NEW TEMPORARY ATRIAL AND VENTRICULAR PACING LEADS FOR PATIENTS AFTER CARDIAC OPERATIONS

Size: px
Start display at page:

Download "NEW TEMPORARY ATRIAL AND VENTRICULAR PACING LEADS FOR PATIENTS AFTER CARDIAC OPERATIONS"

Transcription

1 I I NEW TEMPORARY ATRIAL AND VENTRICULAR PACING LEADS FOR PATIENTS AFTER CARDIAC OPERATIONS We have studied two new temporary pacing leads (Medtronic 6491 and 6492) intended for pacing after cardiac operations. The conductor has stainless steel strands coated with polyethylene connected to a 4 mm 2 surface area, stainless steel, smooth, tapered electrode. A soft 4-0 coiled polypropylene fiber served as a fixation mechanism in the heart. The study included 15 children (aged 3 months to 7 years, body weight 4.4 to 20 kg) with a variety of congenital heart defects and 15 adults (aged 45 to 78 years) with coronary artery disease (n = 13) and aortic valve disease (n = 2). A pair of leads each was placed in a bipolar fashion in the right atrial wall and nonsystemic ventricle in the children (median implant duration 12 days) and in the right atrial wall only in the adults (median implant duration 9 days). The atrial current threshold values in children increased from ma immediately after implant to ma at explant (p < 0.002). In the adults the threshold values increased from ma immediately after implant to ma at explant (p < 0.002). In the ventricle the threshold values increased from ma immediately after implant to ma at explant (p < 0.002). Tissue resistance immediately after implant measured ~ in the atrium and increased to ~ at explant (children, p = not significant). Corresponding values in adults were ~ and l~l (p < 0.004). In the ventricle resistances changed from 1019 _ 143 to ~ (p < 0.05). P wave amplitudes measured mv immediately after implant and decreased to 1.6 _+ 1.2 mv at explant (p = not significant, children) and mv to mv (p = not significant, adults). R wave amplitudes were mv immediately after implant and fell to mv at explant (p < 0.005). Thus threshold values, tissue resistances, and electrogram amplitudes assured a safe pacemaker function. The small diameter and pliable texture of these leads provided a smooth surgical handling. They were found particularly suitable in children. (J THORAC CARDIOVASC SURG 1995;110: ) Ole-J6rgen Ohm, MD, a Kjell Breivik, MD, a Leidulf Segadal, MD, b and Hogne Engedal, MD, b Bergen, Norway emporary pacing leads are routinely used after T cardiac operations. The electrical properties of these leads are important for a safe pacemaker function. We have previously shown that the braided From Medical Department A a and the Surgical Institute, b University of Bergen, School of Medicine, Haukeland Sykehus, Bergen, Norway. Received for publication July 31, Accepted for publication Jan. 5, Address for reprints: Ole-J6rgen Ohm, MD, FACC, Professor of Cardiology, Medical Department A, University of Bergen, School of Medicine, Haukeland Sykehus, N-5021 Bergen, Norway. Copyright 1995 by Mosby-Year Book, Inc /95 $ /1/63337 multifilament wire electrodes have a high failure rate both for sensing of spontaneous heart activity and safe heart stimulationj' 2 The introduction of a temporary lead with a localized fixed surface area was a great improvement that offered superior pacing and sensing performance and minimized chances of dislodgement.3, 4 Still, commercially available leads have not been found suitable for small children because of a large diameter and poor pliability. The purpose of the present study was to evaluate the efficacy of two new temporary pacing leads designed specifically for implantation in the atrial and ventricular muscle of the smaller pediatric heart and in the adult to be afl]xed directly within the 1725

2 1726 Ohm et al. The Journal of Thoracic and December 1995 ~--~,.~ Model 6491 Tissue resistqnce -z O ~ ~ V- chddren ~ s A- child ren Fig. 1. Pediatric 6491 (top) and adult 6492 (bottom) leads used in study with small distal needle and breakaway chest needle slightly curved distal to breakaway marking. Inset is close-up view of polypropylene fixation coil and discrete electrode (diameter 0.5 mm, surface area 4 mm2) /,, ' ' ' ' ' > Days I me~lian ExptantQfion Table I. Congenital heart defects in the 15 children studied Tetralogy of Fallot Complete AV septal defect + tetralogy of Fallot Complete AV septal defect + pulmonary stenosis Complete AV septal defect Partial AV septal defect Gerbode defect TGA TGA + VSD (n = 2) Double outlet right ventricle + pulmonary stenosis + multiple VSDs Situs inversus + tetralogy of Fallot + AV septal defect + PAPVR VSD VSD + dextrocardia ASD + PAPVR ASD II AV,, Atriovenbicular; TGA, tr~insposition of the.great arteries; VSD, ventricular septal defect; PAPVR, partial anomalous pulmonary venous return~ ASD, atrial septal defect. atrial myocardium without the need for sutures or special techniques suchas use of a silicon disk or atrial piication technique, g' 5 Patients and methods Patients. The patients included in the study were 15 children and 15 adults undergoing cardiac operations. To qualify for the pediatric portion of the study the patients had to weigh 25 kg or less. They had a variety of congenital heart diseases (Table I). The age range was 3 months to 7 years, and the body weight varied between 4.4 and 20 kg. The 15 adult patients consisted of 14 men and 1 woman aged _ 10.4 years (range 45 to 78 years). Thirteen patients underwent coronary bypass grafting (7 patients had 3-vessel disease, 6 had 2-vessel disease) and Fig. 2. Tissue resistance calculated from voltage/current ratio 90 tzsec into pulse at myocardial threshold level using pacing system analyzer (Medtronic 5300). I, Implantation; V, ventricle; A, atrium. two Underwent aortic valve replacement (1 aortic stenosis, 1 combined aortic stenosis and insufficiency). The study was approved by the regional ethical committee and the patients or parents (for the children) gave written informed Consent to participate in the investigation. Methods Leads studied (Fig. 1). The leads used have a small 135-degree (3~ circle) curved needle with a length of 18 mm and a diameter of 0.4 mm. A soft 4-0 coiled polypropylene fiber is connected to the distal end of the electrode to serve as a fixation mechanism in the heart. The conductor consists of multifilament stainless steel strands coated with color-coded polyethylene insulation (yellow for pediatric Wire [6491] and purple for adult wire [6492]). The diameter measured over the insulation is 0.5 mm. The stainless steel electrode diameter is 0.5 mm and its surface area is 4 mm 2. A stainless steel needle is connected to the proximal end portion of the conductor to bring the conductor wire outside the thorax. The chest needle has a diameter of 0,7 mm and a length of 60 mm (pediatric) or 90 mm (adult). The needle is grooved to enable breakaway of the sharp section and is slightly curved distal to the breakaway marking to ease smooth chest passage, The remaining part functions as a contact sleeve for an electrocardiographic recorder or temporary pulse generator, Instrumentation. The myocardial excitability threshold, taken at 1 msec pulse width, was measured with a constant voltage source of the condenser discharge type, which also gives the current 90/xsec into the pulse (Medtronic 5300 pacing system analyzer, Medtr0nic, Inc., Minneapolis, Minn.). Thus the initial tissue resistance could be calculated. Additionally, the myocardial ichreshold was measured with a constant current source (Medtronic 5345 DDD temporary pulse generator). This device also deter-

3 The Journal of Thoracic and Volume 110, Number 6 Ohm et al Table II. Myocardial excitabifity thresholds in milliamperes during constant current stimulation at I msec pulse width Implantation Day 1 Day 3 Day 7 Explantation* Atrial, children Mean _+ SD 0.61 _ i _ _ Median 0.5 0, Range Ventricular, chil[dren Mean + SD _ _ Median Range Atrial, adults Mean + SD Median 0, Range Thresholds measured with Medtronic 5345 DDD temporary pulse generator, p Values < from implantation to explantation. SD, Standard deviation. *Median time to lead explant 9 days (range 7 to 13) in the adults and 12 days (2 to 23) in the children. mines the sensing potentials in millivolts with a lowest value of 0.5 mv and a highest value of 15 mv. Myocardial capture was confirmed on an oscilloscope or electrocardiographic recorder, or both. The lead combination that gave the lowest threshold at the initial measurement was connected to the negative terminal of the pulse generator for all subsequent recordings. Myocardial threshold was defined as the lowest voltage or current value that resulted in 1:1 capture when the pulse generator output was gradually increased from below. A dynamometer was used to measure the force in kilograms during lead removal. Operative procedure. One pair of unipolar leads each was implanted in the interatrial groove or posteriorly in the right atrium and on the nonsystemic ventricle with an approximate spacing of 1 cm to obtain a bipolar configuration. All leads of each kind were implanted by the same surgeon (L. S. for the children, H. E. for the adults). Measurements. Measurements were taken at the end of operation or immediately after admission to the intensive care unit, 3 to 6 hours after operation, and thereafter daily through explant. Statistical methods. Threshold values are given both as mean plus or minus the standard deviation and as median values because of skewed distribution of the data with a few high values. Electrogram amplitudes, initial tissue resistance, and force used during lead removal are given as mean plus or minus the standard deviation. Student's t test was used for comparing the development in myocardial threshold, tissue resistance, and electrogram amplitudes over time. Differences were considered statistically significant at p < Results Overall, the results were favorable both for cardiac stimulation and sensing of spontaneous heart activity. The leads were used exclusively on the atria of the adult population. Only one male adult patient had an obvious, dislocation, which occurred the first postoperative day as evidenced by both sensing failure and exit block. Data for this patient are excluded from further analyses. In four children intermittent sensing failures were observed in the atrial position in the early postoperative phase, and the ventricular leads functioned properly in all cases. Myocardial threshold at constant current pacing (Table II). With use of the constant current pulse generator, there was a threefold increase in myocardial threshold in the atrial position in children during the observation period, from a mean of 0.61 ma at implantation to 2.08 ma at explantation (p < 0.002). Although the values at implant were lower in the ventricle (mean 0.38 ma), the threshold increased about six times to a mean of 2.22 ma at explantation, which was also highly statistically significant (p < 0.002) and slightly higher compared with values in the atrium. The highest initial threshold values were found in the atrial position in the adults (mean 0.95 ma), and also in this group there was a threefold increase during the observation period (mean 2.76 ma at explantation, p < 0.002). Myocardial threshold at constant voltage (Table III). The current values measured with the constant voltage pulse generator were somewhat higher at implant compared with the values obtained with the constant current source. The reason for this discrepancy is the difference in the output characteristics for the two measuring devices. The increase was, however, of a similar magnitude to that for the constant current pulse generator during the observation period. Mean atrial current values in the infants increased from 0.88 ma to 3.24 ma and ventricular current values from 0.63 ma at implantation to 3.1 ma at explantation. Corresponding values in the atrium in the adults were 1.52 ma and

4 1728 Ohm et al. The Journal of Thoracic and December 1995 Table III. Myocardial excitability thresholds during constant voltage stimulation at 1 msec pulse width Atrial leads, children Mean ± SD Median Range Ventricular leads, children Mean ± SD Median Range Atrial leads, adults Mean ± SD Median Range Implantation Day 1 Day 3 Day 7 Explantation V ma V ma V ma V ma V rna 0.70± ± ± ± ± ± ± ± ±1.~ 3.24± ± ± ± ± ± ± ± ± ± ±2, ±1.534,01± ± ± ± ± ± ±2.52 0, % , % Thresholds measured with Medtronic 5300 pacing system analyzer, p Values < from implantation to explantation. SD, Standard deviation. Table IV. Electrogram amplitudes Atrial, children Ventricular, Atrial, (mv) children (mv) adults (mv) Implantation 1.8 ± ± ± 1.3 (0.5-6) (6-->15) (0.5-6) Day ± ± ± 2.1 ('<0.5-7) (12.5-~>15) (0.5-8) Day _ ± 1.7 ('<0.5-7) (2-_>15) ('<0.5-6) Day ± ('<0.5-5) (3-15) (0.5-4) Explantation 1.6 ± ± ± 1,1 ('<0.5-4) (3-->15) (0.5-4) Electrograms measured with Medtronic 5345 DDD temporary pulse generator. Values given as mean plus or minus the standard deviation with the range given in parentheses ma. The voltage threshold showed a threefold to fourfold increase during the study period. In the children the atrial threshold increased from a mean value of 0.70 V to 2.37 V and the threshold in the ventricle increased from 0.63 V at implantation to 2.52 V at explantation. Corresponding values in the atrial position in the adults were 1.06 V and 2.76 V (all p values <0.002). Initial tissue resistance (Fig. 2). Included in the tissue resistance is also the resistance of the lead wire, which is Because the leads were applied in the bipolar fashion, the wire resistance equals 32 [1 and thus represents a minor portion of the total resistance. The atrial tissue resistance was almost equal in children and in adults at implant. During the observation period there was a more marked and statistically significant drop in tissue resistance in the adults (p < 0.004) to a level of about 150 ~1 lower than that in the children and the value showed little variation after this initial drop through the study period. Initially the tissue resistance was about 200 f~ higher in the ventricle than in the atrium and it demonstrated a marked and statistically significant drop during the observation period (p < 0.05). Throughout the study the tissue resistances were about 100 fl higher in the ventricle than in the atrium for the children. Electrogram characteristics (Table IV and Fig. 3). The atrial electrograms were of similar amplitude in the children and adults and showed little change in mean values over time (p = not significant). The ventricular electrograms showed a continuous and statistically significant drop through the observation period (p < 0.005). In Fig. 3 an example is given of the use of the atrial leads both for diagnosis and treatment of an arrhythmia during the postoperative phase in one of the children with complex congenital heart disease. Removal of leads. At explant a dynamometer was used to measure the kilograms of force applied during removal of the leads. Because each pair of leads was tied together, they were removed simultaneously. The extraction force in the atrium varied between 0.08 and 0.83 kg ( kg, children) and 0.08 and 0.60 kg (0.34 _ kg, adults). In the ventricle the force varied from 0.08 to 0.51 kg (0.26 _ kg).

5 The Journal of Thoracic and Volume 110, Number 6 Ohm et al 1729 Fig. 3. Electrocardiographic and invasive blood pressure recordings on postoperative day 6 from 2-year-.old girl operated on because of situs inversus, tetralogy of Fallot, atrioventricular septal defect, and partial anomalous pulmonary venous return. During postoperative phase heart rhythm changed between sinus rhythm and accelerated atrioventricular (AV) junctional rhythm. Recording directly from atrial leads (R/l) demonstrated atrial rhythm at rate of 60 beats/rain and accelerated AV junctional rhythm at rate of 120 beats/rain (left part of figure). This is not obvious from monitoring lead (ML). In this case 1:1 AV conduction was obtained with atrial pacing (right part of figure). Atrial pacing (AP) close to spontaneous rate resulted in marked blood pressure elevation (P1) and general improvement in patient's clinical condition. Discussion The primary goal of this study was to investigate the handling and mechanical and electrical properties of these new leads. Both in the atrium and ventricle adequate sensing and secure pacing could be obtained. With the exception of a 30 mm longer chest needle for the adults and color coding of the leads, the characteristics of the pediatric and adult versions of the leads are similar. Handling and mechanical properties. The handling properties of the leads were advantageous. The maximal diameter of 0.5 mm of the lead, introduced with an even thinner distal needle, assured minimal trauma when the electrode was placed within the myocardial wall. No bleeding or dislocation was observed immediately after placement of the leads. This has been of special concern in infants and with thin atrial walls in adults. In some earlier models a silicon rubber disk has been recommended for the fixation of similar leads to the epicardium.4, 5 With the present models no additional fixation mechanism had to be used except for the retaining coil fastened to the distal end of the electrode. Obviously this coil did not result in additional damage to the tissue. Furthermore, the low force that had to be used during removal indicates a good balance in the memory of the retaining coil. A similar but somewhat higher extraction force had to be used when removing the leads from the atrium, both in children and adults, compared with the force required to remove the leads from the ventricles of the children. There was no indication of bleeding after lead removal. Although it has not been tested in this study, it is expected that the lead will function favorably also in the ventricle in the adult inasmuch as the electrode geometry and fixation mechanism are identical for the pediatric and adult lead versions. Myocardial threshold. In none of the patients was there indication of wire breakage or insulation defects after a maximal observation period of 23 days. Compared with those in earlier studies these new leads showed similar values of myocardial thresholds for temporary heart wires with a controlled surface area. 4-7 Also, the myocardial threshold values showed little difference in the atrial and ventricular positions. Although there was a continuous increase in myocardial threshold values, the values were low with a high safety margin for pacing (20 ma for the constant current pulse generator, 10 V for the constant voltage pulse generator). There was somewhat higher atrial threshold values in the

6 1730 Ohm et ai. The Journal of Thoracic and December 1995 adult than in the pediatric hearts, which can possibly be explained by structural changes in the myocardium caused by long-standing heart disease or less direct contact with excitable myocardial tissue as a result of more fatty tissue in the interatrial groove. Tissue resistance. The tissue resistance was of similar magnitude to that found for permanent pacemaker leads with a comparable surface area 8 and on average 200 to 300 Ft higher than that found for earlier temporary electrode models with a localized stainless steel electrode with a surface area of 7.5 mm2. 3' 4 This is of great importance to obtain an adequate current waveform during heart stimulation and thus secure optimal pacemaker function. This will furthermore decrease the current drain on the pulse generator, although this latter factor is of lesser importance in a temporary system. Also, the stable tissue resistance values throughout the study period indicated the integrity of the leads. There is no obvious explanation for the more marked drop and lower tissue resistance in the adults, because the electrodes are identical. Of note also is the higher tissue resistance found in the ventricular myocardium compared with that in the atrial myocardium. Eleetrogram amplitudes. Atrial electrogram amplitudes were of similar size to those found for earlier models of temporary pacing leads in which different application techniques had been used. 3-~ In four children the atrial electrograms intermittently showed borderline amplitudes for adequate sensing ( mv) with the pulse generator used. In two or three of these cases this could be ascribed to temporary instability in the childrens' clinical condition. The ventricular electrogram findings were comparable to values found for permanent and temporary pacemaker leads. 3' 9 In the ventricle there was a high sensitivity margin although there was a continuous drop in amplitude during the observation period. Applicability of the leads. Heart wires can effectively be used for diagnosis and treatment of cardiac arrhythmias in the postoperative phase of cardiac operations. Not only can pacing be used in case of bradyarrhythmia to improve the hemodynamic situation, but also by use of these wires episodes of supraventricular and ventricular tachycardias can be treated immediately without additional invasive procedures. Furthermore, antiarrhythmic drugs with negative inotropic and potential proarrhythmic effects and direct cardioversion necessitating the use of general anesthesia can be avoided. These wires can also be used for electrophysiologic testing before hospital discharge: for example, in patients who have had surgery for ventricular tachycardia. Recognition of atrioventrieular synchrony. With the now available external DDD pacemaker it is important to have a reliable atrial lead to take advantage of improved hemodynamics of atrioventricular synchrony, especially for the patient with poor left ventricular function, reduced compliance, or complex congenital heart corrections (Fig. 3). Because the leads are applied in the bipolar fashion a high pulse generator sensitivity can be used with little risk for oversensing or cross-talk between the atrial and ventricular channels. Furthermore, interference from external sources seems unlikely. Conclusion. The new temporary pacing leads have shown good performance in patients undergoing cardiac operations. Because of the slender construction and pliable texture of the leads trauma to the heart is minimal. The special memory coil secures safe positioning without additional fixation mechanisms. The leads seem particularly favorable for application in small children and for atrial application in the adult. The support of Mr. Wire van Driel, Manager Heart Valve Department, Bakken Research Center, Maastricht, The Netherlands, is highly appreciated. REFERENCES 1. Ohm O-J, M6rkrid L, Skagseth E. Temporary pacemaker treatment in open heart surgery: variation in myocardial threshold, tissue and interface impedances in man. PACE 1979;2: Ohm O-J, Skagseth E. Temporary pacemaker treatment in open heart surgery: pre- to postoperative changes in the electrogram characteristics. PACE 1980; 3: Breivik K, Engedal H, Segadal L, Ohm 04. New temporary pacing lead for use after cardiac operations. J THORAC CARDIOVASC SURG 1982;84: Breivik K, Engedal H, Resch F, Segadal L, Ohm O-J. Bipolar atrial application of a new temporary pacing lead after cardiac operations. J TI~ORaC CARDIOVASC SURG 1983;85: Ferguson TB, Cox JL. Temporary external DDD pacing after cardiac operations. Ann Thorac Surg 1991;51: Wigneswaran WT, Jamieson MPG. Temporary pacing leads in cardiac surgery: a comparison of multifilament braided electrode and localized solitary stainless steel electrode. J Cardiovasc Surg 1986;27:

7 The Journal of Thoracic and Volume 110, Number 6 Ohm et al Wirtz St, Schulte HD, Winter J, Godehardt E, Kunert J. Reliability of different temporary myocardial pacing leads. Thorac Cardiovasc Surg 1989;37: Hoff PI, Breivik K, Tronstad A, Andersen KS, Ohm O-J. A new steroid-eluting electrode for low-threshold pacing. In: P6rez Gdmez F, ed. Cardiac pacing, electrophysiology, tachyarrhythmias. Mount Kisco, New York: Futura Media Services, 1985: Ohm O-J, Breivik K, Hammer EA, Hoff PI. Intraoperative electrical measurements during pacemaker implantation. Clin Prog Pacing Electrophysiol 1984;2:1-23.

Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children

Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children ERALD A. SERWER, M.D., MACDONALD DICK 11, M.D., KAREN UZARK, R.N., Ph.D., WILLIAM A. SCOTT, M.D., and EDWARD L.

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a

More information

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II). Cardiac Pacemaker Premature Beats When one of ectopic foci becomes irritable, it may spontaneously fire, leading to one or more premature beats. Atrial and junctional foci may become irritable from excess

More information

Pediatric Pacemaker Implantation Endocardial or Epicardial

Pediatric Pacemaker Implantation Endocardial or Epicardial Pediatric Pacemaker Implantation Endocardial or Epicardial HAITHAM BADRAN, MD, FEHRA CONSULTANT OF INTERVENTIONAL CARDIOLOGY CONSULTANT OF CARDIAC PACING AND ELECTROPHYSIOLOGY LECTURER OF CARDIOLOGY AIN

More information

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment Karen L. Booth, MD, Lucile Packard Children s Hospital Arrhythmias are common after congenital heart surgery [1]. Postoperative electrolyte

More information

Case 1 Left Atrial Tachycardia

Case 1 Left Atrial Tachycardia Case 1 Left Atrial Tachycardia A 16 years old woman was referred to our institution because of recurrent episodes of palpitations and dizziness despite previous ablation procedure( 13 years ago) of postero-septal

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

MEDICAL CORPORATION Asbury Rd., P.O. Box 758, Farmingdale, NJ USA Fax

MEDICAL CORPORATION Asbury Rd., P.O. Box 758, Farmingdale, NJ USA Fax MEDICAL CORPORATION 5206 Asbury Rd., P.O. Box 758, Farmingdale, NJ 07727 USA 732-938-2266 800-323-4035 Fax 732-938-2399 MYO/WIRE Temporary Cardiac Pacing Wire System MYO/WIRE temporary cardiac pacing wires.

More information

SUPPLEMENTAL MATERIAL

SUPPLEMENTAL MATERIAL SUPPLEMENTAL MATERIAL Table S1. Patient Selection Codes, CIED Generator Procedures Code Type Code Description ICD9 Proc 00.51 Implantation of cardiac resynchronization defibrillator, total system [CRT-D]

More information

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! What is a

More information

Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases

Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases Case Report Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases Koji Tsutsumi MD, Tatsuru Niibori MD, Keiichiro Katsumoto MD Department

More information

Cardiovascular Nursing Practice: A Comprehensive Resource Manual and Study Guide for Clinical Nurses 2 nd Edition

Cardiovascular Nursing Practice: A Comprehensive Resource Manual and Study Guide for Clinical Nurses 2 nd Edition Cardiovascular Nursing Practice: A Comprehensive Resource Manual and Study Guide for Clinical Nurses 2 nd Edition Table of Contents Volume 1 Chapter 1: Cardiovascular Anatomy and Physiology Basic Cardiac

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

PART I. Disorders of the Heart Rhythm: Basic Principles

PART I. Disorders of the Heart Rhythm: Basic Principles PART I Disorders of the Heart Rhythm: Basic Principles FET01.indd 1 1/11/06 9:53:05 AM FET01.indd 2 1/11/06 9:53:06 AM CHAPTER 1 The Cardiac Electrical System The heart spontaneously generates electrical

More information

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION 1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,

More information

Clinical Data Summary: Avoid FFS Study

Clinical Data Summary: Avoid FFS Study Atrial Pacing Lead with 1.1 mm Tip-to-Ring Spacing Clinical Data Summary: Avoid FFS Study A Multi-center, Randomized, Prospective Clinical Study Designed to Evaluate the 1699T Lead Clinical Data Summary:

More information

RN-BC, MS, CCRN, FAHA

RN-BC, MS, CCRN, FAHA Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! 1 A. Atropine

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

Transcoronary Chemical Ablation of Atrioventricular Conduction

Transcoronary Chemical Ablation of Atrioventricular Conduction 757 Transcoronary Chemical Ablation of Atrioventricular Conduction Pedro Brugada, MD, Hans de Swart, MD, Joep Smeets, MD, and Hein J.J. Wellens, MD In seven patients with symptomatic atrial fibrillation

More information

Dual-Chamber Implantable Cardioverter-Defibrillator

Dual-Chamber Implantable Cardioverter-Defibrillator February 1998 9 Dual-Chamber Implantable Cardioverter-Defibrillator A.SH. REVISHVILI A.N. Bakoulev Research Center for Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow, Russia Summary

More information

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

The heart's natural pacemaker is called the sinoatrial (SA) node or sinus node. PACEMAKER Natural pacemaker: The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node. Artificial pacemaker: It is a small, battery-operated device that helps the heart beat in

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

Positive Results with a New Screw-in Electrode with Isolated Screw and Fractal Coating of the Ring: the RETROX Electrode in the Atrial Position

Positive Results with a New Screw-in Electrode with Isolated Screw and Fractal Coating of the Ring: the RETROX Electrode in the Atrial Position 314 June 1999 Positive Results with a New Screw-in Electrode with Isolated Screw and Fractal Coating of the Ring: the RETROX Electrode in the Atrial Position M. BOKERN Waterland Ziekenhuis, Purmerend,

More information

Cardiology. Objectives. Chapter

Cardiology. Objectives. Chapter 1:44 M age 1121 Chapter Cardiology Objectives art 1: Cardiovascular natomy and hysiology, ECG Monitoring, and Dysrhythmia nalysis (begins on p. 1127) fter reading art 1 of this chapter, you should be able

More information

The incidence and risk factors of arrhythmias in the early period after cardiac surgery in pediatric patients

The incidence and risk factors of arrhythmias in the early period after cardiac surgery in pediatric patients The Turkish Journal of Pediatrics 2008; 50: 549-553 Original The incidence and risk factors of arrhythmias in the early period after cardiac surgery in pediatric patients Selman Vefa Yıldırım 1, Kürşad

More information

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable

More information

Transvenous Pacemaker Implantation 22 years after the Mustard Procedure

Transvenous Pacemaker Implantation 22 years after the Mustard Procedure Case Report Transvenous Pacemaker Implantation 22 years after the Mustard Procedure Masato Sakamoto MD, Yoshie Ochiai MD, Yutaka Imoto MD, Akira Sese MD, Mamie Watanabe MD, Kunitaka Joo MD Department of

More information

Project Title Temporary Pacemaker Training Simulator

Project Title Temporary Pacemaker Training Simulator Project Title Temporary Pacemaker Training Simulator Project Description Problem: There is no available training device for temporary pacemakers (pacemakers). A training device will have to essentially

More information

Pacing troubleshooting. NASPE training Lancashire & South Cumbria Cardiac Network

Pacing troubleshooting. NASPE training Lancashire & South Cumbria Cardiac Network Pacing troubleshooting NASPE training Lancashire & South Cumbria Cardiac Network Pacing stimulus present failure to capture Lead dislodgement Early unstable position Late twiddlers syndrome Abnormal ECG

More information

Biatrial Stimulation and the Prevention of Atrial Fibrillation

Biatrial Stimulation and the Prevention of Atrial Fibrillation 374 September 2001 Biatrial Stimulation and the Prevention of Atrial Fibrillation L. MELCZER Heart Institute, Faculty of Medicine, University of Pécs, Pécs, Hungary I. LORINCZ 1 st Internal Department,

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

Active Fixation Models: 7740, 7741 and 7742 Passive Fixation Models: 7731, 7732, 7735 and 7736

Active Fixation Models: 7740, 7741 and 7742 Passive Fixation Models: 7731, 7732, 7735 and 7736 The INGEVITY MRI pacing leads are 6F (2.0 mm) steroid-eluting, endocardial pace / sense leads designed for permanent implantation for either atrial or ventricular applications. INGEVITY MRI is the only

More information

Patient Resources: Arrhythmias and Congenital Heart Disease

Patient Resources: Arrhythmias and Congenital Heart Disease Patient Resources: Arrhythmias and Congenital Heart Disease Overview Arrhythmias (abnormal heart rhythms) can develop in patients with congenital heart disease (CHD) due to thickening/weakening of their

More information

Chapter 12: Cardiovascular Physiology System Overview

Chapter 12: Cardiovascular Physiology System Overview Chapter 12: Cardiovascular Physiology System Overview Components of the cardiovascular system: Heart Vascular system Blood Figure 12-1 Plasma includes water, ions, proteins, nutrients, hormones, wastes,

More information

Index. cardiology.theclinics.com. Note: Page numbers of article titles are in boldface type.

Index. cardiology.theclinics.com. Note: Page numbers of article titles are in boldface type. Index Note: Page numbers of article titles are in boldface type. A ACHD. See Adult congenital heart disease (ACHD) Adult congenital heart disease (ACHD), 503 512 across life span prevalence of, 504 506

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

Procurement Support to Ministry of Health, Ukraine. Invitation to Bid for:

Procurement Support to Ministry of Health, Ukraine. Invitation to Bid for: Procurement Support to Ministry of Health, Ukraine Invitation to Bid for: PROGRAMME XIX: MEDICAL PRODUCTS FOR HEALTHCARE INSTITUTIONS FOR TREATMENT OF PATIENTS WITH CARDIOVASCULAR AND CEREBROVASCULAR DISEASES

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL

More information

Cardiac Radiology In-Training Test Questions for Diagnostic Radiology Residents

Cardiac Radiology In-Training Test Questions for Diagnostic Radiology Residents Cardiac Radiology In-Training Test Questions for Diagnostic Radiology Residents March, 2013 Sponsored by: Commission on Education Committee on Residency Training in Diagnostic Radiology 2013 by American

More information

Cardiovascular Physiology. Heart Physiology. Introduction. The heart. Electrophysiology of the heart

Cardiovascular Physiology. Heart Physiology. Introduction. The heart. Electrophysiology of the heart Cardiovascular Physiology Heart Physiology Introduction The cardiovascular system consists of the heart and two vascular systems, the systemic and pulmonary circulations. The heart pumps blood through

More information

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure Inappropriate Discharges of Intravenous Implantable Cardioverter Defibrillators Owing to Lead Failure Takashi WASHIZUKA, 1 MD, Masaomi CHINUSHI, 1 MD, Ryu KAZAMA, 1 MD, Takashi HIRONO, 1 MD, Hiroshi WATANABE,

More information

Electrocardiography Abnormalities (Arrhythmias) 7. Faisal I. Mohammed, MD, PhD

Electrocardiography Abnormalities (Arrhythmias) 7. Faisal I. Mohammed, MD, PhD Electrocardiography Abnormalities (Arrhythmias) 7 Faisal I. Mohammed, MD, PhD 1 Causes of Cardiac Arrythmias Abnormal rhythmicity of the pacemaker Shift of pacemaker from sinus node Blocks at different

More information

A Fractal Coated, 1.3 mm 2 High Impedance Pacing Electrode: Results from a Multicenter Clinical Trial

A Fractal Coated, 1.3 mm 2 High Impedance Pacing Electrode: Results from a Multicenter Clinical Trial 140 April 2000 A Fractal Coated, 1.3 mm 2 High Impedance Pacing Electrode: Results from a Multicenter Clinical Trial G. PIOGER Clinique Alleray, Paris, France A. LAZARUS Clinique du Val d'or, Saint-Cloud,

More information

Cardiac Catheterization Cases Primary Cardiac Diagnoses Facility 12 month period from to PRIMARY DIAGNOSES (one per patient)

Cardiac Catheterization Cases Primary Cardiac Diagnoses Facility 12 month period from to PRIMARY DIAGNOSES (one per patient) PRIMARY DIAGNOSES (one per patient) Septal Defects ASD (Atrial Septal Defect) PFO (Patent Foramen Ovale) ASD, Secundum ASD, Sinus venosus ASD, Coronary sinus ASD, Common atrium (single atrium) VSD (Ventricular

More information

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΖΩΡΤΖ ΔΑΔΟΥΣ ΕΠΙΚΟΥΡΟΣ ΚΑΘΗΓΗΤΗΣ Α.Π.Θ. ΜΑΡΙΑ ΚΑΡΑΛΙΟΛΙΟΥ ΕΙΔΙΚΕΥΟΜΕΝΗ ΙΑΤΡΟΣ ΚΑΡΔΙΟΛΟΓΟΣ Β ΚΑΡΔΙΟΛΟΓΙΚΗ ΚΛΙΝΙΚΗ Α.Π.Θ. ΙΠΠΟΚΡΑΤΕΙΟ Γ.Ν.Θ. 5 Why a Pacemaker is Implanted

More information

Automatic Atrial Threshold Measurement and Adjustment in Pediatric Patients

Automatic Atrial Threshold Measurement and Adjustment in Pediatric Patients Automatic Atrial Threshold Measurement and Adjustment in Pediatric Patients ANITA HIIPPALA, M.D.,* GERALD A. SERWER, M.D., EVA CLAUSSON, M.Sc., LYNN DAVENPORT, M.S., TRINA BRAND, PH.D., and JUHA-MATTI

More information

Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict

Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict DOI 10.1007/s12471-011-0158-5 ORIGINAL ARTICLE Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict M. C. Post & W. Budts & A. Van de Bruaene & R. Willems

More information

Dear Parent/Guardian,

Dear Parent/Guardian, Dear Parent/Guardian, You have indicated on school records that your child has an ongoing health problem that may require medication and/or treatment during the school day with rescue medication. Attached

More information

Techniques for repair of complete atrioventricular septal

Techniques for repair of complete atrioventricular septal No Ventricular Septal Defect Patch Atrioventricular Septal Defect Repair Carl L. Backer, MD *, Osama Eltayeb, MD *, Michael C. Mongé, MD *, and John M. Costello, MD For the past 10 years, our center has

More information

3/17/2014. WS # 3 ICD Registry Case Scenarios with Structural Abnormalities. Objectives. Denise Pond BSN, RN

3/17/2014. WS # 3 ICD Registry Case Scenarios with Structural Abnormalities. Objectives. Denise Pond BSN, RN WS # 3 ICD Registry Case Scenarios with Structural Abnormalities Denise Pond BSN, RN The following relationships exist related to this presentation: No Disclosures Objectives Discuss specific coding instructions

More information

The Double Switch Using Bidirectional Glenn and Hemi-Mustard. Frank Hanley

The Double Switch Using Bidirectional Glenn and Hemi-Mustard. Frank Hanley The Double Switch Using Bidirectional Glenn and Hemi-Mustard Frank Hanley No relationships to disclose CCTGA Interesting Points for Discussion What to do when. associated defects must be addressed surgically:

More information

Arrhythmias and congenital heart disease

Arrhythmias and congenital heart disease Arrhythmias and congenital heart disease Jolien Roos-Hesselink ErasmusMC Rotterdam Netherlands Patient W, born in 1969 Tetralogy of Fallot 1975 at the age of 6 years surgical correction 2002 Patient W:

More information

National Coverage Determination (NCD) for Cardiac Pacemakers (20.8)

National Coverage Determination (NCD) for Cardiac Pacemakers (20.8) Page 1 of 12 Centers for Medicare & Medicaid Services National Coverage Determination (NCD) for Cardiac Pacemakers (20.8) Tracking Information Publication Number 100-3 Manual Section Number 20.8 Manual

More information

Left ventricular guidewire pacing for transcatheter aortic valve. implantation

Left ventricular guidewire pacing for transcatheter aortic valve. implantation Page 1 of 8 Left ventricular guidewire pacing for transcatheter aortic valve implantation Ênio E. Guérios, MD 1, 2, Peter Wenaweser, MD 1, Bernhard Meier, MD 1 1 Department of Cardiology, Bern University

More information

Ventricular Arrhythmias

Ventricular Arrhythmias Presenting your most challenging cases Venice Arrhythmias Ventricular Arrhythmias Gioia Turitto, MD Presenter Disclosure Information A questionable indication for CRT-D in a patient with VT after successful

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

Anatomy of Atrioventricular Septal Defect (AVSD)

Anatomy of Atrioventricular Septal Defect (AVSD) Surgical challenges in atrio-ventricular septal defect in grown-up congenital heart disease Anatomy of Atrioventricular Septal Defect (AVSD) S. Yen Ho Professor of Cardiac Morphology Royal Brompton and

More information

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie 268 Case Report Cardiac Resynchronization Therapy in a Patient with Persistent Left Superior Vena Cava Draining into the Coronary Sinus and Absent Innominate Vein: A Case Report and Review of Literature

More information

CARDIOVASCULAR SYSTEM

CARDIOVASCULAR SYSTEM CARDIOVASCULAR SYSTEM Overview Heart and Vessels 2 Major Divisions Pulmonary Circuit Systemic Circuit Closed and Continuous Loop Location Aorta Superior vena cava Right lung Pulmonary trunk Base of heart

More information

Practice Questions.

Practice Questions. IBHRE Prep Practice Questions Question 1 The relative refractory yperiod of the ventricular myocardium corresponds to which of the following phases of the action potential? A. (0) B. (1) C. (2) D. (3)

More information

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Case Report Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Taku Nishida MD, Tamio Nakajima MD, PhD, Yutaka Goryo MD, Ken-ichi Ishigami MD, PhD, Hiroyuki Kawata

More information

EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology

EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important

More information

Arrhythmias in Post-operative VSD. Jing-Ming Wu, M.D. Professor & Chairman of Pediatrics, National Cheng Kung University Hospital Tainan, Taiwan

Arrhythmias in Post-operative VSD. Jing-Ming Wu, M.D. Professor & Chairman of Pediatrics, National Cheng Kung University Hospital Tainan, Taiwan Arrhythmias in Post-operative VSD Jing-Ming Wu, M.D. Professor & Chairman of Pediatrics, National Cheng Kung University Hospital Tainan, Taiwan Arrhythmias in Post-operative VSD Not uncommon (30%), and

More information

Long Term Monitoring of the Intrinsic Ventricular Monophasic Action Potential with an Implantable DDD Pacemaker

Long Term Monitoring of the Intrinsic Ventricular Monophasic Action Potential with an Implantable DDD Pacemaker May 1998 79 Long Term Monitoring of the Intrinsic Ventricular Monophasic Action Potential with an Implantable DDD Pacemaker T. LAWO, J. BARMEYER Abteilung für Kardiologie, Universitätsklinik Bergmannsheil,

More information

The Electrocardiogram

The Electrocardiogram The Electrocardiogram Chapters 11 and 13 AUTUMN WEDAN AND NATASHA MCDOUGAL The Normal Electrocardiogram P-wave Generated when the atria depolarizes QRS-Complex Ventricles depolarizing before a contraction

More information

Differentiation of Ventricular Tachycardia from Junctional Tachycardia with Aberrant Conduction

Differentiation of Ventricular Tachycardia from Junctional Tachycardia with Aberrant Conduction Differentiation of Ventricular Tachycardia from Junctional Tachycardia with Aberrant Conduction The Use of Competitive Atrial Pacing By ROBERT M. EASLEY, JR., M.D., AND SIDNEY GOLDSTEIN, M.D. SUMMARY A

More information

OLBI Stimulation in Biatrial Pacing? A Comparison of Acute Pacing and Sensing Conditions for Split Bipolar and Dual Cathodal Unipolar Configurations

OLBI Stimulation in Biatrial Pacing? A Comparison of Acute Pacing and Sensing Conditions for Split Bipolar and Dual Cathodal Unipolar Configurations 236 June 1999 OLBI Stimulation in Biatrial Pacing? A Comparison of Acute Pacing and Sensing Conditions for Split Bipolar and Dual Cathodal Unipolar Configurations A. KUTARSKI, M. SCHALDACH*, M. WÓJCIK,

More information

Introduction to Electrocardiography

Introduction to Electrocardiography Introduction to Electrocardiography Class Objectives: Introduction to ECG monitoring Discuss principles of interpretation Identify the components and measurements of the ECG ECG analysis ECG Monitoring

More information

DEVELOPMENT OF THE CIRCULATORY SYSTEM L E C T U R E 5

DEVELOPMENT OF THE CIRCULATORY SYSTEM L E C T U R E 5 DEVELOPMENT OF THE CIRCULATORY SYSTEM L E C T U R E 5 REVIEW OF CARDIAC ANATOMY Heart 4 chambers Base and apex Valves Pericardial sac 3 layers: epi, myo, endo cardium Major blood vessels Aorta and its

More information

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Jason Collinson & Stuart Tan Essex Cardiothoracic Centre, Basildon and Thurrock University Hospital. Contact: jason.collinson@nhs.net

More information

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine

More information

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds

More information

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Patient Scheduling and Screening Q: Is there any special (other than regular MRI consent) consent form that should be signed by the patient?

More information

Catheter Interruption of Atrioventricular Conduction Using Radiofrequency Energy in a Patient with Transposition ofthe Great Arteries

Catheter Interruption of Atrioventricular Conduction Using Radiofrequency Energy in a Patient with Transposition ofthe Great Arteries Catheter Interruption of Atrioventricular Conduction Using Radiofrequency Energy in a Patient with Transposition ofthe Great Arteries MARK W. RUSSELL, PARVIN C. DOROSTKAR, MACDONALD DICK, II, - JOSEPHA

More information

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program Electrophysiology Implant Classification Table The table below contains the codes that apply to our UnitedHealthcare Medicare Advantage cardiology prior Description Includes Generator Placement Includes

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

Heart and Lungs. LUNG Coronal section demonstrates relationship of pulmonary parenchyma to heart and chest wall.

Heart and Lungs. LUNG Coronal section demonstrates relationship of pulmonary parenchyma to heart and chest wall. Heart and Lungs Normal Sonographic Anatomy THORAX Axial and coronal sections demonstrate integrity of thorax, fetal breathing movements, and overall size and shape. LUNG Coronal section demonstrates relationship

More information

Outcomes of Defibrillator Lead Implants Performed by High Volume Operators vs. Low Volume Operators:

Outcomes of Defibrillator Lead Implants Performed by High Volume Operators vs. Low Volume Operators: Outcomes of Defibrillator Lead Implants Performed by High Volume Operators vs. Low Volume Operators: Results from the Pacemaker and Implantable Defibrillator Leads Survival Study ( PAIDLESS ) Partially

More information

ISO INTERNATIONAL STANDARD. Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers

ISO INTERNATIONAL STANDARD. Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers INTERNATIONAL STANDARD ISO 14708-2 Second edition 2012-08-15 Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers Implants chirurgicaux Dispositifs médicaux implantables actifs

More information

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work Electrical Activity of the Human Heart Oguz Poroy, PhD Assistant Professor Department of Biomedical Engineering The University of Iowa Outline Basic Facts about the Heart Heart Chambers and Heart s The

More information

BIPN100 F15 Human Physiology I (Kristan) Problem set #5 p. 1

BIPN100 F15 Human Physiology I (Kristan) Problem set #5 p. 1 BIPN100 F15 Human Physiology I (Kristan) Problem set #5 p. 1 1. Dantrolene has the same effect on smooth muscles as it has on skeletal muscle: it relaxes them by blocking the release of Ca ++ from the

More information

Case Report Left Ventricular Dysfunction Caused by Unrecognized Surgical AV block in a Patient with a Manifest Right Free Wall Accessory Pathway

Case Report Left Ventricular Dysfunction Caused by Unrecognized Surgical AV block in a Patient with a Manifest Right Free Wall Accessory Pathway 109 Case Report Left Ventricular Dysfunction Caused by Unrecognized Surgical AV block in a Patient with a Manifest Right Free Wall Accessory Pathway Rakesh Gopinathannair, MD, MA 1, Dwayne N Campbell,

More information

Mitral Valve Disease, When to Intervene

Mitral Valve Disease, When to Intervene Mitral Valve Disease, When to Intervene Swedish Heart and Vascular Institute Ming Zhang MD PhD Interventional Cardiology Structure Heart Disease Conflict of Interest None Current ACC/AHA guideline Stages

More information

CORONARY ARTERIES HEART

CORONARY ARTERIES HEART CARDIAC/ECG MODULE THE HEART CORONARY ARTERIES FIBRILLATING HEART CORONARY ARTERIES HEART PRACTICE RHYTHMS PRACTICE RHYTHMS ELECTRICAL CONDUCTION SA Node (60 100) Primary pacemaker AV Node (40 60) ***Creates

More information

ELECTROCARDIOGRAPHY (ECG)

ELECTROCARDIOGRAPHY (ECG) ELECTROCARDIOGRAPHY (ECG) The heart is a muscular organ, which pumps blood through the blood vessels of the circulatory system. Blood provides the body with oxygen and nutrients, as well as assists in

More information

Diploma in Electrocardiography

Diploma in Electrocardiography The Society for Cardiological Science and Technology Diploma in Electrocardiography The Society makes this award to candidates who can demonstrate the ability to accurately record a resting 12-lead electrocardiogram

More information

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study ANEESH V. TOLAT, M.D.,* MELISSA WOICIECHOWSKI, M.S.N.,* ROSEMARIE KAHR, R.C.I.S.,* JOSEPH

More information

Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS

Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS NBG Codes 1st Letter 2nd Letter 3rd Letter A V D Chamber(s) Paced = atrium = ventricle = dual (both atrium and ventricle) Chamber(s) Sensed A = atrium

More information

Adult Congenital Heart Disease: What All Echocardiographers Should Know Sharon L. Roble, MD, FACC Echo Hawaii 2016

Adult Congenital Heart Disease: What All Echocardiographers Should Know Sharon L. Roble, MD, FACC Echo Hawaii 2016 1 Adult Congenital Heart Disease: What All Echocardiographers Should Know Sharon L. Roble, MD, FACC Echo Hawaii 2016 DISCLOSURES I have no disclosures relevant to today s talk 2 Why should all echocardiographers

More information

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing

More information

Temporary Myocardial Electrodes (TME) In Theory and Practice

Temporary Myocardial Electrodes (TME) In Theory and Practice Version 12.09.05 Temporary Myocardial Electrodes (TME) In Theory and Practice Summary of 25 Years of Experience Dr. Peter Osypka This work and the information contained within are the property of Dr. Osypka

More information

Interpretation of complex demand pacemaker arrhythmias

Interpretation of complex demand pacemaker arrhythmias British Heart Journal, I972, 34, 312-3I7. Interpretation of complex demand pacemaker arrhythmias S. Serge Barold, John J. Gaidula, Richard L. Banner, George I. Litman, and Sidney Goldstein From the Department

More information

Acute Evaluation of the Post Atrial Stimulation Evoked Response at Various Sites

Acute Evaluation of the Post Atrial Stimulation Evoked Response at Various Sites June 2000 311 Acute Evaluation of the Post Atrial Stimulation Evoked Response at Various Sites Y. GUYOMAR, P. GRAUX, R. CARLIOZ, C. MOULIN, A. DUTOIT Catholic University of Lille, St-Philibert Hospital,

More information

Where are the normal pacemaker and the backup pacemakers of the heart located?

Where are the normal pacemaker and the backup pacemakers of the heart located? CASE 9 A 68-year-old woman presents to the emergency center with shortness of breath, light-headedness, and chest pain described as being like an elephant sitting on her chest. She is diagnosed with a

More information

Cardiac arrhythmias in the PICU

Cardiac arrhythmias in the PICU Cardiac arrhythmias in the PICU Paolo Biban, MD Director, Neonatal and Paediatric Intensive Care Unit Division of Paediatrics, Major City Hospital Azienda Ospedaliera Universitaria Integrata Verona, Italy

More information

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular

More information

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART December 2015 This chart encompasses all Medtronic cardiac devices FDA-Approved as MR Conditional and included in the MRI SureScan portfolio. If a device

More information