Outcome of single-chamber, ventricular pacemakers with transvenous leads implanted in children

Size: px
Start display at page:

Download "Outcome of single-chamber, ventricular pacemakers with transvenous leads implanted in children"

Transcription

1 Europace (2007) 9, doi: /europace/eum164 Outcome of single-chamber, ventricular pacemakers with transvenous leads implanted in children Massimo Stefano Silvetti 1 *, Fabrizio Drago 1, Simona Marcora 1, and Lucilla Ravà 2 1 Servizio di Aritmologia, Dipartimento Medico-Chirurgico di Cardiologia Pediatrica, Bambino Gesù Pediatric Hospital, IRCCS, Piazza S. Onofrio 4, Roma, Italy; and 2 Unità di Epidemiologia, Bambino Gesù Pediatric Hospital, Roma, Italy Received 2 May 2007; accepted after revision 12 July 2007; online publish-ahead-of-print 10 August 2007 KEYWORDS Cardiac pacing; Children; Endocardial pacing; Ventricular demand pacing; Pacemaker; Pacing complications Introduction In children with advanced or complete atrioventricular block (AVB) and normal ventricular function, ventricular demand pacing with or without a rate-responsive function (VVI/R) is generally effective, safe, and easy to perform. 1 3 VVIR pacing is sufficient for maintaining good cardiac function in most children, and the transvenous implantation of a single lead reduces the surgical risk, the operative time, and the late risk of venous thrombosis 4 6 and lead malfunction. When the child becomes an adolescent or young adult, an upgrade to a dual-chamber (DDD) or to a VDD pacing system may be indicated. 7 9 There are few recent studies which described the outcome of transvenous permanent pacing in children, especially in comparison with epicardial pacing, but to our knowledge there are no large or recent studies entirely dedicated to the outcome of VVI/R pacemaker (PM) implanted with transvenous leads in the paediatric age. We sought to address this issue. * Corresponding author. Tel: þ ; fax: þ address: silvetti@opbg.net Aims In children with bradyarrhythmias, ventricular demand, rate-responsive pacemakers (VVI/R PM) are often indicated, but no study is entirely dedicated to their outcome. Methods and results We evaluated the outcomes of children with VVI/R PM implanted at our centre, with a retrospective analysis. Between 1990 and 2005, 117 children (63 with congenital heart defects), received VVI/R PM with endocardial lead at years of age for atrioventricular block (n ¼ 105) and sinus node dysfunction. The majority of the leads were unipolar (n ¼ 78), tined (n ¼ 110), and steroid eluting (n ¼ 89). The leads were fixed to subcutaneous tissues by absorbable suture in all patients; in 17 patients, also an atrial loop was created. Follow-up (FU) was years. There were 22 system failures (19%), due to lead malfunction (n ¼ 20) and system erosion/ infection. The log-rank test for equality of survivor functions showed no significant risk factor. However, lead malfunction occurred only in the group without loop, but FU duration was longer in these patients. Complications at implantation were haemothorax (2.5%) and lead dislodgement (5%). Clinically silent occlusion of the subclavian vein was documented at FU by Echo-Doppler in 5%. Conclusion In this particular group of patients, VVI/R pacing has good results, but after long-term FU shows 19% of failures, mainly lead-related. Methods Since 1982 and 2005, 381 patients have undergone permanent PM implantation and are followed at the Cardiac Arrhythmias Service of Bambino Gesù Pediatric Hospital. 16 Endocardial pacing systems were implanted in 237 patients and epicardial pacing in the others. With a retrospective study, we reviewed the records of these patients and identified all those who underwent an implantation of a VVI/R PM at the age 15 years. We recorded the techniques used for the initial implantation, the generator and lead characteristics, and any complication. We also recorded the clinical status at most recent follow-up (FU). The study complies with the Declaration of Helsinki. Implantation procedure The indication for endocardial pacing 16 and the implant technique used in our institution has already been described. 7,8,16 The parents of children gave informed consent to the procedure using standard local consent procedure. All procedures were performed under general anaesthesia. Briefly, at initial PM implant, the endocardial pacing lead was inserted by transcutaneous puncture of the subclavian vein and fixed to the subcutaneous tissue with a slowly absorbable ligature. 19 In the last 6 years, an atrial loop was created as well. 20 Ventricular leads were positioned in the & The European Society of Cardiology All rights reserved. For permissions, please journals.permissions@oxfordjournals.org

2 Outcome of VVI/R pacemaker in children 895 non-systemic ventricular apex (left ventricle in congenitally corrected transposition of the great arteries, TGA fs, L, Lg, and in TGA fs, D, Dg after Mustard palliation), where adequate values of pacing threshold (pulse amplitude, 1 V at 0.50 ms of pulse duration), R wave sensing (.5 mv), and impedance (.300 and,1000 ohms) were achieved. A trial of high output ventricular pacing (10 V at ms) was performed during general anaesthesia when muscular paralysis was terminated to exclude inappropriate diaphragmatic stimulation. Pacing threshold, impedance, and sensing of atrial and ventricular electrograms were evaluated during the implant procedure with a Medtronic, Inc., Minneapolis, MN, USA. Pacing System Analyzer Model 5311 B and 8090 Analyzer and during FU telemetric interrogation with an appropriate Programmer. Antibiotic prophylaxis was routinely given according to our Hospital protocol. 16 Definitions Pacing system implantation was defined as the placement of a new PM generator and a lead. PM replacement was defined as the placement of only the PM generator without the insertion of new leads. Complications recorded were divided into two categories: early (occurring in the first 3 months after implantation) and late (.3 months). 16 Lead malfunction requiring new pacing system implantation was defined as: (i) exit block, (ii) abnormal threshold increase with the need of high output values causing early battery depletion and/or partial loss of capture, (iii) lead fracture, and (iv) insulation break. System infection was defined according to published criteria. 21 Follow-up The FU schedule in use in our institution has been described previously. 7,8,16 Patients were generally followed-up at 1, 3, and 6 months after implantation, and then every 6 months or as needed. Each patient underwent clinical examination, telemetric PM interrogation, and standard electrocardiogram at every FU visit. An echocardiogram, Holter monitoring, and exercise testing were performed yearly and a chest X-ray every 1 2 years. A Doppler-echo evaluation of the subclavian vein was performed in 56 patients in the last 3 years and before upgrading procedures. 7,8 Statistical analysis Statistical analysis was carried out using the Stata package, version 8.0 [StataCorp. (2003) Stata Statistical Software: Release 8.0. College Station, TX, USA: Stata Corporation]. Proportions or, when appropriate, means or medians together with standard deviation and range were computed. The median was specified if different from the mean. A P-value of,0.05 was considered significant. The Kaplan Meier method and log-rank test were used to study the longevity of the pacing system implanted. Failures included lead malfunction and pacing system infections or erosions requiring the removal of all the pacing system. Complications that did not require a new pacing system implantation (i.e. infections treated with drugs, early postoperative lead dislodgement that required only the repositioning of the lead, haemothorax at first PM implantation) were recorded as complications but not included in the analysis of the outcome of the first pacing system implanted, as this was not replaced. 13 Cases of lead removal because of extraneous causes (e.g. new cardiac surgery), cardiac transplantation and patient s death unrelated to pacing, were considered as censored or lost to FU. A Cox multivariate proportional hazard model was used to explore factors associated with the duration of the pacing system. Variables included in the model as predictors were patient s age and gender, presence or absence and severity of other congenital heart defects (CHD), number of procedures, lead type (unipolar or bipolar, steroid-eluting or not, and active or passive fixation), implantation procedure (i.e. absorbable ligature of the leads to subcutaneous tissues only or the addiction of the atrial loop), and, as a continuous variable, year of implantation. When venous occlusion was found at Doppler-echo examination, it was compared with other variables (age at implantation, sex, arrhythmia requiring PM implantation, CHD, number of procedures performed, lead dimension, lead polarity, and FU duration) with two-sample Wilcoxon Rank-sum (Mann Whitney) test. Results Population Between 1990 and the end of 2005, 117 children (47 girls) underwent the implantation of a VVI/R PM with an endocardial lead at years of age (median 5 years, range 2 days to 15 years). In 25 patients (21%), a previous epicardial pacing system failed. Fifty-four patients (46%) had no structural heart disease and 63 patients (54%) had CHD (Table 1). No residual shunts have been documented in patients who underwent heart surgery. For purposes of statistical analysis, patients were divided into three groups based on presence and severity of CHD: Group 1, composed of patients without CHD (n ¼ 54 patients) and patients with mild CHD not requiring surgical treatment (n ¼ 7 patients); Group 2, including 40 patients with simple CHD (e.g. isolated atrial, ventricular and AV septal defect, pulmonary stenosis, and tetralogy of Fallot); and Group 3, including all others with complex CHD (n ¼ 16 patients). Indications The main arrhythmia requiring PM implantation was complete AVB in 105 patients (89%), postoperative in 37 patients, and sinus node dysfunction (SND) in 12 patients. Table 1 CHD Main congenital heart defects of the patients Ventricular septal defect 22 Tetralogy of Fallot 11 AV septal defect 10 TGA status/post-mustard 4 CCTGA 3 Pulmonary valve stenosis 2 Aortic valve stenosis 2 Ostium secundum atrial septal defect 2 TGA status/post-arterial switch 2 Cardiomyopathy 2 Totally anomalous pulmonary venous return 1 Patent ductus arteriosus 1 Mitral valve disease 1 Number of patients AV, atrioventricular; CCTGA, congenitally corrected TGA fs, L, Lg without other associated anomalies; CHD, other congenital heart defects; TGA, transposition of the great arteries fs, D, Dg.

3 896 M.S. Silvetti et al. In seven patients, SND was a complication of previous heart surgery (four after Mustard palliation for transposition of the great arteries fs, D, Dg with median age at implantation 6 years) and five patients had SND in the absence of structural abnormalities. Causes of ventricular pacing in patients with SND were: unsuccessful positioning of an endocardial atrial lead (four patients, three of whom after Mustard palliation or atrial septation), association with II degree AVB (three patients, one with Mustard palliation), and use of backup ventricular pacing during symptomatic sinus pauses or intermittent severe bradycardia (five patients). In 2 of this 12 patients, a previous atrial epicardial pacing system failed. Leads implanted The leads implanted (Table 2) were: unipolar (n ¼ 78, 67%) and bipolar (n ¼ 39, 33%); tined (n ¼ 110, 94%) and screw-in (n ¼ 7, 6%); and steroid-eluting (n ¼ 89, 76%) and non-steroid-eluting (n ¼ 28, 24%). The leads were fixed to subcutaneous tissue by absorbable suture in all patients, but in the last 6 years, in 17 patients (12%), an atrial loop was also created. Follow-up Follow-up duration (until end of 2006) was (median 8, range 1 15) years. Three patients died for causes unrelated to pacing, 15 were lost to FU, 4 underwent heart transplantation, and 2 had the leads removed and substituted with epicardial leads at the time of elective cardiac surgery for CHD. An upgrading procedure with the addition of an atrial lead in the same subclavian vein (VVIR DDD) was performed in 16 patients, and some of these cases have already been published. 7 Complications There were 22 system failures (19%), due to lead malfunction (20 patients) and system erosion/infection (2 patients) (Figure 1), all in patients with AVB (Table 3). Only one of these failures was early, the erosion and infection of the PM pocket occurring in a neonate 1 month after implantation on the second day of life. In this patient, the pacing system was switched to an epicardial pacing system after lead extraction. The other patient, who developed a Table 2 Leads implanted pocket erosion 14 years after the first implantation, was treated with pocket revision and PM replacement. In the 20 patients with lead malfunction, the leads were: abandoned and a new transvenous lead was implanted in 14 patients; 7,8 replaced by epicardial leads after surgical extraction (two patients); replaced by endocardial leads after transvenous extraction by means of mechanical traction in our institution (two patients); and extracted in other institutions in two patients. The log-rank test for equality of survivor functions for the first endocardial pacing systems implanted showed no significant association with age, sex, arrhythmia requiring PM implantation, presence of CHD or not, lead polarity, steroid-eluting leads or not, type of endocardial lead fixation, procedure used for lead fixation to subcutaneous tissues with or without atrial loop. However, the 20 cases of lead malfunction occurred only in the group with no atrial loop, but the FU was significantly longer in these patients ( for patient with no loop vs years, for patients with atrial loop, P, 0.05) (Figure 2). There were no differences in the statistical analysis if the outcome was limited only to failures due to lead malfunction or if systems infection/erosion was included as well. The number of procedures performed (generator replacements, lead repositioning for early dislodgement) had no negative impact on the longevity of the pacing system. Complications not requiring pacing system replacement included 11 early events [haemothorax (2.5%), lead dislodgement (5%), infections treated only with antibiotics (nearly 2%)] and one late event (initial pocket inflammation/erosion treated by PM replacement with a smaller generator at battery depletion) (Table 3). Doppler assessment in 56 patients, performed years after the implantation procedure, documented occlusion of the subclavian vein in three cases. Although patients who developed venous occlusion, in comparison with patients who did not, showed a trend towards older age at implantation ( vs years), the lead size ( vs Fr) and the time between implantation and Doppler-echo assessment of the sublavian vein ( years vs ) were not statistically significantly different. In fact, we found no parameter that was associated with the occurrence of venous occlusion. AV valve damage was not documented during the study period. Lead n French Polarity Steroid Fixation Failures Telectronics Unipolar No Passive 2 Vitatron ISP19F Unipolar No Passive 3 Medtronic Bipolar No Passive 1 Medtronic Bipolar No Passive 0 Medtronic Bipolar Yes Passive 5 Medtronic4004M Bipolar Yes Passive 4 Medtronic Unipolar Yes Passive 5 Medtronic Unipolar Yes Passive 0 Medtronic Bipolar Yes Active 0 Osypka KY5 UP Unipolar No Active 0 Osypka KY5 BP Bipolar No Active 0

4 Outcome of VVI/R pacemaker in children 897 Figure 1 children. Table 3 Discussion Total survival of VVIR pacing systems implanted in Complications Early (number of pts, %) Late (n,%) Abnormal threshold 15 (13%) increase Insulation break 1 (0.9%) Exit block 3 (2.5%) Lead fracture 1 (0.9%) Pocket erosion 1 (0.9%) 1 (0.9%) þ1 (0.9%) Haemothorax 3 (2.5%) Lead dislodgement 6 (5%) PM pocket infection 2 (1.8%) treated with antibiotics Total 12 (10%) 22 (19%) Total pts requiring system replacement 1 21 Pts, patients. The numbers italicized are patients who underwent a new system implantation due to complications (see text for further details). The implantation of a permanent PM in children is a procedure with a generally favorable outcome, but children are more prone to complications because of the body growth, their more active lifestyle, their higher frequency of traumatic events, and localized or systemic infections that could affect the pacing system, with complication rates around 10 30%. 8,10,12 16 However, in adults, a rate of 14% of complication has been described. 21 Pacing leads remain the weakest link of the pacing system, 22 particularly in a growing patient. Concerns have been raised about long-term efficacy and safety of endocardial leads in children, particularly regarding lead abandonment, 10 valvular integrity, and venous obstruction. 4,5 For these reasons, most paediatric institutions still mainly use epicardial pacing for young children. In general, long-term complications are lead-related. As a rule, the fewer leads that are implanted, the fewer complications that will occur. 16 Figure 2 Lead survival according to the implantation procedure used. 1: Leads fixed to subcutaneous tissue by absorbable suture alone; 2: leads fixed by absorbable suture and with an atrial loop. P ¼ not significant (see text for further details). In children with AVB and normal ventricular function, it has been demonstrated that the increase in heart rate obtained with VVI/R pacing is an adequate response to their physiological needs 1 3 and that DDD pacing is not clearly superior to VVI/R pacing. 2 For this reason, it seems prudent to implant a VVI/R PM in children who need a transvenous pacing system. DDD pacing with transvenous leads is technically feasible in small children, but may not be needed. 16 Our data demonstrate that VVIR endocardial pacing in these particular patients (young children with median age at implantation 5 years) has good results, although the rate of complications requiring reoperation was nearly 20% during a median FU of 8 years. Kaplan Meier analysis indicated 75% survival of the pacing system at 12 years, decreasing to 50% at 15 years (Figure 1). Most of the complications were due to the lead problems, but we could not identify specific risk factors for these failures. The addition of the atrial loop to the absorbable ligature seems promising (Figure 2) probably for the prevention of the outgrowth of leads, as patients with this procedure have not experienced any lead failure, but their FU is short. We believe that this complication rate is high but acceptable in these patients: pacing prevents the risk of syncope, sudden death, heart failure and allows adequate heart rates and good quality of life in absence of other associated CHD. 2 A comparison of VVI/R pacing with other pacing modes has not been performed in our population, as AAI/R and DDD/R PM have been generally implanted in older patients (median age 11 and 12 years, respectively). 16 Atrial pacing is indicated in SND 7,16 to preserve AV synchrony and to avoid the deleterious effect of right ventricular pacing on the left ventricle. Sometimes, VVI/R PM with transvenous leads may be implanted also in children with SND, especially in the rare conditions of SND in absence of structural abnormalities. Moreover, in children with small atria leaving lead slack or a complete loop in the atrium might increase the risk of lead dislodgement. 10,16 However, AAI/R pacing is still the preferred route in SND and epicardial pacing is an option: even though AAI/R epicardial pacing may be more aggressive, the potential benefit of

5 898 M.S. Silvetti et al. atrial pacing is important. The risk of failure of epicardial leads might be higher, 13,14,16,23 although it has been shown that epicardial and endocardial leads have comparable performances in the paediatric population. 15,17,18 Nearly 14% of these patients underwent successful upgrading procedures during adolescence or young adulthood. Indications (ventricular dysfunction, syncope, PM syndrome, and age), timing, and results of the upgrading procedures have been already published. 7 Anyhow, in absence of symptoms, we generally defer the upgrade, as the procedure might be technically challenging and it has been suggested that the establishment of AV synchrony in the young asymptomatic patient with VVIR pacing and normal ventricular function may not be warranted. 3 This cohort of patients is difficult comparable with other published series, as there are no recent studies directly evaluating VVI/R PM in children. For example, the recent paper of Fortescue et al. 18 described patients with thin transvenous leads compared with patients with epicardial pacing. Ventricular leads were 114, and the median age at implantation was higher (16.7 years) than in our study. Thus, it is not surprising their low number of ventricular lead failures during the study period: seven leads, mostly due to fracture or insulation break. Other recent papers, instead, focused on specific group of patients, as infants. 12,15,23 Infection accounted for nearly 2% in our series and not all infections required system revision (Table 3). Previous reports have indicated an incidence of pacing system infection ranging from 2.3% (superficial infection) to 4.9% (deep PM infection) 24 to 5.5%. 25 These reports also indicate that trisomy 21 and the need for PM revision 24 were significant risk factors for the development of infection after PM implantation. Conversely, the number of previous procedures performed had no negative impact on system infections and pacing system duration in our patients, as previously described in our total population. 16 Implant-related complications (haemothorax and acute lead dislodgement) account for 7.5% of the total population. This is similar to our total population with transvenous leads, 16 where the frequency of implant-related complications accounts for 8.5%, and that of infectious complications for 2%. The assessment of venous patency was generally performed in the last 3 years of FU, in nearly half of the patients. Clinically silent subclavian vein occlusion was documented in 5% of patients. In one of these three patients, a new DDD PM was implanted via the contralateral vein with the abandonment of the old lead, 7 whereas in another the upgrading was not performed. 7 In the last patient, aged 15 years, a new pacing system has not been planned yet. Although patients who developed venous occlusion had a trend towards larger lead size and longer time of FU to Doppler assessment, none of the parameters evaluated had a significant association with the occurrence of venous occlusion. Moreover, the patients with venous occlusion tended to be older at implantation, although we would expect this complication to occur with younger patients and larger ratio of lead/vein dimension. However, our search might have been hampered by the small number of patients with venous occlusion. Recently, other authors were unable to identify specific risk factors (age, size, growth, and lead factors) for venous occlusion 26 as well. Previous published data seem to document an incidence of 10 25%, 6,12,26 although the variability is large, ranging from 0% 23,27 to 100%. 4 Although used for venous patency evaluation, 5 8,23 Doppler-echo is not as reliable for evaluation of innominate/superior vena cava occlusion as venography or intravascular ultrasound. 28 In our experience, 7,8,23 Doppler-echo examination of the subclavian vein before upgrading procedures accurately identified subclavian patency/occlusion subsequently confirmed by wire/ lead insertion and/or venography during upgrading procedures, without false positive or negative results. Thus, although our approach may be less sensitive, and our finding of venous occlusion might be underestimated, we think that an invasive procedure is not indicated simply for FU examination of children. 6,23 In this selected population, as well as in our total 16 or in other study groups, 8,18,23 AV valve damage was not documented. Thus, we can exclude significant AV valve regurgitation/stenosis from the list of likely complications of transvenous leads in use nowadays. Limitations The study has some limitations: it is a retrospective analysis, based on a single-centre experience and the number of patients is relatively small. Doppler-echo evaluation of the subclavian vein, performed in a subgroup of our patients, is less sensitive than angiography or intravascular ultrasound for vessel patency determination. Conclusions The use of VVI/R PM in children has good results, but during the FU, a significant number of failures occurs, mainly related to the lead. The frequency of infection and that of subclavian vein occlusion is low; AV valve damage has not been demonstrated. As for all paediatric studies, larger populations and longer FU might demonstrate a higher frequency of complication. These data might be a base for comparison with groups with alternative sites of pacing, whose indications and numbers will probably increase in the future. Acknowledgements The authors thank Prof. Stephen P. Sanders, MD, for his help in reviewing the manuscript. Conflict of interest: none declared. References 1. Ragonese P, Guccione P, Drago F, Turchetta A, Calzolari A, Formigari R. Efficacy and safety of ventricular rate responsive pacing in children with complete atrioventricular block. Pacing Clin Electrophysiol 1994; 17: Friedman RA, Fenrich AL, Kertesz NJ. Congenital complete atrioventricular block. Pacing Clin Electrophysiol 2001;24: Horenstein MS, Karpawich PP, Tantengco MV. Single versus dual chamber pacing in the young: noninvasive comparative evaluation of cardiac function. Pacing Clin Electrophysiol 2003;26: Campbell RM, Raviele AA, Hulse EJ, Auld DO, McRae GJ, Tam VK et al. Experience with a low profile bipolar, active fixation pacing lead in pediatric patients. Pacing Clin Electrophysiol 1999;22:

6 Outcome of VVI/R pacemaker in children Figa FH, Mccrindle BW, Bigras BJ, Hamilton RM, Gow RM. Risk factors for venous obstruction in children with transvenous pacing leads. Pacing Clin Electrophysiol 1997;20: Stojanov P, Vranes M, Velimirovic D, Zivkovic M, Kocica MJ, Davidovic L et al. Prevalence of venous obstruction in permanent endovenous pacing in newborn and infants: follow-up study. Pacing Clin Electrophysiol 2005;28: Silvetti MS, Drago F. Upgrade of single chamber pacemakers with transvenous leads to dual chamber pacemakers in pediatric and young adults patients. Pacing Clin Electrophysiol 2004;27: Silvetti MS, Drago F. Upgrading of VVIR pacemakers with nonfunctional endocardial ventricular leads to VDD pacemakers in adolescents. Pacing Clin Electrophysiol 2006;29: Rosenthal E, Bostock J, Qureshi SA, Baker EJ, Tynan M. Single pass VDD pacing in children and adolescents. Pacing Clin Electrophysiol 1997;20: Lau YR, Gillette PC, Buckles DS, Zeigler VL. Actuarial survival of transvenous pacing leads in a pediatric population. Pacing Clin Electrophysiol 1993;16: Molina JE, Dunnigan AC, Crosson JE. Implantation of transvenous pacemakers in infants and small children. Ann Thorac Surg 1995;59: Kammeraad JAE, Rosenthal E, Bostock J, Rogers J, Sreeram N. Endocardial pacemaker implantation in infants weighing 10 kilograms. Pacing Clin Electrophysiol 2004;27: Fortescue EB, Berul CI, Cecchin F, Walsh EP, Triedman JK, Alexander ME. Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart disease. Heart Rhythm 2004;1: Sachweh JS, Vazquez-Jimenez JF, Schöndube FA, Daebritz SH, Dorge H, Muhler EG et al. Twenty years experience with pediatric pacing: epicardial and transvenous stimulation. Eur J Cardiothorac Surg 2000;17: Udink ten Cate F, Breur J, Boramanand N, Crosson J, Friedman A, Brenner J et al. Endocardial and epicardial steroid lead pacing in the neonatal and paediatric age group. Heart 2002;88: Silvetti MS, Drago F, Grutter G, De Santis A, Di Ciommo V, Ravà L. Twenty years of cardiac pacing in paediatric age: 515 pacemakers and 480 leads implanted in 292 patients. Europace 2006;8: Beaufort-Krol GCM, Mulder H, Nagelkerke D, Waterbolk TW, Bink-Boelkens MTE. Comparison of longevity, pacing, and sensing characteristics of steroid-eluting epicardial versus conventional endocardial pacing leads in children. J Thorac Cardiovasc Surg 1999;117: Fortescue EB, Berul CI, Cecchin F, Walsh EP, Triedman JK, Alexander ME. Comparison of modern steroid-eluting epicardial and thin transvenous pacemaker leads in pediatric and congenital heart disease patients. J Interv Card Electrophysiol 2005;14: Stojanov P, Velimirovic D, Hrnjak V, Pavlovic SU, Zivkovic M, Djordjevic Z. Absorbable suture technique: solution to the growth problem in pediatric pacing with endocardial leads. Pacing Clin Electrophysiol 1998;21: Rosenthal E, Bostock J. Use of an atrial loop to extend the duration of endocardial pacing in a neonate. Pacing Clin Electrophysiol 1997;20: Kiviniemi MS, Pirnes MA, Eranen HJK, Kettunen RVJ, Hartikainen JEK. Complications related to permanent pacemaker therapy. Pacing Clin Electrophysiol 1999;22: De Voogt WG. Pacemaker leads: performance progress. Am J Cardiol 1999;83:187D 91D. 23. Silvetti MS, Drago F, De Santis A, Grutter G, Ravà L, Monti L et al. Singlecentre experience on endocardial and epicardial pacemaker system function in neonates and infants. Europace 2007;9: Cohen MI, Bush DM, Gaynor JW, Vetter VL, Thanel RE, Rhodes LA. Pediatric pacemaker infections: twenty years of experience. J Thorac Cardiovasc Surg 2002;124: Klug D, Vaksmann G, Jarwé M, Wallet F, Francart C, Kacet S et al. Pacemaker lead infection in young patients. Pacing Clin Electrophysiol 2003; 26: Bar-Cohen Y, Berul CI, Alexander ME, Fortescue EB, Walsh EP, Triedman JK et al. Age, size, and lead factors alone do not predict venous obstruction in children and young adults with transvenous lead systems. J Cardiovasc Electrophysiol 2006;17: Gillette PC, Zeigler V, Bradham GB, Kinsella P. Pediatric transvenous pacing: a concern for venous thrombosis? Pacing Clin Electrophysiol 1988;11: Chintala K, Forbes TJ, Karpawich PP. Effectiveness of transvenous pacemaker leads placed through intravascular stents in patients with congenital heart disease. Am J Cardiol 2005;95:424 7.

Epicardial vs Endocardia Pacing System. Department of Pediatrics, Sejong General Hospital, Bucheon-si, Gyeonggi-do, Republic of Korea

Epicardial vs Endocardia Pacing System. Department of Pediatrics, Sejong General Hospital, Bucheon-si, Gyeonggi-do, Republic of Korea Epicardial vs Endocardia Pacing System Lee Sang-Yun MD, PhD Department of Pediatrics, Sejong General Hospital, Bucheon-si, Gyeonggi-do, Republic of Korea The number of pediatric pacemakers implanted is

More information

Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict

Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict DOI 10.1007/s12471-011-0158-5 ORIGINAL ARTICLE Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict M. C. Post & W. Budts & A. Van de Bruaene & R. Willems

More information

Pediatric Pacemaker Implantation Endocardial or Epicardial

Pediatric Pacemaker Implantation Endocardial or Epicardial Pediatric Pacemaker Implantation Endocardial or Epicardial HAITHAM BADRAN, MD, FEHRA CONSULTANT OF INTERVENTIONAL CARDIOLOGY CONSULTANT OF CARDIAC PACING AND ELECTROPHYSIOLOGY LECTURER OF CARDIOLOGY AIN

More information

T he introduction of steroid eluting pacing leads in the paediatric

T he introduction of steroid eluting pacing leads in the paediatric 392 CONGENITAL HEART DISEASE Endocardial and epicardial steroid lead pacing in the neonatal and paediatric age group F Udink ten Cate, J Breur, N Boramanand, J Crosson, A Friedman, J Brenner, E Meijboom,

More information

Ventricular Capture Management in pediatric pacing: efficacy and safety

Ventricular Capture Management in pediatric pacing: efficacy and safety Ventricular Capture Management in pediatric pacing: efficacy and safety Massimo S. Silvetti, Antonella De Santis, Giorgia Grutter, Vincenzo Di Ciommo*, Fabrizio Drago Cardiac Arrhythmias Service, *Epidemiology

More information

Transvenous Pacemaker Implantation 22 years after the Mustard Procedure

Transvenous Pacemaker Implantation 22 years after the Mustard Procedure Case Report Transvenous Pacemaker Implantation 22 years after the Mustard Procedure Masato Sakamoto MD, Yoshie Ochiai MD, Yutaka Imoto MD, Akira Sese MD, Mamie Watanabe MD, Kunitaka Joo MD Department of

More information

Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children

Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children ERALD A. SERWER, M.D., MACDONALD DICK 11, M.D., KAREN UZARK, R.N., Ph.D., WILLIAM A. SCOTT, M.D., and EDWARD L.

More information

Implantation of Transvenous Pacemakers in Infants and Small Children

Implantation of Transvenous Pacemakers in Infants and Small Children Implantation of Transvenous Pacemakers in Infants and Small Children J. Ernesto Molina, MD, PhD, Ann C. Dunnigan, MD, and Jane E. Crosson, MD Divisions of Cardiovascular and Thoracic Surgery and Pediatric

More information

Equivalent Performance of Epicardial Versus Endocardial Permanent Pacing in Children: A Single Institution and Manufacturer Experience

Equivalent Performance of Epicardial Versus Endocardial Permanent Pacing in Children: A Single Institution and Manufacturer Experience Equivalent Performance of Epicardial Versus Endocardial Permanent Pacing in Children: A Single Institution and Manufacturer Experience Jonah Odim, MD, PhD, Bjoern Suckow, MD, Babak Saedi, MD, Hillel Laks,

More information

Automatic Atrial Threshold Measurement and Adjustment in Pediatric Patients

Automatic Atrial Threshold Measurement and Adjustment in Pediatric Patients Automatic Atrial Threshold Measurement and Adjustment in Pediatric Patients ANITA HIIPPALA, M.D.,* GERALD A. SERWER, M.D., EVA CLAUSSON, M.Sc., LYNN DAVENPORT, M.S., TRINA BRAND, PH.D., and JUHA-MATTI

More information

Predictors affecting durability of epicardial pacemaker leads in pediatric patients

Predictors affecting durability of epicardial pacemaker leads in pediatric patients Murayama et al Surgery for Congenital Heart Disease Predictors affecting durability of epicardial pacemaker leads in pediatric patients a a a b Hiroomi Murayama, MD, Masanobu Maeda, MD, Hajime Sakurai,

More information

A Case of Transvenous Pacemaker Implantation in a 10-year-old Patient

A Case of Transvenous Pacemaker Implantation in a 10-year-old Patient J Rural Med 2014; 9(1): 32 36 Case report A Case of Transvenous Pacemaker Implantation in a 10-year-old Patient Jiajia Liu and Yasuyuki Shimada Department of Cardiovascular Surgery, Yuri-Kumiai General

More information

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt Leadless Pacing Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt The weakest link in Pacemaker system the lead. The more the leads the more the complications Dislodgement Fracture Insulation

More information

Keywords: Pacemaker, transvenous pacemaker, thrombosis, intracardiac thrombosis, transesophageal echocardiography, paediatrics

Keywords: Pacemaker, transvenous pacemaker, thrombosis, intracardiac thrombosis, transesophageal echocardiography, paediatrics IMAGES in PAEDIATRIC CARDIOLOGY Wittekind SG, 1 Salerno JC, 2 Rubio AE. 2 Pacemaker-associated cyanosis in an adolescent: The answer hiding behind 1 Pediatric Residency Program, University of Washington,

More information

Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases

Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases Case Report Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases Koji Tsutsumi MD, Tatsuru Niibori MD, Keiichiro Katsumoto MD Department

More information

Long-Term Outcomes of Cardiac Pacing in Adults With Congenital Heart Disease

Long-Term Outcomes of Cardiac Pacing in Adults With Congenital Heart Disease Journal of the American College of Cardiology Vol. 43, No. 10, 2004 2004 by the American College of Cardiology Foundation ISSN 0735-1097/04/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2003.12.044

More information

Antitachycardia Pacemakers in Congenital Heart Disease

Antitachycardia Pacemakers in Congenital Heart Disease 180 Antitachycardia Pacemakers in Congenital Heart Disease Anna N. Kamp, MD, MPH,* Martin J. LaPage, MD, MS, Gerald A. Serwer, MD, Macdonald Dick II, MD, and David J. Bradley, MD *Cardiology, The Heart

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

Permanent Pacemaker Implantation via the Iliac Vein: An Alternative in 4 Cases with Contraindications to the Pectoral Approach

Permanent Pacemaker Implantation via the Iliac Vein: An Alternative in 4 Cases with Contraindications to the Pectoral Approach Case Report Permanent Pacemaker Implantation via the Iliac Vein: An Alternative in 4 Cases with Contraindications to the Pectoral Approach Koji Tsutsumi MD, Kenichi Hashizume MD, Naritaka Kimura MD, Shinichi

More information

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs).

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs). Management of Children with Cardiac Devices Guideline originally developed by Leann Miles, APRN; Lindsey Pumphrey, RN; Srikant Das, MD, and the ANGELS Team. Last reviewed by Lindsey Pumphrey, RN, Srikant

More information

New generations pacemakers and ICDs: an update

New generations pacemakers and ICDs: an update Advances in Cardiac Arrhythmias and Great Innovations in Cardiology XXVII Giornate Cardiologiche Torinesi New generations pacemakers and ICDs: an update Prof. Fiorenzo Gaita, MD Division of Cardiology

More information

Index. cardiology.theclinics.com. Note: Page numbers of article titles are in boldface type.

Index. cardiology.theclinics.com. Note: Page numbers of article titles are in boldface type. Index Note: Page numbers of article titles are in boldface type. A ACHD. See Adult congenital heart disease (ACHD) Adult congenital heart disease (ACHD), 503 512 across life span prevalence of, 504 506

More information

Trans-Fontan baffle placement of an endocardial systemic ventricular pacing lead

Trans-Fontan baffle placement of an endocardial systemic ventricular pacing lead Trans-Fontan baffle placement of an endocardial systemic ventricular pacing lead The Harvard community has made this article openly available. Please share how this access benefits you. Your story matters.

More information

Automatic assessment of atrial pacing threshold in current medical practice

Automatic assessment of atrial pacing threshold in current medical practice Europace (212) 14, 1615 1619 doi:1.193/europace/eus76 CLINICAL RESEARCH Leads and lead extraction Automatic assessment of atrial pacing threshold in current medical practice Jean Luc Rey 1, Serge Quenum

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL

More information

When to implant an ICD in systemic right ventricle?

When to implant an ICD in systemic right ventricle? When to implant an ICD in systemic right ventricle? Département de rythmologie et de stimulation cardiaque Nicolas Combes n.combes@clinique-pasteur.com Pôle de cardiologie pédiatrique et congénitale Risk

More information

Pacing Codes and Modes Concepts

Pacing Codes and Modes Concepts Pacing Codes and Modes Concepts Pacing codes and modes concepts Objectives Upon completion of this program the participant will be able to: State what the first four positions of the NBG code represent.

More information

Implantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect

Implantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect The Ochsner Journal 10:27 31, 2010 f Academic Division of Ochsner Clinic Foundation Implantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect Anas Bitar, MD, Maria Malaya

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing

More information

Epicardial pacing: a single-centre study on 321 leads in 138 patients

Epicardial pacing: a single-centre study on 321 leads in 138 patients 343 Epicardial pacing: a single-centre study on 321 leads in 138 patients Bavo ECTOR 1, Rik WILLEMS 1, Hein HEIDBÜCHEL 1, Marc GEWILLIG 2, Luc MERTENS 2, Bart MEYNS 3, Willem DAENEN 3, Hugo ECTOR 1 1 Department

More information

Leadless pacing: Technical challenges and concerns. Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital

Leadless pacing: Technical challenges and concerns. Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital Leadless pacing: Technical challenges and concerns Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital Modest consultancy fees/honoraria from Bayer Boehringer Ingelheim

More information

EPICARDIAL PACEMAKER IMPLANTATION AND FOLLOW-UP IN PATIENTS WITH A SINGLE VENTRICLE AFTER THE FONTAN OPERATION

EPICARDIAL PACEMAKER IMPLANTATION AND FOLLOW-UP IN PATIENTS WITH A SINGLE VENTRICLE AFTER THE FONTAN OPERATION EPICARDIAL PACEMAKER IMPLANTATION AND FOLLOW-UP IN PATIENTS WITH A SINGLE VENTRICLE AFTER THE FONTAN OPERATION Mitchell I. Cohen, MD a Victoria L. Vetter, MD a Gil Wernovsky, MD a David M. Bush, MD, PhD

More information

SUPPLEMENTAL MATERIAL

SUPPLEMENTAL MATERIAL SUPPLEMENTAL MATERIAL Table S1. Patient Selection Codes, CIED Generator Procedures Code Type Code Description ICD9 Proc 00.51 Implantation of cardiac resynchronization defibrillator, total system [CRT-D]

More information

Pediatric Echocardiography Examination Content Outline

Pediatric Echocardiography Examination Content Outline Pediatric Echocardiography Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 Anatomy and Physiology Normal Anatomy and Physiology 10% 2 Abnormal Pathology and Pathophysiology

More information

Heart and Lungs. LUNG Coronal section demonstrates relationship of pulmonary parenchyma to heart and chest wall.

Heart and Lungs. LUNG Coronal section demonstrates relationship of pulmonary parenchyma to heart and chest wall. Heart and Lungs Normal Sonographic Anatomy THORAX Axial and coronal sections demonstrate integrity of thorax, fetal breathing movements, and overall size and shape. LUNG Coronal section demonstrates relationship

More information

Surgical Approaches to Epicardial Pacemaker Placement: Does Pocket Location Affect Lead Survival?

Surgical Approaches to Epicardial Pacemaker Placement: Does Pocket Location Affect Lead Survival? Pediatr Cardiol (2010) 31:1016 1024 DOI 10.1007/s00246-010-9754-1 ORIGINAL ARTICLE Surgical Approaches to Epicardial Pacemaker Placement: Does Pocket Location Affect Lead Survival? Brian J. Lichtenstein

More information

Reduction of Mitral Regurgitation by Endocardial Right Ventricular Bifocal Pacing in Cases of Dilated Cardiomyopathy

Reduction of Mitral Regurgitation by Endocardial Right Ventricular Bifocal Pacing in Cases of Dilated Cardiomyopathy June 2000 233 Reduction of Mitral Regurgitation by Endocardial Right Ventricular Bifocal Pacing in Cases of Dilated Cardiomyopathy J. C. PACHON M., R. N. ALBORNOZ, E. I. PACHON M., V. M. GIMENES, J. PACHON

More information

Different indications for pacemaker implantation are the following:

Different indications for pacemaker implantation are the following: Patient Resources: ICD/Pacemaker Overview ICD/Pacemaker Overview What is a pacemaker? A pacemaker is a device that uses low energy electrical pulses to prompt the heart to beat whenever a pause in the

More information

Adult Echocardiography Examination Content Outline

Adult Echocardiography Examination Content Outline Adult Echocardiography Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 2 3 4 5 Anatomy and Physiology Pathology Clinical Care and Safety Measurement Techniques, Maneuvers,

More information

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie 268 Case Report Cardiac Resynchronization Therapy in a Patient with Persistent Left Superior Vena Cava Draining into the Coronary Sinus and Absent Innominate Vein: A Case Report and Review of Literature

More information

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program:

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: Notification and Prior Authorization Program: Electrophysiology Implant Classification Table The following chart contains the codes that require notification or prior authorization as part of UnitedHealthcare

More information

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program Electrophysiology Implant Classification Table The table below contains the codes that apply to our UnitedHealthcare Medicare Advantage cardiology prior Description Includes Generator Placement Includes

More information

Pacing troubleshooting. NASPE training Lancashire & South Cumbria Cardiac Network

Pacing troubleshooting. NASPE training Lancashire & South Cumbria Cardiac Network Pacing troubleshooting NASPE training Lancashire & South Cumbria Cardiac Network Pacing stimulus present failure to capture Lead dislodgement Early unstable position Late twiddlers syndrome Abnormal ECG

More information

2:1 Block with Wenckebach Mechanism in Children Due to Different Etiologies F Laloğlu 1, N Ceviz 1, H Keskin 2, H Olgun 1 ABSTRACT

2:1 Block with Wenckebach Mechanism in Children Due to Different Etiologies F Laloğlu 1, N Ceviz 1, H Keskin 2, H Olgun 1 ABSTRACT 2:1 Block with Wenckebach Mechanism in Children Due to Different Etiologies F Laloğlu 1, N Ceviz 1, H Keskin 2, H Olgun 1 ABSTRACT Objective: In children 2:1 Atrioventricular Block (AVB) with Wenckebach

More information

Patient Resources: Arrhythmias and Congenital Heart Disease

Patient Resources: Arrhythmias and Congenital Heart Disease Patient Resources: Arrhythmias and Congenital Heart Disease Overview Arrhythmias (abnormal heart rhythms) can develop in patients with congenital heart disease (CHD) due to thickening/weakening of their

More information

Pacing in patients with congenital heart disease: part 1

Pacing in patients with congenital heart disease: part 1 Pacing in patients with congenital heart disease: part 1 September 2013 Br J Cardiol 2013;20:117 20 doi: 10.5837/bjc/2013.028 Authors: Khaled Albouaini, Archana Rao, David Ramsdale View details Only a

More information

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΖΩΡΤΖ ΔΑΔΟΥΣ ΕΠΙΚΟΥΡΟΣ ΚΑΘΗΓΗΤΗΣ Α.Π.Θ. ΜΑΡΙΑ ΚΑΡΑΛΙΟΛΙΟΥ ΕΙΔΙΚΕΥΟΜΕΝΗ ΙΑΤΡΟΣ ΚΑΡΔΙΟΛΟΓΟΣ Β ΚΑΡΔΙΟΛΟΓΙΚΗ ΚΛΙΝΙΚΗ Α.Π.Θ. ΙΠΠΟΚΡΑΤΕΙΟ Γ.Ν.Θ. 5 Why a Pacemaker is Implanted

More information

Outcomes of Defibrillator Lead Implants Performed by High Volume Operators vs. Low Volume Operators:

Outcomes of Defibrillator Lead Implants Performed by High Volume Operators vs. Low Volume Operators: Outcomes of Defibrillator Lead Implants Performed by High Volume Operators vs. Low Volume Operators: Results from the Pacemaker and Implantable Defibrillator Leads Survival Study ( PAIDLESS ) Partially

More information

Technical option of surgical approach for trouble-shooting

Technical option of surgical approach for trouble-shooting JHRS Corner Device and lead trouble-shooting - standard strategy and technical option - Technical option of surgical approach for trouble-shooting Katsuhiko IMAI Department of Cardiovascular surgery, Hiroshima

More information

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine

More information

How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France

How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France Presenter Disclosure Information Christophe Leclercq, MD,

More information

Pediatric Post-Operative Atrio-Ventricular Block Meets the Affordable Care Act: A New Strategy for Management

Pediatric Post-Operative Atrio-Ventricular Block Meets the Affordable Care Act: A New Strategy for Management Open Journal of Pediatrics, 2017, 7, 118-127 http://www.scirp.org/journal/ojped ISSN Online: 2160-8776 ISSN Print: 2160-8741 Pediatric Post-Operative Atrio-Ventricular Block Meets the Affordable Care Act:

More information

Leadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece

Leadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece Leadless pacing P.E. Vardas Professor of Cardiology University of Crete, Greece DISCLOSURES Small teaching fees from: Bayer/J&J, Boeringer, Menarini Small consulting fees from: Medtronic 1957: First External

More information

NOVEL DEVICE TECHNOLOGIES

NOVEL DEVICE TECHNOLOGIES NOVEL DEVICE TECHNOLOGIES Leadless Pacemakers and Subcutaneous ICDs Do the Benefits Outweigh MRI Incompatibility? Disclosures None Background PPMs and ICDs are very effective therapy for treating bradyarrhythmias

More information

Practice Questions.

Practice Questions. IBHRE Prep Practice Questions Question 1 The relative refractory yperiod of the ventricular myocardium corresponds to which of the following phases of the action potential? A. (0) B. (1) C. (2) D. (3)

More information

Index. Note: Page numbers of article titles are in boldface type.

Index. Note: Page numbers of article titles are in boldface type. Index Note: Page numbers of article titles are in boldface type. A Acute coronary syndrome(s), anticoagulant therapy in, 706, 707 antiplatelet therapy in, 702 ß-blockers in, 703 cardiac biomarkers in,

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment Karen L. Booth, MD, Lucile Packard Children s Hospital Arrhythmias are common after congenital heart surgery [1]. Postoperative electrolyte

More information

Congenitally Corrected Transposition of the Great Arteries (cctga or l-loop TGA)

Congenitally Corrected Transposition of the Great Arteries (cctga or l-loop TGA) Congenitally Corrected Transposition of the Great Arteries (cctga or l-loop TGA) Mary Rummell, MN, RN, CPNP, CNS Clinical Nurse Specialist, Pediatric Cardiology/Cardiac Surgery Doernbecher Children s Hospital,

More information

2017 HRS Expert Consensus Statement on Cardiovascular Implantable Electronic Device Lead Management and Extraction

2017 HRS Expert Consensus Statement on Cardiovascular Implantable Electronic Device Lead Management and Extraction Summary of Expert Consensus Statement for CLINICIANS 2017 HRS Expert Consensus Statement on Cardiovascular Implantable Electronic Device Lead Management and Extraction This is a summary of the Heart Rhythm

More information

Slide 1. Slide 2. Slide 3 CONGENITAL HEART DISEASE. Papworth Hospital NHS Trust INTRODUCTION. Jakub Kadlec/Catherine Sudarshan INTRODUCTION

Slide 1. Slide 2. Slide 3 CONGENITAL HEART DISEASE. Papworth Hospital NHS Trust INTRODUCTION. Jakub Kadlec/Catherine Sudarshan INTRODUCTION Slide 1 CONGENITAL HEART DISEASE Jakub Kadlec/Catherine Sudarshan NHS Trust Slide 2 INTRODUCTION Most common congenital illness in the newborn Affects about 4 9 / 1000 full-term live births in the UK 1.5

More information

SURGICAL TREATMENT AND OUTCOME OF CONGENITAL HEART DISEASE

SURGICAL TREATMENT AND OUTCOME OF CONGENITAL HEART DISEASE SURGICAL TREATMENT AND OUTCOME OF CONGENITAL HEART DISEASE Mr. W. Brawn Birmingham Children s Hospital. Aims of surgery The aim of surgery in congenital heart disease is to correct or palliate the heart

More information

Adult Congenital Heart Disease: What All Echocardiographers Should Know Sharon L. Roble, MD, FACC Echo Hawaii 2016

Adult Congenital Heart Disease: What All Echocardiographers Should Know Sharon L. Roble, MD, FACC Echo Hawaii 2016 1 Adult Congenital Heart Disease: What All Echocardiographers Should Know Sharon L. Roble, MD, FACC Echo Hawaii 2016 DISCLOSURES I have no disclosures relevant to today s talk 2 Why should all echocardiographers

More information

5.8 Congenital Heart Disease

5.8 Congenital Heart Disease 5.8 Congenital Heart Disease Congenital heart diseases (CHD) refer to structural or functional heart diseases, which are present at birth. Some of these lesions may be discovered later. prevalence of Chd

More information

TGA atrial vs arterial switch what do we need to look for and how to react

TGA atrial vs arterial switch what do we need to look for and how to react TGA atrial vs arterial switch what do we need to look for and how to react Folkert Meijboom, MD, PhD, FES Dept ardiology University Medical entre Utrecht The Netherlands TGA + atrial switch: Follow-up

More information

COMPREHENSIVE EVALUATION OF FETAL HEART R. GOWDAMARAJAN MD

COMPREHENSIVE EVALUATION OF FETAL HEART R. GOWDAMARAJAN MD COMPREHENSIVE EVALUATION OF FETAL HEART R. GOWDAMARAJAN MD Disclosure No Relevant Financial Relationships with Commercial Interests Fetal Echo: How to do it? Timing of Study -optimally between 22-24 weeks

More information

The incidence and risk factors of arrhythmias in the early period after cardiac surgery in pediatric patients

The incidence and risk factors of arrhythmias in the early period after cardiac surgery in pediatric patients The Turkish Journal of Pediatrics 2008; 50: 549-553 Original The incidence and risk factors of arrhythmias in the early period after cardiac surgery in pediatric patients Selman Vefa Yıldırım 1, Kürşad

More information

NEW TEMPORARY ATRIAL AND VENTRICULAR PACING LEADS FOR PATIENTS AFTER CARDIAC OPERATIONS

NEW TEMPORARY ATRIAL AND VENTRICULAR PACING LEADS FOR PATIENTS AFTER CARDIAC OPERATIONS I I NEW TEMPORARY ATRIAL AND VENTRICULAR PACING LEADS FOR PATIENTS AFTER CARDIAC OPERATIONS We have studied two new temporary pacing leads (Medtronic 6491 and 6492) intended for pacing after cardiac operations.

More information

DIAGNOSIS, MANAGEMENT AND OUTCOME OF HEART DISEASE IN SUDANESE PATIENTS

DIAGNOSIS, MANAGEMENT AND OUTCOME OF HEART DISEASE IN SUDANESE PATIENTS 434 E AST AFRICAN MEDICAL JOURNAL September 2007 East African Medical Journal Vol. 84 No. 9 September 2007 DIAGNOSIS, MANAGEMENT AND OUTCOME OF CONGENITAL HEART DISEASE IN SUDANESE PATIENTS K.M.A. Sulafa,

More information

Heart and Soul Evaluation of the Fetal Heart

Heart and Soul Evaluation of the Fetal Heart Heart and Soul Evaluation of the Fetal Heart Ivana M. Vettraino, M.D., M.B.A. Clinical Associate Professor, Michigan State University College of Human Medicine Objectives Review the embryology of the formation

More information

Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS

Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS NBG Codes 1st Letter 2nd Letter 3rd Letter A V D Chamber(s) Paced = atrium = ventricle = dual (both atrium and ventricle) Chamber(s) Sensed A = atrium

More information

Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement

Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement 217 Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement Partha Prateem Choudhury, Vivek Chaturvedi, Saibal Mukhopadhyay, Jamal Yusuf Department

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

Anatomy & Physiology

Anatomy & Physiology 1 Anatomy & Physiology Heart is divided into four chambers, two atrias & two ventricles. Atrioventricular valves (tricuspid & mitral) separate the atria from ventricles. they open & close to control flow

More information

Cigna - Prior Authorization Procedure List Cardiology

Cigna - Prior Authorization Procedure List Cardiology Cigna - Prior Authorization Procedure List Cardiology Category CPT Code CPT Code Description 33206 Insertion of new or replacement of permanent pacemaker with transvenous electrode(s); atrial 33207 Insertion

More information

Finally, a pacemaker may be either permanent or temporary, which will also factor into your code selections.

Finally, a pacemaker may be either permanent or temporary, which will also factor into your code selections. 2015 Cardiology Survival Guide Chapter 15: Pacemakers With more than 30 codes and many variables to choose from, you must weigh your options carefully when reporting insertion, revision, or removal of

More information

In children <10 kg, transvenous pacing developed in our unit

In children <10 kg, transvenous pacing developed in our unit Original Article Twenty-Seven Years Experience With Transvenous Pacemaker Implantation in Children Weighing

More information

CRF procedure PROCEDURE FLOW. PATIENT ASSESSMENT Symptoms and indication. Pacemaker & ICD registration. Procedures. Procedure ICD.

CRF procedure PROCEDURE FLOW. PATIENT ASSESSMENT Symptoms and indication. Pacemaker & ICD registration. Procedures. Procedure ICD. Pacemaker & ICD registration Procedures CRF procedure PROCEDURE FLOW Procedure ICD ICD PM ICM lead only PATIENT ASSESSMENT Symptoms and indication Symptoms and events (multiple possibilities) Asymptomatic

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Tseng ZH, Hayward RM, Clark NM, et al. Sudden death in patients with cardiac implantable electronic devices. JAMA Intern Med. Published online June 22, 2015. doi:10.1001/jamainternmed.2015.2641.

More information

Chapter 80 Implantable Cardiac Devices

Chapter 80 Implantable Cardiac Devices Chapter 80 Implantable Cardiac Devices Episode Overview: 1. List 5 Indications for permanent pacing 2. Pacemaker nomenclature - what do the 5 letters mean a. What are common Pacemaker types? 3. List the

More information

Changing Indications for Pacemakers in Children

Changing Indications for Pacemakers in Children Changing Indications for Pacemakers in Children William H. Fleming, M.D., Lynne B. Sarafian, M.N., John D. Kugler, M.D., Philip J. Hofschire, M.D., and Edward B. Clark, M.D. ABSTRACT Permanent pacemakers

More information

INNOVATIONS IN DEVICE THERAPY:

INNOVATIONS IN DEVICE THERAPY: INNOVATIONS IN DEVICE THERAPY: Subcutaneous ICDs, Leadless Pacemakers, CRT Indications David J Wilber MD Loyola University Medical Center Disclosures: ACC Foundation: Consultant; Biosense / Webster: Consultant,

More information

Case Report Left Ventricular Dysfunction Caused by Unrecognized Surgical AV block in a Patient with a Manifest Right Free Wall Accessory Pathway

Case Report Left Ventricular Dysfunction Caused by Unrecognized Surgical AV block in a Patient with a Manifest Right Free Wall Accessory Pathway 109 Case Report Left Ventricular Dysfunction Caused by Unrecognized Surgical AV block in a Patient with a Manifest Right Free Wall Accessory Pathway Rakesh Gopinathannair, MD, MA 1, Dwayne N Campbell,

More information

Lead extraction. Dr. Mervat Abo El Maaty Professor of Cardiology Ain shams University 2013

Lead extraction. Dr. Mervat Abo El Maaty Professor of Cardiology Ain shams University 2013 Lead extraction Dr. Mervat Abo El Maaty Professor of Cardiology Ain shams University 2013 Agenda Introduction History of consensus Definitions Complications Indications Lead management environment Extraction

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

DON T FORGET TO OPTIMISE DEVICE PROGRAMMING

DON T FORGET TO OPTIMISE DEVICE PROGRAMMING CRT:NON-RESPONDERS OR NON-PROGRESSORS? DON T FORGET TO OPTIMISE DEVICE PROGRAMMING Prof. ALİ OTO,MD,FESC,FACC,FHRS Chairman,Department of Cardiology Hacettepe University Faculty of Medicine,Ankara Causes

More information

Journal of American Science 2014;10(9) Congenital Heart Disease in Pediatric with Down's Syndrome

Journal of American Science 2014;10(9)  Congenital Heart Disease in Pediatric with Down's Syndrome Journal of American Science 2014;10(9) http://www.jofamericanscience.org Congenital Heart Disease in Pediatric with Down's Syndrome Jawaher Khalid Almaimani; Maryam Faisal Zafir; Hanan Yousif Abbas and

More information

Prenatal diagnosis of long QT syndrome: Implications for delivery room and neonatal management

Prenatal diagnosis of long QT syndrome: Implications for delivery room and neonatal management Himmelfarb Health Sciences Library, The George Washington University Health Sciences Research Commons Pediatrics Faculty Publications Pediatrics 2-2013 Prenatal diagnosis of long QT syndrome: Implications

More information

Pattern of Congenital Heart Disease A Hospital-Based Study *Sadiq Mohammed Al-Hamash MBChB, FICMS

Pattern of Congenital Heart Disease A Hospital-Based Study *Sadiq Mohammed Al-Hamash MBChB, FICMS Pattern of Congenital Heart Disease A Hospital-Based Study *Sadiq Mohammed Al-Hamash MBChB, FICMS ABSTRACT Background: The congenital heart disease occurs in 0,8% of live births and they have a wide spectrum

More information

Left ventricular pacing in neonates and infants with isolated congenital complete or advanced atrioventricular block: short- and medium-term outcome

Left ventricular pacing in neonates and infants with isolated congenital complete or advanced atrioventricular block: short- and medium-term outcome Europace (2015) 17, 603 610 doi:10.1093/europace/euu180 CLINICAL RESEARCH Cardiac electrophysiology Left ventricular pacing in neonates and infants with isolated congenital complete or advanced atrioventricular

More information

Long-term Preservation of Left Ventricular Function and Heart Failure Incidence with Ablate and Pace Therapy Utilizing Biventricular Pacing

Long-term Preservation of Left Ventricular Function and Heart Failure Incidence with Ablate and Pace Therapy Utilizing Biventricular Pacing The Journal of Innovations in Cardiac Rhythm Management, 3 (2012), 976 981 HEART FAILURE RESEARCH ARTICLE Long-term Preservation of Left Ventricular Function and Heart Failure Incidence with Ablate and

More information

MONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES

MONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES PURPOSE Guidelines for caring for and monitoring patients with a temporary external epicardial pacemaker. Epicardial pacemaker monitoring is performed on a continuous basis by a nurse to evaluate the effectiveness

More information

Left ventricular guidewire pacing for transcatheter aortic valve. implantation

Left ventricular guidewire pacing for transcatheter aortic valve. implantation Page 1 of 8 Left ventricular guidewire pacing for transcatheter aortic valve implantation Ênio E. Guérios, MD 1, 2, Peter Wenaweser, MD 1, Bernhard Meier, MD 1 1 Department of Cardiology, Bern University

More information

Dipartimento di Scienze Cardiovascolari Università Campus Bio-Medico di Roma Dott. Vito Calabrese

Dipartimento di Scienze Cardiovascolari Università Campus Bio-Medico di Roma Dott. Vito Calabrese Dipartimento di Scienze Cardiovascolari Università Campus Bio-Medico di Roma Dott. Vito Calabrese Because the primary objective was cure symptomatic bradicardya due to syncope Because this is the common

More information

Recent technical advances and increasing experience

Recent technical advances and increasing experience Pediatric Open Heart Operations Without Diagnostic Cardiac Catheterization Jean-Pierre Pfammatter, MD, Pascal A. Berdat, MD, Thierry P. Carrel, MD, and Franco P. Stocker, MD Division of Pediatric Cardiology,

More information

ECHOCARDIOGRAPHIC APPROACH TO CONGENITAL HEART DISEASE: THE UNOPERATED ADULT

ECHOCARDIOGRAPHIC APPROACH TO CONGENITAL HEART DISEASE: THE UNOPERATED ADULT ECHOCARDIOGRAPHIC APPROACH TO CONGENITAL HEART DISEASE: THE UNOPERATED ADULT Karen Stout, MD, FACC Divisions of Cardiology University of Washington Medical Center Seattle Children s Hospital NO DISCLOSURES

More information

Outline. Congenital Heart Disease. Special Considerations for Special Populations: Congenital Heart Disease

Outline. Congenital Heart Disease. Special Considerations for Special Populations: Congenital Heart Disease Special Considerations for Special Populations: Congenital Heart Disease Valerie Bosco, FNP, EdD Alison Knauth Meadows, MD, PhD University of California San Francisco Adult Congenital Heart Program Outline

More information