APPLICATIONS TN Determination of Pesticide Residues in Kale and Grapes using QuEChERS and LC-MS/MS. Introduction
|
|
- Oscar Small
- 5 years ago
- Views:
Transcription
1 APPLICATIONS Determination of Pesticide Residues in Kale and Grapes using QuEChERS and LC-MS/MS Scott Krepich, Matt Trass, and Laura Snow Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA Scott Krepich Senior Field Application Scientist Scott enjoys surfing and eating. He is crazy about chromatography, because his mom is really into CSI and thinks that is what he does. Introduction Kale and grape consumption is increasing globally and is included in the USDA s national pesticide residue monitoring program for produce. They are also commonly listed annually on the Environmental Working Group (EWG) report of the most popular pesticide contaminated produce. The use of a QuEChERS technique followed by LC-MS/MS has proven to be an effective approach for screening multiresidue pesticides from various produce items including kale and grapes. Following the EN pesticide residue method using roq QuEChERS extraction and clean-up dspe kits, along with the use of a highly efficient and broad polarity Kinetex 5 µm Biphenyl Core-Shell HPLC column, a pesticide screening method was developed using LC-MS/MS in under 16 minutes. Over 200 pesticides were screened at concentration ranges between 1 ppm to 1 ppm with the majority of analytes having a recovery range of %. The method produced excellent selectivity and reproducibility for the earlier eluting polar pesticides owing to the use of QuEChERS for removing sample matrix interferences and the biphenyl bonded-phase chemistry of the Kinetex Biphenyl column for the chromatographic separation across the entire range of pesticides. Materials and Methods Materials and Equipment Shimadzu Nexera LC30 pumps and autosampler were used along with a SCIEX 4000 QTRAP mass spectrometer in electrospray ionization (ESI) positive polarity using scheduled MRM mode. SCIEX idquant Standards kits for pesticide analysis were used to prepare calibrators and to spike samples. A Sartorius arium comfort II was used for ASTM Type 1 ultrapure water. All other solvents were purchased from Honeywell. Kale and grapes were purchased from local grocers. Sample Preparation A calibration curve was prepared in purified water at concentrations of 1, 5, 0.1, 0.5 and 1 ppm. Kale and grapes samples were prepared at four concentrations: no spike ( ppm), low spike (0.1 ppm), medium spike (0.5 ppm), and high spike (1 ppm) Procedure: 1. HOMOGENIZE sample using a blender or similar apparatus. 2. WEIGH 10 g of homogenized sample into a clean 50 ml tube (provided in roq Extraction Kits). 3. ADD 10 ml of Acetonitrile containing internal standard. 4. DISPENSE contents of the included QuEChERS salt packet into the 50 ml tube containing homogenized sample. 5. SHAKE vigorously by hand for 1 minute. 6. CENTRIFUGE for rpm, making sure that the solid material is at the bottom of the tube and a liquid layer forms on top of the solid material. Step 2: Dispersive SPE (dspe) QuEChERS Kit: roq QuEChERS EN ml DSPE Kit Description: 15 ml CT, 900 mg MgSO 4, 150 mg PSA, 15 mg GCB Part No.: KS Procedure: 1. TRANSFER 6 ml of supernatant from Step 6 of the extraction process into the 15 ml tube containing the QuEChERS dspe sorbents. 2. SHAKE vigorously by hand for 30 seconds. 3. CENTRIFUGE for rpm to separate solid material from the liquid layer. 4. TRANSFER desired supernatant to an autosampler vial. 5. DILUTE samples 1:10 in mobile phase A prior to injection or LC-MS analysis. QuEChERS Procedure (EN Method) Pretreatment: Kale and grapes were frozen at ~-80 C overnight Step 1: Extraction QuEChERS Kit: roq QuEChERS EN Method Extraction kit Description: 4.0 g MgSO 4, 1.0 g NaCl, 1.0 g SCTD, 0.5 g SCDS Part No.: KS For additional technical notes, visit Page 1 of 8
2 Figure 1. Standard Chromatogram (0.5 ppm standard) 4.5e6 3.5e6 3.0e6 2.5e6 LC-MS/MS Method Parameters Column: Kinetex 5 µm Biphenyl Dimensions: 50 x 4.6 mm Part No.: 00B-4627-E0 Guard Cartridges: ULTRA Cartridges Part No.: AJ Mobile Phase: A: 10mM Ammonium Formate in Water B: Methanol Gradient: Time (min) % B Flow Rate: 0.5 ml/min Inj. Volume: 20 µl Temperature: 35 C Detection: MS/MS (ESI+), scheduled MRM Detector: SCIEX 4000 QTRAP 1.0e min App ID Kale Sample (1 ppm spike) 6.4e6 6.0e6 5.5e6 4.5e6 3.5e6 3.0e6 Grapes Sample (0.5 ppm spike) 5.4 e6 4.5e6 3.5e6 3.0e6 2.5e6 2.5e6 1.0e min App ID e min App ID Page 2 of 8
3 Table 1. Mass Spectrometer Conditions Analyte RT Q1 Q3 Analyte RT Q1 Q3 Methamidophos Aldicarb Acephate Fuberidazole Cyromazine Thiabendazole Omethoate Thidiazuron Mesotrione Fluometuron Dinotefuran Acetamiprid Aldicarbsulfoxide Clethodim E Propamocarb Tebuthiuron Butoxycarboxim Propoxur Aldicarbsulfone Ethirimol Flonicamid Aminocarb Formetanate Bendiocarb Methabenzthiazuron Monolinuron Monocrotophos Propham OxamyI Chlortoluron Methomyl Metribuzin Nitenpyram Forchlorfenuron Clothianidin Diuron Fenuron Pyracarbolid Dicrotophos Sulfentrazone Hydroxycarbofuran Isoprocarb Thiamethoxam Carbofuran Acibenzolar-S-methyl Carbaryl Pymetrozine Thiofanox Dimethoate Metobromuron Carbendazim Isoproturon Mevinphos E Thiacloprid Dioxacarb Ethiofencarb Vamidothion Simetryn Cymoxanil Thiophanate-methyl Carbetamide Prometon Mevinphos Z Cycluron Imidacloprid Secbumeton Ethiprole For additional technical notes, visit Page 3 of 8
4 Table 1. con't Analyte RT Q1 Q3 Analyte RT Q1 Q3 Carboxin Methoxyfenozide Linuron Terbutryn Desmedipham Chloroxuron Fludioxinil Fenamidone Isocarbophos Myclobutanil Phenmedipham Tricyclazole Halofenozide Triticonazole Fipronil Boscalid Paclobutrazol Ethofumesate Terbumeton Hexaconazole Ametryn Tebuconazole Triadimenol Prothioconazole Clethodim Z Tebufenozide Diethofencarb Fenarimol Promecarb Diflubenzuron Oxadixyl Diniconazole Pirimicarb Spiroxamine Methoprotryne Zoxamide Iprovalicarb Triflumuron Flutolanil Flufenacet Cyproconazole (isomer 2) Novaluron Cyproconazole (isomer 1) Bifenazate Methiocarb Mandipropamid Mexacarbate Metconazole Nuarimol Penconazole Neburon Mepanipyrim Fenhexamid Furalaxyl Prometryn Flusilazole Pyrimethanil Ipconazole Tetraconazole Fenoxycarb Diclobutrazol Bupirimate Etaconazole (isomer 1) Cyprodinil Triadimefon Epoxiconazole Mepronil Hydramethylnon Mepronil Bitertanol Page 4 of 8
5 Table 1. con't Analyte RT Q1 Q3 Analyte RT Q1 Q3 Spirotetramat Chlorfluazuron Lufenuron Fenpropimorph Dimoxystrobin Piperonyl butoxide Etaconazole (isomer 2) Difenoconazole (isomer 2) Metaflumizone Difenoconazole (isomer 1) Fenbuconazole Pyraclostrobin Hexaflumuron Alanycarb Butafenacil Spiromesifen Picoxystrobin Benzoximate Carfentrazone-ethyl Etoxazole Triflumizole Benfuracarb Fluquinconazole Propargite Bromuconazole-cis Quinoxyfen Dimethomorph (isomer 1) Furathiocarb Cyazofamid Pyriproxyfen Dimethomorph (isomer 2) Spirodiclofen Pencycuron Avermectin B1b Imazalil Amitraz Tebufenpyrad Hexythiazox Thiobencarb Avermectin B1a Propiconazole Ivermectin B1b Kresoxim-methyl Fenazaquin Mefenacet Temephos Azoxystrobin Chlorantranilinprole Famoxadone Fenpyroximate Flufenoxuron Pyridaben Benalaxyl Ivermectin B1a Trifloxystrobin Moxidectin Buprofezin Doramectin Bromuconazole-trans Rotenone Indoxacarb Emamectin B1b Prochloraz Emamectin B1a Clofentezine Spinosyn A Fluoxastrobin Spinosyn D Spinetoram A Spinetoram B For additional technical notes, visit Page 5 of 8
6 Kale (not spiked) Grapes (not spiked) 5.5e5 6.0e5 5.5e5 4.5e5 4.0e5 4.5e5 3.5e5 3.0e5 2.5e5 2.0e5 4.0e5 3.5e5 3.0e5 2.5e5 1.5e5 2.0e5 1.0e5 5.0e4 App ID e5 1.0e5 5.0e4 App ID min min Imidacloprid Acetamiprid 2.2e6 8.4e6 2.1e6 8.0e6 7.5e6 1.9e6 1.8e6 7.0e6 1.7e6 6.5e6 1.6e6 6.0e6 1.4e6 5.5e6 Area, counts 1.3e6 1.2e6 1.1e6 9.0e5 Calibration Standard Conc (ppm) Accuracy (%) %CV Area, counts 4.5e6 3.5e6 Calibration Standard Conc (ppm) Accuracy (%) %CV 8.0e5 7.0e5 6.0e5 3.0e5 2.0e5 1.0e Concentration, ng/ml App ID e6 2.5e6 1.0e Concentration, ng/ml App ID Azoxystrobin 3.1e7 3.0e7 2.8e7 2.6e7 2.4e7 2.2e7 2.0e7 Area, counts 1.8e7 1.6e7 1.4e7 1.2e7 1.0e7 8.0e6 6.0e Calibration Standard Conc (ppm) Accuracy (%) %CV Concentration, ng/ml App ID Page 6 of 8
7 Results & Discussion Described is a dependable LC-MS/MS method covering 212 pesticides in the SCIEX idquant Standards Kit using a QuEChERS clean-up sample preparation and a Kinetex Biphenyl LC Column Chemistry. The quick, easy, cheap, effective, rugged and safe (QuEChERS) method offers a relatively simple solution for the determination of a wide range and extensive list of pesticide compounds from many different matrices. To add method consistency and throughput, roq QuEChERS kits feature pre-weighed salt packets and pre-filled dispersive SPE sorbents in centrifuge tubes to increase throughput and improve consistency. Over 200 pesticides were screened at concentration ranges between 1 ppm to 1 ppm with the majority of analytes having a recovery range between %. In reviewing the top pesticides tested in kale and grapes, imidacloprid, azoxystrobin, and acetamiprid all demonstrated excellent linearity across the range with r 2 > The highest and lowest calibrators were excluded for weak or strong ionizing compounds, respectively. Precision was good with CV < 20 %. Unspiked kale and grape samples were screened. The kale sample was found to have imidacloprid, tebuconazole, indoxacarb, Spinetoram A, and Spinetoram B. Acetamiprid, myclobutanil, fenhexamid, tebuconazole, and difenconazole were present in the grape sample. Conclusion Kinetex Biphenyl has proven to be an excellent column chemistry for multiresidue pesticide LC-MS/MS screening. The Biphenyl stationary phase yields improved polar retention through dipole-dipole interactions with conjugated biphenyl rings, improved aromatic selectivity, and a large enough carbon skeleton for aliphatic selectivity suitable for the range of polarities and functionalities in multiresidue analyses. Kale: Pesticide Concentration (ppm) CV (%) Imidacloprid 33 6 % Tebuconazole % Indoxacarb 54 1 % Spinetoram A 23 3 % Spinetoram B 16 1 % Grapes: Pesticide Concentration (ppm) CV (%) Acetamiprid % Myclobutanil % Fenhexamid % Tebuconazole % Difenoconazole 23 2 % For additional technical notes, visit Page 7 of 8
8 APPLICATIONS Kinetex Core-Shell LC Column Ordering Information Kinetex 5 μm Columns (mm) Phases 50 x 2.1 3/pk 50 x 3.0 3/pk 50 x x x x 4.6 3/pk Biphenyl 00B-4627-AN AJ B-4627-Y0 AJ B-4627-E0 00D-4627-E0 00F-4627-E0 00G-4627-E0 AJ for 2.1 mm ID for 3.0 mm ID for 4.6 mm ID Kinetex 2.6 μm Columns (mm) Phases 50 x x 2.1 3/pk 50 x 3.0 3/pk 50 x x x 4.6 3/pk Biphenyl 00B-4622-AN 00D-4622-AN AJ B-4622-Y0 AJ B-4622-E0 00D-4622-E0 00F-4622-E0 AJ for 2.1 mm ID for 3.0 mm ID for 4.6 mm ID SecurityGuard Kinetex 1.7 μm Minibore Columns (mm) Phases 50 x x x 2.1 3/pk Biphenyl 00B-4628-AN 00D-4628-AN 00F-4628-AN AJ SecurityGuard ULTRA cartridges require holder, Part No. AJ for 2.1 mm ID Australia t: +61 (0) f: +61 (0) auinfo@phenomenex.com Austria t: +43 (0) f: +43 (0) anfrage@phenomenex.com Belgium t: +32 (0) (French) t: +32 (0) (Dutch) f: +31 (0) beinfo@phenomenex.com Canada t: +1 (800) f: +1 (310) info@phenomenex.com China t: f: +86 (0) phen@agela.com Denmark t: f: nordicinfo@phenomenex.com Finland t: +358 (0) f: nordicinfo@phenomenex.com France t: +33 (0) f: +33 (0) franceinfo@phenomenex.com Germany t: +49 (0) f: +49 (0) anfrage@phenomenex.com India t: +91 (0) f: +91 (0) indiainfo@phenomenex.com Ireland t: +353 (0) f: eireinfo@phenomenex.com Italy t: f: italiainfo@phenomenex.com Luxembourg t: +31 (0) f: +31 (0) nlinfo@phenomenex.com Mexico t: f: tecnicomx@phenomenex.com The Netherlands t: +31 (0) f: +31 (0) nlinfo@phenomenex.com New Zealand t: +64 (0) f: +64 (0) nzinfo@phenomenex.com Norway t: f: nordicinfo@phenomenex.com Puerto Rico t: +1 (800) 541-HPLC f: +1 (310) info@phenomenex.com Spain t: f: espinfo@phenomenex.com Sweden t: +46 (0) f: nordicinfo@phenomenex.com United Kingdom t: +44 (0) f: +44 (0) ukinfo@phenomenex.com USA t: +1 (310) f: +1 (310) info@phenomenex.com All other countries Corporate Office USA t: +1 (310) f: +1 (310) info@phenomenex.com Phenomenex products are available worldwide. For the distributor in your country, contact Phenomenex USA, International Department at international@phenomenex.com roq Extraction Kits Extraction Kits contain fifty easy-pour salt packets and fifty 50 ml stand-alone centrifuge tubes Description Unit Part No. EN Method Extraction Kits 4.0 g MgSO 4, 1.0 g NaCl, 1.0 g SCTD, 0.5 g SCDS 50/pk KS0-8909* AOAC Method Extraction Kits 6.0 g MgSO 4, 1.5 g NaOAc 50/pk KS0-8911* Original Non-Buffered Method Extraction Kits 4.0 g MgSO 4, 1.0 g NaCl 50/pk 6.0 g MgSO 4, 1.5 g NaCl 50/pk *AOAC and EN Extraction Kits also available in traditional non-collared 50 ml centrifuge tubes, Part No.: KSO-8911-NC and KSO-8909-NC KS KS roq dspe Kits dspe Kits contain pre-weighed sorbents/salts inside 2 ml or 15 ml centrifuge tubes Description Unit Part No. 2 ml dspe Kits, 25 mg PSA, 25 mg C18-E 100/pk KS0-8913, 25 mg PSA, 2.5 mg GCB 100/pk KS mg, MgSO 4, 25 mg PSA, 7.5 mg GCB 100/pk KS0-8915, 25 mg PSA 100/pk KS0-8916, 50 mg PSA, 50 mg C18-E, 50 mg GCB 100/pk KS0-8917, 50 mg PSA, 50 mg C18-E 100/pk KS0-8918, 50 mg PSA, 50 mg GCB 100/pk KS0-8919, 50 mg PSA 100/pk KS ml dspe Kits 900 mg MgSO 4, 150 mg PSA, 150 mg C18-E 50/pk KS mg MgSO 4, 150 mg PSA, 15 mg GCB 50/pk KS mg MgSO 4, 150 mg PSA, 45 mg GCB 50/pk KS mg MgSO 4, 150 mg PSA 50/pk KS mg MgSO 4, 400 mg PSA, 400 mg C18-E, 400 mg GCB 50/pk KS mg MgSO 4, 400 mg PSA, 400 mg C18-E 50/pk KS mg MgSO 4, 400 mg PSA, 400 mg GCB 50/pk KS mg MgSO 4, 400 mg PSA 50/pk KS Bulk roq QuEChERS Sorbents Phases 10 g 100 g C18-E 04G-4348 GCB (Graphitized Carbon Black) 04D G-4615 PSA 04G-4610 If Phenomenex products in this technical note do not provide at least an equivalent separation as compared to other products of the same phase and dimensions, return the product with comparative data within 45 days for a FULL REFUND. Terms and Conditions Subject to Phenomenex Standard Terms and Conditions which may be viewed at Trademarks Luna and Gemini are registered trademarks, Krudcatcher, roq, MidBore and SecurityGuard are trademarks of Phenomenex. QTRAP is a registered trademark and idquant is a trademark of AB SCIEX Pte. Ltd. AB SCIEX is being used under license. Arium is a registered trademark of Sartorius AG Phenomenex, Inc. All rights reserved. TN _W Page 8 of 8
[ APPLICATION NOTE ] Multiresidue Analysis of Pesticides in Fruits and Vegetables Using UPLC-MS/MS APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS
Multiresidue Analysis of Pesticides in Fruits and Vegetables Using UPLC-MS/MS Dimple Shah, Naren Meruva, and Gareth Cleland Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Multiresidue method
More informationMultiresidue Analysis of 204 Pesticides in Fresh Produce Using the QuEChERS Method Followed by LC-MS/MS
Multiresidue Analysis of 204 Pesticides in Fresh Produce Using the QuEChERS Method Followed by LC-MS/MS 2015 ACIL P2 Conference April 21, 2015 Crystal City, VA Kelli Sikorski Simon, PhD Food and Drug Administration
More informationChromatography. NUCLEOSHELL Bluebird RP 18 MACHEREY-NAGEL. Fast analyses under highly aqueous conditions
MACHEREY-NAGEL NUCLEOSHELL Bluebird RP 18 Chromatography Fast analyses under highly aqueous conditions nspecial octadecyl core-shell phase with hydrophilic endcapping nsuitable for LC/MS due to low bleeding
More informationAPPLICATIONS TN Determination of Sugars in Animal Feed using HPLC-ELSD
Determination of Sugars in Animal Feed using HPLC-ELSD Laura Snow, Zeshan Aqeel, Allen Misa and Scott Krepich Phenomenex, Inc., Madrid Ave., Torrance, CA 9050 USA Laura Snow Scientist- Phenologix Outside
More informationRun 1 Run 2 TDP PLP TDP PLP. 3.0e5 2.8e5 2.2e5. 2.6e5 2.4e5. 2.2e5 2.0e5. Intensity, cps. 1.8e5 1.6e5 1.4e5 1.2e5 1.0e5 4.0e4. 2.0e e5 1.
APPLICATIONS A Simple and Effective High ph LC/MS/MS Method for Determination of Underivatized Vitamin B1 and B6 in Human Whole Blood Ramkumar Dhandapani Technical Manager Ramkumar loves to write poems,
More informationEURL-FV Multiresidue Method Using QuEChERS
EURL-FV Multiresidue Method Using QuEChERS by GC-QqQ/MS/MS & LC-QqQ/MS/MS for Fruits & Vegetables UCT Products: ECQUEU750CT 50 ml centrifuge tube contains: (4 grams MgSO 4, 1 gram NaCl, 0.5 grams Na Citrate
More informationAPPLICATIONS TN Analytical Method
APPLICATIONS Simultaneous Determination of Antidepressants and Metabolites in Urine Using Ultra-High Performance Liquid Chromatography Tandem Mass Spectrometry (UHPLC-MS/MS) using a Kinetex Core-Shell
More informationEuropean Reference Laboratory - FV. Amadeo R. Fernández-Alba
Summary of Results z-score results: 23764 points z-score FFP RSD 50% 4% of data 3% of data 25% 25% 69% of data 88% of data 50% 2% of data 3% of data ............ Almería Valencia.....................................................................................
More informationEUPT-Panel. DG SANTE Mrs. Almut Bitterhof Mrs. Veerle Vanheusden ORGANISING GROUP Amadeo R. Fernández-Alba Octavio Malato Carmen Ferrer Ana Lozano
EUPT-Panel DG SANTE Mrs. Almut Bitterhof Mrs. Veerle Vanheusden ORGANISING GROUP Amadeo R. Fernández-Alba Octavio Malato Carmen Ferrer Ana Lozano STATISTICAL GROUP Carmelo Rodríguez QUALITY CONTROL GROUP
More informationAchieving the California Pesticide Regulations in Cannabis Using Optimized APCI and ESI Techniques
Achieving the California Pesticide Regulations in Cannabis Using Optimized APCI and ESI Techniques Diana Tran 1, Karl Oetjen 1, Robert Di Lorenzo 2, Paul Winkler 1, Scott Krepich 3, KC Hyland 1, Christopher
More informationProtect Your Mass Spec
Protect Your Mass Spec Sample Preparation Solutions for Biological Samples www.phenomenex.com/bioms Dirty Samples Could be Costing You Money and Sensitivity Biological samples contain a wide variety of
More informationEnantiomeric and Diastereoisomeric Resolutions of Chiral Triazole Fungicides using Lux Polysaccharide-Based Chiral Stationary Phases
T-1164 APPLICATIS Enantiomeric and Diastereoisomeric Resolutions of Chiral Triazole Fungicides using Lux Polysaccharide-Based Chiral Stationary Phases Marc Jacob, Liming Peng, Michael Klein, Tom eveland,
More informationAn Easier QuEChERS Solution for Diverse Matrices. Learn more at
An Easier QuEChERS Solution for Diverse Matrices Learn more at www.phenomenex.com/roq What is roq QuEChERS? QuEChERS is a descriptive acronym that stands for Quick, Easy, Cheap, Effective, Rugged, and
More informationRemoval of Purple Pigmentation from Cannabis using QuEChERS Extraction, ChloroFiltr dspe Clean-up and LC-MS/MS
Removal of Purple Pigmentation from Cannabis using QuEChERS Extraction, ChloroFiltr dspe Clean-up and LC-MS/MS UCT Part Numbers ECMSSC-MP Mylar pouch containing 4g MgSO4 and 1g NaCl CUMPSGGC182CT 2mL dspe
More informationA Fast, Accurate, and Sensitive Quantitiation of Vitamin C in Plasma by Impact Protein Precipitation and Kinetex 5 µm Core-Shell XB-C18 HPLC Columns
A Fast, Accurate, and Sensitive Quantitiation of Vitamin C in Plasma by Impact Protein Precipitation and Kinetex 5 µm Core-Shell XB-C18 HPLC Columns Shuguang Li, Sky Countryman, and Erica Pike Phenomenex,
More informationAPPLICATIONS TN A Sensitive Extraction Method for Nicotine and Metabolites from Urine using Simplified Liquid Extraction (SLE)
APPLICATIONS A Sensitive Extraction Method for Nicotine and Metabolites from Urine using Simplified Liquid Extraction (SLE) Ann Montross Technical Specialist Ann enjoys cooking/baking and roasting her
More informationAPPLICATIONS TN Dramatically Improve Existing 5 µm and 3 µm Fully Porous Methods with Kinetex 5 µm Core-shell Technology.
Dramatically Improve Existing 5 µm and 3 µm Fully Porous Methods with Kinetex 5 µm Core-shell Technology Jeff Layne, Phenomenex, Inc. Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 951 USA With virtually
More informationFinally, a Solid Phase Extraction product for Drugs of Abuse
n io t c a r t x E e s a h P d li Finally, a So e s u b A f o s g u r D r fo product Imagine Only Needing 1 SPE Sorbent For All of Your Drugs of Abuse Extractions Extract over 11 drugs using the same sorbent
More informationMANDATORY CANNABIS TESTING FOR PESTICIDE ACTIVE INGREDIENTS. List and limits
MANDATORY CANNABIS TESTING FOR PESTICIDE ACTIVE INGREDIENTS List and limits Disclaimer: This document must be used in conjunction with the Mandatory cannabis testing for pesticide active ingredients Requirements,
More informationPesticide Residue Testing Laboratory, Krishibhavan, Shivajinagar, Pune, Maharashtra. Discipline Chemical Testing Issue Date
Pesticide Residue Testing, Krishibhavan, Shivajinagar, Last Amended on - Page 1 of 6 I. FOODS & AGRICULTURAL PRODUCTS 1. Organochlorine Group Aldrin AOAC 2012 Method (970.52) 0.01 mg/kg to Endrin 4,4-DDE(PPDDE)
More informationCertificate of Analysis
Page 1 of 9 To: AL-TAZAJ CO. mansoormirza@taza.com.sa Received Date: Location of Test: (except where noted) Kharj - Riyadh Tap Water Analyte Result Units Limits LOD Method Reference Test Date Arsanic ppm
More informationHow Are We Doing? January 2014
How Are We Doing? January 2014 20 produce samples were tested (at an average of 554 different pesticides per product) and we found 31 residues, two of which exceeded the MRL set for In total, we found
More informationPesticide analysis in Hops and Cannabis by GC-MS-MS
PO-CON1752E Pesticide analysis in Hops and Cannabis by GC-MS-MS ASMS 17 TP-191 Jeffrey H. Dahl, 1 Riki Kitano, 1 Vikki Johnson, 1 and Julie Kowalski 2 1 Shimadzu Scientific Instruments; Columbia, Maryland
More informationAPPLICATIONS TN A Comparison of Various Kinetex C18 Phases USP: L1
APPLICATIONS A Comparison of Various Kinetex C18 Phases Zeshan Aqeel and Jeff Layne Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 91 USA Introduction When performing reversed phase LC separations, there
More informationAPPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology
Determination of Impurities and Related Substances for Acetylsalicylic Acid (Ph. Eur. Monograph 9): Increased Sensitivity and Resolution, Faster Analysis, and Reduced Solvent Usage Using Kinetex 2.6 µm
More informationRapid Analysis of Synthetic Cannabinoids and their Metabolites in Urine using Solid Phase Extraction and LC/MS/MS. JWH018-4-Hydroxypentyl metabolite
T-1154 APPLICATIS Rapid Analysis of Synthetic Cannabinoids and their Metabolites in Urine using Solid Phase Extraction and LC/MS/MS Matthew Trass, Shahana Huq, Seyed Sadjadi, Sky Countryman and Zdravko
More informationBroad beans V2.0-07/12/2007
MRL Status report Broad beans V2.0 - Comments and interpretation Where an MRL column contains no data, the particular data source either does not recognise this as an independent crop type, or no information
More informationLabs Across the World Are Advancing Their Protein Precipitation
Phospholipid Removal Solutions Labs Across the World Are Advancing Their Protein Precipitation Achieve Cleaner Samples Extend HPLC Column Lifetime Improve Analyte Sensitivity Reduce Mass Spec Maintenance
More informationAPPLICATIONS Improving Intact Biogeneric Protein Separations with Aeris WIDEPORE Core-Shell Columns
Improving Intact Biogeneric Protein Separations with Aeris WIDEPORE Core-Shell Columns Michael McGinley, Deborah Jarrett, and Jeff Layne Phenomenex, Inc., 411 Madrid Avenue, Torrance, CA 91 USA Aeris WIDEPORE
More informationNEW. Kinetex. Biphenyl TMS TMS TMS TMS. Selectivity that a C18 just can t give you!
Core-Shell Technology NEW Kinetex enyl Biph Biphenyl TM S TM Selectivity that a C8 just can t give you! S Bip he ylhenyl Bnip Remarkable Separation Power Rugged and Reliable % Aqueous Stable Kinetex Core-Shell
More informationMETHAMPHETAMINE & AMPHETAMINE
METHAMPHETAMINE & AMPHETAMINE Enantiomers from Urine Rapid and Accurate Chiral Separation Polysaccharide Chiral Chiral LC LC Columns and SFC PHEN-RUO-49 Revision: Lux 3 µm AMP Column High Efficiency LC/MS/MS
More informationPhospholipid Removal Solutions
Remove Phospholipids Remove Protein No Method Development www.phenomenex.com/phree One Quick Method, Three Big Advantages 1 2 Remove Proteins that can clog HPLC columns Eliminate Phospholipids that cause
More informationCore-Shell Technology for Proteins and Peptides
Core-Shell Technology for Proteins and Peptides Better BioSeparations on HPLC and UHPLC Systems www.phenomenex.com/aeris Aeris Core-Shell Technology Core-Shell Particles Precision Engineered for Protein
More informationAPPLICATIONS TN Materials and Methods
APPLICATIONS A Simple, Fast, and Efficient HPLC Method for Sensitive, Accurate, and Reproducible Quantitation of Vitamin C in Human Plasma by Applying Impact Protein Precipitation Plate Technology with
More informationEuropean Reference Laboratory - FV. 53 rd Annual North American Chemical Residue Workshop. 18 th July 2016
18 th July 2016 Pyrimethanil at 0.01 mg/kg 200.1182 ± 1 Da 200.1182 ± 0.001 Da Pyrimethanil [M+H + ] m/z 200.1182 0.01 mg/ kg in onion RT = 7.05 min Dilution 5x MILLING SAMPLE EXTRACTION LC MS/MS REPORT
More informationFor Far. Greater. UHPLC Column Performance You Must Have Kinetex Core-Shell UHPLC Columns.
For Far Greater UHPLC Column Performance... You Must Have Kinetex Core-Shell UHPLC Columns www.phenomenex.com/kinetex The UHPLC Core-Shell Advantage Kinetex Core-Shell Technology produces increased efficiencies
More informationPesticideScreener. Innovation with Integrity
PesticideScreener A complete multi-residue solution for pesticide screening using high resolution full scan accurate mass data Innovation with Integrity LC-MS The Challenge of Ever Increasing Number of
More informationAPPLICATION TN Materials and Methods Reagents and Chemicals
TN-8 APPLICATION Fast Analysis of Sucrose,, and Composition in Fruit Juices and Processed Beverages using Simplified HPLC Methodology Brian Rivera Product Manager In addition to chromatography, Brian also
More informationRapid Screening and Quantitation of Pesticide Residues in Cannabis by Modified QuEChERS and LC-MS-MS
PO-CON689E Rapid Screening and Quantitation of Pesticide Residues in Cannabis by Modified QuEChERS and LC-MS-MS ASMS 206 TP 20 Jeffrey H. Dahl, Julie Kowalski 2, Derek Laine, and Jason Zitzer Shimadzu
More informationAPPLICATIONS TN µm Core-Shell Particle. (continued on page 2) Page 1 of 8. For additional technical notes, visit
Kinetex 5 µm Core-Shell Columns Deliver Significant Improvements in Chromatographic Efficiency and Resolution versus Traditional Fully Porous 5 µm Columns Philip J. Koerner and Jeff Layne Phenomenex, Inc.,
More informationSIMPLIFIED CHIRAL SEPARATIONS. Lux Cellulose-1. Guaranteed Alternative. to CHIRALCEL OD-H. Exceptional Quality Sensibly Priced
SIMPLIFIED IRAL SEPARATIS Lux ellulose- Guaranteed Alternative to IRALEL D- Exceptional Quality Sensibly Priced Lux ellulose- ellulose tris(3,5-dimethylphenylcarbamate) This dimethyl cellulose selector
More informationPHYTO+ Certificate of Analysis:
Sunshine Trading PHYTO Plus Cannabinoid Profile + O PHYTO Plus % Total CBD: Cannabinoid Profile Component Mass (%) CBD Amount (mg/g) Limit 71.20 CBDA CBG CBGA 0.12 1.20 D9THC THCA 71.20 CBN Total CBD HPLC
More informationQuEChERS Quick, Easy, Cheap, Effective, Rugged and Safe Method for Determining Pesticide Residues. Fruits and Vegetables
QuEChERS Quick, Easy, Cheap, Effective, Rugged and Safe Method for Determining Pesticide Residues 1 Fruits and Vegetables 2 Milk and Honey 3 Meat 4 Tobacco 5 QuEChERS Multiclass, multiresidue method (MRM)
More informationPreparative Chromatography Meets UHPLC
Preparative Chromatography Meets UHPLC The Power of Kinetex Core-Shell Technology with the Performance of Piston Packed Prep Columns Ultra-High Efficiencies Ultra-Long Lifetimes NO Added Pressure Two Award
More informationShuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W
Simple, Fast and Accurate Quantitation of Human Plasma Vitamins and Their Metabolites by Protein Precipitation Combined with Columns Using HPLC-UV, HPLC-FLD or LC/MS/MS Shuguang Li, Jason Anspach, Sky
More informationSKAL:100364, ISO 14001: 2004 certified; ISO 9001: 2008 certified, Organic certified: NL-BIO-01, HACCP certified; GMP certified
Cannabinoid Profile PHYTO Plus 99.23% Total CBD: Cannabinoid Profile CBD 99.23 992.30 N/A CBDA
More informationAnalysis of Pesticide Residues, Mycotoxins and Potency in Cannabis using QuEChERS Extraction, ChloroFiltr dspe Cleanup and LC-MS/MS
Analysis of Pesticide Residues, Mycotoxins and Potency in Cannabis using QuEChERS Extraction, ChloroFiltr dspe Cleanup and LC-MS/MS UCT Part Numbers ECMSSC-MP Mylar pouch containing 4g MgSO4 and 1g NaCl
More informationPAIN PANEL. Research and Solutions
PAIN PANEL Research and Solutions Pain management drug use and abuse has increased significantly in recent years. The 21 National Survey on Drug Use & Health (NSDUH) reports that 17.3% of first-time drug
More informationACE is a registered trademark of Advanced Chromatography Technologies Limited. HALO and Fused-Core are registered trademarks of Advanced Materials
ACE is a registered trademark of Advanced Chromatography Technologies Limited. HALO and Fused-Core are registered trademarks of Advanced Materials Technology. TABLE OF CONTENTS CANNABINOID Different Cannabinoids
More informationValidation Report 8. EURL for Cereals and Feeding stuff National Food Institute Technical University of Denmark
EURL for Cereals and Feeding stuff National Food Institute Technical University of Denmark Validation Report 8 Determination of pesticide residues in wheat, oat, rye, rice and barley by LC-MS/MS (QuEChERS
More informationPesticide Residue Analysis in Whole Milk by QuEChERS and LC-MS/MS
Pesticide Residue Analysis in Whole Milk by QuEChERS and LC-MS/MS UCT Part Numbers: Enviro-Clean RFV5CT (5 ml centrifuge tubes) Enviro-Clean ECMSSA5CT-MP (Mylar pouch containing 6 g MgSO 4 and 1.5 g NaOAc)
More informationFigure 1. Core-Shell Particle Technology
Purification of Chlorogenic cid from Green Coffee using Kinetex Core-Shell Technology in xia Preparative Formats Zeshan qeel Technical Magician Zeeshan loves to collect watches and the ack to the Future
More informationColumns The FIRST and ONLY sub-2 µm core-shell columns on the market
Better UHPLC Performance Starts NW.7 µm Core-Shell * Columns The IRST and NLY sub- µm core-shell columns on the market Phenomenex India Laxmi Cyber City, Ground loor, B Block, Survey no: Kondapur, Hyderabad
More informationSKAL:100364, ISO 14001: 2004 certified; ISO 9001: 2008 certified, Organic certified: NL-BIO-01, HACCP certified; GMP certified
Cannabinoid Profile PHYTO Plus 99.26% Total CBD+CBDa: Cannabinoid Profile Component Mass (%) Amount (mg/g) Limit CBD 99.26 992.60 N/A CBDA
More informationAnalytical Report. Environmental Micro Analysis, Inc. ELAP Certificate #2211. PO No: Client Sample ID: EMA Sample No: Date Received: Sample Matrix:
FDA IMP-OC1 4/24/2009 = Not Detected at or above Environmental Micro Anlaysis, Inc. Acrinathrin Aldrin alpha-bhc Atrazine Benfluralin beta Chlordene Bifenthrin Boscalid Bromopropylate Captan Chlorodimeform
More informationResults for Chemical Residue and Microbiological Testing
Results for Chemical Residue and Microbiological Testing Customer: Biodynamic Marketing Date Batch Product Grower Test Code Result 14/11/08 RT1616 Zucchini Murray & Blackley AT3 No detection Batch No Substance
More informationEndoca Certificate of Analysis:
Endoca Certificate of Analysis: Total CBD+CBDA.9% Cannabinoid Profile: HPLC Chromatograph Raw Data Component Mass (%) Amount (mg/g) CBD.8.8 CBDA.7 7. CBDV.9.9 CBG CBGA CBN.. CBC Δ9THC THCA Δ8THC CBD+CBDA.9.9
More informationAgilent 6470 Triple Quadrupole LC/MS System and EMR-Lipid. The Combination for Confident Pesticide Quantitation. Christoph Mueller
Agilent 670 Triple Quadrupole LC/MS System and EMR-Lipid The Combination for Confident Pesticide Quantitation Christoph Mueller Agilent Technologies LCMS Application Scientist EMEAI Food Market Program
More informationCertificate of Analysis: Organic Hemp CO2 Extract Cannabinoid Profile
Cannabinoid Profile Eczane Group 99.08% Total CBG: Cannabinoid Profile Component Mass (%) Amount (mg/g) Limit CBD 00.26 02.60 N/A CBDA
More informationValidation Report 1. CRL for Cereals and Feeding stuff National Food Institute Danish Technical University
CRL for Cereals and Feeding stuff National Food Institute Danish Technical University Validation Report 1 Determination of pesticide residues in cereals by LC- MS/MS (internal method no. FP086.1) Hanne
More informationAnalysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS
Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS Food Application Author Fadwa Al-Taher Institute for Food Safety and Health Illinois Institute of Technology
More informationSUBSTANCES AUTHORISED FOR AGRICULTURAL USE IN THE EU
Authorised in the EU Updated April 2008 SUBSTANCES AUTHORISED FOR AGRICULTURAL USE IN THE EU Since the beginning of the EU s revision process in 1993, 190 existing and new active substances have been approved
More informationDetermination of Patulin in Apple Juice Using SPE and UHPLC-MS/MS Analysis
Determination of Patulin in Apple Juice Using SPE and UHPLC-MS/MS Analysis UCT Part Numbers ECHLD126-P ENVIRO-CLEAN HLDVB mg/6ml SPE cartridge PE Frit SLDAID21-18UM Selectra DA UHPLC column ( 2.1 mm, 1.8
More informationAnalysis of Limonin in Citrus Juice Using QuEChERS and LC-MS/MS
Analysis of Limonin in Citrus Juice Using QuEChERS and LC-MS/MS UCT Part Numbers ECQUEU7-MP Mylar pouch containing 4g MgSO4, 1g NaCl, 1g Na3Cit 2H2O and 0.5g Na2Cit 1.5H2O CUMPSC1875CB2CT 2mL dspe tube
More informationPure Chromatography. Foods, Flavors & Fragrances Applications. Abstract. Introduction
Foods, Flavors & Fragrances Applications A Comprehensive Approach to Pesticide Residue Testing, Including Non-Target Analysis, for Fruits, Vegetables, and Nuts, Using QuEChERS, LC-MS/MS, and GCxGC-TOFMS
More informationMethod validation report Determination of pesticides in cereals using the QuEChERS method and GC-ITD
CRL for Cereals and Feeding stuff National Food Institute Danish Technical University Method validation report Determination of pesticides in cereals using the QuEChERS method and GC-ITD Susan Strange
More informationDetected Substance Level Detected Permitted Level. Peach MB4 - Ecoli/Listeria Monocytogenes/ Salmonella/ Staphylococci / Thermotolerant Coliforms
P.O. Box 4032 www.rudge.com.au Phone +613 9408 4645 Fax +613 9401 4625 1 Cutri Fruit 38 Byrnes Road Woorinen Vic 1 Product Testing s Customer : Cutri Fruit Test Date Ref Item Test Type 17/11/2016 Peach
More informationThis annex is valid from: to Replaces annex dated: Location(s) where activities are performed under accreditation
Distributieweg 1 2645 EG Delfgauw The Netherlands Location(s) where activities are performed under accreditation Head Office Location Abbreviation/ location code Head fice Distributieweg 1 2645 EG Delfgauw
More informationATTENTION. MRL LOR Result N/A <0.01 <0.01 N/A <0.05 <0.050 < 0.50 <0.01 <0.010 N/A <0.05 <0.05
ACCREr.)lTE[) FOR AtCJdit:Jtiol1 Number: 272& BernadeUe Bulmer DATE RECEIVED OUR SAMPLE NUMBER YOUR REFERENCE SAMPLE TYPE DESCRIPTION TEST C3 Chemicals Test List ( Acc No 2726) TP/311 Acephate (mg/kg)
More informationDevelopment and Validation of Multiresidue Pesticide Methods at FDA/CFSAN
Development and Validation of Multiresidue Pesticide Methods at FDA/CFSAN Jon W. Wong U.S. Food and Drug Administration Center for Food Safety and Applied Nutrition American Council of Independent Laboratories
More informationKey Advantages of Comprehensive Cannabis Analysis
Comprehensive Cannabis Analysis: Pesticides, Aflatoxins, Terpenes, and High Linear Dynamic Range Potency from One Extract Using One Column and One Solvent System Robert Di Lorenzo 1, Diana Tran 2, KC Hyland
More informationCustomer Sample Type Method Technique
Sample ID 18A23000003 Cannabis Potency HPLC-UV Chemistry Cannabidivarin (CBD-V) 1.29 mg/dose Cannabidiol (CBD) 646.50 mg/dose Cannabigerol (CBG) Tetrahydrocannabivarin (THC-V) 24.7 mg/dose
More informationProduct Testing Results
P.O. Box 4032 www.rudge.com.au Phone +613 9408 4645 Fax +613 9401 4625 Product Testing s Customer : NATURES HAVEN Item : ZUCCHINI (GREEN) Ref : Test Date Lab Ref Test Type 01/11/2017 AT3 - Full Residue
More informationAPPLICATION TN Rapid Analysis of a 48 Drug Research Panel in Less Than 4 Minutes Using Kinetex Biphenyl Core-Shell Technology HPLC/UHPLC Columns
APPLICATION Rapid Analysis of a 48 Drug Research Panel in Less Than 4 Minutes Using Kinetex Biphenyl Core-Shell Technology HPLC/UHPLC Columns Sean Orlowicz Manager, PhenoLogix When not in the lab, Sean
More informationPAIN PANEL. Application Guide. Screen 41 compounds in under 4 minutes
Application Guide PAIN PANEL Screen 41 compounds in under 4 minutes Solutions for large compound screens and individual compound classes such as benzodiazepines, opiates, and more Achieve resolution of
More informationComparison of Different Whole Blood Sample Pretreatment Methods for Targeted Analysis of Basic Drugs. Table 1. List of pain panel drugs
APPLICATION Comparison of Different Whole Blood Sample Pretreatment Methods for Targeted Analysis of Basic Drugs Seyed Sadjadi, Shahana Huq, Sean Orlowicz, and Laura Snow Phenomenex, 4 Madrid Avenue, Torrance,
More informationAgilent Sample Preparation The Pesticide Analysis Workflow
Agilent Sample Preparation The Pesticide Analysis Workflow Insights into Preparing your Cannabis Sample and Triple Quad Mass Spec Analysis Joan Stevens, Ph.D. Sample Preparation Application Specialist
More informationProduct Testing Results
P.O. Box 4032 www.rudge.com.au Phone +613 9408 4645 Fax +613 9401 4625 Product Testing s Customer : NATURES HAVEN Item : ONION - BROWN Ref : Test Date Lab Ref Test Type 18/12/2017 AT3 - Full Residue Analysis
More informationAPPLICATIONS TN Rapid Analysis of Hop Acids in Beer using Strata -X Solid Phase Extraction and Kinetex 2.6 µm Core-Shell Technology Column
Rapid Analysis of Hop Acids in Beer using Strata -X Solid Phase Extraction and Kinetex.6 µm Core-Shell Technology Column Philip J. Koerner, Matthew Trass, and Jeff Layne Phenomenex, Inc., Madrid Ave.,
More informationZero Cross-talk on the TSQ Quantum
Application ote: 351 Zero Cross-talk on the TQ Quantum Dipankar Ghosh, Mark Churchill, Eric Genin, and Alan choen; Thermo Fisher cientific, an Jose, CA Key Words TQ Quantum urveyor HPLC AQUAIL Environmental
More informationFast Screening and Quantification of Pesticide Residues Using a Comprehensive LC-MS Solution: The Pesticide Explorer Collection Standard Quantitation
Fast Screening and Quantification of Pesticide Residues Using a Comprehensive LC-MS Solution: The Pesticide Explorer Collection Standard Quantitation Katerina Bousova 1, Ebru Sarikaya 1, Michal Godula
More informationEurofins Agroscience Testing Pty Ltd CHEMICAL TESTING LABORATORY Accreditation Number: Site Number: 14691
Eurofins Agroscience Testing Pty Ltd CHEMICAL TESTING LABORATORY Accreditation Number: 15028 Site Number: 14691 Address Details: Unit F6, Building F 16 Mars Road LANE COVE, NSW 2066 AUSTRALIA Website:
More informationA Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS
A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS By Shun-Hsin Liang and Frances Carroll Abstract Vitamin K₁ and K₂ analysis is typically complex and time-consuming because these
More informationKeep It Clean. Solid Phase Extraction Solutions for Drugs of Abuse Analysis. Neutral Drugs of Abuse Basic Drugs of Abuse
Keep It Clean Solid Phase Extraction Solutions for Drugs of Abuse Analysis Neutral Drugs of Abuse Basic Drugs of Abuse Strata -X-Drug B: Basic Drugs of Abuse Working with basic drugs? Here are 3 reasons
More informationA Sensitive and Robust Workflow to Measure Residual Pesticides and Mycotoxins from the Canadian Target List in Dry Cannabis Flower
Application Note Cannabis Testing, Food A Sensitive and Robust Workflow to Measure Residual Pesticides and Mycotoxins from the Canadian Target List in Dry Cannabis Flower Authors Jean-Francois Roy 1, Christophe
More informationApplication Note: 425. Introduction. Goal. Experimental Conditions
Application Note: 425 Determination of Different Classes of Pesticide Residues in Processed Fruits and Vegetables by LC-MS Using the TSQ Quantum Ultra According to EU Directive 91/414 EEC Eleni Botitsi,
More informationUHPLC / HPLC Columns Developed for Mass Spectrometry
ACE UltraCore SuperC18 UHPLC / HPLC s Developed for Mass Spectrometry Water analysis Pharmaceutical Environmental Toxicology Food safety Clinical Forensics Metabolism Trace analysis Ultra-inert 2.5µm and
More informationPhenomenex roq QuEChERS Kits. Overview. Spinach Facts
Optimisation of LC/MS Analysis of Pesticide Residues in Food Products When Using QuEChERS Technique Phenomenex roq QuEChERS Kits Aman Sharma Phenomenex ANZ amans@phenomenex.com 1 2 Background Overview
More informationGeo-Chem Laboratories (P) Ltd., Pragati Building, Near Crompton Greaves, Kanjur Marg (E), Mumbai, Maharashtra
Last Amended on 24.05.2014 Page 1 of 171 I. FOODS AND AGRICULTURAL PRODUCTS 1 Cereals and Pulses Whole & Ground 1000 Kernel Weight IS 4333 (Part 4) -2002 RA 2009 1 10 gm Total Ash GAFTA 12 : 0 2003 0.1
More informationCAN T FIND THE STANDARDS YOU ARE LOOKING FOR?
Guide to Chemical pesticides have become an integral part of the agricultural toolbox, offering protection to crops from destructive pests. However, some unfortunate side effects of their use include potential
More informationCannabis Reference Standards
Cannabis Reference Standards As regulation intensifies and the need for more testing grows, many cannabis producers and manufacturers are relying on accurate test results from state certified testing labs
More informationTN-0116 Scott Krepich
TN-116 APPLICATIONS Analysis of Perfluorinated Compounds (PFCs) in Aqueous Matrices, Evaluating Various Online SPE Sorbents by LC-MS/MS Scott Krepich Senior Field Application Scientist Scott enjoys surfing
More informationFast and simultaneous analysis of ethanol metabolites and barbiturates using the QTRAP 4500 LC-MS/MS system
Fast and simultaneous analysis of ethanol metabolites and barbiturates using the QTRAP 4500 LC-MS/MS system Xiang He 1, Adrian Taylor 2 and Alexandre Wang 1 1 SCIEX, Redwood City, USA. 2 SCIEX, Concord,
More informationYes You Can! New! Amazing EVO C18. Core-Shell Technology
Core-Shell Technology Yes You Can! New! Amazing EVO C8 Develop robust methods from ph - Get improved peak shape for bases Easily reduce run times and increase sensitivity THE PAST Fully Porous You used
More informationWaters Mass Spectrometry Systems
Waters Mass Spectrometry Systems Mark N Groudas VP Americas Field Operations 25 th, July 2017 2016 Waters Corporation 1 The Science of What s Possible 2016 Waters Corporation 2 Waters MS Systems Portfolio
More informationModified QuEChERS for HILIC LC/MS/MS Analysis of Nicotine and Its Metabolites in Fish
Modified QuEChERS for HILIC LC/MS/MS Analysis of Nicotine and Its Metabolites in Fish Application Note Food Testing & Agriculture Author Mike Chang Agilent Technologies, Inc. Abstract Nicotine is one of
More informationIntegration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation
PO-CON69E Integration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation MSACL 6 EU Stéphane Moreau, Daisuke Kawakami, Toshikazu Minohata Shimadzu Europe GmbH, Duisburg,
More informationAPPLICATION OF ELEVATED RESOLUTION GC-TOF-MS FOR THE MULTIRESIDUE ANALYSIS OF PESTICIDES IN FOOD
APPLICATION OF ELEVATED RESOLUTION GC-TOF-MS FOR THE MULTIRESIDUE ANALYSIS OF PESTICIDES IN FOOD Peter Hancock 1 Waters Corporation, Manchester M22 5PP, UK. INTRODUCTION The inappropriate or unlawful use
More information