Strength in Technology, Proven Reliability CELL-DYN 3700 Strength in Technology, Proven Reliability Optical WBC Technology Patented MAPSS Differential Multiple Technologies CELL-DYN 3700 Multiple Technologies One Superior Result Multiple Technologies

More information

RED BLOOD CELLS IMPORTANCE OF FILM EXAMINATION: PART ONE

RED BLOOD CELLS IMPORTANCE OF FILM EXAMINATION: PART ONE Vet Times The website for the veterinary profession https://www.vettimes.co.uk RED BLOOD CELLS IMPORTANCE OF FILM EXAMINATION: PART ONE Author : Mark Richer Categories : Vets Date : March 25, 2013 Mark

More information

Hematology Unit Lab 1 Review Material

Hematology Unit Lab 1 Review Material Hematology Unit Lab 1 Review Material - 2018 Objectives Laboratory instructors: 1. Assist students during lab session Students: 1. Review the introductory material 2. Study the case histories provided

More information

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation.

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Sysmex America White Paper One Nelson C. White Parkway,

More information

CELL-DYN Strength in Technology, Proven Reliability. Optical WBC Technology. Patented M.A.P.S.S. Differential. Multiple Technologies

CELL-DYN Strength in Technology, Proven Reliability. Optical WBC Technology. Patented M.A.P.S.S. Differential. Multiple Technologies CELL-DYN 3700 Strength in Technology, Proven Reliability Optical WBC Technology Patented M.A.P.S.S. Differential Multiple Technologies CELL-DYN 3700 Multiple Technologies One Superior Result Multiple Technologies

More information

The Evolving Role of Reticulated Platelets

The Evolving Role of Reticulated Platelets The Evolving Role of Kyle Riding, PhD, MLS(ASCP) CM University of Central Florida Session Objectives Platelets Recall the process involved in megakaryopoiesis and thrombocytopoiesis Explain the various

More information

Blood DLC, Retic count, PCV, Hb and ESR. Dr. Tamara Alqudah

Blood DLC, Retic count, PCV, Hb and ESR. Dr. Tamara Alqudah Blood DLC, Retic count, PCV, Hb and ESR Dr. Tamara Alqudah Differential Leukocyte Count (DLC) There are 5 main types of WBCs: 1. Neutrophils: 40-80% 2. Eosinophils: 1-6 % 3. Basophils: < 1-2% 4. Lymphocytes:

More information

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes A. M. CENCI *1, M. MACONI *2, and B. CASOLARI *1 *1 Laboratory of

More information

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100 Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100 R. HERKLOTZ and A. R. HUBER Zentrum für Labormedizin, Kantonsspital Aarau, 5001 Aarau, Switzerland. The automated leukocyte(wbc)

More information

Detection and Classification of Acute Leukemia by the Coulter STKS Hematology Analyzer

Detection and Classification of Acute Leukemia by the Coulter STKS Hematology Analyzer HEMATOPATHOLOGY Detection and Classification of Acute Leukemia by the Coulter STKS Hematology Analyzer JAMES D. HOYER, MD, 1 CATHERINE P. FISHER, MD, 1 VICKI M. SOPPA, BS, 1 KAY L. LANTIS, SH, MT(ASCP),

More information

3. Blood Cell Histograms:

3. Blood Cell Histograms: LECTURE MODULE 6c: ELECTRONIC CELL COUNTING PART III 3. Blood Cell Histograms: a. The Coulter cell counters today provides size distributions of the cellular content: 1) volume given in µm 3 or fl vs relative

More information

Performance Evaluation of the CellaVision DM96 System WBC Differentials by Automated Digital Image Analysis Supported by an Artificial Neural Network

Performance Evaluation of the CellaVision DM96 System WBC Differentials by Automated Digital Image Analysis Supported by an Artificial Neural Network Hematopathology / DIGITAL IMAGE ANALYSIS IN HEMATOLOGY Performance Evaluation of the CellaVision DM96 System WBC Differentials by Automated Digital Image Analysis Supported by an Artificial Neural Network

More information

ADVANCED HAEMATOLOGY BATTLE OF THE BANDS. Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND

ADVANCED HAEMATOLOGY BATTLE OF THE BANDS. Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND ADVANCED HAEMATOLOGY BATTLE OF THE BANDS Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND The identification of immature neutrophils (bands, metamyelocytes,

More information

The measurements of pleural fluid WBC count

The measurements of pleural fluid WBC count Variations in Pleural Fluid WBC Count and Differential Counts With Different Sample Containers and Different Methods* Barrett D. Conner, MD; Y. C. Gary Lee, MD; Paul Branca, MD, FCCP; Jeffrey T. Rogers,

More information

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S 111 Riv Med Lab - JLM, Vol. 4, N. 2, 2003 First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S H. Vandeputte a, D. Darre

More information

Evaluation of a Method for Counting Absolute Numbers of Cells with a Flow Cytometer

Evaluation of a Method for Counting Absolute Numbers of Cells with a Flow Cytometer CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, May 1997, p. 309 313 Vol. 4, No. 3 1071-412X/97/$04.00 0 Copyright 1997, American Society for Microbiology Evaluation of a Method for Counting Absolute Numbers

More information

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags.

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags. WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags www.sysmex.com/us Introduction For many decades, physicians have relied on

More information

Lavender Top Management SUCCESS BEYOND FINANCES

Lavender Top Management SUCCESS BEYOND FINANCES Lavender Top Management Alfonso Ziccardi Ass.Lab.Operation Manager for AP/CP/Safety Officer New York Hospital Medical Center of Queens SUCCESS BEYOND FINANCES 1 . VISION Mission/ Vision To successfully

More information

Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature

Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature Gene L. Gulati, PhD; Lawrence J. Hyland, MD; William Kocher, MD;

More information

Lavender Top Management SUCCESS BEYOND FINANCES. Mission/ Vision

Lavender Top Management SUCCESS BEYOND FINANCES. Mission/ Vision Lavender Top Management Alfonso Ziccardi Lab.Operation Manager for AP/CP/Safety Officer New York Hospital Medical Center of Queens SUCCESS BEYOND FINANCES Mission/ Vision. VISION To successfully develop

More information

The LaboratoryMatters

The LaboratoryMatters Laboratory Medicine Newsletter for clinicians, pathologists & clinical laboratory technologists. A Initiative. Complete Blood Count This issue highlights: CBC, while ubiquitous, is an excellent diagnostic

More information

LYDIA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE, M.D., SANDRA WINGRELD, MT(ASCP), AND EDWARD KEYSER, MT(ASCP)

LYDIA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE, M.D., SANDRA WINGRELD, MT(ASCP), AND EDWARD KEYSER, MT(ASCP) Laboratory Evaluation of Differential White Blood Cell Count nformation from the S-Plus V and H-1 in Patient Populations Requiring Rapid "Turnaround" Time LYDA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE,

More information

Validation of the Efficacy of a Practical Method for Neutrophils Isolation from Peripheral Blood

Validation of the Efficacy of a Practical Method for Neutrophils Isolation from Peripheral Blood Validation of the Efficacy of a Practical Method for Neutrophils Isolation from Peripheral Blood JONATHAN DEGEL, MASIH SHOKRANI OBJECTIVE: The objectives of this study were to validate the Polymorphprep

More information

BACKGROUND. Absolute neutrophil counts (ANCs) and absolute phagocyte counts

BACKGROUND. Absolute neutrophil counts (ANCs) and absolute phagocyte counts 2681 Utility of Automated Counting to Determine Absolute Neutrophil Counts and Absolute Phagocyte Counts for Pediatric Cancer Treatment Protocols Nobuko Hijiya, M.D. 1,2 Mihaela Onciu, M.D. 3 Scott C.

More information

Detection of important abnormalities of the differential

Detection of important abnormalities of the differential J Clin Pathol 1989;42:772-776 Detection of important abnormalities of the differential count using the Coulter STKR blood counter D F BARNARD, S A BARNARD, A B CARTER, S J MACHN, K G PATTERSON, A YARDUMAN

More information

Concentration Estimation from Flow Cytometry Exosome Data Protocol

Concentration Estimation from Flow Cytometry Exosome Data Protocol Concentration Estimation from Flow Cytometry Exosome Data Protocol 1. STANDARD CURVE Create a standard curve for the target exosome by plotting the mean fluorescence (y axis) against the protein concentration

More information

Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How?

Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How? Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How? Dennis B. DeNicola, DVM, PhD, DACVP Chief Veterinary Educator IDEXX Laboratories, Inc. Westbrook, ME USA Adjunct Professor of Veterinary

More information

Blood Cell Identification Graded

Blood Cell Identification Graded Blood Cell Identification Graded Case History The patient was a five-day-old girl with an elevated unconjugated bilirubin and a weakly positive direct antiglobulin test (DAT). Her CBC showed: WBC = 11.0

More information

Incorporating Differentials Into Every Complete Blood Count. Paige Flowers, LVT Dogwood Veterinary Internal Medicine

Incorporating Differentials Into Every Complete Blood Count. Paige Flowers, LVT Dogwood Veterinary Internal Medicine Incorporating Differentials Into Every Complete Blood Count Paige Flowers, LVT Dogwood Veterinary Internal Medicine Complete Blood Count Diagnostic performed to evaluate the quantity and morphology of

More information

Evaluation of the QBC Star centrifugal three-part differential haematology system

Evaluation of the QBC Star centrifugal three-part differential haematology system ORIGINAL ARTICLE 67 Evaluation of the QBC Star centrifugal three-part differential haematology system O. ERHABOR, G. RICHARDSON, I. MOHAMMED, C. THORNTON, J. BARK, M. HURST, D. HAMER and P. KINSELLA Blood

More information

Identification and Enumeration of Nucleated Red Blood Cells in Peripheral Blood

Identification and Enumeration of Nucleated Red Blood Cells in Peripheral Blood Identification and Enumeration of Nucleated Red Blood Cells in Peripheral Blood Bernard J FERNANDES Department of Laboratory Medicine, Mount Sinai Hospital and University of Toronto, Toronto, 6 University

More information

by Automated Cell Counters

by Automated Cell Counters Cytometry (Communications in Clinical Cytometry) 22:26-34 (1995) Variability in Absolute Lymphocyte Counts Obtained by Automated Cell Counters E. Simson and W. Groner Department of Pathology, Long Island

More information

PointCare NOW TM Technical Note

PointCare NOW TM Technical Note PointCare NOW TM Technical Note Introduction to our Technology And Summary of Evaluation and Performance data 1 of 19 Introduction The PointCare NOW is a portable, and very fast, diagnostic instrument

More information

Changes of hematological references depends on storage period and temperature conditions in rats and dogs

Changes of hematological references depends on storage period and temperature conditions in rats and dogs ISSN 1738-6055 (Print) ISSN 2233-7660 (Online) Lab Anim Res 2016: 32(4), 241-248 https://doi.org/10.5625/lar.2016.32.4.241 Changes of hematological references depends on storage period and temperature

More information

10/30/2015. XN Case Studies: Every Picture Tells a Story

10/30/2015. XN Case Studies: Every Picture Tells a Story XN Case Studies: Every Picture Tells a Story Jill Crist MT(ASCP)Field Product Specialist OBJECTIVES Explain how scattergrams and histogram pictures can provide great insight into abnormal hematology samples

More information

Literature List Body Fluids

Literature List Body Fluids Literature List Body Fluids Customer Information March 2018 Date: March 2018 Subject: Literature List body fluids Issued by: Scientific Customer Services Number: 180312 Note: Whether references are given

More information

Hematology & Coagulation Practicum Objectives CLS - 647

Hematology & Coagulation Practicum Objectives CLS - 647 Hematology & Coagulation Practicum Objectives CLS - 647 The following objectives are to be completed by the student for successful completion of this clinical rotation. The objectives within the psychomotor

More information

Use of the XE-2100 in a Patient with Cold Auto-immune Hemolytic Anemia

Use of the XE-2100 in a Patient with Cold Auto-immune Hemolytic Anemia Use of the XE-21 in a Patient with Cold Auto-immune Hemolytic Anemia Gudrun STAMMINGER and Lothar BEIER Klinikum Chemnitz ggmbh, Institut für Laboratoriumsmedizin, Flemmingstraße 2, Chemnitz, Germany.

More information

University of Miami Miller School of Medicine, Miami, FL 33136, USA, 3 State Key Laboratory

University of Miami Miller School of Medicine, Miami, FL 33136, USA, 3 State Key Laboratory Supplementary File ASXL1 plays an important role in erythropoiesis Hui Shi 1,2,3, Shohei Yamamoto 1,2,4, Mengyao Sheng 3, Jie Bai 3, Peng Zhang 1,2, Runze Chen 1,2, Shi Chen 1,2, Lihong Shi 3, Omar Abdel-Wahab

More information

Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential

Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential H e m a t o p a t h o l o g y / REFRIGERATION IMPROVES CBC COUNT STABILITY Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential Brent L. Wood, MD, PhD,

More information

Interpreting the CBC. Robert Miller PA Assistant Professor of Clinical Pediatrics and Family Medicine USC Keck School of Medicine Retired

Interpreting the CBC. Robert Miller PA Assistant Professor of Clinical Pediatrics and Family Medicine USC Keck School of Medicine Retired Interpreting the CBC Robert Miller PA Assistant Professor of Clinical Pediatrics and Family Medicine USC Keck School of Medicine Retired The CBC 3 Cell Lines RBCs WBCs Platelets Assess general health Make

More information

Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples

Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples Hematopathology / HEMOGLOBIN-BASED BLOOD SUBSTITUTES Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples Jolanta Kunicka, PhD, 1 Michael Malin, PhD, 1 David Zelmanovic,

More information

Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities

Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities JOANNE CORNBLEET, M.D. AND SALLY KESSINGER, MT(ASCP)SH The authors evaluated the Coulter 15 S-Plus

More information

DR SUDHIR MEHTA MD,MNAMS,FICP. Senior Professor & Head Medical Unit SMS Medical College & Hospital Jaipur

DR SUDHIR MEHTA MD,MNAMS,FICP. Senior Professor & Head Medical Unit SMS Medical College & Hospital Jaipur DR SUDHIR MEHTA MD,MNAMS,FICP Senior Professor & Head Medical Unit SMS Medical College & Hospital Jaipur s.smehta@hotmail.com CBC..What is the Utility of performing this basic Hematology Test? 10/31/2010

More information

Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital

Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital HEMATOPATHOLOGY Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital JOAN-LLUIS VIVES-CORRONS, MD, ISABEL BESSON, PHD, JOSEP MARIA JOU, MD, AND GABRIELA GUTIERREZ, MB The

More information

BD Leucocount Kit. Rx Only IVD

BD Leucocount Kit. Rx Only IVD 2/217 23-19381- IVD BD Leucocount Kit For enumeration of residual leucocytes in leucoreduced blood products 5 Tests Catalog No. 662415 217 BD. BD, the BD logo and all other trademarks are property of Becton,

More information

Do Your Flow Cytometric LDTs. Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine

Do Your Flow Cytometric LDTs. Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine Do Your Flow Cytometric LDTs Conform to the ICSH ICCS Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine How should LDTs be validated? Accuracy Specificity Sensitivity

More information

Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. ABSTRACT

Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. ABSTRACT OriginalArticle Comparison of Leukocyte Differential Count by Beckman Coulter Unicel DxH800, Beckman Coulter LH780 and Sysmex XE-5000 Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. Department

More information

Hematology 101. Cindy Rogers, MT(ASCP) Diagnostics System Specialist

Hematology 101. Cindy Rogers, MT(ASCP) Diagnostics System Specialist Hematology 101 Cindy Rogers, MT(ASCP) Diagnostics System Specialist More Acronyms...» CBC» RBC» HGB» HCT» WBC» MPV» PLT» RDW» DIFF» H&H» Complete Blood Count» Red Blood Cell» Hemoglobin» Hematocrit» White

More information

Automated Urinalysis Evaluation of the Sysmex UF-50

Automated Urinalysis Evaluation of the Sysmex UF-50 Clinical Chemistry / FLOW CYTOMETRY BASED URINALYSIS Automated Urinalysis Evaluation of the Sysmex UF-50 Hiroshi Okada, MD, 1 Yutaka Sakai, MD, 1 Gaku Kawabata, MD, 1 Masato Fujisawa, MD, 1 Soichi Arakawa,

More information

We noted a problem with the use of the hematology

We noted a problem with the use of the hematology Standard Article J Vet Intern Med 2017;31:1708 1716 Feline Differential Leukocyte Count with ProCyte Dx: Frequency and Severity of a Neutrophil-Lymphocyte Error and How to Avoid It H.W. Tvedten, V. Andersson,

More information

Human Progenitor Cell Enumeration by Flow Cytometry Practical Considerations

Human Progenitor Cell Enumeration by Flow Cytometry Practical Considerations Human Progenitor Cell Enumeration by Flow Cytometry Practical Considerations Bruce Briggs, M.S., S.I. (A.S.C.P.) OSU Medical Center Clinical Flow Cytometry Objectives Conceptual overview: How the Flow

More information

White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation

White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation ISLH Laboratory Hematology 7:255 266 2001 Carden Jennings Publishing Co., Ltd. Official Publication White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation BERNARD

More information

EDUCATIONAL COMMENTARY DISTINGUISHING MORPHOLOGIC LOOK-ALIKES

EDUCATIONAL COMMENTARY DISTINGUISHING MORPHOLOGIC LOOK-ALIKES EDUCATIONAL COMMENTARY DISTINGUISHING MORPHOLOGIC LOOK-ALIKES Educational commentary is provided through our affiliation with the American Society for Clinical Pathology (ASCP). To obtain FREE CME/CMLE

More information

Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer

Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer Science Gene Gulati, PhD, Sean Cote, DO, Eric Behling, MD, William Kocher, MD

More information

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Alert or critical values represent those assay results that require prompt, rapid clinical attention to avert significant study-participant

More information

USING RETICULOCYTE INDICES TO IDENTIFY α-thalassemia A PRELIMINARY REPORT

USING RETICULOCYTE INDICES TO IDENTIFY α-thalassemia A PRELIMINARY REPORT USING RETICULOCYTE INDICES TO IDENTIFY α-thalassemia A PRELIMINARY REPORT Suphan Soogarun 1, Jamsai Suwansaksri 2 and Viroj Wiwanitkit 3 1 Department of Clinical Microscopy, Faculty of Allied Health Sciences,Chulalongkorn

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

7-AAD/CFSE Cell-Mediated Cytotoxicity Assay Kit

7-AAD/CFSE Cell-Mediated Cytotoxicity Assay Kit 7-AAD/CFSE Cell-Mediated Cytotoxicity Assay Kit Catalog Number KA1293 96 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of

More information

Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population

Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population Clinical immunology Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population EL BIETA GÓRSKA, URSZULA DEMKOW, ROMAN PIÑKOWSKI, BARBARA JAKUBCZAK,

More information

Customer Information Literature List White Blood Cells

Customer Information Literature List White Blood Cells Customer Information Literature List White Blood Cells Date: September 2015 Subject: Literature List White Blood Cells Issued by: Scientific Customer Services Number: 150901 Note: Whether references are

More information

Sue Jung Kim, MD, MS; Yoonjung Kim, MD; Saeam Shin, MD; Jaewoo Song, MD, MS; Jong Rak Choi, MD, PhD

Sue Jung Kim, MD, MS; Yoonjung Kim, MD; Saeam Shin, MD; Jaewoo Song, MD, MS; Jong Rak Choi, MD, PhD Comparison Study of the Rates of Manual Peripheral Blood Smear Review From 3 Automated Hematology Analyzers, Unicel DxH 800, ADVIA 2120i, and XE 2100, Using International Consensus Group Guidelines Sue

More information

Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis

Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis Hematopathology / LEFT SHIFT MODEL PREDICTS INFECTION Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis Michael J. Wile, MD, 1 Louis D. Homer, MD, PhD, 2 Stede Gaehler,

More information

Insulin (Porcine/Canine) ELISA

Insulin (Porcine/Canine) ELISA Insulin (Porcine/Canine) ELISA For the quantitative measurement of insulin in Porcine/Canine serum and plasma (EDTA) For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSPO-E01

More information

Supplemental Figure 1. Activated splenocytes upregulate Serpina3g and Serpina3f expression.

Supplemental Figure 1. Activated splenocytes upregulate Serpina3g and Serpina3f expression. Relative Serpin expression 25 2 15 1 5 Serpina3f 1 2 3 4 5 6 8 6 4 2 Serpina3g 1 2 3 4 5 6 C57BL/6 DBA/2 Supplemental Figure 1. Activated splenocytes upregulate Serpina3g and Serpina3f expression. Splenocytes

More information

Scoring System for Detecting Spurious Hemolysis in Anticoagulated Blood Specimens

Scoring System for Detecting Spurious Hemolysis in Anticoagulated Blood Specimens Original Article Laboratory Informatics Ann Lab Med 2015;35:341-347 http://dx.doi.org/10.3343/alm.2015.35.3.341 ISSN 2234-3806 eissn 2234-3814 Scoring System for Detecting Spurious Hemolysis in Anticoagulated

More information

Hematology Revision. By Dr.AboRashad . Mob

Hematology Revision. By Dr.AboRashad  . Mob 1 1- Hb A2 is consisting of: a) 3 ά chains and 2 γ chains b) 2 ά chains and 2 β chains c) 2 ά chains and 2 δ chains** d) 2 ά chains and 3 δ chains e) 3 ά chains and 2 δ chains 2- The main (most) Hb found

More information

JMSCR Volume 03 Issue 05 Page May 2015

JMSCR Volume 03 Issue 05 Page May 2015 www.jmscr.igmpublication.org Impact Factor 3.79 ISSN (e)-2347-176x Time Effect Storage Artifacts of Anticoagulant EDTA on Peripheral Blood Cells Authors Dr. Ramu Thakur 1, Dr.O.P.Moorjani 2 1 Assitant

More information

EXOTESTTM. ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids

EXOTESTTM. ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids DATA SHEET EXOTESTTM ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids INTRODUCTION Exosomes are small endosome-derived lipid nanoparticles

More information

5/1/2017 DISCUSSION POINTS. Clinical Utility of Immature Cell Indices Beyond the Routine CBC John E. Donnelly BSN, RN

5/1/2017 DISCUSSION POINTS. Clinical Utility of Immature Cell Indices Beyond the Routine CBC John E. Donnelly BSN, RN DISCUSSION POINTS Importance of hematological immature cell indices Clinical Utility of Immature Cell Indices Beyond the Routine CBC John E. Donnelly BSN, RN Investigate the evidence for clinical utility:

More information

How can Blood Stabilizers be used for Validation of New Instruments? UK NEQAS for Leucocyte Immunophenotyping

How can Blood Stabilizers be used for Validation of New Instruments? UK NEQAS for Leucocyte Immunophenotyping How can Blood Stabilizers be used for Validation of New Instruments? Quality Pathology is essential for about 60% of all diagnoses. Thus, quality of pathology services is critical for good care and management.

More information

Blood Cell Identification Graded

Blood Cell Identification Graded BCP-21 Blood Cell Identification Graded Case History The patient is a 37-year-old female with a history of multiple sickle cell crises. She now presents with avascular necrosis of the left hip. Laboratory

More information

Comparative Evaluation of Two Flowcytometric Analysers as Diagnostic Tools for the Automated Detection of Malaria

Comparative Evaluation of Two Flowcytometric Analysers as Diagnostic Tools for the Automated Detection of Malaria 82 Available online at www.annclinlabsci.org Comparative Evaluation of Two Flowcytometric Analysers as Diagnostic Tools for the Automated Detection of Malaria S. Mohapatra 1, J. C. Samantaray 1, S. Arulselvi

More information

Complete Blood Count (CBC)

Complete Blood Count (CBC) Complete Blood Count (CBC) Objectives 1. To estimate the number of RBC in blood sample 2. To estimate the number of total WBC in blood sample 3. To perform a differential count for a blood sample Introduction

More information

Pseudobasophilia in Pediatric Age Group

Pseudobasophilia in Pediatric Age Group Original Article DOI: 10.17354/ijss/2015/259 Pseudobasophilia in Pediatric Age Group Femela Muniraj 1, Vijay Amritraj 2 1 Assistant Professor, Department of Pathology, Chettinad Hospital & Research Institute,

More information

METHOD VALIDATION: WHY, HOW AND WHEN?

METHOD VALIDATION: WHY, HOW AND WHEN? METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test

More information

Flow cytometry leukocyte differential : a critical appraisal

Flow cytometry leukocyte differential : a critical appraisal Flow cytometry leukocyte differential : a critical appraisal Francis Lacombe Flow cytometry department University Hospital of Bordeaux, Pessac, France francis.lacombe@chu-bordeaux.fr 2008 HORIBA ABX, All

More information

Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana

Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana IS THERE STILL A ROLE FOR THE MANUAL BLOOD COUNT? Indication for examination

More information

Blood smear analysis in the emergency veterinary patient

Blood smear analysis in the emergency veterinary patient Vet Times The website for the veterinary profession https://www.vettimes.co.uk Blood smear analysis in the emergency veterinary patient Author : Ashley Wemple Categories : RVNs Date : April 1, 2010 Ashley

More information

Evaluation of Early Post-transplant Leukocyte Recovery Using the Undiluted Erythrocyte Lysing Technique

Evaluation of Early Post-transplant Leukocyte Recovery Using the Undiluted Erythrocyte Lysing Technique Annals of Clinical & Laboratory Science, vol. 32, no. 2, 2002 159 Evaluation of Early Post-transplant Leukocyte Recovery Using the Undiluted Erythrocyte Lysing Technique Myungshin Kim, 1 Ja Young Kim,

More information

The Egyptian Journal of Hospital Medicine (July 2017) Vol.68 (3), Page

The Egyptian Journal of Hospital Medicine (July 2017) Vol.68 (3), Page The Egyptian Journal of Hospital Medicine (July 2017) Vol.68 (3), Page 1358-1363 Study of Automated Red Blood Cells Parameters in Correlation with Routine RBCs Morphology by Smear Review Tahany Ali El

More information

Evidence-Based Hematological Solutions

Evidence-Based Hematological Solutions Evidence-Based Hematological Solutions Beyond the Routine CBC Objectives Describe novel hematology parameters and their derivation. Investigate the evidence for their clinical utility. Discuss how new

More information

Interpreting Hematology Scatter-Plots; One Cancer Center s Keys to Seeing the BIG Picture

Interpreting Hematology Scatter-Plots; One Cancer Center s Keys to Seeing the BIG Picture Interpreting Hematology Scatter-Plots; One Cancer Center s Keys to Seeing the BIG Picture Barbara L. Burch, MHA MT (ASCP) Laboratory Manager New York University Clinical Cancer Center Disclosure Ms Burch

More information

PARASITOLOGY CASE HISTORY #14 (BLOOD PARASITES) (Lynne S. Garcia)

PARASITOLOGY CASE HISTORY #14 (BLOOD PARASITES) (Lynne S. Garcia) PARASITOLOGY CASE HISTORY #14 (BLOOD PARASITES) (Lynne S. Garcia) A 37-year-old woman, who had traveled to New Guinea for several weeks, presented to the medical clinic with fever, chills, and rigors within

More information

Requirements: Glass slides Leishman stain Microscopes Disposable needles Vials containing anticoagulants Methylated-spirit Staining rack

Requirements: Glass slides Leishman stain Microscopes Disposable needles Vials containing anticoagulants Methylated-spirit Staining rack Aim: To make a blood smear and to count the different types of leucocytes present in a stained blood smear and express their relative counts in percentage Principle - The blood contains various types of

More information