1 Hematopathology / DIGITAL IMAGE ANALYSIS IN HEMATOLOGY Performance Evaluation of the CellaVision DM96 System WBC Differentials by Automated Digital Image Analysis Supported by an Artificial Neural Network Alexander Kratz, MD, PhD, MPH, 1,2 Hans-Inge Bengtsson, MSc, 3 Jeanne E. Casey, H(ASCP), 1* Joan M. Keefe, MT(ASCP), 1 Gail H. Beatrice, MLT(ASCP), 1 Debera Y. Grzybek, H(ASCP), 1 Kent B. Lewandrowski, MD, 1,2 and Elizabeth M. Van Cott, MD 1,2 Key Words: Image analysis; Differentials; Hematology analyzers; Laboratory automation Abstract We evaluated the CellaVision DM96 (CellaVision AB, Lund, Sweden), an automated digital cell morphology and informatics system for peripheral blood smears. Technologists agreed with 82% of the instrument s preclassifications. Correlation coefficients between final results released from the CellaVision and results obtained by direct microscopy were 0.96 (all neutrophils), 0.94 (lymphocytes), 0.88 (segmented neutrophils), 0.73 (eosinophils), 0.69 (bands), and 0.67 (monocytes). After correction for statistically and clinically insignificant variations, the CellaVision DM96 had 95% sensitivity and 88% specificity for immature myeloid cells. It was 100% sensitive and 94% specific for blasts, and 100% sensitive and 97% specific for unusual WBCs and nucleated RBCs. Advantages of the CellaVision DM96 over direct microscopy include the ability to review slides from a remote location, consultation and quality control on a cell-by-cell basis, and potential labor savings. The differential counting of leukocytes, introduced almost 115 years ago by Ehrlich, 1 has remained a clinically important and frequently ordered laboratory test. 2 Progress in the design of laboratory instrumentation has replaced the microscope with automated cell counters as the instrument of choice for the majority of differential counts (differentials) in most laboratories. 3 Modern automated cell counters can provide a reliable WBC differential count for samples that are normal or that exhibit only a quantitative abnormality. Qualitatively abnormal samples, such as those with immature or abnormal cells, still require the preparation of a slide and microscopic analysis. In our laboratory, approximately 28% of differentials are flagged by the automated cell counter and require the preparation of a blood smear for microscopic review. Difficult differentials with unusual or abnormal cells may be a relatively rare event in many clinical laboratories; nevertheless, the significant clinical impact of a wrong diagnosis necessitates the around-the-clock on-site presence of highly trained personnel for the microscopic review of these blood smears. This increases labor costs, a major issue in today s cost-sensitive health care environment. Other disadvantages of direct microscopic examination of a stained blood smear include interobserver and intraobserver variation and the fact that the individual cells used for the differential count are not easily retrievable for subsequent review. These issues lead to difficulties in quality control and maintaining interobserver and intraobserver consistency in interpretations. Automated image processing systems have been developed to address these difficulties. These systems generally obtain digital images of objects on the blood smear and then use sophisticated software to identify and preclassify cells; the images are stored for possible later retrieval. A number of such 770 Am J Clin Pathol 2005;124: Downloaded 770 from

2 Hematopathology / ORIGINAL ARTICLE devices have been developed (recently reviewed by Tatsumi and Pierre 2 ). Capitalizing on advances in digital photography and data processing and storage, these devices have made it possible to envision a virtual blood film that can be reviewed, shared with experts in distant locations, stored, retrieved, and rereviewed at low cost. 4,5 The latest such automated image analysis system to become commercially available is the CellaVision DM96 (CellaVision AB, Lund, Sweden; distributed in the United States by Sysmex America, Mundelein, IL). The successor model to the DiffMaster Octavia (CellaVision AB), 6 the CellaVision DM96 is an automated image analysis system for peripheral blood smears. Barcode-labeled, stained glass slides are placed into a magazine. The instrument can be loaded with up to 8 magazines, each containing up to 12 slides, and operates with a continuous feed that allows magazines to be added constantly. By scanning the glass slides at low power, the instrument identifies potential WBCs and then takes digital images at high magnification. The images are analyzed by an artificial neural network based on a large database of cells and preclassified according to WBC class. The WBCs and their suggested classification are presented to the user on a customizable computer display for confirmation or reclassification Figure 1. The system also provides functionality for the review of RBC and platelet morphologic features and for estimation of the platelet count. Advantages of this approach over slide review with a microscope include possible labor savings owing to the localization and preclassification of the WBCs by the instrument, more reproducible results, and the ability to review cases from a remote location or to rereview cells at a later date. Direct microscopic review of blood smears is a well-established procedure; clinicians and laboratory workers have extensive experience with this method, are comfortable with it, and have confidence in the results obtained. Introduction of an instrument for automated image analysis, such as the CellaVision DM96, into the clinical laboratory will necessitate a change in work habits, retraining of personnel, and increased spending on hardware and software. For automated image analysis to replace direct microscopic review, 2 conditions will have to be fulfilled: the automated method will have to be shown to be at least as reliable clinically as the direct microscopic review of slides, and there will have to be significant labor savings associated with the new method to justify the additional expenditure associated with the purchase of the equipment. In this study, we compare the clinical performance of the CellaVision DM96 with direct manual microscopy. We also show the results of workflow timing studies comparing direct microscopy with the CellaVision DM96. Materials and Methods Study Site and Personnel The study was performed in the Clinical Hematology Laboratory, Massachusetts General Hospital, Boston, a tertiary care academic medical center serving a large inpatient and outpatient population, including an active hematology-oncology service. All studies were performed by hospital laboratory employees who had been trained in the use of the instrument on-site by a CellaVision company representative. A reliability log for the instrument was kept for the duration of the study. Figure 1 Computer screen on which WBCs preclassified by the CellaVision DM96 are presented to a reviewer for approval or reclassification. The reviewer can enlarge single cells as needed and can leave individual cells in the categories suggested by the instrument (accept the preclassification) or move individual cells into other groups (reclassify the cells). Digital images and the actions taken by the reviewer are stored by the instrument and can be reviewed later for quality control, confirmation of diagnosis, consultation, or comparison with later findings. Sample Selection and Preparation Routine patient samples on which WBC differentials had been ordered by the clinician and on which a microscopic slide review was necessary were used in the study. In our laboratory, the determination that a microscopic slide review is necessary is made when a sample fails criteria for the direct release of results from an ADVIA 120 Hematology System (Bayer HealthCare, Diagnostics Division, Tarrytown, NY). These criteria have been set to conform with the manufacturer s recommendations and the clinical needs of the patient population seen at our institution. All samples that were included in the study had been screened by the ADVIA 120 Downloaded from Am J Clin Pathol 2005;124:

3 Kratz et al / DIGITAL IMAGE ANALYSIS IN HEMATOLOGY Hematology System and flagged for microscopic review. The use of cases prescreened by an automated cell counter was intended to mimic the standard workflow in a hematology laboratory, where usually only cases flagged by the automated cell counter are subject to slide review. Wedge blood smears were prepared manually with the push-pull method with a spreader slide or with a semiautomated method, the Miniprep blood smearing instrument (Geometric Data, Wayne, PA), air dried, and stained with a Sakura Stainer RSG-61 (Sakura Fineteck USA, Torrance, CA) with a Wright-Giemsa stain. 7 Manual Differential Counts (Reference Method) Manual differentials using standard microscopic technique were performed following the laboratory s standard operating procedure, which is based on National Committee for Clinical Laboratory Standards guidelines. 8 Analysis With the CellaVision DM96 After analysis by direct microscopy, slides were labeled with a barcode and loaded into the CellaVision DM96. This instrument scans slides, identifies potential WBCs, takes digital images of them, and uses artificial neural network based software to analyze the cells. Digital images of preclassified cells are presented to the technologist on a computer display (Figure 1). The technologist is asked to review the images and to accept the preclassification provided by the software or to reclassify cells. For this study, technologists using the CellaVision DM96 were provided with the same information about the cases as was available to them during direct microscopic analysis. This included the results from the automated cell counter on the present sample and previous CBC counts, WBC differentials, and clinical results for the patient, if available. Technologists also were asked to evaluate whether the areas presented by the CellaVision DM96 for the evaluation of RBC and platelet morphologic features and platelet count were adequate. Studies Performed Three sets of studies were performed: a clinical performance evaluation, a manual differential timing study, and a standard workflow timing study. The clinical performance evaluation was intended to determine the performance of the CellaVision DM96 in difficult cases; for this study, cases that met certain criteria (eg, low WBC count, unusual WBCs) were obtained during a 2-month period and used for this study. For the manual differential timing study, 10 cases of average difficulty were selected. Last, for the standard workflow timing studies, a random sample of cases for which slides were prepared in the laboratory during the study period was used. Clinical Performance Evaluation We selected 120 slides with abnormal blood smear findings thought to be difficult cases for evaluation of the clinical performance of the CellaVision DM96. These cases included slides with increased numbers of immature WBCs and the presence of abnormal WBCs and some cases with very low WBC counts. All slides had required a full microscopic differential count. Three laboratory employees (2 certified laboratory technologists [J.E.C. and J.M.K.] and 1 certified medical laboratory technician [G.H.B.]) were asked to participate in this part of the study; they were selected for representing the range of expertise in the performance of manual differentials found in the laboratory. Expertise of employees was determined by their level of formal training (certified technologists vs certified technicians), the length of their work experience (between 1.5 and 35 years), and whether the employee was training new employees in the performance of WBC differentials and was a designated resource person in the laboratory for difficult WBC differential cases. Each employee was asked to perform manual differentials on 40 slides with the microscope and approximately 1 week later on the same 40 slides with the CellaVision DM96. Manual Differential Timing Studies Ten slides that had required full microscopic differential count were used for a manual differential timing study. Five technologists and a technician, representing the range of expertise in the performance of manual differentials found in the laboratory, were timed while performing manual differentials on these slides with the microscope and then approximately 1 week later while performing the same task with the CellaVision DM96. Cases were considered completed on finalization of the results in the laboratory information system or on conclusion of analysis with the CellaVision DM96. Standard Workflow Timing Studies We used 236 additional randomly selected cases on which a manual slide review was necessary for a workflow study. Approximately half of these slides required the performance of a full microscopic differential count; the other half required only microscopic review ( scan ) for confirmation of the results of the differential obtained from the automated cell counter. These cases first were analyzed by a medical laboratory technician (G.H.B.) with 1.5 years of experience in reading blood smears (n = 160 cases) or a very experienced technologist (J.E.C.) with 35 years of experience (n = 76 cases) with the direct microscopic method as part of their normal work assignments in the clinical laboratory. The employees then were asked to reanalyze the same samples with the CellaVision DM96. These employees also had participated in the other studies. Statistical Analysis Statistical analysis was performed using Microsoft Excel software (Microsoft, Redmond, WA). Two-tailed paired t tests were used to evaluate differences between the time needed for 772 Am J Clin Pathol 2005;124: Downloaded 772 from

4 Hematopathology / ORIGINAL ARTICLE fully manual differentials with the CellaVision DM96 and with direct microscopy. Clinical sensitivity and specificity of the CellaVision DM96 were defined as its ability to obtain positive and negative results concordant with the results obtained by direct microscopy. Results System Reliability The CellaVision DM96 was in use in our laboratory for 2 weeks. During this time, a critical oil error required 5 minutes for resolution by priming the immersion oil pump to remove air bubbles. There were 4 rack jams; each of these incidents required approximately 1 or 2 minutes to resolve; 2 of the jams were caused by trying to load slides not appropriate for the instrument. There were 3 cases of software failure that required the viewing program to be restarted, leading to the loss of approximately 1 minute per incident. In total, approximately 12 to 20 minutes were lost in these 8 downtime incidents during 2 weeks. Ability of the CellaVision DM96 to Present to the Reviewer WBCs That Are Qualitatively and Quantitatively Adequate for a Differential Count We used 120 cases that required a full microscopic differential to evaluate the clinical performance of the CellaVision DM96. A laboratory employee first performed a differential on these cases using the microscope; the same person then performed a differential on the same slides using the CellaVision DM96. The slide readers were asked to note when they thought that the images presented by the CellaVision DM96 were of insufficient quality to allow them to perform a differential. Such a notation was made in 11 (9.2%) of the cases. For 10 of these 11 cases, the same technologists had been able to do a full 100-cell differential with the microscope; in the remaining case, the microscopic differential also was inadequate (it was based on 5 WBCs). Of the 11 slides, 10 had been prepared with the manual push-pull method with a spreader slide; 1 had been prepared with a semiautomated method, the Miniprep blood smearing instrument Table 1. These 11 cases in which no differential was available from the CellaVision DM96 were excluded from subsequent correlation analysis. The precision of a WBC differential is affected negatively if only a few cells are counted. 9,10 We compared the total number of cells found by the technologists when performing WBC differentials using the microscope with the number of adequate WBCs presented for review for each case by the CellaVision DM96 Figure 2. The ability of both methods to find identifiable cells on a blood smear was similar, with a slightly higher number of cases with very few (< 20) WBCs encountered with Table 1 Adequacy of Images From the CellaVision DM96 for WBC and RBC Analysis According to Slide Preparation Method * Manual Method Semiautomated Images Inadequate for (n = 90) Method (n = 30) WBC analysis 10 (11) 1 (3) RBC analysis 18 (20) 10 (33) * The manual method refers to slides prepared with the push-pull method with a spreader slide; the semiautomated method refers to slides prepared with the Miniprep slide preparation instrument. A total of 120 slides were analyzed. Numbers represent the total number of slides. Data are given as number (percentage). the microscopic method (Figure 2). There were 11 cases (9.2%) with fewer than 20 cells counted when using the manual method; the CellaVision DM96 differential was based on fewer than 20 cells in 7 cases (5.8%). Among the 11 cases in which the manual differential was based on fewer than 20 cells, in 5 cases the CellaVision DM96 was able to find more than 20 cells per slide. There were 2 cases in which the CellaVision DM96 differential had to be based on fewer than 20 cells per slide, whereas the manual differential was based on more than 20 cells per slide. Because of the limited reproducibility of differential results based on such low counts, the 13 cases in which the manual differential, the CellaVision DM96 differential, or both were based on fewer than 20 cells per slide were excluded from subsequent analysis. Ability of the CellaVision DM96 to Provide Images Adequate for Evaluation of RBC and Platelet Morphologic Features and the Platelet Count The CellaVision DM96 allows the user to evaluate RBC and platelet morphologic features and to estimate the platelet No. of Samples < >100 Counted Cells Figure 2 Total number of cells used by technologists for WBC differential counts when using the microscope (white bars) and the CellaVision DM96 (black bars). Only objects classified as WBCs by the technologist were included in the analysis. Downloaded from Am J Clin Pathol 2005;124:

5 Kratz et al / DIGITAL IMAGE ANALYSIS IN HEMATOLOGY count by producing an overview image consisting of 35 patched areas of the slide. We asked technologists to indicate whether they thought that these images were adequate for the evaluation of RBCs and platelets. In 28 cases (23%), the technologists thought the images did not allow them to adequately evaluate RBC and platelet morphologic features. The use of a semiautomated slide preparation method was associated with a higher percentage of inadequate slides (33%) than the manual push-pull method (20%) (Table 1). When reviewing the same slides with the microscope, the technologists thought in 100% of cases that they were able to find areas allowing them to adequately address RBC and platelet morphologic features and to estimate the platelet count. Correlation of the CellaVision DM96 Preliminary Classifications With Final Classifications by Medical Technologists and Technicians The CellaVision DM96 presents, on a computer display, WBCs preclassified according to cell type. The technologist has to review every cell and can move a cell to a different cell class (reclassify the cell) or can leave the cell in the category suggested by the instrument by taking no additional action regarding the cell. Table 2 shows the percentages of the various cell types preclassified correctly by the CellaVision DM96. Overall, the instrument preclassified 82% of all cells correctly (86% if band forms were not differentiated from segmented neutrophils). The percentages of correctly preclassified cells were highest for mature cells and lowest for immature and abnormal cells. Correlation of Direct Microscopic WBC Differential Results With Final Results From the CellaVision DM96 After review and, if necessary, reclassification of the cells presented by the CellaVision DM96, a reviewer releases the final differential results to the laboratory information system. Differential results obtained by direct microscopy were compared with the final results released by the same technologist for the same slide from the CellaVision DM96. The correlation graphs for neutrophils (including segmented neutrophils and band forms), lymphocytes, monocytes, and eosinophils are shown in Figure 3. The correlation coefficients were highest for total neutrophils (0.96), lymphocytes (0.94), and segmented neutrophils (0.88). For eosinophils, the correlation coefficient was The lowest correlations were observed for band forms (0.69) and monocytes (0.67). Ability of Digital Image Analysis by CellaVision DM96 to Identify Clinically Important Abnormalities In clinical practice, the ability to identify the presence of an abnormality on a blood smear, for example, the presence of blasts, is an important determinant of the reliability of a differential counting method. We compared the ability of the Table 2 Percentages of Cells Correctly Preclassified by the CellaVision DM96 * Correct All Verified Results Suggestions by Suggested Correctly Cell Class CellaVision (%) by CellaVision (%) Segmented neutrophils (n = 3,510) Band neutrophils (n = 868) Lymphocytes (n = 2,585) Monocytes (n = 763) Eosinophils (n = 231) Basophils (n = 50) Blasts (n = 395) Immature myeloid cells (n = 627) Nucleated RBCs (n = 165) * The percentage of correct suggestions is the percentage of the cells in the various groups, as preclassified by the CellaVision DM96, in which the technologist agreed with the instrument s preclassification (ie, the percentages of original suggestions that were correct); the percentage of all verified results suggested correctly is the percentage of cells in each group released by the technologist that were preclassified correctly by the CellaVision (ie, percentages of final results preclassified correctly by the CellaVision). A total of 9,194 cells were analyzed. CellaVision DM96 to identify the presence of several clinically important abnormalities with the results of direct microscopy. These abnormalities included the presence of immature and abnormal WBCs and of nucleated RBCs. As shown in Table 3, a significant number of cases showed an abnormality on review with the CellaVision DM96 that were not reported on direct microscopy; some cases did not show an abnormality on review with the CellaVision DM96 that were reported on direct microscopic review. When direct microscopy was used as the reference method ( gold standard ) against which the results of the CellaVision DM96 were compared and when all discrepant results were included in the analysis, the specificity of the CellaVision was 82% to 93%; sensitivity was 25% to 91% Table 4. To resolve the significant percentage of false-positive and false-negative findings of the CellaVision DM96 compared with direct microscopy, digital images from the CellaVision DM96 and microscopic differentials were rereviewed by the medical director of the laboratory (A.K.) for all cases with discrepant results. As shown in Table 5, Table 6, and Table 7, most of the discrepant cases were due to small variations within the 95% confidence interval of the cell count (eg, most cases with a discrepancy involving nucleated RBCs showed 1 nucleated RBC/100 WBCs with one method and none with the other method). Some other discrepancies were due to a slight difference in nomenclature without clinical relevance (eg, the same technologist would call cells blasts with one method and use the expression other cells; immature myeloid cells, query blasts with the other method). In 15 of the discrepant cases, rereview of the digital images showed that the discrepancy was due to 774 Am J Clin Pathol 2005;124: Downloaded 774 from

6 Hematopathology / ORIGINAL ARTICLE A B 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% 0% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100% C 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% 0% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100% D 70% 60% 50% 40% 30% 20% 10% 0% 0% 10% 20% 30% 40% 50% 60% 70% E 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% 0% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100% F 70% 60% 50% 40% 30% 20% 10% 0% 0% 10% 20% 30% 40% 50% 60% 70% 25% 20% 15% 10% 5% 0% 0% 5% 10% 15% 20% 25% Figure 3 Correlation of neutrophil (A), segmented neutrophil (B), band form (C), lymphocyte (D), monocyte (E), and eosinophil (F) counts on direct microscopy and on the CellaVision DM96. A, y = x ; R 2 = B, y = x ; R 2 = C, y = 1.188x ; R 2 = D, y = x ; R 2 = E, y = x ; R 2 = F, y = x ; R 2 = misidentification of cells by the initial reviewer. There were only 3 cases in which the discrepancy could not be resolved by statistical variation, differences in nomenclature, or misclassification of digital images presented by the CellaVision DM96; in these cases, there were clearly cells consistent with immature myeloid cells (promyelocytes, metamyelocytes, and/or myelocytes) present on the CellaVision DM96 images that had not been reported on direct microscopic review. A repeated standard microscopic differential also did not show these immature myeloid cells on the slides. When a third direct microscopic differential based on up to 300 cells was performed, populations of immature myeloid cells (between 2.3% and 8%) were identified. As noted in the preceding paragraph, most of the discrepancies were insignificant statistically (within the 95% confidence Downloaded from Am J Clin Pathol 2005;124:

7 Kratz et al / DIGITAL IMAGE ANALYSIS IN HEMATOLOGY interval of the cell count) or clinically (slightly different nomenclature without clinical relevance). We therefore reanalyzed the data and omitted these cases. The results of the reanalysis are given in Table 8. Sensitivities for the various abnormalities now were between 95% and 100%; specificities varied between 88% and 97%. Timing Studies on Fully Manual Differentials An important issue in the evaluation of a new technology is its cost-effectiveness. For an instrument like the CellaVision DM96, a major determinant of cost-effectiveness is the amount of time it takes for a technologist to perform a manual differential. To compare the time needed to perform fully manual differentials on the CellaVision DM96 with direct microscopy, 10 slides that required a full manual differential (as opposed to a review of the slide and release of the automated differential from the cell counter) were analyzed by 6 technologists with a microscope and with the CellaVision DM96. It took the technologists, on average, 1.3 minutes longer per manual differential when they were using the CellaVision DM96 than when using the standard microscopic method (P <.01) Table 9. Standard Workflow Timing Studies The timing studies using fully manual differentials do not adequately reflect the usual workload of the differential workstation of a clinical hematology laboratory. At our institution, only approximately 50% of specimens that are flagged by the automated cell counter and have a slide prepared require a full manual differential. The other 50% of the slides are reviewed (scanned) by the technologist, and the WBC differential results of the automated cell counter are released, without need for a full manual differential. To capture this aspect of the normal laboratory workflow, 2 technologists (J.E.C. and G.H.B.) were timed on randomly selected slides. These slides consisted of cases that required a full manual differential and of scanned slides that were reviewed only by the technologist. The 2 technologists selected for this study were a senior technologist with a special interest in manual differentials who is the designated trainer for the differential workstation for new employees and an employee who had joined the laboratory more recently and who had no special expertise in manual differentials. By this Table 3 Identification of the Presence of Abnormal Findings on a Blood Smear * time, these 2 reviewers had received more extensive training in the use of the CellaVision DM96 than the technologists participating in the timing study involving only manual differentials. All slides used for workflow timing were prepared with the Miniprep blood smearing instrument. As shown in Table 10, there was no significant difference in the average time per slide for the very experienced technologist. The technician with no special expertise in WBC differentials, in contrast, was 25% faster using the CellaVision DM96 than using the manual method. The average time per slide for the first 77 slides that the technician analyzed with the CellaVision DM96 was 3.1 minutes; for the last 83 slides, her average time per slide using the CellaVision DM96 was 2.8 minutes. The reviewers thought that in all 236 cases, the images presented by the CellaVision DM96 for the WBC differential were adequate. Discussion CellaVision DM96 Manual Microscopy Positive Negative >3% Promyelocytes, myelocytes, and/or metamyelocytes Positive 21 2 Negative Blasts Positive 12 2 Negative 7 75 Other unusual WBCs, eg, promonocytes, plasma cells, prolymphocytes Positive 1 3 Negative 6 86 Nucleated RBCs Positive 9 5 Negative * Based on the analysis of 96 blood smears. Data are given as the number of blood smears positive or negative by the respective method. The CellaVision DM96 is an automated digital image analysis system. Routinely prepared peripheral blood smears are scanned at low power, WBCs are identified, and digital Table 4 Clinical Performance of the CellaVision DM96 Compared With Direct Microscopy as the Reference Method, Including All Discrepant Cases Cells Identified on a Slide Specificity (%) Sensitivity (%) False-Positive (%) False-Negative (%) >3% Promyelocytes, myelocytes, and/or metamyelocytes Blasts Other unusual WBCs Nucleated RBCs Am J Clin Pathol 2005;124: Downloaded 776 from

8 Hematopathology / ORIGINAL ARTICLE Table 5 Rereview of Cases With Positive Findings by the CellaVision DM96 and Negative Findings by Direct Microscopy (False-Positives on the CellaVision DM96) * Case No. Resolution Category Presence of immature myeloid cells (>3% promyelocytes, myelocytes, and/or metamyelocytes) on CellaVision DM96 differential and not on manual differential 004 Misclassification of CellaVision images by technologist % immature myeloid cells present on CellaVision differential; 2% immature myeloid cells present on microscopic 1 differential 019 Original review and rereview of CellaVision images showed presence of 17% immature myeloid cells; original 4 microscopic review and first rereview showed no immature myeloid cells; third microscopic differential based on 200 cells showed 6.5% immature myeloid cells 039 Misclassification of CellaVision images by technologist Misclassification of CellaVision images by technologist Misclassification of CellaVision images by technologist Original review and rereview of CellaVision images showed presence of 4% immature myeloid cells; original 1 microscopic review showed 0% immature myeloid cells; microscopic rereview of the slide showed 2% immature myeloid cells 098 Misclassification of CellaVision images by technologist Original review and rereview of CellaVision images showed presence of 13% immature myeloid cells; original 4 microscopic review and first rereview showed no immature myeloid cells; third microscopic differential showed 8% immature myeloid cells 118 Cells classified as promyelocytes by CellaVision classified as Others, immature myeloid cells on the microscopic 2 differential 119 Original review and rereview of CellaVision images showed presence of 4% immature myeloid cells; original 1 microscopic review and rereview showed no immature myeloid cells 120 Original review and rereview of CellaVision images showed presence of 8% immature myeloid cells; original 4 microscopic review and first rereview showed no immature myeloid cells; third microscopic differential based on 300 cells showed 2.3% immature myeloid cells Presence of blasts on CellaVision DM96 differential and not on manual differential 013 Misclassification of CellaVision images by technologist Misclassification of CellaVision images by technologist Misclassification of CellaVision images by technologist Original review and rereview of CellaVision images showed presence of 1% blasts; microscopic review and 1 rereview showed no blasts 083 Misclassification of CellaVision images by technologist Misclassification of CellaVision images by technologist Blasts had been classified as Other, immature myeloid cells on manual differential 2 Presence of unusual cells on CellaVision DM96 differential and not on manual differential 046 Misclassification of CellaVision images by technologist Original review and rereview of CellaVision DM96 images showed presence of 1% plasma cells; microscopic 1 review and rereview showed no plasma cells 094 Original review and rereview of CellaVision images showed 1% prolymphocytes; microscopic review and rereview 1 showed no prolymphocytes 101 Misclassification of CellaVision images by technologist Cells classified as Other, immature myeloid cells on CellaVision classified as blasts on microscopic review Misclassification of CellaVision images by technologist 3 Presence of nucleated RBCs (nrbcs) on CellaVision DM96 differential and not on manual differential nrbc/100 WBCs on CellaVision; no nrbcs on manual differential Misclassification of CellaVision images by technologist Misclassification of CellaVision images by technologist nrbcs/100 WBCs on CellaVision; no nrbcs on manual differential nrbcs/100 WBCs on CellaVision; no nrbcs on manual differential nrbc/100 WBCs on CellaVision; no nrbcs on manual differential nrbc/100 WBCs on CellaVision; no nrbcs on manual differential nrbc/100 WBCs on CellaVision; no nrbcs on manual differential nrbc/100 WBCs on CellaVision; no nrbcs on manual differential nrbcs/100 WBCs on CellaVision; no nrbcs on manual differential nrbcs/100 WBCs on CellaVision; no nrbcs on manual differential nrbcs/100 WBCs on CellaVision; no nrbcs on manual differential nrbc/100 WBCs on CellaVision; no nrbcs on manual differential nrbc/100 WBCs on CellaVision; no nrbcs on manual differential nrbcs/100 WBCs on CellaVision; no nrbcs on manual differential 1 * Category 1, the discrepancy could be explained by random variation (ie, results of both methods were within the 95% confidence interval of the number of cells counted); category 2, cases with essentially identical diagnoses (eg, the technologist used the expression blasts with one method and language such as other cells; immature myeloid cells, query blasts with the other); category 3, rereview of the digital images obtained by the CellaVision indicated a misclassification by the technologist of cells shown on the images; category 4, an abnormality identified by one method was not identified with the other and the discrepancy could not be explained by random variation within the 95% confidence interval, differences in nomenclature, or misclassification of cells on digital images. Analysis was performed twice for all cases listed (once as part of the original analysis and then as rereview of cases with discrepant findings between direct microscopy and the CellaVision). As described in the Results section, a third microscopic review then was performed on cases 019, 104, and 120 to resolve the remaining discrepancy. Downloaded from Am J Clin Pathol 2005;124:

9 Kratz et al / DIGITAL IMAGE ANALYSIS IN HEMATOLOGY Table 6 Rereview of Cases With Negative Findings by the CellaVision DM96 and Positive Findings by Direct Microscopy (False-Negatives on the CellaVision DM96) * Case No. Resolution Category Presence of immature myeloid cells (>3% promyelocytes, myelocytes, and/or metamyelocytes) on microscopic differential and not on CellaVision DM96 differential 046 3% immature myeloid cells present on CellaVision; 4% immature myeloid cells present on microscopic 1 differential 052 Misclassification of CellaVision images by technologist 3 Presence of blasts on microscopic differential and not on CellaVision DM96 differential 052 No blasts seen on CellaVision; 2% blasts seen on microscopic differential Blasts classified as Others (immature myeloid cells) on microscopic differential 2 Presence of unusual cells on microscopic differential and not on CellaVision DM96 differential 089 Others (immature myeloid cells) called blasts on the CellaVision Review and rereview of CellaVision differential showed no unusual cells; 1% prolymphocytes seen on 1 microscopic differential 108 Others (immature myeloid cells) called blasts on the CellaVision 2 Presence of nucleated RBCs (nrbcs) on microscopic differential but not on CellaVision DM96 differential 094 No nrbcs seen on CellaVision, 1 nrbc/100 WBCs seen on microscopic differential No nrbcs seen on CellaVision, 1 nrbc/100 WBCs seen on microscopic differential No nrbcs seen on CellaVision, 2 nrbcs/100 WBCs seen on microscopic differential No nrbcs seen on CellaVision, 1 nrbc/100 WBCs seen on microscopic differential No nrbcs seen on CellaVision, 1 nrbc/100 WBCs seen on microscopic differential 1 * See Table 5 for an explanation of the categories. images of these cells are taken at high magnification. An artificial neural network analyzes the pictures and preclassifies them according to WBC class. Preclassified WBCs are presented to a reviewer for confirmation or reclassification (Figure 1). We compared the clinical performance of this system with the standard direct microscopic method. In 11 (9.2%) of 120 cases, the technologists thought the cells presented by the CellaVision DM96 were inadequate for a reliable WBC differential. Such a high rejection rate would seriously interfere with the ability of a clinical laboratory to report differentials in a timely manner; however, 10 of these 11 cases had been prepared with the manual push-pull method with a spreader slide; only 1 of the cases had been prepared with a semiautomated slide maker. In the standard workflow timing studies, the technologists were asked to use the CellaVision DM96 to review 236 slides prepared with the semiautomated method. They thought that for all cases the images presented for the WBC differential were adequate. This indicates that the use of an automated or semiautomated slide maker significantly Table 7 Summary of Rereview of Cases With Differing Results on Manual Microscopy vs CellaVision DM96 * Category CellaVision false-positive cases Immature myeloid cells Blasts Unusual cells Nucleated RBCs CellaVision false-negative cases Immature myeloid cells Blasts Unusual cells Nucleated RBCs * Data are given as number of cases. See Table 5 for an explanation of the categories. decreases the number of cases with inadequate digital images presented by the CellaVision DM96 and allows the performance of the overwhelming majority of manual differentials using the CellaVision DM96. Table 8 Clinical Performance of the CellaVision DM96 Compared With Direct Microscopy as the Reference Method, Excluding Discrepancies Due to Statistically Insignificant Variations (Within the 95% Confidence Interval of the Cell Count) and to Clinically Insignificant Variations in Nomenclature Cells Identified on a Slide Specificity (%) Sensitivity (%) False-Positive (%) False-Negative (%) >3% Promyelocytes, myelocytes, and/or metamyelocytes Blasts Other unusual WBCs Nucleated RBCs Am J Clin Pathol 2005;124: Downloaded 778 from

10 Hematopathology / ORIGINAL ARTICLE Table 9 Average Time per Slide for 6 Technologists Performing 10 Fully Manual Differential Counts With the Microscope or the CellaVision DM96 * Technologist Microscope CellaVision A B C D E F Average for all technologists * Times are given in minutes. It is possible that in laboratories that decide to maintain the manual push-pull method for slide preparation, feedback provided to technologists will improve the quality of smear preparation. The importance of the enforcement of rigorous quality standards for slide preparation was recently stressed by Sandhaus and coworkers. 11 Among cases with an appropriate smear quality, the ability of the CellaVision DM96 to find a sufficient number of WBCs suitable for a reliable differential was at least equivalent if not better than the direct microscopic method (Figure 2). The number of cases in which the instrument provided images inadequate for RBC and platelet analysis was high (23%) and did not improve with the use of a semiautomated slide maker. By using the microscope, the reviewers were able to find areas appropriate for the evaluation of RBC and platelet morphologic features on 100% of these slides. It is possible that the percentage of slides with inadequate RBC areas presented by the CellaVision DM96 can be decreased by better slide preparation, better training of technologists in the use of the CellaVision DM96 for this purpose, and/or adjustments in the CellaVision DM96. Otherwise, laboratories using the CellaVision DM96 may have to decide not to provide information on RBC and platelet morphologic features as part of their standard manual differentials, unless there is reason to suspect an RBC or a platelet abnormality (ie, an RBC- or a plateletspecific flag from the automated cell counter). Such a development would be in line with the fact that already, neither RBC nor platelet morphologic features are reviewed microscopically for differential samples that are reported without preparation of a slide directly from the automated cell counter. The CellaVision DM96 presents preclassified WBCs to a reviewer for approval or reclassification. We found that the instrument overall correctly preclassified 82% of all cells. Preclassification of WBCs was most accurate for segmented neutrophils (92.5% correctly classified), lymphocytes (96.4%), and monocytes (81.4%); it was less reproducible for basophils (80.0%) and for eosinophils (63.2%) (Table 2). A study based on images of WBCs sent to 2,400 laboratories in surveys of the College of American Pathologists and classified by human observers was published in ; 99% of segmented neutrophils, 96% of lymphocytes, 87% of monocytes, 95% of basophils, and 96% of eosinophils were identified correctly by the participants. This indicates that the ability of the CellaVision DM96 to preclassify segmented neutrophils, lymphocytes, and monocytes is similar to that of the human observer, whereas improvements to the neural network are needed for eosinophils and basophils. After the WBCs presented by the instrument are reclassified or confirmed by the user, final results are released. We compared the final results obtained with digital image analysis to differentials from direct microscopy. The correlation coefficients for differential counts for the WBC classes found in blood samples with a qualitatively normal differential were between 0.67 and 0.96 (Figure 3). Koepke and colleagues 13 evaluated the performance of differential WBC counting by 73 technologists and technicians in 5 different laboratories in a large medical center by comparing them with differentials performed by referees. Correlation coefficients were lowest for basophils (0.32) and monocytes (0.41) and highest for lymphocytes (0.73), eosinophils (0.83), and segmented neutrophils (0.87). The correlation coefficients for the cell counts from the CellaVision DM96 compare quite favorably with these results. Correlation for the various cell classes with the direct microscopic differential also was very similar to the findings reported by Swolin and colleagues 6 for the DiffMaster Octavia, the predecessor of the CellaVision DM96. The low correlation for band forms might be due to the known high variability in band counts. 14 There was a significant percentage of cases that were false-positive or false-negative for clinically significant abnormalities on the CellaVision DM96 compared with the results of direct microscopy. Similar findings were reported by Koepke and coworkers, 13 who found that the sensitivity of Table 10 Average Time per Slide Using the Standard Workload of the Differential Workstation (Approximately 50% Fully Manual Differential Counts and 50% Scanned Slides With Release of the Automated Differential Count) Staff Member/No. of Slides Microscope CellaVision DM96 Very experienced technologist with special expertise in WBC differentials (n = 76) Technician with no special expertise in WBC differentials (n = 160) Downloaded from Am J Clin Pathol 2005;124:

11 Kratz et al / DIGITAL IMAGE ANALYSIS IN HEMATOLOGY slide review by technologists ranged from 34% to 100%, depending on the abnormality. Review of the discrepant cases in our study showed that the majority of discrepant cases were due to variation that was not statistically significant (within the 95% confidence interval of the number of cells on which the differential was based) or to clinically insignificant variations in nomenclature (eg, blasts vs other cells, query blasts). When we corrected for these statistically or clinically insignificant discrepancies, the sensitivity of the CellaVision DM96 was between 95% and 100% and the specificity between 88% and 97%, depending on the abnormality. The remaining discrepancies were mostly (16/19 discrepant cases) due to misidentification of cells on digital images by the technologists. In the remaining 3 discrepant cases, cells that were clearly present on the digital images obtained with the CellaVision DM96 had not been reported on direct microscopy. A rereview of these slides by experienced reviewers at first did not allow the visualization of these cells with the microscope. When an additional microscopic differential based on up to 300 cells was performed on these slides, the immature myeloid cells seen with the CellaVision DM96 were seen with the microscope. Therefore, these discrepant cases were most likely due to random variation outside the 95% confidence interval of a 100-cell differential count. As described in the preceding paragraphs, when we found discrepancies between the findings with direct microscopy and with the CellaVision DM96, we rereviewed the digital images provided by the CellaVision DM96 and determined that in a number of cases, the initial reviewers had misidentified cells. This episode underlines a strength of digital image analysis compared with direct microscopy: we were able to do quality control on a cell-by-cell basis, identify errors, correct them, and perform corrective action by educating the technologists who had made the misidentification. Such interventions are much more difficult or impossible using direct microscopy. Even after correction for misclassified cells, there were more positive findings on the CellaVision DM96 than on direct microscopy. Most of these false-positive findings involved very small populations of cells (<5%). Since the presence of abnormal WBCs will only be recognized with greater than 95% certainty in a 100-cell differential if the percentage of the abnormal cell population is at least 5%, random variation is the most likely explanation for these false positives. 15 Another possibility is that some cells were misclassified or missed on direct microscopy. Finally, it is very plausible that some of the positive findings identified with the CellaVision DM96 and not reported on direct microscopy were due to the fact that it is not possible for the technologist to skip cells on the CellaVision DM96. Every cell has to be classified (and this classification remains subject to rereview), leading to a higher likelihood of reporting rare or unusual cells when using the CellaVision DM96. To evaluate the impact of the CellaVision DM96 on labor costs, we performed 2 studies. First, we timed 6 technologists performing 10 fully manual differentials with the microscope and with the CellaVision DM96. It took the technologists, on average, 5.1 minutes per slide to perform a manual differential; the time required with the CellaVision DM96 was longer, 6.4 minutes. The limitations of this study are that the technologists had years of experience in performing differentials with a microscope; in contrast, the CellaVision DM96 was new to them, and they lacked any prolonged experience on this instrument. It therefore is possible that with increased exposure to the CellaVision DM96, the average time per manual differential would have decreased. The second limitation of this study was that it addressed only fully manual differentials; the usual workflow at the differential workstation in our laboratory consists of approximately 50% manual differentials and 50% slides that only need to be reviewed for release of the differential results from the automated cell counter (scanned slides). It is possible that the major labor saving that can be derived from the CellaVision DM96 will be from these scanned slides, for which the technologist is conveniently presented with a large number of WBCs on a computer screen that can be reviewed quickly for the presence of any abnormality, with subsequent release of the differential results from the automated cell counter. To address these possibilities, we performed a second timing study, in which 2 employees were trained more extensively in the use of the CellaVision DM96. They were asked to result randomly selected cases with the microscope and to process the same cases with the CellaVision DM96. These cases consisted of both fully manual differentials and of slides that required only a scan. As shown in Table 10, it took a very experienced technologist approximately the same time per slide to result differentials from the CellaVision DM96 (1.5 minutes per slide) as from the microscope (1.4 minutes per slide). However, a less experienced employee was faster using the CellaVision DM96 (3.0 minutes per case) than using the microscope (4.1 minutes per case), indicating that after training on the CellaVision DM96 and with a caseload similar to the normal workflow in the laboratory, an average technologist is likely to be more productive using the CellaVision DM96 than using the microscope. The possibility of a learning curve was also indicated by the fact that the less experienced employee was 10% faster using the CellaVision DM96 when analyzing the second half of her assigned slides than when analyzing the first half. The cost-effectiveness of the CellaVision DM96 will be determined by a variety of factors. Our data indicate that the instrument may allow some labor savings by decreasing the time for technologists trained in the use of the instrument to release a differential. Increased efficiencies also could be achieved by centralizing the expertise for morphologic diagnosis and by making it less time-consuming to obtain confirmation of abnormal findings by referees. These cost savings 780 Am J Clin Pathol 2005;124: Downloaded 780 from

Performance evaluation and relevance of the CellaVision TM DM96 system in routine analysis and in patients with malignant hematological diseases

Performance evaluation and relevance of the CellaVision TM DM96 system in routine analysis and in patients with malignant hematological diseases ORIGINAL ARTICLE INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY Performance evaluation and relevance of the CellaVision TM DM96 system in routine analysis and in patients with malignant hematological diseases

More information

Impact of Integrating Rumke Statistics to Assist with Choosing Between Automated Hematology Analyzer Differentials vs Manual Differentials

Impact of Integrating Rumke Statistics to Assist with Choosing Between Automated Hematology Analyzer Differentials vs Manual Differentials Impact of Integrating Rumke Statistics to Assist with Choosing Between Automated Hematology Analyzer Differentials vs Manual Differentials Laura Stephens, 1 Wendy Hintz-Prunty, 1 Hans-Inge Bengtsson, 2

More information

BACKGROUND. Absolute neutrophil counts (ANCs) and absolute phagocyte counts

BACKGROUND. Absolute neutrophil counts (ANCs) and absolute phagocyte counts 2681 Utility of Automated Counting to Determine Absolute Neutrophil Counts and Absolute Phagocyte Counts for Pediatric Cancer Treatment Protocols Nobuko Hijiya, M.D. 1,2 Mihaela Onciu, M.D. 3 Scott C.

More information

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa The Role of the FBC in clinical decision making History Examination Investigations Decision 70% FBC Laboratory

More information

Sue Jung Kim, MD, MS; Yoonjung Kim, MD; Saeam Shin, MD; Jaewoo Song, MD, MS; Jong Rak Choi, MD, PhD

Sue Jung Kim, MD, MS; Yoonjung Kim, MD; Saeam Shin, MD; Jaewoo Song, MD, MS; Jong Rak Choi, MD, PhD Comparison Study of the Rates of Manual Peripheral Blood Smear Review From 3 Automated Hematology Analyzers, Unicel DxH 800, ADVIA 2120i, and XE 2100, Using International Consensus Group Guidelines Sue

More information

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes A. M. CENCI *1, M. MACONI *2, and B. CASOLARI *1 *1 Laboratory of

More information

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer Hematopathology / EVALUATION OF THE ABBOTT CELL-DYN 4 HEMATOLOGY ANALYZER Evaluation of the Abbott CELL-DYN 4 Hematology Analyzer Ernesto Grimaldi, MD, and Francesco Scopacasa, PhD Key Words: Abbott CD

More information

2008 CAP TODAY Q & A

2008 CAP TODAY Q & A 2008 CAP TODAY Q & A Q. How often should we document hematology competencies to ensure consistency of morphologic observations for manual differentials and fluids? Should we do this every six months or

More information

LYDIA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE, M.D., SANDRA WINGRELD, MT(ASCP), AND EDWARD KEYSER, MT(ASCP)

LYDIA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE, M.D., SANDRA WINGRELD, MT(ASCP), AND EDWARD KEYSER, MT(ASCP) Laboratory Evaluation of Differential White Blood Cell Count nformation from the S-Plus V and H-1 in Patient Populations Requiring Rapid "Turnaround" Time LYDA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE,

More information

Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature

Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature Gene L. Gulati, PhD; Lawrence J. Hyland, MD; William Kocher, MD;

More information

Combining. and New Diagnostic. to Help Clinicians Achieve. Patient Outcomes at. per Healthcare Encounter

Combining. and New Diagnostic. to Help Clinicians Achieve. Patient Outcomes at. per Healthcare Encounter Combining and New Diagnostic to Help Clinicians Achieve Patient Outcomes at per Healthcare Encounter Holly McDaniel, MD hmcdaniel@clinpath.com Holly.mcdaniel@bannerhealth.com Holly McDaniel, MD AP/CP and

More information

Vision Hema. Additional modules

Vision Hema. Additional modules Vision Hema Additional modules Vision Hema RTC Digital morphology of reticulocytes Innovative solution for the study of reticulocytes Reticulocytes Reticulocytes young erythrocytes that appear in the bone

More information

HISTOLOGY VIRTUAL LABORATORY BLOOD AND LYMPHATICS SYSTEM

HISTOLOGY VIRTUAL LABORATORY BLOOD AND LYMPHATICS SYSTEM HISTOLOGY VIRTUAL LABORATORY BLOOD AND LYMPHATICS SYSTEM Login: http://histopath.westernu.edu Histology Atlas AND Virtual Histology links. I. HEMATOLOGY - PERIPHERAL BLOOD Purpose: To be able to identify

More information

Ordering Physician CLIENT,CLIENT. Collected REVISED REPORT

Ordering Physician CLIENT,CLIENT. Collected REVISED REPORT HPWET Hematopathology Consultation, MML Embed Client Hematopathology Consult REVISED INAL DIAGNOSIS Interpretation Peripheral blood, bone marrow aspirate and biopsies, bilateral iliac crests: 1. Normocellular

More information

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100 Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100 R. HERKLOTZ and A. R. HUBER Zentrum für Labormedizin, Kantonsspital Aarau, 5001 Aarau, Switzerland. The automated leukocyte(wbc)

More information

A Look Into the Determination of Cell Morphology in Hematology in the 21 st Century. Ramon Simon-Lopez, MD Global Scientific Director Beckman Coulter

A Look Into the Determination of Cell Morphology in Hematology in the 21 st Century. Ramon Simon-Lopez, MD Global Scientific Director Beckman Coulter A Look Into the Determination of Cell Morphology in Hematology in the 21 st Century Ramon Simon-Lopez, MD Global Scientific Director Beckman Coulter Is cell morphology important? AML M7 CLL CD5 CD19 NHL

More information

A HemoCue Novel Technology Product for 5-Part Differentiation of Leukocytes Point-of-Care

A HemoCue Novel Technology Product for 5-Part Differentiation of Leukocytes Point-of-Care A HemoCue Novel Technology Product for -Part Differentiation of Leukocytes Point-of-Care Stellan Lindberg, M Sc Medical Director HemoCue AB Kuvettgatan 1 SE-262 71 Ängelholm Sweden Stellan.Lindberg@hemocue.se

More information

Literature List Body Fluids

Literature List Body Fluids Literature List Body Fluids Customer Information March 2018 Date: March 2018 Subject: Literature List body fluids Issued by: Scientific Customer Services Number: 180312 Note: Whether references are given

More information

ADVANCED HAEMATOLOGY BATTLE OF THE BANDS. Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND

ADVANCED HAEMATOLOGY BATTLE OF THE BANDS. Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND ADVANCED HAEMATOLOGY BATTLE OF THE BANDS Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND The identification of immature neutrophils (bands, metamyelocytes,

More information

Validation of Criteria for Smear Review Following Automated Blood Cell Analysis in Ain Shams University Laboratory

Validation of Criteria for Smear Review Following Automated Blood Cell Analysis in Ain Shams University Laboratory Validation of Criteria for Smear Review Following Automated Blood Cell Analysis in Ain Shams University Laboratory Azza S Eldanasoury 1, Noha H Boshnak 2, Raghda E Abd El Monem 3 1 MD, Professor of Clinical

More information

Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer

Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer Science Gene Gulati, PhD, Sean Cote, DO, Eric Behling, MD, William Kocher, MD

More information

Detection of important abnormalities of the differential

Detection of important abnormalities of the differential J Clin Pathol 1989;42:772-776 Detection of important abnormalities of the differential count using the Coulter STKR blood counter D F BARNARD, S A BARNARD, A B CARTER, S J MACHN, K G PATTERSON, A YARDUMAN

More information

Hematology & Coagulation Practicum Objectives CLS - 647

Hematology & Coagulation Practicum Objectives CLS - 647 Hematology & Coagulation Practicum Objectives CLS - 647 The following objectives are to be completed by the student for successful completion of this clinical rotation. The objectives within the psychomotor

More information

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation.

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Sysmex America White Paper One Nelson C. White Parkway,

More information

Performance evaluation of Celltac G: a new automated hematology analyzer

Performance evaluation of Celltac G: a new automated hematology analyzer Original Article Performance evaluation of Celltac G: a new automated hematology analyzer Masaaki Sugiyama 1, Tsukasa Kobayashi 1, Yuki Jisyage 2, Shigeko Yamamoto 2, Yutaka Nagai 2,3 and Hiroshi Kondo

More information

LEANING Away Waste and Improving Patient Care with Advances in Hematology

LEANING Away Waste and Improving Patient Care with Advances in Hematology LEANING Away Waste and Improving Patient Care with Advances in Hematology Selke Mantie, MLS (ASCP), CLS (CSMLS), SSGBC Disclosure I am receiving an honorarium from Sysmex for today s presentation Objectives

More information

Kathleen Finnegan MS MT(ASCP)SHCM

Kathleen Finnegan MS MT(ASCP)SHCM Kathleen Finnegan MS MT(ASCP)SHCM Discuss the history of hematology automation and digital differentials. Discuss the HemoFAXS Hematology Analysis System by Tissue Gnostics. Review automated microscopy

More information

Detection and Classification of Acute Leukemia by the Coulter STKS Hematology Analyzer

Detection and Classification of Acute Leukemia by the Coulter STKS Hematology Analyzer HEMATOPATHOLOGY Detection and Classification of Acute Leukemia by the Coulter STKS Hematology Analyzer JAMES D. HOYER, MD, 1 CATHERINE P. FISHER, MD, 1 VICKI M. SOPPA, BS, 1 KAY L. LANTIS, SH, MT(ASCP),

More information

Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities

Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities JOANNE CORNBLEET, M.D. AND SALLY KESSINGER, MT(ASCP)SH The authors evaluated the Coulter 15 S-Plus

More information

Flow Cytometric White Blood Cell Differential Using CytoDiff is Excellent for Counting Blasts

Flow Cytometric White Blood Cell Differential Using CytoDiff is Excellent for Counting Blasts Original Article Diagnostic Hematology Ann Lab Med 215;35:28-34 http://dx.doi.org/1.3343/alm.215.35.1.28 ISSN 2234-386 eissn 2234-3814 Flow Cytometric White Blood Cell Differential Using is Excellent for

More information

3. Blood Cell Histograms:

3. Blood Cell Histograms: LECTURE MODULE 6c: ELECTRONIC CELL COUNTING PART III 3. Blood Cell Histograms: a. The Coulter cell counters today provides size distributions of the cellular content: 1) volume given in µm 3 or fl vs relative

More information

Blood Cell Identification Graded

Blood Cell Identification Graded BCP-21 Blood Cell Identification Graded Case History The patient is a 37-year-old female with a history of multiple sickle cell crises. She now presents with avascular necrosis of the left hip. Laboratory

More information

NEW YORK STATE CYTOHEMATOLOGY PROFICIENCY TEST PROGRAM Glass Slide - November 2016

NEW YORK STATE CYTOHEMATOLOGY PROFICIENCY TEST PROGRAM Glass Slide - November 2016 NEW YORK STATE CYTOHEMATOLOGY PROFICIENCY TEST PROGRAM Glass Slide - November 2016 Results from this proficiency test event are available at: http://www.wadsworth.org/regulatory/clep/pt/summaries SLIDE

More information

Interpreting Hematology Scatter-Plots; One Cancer Center s Keys to Seeing the BIG Picture

Interpreting Hematology Scatter-Plots; One Cancer Center s Keys to Seeing the BIG Picture Interpreting Hematology Scatter-Plots; One Cancer Center s Keys to Seeing the BIG Picture Barbara L. Burch, MHA MT (ASCP) Laboratory Manager New York University Clinical Cancer Center Disclosure Ms Burch

More information

CELL-DYN Strength in Technology, Proven Reliability. Optical WBC Technology. Patented M.A.P.S.S. Differential. Multiple Technologies

CELL-DYN Strength in Technology, Proven Reliability. Optical WBC Technology. Patented M.A.P.S.S. Differential. Multiple Technologies CELL-DYN 3700 Strength in Technology, Proven Reliability Optical WBC Technology Patented M.A.P.S.S. Differential Multiple Technologies CELL-DYN 3700 Multiple Technologies One Superior Result Multiple Technologies

More information

Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital

Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital HEMATOPATHOLOGY Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital JOAN-LLUIS VIVES-CORRONS, MD, ISABEL BESSON, PHD, JOSEP MARIA JOU, MD, AND GABRIELA GUTIERREZ, MB The

More information

Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How?

Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How? Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How? Dennis B. DeNicola, DVM, PhD, DACVP Chief Veterinary Educator IDEXX Laboratories, Inc. Westbrook, ME USA Adjunct Professor of Veterinary

More information

Morfologia normale e patologica

Morfologia normale e patologica Morfologia normale e patologica Gina Zini Centro di Ricerca ReCAMH Dpt. Ematologia Università Cattolica S. Cuore - Roma EMATOLOGIA DI LABORATORIO: percorsi diagnostici e obiettivi clinici. Milano 11-12

More information

Collected: , PM Sent: , PM Received: , PM Preliminary: , PM. Notification Status: COMPREHENSIVE DIAGNOSIS

Collected: , PM Sent: , PM Received: , PM Preliminary: , PM. Notification Status: COMPREHENSIVE DIAGNOSIS PATIENT Name:HIPAA, Compliant DOB: 03-25-1945 (60 yr) ID#: xxx-xx-0000 Sex: M Tel: xxx-xxx-xxxx SPECIMEN Your No:WS05-xxxx Case No:C05-00xxx Req. No:Txxxxx Collected: 06-08-05, PM Sent: 06-09-05, PM Received:

More information

Lessons learned in the use of digital imaging at Memorial Sloan Kettering Cancer Center

Lessons learned in the use of digital imaging at Memorial Sloan Kettering Cancer Center Lessons learned in the use of digital imaging at Memorial Sloan Kettering Cancer Center Executive War College May 3, 2018 Victor E. Reuter, M.D. Vice-Chair, Department of Pathology Medical Director, Warren

More information

Requirements: Glass slides Leishman stain Microscopes Disposable needles Vials containing anticoagulants Methylated-spirit Staining rack

Requirements: Glass slides Leishman stain Microscopes Disposable needles Vials containing anticoagulants Methylated-spirit Staining rack Aim: To make a blood smear and to count the different types of leucocytes present in a stained blood smear and express their relative counts in percentage Principle - The blood contains various types of

More information

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S 111 Riv Med Lab - JLM, Vol. 4, N. 2, 2003 First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S H. Vandeputte a, D. Darre

More information

It s Easier Than you Think How to Implement the Advanced Clinical Parameters

It s Easier Than you Think How to Implement the Advanced Clinical Parameters It s Easier Than you Think How to Implement the Advanced Clinical Parameters Implementation of our New Hematology Platform ( 9 HST Lines - 9 Facilities - 9 Month) Outline The Organization Banner Health

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

Notes for the 2 nd histology lab

Notes for the 2 nd histology lab Notes for the 2 nd histology lab Note : Please refer to the slides and see the morphological characteristics of each cell, as the practical exam will be in the form of figures. SLIDE #2 Erythropoiesis

More information

Differential Blood Smear H3

Differential Blood Smear H3 Verein für Association pour le Associazione per il medizinische Qualitätskontrolle contrôle de qualité médical controllo di qualità medico Report Differential Blood Smear H3 MQ 2015-4 MQ, Institut für

More information

Detail PRINCIPLE: Body fluids other than blood and urine will be analyzed according to their site of origin, and the providers specific orders.

Detail PRINCIPLE: Body fluids other than blood and urine will be analyzed according to their site of origin, and the providers specific orders. Subject Body Fluid Analysis - Affiliate Index Number Lab-8760 Section Laboratory Subsection Regional/Affiliates Category Departmental Contact Munson, Karen Last Revised 6/28/2017 References Required document

More information

Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential

Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential H e m a t o p a t h o l o g y / REFRIGERATION IMPROVES CBC COUNT STABILITY Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential Brent L. Wood, MD, PhD,

More information

Validation of the Efficacy of a Practical Method for Neutrophils Isolation from Peripheral Blood

Validation of the Efficacy of a Practical Method for Neutrophils Isolation from Peripheral Blood Validation of the Efficacy of a Practical Method for Neutrophils Isolation from Peripheral Blood JONATHAN DEGEL, MASIH SHOKRANI OBJECTIVE: The objectives of this study were to validate the Polymorphprep

More information

XN series. Case interpretation. Gebruikersdag Vlaanderen- 6 oktober 2016

XN series. Case interpretation. Gebruikersdag Vlaanderen- 6 oktober 2016 XN series Case interpretation Gebruikersdag Vlaanderen- 6 oktober 2016 Fluorescence flow cytometry RET channel PLT-F channel WDF channel WPC channel WNR channel Case 1 Case 1: Initial measurement Patient

More information

Course Title: Thin Prep PAP Smears. Number of Clock Hours: 3 Course Title #

Course Title: Thin Prep PAP Smears. Number of Clock Hours: 3 Course Title # Course Title: Thin Prep PAP Smears Number of Clock Hours: 3 Course Title #5240303 Course Introduction There is a new laboratory test that is gaining popularity as a cancer screening tool for cervical cancer.

More information

CELL-DYN 3700 Strength in Technology, Proven Reliability

CELL-DYN 3700 Strength in Technology, Proven Reliability CELL-DYN 3700 Strength in Technology, Proven Reliability Optical WBC Technology Patented MAPSS Differential Multiple Technologies CELL-DYN 3700 Multiple Technologies One Superior Result Multiple Technologies

More information

An Automated Blood Smear Analysis System

An Automated Blood Smear Analysis System An Automated Blood Smear Analysis System Clinical Experience and Performance JEROME S. NOSANCHUK, M.D., PAT DAWES, M.T.(ASCP), ALLISON KELLY, B.S., AND CHARLES HECKLER, B.S. Nosanchuk, Jerome S., Dawes,

More information

CAP Laboratory Improvement Programs. Utility of Repeat Testing of Critical Values. A Q-Probes Analysis of 86 Clinical Laboratories

CAP Laboratory Improvement Programs. Utility of Repeat Testing of Critical Values. A Q-Probes Analysis of 86 Clinical Laboratories CAP Laboratory Improvement Programs Utility of Repeat Testing of Critical Values A Q-Probes Analysis of 86 Clinical Laboratories Christopher M. Lehman, MD; Peter J. Howanitz, MD; Rhona Souers, MS; Donald

More information

Investigating the recheck rules for urine analysis in children

Investigating the recheck rules for urine analysis in children Investigating the recheck rules for urine analysis in children Y.M. He*, S.W. Yao*, Y.J. Huang, B.S. Liang and H.Y. Liu Clinical Laboratory, Guangzhou Women and Children s Medical Center, Guangzhou Medical

More information

Implementation Guide for the DoseControl Dosimetry System

Implementation Guide for the DoseControl Dosimetry System 1.0 SCOPE The GEX DoseControl System is intended for use as a single dosimetry system solution, specifically developed to satisfy the generic dosimetry needs of radiation sterilization application users

More information

VETERINARY HEMATOLOGY ATLAS OF COMMON DOMESTIC AND NON-DOMESTIC SPECIES COPYRIGHTED MATERIAL SECOND EDITION

VETERINARY HEMATOLOGY ATLAS OF COMMON DOMESTIC AND NON-DOMESTIC SPECIES COPYRIGHTED MATERIAL SECOND EDITION VETERINARY HEMATOLOGY ATLAS OF COMMON DOMESTIC AND NON-DOMESTIC SPECIES SECOND EDITION COPYRIGHTED MATERIAL CHAPTER ONE HEMATOPOIESIS GENERAL FEATURES All blood cells have a finite life span, but in normal

More information

Standard Operating Procedure

Standard Operating Procedure Subject Differential: Counting and Morphology Index Number Lab-5074 Section Laboratory Subsection Hematology Automated Laboratory Category Departmental Contact Moldenhauer, Emily C Last Revised 11/28/2017

More information

Myelodysplastic Syndrome Case 158

Myelodysplastic Syndrome Case 158 Myelodysplastic Syndrome Case 158 Dong Chen MD PhD Division of Hematopathology Mayo Clinic Clinical History 86 year old man Persistent borderline anemia and thrombocytopenia. His past medical history was

More information

Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis

Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis Hematopathology / LEFT SHIFT MODEL PREDICTS INFECTION Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis Michael J. Wile, MD, 1 Louis D. Homer, MD, PhD, 2 Stede Gaehler,

More information

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags.

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags. WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags www.sysmex.com/us Introduction For many decades, physicians have relied on

More information

of s and s Disclosure

of s and s Disclosure The s of s and s Successfully implement the Advanced Clinical Parameters to benefit the patient, the lab and the Accountable Care Organization Model Disclosure Honorarium from Sysmex for this lecture 1

More information

The White Blood Cell Differential. Evaluation of Rapid Impression Scanning Versus the Routine Manual Count

The White Blood Cell Differential. Evaluation of Rapid Impression Scanning Versus the Routine Manual Count 628 BRODY ET AL. A.J.C.P. May 987 culating lymphocytes may reflect their release from reticuloendothelial organs, entry of new lymphocyte cohorts from lymphoid tissue, or spontaneous reexpression of epitopes

More information

Blood smear analysis in the emergency veterinary patient

Blood smear analysis in the emergency veterinary patient Vet Times The website for the veterinary profession https://www.vettimes.co.uk Blood smear analysis in the emergency veterinary patient Author : Ashley Wemple Categories : RVNs Date : April 1, 2010 Ashley

More information

Flow cytometry leukocyte differential : a critical appraisal

Flow cytometry leukocyte differential : a critical appraisal Flow cytometry leukocyte differential : a critical appraisal Francis Lacombe Flow cytometry department University Hospital of Bordeaux, Pessac, France francis.lacombe@chu-bordeaux.fr 2008 HORIBA ABX, All

More information

Open Range Software, LLC

Open Range Software, LLC Comprehensive Tracking System CTS User Manual CTS IH/Medical Surveillance Interface Open Range Software, LLC Revised: 09/04/2013 CONTENTS Introduction...3 WHAT IS THE CTS IH/MEDICAL SURVEILLANCE INTERFACE?...3

More information

Peripheral Blood Smear: Diagnostic Clues and Algorithms

Peripheral Blood Smear: Diagnostic Clues and Algorithms Transcript Details This is a transcript of a continuing medical education (CME) activity accessible on the ReachMD network. Additional media formats for the activity and full activity details (including

More information

The measurements of pleural fluid WBC count

The measurements of pleural fluid WBC count Variations in Pleural Fluid WBC Count and Differential Counts With Different Sample Containers and Different Methods* Barrett D. Conner, MD; Y. C. Gary Lee, MD; Paul Branca, MD, FCCP; Jeffrey T. Rogers,

More information

MALARIA PARASITE COUNTING

MALARIA PARASITE COUNTING VERSION 1 EFFECTIVE DATE: 01/01/2016 MALARIA PARASITE COUNTING MALARIA MICROSCOPY STANDARD OPERATING PROCEDURE MM-SOP-09 1. PURPOSE AND SCOPE To describe the procedure for counting malaria parasites on

More information

The LaboratoryMatters

The LaboratoryMatters Laboratory Medicine Newsletter for clinicians, pathologists & clinical laboratory technologists. A Initiative. Complete Blood Count This issue highlights: CBC, while ubiquitous, is an excellent diagnostic

More information

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Alert or critical values represent those assay results that require prompt, rapid clinical attention to avert significant study-participant

More information

Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana

Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana IS THERE STILL A ROLE FOR THE MANUAL BLOOD COUNT? Indication for examination

More information

Combination of life science entrepreneurs, software development and machine vision experts & recognized scientists.

Combination of life science entrepreneurs, software development and machine vision experts & recognized scientists. Combination of life science entrepreneurs, software development and machine vision experts & recognized scientists. Kaisa Helminen CEO Previously: Sartorius Thermo Scientific Finnzymes Johan Lundin MD,

More information

Hematopathology Case Study

Hematopathology Case Study Hematopathology Case Study AMP Outreach Course 2009 AMP Annual Meeting John Greg Howe Ph.D. Department of Laboratory Medicine Yale University School of Medicine November 19, 2009 HISTORY Case History An

More information

Telepathology 2/17/2011. and The Future of Pathology (or Why did we change our Practice Model) University Health Network IMAGINE3: 3 IMAGINE 2!

Telepathology 2/17/2011. and The Future of Pathology (or Why did we change our Practice Model) University Health Network IMAGINE3: 3 IMAGINE 2! Telepathology and The Future of Pathology (or Why did we change our Practice Model) Jagdish Butany, MBBS, MS, FRCPC (a)why did we implement telepathology (WSI) at UHN? Consultant Cardiovascular Pathologist/Director

More information

HumaCount 5D. Outperforming 5-part Hematology System. CoreLab DX. Hematology

HumaCount 5D. Outperforming 5-part Hematology System. CoreLab DX. Hematology HumaCount 5D Outperforming 5-part Hematology System Direct Capillary Process by OptimalCount Technology Distinct 5-part Differentiation Definite Immature Cell Count (LIC, ALY) CoreLab DX Hematology 5-part

More information

10/30/2015. XN Case Studies: Every Picture Tells a Story

10/30/2015. XN Case Studies: Every Picture Tells a Story XN Case Studies: Every Picture Tells a Story Jill Crist MT(ASCP)Field Product Specialist OBJECTIVES Explain how scattergrams and histogram pictures can provide great insight into abnormal hematology samples

More information

January Testing Delays and Spurious Results Caused by Improper Specimen Collection

January Testing Delays and Spurious Results Caused by Improper Specimen Collection January 2011 Testing Delays and Spurious Results Caused by Improper Specimen Collection Lauren Anthony, MD, MT(ASCP)SBB Medical Director, Allina Medical Laboratories Collecting a specimen may seem routine,

More information

We noted a problem with the use of the hematology

We noted a problem with the use of the hematology Standard Article J Vet Intern Med 2017;31:1708 1716 Feline Differential Leukocyte Count with ProCyte Dx: Frequency and Severity of a Neutrophil-Lymphocyte Error and How to Avoid It H.W. Tvedten, V. Andersson,

More information

Urine Sediment Photomicrographs/Photographs

Urine Sediment Photomicrographs/Photographs Urine Sediment Photomicrographs/Photographs Case History CMP-17 This urine sample is from a 48-year-old male with a 30-year history of diabetes mellitus and new onset renal failure. Laboratory data include:

More information

Auto Validation and Management Reports Using Sysmex WAM. Alfonso Ziccardi Laboratory Operation Manager for AP/CP/Safety Officer

Auto Validation and Management Reports Using Sysmex WAM. Alfonso Ziccardi Laboratory Operation Manager for AP/CP/Safety Officer Auto Validation and Management Reports Using Sysmex WAM Alfonso Ziccardi Laboratory Operation Manager for AP/CP/Safety Officer SUCCESS BEYOND FINANCES Hematology Testing Breakdown NYHQ Descrete testing

More information

Laboratory for diagnosis of THALASSEMIA

Laboratory for diagnosis of THALASSEMIA SCBM343 CLINICAL PATHOLOGY 2(1-2-3) Laboratory for diagnosis of THALASSEMIA PORNTHIP CHAICHOMPOO pornthip.chh@mahidol.ac.th Acknowledgements Dr. Pranee Winichagoon Fucharoen Ms. Pornnapa Khampan Thalassemia

More information

Digital Pathology Diagnosis Assistance System

Digital Pathology Diagnosis Assistance System August 3, 2011 Digital Pathology Diagnosis Assistance System Computer s s eye and Pathologist s s eye Ogura M, Saito A, Graf HP, Marugame A, Kiyuna T, Yamashita Y, Cosatto E, Malon C & Fukumoto M. Innovative

More information

DR SUDHIR MEHTA MD,MNAMS,FICP. Senior Professor & Head Medical Unit SMS Medical College & Hospital Jaipur

DR SUDHIR MEHTA MD,MNAMS,FICP. Senior Professor & Head Medical Unit SMS Medical College & Hospital Jaipur DR SUDHIR MEHTA MD,MNAMS,FICP Senior Professor & Head Medical Unit SMS Medical College & Hospital Jaipur s.smehta@hotmail.com CBC..What is the Utility of performing this basic Hematology Test? 10/31/2010

More information

difficult and may not be practical. Nevertheless, the performance characteristics of rapid screening have not been completely characterized.

difficult and may not be practical. Nevertheless, the performance characteristics of rapid screening have not been completely characterized. Anatomic Pathology / PERFORMANCE CHARACTERISTICS OF RAPID PRESCREENING Performance Characteristics of Rapid (0-Second) Prescreening Implications for Calculating the False-Negative Rate and Comparison With

More information

Haematologist-reviewed peripheral blood smear in paediatric practice

Haematologist-reviewed peripheral blood smear in paediatric practice Singapore Med J 2018; 59(2): 64-68 https://doi.org/10.11622/smedj.2018013 Haematologist-ed peripheral blood smear in paediatric practice Anselm Chi-wai Lee, MBBS, FAMS ABSTRACT Manual examination of the

More information

Myelodysplasia/Myeloproliferative Neoplasms (MDS/MPN) Post-HCT Data

Myelodysplasia/Myeloproliferative Neoplasms (MDS/MPN) Post-HCT Data Instructions for Myelodysplasia/Myeloproliferative Neoplasms (MDS/MPN) Post-HCT Data (Form 2114) This section of the CIBMTR Forms Instruction Manual is intended to be a resource for completing the Myelodysplasia/Myeloproliferative

More information

SEED HAEMATOLOGY. Medical statistics your support when interpreting results SYSMEX EDUCATIONAL ENHANCEMENT AND DEVELOPMENT APRIL 2015

SEED HAEMATOLOGY. Medical statistics your support when interpreting results SYSMEX EDUCATIONAL ENHANCEMENT AND DEVELOPMENT APRIL 2015 SYSMEX EDUCATIONAL ENHANCEMENT AND DEVELOPMENT APRIL 2015 SEED HAEMATOLOGY Medical statistics your support when interpreting results The importance of statistical investigations Modern medicine is often

More information

g/dl g/dl

g/dl g/dl Bio 151 Laboratory Manual Human Anatomy & Physiology II #3 - Objectives: Determine the hemoglobin content of sheep blood Know the normal values of hemoglobin for males and females Be able to calculate

More information

Hematology 101. Cindy Rogers, MT(ASCP) Diagnostics System Specialist

Hematology 101. Cindy Rogers, MT(ASCP) Diagnostics System Specialist Hematology 101 Cindy Rogers, MT(ASCP) Diagnostics System Specialist More Acronyms...» CBC» RBC» HGB» HCT» WBC» MPV» PLT» RDW» DIFF» H&H» Complete Blood Count» Red Blood Cell» Hemoglobin» Hematocrit» White

More information

Top 5 Pitfalls of Spirometry And How to Avoid Them

Top 5 Pitfalls of Spirometry And How to Avoid Them Don t Blow It Make Every Breath Count Top 5 Pitfalls of Spirometry And How to Avoid Them Is your company moving into the respiratory clinical trials market? Maybe you ve ventured into spirometry, learned

More information

Clinical Utility. Clinical Utility of the Automated IG-Count Xavier Troussard

Clinical Utility. Clinical Utility of the Automated IG-Count Xavier Troussard 5. Clinical Utility Clinical Utility of the Automated IG-Count Xavier Troussard Introduction Quantitative analysis of immature granulocytes (IG) is useful for the optimal clinical and biological management

More information

MLAB 1211 Body Fluids BF Differentials - pretest

MLAB 1211 Body Fluids BF Differentials - pretest MLAB 1211 BF Differentials - pretest Instructor: Carolyn Ragland MLAB 1211 pre-test #2 Spring No Extra Materials Needed Name Date Instructions: Neatly write your answer in the space provided. Unless otherwise

More information

EDUCATIONAL COMMENTARY DIFFERENTIATING IMMATURE PERIPHERAL BLOOD CELLS

EDUCATIONAL COMMENTARY DIFFERENTIATING IMMATURE PERIPHERAL BLOOD CELLS Educational commentary is provided through our affiliation with the American Society for Clinical Pathology (ASCP). To obtain FREE CME/CMLE credits click on Continuing Education on the left side of the

More information

Lunar Prodigy Advance

Lunar Prodigy Advance GE Medical Systems Lunar Prodigy Advance Direct-Digital Densitometry imagination at work Your practice needs to move fast, yet you want peace of mind. A partnership is a journey - expertise, support and

More information

When in doubt, ask an expert

When in doubt, ask an expert When in doubt, ask an expert MARC BEISHON No pathologist can be an expert in every type of cancer. But there is a lot that can be done to greatly improve the accuracy of diagnoses, particularly in rare

More information

2007 Workshop of Society for Hematopathology & European Association for Hematopathology Indianapolis, IN, USA Case # 228

2007 Workshop of Society for Hematopathology & European Association for Hematopathology Indianapolis, IN, USA Case # 228 2007 Workshop of Society for Hematopathology & European Association for Hematopathology Indianapolis, IN, USA Case # 228 Vishnu V. B Reddy, MD University of Alabama at Birmingham Birmingham, AL USA 11/03/07

More information

JMSCR Volume 03 Issue 05 Page May 2015

JMSCR Volume 03 Issue 05 Page May 2015 www.jmscr.igmpublication.org Impact Factor 3.79 ISSN (e)-2347-176x Time Effect Storage Artifacts of Anticoagulant EDTA on Peripheral Blood Cells Authors Dr. Ramu Thakur 1, Dr.O.P.Moorjani 2 1 Assitant

More information

Hematopathology Case Study

Hematopathology Case Study www.medfusionservices.com Hematopathology Case Study CV3515-14 JUNE Clinical Presentation: Clinical Information: A 42 year old male with history of chronic myelogenous leukemia (CML) presents with an elevated

More information