Quality Control for Point-of-Care Testing

Size: px
Start display at page:

Download "Quality Control for Point-of-Care Testing"

Transcription

1 Quality Control for Point-of-Care Testing James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology and Immunology Medical Director of Clinical Chemistry and Point-of-Care Testing Vanderbilt University School of Medicine Medical Director, Clinical Chemistry Nashville, Tennessee, USA 1

2 Objectives 1. Review quality control (QC) as a means of controlling laboratory error 2. Identify risk management as a total quality assurance of the testing process 3. Describe regulatory changes adopting risk management in the laboratory 2

3 Historical Quality Control Quality control historically used to reduce errors and prevent the release of incorrect results from the laboratory! Born from the 1950 s industrial model of quality in analytical and factory processes Quality control is a stabilized surrogate sample analyzed like a patient sample containing a known amount of measured analyte. If the analytical test system can achieve the desired result using a QC sample, then the system is stable and quality patient results are assumed to be produced. 3

4 Quality Control Advantages QC has target values, if assay recovers target, then system is assumed to be stable (instrument, reagent, operator, sample) Shift in QC = Shift in quality patient results! Warns of test result problems before release to physician 4

5 Cape Clinic Laboratory L-J Chart for XYZ Chemistry Analyzer Analyte: Glucose Control Material: Level 1 Units: mmol/l Lot #: Exp Date: 19/4/20XX 3.48 SD assign 0.05 From: June 1, 20XX Through: June 30, 20XX Target Value 3.39 TEa 6.90% SD C o n t r o l V a l u e SD + 1 SD X - 1 SD SD SD Run

6 Quality Control Disadvantages If QC not analyzed frequently enough or lab fails to take action when QC fails, quality of patient results affected! With auto-verification or continuous release of results by automated analyzers - Possible to release patient results before a problem is detected When a problem is detected, lab must reanalyze patients since last good QC and possibly send out corrected results. Clinical action could be taken on erroneous results before correcting an unrecognized issue! Despite its limitations, QC remains an important means of detecting and preventing laboratory errors! 6

7 Systematic Errors Liquid QC does a good job at detecting systematic errors, like: Reagent deterioration or preparation Improper storage or shipment conditions Incorrect operator technique (dilution, pipette setting) Calibration errors wrong setpoint, factors Errors which affect every test in a constant and predictable manner Errors that occur from one point forward or for a limited period of time 7

8 Random Errors Liquid QC does a poor job at detecting random errors unless the error specifically occurs with the QC sample or increased result variability is noted. Errors which affect individual samples in a random and unpredictable fashion, like: Clots Bubbles Interfering substances 8

9 QC and Lab Errors QC Hemolyzed sample QC 09:00 11:00 01:15 09:00 Line Leak QC daily at 09:00, two levels Line leak occurs at 11:00, dispenses partial reagent causing 25% decrease in values, not detected until next day QC at 09:00. (Systematic Error), must reanalyze all specimens since previous QC at 09:00, the previous day. Hemolyzed specimen (Random error) affects one sample 9

10 History of Quality Control Quality control has been traditional means of managing error in the laboratory! CLIA 67 required daily QC as integral part of testing process CLIA 88 - reaffirmed two levels of QC per test per day (blood gases and coagulation testing every 8 hours) CAP advocates daily QC The Joint Commission focuses on patient safety, daily QC and preanalytic, analytic and postanalytic processes 10

11 ISO QC Recommendations ISO The laboratory shall design internal control systems that verify the attainment of the intended quality of results. Special attention should be paid to elimination of mistakes in the process of handling samples, requests, examinations, reports, etc. Documentation should include quality control procedures based on manufacturer instructions for use. Internal Quality Control (internal to the laboratory) is defined as a set of procedures undertaken by laboratory staff for the continuous monitoring of operation and the results of measurements in order to decide whether results are reliable enough to be released. The regular analysis of QC materials can serve as an essential component of a laboratory s internal control system. 11

12 Quality in Laboratory Testing Despite analyzing the minimum required 2 levels QC/day, we have all experienced erroneous laboratory results! Quality is more than QC! Quality is: Testing that is safe and reliable Appropriate use of technology to meet medical needs Test results that can be trusted for medical management A lab s reputation depends on quality of its results! What should be the lab goals for errors? 12

13 Sigma Metrics Sigma metrics are an industrial measure of quality in a factory process - estimates rate of defects in products Goal of sigma metrics: Improve customer satisfaction Balance number of rejected products with the increased cost required to improve processes (to lower defect rates) Six sigma consensus for industrial processes Six sigma was adopted and popularized into the clinical laboratories in 1990 s 13

14 Sigma Performance

15 Need for Quality Laboratory Testing Laboratory testing influences >70% of all medical decisions. >7 billion lab tests conducted in the U.S. annually Silverstein MD. An approach to medical errors and patient safety in laboratory services. A white paper prepared for the Quality Institute Meeting, Making the Laboratory a Partner in Patient Safety. Division of Laboratory Systems, Centers for Disease Control and Prevention, April

16 Beyond Six Sigma With > 7 billion tests conducted in the US annually, six sigma (3.4 defects/million tests) means releasing 23,800 incorrect lab tests annually in the US! Any single result error could mean a misdiagnosis, failure to treat an illness or incorrect treatment selection! Virus contamination of blood product, Incorrect cancer diagnosis Missed heart attack! What if that wrong test result belonged to you or a family member? For a medical laboratory, six sigma is an unacceptable goal! We need to go beyond six sigma for laboratory performance, with a goal of ZERO errors that reach the patient! The key is risk management! 16

17 Identifying Risk in the Laboratory There is no perfect device! Any device can and will fail under the right conditions Every device is different! (universal 2 levels QC/day doesn t fit every situation, may need more frequent QC especially if frequent staff turnover or unstable reagent) Any discussion of risk must start with what can go wrong with a test (errors) Lab tests are not fool-proof, for a sufficiently talented fool!

18 CLSI Project: EP23 Laboratory Quality Control Based on Risk Management. James H. Nichols, Ph.D., Chairholder EP23 describes good laboratory practice for developing a quality control plan based on manufacturer s information, applicable regulatory and accreditation requirements, and the individual healthcare and laboratory setting 18

19 EP23 Laboratory QC Based on Risk Management Input Information Medical Requirements for Test Results Regulatory and Accreditation Requirements Test System Information: Provided by the manufacturer Obtained by the Laboratory Information about Health Care and Test- Site Setting Process Risk Assessment Continuous Improvement Output Laboratory Director s QC Plan Post Implementation Monitoring CLSI EP23 Table 19

20 Where is the Risk in the Process? What Could Possibly Go Wrong? 20

21 Identifying Risk in the Laboratory Follow the sample through the testing process! At each step, ask What could possibly go wrong? For each hazard identified, define a mitigation step, a way to minimize the risk of error occurring! Summarize the hazards and actions in a quality control plan! 21

22 Steps of the Testing Process Order Treatment Treatment Postanalytical Patient Need for Test Preanalytical Order Patient Preparation Acknowledge Result Analytical Specimen Collection Processing Result Verification Analysis Lab Receipt Transport to Lab

23 1 Samples 2 Operator 4 Laboratory Environment Sample Integrity - Lipemia - Hemolysis - Interfering subtances - Clotted - Incorrect tube Sample Presentation - Bubbles - Inadequate volume Identify Potential Hazards Reagent Degradation Operator Capacity - Shipping - Storage - Used past expiration - Preparation Quality Control Material Degradation - Shipping - Storage - Used past expiration - Preparation - Training - Competency Operator staffing 3 Reagents - Short staffing - Correct staffing Atmospheric Environment - Dust - Temperature - Humidity Utility Environment - Electrical - Water quality - Pressure Calibrator Degradation - Shipping - Storage - Use past expiration - Preparation 5 Measuring System Instrument Failure - Software failure - Optics drift - Electronic instability Incorrect Test Result Inadequate Instrument Maintenance - Dirty optics - Contamination - Scratches Fishbone Diagram from CLSI EP23 Refer to Appendix A in CLSI EP18 for more comprehensive list of error sources

24 Risk Management Holistic or global approach to quality assurance incorporating preanalytic, analytic and postanalytic phases of testing AND considers the major sources of laboratory error Every instrument or device is different Hazard analysis and mitigation Hazard analysis defines the sources of potential error for an instrument or device and a means of detecting or preventing the error (mitigation) 24

25 Manufacturer Checks Alternative QC Some devices have internal checks which are performed automatically with every specimen: Development of a line (Pregnancy test, Occult blood) Sensor signal (blood gas analyzer, clots) Flow resistance and liquid sensors (clots or bubbles in chemistry and hematology analyzer pipettes) Other checks engineered into instrument: Barcoding of reagent expiration dates (prevents use) Lockout features that require successful QC Disposable analyzer cuvettes/pipette tips (carry-over) Temperature indicator in shipping carton 25

26 Unit-Use (Single Use) POCT QC Device Test Cartridge Chemistry of reaction Reads test cartridge signal Internal electronic checks maintain reader in narrow range Locks out testing if outside specs Acts as analyzer contains fluidics, calibrant, int controls QC consumes test, no guarantee next test performs same Additional control processes needed for specimen and operator technique 26

27 QC and Risk Management QC is expensive, consumes POCT reagent and labor without producing results that can be billed. The presence of engineered control processes on newer POCT devices raise the question of when and how these controls can substitute for two levels of liquid QC? How to best control unit-use POCT devices? Lab Director must balance requirements for QC (2 levels/day) against cost and availability of alternative controls on the test system Risk management steps through the testing process to identify sources of error that can help labs identify the right balance of liquid QC with built in control processes on the POCT devices 27

28 Individualized Quality Control Plan Risk Assessment Quality Control Plan Quality Assessment Individualized Quality Control Plan 28 CLIA

29 29

30 New IQCP CLIA, CAP and Joint Commission Requirements Two levels of liquid QC required each day of testing OR Laboratory develops an Individualized Quality Control Plan (IQCP): Balance built-in control processes with liquid QC Reduce frequency of liquid QC to minimum recommended by manufacturer Maximize clinical outcome, available staff resources and cost effectiveness in the lab Considers the laboratory process for testing and risk of errors at each step of the testing process! 30

31 What Have We Learned From Our IQCPs? 31

32 What Have We Learned From Our IQCPs? Processes on different units were not uniform Some barcoded BG bedside, others waited to satellite lab IQCP supports QC rationale and resources Each action is linked to a specific hazard Gives meaning for why we do what we do rather than simply meeting a regulation Opportunity for improving efficiency QC the device versus QC the reagent (i-stat) Multi-site validations of reagent shipments Monthly 3 level QC versus 6 month cal verifications 32

33 What Have We Learned From Our IQCPs? Before: (QC the device) Shipments = 10 shipments/yr x 2 QC x 7 sites = 140 tests Lot validations = 5 x/yr x 2 levels x 8 meters = 80 tests QC monthly = 2 QC x 8 i-stats x 12 mos = 192 tests 6 mo cal-ver = 8 i-stats x 3 levels x 3 reps x 2x/yr = 144 tests 6 mo correlations = 10 patients x 8 i-stats x 2x/yr = 160 tests TOTAL = 716 tests After: (QC the cartridge lots) Shipments = 4 shipments/yr x 3 QC x 1 site = 12 tests Lot validations = QC shipment, max 4x/yr x 5 pts x 2(old/new) 40 tests QC monthly = 3 QC x 7 sites x 12 mos = 252 tests If additional lot: 3 QC x 7 sites x 4 mos 84 tests 6 mo cal ver and pt correl already done monthly QC/lot val = 0 tests TOTAL = 304/(388) tests 33 Savings of nearly half each year!

34 What Have We Learned From Our IQCPs? Improved quality - Operators now perform all the required testing before the POCT staff would analyze linearities and perform 6 mo comparisons! Enhanced efficiency fewer cartridges required for non-patient testing, saves cost and resources Better quality assurance of cartridges QC every lot of cartridges monthly (the i-stat has internal checks)! 34

35 Falsely Decreased Glucose Meter Results Complaint from an intensive care unit of sporadic falsely decreased glucose meter results Immediate repeat test on same meter, gave significantly higher clinically sensible values Inspection of unit found nurses taking procedural shortcuts to save time Bottles of test strips dumped on counter in spare utility room Some strips not making it into trash, falling back on counter and being REUSED 35

36 Risk of Error from Open Reagents Glucose test strips exposed to air for as little as 2 hours have been shown to cause -26% bias. 1 Strips left on counters pose risk of reuse, leading to falsely low results. Some meters catch reuse and error preventing a result. Other meters do not! 2 1. Keffer P, Kampa IS. Diabetes 1998; 47; abs Silverman BC, Humbertson SK, Stem JE, Nichols JH. Operational errors cause inaccurate glucose results. Diabetes Care 2000;23:

37 Reagent Errors: Expired Reagents Centers for Disease Control Check and record expiration dates of reagents/kits, and discard any reagents or tests that have expired. 1 U.S. Food and Drug Administration Check the expiration date on the test strips. As a test strip ages, its chemical coating breaks down. If the strip is used after this time, it may give inaccurate results Ready? Set? Test! Centers for Disease Control booklet 2. Useful Tips to Increase Accuracy and Reduce Errors in Test Results from Glucose Meters, U.S. Food and Drug Administration 37

38 Strip Wastage When Outdated Operator must check manufacturer s expiration date prior to testing. Vials/strips and controls must be manually dated when opened by operator (prematurely expires once opened) Undated, opened vials must be discarded. (? expiration) Discarded strips due to no date 1 1. Undated vials between September, 2010 and May, 2011, Willis- Knighton Medical Center, Shreveport, Louisiana 38

39 Reagent Errors: Expired Reagents Serialized vials/strips and controls barcoded for lot number and expiration date (good to stamped expiration date) can recognize individual vials on opening (30, 60 or 90 day open expiration) Automatic lockout for expired test strips and controls Some devices can also recognize exposure to humidity (few hours), wet or reused strips as additional control measure 39

40 Manufacturer Mitigations Operator training/competency Operator lock-out Failure to perform QC QC lock-out Result documentation Data management/wireless Patient ID errors barcoding/positive patient ID Mixing sample bubbles/clots clot detection Overfill/Underfill Specimen trigger detection Calibration errors auto calibration with lot barcode 40

41 Resources for Reducing Errors Clinical Chemistry book recently released! Focus on errors in the Chemistry Laboratory including POCT Discussion of real-world errors and what can be done to manage risk by detecting and preventing errors. 41

42 Summary Many POCT devices have built-in control processed POCT requires different approach to QC Risk management is a process to reduce error in the clinical laboratory that can help develop an individualized QC plan for POCT CLSI EP23 is a resource that introduces industrial risk management principles to the laboratory We have reduced errors, saved cost, improved staff efficiency, and enhanced patient care by developing a quality control plan based on risk management for our POCT systems 42

Individual Quality Control Plans (IQCP) Risk Assessment

Individual Quality Control Plans (IQCP) Risk Assessment Individual Quality Control Plans (IQCP) Risk Assessment James H. Nichols, PhD, DABCC, FACB Professor of Clinical Pathology, Microbiology and Immunology Medical Director of Clinical Chemistry and Point-of-Care

More information

Errors in POCT: When Improbable Situations Become Possible

Errors in POCT: When Improbable Situations Become Possible Errors in POCT: When Improbable Situations Become Possible James H. Nichols, Ph.D., DABCC, FACB Chairholder CLSI EP23 Document Development Committee Director, Section of Clinical Chemistry Department of

More information

Patient Safety: A Quality System Approach To POCT QC/QA

Patient Safety: A Quality System Approach To POCT QC/QA Patient Safety: A Quality System Approach To POCT QC/QA Ellis Jacobs, Ph.D., DABCC New York University School of Medicine Coler-Goldwater Specialty Hospital & Nursing Facility New York, New York Point-of-Care

More information

The Virtues and Pitfalls of Implementing a New Test

The Virtues and Pitfalls of Implementing a New Test The Virtues and Pitfalls of Implementing a New Test James H. Nichols, Ph.D., DABCC, FACB Professor of Clinical Pathology, Microbiology and Immunology Associate Medical Director for Clinical Operations

More information

Patient Safety: A Quality System Approach To POCT QC/QA

Patient Safety: A Quality System Approach To POCT QC/QA Patient Safety: A Quality System Approach To POCT QC/QA Ellis Jacobs, Ph.D., DABCC New York University School of Medicine Coler-Goldwater Specialty Hospital & Nursing Facility New York, New York Point-of-Care

More information

It s a Gas! Issues in the Blood Gas Laboratory. D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC

It s a Gas! Issues in the Blood Gas Laboratory. D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC It s a Gas! Issues in the Blood Gas Laboratory D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC Learning Objectives After participating in this session, you will be able

More information

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009 Basic knowledge of POCT instuments The Tide Resort Bangsan Beach 18 June 2009 1 POCT instruments concepts Point of care testing Diagnostic medical devices used at or near the site of patient care Bed site,

More information

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved: PROCEDURE TITLE: Bedside Glucose Monitoring Issuing Department: Clinical Director Signature: Departments Involved: Laboratory Nursing Effective Date: 10/97 Review Dates: 09/01, 07/02, 05/13 Revision Dates:

More information

LeadCare BLOOD LEAD ANALYZER. Quick Reference Guide

LeadCare BLOOD LEAD ANALYZER. Quick Reference Guide LeadCare II BLOOD LEAD ANALYZER Quick Reference Guide Precautions Precautions Caution The LeadCare II Blood Lead Analyzer is a CLIA-waived device. Facilities that perform tests with the LeadCare II System

More information

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Polymer Technology Systems, Inc. CardioChek PA Comparison Study Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory

More information

Point of Care Testing (POCT)

Point of Care Testing (POCT) Accounting for Pre-analytical Variables in POCT L.V. Rao, PhD Senior Clinical Laboratory Director Associate Professor, Pathology UMass Memorial Medical Center (Worcester, MA) Senior Scientific Director

More information

Tips for Training New Customers

Tips for Training New Customers Tips for Training New Customers Good Habits: Test Strips Recap vial immediately and tightly after removing a single test strip. Never remove the desiccant from the vial. Do not lay test strips out ahead

More information

2014 Continuing Compliance Master Series Best Practices in Alternative Assessment of Performance

2014 Continuing Compliance Master Series Best Practices in Alternative Assessment of Performance 2014 Continuing Compliance Master Series Best Practices in Alternative Assessment of Performance Brad S. Karon, MD, PhD, FCAP November 19, 2014 www.cap.org Today s Presenter Brad S. Karon, MD, PhD, FCAP

More information

NEPHROCHECK Calibration Verification Kit Package Insert

NEPHROCHECK Calibration Verification Kit Package Insert NEPHROCHECK Calibration Verification Kit Package Insert Manufactured for Astute Medical, Inc. 3550 General Atomics Ct. Building 2 San Diego, CA 92121 USA Intended Use The NEPHROCHECK Calibration Verification

More information

Laboratory Procedure Manual

Laboratory Procedure Manual Laboratory Procedure Manual Analyte: Matrix: Method: as performed by: Fasting Glucose Plasma Hexokinase-mediated reaction University of Minnesota Medical Center, Fairview Collaborative Studies Clinical

More information

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide LESSON 6-3 Laboratory Reagent Preparation and Calculations Worksheet LESSON 6-4 Chemistry Instrumentation in the Physician Office Laboratory Worksheet LESSON 6-6 Measuring Blood Glucose Worksheet LESSON

More information

SAMPLE. Glucose Monitoring in Settings Without Laboratory Support

SAMPLE. Glucose Monitoring in Settings Without Laboratory Support POCT13c Glucose Monitoring in Settings Without Laboratory Support This guideline focuses on performance of point-of-care glucose monitoring systems, with an emphasis on safety practices, quality control,

More information

Method Comparison Report Semi-Annual 1/5/2018

Method Comparison Report Semi-Annual 1/5/2018 Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community

More information

Procine sphingomyelin ELISA Kit

Procine sphingomyelin ELISA Kit Procine sphingomyelin ELISA Kit For the quantitative in vitro determination of Procine sphingomyelin concentrations in serum - plasma - celiac fluid - tissue homogenate - body fluid FOR LABORATORY RESEARCH

More information

Bovine Insulin ELISA

Bovine Insulin ELISA Bovine Insulin ELISA For quantitative determination of insulin in bovine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSBO-E01 Size: 96 wells Version:

More information

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff

More information

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date: BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET Revision date: 2017-02-24 BÜHLMANN LABORATORIES AG Baselstrasse 55 4124 Schönenbuch, Switzerland Tel.:

More information

METHOD VALIDATION: WHY, HOW AND WHEN?

METHOD VALIDATION: WHY, HOW AND WHEN? METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test

More information

Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan

Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan 729488DSTXXX10.1177/1932296817729488Journal of Diabetes Science and TechnologyWestgard et al research-article2017 Original Article Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan Journal

More information

INTERNAL QUALITY CONTROL ( Q I C) QC)

INTERNAL QUALITY CONTROL ( Q I C) QC) EXTERNAL QUALITY ASSESSMENT PROGRAM (EQAP) BIOCHEMISTRY DEPARTMENT R. Mohammadi Biochemist (Ph.D.) Faculty member of Medical Faculty TWO COMPLEMENTARY COMPONENETS OF TQM ARE Internal Quality Control (IQC)

More information

C30-A2 ISBN ISSN Point-of-Care Blood Glucose Testing in Acute and Chronic Care Facilities; Approved Guideline Second Edition

C30-A2 ISBN ISSN Point-of-Care Blood Glucose Testing in Acute and Chronic Care Facilities; Approved Guideline Second Edition ISBN 1-56238-471-6 ISSN 0273-3099 Point-of-Care Blood Glucose Testing in Acute and Chronic Care Facilities; Approved Guideline Second Edition Volume 22 Number 17 David B. Sacks, M.D., Chairholder Patricia

More information

Point of Care Testing for INR using CoaguChek XS Plus

Point of Care Testing for INR using CoaguChek XS Plus Point of Care Testing, Pathology Page 1 of 18 Point of Care Testing for INR using CoaguChek XS Plus EDITION No 1.4 DATE OF ISSUE Feb 2014 REVIEW INTERVAL AUTHOR LOCATION OF COPIES 3 YEARS D O Neill 1.

More information

HbA1c (Human) ELISA Kit

HbA1c (Human) ELISA Kit HbA1c (Human) ELISA Kit Cat. No.:DEIA3509 Pkg.Size:96T Intended use GHbA1c (Human) ELISA Kit is a sandwich enzyme immunoassay for the quantitative measurement of human GHbA1c. General Description vhemoglobin,

More information

Abbott PXP Glucose Meters Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory

Abbott PXP Glucose Meters Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory Abbott PXP Glucose Meters 2015 Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory Objectives After you complete this Computer-Based Learning (CBL) module, you should be able to: Explain

More information

See external label 2 C-8 C 96 tests CHEMILUMINESCENCE. CMV IgG. Cat # Step (20-25 C Room temp.) Volume

See external label 2 C-8 C 96 tests CHEMILUMINESCENCE. CMV IgG. Cat # Step (20-25 C Room temp.) Volume DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum Rat C-peptide ELISA For the quantitative determination of C-peptide in rat serum Please read carefully due to Critical Changes, e.g., see Calculation of Results. For Research Use Only. Not For Use In Diagnostic

More information

3. System Accuracy Evaluation LUX Hgb test 1. Samples

3. System Accuracy Evaluation LUX Hgb test 1. Samples 3. System Accuracy Evaluation LUX Hgb test 1. Samples - Type of sample: Capillary Blood System accuracy evaluation was performed with 200 fresh capillary blood samples, each with sufficient volume to be

More information

SOLUTION TEST STRIPS

SOLUTION TEST STRIPS SOLUTION TEST STRIPS Patented AHPHLD8 REF066 (17/055) Virox_InsertBook_PreventionHLD8_06F.indd 1 2017-04-21 11:14 Virox_InsertBook_PreventionHLD8_06F.indd 2 2017-04-21 11:14 INTENDED USE: The Prevention

More information

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual Page 1 / 12 25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual ZV-4027-0500-10 500 2-8 C Page 2 / 12 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. IVD SYMBOLS...

More information

Insulin (Porcine/Canine) ELISA

Insulin (Porcine/Canine) ELISA Insulin (Porcine/Canine) ELISA For the quantitative measurement of insulin in Porcine/Canine serum and plasma (EDTA) For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSPO-E01

More information

Rapid Detection FAQs

Rapid Detection FAQs Rapid Detection FAQs 3M Rapid Detection Reader What are the dimensions of the Rapid Detection Reader? The Rapid Detection Reader includes the Control Module and the Test Module. The Control Module is 185

More information

Quality Specifications for POCT

Quality Specifications for POCT Quality Specifications for POCT the patient journey perspective Christopher P Price Department of Clinical Biochemistry, University of Oxford, Oxford, UK Quality Specifications fit-for-purpose QUALITY

More information

Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient.

Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient. 1 2 Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient. All such tests are considered lab tests and are

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole

More information

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Rat Insulin ELISA For the quantitative determination of insulin in rat serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: Size: 80-INSRT-E01, E10 96 wells, 10

More information

FREND Total T3 Total Triiodothyronine

FREND Total T3 Total Triiodothyronine NESPI-FRTT3-001EN(V.0.1) FREND Total T3 Total Triiodothyronine An elevated TSH in conjunction with a low T4, is encountered in hypothyroidism. The T3 test is performed as part of an evaluation of thyroid

More information

Quality Corner: Inspection Preparation By Rhonda Burgard, Client Services Supervisor

Quality Corner: Inspection Preparation By Rhonda Burgard, Client Services Supervisor Volume 12 Issue 59 401 North 9 th Street, Bismarck ND 58501-4507 June 2016 Quality Corner: Inspection Preparation To prepare for a laboratory inspection begin by obtaining the appropriate inspection checklist

More information

Porcine/Canine Insulin ELISA

Porcine/Canine Insulin ELISA Porcine/Canine Insulin ELISA For the quantitative determination of insulin in porcine or canine serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research

More information

Mouse Ultrasensitive Insulin ELISA

Mouse Ultrasensitive Insulin ELISA Mouse Ultrasensitive Insulin ELISA For the quantitative determination of insulin in mouse serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research Use

More information

QUICK REFERENCE INSTRUCTIONS For use with Sofia and Sofia 2. Rx only

QUICK REFERENCE INSTRUCTIONS For use with Sofia and Sofia 2. Rx only QUICK REFERENCE INSTRUCTIONS For use with Sofia and Sofia 2. Rx only Test Procedure Study the Package Insert and User Manual thoroughly before using Quick Reference Instructions. This is not a complete

More information

READ HIGHLIGHTED CHANGES

READ HIGHLIGHTED CHANGES BIO-FLASH anti-hbs 3000-8581 100 tests The BIO-FLASH anti-hbs is a fully automated chemiluminescent simultaneous immunoassay for quantitative measurement of antibodies to Hepatitis B surface antigen (anti-hbs)

More information

Auto HbA1C ANALYZER HbA1LC HPLC 3660

Auto HbA1C ANALYZER HbA1LC HPLC 3660 Turnkey Laboratories Solutions Auto HbA1C ANALYZER HbA1LC HPLC 3660 Viewable Accuracy Real Time Summarized Chromatogram Authentic Results EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES

More information

QUICK REFERENCE INSTRUCTIONS

QUICK REFERENCE INSTRUCTIONS QUICK REFERENCE INSTRUCTIONS For use with the Sofia Analyzer only. CLIA Complexity: WAIVED Nasal Swab and Nasopharyngeal Swab specimens ONLY. Study the Package Insert and User Manual thoroughly before

More information

Internal Quality Control in the Haemostasis laboratory. Dr Steve Kitchen Sheffield Haemophilia and Thrombosis centre & UK NEQAS Blood Coagulation

Internal Quality Control in the Haemostasis laboratory. Dr Steve Kitchen Sheffield Haemophilia and Thrombosis centre & UK NEQAS Blood Coagulation Internal Quality Control in the Haemostasis laboratory Dr Steve Kitchen Sheffield Haemophilia and Thrombosis centre & UK NEQAS Blood Coagulation Disclosures None What is Quality Control? In a medical laboratory

More information

A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls

A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls Shyamali Pal R B Diagnostic Private Limited, Lake Town, Kolkata, India INFO ABSTRACT Corresponding

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th

More information

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum. For Research Use Only. Not For Use In Diagnostic Procedures.

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum. For Research Use Only. Not For Use In Diagnostic Procedures. Rat C-peptide ELISA For the quantitative determination of C-peptide in rat serum. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: Size: 80-CPTRT-E01 96 wells Version: May 26,

More information

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. GLUH Glucose REF B24985 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE

More information

DIAGNOSTIC AUTOMATION, INC.

DIAGNOSTIC AUTOMATION, INC. DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information

See external label 2 C-8 C 96 tests Chemiluminescence. CMV IgM. Cat # Diluted samples, controls & calibrator 100 µl 30 minutes

See external label 2 C-8 C 96 tests Chemiluminescence. CMV IgM. Cat # Diluted samples, controls & calibrator 100 µl 30 minutes DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information

ARK Methotrexate Assay Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 4600 Chemistry System

ARK Methotrexate Assay Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 4600 Chemistry System C ARK Methotrexate Assay Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 4600 Chemistry System Reference No. 5026000100 Intended for the Quantitative Determination of Methotrexate

More information

This is a CLIA-waived system. Policies and Procedures Manual

This is a CLIA-waived system. Policies and Procedures Manual This is a CLIA-waived system. Policies and Procedures Manual Approval date Approved by Reviewed/Revised: Date By whom Title Roche Diagnostics 9115 Hague Road Indianapolis, IN 46256 www.poc.roche.com COAGUCHEK

More information

BLOOD IS COMPLEX ANALYZING IT SHOULDN T BE DxH 500 * *Not available for sale in the U.S.

BLOOD IS COMPLEX ANALYZING IT SHOULDN T BE DxH 500 * *Not available for sale in the U.S. BLOOD IS COMPLEX ANALYZING IT SHOULDN T BE DxH 500 * *Not available for sale in the U.S. Introducing the DxH 500, an Open-vial Hematology System *. Quality 5-part Differential in a Compact Design The first

More information

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Alert or critical values represent those assay results that require prompt, rapid clinical attention to avert significant study-participant

More information

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014 Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The

More information

Rat Proinsulin ELISA

Rat Proinsulin ELISA Rat Proinsulin ELISA For the quantitative determination of proinsulin in rat serum For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-PINRT-E01 Size: 96 wells Version: June

More information

Tips for Successful Point-of-Care Testing

Tips for Successful Point-of-Care Testing Tips for Successful Point-of-Care Testing James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology and Immunology Medical Director of Clinical Chemistry and Point-of-Care Testing Vanderbilt

More information

Mouse C-peptide ELISA

Mouse C-peptide ELISA Mouse C-peptide ELISA For the quantitative determination of C-peptide in mouse serum. For Research Use Only. Not for use in Diagnostic Procedures. Please read carefully due to Critical Changes, e.g., Preparation

More information

Insulin ELISA. For the quantitative determination of insulin in serum and plasma

Insulin ELISA. For the quantitative determination of insulin in serum and plasma Insulin ELISA For the quantitative determination of insulin in serum and plasma For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United

More information

Management of Central Venous Access Devices. Blood Glucose Monitoring

Management of Central Venous Access Devices. Blood Glucose Monitoring Management of Central Venous Access Devices Blood Glucose Monitoring Purpose To provide education on the standard of care regarding the use and monitoring of the Accu- Chek Blood glucose machine, including

More information

Mouse C-peptide ELISA

Mouse C-peptide ELISA Mouse C-peptide ELISA For the quantitative determination of C-peptide in mouse serum. For Research Use Only. Not for use in Diagnostic Procedures. Please read carefully due to Critical Changes, e.g., Calculation

More information

Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual

Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual Page 1 / 14 Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual ZV-3051-0200-15 200 Page 2 / 14 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST

More information

Exendin-4 (Exenatide) ELISA Kit

Exendin-4 (Exenatide) ELISA Kit Exendin-4 (Exenatide) ELISA Kit Catalog: DEIABL227 For the quantitative determination of Exendin-4 in serum or plasma using competitive ELISA method For Research Use Only. Protocol Provided for Informational

More information

SARASOTA MEMORIAL HOSPITAL

SARASOTA MEMORIAL HOSPITAL SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE BLOOD GLUCOSE TESTING: ACCUCHEK INFORM II/ QUALITY CONTROL DATE: REVIEWED: PAGES: 1/95 3/18 1 of 8 PS1094 ISSUED FOR: Nursing RESPONSIBILITY: RN, LPN,

More information

Items in the package:

Items in the package: Intended Use: The EasyLife Hb Monitoring System is designed for in vitro diagnostic use only (external use only), and is suitable for self-testing. The system is for healthcare professionals and persons

More information

Precision ceedpro. Online Training for. Glucometer & Capillary Collection

Precision ceedpro. Online Training for. Glucometer & Capillary Collection Online Training for Precision ceedpro Glucometer & Capillary Collection How long will it take? This module takes approximately 30 minutes to complete. What if I have to leave before I finish? Progress

More information

The challenges of quality assessment for point-of-care testing. Anne Stavelin. EQALM Symposium, Bergen 8 October General challenges

The challenges of quality assessment for point-of-care testing. Anne Stavelin. EQALM Symposium, Bergen 8 October General challenges The challenges of quality assessment for point-of-care testing Anne Stavelin EQALM Symposium, Bergen 8 October 2015 www.noklus.no Norwegian quality improvement of primary care laboratories General challenges

More information

Insulin ELISA. For the quantitative determination of insulin in serum and plasma.

Insulin ELISA. For the quantitative determination of insulin in serum and plasma. Insulin ELISA For the quantitative determination of insulin in serum and plasma. For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United

More information

EpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric)

EpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric) EpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric) Base Catalog # PLEASE READ THIS ENTIRE USER GUIDE BEFORE USE Uses: The EpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric)

More information

Dear HighQ Check System Owner :

Dear HighQ Check System Owner : Dear HighQ Check System Owner : Thank you for purchasing the HighQ Check Blood Glucose Monitoring System. This manual provides important information to help you to use the system properly. Before using

More information

Disclosures. Relevant Financial Relationship(s): Nothing to Disclose. Off Label Usage: Nothing to Disclose 9/19/2017. Proficiency Testing

Disclosures. Relevant Financial Relationship(s): Nothing to Disclose. Off Label Usage: Nothing to Disclose 9/19/2017. Proficiency Testing September 11, 2017 Teri Ross, Quality Specialist II Doreen Ryan, MA, MT; Quality Operations Supervisor Deborah J. Wells, MPA, MT(ASCP)SH, Quality Management Coordinator Disclosures Relevant Financial Relationship(s):

More information

Herpes Simplex Virus 2 IgM HSV 2 IgM

Herpes Simplex Virus 2 IgM HSV 2 IgM DIAGNOSTIC AUTOMATION, INC. 21250 Califa Street, Suite 102 and 116, Woodland Hills, CA 91367 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com

More information

NEPHROCHECK Liquid Control Kit Package Insert

NEPHROCHECK Liquid Control Kit Package Insert NEPHROCHECK Liquid Control Kit Package Insert Manufactured for Astute Medical, Inc. 3550 General Atomics Ct. Building 2 San Diego, CA 92121 USA Intended Use The NEPHROCHECK Liquid Controls are used for

More information

See external label 96 tests HSV 2 IgA. Cat #

See external label 96 tests HSV 2 IgA. Cat # DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information

Hepatitis C Virus (HCV) Antibody Test

Hepatitis C Virus (HCV) Antibody Test Hepatitis C Virus (HCV) Antibody Test Instructions For Use Format: Cassette For: Catalog Number: VEL-001-HCV Specimen: Serum/Plasma * Please read the instructions carefully before use INTENDED USE Hepatitis

More information

your bioassay is in good hands: transfer from a CRO perspective MIKE MERGES March 4, 2013

your bioassay is in good hands: transfer from a CRO perspective MIKE MERGES March 4, 2013 your bioassay is in good hands: transfer from a CRO perspective MIKE MERGES March 4, 2013 Contact Information Michael Merges, M.S. Director Analytical Development Solutions Development & Clinical Services

More information

PROFICIENCY TESTING POLICY

PROFICIENCY TESTING POLICY Supersedes Prepared by: APPROVALS in this section Approved by: Date: Laboratory Director RECORD OF REVIEWS Date Signature Title Procedural Changes/Review VERSION HISTORY Revision # 0 Section #/Changes

More information

Interference and Point-of-Care Testing Devices

Interference and Point-of-Care Testing Devices Interference and Point-of-Care Testing Devices Nam K. Tran, PhD, HCLD, (ABB), FACB Associate Clinical Professor Director of Clinical Chemistry, Special Chemistry/Toxicology and POCT 1 Learning Objectives

More information

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Original: 5/02 Minor Revision: 9/15 Full Review: 1/15 Replaces: 1/15 Responsible Person: Point of Care Laboratory Specialist Approving Committee:

More information

Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter

Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter The Nova Stat Strip Blood Glucose meter is UHB s approved Glucometer

More information

Standard Operating Procedure

Standard Operating Procedure Subject Urine Microalbumin/Creatinine by Siemens DCA Vantage - Boscobel Index Number Lab-8764 Section Laboratory Subsection Regional and/or Affiliates Category Departmental Contact Judy Dayton Last Revised

More information

Using Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues

Using Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues Using Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues Stephen E. Kahn, PhD, DABCC, FACB Professor and Vice Chair, Clinical Services, Pathology Loyola University

More information

Blood Glucose Monitoring System. Owner s Booklet

Blood Glucose Monitoring System. Owner s Booklet Blood Glucose Monitoring System Owner s Booklet The FreeStyle Freedom Lite Meter and How It Works System Check Screen This display always appears when the meter is turned on. You should check that your

More information

ANTIBODY SCREENING by Uni-Gold Recombigen HIV

ANTIBODY SCREENING by Uni-Gold Recombigen HIV ANTI-HIV SPECIMEN 1 REQUIREMENTS ANTIBODY SCREENING by Uni-Gold Recombigen HIV PRINCIPLE: The Uni-Gold Recombigen HIV was designed as a rapid immunoassay and is intended to detect antibodies to HIV- 1

More information

Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised:

Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised: Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised: PURPOSE The Abbott FreeStyle Precision Pro (FSPP) Blood Glucose

More information

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2 For detection and measurement of human interleukin 2 Catalog #02006 Catalog #02007 2 Plates 10 Plates Product Description The Human Interleukin 2 (IL-2) ELISA Kit is designed for the quantitative detection

More information

Rapid-VIDITEST. Influenza A

Rapid-VIDITEST. Influenza A Rapid-VIDITEST Influenza A (One step Influenza A Card test for the detection of Influenza type A antigen from human nasopharyngeal specimens (swab, nasopharyngeal wash and aspirate). Instruction manual

More information

Blood Glucose Monitoring System. User Guide

Blood Glucose Monitoring System. User Guide Blood Glucose Monitoring System User Guide Table of Contents Introduction...2 Important Safety Instructions...2 About ipet PRO Blood Glucose Monitoring System...3 About ipet PRO Meter...4 About the ipet

More information

It s Just a Waived Glucose, Isn t It?

It s Just a Waived Glucose, Isn t It? It s Just a Waived Glucose, Isn t It? What Is the Next Step? Becky Damiani, MT (ASCP), Senior Inspection Specialist Laboratory Accreditation Program Objectives Understand the CLIA requirements surrounding

More information

Vedolizumab Drug Level ELISA

Vedolizumab Drug Level ELISA Vedolizumab Drug Level ELISA For the quantitative determination of free Entyvio concentration in human serum and EDTA plasma. Please read carefully due to Critical Changes, e.g., Updates to specificity

More information

CYTOMEGALOVIRUS (CMV) IgM ELISA Kit Protocol

CYTOMEGALOVIRUS (CMV) IgM ELISA Kit Protocol CYTOMEGALOVIRUS (CMV) IgM ELISA Kit Protocol (Cat. No.:EK-310-91) 330 Beach Road, Burlingame CA Tel: 650-558-8898 Fax: 650-558-1686 E-Mail: info@phoenixpeptide.com www.phoenixpeptide.com INTENDED USE The

More information

GlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System

GlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System GlucCell TM SYSTEM USER S GUIDE Ver 2.1 Intended Use The GlucCell TM Glucose Monitoring System (The GlucCell TM System) is designed to quantitatively measure the concentration of glucose during cell culture.

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Lactate REF A95550 For In Vitro Diagnostic Use ANNUAL REVIEW Refer to Review and Revision Coversheet at the front of each method. PRINCIPLE INTENDED USE reagent,

More information