Analysis of Peptides via Capillary HPLC and Fraction Collection Directly onto a MALDI Plate for Off-line Analysis by MALDI-TOF
|
|
- Alexis Hubbard
- 5 years ago
- Views:
Transcription
1 Analysis of Peptides via Capillary HPLC and Fraction Collection Directly onto a MALDI Plate for Off-line Analysis by MALDI-TOF Application Note 219 Joan Stevens, PhD; Luke Roenneburg; Kevin Fawcett (Gilson, Inc.) Introduction Off-line coupling of capillary HPLC to mass spectrometry has significant advantages to current on-line methodology. Both the HPLC and MS parameters can be individually optimized and off-line coupling provides a means of capturing the HPLC run in time. All proteins and peptides that are separated through the capillary HPLC system are captured on a MALDI target and are analyzed through the MALDI-TOF MS at a convenient time. MALDI-TOF MS provides a means of freezing the sample in time, which allows re-analyzing of the collected fractions from a chromatographed sample as many times as are needed. Hence, fraction collecting the entire sample to a MALDI plate ensures that all information from the chromatographic analysis is retained (i.e., nothing is lost). The MALDILC System described below provides a solution to both micro and capillary HPLC. The detector output from the HPLC is coupled with on-line matrix addition through a micro Tee for direct fraction collection onto a Bruker MALDI plate. The column eluent, after passing through the detector, is mixed with matrix in the micro Tee and deposited onto the MALDI plate. The flow rate of the matrix addition is user-selectable, allowing for optimization of the sample/matrix ratio. November 2003 Page 1 of
2 Matrix-Assisted Laser Desorption Ionization (MALDI) MALDI is a method that allows for the vaporization and ionization of non-volatile biological samples from a solid-state phase directly into the gas phase. The sample (analyte) is suspended or dissolved in a matrix (usually in a 1000x molar excess). Matrices are small organic compounds that are co-crystallized with the analyte. The presence seems to spare the analyte from degradation, resulting in the detection of intact molecules as large as 1 million Da. A laser beam serves as the desorption and ionization source. The matrix absorbs the laser light energy, causing part of the illuminated substrate to vaporize. The matrix plume carries the analyte into the vacuum, which aids the sample ionization process. The matrix molecules absorb most of the incident laser energy, minimizing sample damage and ion fragmentation. Once the sample molecules are vaporized and ionized, they are transferred electrostatically into a time-of-flight mass spectrometer (TOF-MS) where they are separated from the matrix ions. The molecules are then individually detected, based on their mass-to-charge (m/z) ratios, and analyzed. Materials & Methods Instruments and Accessories Gilson 350 Micro Pumps for binary gradient Nano Mixer, equipped with: 360-µm OD tubing Gilson 235 Autoinjector, equipped with: 100-µL syringe and rack for 12 x 32-mm vials Gilson 155 UV/VIS Dual-wavelength Detector, equipped with: capillary 35 nl, 8-mm flow cell Gilson 223 Fraction Collector, equipped with: 56-mm arm, spring-loaded probe (1/32 OD x ID x 90 µm PEEK) with Upchurch MicroTee, and rack holder for Bruker AnchorChip 384-well MALDI plate Gilson 402 Syringe Pump, equipped with: 250-µL syringe and matrix valve with matrix/wash solution Discovery BioWidePore C-18 capillary column, 15 cm x 0.32 mm x 5 µ UniPoint System Software version 5.1, 506C Interface, Pentium 4, >2 GHz, 512 MB RAM, 80 GB Hard drive Bruker MS MALDI-TOF XACO, 4.04 acquisition software, Xmass 5.1 data processing software and AutoXecute 5.0 Description of the Procedure: Application of the Sample/Matrix to the MALDI Plate A stock solution of α-cyano-4-hydroxycinnamic acid (matrix) was prepared by dissolving 5.0 mg/ml in 60:40 water:acn An external 402 Syringe Pump pumps the matrix solution at a constant flow through the matrix valve into the Upchurch MicroTee (the rate is adjustable) An Upchurch MicroTee was used to mix the eluent from the detector with the matrix solution (at the very top of the spring-loaded probe) The 223 Fraction Collector with spring-loaded probe was used to spot a mixture of the matrix and eluent on the Bruker AnchorChip 384-well plate and allowed to dry Stock solution of peptides was used for the testing of the HPLC and MALDI FC, containing Val-Tyr-Val, methionine enkephalin, leucine enkephalin, angiotensin II (Simga H-2016 HPLC Peptide Standard Mixture) The total concentration of the peptide mixture in water 0.05% TFA was 220 ng/µl November 2003 Page 2 of
3 Photo 1: Gilson 223 Fraction Collector Collection of the fractions directly on the MALDI plate is accomplished with a spring-loaded probe on the 223 (1/32 OD x ID x 90 µm capillary PEEK tubing). Photo 2: Collected Fractions on the MALDI Plate Bruker AnchorChip MALDI plate after fraction collection directly onto the plate. Two different fraction volumes were collected onto the plate: 750 nl was spotted for the first 10+ rows, 1.5 µl was spotted for the last 5 rows. The plate was allowed to dry and then was evaluated by a Bruker MALDI-TOF XACO. November 2003 Page 3 of
4 Figure 1: Spring-loaded Probe and Mixing Tee The output flow rate from the detector is introduced into the Upchurch mixing tee at the top of the spring-loaded probe. The matrix solution is introduced at a constant flow via a 402 Syringe Pump to the mixing tee from the side. The ratio of mobile phase to matrix solution is an adjustable parameter within the software. The length of tubing is optimized between the detector and the spring-loaded probe to minimize band broadening; there is a 3.0-µL delay from the output of the UV detector to the tip of the probe. Figure 2: Examples of Band Broadening of Peaks Directly Related to Delay Volume The peak width (PW) is doubled through 4 ft. of 100 µm ID tubing with a delay of 40 µl. November 2003 Page 4 of
5 Figure 3: Band Broadening of Peaks Through 2 ft. and 16 in. (respectively) of 100-µm ID Tubing The peak overlays show the amount of peak spreading through 1) 2 ft. of 100-µm ID tubing with a delay volume of 5 µl, and 2) 16 in. of 100-µm ID tubing with a delay volume of 3.0 µl (which is the same tubing delay volume on the MALDI injection collection system presented in this application, hence the peak shape is maintained). Control and Data Handling The MALDILC System was controlled with UniPoint System Software version 5.1. The collection of fractions directly onto the MALDI plate was defined by the control method in the software. The rate at which the MALDI matrix was dispensed is a user-defined parameter. Fractions were collected by time, and multiple collection windows within a run is an available option. Figure 4: UniPoint Method for the MALDILC System The table represents the method used to accomplish MALDI fraction collection and spotting using the Gilson MALDILC System. November 2003 Page 5 of
6 Figure 5: MALDI Tray File The tray represents the layout for the Bruker MALDI plate. Wicking positions are available either on or off the plate. The wicking positions are used as a location for the probe to go to in between collection windows. Results Figure 6: Operations List The Operations List represents the flexibility of the MALDILC System. Rates associated with the matrix can be user defined. Time-per-fraction spot can also be optimized for the samples and can vary within a series of samples. Automatic priming of the matrix pump and cleaning of the matrix valve and pump are available options and are automated through the software. Figure 8: Fractions Collected onto a MALDI Plate Graphical depiction of the positions where fractions were collected onto the MALDI plate. Figure 7: Chromatogram of the MALDI Standards The chromatogram is a 1-μL injection (internal loop) of the 220 ng/µl peptide mix (Val-Try-Val, methionine enkephalin, angiotensin II and leucine enkephalin); approximately 55 ng/µl of each peptide. November 2003 Page 6 of
7 Figures 8 10: Resulting Spectra from the Collected Fractions The MS spectra in figures 8 through 10 are from the MALDI plate that was spotted with the collected fractions. A spectrum was not obtained for Val-Try-Val (MH+ = 379) because of background interference within the MS. November 2003 Page 7 of
8 Summary The 233 Fraction Collector equipped with spring-loaded probe allows direct spotting of fractions on the Bruker AnchorChip 384-well plate; its smaller footprint offers a solution for fraction collection directly onto MALDI plates at a fraction of the cost of larger MALDI spotters The 350 Micro Pumps offer a splitless flow with a flow rate range from 300 nl/min to 50 µl/min UniPoint System Software offers control over the instruments, while still allowing the flexibility to customize the flow rate, rate of matrix addition, and time collection per spot to accommodate the needs of the researcher Conclusion Gilson s MALDILC System offers researchers a total solution for the analysis of proteins/peptides via capillary HPLC Splitless Flow and direct collection of fractions onto a Bruker AnchorChip 384-well MALDI plate. Implementing direct collection of fractions on a MALDI plate eliminates the requirement of manual and/or automated formatting of the MALDI plate prior to analysis. Using an innovative spring-loaded probe (internal peek tubing) eliminates the possibility of damaging the surface or substrate of the MALDI plate. Gilson s 223 Fraction Collector allows incrementing from spot to spot in as little as 0.03 minutes (1.8 seconds per collection spot); the rapid arm movement (<25 cm/sec) of the 223 Fraction Collector is well suited for fraction collection onto a MALDI plate. The springloaded probe demonstrated a repeatability of ±0.25 mm in the X/Y/Z dimensions. The MALDILC System represents Gilson s commitment to offer a complete solution for researchers at a fraction of the investment of other comparable systems. Gilson, Inc. World Headquarters 3000 W. Beltline Hwy., P.O. Box , Middleton, WI USA Telephone: or Fax: Gilson S.A.S. 19, avenue des Entrepreneurs, BP 145, F VILLIERS LE BEL France sales@gilson.com, service@gilson.com, training@gilson.com November 2003 Page 8 of
Automated Sample Preparation/Concentration of Biological Samples Prior to Analysis via MALDI-TOF Mass Spectroscopy Application Note 222
Automated Sample Preparation/Concentration of Biological Samples Prior to Analysis via MALDI-TOF Mass Spectroscopy Application Note 222 Joan Stevens, Ph.D.; Luke Roenneburg; Tim Hegeman; Kevin Fawcett
More informationThe Ephedra Story: An Automated SPE Procedure and Analysis for the Determination of Ephedra and Ephedra-like Compounds in Dietary Supplements
The Ephedra Story: An Automated SPE Procedure and Analysis for the Determination of Ephedra and Ephedra-like Compounds in Dietary Supplements Application Note 215 Joan Stevens, PhD, Kirby Reed (Gilson,
More informationBiological Mass Spectrometry. April 30, 2014
Biological Mass Spectrometry April 30, 2014 Mass Spectrometry Has become the method of choice for precise protein and nucleic acid mass determination in a very wide mass range peptide and nucleotide sequencing
More informationSolving practical problems. Maria Kuhtinskaja
Solving practical problems Maria Kuhtinskaja What does a mass spectrometer do? It measures mass better than any other technique. It can give information about chemical structures. What are mass measurements
More informationMatrix Assisted Laser Desorption Ionization Time-of-flight Mass Spectrometry
Matrix Assisted Laser Desorption Ionization Time-of-flight Mass Spectrometry Time-of-Flight Mass Spectrometry. Basic principles An attractive feature of the time-of-flight (TOF) mass spectrometer is its
More informationAutomated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System
Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System Keywords Introduction Application Note PHA0413 High Pressure Liquid Chromatography
More informationPTM Discovery Method for Automated Identification and Sequencing of Phosphopeptides Using the Q TRAP LC/MS/MS System
Application Note LC/MS PTM Discovery Method for Automated Identification and Sequencing of Phosphopeptides Using the Q TRAP LC/MS/MS System Purpose This application note describes an automated workflow
More informationWolf Laboratories Limited
Wolf Laboratories Limited www.wolflabs.co.uk Tel: 01759 301142 Fax:01759 301143 sales@wolflabs.co.uk Use the above details to contact us if this literature doesn't answer all your questions. Pricing on
More informationLOCALISATION, IDENTIFICATION AND SEPARATION OF MOLECULES. Gilles Frache Materials Characterization Day October 14 th 2016
LOCALISATION, IDENTIFICATION AND SEPARATION OF MOLECULES Gilles Frache Materials Characterization Day October 14 th 2016 1 MOLECULAR ANALYSES Which focus? LOCALIZATION of molecules by Mass Spectrometry
More informationIonization Methods. Neutral species Charged species. Removal/addition of electron(s) Removal/addition of proton(s)
Ionization Methods Neutral species Charged species Removal/addition of electron(s) M + e - (M +. )* + 2e - electron ionization Removal/addition of proton(s) M + (Matrix)-H MH + + (Matrix) - chemical ionization
More information2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry
Dr. Sanjeeva Srivastava 1. Fundamental of Mass Spectrometry Role of MS and basic concepts 2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry 2 1 MS basic concepts Mass spectrometry - technique
More informationOperating Instructions
Sequazyme C-Peptide Sequencing Kit Operating Instructions 1 Product Description The Sequazyme C-Peptide Sequencing Kit enables peptide digestion and the analysis of sequentially truncated peptides using
More informationOptimizing Chiral Separations and Purification of Enantiomers Through Selective Sample Slicing and Column Switching
Optimizing Chiral Separations and Purification of Enantiomers Through Selective Sample Slicing and Column Switching Application Note PHA0113 Keywords Enantiomers, Purification, Chiral Pharmaceuticals,
More informationMALDI-TOF. Introduction. Schematic and Theory of MALDI
MALDI-TOF Proteins and peptides have been characterized by high pressure liquid chromatography (HPLC) or SDS PAGE by generating peptide maps. These peptide maps have been used as fingerprints of protein
More informationSupporting information for: Memory Efficient. Principal Component Analysis for the Dimensionality. Reduction of Large Mass Spectrometry Imaging
Supporting information for: Memory Efficient Principal Component Analysis for the Dimensionality Reduction of Large Mass Spectrometry Imaging Datasets Alan M. Race,,, Rory T. Steven,, Andrew D. Palmer,,,
More informationCAMAG TLC-MS INTERFACE
CAMAG TLC-MS INTERFACE 93.1 249.2 40 30 97.1 20 10 250.2 0 200 400 m/z WORLD LEADER IN PLANAR-CHROMATOGRAPHY Identification and elucidation of unknown substances by hyphenation of TLC / HPTLC and MS The
More information2D-LC as an Automated Desalting Tool for MSD Analysis
2D-LC as an Automated Desalting Tool for MSD Analysis Direct Mass Selective Detection of a Pharmaceutical Peptide from an MS-Incompatible USP Method Application Note Biologics and Biosimilars Author Sonja
More informationData Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section
Data Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section Emmanuelle Claude, Mark Towers, and Kieran Neeson Waters Corporation, Manchester,
More informationMass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System
Application Note Food Testing and Agriculture Mass-Based Purification of Natural Product Impurities Using an Agilent 126 Infinity II Preparative LC/MSD System Authors Florian Rieck and Jörg Hippler Agilent
More informationIon Source. Mass Analyzer. Detector. intensity. mass/charge
Proteomics Informatics Overview of spectrometry (Week 2) Ion Source Analyzer Detector Peptide Fragmentation Ion Source Analyzer 1 Fragmentation Analyzer 2 Detector b y Liquid Chromatography (LC)-MS/MS
More informationMass Spectrometry. Mass spectrometer MALDI-TOF ESI/MS/MS. Basic components. Ionization source Mass analyzer Detector
Mass Spectrometry MALDI-TOF ESI/MS/MS Mass spectrometer Basic components Ionization source Mass analyzer Detector 1 Principles of Mass Spectrometry Proteins are separated by mass to charge ratio (limit
More information[ Care and Use Manual ]
MALDI Calibration Kit I. Introduction The MALDI Calibration Kit is a conveniently packaged selection of MALDI matrices and calibration standards (includes high-purity Neg Ion Mode Calibrant for accurate
More informationFundamentals of Soft Ionization and MS Instrumentation
Fundamentals of Soft Ionization and MS Instrumentation Ana Varela Coelho varela@itqb.unl.pt Mass Spectrometry Lab Analytical Services Unit Index Mass spectrometers and its components Ionization methods:
More informationQuickSplitTM Flow Splitters
QuickTM Flow ters between 2 or more detectors such as Mass Spectroscopy and evaporative light scattering, UV, or IR. This is especially important when one of the detectors is destructive, such as Mass
More informationMass Spectrometry. Actual Instrumentation
Mass Spectrometry Actual Instrumentation August 2017 See also http://www.uni-bielefeld.de/chemie/analytik/ms f additional infmation 1. MALDI TOF MASS SPECTROMETRY ON THE ULTRAFLEX 2 2. ESI MASS SPECTROMETRY
More informationMALDI-TOF Mass Spectrometry: A New Rapid ID Method in Clinical Microbiology
MALDI-TOF Mass Spectrometry: A New Rapid ID Method in Clinical Microbiology Patrick R. Murray, PhD WW Director, Scientific Affairs BD Diagnostic Systems Outline MALDI-TOF is the most important innovation
More informationHigh Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column
High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column Application Note BioPharma Authors James Martosella, Phu Duong, and Alex Zhu Agilent Technologies, Inc. Abstract
More informationMALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels
MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels Jane Y. Yang, David A. Herold Department of Pathology, University of California San Diego, 9500 Gilman
More informationAgilent MALDI Protein Identification Solution
Agilent MALDI Protein Identification Solution MS Application Guide Agilent Technologies Notices Agilent Technologies, Inc. 2005 No part of this manual may be reproduced in any form or by any means (including
More informationSUPPORTING INFORMATION:
1S SUPPORTING INFORMATION: The discovery of the once weekly glucagon like peptide 1 (GLP-1) analog semaglutide Jesper Lau*, Paw Bloch, Lauge Schäffer, Ingrid Pettersson, Jane Spetzler, Jacob Kofoed, Kjeld
More informationMethods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University
Methods in Mass Spectrometry Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University Sample Engineering Chemistry Biology Life Science Medicine Industry IDF / Police Sample
More informationApplication of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices
Application ote #LCMS-2 esquire series Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices Introduction The simple monitoring
More informationCharacterization of Noncovalent Complexes of Polyelectrolytes by Mass Spectrometry
The University of Akron IdeaExchange@UAkron Honors Research Projects The Dr. Gary B. and Pamela S. Williams Honors College Spring 2015 Characterization of Noncovalent Complexes of Polyelectrolytes by Mass
More informationIntroduction to the Oligo HTCS Systems. Novatia, LLC
Introduction to the Oligo HTCS Systems Novatia, LLC What is the Oligo HTCS and what can it do? Automated ESI/MS system that can analyze up to 14 oligonucleotide samples/day (>3/day with twin autosampler)
More informationJohn Haselden 1, Gordon Dear 1, Jennifer H. Granger 2, and Robert S. Plumb 2. 1GlaxoSmithKline, Ware, UK; 2 Waters Corporation, Milford, MA, USA
Challenges of Polar Compound Analysis Routine LC/MS-based analyses usually rely on using reversed-phase column chemistries (e.g. C 18, C 8, C 4 ) in their LC methodology. However, one of the shortcomings
More informationAgilent 7700x ICP-MS
Agilent 7700x ICP-MS Metals analysis Mid ppt to low ppb detection range Option to detect in four different gas modes to gain optimum detection of multiple metals or interferences in a sample Autosampler
More information[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS
Yun Wang Alelyunas, Henry Shion, Mark Wrona Waters Corporation, Milford, MA, USA APPLICATION BENEFITS mab LC-MS method which enables users to achieve highly sensitive bioanalysis of intact trastuzumab
More informationSmall Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS
Bruker Daltonics Application Note # MT-93/FTMS-38 Small Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS Introduction Matrix Assisted Laser Desorption Ionization (MALDI)
More informationO O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions
Simplifying Qual/Quan Analysis in Discovery DMPK using UPLC and Xevo TQ MS Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION The determination of the drug metabolism
More informationSmall Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application
Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application Pharmaceutical Author Cliff Woodward Agilent Technologies, Inc. 285 Centerville
More informationUltra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification
22 SEPARATION SCIENCE REDEFINED MAY 2005 Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification In the drug discovery process the detection and
More information1. Sample Introduction to MS Systems:
MS Overview: 9.10.08 1. Sample Introduction to MS Systems:...2 1.1. Chromatography Interfaces:...3 1.2. Electron impact: Used mainly in Protein MS hard ionization source...4 1.3. Electrospray Ioniztion:
More informationMS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine
Application Note: 346 MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine Gargi Choudhary and Diane Cho, Thermo Fisher Scientific, San Jose, CA Wayne Skinner and
More informationRapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column
APPLICATION NOTE 21673 Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column Authors Aaron Lamb and Brian King, Thermo Fisher Scientific, Runcorn, UK
More informationMALDI Sepsityper Kit
Instructions for Use MALDI Sepsityper Kit Kit for identification of microorganisms from positive blood cultures using the MALDI Biotyper system CARE products are designed to support our worldwide customers
More informationShotgun Proteomics MS/MS. Protein Mixture. proteolysis. Peptide Mixture. Time. Abundance. Abundance. m/z. Abundance. m/z 2. Abundance.
Abundance Abundance Abundance Abundance Abundance Shotgun Proteomics Protein Mixture 1 2 3 MS/MS proteolysis m/z 2 3 Time µlc m/z MS 1 m/z Peptide Mixture m/z Block Diagram of a Mass Spectrometer Sample
More informationComponents of a Mass Spectrometer
Components of a Mass Spectrometer Sample Introduction Inlet GC LC Direct Insertion (Syringe/Probe) Ionization Ion Separation Ion Detection Ion Source EI,CI,,, MALDI Mass Analyzer Under vacuum TOF, Quadrupole,
More informationInfinityLab Purification Capillary Kits
InfinityLab Purification Capillary Kits Technical Note This note describes the capillary connections for Agilent preparative systems. Agilent Technologies Contents Abbreviations 3 List of Modules 4 Overview
More informationMALDI Imaging Drug Imaging Detlev Suckau Head of R&D MALDI Bruker Daltonik GmbH. December 19,
MALDI Imaging Drug Imaging Detlev Suckau Head of R&D MALDI Bruker Daltonik GmbH December 19, 2014 1 The principle of MALDI imaging Spatially resolved mass spectra are recorded Each mass signal represents
More informationComparison of mass spectrometers performances
Comparison of mass spectrometers performances Instrument Mass Mass Sensitivity resolution accuracy Quadrupole 1 x 10 3 0.1 Da* 0.5-1.0 pmol DE-MALDI 2 x 10 4 20 ppm 1-10 fmol peptide 1-5 pmol protein Ion
More informationPosterREPRINT A NOVEL APPROACH TO MALDI-TOF-MS SAMPLE PREPARATION. Presented at ABRF 2002, Austin, Texas, USA, 9th - 12th March 2002.
Introduction A NOVEL APPROACH TO MALDI-TOF-MS SAMPLE PREPARATION Ed Bouvier 2, Jeff Brown 1, Emmanuelle Claude 1, John L. Gebler 2, Weibin Chen 2, *Dominic Gostick 1, Kevin Howes 1, James Langridge 1,
More information(III) MALDI instrumentation
Dr. Sanjeeva Srivastava (I) Basics of MALDI-TF (II) Sample preparation In-gel digestion Zip-tip sample clean-up Matrix and sample plating (III) MALDI instrumentation 2 1 (I) Basics of MALDI-TF Analyte
More informationSCS Mass Spectrometry Laboratory
SCS Mass Spectrometry Laboratory Contact Information Staff 31 Noyes Laboratory (8:00-5:00 M-F) 217-333-2545 http://scs.illinois.edu/massspec/ Furong Sun (frs@illinois.edu) Furong Sun Director Training
More informationMass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010
Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010 Introduction, Ionization Methods Mass Analyzers, Ion Activation Methods
More informationTriptycene-Based Small Molecules Modulate (CAG) (CTG) Repeat Junctions
Electronic Supplementary Material (ESI) for Chemical Science. This journal is The Royal Society of Chemistry 2015 Triptycene-Based Small Molecules Modulate (CAG) (CTG) Repeat Junctions Stephanie A. Barros
More informationLipidomic Analysis by UPLC-QTOF MS
Lipidomic Analysis by UPLC-QTOF MS Version: 1 Edited by: Oliver Fiehn Summary Reagents and Materials Protocol Summary:Lipidomic analysis by UPLC-QTOF mass spectrometry Reagents and Materials: Reagent/Material
More informationEdgar Naegele. Abstract
Simultaneous determination of metabolic stability and identification of buspirone metabolites using multiple column fast LC/TOF mass spectrometry Application ote Edgar aegele Abstract A recent trend in
More informationAdvancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations
Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations For Research Use Only. Not for use in diagnostic procedures. 1 2015 AB Sciex RUO-MKT-11-1018-A For research
More informationImpurity Identification using a Quadrupole - Time of Flight Mass Spectrometer QTOF
Impurity Identification using a Quadrupole - Time of Flight Mass Spectrometer QTOF PUSHER TOF DETECTOR ZSPRAY TM Ion Source SAMPLING CONE SKIMMER RF HEXAPOLE RF HEXAPOLE QUADRUPOLE IN NARROW BANDPASS MODE
More informationA NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.
A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS. Martin R Green*, Keith Richardson, John Chipperfield, Nick Tomczyk, Martin
More informationSequence Identification And Spatial Distribution of Rat Brain Tryptic Peptides Using MALDI Mass Spectrometric Imaging
Sequence Identification And Spatial Distribution of Rat Brain Tryptic Peptides Using MALDI Mass Spectrometric Imaging AB SCIEX MALDI TOF/TOF* Systems Patrick Pribil AB SCIEX, Canada MALDI mass spectrometric
More informationMass spectrometry Technologies in Lipid chemistry
Mass spectrometry Technologies in Lipid chemistry Rabah Soliymani University Of Helsinki Protein Chemistry Unit Biomedicum Helsinki Rabah.soliymani@helsinki.fi Complex_&_dynamic_mixtures (few copies to
More informationAdvances in Hybrid Mass Spectrometry
The world leader in serving science Advances in Hybrid Mass Spectrometry ESAC 2008 Claire Dauly Field Marketing Specialist, Proteomics New hybrids instruments LTQ Orbitrap XL with ETD MALDI LTQ Orbitrap
More informationShuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W
Simple, Fast and Accurate Quantitation of Human Plasma Vitamins and Their Metabolites by Protein Precipitation Combined with Columns Using HPLC-UV, HPLC-FLD or LC/MS/MS Shuguang Li, Jason Anspach, Sky
More informationNew Instruments and Services
New Instruments and Services http://planetorbitrap.com/orbitrap fusion Combining the best of quadrupole, Orbitrap, and ion trap mass analysis in a revolutionary Tribrid architecture, the Orbitrap Fusion
More informationapplication Natural Food Colorants Analysis of Natural Food Colorants by Electrospray and Atmospheric Pressure Chemical Ionization LC/MS
application LC/MS Natural Food Colorants Analysis of Natural Food Colorants by Electrospray and Atmospheric Pressure Chemical Ionization LC/MS Introduction Many kinds of natural colors are used in beverages,
More informationTechnical Note # TN-31 Redefining MALDI-TOF/TOF Performance
Bruker Daltonics Technical Note # TN-31 Redefining MALDI-TOF/TOF Performance The new ultraflextreme exceeds all current expectations of MALDI-TOF/TOF technology: A proprietary khz smartbeam-ii TM MALDI
More informationELECTRONIC PIPETTE. The electronic pipette with guaranteed performance in both standard and repetitive pipetting. PRECISE EASY-TO-USE RELIABLE
ELECTRONIC PIPETTE The electronic pipette with guaranteed performance in both standard and repetitive pipetting. PRECISE EASY-TO-USE RELIABLE DELIVER ON-TARGET PERFORMANCE PIPETMAN M has been designed
More informationMass Directed Flash Chromatography by Biotage. Biotage Dalton Mass Detectors
Mass Directed Flash Chromatography by Biotage Biotage Dalton Mass Detectors Why Do Companies Around the World Choose the Biotage Approach to Flash Purification? As a pioneer of automated Flash purification,
More informationSupporting Information
Supporting Information Cyclic Peptidyl Inhibitors against Human Peptidyl-Prolyl Isomerase Pin1 Tao Liu, Yu Liu, Hung-Ying Kao, *,, and Dehua Pei Table of Contents: Table S1. Structures of amino acid building
More informationNew Solvent Grade Targeted for Trace Analysis by UHPLC-MS
New Solvent Grade Targeted for Trace Analysis by UHPLC-MS Subhra Bhattacharya, Deva H. Puranam, and Stephen C. Roemer Thermo Fisher Scientific Fisher Chemical, One Reagent Lane, Fair Lawn, NJ Material
More informationDoubling the throughput of long chromatographic methods by using a novel Dual LC workflow
APPLICATIN NTE 7601 Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow Authors Sylvia Grosse, Mauro De Pra, Frank Steiner Thermo Fisher Scientific, Germering, Germany
More informationAnalysis of Triglycerides in Cooking Oils Using MALDI-TOF Mass Spectrometry and Principal Component Analysis
Analysis of Triglycerides in Cooking Oils Using MALDI-TOF Mass Spectrometry and Principal Component Analysis Kevin Cooley Chemistry Supervisor: Kingsley Donkor 1. Abstract Triglycerides are composed of
More informationAuto HbA1C ANALYZER HbA1LC HPLC 3660
Turnkey Laboratories Solutions Auto HbA1C ANALYZER HbA1LC HPLC 3660 Viewable Accuracy Real Time Summarized Chromatogram Authentic Results EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES
More informationIdentification of Steroids in Water by Ion Trap LC/MS/MS Application
Identification of Steroids in Water by Ion Trap LC/MS/MS Application Environmental Author Paul Zavitsanos and Christine Miller Agilent Technologies, Inc. Centerville Road Wilmington, DE 9- USA Abstract
More informationPHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX
PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX E. Claude 1, E. Emekwue 2, M. Snel 1, T. McKenna 1 and J. Langridge 1 1: Waters Corporation, Manchester, UK 2:
More informationMALDI-IMS (MATRIX-ASSISTED LASER DESORPTION/IONIZATION
MALDI-IMS (MATRIX-ASSISTED LASER DESORPTION/IONIZATION IMAGING MASS SPECTROMETER) IN TISSUE STUDY YANXIAN CHEN MARCH 8 TH, WEDNESDAY. SEMINAR FOCUSING ON What is MALDI imaging mass spectrometer? How does
More information4-Plex itraq Based Quantitative Proteomic Analysis Using an Agilent Accurate -Mass Q-TOF
4-Plex itraq Based Quantitative Proteomic Analysis Using an Agilent Accurate -Mass Q-TOF Application Note Authors H. C. Harsha, G. S. S. Kumar, and A. Pandey Institute of Bioinformatics Bangalore India
More informationChromatography Vacuum Ultraviolet Spectroscopy
Application Note Differentiation and Determination Differentiation and Determination of Fatty Acid Methyl of Fatty Esters Acid by Gas Methyl Chromatography Esters by Vacuum Gas Ultraviolet Spectroscopy
More informationHiroya Hidaka *1), Masaki Takiwaki 2), Mine Yamashita 2), Shinya Otsuki 1), Kenji Kawasaki 3), Mitsutoshi Sugano 3) and Takayuki Honda 4)
Mild acid hydrolysis of sphingolipids yields lysosphingolipids: a matrix-assisted laser desorption and ionization time-of-flight mass spectrometry study Hiroya Hidaka *1), Masaki Takiwaki 2), Mine Yamashita
More informationApplication of a new capillary HPLC- ICP-MS interface to the identification of selenium-containing proteins in selenized yeast
Application of a new capillary HPLC- ICP-MS interface to the identification of selenium-containing proteins in selenized yeast Application note Food supplements Authors Juliusz Bianga and Joanna Szpunar
More informationMass spectrometry imaging. What does MS imaging offer?
BMG 744 Mass spectrometry imaging Stephen Barnes, PhD With sincere acknowledgments to David Stella, PhD and Kyle A. Floyd, MS, former students in the Barnes Laboratory (2005 2012) and Kevin Schey, PhD,
More informationAnalyzing Trace-Level Impurities of a Pharmaceutical Intermediate Using an LCQ Fleet Ion Trap Mass Spectrometer and the Mass Frontier Software Package
Analyzing Trace-Level Impurities of a Pharmaceutical Intermediate Using an LCQ Fleet Ion Trap Mass Spectrometer and the Mass Frontier Software Package Chromatography and Mass Spectrometry Application Note
More informationProtein Analysis using Electrospray Ionization Mass Spectroscopy *
OpenStax-CNX module: m38341 1 Protein Analysis using Electrospray Ionization Mass Spectroscopy * Wilhelm Kienast Andrew R. Barron This work is produced by OpenStax-CNX and licensed under the Creative Commons
More informationImpact of Chromatography on Lipid Profiling of Liver Tissue Extracts
Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts Application Note Clinical Research Authors Mark Sartain and Theodore Sana Agilent Technologies, Inc. Santa Clara, California, USA Introduction
More informationSingle Quadrupole Compact Mass Spectrometer for Liquid Sample
Single Quadrupole Compact Mass Spectrometer for Liquid Sample YL INSTRUMENT CO., LTD. YL9900 MS quickly provides essential information and improves the chemist's workflow. Mordern organic labs have become
More informationFully Automated Online Sample Preparation AND Quantification of Amiodarone from Whole Blood using CLAM-LC-MS/MS
PO-CON177E Fully Automated Online Sample Preparation AND Quantification of Amiodarone from Whole Blood using CLAM-LC-MS/MS ASMS 17 ThP 93 Christian Bunse 1, Lars Kröner 1 Shimadzu Deutschland GmbH, Duisburg,
More informationRapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit
Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit Application Note General Chromatography, Food Industry Authors John W. Henderson Jr., Judy Berry, Anne Mack, William Long Agilent
More informationAn optical dosimeter for the selective detection of gaseous phosgene with ultra-low detection limit
Supporting information for An optical dosimeter for the selective detection of gaseous phosgene with ultra-low detection limit Alejandro P. Vargas, Francisco Gámez*, Javier Roales, Tània Lopes-Costa and
More informationMass Spectrometry at the Laboratory of Food Chemistry. Edwin Bakx Laboratory of Food Chemistry Wageningen University
Mass Spectrometry at the Wageningen University Mass Spectrometry at the 3 UPLC/CE - ESI - Ion trap MS systems UPLC Thermo Acella with a Velos or VelosPro CE Beckman PA800 with a Thermo VelosPro 1 UPLC-
More informationComprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS
Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS SCIEX QTRAP 4500 LC-MS/MS System and Spark Holland PICO Adrian M. Taylor 1, Peter Ringeling 2, Martin Sibum 2, Stefan Sturm
More informationSupporting Information. 58 Pages. 6 Figures 4 Tables
Light-Source-Dependent Effects of Main Water Constituents on Photodegradation of Phenicol Antibiotics: Mechanism and Kinetics Linke Ge, Jingwen Chen, * Xianliang Qiao, Jing Lin, Xiyun Cai Key Laboratory
More informationBrooke Dilmetz. Quantitation of PFAS using MALDI-TOF MS
Brooke Dilmetz Quantitation of PFAS using MALDI-TOF MS Per and polyfluoroalkyl substances PFOS and PFOA manufactured from early 1950 s present 1 Highly persistent and bioaccumulative PFOS (Perfluorooctanesulfonate)
More informationUPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes
UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes Catalin E. Doneanu, Weibin Chen, and Jeffrey R. Mazzeo Waters Corporation, Milford, MA, U.S. A P P L I C AT ION B E N E F
More informationSystematic analysis of protein-detergent complexes applying dynamic light scattering to optimize solutions for crystallization trials
Supporting information 1 2 3 Volume 71 (2015) Supporting information for article: 4 5 6 7 8 Systematic analysis of protein-detergent complexes applying dynamic light scattering to optimize solutions for
More informationExtended Application Note
Extended Application Note Comparison of 4 μm vs. 2.o Columns in Analysis of Drospirenone/Ethinyl Estradiol Tablet APP A-339 www.mtc-usa.com 1-732-578-1777 INTRODUCTION The formulation of drospirenone and
More informationQuantitation of Protein Phosphorylation Using Multiple Reaction Monitoring
Quantitation of Protein Phosphorylation Using Multiple Reaction Monitoring Application Note Authors Ning Tang, Christine A. Miller and Keith Waddell Agilent Technologies, Inc. Santa Clara, CA USA This
More informationCannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system
Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 9 Infinity II/63B LC/TOF system Application Brief Authors Mike Adams, Karen Kaikaris, and A. Roth CWC Labs Joan Stevens, Sue D
More informationDevelopment of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid
Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid Joanne ( 乔安妮 ) Mather Senior Scientist Waters Corporation Data courtesy of Erin Chambers and Mary
More informationANALYTICAL REPORT 1 3C-P (C14H23NO3) 1-(3,5-dimethoxy-4-propoxylphenyl)propan-2-amine.
Vodovodna 95, 1000 LJUBLJANA, SLOVENIA T: +386 (0)1 428 44 93 E: nfl@policija.si www.policija.si Remark other NPS detected: none ANALYTICAL REPORT 1 3C-P (C14H23NO3) 1-(3,5-dimethoxy-4-propoxylphenyl)propan-2-amine
More information