1 Clinical Chemistry 53: (2007) Proteomics and Protein Markers Standardized Peptidome Profiling of Human Urine by Magnetic Bead Separation and Matrix-Assisted Laser Desorption/Ionization Time-of-Flight Mass Spectrometry Georg Martin Fiedler, Sven Baumann, Alexander Leichtle, Anke Oltmann, Julia Kase, Joachim Thiery, * and Uta Ceglarek Background: Peptidome profiling of human urine is a promising tool to identify novel disease-associated biomarkers; however, a wide range of preanalytical variables influence the results of peptidome analysis. Our aim was to develop a standardized protocol for reproducible urine peptidome profiling by means of magnetic bead (MB) separation followed by matrix-assisted laser desorption/ionization time-of-flight (MALDI- TOF) mass spectrometry (MS). Methods: MBs with defined surface functionalities (hydrophobic interaction, cation exchange, and metal ion affinity) were used for peptide fractionation of urine. Mass accuracy and imprecision were calculated for 9 characteristic mass signals (M r, ). Exogenous variables (instrument performance, urine sampling/storage conditions, freezing conditions, and freeze-thaw cycles) and endogenous variables (ph, urine salt and protein concentrations, and blood and bacteria interferences) were investigated with urine samples from 10 male and 10 female volunteers. Results: We detected 427 different mass signals in the urine of healthy donors. Within- and between-day imprecision in relative signal intensities ranged from 1% to 14% and from 4% to 16%, respectively. Weak cationexchange and metal ion affinity MB preparations required adjustment of the urinary ph to 7. Storage time, Institute of Laboratory Medicine, Clinical Chemistry and Molecular Diagnostics, University Hospital, Leipzig, Germany. These authors contributed equally to this work. * Address correspondence to this author at: Institute of Laboratory Medicine, Clinical Chemistry and Molecular Diagnostics, University Hospital Leipzig, Liebigstr. 27, D Leipzig, Germany. Fax ; thiery@medizin.uni-leipzig.de. Received August 7, 2006; accepted December 21, Previously published online at DOI: /clinchem storage temperature, the number of freeze-thaw cycles, and bacterial and blood contamination significantly influenced urine peptide patterns. Individual urine peptide patterns differed significantly within and between days. This imprecision was diminished by normalization to a urinary protein content of 3.5 g. Conclusion: This reliable pretreatment protocol allows standardization of preanalytical modalities and facilitates reproducible peptidome profiling of human urine by means of MB separation in combination with MALDI-TOF MS American Association for Clinical Chemistry Peptidome analysis based on mass spectrometric screening methods, such as surface-enhanced laser desorption/ ionization time-of-flight (SELDI-TOF) 1 and matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) mass spectrometry (MS), is a promising high-throughput approach for identifying new potential biomarkers in various body fluids (1 6). By virtue of its noninvasiveness and availability, peptidome profiling of human urine is now becoming an important method for investigating kidney physiology and detecting novel disease-associated markers of renal and bladder diseases (7 13). Analyses based on Western blotting or immunologic methods have been successfully used for identifying proteins in human urine (14). Studies with 2-dimensional gel electrophoresis techniques have been carried out in the fields of metabolism, toxicology, and renal diseases (15 19). Liquid chro- 1 Nonstandard abbreviations: SELDI-TOF, surface-enhanced laser desorption/ionization time-of-flight; MALDI-TOF, matrix-assisted laser desorption/ ionization time-of-flight; MS, mass spectrometry; MB, magnetic bead; MB-HIC C8, magnetic bead hydrophobic-interaction chromatography; MB-IMAC Cu, magnetic bead immobilized metal ion affinity chromatography; MB WCX, magnetic bead weak cation-exchange chromatography; S/N, signal-to-noise. 421

2 422 Fiedler et al.: Magnetic Bead Separation and MALDI-TOF MS of Urine matography tandem MS has also been applied for the identification of urinary peptides after tryptic digestion (20). In contrast to these complex methods, SELDI-TOF MS and magnetic bead (MB) separation followed by MALDI-TOF MS enable high-throughput investigations of proteins and peptides without the laborious samplepretreatment procedures (3, 7, 21). In addition, SELDI- TOF MS and MALDI-TOF MS especially focus on the region of low M r ( ), which includes the small proteins, protein/peptide fragments, and peptides that constitute the peptidome or fragmentome and that are overlooked by traditional techniques (5). Before urinary peptidomics can proceed from the bench to the bedside, however, a large number of important barriers need to be overcome (2, 22). The preanalytical phase especially is affected by a wide range of preanalytical variables, both exogenous (instrument settings, urine collection and storage methods, freezing conditions, and the number of freeze-thaw cycles) and endogenous (ph, urine concentrations of salts and proteins, and blood and bacterial interferences), that can markedly influence the results of peptide profiling (8, 21 26). Therefore, before clinical implementation of the method can be warranted, the importance of these variables on peptidome profiling has to be investigated, and the preanalytical and analytic steps have to be standardized (22, 27 30). The aim of our study was to investigate instrument performance and the effect of preanalytical variables on urinary peptidome profiling with MB separation and MALDI-TOF MS and to establish a reliable pretreatment protocol for collecting and storing urine samples to permit valid and reproducible analyses of urinary peptide patterns. Materials and Methods chemicals, standards, and consumables Gradient-grade acetonitrile and ethanol were obtained from J.T. Baker, and pro analysi grade trifluoroacetic acid, urea, sodium chloride, and acetone were purchased from Sigma-Aldrich. Peptide Calibration Standard I, Protein Calibration Standard I, and -cyano-4-hydroxycinnamic acid were purchased from Bruker Daltonics. We used 0.2-mL polypropylene tubes (8-tube strips) for manually operated MB preparations. Automated MB preparations were performed with 96-well plates and TubePlates from Biozym, low-profile polypropylene tubes from ABgene, and Modular Reservoir Quarter Modules from Beckman Coulter. For sample storage, 450- L CryoTubes were purchased from Sarstedt. urine collection and storage First- and 2nd-morning urine samples were collected from 20 healthy volunteers (10 women and 10 men, ages years) in 100-mL urine cups (Sarstedt). Multivariable strips (Combur10 Test ; Roche Diagnostics) were used to analyze samples for specific gravity, ph, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, and blood, and a Hitachi 917 analyzer (Roche Diagnostics) was used for assays of electrolytes, total protein, creatinine (enzymatic test), and urea (for results, see Table 1 in the Data Supplement that accompanies the online version of this article at org/content/vol53/issue3). Unless otherwise stated, samples were centrifuged at g for 10 min at room temperature. Aliquots of 450 L were stored at 80 C. Urine samples were thawed at room temperature for 30 min, adjusted to ph 7, centrifuged again, and immediately processed for peptidome separation after the sample was adjusted for total protein content. Variations in sample pretreatment are described in the experimental section. The samples from the 10 female and 10 male volunteers were pooled for precision and mass accuracy analysis. peptidome separation We used ClinProt purification reagent sets (Bruker Daltonics) for peptidome separation of samples and magnetic particles (particle size, 1 m; mean pore size, 40 nm; specific surface area, 100 cm 2 /g) with defined surface functionalities [MB hydrophobic-interaction chromatography (MB HIC C8), MB immobilized metal ion affinity chromatography (MB IMAC Cu) beads, and MB weak cation-exchange chromatography (MB WCX)]. MB purifications were performed according to the manufacturer s protocol for serum and with the ClinProt liquid-handling robot. The protocol was adapted with regard to sample and binding-solution volumes. A 30- L volume of the urine sample was diluted in 60 L binding solution and added to the bead slurry (5 L MB HIC C8 beads, 5 L MB IMAC Cu beads, or 10 L MB WCX beads). After thorough stirring, samples were incubated for 1 min (MB HIC C8 and MB WCX beads) or 5 min (MB IMAC Cu beads) at room temperature. The protein content was normalized by increasing the loaded urine sample volume in a stepwise manner. After the 1st fractionation step, residual urine was removed, and the procedure was repeated with a 2nd urine fraction (60 L maximum). Following the stepwise application of sample and MB separation, the peptide fraction was eluted from MB HIC C8 beads with 5 L acetonitrile/ water (1:1, volume/volume). For MB WCX beads, we used 5 L elution solution and 4 L stabilization buffer, and we eluted the peptide fraction from MB IMAC Cu beads with 10 L of elution solution from Bruker Daltonics. To prepare the MALDI target, we spotted 1 L ofa mixture containing 10 L 0.3 g/l -cyano-4-hydroxycinnamic acid in 2:1 ethanol/acetone (volume/volume) and 1 L of the eluted peptide fraction onto the Anchor- Chip 600- m target (Bruker Daltonics). mass spectrometry A linear MALDI-TOF mass spectrometer (Autoflex; Bruker Daltonics) was used for peptidome profiling with the following settings: linear positive mode; ion source 1,

3 Clinical Chemistry 53, No. 3, kv; ion source 2, kv; lens, 9.00 kv; pulsed ion extraction, 120 ns. For matrix suppression, we used a high gating factor with signal suppression up to an M r of 500. Mass calibration was performed with the standard calibration mixture of peptides and proteins (M r range, ). Four MALDI preparations (MALDI spots) from each sample were measured. For each MALDI spot, 450 spectra were acquired (30 laser shots at 15 different spot positions). To increase detection sensitivity, we removed excess matrix with 6 shots at a laser power of 45% before data acquisition at a power of 25%. All signals with a signal-to-noise (S/N) ratio 3 inanm r range of were collected with the AutoXecute tool of the flexcontrol acquisition software (version 2.0; Bruker Daltonics), and Bruker Daltonics flexanalysis software was applied for data analysis. ClinProTools bioinformatics software (version 2.0 beta; Bruker Daltonics) was used for the recognition of peptide patterns. evaluation of exogenous variables Instrument performance. Mass accuracy and within-day and between-day reproducibility were determined by the relative peak intensities of 9 characteristic signals of the urine sample. MB HIC C8, MB WCX, and MB IMAC Cu beads were used for 10 bead preparations on 3 consecutive days. Urine storage and freeze-thaw cycles. The influence of storage temperature and time on the peptide profiles was investigated with the pooled urine sample after separation by MB HIC C8. Pooled urine was divided into 2 aliquots. One milliliter of the fresh urine aliquot and 1 ml of a deep-frozen aliquot (thawed at room temperature for 30 min) were incubated in 1.5-mL vials at 4 C, 25 C, and 40 C for 0, 1, 3, 6, 24, 72, and 168 h. At each defined time point, L aliquots were removed and prepared immediately. We also investigated the effect of up to 3 freeze-thaw cycles on the peptide pattern. A 1-mL aliquot of a pooled urine sample was frozen for 24 h and then thawed for 30 min at room temperature. Thirty microliters of urine was removed and processed immediately. The remaining sample was frozen again. The procedure was repeated twice. evaluation of endogenous variables Urine-collecting conditions. We evaluated differences between urine samples collected on the 1st and 2nd mornings with urine aliquots from 5 female and 5 male volunteers. Samples were prepared with MB HIC C8 beads according to the standard protocol. The peptide profiles were compared with respect to relative peak intensities. Influence of urea and salt. The influence of urea and salt content on peptide profiles was examined by adding urea and sodium chloride stock solutions to the pooled urine sample to achieve final concentrations of 50, 100, 250, 500, and 1000 mmol/l. A blank sample was diluted in the same way with water and analyzed to avoid misinterpretation of spectra caused by sample dilution. Blood and bacterial interferences. To investigate the effect of blood contamination on peptide pattern, we added whole blood to the pooled urine samples in blood urine ratios of 1:100, 1:500, 1:1000, 1:2500, 1:5000, 1:10 000, and 1: (vol/vol). Results were compared with uncontaminated urine samples. In addition, the influence of bacteria was determined in an ex vivo simulation. One hundred microliters of an Escherichia coli suspension ( colonyforming units/l) was added to 900 L of pooled urine samples without significant bacteriuria. These mixtures were incubated at room temperature for 0, 30, and 60 min. Urine samples were diluted 1:1 (volume/volume) with acetonitrile containing 5 g/l trifluoroacetic acid for inactivation of bacteria. The samples were then lyophilized, resuspended with 35 L water, and processed by MB MALDI-TOF MS (30 L). The results were compared with sterile urine samples. Creatinine, protein, and ph normalization. Urine samples collected midstream from 10 volunteers on the 1st and 2nd mornings were normalized to the creatinine concentration and to the protein content. The creatinine concentration was adjusted to the lowest measured creatinine concentration (4 mmol/l for the 1st morning s sample and 2.5 mmol/l for the 2nd morning s urine sample) by diluting each sample with 0.15 mol/l sodium chloride. The total protein amounts of the 1st and 2nd morning s urine samples were adjusted to 3.5 g by varying sample volume from 30 to 153 L. To evaluate the optimal ph value for MB purification, we adjusted the ph of pooled urine samples to ph 6, ph 7, ph 8, and ph 9 with 1 mol/l NH 4 HCO 3 or 1 mol/l (NH 4 ) 2 SO 4. Results evaluation of exogenous variables Instrument performance. After separation with MB HIC C8, MB WCX, and MB IMAC Cu beads, we detected 427 peaks in human urine with an S/N ratio 3 (MB HIC C8, 205 signals/spectrum; MB WCX, 217 signals/spectrum; MB IMAC Cu, 197 signals/spectrum) with an overlap frequency of 42.6%. We detected 244 signals overall with an S/N ratio 6 and 158 signals overall with an S/N ratio 10. The reproducibility of mass spectra generation was determined by evaluating relative peak intensities. For the pooled urine sample, we obtained CVs in the ranges of 1% 14% (within days) and 4% 16% (between days; see Table 2 in the online Data Supplement). A mass accuracy of 0.035% (350 ppm) was obtained in the M r range of This result corresponds well to our previous results with standardization of the serum proteome (29). We used MB HIC C8 beads for all further investigations of the effects of preanalytical variables because these MBs

4 424 Fiedler et al.: Magnetic Bead Separation and MALDI-TOF MS of Urine have a separation capability that is more versatile than the other bead modifications. Urine storage and freeze-thaw cycles. The influence of storage temperature and time on the stability of the peptide pattern was investigated by incubating fresh and once-frozen pooled urine samples. Once-frozen samples showed greater reproducibility and improved stability against increased temperatures compared with fresh urine (data not shown). We also investigated the effect of 3 freeze-thaw cycles on the peptide pattern. Generally, freezing only once decreased the relative peak intensity of some mass signals (e.g., M r s , , and ), whereas other mass signals (M r s , , , and ) increased. The variation in the relative intensities of mass signals in fresh and frozen urine is shown in Fig. 1. Compared with once-frozen samples, recurrent freeze-thaw cycles (3 ) did not further affect urine peptide patterns. The peptide patterns of once-frozen urine samples were stable up to 24 h at 4 C with no detectable degradation; however, we observed significant changes in peptide patterns at 25 C after 6 h of storage. At 40 C, peptide patterns showed significant decreases and increases in signal intensities for numerous signals by 3 h (Fig. 2). evaluation of endogenous variables Urine-collecting conditions. Urine samples collected on the 1st and 2nd mornings showed significant differences in relative signal intensities. The peptide profiles of 2nd-morning urine samples obtained with MB HIC C8 beads featured reduced intensities for several low-m r signals (e.g., , , , , and ) and by increases in other mass signals (e.g., M r s , , and ). Therefore, the peptide patterns of urine Fig. 2. Characteristic mass signals after MB HIC C8 purification following sample storage for different times and at different temperatures. Presented are results for storage at 4 C (A), 25 C (B), and 40 C (C) for the indicated times. Fig. 1. Two-dimensional mass plot comparing relative peak areas of 2 characteristic mass signals after MB HIC C8 purification with respect to the effect of freeze-thaw cycles on urine peptide pattern. The ellipses represent the standard deviations of the selected M r s of each group. samples collected on the 1st and 2nd mornings were different (Fig. 3). We also investigated the variation in individual urinary-peptide patterns by analyzing midstream urine samples collected on the 1st and 2nd mornings for 5 consecutive days. The variation in relative signal intensities for 9 characteristic signals ranged from 4% to 31% for samples collected on the 1st morning and

5 Clinical Chemistry 53, No. 3, Fig. 3. Two-dimensional mass plot comparing relative peak areas of 2 characteristic mass signals after MB HIC C8 purification with respect to 1st- and 2nd-morning urine samples. The ellipses represent the standard deviations of the selected M r s of each group. from 5% to 43% for samples collected on the 2nd morning. Compared with these peptidomic data, the median between-day variations in creatinine concentration for samples collected on the 1st and 2nd mornings were 31.9% and 28.2%, respectively, and those for total protein were 41.0% and 39.5%, respectively. Influence of urea and salt. Increasing urea and sodium chloride concentrations did not disturb sample preparation or MS analyses up to a concentration of 1000 mmol/l. Blood and bacterial interferences. Blood contamination influenced the peptide patterns of urine samples. New mass signals were detected up to a urine blood ratio of 2500:1 (see Fig. 1 in the online Data Supplement). For samples with a urine blood ratio of 100:1, we detected both completely new signals and ion-suppression effects on previously detected signals. We evaluated the effect of bacterial interference on peptide profiles with an ex vivo simulation. Comparison of the 4 sample groups (control urine and urine incubated at room temperature for 0, 30, and 60 min with colony-forming E. coli units/l) showed decreases in intensity for several signals in all urine samples incubated with a bacterial suspension for 30 min. The most significant differences were observed for mass signals of M r s and (Fig. 4). This effect was not observed immediately after the addition of the bacteria. ph, creatinine, and protein normalization. Urine ph influenced protein separation with the MB WCX and MB IMAC Cu beads. The best signal yield and highest signal intensities were obtained at ph 7 with MB WCX beads (217 signals with an S/N ratio 3) and MB IMAC Cu beads (197 signals with an S/N ratio 3). The numbers and intensities of signals with M r s 3500 were decreased Fig. 4. Characteristic mass signals (A and B) after incubating urine samples with E. coli suspension and subsequent MB HIC C8 purification. above ph 7. We observed no ph-dependent differences in peptide patterns with MB HIC C8 beads. We normalized creatinine concentrations by diluting urine samples to a particular creatinine concentration. Creatinine concentrations varied from 4.0 to 29.2 mmol/l for 1st-morning samples and from 2.5 to 15.9 mmol/l for 2nd-morning samples. We adjusted creatinine concentrations to 4 mmol/l (1st-morning samples) and 2.5 mmol/l (2nd-morning samples). A comparison of the spectra for normalized and nonnormalized urine samples after MB purification showed a decline in signals with an S/N ratio 10 of up to 15% in diluted samples, compared with undiluted samples. Signal losses of up to 35% were detected after normalizing the creatinine concentrations of 2nd-morning samples. Generally, adjustments in creatinine concentration via dilution did not decrease the variation in individual signals but produced a loss of mass signals. Concentrations of total protein for the 20 analyzed urine samples ranged from 29.0 to mg/l for 1stmorning samples and from 8.2 to mg/l for 2ndmorning samples. Sample dilution to the lowest protein concentration produced a loss of mass signals similar to that observed for creatinine. Therefore, we varied the urine sample volume from 30 to 60 L to obtain a maximal loading of the MBs. The capacity of the pipetting robot limited the maximum sample volume to 60 L.

6 426 Fiedler et al.: Magnetic Bead Separation and MALDI-TOF MS of Urine Fig. 5. Normalization effects on MB HIC C8 urine peptide patterns for 2 healthy volunteers (A and ). Without normalization, 3.4 mmol/l creatinine, 22.8 mg/l total protein, 30- L sample volume (0.7 g adsorbed; A1); without normalization, 14.4 mmol/l creatinine, mg/l total protein, 30- L sample volume (3.3 g adsorbed; B1); normalized by dilution, 2.5 mmol/l creatinine, 16.8 mg/l total protein, 30- L sample volume (0.5 g adsorbed; A2); normalized by dilution, 2.5 mmol/l creatinine, 19.1 mg/l total protein, 30- L sample volume (0.6 g adsorbed; B2); normalized by applied sample volume, 3.4 mmol/l creatinine, 22.8 mg/l total protein, 153- L sample volume (3.5 g adsorbed; A3); normalized by applied sample volume, 14.4 mmol/l creatinine, mg/l total protein, 33- L sample volume (3.5 g adsorbed; B3). Therefore, we normalized the total protein content of urine samples to 3.5 g for the 1st- and 2nd-morning samples. With MB HIC C8 purification, the number of mass signals in the urine with an S/N ratio 10 increased from 60 before protein normalization to 90 after normalization. In addition, interindividual variation in the urine peptide patterns of the 20 healthy volunteers was minimized (Fig. 5). Discussion Urinary peptidome profiling with high-throughput methods such as MB MALDI-TOF MS or SELDI-TOF MS is a promising tool in nephrology research (9, 12, 13). Both methods are of particular interest because they enable rapid analyses of many individual samples in large-scale clinical studies, but the preanalytical and analytic steps require thorough validation before clinical implementation can be warranted (8, 21, 22, 26 29). Therefore, we used MB MALDI-TOF MS to investigate instrument performance and the effects of various exogenous and endogenous variables on urine peptide profiles and established a reliable pretreatment protocol. MB MALDI-TOF MS seems to be more sensitive than SELDI-TOF MS. With 3 different MB preparations (MB HIC C8, MB WCX, and MB IMAC Cu beads), we detected 427 mass peaks with an S/N ratio 3 (M r range, ). Previous investigations of urine with SELDI-TOF MS yielded fewer mass peaks. The WCX2 ProteinChip detected 180 peaks in 138 samples (7), the CM10 (cationic) and the IMAC30 (anionic) ProteinChips detected 133 peaks in the M r range of (26), and use of the NP20 ProteinChip demonstrated only 25 peaks/spectrum (21). One reason for the improved sensitivity might be the higher adsorption capability of the beads porous surface structure compared with on-layer chips (3). Analysis of urine with MB HIC C8 beads produced fewer mass peaks with an S/N ratio 3 than serum analysis (serum, 250 peaks; urine, 200 peaks) (29), but the 20% higher proportion of signals with an S/N ratio 10 in urine samples might improve data analysis. We also determined the reproducibility and mass accuracy of the MB MALDI-TOF MS method. The within-day and between-day imprecision of urine peptidome profiling was 16%. SELDI-TOF MS produced imprecision values of up to 30% (21). One reason for the improved precision of MB MALDI-TOF MS might be the application of 4-fold spotting of the MALDI target for each urine sample. The determination of the mass accuracy (M r range, ) produced absolute differences in M r of , which are comparable to our data for serum and in accordance with the known mass accuracy of MALDI-TOF MS (29, 31). Differences in collection, sample storage, blood and bacterial interferences, and modes of normalization (creatinine, protein, and ph) of urine samples had significant effects on the variation in the peptide patterns. The time of urine collection (1st or 2nd morning) significantly influenced the peptide pattern because of the varying concentrations of urine components; therefore, it is important that collection times be defined exactly for all samples according to the requirements of the clinical study. Compared with our peptide analysis of serum, the urine

7 Clinical Chemistry 53, No. 3, Fig. 6. Recommended sample pretreatment protocol for reproducible MB-based urinary peptidome profiling. peptide pattern is much more stable at room temperature (29); however, a storage time 6 h induced significant changes in peptide patterns and therefore should be avoided. In contrast to the results of Schaub et al. (21), we found differences between fresh and once-frozen samples. The peptide profiles of once-frozen urine samples showed better reproducibility. Therefore, we recommend that urine samples be frozen at 80 C before processing. We demonstrated that bacterial (E. coli) contamination affected the time-dependent alteration of several urine peptide mass signals (M r range, ). Contamination with blood also interfered with the peptide pattern down to a urine blood ratio of 2500:1, a contamination level that produces no change in urine color. It is important, therefore, that each urine sample be checked with a urine test strip for the presence of leukocytes, nitrite, and blood. The use of MB WCX and MB IMAC Cu beads requires adjusting the urine ph to ph 7. In contrast, no ph adjustment is necessary for MB HIC C8 separation because of the strongly acidic fractionation conditions and the hydrophobic interactions between analytes and the beads porous surface structure. A sensitive point in urine proteome profiling is the dependence of the urinary protein concentration on the fluid intake and whether urine samples should be normalized to the protein content. We showed that adjusting the protein concentration by varying the volume of the applied sample minimized the influence of variation in urine protein concentration. Sample preparation for the MBs enables stepwise sample loading on the bead surfaces up to 5 g total protein (the maximum loading capacity). In conclusion, we showed that MB separation in combination with MALDI-TOF MS is a sensitive and reproducible analytic platform for peptidome profiling of human urine. The influence of various preanalytical variables must be taken into account, however. The reliable pretreatment protocol we have presented (Fig. 6) allows standardization of the critical preanalytical phase and facilitates the reproducible peptidome profiling of human urine. This work was supported by a grant from the Sächsische Aufbaubank, by a formel-1 grant of the Medical Faculty of the University Leipzig, and by Grant Th 374/2-3 of the Deutsche Forschungsgemeinschaft. References 1. Aldred S, Grant MM, Griffiths HR. The use of proteomics for the assessment of clinical samples in research. Clin Biochem 2004; 37: Diamandis EP. Mass spectrometry as a diagnostic and a cancer biomarker discovery tool: opportunities and potential limitations. Mol Cell Proteomics 2004;3:

8 428 Fiedler et al.: Magnetic Bead Separation and MALDI-TOF MS of Urine 3. Villanueva J, Philip J, Entenberg D, Chaparro CA, Tanwar MK, Holland EC, et al. Serum peptide profiling by magnetic particleassisted, automated sample processing and MALDI-TOF mass spectrometry. Anal Chem 2004;76: Zhang X, Leung SM, Morris CR, Shigenaga MK. Evaluation of a novel, integrated approach using functionalized magnetic beads, bench-top MALDI-TOF-MS with prestructured sample supports, and pattern recognition software for profiling potential biomarkers in human plasma. J Biomol Tech 2004;15: Hortin GL. The MALDI-TOF mass spectrometric view of the plasma proteome and peptidome. Clin Chem 2006;52: Hortin GL, Jortani SA, Ritchie JC Jr., Valdes R Jr., Chan DW. Proteomics: a new diagnostic frontier. Clin Chem 2006;52: Rogers MA, Clarke P, Noble J, Munro NP, Paul A, Selby PJ, et al. Proteomic profiling of urinary proteins in renal cancer by surface enhanced laser desorption ionization and neural-network analysis: identification of key issues affecting potential clinical utility. Cancer Res 2003;63: Hortin GL, Meilinger B, Drake SK. Size-selective extraction of peptides from urine for mass spectrometric analysis. Clin Chem 2004;50: Thongboonkerd V. Proteomics in nephrology: current status and future directions. Am J Nephrol 2004;24: Haubitz M, Wittke S, Weissinger EM, Walden M, Rupprecht HD, Floege J, et al. Urine protein patterns can serve as diagnostic tools in patients with IgA nephropathy. Kidney Int 2005;67: Jurgens M, Appel A, Heine G, Neitz S, Menzel C, Tammen H, et al. Towards characterization of the human urinary peptidome. Comb Chem High Throughput Screen 2005;8: Thongboonkerd V, Malasit P. Renal and urinary proteomics: current applications and challenges. Proteomics 2005;5: Vidal BC, Bonventre JV, I-Hong Hsu S. Towards the application of proteomics in renal disease diagnosis. Clin Sci (Lond) 2005;109: McKee JA, Kumar S, Ecelbarger CA, Fernandez-Llama P, Terris J, Knepper MA. Detection of Na( ) transporter proteins in urine. J Am Soc Nephrol 2000;11: Marshall T, Williams K. Two-dimensional electrophoresis of human urinary proteins following concentration by dye precipitation. Electrophoresis 1996;17: Sarto C, Marocchi A, Sanchez JC, Giannone D, Frutiger S, Golaz O, et al. Renal cell carcinoma and normal kidney protein expression. Electrophoresis 1997;18: Celis JE, Wolf H, Ostergaard M. Bladder squamous cell carcinoma biomarkers derived from proteomics. Electrophoresis 2000;21: Thongboonkerd V, McLeish KR, Arthur JM, Klein JB. Proteomic analysis of normal human urinary proteins isolated by acetone precipitation or ultracentrifugation. Kidney Int 2002;62: Zerefos PG, Vougas K, Dimitraki P, Kossida S, Petrolekas A, Stravodimos K, et al. Characterization of the human urine proteome by preparative electrophoresis in combination with 2-DE. Proteomics 2006;6: Spahr CS, Davis MT, McGinley MD, Robinson JH, Bures EJ, Beierle J, et al. Towards defining the urinary proteome using liquid chromatography-tandem mass spectrometry. I. Profiling an unfractionated tryptic digest. Proteomics 2001;1: Schaub S, Wilkins J, Weiler T, Sangster K, Rush D, Nickerson P. Urine protein profiling with surface-enhanced laser-desorption/ ionization time-of-flight mass spectrometry. Kidney Int 2004;65: Hortin GL. Can mass spectrometric protein profiling meet desired standards of clinical laboratory practice? Clin Chem 2005;51: Kouri T, Harmoinen A, Laurila K, Ala-Houhala I, Koivula T, Pasternack A. Reference intervals for the markers of proteinuria with a standardised bed-rest collection of urine. Clin Chem Lab Med 2001;39: Kumps A, Duez P, Mardens Y. Metabolic, nutritional, iatrogenic, and artifactual sources of urinary organic acids: a comprehensive table. Clin Chem 2002;48: Thongboonkerd V, Chutipongtanate S, Kanlaya R. Systematic evaluation of sample preparation methods for gel-based human urinary proteomics: quantity, quality, and variability. J Proteome Res 2006;5: Traum AZ, Wells MP, Aivado M, Libermann TA, Ramoni MF, Schachter AD. SELDI-TOF MS of quadruplicate urine and serum samples to evaluate changes related to storage conditions. Proteomics 2006;6: Boguski MS, McIntosh MW. Biomedical informatics for proteomics. Nature 2003;422: White CN, Chan DW, Zhang Z. Bioinformatics strategies for proteomic profiling. Clin Biochem 2004;37: Baumann S, Ceglarek U, Fiedler GM, Lembcke J, Leichtle A, Thiery J. Standardized approach to proteome profiling of human serum based on magnetic bead separation and matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Clin Chem 2005;51: Findeisen P, Sismanidis D, Riedl M, Costina V, Neumaier M. Preanalytical impact of sample handling on proteome profiling experiments with matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Clin Chem 2005;51: Mengel-Jorgensen J, Sanchez JJ, Borsting C, Kirpekar F, Morling N. MALDI-TOF mass spectrometric detection of multiplex single base extended primers. A study of 17 y-chromosome single-nucleotide polymorphisms. Anal Chem 2004;76:

Sven Baumann, Uta Ceglarek, Georg Martin Fiedler, Jan Lembcke, Alexander Leichtle, and Joachim Thiery *

Sven Baumann, Uta Ceglarek, Georg Martin Fiedler, Jan Lembcke, Alexander Leichtle, and Joachim Thiery * Clinical Chemistry 5:6 97 9 (5) Proteomics and Protein Markers Standardized Approach to Proteome Profiling of Human Serum Based on Magnetic Bead Separation and Matrix-Assisted Laser Desorption/Ionization

More information

Using CART to Mine SELDI ProteinChip Data for Biomarkers and Disease Stratification

Using CART to Mine SELDI ProteinChip Data for Biomarkers and Disease Stratification Using CART to Mine SELDI ProteinChip Data for Biomarkers and Disease Stratification Kenna Mawk, D.V.M. Informatics Product Manager Ciphergen Biosystems, Inc. Outline Introduction to ProteinChip Technology

More information

Agilent Protein In-Gel Tryptic Digestion Kit

Agilent Protein In-Gel Tryptic Digestion Kit Agilent 5188-2749 Protein In-Gel Tryptic Digestion Kit Agilent Protein In-Gel Tryptic Digestion Kit Instructions Kit Contents The Protein In-Gel Tryptic Digestion Kit includes sufficient reagents for approximately

More information

Small Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS

Small Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS Bruker Daltonics Application Note # MT-93/FTMS-38 Small Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS Introduction Matrix Assisted Laser Desorption Ionization (MALDI)

More information

TECHNICAL BULLETIN. R 2 GlcNAcβ1 4GlcNAcβ1 Asn

TECHNICAL BULLETIN. R 2 GlcNAcβ1 4GlcNAcβ1 Asn GlycoProfile II Enzymatic In-Solution N-Deglycosylation Kit Product Code PP0201 Storage Temperature 2 8 C TECHNICAL BULLETIN Product Description Glycosylation is one of the most common posttranslational

More information

Proteomics of body liquids as a source for potential methods for medical diagnostics Prof. Dr. Evgeny Nikolaev

Proteomics of body liquids as a source for potential methods for medical diagnostics Prof. Dr. Evgeny Nikolaev Proteomics of body liquids as a source for potential methods for medical diagnostics Prof. Dr. Evgeny Nikolaev Institute for Biochemical Physics, Rus. Acad. Sci., Moscow, Russia. Institute for Energy Problems

More information

In-Gel Tryptic Digestion Kit

In-Gel Tryptic Digestion Kit INSTRUCTIONS In-Gel Tryptic Digestion Kit 3747 N. Meridian Road P.O. Box 117 Rockford, IL 61105 89871 1468.2 Number Description 89871 In-Gel Tryptic Digestion Kit, sufficient reagents for approximately

More information

Tissue and Fluid Proteomics Chao-Cheng (Sam) Wang

Tissue and Fluid Proteomics Chao-Cheng (Sam) Wang Tissue and Fluid Proteomics Chao-Cheng (Sam) Wang UAB-03/09/2004 What is proteomics? A snap shot of the protein pattern!! What can proteomics do? To provide information on functional networks and/or involvement

More information

Chapter 10apter 9. Chapter 10. Summary

Chapter 10apter 9. Chapter 10. Summary Chapter 10apter 9 Chapter 10 The field of proteomics has developed rapidly in recent years. The essence of proteomics is to characterize the behavior of a group of proteins, the system rather than the

More information

Solving practical problems. Maria Kuhtinskaja

Solving practical problems. Maria Kuhtinskaja Solving practical problems Maria Kuhtinskaja What does a mass spectrometer do? It measures mass better than any other technique. It can give information about chemical structures. What are mass measurements

More information

Automated Sample Preparation/Concentration of Biological Samples Prior to Analysis via MALDI-TOF Mass Spectroscopy Application Note 222

Automated Sample Preparation/Concentration of Biological Samples Prior to Analysis via MALDI-TOF Mass Spectroscopy Application Note 222 Automated Sample Preparation/Concentration of Biological Samples Prior to Analysis via MALDI-TOF Mass Spectroscopy Application Note 222 Joan Stevens, Ph.D.; Luke Roenneburg; Tim Hegeman; Kevin Fawcett

More information

Serum Peptidome Profiling Revealed Platelet Factor 4 as a Potential Discriminating Peptide

Serum Peptidome Profiling Revealed Platelet Factor 4 as a Potential Discriminating Peptide 1 Serum Peptidome Profiling Revealed Platelet Factor 4 as a Potential Discriminating Peptide Associated With Pancreatic Cancer Authors Georg Martin Fiedler 1*, Alexander Benedikt Leichtle 1*, Julia Kase

More information

The Detection of Allergens in Food Products with LC-MS

The Detection of Allergens in Food Products with LC-MS The Detection of Allergens in Food Products with LC-MS Something for the future? Jacqueline van der Wielen Scope of Organisation Dutch Food and Consumer Product Safety Authority: Law enforcement Control

More information

The NGAL Turbidimetric Immunoassay Reagent Kit

The NGAL Turbidimetric Immunoassay Reagent Kit Antibody and Immunoassay Services Li Ka Shing Faculty of Medicine, University of Hong Kong The NGAL Turbidimetric Immunoassay Reagent Kit Catalogue number: 51300 For the quantitative determination of NGAL

More information

(III) MALDI instrumentation

(III) MALDI instrumentation Dr. Sanjeeva Srivastava (I) Basics of MALDI-TF (II) Sample preparation In-gel digestion Zip-tip sample clean-up Matrix and sample plating (III) MALDI instrumentation 2 1 (I) Basics of MALDI-TF Analyte

More information

SCIENCE CHINA Life Sciences

SCIENCE CHINA Life Sciences SCIENCE CHINA Life Sciences SPECIAL TOPIC January 2011 Vol.54 No.1: 48 53 RESEARCH PAPERS doi: 10.1007/s11427-010-4119-9 Whole-cell matrix-assisted laser desorption/ionization time-offlight mass spectrometry

More information

Rapid identification and resistance assessment: The future is mass spectrometry

Rapid identification and resistance assessment: The future is mass spectrometry Rapid identification and resistance assessment: The future is mass spectrometry Dr Sanmarié Schlebusch Director of Microbiology Mater Pathology Brisbane Outline Introduction Plug and play Pre-prep and

More information

Trypsin Mass Spectrometry Grade

Trypsin Mass Spectrometry Grade 058PR-03 G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name Trypsin Mass Spectrometry Grade A Chemically Modified, TPCK treated, Affinity Purified

More information

MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels

MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels Jane Y. Yang, David A. Herold Department of Pathology, University of California San Diego, 9500 Gilman

More information

MALDI Sepsityper Kit

MALDI Sepsityper Kit Instructions for Use MALDI Sepsityper Kit Kit for identification of microorganisms from positive blood cultures using the MALDI Biotyper system CARE products are designed to support our worldwide customers

More information

Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform

Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform Abstract Targeted proteomics for biomarker verification/validation

More information

Research Article Serum Peptidome Patterns of Colorectal Cancer Based on Magnetic Bead Separation and MALDI-TOF Mass Spectrometry Analysis

Research Article Serum Peptidome Patterns of Colorectal Cancer Based on Magnetic Bead Separation and MALDI-TOF Mass Spectrometry Analysis Journal of Biomedicine and Biotechnology Volume 2012, Article ID 985020, 8 pages doi:10.1155/2012/985020 Research Article Serum Peptidome Patterns of Colorectal Cancer Based on Magnetic Bead Separation

More information

Bruker Daltonics. autoflex III smartbeam. The Standard in MALDI-TOF Performance MALDI-TOF/TOF. think forward

Bruker Daltonics. autoflex III smartbeam. The Standard in MALDI-TOF Performance MALDI-TOF/TOF. think forward Bruker Daltonics autoflex III smartbeam The Standard in MALDI-TOF Performance think forward MALDI-TOF/TOF Designed for a Routine High Level of Performance The autoflex III smartbeam brings the power of

More information

Systematic analysis of protein-detergent complexes applying dynamic light scattering to optimize solutions for crystallization trials

Systematic analysis of protein-detergent complexes applying dynamic light scattering to optimize solutions for crystallization trials Supporting information 1 2 3 Volume 71 (2015) Supporting information for article: 4 5 6 7 8 Systematic analysis of protein-detergent complexes applying dynamic light scattering to optimize solutions for

More information

RayBio DPP4 Inhibitor Screening Kit

RayBio DPP4 Inhibitor Screening Kit RayBio DPP4 Inhibitor Screening Kit User Manual Version 1.0 Mar 25, 2013 RayBio DPP4 Inhibitor Screening (Cat#: 68SR-DPP4-S100) RayBiotech, Inc. We Provide You With Excellent Support And Service Tel:(Toll

More information

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction Clin Chem Lab Med 2015; 53(11): 1737 1743 Beryl E. Jacobson, David W. Seccombe*, Alex Katayev and Adeera Levin A study examining the bias of albumin and albumin/creatinine ratio measurements in urine DOI

More information

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Cell culture

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Cell culture Version 1 Last updated 14 March 2018 ab228564 Exosome Isolation and Analysis Kit - Flow Cytometry, Cell culture For the measurement of human exosomes in cell culture. This product is for research use only

More information

PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX

PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX E. Claude 1, E. Emekwue 2, M. Snel 1, T. McKenna 1 and J. Langridge 1 1: Waters Corporation, Manchester, UK 2:

More information

MALDI-TOF Mass Spectrometry: A New Rapid ID Method in Clinical Microbiology

MALDI-TOF Mass Spectrometry: A New Rapid ID Method in Clinical Microbiology MALDI-TOF Mass Spectrometry: A New Rapid ID Method in Clinical Microbiology Patrick R. Murray, PhD WW Director, Scientific Affairs BD Diagnostic Systems Outline MALDI-TOF is the most important innovation

More information

Procine sphingomyelin ELISA Kit

Procine sphingomyelin ELISA Kit Procine sphingomyelin ELISA Kit For the quantitative in vitro determination of Procine sphingomyelin concentrations in serum - plasma - celiac fluid - tissue homogenate - body fluid FOR LABORATORY RESEARCH

More information

Comparison of mass spectrometers performances

Comparison of mass spectrometers performances Comparison of mass spectrometers performances Instrument Mass Mass Sensitivity resolution accuracy Quadrupole 1 x 10 3 0.1 Da* 0.5-1.0 pmol DE-MALDI 2 x 10 4 20 ppm 1-10 fmol peptide 1-5 pmol protein Ion

More information

Mass Spectrometry. - Introduction - Ion sources & sample introduction - Mass analyzers - Basics of biomolecule MS - Applications

Mass Spectrometry. - Introduction - Ion sources & sample introduction - Mass analyzers - Basics of biomolecule MS - Applications - Introduction - Ion sources & sample introduction - Mass analyzers - Basics of biomolecule MS - Applications Adapted from Mass Spectrometry in Biotechnology Gary Siuzdak,, Academic Press 1996 1 Introduction

More information

Microalbuminuric Diabetic patients N=18

Microalbuminuric Diabetic patients N=18 ONLINE APPENDIX Table A1 Clinical and laboratory features of healthy subjects, type 2 diabetic patients and NDCKD patients enrolled in the present study. Healthy Subjects N= 2 Normoalbuminuric Diabetic

More information

Supplementary Material (ESI) for Chemical Communications This journal is (c) The Royal Society of Chemistry 2011

Supplementary Material (ESI) for Chemical Communications This journal is (c) The Royal Society of Chemistry 2011 Supporting Information Experimental details 1. Materials Ferric chloride (FeCl 3 6H 2 O), Ethylene glycol (EG) and Ammonium hydroxide (NH 3 H 2 O) were obtained from Beijing Chemical Regent Co. Ltd. (Beijing,

More information

Sequence Identification And Spatial Distribution of Rat Brain Tryptic Peptides Using MALDI Mass Spectrometric Imaging

Sequence Identification And Spatial Distribution of Rat Brain Tryptic Peptides Using MALDI Mass Spectrometric Imaging Sequence Identification And Spatial Distribution of Rat Brain Tryptic Peptides Using MALDI Mass Spectrometric Imaging AB SCIEX MALDI TOF/TOF* Systems Patrick Pribil AB SCIEX, Canada MALDI mass spectrometric

More information

HbA1c (Human) ELISA Kit

HbA1c (Human) ELISA Kit HbA1c (Human) ELISA Kit Cat. No.:DEIA3509 Pkg.Size:96T Intended use GHbA1c (Human) ELISA Kit is a sandwich enzyme immunoassay for the quantitative measurement of human GHbA1c. General Description vhemoglobin,

More information

Supporting Information for: Daniel Knappe, Stefania Piantavigna, Anne Hansen, Adam Mechler, Annegret Binas,

Supporting Information for: Daniel Knappe, Stefania Piantavigna, Anne Hansen, Adam Mechler, Annegret Binas, Supporting Information for: Oncocin (VDKPPYLPRPRPPRRIYNR-NH 2 ): a novel antibacterial peptide optimized against Gram-negative human pathogens Daniel Knappe, Stefania Piantavigna, Anne Hansen, Adam Mechler,

More information

Work-flow: protein sample preparation Precipitation methods Removal of interfering substances Specific examples:

Work-flow: protein sample preparation Precipitation methods Removal of interfering substances Specific examples: Dr. Sanjeeva Srivastava IIT Bombay Work-flow: protein sample preparation Precipitation methods Removal of interfering substances Specific examples: Sample preparation for serum proteome analysis Sample

More information

Urine protein patterns can serve as diagnostic tools in patients with IgA nephropathy

Urine protein patterns can serve as diagnostic tools in patients with IgA nephropathy Kidney International, Vol. 67 (2005), pp. 2313 2320 CLINICAL NEPHROLOGY EPIDEMIOLOGY CLINICAL TRIALS Urine protein patterns can serve as diagnostic tools in patients with IgA nephropathy MARION HAUBITZ,

More information

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Cell Culture (CD63 / CD81)

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Cell Culture (CD63 / CD81) Version 1 Last updated 26 September 2018 ab239682 Exosome Isolation and Analysis Kit - Flow Cytometry, Cell Culture (CD63 / For the isolation and analysis of exosome from cell culture. This product is

More information

Analysis of Triglycerides in Cooking Oils Using MALDI-TOF Mass Spectrometry and Principal Component Analysis

Analysis of Triglycerides in Cooking Oils Using MALDI-TOF Mass Spectrometry and Principal Component Analysis Analysis of Triglycerides in Cooking Oils Using MALDI-TOF Mass Spectrometry and Principal Component Analysis Kevin Cooley Chemistry Supervisor: Kingsley Donkor 1. Abstract Triglycerides are composed of

More information

MALDI-TOF. Introduction. Schematic and Theory of MALDI

MALDI-TOF. Introduction. Schematic and Theory of MALDI MALDI-TOF Proteins and peptides have been characterized by high pressure liquid chromatography (HPLC) or SDS PAGE by generating peptide maps. These peptide maps have been used as fingerprints of protein

More information

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Plasma

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Plasma Version 1 Last updated 25 May 2018 ab228565 Exosome Isolation and Analysis Kit - Flow Cytometry, Plasma For the isolation/detection of exosomes from human plasma, urine or cell culture media. This product

More information

The Adiponectin Turbidimetric Immunoassay Reagent Kit

The Adiponectin Turbidimetric Immunoassay Reagent Kit The Adiponectin Turbidimetric Immunoassay Reagent Kit Catalogue number: 51010 For the quantitative determination of Adiponectin in human serum and plasma This package insert must be read in its entirety

More information

PosterREPRINT A NOVEL APPROACH TO MALDI-TOF-MS SAMPLE PREPARATION. Presented at ABRF 2002, Austin, Texas, USA, 9th - 12th March 2002.

PosterREPRINT A NOVEL APPROACH TO MALDI-TOF-MS SAMPLE PREPARATION. Presented at ABRF 2002, Austin, Texas, USA, 9th - 12th March 2002. Introduction A NOVEL APPROACH TO MALDI-TOF-MS SAMPLE PREPARATION Ed Bouvier 2, Jeff Brown 1, Emmanuelle Claude 1, John L. Gebler 2, Weibin Chen 2, *Dominic Gostick 1, Kevin Howes 1, James Langridge 1,

More information

Double charge of 33kD peak A1 A2 B1 B2 M2+ M/z. ABRF Proteomics Research Group - Qualitative Proteomics Study Identifier Number 14146

Double charge of 33kD peak A1 A2 B1 B2 M2+ M/z. ABRF Proteomics Research Group - Qualitative Proteomics Study Identifier Number 14146 Abstract The 2008 ABRF Proteomics Research Group Study offers participants the chance to participate in an anonymous study to identify qualitative differences between two protein preparations. We used

More information

Mammalian Membrane Protein Extraction Kit

Mammalian Membrane Protein Extraction Kit Mammalian Membrane Protein Extraction Kit Catalog number: AR0155 Boster s Mammalian Membrane Protein Extraction Kit is a simple, rapid and reproducible method to prepare cellular protein fractions highly

More information

Supplementary material: Materials and suppliers

Supplementary material: Materials and suppliers Supplementary material: Materials and suppliers Electrophoresis consumables including tris-glycine, acrylamide, SDS buffer and Coomassie Brilliant Blue G-2 dye (CBB) were purchased from Ameresco (Solon,

More information

Biological Mass Spectrometry. April 30, 2014

Biological Mass Spectrometry. April 30, 2014 Biological Mass Spectrometry April 30, 2014 Mass Spectrometry Has become the method of choice for precise protein and nucleic acid mass determination in a very wide mass range peptide and nucleotide sequencing

More information

Babu Antharavally, Ryan Bomgarden, and John Rogers Thermo Fisher Scientific, Rockford, IL

Babu Antharavally, Ryan Bomgarden, and John Rogers Thermo Fisher Scientific, Rockford, IL A Versatile High-Recovery Method for Removing Detergents from Low-Concentration Protein or Peptide Samples for Mass Spectrometry Sample Preparation and Analysis Babu Antharavally, Ryan Bomgarden, and John

More information

Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase

Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase Clinical Chemistry 43:10 1932 1936 (1997) Automation and Analytical Techniques Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase Yoshitaka Morishita, 1 * Kiyoshi Nakane,

More information

[ Care and Use Manual ]

[ Care and Use Manual ] MALDI Calibration Kit I. Introduction The MALDI Calibration Kit is a conveniently packaged selection of MALDI matrices and calibration standards (includes high-purity Neg Ion Mode Calibrant for accurate

More information

The study of phospholipids in single cells using an integrated microfluidic device

The study of phospholipids in single cells using an integrated microfluidic device Supporting Information: The study of phospholipids in single cells using an integrated microfluidic device combined with matrix-assisted laser desorption/ionization mass spectrometry Weiyi Xie,, Dan Gao,

More information

RESEARCH ARTICLE. Biomarkers Screening Between Preoperative and Postoperative Patients in Pancreatic Cancer

RESEARCH ARTICLE. Biomarkers Screening Between Preoperative and Postoperative Patients in Pancreatic Cancer RESEARCH ARTICLE Biomarkers Screening Between Preoperative and Postoperative Patients in Pancreatic Cancer Pei Li 1&, Juan Yang 2&, Qing-Yong Ma 1, Zheng Wu 1, Chen Huang 2, Xu-Qi Li 1, Zheng Wang 1 *

More information

Improve Protein Analysis with the New, Mass Spectrometry- Compatible ProteasMAX Surfactant

Improve Protein Analysis with the New, Mass Spectrometry- Compatible ProteasMAX Surfactant Improve Protein Analysis with the New, Mass Spectrometry- Compatible Surfactant ABSTRACT Incomplete solubilization and digestion and poor peptide recovery are frequent limitations in protein sample preparation

More information

Supplementary Data. Different volumes of ethanol or calcium solution were slowly added through one of four

Supplementary Data. Different volumes of ethanol or calcium solution were slowly added through one of four Supplementary Data METHODS Liposome preparation Different volumes of ethanol or calcium solution were slowly added through one of four methods: Method I, no ethanol or calcium solution; Method II, exactly

More information

Electronic Supplementary Information

Electronic Supplementary Information Electronic Supplementary Information Microwave-assisted Kochetkov amination followed by permanent charge derivatization: A facile strategy for glycomics Xin Liu a,b, Guisen Zhang* a, Kenneth Chan b and

More information

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS. A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS. Martin R Green*, Keith Richardson, John Chipperfield, Nick Tomczyk, Martin

More information

For Research Use Only Ver

For Research Use Only Ver INSTRUCTION MANUAL Quick-cfDNA Serum & Plasma Kit Catalog No. D4076 Highlights High-quality DNA, including cell-free, is easily and robustly purified from up to 10 ml of serum/plasma, up to 1 ml amniotic

More information

Ionization Methods. Neutral species Charged species. Removal/addition of electron(s) Removal/addition of proton(s)

Ionization Methods. Neutral species Charged species. Removal/addition of electron(s) Removal/addition of proton(s) Ionization Methods Neutral species Charged species Removal/addition of electron(s) M + e - (M +. )* + 2e - electron ionization Removal/addition of proton(s) M + (Matrix)-H MH + + (Matrix) - chemical ionization

More information

Oxalate (urine, plasma)

Oxalate (urine, plasma) Oxalate (urine, plasma) 1 Name and description of analyte 1.1 Name of analyte Oxalate 1.2 Alternative names 1.3 NLMC code To follow 1.4. Function of analyte Oxalate is a metabolic end product primarily

More information

Bovine Insulin ELISA

Bovine Insulin ELISA Bovine Insulin ELISA For quantitative determination of insulin in bovine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSBO-E01 Size: 96 wells Version:

More information

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Rat Insulin ELISA For the quantitative determination of insulin in rat serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: Size: 80-INSRT-E01, E10 96 wells, 10

More information

Peptide sequencing using chemically assisted fragmentation (CAF) and Ettan MALDI-ToF Pro mass spectrometry

Peptide sequencing using chemically assisted fragmentation (CAF) and Ettan MALDI-ToF Pro mass spectrometry application note Ett MALDI-ToF Pro Peptide sequencing using chemically assisted fragmentation (CAF) d Ett MALDI-ToF Pro mass spectrometry Key words: MALDI-ToF, PSD, peptide sequencing, chemically assisted

More information

Fundamentals of Soft Ionization and MS Instrumentation

Fundamentals of Soft Ionization and MS Instrumentation Fundamentals of Soft Ionization and MS Instrumentation Ana Varela Coelho varela@itqb.unl.pt Mass Spectrometry Lab Analytical Services Unit Index Mass spectrometers and its components Ionization methods:

More information

Analysis of Peptides via Capillary HPLC and Fraction Collection Directly onto a MALDI Plate for Off-line Analysis by MALDI-TOF

Analysis of Peptides via Capillary HPLC and Fraction Collection Directly onto a MALDI Plate for Off-line Analysis by MALDI-TOF Analysis of Peptides via Capillary HPLC and Fraction Collection Directly onto a MALDI Plate for Off-line Analysis by MALDI-TOF Application Note 219 Joan Stevens, PhD; Luke Roenneburg; Kevin Fawcett (Gilson,

More information

Human Obestatin ELISA

Human Obestatin ELISA K-ASSAY Human Obestatin ELISA For the quantitative determination of obestatin in human serum and plasma Cat. No. KT-495 For Research Use Only. 1 Rev. 081309 K-ASSAY PRODUCT INFORMATION Human Obestatin

More information

MALDI-IMS (MATRIX-ASSISTED LASER DESORPTION/IONIZATION

MALDI-IMS (MATRIX-ASSISTED LASER DESORPTION/IONIZATION MALDI-IMS (MATRIX-ASSISTED LASER DESORPTION/IONIZATION IMAGING MASS SPECTROMETER) IN TISSUE STUDY YANXIAN CHEN MARCH 8 TH, WEDNESDAY. SEMINAR FOCUSING ON What is MALDI imaging mass spectrometer? How does

More information

Supporting information

Supporting information Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 2014 Supporting information Glycan Reductive Isotope-coded Amino Acid Labeling (GRIAL) for Mass Spectrometry-based

More information

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products) Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products) The target compound to be determined is 2, 4, 5-T. 1. Instrument Liquid Chromatograph-tandem mass spectrometer (LC-MS/MS)

More information

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research Dominic Foley, Michelle Wills, and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION

More information

Introduction to Proteomics 1.0

Introduction to Proteomics 1.0 Introduction to Proteomics 1.0 CMSP Workshop Pratik Jagtap Managing Director, CMSP Objectives Why are we here? For participants: Learn basics of MS-based proteomics Learn what s necessary for success using

More information

Ion Source. Mass Analyzer. Detector. intensity. mass/charge

Ion Source. Mass Analyzer. Detector. intensity. mass/charge Proteomics Informatics Overview of spectrometry (Week 2) Ion Source Analyzer Detector Peptide Fragmentation Ion Source Analyzer 1 Fragmentation Analyzer 2 Detector b y Liquid Chromatography (LC)-MS/MS

More information

BabyBio IMAC columns DATA SHEET DS

BabyBio IMAC columns DATA SHEET DS BabyBio IMAC columns DATA SHEET DS 45 655 010 BabyBio columns for Immobilized Metal Ion Affinity Chromatography (IMAC) are ready-to-use for quick and easy purification of polyhistidine-tagged (His-tagged)

More information

SensoLyte Rh110 Cathepsin K Assay Kit *Fluorimetric* Revision#1.2 Last Updated: May 2017 Catalog # Kit Size

SensoLyte Rh110 Cathepsin K Assay Kit *Fluorimetric* Revision#1.2 Last Updated: May 2017 Catalog # Kit Size SensoLyte Rh110 Cathepsin K Assay Kit *Fluorimetric* Revision#1.2 Last Updated: May 2017 Catalog # 72152 Kit Size 100 Assays (96-well plate) Optimized Performance: This kit detects Cathepsin K activity.

More information

A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography

A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography Application Note LCMS-109 A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography Time of Flight (LC-TOF) Mass Spectrometry Introduction Many clinical

More information

Apolipoprotein A-1 ELISA

Apolipoprotein A-1 ELISA Apolipoprotein A-1 ELISA For the quantitative determination of apolipoprotein A1 in serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Please read carefully due to Critical

More information

Supporting information for

Supporting information for Supporting information for Nitrogen and Sulfur Co-doped Carbon Dots-Assisted Laser Desorption/Ionization Time-of-Flight Mass Spectrometry Imaging for Profiling Bisphenol S Distribution in Mouse Tissues

More information

Tunable Hydrophobicity in DNA Micelles Anaya, Milena; Kwak, Minseok; Musser, Andrew J.; Muellen, Klaus; Herrmann, Andreas; Müllen, Klaus

Tunable Hydrophobicity in DNA Micelles Anaya, Milena; Kwak, Minseok; Musser, Andrew J.; Muellen, Klaus; Herrmann, Andreas; Müllen, Klaus University of Groningen Tunable Hydrophobicity in DNA Micelles Anaya, Milena; Kwak, Minseok; Musser, Andrew J.; Muellen, Klaus; Herrmann, Andreas; Müllen, Klaus Published in: Chemistry DOI: 10.1002/chem.201001816

More information

DetergentOUT Tween. DetergentOUT GBS10. OrgoSol DetergentOUT

DetergentOUT Tween. DetergentOUT GBS10. OrgoSol DetergentOUT 252PR 01 G-Biosciences, St Louis, MO. USA 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name DetergentOUT Detergent Removal Systems For the Removal of Detergents

More information

Determination of Benzodiazepines in Urine by CE-MS/MS

Determination of Benzodiazepines in Urine by CE-MS/MS Determination of Benzodiazepines in Urine by CE-MS/MS Application ote Forensic Toxicology Authors audimir Lucio do Lago Department of Fundamental Chemistry, Institute of Chemistry University of São Paulo,

More information

DetergentOUT Detergent Removal Systems

DetergentOUT Detergent Removal Systems 252PR-04 G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name DetergentOUT Detergent Removal Systems For the Removal of Detergents from Peptide

More information

Detection of Cotinine and 3- hydroxycotine in Smokers Urine

Detection of Cotinine and 3- hydroxycotine in Smokers Urine Detection of Cotinine and 3- hydroxycotine in Smokers Urine Behavioural and Situational Research Group School of Medicine, University of Tasmania Version number: 2 Effective date: 01/12/2015 Review due:

More information

Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification

Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification 22 SEPARATION SCIENCE REDEFINED MAY 2005 Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification In the drug discovery process the detection and

More information

Galactose and Lactose Assay Kit

Galactose and Lactose Assay Kit Galactose and Lactose Assay Kit Catalog Number KA0842 100 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials

More information

Porcine/Canine Insulin ELISA

Porcine/Canine Insulin ELISA Porcine/Canine Insulin ELISA For the quantitative determination of insulin in porcine or canine serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research

More information

Application Note # FTMS-46 solarix XR: Analysis of Complex Mixtures

Application Note # FTMS-46 solarix XR: Analysis of Complex Mixtures Application Note # FTMS-46 solarix XR: Analysis of Complex Mixtures Introduction Natural organic matter (NOM) as well as crude oil and crude oil fractions are very complex mixtures of organic compounds

More information

Insulin ELISA. For the quantitative determination of insulin in serum and plasma.

Insulin ELISA. For the quantitative determination of insulin in serum and plasma. Insulin ELISA For the quantitative determination of insulin in serum and plasma. For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United

More information

Measuring Lipid Composition LC-MS/MS

Measuring Lipid Composition LC-MS/MS Project: Measuring Lipid Composition LC-MS/MS Verification of expected lipid composition in nanomedical controlled release systems by liquid chromatography tandem mass spectrometry AUTHORED BY: DATE: Sven

More information

Characterization of Noncovalent Complexes of Polyelectrolytes by Mass Spectrometry

Characterization of Noncovalent Complexes of Polyelectrolytes by Mass Spectrometry The University of Akron IdeaExchange@UAkron Honors Research Projects The Dr. Gary B. and Pamela S. Williams Honors College Spring 2015 Characterization of Noncovalent Complexes of Polyelectrolytes by Mass

More information

DetergentOUT GBS10 Spin Plates

DetergentOUT GBS10 Spin Plates G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name DetergentOUT GBS10 Spin Plates 96-Well Plates for the Removal of Detergents from Peptide

More information

Human TSH ELISA Kit. User Manual

Human TSH ELISA Kit. User Manual Human TSH ELISA Kit User Manual Catalog number: GTX15585 GeneTex Table of Contents A. Product Description... 2 B. Kit Components... 3 C. Additional Required Materials (not included)... 3 D. Reagent Preparation...

More information

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids Jeremy Netto, 1 Stephen Wong, 1 Federico Torta, 2 Pradeep Narayanaswamy, 2

More information

Brooke Dilmetz. Quantitation of PFAS using MALDI-TOF MS

Brooke Dilmetz. Quantitation of PFAS using MALDI-TOF MS Brooke Dilmetz Quantitation of PFAS using MALDI-TOF MS Per and polyfluoroalkyl substances PFOS and PFOA manufactured from early 1950 s present 1 Highly persistent and bioaccumulative PFOS (Perfluorooctanesulfonate)

More information

Mouse C-peptide EIA. Cat. No. YII-YK013-EX FOR LABORATORY USE ONLY

Mouse C-peptide EIA. Cat. No. YII-YK013-EX FOR LABORATORY USE ONLY Mouse C-peptide EIA Cat. No. YII-YK013-EX FOR LABORATORY USE ONLY TOYO 2CHOME, KOTO-KU, TOKYO, 135-0016, JAPAN http://www.cosmobio.co.jp e-mail : export@cosmobio.co.jp Phone : +81-3-5632-9617 FAX : +81-3-5632-9618

More information

Blood Urea Nitrogen Detection Kit

Blood Urea Nitrogen Detection Kit Discovery through partnership Excellence through quality StressXpress Blood Urea Nitrogen Detection Kit Catalog# SKT-213 (2 Plate kit) Quantitative colorimetric measurement of urea nitrogen TABLE OF CONTENTS

More information

Nori Rabbit IL-2 ELISA Kit DataSheet

Nori Rabbit IL-2 ELISA Kit DataSheet Nori Rabbit IL-2 ELISA Kit DataSheet IL-2 is an interleukin, a type of cytokine immune system signaling molecule. IL-2 is a T cell stimulatory cytokine best known for inducing T cell proliferation, the

More information

UREA CE Insert. 01 English - Ref.: 27. Ref.:27

UREA CE Insert. 01 English - Ref.: 27. Ref.:27 UREA CE Insert Ref.:27 Intended use. Colorimetric and enzymatic system for urea determination in samples of blood and urine, by end point reaction. Professional use. [Only for in vitro diagnostic use.]

More information

MagCapture Exosome Isolation Kit PS Q&A

MagCapture Exosome Isolation Kit PS Q&A MagCapture Exosome Isolation Kit PS Q&A Specifications and performance P.1 Comparison of the conventional method P.2 Operation methods and composition P.4 Amount of starting sample P.5 Analysis after exosomes

More information

Item Catalog Number Manufacturer 1,4-Dithioerythritol (1 g) D9680 Sigma-Aldrich

Item Catalog Number Manufacturer 1,4-Dithioerythritol (1 g) D9680 Sigma-Aldrich SOP: Nuclei isolation from fresh mouse tissues and DNaseI treatment Date modified: 01/12/2011 Modified by: E. Giste/ T. Canfield (UW) The following protocol describes the isolation of nuclei and subsequent

More information