George X. Ding a) Department of Radiation Oncology, Vanderbilt University School of Medicine, Nashville, TN 37232, USA

Size: px
Start display at page:

Download "George X. Ding a) Department of Radiation Oncology, Vanderbilt University School of Medicine, Nashville, TN 37232, USA"

Transcription

1 Image guidance doses delivered during radiotherapy: Quantification, management, and reduction: Report of the AAPM Therapy Physics Committee Task Group 180 George X. Ding a) Department of Radiation Oncology, Vanderbilt University School of Medicine, Nashville, TN 37232, USA Parham Alaei University of Minnesota, Minneapolis, MN 55455, USA Bruce Curran Virginia Commonwealth University, Richmond, VA 23284, USA Ryan Flynn University of Iowa, Iowa City, IA 52242, USA Michael Gossman Tri-State Regional Cancer Center, Ashland, KY 41101, USA T. Rock Mackie University of Wisconsin, Madison, WI 53715, USA Moyed Miften University of Colorado, Aurora, CO 80045, USA Richard Morin Mayo Clinic, Jacksonville, FL 32224, USA X. George Xu Rensselaer Polytechnic Institute, Troy, NY 12180, USA Timothy C. Zhu University of Pennsylvania, Philadelphia, PA 19104, USA (Received 18 July 2017; revised 10 January 2018; accepted for publication 10 January 2018; published 24 March 2018) Background: With radiotherapy having entered the era of image guidance, or image-guided radiation therapy (IGRT), imaging procedures are routinely performed for patient positioning and target localization. The imaging dose delivered may result in excessive dose to sensitive organs and potentially increase the chance of secondary cancers and, therefore, needs to be managed. Aims: This task group was charged with: a) providing an overview on imaging dose, including megavoltage electronic portal imaging (MV EPI), kilovoltage digital radiography (kv DR), Tomotherapy MV-CT, megavoltage cone-beam CT (MV-CBCT) and kilovoltage cone-beam CT (kv- CBCT), and b) providing general guidelines for commissioning dose calculation methods and managing imaging dose to patients. Materials & Methods: We briefly review the dose to radiotherapy (RT) patients resulting from different image guidance procedures and list typical organ doses resulting from MV and kv image acquisition procedures. Results: We provide recommendations for managing the imaging dose, including different methods for its calculation, and techniques for reducing it. The recommended threshold beyond which imaging dose should be considered in the treatment planning process is 5% of the therapeutic target dose. Discussion: Although the imaging dose resulting from current kv acquisition procedures is generally below this threshold, the ALARA principle should always be applied in practice. Medical physicists should make radiation oncologists aware of the imaging doses delivered to patients under their care. Conclusion: Balancing ALARA with the requirement for effective target localization requires that imaging dose be managed based on the consideration of weighing risks and benefits to the patient American Association of Physicists in Medicine [ Key words: IGRT, image dose management, image dose reduction techniques, image guidance doses, recommended dose threshold e84 Med. Phys. 45 (5), May /2018/45(5)/e84/ American Association of Physicists in Medicine e84

2 e85 Ding et al.: Image Guidance Doses to Radiotherapy Patients e85 TABLE OF CONTENTS 1. INTRODUCTION 2. OVERVIEW OF DOSE RESULTING FROM IMAGE GUIDANCE PROCEDURES 2.A. Megavoltage beam imaging 2.B. Kilovoltage beam imaging 3. DOSE CALCULATION ALGORITHMS FOR KVAND MV IMAGING DOSE 3.A. Monte Carlo-based methods 3.B. Model-based methods 4. KILOVOLTAGE IMAGING BEAM DOSIMETRY 4.A. Input data for kv x-ray beam characterization 4.B. Output of a kv imager 4.C. In vivo dosimetry 5. METHODS OF ACCOUNTING FOR IMAGING DOSE 5.A. Patient-specific imaging dose calculations 5.B. Nonpatient-specific imaging dose estimations 6. RECOMMENDATIONS 6.A. General recommendations 6.B. Imaging dose output and consistency checks 6.C. Accounting for imaging dose to RT patients 6.D. Available techniques to reduce imaging dose to patients 7. ACKNOWLEDGMENTS 1. INTRODUCTION Image-guided radiation therapy (IGRT) has rapidly been adopted as the standard of care to improve the geometric accuracy of patient positioning during radiotherapy. 1 6 IGRT can significantly reduce target positioning errors, therefore, enabling highly conformal treatments. The acquired images during treatment delivery can be used for monitoring patient and target geometry changes, potential adaptive planning, 7 9 or margin reduction. 10 During the course of IGRT, the image guidance procedure is typically performed for each treatment fraction. The patient may occasionally be imaged more than once during any fraction in order to ensure that the patient s position is correct, and to reposition if necessary. Since these imaging procedures deliver additional radiation dose to patients, understanding the magnitude of this dose has become increasingly important in order to minimize its risk. While the commonly adopted radiation protection safety philosophy of As Low As Reasonably Achievable (ALARA) is also applicable for imaging dose, minimizing imaging doses by employing reasonable methods should not compromise target localization. Currently, imaging dose is not accounted for in RT treatment planning, 11 and the purpose of this report is to provide imaging dose data and develop guidelines for clinicians to make informed decisions regarding the risk and benefits of x-ray image guidance. The report of AAPM TG-75 1 provided dose estimates for a variety of image guidance techniques and recommended strategies for minimizing imaging dose while improving treatment delivery. TG-75 also identified the need to manage imaging dose for the large number of current and emerging imaging techniques in RT. These techniques include CT; 4D- CT; diagnostic x-ray imaging; in-room CT; dual radiographic x-ray imaging; fluoroscopy; and portal imaging, either using film or an electronic portal imaging device (EPID), in various modes. This report is intended to complement the AAPM TG-75 report. 1 It contains updated dose data resulting from current image acquisition procedures and modalities and addresses current issues in estimating and accounting for imaging dose during treatment planning when required. It also addresses how to minimize dose due to imaging procedures, and provides recommendations for incorporating suggestions made in AAPM TG-75 1 and ICRP This report also offers guidelines for commissioning imaging beams in order to provide patient-specific imaging dose calculations when needed. The imaging dose referred to in this report is absorbed dose to medium (e.g., dose to bone or dose to soft tissue), which differs from the effective dose metric used in TG-75. Effective dose, as defined by the ICRP, 13 is based on an estimate of biological effect integrated over the entire patient body, requiring a detailed knowledge of the radiation energy spectrum and irradiation geometry. Most treatment planning systems (TPSs) are not capable of calculating and displaying kv energy range dose distributions without special research tools, nor are they capable of converting calculated absorbed dose to effective dose and displaying the results. Thus, in order to avoid this additional level of complexity, absorbed dose is used in this report rather than effective dose. The magnitude of imaging dose is dependent on many factors, including the frequency of imaging and the technique used. For a single treatment fraction, two or more planar images or one or more volumetric acquisitions may be acquired. In the case of the Brainlab AG (Feldkirchen, GER- MANY) ExacTrac 14,15 or Accuray, Inc. (Sunnyvale, CA) CyberKnife systems, 16,17 the number of planar image acquisitions per session can potentially be well over 80, even for none of the SRS/SBRT treatments, as frequent imaging is typical for monitoring patient position. Each of these imaging procedures deliver additional radiation dose to normal tissue. 1,18 25 Depending on the imaging protocols and techniques used there are large variations in the imaging dose delivered to the patient. In general, MV imaging delivers higher doses than kv imaging. 26 With the exception of MV volumetric imaging, a single image acquisition can deliver a dose of cgy to the patient depending on the imaging modality. However, even with demonstrated progress in dose reduction, 21,25,27 41 the kv-cbct procedure employed for pelvic imaging can add a cumulative dose of 1 3% of the prescription dose during the course of treatment. Since the photoelectric effect is the dominant photon interaction process for kv imaging, the imaging dose to bony structures is a factor of 2 4 greater than that to soft tissue. 20,21,29 For an MV-CBCT image acquisition, the imaging dose can be greater than 10 cgy, depending on the imaging site and clinical protocol. 23,42 45 With all imaging procedures, the imaged volume is generally larger than the treatment volume, and tissues and organs outside the

3 e86 Ding et al.: Image Guidance Doses to Radiotherapy Patients e86 therapeutic beams are exposed to imaging radiation. These imaging doses to organs outside the treatment volume also need to be managed, as they may present an increased risk, especially in the case of pediatric patients. In this report, we review a variety of methods for determining x-ray imaging dose. For kv-cbct imaging dose, these include experimental phantom measurements, 19,38,46,47 in vivo measurements on patients, 48 and Monte Carlo (MC) calculations. 21,22,25 27,29,30,40,49 52 Commercially available treatment planning systems with user modifications, as well as mathematical models, 39 have been used to calculate the MV-CBCT 42,45 and kv-cbct 53,54 dose. Measurements have been used to estimate doses from 2D kv radiographs, 32,38 kv-cbct, 21,27 41 MV portal images, 26,32,38 MV- CBCT, 23,42,44 and MVCT. 55,56 In radiation therapy the prescribed therapeutic dose represents the minimum dose to part or all of the planning target volume. In developing strategies for managing the imaging dose, this task group considers 5% of the therapeutic target dose to be the threshold beyond which imaging dose should be accounted for in the treatment planning process. Dische et al. 57 stated that there was evidence from published clinical data and a suggestion from an analysis of the Continuous Hyperfractionated, Accelerated Radiotherapy (CHART) pilot study data 58 that dose variations as small as 5% may lead to real variations in both tumor response and the risk of morbidity. 59 Many studies on accuracy requirements in radiotherapy have recommended an accuracy level of 5% in the delivery and determination of dose to tumors and normal tissue Hence, the choice of a 5% threshold is based on considerations of clinical relevance, accuracy of dose calculation and delivery, dose tolerances for critical organs, and feasibility in clinical practice. Data currently available in the literature, and quoted throughout this report, for patient populations undergoing IGRT indicate that imaging dose is generally less than 5% of the therapeutic target dose, 26,32,66,67 except for some imaging procedures that use MV beams, particularly MV-CBCT. 42 Balancing ALARA principles with the requirement for effective target localization, however, requires the imaging dose to be managed on the consideration of weighing risks and benefits to the patient. 2. OVERVIEW OF DOSE RESULTING FROM IMAGE GUIDANCE PROCEDURES 2.A. Megavoltage beam imaging Megavoltage imaging modalities capture projection images using either electronic portal imaging devices (EPIDs) [Fig. 1(a)] or, in the case of MVCT in the Tomotherapy Hi- Art Radixact system (Accuray Inc., Sunnyvale, CA), 55,68,69 a single-row CT detector. A typical pair of orthogonal 6 MV portal images acquired using an EPID results in a dose distribution like the one shown in Fig. 1(b), and organ doses of 1 5 cgy (Tables Ia Ic) while the imaging dose from a 2.5-MV image beam 70 is about 50% of that from a 6-MV beam. Volumetric MV-CBCT images are reconstructed using projections acquired with the EPID and result in a greater dose than a pair of orthogonal MV portal images. Reported doses per monitor unit for MV-CBCT delivered to regions and organs typically considered are listed in Table Id. 45 Monitor unit values ranging from 2 to 15 have been reported in the literature ,71 Typically the head and neck region is imaged with lower monitor unit protocols (2 5 MU) than the thoracic or pelvic regions, which may be imaged at up to 15 MU. Figure 2 shows an MV-CBCT dose distribution for a pelvic patient imaged at 15 MU for reference. The Siemens kview system provides the option to improve MV-CBCT image quality per unit dose by generating the imaging beam with a low Z (carbon) electron target and a 4.2-MeV electron beam, increasing the percentage of kilovoltage photons in the imaging beam considerably relative to MV-CBCT acquired with a conventional 6-MV beam generated with a tungsten target. 72 With the kview system it is possible to obtain the same contrast-to-noise ratio as conventional MV-CBCT at about one-third the imaging dose. 73 Organ-specific kview CBCT doses reported by Beltran 74 and Dzierma et al. 75 ranged from 0.6 cgy to 1.2 cgy/ MU. Image doses from kview based MV-CBCT per monitor unit are generally less than those of conventional 6-MV beams. 44,73 80 Table Ie shows measured Tomo MVCT doses at the center of a 30-cm water phantom. Doses range from 0.8 to 2.5 cgy and depend on the acquisition mode. 2.B. Kilovoltage beam imaging Current kv imaging devices are generally integrated into linear accelerators and are capable of acquiring both 2D radiographs and 3D volumetric kv-cbct images. 8,81 83 Examples of kv-cbct scanners integrated into a Varian Medical Systems, Inc. (Palo Alto, CA) On Board Imaging (OBI) system and an Elekta (Stockholm, Sweden) X-Ray Volume Imaging (XVI) system are shown in Fig. 3. Using MC methods to simulate an earlier version of the kv-cbct scanner on the Varian OBI system (Varian OBI 1.3), Ding et al. reported 21 the imaging doses received by patients from kv-cbct scans of different treatment sites, including head and neck, chest, and abdomen. The imaging doses resulting from a single kv-cbct procedure were 1 9 cgy to soft tissues and 6 29 cgy to bones and depended on the patient size and the site of the scan. 21 Since the introduction of kv imaging systems, progress has been made by the vendors to reduce the imaging dose 29 while maintaining or improving image quality. Examples of these efforts include better reconstruction techniques, improved software implementation, and the use of lower beam energies (100 and 110 kvp, in addition to the standard 125 kvp x rays), where beam energy is optimized for the size of the patient and the atomic number of the structures being imaged. 85 Newer kv-cbct acquisition techniques, which utilize 200 degree scans (instead of 370 degree ones), were introduced to reduce the imaging dose to patients for head and pelvis scans. 29

4 e87 Ding et al.: Image Guidance Doses to Radiotherapy Patients e87 (a) (b) FIG. 1. (a) Conventional electronic portal imaging device (EPID) and (b) typical dose distributions and organ dose-volume histograms (DVHs) resulting from an orthogonal pair of 6 MV portal images (2 MU per image). Reproduced from Ding and Munro. 26 TABLE IA. Typical organ doses for the head and neck treatment site with MV EPID portal imaging. D50 is minimum dose delivered to 50% of the organ volume (from Reference [26] for 6 MV beam and Reference [70] for 2.5 MV beam). These are for a typical pair of orthogonal setup fields (2 MU/field for 6 MV and 1 MU for 2.5 MV). Brain TABLE IB. Typical organ doses for the chest treatment site with MV EPID portal imaging (from References [26] and [70]) for a typical pair of orthogonal setup fields (2 MU/field for 6 MV and 1 MU for 2.5 MV). Chest D50 range (cgy) D50 range (cgy) Organ 6 MV 2.5 MV Organ 6 MV 2.5 MV Brain Brainstem Chiasm Eyes Optic Nerves Pituitary Aorta Lungs Esophagus Kidney Heart Liver Spinal Cord This not only reduces CBCT acquisition times, but results in a nonuniform exposure of the patient with the potential to avoid irradiating superficial critical structures. 29 These improvements have reduced imaging dose by more than an order of magnitude in some cases. Figures 4 6 compare dose distributions between kv-cbct scanners on newer

5 e88 Ding et al.: Image Guidance Doses to Radiotherapy Patients e88 TABLE IC. Typical organ doses for the pelvis treatment site with MV EPID portal imaging (from Reference [26] and [70]) for a typical pair of orthogonal setup fields (2 MU/field for 6 MV and 1 MU for 2.5 MV). Pelvis Organ D50 range (cgy) 6 MV 2.5 MV Bladder Bowel Femoral heads Prostate Rectum TABLE ID. MV-CBCT doses per monitor unit using a 6 MV treatment beam with an acquisition arc of 200 degrees, starting at 270 degrees and stopping at 110 degrees (from Reference [45]). Location Isocenter dose (cgy/mu) Average organ dose (cgy/mu) Maximum organ dose (cgy/mu) Cranium Total-brain Left lens Right lens Left eye Right eye Thorax Left lung Right lung Total lung Spinal canal Heart Vertebral bodies Soft Tissue Pelvis Femoral heads TABLE IE. Tomo MVCT dose at the center of a 30-cm water phantom and its dependency on acquisition protocols. MVCT in Tomo Acquisition mode Fine pitch (4 mm couch travel/rotation) Normal pitch (8 mm couch travel/rotation) Coarse pitch (12 mm couch travel/rotation) Dose (cgy) 2.5 cgy 1.2 cgy 0.8 cgy From Edward Chao, Accuray Incorporated and T. Rock Mackie, UW, Madison, WI. Varian OBI systems (OBI 1.4/1.5) and on a previous version of the Varian OBI system (OBI 1.3). 26,29 When two orthogonal planar kv images are sufficient for an image guidance task, 2D radiographs are often used. The doses resulting from 2D kv imaging, investigated using multiple methods, 26,32,38 have been shown to be much lower compared to those from volumetric kv-cbct 26 procedures. Figure 7 shows typical dose distributions resulting from a pair of orthogonal planar images in head, thorax, and pelvis scans. 26 Similar lower doses, on the order of 0.1 cgy, were reported for fixed double x-ray tube systems that use projected images for patient localization, such as the Brainlab ExacTrac 14,15 and the Accuray CyberKnife. 16,17 3. DOSE CALCULATION ALGORITHMS FOR KV AND MV IMAGING DOSE 3.A. Monte Carlo-based methods Monte Carlo (MC) techniques for simulating dosimetry problems have evolved considerably over the last three decades 86 and are regarded as the gold standard in dose calculations. The development of a special purpose MC code, BEAM, 87,88 made it practical to simulate megavoltage and kilovoltage beams. With many improvements in both accuracy and computational efficiency, 87,89 MC techniques have been used to characterize therapeutic megavoltage beams from linear accelerators and kilovoltage photon beams from x-ray units. 21,25,51,84,86,88,90 92 Given these capabilities, MC simulations have been used to calculate realistic imaging dose distributions in patients resulting from different x-ray imaging systems. 21,22,25,27,29,30,40,49 51 These studies provide detailed information about patient organ dose resulting from different image guidance procedures. Although a number of commercial treatment planning systems incorporate MC calculations for MV beams, none currently allow MC calculations for kv beams. 3.B. Model-based methods FIG. 2. Dose distribution resulting from an MV-CBCT localization procedure of a prostate cancer patient using a 15 MU imaging protocol with a 6 MV beam. Reproduced from Miften et al. 42 It is worth noting that the asymmetric dose distribution is because the scan is acquired with a gantry rotation of 200 degrees. Model-based dose calculation algorithms were developed for accurate MV beam dose calculations and are commonly implemented in commercial treatment planning systems. 93 When an imaging procedure uses MV beams, these algorithms are capable of accurately calculating the imaging dose.

6 e89 Ding et al.: Image Guidance Doses to Radiotherapy Patients e89 (a) (b) wedge to mimic the isodose tilt of a half fan/half-bow tie imaging beam in the treatment planning system. Further work in this area has demonstrated that this could also be achieved by inserting a compensator in the beam. 99 The same commercial TPS has also been used to model kv beams from Elekta XVI and Siemens kvision imaging systems. 100,101 In general, use of these algorithms produces dose distributions of sufficient accuracy, except in and around bony structures. A proposed correction method, the Medium-Dependent Correction (MDC) approach, 102 accounts for atomic number dependency when computing kv dose distributions and can potentially increase the accuracy of kv imaging dose calculations to an acceptable 10 20%. 94,95 With additional improvements in commercial TPSs, therefore, it may become feasible to use the same model-based algorithms to calculate doses from both an MV therapeutic beam and a kv imaging beam. 4. KILOVOLTAGE IMAGING BEAM DOSIMETRY Unlike MV imaging beams that can be easily added to treatment plans for dose calculation purposes, kv imaging beams generally require commissioning in treatment planning systems. As indicated above, current commercial treatment planning systems do not accommodate the addition of kv imaging beams in routine clinical practice. Considering that kv imaging dose calculation in treatment planning systems may become possible in the future, however, guidance is provided in Sections 4.A and 4.B to address kv imaging beam data acquisition. FIG. 3. kv image devices integrated into linear accelerators: (a) Varian OBI system on a Varian Trilogy treatment unit, Reproduced from Ding at el. 84 ; (b) Elekta XVI on an Elekta Synergy treatment unit. Morin et al. 43 and Miften et al. 42 calculated the imaging dose from MV-CBCT using two different treatment planning systems and reported dose calculation accuracy of better than 3%. The dose from the Siemens kview system has been computed using a commercial treatment planning system 44,76 by implementing the kview beam spectrum given by Faddegon et al. 72 in the TPS. It is feasible to use model-based dose calculation algorithms to perform dose calculations for kv-range beams, but there are inherent inaccuracies in the approach. Commercially available model-based algorithms underestimate dose to bone by up to 300% 94,95 when used for kv beams due to the fact that they do not account for photoelectric effect. 21,93,96,97 Alaei et al. demonstrated the feasibility of modifying an existing MV model-based dose calculation algorithm in a commercial TPS to calculate kv-cbct dose. 53 This required the addition of low-energy deposition kernels 98 which accounted for the density effect of bone but not the photoelectric effect. In addition, unlike MV beams, modeling kv beams in a planning system requires beam data for the respective kv system, a topic which is further discussed in Section 5. In reference 51, Alaei et al. used a 4.A. Input data for kv x-ray beam characterization Characterizing kv imaging beams in a planning system requires collecting beam data, including depth-dose curves, cross-profiles, and the absolute dose resulting from an image acquisition procedure. There are inherent challenges in collecting data for kv imaging beams due to the low radiation dose rate and the strong dependence on the medium in which the measurements are made. To overcome these challenges, experimentally validated MC-simulated beam data can be used. 53,84,103 Caution should be taken when using the beam data generated from MC simulations, as they will depend on the simulation parameters. 4.B. Output of a kv imager The beam output should be collected by the clinical physicist using a proper detector calibration and data collection protocol. Although a method for determining kv absorbed dose with a calibrated ionization chamber based on x-ray beam specifiers, such as half-value layer (HVL) and kvp, is available from dosimetry protocols, the calibration conditions recommended in these dosimetry protocols are often not applicable to imaging acquisition procedures, especially when the x-ray source is moving during the scan. It is known that water is the most suitable medium for kv x-ray beam measurements. However, plastic phantom materials are more convenient. A method to determine the output

7 e90 Ding et al.: Image Guidance Doses to Radiotherapy Patients e90 (a) (b) (c) % volume (d) (e) (f) Head Head Scan: OBI 1.3 Standard Dose Head: OBI OBI Spinal cord 80 Spinal cord Left eye Bone Right eye Left eye Brain Brain Right eye Body 40 Right eye Bone 40 Spinal cord Body Left eye dose /cgy dose /cgy % volume % volume dose /cgy FIG. 4. Typical doses for a head scan shown in color wash along with dose-volume histograms (DVHs) for radiosensitive organs for the Varian OBI 1.3 (a,d); for the OBI 1.4 during a Standard-Dose Head scan (b,e), where the x-ray source is positioned below the patient; and for the OBI 1.4 during a head scan with the x- ray source positioned above the patient (c,f dashed lines). The solid lines in (f) are reproduced from (e) and show the quantitative effect of rotating the x-ray source from the back to the front of the patient. Note that the abscissa in (d) is 10 times larger than in (e). Reproduced from Ding et al. 29 (Scanning parameters are listed Table IId). resulting from a specific image acquisition procedure has been described 103 where the effect of using different plastic phantom materials in the kv energy range is investigated. This study includes two water equivalents, Plastic Water Low Energy Range (PW-LR CIRS, Inc., Norfolk, VA) and Solid Water (Gammex, Inc., Middleton, WI), along with the less water-equivalent polymethyl methacrylate (PMMA). Caution should be used when interpreting the measured data from solid water-equivalent and PMMA phantoms to determine the output of a kv beam, as the uncertainties can be significant (8 20%). 103 The considerations for a suitable phantom include phantom size relative to x-ray field size, beam attenuation by the phantom, and availability. For cases where fluoroscopy mode is used instead of radiography mode for acquiring 2D kv planar images, the imaging dose is proportional to the x-ray exposure times for the selected imaging protocol parameters (kvp and ma), and doses can be estimated based on the product of dose rate and scan time for the kv beam. The output of a kv beam can be obtained in air or in a phantom as recommended in the AAPM dosimetry protocol for kv beams. 106 When measured in air, the beam output is insensitive to the field size used. When measured in a phantom, the beam output is affected by the phantom scatter, which depends on the field size. If the beam cannot be delivered in static mode, rotational absorbed dose can be measured utilizing a cylindrical chamber placed at the isocenter. 4.C. In vivo dosimetry If patient-specific imaging dose verification is desired, in vivo patient dose measurements can be performed using available detectors such as diodes, thermoluminescent dosimeters (TLDs), or optically stimulated luminescent dosimeters (OSLDs). 48,107,108 Dosimeters, such as diodes, that are intended for use in MV beam in vivo measurements are not suitable for kv beams as they include inherent buildup. In addition, their response to kv radiation is significantly different from MV beams. Dosimeters that are used for kv beams should be calibrated for the beam energy used. 48 In vivo measurements using detectors placed on the patient s skin may be used to estimate organ dose once detector response has been scaled by the known dose distribution inside the patient. 48 There have been many publications discussing the use of TLDs, OSLDs, MOSFETs, and other detectors in kv beams. 48, METHODS OF ACCOUNTING FOR IMAGING DOSE Many different methods have been used to calculate the imaging dose delivered to RT patients. 1,11,18,19,21 23,25,51 When there is reasonable expectation that the imaging dose will exceed 5% of the total prescribed dose, two methods can

8 e91 Ding et al.: Image Guidance Doses to Radiotherapy Patients e91 (a) (b) % volume (c) (d) (e) 100 Pelvis: Full scan length Pelvis: 5 cm scan length Left femur head Righ femur head Body Left femur head Righ femur head Body Rectum Bladder 30 Rectum Bladder Prostate 10 Prostate dose /cgy % volume dose /cgy dose /cgy Dose profiles in axial direction Standard scan length (17 cm) Scan length set to 10 cm Scan length set to 5 cm z-axis /cm FIG. 5. The effect of reducing the scanned length: (a) Dose distributions shown in colorwash for default Pelvis scan length (16 cm); (b) reduced Pelvis scan length (5 cm); (c, d) Corresponding dose-volume histograms for the specific organs resulting from respective scans; (e) Dose profiles in the inferior superior direction along the line AB shown in Figure (a,b) across the irradiated volume for 16 cm (standard scan length), 10 cm, and 5 cm, respectively. The direction of z-axis is from inferior to superior in (e). The peak at the right in (e) represents the dose as the line AB crosses into the sacral vertebral body (bone). Note that reducing the scan length of the CBCT scan reduces both the maximum dose and the volume that is exposed to radiation. Reproduced from Ding et al. 29 (Scanning parameters are listed Table IId). be used to manage it: patient-specific dose calculations and nonpatient-specific dose estimations. 5.A. Patient-specific imaging dose calculations Patient-specific imaging dose calculations are based on patient CT images 67,113 and provide individualized organ doses, since the dose resulting from image guidance procedures will vary depending on the patient size as well as image location. 21,54,67 In order to perform patient-specific imaging dose calculations in a treatment planning system, the beams used for imaging procedures must be characterized as part of the commissioning process. The process of characterizing MV imaging beams in a treatment planning system is straightforward in cases where the imaging beam is the same as the therapeutic beam, allowing imaging dose to be calculated according to the monitor units and field sizes used in image acquisition. Imaging dose can then be added to therapeutic dose directly during treatment planning. Moreover, if the total number of imaging procedures is known at the planning stage, the imaging dose can be accounted for when optimizing patient treatment plans for total dose to the target and OARs. Miften et al. demonstrated this approach 42 by showing optimized IMRT plans with and without MV-CBCT included in the process. This, along with the lack of MC simulations or model-based algorithms capable of handling kv beams in commercial TPSs, 53 currently precludes patient-specific imaging dose calculations for any other than therapeutic MV beams.

9 e92 Ding et al.: Image Guidance Doses to Radiotherapy Patients e92 (a) (b) (c) Low Dose Thorax: OBI 1.4 (d) Bone Chest: OBI 1.3 % volume Body Bone Heart Left lung Right lung Spinal cord % volume Body Heart Left lung Right lung Spinal cord dose /cgy dose /cgy FIG. 6. Axial, frontal, and sagittal views showing dose distributions resulting from (a) OBI 1.4 low-dose thorax scan and (b) OBI 1.3 scan. The corresponding dose-volume histograms for different organs are shown for the low-dose scan (c) and the OBI 1.3 scan (d). Note that the horizontal scales in (c) and (d) differ by a factor of 5. Reproduced from Ding et al. 29 (Scanning parameters are listed Table IId). 5.B. Nonpatient-specific imaging dose estimations Given the small magnitude of imaging dose relative to therapeutic dose, 26,113 it may be adequate to use simpler approaches that provide reasonable estimates of the imaging dose. It has been shown Ref. [54,67] that interpatient variation and geometry dependence are small in most cases and that dose estimates could be provided in the form of simple look-up tables, which may be accurate enough to estimate the dose from repeat imaging procedures. Such tabulated values 67,113 can provide clinicians with adequate estimates of imaging dose to organs. Tables IIa, IIb, IIc, IId, IIe, and IIIa, IIIb, IIIc list representative organ doses resulting from kv- CBCT scans in the Varian OBI and Elekta XVI systems for different treatment sites and scan protocols. Note that bow tie filters used in the kv-cbct acquisition not only improve image quality but also significantly reduce imaging dose. 26 These tabulated values, when scaled by the mas used for image acquisition, are sufficient to estimate imaging dose to within 20% 67 and can assist the clinician in: (a) determining if the imaging doses are expected to be close to the 5% threshold, (b) choosing a suitable IGRT protocol, and (c)

10 e93 Ding et al.: Image Guidance Doses to Radiotherapy Patients e93 (a) (b) (c) FIG. 7. Imaging dose from a pair of orthogonal planar kv images for (a) head, (b) chest, and (c) pelvis images using the Varian OBI system. 26 (Parameters for kv radiographs for specified acquisition techniques are listed in Table IId).

11 e94 Ding et al.: Image Guidance Doses to Radiotherapy Patients e94 TABLE IIA. Organ doses for the head and neck and brain treatment sites from Varian OBI v1.4 using Standard Head kv-cbct scan. D50 and D10 are minimum dose delivered to 50% and 10% of the organ volume, respectively. (From Reference [67] and kv-cbct scan parameters for Varian OBI 1.4 shown in Table IId). Organ Standard head, brain D50 range (cgy) D10 range (cgy) Standard head, head and neck Organ D50 range (cgy) D10 range (cgy) Brain Brain Brainstem Larynx Chiasm Oral cavity Eyes Parotids Optic Spinal cord Nerves Pituitary Thyroid Spinal Esophagus Cord Skin Skin Bones Bones TABLE IID. Parameters for kv-cbct-specified acquisition techniques in Varian OBI 1.4 (from Reference [26]). kv-cbct Name Bow tie filter (kv) (mas) Gantry rotation (degrees) OBI Standard-dose Full fan head OBI Low-dose Full fan head OBI High-quality Full fan head OBI Pelvis Half fan OBI Pelvis spot Full fan light OBI Low-dose Half fan thorax TrueBeam Head Full fan TrueBeam Pelvis Half fan TrueBeam Spotlight Full fan TrueBeam Thorax Half fan TABLE IIB. Organ doses for the chest treatment site from Varian OBI v1.4 using the low-dose thorax kv-cbct scan (from Reference [114] and kv- CBCT scan parameters for Varian OBI 1.4 shown in Table IId). Low-dose thorax Organ D50 range (cgy) D10 range (cgy) Aorta Lungs Small bowel Esophagus Kidney Heart Liver Spinal cord Spleen Stomach Trachea Skin Bones TABLE IIC. Organ doses for the pelvis treatment site from Varian OBI v1.4 using Pelvis kv-cbct scan. (From Reference [114] and kv-cbct scan parameters for Varian OBI 1.4 shown in Table IId). Pelvis scan, prostate isocenter Organ D50 range (cgy) D10 range (cgy) Bladder Bowel Femoral heads Prostate Rectum Skin Bone TABLE IIE. Parameters for kv radiographs for specified acquisition techniques in Varian OBI 1.4. Name (kv) (mas) Head-AP Head-Lat 70 5 Thorax-AP 75 5 Thorax-Lat Pelvis-AP-Med Pelvis-Lat-Med The clinical default OBI blades are set to X1 = X2 = 13.3 cm and Y1 = Y2 = 10.3 cm in all acquisition techniques. All six techniques were modeled with and without a full fan bow tie filter. TABLE IIIA. Organ doses for the head and neck treatment site from Elekta XVI kv-cbct scan using S cassettes, 100 kvp, 0.1 mas/acquisition, 360 acquisitions, degree (IEC) rotation (from Reference [54]). Head and neck Organ D50 range (cgy) Brainstem Rt eye Lt eye Rt parotid Lt parotid Rt cochlea Lt cochlea Oral cavity accounting for the organ dose resulting from a specific image acquisition procedure over the course of treatment. Until kv dose calculations become available in commercial treatment planning systems, using estimated organ dose

12 e95 Ding et al.: Image Guidance Doses to Radiotherapy Patients e95 TABLE IIIB. Organ doses for the pelvis treatment site from Elekta XVI kv- CBCT scan using an M cassette without a bow tie filter, 120 kvp, 1.0 mas/ acquisition, 650 acquisitions, full 360 degree rotation (from Reference [54]). Pelvis Organ D50 range (cgy) Bladder Rectum Small bowel TABLE IIIC. Organ doses for the pelvis treatment site from Elekta XVI kv- CBCT scan using an M cassette with bow tie filter, 120 kvp, 1.6 mas/acquisition, 650 acquisitions, full 360 degree rotation (from Reference [54]). Pelvis Organ D50 range (cgy) Bladder Rectum Small Bowel that measured dose is within the manufacturer-stated specifications at the time of acceptance of the image device. Image acquisition procedures should include both those at MV and kv energies. The phantom and detectors used should be appropriate for the beam energy. Phantom sizes should be large enough to provide full x-ray scattering. (b) Consistency checks should be performed annually and after each system upgrade, and the recommendations from AAPM quality assurance reports, such as those from AAPM Task Group 142, 115 should be used. Checks for imaging dose consistency in air can also be performed using commercially available tools for measuring beam parameters (i.e., kvp, mas, etc.) that uniquely define the specific procedure. (c) If patient-specific imaging dose verification is desired for a particular patient, in vivo patient dose measurements should be performed with suitable detectors. The limited accuracy of patient-specific measurements should be taken into account in the review of the measured data. tables from various imaging procedures may be an acceptable option given the expected low-dose level. 6. RECOMMENDATIONS Unlike diagnostic imaging procedures, IGRT image acqusitions are more frequent, repeated on a daily basis, and include a volume that is larger than the treated one. Proper management of imaging dose in IGRT includes adherence to ALARA principles by minimizing the dose as much as possible and accounting for it when necessary. 6.A. General recommendations (a) Create local imaging protocols, including imaging modality, technique, and frequency, that are suitable for the imaging requirements of the clinic. 26 Consulting with a diagnostic imaging physicist may be helpful in this process. (b) Develop protocols that are specific for pediatric patients. 29,40 (c) Communicate the imaging dose associated with IGRT protocols by site (head, thorax, abdomen, pelvis) to radiation oncologists. This enables informed decision-making for selecting imaging protocols and ensures the clinicians are aware of the imaging doses being delivered to their patients. 6.B. Imaging dose output and consistency checks (a) The anticipated imaging dose for each image acquisition procedure, with specified protocol parameters, should be measured in air or in-phantom, according to the AAPM dosimetry protocols for kv and MV beams, to confirm 6.C. Accounting for imaging dose to RT patients It is recommended that imaging dose be considered part of the total dose at the treatment planning stage if the dose from repeated imaging procedures is expected to exceed 5% of the prescribed target dose. Patient organ doses can be calculated or estimated by using either patient-specific or nonpatient-specific methods. Facilitation of patient-specific imaging dose calculations may require implementation of new algorithms for MV and kv beams in commercial treatment planning systems. It is acceptable for the uncertainties of calculated imaging doses to reach 20%, because imaging dose is generally only a few percent of the prescribed target dose. As a result, the uncertainty of the summed dose (therapeutic + imaging) will still be at a level of 2 3%. 6.D. Available techniques to reduce imaging dose to patients Depending on the imaging modality, a variety of techniques are available to reduce the imaging dose to organs at risk, as recommended by AAPM TG-75 1 and AAPM TG ALARA should always be the guiding principle applied in practice. At the time of this report, the following techniques are recommended: (a) Reduce the imaging field size as much as possible. This will reduce the volume of irradiated tissue surrounding the target. Reducing the cranial-caudal extent of kv- CBCT scans can also significantly reduce the integral as well as scattered dose in the volume. 29 (b) During patient setup for MV portal images, minimize the imaging field size without removing reference

13 e96 Ding et al.: Image Guidance Doses to Radiotherapy Patients e96 structures needed for patient alignment. For images acquired for documenting delivery, select image during treatment to avoid adding additional dose to the patient. (c) For Tomotherapy units, select MVCT scan pitch parameters that balance imaging dose with clinical need (i.e. patient positioning or adaptive planning). The imaging dose differs significantly when different pitch parameters are selected (Table Ie). 56 (d) For MV-CBCT, select a patient-specific MV imaging protocol and restrict the imaging field of view (FOV). The imaging dose can be reduced if bony anatomy rather than soft tissue is used for treatment localization. 78 Note that the degree of dose reduction possible will depend upon the image quality requirements of the clinicians. (e) When deciding between 2D radiographs or 3D volumetric images, consider the image requirements. As ALARA is the guiding principle, consider 2D if two planar orthogonal kv images are sufficient for the task. The organ doses from image guidance can be reduced by a factor of 10 using 2D kv imaging compared to 3D kv-cbct. 26 (f) Optimize imaging parameters (e.g., kvp, mas) and select appropriate manufacturer-provided default clinical protocols (pelvis, abdomen, thorax, head and neck) for different normal adult body sites. In the case of pediatric patients with a small body size, default low-dose protocols for a head and neck kv-cbct scan can be used to image a pelvic site. This reduces imaging dose by a factor of 2 3 without compromising the image quality. 29 (g) The kv-cbct scan protocols that use partial rotation provide the opportunity to selectively avoid irradiating superficial organs. Partial rotation during a head scan can be used to dramatically reduce the dose to the eyes. 29 The technique can also be applied to reduce dose to the bladder or rectum for kv-cbct scans. (h) Since the beam exit dose is only a few percent of the entry dose for kv x rays, the beam directions used for orthogonal planar images can be selected to minimize dose to critical organs. 26 For a fixed orthogonal pair, consider not only 0 and 90 but also 180 and 90, 0 and 270, and 180 and 270 beam angles to minimize dose to organs at risk. (i) Consider the use of full bow tie filters when acquiring planar kv images. Bow tie filters can significantly reduce skin dose and dose to organs at risk. Always use the appropriate bow tie filter for kv-cbct acquisition when manual placement of the filter is required, since failure to employ the filter increases the imaging dose by factors of ACKNOWLEDGMENTS The members of this task group thank Greg Sharp (TISC Lead Reviewer), Ping Xia (TPC Lead Reviewer), Jeffrey Siebers (External Lead Reviewer), Jeff Colvin, and Sonja Dieterich for their very helpful comments and suggestions. We thank Gig Mageras and Debra Brinkmann for their helpful guidance over the years in which this report was developed. We also thank the referees for their very helpful and constructive comments. We wish to acknowledge the support from AAPM TPC leadership and extend our thanks to the science editor for the professional proof reading of this report. Finally, we want to thank our consultants: Walter Bosch, Jun Deng, Choonsik Lee, Peter Munro and Ying Xiao for their invaluable contributions to this report. a) Author to whom correspondence should be addressed. Electronic mail: george.ding@vanderbilt.edu. REFERENCES 1. Murphy M, Balter JM, BenComo J, et al. The management of imaging dose during image-guided radiotherapy: Report of the AAPM Task Group 75. Med Phys. 2007;34: Duggan DM, Ding GX, Coffey 2nd CW, et al. Deep-inspiration breath-hold kilovoltage cone-beam CT for setup of stereotactic body radiation therapy for lung tumors: initial experience. Lung Cancer. 2007;56: Jaffray DA, Siewerdsen JH. Cone-beam computed tomography with a flat-panel imager: initial performance characterization. Med Phys. 2000;27: Letourneau D, Wong JW, Oldham M, et al. Cone-beam-CT guided radiation therapy: technical implementation. Radiother Oncol. 2005;75: Oldham M, Letourneau D, Watt L, et al. Cone-beam-CT guided radiation therapy: a model for on-line application. Radiother Oncol. 2005;75:271 E1 271 E Thilmann C, Nill S, Tucking T, et al. Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences. Radiat Oncol. 2006;1: Ding GX, Duggan DM, Coffey CW, et al. A study on adaptive IMRT treatment planning using kv cone-beam CT. Radiother Oncol. 2007;85: Jaffray DA, Drake DG, Moreau M, Martinez AA, Wong JW. A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft-tissue targets. Int J Radiat Oncol Biol Phys. 1999;45: Nijkamp J, Pos FJ, Nuver TT, et al. Adaptive radiotherapy for prostate cancer using kilovoltage cone-beam computed tomography: first clinical results. Int J Radiat Oncol Biol Phys. 2008;70: Pawlowski JM, Yang ES, Malcolm AW, Coffey CW, Ding GX. Reduction of dose delivered to organs at risk in prostate cancer patients via image-guided radiation therapy. Int J Radiat Oncol Biol Phys. 2010;76: Ding GX, Coffey CW. Is it time to include imaging guidance doses in the reportable total radiation doses of radiotherapy patients? Int J Radiat Oncol Biol Phys. 2008;72:S145 S ICRP-102. Managing patient dose in multi-detector computed tomography (MDCT). Ann ICRP. 2007;37: ICRP-103. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103, vol. Ann. ICRP 37; Lee SW, Jin JY, Guan H, Martin F, Kim JH, Yin FF. Clinical assessment and characterization of a dual tube kilovoltage X-ray localization system in the radiotherapy treatment room. J Appl Clin Med Phys. 2008;9: Ding G. Characteristics of x-rays from ExacTrac and patient dose from imaging procedures. Med Phys. 2015;42: Dieterich S, Cavedon C, Chuang CF, et al. Report of AAPM TG 135: quality assurance for robotic radiosurgery. Med Phys. 2011;38:

14 e97 Ding et al.: Image Guidance Doses to Radiotherapy Patients e Sullivan A, Ding G. Additional imaging guidance dose to patient organs resulting from x-ray tubes used in CyberKnife image guidance system. Med Phys. 2015;42: Islam MK, Purdie TG, Norrlinger BD, et al. Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy. Med Phys. 2006;33: Wen N, Guan H, Hammoud R, et al. Dose delivered from Varian s CBCT to patients receiving IMRT for prostate cancer. Phys Med Biol. 2007;52: Walters BR, Ding GX, Kramer R, Kawrakow I. Skeletal dosimetry in cone beam computed tomography. Med Phys. 2009;36: Ding GX, Coffey CW. Radiation dose from kilovoltage cone beam computed tomography in an image-guided radiotherapy procedure. Int J Radiat Oncol Biol Phys. 2009;73: Downes P, Jarvis R, Radu E, Kawrakow I, Spezi E. Monte Carlo simulation and patient dosimetry for a kilovoltage cone-beam CT unit. Med Phys. 2009;36: Gayou O, Parda DS, Johnson M, Miften M. Patient dose and image quality from mega-voltage cone beam computed tomography imaging. Med Phys. 2007;34: Ding GX, Duggan DM, Coffey C. Additional doses to critical organs from CBCT scan in image guided radiation therapy. Int J Radiat Oncol Biol Phys. 2007;69:S Ding GX, Duggan DM, Coffey CW. Accurate patient dosimetry of kilovoltage cone-beam CT in radiation therapy. Med Phys. 2008;35: Ding GX, Munro P. Radiation exposure to patients from image guidance procedures and techniques to reduce the imaging dose. Radiother Oncol. Jul 2013;108: Chow JC, Leung MK, Islam MK, Norrlinger BD, Jaffray DA. Evaluation of the effect of patient dose from cone beam computed tomography on prostate IMRT using Monte Carlo simulation. Med Phys. 2008;35: Ding GX, Coffey CW. Dosimetric evaluation of the OneDoseTM MOSFET for measuring kilovoltage imaging dose from image-guided radiotherapy procedures. Med Phys. 2010;37: Ding GX, Munro P, Pawlowski J, Malcolm A, Coffey CW. Reducing radiation exposure to patients from kv-cbct imaging. Radiother Oncol. 2010;97: Deng J, Chen Z, Roberts KB, Nath R. Kilovoltage imaging doses in the radiotherapy of pediatric cancer patients. Int J Radiat Oncol Biol Phys. 2012;82: Ding G, Munro P. Reduced CBCT imaging dose due to the new x-ray source in TrueBeam TM. Med Phys. 2011;38: Stock M, Palm A, Altendorfer A, Steiner E, Georg D. IGRT induced dose burden for a variety of imaging protocols at two different anatomical sites. Radiother Oncol. 2011;102: Hyer DE, Hintenlang DE. Estimation of organ doses from kilovoltage cone-beam CT imaging used during radiotherapy patient position verification. Med Phys. 2010;37: Palm A, Nilsson E, Herrnsdorf L. Absorbed dose and dose rate using the Varian OBI 1.3 and 1.4 CBCT system. J Appl Clin Med Phys. 2010;11: Kim S, Yoo S, Yin FF, Samei E, Yoshizumi T. Kilovoltage cone-beam CT: comparative dose and image quality evaluations in partial and fullangle scan protocols. Med Phys. 2010;37: Kan MW, Leung LH, Wong W, Lam N. Radiation dose from cone beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys. 2008;70: Shah A, Aird E, Shekhdar J. Contribution to normal tissue dose from concomitant radiation for two common kv-cbct systems and one MVCT system used in radiotherapy. Radiother Oncol. 2012;105: Walter C, Boda-Heggemann J, Wertz H, et al. Phantom and in-vivo measurements of dose exposure by image-guided radiotherapy (IGRT): MV portal images vs. kv portal images vs. cone-beam CT. Radiother Oncol. 2007;85: Jeng SC, Tsai CL, Chan WT, Tung CJ, Wu JK, Cheng JC. Mathematical estimation and in vivo dose measurement for cone-beam computed tomography on prostate cancer patients. Radiother Oncol. 2009;92: Zhang Y, Yan Y, Nath R, Bao S, Deng J. Personalized assessment of kv cone beam computed tomography doses in image-guided radiotherapy of pediatric cancer patients. Int J Radiat Oncol Biol Phys. 2012;83: Perks JR, Lehmann J, Chen AM, Yang CC, Stern RL, Purdy JA. Comparison of peripheral dose from image-guided radiation therapy (IGRT) using kv cone beam CT to intensity-modulated radiation therapy (IMRT). Radiother Oncol. 2008;89: Miften M, Gayou O, Reitz B, Fuhrer R, Leicher B, Parda DS. IMRT planning and delivery incorporating daily dose from mega-voltage cone-beam computed tomography imaging. Med Phys. 2007;34: Morin O, Gillis A, Descovich M, et al. Patient dose considerations for routine megavoltage cone-beam CT imaging. Med Phys. 2007;34: Flynn RT, Hartmann J, Bani-Hashemi A, et al. Dosimetric characterization and application of an imaging beam line with a carbon electron target for megavoltage cone beam computed tomography. Med Phys. 2009;36: VanAntwerp AE, Raymond SM, Addington MC, Gajdos S, Vassil A, Xia P. Dosimetric evaluation between megavoltage cone-beam computed tomography and body mass index for intracranial, thoracic, and pelvic localization. Med Dosim. 2011;36: Tomic N, Devic S, DeBlois F, Seuntjens J. Comment on reference radiochromic film dosimetry in kilovoltage photon beams during CBCT image acquisition [Med. Phys. 37, (2010)]. Med Phys. 2010;37: Hyer DE, Serago CF, Kim S, Li JG, Hintenlang DE. An organ and effective dose study of XVI and OBI cone-beam CT systems. J Appl Clin Med Phys. 2010;11: Ding GX, Malcolm AW. An optically stimulated luminescence dosimeter for measuring patient exposure from imaging guidance procedures. Phys Med Biol. 2013;58: Zhang Y, Yan Y, Nath R, Bao S, Deng J. Personalized estimation of dose to red bone marrow and the associated leukaemia risk attributable to pelvic kilo-voltage cone beam computed tomography scans in image-guided radiotherapy. Phys Med Biol. 2012;57: Deng J, Chen Z, Yu JB, Roberts KB, Peschel RE, Nath R. Testicular doses in image-guided radiotherapy of prostate cancer. Int J Radiat Oncol Biol Phys. 2011;82:e39 e Spezi E, Downes P, Radu E, Jarvis R. Monte Carlo simulation of an x- ray volume imaging cone beam CT unit. Med Phys. 2009;36: Ding A, Gu J, Trofimov AV, Xu XG. Monte Carlo calculation of imaging doses from diagnostic multidetector CT and kilovoltage cone-beam CT as part of prostate cancer treatment plans. Med Phys. 2010;37: Alaei P, Ding GX, Gua H. Inclusion of the dose from kilovoltage cone beam CT in the radiation therapy treatment plans. Med Phys. 2010;37: Alaei P, Spezi E, Reynolds M. Dose calculation and treatment plan optimization including imaging dose from kilovoltage cone beam computed tomography. Acta Oncol. 2014;53: Mackie TR, Kapatoes J, Ruchala K, et al. Image guidance for precise conformal radiotherapy. Int J Radiat Oncol Biol Phys. 2003;56: Meeks SL, Harmon Jr JF, Langen KM, Willoughby TR, Wagner TH, Kupelian PA. Performance characterization of megavoltage computed tomography imaging on a helical tomotherapy unit. Med Phys. 2005;32: Dische S, Saunders MI, Williams C, Hopkins A, Aird E. Precision in reporting the dose given in a course of radiotherapy. Radiother Oncol. 1993;29: Saunders MI, Dische S, Grosch EJ, et al. Experience with CHART. Int J Radiat Oncol Biol Phys. 1991;21: ICRU-62. Prescribing, recording and reporting photon beam therapy (supplement to ICRU report 50), Bethesda. USA; Brahme A. Dosimetric precision requirements in radiation therapy. Acta Radiol Oncol. 1984;23: Dutreix A. When and how can we improve precision in radiotherapy? Radiother Oncol. 1984;2: Mijnheer BJ, Battermann JJ, Wambersie A. What degree of accuracy is required and can be achieved in photon and neutron therapy? Radiother Oncol. 1987;8:

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy George Ding, Ron Price, Charles Coffey Vanderbilt-Ingram Cancer Center Vanderbilt University Medical Center, Nashville, TN Introduction

More information

Managing the imaging dose during image-guided radiation therapy

Managing the imaging dose during image-guided radiation therapy Managing the imaging dose during image-guided radiation therapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Richmond VA Imaging during radiotherapy Radiographic

More information

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using

More information

Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations

Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations

More information

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1 1 2 KV CBCT Imaging Part I Rabih Hammoud, MS, DABR Henry Ford Health System Detroit, Michigan Acknowledgements Indrin Chetty, PhD Teamour Nurushev, PhD Harrison Guan, PhD Jinkoo Kim, PhD JianYue Jin, PhD

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

Feasibility of improving cone-beam CT number consistency using a scatter correction algorithm.

Feasibility of improving cone-beam CT number consistency using a scatter correction algorithm. Thomas Jefferson University Jefferson Digital Commons Department of Medicine Faculty Papers Department of Medicine 11-4-2013 Feasibility of improving cone-beam CT number consistency using a scatter correction

More information

Image quality & dosimetric property of an investigational Imaging Beam Line MV-CBCT

Image quality & dosimetric property of an investigational Imaging Beam Line MV-CBCT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 3, Summer 2009 Image quality & dosimetric property of an investigational Imaging Beam Line MV-CBCT Chris Beltran, 1,a Renin Lukose, 1 Bijumon

More information

Concept for quantifying the dose from image guided radiotherapy

Concept for quantifying the dose from image guided radiotherapy Schneider et al. Radiation Oncology (2015) 10:188 DOI 10.1186/s13014-015-0492-7 RESEARCH Open Access Concept for quantifying the dose from image guided radiotherapy Uwe Schneider 1,2*, Roger Hälg 1,2 and

More information

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of Medicine Indianapolis, IN, USA Indra J. Das,

More information

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland IGRT1 technologies Paweł Kukołowicz Warsaw, Poland Minimal prerequisite for good, efficient radiotherapy ICTP 2015 Paweł Kukołowicz 2/29 Minimal prerequisite for good, efficient radiotherapy Well trained

More information

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor IMRT/IGRT Patient Treatment: A Community Hospital Experience Charles M. Able, Assistant Professor Disclosures I have no research support or financial interest to disclose. Learning Objectives 1. Review

More information

TESTICULAR DOSES IN IMAGE-GUIDED RADIOTHERAPY OF PROSTATE CANCER

TESTICULAR DOSES IN IMAGE-GUIDED RADIOTHERAPY OF PROSTATE CANCER doi:10.1016/j.ijrobp.2011.01.071 Int. J. Radiation Oncology Biol. Phys., Vol. 82, No. 1, pp. e39 e47, 2012 Copyright Ó 2012 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/$ - see front

More information

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital EBT Process Diagnosis Simulation Tx Planning Tx Verification Tx Delivery X-ray CT MRI NM Patho Process

More information

Unrivaled, End-to-End

Unrivaled, End-to-End PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

Reducing excess radiation from portal imaging of pediatric brain tumors

Reducing excess radiation from portal imaging of pediatric brain tumors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 5, 2013 Reducing excess radiation from portal imaging of pediatric brain tumors Moses Tam, 1 Maya Mathew, 1 Christine J. Hitchen, 1 Ashwatha

More information

Collapsed Cone Convolution 2D illustration

Collapsed Cone Convolution 2D illustration Collapsed Cone Convolution 2D illustration 8 cones Energy desposition decreases very quickly with distance Energy is absorber in blue pixels only. IGRT1 technologies Paweł Kukołowicz Warsaw, Poland IGRT

More information

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by: SRS PHANTOM SCAN PLAN LOCALIZE TREAT THE COMPREHENSIVE END-TO-END SRS PHANTOM distributed by: Tel: +33 (0) 42 88 68 41 info@orion-france.com www.orion-france.com 2, Avenue du General Balfourier 75016 Paris,

More information

Quanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac.

Quanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac. IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-4861.Volume 9, Issue 5 Ver. IV (Sep. - Oct. 2017), PP 46-52 www.iosrjournals.org Quanitative and qualitative analysis of Image Quality and Imaging

More information

Monte Carlo Dose Calculation for Radiotherapy Treatment Planning

Monte Carlo Dose Calculation for Radiotherapy Treatment Planning Monte Carlo Dose Calculation for Radiotherapy Treatment Planning Monte Carlo Dose Calculation for Radiotherapy Treatment Planning C-M M Charlie Ma, Ph.D. Department of Radiation Oncology Fox Chase Cancer

More information

Helical Tomotherapy Experience. TomoTherapy Whole Brain Head & Neck Prostate Lung Summary. HI-ART TomoTherapy System. HI-ART TomoTherapy System

Helical Tomotherapy Experience. TomoTherapy Whole Brain Head & Neck Prostate Lung Summary. HI-ART TomoTherapy System. HI-ART TomoTherapy System The Challenges Associated with Differential Dose Delivery using IMRT Chester Ramsey, Ph.D. Director of Medical Physics Thompson Cancer Center Knoxville, Tennessee, U.S.A Collaborators Chester Ramsey, Ph.D.

More information

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations Aiping Ding, X. George Xu Rensselaer Polytechnic Institute Troy, NY USA http://rrmdg.rpi.edu 1 Acknowledgements

More information

ph fax

ph fax This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

biij Initial experience in treating lung cancer with helical tomotherapy

biij Initial experience in treating lung cancer with helical tomotherapy Available online at http://www.biij.org/2007/1/e2 doi: 10.2349/biij.3.1.e2 biij Biomedical Imaging and Intervention Journal CASE REPORT Initial experience in treating lung cancer with helical tomotherapy

More information

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and

More information

SRS Uncertainty: Linac and CyberKnife Uncertainties

SRS Uncertainty: Linac and CyberKnife Uncertainties SRS Uncertainty: Linac and CyberKnife Uncertainties Sonja Dieterich, PhD Linac/CyberKnife Technological Uncertainties 1 Linac Mechanical/Radiation Isocenters Depuydt, Tom, et al. "Computer aided analysis

More information

1. Murphy, Martin J., et al. "The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75.

1. Murphy, Martin J., et al. The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75. 1 2 1. Murphy, Martin J., et al. "The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75." Medical physics 34.10 (2007): 4041-4063. 3 1. Murphy, Martin J., et

More information

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA Independent Dose Verification for IMRT Using Monte Carlo C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA Outline Why Monte Carlo for IMRT QA? Experimental verification

More information

IMAGING DOSES IN RADIATION THERAPY FROM KILOVOLTAGE CONE-BEAM COMPUTED TOMOGRAPHY

IMAGING DOSES IN RADIATION THERAPY FROM KILOVOLTAGE CONE-BEAM COMPUTED TOMOGRAPHY IMAGING DOSES IN RADIATION THERAPY FROM KILOVOLTAGE CONE-BEAM COMPUTED TOMOGRAPHY By DANIEL ELLIS HYER A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT

More information

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

7/31/2018. Research Activities: Imaging for RT Image Guidance for Pediatric Radiotherapy. Chia-ho Hua, PhD. Outline

7/31/2018. Research Activities: Imaging for RT Image Guidance for Pediatric Radiotherapy. Chia-ho Hua, PhD. Outline Research Activities: Imaging for RT Image Guidance for Pediatric Radiotherapy Chia-ho Hua, PhD Chia-ho Associate Hua, Member PhD St. Jude Children s Research Hospital AAPM-COG Joint Session, 60 th Annual

More information

Feasibility study on effect and stability of adaptive radiotherapy on kilovoltage cone beam CT

Feasibility study on effect and stability of adaptive radiotherapy on kilovoltage cone beam CT 220 research article Feasibility study on effect and stability of adaptive radiotherapy on kilovoltage cone beam CT Poonam Yadav 1,2,3, Velayudham Ramasubramanian 3, Bhudatt R. Paliwal 1,2 1 Department

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

Leila E. A. Nichol Royal Surrey County Hospital

Leila E. A. Nichol Royal Surrey County Hospital 2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry

More information

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

Eric E. Klein, Ph.D. Chair of TG-142

Eric E. Klein, Ph.D. Chair of TG-142 Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing

More information

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

Investigation of the radiation dose from cone-beam CT for image-guided radiotherapy: A comparison of methodologies

Investigation of the radiation dose from cone-beam CT for image-guided radiotherapy: A comparison of methodologies Received: 21 September 217 Revised: 31 October 217 Accepted: 3 November 217 DOI: 1.12/acm2.12239 RADIATION ONCOLOGY PHYSICS Investigation of the radiation dose from cone-beam CT for image-guided radiotherapy:

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Impact of different CBCT imaging monitor units, reconstruction slice thicknesses, and planning CT slice thicknesses on the positioning

More information

Stereotactic Radiosurgery. Extracranial Stereotactic Radiosurgery. Linear accelerators. Basic technique. Indications of SRS

Stereotactic Radiosurgery. Extracranial Stereotactic Radiosurgery. Linear accelerators. Basic technique. Indications of SRS Stereotactic Radiosurgery Extracranial Stereotactic Radiosurgery Annette Quinn, MSN, RN Program Manager, University of Pittsburgh Medical Center Using stereotactic techniques, give a lethal dose of ionizing

More information

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

8/3/2016. Outline. Site Specific IGRT Considerations for Clinical Imaging Protocols. Krishni Wijesooriya, PhD University of Virginia

8/3/2016. Outline. Site Specific IGRT Considerations for Clinical Imaging Protocols. Krishni Wijesooriya, PhD University of Virginia Site Specific IGRT Considerations for Clinical Imaging Protocols Krishni Wijesooriya, PhD University of Virginia Outline Image registration accuracies for different modalities What imaging modality best

More information

Simple methods to reduce patient dose in a Varian cone beam CT system for delivery verification in pelvic radiotherapy

Simple methods to reduce patient dose in a Varian cone beam CT system for delivery verification in pelvic radiotherapy The British Journal of Radiology, 82 (2009), 855 859 SHORT COMMUNICATION Simple methods to reduce patient dose in a Varian cone beam CT system for delivery verification in pelvic radiotherapy 1 P ROXBY,

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy Turkish Journal of Cancer Volume 37, No. 2, 2007 59 Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy MUHAMMAD BASIM

More information

Role of adaptive radiation therapy for pediatric patients with diffuse pontine glioma

Role of adaptive radiation therapy for pediatric patients with diffuse pontine glioma JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 2, spring 2011 Role of adaptive radiation therapy for pediatric patients with diffuse pontine glioma Chris Beltran, a Saumya Sharma, and Thomas

More information

Contributors to this Talk

Contributors to this Talk Tomotherapy Thomas Rockwell Mackie Tomotherapy Research Group Depts. Of Medical Physics, Human Oncology and Biomedical Engineering University of Wisconsin Madison WI 53706 Phone: (608) 262-7358 Email:

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS ICTP SCHOOL ON MEDICAL PHYSICS FOR RADIATION THERAPY DOSIMETRY AND TREATMENT PLANNING FOR BASIC AND ADVANCED APPLICATIONS March

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session Stereotactic Body Radiation Therapy Quality Assurance Educational Session J Perks PhD, UC Davis Medical Center, Sacramento CA SBRT fundamentals Extra-cranial treatments Single or small number (2-5) of

More information

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient Number of Publications Number of publications 8/3/2016 The Strikes Back! - In-Vivo Dosimetry AAPM, Washington D.C, USA, 2016 Peter Greer 1,2 (1) University of Newcastle, Australia, (2) Calvary Mater Newcastle,

More information

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment

More information

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments Jean M. Moran, Ph.D. Department of Radiation Therapy, University of Michigan, 1500 E. Medical Center Drive, Ann Arbor

More information

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases CASE STUDY Institution: Odette Cancer Centre Location: Sunnybrook

More information

The effect of region extraction accuracy on the dose calculation with CBCT images combined with MSCT images.

The effect of region extraction accuracy on the dose calculation with CBCT images combined with MSCT images. The effect of region extraction accuracy on the dose calculation with CBCT images combined with MSCT images. Poster No.: C-0648 Congress: ECR 2013 Type: Scientific Exhibit Authors: K. Usui, E. Kunieda,

More information

Intensity-Modulated and Image- Guided Radiation Treatment. Outline. Conformal Radiation Treatment

Intensity-Modulated and Image- Guided Radiation Treatment. Outline. Conformal Radiation Treatment Intensity-Modulated and Image- Guided Radiation Treatment J. Daniel Bourland, PhD Professor Departments of Radiation Oncology, Physics, and Biomedical Engineering Wake Forest University School of Medicine

More information

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011 MVCT Image Guidance and QA Robert Staton, PhD DABR MD Anderson Cancer Center Orlando ACMP Annual Meeting 2011 Disclosures MDACCO has received grant funding from TomoTherapy, Inc. Overview TomoTherapy MVCT

More information

Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*

Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark* Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark* Purpose: To evaluate an institution s 3D treatment planning process and the institution

More information

Investigation of the radiation dose from cone-beam CT for image-guided radiotherapy: A comparison of methodologies

Investigation of the radiation dose from cone-beam CT for image-guided radiotherapy: A comparison of methodologies University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part B Faculty of Engineering and Information Sciences 2018 Investigation of the radiation dose from cone-beam

More information

IGRT Protocol Design and Informed Margins. Conflict of Interest. Outline 7/7/2017. DJ Vile, PhD. I have no conflict of interest to disclose

IGRT Protocol Design and Informed Margins. Conflict of Interest. Outline 7/7/2017. DJ Vile, PhD. I have no conflict of interest to disclose IGRT Protocol Design and Informed Margins DJ Vile, PhD Conflict of Interest I have no conflict of interest to disclose Outline Overview and definitions Quantification of motion Influences on margin selection

More information

Transition to Heterogeneity Corrections. Why have accurate dose algorithms?

Transition to Heterogeneity Corrections. Why have accurate dose algorithms? Transition to Heterogeneity Corrections Eric E. Klein, M.S., Washington University, St. Louis, MO Craig Stevens, M.D., Ph.D., MD Anderson Cancer Center, Houston, TX Nikos Papinikolou, Ph.D., University

More information

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation Online MR-IG-ART Dosimetry and Dose Accumulation Deshan Yang, PhD, Associate Professor Department of Radiation Oncology, School of Medicine Washington University in Saint Louis 1 Disclosure Received research

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT April 25 nd, 2015 Lung Cancer Lung cancer is the most lethal cancer: Over 224,000 new diagnoses in the U.S. predicted

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

In-Room Radiographic Imaging for Localization

In-Room Radiographic Imaging for Localization In-Room Radiographic Imaging for Localization Fang-Fang Yin, Zhiheng Wang, Sua Yoo, Devon Godfrey, Q.-R. Jackie Wu Department of Radiation Oncology Duke University Medical Center Durham, North Carolina

More information

Overview of Advanced Techniques in Radiation Therapy

Overview of Advanced Techniques in Radiation Therapy Overview of Advanced Techniques in Radiation Therapy Jacob (Jake) Van Dyk Manager, Physics & Engineering, LRCP Professor, UWO University of Western Ontario Acknowledgements Glenn Bauman Jerry Battista

More information

Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient Using Electronic Portal Imaging Device (EPID)

Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient Using Electronic Portal Imaging Device (EPID) IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-4861.Volume 10, Issue 2 Ver. I (Mar. Apr. 2018), PP 55-59 www.iosrjournals.org Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient

More information

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,

More information

Potential systematic uncertainties in IGRT when FBCT reference images are used for pancreatic tumors

Potential systematic uncertainties in IGRT when FBCT reference images are used for pancreatic tumors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Potential systematic uncertainties in IGRT when FBCT reference images are used for pancreatic tumors Ahmad Amoush, May Abdel-Wahab,

More information

FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE. Geoffrey S. Ibbott, Ph.D. June 20, 2017

FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE. Geoffrey S. Ibbott, Ph.D. June 20, 2017 FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE Geoffrey S. Ibbott, Ph.D. June 20, 2017 1 DISCLOSURES My institution holds Strategic Partnership Research Agreements with Varian, Elekta, and Philips

More information

Practical Considerations in Preparing an Institutional Procedure of Image Guided Radiation Therapy

Practical Considerations in Preparing an Institutional Procedure of Image Guided Radiation Therapy Review Article PROGRESS in MEDICAL PHYSICS Vol. 24, No. 4, December, 2013 http://dx.doi.org/10.14316/pmp.2013.24.4.205 Practical Considerations in Preparing an Institutional Procedure of Image Guided Radiation

More information

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System Varian Edge Experience Jinkoo Kim, Ph.D Henry Ford Health System Disclosures I participate in research funded by Varian Medical Systems. Outline of Presentation Review advanced imaging in Varian Edge Linear

More information

Subject: Image-Guided Radiation Therapy

Subject: Image-Guided Radiation Therapy 04-77260-19 Original Effective Date: 02/15/10 Reviewed: 01/25/18 Revised: 01/01/19 Subject: Image-Guided Radiation Therapy THIS MEDICAL COVERAGE GUIDELINE IS NOT AN AUTHORIZATION, CERTIFICATION, EXPLANATION

More information

Guideline & Reports 医学物理学会教育委員会資料

Guideline & Reports 医学物理学会教育委員会資料 Guideline & Reports 医学物理学会教育委員会資料 2017/11 更新 report title year keyword 1 keyword 2 AAPM TG211 Classification and evaluation strategies of auto-segmentation approaches for PET 2017 PET autosegmentation

More information

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology Conflict of interest None Acknowledgement: Jim

More information

In-Room Radiographic Imaging for Localization

In-Room Radiographic Imaging for Localization In-Room Radiographic Imaging for Localization Fang-Fang Yin, Zhiheng Wang, Sua Yoo, Devon Godfrey, Q.-R. Jackie Wu Department of Radiation Oncology Duke University Medical Center Durham, North Carolina

More information

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,

More information

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit MR-Linac is a research programme. It is not available for sale and its future availability cannot be guaranteed 8/2/2017 Clinical Implementation of an MR-Guided Treatment Unit Geoffrey S. Ibbott, Ph.D.

More information

An organ and effective dose study of XVI and OBI cone-beam CT systems

An organ and effective dose study of XVI and OBI cone-beam CT systems JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 An organ and effective dose study of XVI and OBI cone-beam CT systems Daniel E. Hyer, 1a Christopher F. Serago, 2 Siyong Kim,

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D. Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation

More information

Presentation Outline. Patient Setup Imaging in RT: Getting the Most Bang for your Buck. Main Errors in RT. Hypothetical Patient Examples

Presentation Outline. Patient Setup Imaging in RT: Getting the Most Bang for your Buck. Main Errors in RT. Hypothetical Patient Examples Patient Setup Imaging in RT: Getting the Most Bang for your Buck Olivier Morin, PhD UC SF Comprehensive Cancer Center San Francisco, CA Presentation Outline Errors in RT. Radiographic films in RT. On-board

More information

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY doi:10.1016/j.ijrobp.2009.03.029 Int. J. Radiation Oncology Biol. Phys., Vol. 75, No. 1, pp. 253 259, 2009 Copyright Ó 2009 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/09/$ see front

More information

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose Petrovic B Comparison of dose calculation algorithms for stereotaxy Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in and RT Dose Borislava Petrovic 1, Aleksandra Grządziel

More information

Measurement of Dose to Critical Structures Surrounding the Prostate from. Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional

Measurement of Dose to Critical Structures Surrounding the Prostate from. Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional Measurement of Dose to Critical Structures Surrounding the Prostate from Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional Conformal Radiation Therapy (3D-CRT); A Comparative Study Erik

More information