Accuracy of Point-of-Care Blood Glucose Measurements in Critically Ill Patients in Shock

Size: px
Start display at page:

Download "Accuracy of Point-of-Care Blood Glucose Measurements in Critically Ill Patients in Shock"

Transcription

1 538608DSTXXX / Journal of Diabetes Science and TechnologyGaringarao et al research-article2014 Original Article Accuracy of Point-of-Care Blood Glucose Measurements in Critically Ill Patients in Shock Journal of Diabetes Science and Technology 2014, Vol. 8(5) Diabetes Technology Society Reprints and permissions: sagepub.com/journalspermissions.nav DOI: / dst.sagepub.com Carlo Jan Pati-an Garingarao, MD 1, Myrna Buenaluz-Sedurante, MD 1, and Cecilia Alegado Jimeno, MD 1 Abstract Background: A widely used method in monitoring glycemic status of ICU patients is point-of-care (POC) monitoring devices. A possible limitation to this method is altered peripheral blood flow in patients in shock, which may result in over/ underestimations of their true glycemic status. Methods: This study aims to determine the accuracy of blood glucose measurements with a POC meter compared to laboratory methods in critically ill patients in shock. POC blood glucose was measured with a glucose-1-dehydrogenasebased reflectometric meter. The reference method was venous plasma glucose measured by a clinical chemistry analyzer (glucose oxidase-based). Outcomes assessed were concordance to ISO 15197:2003 minimum accuracy criteria for glucose meters, bias in glucose measurements obtained by the 2 methods using Bland Altman analysis, and clinical accuracy through modified error grid analysis. Results: A total of 186 paired glucose measurements were obtained. ISO 2003 accuracy criteria were met in 95.7% and 79.8% of POC glucose values in the normotensive and hypotensive group, respectively. Mean bias for the normotensive group was 12.4 mg/dl, while mean bias in the hypotensive group was 34.9 mg/dl. POC glucose measurements within the target zone for clinical accuracy were 90.2% and 79.8% for the normotensive and hypotensive group, respectively. Conclusions: POC blood glucose measurements were significantly less accurate in the hypotensive subgroup of ICU patients compared to the normotensive group. We recommend a lower threshold in confirming POC blood glucose with a central laboratory method if clinically incompatible. In light of recently updated accuracy standards, we also recommend alternative methods of glucose monitoring for the ICU population as a whole regardless of blood pressure status. Keywords accuracy, blood glucose, critically ill, hypotension, point-of-care glucose meter, shock Maintaining normoglycemia is an important aspect in critical care, for patients with established diabetes or stress-induced hyperglycemia. Hyperglycemia has been shown to have detrimental effects in such patients, 1-3 and appropriate treatment has been shown to reduce mortality and admission duration, prevent acute renal injury, and provide faster weaning from mechanical ventilation. 1,2,4 On the other hand, hypoglycemia also has its associated adverse effects 5 and has been shown to be an independent predictor of mortality in several studies. 6,7 Accordingly, several organizations have proposed their own target blood glucose levels for ICU patients. 8,9 To achieve these goals, many local ICU s and hospital wards utilize point-of-care (POC) glucose meters in monitoring glycemic status. These provide portability, ease of use, and readily available results; important in patients requiring tight glycemic control where delayed central laboratory results make rapid adjustments in therapy impractical. Most of these meters have established accuracy in stable outpatients. For critically ill patients, various confounding factors have been identified that may interfere with POC glucometry, including hematocrit, oxygenation, acid-base disturbances, temperature, and shock states. 10,11 With regard to shock, several possible mechanisms may explain the discrepancies in blood glucose measurements obtained by different methods. Peripheral vasoconstriction 1 Section of Endocrinology, Diabetes and Metabolism, Department of Medicine, University of the Philippines Philippine General Hospital, Manila, Philippines Corresponding Author: Carlo Jan Pati-an Garingarao, MD, Section of Endocrinology, Diabetes and Metabolism, Department of Medicine, University of the Philippines, Endocrine Office, Medical Research Laboratory, 2nd Fl, UP-Philippine General Hospital, Taft Ave, Manila, Philippines. jinggoy83@yahoo.com

2 938 Journal of Diabetes Science and Technology 8(5) in hypoperfusion states leads to increased glucose extraction by tissues due to low capillary flow, resulting in falsely decreased glucose measurements with capillary blood. 12,13 Some studies on critically ill patients include only a small number of measurements taken from patients in shock, resulting in a heterogeneous cohort, or did not perform analysis on compliance with international accuracy standards. 18,21,23 For those that performed a separate analysis on hypotensive patients, limitations include a small sample size, 12,16 relatively old glucose meter technology, 12 and lack of a reference standard of glucose measurement for comparison of POC glucose meter readings. 13 Thus, this study aims to investigate the accuracy of POC glucose meter measurements in critically ill, hypotensive patients in terms of ISO 15197:2003 criteria, Bland Altman agreement analysis, and modified error grid analysis. Results of this study may help guide future policies on ICU glucose monitoring and insulin infusion protocols in determining intervention thresholds more suitable for patients in shock. Methods Study Participants and Recruitment This is a cross-sectional, criterion-referenced study conducted in the University of the Philippines Philippine General Hospital. Subjects were recruited from adult, nonpregnant patients referred for intensive care after obtaining informed consent from primary decision makers. Patients were consecutively classified according to perfusion status. The hypotensive group included those in shock with mean arterial pressure (MAP) <70 mmhg or those requiring vasopressor support. The normotensive group included patients with MAP 70 mmhg without vasopressor support. Exclusion criteria were hypovolemic shock due to severe active bleeding, bleeding disorders, platelet count <25,000/ ul, hematocrit <20%, the use of substances known to interfere with the POC glucose meter technology used (icodextrin-containing solutions, intravenous immunoglobulins, abatacept, maltose), and lack of consent. Participation in the study continued throughout ICU stay; provided that consent was not withdrawn, none of the exclusion criteria were fulfilled, and the maximum limit of glucose determinations was not exceeded. For patients initially recruited in the normotensive group, participation was terminated if they suffered from hypotension or require vasopressor support during their ICU stay, and vice versa. Sample Size Calculation A sample size of 89 patients for each group was needed to determine the percentage of POC blood glucose values concordant with the laboratory whole-blood glucose method with ±10% margin of error. This computation was based on the data from Atkin et al 12 using OpenEpi opensource calculator (version 3), with the following formula, n = [DEFF*Np(1-p)]/ [(d 2 /Z 2 *(N-1)+p*(1-p)]; where 1-α/2 N = total population (assumed infinite), p = proportion of hypotensive POC glucose values meeting accuracy criteria in the reference study (36%), DEFF = design effect (assumed 1), d = error (10%), α = level of significance (.05). Patient Data Collection Pertinent patient and laboratory data were collected at baseline from chart review. Demographic data included age, sex, and anthropometric measurements. Historical data included chronic comorbidities and maintenance medications taken for the past month. Baseline clinical data noted were initial vital signs, the Glasgow Coma Scale score with derived verbal scores for intubated patients, 24 the need for ventilatory support or dialysis, and the presence of acute significant arrhythmias. The medications given during admission and admitting diagnosis were obtained by systematic chart review. Laboratory parameters recorded were arterial blood gas analysis, serum creatinine, albumin, and complete blood count. The laboratory values taken at the time most proximate to the blood glucose determination were recorded. Blood Glucose Determinations A pair of blood glucose measurements was performed upon recruitment, comprising 1 whole-blood specimen from a peripheral vein for central laboratory glucose determination and 1 POC glucose meter measurement. These were obtained from an extremity with no vasopressors or glucose-containing fluids being infused, if feasible. Other noncritical intravenous fluids were temporarily discontinued for at least 1 minute prior and during blood extraction. If any indwelling central venous access was already in place, these was used to obtain venous whole-blood specimens after discarding initial dead-space blood (3-4 ml). The venous whole-blood specimen (~3 ml) underwent centrifugation and separation from red blood cells within 30 minutes to avoid falsely low glucose values due to glycolysis, refrigerated, and later measured using a glucose oxidasebased laboratory method (I-Lab 300 Plus Clinical Chemistry Analyzer, Diamond Diagnostics). This central laboratory analyzer has 2.5% CV (coefficient of variation) at mean glucose of 79 mg/dl, and 2.2% CV at mean glucose of 273 mg/ dl (I-Lab Application manual, rev /11). Capillary blood was obtained by fingertip puncture with a lancing device and 1 drop of blood (~1 ul) tested using an Accu-Chek Active Meter (Roche Diagnostics, Indianapolis, Indiana, USA), a glucose-1-dehydrogenase pyrroloquinolone quinone (GDH-PQQ) enzyme reaction-based, reflectometric glucose meter with plasma-corrected results. An accuracy study on outpatients with diabetes using this meter showed 100% of measurements falling within ISO minimum accuracy criteria. Intermediate precision showed a % CV; and repeatability results showed a % CV (data provided by Roche Diagnostics). This meter has

3 Garingarao et al 939 also been tested on stable inpatients, with 100% concordance to ISO standards for meter accuracy and 100% clinically accurate results by Clarke error grid analysis. 25 General advantages of a GDH-based meter include decreased susceptibility to altered results due to hypoxia or oxygen therapy and ph derangements compared to glucose-oxidase based meters. 10 All blood extractions in this study followed standard procedures set by the Clinical and Laboratory Standards Institute (CLSI). 26 Important data during each blood glucose measurement such as blood pressure and ongoing vasopressor infusion rates were recorded. Aside from baseline blood extractions, additional measurements were performed during any glucose <70 mg/dl or 180 mg/dl detected during routine monitoring. A maximum limit of 4 additional measurements during dysglycemia was set for each patient. The ICU physicians and staff were oriented monthly on the study procedures and the attending physicians were also provided a summary of recommendations on blood glucose monitoring for inpatients by the American Diabetes Association (2009). The POC glucose meters were calibrated monthly according to manufacturer s instructions, with all tests falling within specified calibration values. No significant meter errors were encountered during the course of the study. Outcome Measures Concordance to minimum accuracy criteria for POC blood glucose meters set by the International Organization for Standardization (ISO) in was evaluated. This requires that at least 95% of POC glucose meter readings should fall within 20% of true blood glucose values if 75 mg/dl, or should fall within 15 mg/dl of true blood glucose values if < 75 mg/dl. Technical accuracy was assessed by Bland Altman graphs plotting the average between paired values of glucose values measured by the 2 methods (x-axis) versus the difference between the paired measurements (y-axis). 28 Separate plots and calculations were done for each group. Clinical accuracy was evaluated by adapting the modified error grid analysis table of Kanji et al, 16 which categorizes measurements into 1 of several error grid analysis zones according to clinical action. The analysis zones based on the capillary blood glucose results were labeled as follows: target zone (acceptable error), hypoglycemia zone (<70 mg/dl), normoglycemia zone ( mg/dl), and hyperglycemic zone ( 180 mg/dl), with subclassification whether these were over- or underestimations of the reference method values. Classification as over/underestimation was based on the intervention thresholds set by our local insulin titration protocol (Josol CV et al, unpublished data); with limits at <50 mg/dl, mg/dl, mg/dl, mg/dl, and 250 mg/dl and corresponding interventions based on magnitude of change from the previous glucose measurement. Data Analysis The comparison of clinical characteristics between groups was done by Fisher s exact test or z test for discrete variables and 2-tailed t test for continuous variables with α of.05. All continuous data were expressed as mean (SD). Concordance to ISO 15197:2003 criteria for accuracy was expressed in proportions of values falling within various intervals following the ISO recommended format. Bland Altman graphs for agreement were constructed using Stata/ SE (version 12.0) statistical software. Tabulation of modified error grid analysis results were patterned according the study of Kanji et al, 16 and the proportions falling to the different analysis zones compared between normotensive and hypotensive groups. Ethical Considerations The study protocol was approved by the University of the Philippines Manila, Research Ethics Board. Implementation of the study commenced only after approval. The investigators had no conflict of interest in any form with the manufacturers of any equipment used and adhered to the guiding principles of good clinical practice throughout the conduct of the study. A maximum of 4 additional measurements during dysglycemia was set to minimize iatrogenic blood loss. Attending physicians were immediately notified of both POC and laboratory whole-blood glucose results once available. Results A total of 180 patients were recruited from July 2013 to January 2014, composed of 89 normotensive and 91 hypotensive patients. This resulted in 186 paired glucose measurements (92 normotensive and 94 hypotensive) for analysis. The majority of the normotensive patients had venous blood specimens obtained through peripheral venipuncture (85.4%). Most subjects in the hypotensive group also had venous blood specimens collected via peripheral venipuncture (73.6%), the remainder from central venous pressure lines (14.3%), or central vein catheters for hemodialysis access (12.1%). The hypotensive group showed a significantly lower MAP, higher heart rate, higher respiratory rate, lower GCS score, and higher proportion requiring mechanical ventilation (Table 1), which understandably reflects a greater severity of illness compared to the normotensive group. For the baseline laboratory parameters, arterial ph was significantly lower in the hypotensive group probably due to lactic acidosis from systemic hypoperfusion. Serum albumin levels were significantly lower and leukocyte counts were significantly higher in this group, probably correlating with the degree of systemic inflammation in more critically ill patients. The rest

4 940 Journal of Diabetes Science and Technology 8(5) Table 1. Clinical Characteristics of Included ICU Patients in both Normotensive and Hypotensive Groups. Clinical characteristic Normotensive group (n = 89) of the clinical and laboratory parameters were comparable between the 2 groups. The hypotensive group (Table 2) was composed mainly of cases of cardiogenic shock (69.2%) from acute myocardial infarction, decompensated heart failure, or bradyarrhythmias, followed by septic shock (51.7%) from severe pneumonia or urosepsis. Nearly a third (29.7%) had multiple causes of shock, usually both cardiogenic and septic shock. More than half were on a single vasopressor (58.2%), and dopamine was most commonly used vasopressor (51.7%) at a mean dose of 13.1 ug/kg/min. Concordance to ISO 2003 Criteria Hypotensive group (n = 91) P value Age (years) 51.3 (16.1) 54.7 (16.2).158 Sex Male Female Mean arterial pressure 87.8 (15.7) 70.8 (22.0) <.001* (mmhg) Heart rate (beats per 86.3 (21.4) (26.5) <.001* minute) Temperature (degrees 36.6 (1.2) 36.8 (1.5).282 centigrade) Respiratory rate (cycles per 22.1 (5.9) 25.3 (8.6).005* minute) Estimated weight (kg) 58.5 (11.8) 60.3 (12.5).303 Glasgow Coma Scale score 14.0 (1.9) 12.7 (3.6).004* Need for mechanical 32 (36.0%) 50 (55.0%).011* ventilation Need for hemodialysis 11 (12.4%) 15 (16.5%).770 Type 2 diabetes, previous 20 (22.5%) 16 (17.6%).717 diagnosis Significant arrhythmia/s 20 (22.5%) 17 (18.7%).777 Serum creatinine (umol/l) (213.8) (256.0).196 Albumin (mg/dl) 29.8 (8.8) 26.8 (9.1).044* Hematocrit (%) (0.073) (0.077).354 Hemoglobin (g/l) (24.3) (26.4).920 Leukocyte count ( 10 9 /L) 12.4 (6.3) 16.0 (15.0).040* Arterial blood ph 7.46 (0.091) 7.39 (0.129) <.001* Arterial blood po2 (mmhg) (83.5) (108.6).437 Fraction inspired oxygen (%) 44.2 (30.5) 53.4 (33.5).055 Continuous values expressed in mean (SD); n, number of patients recruited. *P <.05. The POC glucose measurements in the normotensive group met the minimum ISO 2003 accuracy criteria (Table 3). All glucose values <75 mg/dl were within 15 mg/dl of the reference method results, while 95.5% of values 75 mg/dl were within 20% of the reference method measurements. Grouping all measurements resulted in a 95.65% concordance to the said criteria. The POC glucose values in the hypotensive group, however, failed to meet the minimum accuracy criteria. For Table 2. Clinical Characteristics Specific for the Hypotensive Group. Clinical characteristic Table 3. System Accuracy Results of Capillary Blood POC Glucose Measurements Grouped According to Glucose Concentrations <75 mg/dl and 75 mg/dl. Normotensive group (n = 92) Hypotensive group (n = 94) Concentration n (%) n (%) Glucose concentration <75 mg/dl Within ± 5 mg/dl 1 (33.3) 2 (50) Within ± 10 mg/dl 3 (100) 2 (50) Within ± 15 mg/dl a 3 (100) 2 (50) Glucose concentration 75 mg/dl Within ± 5% 27 (30.3) 25 (27.8) Within ± 10% 59 (66.3) 43 (47.8) Within ± 15% 75 (84.3) 62 (68.9) Within ± 20% a 85 (95.5) 73 (81.1) Overall concordance (%) n, number of paired glucose measurements analyzed. a Criteria used for determining acceptability according to ISO 15197:2003. hypoglycemic values, only half met the accuracy criteria; for glucose values >75 mg/dl, only 81.1% of values fell within 20% of the reference method values, resulting in an overall concordance of 79.8% for the hypotensive group. Technical Accuracy Number of patients (%) Mechanism of shock 63 (69.2) Cardiogenic Septic 47 (51.7) Hypovolemic 11 (12.1) Others 3 (3.3) Multiple (>1) 27 (29.7) Number of vasopressors 0 3 (3.3) 1 53 (58.2) 2 24 (26.4) 3 11 (12.1) Type of vasopressor support Dopamine 47 (51.7) Norepinephrine 37 (40.7) Dobutamine 34 (37.4) Epinephrine 16 (17.6) Bland Altman difference plot for the normotensive group (Figure 1) including all paired measurements showed a mean bias of 12.4 mg/dl, with mean POC glucose slightly lower than mean venous glucose and limits of agreement (LOA) at 86.0 and 61.2 mg/dl. The hypotensive group (Figure 2)

5 Garingarao et al 941 absolute bias in either group, but did not change the results when comparing the mean bias and confidence limits between groups. Figure 1. Bland Altman difference plot of paired glucose measurements (POC glucose and venous whole-blood glucose) in the normotensive group (n = 92). LOA, limits of agreement; POC, point of care. Clinical Accuracy The normotensive group had 90.2% of POC glucose measurements within the target zone (Table 4). A small proportion of apparently normoglycemic (4.3%) and apparently hyperglycemic POC glucose values (3.3%) were considered underestimations, and a few hyperglycemic values would have resulted in unnecessary insulin administration (2.2%) due to overestimation. For the hypotensive group, a smaller proportion of glucose measurements were in the target zone (79.8%). One hypoglycemic measurement each was an over- and underestimation, resulting in inappropriate intervention. A larger proportion of apparently normoglycemic values (11.7%) were underestimations. For apparently hyperglycemic values, 4.3% were overestimations, which would have resulted in unnecessary insulin administration, while 2.1% were underestimations. Discussion Figure 2. Bland Altman difference plot of paired glucose measurements (POC glucose and venous whole-blood glucose) in the hypotensive group (n = 94). LOA, limits of agreement; POC, point of care. also showed a lower mean POC glucose compared to venous glucose values, with a mean bias of 34.9 mg/dl (LOA: 207.1, mg/dl). Comparing between groups showed a larger absolute mean bias in the hypotensive group. The confidence limits (range between upper and lower LOA) are also wider for the hypotensive group versus the normotensive group (344 mg/ dl vs 147 mg/dl). Paired glucose measurements with bias lying outside the 95% CI were reviewed for any potential confounders in accuracy and technical difficulties during sampling, but none were found. Performing Bland Altman analysis after excluding these samples (not shown) slightly decreased the mean Only a few published studies focused on shock as a possible factor affecting POC glucose meter accuracy, with various design limitations cited earlier ,16-19,21-23 For concordance with ISO 2003 criteria, the POC glucose meter used met the minimum accuracy criteria for the normotensive group (95.6% of all values). This was not attained for the hypotensive group, with an overall concordance of only 79.8%. This reflects the overall lower accuracy of glucose measurements, both technical and clinical, taken from this subset of ICU patients. Other studies found similar results when using various accuracy criteria for POC glucose meters. Atkin et al 12 noted 36% concordance in hypotensive patients compared to 90% in a normotensive group; using a range of 20% as acceptable bias similar to the ISO. Both groups would still fail to meet ISO standards using the current minimum cut-off of 95%. Critchell et al 19 studied a mixed cohort of ICU patients; with findings of 81% concordance with CLSI accuracy criteria. Desachy et al, 15 with around one-third of their ICU cohort in shock, found that 85% of POC glucose values fell within 20% of reference method values. Lonjaret et al, 17 with another heterogeneous ICU cohort, reported 74.8% ISO concordance with POC blood glucose. Pulzi et al, 29 with a separate analysis for a small subset of ICU patients on norepinephrine, found a 63.6% concordance versus 93.8% for those without norepinephrine support. Our own data show similar trends, with the normotensive group passing minimum ISO accuracy criteria (95.7%) while the hypotensive group failing to do so (79.8%). The effects of low MAP

6 942 Journal of Diabetes Science and Technology 8(5) Table 4. Distribution of POC Glucose Measurements According to Modified Error Grid Analysis Zones and Effect on Clinical Action. Distribution of paired differences from modified error grid analysis Modified error grid analysis zone Comparison with reference standard Normotensive group (n = 92) Hypotensive group (n = 94) n (%) n (%) Clinical interpretation Target zone 83 (90.2) 75 (79.8) Target (acceptable error) Hypoglycemia zone (CBG < 70 mg/dl) Overestimation 0 1 (1.1) Risk of failure to detect and treat hypoglycemia Underestimation 0 1 (1.1) Risk of unnecessary treatment of hypoglycemia Normoglycemia zone (CBG mg/dl) Hyperglycemia zone (CBG 180 mg/dl) n = number of paired glucose measurements analyzed. Overestimation 0 0 Risk of failure to detect and treat hypoglycemia Underestimation 4 (4.3) 11 (11.7) Risk of failure to detect and treat hyperglycemia Overestimation 2 (2.2) 4 (4.3) Risk of overtreatment of hyperglycemia Underestimation 3 (3.3) 2 (2.1) Risk of failure to detect and treat hyperglycemia or vasopressor use on glucose meter accuracy may be more emphasized in our study since we did separate analysis on a well-defined hypotensive group with a relatively larger sample size compared to previous studies. It appears POC glucose measurements in critically ill but normotensive patients provides acceptable accuracy (at least for the meter technology used), provided that known confounding factors are not present and standardized collection techniques are followed. Two new glucose meter accuracy criteria were released during the course of the study which deserve mention; the revised ISO 15197:2013 criteria, 30 and draft guidance from the US Food and Drug Administration (US-FDA). 31 The former sets acceptable accuracy if 95% of POC values fall within ± 15% of the reference method for glucose values 100 mg/dl, and fall within ± 15 mg/dl for glucose values <100 mg/dl. With this new criterion as reference, both groups would fail to meet minimum accuracy requirements, with only 86.9% of normotensive POC values and 68.1% of hypotensive POC values having acceptable bias. The much stricter US-FDA draft guidelines requires 99% of POC values within ± 10% of the reference method for glucose values 70 mg/dl and within ± 7 mg/dl for glucose values <70 mg/dl. None of the POC glucose measurements should exceed ± 20% of the reference method for glucose values 70 mg/dl or ± 15 mg/dl for values <70 mg/dl. Using this standard, only 65.2% of normotensive POC values and 46.8% of hypotensive POC values met accuracy criteria. Moreover, 4.3% of normotensive POC values and 18.1% of hypotensive POC values exceeded the maximum allowable bias. Bland Altman agreement analysis showed lower mean POC glucose values compared to venous blood glucose in both groups. The confidence limits were also wide in both groups, more so for the hypotensive group (147.2 mg/dl vs mg/dl). After excluding outlying values, the confidence limits became narrower but still spanned across zero, consistent with most method-comparison studies revie wed. 13,17-19,23,29 For studies with separate analysis on hypotensive subgroups, those on vasopressor support also had wider confidence limits compared to normotensive patients. 23,29 This lack of a consistent directional bias precluded us from proposing a safe correction factor for capillary glucose values in either group on which to make adjusted intervention thresholds. Modified error grid analysis for clinical accuracy showed fewer values in the hypotensive group within the target zone (79.8% vs 90.2%). For both groups, the largest proportion of clinical errors was apparently normoglycemic capillary glucose values that were underestimations (4.3% normotensive and 11.7% hypotensive glucose values). This meant undetected hyperglycemia ideally requiring supplemental insulin. No normoglycemic range values were overestimations in either group, which would translate to unnecessary insulin administration and potentially life-threatening hypoglycemia. A limitation in our data is the small number of glucose measurements <75 mg/dl (only 7), which prevented us from making generalizations for this range. This was despite our provision of obtaining additional measurements during any dysglycemia detected during ICU stay. Comparison with other published data on clinical accuracy show conflicting results. Kanji et al 16 reported an overall agreement in clinical action of 61.1% for patients on vasopressors versus 53.8% agreement for a postsurgical normotensive group. The most common errors in this study were

7 Garingarao et al 943 apparently hypoglycemic (9%) and normoglycemic (9%) values that were overestimations compared to the reference method. Another study 17 in a mixed cohort reported 75% of POC measurements in the target zone, with the most common error being underestimation compared to reference method values (22%), consistent in all prespecified glucose ranges. Petersen et al, 22 involving another heterogeneous ICU cohort, reported mild overestimation as being the most common error using Parkes error grid analysis, occurring in 17.4% of POC measurements. For the study of Pulzi et al, 29 their subgroup analysis of patients on norepinephrine showed fewer measurements on target (71%) versus those without vasopressors (81.5%). The most common errors in this study were overestimations resulting in undetected hypoglycemia. In comparing these various studies in terms of clinical accuracy, it should be taken into consideration that the labeling of inappropriate intervention depends largely on the local ICU protocol being practiced, with various thresholds for change in clinical action. Conclusions In our population of hypotensive, critically ill patients on vasopressor support, POC glucose meter values had lower accuracy in terms of concordance to ISO 2003:15197 criteria, Bland Altman agreement analysis, and modified error grid analysis, compared to normotensive ICU patients. Other methods of glucose monitoring may be more appropriate in hypotensive ICU patients, possibly central catheter blood POC measurements, 12,15 automated blood/gas analyzers, 16 or other more advanced methods. Confirmatory laboratory determinations should immediately be performed in hypotensive patients if POC glucose values appear clinically incompatible. Glycemic targets for this type of patients should perhaps be less stringent to avoid overtreatment and potentially fatal hypoglycemia. A limitation of our study is the small number of hypoglycemic measurements obtained, precluding us from making strong conclusions for the accuracy of measurements in this particular range. For normotensive ICU patients, the present glucose meter used in the study appears accurate according to ISO 2003:15197 criteria, as long as identified confounding factors are excluded and specimen collection is performed in a standardized manner to minimize accuracy errors. Recently published ISO 15197:2013 accuracy standards will render the POC glucose meter in this study inaccurate for both normotensive and hypotensive cohorts and point out the need for alternative methods of glucose monitoring. These tighter criteria will undoubtedly impact the development and costs of future POC glucose monitoring methods once compliance becomes mandatory after the 3-year transitional period, more so for the critically ill population. Abbreviations CLSI, Clinical and Laboratory Standards Institute; GCS, Glasgow Coma Scale; GDH, glucose-1-dehydrogenase; ISO, International Organization for Standardization; LOA, limits of agreement; MAP, mean arterial pressure; POC, point of care; US-FDA, US Food and Drug Administration. Declaration of Conflicting Interests The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article. Funding The author(s) received no financial support for the research, authorship, and/or publication of this article. References 1. Clement S, Braithwaite SS, Magee MF, et al. Management of diabetes & hyperglycemia in hospitals. Diabetes Care. 2004;27(2): Van den Berghe G, Wouters P, Weekers F, et al. Intensive insulin therapy in critically ill patients. N Engl J Med. 2001;345(19): Malmberg K, Norhammar A, Wedel H, Ryden L. Glycometabolic state at admission: important risk marker of mortality in conventionally-treated patients with diabetes mellitus and acute myocardial infarction: long-term results from the DIGAMI study. Circulation. 1999;99: Van den Berghe G, Wilmer A, Milants I, et al. Intensive insulin therapy in mixed medical/ surgical intensive care units. Diabetes. 2006;55: Melmed S, Polonsky KS, Reed Larsen PR, Kronenberg HM. Williams Textbook of Endocrinology. 12th ed. Philadelphia, PA: Saunders; Krinsley J, Grover A. Severe hypoglycemia in critically ill patients: Risk factors and outcomes. Crit Care Med. 2007;35(10): Griesdale DEG, de Souza RJ, vam Dam RM, et al. Intensive insulin therapy and mortality among critically ill patients: a meta-analysis including NICE-SUGAR study data. CMAJ. 2008;180(8): American Diabetes Association. Standards of medical care of diabetes Diabetes Care. 2012;35(s1): Dellinger RP, Levy MM, Carlet JM, et al. Surviving sepsis campaign: international guidelines for management of severe sepsis and septic shock. Crit Care Med. 2008;36(1): Dungan K, Chapman J, Braithwaite SS, Buse J. Glucose measurements: confounding issues in setting targets for inpatient management. Diabetes Care. 2007;30(2): Heinemann L. Quality of glucose measurement with blood glucose meters at the point-of-care: relevance of interfering factors. Diabetes Technol Ther. 2010;12(11): Atkin SH, Dasmahapatra A, Jaker MA, Chorost MI, Reddy S. Fingerstick glucose determination in shock. Ann Int Med. 1991;114: Juneja D, Pandey R, Singh O. Comparison between arterial and capillary blood glucose monitoring in patients with shock. Eur J Intern Med. 2011;22:

8 944 Journal of Diabetes Science and Technology 8(5) 14. Cook A, Laughlin D, Moore M, et al. Differences in glucose values obtained from point-of-care glucose meters and laboratory analysis in critically ill patients. Am J Crit Care. 2009;18: Desachy A, Vuagnat AC, Ghazali AD, et al. Accuracy of bedside glucometry in critically ill patients: influence of clinical characteristics & perfusion index. Mayo Clin Proc. 2008;83(4): Kanji S, Buffie J, Hutton B, et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med. 2005;33(12): Lonjaret L, Claverie V, Berard E, et al. Relative accuracy of arterial and capillary glucose meter measurements in critically ill patients. Diabetes Metab. 2012;38: Shearer A, Boehmer M, Closs M, et al. Comparison of glucose point-of-care values with laboratory values in critically ill patients. Am J Crit Care. 2009;18: Critchell CD, Savarese V, Callahan A, Aboud C, Jabbour S, Parik P. Accuracy of bedside capillary blood glucose measurements in critically ill patients. Intensive Care Med. 2007;33: Hoedemaekers CWE, Klein Gunnewiek JMT, Prinsen MA, Willems JL, Van der Hoeven JG. Accuracy of bedside glucose measurement from three glucometers in critically ill patients. Crit Care Med. 2008;36(11): Lacara T, Domagtoy C, Lickliter D, et al. Comparison of pointof-care and laboratory glucose analysis in critically ill patients. Am J Crit Care. 2007;16: Petersen JR, Graves DF, Tacker DH, Okorodudu AO, Mohammad AA, Cardenas VJ Jr. Comparison of POCT and central laboratory blood glucose results using arterial, capillary and venous samples from MICU patients on a tight glycemic protocol. Clin Chim Acta. 2008;396: Fekih Hassen M, Ayed S, Gharbi R, Ben Sik Ali H, Marghli S, Elatrous S. Bedside capillary blood glucose measurements in critically ill patients: Influence of catecholamine therapy. Diabetes Res Clin Pract. 2009;87: Meredith W, Rutledge R, Fakhry S, Emery S, Kromhout-Schiro S. The conundrum of the Glasgow Coma Scale in intubated patients: a linear regression prediction of the Glasgow Verbal Score from the Glasgow Eye and Motor Scores. J Trauma. 1998;44(5): Freckman G, Baumstark A, Jendrike N, et al. System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO Diabetes Technol Ther. 2010;12(3): National Committee for Clinical Laboratory Standards. Procedures and devices for the collection of diagnostic capillary blood specimens; approved standard fifth edition Jun; NCCLS document H4-A International Organization for Standardization. In vitro diagnostic test systems requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197:2003 (E). 28. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;i: Pulzi SA Jr, de Assuncao MSC, Mazza BF, et al. Accuracy of different methods for blood glucose measurement in critically ill patients. Sao Paolo Med J. 2009;127(5): International Organization for Standardization. In vitro diagnostic test systems requirements for blood glucose monitoring system for self-testing in managing diabetes mellitus. ISO 15197:2013 (E). 31. US Department of Health and Human Services, Food and Drug Administration. Blood glucose monitoring test systems for prescription point-of-care use: draft guidance for industry and Food and Drug Administration staff FDA-2013-D

Relationship between glucose meter error and glycemic control efficacy

Relationship between glucose meter error and glycemic control efficacy Relationship between glucose meter error and glycemic control efficacy Brad S. Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Department of Laboratory Medicine and Pathology Mayo Clinic

More information

Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES

Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES Accuracy of Roche Accu-Chek Inform Whole Blood Capillary, Arterial, and Venous Glucose Values in Patients Receiving Intensive Intravenous

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th

More information

VANDERBILT UNIVERSITY MEDICAL CENTER MULTIDISCIPLINARY SURGICAL CRITICAL CARE TRAUMA INTENSIVE CARE UNIT GLYCEMIC CONTROL PROTOCOL

VANDERBILT UNIVERSITY MEDICAL CENTER MULTIDISCIPLINARY SURGICAL CRITICAL CARE TRAUMA INTENSIVE CARE UNIT GLYCEMIC CONTROL PROTOCOL VANDERBILT UNIVERSITY MEDICAL CENTER MULTIDISCIPLINARY SURGICAL CRITICAL CARE TRAUMA INTENSIVE CARE UNIT GLYCEMIC CONTROL PROTOCOL Background: For some time, the presence of diabetes and hyperglycemia

More information

A comparative study between capillary and venous blood glucose levels of type 2 diabetes mellitus patients in intensive care units

A comparative study between capillary and venous blood glucose levels of type 2 diabetes mellitus patients in intensive care units Available online at www.medicinescience.org ORIGINAL ARTICLE Medicine Science International Medical Journal Medicine Science 2018; ( ): A comparative study between capillary and venous blood glucose levels

More information

Disclosures. Glycemic Control in the Intensive Care Unit. Objectives. Hyperglycemia. Hyperglycemia. History. No disclosures

Disclosures. Glycemic Control in the Intensive Care Unit. Objectives. Hyperglycemia. Hyperglycemia. History. No disclosures Disclosures Glycemic Control in the Intensive Care Unit No disclosures Jorie Frasiolas, Pharm.D., BCPS Clinical Pharmacy Manager, CTICU NewYork-Presbyterian Hospital Columbia University Medical Center

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess

More information

It s Just a Waived Glucose, Isn t It?

It s Just a Waived Glucose, Isn t It? It s Just a Waived Glucose, Isn t It? What Is the Next Step? Becky Damiani, MT (ASCP), Senior Inspection Specialist Laboratory Accreditation Program Objectives Understand the CLIA requirements surrounding

More information

Deepika Reddy MD Department of Endocrinology

Deepika Reddy MD Department of Endocrinology Deepika Reddy MD Department of Endocrinology Management of hyperglycemic crisis Review need for inpatient glycemic control Brief overview of relevant trials Case based review of diabetes management strategies/review

More information

FDA Perspective - Public Health Notification: Potentially Fatal Errors with GDH-PQQ Glucose Monitoring Technology Courtney C. Harper, Ph.D.

FDA Perspective - Public Health Notification: Potentially Fatal Errors with GDH-PQQ Glucose Monitoring Technology Courtney C. Harper, Ph.D. FDA Perspective - Public Health Notification: Potentially Fatal Errors with GDH-PQQ Glucose Monitoring Technology Courtney C. Harper, Ph.D., Director, Division of Chemistry and Toxicology Devices Office

More information

Performance of three portable blood glucose monitoring devices used in a veterinary application

Performance of three portable blood glucose monitoring devices used in a veterinary application Performance of three portable blood glucose monitoring devices used in a veterinary application Prepared by Michael I. Lindinger, PhD, Vice-President of Research, The Nutraceutical Alliance, Campbellville,

More information

10.4.a. Optimal glucose control: Insulin therapy March 2013

10.4.a. Optimal glucose control: Insulin therapy March 2013 10.4.a. Optimal glucose control: Insulin therapy March 2013 2013 Recommendation: Based on 26 level 2 studies, we recommend that hyperglycemia (blood sugars > 10 mmol/l) be avoided in all critically ill

More information

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing

More information

EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS

EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:23 ACCURACY AND USER EVALUATION REQUIREMENTS Douglas E. Bell, PhD Teri A. Sasse, RN, MS BACKGROUND: The

More information

Inpatient Management of Diabetes Mellitus. Jessica Garza, Pharm.D. PGY-1 Pharmacotherapy Resident TTUHSC School of Pharmacy

Inpatient Management of Diabetes Mellitus. Jessica Garza, Pharm.D. PGY-1 Pharmacotherapy Resident TTUHSC School of Pharmacy Inpatient Management of Diabetes Mellitus Jessica Garza, Pharm.D. PGY-1 Pharmacotherapy Resident TTUHSC School of Pharmacy 2 Disclosure Jessica Garza does not have any actual or potential conflicts of

More information

Surviving Sepsis Campaign: International Guidelines for Management of Sepsis and Septic Shock: 2016

Surviving Sepsis Campaign: International Guidelines for Management of Sepsis and Septic Shock: 2016 Surviving Sepsis Campaign: International Guidelines for Management of Sepsis and Septic Shock: 2016 Mitchell M. Levy MD, MCCM Professor of Medicine Chief, Division of Pulmonary, Sleep, and Critical Care

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,

More information

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical

More information

Control of Blood Glucose in the ICU: Reconciling the Conflicting Data

Control of Blood Glucose in the ICU: Reconciling the Conflicting Data Control of Blood Glucose in the ICU: Reconciling the Conflicting Data Steven E. Nissen MD Disclosure Consulting: Many pharmaceutical companies Clinical Trials: AbbVie, Amgen, Astra Zeneca, Esperion, Eli

More information

Objectives. Management of Septic Shock. Definitions Progression of sepsis. Epidemiology of severe sepsis. Major goals of therapy

Objectives. Management of Septic Shock. Definitions Progression of sepsis. Epidemiology of severe sepsis. Major goals of therapy Objectives Management of Septic Shock Review of the Evidence and Implementation of Pediatric Guidelines at Christus Santa Rosa Manish Desai, M.D. PL 5 2 nd year Pediatric Critical Care Fellow Review of

More information

Results of recent studies examining

Results of recent studies examining Relationship between hypoglycemia and mortality in critically ill children* Edward Vincent S. Faustino, MD; Clifford W. Bogue, MD Objectives: To determine the prevalence of hypoglycemia in critically ill

More information

There has been increasing interest in glucose measurement

There has been increasing interest in glucose measurement J Vet Intern Med 28;22:1189 1195 Comparison of the Accu-Chek Aviva Point-of-Care Glucometer with Blood Gas and Laboratory Methods of Analysis of Glucose Measurement in Equine Emergency Patients A.R. Hollis,

More information

Validity and Reliability of a Glucometer Against Industry Reference Standards

Validity and Reliability of a Glucometer Against Industry Reference Standards 514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference

More information

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose

More information

Parenteral Nutrition The Sweet and Sour Truth. From: Division of Endocrinology, Diabetes and Bone Disease Icahn School of Medicine at Mount Sinai

Parenteral Nutrition The Sweet and Sour Truth. From: Division of Endocrinology, Diabetes and Bone Disease Icahn School of Medicine at Mount Sinai ENDOCRINE PRACTICE Rapid Electronic Article in Press Rapid Electronic Articles in Press are preprinted manuscripts that have been reviewed and accepted for publication, but have yet to be edited, typeset

More information

Frank Sebat, MD - June 29, 2006

Frank Sebat, MD - June 29, 2006 Types of Shock Hypovolemic Shock Low blood volume decreasing cardiac output. AN INTEGRATED SYSTEM OF CARE FOR PATIENTS AT RISK SHOCK TEAM and RAPID RESPONSE TEAM Septic or Distributive Shock Decrease in

More information

Early-goal-directed therapy and protocolised treatment in septic shock

Early-goal-directed therapy and protocolised treatment in septic shock CAT reviews Early-goal-directed therapy and protocolised treatment in septic shock Journal of the Intensive Care Society 2015, Vol. 16(2) 164 168! The Intensive Care Society 2014 Reprints and permissions:

More information

Journal Club ICU

Journal Club ICU Journal Club 2018.9.4 ICU Crit Care Med. 2018 Aug;46(8):1224-1229 Introduction Stress hyperglycemia SH SH >124 mg/dl or >200 mg/dl Lancet 2009;373:1798. stress hyperglycemia Lancet 2009;373:1798. SH Critical

More information

The Ever Changing World of Sepsis Management. Laura Evans MD MSc Medical Director of Critical Care Bellevue Hospital

The Ever Changing World of Sepsis Management. Laura Evans MD MSc Medical Director of Critical Care Bellevue Hospital The Ever Changing World of Sepsis Management Laura Evans MD MSc Medical Director of Critical Care Bellevue Hospital COI Disclosures No financial interests to disclose Learning Objectives Review the evolution

More information

Normal glucose values are associated with a lower risk of mortality in hospitalized patients

Normal glucose values are associated with a lower risk of mortality in hospitalized patients Diabetes Care Publish Ahead of Print, published online August Hyperglycemia 20, 2008 in hospital Normal glucose values are associated with a lower risk of mortality in hospitalized patients Alberto Bruno

More information

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced

More information

Sepsis overview. Dr. Tsang Hin Hung MBBS FHKCP FRCP

Sepsis overview. Dr. Tsang Hin Hung MBBS FHKCP FRCP Sepsis overview Dr. Tsang Hin Hung MBBS FHKCP FRCP Epidemiology Sepsis, severe sepsis, septic shock Pathophysiology of sepsis Recent researches and advances From bench to bedside Sepsis bundle Severe sepsis

More information

Controversies in Hospital Medicine: Critical Care. Vasopressors, Steroids, and Insulin Therapy

Controversies in Hospital Medicine: Critical Care. Vasopressors, Steroids, and Insulin Therapy Controversies in Hospital Medicine: Critical Care Vasopressors, Steroids, and Insulin Therapy Douglas Fish, Pharm.D. Professor of Pharmacy, University of Colorado Denver Clinical Specialist in Critical

More information

OHSU. Update in Sepsis

OHSU. Update in Sepsis Update in Sepsis Jonathan Pak, MD June 1, 2017 Structure of Talk 1. Sepsis-3: The latest definition 2. Clinical Management - Is EGDT dead? - Surviving Sepsis Campaign Guidelines 3. A novel therapy: Vitamin

More information

Using Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues

Using Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues Using Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues Stephen E. Kahn, PhD, DABCC, FACB Professor and Vice Chair, Clinical Services, Pathology Loyola University

More information

EFFECT OF EARLY VASOPRESSIN VS NOREPINEPHRINE ON KIDNEY FAILURE IN PATIENTS WITH SEPTIC SHOCK. Alexandria Rydz

EFFECT OF EARLY VASOPRESSIN VS NOREPINEPHRINE ON KIDNEY FAILURE IN PATIENTS WITH SEPTIC SHOCK. Alexandria Rydz EFFECT OF EARLY VASOPRESSIN VS NOREPINEPHRINE ON KIDNEY FAILURE IN PATIENTS WITH SEPTIC SHOCK Alexandria Rydz BACKGROUND- SEPSIS Sepsis is defined as life-threatening organ dysfunction caused by a dysregulated

More information

Lung Day University of Washington. June 17, Translating Data Into Practice in the MICU. Richard K. Albert, M.D.

Lung Day University of Washington. June 17, Translating Data Into Practice in the MICU. Richard K. Albert, M.D. Lung Day 2011 University of Washington June 17, 2011 Translating Data Into Practice in the MICU Richard K. Albert, M.D. Chief of Medicine Denver Health Professor of Medicine University of Colorado Adjunct

More information

Αναγκαιότητα και τρόπος ρύθμισης του διαβήτη στους νοσηλευόμενους ασθενείς

Αναγκαιότητα και τρόπος ρύθμισης του διαβήτη στους νοσηλευόμενους ασθενείς Αναγκαιότητα και τρόπος ρύθμισης του διαβήτη στους νοσηλευόμενους ασθενείς Αναστασία Θανοπούλου Επίκουρη Καθηγήτρια Β Παθολογικής Κλινικής Πανεπιστημίου Αθηνών Διαβητολογικό Κέντρο, Ιπποκράτειο Νοσοκομείο

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole

More information

What Should Be the Therapeutic Glycemic Target in Intensive Care Units?

What Should Be the Therapeutic Glycemic Target in Intensive Care Units? What Should Be the Therapeutic Glycemic Target in Intensive Care Units? Irl B. Hirsch, M.D. Professor of Medicine University of Washington School of Medicine Disclosures Research/Grants: Sanofi, Halozyme

More information

4/5/2018. Update on Sepsis NIKHIL JAGAN PULMONARY AND CRITICAL CARE CREIGHTON UNIVERSITY. I have no financial disclosures

4/5/2018. Update on Sepsis NIKHIL JAGAN PULMONARY AND CRITICAL CARE CREIGHTON UNIVERSITY. I have no financial disclosures Update on Sepsis NIKHIL JAGAN PULMONARY AND CRITICAL CARE CREIGHTON UNIVERSITY I have no financial disclosures 1 Objectives Why do we care about sepsis Understanding the core measures by Centers for Medicare

More information

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX AIR Blood Glucose Monitoring System, from Trividia

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia

More information

Staging Sepsis for the Emergency Department: Physician

Staging Sepsis for the Emergency Department: Physician Staging Sepsis for the Emergency Department: Physician Sepsis Continuum 1 Sepsis Continuum SIRS = 2 or more clinical criteria, resulting in Systemic Inflammatory Response Syndrome Sepsis = SIRS + proven/suspected

More information

ACCORDING to the Centers for Disease Control and

ACCORDING to the Centers for Disease Control and Accuracy of Capillary and Arterial Whole Blood Glucose Measurements Using a Glucose Meter in Patients under General Anesthesia in the Operating Room Brad S. Karon, M.D., Ph.D., Leslie J. Donato, Ph.D.,

More information

Disclosures. Objectives. Balancing nutrition and elevated blood glucose. Hyperglycemia in Critical Illness. Hyperglycemia in Critical Illness

Disclosures. Objectives. Balancing nutrition and elevated blood glucose. Hyperglycemia in Critical Illness. Hyperglycemia in Critical Illness Strategies for Managing Hyperglycemia in the Critically Ill Patient on Nutrition Support Balancing nutrition and elevated blood glucose Tuesday, February 3, 2009 11-12:30 Sarah Peterson, MS, RD, CNSD Gourang

More information

Surviving Sepsis Campaign. Guidelines for Management of Severe Sepsis/Septic Shock. An Overview

Surviving Sepsis Campaign. Guidelines for Management of Severe Sepsis/Septic Shock. An Overview Surviving Sepsis Campaign Guidelines for Management of Severe Sepsis/Septic Shock An Overview Mechanical Ventilation of Sepsis-Induced ALI/ARDS ARDSnet Mechanical Ventilation Protocol Results: Mortality

More information

ACUTE ABDOMEN IN DIABETIC PATIENTS ANALYSIS OF COMPLICATIONS AND MORTALITY

ACUTE ABDOMEN IN DIABETIC PATIENTS ANALYSIS OF COMPLICATIONS AND MORTALITY 2014 ILEX PUBLISHING HOUSE, Bucharest, Roumania http://www.jrdiabet.ro Rom J Diabetes Nutr Metab Dis. 21(4):277-284 doi: 10.2478/rjdnmd-2014-0034 ACUTE ABDOMEN IN DIABETIC PATIENTS ANALYSIS OF COMPLICATIONS

More information

Sepsis Bundle Project (SEP) Kathy Wonderly RN, MSEd, CPHQ Consultant Developed: April 2015 Most recent Revision: December 2018

Sepsis Bundle Project (SEP) Kathy Wonderly RN, MSEd, CPHQ Consultant Developed: April 2015 Most recent Revision: December 2018 Sepsis Bundle Project (SEP) Kathy Wonderly RN, MSEd, CPHQ Consultant Developed: April 2015 Most recent Revision: December 2018 Objectives 1. To identify the symptom of severe sepsis and septic shock syndrome.

More information

Inflammation. Sepsis Ladder

Inflammation. Sepsis Ladder Maureen Maloney-Poldek MSN, RN Chamberlain College of Nursing Pathophysiology of sepsis and septic shock How sepsis affects the endocrine system Pathophysiology of adrenal insufficiency Clinical manifestations

More information

Point-of-Care Blood Glucose Meter Accuracy in the Hospital Setting

Point-of-Care Blood Glucose Meter Accuracy in the Hospital Setting In Brief Point-of-care (POC) tests provide analytical information that can be used to make decisions at patients bedside, as opposed to laboratory tests that must be run at a central laboratory. POC testing

More information

Patient safety has become a major global social health problem.

Patient safety has become a major global social health problem. REVIEWS IMAJ VOL 14 april 2012 Glycemic Control in the Intensive Care Unit: Between Safety and Benefit Alexander Samokhvalov MD 1, Raymond Farah MD 2 and Nicola Makhoul MD 1 1 Intensive Care Unit, Western

More information

Effects of Measurement Frequency on Analytical Quality Required for Glucose Measurements in Intensive Care Units: Assessments by Simulation Models

Effects of Measurement Frequency on Analytical Quality Required for Glucose Measurements in Intensive Care Units: Assessments by Simulation Models Clinical Chemistry 60:4 644 650 (2014) Endocrinology and Metabolism Effects of Measurement Frequency on Analytical Quality Required for Glucose Measurements in Intensive Care Units: Assessments by Simulation

More information

Evidence-Based. Management of Severe Sepsis. What is the BP Target?

Evidence-Based. Management of Severe Sepsis. What is the BP Target? Evidence-Based Management of Severe Sepsis Michael A. Gropper, MD, PhD Professor and Vice Chair of Anesthesia Director, Critical Care Medicine Chair, Quality Improvment University of California San Francisco

More information

Online Supplementary Appendix

Online Supplementary Appendix Online Supplementary Appendix This appendix has been provided by the authors to give readers additional information about their work. Supplement to: Lehman * LH, Saeed * M, Talmor D, Mark RG, and Malhotra

More information

John Park, MD Assistant Professor of Medicine

John Park, MD Assistant Professor of Medicine John Park, MD Assistant Professor of Medicine Faculty photo will be placed here park.john@mayo.edu 2015 MFMER 3543652-1 Sepsis Out with the Old, In with the New Mayo School of Continuous Professional Development

More information

FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols

FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society CLINICAL APPLICATIONS FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in

More information

Postoperative Glucose Control and SCIP Measures. Gorav Ailawadi, MD Chief, Adult Cardiac Surgery University of Virginia April 25, 2015

Postoperative Glucose Control and SCIP Measures. Gorav Ailawadi, MD Chief, Adult Cardiac Surgery University of Virginia April 25, 2015 Postoperative Glucose Control and SCIP Measures Gorav Ailawadi, MD Chief, Adult Cardiac Surgery University of Virginia April 25, 2015 Diabetes in CABG Incidence of Diabetes in cardiac surgery increased

More information

Sepsis Early Recognition and Management. Therese Hughes, PhD, MPA, RN

Sepsis Early Recognition and Management. Therese Hughes, PhD, MPA, RN Sepsis Early Recognition and Management Therese Hughes, PhD, MPA, RN 1 Sepsis a Deadly Progression Affects millions around the world each year, killing one in four Contributes to approximately 50% of all

More information

C CONFERENCIAS MAGISTRALES Vol. 36. Supl. 1 Abril-Junio 2013 pp S61-S68 Management of hyperglycemia in the perioperative patient. 39 th Annual Refresher Course on Anesthesiology and Perioperative Medicine,

More information

Accurate glucose measurement is an essential requirement

Accurate glucose measurement is an essential requirement The Journal of TRAUMA Injury, Infection, and Critical Care Error Rates Resulting From Anemia can be Corrected in Multiple Commonly Used Point-of-Care Glucometers Elizabeth A. Mann, MS, RN, Jose Salinas,

More information

GLYCEMIC CONTROL SURVEY

GLYCEMIC CONTROL SURVEY GLYCEMIC CONTROL SURVEY Objective: To gain an understanding of the current state of glycemic control (ie, intensive insulin therapy and frequent blood glucose testing) protocol use in hospital inpatients.

More information

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D. Journal of Diabetes Science and Technology Volume 5, Issue 4, July 2011 Diabetes Technology Society TECHNOLOGY REPORTS Accuracy of the CONTOUR Blood Glucose Monitoring System Bern, B.A., Cheryl Leazenby,

More information

Contraindications to time critical surgery; when not to proceed from the perspective of: The Physician A/Prof Peter Morley

Contraindications to time critical surgery; when not to proceed from the perspective of: The Physician A/Prof Peter Morley Contraindications to time critical surgery; when not to proceed from the perspective of: The Physician A/Prof Peter Morley British Journal of Surgery 2013; 100: 1045 1049 The risk of 30 day mortality

More information

CRRT Fundamentals Pre- and Post- Test. AKI & CRRT Conference 2018

CRRT Fundamentals Pre- and Post- Test. AKI & CRRT Conference 2018 CRRT Fundamentals Pre- and Post- Test AKI & CRRT Conference 2018 Question 1 Which ONE of the following statements regarding solute clearance in CRRT is MOST correct? A. Convective and diffusive solute

More information

INTENSIVE INSULIN THERAPY: A Long History of Conflicting Data.

INTENSIVE INSULIN THERAPY: A Long History of Conflicting Data. INTENSIVE INSULIN THERAPY: A Long History of Conflicting Data. Candice Preslaski, PharmD BCPS Clinical Pharmacist Specialist SICU Denver Health Medical Center December 2014 OBJECTIVES Review the risk factors

More information

Implementing Glucose Control in 2009 and Beyond: Changes in Patterns and Perceptions

Implementing Glucose Control in 2009 and Beyond: Changes in Patterns and Perceptions Implementing Glucose Control in 2009 and Beyond: Changes in Patterns and Perceptions Charles C. Reed MSN, RN, CNRN Patient Care Coordinator Surgical Trauma ICU University Hospital San Antonio, Texas Relationships

More information

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood

More information

Sepsis: Mitigating Denials Amid Definition Disparity

Sepsis: Mitigating Denials Amid Definition Disparity Sepsis: Mitigating Denials Amid Definition Disparity White Paper - April 2017 Sepsis Criteria at a Glance The Society of Critical Care Medicine (SCCM) met in 2016 to update the definition of sepsis. During

More information

A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting

A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting Original papers A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting Güzin Aykal*, Ayşenur Yegin, Özgür Tekeli, Necat Yilmaz Central Laboratories, Antalya

More information

Critical Care Treatment Guidelines

Critical Care Treatment Guidelines Critical Care Treatment Guidelines West Virginia Office of Emergency Medical Services CCT Guidelines CCT Guidelines TABLE OF CONTENTS Preface Acknowledgments Using the Guidelines INITIAL TREATMENT / UNIVERSAL

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

123 Are You Providing Evidence-Based Diabetes Care? - Martin

123 Are You Providing Evidence-Based Diabetes Care? - Martin Donna Martin, DNP, RN, CDE, CMSRN Lewis University Learner will be able to: Identify current inpatient standards of care for patients with diabetes Describe causes of hyperglycemia / hypoglycemia in the

More information

Aina Blood Monitoring System

Aina Blood Monitoring System Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale

More information

VOLUME RESPONSIVENESS IS DIFFERENT FROM NEED FOR FLUIDS BLOOD PRESSURE TARGETS IN SEPSIS

VOLUME RESPONSIVENESS IS DIFFERENT FROM NEED FOR FLUIDS BLOOD PRESSURE TARGETS IN SEPSIS Department of Intensive Care Medicine VOLUME RESPONSIVENESS IS DIFFERENT FROM NEED FOR FLUIDS BLOOD PRESSURE TARGETS IN SEPSIS SEPTIC SHOCK : THE CLINICAL SCENARIO HYPOTENSION DESPITE ADEQUATE VOLUME RESUSCITATION

More information

SEPSIS 2015 DISCLOSURES FINANCIAL DISCLOSURES 9/1/2015. William M. Johnson, MD Nebraska Pulmonary Specialties. William Johnson

SEPSIS 2015 DISCLOSURES FINANCIAL DISCLOSURES 9/1/2015. William M. Johnson, MD Nebraska Pulmonary Specialties. William Johnson SEPSIS 2015 William M. Johnson, MD Nebraska Pulmonary Specialties 1 DISCLOSURES William Johnson No financial interests related to this presentation 2 FINANCIAL DISCLOSURES I do however have 3 children

More information

ESPEN Congress Madrid 2018

ESPEN Congress Madrid 2018 ESPEN Congress Madrid 2018 Dysglycaemia In Acute Patients With Nutritional Therapy Mechanisms And Consequences Of Dysglycaemia In Patients Receiving Nutritional Therapy M. León- Sanz (ES) Mechanisms and

More information

Vasopressors for shock

Vasopressors for shock Vasopressors for shock Background Reviews and Observational Studies Holler 2015. Nontraumatic Hypotension and Shock in the Emergency Department and Prehospital Setting Prevalence, Etiology and Mortality:

More information

Management of Inpatient Hyperglycemia: 2011 Endocrine Society Meeting Hyperglycemia in Critically ill patients in ICU Settings.

Management of Inpatient Hyperglycemia: 2011 Endocrine Society Meeting Hyperglycemia in Critically ill patients in ICU Settings. Management of Inpatient Hyperglycemia: 2011 Endocrine Society Meeting Guillermo E. Umpierrez, MD, Emory University School of Medicine and Jack Leahy, MD, University of Connecticut Hyperglycemia in Critically

More information

Septic Shock. Rontgene M. Solante, MD, FPCP,FPSMID

Septic Shock. Rontgene M. Solante, MD, FPCP,FPSMID Septic Shock Rontgene M. Solante, MD, FPCP,FPSMID Learning Objectives Identify situations wherein high or low BP are hemodynamically significant Recognize complications arising from BP emergencies Manage

More information

The Art and Science of Infusion Nursing Gwen Klinkner, MS, RN, APRN, BC-ADM, CDE

The Art and Science of Infusion Nursing Gwen Klinkner, MS, RN, APRN, BC-ADM, CDE The Art and Science of Infusion Nursing Gwen Klinkner, MS, RN, APRN, BC-ADM, CDE The Importance of Glycemic Control in the Hospital and the Role of the Infusion Nurse ABSTRACT Diabetes is reaching epidemic

More information

The Impact of Measurement Frequency on the Domains of Glycemic Control in the Critically Ill-A Monte Carlo Simulation

The Impact of Measurement Frequency on the Domains of Glycemic Control in the Critically Ill-A Monte Carlo Simulation 566507DSTXXX10.1177/1932296814566507Journal of Diabetes Science and TechnologyKrinsley et al research-article2015 Original Article The Impact of Measurement Frequency on the Domains of Glycemic Control

More information

Hyperglycemia occurs frequently in critically ill patients.

Hyperglycemia occurs frequently in critically ill patients. Mayo Clin Proc, December 2003, Vol 78 Hyperglycemia and Increased Hospital Mortality 1471 Original Article Association Between Hyperglycemia and Increased Hospital Mortality in a Heterogeneous Population

More information

Sepsis: Identification and Management in an Acute Care Setting

Sepsis: Identification and Management in an Acute Care Setting Sepsis: Identification and Management in an Acute Care Setting Dr. Barbara M. Mills DNP Director Rapid Response Team/ Code Resuscitation Stony Brook University Medical Center SEPSIS LECTURE NPA 2018 OBJECTIVES

More information

Hyperglycemia in ACS. Dr. Imhemed Eljazwi

Hyperglycemia in ACS. Dr. Imhemed Eljazwi Hyperglycemia in ACS 2012-5-8 Dr. Imhemed Eljazwi Percentage of Population (n = 1181) Prevalence of Hyperglycemia in 181 Cardiac Patients Without Known Diabetes 100% 75% 50% 66% of AMI patients have

More information

Fluids in Sepsis: How much and what type? John Fowler, MD, FACEP Kent Hospital, İzmir Eisenhower Medical Center, USA American Hospital Dubai, UAE

Fluids in Sepsis: How much and what type? John Fowler, MD, FACEP Kent Hospital, İzmir Eisenhower Medical Center, USA American Hospital Dubai, UAE Fluids in Sepsis: How much and what type? John Fowler, MD, FACEP Kent Hospital, İzmir Eisenhower Medical Center, USA American Hospital Dubai, UAE In critically ill patients: too little fluid Low preload,

More information

CORTICOSTEROID USE IN SEPTIC SHOCK THE ONGOING DEBATE DIEM HO, PHARMD PGY1 PHARMACY RESIDENT VALLEY BAPTIST MEDICAL CENTER BROWNSVILLE

CORTICOSTEROID USE IN SEPTIC SHOCK THE ONGOING DEBATE DIEM HO, PHARMD PGY1 PHARMACY RESIDENT VALLEY BAPTIST MEDICAL CENTER BROWNSVILLE CORTICOSTEROID USE IN SEPTIC SHOCK THE ONGOING DEBATE DIEM HO, PHARMD PGY1 PHARMACY RESIDENT VALLEY BAPTIST MEDICAL CENTER BROWNSVILLE 1 ABBREVIATIONS ACCP = American College of Chest Physicians ARF =

More information

A Children s Bedtime Story

A Children s Bedtime Story A Children s Bedtime Story Setting: University Medical Center, Big Town, USA Scenario: 0500, last admission of the night, 10 previous admissions, all tucked in for the night Patient: 75 year old male with

More information

Hyperglycemia is common among medical and surgical. Clinical Guideline

Hyperglycemia is common among medical and surgical. Clinical Guideline Clinical Guideline Annals of Internal Medicine Intensive Insulin Therapy in Hospitalized Patients: A Systematic Review Devan Kansagara, MD, MCR; Rongwei Fu, PhD; Michele Freeman, MPH; Fawn Wolf, MD; and

More information

Glucose Management in the ICU: The Role of the Pharmacist

Glucose Management in the ICU: The Role of the Pharmacist Objectives Glucose Management in the ICU: The Role of the Pharmacist James Gilmore PharmD, BCPS Senior Pharmacist- Surgical Intensive Care Unit Brigham and Women s Hospital Boston, MA Evaluate primary

More information

escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE

escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE No conflict of interest u Technology, based on Pulse Wave Transit Time u Limits of the technology u Precision

More information

April Dear (Editor):

April Dear (Editor): April 2014 Dear (Editor): Registered Dietitians (RD) play an integral role in patient care in the medical intensive care unit. RD s have increased knowledge in blood glucose control and reducing adverse

More information

NYSDOH Sepsis Q&A Session from February 2018 Data Abstraction Meetings Table of Content

NYSDOH Sepsis Q&A Session from February 2018 Data Abstraction Meetings Table of Content NYSDOH Sepsis Q&A Session from February 2018 Data Abstraction Meetings Table of Content Adherence variables Q: within 3 hours of the start datetime. How can we document that monotherapy was started before

More information

Update in Critical Care Medicine

Update in Critical Care Medicine Update in Critical Care Medicine Michael A. Gropper, MD, PhD Professor and Executive Vice Chair Department of Anesthesia and Perioperative Care Director, Critical Care Medicine UCSF Disclosure None Update

More information

Sepsis Management Update 2014

Sepsis Management Update 2014 Sepsis Management Update 2014 Laura J. Moore, MD, FACS Associate Professor, Department of Surgery The University of Texas Health Science Center, Houston Medical Director, Shock Trauma ICU Texas Trauma

More information

Intensive Insulin in the Intensive Care Unit

Intensive Insulin in the Intensive Care Unit TABLE OF CONTENTS Introduction to Intensive Insulin in Adult Critical Care Patients - UIMCC Guideline for Insulin Infusion in Adult ICU Patients - P&T Committee Formulary Action Intensive Insulin in the

More information

Initial Resuscitation of Sepsis & Septic Shock

Initial Resuscitation of Sepsis & Septic Shock Initial Resuscitation of Sepsis & Septic Shock Dr. Fatema Ahmed MD (Critical Care Medicine) FCPS (Medicine) Associate professor Dept. of Critical Care Medicine BIRDEM General Hospital Is Sepsis a known

More information

IDENTIFYING SEPSIS IN THE PREHOSPITAL SETTING

IDENTIFYING SEPSIS IN THE PREHOSPITAL SETTING IDENTIFYING SEPSIS IN THE PREHOSPITAL SETTING Christopher Hunter, MD, PhD, FACEP Director, Health Services Department Associate Medical Director, Orange County EMS System Medical Director, Orlando Health

More information

Nothing to disclose 9/25/2017

Nothing to disclose 9/25/2017 Jessie O Neal, PharmD, BCCCP Critical Care Clinical Pharmacist University of New Mexico Hospital New Mexico Society of Health-System Pharmacists 2017 Balloon Fiesta Symposium Nothing to disclose 1 Explain

More information

9/25/2017. Nothing to disclose

9/25/2017. Nothing to disclose Nothing to disclose Jessie O Neal, PharmD, BCCCP Critical Care Clinical Pharmacist University of New Mexico Hospital New Mexico Society of Health-System Pharmacists 2017 Balloon Fiesta Symposium Explain

More information