Accuracy of Portable Blood Glucose Monitoring

Size: px
Start display at page:

Download "Accuracy of Portable Blood Glucose Monitoring"

Transcription

1 CLINICAL CHEMISTRY Accuracy of Portable Blood Glucose Monitoring Effect of Glucose Level and Prandial State DINO A. VALLERA, M.D., MICHAEL G. BISSELL, M.D., PH.D., M.P.H., AND WILLIAM BARRON, M.D. Glucose was determined on capillary and venous blood samples from 274 adult diabetics by three different methodologies: the Glucoscan 2000 and Accu-Check II portable glucose meters (capillary) and the Kodak Ektachem 700 analyzer (venous). Both glucose meters correlated significantly with the Ektachem results. A significant positive bias was found for the Glucoscan compared with Ektachem, not found with the Accu-Check II. The Accu-Check performed better than the Glucoscan at venous plasma glucose levels < 1 g/l. The mean error of Glucoscan determinations was significantly greater and biased positive when the measurement was performed within 4 hours of eating, whereas no such effect was seen with Accu-Check error. Multiple regression analysis revealed that the Glucoscan measurement was independently influenced by both venous plasma glucose and prandial state, whereas the Accu-Check II measurement was not dependent on either variable. The within-run precision for both glucose meters were comparable. (Key Words: Portable glucose monitoring; Accuracy.) Am J Clin Pathol 1991;95: Portable capillary blood glucose meters have been used extensively in both inpatient and outpatient settings for many reasons, including patient and physician convenience and cost. 1,2 Although the accuracy and precision of this technology have been addressed by a number of previous investigators, 3,4 many of these studies have had considerable limitations. For example, sample sizes are often small, 5 " 7 meter accuracy frequently is addressed only through determination of the correlation coefficient between meter and reference laboratory methods, 5,8 " 12 and the impact of ambient plasma glucose levels on accuracy often is neglected. Furthermore, we are aware of only one study that has addressed the impact of prior feeding on the concordance between capillary blood glucose measured by a portable meter and venous plasma glucose de- termined by a laboratory reference methodology. 12 Consideration of this factor is important because there may be significant peripheral tissue glucose extraction in the postprandial state, 13 resulting in venous plasma levels that are considerably lower than capillary values. The present investigation was designed to address a number of these issues by performing a large number of capillary blood glucose determinations on two currently popular instruments, analyzing differences between portable meter and standard laboratory measurements as a function of both ambient plasma glucose level and the time since the patient's last meal. METHODS Study subjects included 274 adult diabetics on whom a blood glucose level had been requested by their primary From the Departments of Pathology, Medicine, and Obstetrics care and physician in the University of Chicago Health Service, Gynecology, University of Chicago Hospitals, Chicago, Illinois. a large internal medicine clinic with more than 50,000 Received September 18, 1989; accepted for publication July 24, outpatient visits per year. Participants included patients Dr. Vallera is currently with the Department of Pathology, Hinsdale on dietary therapy only, on oral hypoglycemic agents and Hospital, Hinsdale, Illinois; Dr. Bissell is currently with the Department on insulin. Each subject was asked when he had last eaten, of Clinical Pathology, City of Hope National Medical Center, Duarte, California. and the time was recorded to the nearest 30 minutes. Supported by Boehringer Mannheim Diagnostics, Indianapolis, Indiana. immediately into a tube containing potassium oxalate Blood was obtained from an antecubital vein and placed Address reprint requests to Dr. Bissell: Department of Clinical Pathology, City of Hope National Medical Center, 1500 East Duarte Road, (14.0 mg) and sodium fluoride (17.5 mg) (Becton Dickinson Vacutainer Systems, Rutherford, NJ). Plasma Duarte, California ob- 247

2 248 CLINICAL CHEMISTRY tained from the specimen was analyzed for glucose in the General Clinical Chemistry Laboratory of the University of Chicago Hospitals on the Ektachem 700 analyzer (Eastman Kodak Co., Rochester, NY), which uses a glucose oxidase-based reflection densitometric method. Within 4 minutes after venipuncture, the patient's finger was punctured with a spring-loaded lancet to obtain a drop of capillary whole blood. This was analyzed for glucose with either the Glucoscan 2000 with Glucoscan test strips (Lifescan, Inc., Mountain View, CA) or the Accu- Check II with Chemstrip bg test strips (Boehringer Mannheim Diagnostics, Indianapolis, IN). All determinations were performed by a single nurse who was trained in the operation and quality control of both instruments as directed by the manufacturers. One instrument of each type was used for the entire study. The precision of the Ektachem 700 was assessed by performing 20 repeated glucose determinations on a single sample over 20 consecutive minutes and by analyzing the same sample once daily on 20 consecutive days. Precision of each portable meter was evaluated by daily determinations of the glucose concentration of a standard test solution (provided by the manufacturer) for 20 days. In addition, the precision of the entire portable glucose meter methodology was assessed for each instrument by performing 20 repeated capillary blood glucose determinations on a single subject over a minute period. The statistical approach used to assess the agreement between each meter and the laboratory reference method is a modification of that suggested by Bland and Altman. 14 First, least squares linear regression analysis was used to assess correlations between meter and Ektachem determinations. Next, two-tailed paired /-tests were performed to evaluate the difference between meter and Ektachem determinations (referred to here as "meter bias" or "error") and to compare the magnitude of these two differences for the two meters. In addition, 95 confidence intervals (CI) were calculated for the mean bias for each meter. To determine whether meter bias was dependent on plasma glucose level, we performed a linear regression of the difference between meter and Ektachem measurements on the Ektachem value. To further clarify this relationship, data were stratified in 1 g/l increments according to the Ektachem value, and an analysis of variance was performed, followed by two-tailed two sample Mests to assess differences between the meters at various levels of venous plasma glucose. To analyze the impact of prior meal ingestion on monitor accuracy, meter and Ektachem values were grouped by the length of time that had passed since the last meal. Preliminary analysis separating data into four groups (0-2.0 hours, hours, hours, and >6.0 hours) indicated that results for groups 1 and 2 were similar, as were those for groups 3 and 4; therefore, data were combined into two groups, hours and >4.0 hours since the last meal, and two-tailed, paired and two-sample t- tests were performed to assess group differences. Finally, a multiple regression analysis was performed to assess the independent contributions of (1) the Ektachem value, and (2) the time since the last meal, on meter error. Data are reported as mean ± 1 SD. A P value of <0.05 was considered significant, except where multiple testing was performed, in which case the critical P value was adjusted by the Bonferroni method (both uncorrected and Bonferroni P values are presented). All calculations were performed on a microcomputer using SYSTAT software. Precision evaluation experiments were performed on both the Accu-Check II and the Ektachem according to the National Committee for Clinical Laboratory Standards (NCCLS) Guidelines EP5T. Briefly, manufacturers' supplied glucose control material was analyzed in duplicate, in two runs per day, for 20 days. Two analyte concentrations were tested for each method. This data was analyzed to provide an estimate of within-run and total imprecision. For the Accu-Check II, the determinations were performed by the same trained nurse who performed the patient determinations. The Ektachem study was performed by trained medical technologists in the General Clinical Chemistry Laboratory. The Glucoscan 2000 could not be included in this phase of the study because of the nature of its control materials. These were packaged in small plastic vials and, according to the manufacturer's instructions, are to be used within 24 hours of opening. This would necessitate the use of 20 different vials of control material for each analyte concentration to complete the study and would introduce another source of variability into the data. RESULTS A total of 277 determinations were made with the Glucoscan 2000, and 232 were performed on the Accu-Check II. There was a highly significant correlation between capillary (meter) and venous (Ektachem) glucose values for both instruments: Glucoscan: meter value = 0.84 X (Ektachem value) , r = 0.94 Accu-Check: meter value = 0.92 X (Ektachem value) , r = 0.94 Despite these results, further evaluation disclosed significant differences in performance of the two meters. Capillary glucose values measured with the Glucoscan averaged 1.91 ± 0.84 g/l (range = g/l; n = 277), whereas venous plasma glucose values on the same A.J.C.P. February 1991

3 VALLERA, BISSELL, AND BARRON 249 Accuracy of Portable Blood Glucose Monitoring subjects measured with the Ektachem averaged 1.85 ± 0.96 g/l (range = g/l), a highly significant difference (P = 0.001). The average absolute difference between the Glucoscan and Ektachem was ± g/l (95 confidence interval = ), and the average percentage difference was 8.8 ±21.6 (95 confidence interval = ). The observation that neither of these 95 confidence intervals includes 0 supports the given result (i.e., the mean Glucoscan 2000 and Ektachem values are significantly different from one another). Mean capillary glucose values determined with the Accu-Check II were 1.95 ± 0.99 g/l (range = g/l; n = 232), and paired venous samples measured on the Ektachem averaged 1.94 ± 1.01 g/l (range = g/l), a difference that was not significant (P = 0.93). The mean difference between the Accu-Check and Ektachem was ± 0.35 g/l (95 confidence interval = ), data which confirm the lack of difference between Accu-Check and Ektachem determinations. For the Glucoscan, 79.8 of determinations were within 20 of the correct value, whereas 87.9 of the Accu-Check measurements were in this range. Fourteen Glucoscan (5.0) and five Accu-Check (2.2) values deviated more than 50 from that determined on the Ektachem. Direct comparison of the average bias of the two meters indicated a highly significant difference when measured as a percentage of the Ektachem value. (Glucoscan: 8.8 ± 21.6; Accu-Check: 1.1 ± 21.8, P < 0.001), and a trend in the same direction when the bias was expressed in g/l (Glucoscan: ± g/l; Accu- Check: ± g/l, P = 0.038, Bonferroni critical value: P = 0.025). Next, we analyzed the effect of the venous plasma glucose level (Ektachem measurement) on the accuracy of each meter. For the Glucoscan there was a significant linear correlation between the meter error expressed in either absolute [(Glucoscan value) - (Ektachem value) = X (Ektachem value) , r = 0.486, P < 0.001] or percentage terms [ difference between Glucoscan and Ektachem = X (Ektachem) , r = 0.477, P < 0.001]. This effect was much less evident for the Accu- Check II, and was significant only for the error measured in absolute terms [difference (g/l) = X (Ektachem) , r = 0.229, P< 0.001; difference = X (Ektachem) + 4.8, r = 0.09, P = 0.18]. This effect is better appreciated when the data are grouped in 1-g/L intervals of the Ektachem value (Table 1). The average bias of the Glucoscan is more than 20 (95 confidence interval = 19-28) when the Ektachem value is less than 1 g/l, becomes minimal in the range 2-3 g/l, and is more than -10 when the venous plasma level is >4 g/l. In contrast, the error of the Accu-Check at the lowest glucose values averaged -6 ± 15 (95 confidence interval = -2 11; P < vs. Glucoscan error, Bonferroni critical P = 0.01) and varied from 6.3 to +7.8 in other ranges of venous glucose (= not significant vs. Glucoscan for each of these ranges). Thus, the Accu-Check performed better (i.e., had less bias) than the Glucoscan primarily when the venous plasma glucose was <lg/l. It also is of interest to note that all meter determinations with a difference from the Ektachem value of greater than 90 occurred in the most commonly encountered venous plasma glucose range (1-2 g/l) for both the Glucoscan (n = 4) and Accu-Check (n = 3). These seven determinations were made on four different subjects; on two of these, other meter measurements were within 20 of Ektachem values. When these seven outliers were excluded, the mean measurement error in the 1-2 g/l range was 8 ± 13 (95 confidence interval = 6-11) for the Glucoscan and 3 ± 16 (95 confidence interval = ) for the Accu-Check (P = 0.023, Bonferroni critical P = 0.01). Assessment of the effect of prior meal ingestion on measurement bias indicated further differences in meter performance (Table 2). The mean error of Glucoscan determinations was significantly greater and biased in a positive direction when the measurement was performed within 4 hours of eating, whereas recent food intake did TABLE 1. METER BIAS IN ABSOLUTE (mg/dl) AND PERCENTAGE TERMS AS A FUNCTION OF VENOUS PLASMA GLUCOSE LEVEL Venous Plasma Glucose (q/l) 3'L Glucoscan 2000 Bias H 9IL Accu-Check II Bias n < ^ ± * ± ± ± ± ±25* ±8-11 ± ± ± ± ± ±0.67-6± 15 8±30f 1 ± ± When four outliers { difference from Ektachem >90) were excluded, differences were 0.11 ± 0.20 g/l and 8 ± 13. t When three outliers were excluded, differences were 0.05 ± 0.25 g/l and 3 ± 16. Results are expressed as mean ± SD; n = number of paired determinations. Vol. 95 No. 2

4 250 CLINICAL CHEMISTRY TABLE 2. EFFECT OF PRIOR MEAL INGESTION ON METER BIAS GLUCOSCAN 2000 Time Since Last Meal Instrument Units 0-4 hours n >4 hours n Glucoscan Accu-Check g/l g/l ± ± ± 0.38* 2.1 ± ± 0.28* 4.5 ± 16.9f ± ± I * P < 0.02 for the comparison with the 0-4 hour time period. t P < for the comparison with the 0-4 hour time period. X P = for the comparison with Glucoscan. P < for the comparison with Glucoscan. Bonferroni critical P = Results are expressed as mean ± SD; n = number of paired determinations DIFFERENCE (X) not appear to have a significant effect on the Accu-Check meter error. Furthermore, the Accu-Check performed significantly better (i.e., less bias) compared with the Glucoscan only for measurements made less than 4 hours after the last meal. Finally, multiple regression analysis was performed to determine whether the level of venous plasma glucose and the time since the last meal had independent effects on meter bias. Regression coefficients (Table 3) demonstrate that the bias of the Glucoscan, in both g/l and percent, is independently influenced by both plasma glucose and prandial state. In contrast, the bias of the Accu- Check, as a percentage of the Ektachem value, was dependent upon neither of these variables, although the bias in g/l was significantly influenced, albeit to a smaller degree than for the Glucoscan, by the venous plasma glucose. For the Accu-Check II meter, the overall mean of all determinations using the high glucose control material was g/l. The within-run precision was g/l, whereas the between-run precision was g/l (coefficient of variation = 8.7). With the low concentration glucose control, the overall mean was g/l, with the within-run precision being g/l and the betweenrun precision being g/l (coefficient of variation = 6.4). TABLE 3. COEFFICIENT OF MULTIPLE REGRESSION ANALYSIS EVALUATING THE DEPENDENCE OF METER BIAS ON PLASMA VENOUS GLUCOSE AND TIME SINCE LAST MEAL Instrument Units Glucoscan Accu-Check Glucose Level g/l ± ±0.01 g/l ± Time Interval < ± < ± < ± ± Results are expressed as mean ± SE. "Time period since last meal" is categorized into "0-4 hours" or ">4 hours." ACCUCHECK II B DIFFERENCE () FIG. 1. A (upper). Percent differences between EKTACHEM and GLUCOSCAN 2000 values grouped according to the level of venous glucose determined by the EKTACHEM. Note that the ambient level of venous glucose has a significant effect on meter bias and that when venous glucose is less than 1 g/l, bias is approximately +20. B (lower). Percent differences between EKTACHEM and ACCUCHECK II values grouped according to the level of venous glucose determined by the Ektachem. There is no significant relationship between meter bias and venous plasma glucose levels. With the Ektachem, using the high concentration glucose control, the overall mean was 4.28 g/l, with the within-run precision being g/l, between-run precision being 4.08 g/l (coefficient of variation = 0.95). For the low glucose control, the overall mean was g/l, with within-run precision being g/l and between-run precision being g/l (coefficient of variation = 5.5). In the whole method precision study, both the Accu- Check II and Glucoscan 2000 were tested. Using the Glucoscan 2000, the first volunteer's mean blood glucose concentration was 0.94 g/l, with SD of g/l (coefficient of variation = 6.5). The second volunteer was tested with the Accu-Check II, yielding a mean blood glu- A.J.C.P. February 1991

5 VALLERA, BISSELL, AND BARRON 251 Accuracy of Portable Bh <d Glucose Monitoring cose concentration of g/l, with SD of g/l and coefficient of variation of 5.5. Repetitive sampling (n = 20) of the Ektachem high glucose control resulted in a mean concentration of g/l, with SD of g/l and coefficient of variation of For the low control, the mean concentration was g/l, with SD of g/l and coefficient of variation of DISCUSSION This study critically evaluated the performance of two popular blood glucose meters in a busy clinical setting and attempted to identify variables that might account for any observed bias. As reported by others, 815 we found that both meters' readings were highly correlated (r = 0.94) with the laboratory reference method; however, additional analysis disclosed that, on average, the Accu-Check II provided more accurate determinations, with a mean bias of only 1. i compared with 8.8 for the Glucoscan Yet more detailed evaluation disclosed that the greater bias of the latter instrument occurred primarily at lower levels of plasma glucose and when the patient had eaten within several hours of capillary glucose measurement. There have been numerous published reports that purportedly have investigated the accuracy of portable blood glucose monitors. 5 " 8,10,11,15 " 20 However, as discussed in detail by Bland and Altman, 14 many investigators inappropriately report only the correlation coefficient between measurements made by meter and some reference methodology. 5 " 12 Such a statistic fails to provide clinically relevant information {i.e., how much does the meter determination deviate from the true [laboratory reference] value, and does this deviation change at various levels of venous plasma glucose). Godine and colleagues 15 evaluated the accuracy of the Accu-Check meter when used by staff nurses and found a mean error of 8 ± 7, whereas Landon and colleagues, 20 studying the same meter, observed a mean deviation of approximately 15. Brooks and co-workers' 8 found a mean bias of 0.06 ± 0.30 g/l for the Accu-Check II and 0.08 ± 0.27 g/l for the Glucoscan 2000 (percentage data not available). Results of average bias for the two meters in the present study were in the ranges cited, with a mean error for the Accu-Check II of 1 ± 22 (0 ± 0.35 g/l) and 9 ± 22 (0.06 ± 0.33 g/l) for the Glucoscan Although the mean measurement error (as a of the venous glucose level) of both glucose meters was relatively small (<10; 95 confidence interval for mean bias: Glucoscan: ; Accu-Check II: ), this does not provide an estimate of a given capillary glucose determination, a parameter of considerable clinical importance. The latter is better estimated by the 95 limits of agreement (i.e., mean error ± 2 SD), which in the pres- ent study allow one to be 95 certain that a given glucoscan measurement was between -34 and +52 of the laboratory reference value and that an Accu-Check determination was between 43 and +45 of this value. When viewed in this light, the two meters do not appear, on average, to differ as much as suggested by comparison of the mean bias. Furthermore, the range of these limits make it clear that portable blood glucose meter results must be interpreted with caution, particularly in those ranges where a change in therapy may have substantial adverse effects. 21 Overall estimates of glucose meter bias as described may be misleading if this bias is significantly dependent on the level of venous plasma glucose (e.g., the bias may be large at one glucose level and minimal at another). Indeed, we found a significant linear relationship between Glucoscan 2000 meter error and venous plasma glucose, with highest errors at the lowest levels of plasma glucose, a phenomenon not observed for the Accu-Check II. Two previous studies of earlier models of the Accu-Check meter provided similar results. 6,20 When we stratified venous glucose levels (Table 1), it was discovered that the major difference between the two meters occurred at concentrations of <1 g/l, at which level the mean bias of the Glucoscan was +23 (95 limits of agreement: ), whereas the average error for the Accu-Check II was 6 (95 limits of agreement: ). These observations may have important clinical implications. For example, when rigid metabolic control is being attempted, such as is standard during diabetic pregnancy 10,22 (for which desired fasting and postprandial glucose levels are <1 and <1.2 g/l, respectively), Glucoscan 2000 results may lead to excessive insulin administration. On the other hand, use of Accu-Check II would be somewhat more likely to lead to unacceptably high blood glucose levels. Such conclusions highlight the need to understand glucose meter performance at levels of blood glucose that are likely to lead to changes in therapy. Another factor that might affect meter accuracy is the proximity of glucose measurement to the patient's last meal because significant forearm glucose uptake occurs in the postprandial state. 13 This could lead to capillary (meter) glucose levels that are significantly higher than venous (Ektachem) if food had been consumed shortly before measurement. 23 We are aware of only one study that addressed the impact of this phenomenon on the apparent bias of portable meter capillary glucose determinations. 12 In the fasting state, capillary determinations exceeded venous glucose levels by 14; however, 1 and 2 hours after a 100-g glucose load, capillary measurements were, respectively, 20 and 19 higher than venous determinations. Surprisingly, we found a similar effect only with the Glucoscan 2000 and not with the Accu-Check Vol. 95 No. 2

6 252 CLINICAL CHEMISTRY II (Table 2). Furthermore, multiple regression analysis indicated that this was not simply due to the positive bias of the Glucoscan 2000 at certain levels of venous glucose. Thus, we have no explanation for this observation, but the data indicate that a greater allowance for error must be taken into account if the Glucoscan 2000 is used to measure capillary glucose in the postprandial state. In considering all of these statistics, one must balance factors such as availability, turnaround time, convenience, and portability against the difference in accuracy and precision in various testing systems. These factors may become more or less important in different clinical settings. For example, accuracy and precision may be more important in measurements used to adjust insulin doses in a "rigidly" controlled diabetic, whereas ease of operation and availability may have increased importance in less tightly controlled patients. REFERENCES 1. American Diabetes Association. Bedside blood glucose monitoring in hospitals. Diabetes Care 1986;9: Belsey R, Morrison JI, Whitlow KJ, Baer DM, Nelson S, Hardwick DF. Managing bedside glucose testing in the hospital. JAMA 1987;258: Rasaiah B. Self-monitoring of the blood glucose level: potential sources of inaccuracy. Can Med Assoc J 1985;132: Villeneuve ME, Murphy J, Mazze RS. Evaluating blood glucose monitors. Am J Nurs 1985;11: Gifford-Jorgensen RA, Borchert J, Hassanein L, Tilzer GA, Eaks GA, Moore WV. Comparison offiveglucose meters for self-monitoring of blood glucose by diabetic patients. Diabetes Care 1986;9: James K, Latham D, Marrero J, Barrett DA. Capillary blood glucose measured by the laboratory phlebotomy team. Laboratory Medicine 1985;16: Webb DJ, Lovesay JM, Ellis A, Knight AH. Blood glucose monitors: a laboratory and patient assessment. Br Med J 1980;1: Chiasson JL, Morrisset R, Hamet P. Precision and costs of techniques for self-monitoring of serum glucose levels. Can Med Assoc J 1984;130: Drucker RF, Williams DRR, Price CP. Quality assessment of blood glucose monitors in use outside the hospital laboratory. J Clin Pathol 1983;36: Marquette GP, Dillard T, Bietla S, Niebyl JR. The accuracy of visual and meter determinations of blood glucose with the use of Chemstrip bg. Am J Obstet Gynecol 1985;153: Shapiro B, Savage PJ, Lomatch D, et al. A comparison of accuracy and estimated cost of methods for home blood glucose monitoring. Diabetes Care 1981;4: Weiner CP, Faustich M, Burns J, Fraser L, Whitaker L, KJugman M. The relationship between capillary and venous glucose concentration during pregnancy. Am J Obstet Gynecol 1986; 155: Jackson RA, Peters N, Advani U, et al. Forearm glucose uptake during the oral glucose tolerance test in normal subjects. Diabetes 1973;22: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: Godine JE, Hurxthal K, Nathan DM. Bedside capillary glucose measurement by staff nurses in a general hospital. Am J Med 1986;80: Aziz S, Hsiang YH. Comparative study of home blood glucose monitoring devices: Visidex, Chemstrip bg, Glucometer, and Accu- Check bg. Diabetes Care 1983;6: Barr B, Leichter SB, Taylor L. Bedside capillary glucose monitoring in the general hospital. Diabetes Care 1984;7: Brooks KE, Rawal N, Henderson AR. Laboratory assessment of three new monitors of blood glucose: Accu-Check II, Glucometer II, and Glucoscan Clin Chem 1986;32: Clements RS, Keane NA, Kirk KA, Boshell BR. Comparison of various methods for rapid glucose estimation. Diabetes Care 1981;4: Landon MB, Cembrowski GS, Gabbe SG. Capillary blood glucose screening for gestational diabetes: a preliminary investigation. Am J Obstet Gynecol 1986;155: Cohen FE, Sater B, Feingold KR. Potential danger of extending SMBG techniques to hospital wards. Diabetes Care 1986;9: : 22. Langer O, Mazze RS. Diabetes in pregnancy: evaluating self-monitoring performance and glycemic control with memory-based reflectance meters. Am J Obstet Gynecol 1986;155: Eriksson KF, Fex G, Trell E. Capillary-venous differences in blood glucose values during oral glucose tolerance test. Clin Chem 1983;29:993. A.J.C.P. February 1991

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th

More information

What Do Statistics Really Tell Us About the Quality of the Data from Self-monitoring of Blood Glucose?

What Do Statistics Really Tell Us About the Quality of the Data from Self-monitoring of Blood Glucose? What Do Statistics Really Tell Us bout the Quality of the Data from Self-monitoring of lood Glucose? R. F. Dedrick, W. K. Davis Michigan Diabetec Research and Training Center, University of Michigan Medical

More information

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose

More information

Validity and Reliability of a Glucometer Against Industry Reference Standards

Validity and Reliability of a Glucometer Against Industry Reference Standards 514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference

More information

Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES

Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES Accuracy of Roche Accu-Chek Inform Whole Blood Capillary, Arterial, and Venous Glucose Values in Patients Receiving Intensive Intravenous

More information

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803) Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole

More information

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff

More information

Chapter 14 Elderly, Home and Long-term Care Collections. Objectives:

Chapter 14 Elderly, Home and Long-term Care Collections. Objectives: EXERCISE 11: BEDSIDE GLUCOSE TESTING Textbook: Skills: Chapter 14 Elderly, Home and Long-term Care Collections 15 points Objectives: 1. Define diabetes mellitus. 2. Compare and contrast: Type 1 Diabetes,

More information

Clinical Impact of Prandial State, Exercise, and Site Preparation on the Equivalence of Alternative-Site Blood Glucose Testing

Clinical Impact of Prandial State, Exercise, and Site Preparation on the Equivalence of Alternative-Site Blood Glucose Testing Clinical Care/Education/Nutrition O R I G I N A L A R T I C L E Clinical Impact of Prandial State, Exercise, and Preparation on the Equivalence of Alternative- Blood Glucose Testing DAWN M. BINA, CMT ROBYN

More information

Ignell, Claes; Berntorp, Kerstin. Published in: Scandinavian Journal of Clinical & Laboratory Investigation DOI: /

Ignell, Claes; Berntorp, Kerstin. Published in: Scandinavian Journal of Clinical & Laboratory Investigation DOI: / Evaluation of the relationship between capillary and venous plasma glucose concentrations obtained by the HemoCue Glucose 201+ system during an oral glucose tolerance test Ignell, Claes; Berntorp, Kerstin

More information

Performance of three portable blood glucose monitoring devices used in a veterinary application

Performance of three portable blood glucose monitoring devices used in a veterinary application Performance of three portable blood glucose monitoring devices used in a veterinary application Prepared by Michael I. Lindinger, PhD, Vice-President of Research, The Nutraceutical Alliance, Campbellville,

More information

Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2

Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2 Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2 A.W. Jones and KÄ. Jönsson Departments of Alcohol Toxicology and Internal Medicine, University Hospital,

More information

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX AIR Blood Glucose Monitoring System, from Trividia

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia

More information

Technical Evaluation of Five Glucose Meters With Data Management Capabilities

Technical Evaluation of Five Glucose Meters With Data Management Capabilities Clinical Chemistry / LABORATORY EVALUATION OF GLUCOSE METERS Technical Evaluation of Five Glucose Meters With Data Management Capabilities Jeffrey]. Chance, PhD, Dai J. Li,MD, PhD, Kerrie A. Jones, MS,

More information

FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols

FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society CLINICAL APPLICATIONS FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in

More information

Ultra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC

Ultra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC VS UNISTRIP TECHNOLOGIES,LLC Charlotte, NC 28269 1-866-889-6229 V2-10/25/2018 UniStrip Generic Blood Glucose VS OneTouch Ultra Blue 1. Purpose This study is intended to evaluate the difference in values

More information

Journal of Diabetes and Metabolic Disorders; 2011; Vol 10, pp 1-6. Keywords: Glycated hemoglobin, Paper filter, Diabetes, Dried blood spot.

Journal of Diabetes and Metabolic Disorders; 2011; Vol 10, pp 1-6. Keywords: Glycated hemoglobin, Paper filter, Diabetes, Dried blood spot. Journal of Diabetes and Metabolic Disorders; 2011; Vol 10, pp 1-6 Glycated hemoglobin measurements from dried blood spots: reliability and relation to results obtained from whole blood samples Ozra Tabatabaei-Malazy

More information

Gestational Diabetes. Gestational Diabetes:

Gestational Diabetes. Gestational Diabetes: Gestational Diabetes Detection and Management Steven Gabbe, MD The Ohio State University Medical Center Gestational Diabetes: Detection and Management Learning Objectives: At the conclusion of this presentation,

More information

MAGNITUDE OF SAMPLING AND ANALYTICAL VARIATIONS IN BLOOD AND BREATH ALCOHOL MEASUREMENTS

MAGNITUDE OF SAMPLING AND ANALYTICAL VARIATIONS IN BLOOD AND BREATH ALCOHOL MEASUREMENTS MAGNITUDE OF SAMPLING AND ANALYTICAL VARIATIONS IN BLOOD AND BREATH ALCOHOL MEASUREMENTS A. W. Jones, K. Jensson and P.M. Williams Department of alcohol Toxicology,j National Laboratory of Forensic Chemistry

More information

Assessment of a Critical Limit Protocol for Point-of-Care Glucose Testing

Assessment of a Critical Limit Protocol for Point-of-Care Glucose Testing CLINICAL CHEMISTRY Original Article Assessment of a Critical Limit Protocol for Point-of-Care Glucose Testing GIFFORD LUM, MD A critical limit protocol requiring that all point of care glucose meter readings

More information

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff

More information

Accuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method

Accuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method Accuracy and Precision Performance of the Accu-Chek Mobile System I. ACCURACY The accuracy of the system was assessed according to ISO 15197. Introduction The purpose of this study was to determine the

More information

Multicenter Evaluation of the Glucometer Elite XL Meter, an Instrument Specifically Designed for Use With Neonates

Multicenter Evaluation of the Glucometer Elite XL Meter, an Instrument Specifically Designed for Use With Neonates Emerging Treatments and O R I G I N A L A R T I C L E Technologies Multicenter Evaluation of the Glucometer Elite XL Meter, an Instrument Specifically Designed for Use With Neonates JOËL GIROUARD, MD JEAN-CLAUDE

More information

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,

More information

Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase

Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase Washington University School of Medicine Digital Commons@Becker Open Access Publications 2013 Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase Dennis J. Dietzen

More information

Alternate Site Testing for Hemoglobin A1c in Children with Diabetes. Sarah L. Flores, MS, RN, BC Manager, Specialty Clinics

Alternate Site Testing for Hemoglobin A1c in Children with Diabetes. Sarah L. Flores, MS, RN, BC Manager, Specialty Clinics Alternate Site Testing for Hemoglobin A1c in Children with Diabetes Sarah L. Flores, MS, RN, BC Manager, Specialty Clinics Background Literature supports use of alternate site testing (AST) for blood glucose

More information

Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes

Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes Citation for published version (APA): Berkel, van, M., & Scharnhorst,

More information

Chapter 11: Advanced Remedial Measures. Weighted Least Squares (WLS)

Chapter 11: Advanced Remedial Measures. Weighted Least Squares (WLS) Chapter : Advanced Remedial Measures Weighted Least Squares (WLS) When the error variance appears nonconstant, a transformation (of Y and/or X) is a quick remedy. But it may not solve the problem, or it

More information

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Institute of Diabetes Gerhardt Katsch Karlsburg (IDK) Report Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Sponsor: The study was

More information

Multiple studies are available concerning the use of

Multiple studies are available concerning the use of SPECIAL ARTICLE Evaluation of the Biosite Ò Quantitative Whole Blood D-dimer Assay and Comparison With the biomérieux VIDAS Ò D-dimer Exclusion Test Validation and Utility For Use in the Central Laboratory

More information

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D. Journal of Diabetes Science and Technology Volume 5, Issue 4, July 2011 Diabetes Technology Society TECHNOLOGY REPORTS Accuracy of the CONTOUR Blood Glucose Monitoring System Bern, B.A., Cheryl Leazenby,

More information

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System Embrace No Code Report Clinical Evaluation for Embrace No Code Blood Glucose Monitoring System Applicant: Apex Biotechnology Corp. Test Date: 2007/8/28~2007/9/21 Clinical Site: SCHMIDT Group Practice Clinic

More information

Testing Blood Glucose at the Bedside in a Chronic Care Hospital

Testing Blood Glucose at the Bedside in a Chronic Care Hospital CHEMISTRY Robert M. Greendyke, MD F. Ronald Gifford, MT(ASCP) Testing Blood Glucose at the Bedside in a Chronic Care Hospital We can provide tests that are fast, inexpensive, or accurate. Pick two. Anonymous

More information

3. System Accuracy Evaluation LUX Hgb test 1. Samples

3. System Accuracy Evaluation LUX Hgb test 1. Samples 3. System Accuracy Evaluation LUX Hgb test 1. Samples - Type of sample: Capillary Blood System accuracy evaluation was performed with 200 fresh capillary blood samples, each with sufficient volume to be

More information

The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary Blood Glucose Values

The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary Blood Glucose Values University of New Hampshire University of New Hampshire Scholars' Repository Honors Theses and Capstones Student Scholarship Spring 2017 The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary

More information

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE Brooke Ann Spaeth, Anne K Shephard, Mark DS Shephard, Timothy H Mathew ABSTRACT

More information

Analytical error of home glucose monitors: a comparison of 18 systems

Analytical error of home glucose monitors: a comparison of 18 systems Original Article Ann Clin Biochem 1999; 36: 72-79 Analytical error of home glucose monitors: a comparison of 18 systems Roger N Johnson and John R Baker From the Department of Clinical Biochemistry, Green

More information

Aina Blood Monitoring System

Aina Blood Monitoring System Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale

More information

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Thomas J. Spira, M.D. International Laboratory Branch Global AIDS Program Centers for Disease

More information

Relationship between glucose meter error and glycemic control efficacy

Relationship between glucose meter error and glycemic control efficacy Relationship between glucose meter error and glycemic control efficacy Brad S. Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Department of Laboratory Medicine and Pathology Mayo Clinic

More information

estimated to be complicated by diabetes, and up to 0.5 percent of pregnancies occur in mothers who already have diabetes (Ang, Howe, &

estimated to be complicated by diabetes, and up to 0.5 percent of pregnancies occur in mothers who already have diabetes (Ang, Howe, & Title Subtitle author byline Insulin Revisited f in the Maternity Setting For many women who would previously have been instructed to avoid pregnancy, the use of insulin has allowed them to become pregnant

More information

2/27/2017. Point of Care Testing- current and future opportunities for pharmacists in Virginia. Financial Disclosures. Pre-Assessment.

2/27/2017. Point of Care Testing- current and future opportunities for pharmacists in Virginia. Financial Disclosures. Pre-Assessment. Point of Care Testing- current and future opportunities for pharmacists in Virginia Objectives Define the Pharmacy Practice Act in Virginia M AR G AR E T L AN D I S, P H AR M. D. K R O G E R P H AR M AC

More information

A Clinical Evaluation of Various Delta Check Methods

A Clinical Evaluation of Various Delta Check Methods CLIN. CHEM. 27/1, 5-9 (1981) A Clinical Evaluation of Various Delta Check Methods Lawrence A. Wheeler1 and Lewis B. Sheiner2 To evaluate the performance of delta check techniques, we analyzed 707 unselected

More information

Advances in Diabetes Care Technologies

Advances in Diabetes Care Technologies 1979 Advances in Diabetes Care Technologies 2015 Introduction Insulin pump use: ~ 20% - 30% of patients with T1DM < 1% of insulin-treated patients with T2DM 2007 FDA estimates ~375,000 insulin pumps for

More information

associated with serious complications, but reduce occurrences with preventive measures

associated with serious complications, but reduce occurrences with preventive measures Wk 9. Management of Clients with Diabetes Mellitus 1. Diabetes Mellitus body s inability to metabolize carbohydrates, fats, proteins hyperglycemia associated with serious complications, but reduce occurrences

More information

A comparison of fluoride-oxalate and plain (serum gel) tube on glucose measurement

A comparison of fluoride-oxalate and plain (serum gel) tube on glucose measurement ORIGINAL ARTICLE A comparison of fluoride-oxalate and plain (serum gel) tube on glucose measurement E.A. Amegashie 1, N. Amidu 2 and W.K.B.A. Owiredu 3 Journal of Medical and Biomedical Sciences (2015)

More information

Current Trends in Diagnosis and Management of Gestational Diabetes

Current Trends in Diagnosis and Management of Gestational Diabetes Current Trends in Diagnosis and Management of Gestational Diabetes Shreela Mishra, MD Assistant Clinical Professor UCSF Fresno Medical Education Program 2/2/2019 Disclosures No disclosures 2/2/19 Objectives

More information

A Comparison between Venous and Finger-Prick Blood Sampling on Values of Blood Glucose

A Comparison between Venous and Finger-Prick Blood Sampling on Values of Blood Glucose 2012 International Conference on Nutrition and Food Sciences IPCBEE vol. 39 (2012) (2012) IACSIT Press, Singapore A Comparison between Venous and Finger-Prick Blood Sampling on Values of Blood Glucose

More information

Pseudohyperglycemia: Effects of Unwashed Hand after Fruit Peeling or Handling on Fingertips Blood Glucose Monitoring Results

Pseudohyperglycemia: Effects of Unwashed Hand after Fruit Peeling or Handling on Fingertips Blood Glucose Monitoring Results Original Article Pseudohyperglycemia: Effects of Unwashed Hand after Fruit Peeling or Handling on Fingertips Blood Glucose Monitoring Results Olamoyegun MA, Oloyede T 1, Adewoye OG 2, Abdulkarim SO 2,

More information

The Second Report of the Expert Panel on Detection,

The Second Report of the Expert Panel on Detection, Blood Cholesterol Screening Influence of State on Cholesterol Results and Management Decisions Steven R. Craig, MD, Rupal V. Amin, MD, Daniel W. Russell, PhD, Norman F. Paradise, PhD OBJECTIVE: To compare

More information

VALIDATION OF CONTINUOUS GLUCOSE MONITORING IN CHILDREN AND ADOLESCENTS WITH CYSTIC FIBROSIS - A PROSPECTIVE COHORT STUDY

VALIDATION OF CONTINUOUS GLUCOSE MONITORING IN CHILDREN AND ADOLESCENTS WITH CYSTIC FIBROSIS - A PROSPECTIVE COHORT STUDY Diabetes Care Publish Ahead of Print, published online March 11, 2009 Validation of CGM in Children with CF VALIDATION OF CONTINUOUS GLUCOSE MONITORING IN CHILDREN AND ADOLESCENTS WITH CYSTIC FIBROSIS

More information

Method Comparison Report Semi-Annual 1/5/2018

Method Comparison Report Semi-Annual 1/5/2018 Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community

More information

Assessment of "Average of Normals" Quality Control Procedures and Guidelines for Implementation

Assessment of Average of Normals Quality Control Procedures and Guidelines for Implementation Assessment of "Average of Normals" Quality Control Procedures and Guidelines for Implementation GEORGE S. CEMBROWSKI, M.D., PH.D., ELLIOT P. CHANDLER, M.D., AND JAMES O. WESTGARD, PH.D. The capabilities

More information

Quality Specifications for POCT

Quality Specifications for POCT Quality Specifications for POCT the patient journey perspective Christopher P Price Department of Clinical Biochemistry, University of Oxford, Oxford, UK Quality Specifications fit-for-purpose QUALITY

More information

Diabetes Care Publish Ahead of Print, published online February 2, 2011

Diabetes Care Publish Ahead of Print, published online February 2, 2011 Clinical Care/Education/Nutrition/Psychosocial Research O R I G I N A L A R T I C L E Diabetes Care Publish Ahead of Print, published online February 2, 2011 Self-Monitoring of Blood Glucose: The Use of

More information

Evaluation of Models to Estimate Urinary Nitrogen and Expected Milk Urea Nitrogen 1

Evaluation of Models to Estimate Urinary Nitrogen and Expected Milk Urea Nitrogen 1 J. Dairy Sci. 85:227 233 American Dairy Science Association, 2002. Evaluation of Models to Estimate Urinary Nitrogen and Expected Milk Urea Nitrogen 1 R. A. Kohn, K. F. Kalscheur, 2 and E. Russek-Cohen

More information

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide LESSON 6-3 Laboratory Reagent Preparation and Calculations Worksheet LESSON 6-4 Chemistry Instrumentation in the Physician Office Laboratory Worksheet LESSON 6-6 Measuring Blood Glucose Worksheet LESSON

More information

Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration

Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration Clinical Chemistry / Whole Blood Creatinine for egfr Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration Nichole

More information

Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube

Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube Coagulation and Transfusion Medicine / Discard Tube Unnecessary for Coagulation Assays Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube Maarten

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

Recent Advances in the Analysis of Ethanol in Saliva: Evaluation of the QED Device

Recent Advances in the Analysis of Ethanol in Saliva: Evaluation of the QED Device Recent Advances in the Analysis of Ethanol in Saliva: Evaluation of the QED Device A.W. Jones and k A. Jünsson Departments of Alcohol Toxicology and Internal Medicine, University Hospital, 581 85 Linköping,

More information

APEC Guidelines Gestational Diabetes Mellitus

APEC Guidelines Gestational Diabetes Mellitus Gestational diabetes mellitus (GDM) is defined as insulin resistance of variable severity with onset or first recognition during pregnancy. The prevalence of diabetes mellitus (DM) in the US is growing

More information

Figure 2.1: Glucose meter

Figure 2.1: Glucose meter CHAPTER TWO: MONITORING TECHNOLOGIES 2.1 Introduction Glucose monitoring is a method of self-testing glucose (blood sugar) levels for the management of diabetes. Traditionally, it involves pricking the

More information

Research Article Comparison of Full Blood Count Parameters Using Capillary and Venous Samples in Patients Presenting to the Emergency Department

Research Article Comparison of Full Blood Count Parameters Using Capillary and Venous Samples in Patients Presenting to the Emergency Department International Scholarly Research Network ISRN Emergency Medicine Volume 2012, Article ID 508649, 6 pages doi:10.5402/2012/508649 Research Article Comparison of Full Blood Count Parameters Using Capillary

More information

Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions

Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions Clinica Chimica Acta 300 (2000) 13 21 www.elsevier.com/ locate/ clinchim Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions L.V. Rao *, A.O.

More information

CJASN epress. Published on November 5, 2008 as doi: /CJN

CJASN epress. Published on November 5, 2008 as doi: /CJN CJASN epress. Published on November 5, 2008 as doi: 10.2215/CJN.03360708 Correlation of Point-of-Care International Normalized Ratio to Laboratory International Normalized Ratio in Hemodialysis Patients

More information

NICO /14 Nipro Diagnostics, Inc.

NICO /14 Nipro Diagnostics, Inc. Clinical Performance of the TRUEresult twist Blood Glucose Monitoring System () Using ISO 15197:13 Criteria for Accuracy Utilizing GDH-FAD Enzyme NICO-8 1/1 Nipro Diagnostics, Inc. NICO_8_mmol.indd CLINICAL

More information

Gestational Diabetes Mellitus: Screening and Diagnosis

Gestational Diabetes Mellitus: Screening and Diagnosis SYMPOSIUM Gestational Diabetes Mellitus: Screening and Diagnosis Gestational diabetes mellitus (GDM) represents a major health problem, affecting 2% to 3% of the obstetric population, or greater than 100,000

More information

BLOOD GLUCOSE MONITORING: CLINICAL NEED AND TECHNOLOGY

BLOOD GLUCOSE MONITORING: CLINICAL NEED AND TECHNOLOGY BLOOD GLUCOSE MONITORING: CLINICAL NEED AND TECHNOLOGY DAVID C KLONOFF, MD Medical Director Diabetes Research Institute of Mills-Peninsula Health Services San Mateo, California Clinical Professor of Medicine,

More information

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood

More information

Accuracy of Three Dry-Chemistry Methods for Lipid Profiling and Risk Factor Classification

Accuracy of Three Dry-Chemistry Methods for Lipid Profiling and Risk Factor Classification 358 International / Rubin Journal et al. of Sport Nutrition and Exercise Metabolism, 2003, 13, 358-368 2003 Human Kinetics Publishers, Inc. Accuracy of Three Dry-Chemistry Methods for Lipid Profiling and

More information

Cover Page. The handle holds various files of this Leiden University dissertation

Cover Page. The handle   holds various files of this Leiden University dissertation Cover Page The handle http://hdl.handle.net/1887/37409 holds various files of this Leiden University dissertation Author: Engbers, Marissa Title: Conventional and age-specific risk factors for venous thrombosis

More information

Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous Glucose Sensors

Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous Glucose Sensors Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society ORIGINAL ARTICLES Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous

More information

Evaluation Report of the Pneumatic Tube Transport System (PEVCO) connecting Dialysis Hospital to. Mubarak Hospital. Dr.

Evaluation Report of the Pneumatic Tube Transport System (PEVCO) connecting Dialysis Hospital to. Mubarak Hospital. Dr. 5 Evaluation Report of the Transport System (PEVCO) connecting Dialysis Hospital to Mubarak Hospital Dr. Anwar AlAnjeri Senior Registrar Clinical Biochemistry Laboratory Mubarak Hospital Introduction:

More information

Statistical techniques to evaluate the agreement degree of medicine measurements

Statistical techniques to evaluate the agreement degree of medicine measurements Statistical techniques to evaluate the agreement degree of medicine measurements Luís M. Grilo 1, Helena L. Grilo 2, António de Oliveira 3 1 lgrilo@ipt.pt, Mathematics Department, Polytechnic Institute

More information

The moderating impact of temporal separation on the association between intention and physical activity: a meta-analysis

The moderating impact of temporal separation on the association between intention and physical activity: a meta-analysis PSYCHOLOGY, HEALTH & MEDICINE, 2016 VOL. 21, NO. 5, 625 631 http://dx.doi.org/10.1080/13548506.2015.1080371 The moderating impact of temporal separation on the association between intention and physical

More information

Optimization of urinary dipstick ph: Are multiple dipstick ph readings reliably comparable to commercial 24-hour urinary ph?

Optimization of urinary dipstick ph: Are multiple dipstick ph readings reliably comparable to commercial 24-hour urinary ph? Original Article - Basic and Translational Research https://doi.org/10.4111/icu.201.378 pissn 2466-0493 eissn 2466-054X Optimization of urinary dipstick ph: Are multiple dipstick ph readings reliably comparable

More information

Diabetes Control In Pregnancy Part I.

Diabetes Control In Pregnancy Part I. Diabetes Control in Pregnancy - Part I Expires Monday, April 30, 2018 Nursing Tamerou Asrat, M.D. Objectives 1. Discuss the different classifications of diabetes and the complications that may be seen

More information

GB/T Translated English of Chinese Standard: GB/T

GB/T Translated English of Chinese Standard: GB/T Translated English of Chinese Standard: GB/T 19634-2005 www.chinesestandard.net Sales@ChineseStandard.net GB ICS 11.100 C 44 National Standard of the People s Republic of China GB/T 19634-2005 In vitro

More information

GESTATIONAL DIABETES TESTING AND TREAMENT

GESTATIONAL DIABETES TESTING AND TREAMENT Boston Medical Center Maternity Care Guideline: GESTATIONAL DIABETES TESTING AND TREAMENT Accepted: August 2015 Updated: December 2018 Contributors: Aviva Lee-Parritz, MD, Sara Alexanian, MD, Kari Radoff,

More information

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Polymer Technology Systems, Inc. CardioChek PA Comparison Study Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory

More information

Phlebotomy Top Gun Order of Draw: Do We Still Care?

Phlebotomy Top Gun Order of Draw: Do We Still Care? Phlebotomy Top Gun Order of Draw: Do We Still Care? 2018 MFMER 3793435-1 Presenter: Brad Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Division of Clinical Core Laboratory Services

More information

Original Article. Keiko KOHNO 1), Kazuhiko HOSHI 1), Motoi TAKIZAWA 1), Takashi KANEKO 2), and Shuji HIRATA 1)

Original Article. Keiko KOHNO 1), Kazuhiko HOSHI 1), Motoi TAKIZAWA 1), Takashi KANEKO 2), and Shuji HIRATA 1) Yamanashi Med. J. 21(3), 53 ~ 58, 2006 Original Article Usefulness of the 50-g Glucose Challenge Test for Screening of Patients with Gestational Diabetes Mellitus and an Analysis of the Timing of Administration

More information

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS)

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Type 2 Diabetes DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Clinical Guideline Self-monitoring of blood glucose (SMBG) should be considered but not be routinely offered for adults with type 2 diabetes

More information

Test strip reading - Cats Negative Trace

Test strip reading - Cats Negative Trace WHAT I LEARNED FROM MY DIABETIC CAT: DIAGNOSIS AND MONITORING Ellen N. Behrend, VMD, PhD, DACVIM Although diagnosis of diabetes mellitus (DM) is usually a no-brainer, it can sometimes be difficult, especially

More information

Assessing Agreement Between Methods Of Clinical Measurement

Assessing Agreement Between Methods Of Clinical Measurement University of York Department of Health Sciences Measuring Health and Disease Assessing Agreement Between Methods Of Clinical Measurement Based on Bland JM, Altman DG. (1986). Statistical methods for assessing

More information

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS)

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Type 2 Diabetes Clinical Guideline Self-monitoring of blood glucose (SMBG) should be considered but not routinely offered for adults with type 2 diabetes

More information

OGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking.

OGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking. ITALIAN PRODUCT S E L F T E S T I N G Personal medical devices for metabolic disorders MONITORING and PREVENTION, taking advantage of reflectometric and electrochemical technologies. Blood glucose monitoring

More information

Analytical and Clinical Performance of the epoc Blood Analysis System. Experience at a Large Tertiary Academic Medical Center

Analytical and Clinical Performance of the epoc Blood Analysis System. Experience at a Large Tertiary Academic Medical Center AJCP / Original Article Analytical and Clinical Performance of the epoc Blood Analysis System Experience at a Large Tertiary Academic Medical Center Brie A. Stotler, MD, MPH, 1,2 and Alexander Kratz, MD,

More information

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including Week 17 and 21 Comparing two assays and Measurement of Uncertainty 2.4.1.4. Explain tools used to compare the performance of two assays, including 2.4.1.4.1. Linear regression 2.4.1.4.2. Bland-Altman plots

More information

REPORT INTERPRETATION

REPORT INTERPRETATION REPORT INTERPRETATION: Interpreting ipro Professional Continuous Glucose Monitoring (CGM) Reports and Making Therapy Adjustments TARGET AUDIENCE The audience for this section is physicians, mid-level practitioners,

More information

Blood Sugar Screening Levels*

Blood Sugar Screening Levels* April, 1951 A Preliminary Report on a Study of Blood Sugar Screening Levels HUGH L. C. WILKERSON, M.D., F.A.P.H.A., AND BARBARA GLENN Chief; and Statistician, Diabetes Branch, Division of Chronic Diseases,

More information

Do fat and protein affect post prandial glycemia in the outpatient real world setting? Neha Verma, MD Robert Melfi, MD

Do fat and protein affect post prandial glycemia in the outpatient real world setting? Neha Verma, MD Robert Melfi, MD Do fat and protein affect post prandial glycemia in the outpatient real world setting? Neha Verma, MD Robert Melfi, MD Objectives Understand the effect of fat and protein on glycemic excursion in the post-prandial

More information

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing

More information