1 Instructions for use PROPHYflex perio tip Starter Set Refill PROPHYflex perio tip
2 Distributed by: KaVo Dental GmbH Bismarckring 39 D Biberach Phone Fax Manufacturer: Kaltenbach & Voigt GmbH Bismarckring 39 D Biberach
3 Table of contents Table of contents 1 User instructions Safety Description of safety instructions Warning symbol Structure Description of hazard levels Safety instructions Product description Purpose Intended use Transportation and storage conditions Startup and shut down Inserting the adaptor into the PROPHYflex Inserting the PROPHYflex perio tip into the wrench Screwing the PROPHYflex perio tip onto the adaptor using the wrench Removing the PROPHYflex perio tip Operation s for application Preparation methods according to ISO Preparations at the site of use Preparations before reprocessing Non-fixing pre-cleaning of the PROPHYflex perio tip Cleaning Manual external and internal cleaning Automated external and internal cleaning Disinfection Manual external disinfection Manual disinfection - internal Machine disinfection - external and internal Drying Visual inspection and functional check Packaging Sterilisation Storage Tools Terms and conditions of warranty / 22
4 1 User instructions 1 User instructions Dear User Congratulations on purchasing this KaVo quality product. By following the instructions below you will be able to work smoothly, economically and safely. Symbols Refer to the chapter on Safety/Warning symbol Important information for users and service technicians Can be steam-sterilised at 134 o C -1 o C / +4 o C (273 o F -1.6 o F / +7.4 o F) Thermodisinfectable CE-Zeichen (Communauté Européenne). Ein Produkt mit diesem Zeichen entspricht den Anforderungen der anwendbaren EG-Richtlinie. Action request Target group This document is intended for dentists and their assistants. The section on starting up is also intended for service technicians. 4 / 22
5 2 Safety 2 Safety 2.1 Description of safety instructions 2.1 Description of safety instructions Warning symbol Warning symbol Structure DANGER The introduction describes the type and source of the hazard. This section describes potential consequences of non-compliance. The optional step includes necessary measures for hazard prevention Description of hazard levels The safety instructions listed here, together with the three levels of danger will help avert property damage and injury. indicates a hazardous situation that can cause damage to property or mild to moderate injuries. WARNING WARNING indicates a hazardous situation that can lead to serious or fatal injury. DANGER DANGER indicates a maximal hazard due to a situation that can directly cause death or fatal injury. 2.2 Safety instructions Hazard of air embolism and skin emphysema. There is a danger of insufflation of spray in open wounds and areas of inflammation causing air embolisms and skin emphysema. Avoid the insufflation of spray in open wounds and areas of inflammation. Hazard from the perio tip detaching from the PROPHYflex during the treatment. This can be a hazard for care provider and patient. Check the perio tip for secure attachment before treatment. 5 / 22
6 2 Safety 2.2 Safety instructions Hazard from a bent perio tip due to frequent use or dropping the handpiece. The PROPHYflex perio tip can fracture or be contaminated. The PROPHYflex perio tip may be re-used for up to10 times. Premature wear and malfunctioning from improper storage during long periods of nonuse. Reduced production time. The instrument must be cleaned and stored dry if it has not been used for a long period. When the PROPHYflex is placed on a support, the torque wrench should be placed on the perio tip for safety reasons as protection against injuries. Protective goggles must be worn by both the care provider and patient during treatment. We recommend working with a dust extraction system and mouth protector. After the treatment, unscrew the empty powder container and rinse the PROPHY FLEX perio tip for approx. 10 seconds with air and water. 6 / 22
7 3 Product description 3.1 Purpose Intended use 3 Product description PROPHYflex perio tip Starter Set - REF x adaptor, 2 3 x perio tip, 3 1 x wrench, 4 1 x nozzle pin PROPHYflex perio tip Refill - REF x perio tip, 2 1 x wrench, 3 1 x nozzle pin 8,5 mm 5,5 mm 3,5 mm The markings of the PROPHYflex perio tip are based on the WHO index. 3.1 Purpose Intended use Indications for use: This medical device is intended for dental treatment only. All other types of use or alterations to the product are not permitted and can be hazardous. The PROPHYflex perio tip is used in combination with PROPHYflex perio powder to remove sub- and supragingival plaque. Proper use: According to these provisions, the medical device is only for the described use in conformance with: the applicable health and safety regulations, the applicable accident prevention regulations and these instructions for use. According to these regulations, the user is required to: only use properly operating equipment, use the equipment for the proper purpose, to protect himself, the patient and third parties from danger, to avoid contamination from the product. 7 / 22
8 3 Product description 3.2 Transportation and storage conditions 3.2 Transportation and storage conditions It is hazardous to start up the medical device after it has been stored strongly refrigerated. This can cause the medical device to malfunction. Prior to start-up, very cold products must be heated to a temperature of 20 o C to 25 o C (68 o F to 77 o F). Temperature: -50 C to +80 C (-58 F to +176 F) Relative humidity: 5% RH to 95% RH absence of condensation Air pressure: 700 hpa to 1060 hpa (10 psi to 15 psi) Protect from moisture 8 / 22
9 4 Startup and shut down 4 Startup and shut down 4.1 Inserting the adaptor into the PROPHYflex WARNING Hazard from non-sterile products. Infection hazard for care provider and patient. Before first use and after each use, prepare and sterilise the medical device and accessories accordingly. WARNING Dispose of the product in the appropriate manner. Prior to disposal, the product and accessories must be appropriately prepared or sterilised according to need. Dispose the perio tip like injection needles in a closed container. 4.1 Inserting the adaptor into the PROPHYflex The markings must coincide; otherwise the cannula can detach. If the adaptor detaches during treatment, it can give rise to a substantially hazard for the patient and user. Pull on the adaptor before each treatment and check if it is seated firmly. Before each treatment, make sure that the adaptor works properly. Insert the adaptor into the bayonet closure and lock it by turning it clockwise. Locking is successful of the two markings on the handpiece and the adaptor are situated directly across from each other. 4.2 Inserting the PROPHYflex perio tip into the wrench Slide the PROPHYflex perio tip into the wrench until it locks into place. 4.3 Screwing the PROPHYflex perio tip onto the adaptor using the wrench Hazard from detachment of the perio tip during the treatment. This can lead to a significant hazard for user and patient. Pull on the perio tip before each treatment and check if it is seated firmly. Before each treatment, make sure that the perio tip works properly. 9 / 22
10 4 Startup and shut down 4.4 Removing the PROPHYflex perio tip In order to screw it on, hold the wrench with 2 fingers in the round grasping area (not at the cover for the perio tip). Screw the PROPHYflex perio tip onto the adaptor in clockwise direction holding the wrench with 2 fingers. Pull the wrench off and check if the perio tip is seated firmly by pulling on it. 4.4 Removing the PROPHYflex perio tip Slide the wrench onto the PROPHYflex perio tip. Unscrew the PROPHYflex perio tip in counter-clockwise direction using the wrench. Carefully pull the PROPHYflex perio tip out of the wrench. Take the adaptor off Open the lock of the bayonet closure through a counter-clockwise rotation. Pull the adaptor out of the bayonet closure. 10 / 22
11 5 Operation 5 Operation 5.1 s for application 5.1 s for application For details of the application procedure, please refer to the perio tip User s. Preparation Insert the reduced PROPHYflex nozzle K (see image on the left, with level puncture) (Mat. no ). Add the powder Remove all residual cleansers from the device and container. Shake the refill package before adding powder. Add PROPHYflex Perio powder to the maximal filling level. Add the powder slowly to avoid the formation of dust. Screw the container filled with the PROPHYflex Perio powder onto the PROPHYflex. Settings on the powder jet unit Aim the jet nozzle at a wet washbasin from a distance of approx. 20 cm. Set the quantities of water and air/powder for treatment as described in the Instructions for Use of the unit. Never use the powder jet device without water since renders the mixture of air/powder more difficult to aspirate. Direct the jet into a washbasin until a homogeneous powder/water mixture is produced. Getting the patient ready Apply some Vaseline ointment to the patient's lips. This prevents the lips from becoming dry and cracking. Hook the small saliva ejector into the corner of the patient's mouth such that it aspirates underneath the tongue. Use the large suction cannula for aspiration of the water/powder mixture bouncing off the tooth. Treating the patient Indication Pockets with a probing depth of up to 5 mm Use 1-3 weeks after initial treatment with sonic, ultrasonic or hand-held instruments Root tip with at least 3 mm bony periodontium comprises Record X-ray image for clarification Contraindication Apical focus Root caries 11 / 22
12 5 Operation 5.1 s for application Treatment procedure: 1. Introduce the perio tip to the floor of the pocket, then pull the perio tip out by 1-2 mm. 2. Start-up the PROPHYflex with the foot control. 3. Using slight swinging motions, pull the perio tip out of the pocket and deactivate the foot control. 4. Spray the treated sites with anti-inflammatory agent twice daily for 1 to 4 weeks. The PROPHYflex perio tip must not be used in combination with powder for supragingival plaque. For details of the application procedure, please refer to the perio tip User s. Please advise your patients that the intake of foods (tea, coffee or food items) in the first 2-3 hours after the treatment may lead to discoloration of the teeth. The abrasion can be varied on devices in which the drive air can be adjusted with a foot switch. 12 / 22
13 6 Preparation methods according to ISO Preparations at the site of use 6 Preparation methods according to ISO WARNING Hazard from non-sterile products. Risk of infection to the care provider and patient. Before first use and after each use, prepare and sterilise the medical device. The reprocessing procedures described in the following apply to the PROPHYflex perio tip, adaptor and wrench. 6.1 Preparations at the site of use WARNING Hazard from nonsterile products. There is a risk of infection from contaminated medical devices. Take suitable personal protective measures. Do not place the perio tip in the drill bit bath since the fine capillaries will not be able to be rinsed under running water any longer and strong corrosion occurs. Reprocess the medical device right after treatment. Reprocess the PROPHYflex instrument according to the Instructions for Use of the manufacturer. 6.2 Preparations before reprocessing Use the wrench to remove the PROPHYflex perio tip from the adaptor. See also: Removing the PROPHYflex perio tip, Page 10 Hardened or damaged adaptor O-rings need to be replaced. 6.3 Non-fixing pre-cleaning of the PROPHYflex perio tip Accessories required: Demineralised water 30 C ± 5 C (86 o F +/- 10 o F) Nozzle pin Brush, e.g. medium-hard toothbrush Disposable syringe Brush the PROPHYflex perio tip with a medium-hard toothbrush under demineralised water. Check the patency of the PROPHYflex perio tip and if necessary clean it with the nozzle pin (Mat. no ). Rinse the PROPHYflex perio tip with at least 20 ml demineralised water using a disposable syringe. 13 / 22
14 6 Preparation methods according to ISO Cleaning If the PROPHYflex perio tip cannot be determined to be patent after the manual rinsing procedure or after completion of 10 reprocessing cycles, the medical device must be replaced. 6.4 Cleaning Malfunctions from cleaning in the ultrasonic unit. Defects in the device. Only clean in a thermodisinfector! Manual external and internal cleaning The PROPHYflex perio tip must not be subjected to manual external and internal cleaning. Following the non-fixing preliminary cleaning (item 6.3), the reprocessing must be continued in the thermodisinfector Automated external and internal cleaning Place the PROPHYflex perio tip on the servicing adaptor (Mat. no ) for reprocessing. Slide the PROPHYflex perio tip into the wrench until it locks into place. Screw the PROPHYflex perio tip onto the adaptor in clockwise direction holding the wrench with 2 fingers. Pull the wrench off. Attach the adaptor, with the PROPHYflex perio tip screwed into it, to the servicing adaptor (Mat. no ). 1 servicing adaptor, 2 adaptor, 3 PROPHYflex perio tip Plug the servicing adaptor (Mat. no ) onto the Miele adaptor. KaVo recommends thermodisinfectors in accordance with EN ISO , which are operated with alkaline cleaning agents with a ph value of max. 10 (e.g. Miele G 7781 / G 7881 Validation was carried out with Programme "VARIO-TD", cleaning agent "neodisher mediclean", neutralisation agent "neodisher Z" and rinsing agent "neodisher mielclear"). For program settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the thermodisinfector. 14 / 22
15 6.5 Disinfection 6 Preparation methods according to ISO Disinfection Malfunctioning from using a disinfectant bath or chlorine-containing disinfectant. Defects in the device. Disinfect in the thermodisinfector only! Manual external disinfection KaVo recommends the following products based on material compatibility. The microbiological efficacy must be ensured by the disinfectant manufacturer. Mikrozid AF made by Schülke & Mayr (liquid or cloths) FD 322 made by Dürr CaviCide made by Metrex Consumables required: Cloths for wiping off the medical device. Spray the disinfectant on a cloth, then thoroughly wipe down the medical device and leave the disinfectant to soak in according to the instructions from the disinfectant manufacturer. Follow the instructions for use of the disinfectant Manual disinfection - internal The PROPHYflex perio tip must not be subjected to manual internal disinfection. For effective reprocessing, automated internal cleaning in a washer/disinfector unit in accordance with ISO is required Machine disinfection - external and internal Place the PROPHYflex perio tip on the servicing adaptor (Mat. no ) for reprocessing. Slide the PROPHYflex perio tip into the wrench until it locks into place. Screw the PROPHYflex perio tip onto the adaptor in clockwise direction holding the wrench with 2 fingers. Pull the wrench off. Attach the adaptor, with the PROPHYflex perio tip screwed into it, to the servicing adaptor (Mat. no ). 1 servicing adaptor, 2 adaptor, 3 PROPHYflex perio tip Plug the servicing adaptor (Mat. no ) onto the Miele adaptor. 15 / 22
16 6 Preparation methods according to ISO Drying KaVo recommends thermodisinfectors in accordance with EN ISO , which are operated with alkaline cleaning agents with a ph value of max. 10 (e.g. Miele G 7781 / G 7881 Validation was carried out with Programme "VARIO-TD", cleaning agent "neodisher mediclean", neutralisation agent "neodisher Z" and rinsing agent "neodisher mielclear"). For program settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the thermodisinfector. 6.6 Drying Manual Drying Blow off the outside and inside with compressed air until water drops are no longer visible. Automatic Drying The drying procedure is normally part of the disinfection program of the thermodisinfector. Follow the instructions for use of the thermodisinfector. 6.7 Visual inspection and functional check Plug the adaptor with the PROPHYflex perio tip onto the PROPHYflex with an empty powder container, and check it. 6.8 Packaging The sterilisation bag must be large enough to accommodate the perio tip without stretching the bag. The quality and use of the sterilisation packaging must comply with applicable standards and be suitable for the sterilisation procedure! Individually seal the medical device in the sterilised item packaging. 6.9 Sterilisation Sterilisation in a steam steriliser (autoclave) in accordance with EN / ISO Contact corrosion due to moisture. Damage to product. Immediately remove the product from the steam steriliser after the sterilisation cycle! The KaVo medical device has a maximum temperature resistance up to 138 (280.4 F). 16 / 22
17 6 Preparation methods according to ISO Storage Select a suitable procedure (depending on the available autoclave) from the following sterilisation processes: Autoclave with three times pre-vacuum: - at least 3 minutes at 134 C -1 C / +4 C (273 F -1.6 F / +7.4 F) Autoclave using the gravity method: - at least 10 minutes at 134 C -1 C / +4 C (273 F -1.6 F / +7.4 F) Use according to the manufacturer's Instructions for Use Storage Reprocessed products should be stored protected from dust with minimum exposure to germs in a dry, dark and cool space. Comply with the expiry date of the sterilised items. 17 / 22
18 7 Tools 6.10 Storage 7 Tools Available from dental suppliers. Material summary Mat. No. Adaptor O-Ring adaptor Wrench Nozzle pin Refill PROPHYflex perio tip (10 units) Service adaptor / 22
19 8 Terms and conditions of warranty 8 Terms and conditions of warranty 6.10 Storage The following warranty conditions apply to this KaVo medical device: KaVo provides the end customer with a warranty of proper function and guarantees zero defects in respect of material and processing for a period of 10 reprocessing cycles / applications or 12 months from the date of the invoice, subject to the following conditions: In case of justified complaints, KaVo will honour its warranty with a free replacement or repair. Other claims of any nature whatsoever, in particular with respect to compensation, are excluded. In the event of default, gross negligence or intent, this shall only apply in the absence of mandatory legal regulations to the contrary. KaVo shall not be liable for defects and their consequences that have arisen or may arise from natural wear, improper handling, cleaning or maintenance, non-compliance with operating, maintenance or connection instructions, calcination or corrosion, contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in accordance with KaVo's instructions for use or other manufacturer's instructions. The warranty granted does not usually extend to lamps, optical fibres made of glass and glass fibres, glassware, rubber parts, and the colourfastness of plastic parts. All liability is excluded if defects or their consequences originate from manipulations or changes to the product made by the customer or a third party that is not authorised by KaVo. Warranty claims will only be accepted if the product is submitted along with proof of purchase in the form of a copy of the invoice or note of delivery. The dealer, purchase date, type, and serial number must be clearly evident from this document. 19 / 22
20 8 Terms and conditions of warranty 6.10 Storage 20 / 22
21
22 kb en
Instructions for use RONDOflex plus REF
Instructions for use RONDOflex plus 360 - REF 1.002.2179 Distributed by: KaVo Dental GmbH Bismarckring 39 D-88400 Biberach Phone +49 7351 56-0 Fax +49 7351 56-1488 Manufacturer: Kaltenbach & Voigt GmbH
More informationInstructions for use EXPERTmatic handpiece E10 C - REF
Instructions for use EXPERTmatic handpiece E10 C - REF 1.007.5560 Distributed by: KaVo Dental GmbH Bismarckring 39 D-88400 Biberach Phone +49 7351 56-0 Fax +49 7351 56-1488 Manufacturer: Kaltenbach & Voigt
More informationManufacturer s information
WARNINGS: Observe the standard German accident prevention regulations (UVV) We are not aware of any warnings if the instructions for the devices and disinfection and cleaning agents to be used are followed.
More informationMidwest E Motor. Directions For Use. Doc. No
Midwest E Motor Directions For Use CAUTION: US Federal law restricts this device to sale by or on the order of a licensed dental professional. Doc. No. 87513-0814 All trademarks referred to are the property
More informationINSTRUCTIONS FOR USE AND CARE
DIAMONDS INSTRUCTIONS FOR USE AND CARE DIAMONDS, MULTI-USE, FRICTION GRIP Quala Diamonds are a rotary cutting device made of stainless steel and coated with diamond particles on the working end. It is
More informationE-STATIS Motor. Operator's Manual. E-STATIS Operator s Manual Copyright 2009 SciCan. All rights reserved.
Motor Operator's Manual CAUTION: Federal law restricts this device to sale by or on the order of a dentist. www.scican.com Operator s Manual Copyright 2009 SciCan. All rights reserved. Doc. No. 2.000.3441
More informationE-STATIS. Operator's Manual. Motor. E-STATIS Operator s Manual Copyright 2009 SciCan. All rights reserved.
E-STATIS Motor Operator's Manual CAUTION: Federal law restricts this device to sale by or on the order of a dentist. www.scican.com E-STATIS Operator s Manual Copyright 2009 SciCan. All rights reserved.
More informationHygienic Reprocessing
Hygienic Reprocessing HEINE UniSpec Instrument head General warning and safety information: WARNING! This symbol draws attention to a potentially dangerous situation. Non-observance can result in moderate
More informationEndoscopic Technology. Operating Manual. Cannula system
Endoscopic Technology Operating Manual AlphaPort Cannula system This manual contains proprietary information that is protected by copyright. All rights are reserved. This manual or excerpts thereof may
More informationPreparation Instructions for the CAMLOG /CONELOG Implant System
J8000.0032 Rev.6 EN 06/2015 Preparation Instructions for the CAMLOG /CONELOG Implant System ENGLISH The following descriptions contain detailed instructions on cleaning, disinfection and sterilization
More informationTable of Contents. Introduction Indications For Use Contraindications Warnings Precautions...5
User Manual 3 Table of Contents Introduction....4 1. Indications For Use...4 2. Contraindications...4 3. Warnings...5 4. Precautions...5 5. Adverse Reactions...5 6. Step-By-Step Instructions...6 A. Contents...6
More informationmax airless Operating Manual for Dispenser Read this manual thoroughly before operating the Labmax airless!
Operating Manual for max airless Dispenser Read this manual thoroughly before operating the Labmax airless! witeg Labortechnik GmbH Am Bildacker 16 D-97866 Wertheim/Main Phone: +49 (0)9342 / 9301-0 Fax:
More informationbook Seite 1 Freitag, 21. April : fjmi^kq=omwn. Operating instructions. båöäáëü. lééê~íáåö=fåëíêìåíáçåë
6278761.book Seite 1 Freitag, 21. April 2017 3:54 15 fjmi^kq=omwn Operating instructions båöäáëü lééê~íáåö=fåëíêìåíáçåë 6278761.book Seite 2 Freitag, 21. April 2017 3:54 15 Table of contents Operating
More informationProUltra ULTRASONIC SURGICAL TIPS
ProUltra ULTRASONIC SURGICAL TIPS EN FOR DENTAL USE ONLY DIRECTIONS FOR USE (ULTRASONIC SURGICAL TIPS) A0640 A0650 Satelec M3 x 0,6 EMS M3 x 0,5 ProUltra Surgical tips Ref. A0640 Ref. A0650 1) INDICATIONS
More informationChapter 5: Checking and Maintaining Ultrasound Equipment
Chapter 5: Checking and Maintaining Ultrasound Equipment Care and Cleaning (BB1564-AJ) Chapter 5: Checking and Maintaining Ultrasound Equipment 45 Ultrasound equipment requires regular checks and maintenance.
More informationVistaCam ix Proof interchangeable head
VistaCam ix Proof interchangeable head Installation and Operating Instructions 9000-618-178/30 2014/07 Content Important information 1. About this document 4 1.1 Warnings and Symbols 4 1.2 Copyright notice
More informationCOMPRESSOR NEBULIZER MODEL: JB USER S MANUAL
COMPRESSOR NEBULIZER MODEL: JB0112-062 USER S MANUAL Read this manual before operating the nebulizer Save these instructions for future reference Caution: Federal law restricts this device to sale by or
More informationEVER Pharma D-mine Pen Pen injector for Apomorphine 10 mg/ml
EVER Pharma D-mine Pen Pen injector for Apomorphine 10 mg/ml Instructions for Use with Dacepton 3 ml Cartridges Apomorphine hydrochloride hemihydrate solution for injection Subcutaneous Use TABLE OF CONTENTS
More informationInstructions for Use. Contra-angle handpieces. Endea Endo Cursor EB-62 Endea EB-75 / EB-79. Contra-angle handpieces. Endo NiTi WD-73 M / WD-74 M
Instructions for Use Contra-angle handpieces Endea Endo Cursor EB-62 Endea EB-75 / EB-79 Contra-angle handpieces Endo NiTi WD-73 M / WD-74 M Contents W&H Symbols... 3 1. Introduction... 4 5 2. Before use
More informationENGLISH. USER MANUAL Register your product at: multi-action toothbrush
USER MANUAL Register your product at: www.acteh.com/register ENGLISH multi-action toothbrush To reorder toothbrush heads, please visit: www.acteh.com/heads 15% of your next order at www.acteh.com/store
More informationAS Medizintechnik GmbH Sattlerstraße Tuttlingen Germany
Before using the intramedullary. please ensure that it is functioning properly and free of any defects. In addition, please follow these instructions, including the information regarding repair as wellas
More informationE-type Straight Nose Cone
OPERATION MANUAL E-type Straight Nose Cone Please read this Operation Manual carefully before use and file for future reference. Caution This handpiece is intended for use by dental professionals only.
More informationStraumann ProClean Cassette. Basic Information on the
Straumann ProClean Cassette Basic Information on the Straumann ProClean Cassette Contents 1. Straumann ProClean Cassette Uncompromised hygiene 2 2. Straumann ProClean Cassette product at a glance 3 2.1
More informationStar ETorque TM Attachments User Manual
Attachments Star ETorque TM Attachments User Manual 1:1 Straight 5:1 AL-L Auto Latch 5:1 AL Auto Latch 1:1 AL-L Auto Latch 1:5 AC-L FG Autochuck Rx only ETorque Attachments Copyright 2017 DentalEZ. All
More informationMVKOMNT. kéï=~ë=çñw= bkal=swn. lééê~íáåö=fåëíêìåíáçåë. båöäáëü. Operating Instructions
kéï=~ë=çñw= MVKOMNT bkal=swn lééê~íáåö=fåëíêìåíáçåë båöäáëü Operating Instructions = Table of contents Sirona Dental Systems GmbH Table of contents 1 Before you begin... 4 1.1 Structure of the document...
More informationAltus Product description / Instructions for Use IFU 6 Rev. 1
Content: 1 Introduction... 2 1.1 Legend... 2 1.2 Indication... 2 1.3 Contraindication... 2 1.4 Field of Application... 2 1.5 Sterilization... 2 1.6 Storage... 2 1.7 Allergies... 2 2 SYSTEM DESCRIPTION...
More information2. Before Testing Monitor Checker Test...10 Inserting Lancets into Lancing Device...11 Quality Control Testing...13
Table of Contents 1. About your HemoSmart Haemoglobin Screening System Contents of Kit...4 HemoSmart Haemoglobin Meter...5 HemoSmart Haemoglobin Test Strip...7 Adjustable Lancing Device and Lancets...8
More informationSelectTech 3.1 Bench Assembly / Owner s Manual
SelectTech 3.1 Bench Assembly / Owner s Manual This product is compliant with the applicable CE requirements. Congratulations on your commitment to fitness and your purchase of the Bowflex SelectTech 3.1
More informationExclusively distributed by: RFE UK Ltd The Performance Centre Maidstone Road Kingston, Milton Keynes MK10 0BD, UK +44 (0)1908 793020 sales@goldsgymfitness.com www.goldsgymfitness.com 2010. All rights reserved.
More informationPhonak PilotOne II. User Guide
Phonak PilotOne II User Guide Contents 1. Welcome 4 2. Description 5 3. Using Phonak PilotOne II 6 3.1 Inserting a new battery 6 3.2 Switching On/Off 7 3.3 Holding PilotOne II 7 3.4 Changing hearing aid
More informationINSTRUCTIONS FOR USE
Comfort Injects 50 Comfort-in INSTRUCTIONS FOR USE English user manual 1 Introduction Thank you for choosing the Comfort in Needle free Injection system, developed with your comfort and convenience in
More informationUNICLIP DIRECTIONS FOR USE REF C226U - 0,8 MM / 1 MM FOR DENTAL USE ONLY. FISDR / F EN / 11 / updated 06/2016 1/7
UNICLIP FOR DENTAL USE ONLY DIRECTIONS FOR USE REF C226U - 0,8 MM / 1 MM EN FISDR / F19 02 31.EN / 11 / 2001 - updated 06/2016 1/7 1) Indications for use These products have to be used only in hospital
More informationDip Station G
Exclusively distributed by: RFE UK Ltd The Performance Centre Maidstone Road Kingston, Milton Keynes MK 0BD, UK +44 (0)190 793020 sales@goldsgymfitness.com 20. All rights reserved. Manufactured under license
More informationbook Seite 1 Mittwoch, 26. Juli : qq=ifkb=l=^áê=jçíçê. Operating Instructions / T4 LINE motor kit. båöäáëü. lééê~íáåö=fåëíêìåíáçåë
6397959.book Seite 1 Mittwoch, 26. Juli 2017 2:52 14 qq=ifkb=l=^áê=jçíçê Operating Instructions / T4 LINE motor kit båöäáëü lééê~íáåö=fåëíêìåíáçåë 6397959.book Seite 2 Mittwoch, 26. Juli 2017 2:52 14 Table
More informationDistributed by REF INSTRUCTIONS FOR USE
Distributed by REF 1600911-001 INSTRUCTIONS FOR USE Fig. 1 Fig. 2 Fig. 3 Fig. 4 Fig. 5 Fig. 6 TAblE Of contents I. MEANING OF SYMBOLS II. SCOPE OF USE III. GENERAL SAFETY INFORMATION IV. TECHNICAL CHARACTERISTICS
More informationReprocessing Guide. Autoclavable Arthroscope and Hardware Sterilization Tray
Reprocessing Guide Autoclavable Arthroscope and Hardware Sterilization Tray 0233032116 Contents Introduction...2 Intended Use of Sterilization Trays...4 Warnings...5 Cautions...5 Instructions...6 Sterilization
More informationPilotOne II. User Guide
PilotOne II User Guide Contents 1. Welcome 4 2. Description 5 3. Using Phonak PilotOne II 6 3.1 Inserting a new battery 6 3.2 Switching On / Off 7 3.3 Holding correctly 7 3.4 Changing hearing aid volume
More informationLow Row. User manual E S S E N T I A L S T R E N G T H
E L E M E N T E S S E N T I A L S T R E N G T H User manual 1 The identification plate of and manufacturer, affixed on the back panel of the weight stack, gives the following details: A B C D E Name and
More informationOPERATING INSTRUCTIONS. Orascoptic DK
User Manual Orascoptic DK is a versatile, 3-in-1 dental device that employs a handheld LED light source and three unique, interchangeable diagnostic instruments. Table of Contents 1. Orascoptic DK Kit
More informationinsafe System - Directions for Use
insafe System - Directions for Use 1. Sharps Container Fitting 1a Remove protective covers from the adhesive pads. 1b Position the base on a smooth clean worktop and push down firmly to secure. 1c Slide
More informationINSTRUCTION MANUAL. Orascoptic 3225 Deming Way, Suite 190 Middleton, WI USA T:
INSTRUCTION MANUAL Orascoptic 3225 Deming Way, Suite 190 Middleton, WI 53562 USA T: 800.369.3698 www.orascoptic.com ORASCOPTIC DK LED LIGHT SOURCE INSTRUCTIONS OPERATION AND FIRST TIME USE: Remove the
More informationSelectTech 4.1 Bench Assembly / Owner s Manual
SelectTech 4.1 Bench Assembly / Owner s Manual This product is compliant with the applicable CE requirements. Table of Contents Important Safety Instructions...3 Safety Warning Labels and Serial Number...4
More informationProUltra Ultrasonic Non Surgical Endo Tips
ProUltra Ultrasonic Non Surgical Endo Tips EN FOR DENTAL USE ONLY DIRECTIONS FOR USE (ULTRASONIC NON SURGICAL TIPS) A0620 A0621 A0630 A0631 Satelec M3 x 0,6 EMS M3 x 0,5 ProUltra Endo tips, Zirconium Ref.
More informationCompressor Nebulizer Instruction Manual Part No.: 5055
Compressor Nebulizer Instruction Manual Part No.: 5055 DISTRIBUTED BY: SAVE THESE INSTRUCTIONS. CAUTION - U.S. Federal Law restricts this device to sale by or on the order of a physician. 666002-6310 V1.3
More informationExclusively distributed by: RFE UK Ltd The Performance Centre Maidstone Road Kingston, Milton Keynes MK10 0BD, UK +44 (0)1908 793020 sales@goldsgymfitness.com www.goldsgymfitness.com 2010. All rights reserved.
More informationGoldspeed EVO S1 / S1-L / M5-L /
97050865 Rev. 00 15.05 Goldspeed EVO S1 / S1-L / M5-L / D1 / D1-L / R20-L / E16 ENGLISH..2 EN ENGLISH 1. INTENDED USE... 3 2. IMPORTANT WARNINGS... 3 3. SYMBOLS... 4 4. TECHNICAL SPECIFICATIONS... 4 5.
More informationAesculap Implant Systems
Aesculap Implant Systems Aesculap Implant Systems Spine USA Instructions for use/technical description Page 1 of 9 A 2 1 4 B 3 Page 2 of 9 USA Aesculap Implant Systems Instrument Legend A Rigid Inserter
More informationCA 1:1 EVO.15 1:1 CA 1:5 EVO.15 1:5 CA 10:1
CA 1:1 EVO.15 1:1 CA 1:5 EVO.15 1:5 CA 10:1 ENG Instructions for use other languages available on www.bienair.com/ifu 0120 REF 2100294-0000/2015.02 Set supplied (REF) CA 1:5 1600386-001 1600325-001 1600690-001
More informationInstruction Manual Aerogen, Inc. Part No. AG-AL1010 Rev. C
Instruction Manual 2004 Aerogen, Inc. Part No. AG-AL1010 Rev. C Table of Contents Introduction... 3 System description... 4 Warnings... 5 Cautions... 5 Electromagnetic Susceptibility... 5 Symbols... 6
More informationAdjustable Ab Board G
Exclusively distributed by: RFE UK Ltd The Performance Centre Maidstone Road Kingston, Milton Keynes MK10 0BD, UK + (0)10 73020 sales@goldsgymfitness.com 2010. All rights reserved. Manufactured under license
More informationConvex Array Transducer
User Guide Type 8823 Convex Array Transducer English BB1740-C June 2012 For Professional Users Only BK MEDICAL Mileparken 34 2730 Herlev Denmark Tel.:+45 4452 8100 / Fax:+45 4452 8199 www.bkmed.com Email:
More informationINSTRUCTIONS FOR CLEANING, DISINFECTION AND STERILIZATION OF DFS INSTRUMENTS
Preparation (cleaning, disinfection and sterilization) of reusable rotating dental instruments General principles All instruments must be cleaned, disinfected and sterilized before each use; this applies
More informationDear HighQ Check System Owner :
Dear HighQ Check System Owner : Thank you for purchasing the HighQ Check Blood Glucose Monitoring System. This manual provides important information to help you to use the system properly. Before using
More informationConfidence for you comfort for your patient
Clinpro Glycine Prophy Powder For supra- and subgingival cleaning Clinpro Glycine Prophy Powder with TCP For added sensitivity relief Confidence for you comfort for your patient f TCP functionalized Tri-Calcium
More informationSI-BONE ifuse Implant System Instruments Hospital Cleaning and Sterilization Instructions 0344
HOW SUPPLIED The SI-BONE instruments are supplied non-sterile and must be cleaned and sterilized prior to use. The ifuse Implant is provided separately and supplied sterile. Please refer to the ifuse System
More informationMANUAL FOR USE NO. 5 FLEXIBLE INTRAMEDULLARY REAMER REV ( GB ) non - sterile
MANUAL FOR USE NO. 5 FLEXIBLE INTRAMEDULLARY REAMER REV. 10.10.2013 ( GB ) non - sterile Before using the intramedullary reamer. please ensure that it is functioning properly and free of any defects. In
More informationSomnoDent Sleep Apnea Device. Instructions for use and care. SomnoDent Flex. SomnoDent Edent
TM SomnoDent Sleep Apnea Device Instructions for use and care SomnoDent Flex SomnoDent Classic SomnoDent Edent Contents Treatment with SomnoDent Contents 2 How the SomnoDent oral sleep device works 3 Instructions
More informationSelectTech 4.1 Bench. Assembly Manual
SelectTech 4.1 Bench Assembly Manual Table of Contents Important Safety Instructions...3 Safety Warning Labels and Serial Number...4 Specifications...5 Before Assembly...5 Parts...6 Hardware...7 Tools...7
More informationStratis INSTRUCTIONS FOR USE. Needle-free Injection System. 0.5mL volume (+/-5%)
Stratis Needle-free Injection System INSTRUCTIONS FOR USE 0.5mL volume (+/-5%) Stratis English Symbols Glossary (Note: All symbols are derived from ISO 15223-1, Medical Devices - Symbols to be used with
More informationScope This policy applies to all personnel and departments that clean, prepare and/or sterilize items intended for patient care use.
Dental Sterilization Procedures Policy Number VIM4(4)-10 Purpose The purpose of this policy is to ensure patient and employee safety when using instruments with potential for exposure to bloodborne pathogens
More informationGewässer Umwelt Schutz
Gewässer Umwelt Schutz GmbH CUW Oil protector Assembly instructions 2 Foreword Always study this assembly instruction manual before Installation and initial operation. This assembly instruction manual
More informationTHE TOTALLY WICKED TORNADO NC CONGRATULATIONS ON PURCHASING YOUR TOTALLY WICKED TORNADO NC ELECTRONIC CIGARETTE KIT
USER MANUAL THE TOTALLY WICKED TORNADO NC CONGRATULATIONS ON PURCHASING YOUR TOTALLY WICKED TORNADO NC ELECTRONIC CIGARETTE KIT 3 QUICK START GUIDE For those familiar with e-cigarettes, the steps below
More informationI n s t r u c t i o n M a n u a l
I n s t r u c t i o n M a n u a l TABLE OF CONTENTS 2.5THE STARTER KIT...4 3.USING THE INJEX SYSTEM...5 3.1 PREPARING THE INJECTOR...5 1. INTRODUCTION TO THE INJEX TM SYSTEM...1 1.1PRODUCT DESCRIPTION...1
More informationWEIGHT BENCH EXERCISER User s Manual
Model No. 831.15964.1 Serial No. Write the serial number in the space above for reference. WEIGHT BENCH EXERCISER User s Manual Serial Number Decal (under frame) Assembly Operation Maintenance Part List
More informationPRO-V IDS Kit. Bisco. Instructions for Use. Provisional Restorative System
Bisco PRO-V IDS Kit Provisional Restorative System Instructions for Use U.S. Patent: 7,748,980; 8,106,110 IN-199R3 Rev. 5/16 BISCO, Inc. 1100 W. Irving Park Rd. Schaumburg, IL 60193 U.S.A. 847-534-6000
More informationFor more information visit or contact hearx:
USER MANUAL hearscope - Ground Floor, Building 2, Ashlea Gardens Office Park, 180 Garsfontein Road, Ashlea Gardens, Pretoria, 0081, South Africa hearscope v2. HSCP-MN-EN hearscope IFU v1.0 For more information
More informationImplant Locator. User Manual
Implant Locator User Manual Table of Contents Introduction..... 2 1. Indications for use..... 3 2. Contraindications... 3 3. Warnings........ 3 4. Precautions........ 3 5. Adverse Reactions.... 3 6. Step-by-Step
More informationMultipower. User Guide E S S E N T I A L S T R E N G T H
E L E M E N T E S S E N T I A L S T R E N G T H User Guide The identification plate of and manufacturer, affixed to the frame behind the barbell rack, gives the following details: A B C D E Name and address
More informationHCS70004P, HCS70004G HCS70004C, HCS70004BL HCS70004DP, HCS70004 AEROMIST COLORS NEBULIZER COMPRESSOR KIT. Instruction Manual
HCS70004P, HCS70004G HCS70004C, HCS70004BL HCS70004DP, HCS70004 AEROMIST COLORS NEBULIZER COMPRESSOR KIT Instruction Manual TABLE OF CONTENTS IEC Symbols...2 Important Safeguards...2 Introduction...3 Specifications...4
More informationpiezosteril 6 - piezolight 6
97050228 Rev. 06 15.04 piezosteril 6 - piezolight 6 ENGLISH...2 EN ENGLISH 1. INTENDED USE... 3 2. IMPORTANT WARNINGS... 3 3. SYMBOLS... 4 4. TECHNICAL SPECIFICATIONS... 4 5. CONNECTION TO THE POWER SUPPLY
More informationM6-C Artificial Cervical Disc Surgical Instruments
CARE AND HANDLING INSTRUCTIONS Orthofix Inc. 3451 Plano Parkway Lewisville, Texas 75056-9453 U.S.A. 1-214-937-3199 1-888-298-5700 www.orthofix.com OSI-CustomerService@Orthofix.com Spinal Kinetics LLC,
More informationMULTI USE WORKOUT BENCH Assembling &User Manual
NOTE: Please ensure that you read this instruction carefully before attempting to assemble or use this product MULTI USE WORKOUT BENCH Assembling &User Manual Model :JX-503B Cat No.# 9283471 Model : JX-503B
More informationMedescan Nebuliser Med - S600A Instructions Manual
Medescan Nebuliser Med - S600A Instructions Manual Please read this guidebook carefully before operating this unit Your Nebuliser is intended for use in the treatment of asthma, COPD and other respiratory
More informationCONGRATULATIONS ON PURCHASING YOUR TOTALLY WICKED ARC 5 ELECTRONIC CIGARETTE KIT.
USER MANUAL YOUR ARC 5 BATTERY CS AIR SLIM TANK Tank connection Mouthpiece CONGRATULATIONS ON PURCHASING YOUR TOTALLY WICKED ARC 5 ELECTRONIC CIGARETTE KIT. UP adjustment button DOWN adjustment button
More informationMATERIAL SAFETY DATA SHEET (MSDS) Information about stainless steel for Surgical/Dental/Orthodontic Instruments.
MATERIAL SAFETY DATA SHEET (MSDS) THE INFORMATION BELOW IS BELIEVED TO BE ACCURATE AND REPRESENTS THE BEST INFORMATION CURRENTYLY AVAILABLE TO US. HOWEVER, WE MAKE NO WARRANTY OF MERCHANTABILITY OR ANY
More informationBlood Glucose Monitoring System. User Guide
Blood Glucose Monitoring System User Guide Table of Contents Introduction...2 Important Safety Instructions...2 About ipet PRO Blood Glucose Monitoring System...3 About ipet PRO Meter...4 About the ipet
More information1) INDICATIONS FOR USE These products have to be used only in hospital environments, clinics or dental offices by qualified dental personnel.
ProFile 04/06 & O.S. EN DENTAL USE ONLY DIRECTIONS FOR USE PROFILE 0) COMPOSITION The cutting part of these instruments is made of a nickel-titanium alloy. 1) INDICATIONS FOR USE These products have to
More informationVelocity Orthopedics Instrument Cleaning and Sterilization
Velocity Orthopedics Instrument Cleaning and Sterilization It is important to read the Instructions For Use in its entirety prior to using the product. Caution: Federal law (USA) restricts this device
More informationUSER S MANUAL QUESTIONS? CAUTION. Visit our website at. new products, prizes, fitness tips, and much more!
Model No. WEBE44.0 Serial No. Write the serial number in the space above for future reference. USER S MANUAL Serial Number Decal QUESTIONS? As a manufacturer, we are committed to providing complete customer
More informationPROFESSIONAL CARE OXYJET air. powered. Downloaded from manuals search engine. filter. OxyJet
PROFESSIONAL powered CARE by OXYJET+ 3000 600 500 air filter 5 400 300 200 ml OxyJet 6 English 9 KR If you should require further assistance, please call our Consumer Service line. Helpline GB 0 800 783
More informationTJF-160F/VF Cleaning and Disinfection Checklist
TJF-160F/VF Cleaning and Disinfection Checklist TJF-160F/VF Cleaning and Disinfection Checklist This checklist is used to evaluate and confirm if cleaning and disinfection of the TJF-160F/VF has been performed
More informationDr FuelCell Load Measurement Box
1 (855) 251-0016 sales@fuelcellstore.com Dr FuelCell Load Measurement Box Instruction Manual www.fuelcellstore.com Instruction Manual for Dr FuelCell TM Load Measurement Box Version 1.0.1 November 2008
More informationThe Multi-unit Abutments of the CE-Series are indicated for Nobel Biocare / NobelReplace Tapered. The current implant diameters have to be attended.
Instructions for Use Multi-Unit Abutment with Screw Description: Multi-Unit Abutment with screws for series CC-Series, CE-Series, CV- Series, CF-Series, CH-Series, CK-Series, CL-Series, CN-Series, CR-Series,
More informationConvex Array Transducer
User Guide Type 8802 Convex Array Transducer English BB0334-I August 2012 For Professional Users Only WORLD HEADQUARTERS Mileparken 34 DK-2730 Herlev Denmark Tel.:+45 44528100 / Fax:+45 44528199 www.bkmed.com
More informationVarseoWax CAD/Cast GEBRAUCHSANWEISUNG INSTRUCTIONS FOR USE MODE D EMPLOI INSTRUCCIONES DE USO ISTRUZIONI PER L USO Partners in Progress
www.bego.com VarseoWax CAD/Cast GEBRAUCHSANWEISUNG INSTRUCTIONS FOR USE MODE D EMPLOI INSTRUCCIONES DE USO ISTRUZIONI PER L USO Partners in Progress EN INSTRUCTION FOR USE VarseoWax CAD/Cast VarseoWax
More informationTHE. Curve THE CURVE STRIPPING TOOL
THE Curve THE CURVE STRIPPING TOOL Innovative Product Made in Germany THE Curve Mastering the perfect curve Curvable polishing strips for interproximal reduction Sterilizable, double-sided diamond instruments
More informationD15525 D15513 D15513T D15511
Oral-B Professional Care D15525 D15513 D15513T D15511 timer Professional Care powered by Type 4729 Type 4736 KR If you should require further assistance, please call our Consumer Service line. GB IRL Helpline
More informationA82-LIFE-Dog Nebulizer NBB02
A82-LIFE-Dog Nebulizer NBB02 פיית אוויר A82-LIFE-Dog Nebulizer NBB02 www.mfcl.co.il COMPRESSOR NEBULIZER USER S MANUAL A82-LIFE-Dog Nebulizer NBB02 ILLUSTRATION OF PARTS A. NEBULIZER COMPRESSOR
More informationINSTALLATION MANUAL. VIDEO Camera, Probe and Lightsource OTOSCOPES.
INSTALLATION MANUAL VIDEO Camera, Probe and Lightsource OTOSCOPES www.medrx-int.com Contents Using The Video Otoscope... 3 The Battery Operated LED Light Source... 4 Wiring Diagram - Battery Operated LED
More informationHigh Frequency Linear Array Transducer
User Guide Type 8870 High Frequency Linear Array Transducer English BB1980-C May 2016 For Professional Users Only BK MEDICAL Mileparken 34 2730 Herlev Denmark Tel.:+45 4452 8100 / Fax:+45 4452 8199 www.bkultrasound.com
More informationTABLE OF CONTENTS 1.1. WARRANTY DAMAGED SHIPMENTS SERVICE INTRODUCTION UNPACKING AND INSTALLATION
TABLE OF CONTENTS EDITION 5 Section 1 - WARRANTY & SERVICE 1.1. WARRANTY 2 1.2. DAMAGED SHIPMENTS 2 1.3. SERVICE 2 Section 2 - INTRODUCTION 2.1. INTRODUCTION 3 Section 3 - GENERAL 3.1. UNPACKING AND INSTALLATION
More informationFlex-Point HEATED MUSCLE MASSAGER
Flex-Point HEATED MUSCLE MASSAGER TABLE OF CONTENTS Cautions and Warnings...1 FCC Information...5 Location of Controls...6 Operation...6 Using the Node Covers...7 Care and Maintenance...7 Specifications...7
More informationNovoPen Echo User guide
NovoPen Echo User guide Read this user guide carefully before you use your NovoPen Echo for insulin delivery the first time Find a quick guide at the back of this manual NovoPen Echo Pen Insulin window
More informationInstructions for use. 400 series. Highspeed Air Handpieces. valid since serial no
Instructions for use 400 series Highspeed Air Handpieces valid since serial no. 01001 Contents A-dec W&H symbols... 3 4 1. Introduction... 5 7 2. Safety notes... 8 11 3. Product description...12 13 4.
More informationCURVES. Anatomically Adaptable IPR strips. Brochure Instructions for use Processing instructions
CURVES Anatomically Adaptable IPR strips Brochure Instructions for use Processing instructions Visit us online! To learn more about all our orthodontic products plus many helpful user videos Innovative
More informationALPROLIX Coagulation Factor IX (Recombinant), Fc Fusion Protein INSTRUCTIONS FOR USE Do not Do not YOUR KIT CONTAINS:
ALPROLIX Coagulation Factor IX (Recombinant), Fc Fusion Protein INSTRUCTIONS FOR USE Read the Instructions for Use before you start using ALPROLIX and each time you get a refill. There may be new information.
More informationCompatible with the following products TENS. Replacement Pads. User Manual. Last revised: V Last revised: V
Compatible with the following products TENS Replacement Pads User Manual Last revised: V11-160923 Last revised: V4-170619 Contents Introduction 2-3 Parts 3 General Warnings and Safety 4-10 Using your TENS
More informationIMPORTANT SAFETY INSTRUCTIONS
IMPORTANT SAFETY INSTRUCTIONS When using an electrical appliance, basic precautions should always be followed, including the following. Read all instructions before using this appliance. DANGER - To reduce
More informationLOCATOR F-Tx SEATING AND REMOVAL TOOL PROCEDURES ADDENDUM
1 LOCATOR F-Tx SEATING AND REMOVAL TOOL PROCEDURES ADDENDUM INSTRUCTIONS FOR USE DESCRIPTION Wire and Lever Wrench Tip Wrench Seating and Removal Tool Seating Tip Removal Tip with Threaded Adapter The
More informationCompressor Nebulizer. Model UN-014
Compressor Nebulizer Model UN-014 Instruction Manual Original Manuel d instructions Traduction Manual de Instrucciones Traducción Manuale di Istruzioni Traduzione ITALIANO ESPAÑOL FRANÇAIS ENGLISH 1WMPD4003398
More information