Assessment of a Critical Limit Protocol for Point-of-Care Glucose Testing
|
|
- Erik Barber
- 6 years ago
- Views:
Transcription
1 CLINICAL CHEMISTRY Original Article Assessment of a Critical Limit Protocol for Point-of-Care Glucose Testing GIFFORD LUM, MD A critical limit protocol requiring that all point of care glucose meter readings >22.2 mmol/l (400 mg/dl) and <2.2 mmol/l (40 mg/dl) be immediately confirmed by the laboratory was assessed. A total of 193 (2%) of 9,523 glucose meter determinations (63 patients) were >22.2 or <2.2 mmol/l. One hundred twenty-two (63%) of critically high and low glucose readings were followed up, and 71 (37%) results were not. Seventy-seven percent (55 of 71) of results without follow up were in patients with multiple glucose meter/central lab comparisons, suggesting that users may have thought it unnecessary to confirm such results. Split sample quality control specimens showed good correlation (r = 0.927) between glucose meter and central lab results, whereas correlation for follow-up glucose results was poorer (r = 0.793), perhaps reflecting time delay in obtaining a lab sample. For follow-up results, only 18% of high/low critical limit glucose meter readings were confirmed by drawing a lab specimen within 10 minutes. Fifty-eight percent were in 17 patients with multiple previous glucose meter readings, suggesting that users may have thought it less urgent to confirm a sequence of such results. Eleven follow-up results (9%) showed a >50% discordance between glucose meter/central lab with three (27%) glucose meter errors, emphasizing the need to confirm critically high/low glucose meter results to avoid potential errors. The critical limit protocol now requires that only the initial critically high/low glucose meter reading be confirmed by the lab and that these patients now be followed with lab values until glucose levels are between mmol/l ( mg/dl) before the glucose meter can again be used. (Key words: Point-of-care; Glucose testing; Critical high/low limit; Hyperglycemia; Hypoglycemia) Am J Clin Pathol 1996; 106: Point-of-care glucose testing in some sites has become an integral part of the management of the diabetic patient. Guidelines for point-of-care glucose testing should be established to optimize its use, including the establishment of high and low critical limits designed to provide urgent physician notification of patients who may need immediate diagnostic evaluation and possible therapy for lifethreatening emergencies. Because of the acute deleterious effects of hyperglycemia, which may include acidosis, coma and death, and of hypoglycemia, which may include confusion, dizziness, blurred vision, deterioration in cognitive function, coma and death, medical centers have established critical limits for high and low glucose levels that are used for urgent physician notification. Generally, the critical limit for hyperglycemia varies between mmol/l (110-1,000 mg/dl) and for hypoglycemia between mmol/l (30-70 mg/dl). 1 ' 2 A recent national survey of 92 medical facilities, including 20 trauma centers, found a mean critical limit for hyperglycemia of 26.9 mmol/l (484 mg/dl) (standard deviation [SD] of 8.0 mmol/l [144 mg/dl]) and for hypoglycemia of 2.6 mmol/l (46 mg/dl) (SD of 0.4 mmol/l [7 mg/dl]). 1 At this medical center, for point-of-care glucose meters, the critical limit for hyperglycemia was established as any glucose level > 22.2 mmol/l (400 mg/dl) and for hypoglycemia as any glucose level < 2.2 mmol/l (40 mg/dl). The critical limit glucose protocol requires that all glucose readings >22.2 and <2.2 mmol/l be verified by submitting a STAT specimen to the central clinical laboratory for confirmation by a reference glucose method. To assess the effectiveness of our present critical limit protocol for glucose meters, I decided to study all critically high and low glucose meter determinations for a period of 3 months. MATERIALS AND METHODS From the Pathology and Laboratory Medicine Service, Brockton/ West Roxbury Veteran Affairs Medical Center, Boston, Massachusetts and Harvard Medical School, Boston, Massachusetts. Manuscript received March 12, 1996; revision accepted May 17, Address reprint requests to Dr. Lum: Pathology and Laboratory Medicine Service, Brockton/West Roxbury Veteran Affairs Medical Center VFW Parkway, Boston, MA Patient Samples and Analytical Methods Patient blood samples for point-of-care glucose testing were obtained by nursing staff by fingerprick. Analysis for blood glucose was performed on the ward using a Satellite G (Sat G) Blood Glucose Monitoring System (Medisense, Waltham, MA) and sensor electrode test strips for 390
2 LUM 391 A Glucose Criti Limit Protocol blood glucose, a drug reagent technology using an electrochemical detection technique measuring the electrical current generated from the enzymatic cleavage of glucose. 3 Blood glucose results were reported without correcting the result for the effect of plasma volume and are linear between mmol/l. Patient samples for centralized laboratory testing were obtained by venipuncture in heparinized vacutainer blood collection tubes, mixed and submitted to the central laboratory for glucose determination. Plasma glucose levels were determined on a Beckman Synchron CX-7 automated chemistry analyzer (Beckman Instruments, Brea, CA) using a glucose oxidase method with an oxygen electrode 4 and are linear between mmol/l. Reference Range and Definition of Critical Limits for Hyper and Hypoglycemia The reference range for whole blood glucose levels was mmol/l (65-95 mg/dl) and for plasma glucose levels, mmol/l ( mg/dl). The critical limit for hyperglycemia for the glucose meter was defined as any blood glucose level > 22.2 mmol/l (400 mg/ dl) and for hypoglycemia as any level <2.2 mmol/l (40 mg/dl). Study Subjects This study, involving the use of medical data from human subjects, fell into the category of exempt research in accordance with the federal policy for the protection of human subjects (38 CFR Part 16). All patients studied were adults (> 18 years of age); no pediatric patients were included in this study. Quality Control Program for the Satellite G Glucose Meter A total of 200 trained and certified registered nurses are responsible for operating 21 Sat G meters on 17 in house wards, including ambulatory care sites and emergency room. An ancillary glucose committee (laboratory, nursing, medical staff) is responsible for the overall direction and quality control of the glucose meter program. All Satellite G glucose meter users must be trained and certified before they can use the glucose meter in a clinical situation. Recertification of operators is required on a semi-annual basis or when there is evidence that an operator is not adhering to established medical center guidelines. Quality control material (high and low control) is performed once per day on each day of glucose meter use for patient results. Once a week, the glucose level of an identical heparinized plasma split sample is analyzed on the glucose meter and by the clinical laboratory. Proficiency samples (Ancillary Whole Blood Glucose survey, College of American Pathologists [CAP]) 5 are done 3 times per year for each glucose meter. Critical Limit for Hyperglycemia and Hypoglycemia: Protocol and Assessment The critical limit protocol at this facility requires that all critically high and low glucose meter readings be confirmed by submitting a confirmatory STAT glucose specimen to the central lab, regardless of whether the same patient had multiple previous glucose meter results. All glucose meter results are entered directly into the computer by the user, with notation of time, and for critically high and low glucose readings, the time of drawing a confirmatory laboratory specimen is noted. The critical limit protocol was assessed using the following criteria: (1) glucose meter results with/without follow up; (2) time interval between glucose meter reading and drawing lab specimen; (3) concordance distribution for glucose meter readings and central lab results; (4) review of results and of patient data for glucose meter/ central laboratory follow-up results with >50% discordance; and (5) assessment of patient outcome data for results with no lab follow up. RESULTS Quality Control Program and Satellite G Glucose Meter Data Over a 6-month period, the coefficient of variation (CV) for the low control (2.6 mmol/l, 47 mg/dl) was 7.8% and for the high control (19.4 mmol/l, 350 mg/ dl), the CV was 6.7%. For 200 split samples, the mean whole blood glucose was 5.3 mmol/l (96 mg/dl) for the Satellite G vs a mean plasma glucose of 6.3 mmol/l (114 mg/dl) for the central laboratory (mean difference of 16%, range mmol/l, mg/dl). The correlation coefficient r was 0.927, and linear regression was as follows: Satellite G result (y) = 0.87 X laboratory result For the last three mailings of the Ancillary Whole Blood glucose surveys of the College of American Pathologists WBC A,B,C (1995), 5 only one unacceptable result was found for the 21 glucose meters used at this facility. Critical High/Low Glucose meter Readings During the 3-month period of study, there were a total of 9,523 glucose meter determinations. Figure 1 shows Vol. 1 No. 3
3 392 CLINICAL CHEMISTRY Original Article 193 results inpatients Only 18% of specimens were drawn for lab confirmation within 10 minutes of a critically high and low glucose meter reading, with an additional 20% and 16% within 20 and 30 minutes, respectively. Hence, slightly more than half of the laboratory comparison specimens were drawn within 30 minutes of the initial critical limit high or low glucose meter reading, but in some instances, nurses waited as long as 2 hours to confirm the critical limit result. Seventy-one (58%, 71 of 122) of the followup results were in 17 patients with multiple glucose meter readings, but in only 4 of these patients was the confirmatory laboratory specimen drawn within 10 minutes. >22.2 mmow. 168(51) <22 mmom. 25(12) Concordance Between Glucose Meter Readings and Central Laboratory FQkNV Up-106 (42) NO FOllOWUp-62 (17) Previous- 50(8) Clerical error-5(1) No previous-9 (8) UP-16 (9) NO FOllOWUp-9 (5) Prevlous- 5(2) Clerical error-2(d No previous-2 (2) FIG. I. Flow chart for critical high/low glucose meter readings. the data for the 193 glucose meter results (2.0% of total), which were >22.2 or <2.22 mmol/l. For glucose meter results more than the critical limit of 22.2 mmol/l, operators followed up the high result in 106 instances (42 patients) and did not follow up 62 results (17 patients). For results with no follow up, 81 % (50 of 62) were in eight patients who had previous glucose meter results (consistent with hyperglycemia), 5% (3 of 62) were clerical errors in 1 patient, and 15% (9 of 62) were in eight patients with no previous glucose meter/ laboratory comparisons. Thus, the majority of results that were not followed up were in patients who already had a sequence of high glucose results confirming hyperglycemia. For the 25 glucose meter readings <2.2 mmol/l (12 patients), there was follow up for 16 results (9 patients) and no follow up for 9 results (5 patients). For the results with no follow up, 56% (5 of 9) were in two patients with previous glucose meter/lab results that confirmed hypoglycemia, 22% (2 of 9) were clerical errors in one patient, and 22% (2 of 9) were in two patients with no previous glucose meter/laboratory comparisons. As previously noted, a correlation coefficient (r) of was found for split sample quality control specimens, whereas a r = was found between the glucose meter and central lab glucose results for the 122 follow-up specimens. It is possible that point-of-care glucose meters/central lab comparisons are poor because the initial result is not followed up in a timely manner. Quality control split sample results indicate that when performed with little or no time difference, the glucose meter results have a good correlation with the central lab. Figure 3 shows the concordance distribution for the glucose meter readings versus the central lab for split sample (quality control procedure) results and for follow-up specimens. Lower glucose meter versus laboratory results were designated as negative, whereas higher glucose meter versus laboratory results were designated as positive. For split sample quality control comparisons, 88% of glucose meter results are lower and none of the glucose meter results vary by >50% compared to the central lab. In contrast, for the 122 glucose meter results with follow up, 58% were lower and 9% showed more than 50% variation compared to the central lab. Results Exceeding 50% Concordance Between the Glucose Meter/Laboratory (Table 1) Glucose meter results were erroneous for two patients with critically high glucose meter readings. Patients 7-9 showed relatively large percentage differences between 25 Percent Time Between Critical High/Low Glucose Meter Reading and Laboratory Specimen (Fig. 2) Minutes FIG. 2. Time interval between glucose meter results and time that laboratory specimen was drawn for specimens with follow up. A.J.C.P.-September 1996
4 LUM 393 A Glucose Critical Limit Protocol % of results three died within a month, two with acute myocardial infarction, and one with a lung malignancy (poor outcome not directly related to the hyperglycemia). Two critically low glucose meter readings in two patients were not followed up. Both of these patients were treated for hypoglycemia and subsequently followed by central laboratory glucose results without adverse outcomes. DISCUSSION * ^V" ^V> -yv 1 ^V* ^P & «?V" «?V> ^f ^ Jf -?\P >* J? >* J? <b* N* 1? <!? ** **,** H Split specimen I % Concordance Follow Up specimens FIG. 3. Concordance distribution (10 percentile intervals) for glucose meter versus central laboratory glucose results. the glucose meter reading and the laboratory attributable to treatment for hypoglycemia initiated before the laboratory specimen was drawn for analysis. The hypoglycemic result in patient 10 could not be confirmed by the laboratory, and the glucose meter result was erroneous. Thus, overall, for these 11 discordant results, 3 (27%) were found to be glucose meter errors. Outcome for Patients With No Glucose Meter/Central Laboratory Follow up (Table 2) Eight hyperglycemic patients were appropriately treated with subsequent lab monitoring of glucose levels; The glucose results of the Satellite G glucose meter (whole blood) and the central laboratory's reference glucose method (plasma) are not directly interchangeable. Typically, the glucose level in plasma will be higher than the glucose level in whole blood cells because the average water content of plasma is approximately 93%, whereas it is only about 71 % in red blood cells. 6 A mathematically derived factor of 1.12, assuming a hematocrit of 45% and a red cell:plasma water ratio of 0.76, 7 may be used to convert whole blood glucose to plasma glucose. However, neither of these assumptions may be true in many patients especially in those with dehydration, hyperosmolality, and hyperglycemia. 7 Despite this difference in whole blood/plasma glucose values, for split samples of glucose meter/central laboratory comparisons, a significant correlation (r = 0.927) was found. A less significant correlation (r = 0.793) was found between glucose meter and central lab follow-up glucose results. The relatively poor comparison between glucose meter and central lab in situations where there TABLE 1. SUMMARY OF GLUCOSE METER AND LABORATORY GLUCOSE RESULTS WITH >50% DISCORDANCE Patient Satellite G Laboratory (mmolll) Time Difference No. (mmolll) [% difference] (min) Diagnosis Outcome (-72) 5.8 (+82) 7.2 (+70) 2.2 (-83) 3.1 (-55) 3.3 (-200) 7.4 (-363) 6.4 (-276) 3.8 (-138) 13.5 (-864) 22.6 (-1955) Pancreatic carcinoma Diabetes, end stage renal disease Angina, CABG, no diabetic history Brittle diabetes, hypoglycemic episodes Diabetes with ketoacidosis Insulin-dependent diabetes Diabetes, hypoglycemia thought to be secondary to poor p.o. intake and high-dose glyburide Insulin-dependent diabetes Diabetes, markedly hypoglycemic Metastatic renal cell carcinoma, steroid-induced worsening of diabetes Died 4 days after high glucose noted Patient not treated, no glucose > 11.1 mmol/l, glucose meter error Patient not treated, glucose meter error Patient treated for hypoglycemia Patient treated for hypoglycemia Patient treated for hypoglycemia Glyburide stopped, patient treated with glucose before laboratory specimen drawn Patient treated for low glucose before laboratory specimen drawn Patient treated with dextrose bolus at same time laboratory specimen drawn Hypoglycemia not confirmed by laboratory, no laboratory result < 12.8 mmol/l, patient not treated for hypoglycemia, glucose meter error CABG = coronary artery bypass graft. Vol. 106-No. 3
5 394 CLINICAL CHEMISTRY Original Article TABLE 2. SUMMARY OF OUTCOME IN PATIENTS WITH NO GLUCOSE METER FOLLOW-UP Patient No. Satellite G (mmoi/l) Diagnosis Outcome > <1.1 <1.1 Diabetes, quadriplegic, prostate carcinoma Spinal cord injury patient with diabetes and non-q wave myocardial infarct X 2 Diabetes with below knee amputation and osteomyelitis Non-small cell carcinoma of the lung Diabetic with third-degree burn Spinal cord injury patient with insulin-dependent diabetes Poorly controlled insulindependent diabetes Diabetes and osteomyelitis 84-year-old with cognitive decline, on ADA diet with? hypoglycemia Cirrhosis of liver and hepatitis A, B, and C, alcoholic and drug dependence, diabetes Patient died 20 days after high glucose noted, transmural myocardial infarct, patient treated with insulin but died 22 days after high glucose noted (acute subendocardial MI) and followed by laboratory results but died 2 days after high glucose noted and followed by laboratory and then followed by laboratory and followed by laboratory Patient changed to regular diet, treated with glucose, no evidence of glucose intolerance Patient treated for hypoglycemia and followed by laboratory results ADA = American Dietetic Association; Ml = myocardial infarction. were delays in central lab analysis may well be related to the time factor because, if used as intended (little or no time differential), the glucose meter and central lab glucose results have good correlation. Another factor that may account for the poor correlation for glucose meter/ central lab results for follow-up specimens would be initiation of treatment for hyper or hypoglycemia. There is strong suggestive support for this explanation because there is no correlation (r = 0.107) between time interval and glucose differences (initial glucose meter and central lab glucose result). Approximately one third of critically high and low glucose meter results (37%, 71 of 193) were not followed up, a strikingfindingof this study. Seventy-seven percent (55 of 71) of these results lacking follow up were in patients who already had a sequence of high or low glucose results, suggesting that users may have felt it unnecessary to reconfirm the abnormal glucose reading with a separate lab specimen. Many studies have focused on the acceptable turnaround time between actual test ordering and receipt of results by the clinical team. An acceptable rapid response time during critical emergencies of 5 minutes or less from the ordering of the test to test result receipt has been recently recommended. 8 A recent study found a mean turnaround time of 8.5 minutes for ward glucose meter determinations from specimen collection to test completion. 9 We found that only 18% of specimens were actually drawn for lab confirmation within 10 minutes, a less than optimal response. Fifty-eight percent of the follow-up results were in 17 patients who had multiple glucose meter readings and were known to be hyper or hypoglycemic; hence, this time delay may reflect less urgency to confirm a sequence of such glucose results on the part of users. Other reasons cited for delay included lack of time ("too busy"), waiting for phlebotomist to draw specimen, and results of glucose meter considered reliable for therapeutic action. Our data found that 9% of glucose meter results showed a >50% variation compared to the central lab for those results with follow up, indicating that it is important to confirm critically high and low glucose meter results because approximately 25% of the discordant results were glucose meter errors. No quality control split sample comparisons between the glucose meter and central lab showed a >50% variation. For glucose meter results that were not followed up (patients subsequently followed with central lab glucose A.J.C.P.- September 1996
6 LUM 395 A Glucose Criti Limit Protocol results), outcome data indicated that this course of action resulted in no adverse consequences directly related to hyper or hypoglycemia in 10 such patients. Patients with initially critically high/low glucose results were treated appropriately and then followed by the central lab without adverse outcome. An excellent level of precision for glucose meter quality assurance has been achieved by our nurse users (CVs were 7.8% and 6.7% at low (2.6 mmol/l) and at high (19.4 mmol/l) glucose levels, respectively). A recent consensus conference recommended performance goals for glucose meter testing of a total error (analytical plus user) of < 10% at glucose levels between mmol/ L ( mg/dl) 100% oftime,' oracvof <4%. 7 " A more recent consensus conference recommended a target performance for future glucose meters of an analytical error of 5%.' 2 These target goals clearly may not be achievable in the real world because a recent CAP survey found CVs for glucose meters ranged from 4.2% 35.3%, 5 with CVs for the Medisense Satellite G of 6.8% to 9.7%. 5 Did point-of-care glucose testing result in a decrease in laboratory workload? I compared the total number of glucose levels performed by the central lab before and after the introduction of the glucose meter at this facility. There were 20,406 laboratory glucose measurements for a randomly selected preglucose meter quarter compared to 24,028 glucose determinations for the quarter studied, a 18% increase, (the hospital census and patient population mix were comparable for the 2 quarters). Thus, glucose meters did not result in a decrease in lab glucose workload, a finding similar to those of others who found an increased annual lab glucose workload of approximately 10% after the introduction of glucose meters. 13 The medical center protocol for the critical limit high/ low for glucose meters has been modified to incorporate the findings in this study. These findings included relatively poor compliance with the requirement to follow up all glucose meter results in every instance of a critically high and low glucose meter reading, delay between noting an abnormal glucose reading and drawing a confirmatory lab specimen, the need to confirm glucose meter results because of potential erroneous glucose meter results and no adverse outcome directly related to hyper or hypoglycemia for patients whose glucose meter results were subsequently followed with central lab results. The protocol now requires that critically high/low glucose meter readings be confirmed by drawing a STAT lab specimen but that such patients now be followed with laboratory glucose results until the plasma glucose level is within the range of mmol/l ( mg/ dl), selected for ease of recall of conventional concentration units, at which time the glucose meter can again be used. This protocol should result in reducing the confusion among users regarding the need to confirm repetitive critically high/low glucose meter readings in the same patient and result in reducing the potential for glucose meter errors. REFERENCES 1. Kost GJ. Critical limits for urgent clinician notification at US medical centers. JAMA 1990;263: KostGJ. Using critical limitsto improve patient outcome. Medical Laboratory Observer 1993;25: Marks V, Dawson A. Rapid stick method for determining bloodglucose concentration. BMJ 1965:i: Kadish AH, Litle RL, Sternberg JC. A new and rapid method for determination of glucose by measurement of rate of oxygen consumption. Clin Chem 1968; 14: College of American Pathologists Whole Blood Glucose. Multiple Site Series 1 Survey, 1995 Set WBG A,B,C, Northfield, Illinois. 6. Howanitz PJ, Howanitz JH, Henry JB. Carbohydrates. In: Henry JB, ed. Clinical Diagnosis and Management by Laboratory Methods, ed 18. Philadelphia: WBSaunders. 199l.pp Rainey PM, Jatlow P. Monitoring blood glucose meters [Editorial]. Am J Clin Pathol 1995; 103: Kost GJ. Guidelines for point-of-care testing; Improving patient outcomes. Am J Clin Pathol 1995; 104(Suppl):SI 11-S Winkelman JW, Wybenga DR. Tanasijevic MJ. The fiscal consequences of central vs distributed testing of glucose. Clin Chem 1994;40: American Diabetes Association. Consensus statement on selfmonitoring of blood glucose. Diabetes Care 1987; 10: Fraser CG, Petersen PH. Desirable Performance Standards for Imprecision and Bias in Alternate Sites: The Views of Laboratory Professionals. Arch Pathol Lab Med 1995; 119: American Diabetes Association. Self-monitoring of blood glucose. Diabetes Care 1994; 17: Lee-Lewandrowski E, Laposata M, Eschenbach K, et al. Utilization and cost analysis of bedside capillary glucose testing in a large teaching hospital: Implications for managing point-of-care testing. Am J Med 1994;97: No. 3
Testing Blood Glucose at the Bedside in a Chronic Care Hospital
CHEMISTRY Robert M. Greendyke, MD F. Ronald Gifford, MT(ASCP) Testing Blood Glucose at the Bedside in a Chronic Care Hospital We can provide tests that are fast, inexpensive, or accurate. Pick two. Anonymous
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess
More informationCAP Laboratory Improvement Programs. Comparative Analytical Costs of Central Laboratory Glucose and Bedside Glucose Testing
CAP atory Improvement Programs Comparative Analytical Costs of Central atory Glucose and Bedside Glucose Testing A College of American Pathologists Q-Probes Study Peter J. Howanitz, MD; Bruce A. Jones,
More informationCAP Laboratory Improvement Programs. Utility of Repeat Testing of Critical Values. A Q-Probes Analysis of 86 Clinical Laboratories
CAP Laboratory Improvement Programs Utility of Repeat Testing of Critical Values A Q-Probes Analysis of 86 Clinical Laboratories Christopher M. Lehman, MD; Peter J. Howanitz, MD; Rhona Souers, MS; Donald
More informationTechnical Evaluation of Five Glucose Meters With Data Management Capabilities
Clinical Chemistry / LABORATORY EVALUATION OF GLUCOSE METERS Technical Evaluation of Five Glucose Meters With Data Management Capabilities Jeffrey]. Chance, PhD, Dai J. Li,MD, PhD, Kerrie A. Jones, MS,
More informationEquivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System
Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th
More informationIt s Just a Waived Glucose, Isn t It?
It s Just a Waived Glucose, Isn t It? What Is the Next Step? Becky Damiani, MT (ASCP), Senior Inspection Specialist Laboratory Accreditation Program Objectives Understand the CLIA requirements surrounding
More informationNOTE: The first appearance of terms in bold in the body of this document (except titles) are defined terms please refer to the Definitions section.
TITLE TREATMENT OF HYPERGLYCEMIA - ADULT SCOPE Provincial: Acute Care APPROVAL AUTHORITY Clinical Operations Executive Committee SPONSOR Diabetes, Obesity & Nutrition Strategic Clinical Network PARENT
More informationHyperglycemia Procedure: Management and Treatment Adult Patients
Approved by: Chief Operating Officer; and Chief Medical Officer Hyperglycemia Procedure: Management and Treatment Corporate Policy & Procedures Manual Number: VII-B-465 Date Approved January 23, 2018 Date
More informationRelationship between glucose meter error and glycemic control efficacy
Relationship between glucose meter error and glycemic control efficacy Brad S. Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Department of Laboratory Medicine and Pathology Mayo Clinic
More informationChapter 14 Elderly, Home and Long-term Care Collections. Objectives:
EXERCISE 11: BEDSIDE GLUCOSE TESTING Textbook: Skills: Chapter 14 Elderly, Home and Long-term Care Collections 15 points Objectives: 1. Define diabetes mellitus. 2. Compare and contrast: Type 1 Diabetes,
More informationDipstick Testing of Urine Can It Replace Urine Microscopy?
Dipstick Testing of Urine Can It Replace Urine Microscopy? MARY C. MORRISON, M.D. AND GIFFORD LUM, M.D. One thousand consecutive urine specimens were studied to assess the sensitivity of a commercially
More informationAina Blood Monitoring System
Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale
More informationDetermination of Delay in :flirn Around Time (TAT) of Stat Tests and its Causes: an AKUH Experience
Determination of Delay in :flirn Around Time (TAT) of Stat Tests and its Causes: an AKUH Experience F. Bilwani,I. Siddiqui,S. Vaqar ( Section of Chemical Pathology, Department of Pathology, Aga Khan University
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More informationMultiple studies are available concerning the use of
SPECIAL ARTICLE Evaluation of the Biosite Ò Quantitative Whole Blood D-dimer Assay and Comparison With the biomérieux VIDAS Ò D-dimer Exclusion Test Validation and Utility For Use in the Central Laboratory
More informationDIABETES AND RAMADAN FASTING
DIABETES AND RAMADAN FASTING Dr. A. Nigam, M.B.B.S., M.D. (Medicine) Specialist Internal Medicine Al Zahrawi Hospital, Ras Al Khaimah, U.A.E. It is estimated that UAE s population currently stands at approximately
More informationFDA Perspective - Public Health Notification: Potentially Fatal Errors with GDH-PQQ Glucose Monitoring Technology Courtney C. Harper, Ph.D.
FDA Perspective - Public Health Notification: Potentially Fatal Errors with GDH-PQQ Glucose Monitoring Technology Courtney C. Harper, Ph.D., Director, Division of Chemistry and Toxicology Devices Office
More informationComparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration
Clinical Chemistry / Whole Blood Creatinine for egfr Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration Nichole
More informationContractor Name: Novitas Solutions, Inc. Contractor Number: Contractor Type: MAC B. LCD ID Number: L34834 Status: A-Approved
LCD for Blood Glucose Monitoring in a Skilled Nursing Facility (SNF) (L34834) Contractor Name: Novitas Solutions, Inc. Contractor Number: 12502 Contractor Type: MAC B LCD ID Number: L34834 Status: A-Approved
More informationUsing Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues
Using Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues Stephen E. Kahn, PhD, DABCC, FACB Professor and Vice Chair, Clinical Services, Pathology Loyola University
More informationValidity and Reliability of a Glucometer Against Industry Reference Standards
514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationGLYCEMIC CONTROL SURVEY
GLYCEMIC CONTROL SURVEY Objective: To gain an understanding of the current state of glycemic control (ie, intensive insulin therapy and frequent blood glucose testing) protocol use in hospital inpatients.
More informationPoint of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter
Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter The Nova Stat Strip Blood Glucose meter is UHB s approved Glucometer
More informationLaboratory Results Timeliness as a Quality Attribute and Strategy
Laboratory Results Timeliness as a Quality Attribute and Strategy Joan H. Howanitz, MD, and Peter J. Howanitz, MD Key Words: Turnaround time; Strategy; Quality improvement Abstract Although timeliness
More informationFor In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Lactate REF A95550 For In Vitro Diagnostic Use ANNUAL REVIEW Refer to Review and Revision Coversheet at the front of each method. PRINCIPLE INTENDED USE reagent,
More informationFreeStyle Lite A Blood Glucose Meter That Requires No Coding
Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood
More informationDIABETES MANAGEMENT DISCHARGE COMMUNICATION (DM-DC) AUDIT TOOL
DIABETES MANAGEMENT DISCHARGE COMMUNICATION (DM-DC) AUDIT TOOL Facility: Date: Data Collector s name: Email/Phone: Purpose: To evaluate your facility practices regarding communication of requisite diabetes
More informationGB/T Translated English of Chinese Standard: GB/T
Translated English of Chinese Standard: GB/T 19634-2005 www.chinesestandard.net Sales@ChineseStandard.net GB ICS 11.100 C 44 National Standard of the People s Republic of China GB/T 19634-2005 In vitro
More informationThe Many Faces of T2DM in Long-term Care Facilities
The Many Faces of T2DM in Long-term Care Facilities Question #1 Which of the following is a risk factor for increased hypoglycemia in older patients that may suggest the need to relax hyperglycemia treatment
More informationQuality Specifications for POCT
Quality Specifications for POCT the patient journey perspective Christopher P Price Department of Clinical Biochemistry, University of Oxford, Oxford, UK Quality Specifications fit-for-purpose QUALITY
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More informationIt s a Gas! Issues in the Blood Gas Laboratory. D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC
It s a Gas! Issues in the Blood Gas Laboratory D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC Learning Objectives After participating in this session, you will be able
More informationUSING THE ACCESS AMH ASSAY IN YOUR LABORATORY
INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
More informationReport Reference Guide. THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1
Report Reference Guide THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1 How to use this guide Each type of CareLink report and its components are described in the following sections.
More informationChapter Goal. Learning Objectives 9/12/2012. Chapter 25. Diabetic Emergencies
Chapter 25 Diabetic Emergencies Chapter Goal Use assessment findings to formulate field impression & implement treatment plan for patients with diabetic emergencies Learning Objectives Describe pathophysiology
More informationGLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. GLUH Glucose REF B24985 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE
More information1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)
Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch
More informationThe Effect of Alcohol Prep Pads and Blood Drop Number On Capillary Blood Glucose Values
University of New Hampshire University of New Hampshire Scholars' Repository Honors Theses and Capstones Student Scholarship Spring 2017 The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary
More informationMethod Precision and Frequent Causes of Errors Observed in Point-of-Care Glucose Testing. A Proficiency Testing Program Perspective
Method Precision and Frequent Causes of Errors Observed in Point-of-Care Testing A Proficiency Testing Program Perspective Berna Aslan, MD, 1 Julia Stemp, ART, 1 Paul Yip, PhD, 2 and Jane Gun-Munro, MLT,
More informationAnalytical and Clinical Performance of the epoc Blood Analysis System. Experience at a Large Tertiary Academic Medical Center
AJCP / Original Article Analytical and Clinical Performance of the epoc Blood Analysis System Experience at a Large Tertiary Academic Medical Center Brie A. Stotler, MD, MPH, 1,2 and Alexander Kratz, MD,
More informationApril Dear (Editor):
April 2014 Dear (Editor): Registered Dietitians (RD) play an integral role in patient care in the medical intensive care unit. RD s have increased knowledge in blood glucose control and reducing adverse
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More informationSupplementary Data. Correlation analysis. Importance of normalizing indices before applying SPCA
Supplementary Data Correlation analysis The correlation matrix R of the m = 25 GV indices calculated for each dataset is reported below (Tables S1 S3). R is an m m symmetric matrix, whose entries r ij
More informationPrimary Stroke Center Quality & Performance Measures
Primary Stroke Center Quality & Performance Measures This section of the manual contains information related to the quality performance of Primary Stroke Centers. Brain Attack Coalition Definitions Recognition
More informationPOTASSIUM/K. 3.6 mmol/l
POTASSIUM/K Potassium is measured by ion-selective electrode potentiometry. In the calculation of results for potassium, concentration is related to potential through the Nernst equation. The i-stat System
More informationCareLink. software REPORT REFERENCE GUIDE. Management Software for Diabetes
CareLink Management Software for Diabetes software REPORT REFERENCE GUIDE How to use this guide Each type of CareLink report and its components are described in the following sections. Report data used
More informationRN Diabetes Medication Titration Protocol
Metformin (Glucophage) Biguanide Medication titration: Time 0 APC or MD Start Metformin 500mg (500mg in the morning, taken with food) Week 1 Week 2 Metformin 1000mg (500mg in the morning and 500mg in the
More information1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014
Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The
More informationUse of Blood Lactate Measurements in the Critical Care Setting
Use of Blood Lactate Measurements in the Critical Care Setting John G Toffaletti, PhD Director of Blood Gas and Clinical Pediatric Labs Professor of Pathology Duke University Medical Center Chief, VAMC
More informationFour is the Floor Symptoms can be felt at higher levels if control is poor Worth confirmation using BG meter if at all possible
Sandra Coats Diabetes Specialist Nurse 1 Hypoglycaemia Hyperglycaemia Diabetes and Illness sick day Diabetic Ketoacidosis HONK/HHS 2 What is Hypoglycaemia BG levels below 4mmol/l. Four is the Floor Floor
More informationFreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols
Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society CLINICAL APPLICATIONS FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in
More informationDIABETIC KETOACIDOSIS (DKA) K E M I A D E Y E R I, P G Y - 1
DIABETIC KETOACIDOSIS (DKA) K E M I A D E Y E R I, P G Y - 1 QUESTION # 1 7 year old boy comes to the ER with a 2 week history of abdominal pain and weight loss. Further history reveals polyuria and polydipsia,
More informationClinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES
Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES Accuracy of Roche Accu-Chek Inform Whole Blood Capillary, Arterial, and Venous Glucose Values in Patients Receiving Intensive Intravenous
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. HDLD HDL Cholesterol 650207 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by
More informationBern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.
Journal of Diabetes Science and Technology Volume 5, Issue 4, July 2011 Diabetes Technology Society TECHNOLOGY REPORTS Accuracy of the CONTOUR Blood Glucose Monitoring System Bern, B.A., Cheryl Leazenby,
More informationTitle: Glucose Testing: StatStrip Glucose Meter Page 1 of 6
Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Original: 5/02 Minor Revision: 9/15 Full Review: 1/15 Replaces: 1/15 Responsible Person: Point of Care Laboratory Specialist Approving Committee:
More informationBlood Glucose monitoring during extra-corporeal renal therapy and plasmapheresis.
Blood Glucose monitoring during extra-corporeal renal therapy and plasmapheresis. Lead Clinician: Dr. R. Diwakar Implementation date: July 2013 Last updated: August 2017 Last review date: Planned review
More informationZACHARY COMMUNITY SCHOOLS
PARENTAL CONSENT/ RELEASE OF INFORMATION/ AND STUDENT WITH DIABETES CONTRACT Student s Name D.O.B. Parent/Guardian Home Phone # Work/Cell Phone # School Teacher Grade Physician Office # Fax # 1. I give
More informationThe Virtues and Pitfalls of Implementing a New Test
The Virtues and Pitfalls of Implementing a New Test James H. Nichols, Ph.D., DABCC, FACB Professor of Clinical Pathology, Microbiology and Immunology Associate Medical Director for Clinical Operations
More informationSupplement Table 1. Definitions for Causes of Death
Supplement Table 1. Definitions for Causes of Death 3. Cause of Death: To record the primary cause of death. Record only one answer. Classify cause of death as one of the following: 3.1 Cardiac: Death
More informationestimated to be complicated by diabetes, and up to 0.5 percent of pregnancies occur in mothers who already have diabetes (Ang, Howe, &
Title Subtitle author byline Insulin Revisited f in the Maternity Setting For many women who would previously have been instructed to avoid pregnancy, the use of insulin has allowed them to become pregnant
More informationReport Reference Guide
Report Reference Guide How to use this guide Each type of CareLink report and its components are described in the following sections. Report data used to generate the sample reports was from sample patient
More informationJanice Lazear, DNP, FNP-C, CDE DIAGNOSIS AND THE OLDER ADULT
Janice Lazear, DNP, FNP-C, CDE DIAGNOSIS AND THE OLDER ADULT Objectives u At conclusion of the presentation the participant will: 1. Discuss challenges to glycemic control unique in the older population
More information3. Screening Subject Identification Screening Overview
3. Screening 3.1 Subject Identification Each site will be responsible for identifying and recruiting participants into the study. It is known that screening methods vary across sites. It is, however, important
More informationDisclosures. Glycemic Control in the Intensive Care Unit. Objectives. Hyperglycemia. Hyperglycemia. History. No disclosures
Disclosures Glycemic Control in the Intensive Care Unit No disclosures Jorie Frasiolas, Pharm.D., BCPS Clinical Pharmacy Manager, CTICU NewYork-Presbyterian Hospital Columbia University Medical Center
More informationHEMOCHRON. Whole Blood Coagulation Systems
HEMOCHRON Whole Blood Coagulation Systems Citrated Activated Partial Thromboplastin Time (APTT) Cuvette Correlation Protocol for HEMOCHRON Microcoagulation Instruments MSIG:131 10/06 Dear Medical Professional:
More informationAccuracy of Portable Blood Glucose Monitoring
CLINICAL CHEMISTRY Accuracy of Portable Blood Glucose Monitoring Effect of Glucose Level and Prandial State DINO A. VALLERA, M.D., MICHAEL G. BISSELL, M.D., PH.D., M.P.H., AND WILLIAM BARRON, M.D. Glucose
More informationIn Vitro Diagnostic Glucose Test System
CDRH Final Guidance 2-Page Cover Sheet Guidance for Industry In Vitro Diagnostic Glucose Test System Document issued on: July 6, 1998 U.S. Department Of Health and Human Services Food and Drug Administration
More informationObjectives / Learning Targets: The learner who successfully completes this course will be able to demonstrate understanding of the following concepts:
Objectives / Learning Targets: The learner who successfully completes this course will be able to demonstrate understanding of the following concepts: Insulin s function in the body. The basics of diabetes
More informationPatient: 55 y female (ambulatory)
Disclosures Speaking Honoraria Radiometer (Canada) Nova Biomedical, Draeger Roche Diagnostics (Canada) Research Support (Reagents, Instrumentation, Travel) Nova Biomedical Abbott Laboratories (Canada)
More informationCAP Laboratory Improvement Programs. Clinical Consequences of Specimen Rejection
CAP Laboratory Improvement Programs Clinical Consequences of Specimen Rejection A College of American Pathologists Q-Probes Analysis of 78 Clinical Laboratories Donald S. Karcher, MD; Christopher M. Lehman,
More informationImproved IPGM: Demonstrating the Value to both Patients and Hospitals
Improved IPGM: Demonstrating the Value to both Patients and Hospitals Osama Hamdy, MD, PhD, FACE Medical Director, Inpatient Diabetes Program Joslin Diabetes Center Harvard Medical School, Boston, MA Cost
More informationDiabetic Ketoacidosis (DKA) Critical Care Guideline Two Bag System
Critical Care Guideline Two Bag System Inclusion Criteria (Definition of DKA): Blood glucose (BG) > 200 mg/dl Acidosis (bicarbonate < 15 or blood gas ph < 7.3) Associated glycosuria, ketonuria &/or ketonemia
More informationSARASOTA MEMORIAL HOSPITAL
SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE CARE OF THE INTRAPARTUM PATIENT RECEIVING CONTINUOUS INTRAVENOUS INSULIN ADMINISTRATION (obs25) DATE: REVIEWED: PAGES: 9/93 12/17 1 of 5 PS1094 ISSUED
More informationMethod Comparison Report Semi-Annual 1/5/2018
Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community
More informationBLOOD IS COMPLEX ANALYZING IT SHOULDN T BE DxH 500 * *Not available for sale in the U.S.
BLOOD IS COMPLEX ANALYZING IT SHOULDN T BE DxH 500 * *Not available for sale in the U.S. Introducing the DxH 500, an Open-vial Hematology System *. Quality 5-part Differential in a Compact Design The first
More informationWhite Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard
White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced
More informationSODIUM/NA. See below for information on factors affecting results. Certain substances, such as drugs, may affect analyte levels in vivo.
SODIUM/NA Sodium is measured by ion-selective electrode potentiometry. In the calculation of results for sodium, concentration is related to potential through the Nernst equation. The i-stat System uses
More informationMEDICAL DEFINITIONS REFERENCE GUIDES
MEDICAL DEFINITIONS REFERENCE GUIDES What do the reference guides do? The reference guides provide updated definitions for certain claimable medical conditions under the following policies: St Andrew s
More informationNHS Greater Glasgow & Clyde Managed Clinical Network for Diabetes
Guidelines for the Diagnosis of Diabetes Mellitus NHS Greater Glasgow & Clyde Managed Clinical Network for Diabetes Lead Authors: Dr Brian Kennon, Dr David Carty June 2015 Review due: December 2016 Diagnosis
More informationDKA : Diabetic Ketoacidosis & HHS: Hyperlgycemic Hyperosmolar Syndrome Protocol. Glycemic Task Force September 2014
DKA : Diabetic Ketoacidosis & HHS: Hyperlgycemic Hyperosmolar Syndrome Protocol Glycemic Task Force September 2014 Hyperglycemic Crises: Pathophysiology DKA HHS Hyperglycemia DKA HHS Umpierrez, In Shoemaker,
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System
Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,
More informationHEALTH SERVICES POLICY & PROCEDURE MANUAL
PAGE 1 of 5 PURPOSE To assure that DOP inmates with Diabetes are receiving high quality Primary Care for their condition. POLICY All DOP Primary Care Providers are to follow these guidelines when treating
More informationFigure 2.1: Glucose meter
CHAPTER TWO: MONITORING TECHNOLOGIES 2.1 Introduction Glucose monitoring is a method of self-testing glucose (blood sugar) levels for the management of diabetes. Traditionally, it involves pricking the
More informationBlood Glucose Monitoring
Blood Glucose Monitoring What is Glucose? A simple sugar that enters the diet as part of sucrose, lactose, or maltose Part of a polysaccharide called dietary starch Most of the body s energy comes from
More informationNPQR 2018 Qualified Clinical Data Registry (QCDR) Measures 21_18247_LS.
NPQR 2018 Qualified Clinical Data Registry (QCDR) Measures 21_18247_LS MEASURE ID: NPQR1 MEASURE TITLE: Notification to the Ordering Provider Requesting Myoglobin or CK-MB in the Diagnosis of Suspected
More informationPharmacy Coverage Guidelines are subject to change as new information becomes available.
Metformin tablet SR 24-hour modified release oral tablet Coverage for services, procedures, medical devices and drugs are dependent upon benefit eligibility as outlined in the member's specific benefit
More informationCoding spotlight: diabetes provider guide to coding the diagnosis and treatment of diabetes
Medicaid Managed Care December 2018 provider guide to coding the diagnosis and treatment of diabetes Diabetes mellitus is a chronic disorder caused by either an absolute decrease in the amount of insulin
More informationState of the Art of HbA1c Measurement
State of the Art of HbA1c Measurement XXI Congreso Latinoamericano de Patologia Clinica Y XLII Congreso Mexicano de Patologia Clinica, Cancun October 2012 Randie R. Little, Ph.D. NGSP Network Laboratory
More informationGlucose, glucose oxidase/peroxidase method, photometry, diabetes mellitus.
Quantitative Determination TEAS Related topics Glucose, glucose oxidase/peroxidase method, photometry, diabetes mellitus. Principle Blood sugar is the key energy provider for our cells. The primary hormone
More informationDiabetic Emergencies. Chapter 15
Diabetic Emergencies Chapter 15 Diabetes- is a disorder of glucose metabolism or difficulty metabolizing carbohydrates, fats and proteins Full name is diabetes mellitus which refers to the presence of
More informationc onsensus Statement on
c onsensus Statement on Self-Monitor ing of Blood Glucose Self-monitoring of blood glucose (SMBG) has become a major adjunct to the care of individuals with diabetes mellitus in the past decade. It is
More informationADAG Study Group Data Links A1C Levels with Empirically Measured Blood Glucose Values - New Treatment Guidelines Will Now be Needed
529638DSTXXX10.1177/1932296814529638Journal of Diabetes Science and TechnologyKlonoff research-article2014 Editorial ADAG Study Group Data Links A1C Levels with Empirically Measured Blood Glucose Values
More informationThe Second Report of the Expert Panel on Detection,
Blood Cholesterol Screening Influence of State on Cholesterol Results and Management Decisions Steven R. Craig, MD, Rupal V. Amin, MD, Daniel W. Russell, PhD, Norman F. Paradise, PhD OBJECTIVE: To compare
More information