TESTICULAR DOSES IN IMAGE-GUIDED RADIOTHERAPY OF PROSTATE CANCER

Size: px
Start display at page:

Download "TESTICULAR DOSES IN IMAGE-GUIDED RADIOTHERAPY OF PROSTATE CANCER"

Transcription

1 doi: /j.ijrobp Int. J. Radiation Oncology Biol. Phys., Vol. 82, No. 1, pp. e39 e47, 2012 Copyright Ó 2012 Elsevier Inc. Printed in the USA. All rights reserved /$ - see front matter CLINICAL INVESTIGATION Genitourinary Cancer TESTICULAR DOSES IN IMAGE-GUIDED RADIOTHERAPY OF PROSTATE CANCER JUN DENG, PH.D., ZHE CHEN, PH.D., JAMES B. YU, M.D., KENNETH B. ROBERTS, M.D., RICHARD E. PESCHEL, M.D., AND RAVINDER NATH, PH.D. Department of Therapeutic Radiology, Yale University, New Haven, CT Purpose: To investigate testicular doses contributed by kilovoltage cone-beam computed tomography (kvcbct) during image-guided radiotherapy (IGRT) of prostate cancer. Methods and Materials: An EGS4 Monte Carlo code was used to calculate three-dimensional dose distributions from kvcbct on 3 prostate cancer patients. Absorbed doses to various organs were compared between intensity-modulated radiotherapy (IMRT) treatments and kvcbct scans. The impact of CBCT scanning mode, kilovoltage peak energy (kvp), and CBCT field span on dose deposition to testes and other organs was investigated. Results: In comparison with one 10-MV IMRT treatment, a 125-kV half-fan CBCT scan delivered 3.4, 3.8, 4.1, and 5.7 cgy to the prostate, rectum, bladder, and femoral heads, respectively, accounting for 1.7%, 3.2%, 3.2%, and 8.4% of megavoltage photon dose contributions. However, the testes received 2.9 cgy from the same CBCT scan, a threefold increase as compared with 0.7 cgy received during IMRT. With the same kvp, full-fan mode deposited much less dose to organs than half-fan mode, ranging from 9% less for prostate to 69% less for testes, except for rectum, where full-fan mode delivered 34% more dose. As photon beam energy increased from 60 to 125 kv, kvcbct-contributed doses increased exponentially for all organs, irrespective of scanning mode. Reducing CBCT field span from 30 to 10 cm in the superior inferior direction cut testicular doses from 5.7 to 0.2 cgy in half-fan mode and from 1.5 to 0.1 cgy in full-fan mode. Conclusions: Compared with IMRT, kvcbct-contributed doses to the prostate, rectum, bladder, and femoral heads are clinically insignificant, whereas dose to the testes is threefold more. Full-fan CBCT usually deposits much less dose to organs (except for rectum) than half-fan mode in prostate patients. Kilovoltage CBCT contributed doses increase exponentially with photon beam energy. Reducing CBCT field significantly cuts doses to testes and other organs. Ó 2012 Elsevier Inc. Prostate cancer, Testicular doses, Image-guided radiotherapy, kvcbct, Monte Carlo simulation. INTRODUCTION Recently, kilovoltage cone-beam computed tomography (kvcbct) has been introduced into the radiotherapy community as a useful X-ray imaging tool for daily patient setup and tumor localization (1, 2). Consequently, our radiotherapy paradigms for prostate cancer have been shifting dramatically from three-dimensional conformal radiotherapy and intensity-modulated radiotherapy (IMRT) to most recent image-guided radiotherapy (IGRT). Imageguided radiotherapy of prostate cancer has been implemented with a variety of imaging techniques, such as kv radiographs and fluoroscopy, megavoltage (MV) electronic portal imaging, magnetic resonance imaging, and radiofrequency transponders (1, 2). One of the commonly used techniques usually involves daily kvcbct for patient setup verification before each fraction of radiation treatment using a kv X-ray imager attached to the linear accelerator. However, frequent use of kvcbct can potentially add a considerable amount of radiation dose to cancer patients who have already received a significant amount of radiation from MV photon treatments (3 11). Yet, this portion of kvcbct-contributed radiation dose has not been included appropriately in the calculation of total doses to the patients, owing largely to the lack of a dose calculation engine in commercial treatment planning systems for kv beams. Because the testes are one of the most radiosensitive tissues, with significant impairment of spermatogenesis occurring at doses as low as 10 cgy (12), dose deposition to the testes in radiotherapy of prostate cancer (13 18) and rectal cancer (19 22) has been a major concern for many years. Earlier studies suggested that clinically significant hypogonadism could occur at doses as low as 2 4 Gy (15, 22). Although testicular doses were found to be <2 Gy in most situations, recent frequent application of kvcbct in IGRT treatments of prostate cancer has increased the concern for testes sparing Reprint requests to: Jun Deng, Ph.D., Department of Therapeutic Radiology, Yale University, 15 York Street, LL508-Smilow, New Haven, CT Tel: (203) ; Fax: (203) ; jun.deng@yale.edu e39 Conflicts of interest: none. Received Oct 29, 2010, and in revised form Jan 24, Accepted for publication Jan 29, 2011.

2 e40 I. J. Radiation Oncology d Biology d Physics Volume 82, Number 1, 2012 because of enhanced doses from kv beams. In fact, our recent study indicated that kvcbct-contributed testicular dose has increased to 1.2 Gy during a typical IGRT treatment of prostate cancer, up by 330% compared with regular IMRT without kvcbct, yet this portion of dose was neglected in the radiotherapy design and decision making (11). Hence, the goal of this work was to apply a Monte Carlo method to systematically investigate excessive dose to the testes and other organs due to kvcbct for prostate cancer patients undergoing IGRT. Specifically, we investigated On-Board Imager (OBI; Varian Medical Systems, Palo Alto, CA) kvcbct. In this work, an EGS4/BEAM user code, MCSIM, was used for Monte Carlo treatment planning of IMRT plans for prostate cancer treated with five-field 10-MV photon beams. We also calculated three-dimensional dose distributions of a 364 arc from kvcbct scanned with pelvis protocol and a 204 arc scanned with high-quality head protocol using Monte Carlo. The impact of such clinical parameters as scanning mode, kv peak energy, and scanning field on testicular doses was also investigated. Finally, a dosimetric comparison between whole-pelvis irradiation, prostate-only IMRT, and IGRT was conducted. METHODS AND MATERIALS IGRT treatments of prostate cancer Three male adult patients diagnosed with prostate carcinoma and treated with IGRT were selected in this study. The prostate radiotherapy at Yale included a primary irradiation of 66.6 Gy (1.8 Gy 37 fractions) to the planning target volume (PTV) followed by a boost of 12.6 Gy (1.8 Gy 7 fractions) using five-field IMRT and 10-MV photons. The prostate gland and all the critical structures, such as rectum, bladder, femoral heads, and testes, were contoured and/or reviewed by the physicians consistently. Planning target volumes for primary and boost plans were generated on an Eclipse treatment planning system according to the following margin specifications: 0.8 cm all around the prostate gland except for 0.6 cm between rectum prostate interface for primary plan, and 0.6 cm all around for boost plan. No seminal vesicles were included in the PTV volume. Before each fraction, a half-fan kvcbct scan in pelvis protocol was performed for patient setup. For the purpose of this study, both half-fan scan in pelvis protocol and full-fan scan in high-quality head protocol were simulated for comparison. The contours of PTVs and critical structures in kvcbct scans were identical to those defined in IMRT treatment plans. Monte Carlo treatment planning and patient dose calculations The Monte Carlo code EGS4/BEAM was used to simulate the 10- MV photon beams from a Varian Trilogy (Varian Medical Systems) (23, 24). On the basis of our previous study, we have constructed a three-source model to represent 10-MV photon beams emanating from the linear accelerator treatment head for this work: an extended annular source for the target, a planar ring source for the primary collimator, and a planar annular source for the flattening filter (25). Likewise, we have also characterized kv photon beams from the Varian Trilogy OBI system in both half-fan and full-fan modes using EGS4/BEAM code. On the basis of beam characteristics, it was determined that a four-source model could be used to best represent the four most important components of OBI in producing kv photons, that is, the target, the exit window, the blades, and the bowtie filter (11). We developed eight individual four-source models in this study to represent the eight clinically viable configurations (i.e., half-fan pelvis protocol or full-fan high-quality head protocol with 60, 80, 100, or 125 kv peak energy [kvp]). An EGS4/BEAM user code, MCSIM, was used for patient dose calculations with multiple source models as beam input. The dose calculation accuracy of MCSIM has been benchmarked against EGS4/BEAM/DOSXYZ and verified by extensive phantom measurements (26, 27). Specifically, in simulating IMRT treatments, each dynamic multileaf collimeter (MLC) field was simulated by changing particle weighting factors using an intensity distribution reconstructed from the MLC leaf sequence file for that field. When simulating the half-fan kvcbct in a pelvis protocol (variable kv, 80 ma, 12 ms, 364 arc, half-bowtie filter), a series of coplanar fields around the gantry rotation axis was simulated to mimic the 364 arc. Similarly, the full-fan mode with default settings of a high-quality head protocol (variable kv, 80 ma, 25 ms, 204 arc, full-bowtie filter) was also simulated. In MCSIM patient dose calculations, the energy cut-offs for electrons (ECUT) and for photons (PCUT), and the energy thresholds for d-ray production (AE) and for bremsstrahlung production (AP) were set as ECUT = AE = 521 kev and PCUT = AP = 10 kev for both 10-MV and kv photon beams. A statistical uncertainty (1s) of 0.5% and 2% has been achieved for 10-MV and kv beam dose calculations, respectively. To convert Monte Carlo simulations into absolute doses, for 10-MV photons, Monte Carlo simulation was performed to a calibration point in a water phantom at 100-cm source-to-surface distance (SSD) for a 10 cm 10 cm field using MCSIM with three-source model as beam input. The calibration point was at 2.4-cm depth along the central axis where linear accelerator beam output has been tuned to be 1 cgy per monitor unit, per TG-51 protocol (28). The reciprocal of the dose to the calibration point gave a unique conversion factor to this three-source model. For kv photons, absorbed dose was first measured at isocenter of an acrylic ball phantom of 5-cm diameter per TG-61 protocol with an EXRADIN A12 ionization chamber (Standard Imaging, Middleton, WI) for a 125-kV CBCT acquisition in pelvis protocol (29). Monte Carlo simulation was then performed to a chamber volume inside the ball phantom with the same beam setup using a corresponding four-source model. The ratio of the two yielded a conversion factor, which was unique to the clinical setup and beam configuration. We repeated the process for eight different scanning scenarios (i.e., half-fan or full-fan mode with 60, 80, 100, or 125 kvp) and obtained eight unique conversion factors. With all those conversion factors, all the Monte Carlo simulations in patient CT geometry can be converted into absolute doses to water. Dose measurements of MV and kv photon beams To validate Monte Carlo simulations, we have first measured central axis percent depth doses (PDD) and dose profiles for both 10-MV and 125-kV photons with an IBA blue phantom water scanning system (IBA Dosimetry America, Bartlett, TN). Specifically, for 10-MV photons, the PDDs were measured with a CC-13 ionization chamber while the dose profiles were scanned with a photon diode. The field size was 10 cm 10 cm at 100-cm SSD, and the three depths for profile measurements were 2.4, 10, and 20 cm. For 125-kV photons, both PDD and dose profiles were scanned with two CC-13 ionization chambers with one as field detector and the other as reference detector. Default blade settings in OBI were used (i.e., X1 = 8.3 cm, X2 = 24.9 cm, Y1 = Y2 = 11.8 cm for half-fan mode; and X1 = X2 = 15.1 cm and Y1 = Y2 = 11.8 cm for full-fan mode at 100-cm SSD, respectively).

3 Testicular doses during prostate IGRT d J. DENG et al. e41 Furthermore, we measured point doses at the center of a rectilinear solid water phantom (25 cm 25 cm 23 cm) using a Capintec PR-05P mini-chamber (Capintec, Ramsey, NJ) for the five-field 10-MV IMRT plans used for radiotherapy of three prostate cancer patients. For kv photon beams, we also measured absolute doses in the same way with an A12 ionization chamber for the eight scanning scenarios being investigated in this study. kvcbct scanning mode, beam energy, and scanning field To study the impact of CBCT scanning mode, photon beam energy, and scanning field on testicular dose during IGRT treatments of prostate cancer, we performed a series of Monte Carlo simulations on patient CT anatomy with variable clinical parameters while keeping the rest of the parameters intact. Those clinical parameters being investigated included half-fan vs. full-fan mode, peak photon energy of 60, 80, 100, or 125 kv, and scanning field along the superior inferior direction ranging from 10 to 30 cm. Testicular doses in whole-pelvis irradiation, prostate-only IMRT, and IGRT To evaluate testicular doses in three treatment modalities of prostate cancer, Monte Carlo simulations were performed on patient CT anatomy with beam arrangements, MLC leaf sequence patterns, and monitor unit settings identical to the actual radiotherapy treatments. Specifically, whole-pelvis irradiation was done with a fourfield box technique with photon jaws defining 14 cm 15.4 cm field size and MLC blocking four corners. RESULTS Validation of Monte Carlo simulations As shown in Figs. 1a and 1c, the central axis PDD and dose profiles in water at three depths (i.e., 2.4, 10, and 20 cm) of a 10-MV photon beam calculated by Monte Carlo were compared with measurements for a 10 cm 10 cm field at 100-cm SSD. The maximum differences in measured and calculated PDDs were <1.1% from the d max of 2.4 cm to the depth of 25 cm and <2% between the depth of 0.3 cm and d max. The dose profile comparisons indicated a better than 1.1%/1 mm agreement between the two. A similar comparison between Monte Carlo simulations and measurements in water for 125-kV photon beams is also shown in Figs. 1b and 1d. The PDD comparison showed a <2.3% differences from 0.1 cm down to 20 cm, whereas the dose profile comparisons along the x axis at depths of 1, 5, and 10 cm yielded a better than 3%/1 mm agreement between the two. The overall agreement between the measurements and Monte Carlo simulations has verified the accuracy of our Monte Carlo Fig. 1. Comparison of Monte Carlo simulations with the measurements for central axis percent depth doses of (a) 10-MV photons and (b) 125-kV photons at 100-cm source-to-skin distance, and for dose profiles along x axis of (c) 10-MV photons at depths of 2.4, 10, and 20 cm and (d) 125-kV photons at depths of 1, 5, and 10 cm. Error bars representing 1s standard deviation were approximately 0.5% for Monte Carlo simulations and hardly visible in the graphs.

4 e42 I. J. Radiation Oncology d Biology d Physics Volume 82, Number 1, 2012 Table 1. Comparison between ionization chamber measurements and Monte Carlo simulations in solid water phantoms for 10-MV five-field IMRT prostate plans and kvcbct scans in half-fan and full-fan modes with 60, 80, 100, and 125 kvp IMRT kvcbct half-fan pelvis protocol kvcbct full-fan high-quality head protocol Parameter Pt 1 Pt 2 Pt 3 60 kv 80 kv 100 kv 125 kv 60 kv 80 kv 100 kv 125 kv Measurements (cgy) Monte Carlo (cgy) (MC-Mea)/Mea (100%) Abbreviations: IMRT = intensity-modulated radiotherapy; kvcbct = kilovoltage cone-beam computed tomography; kvp = kilovoltage peak energy; Pt = patient; MC = Monte Carlo; Mea = measurement. All doses in table are in cgy per fraction for IMRT or per scan for kvcbct. simulations using multiple source models for both 10-MV and 125-kV photon beams. Table 1 lists the absolute dose comparison between Monte Carlo simulations and ionization chamber measurements in solid water phantom for both IMRT plans and kvcbct scans. Overall, the Monte Carlo predicted absorbed doses at isocenter of the phantom were within 1% of the measurements for the three IMRT plans and within 3.1% of the measurements for kvcbct scans, with kvp ranging from 60 to 125 kv in either half-fan or full-fan mode. The excellent agreement (approximately 1%) between the measurements and Monte Carlo simulations for 10-MV IMRT plans verified our Monte Carlo multiple source modeling, MLC leaf sequencing, and absolute dose calculation accuracy. Considering the statistical uncertainty of Monte Carlo simulations (approximately 2%) and measurement uncertainty (approximately 2%) for kv photon beams, the differences of up to 3.1% between the two were clinically acceptable. On the basis of this validation study, we have demonstrated the accuracy of both MVand kv photon beam dose calculations in patient CT anatomy with Monte Carlo multiple source models. Dosimetric comparison of 10-MV IMRT vs. 125-kV CBCT in prostate IGRT The comparison of Monte Carlo calculated dose distributions for a 10-MV five-field IMRT plan and a 125-kV CBCT scan in half-fan mode is shown in Fig. 2, with the corresponding dose volume histograms compared in Fig. 3. Shown in Fig. 2a is a typical IMRT plan: a highly conformal dose distribution around the prostate gland with adjacent critical structures largely spared. For adult men, because the testes were on average 12 cm apart inferior to the prostate, the testes Fig. 2. Comparison of Monte Carlo calculated dose distributions of a five-field 10-MV intensity-modulated radiotherapy treatment (a, b) and a 125-kV half-fan cone-beam computed tomography scan in pelvis protocol (c, d) on a prostate cancer patient. The arrow indicates the 23.6-cm field span along the superior inferior direction with respect to the isocenter.

5 Testicular doses during prostate IGRT d J. DENG et al. e43 8.4% more dose to prostate, rectum, bladder, and femoral heads, respectively, while inducing a 300% increase in testicular dose, as compared with IMRT. Because of enhanced photoelectric effect in bone for kv X-rays compared with MV photons, the kvcbct-contributed doses were most significant at femoral heads and sacrum (approximately 4.5% of prescribed dose). High doses were also observed on anterior and posterior (AP) sides of patient anatomy (approximately 3% of prescribed dose) owing to the 364 arc, with slightly higher doses on left lateral than on right lateral, consistent with earlier measurements (4, 5) and Monte Carlo simulations (6, 8, 10). Impact of kvcbct scanning mode on testicular dose The most important difference between half-fan and fullfan CBCT scans came from the fact that whereas the half-fan mode involved a 364 arc rotating around patient anatomy, the full-fan mode delivered a 204 arc between patient s left posterior and right posterior in patient supine position. Compared in Figs. 3b and 3c are dose volume histograms of various organs due to half-fan and full-fan CBCTs scanned at 125 kvp, respectively, with mean organ doses shown in Fig. 4a and organ dose ratios shown in Fig. 4b. Overall, full-fan CBCT deposited less dose to almost all the organs than half-fan CBCT, ranging from 9% less for prostate to 69% less for testes. The large sparing of testes in full-fan mode was primarily caused by their anterior and superficial location. The only exception was rectum, where full-fan mode delivered 34% more dose owing to its posterior location. On the basis of our results, a compromise has to be made in choosing kvcbct scanning mode when both rectum and testes needed to be spared simultaneously. Fig. 3. Comparison of dose volume histograms of (a) a five-field 10-MV intensity-modulated radiotherapy treatment, (b) a 125-kV half-fan cone-beam computed tomography (CBCT) scan, and (c) a 125-kV full-fan cone-beam computed tomography scan on a prostate cancer patient. essentially lay 8 cm outside of the field and underneath photon jaws. Because photon jaws blocked almost all of the primary photons, the testicular dose from direct transmission and indirect scatter was reduced significantly, as shown in Fig. 2b and Table 2. The 125-kV CBCT, on the other hand, delivered significantly more dose to the testes, as shown in Figs. 2c and 2d, owing to large field span (23.6 cm) along the superior inferior direction defined in default half-fan mode. According to Table 2, the 125-kV half-fan CBCT added 1.7%, 3.2%, 3.2%, and Impact of kvcbct beam energy on testicular dose An early study indicated that kvcbct-contributed doses were essentially linearly dependent on the setting of milliampere-second (mas) (5). In this study, we demonstrated that kvcbct-contributed doses increased dramatically as kvp increased from 60 to 125 kv for all the organs, regardless of scanning mode, as shown in Fig. 5. On the basis of mean absorbed doses listed in Table 2, our further investigation revealed that kvcbct-contributed mean doses followed an exponential relationship with kvp for all the organs. Specifically, testicular doses were least dependent on kvp as compared with doses to other organs in both half-fan and full-fan modes. Impact of kvcbct scanning field on testicular dose The fact that testicular doses depended heavily on field size has been reported in early studies in which MV photon beams were used in radiotherapy of prostate and rectal cancers (13, 20). In this work, we have manually set a variety of locations for Y blades (responsible for CBCT field span along the superior inferior direction) in our Monte Carlo simulations and calculated dose depositions accordingly. As shown in Fig. 6a, increasing the scanning field from 10 to 30 cm caused significant dose deposition to various organs from half-fan kvcbct, ranging from 82% increase for

6 e44 I. J. Radiation Oncology d Biology d Physics Volume 82, Number 1, 2012 Table 2. Mean absorbed doses to the organs compared between PO-IMRT and kvcbct scans at half-fan and full-fan modes Organ PO-IMRT (10 MV) kvcbct half-fan pelvis protocol kvcbct full-fan high-quality head protocol 60 kv 80 kv 100 kv 125 kv 60 kv 80 kv 100 kv 125 kv Prostate Rectum Bladder Testes Left femoral head Right femoral head Abbreviations: PO-IMRT = prostate-only intensity-modulated radiotherapy; kvcbct = kilovoltage cone-beam computed tomography. All doses in table are in cgy per fraction for PO-IMRT or per scan for kvcbct. rectum and bladder to 3200% increase for testes. A similar relationship is also observed in Fig. 6b for full-fan kvcbct, for which the dose increase ranged from 73% for rectum to 1600% for testes. Further investigation also indicated that whereas CBCT-contributed doses to other organs increased with CBCT field and gradually reached a plateau at a field size of 30 cm, the testicular doses increased exponentially with field size, regardless of scanning mode. Dosimetric comparison of WPI vs. PO-IMRT vs. IGRT Whole-pelvis irradiation (WPI), prostate-only IMRT (PO- IMRT), and IGRT are compared in Table 3 in terms of dose depositions to various organs. Generally speaking, in comparison with IMRT, half-fan kvcbct added only 1.7%, 3.2%, 3.2%, and 8.4% more doses to prostate, rectum, bladder, and femoral heads, respectively, but it contributed 400% more dose to the testes. On the other hand, WPI delivered much more dose to testes, rectum, bladder, and femoral heads as compared with PO-IMRT and IGRT, indicating excellent normal tissue sparing achieved in IMRT. DISCUSSION According to this study, kvcbct contributed significantly more dose to testes compared with PO-IMRT, whereas doses to the rest of the organs at risk were minimal and clinically insignificant. This was mainly due to two reasons: (1) kvcbct scanned with a large superior inferior field whose inferior border touched the superior portion of testes, whereas PO-IMRT focused high doses to PTV with photon jaws and MLC leaves heavily blocking MV photon beams outside of the field, and (2) testes superficial location made them generally accessible to kv photons with large scattering power. During a regular course of prostate IGRT (79.2 Gy/44 fractions), the kvcbct-contributed testicular dose would be approximately 1.3 Gy, whereas rectum, bladder, and femoral heads would receive 1.7, 1.8, and 2.5 Gy, respectively. Although 1 3 Gy doses were added, the total doses for rectum, bladder, and femoral heads were only 54, 57, and 32 Gy, respectively, still within tolerances. Yet, because hypogonadism can be clinically appreciated at doses of 2 4 Gy, the extra 1.3 Gy from kvcbct would create a much higher risk for testicular malfunction in the long term. The testes are very sensitive to radiation, spermatogenesis being more easily affected than androgen production (12). Testicular doses as low as 10 cgy can cause morphologic and quantitative changes to spermatogonia, whereas a total dose to the testicles exceeding 200 cgy may lead to permanent infertility (12). Various studies have reported that testicular doses in the range of 1.5 to 5.5 Gy can result in a significant increase in serum follicle-stimulating hormone Fig. 4. Comparison of (a) kilovoltage cone-beam computed tomography (kv CBCT)-contributed mean organ doses, and (b) organ dose ratios of full-fan mode to half-fan mode. Minimum, mean, and maximum ratios have been shown for each organ.

7 Testicular doses during prostate IGRT d J. DENG et al. e45 Fig. 5. Kilovoltage cone-beam computed tomography (CBCT)-contributed mean organ doses as a function of X-ray peak energy for (a) half-fan mode in pelvis protocol, and (b) full-fan mode in high-quality head protocol. and luteinizing hormone and a significant decrease in testosterone levels, with more prominent impact on men older than 70 years (15, 16, 19 22). These results strongly suggest that prominent and permanent testicular damage is sustained during radiotherapy of prostate and rectum cancers, frequently severe enough to cause hypogonadism (15, 16, 19 22). Although kvcbct-contributed doses were relatively small compared with the large amount of radiation doses from MV treatments, an attempt should be made to avoid any unwanted doses and keep the organ doses as low as reasonably achievable when applying image guidance clinically. According to this study, it is clear that by taking some simple steps, large amounts of unnecessary dose from kvcbct can be avoided. These steps include (1) minimizing CBCT field span and maximizing distances from organs at risk to field border, (2) reducing kvp if possible, and (3) choosing the appropriate scanning mode to reduce radiation exposure to organs at risk. Our recommendations based on this study of the Varian OBI system are largely applicable to other CBCT systems, such as the Elekta X-ray Volumetric Imager system (Elekta, Stockholm, Sweden), except that the absolute dose depositions to various organs due to the Varian OBI are significantly larger than those from Elekta X-ray Volumetric Imager, usually by a factor of 2 or more (3, 5, 6). In light of recent concerns regarding testicular doses (17, 18), our investigation indicated that WPI via four-field 10-MV photons deposited approximately 1% of prescription dose to testes, as compared with 0.4% in PO-IMRT and 2% in IGRT, which were lower than 3 16% of prescription dose to testes measured on a phantom with MV photons (13, 14, 16, 19, 20). The main reason for the large differences was the inferior field border. As shown in Fig. 6, the testicular doses increased exponentially with CBCT field, ranging from 8 cgy in a course of 79.2 Gy for testes 7 cm outside of field to 250 cgy when testes were 3 cm in the field. By setting the inferior field border at the level of ischial tuberosity in WPI, the testes were more than 5 cm outside of the treatment field and therefore largely spared, which was consistent with the Fig. 6. Kilovoltage cone-beam computed tomography (CBCT)-contributed mean organ doses as a function of CBCT field span for (a) half-fan mode in pelvis protocol, and (b) full-fan mode in high-quality head protocol.

8 e46 I. J. Radiation Oncology d Biology d Physics Volume 82, Number 1, 2012 Table 3. Mean absorbed doses to the organs compared between WPI, PO-IMRT, and IGRT (PO-IMRT plus daily kvcbct in half-fan pelvis protocol) Organ WPI (10 MV) PO-IMRT (10 MV) IGRT (PO-IMRT + kvcbct half-fan pelvis protocol) 10 MV + 60 kv 10 MV + 80 kv 10 MV kv 10 MV kv Prostate Rectum Bladder Testes Left femoral head Right femoral head Abbreviations: WPI = whole-pelvis irradiation; PO-IMRT = prostate-only intensity-modulated radiotherapy; IGRT = image-guided radiotherapy; kvcbct = kilovoltage cone-beam computed tomography. All doses in table are in cgy per fraction. observation of Roach et al. (30) in their recent response to an earlier work by King and Kapp (17). Another reason came from the fact that MV photons used in earlier studies produced much more neutrons and secondary particles than 10-MV beams, therefore adding more scatter doses to the testes. CONCLUSIONS In general, kvcbct deposits higher doses on the anterior and posterior sides of patient anatomy, owing to smaller dimension in the anterior posterior direction compared with lateral, and highest at femoral heads and sacrum. Full-fan CBCT usually deposits much less dose to organs (except for rectum) than half-fan mode in prostate patients. Kilovoltage CBCT contributed doses increase exponentially with photon beam energy for both half-fan and full-fan modes. Reducing the CBCT field significantly cuts doses to testes and other organs. For prostate cancer patients, kvcbct likely has a more significant impact on the testes than other critical structures. REFERENCES 1. Jaffray DA, Siewerdsen JH, Wong JW, et al. Flat-panel conebeam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys 2002;53: Dawson LA, Sharpe MB. Image-guided radiotherapy: Rationale, benefits, and limitations. Lancet Oncol 2006;7: Islam MK, Purdie TG, Norrlinger BD, et al. Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy. Med Phys 2006;33: Wen N, Guan H, Hammoud R, et al. Dose delivered from Varian s CBCT to patients receiving IMRT for prostate cancer. Phys Med Biol 2007;52: Song WY, Kamath S, Ozawa S, et al. A dose comparison study between XVI and OBI CBCT systems. Med Phys 2008;35: Chow JC, Leung MK, Islam MK, et al. Evaluation of the effect of patient dose from cone beam computed tomography on prostate IMRT using Monte Carlo simulation. Med Phys 2008;35: Ding GX, Duggan DM, Coffey CW. Characteristics of kilovoltage x-ray beams used for cone-beam computed tomography in radiation therapy. Phys Med Biol 2007;52: Ding GX, Duggan DM, Coffey CW. Accurate patient dosimetry of kilovoltage cone-beam CT in radiation therapy. Med Phys 2008;35: Ding GX, Pawlowski JM, Coffey CW. A correction-based dose calculation algorithm for kilovoltage x rays. Med Phys 2008; 35: Ding GX, Coffey CW. Radiation dose from kilovoltage cone beam computed tomography in an image-guided radiotherapy procedure. Int J Radiat Oncol Biol Phys 2009;73: Deng J, Chen Z, Nath R. Impact of kilo-voltage cone beam computed tomography on image-guided radiotherapy of prostate cancer. IFMBE Proc 2009;25/I: Howell S, Shalet S. Gonadal damage from chemotherapy and radiotherapy. Endocrinol Metab Clin North Am 1998;27: Amies CJ, Mameghan H, Rose A, et al. Testicular doses in definitive radiation therapy for localized prostate cancer. Int J Radiat Oncol Biol Phys 1995;32: Zagars GK, Pollack A. Serum testosterone levels after external beam radiation for clinically localized prostate cancer. Int J Radiat Oncol Biol Phys 1997;39: Daniell HW, Clark JC, Pereira SE, et al. Hypogonadism following prostate-bed radiation therapy for prostate carcinoma. Cancer 2001;91: Boehmer D, Badakhshi H, Kuschke W, et al. Testicular dose in prostate cancer radiotherapy: Impact on impairment of fertility and hormonal function. Strahlenther Onkol 2005;181: King CR, Kapp DS. To treat pelvic nodes or not: Could the greater testicular scatter dose from whole pelvic fields confound results of prostate cancer trials? J Clin Oncol 2009;27: King CR, Maxim PG, Hsu A, et al. Incidental testicular irradiation from prostate IMRT: It all adds up. Int J Radiat Oncol Biol Phys 2010;77: Dueland S, Guren MG, Olsen DR, et al. Radiation therapy induced changes in male sex hormone levels in rectal cancer patients. Radiother Oncol 2003;68: Hermann RM, Henkel K, Christiansen H, et al. Testicular dose and hormonal changes after radiotherapy of rectal cancer. Radiother Oncol 2005;75: Mazonakis M, Damilakis J, Varveris H, et al. Radiation dose to testes and risk of infertility from radiotherapy for rectal cancer. Oncol Rep 2006;15: Bruheim K, Svartberg J, Carlsen E, et al. Radiotherapy for rectal cancer is associated with reduced serum testosterone and increased FSH and LH. Int J Radiat Oncol Biol Phys 2008;70:

9 Testicular doses during prostate IGRT d J. DENG et al. e Nelson W, Hirayama H, Rogers DW. The EGS4 code system. Stanford Linear Accelerator Center Report SLAC-265. Stanford, CA: Stanford University; Rogers DW, Faddegon BA, Ding GX, et al. BEAM: A Monte Carlo code to simulate radiotherapy treatment units. Med Phys 1995;22: Deng J, Jiang SB, Kapur A, et al. Photon beam characterization and modeling for Monte Carlo treatment planning. Phys Med Biol 2000;45: Ma CM, Li JS, Pawlicki T, et al. A Monte Carlo dose calculation tool for radiotherapy treatment planning. Phys Med Biol 2002;47: Li JS, Pawlicki T, Deng J, et al. Validation of a Monte Carlo dose calculation tool for radiotherapy treatment planning. Phys Med Biol 2000;45: Almond PR, Biggs PJ, Coursey BM, et al. AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 1999;26: Ma CM, Coffey CW, DeWerd LA, et al. AAPM protocol for kv x-ray beam dosimetry in radiotherapy and radiobiology. Med Phys 2001;28: Roach M 3rd, Lawton CA, Chen J. To treat pelvic nodes or not? Greater testicular scatter does not explain the results of randomized trials. J Clin Oncol 2010;28:e450 e452.

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy George Ding, Ron Price, Charles Coffey Vanderbilt-Ingram Cancer Center Vanderbilt University Medical Center, Nashville, TN Introduction

More information

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using

More information

Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations

Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations

More information

Feasibility of improving cone-beam CT number consistency using a scatter correction algorithm.

Feasibility of improving cone-beam CT number consistency using a scatter correction algorithm. Thomas Jefferson University Jefferson Digital Commons Department of Medicine Faculty Papers Department of Medicine 11-4-2013 Feasibility of improving cone-beam CT number consistency using a scatter correction

More information

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland IGRT1 technologies Paweł Kukołowicz Warsaw, Poland Minimal prerequisite for good, efficient radiotherapy ICTP 2015 Paweł Kukołowicz 2/29 Minimal prerequisite for good, efficient radiotherapy Well trained

More information

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA Independent Dose Verification for IMRT Using Monte Carlo C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA Outline Why Monte Carlo for IMRT QA? Experimental verification

More information

Managing the imaging dose during image-guided radiation therapy

Managing the imaging dose during image-guided radiation therapy Managing the imaging dose during image-guided radiation therapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Richmond VA Imaging during radiotherapy Radiographic

More information

Dosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer

Dosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer Dosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer Tan Chek Wee 15 06 2016 National University Cancer Institute, Singapore Clinical Care Education Research

More information

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,

More information

Collapsed Cone Convolution 2D illustration

Collapsed Cone Convolution 2D illustration Collapsed Cone Convolution 2D illustration 8 cones Energy desposition decreases very quickly with distance Energy is absorber in blue pixels only. IGRT1 technologies Paweł Kukołowicz Warsaw, Poland IGRT

More information

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of Medicine Indianapolis, IN, USA Indra J. Das,

More information

Concept for quantifying the dose from image guided radiotherapy

Concept for quantifying the dose from image guided radiotherapy Schneider et al. Radiation Oncology (2015) 10:188 DOI 10.1186/s13014-015-0492-7 RESEARCH Open Access Concept for quantifying the dose from image guided radiotherapy Uwe Schneider 1,2*, Roger Hälg 1,2 and

More information

Quanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac.

Quanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac. IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-4861.Volume 9, Issue 5 Ver. IV (Sep. - Oct. 2017), PP 46-52 www.iosrjournals.org Quanitative and qualitative analysis of Image Quality and Imaging

More information

The effect of region extraction accuracy on the dose calculation with CBCT images combined with MSCT images.

The effect of region extraction accuracy on the dose calculation with CBCT images combined with MSCT images. The effect of region extraction accuracy on the dose calculation with CBCT images combined with MSCT images. Poster No.: C-0648 Congress: ECR 2013 Type: Scientific Exhibit Authors: K. Usui, E. Kunieda,

More information

Leila E. A. Nichol Royal Surrey County Hospital

Leila E. A. Nichol Royal Surrey County Hospital 2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry

More information

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall

More information

OPTIMIZATION OF COLLIMATOR PARAMETERS TO REDUCE RECTAL DOSE IN INTENSITY-MODULATED PROSTATE TREATMENT PLANNING

OPTIMIZATION OF COLLIMATOR PARAMETERS TO REDUCE RECTAL DOSE IN INTENSITY-MODULATED PROSTATE TREATMENT PLANNING Medical Dosimetry, Vol. 30, No. 4, pp. 205-212, 2005 Copyright 2005 American Association of Medical Dosimetrists Printed in the USA. All rights reserved 0958-3947/05/$ see front matter doi:10.1016/j.meddos.2005.06.002

More information

Utilization of cone-beam CT for offline evaluation of target volume coverage during prostate image-guided radiotherapy based on bony anatomy alignment

Utilization of cone-beam CT for offline evaluation of target volume coverage during prostate image-guided radiotherapy based on bony anatomy alignment reports of practical oncology and radiotherapy 1 7 (2 0 1 2) 134 140 Available online at www.sciencedirect.com journal homepage: http://www.elsevier.com/locate/rpor Original research article Utilization

More information

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY doi:10.1016/j.ijrobp.2009.03.029 Int. J. Radiation Oncology Biol. Phys., Vol. 75, No. 1, pp. 253 259, 2009 Copyright Ó 2009 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/09/$ see front

More information

Application of asi-kvcbct for Volume Assessment and Dose Estimation: An Offline Adaptive Study for Prostate Radiotherapy

Application of asi-kvcbct for Volume Assessment and Dose Estimation: An Offline Adaptive Study for Prostate Radiotherapy DOI:10.31557/APJCP.2019.20.1.229 Volume and Dose Assessment on Adapted CT RESEARCH ARTICLE Editorial Process: Submission:08/10/2018 Acceptance:01/05/2019 Application of asi-kvcbct for Volume Assessment

More information

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital EBT Process Diagnosis Simulation Tx Planning Tx Verification Tx Delivery X-ray CT MRI NM Patho Process

More information

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331

More information

Measurement of Dose to Critical Structures Surrounding the Prostate from. Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional

Measurement of Dose to Critical Structures Surrounding the Prostate from. Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional Measurement of Dose to Critical Structures Surrounding the Prostate from Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional Conformal Radiation Therapy (3D-CRT); A Comparative Study Erik

More information

Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment

Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment Chinese Journal of Cancer Original Article Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment Suresh Rana 1 and ChihYao Cheng 2 Abstract The volumetric modulated

More information

Image quality & dosimetric property of an investigational Imaging Beam Line MV-CBCT

Image quality & dosimetric property of an investigational Imaging Beam Line MV-CBCT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 3, Summer 2009 Image quality & dosimetric property of an investigational Imaging Beam Line MV-CBCT Chris Beltran, 1,a Renin Lukose, 1 Bijumon

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

Intensity Modulated RadioTherapy

Intensity Modulated RadioTherapy Intensity Modulated RadioTherapy A clinical application of a small accelerator University Medical Center Groningen A.A. van t Veld PhD I. Hoveijn PhD part 1 part2 CERN/KVI accelerator school, Zeegse, June

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

New Technologies for the Radiotherapy of Prostate Cancer

New Technologies for the Radiotherapy of Prostate Cancer Prostate Cancer Meyer JL (ed): IMRT, IGRT, SBRT Advances in the Treatment Planning and Delivery of Radiotherapy. Front Radiat Ther Oncol. Basel, Karger, 27, vol. 4, pp 315 337 New Technologies for the

More information

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence SCIENCE & TECHNOLOGY To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence Alburuj R. Rahman*, Jian Z. Wang, Dr. Z. Huang, Dr. J. Montebello Department of

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer Tony Kin Ming Lam Radiation Planner Dr Patricia Lindsay, Radiation Physicist Dr John Kim, Radiation Oncologist Dr Kim Ann Ung,

More information

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor IMRT/IGRT Patient Treatment: A Community Hospital Experience Charles M. Able, Assistant Professor Disclosures I have no research support or financial interest to disclose. Learning Objectives 1. Review

More information

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1 1 2 KV CBCT Imaging Part I Rabih Hammoud, MS, DABR Henry Ford Health System Detroit, Michigan Acknowledgements Indrin Chetty, PhD Teamour Nurushev, PhD Harrison Guan, PhD Jinkoo Kim, PhD JianYue Jin, PhD

More information

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS PAPER EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS Isra Israngkul Na Ayuthaya *, Sivalee Suriyapee, Phongpheath Pengvanich

More information

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,

More information

Helical Tomotherapy Experience. TomoTherapy Whole Brain Head & Neck Prostate Lung Summary. HI-ART TomoTherapy System. HI-ART TomoTherapy System

Helical Tomotherapy Experience. TomoTherapy Whole Brain Head & Neck Prostate Lung Summary. HI-ART TomoTherapy System. HI-ART TomoTherapy System The Challenges Associated with Differential Dose Delivery using IMRT Chester Ramsey, Ph.D. Director of Medical Physics Thompson Cancer Center Knoxville, Tennessee, U.S.A Collaborators Chester Ramsey, Ph.D.

More information

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D. Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation

More information

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health

More information

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During

More information

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 Lung Spine Phantom Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 The study groups are requesting that each institution keep the phantom for no more than 2 week. During this

More information

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM * Romanian Reports in Physics, Vol. 66, No. 2, P. 401 410, 2014 A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM * M. D. SUDITU 1,2, D. ADAM 1,2, R. POPA 1,2, V. CIOCALTEI

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

IMRT Planning Basics AAMD Student Webinar

IMRT Planning Basics AAMD Student Webinar IMRT Planning Basics AAMD Student Webinar March 12, 2014 Karen Chin Snyder, MS Senior Associate Physicist Department of Radiation Oncology Disclosures The presenter has received speaker honoraria from

More information

Feasibility study on effect and stability of adaptive radiotherapy on kilovoltage cone beam CT

Feasibility study on effect and stability of adaptive radiotherapy on kilovoltage cone beam CT 220 research article Feasibility study on effect and stability of adaptive radiotherapy on kilovoltage cone beam CT Poonam Yadav 1,2,3, Velayudham Ramasubramanian 3, Bhudatt R. Paliwal 1,2 1 Department

More information

A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse

A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse Peters et al. Radiation Oncology 2014, 9:153 RESEARCH Open Access A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse Samuel Peters

More information

Monte Carlo Dose Calculation for Radiotherapy Treatment Planning

Monte Carlo Dose Calculation for Radiotherapy Treatment Planning Monte Carlo Dose Calculation for Radiotherapy Treatment Planning Monte Carlo Dose Calculation for Radiotherapy Treatment Planning C-M M Charlie Ma, Ph.D. Department of Radiation Oncology Fox Chase Cancer

More information

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS 6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types

More information

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system Journal of Physics: Conference Series PAPER OPEN ACCESS Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system To cite this article: S Yani et al 2016 J.

More information

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Tagreed AL-ALAWI Medical Physicist King Abdullah Medical City- Jeddah Aim 1. Simplify and standardize

More information

UNIVERSITY OF WISCONSIN-LA CROSSE Graduate Studies

UNIVERSITY OF WISCONSIN-LA CROSSE Graduate Studies UNIVERSITY OF WISCONSIN-LA CROSSE Graduate Studies A SINGLE INSTITUTION S EXPERIENCE IN DEVELOPING A PURPOSEFUL AND EFFICIENT OFF-LINE TECHNIQUE FOR ADAPTIVE RADIOTHERAPY IN A CLINICAL ENVIRONMENT A Research

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER Bro. Dr. Collie Miller IARC/WHO Based on trends in the incidence of cancer, the International Agency for Research on Cancer (IARC) and WHO

More information

Role of adaptive radiation therapy for pediatric patients with diffuse pontine glioma

Role of adaptive radiation therapy for pediatric patients with diffuse pontine glioma JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 2, spring 2011 Role of adaptive radiation therapy for pediatric patients with diffuse pontine glioma Chris Beltran, a Saumya Sharma, and Thomas

More information

George X. Ding a) Department of Radiation Oncology, Vanderbilt University School of Medicine, Nashville, TN 37232, USA

George X. Ding a) Department of Radiation Oncology, Vanderbilt University School of Medicine, Nashville, TN 37232, USA Image guidance doses delivered during radiotherapy: Quantification, management, and reduction: Report of the AAPM Therapy Physics Committee Task Group 180 George X. Ding a) Department of Radiation Oncology,

More information

Additional Questions for Review 2D & 3D

Additional Questions for Review 2D & 3D Additional Questions for Review 2D & 3D 1. For a 4-field box technique, which of the following will deliver the lowest dose to the femoral heads? a. 100 SSD, equal dmax dose to all fields b. 100 SSD, equal

More information

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-861.Volume 6, Issue Ver. II (May-Jun. 201), PP 2- A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

The Effects of DIBH on Liver Dose during Right-Breast Treatments: A Case Study Abstract: Introduction: Case Description: Conclusion: Introduction

The Effects of DIBH on Liver Dose during Right-Breast Treatments: A Case Study Abstract: Introduction: Case Description: Conclusion: Introduction 1 The Effects of DIBH on Liver Dose during Right-Breast Treatments: A Case Study Megan E. Sullivan, B.S., R.T.(T)., Patrick A. Melby, B.S. Ashley Hunzeker, M.S., CMD, Nishele Lenards, M.S., CMD, R.T. (R)(T),

More information

Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1

Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1 Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1 Medicalconsult SA, Lisboa, Portugal; 2 Escola Superior de Tecnologias da Saúde, Lisboa, Portugal; 3 InsEtuto Tecnológico e Nuclear, InsEtuto Superior

More information

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.111 pissn 2508-4445, eissn 2508-4453 Clinical Implications of High Definition Multileaf Collimator

More information

IMAGING DOSES IN RADIATION THERAPY FROM KILOVOLTAGE CONE-BEAM COMPUTED TOMOGRAPHY

IMAGING DOSES IN RADIATION THERAPY FROM KILOVOLTAGE CONE-BEAM COMPUTED TOMOGRAPHY IMAGING DOSES IN RADIATION THERAPY FROM KILOVOLTAGE CONE-BEAM COMPUTED TOMOGRAPHY By DANIEL ELLIS HYER A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT

More information

Dosimetric consequences of tumor volume changes after kilovoltage cone-beam computed tomography for non-operative lung cancer during adaptive

Dosimetric consequences of tumor volume changes after kilovoltage cone-beam computed tomography for non-operative lung cancer during adaptive Oncology and Translational Medicine October 2015, Vol. 1, No. 5, P195 200 DOI 10.1007/s10330-015-0054-3 ORIGINAL ARTICLE Dosimetric consequences of tumor volume changes after kilovoltage cone-beam computed

More information

Reducing excess radiation from portal imaging of pediatric brain tumors

Reducing excess radiation from portal imaging of pediatric brain tumors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 5, 2013 Reducing excess radiation from portal imaging of pediatric brain tumors Moses Tam, 1 Maya Mathew, 1 Christine J. Hitchen, 1 Ashwatha

More information

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014 IROC Head and Neck Phantom Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014 The study groups are requesting that each institution keep the phantom for a period of time

More information

Image-guided adaptive radiation therapy : retrospective study and assessment of clinical workflow

Image-guided adaptive radiation therapy : retrospective study and assessment of clinical workflow The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2013 Image-guided adaptive radiation therapy : retrospective study and assessment of clinical workflow Jason

More information

Evaluation of Three-dimensional Conformal Radiotherapy and Intensity Modulated Radiotherapy Techniques in High-Grade Gliomas

Evaluation of Three-dimensional Conformal Radiotherapy and Intensity Modulated Radiotherapy Techniques in High-Grade Gliomas 1 Carol Boyd Comprehensive Case Study July 11, 2013 Evaluation of Three-dimensional Conformal Radiotherapy and Intensity Modulated Radiotherapy Techniques in High-Grade Gliomas Abstract: Introduction:

More information

1. Murphy, Martin J., et al. "The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75.

1. Murphy, Martin J., et al. The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75. 1 2 1. Murphy, Martin J., et al. "The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75." Medical physics 34.10 (2007): 4041-4063. 3 1. Murphy, Martin J., et

More information

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Can you currently use direct documentation and measurement of delivered dose to ensure you did not underdose or

More information

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

More information

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Multilayer Gafchromic film detectors for breast skin dose determination in vivo INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination

More information

Evaluation of three APBI techniques under NSABP B-39 guidelines

Evaluation of three APBI techniques under NSABP B-39 guidelines JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 1, WINTER 2010 Evaluation of three APBI techniques under NSABP B-39 guidelines Daniel Scanderbeg, a Catheryn Yashar, Greg White, Roger Rice,

More information

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS ICTP SCHOOL ON MEDICAL PHYSICS FOR RADIATION THERAPY DOSIMETRY AND TREATMENT PLANNING FOR BASIC AND ADVANCED APPLICATIONS March

More information

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES MELISSA C. MARTIN, M.S., FACR, FAAPM PRESIDENT AAPM - 2017 PRESIDENT - THERAPY PHYSICS INC., GARDENA, CA MELISSA@THERAPYPHYSICS.COM AAPM Spring

More information

Implementation of advanced RT Techniques

Implementation of advanced RT Techniques Implementation of advanced RT Techniques Tibor Major, PhD National Institute of Oncology Budapest, Hungary 2. Kongres radiološke tehnologije, Vukovar, 23-25. September 2016. Current RT equipments at NIO,

More information

CALCULATION OF OPTIMAL MARGINS BETWEEN CLINICAL TARGET VOLUME (CTV) AND PLANNING TARGET VOLUME (PTV)

CALCULATION OF OPTIMAL MARGINS BETWEEN CLINICAL TARGET VOLUME (CTV) AND PLANNING TARGET VOLUME (PTV) Available online at http://www.journalijdr.com ISSN: 2230-9926 International Journal of Development Research Vol. 07, Issue, 10, pp.15773-15779, October, 2017 ORIGINAL RESEARCH ARTICLE ORIGINAL RESEARCH

More information

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams Received: 12 May 2017 Revised: 23 October 2017 Accepted: 10 April 2018 DOI: 10.1002/acm2.12358 RADIATION ONCOLOGY PHYSICS Dose rate response of Digital Megavolt Imager detector for flattening filter-free

More information

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis Acknowledgments Thanks to Saiful Huq for an illuminating conversation about the application of TG-100 Jennifer O Daniel, Ph.D. & Fang-Fang Yin, Ph.D. Duke University Medical Center Annual AAPM Meeting,

More information

A Dosimetric Comparison of Whole-Lung Treatment Techniques. in the Pediatric Population

A Dosimetric Comparison of Whole-Lung Treatment Techniques. in the Pediatric Population A Dosimetric Comparison of Whole-Lung Treatment Techniques in the Pediatric Population Corresponding Author: Christina L. Bosarge, B.S., R.T. (R) (T) Indiana University School of Medicine Department of

More information

Intensity modulated radiotherapy (IMRT) for treatment of post-operative high grade glioma in the right parietal region of brain

Intensity modulated radiotherapy (IMRT) for treatment of post-operative high grade glioma in the right parietal region of brain 1 Carol Boyd March Case Study March 11, 2013 Intensity modulated radiotherapy (IMRT) for treatment of post-operative high grade glioma in the right parietal region of brain History of Present Illness:

More information

Organ Contour Adaptor to create new structures to use for adaptive radiotherapy of cervix cancer using Matlab Bridge and 3DSlicer / SlicerRT

Organ Contour Adaptor to create new structures to use for adaptive radiotherapy of cervix cancer using Matlab Bridge and 3DSlicer / SlicerRT Organ Contour Adaptor to create new structures to use for adaptive radiotherapy of cervix cancer using Matlab Bridge and 3DSlicer / SlicerRT Y. Seppenwoolde 1,2, M. Daniel 2, H. Furtado 2,3, D. Georg 1,2

More information

Department of Radiotherapy & Nuclear Medicine, National Cancer Institute, Cairo University, Cairo, Egypt.

Department of Radiotherapy & Nuclear Medicine, National Cancer Institute, Cairo University, Cairo, Egypt. Original article Res. Oncol. Vol. 12, No. 1, Jun. 2016:10-14 Dosimetric comparison of 3D conformal conventional radiotherapy versus intensity-modulated radiation therapy both in conventional and high dose

More information

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 703: Clinical Rotation 2 Inclusive dates of rotation: Feb. 26, 2016 Aug. 25, 2016 Director or Associate

More information

VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT

VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT Poster No.: C-0520 Congress: ECR 2011 Type: Scientific Exhibit Authors: Y. Kawasaki, S. Tadokoro,

More information

Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*

Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark* Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark* Purpose: To evaluate an institution s 3D treatment planning process and the institution

More information

Intensity-Modulated and Image- Guided Radiation Treatment. Outline. Conformal Radiation Treatment

Intensity-Modulated and Image- Guided Radiation Treatment. Outline. Conformal Radiation Treatment Intensity-Modulated and Image- Guided Radiation Treatment J. Daniel Bourland, PhD Professor Departments of Radiation Oncology, Physics, and Biomedical Engineering Wake Forest University School of Medicine

More information

Simple methods to reduce patient dose in a Varian cone beam CT system for delivery verification in pelvic radiotherapy

Simple methods to reduce patient dose in a Varian cone beam CT system for delivery verification in pelvic radiotherapy The British Journal of Radiology, 82 (2009), 855 859 SHORT COMMUNICATION Simple methods to reduce patient dose in a Varian cone beam CT system for delivery verification in pelvic radiotherapy 1 P ROXBY,

More information

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments Jean M. Moran, Ph.D. Department of Radiation Therapy, University of Michigan, 1500 E. Medical Center Drive, Ann Arbor

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

I. Equipments for external beam radiotherapy

I. Equipments for external beam radiotherapy I. Equipments for external beam radiotherapy 5 linear accelerators (LINACs): Varian TrueBeam 6, 10 & 18 MV photons, 6-18 MeV electrons, image-guided (IGRT) and intensity modulated radiotherapy (IMRT),

More information

Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient Using Electronic Portal Imaging Device (EPID)

Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient Using Electronic Portal Imaging Device (EPID) IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-4861.Volume 10, Issue 2 Ver. I (Mar. Apr. 2018), PP 55-59 www.iosrjournals.org Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient

More information

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of

More information

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 Dosimetric and volumetric changes in the rectum and bladder in patients receiving CBCT-guided prostate IMRT: analysis based on daily

More information

Eric E. Klein, Ph.D. Chair of TG-142

Eric E. Klein, Ph.D. Chair of TG-142 Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing

More information

Which Planning CT Should be Used for Lung SBRT? Ping Xia, Ph.D. Head of Medical Physics in Radiation Oncology Cleveland Clinic

Which Planning CT Should be Used for Lung SBRT? Ping Xia, Ph.D. Head of Medical Physics in Radiation Oncology Cleveland Clinic Which Planning CT Should be Used for Lung SBRT? Ping Xia, Ph.D. Head of Medical Physics in Radiation Oncology Cleveland Clinic Outline Image quality and image dose Free breathing CT, 4DCT, and synthetic

More information

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information