1 Vibrational Analysis of Bioprosthetic Heart Valve Leaflets Using Numerical Models: Effects of Leaflet Stiffening, Calcification, and Perforation 687 Mohamed S. Hamid, Hani N. Sabbah, and Paul D. Stein The fundamental natural frequency of the closed cusps of porcine bioprosthetic valves, fabricated from the normal leaflets of pig aortic valves, was estimated using a finite element model. Both normal and stiffened leaflets were considered in the vibrational analysis. The effects of conditions that simulated degeneration, such as stiffening, central perforation, a tear, calcium deposits in the commissural attachments, and combinations of these were determined. The primary frequency of vibration of the normal leaflets was within the range of the dominant frequency of the heart sounds determined clinically by spectral analysis of the recorded phonocardiogram. If only one leaflet was stiffened or calcined, there was only a marginal change of frequency. With stiffening and calcification of the commissures of all 3 leaflets, the frequency of vibration increased. Introduction of a tear in a single leaflet of a stiffened and calcified valve markedly reduced the fundamental frequency. In view of the relation between the frequency content of heart sounds and the frequency of valve vibration, this mathematical simulation establishes a possible basis for the observation of a varying dominant frequency of heart sounds in patients with bioprosthetic valves that are in the process of degenerating. (Circulation Research 1987;61: ) Degeneration of porcine bioprosthetic valves (PB V) several years after insertion is a clinical problem, 1 " 3 and any noninvasive technique to diagnose degeneration before it becomes clinically obvious would be useful. One such technique is based on the dominant frequency of the heart sounds produced by prosthetic valves. 4 " 6 It has been demonstrated that the dominant frequency of bioprosthetic valve sounds increases as the duration of insertion increases, and this change was attributed to stiffening of the leaflets. 4-3 In vitro tests have shown that the leaflets stiffen when subjected to millions of cyclic stresses. 7 An increased frequency content of the heart sounds produced by degenerated porcine bioprosthetic valves also has been shown to occur if the valves are stiffened or calcified. 6 Whether a tear or perforation of the leaflets would modify the frequency content of the sounds has not been determined. This problem of whether tears or perforations of the valve modify the frequency content of the heart sounds can be approached by an analysis of valve vibration based on the evidence that the heart sounds are produced by vibration of the valvular cusps immediately after closure and that the frequency of sound relates to the frequency of vibration. 8 "" The effect on the frequency of valve vibration of a tear or perforation From the Henry Ford Heart and Vascular Institute, Detroit, Mich. Supported in part by National Institutes of Health grant HL Address for reprints: Paul D. Stein, MD, Henry Ford Hospital, 2799 West Grand Boulevard, Detroit, MI Received January 6, 1987; accepted May 20, of one of the leaflets, in combination with a stiffened and calcified valve, can be assessed by mathematical modeling of vibration of the valve leaflets. The purpose of this study was to estimate the fundamental vibrational frequency that occurs following closure of a porcine bioprosthetic valve when one or more leaflets are calcified, stiffened, and torn. The finite element technique was used. ' 2 The characteristics of valve vibration will lead to further understanding of the factors that affect the frequency content of heart sounds when the valves are abnormally stiffened, calcified, or torn. Hopefully, this information will lead to a more knowledgeable application of the frequency analysis of heart sounds for the detection of bioprosthetic valve degeneration. Materials and Methods Assumptions Several assumptions were made in the development of the numerical model to simplify the solution or because of lack of reliable experimental data. The following assumptions were made in our model: 1) the 3 PBV leaflets were assumed to be equal in size, symmetric, and uniformly thick (0.6 mm); 2) the stent posts were assumed to be rigid; 3) valve annulus deformation was neglected; 4) the leaflet tissue was considered to be isotropic; 5) a Poisson's ratio of was used for the tissue to achieve a nearly leaflet incompressibility condition; 6) an "effective" leaflet mass density was used to account for the fluid coupling (blood coupling) during vibration; and 7) the damping was assumed to be very small (<<1).

2 688 Circulation Research Vol 61, No 5, November 1987 FIGURE 1. Three-dimensional view of a Hancock porcine bioprosthetic valve. A, B, and C show the stent post (h, stem height). The Model The geometry of the closed PBV leaflets is complex even in the relaxed state, i.e., under zero pressure difference across the cusps. Based on our observations of various types of PBV, the geometry of a single leaflet approximated that of one half of an elliptic paraboloid. Details of geometric modeling of the closed leaflet have been reported previously. l3 The annular diameter of the valve used in the model was 27 mm, the height of the stent posts 19.0 mm, and the surface area of a single leaflet 5.77 cm 2. The geometry of the PBV, viewed from the side of the valve, is shown in Figure 1. The leaflets are attached to the support structure ABC. The assumption was made that the cloth and sutures, in addition to the stent itself, were rigid. Although this is not exactly true, the deformation of these components can reasonably be expected to be small, and this small deformation is neglected in the determination of vibrational frequencies. As the pressure gradient increases, the stresses in the leaflet increase because of an increase in pressure, and hence, the mechanical property (Young's modulus) of the tissue changes. 7 In addition to this change of mechanical properties (Young's modulus), the coaptation surface area also increases In this analysis, it was assumed that the coaptation surface area and the tissue Young's modulus were constant during vibration. Analysis The finite element technique involves discretizing the structure into small portions called "elements." These elements are assumed to be interconnected at discrete points called "nodes." These nodes are located along the boundary of each element. The governing mathematical relations for each element are established in terms of nodal variable parameters (displacements) by invoking the principles of structural mechanics. 12 This technique permits variation of mechanical properties of each element and allows an incorporation of complex geometries in the analysis. Assembly of each element-governing equation represents the overall mathematical model of the complete structure. After appropriate "boundary" conditions are applied, the overall governing equations are solved. The finite element discretization of all the three leaflets is shown in Figure 2. Triangular thin shell elements are used in the analysis. This discretization resulted in 420 nodal points that connected 828 elements. At each nodal point, three translations and three rotations accounted for the nodal displacements (degrees of freedom). With the damping effect neglected, the governing equations for free vibration of all 3 leaflets together are expressed in matrix form as follows 12 : [K] {8} - co 2 [M] {8} = {0} (1)

3 Hamid et al Bioprosthetic Heart Valve Vibration 689 A B Table 1. No Material Properties Young's modulus Material (MPa) Normal leaflet 0.30 Stiff leaflet 2.10 Calcium 21.0 Poisson's ratio Density (kn/m*) NO. OF EL.EMENTS-B2B NO. OF NODES -420 by blood. Since the "exact" mass density of the vibrating membrane supporting the column of blood is not yet established, an approximate "effective" mass density of 10.4 g/cm 3, which is 10 times that of tissue, was used in the analysis. Equation 1 was solved to obtain the frequency "a>" (eigen values) and the nodal displacements {8}, the eigen vectors." The fundamental vibrational frequencies of the leaflets were obtained for normal leaflets, simulated stiffened leaflets, leaflets with simulated focal calcium, and leaflets with a simulated central perforation and tear. For normal leaflets, the Young's modulus was assumed to be 0.3 MPa (mega Pascal), and this modulus value corresponds to the modulus at low stress values. 7 For the stiffened valve, the leaflet modulus was assumed to be 2.1 MPa. The Young's modulus for the calcified portion of the leaflet was assumed to be 21.0 MPa. The density and the Poisson's ratio used for these various conditions of the leaflet are given in Table 1. FIGURE 2. Top: Finite element discretization of the three leaflets. Triangular shell elements are used. A, B, andc arethe commissures. There are 828 elements connected by 420 nodal points. Bottom: Top view of the discretized leaflets. A, B, and C are commissures. where [K] is the stiffness matrix and [M] is the effective mass matrix of the leaflets. For the assumed closed leaflet geometry, the stiffness matrix is the function of Young's modulus of the leaflet, and the mass matrix is the function of mass density of the leaflet surrounded KORMAL VALVE f - 55 Hi 1 LEAFLET STIFF + C«+ f - 62 Hi 3 f LEAFLETS HI STIFF Results The results are summarized diagrammatically in Figure 3. NORMAL LEAFLETS. The fundamental vibrational frequency was calculated at 55 Hz for the normal valve. The modal shape of a single leaflet vibration is shown in Figure 4 because of symmetric conditions. NORMAL VALVE WITH PERFORATION OF ONE LEAFLET. A central perforation of 1.17 cm 2 area (20% of the single leaflet area) was simulated (Figure 5). The fundamental frequency was calculated as 52 Hz. 1 LEAFLET STIFF + CA + 3 LEAFLETS STIFF + Ca ++ f HI 3 LEAFLETS STIFF + C«++ TEAK III 1 LEAFLET t HI 1 LEAFLET PERFORATION t - 52 HI FIGURE 3. in Hz. Fundamental frequency of vibration of a bioprosthetic valve at normal and various degenerated conditions, f, Frequency

4 690 Circulation Research Vol 61, No 5, November 1987 FUNDAMENTAL FREQUENCY - 39 Hz FIGURE 6. A top view of a valve with normal leaflets with a tear (35% of a single leaflet area). FIGURE 4. Fundamental modal shape of a single normal leaflet. Solid line represents the undeformed boundaries. Broken line represents the shape of the leaflet during vibration. NORMAL VALVE WITH TEAR OF ONE LEAFLET. A tear that involved 35% of the area of a single leaflet was introduced in one leaflet (Figure 6). The frequency of vibration diminished somewhat to 39 Hz. ONE LEAFLET STIFFENED AND CALCIFIED (TWO LEAFLETS NOR- MAL). Focal calcium at the commissural attachments of one of the leaflets (17% of a single leaflet area) and stiffening of that leaflet had little effect on the frequency of vibration of the entire valve (Figure 7). The estimated fundamental frequency of the valve for these conditions was 62 Hz. ONE LEAFLET STIFFENED, CALCIFIED, AND TORN (TWO LEAF- LETS NORMAL). Focal calcification was added to the commissural attachments of one of the leaflets (17% of a single leaflet area), and the leaflet was assumed to be stiff. In addition to calcification and stiffening of one leaflet, a central tear that involved 35% of the area of the leaflet was included (Figure 8). The resultant natural frequency of the entire valve was 60 Hz. STIFFENED VALVE (ALL THREE LEAFLETS). The fundamental vibrational frequency of the stiffened valve was calculated as 145 Hz. STIFFENED VALVE WITH CALCIFICATION (ALL THREE LEAF- LETS). All three leaflets were caused to be stiff in the model, and focal calcifications were introduced at the commissural attachments of all the three leaflets (Figure 9). In this configuration, the stiffened and calcified valve showed a frequency of 176 Hz. STIFFENED AND CALCIFIED VALVE (ALL THREE LEAFLETS) WITH TEAR IN ONE LEAFLET. All three leaflets were modeled as stiffened and calcified at the commissural attachments. In addition, a tear that included 35% of a single leaflet NORMAL - NORMAL STIFF FIGURE 5. Top: Finite element discretization of valve leaflets with a 20% hole in one of the leaflets. Bottom: Top view of the leaflets with hole as viewed from the ventricular side. FUNDAMENTAL FREQUENCY = 62 Hz FIGURE 7. Valve with two leaflets normal and the third leaflet stiff and calcified (17% of the single leaflet area) at the commissural attachments.

5 Hamid et al Btoprosthetic Heart Valve Vibration 691 NORMAL STIFF FUNDAMENTAL FREQUENCY - 60 Hz FIGURE 8. Valve with one leaflet stiff, calcified, and torn. Two leaflets are normal. The calcification is at the commissural attachments and occupies 17% of the single leaflet. The tear occupies 35% of the single leaflet area. area was introduced in one leaflet (Figure 10). The frequency of vibration was 123 Hz. Discussion Numerical simulations of anatomic structures suffer in accuracy as a result of inherent assumptions necessary to obtain a solution. Some of the assumptions made in this study are readily justified, while others represent approximations. The three leaflets of a PBV are usually of different size and seldom symmetric. Lack of symmetry, however, does not affect the static response of the loaded valves, 1314 and this fact can also be extended to the dynamic analysis of the leaflets. The thickness of the leaflets is not uniform, and variability among specimens is possible. Human aortic valve leaflets are not uniformly thick and vary between the relaxed state and pressure stressed state. 16 Since topographic variation of leaflet thickness of a PBV has not yet been determined, a constant thickness of the leaflets was assumed in our model. The leaflet material is incompressible, i.e., the volume of the leaflet does not change during deformation. This incompressibility condition was achieved by approximating the Poisson's ratio (ratio of lateral strain to axial strain) as in our model. The leaflet material of a PBV is invariably anisotropic, and the stress-strain properties of PBV leaflets FUNDAMENTAL FREQUENCY Hz FIGURE 9. Valve with all three leaflets stiff and calcified. Calcification involves 17% of the total valve area. FUNDAMENTAL FREQUENCY Hz FIGURE 10. Valve with all three leaflets stiff and calcified and a tear in one of the leaflets. Calcification was at the commissural attachments and involved 17% of the total valve area. The tear involved 35% of the area of a single leaflet. differ in the circumferential and radial directions. 17 Nevertheless, the assumption of isotropy was made to simplify the problem. Also, even though PBV valves have flexible stents, 18 a rigid stent was assumed in this model. The effective density was assumed to be 10 times that of the leaflet mass density in our model. This approximation reduces the frequency values by a factor of 3.16 times that of the value in vacuum. This estimate of the effective mass density was based on the effects of a liquid in a fluid-filled structure on the vibrational frequency of the empty structure. The natural frequency of vibration of a fluid-filled cylindric shell is lower than the frequency of the cylinder alone." Also, the natural frequency of vibration of a fluid-filled tank is lower than the empty one. 20 These effects are due to the added mass of the fluid and solid-fluid coupling. The frequency of fluid-filled structures may be reduced to 33 to 50% of the frequency in a vacuum. On the basis of these observations, an effective mass density of 10 times the leaflet mass density was assumed, thereby decreasing the frequency to 32% of the frequency that would occur in a vacuum. The Young's modulus of the tissue was chosen as 300 kpa at the moment of closure. This value, which was based on test data of others, 7 depends on factors such as the pressure during mounting of the valve leaflets and chemical treatment of the tissue before mounting. The stress-strain curves of gluteraldehyde-preserved porcine leaflet specimens prior to and following accelerated fatigue cycling are given. 7 The initial modulus of the cycled tissue at 648 x 10 6 cycles and 808 x 10* cycles may vary from 1,500 kpa to 3,000 kpa. In our model, a value of 2,100 kpa was arbitrarily chosen for Young's modulus for a stiffened leaflet. This value is approximately 25% of the highest Young's modulus that would occur before the normal tissue ruptures. 13 The Young's modulus of the calcified tissue was assumed to be in the range of Young's modulus of limestone. Sites of high mechanical stresses in the leaflet were correlated to the sites of calcification and perforation. 21 " 23 The calcium sites were introduced in the

6 692 Circulation Research Vol 61, No 5, November 1987 commissural attachments at maximally stressed sites. Any real structure dissipates energy as it deforms during vibration. This energy dissipation is called damping. It is difficult, if not impossible, to measure the damping of the vibrating valve because of the environment in which the leaflets vibrate. The damping mechanism may include fluid viscosity, turbulence, acoustic radiation, and material damping. To analyze the mechanical response, one has to consider all of these phenomena. However, in practical cases, one or two mechanisms are generally predominant. 24 The major effects of damping are 1) reduction of vibration amplitude at resonance and 2) more rapid decay of free vibrations. In general, the damped free vibration frequency (wj) of any structure can be expressed in terms of the damping ratio (Q and the natural undamped frequency (coj as If the damping ratio is as high as 0.2, the frequency reduces by only 2%. In our computation, the effect of damping was neglected. The natural frequency of the normal valve that was computed in this analysis, 55 Hz, was comparable to the dominant frequency of the heart sounds produced by normal bioprosthetic valves (60 ±17 Hz). 25 An example of Fourier analysis of the first heart sound in a patient with a normally functioning porcine bioprosthetic valve inserted in the mitral position is shown in Figure 11. In view of the close relation between the dominant frequency of heart sounds and the frequency of vibration of the aortic valve, 8 " 10 the computed value is believed to be correct. The introduction of a central perforation in one of the leaflets had a negligible effect on the frequency, reducing it only from 55 to 52 Hz. Similarly, the introduction of tear that involved 35% of one of the normal leaflets caused a reduction of the frequency only from 59 to 39 Hz. The introduction in the mathematical model of stiffening and calcification of a single leaflet had a negligible effect on the frequency of vibration since the frequency increased only from 55 to 62 Hz. If the single stiffened and calcified leaflet became torn, the computer modeling also indicated that there would be a _49 Hz (2) negligible effect on total valve vibration (Figure 3). Thus, if the characteristics of only a single leaflet were modified, whether that leaflet was torn, perforated, stiffened, calcified, or calcified, stiffened, and torn, the frequency of the vibration of the valve as a whole changed little. In contrast to calculations related to modifications of a single leaflet, if all three leaflets were stiffened, then the frequency of vibration increased markedly from 55 to 145 Hz. If all three leaflets were both stiffened and calcified at the commissures, the frequency was even higher at 176 Hz. An example of the frequency spectrum of the first heart sound in a patient with a degenerated porcine bioprosthetic valve in the mitral position is shown in Figure 12. The valve showed stiffening and calcification of all three leaflets, and two of the leaflets were torn at the commissures. The dominant frequency was abnormally elevated at 189 Hz. In the mathematical model, if one of the stiffened and calcified leaflets was torn at the center of the cusp, then there was a prominent reduction of the frequency to 123 Hz, although the frequency did not fall to normal. This finding may explain occasional observations in patients in whom, over the course of years, the dominant frequency of heart sounds increases and then diminishes strikingly at the time that clinical evidence of degeneration is apparent. It may be that stiffening of the leaflet that is not recognized clinically is followed by a tear that lowers the sound frequency. Numerical techniques have been applied to valve vibration to correlate the heart sounds with the vibration of the valve cusps.'' 2 *' 28 These models assumed the leaflets have the geometry of either a circular membrane 2 * 27 or a one-dimensional idealized geometry. " 28 The models did not account for the actual geometry and degenerative conditions of the leaflets. A direct comparison of the results with these models is not possible because of the various assumptions and techniques involved in the analyses. However, the frequency of vibration obtained by our model is of the same order of magnitude as predicted by these models. A number of assumptions were made in our calculations. Mathematical modeling of biologic structures is complex because of a wide variation in the geometry and because of unknown tissue properties. Even FIGURE 11. Frequency spectrum of the first heart sound ' in a patient with a normal porcine bioprosthetic valve in the mitral position FREQUENCY (Hz)

7 Hamid et al Bioprosthetic Heart Valve Vibration Hz a 3 a. 2 1 > u It FIGURE 12. Frequency spectrum of the first heart sound in a patient with a degenerated porcine bioprosthetic valve in the mitral position. The valve was stiffened and calcified, and two of the leaflets were torn at the commissures. I I 150 I FREQUENCY (Hz) I 400 though some of these assumptions represent approximations, it is believed that these calculations are advanced over previous calculations in the literature and represent a significant step toward mathematical modeling of bioprosthetic heart valve vibration. Nevertheless, as with any mathematical model of biologic structures, the conclusions should be interpreted with caution. It should also be kept in mind that the numerical values of the estimated fundamental frequencies reflect the input data, including the valve geometry, material properties of the leaflet tissue, and the coupling of blood hemodynamics with the leaflets. In conclusion, an estimation of the fundamental frequency of vibration of the closed cusps of a porcine bioprosthetic valve, using a numerical model with the finite element technique, indicates that 1) the dominant frequency of vibration of the valve changes little with tears, calcification, or stiffening of a single leaflet; 2) stiffening of all three leaflets causes a prominent increase of frequency, and the addition of calcification increases the frequency even more; and 3) if a stiffened valve becomes torn, then the frequency diminishes. These observations may be applicable to the evaluation of the frequency content of heart sounds in patients with PBV. References 1. Magilligan DJ Jr, Lewis JW Jr, Jara FM, Lee MW, Alam M, Riddle JM, Stein PD: Spontaneous degeneration of porcine bioprosthetic valves. Ann Thorac Surg 198O;3O: Lakier JB, Khaja F, Magilligan DJ Jr, Goldstein S: Porcine xenograft valves: Long-term (60-89 month) follow-up. Circulation 1980;62: Ferrans VJ, Spray TL, BillinghamME, Roberts WC: Structural changes in glutaraldehyde-treated porcine heterografts used as substitute cardiac valves. Am J Cardiol 1978;41: Stein PD, Sabbah HN, Lakier JB, Goldstein S: Frequency spectrum of the aortic component of the second sound in patients with normal valves, aortic stenosis, and aortic porcine xenografts: Potential for the detection of procine xenograft degeneration. Am J Cardiol 1980;46: Stein PD, Sabbah HN, Lakier JB, Magilligan DJ Jr, Goldstein S: Frequency of the first heart sound in the assessment of stiffening of mitral bioprosthetic valves. Circulation 1981; 63: Stein PD, Sabbah HN, Lakier JB, Kemp SR, Magilligan DJ Jr: Frequency spectra of the first heart sound and of the aortic component of the second heart sound in patients with degenerated porcine bioprosthetic valves. Am J Cardiol 1984; 53: Broom ND: Fatigue-induced damage in gluteraldehyde-preserved heart valve tissue. J Thorac Cardiovasc Surg 1978;76: Stein PD: A Physical and Physiological Basis for the Interpretation of Cardiac Auscultation: Evaluations Based Primarily on the Second Sound and Ejection Murmurs. Mount Kisco, NY, Future, 1981, pp Sabbah HN, Stein PD: Investigation of the theory and mechanism of the origin of the second heart sound. Circ Res 1976;39: Sabbah HN, Stein PD: Relation of the second sound to diastolic vibration of the closed aortic valve. Am J Physiol (Heart and Circ Physiol) 1978;3:H696-H Blick EF, Sabbah HN, Stein PD: One-dimensional model of diastolic semilunar valve vibrations productive of heart sounds. JBiomech 1979;12: Zienkiewicz OC: The Finite Element Method. Berkshire, England, McGraw-Hill, Hamid MS, Sabbah HN, Stein PD: Influence of stem height upon stresses on the cusps of closed bioprosthetic valves. J Biomech 1986;19: Christie GW, Medland IC: A non-linear finite element stress analysis of porcine bioprosthetic heart valves, in Gallagher RH, Simon BR, Johnson PC, Gross JF (eds): Finite Elements in Biomechanics. New York, Wiley & Sons, 1982, pp Anderson WJ: MSC/NASTRAN Interactive Training Program. New York, Wiley & Sons, Clark RE, Finke EH: Scanning and light microscopy of human aortic leaflets in stressed and relaxed states. J Thorac Cardiovasc Surg 1974;67: Thubrikar M, Piepgrass WC, Bosher LP, Nolan SP: The elastic modulus of canine aortic valve leaflets in vivo and in vitro. Circ Res 198O;47:792-8O0 18. Reis RL, Hancock WD, Yarbrough JW, Glancy DL, Morrow AG: The flexible stent, a new concept in the fabrication of tissue heart valve prostheses. J Thorac Cardiovasc Surg 1971; 62: Yu IW: Subspace iteration for eigen-solution of fluid structure interaction problems. Fluid-structure Dynamics, Proc 1985 Pressure Vessels and Piping Conference, ASME, 1985, PVP98-7: Brusa L, Ciacci R, Casirati M, Fanelli M, Giuseppetti G: Dynamic analysis of fluid-structure coupled systems: Experimental and numerical comparisons. Fluid-structure Dynamics, Proc 1985 Pressure Vessels and Piping Conference, ASME, 1985, PVP98-7: Sabbah HN, Hamid MS, Stein PD: Estimation of mechanical stresses on closed cusps of porcine bioprosthetic valves: Effects of stiffening, focal calcium and focal thinning. Am J Cardiol 1985;55:

8 694 Circulation Research Vol 61, No 5, November Ferrans VJ, Boyce SW, Billingham ME, Jones M, Ishihara T, Roberts WC: Calcific deposits in porcine bioprostheses: Structure and pathogenesis. Am J Cardiol 1980;46: Deck JD, Thubrikar MJ, Nolan SP, Aouad J: Role of mechanical stress in calcification of bioprostheses, in Cohn LH, Gallucci V (eds): Cardiac Bioprostheses. New York, Yorke Medical Books, 1982, pp Ungar EE: Damping of panels, in Beranek LL (ed), Noise and Vibration Control. New York, McGraw-Hill Book Company, 1971, pp Stein PD, Sabbah HN, Lakier JB, Magilligan DJ Jr, Kemp SR, Alam M, Rosman H, Lewis J: Valve assessment by sound (abstract). Proc Devices and Technology Branch Contractors Meeting, National Heart, Lung and Blood Institute, Bethesda, MD, December 16-18, 1985, p Lim KO, Liew YC, Oh CH: Analysis of mitral and aortic valve vibrations and their role in the production of first and second heart sounds. Phys Med Biol 1980;25: Mazumdar J, Hearn TC: A mathematical analysis of mitral valve leaflets. J Biomech 1978;11: Sikarskie DL, Stein PD, Vable M: A mathematical model of aortic valvular vibration. J Biomech 1984;I7: KEY WORDS heart valves heart sounds frequency analysis vibrational analysis finite element modeling bioprosthetic valves

Relation of Calcification to Tom Leaflets

Relation of Calcification to Tom Leaflets Relation of Calcification to Tom Leaflets of Spontaneously Degenerated Porcine Bioprosthetic Valves Paul D. Stein, M.D., Stephen R. Kemp, Ph.D., Jeanne M. Riddle, Ph.D., Min W. Lee, M.D., Joseph W. Lewis,

More information

Geometrical Stress-Reducing Factors in the Anisotropic Porcine Heart Valves

Geometrical Stress-Reducing Factors in the Anisotropic Porcine Heart Valves X. Y. Luo* Department of Mechanical Engineering University of Sheffield Sheffield, S1 3JD, UK W. G. Li Hydraulic Machinery Division Gansu University of Technology 730050, Lanzhou, P.R. China J. Li School

More information

Numerical Simulation Transcatheter Aortic Valve Implantation and Mechanics of Valve Function

Numerical Simulation Transcatheter Aortic Valve Implantation and Mechanics of Valve Function Numerical Simulation Transcatheter Aortic Valve Implantation and Mechanics of Valve Function M.S. Hamid, Ph.D. Advanced Computational Systems LLC Rochester, MI 48306 acssah027@gmail.com Abstract As the

More information

Degeneration and Cacfi cation of Bioprosthetic Cardiac Valves

Degeneration and Cacfi cation of Bioprosthetic Cardiac Valves ANIMAL MODEL OF HUMAN DISEASE Degeneration and Cacfi cation of Bioprosthetic Cardiac Valves Bioprosthetic Tricuspid Valve Implantation in Sheep GLENN R. BARNHART, MD, MICHAEL JONES, MD, TOKUHIRO ISHIHARA,

More information

Stress Sharing Between the Sinus and Leaflets of Canine Aortic Valve

Stress Sharing Between the Sinus and Leaflets of Canine Aortic Valve Stress Sharing Between the Sinus and Leaflets of Canine Aortic Valve Mano J. Thubrikar, Ph.D., Stanton P. Nolan, M.D., Jaafar Aouad, D.V.M., and J. David Deck, M.D. ABSTRACT A knowledge of the behavior

More information

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction Eann A. Patterson Department of Mechanical Engineering, The University of Sheffield Mappin Street, Sheffield, S1 3JD

More information

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture RAMON BERGUER, a,b JOSEPH L. BULL, a,b AND KHALIL KHANAFER a a Vascular Mechanics Laboratory, Department of Biomedical Engineering,

More information

Finite element modeling of the thoracic aorta: including aortic root motion to evaluate the risk of aortic dissection

Finite element modeling of the thoracic aorta: including aortic root motion to evaluate the risk of aortic dissection Journal of Medical Engineering & Technology, Vol. 32, No. 2, March/April 2008, 167 170 Short Communication Finite element modeling of the thoracic aorta: including aortic root motion to evaluate the risk

More information

An effort is made to analyse the stresses experienced by the human femur. In order

An effort is made to analyse the stresses experienced by the human femur. In order Finite Element Analysis of Human Femur Bone U N Mughal 1, H A Khawaja 2*, M Moatamedi 1, M Souli 3 1. Narvik University College, Norway 2. UiT-The Arctic University of Norway, Norway 3. University of Lille,

More information

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System Contents 1 Computational Haemodynamics An Introduction... 1 1.1 What is Computational Haemodynamics (CHD)... 1 1.2 Advantages of CHD... 3 1.3 Applications in the Cardiovascular System... 4 1.3.1 CHD as

More information

Finite Element Modeling of the Mitral Valve and Mitral Valve Repair

Finite Element Modeling of the Mitral Valve and Mitral Valve Repair Finite Element Modeling of the Mitral Valve and Mitral Valve Repair Iain Baxter A thesis submitted to the Faculty of Graduate and Postdoctoral Studies in partial fulfillment of the requirements for the

More information

Late failure of transcatheter heart valves: An open question

Late failure of transcatheter heart valves: An open question Late failure of transcatheter heart valves: An open question A comparison with surgically implanted bioprosthetic heart valves. A. Rashid The Cardiothoracic Centre Liverpool, UK. Conflict of Interest Statement

More information

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases Proceedings of the 215 2nd International Symposium on Physics and Technology of Sensors, 8-1th March, 215, Pune, India Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

More information

Original. Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging

Original. Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging Original Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging P. Khamdaengyodtai 1, T. Khamdaeng 1, P. Sakulchangsatjatai 1, N. Kammuang-lue

More information

Primary Tissue Valve Degeneration in Glutaraldehvde-Preserved Porcine Biomostheses: Hancock I Vekus Carpentier-Edwards at 4- to 7-Years Follow-up

Primary Tissue Valve Degeneration in Glutaraldehvde-Preserved Porcine Biomostheses: Hancock I Vekus Carpentier-Edwards at 4- to 7-Years Follow-up Primary Tissue Valve Degeneration in Glutaraldehvde-Preserved Porcine Biomostheses: A Hancock I Vekus Edwards at 4- to 7-Years Follow-up Francisco Nistal, M.D., Edurne Artifiano, M.D., and Ignacio Gallo,

More information

Stretching Cardiac Myocytes: A Finite Element Model of Cardiac Tissue

Stretching Cardiac Myocytes: A Finite Element Model of Cardiac Tissue Megan McCain ES240 FEM Final Project December 19, 2006 Stretching Cardiac Myocytes: A Finite Element Model of Cardiac Tissue Cardiac myocytes are the cells that constitute the working muscle of the heart.

More information

I will not discuss off label use or investigational use in my presentation.

I will not discuss off label use or investigational use in my presentation. I will not discuss off label use or investigational use in my presentation. Surgical valves Design and Durability Testing Potential Concerns Real Practice 1952-1962 1963-1966 1967-1969 1969-1977 1977-1984

More information

The University of Sydney Slide 1

The University of Sydney Slide 1 The University of Sydney Slide 1 SIMULATION DRIVEN BIOMEDICAL DESIGN Lecture 4 Presented by Dr Paul Wong AMME4981/9981 Semester 1, 2016 The University of Sydney Slide 2 Simulation Types There is more to

More information

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 6, Issue 12 (May 2013), PP. 119-123 A Mathematical Model to Calculate Contact Stresses

More information

Dvnamic Behavior of Prosthetic Aortic Tissue Vdves as Viewed by High-speed Cinematography

Dvnamic Behavior of Prosthetic Aortic Tissue Vdves as Viewed by High-speed Cinematography Dvnamic Behavior of Prosthetic Aortic Tissue Vdves as Viewed by High-speed Cinematography W. Gerald Rainer, M.D., Robert A. Christopher, Ph.D., Theodore R. Sadler, Jr., M.D., and Alan D. Hilgenberg, M.D.

More information

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Excerpt from the Proceedings of the COMSOL Conference 2010 Paris (COMSOL Conference) Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Vaidehi

More information

The use of LS-DYNA fluid-structure interaction to simulate fluid-driven deformation in the aortic valve

The use of LS-DYNA fluid-structure interaction to simulate fluid-driven deformation in the aortic valve 4 th European LS-DYNA Users Conference Aerospace / Fluid-Struct. Inter. The use of LS-DYNA fluid-structure interaction to simulate fluid-driven deformation in the aortic valve Authors: Chris Carmody*,

More information

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents.

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents. Blucher Mechanical Engineering Proceedings May 2014, vol. 1, num. 1 www.proceedings.blucher.com.br/evento/10wccm AN ANALYSIS OF THE CONTACT BETWEEN THE STENT AND THE ARTERY USING TUBE HIDROFORMING SIMULATION

More information

Biomechanical implications of congenitally bicuspid aortic valves: a finite element analysis of patient-specific geometry

Biomechanical implications of congenitally bicuspid aortic valves: a finite element analysis of patient-specific geometry University of Iowa Iowa Research Online Theses and Dissertations Spring 2013 Biomechanical implications of congenitally bicuspid aortic valves: a finite element analysis of patient-specific geometry Paul

More information

Figure 1 The Human Heart

Figure 1 The Human Heart ARTIFICIAL HEART VALVE MODEL INTRODUCTION The heart is a muscular pump with four chambers and four heart valves. The upper chambers, the right atrium and left, are thin walled filling chambers. Blood flows

More information

A novel mathematical technique to assess of the mitral valve dynamics based on echocardiography

A novel mathematical technique to assess of the mitral valve dynamics based on echocardiography ORIGINAL ARTICLE A novel mathematical technique to assess of the mitral valve dynamics based on echocardiography Mersedeh Karvandi 1 MD, Saeed Ranjbar 2,* Ph.D, Seyed Ahmad Hassantash MD 2, Mahnoosh Foroughi

More information

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties Lin Chen, Shen Jingfeng & Chen Bing School of mechanical engineering, University of Shanghai for Science & Technology,

More information

Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease

Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease TIRONE E. DAVID, MD ; SEMIN THORAC CARDIOVASC SURG 19:116-120c 2007 ELSEVIER INC. PRESENTED BY INTERN 許士盟 Mitral valve

More information

by Age Groups W. R. E. Jamieson, M.D., L. J. Rosado, M.D., A. I. Munro, M.D., A. N. Gerein, M.D., L. H. Burr, M.D., R. T. Miyagishima, M.D.

by Age Groups W. R. E. Jamieson, M.D., L. J. Rosado, M.D., A. I. Munro, M.D., A. N. Gerein, M.D., L. H. Burr, M.D., R. T. Miyagishima, M.D. Carpentier-Edwards Standard Porcine Bioprosthesis: Primary Tissue Failure (Structural Valve Deterioration) by Age Groups W. R. E. Jamieson, M.D., L. J. Rosado, M.D., A. I. Munro, M.D., A. N. Gerein, M.D.,

More information

The Bicuspid AV Surgical Conisiderations

The Bicuspid AV Surgical Conisiderations The Bicuspid AV Surgical Conisiderations Ehud Raanani, MD Cardiothoracic Surgery, Sheba Medical Center Sackler School of Medicine, Tel Aviv University MAY 15, 2014 Homburg BAV Repair Congenital variations

More information

2. The heart sounds are produced by a summed series of mechanical events, as follows:

2. The heart sounds are produced by a summed series of mechanical events, as follows: Heart Sounds. Phonocardiography 1 Objectives 1. Phonocardiography - Definition 2. What produces the heart sounds 3. Where to listen for the heart sounds 4. How to record a phonocardiogram 5. Normal heart

More information

Introduction to soft tissues

Introduction to soft tissues Modelli Costitutivi dei Materiali Corso di Laurea in Ingegneria Biomedica Pavia, 2013 Introduction to soft tissues Ferdinando Auricchio 1 2 3 4 1 Dipartimento di Ingegneria Civile e Architettura, Università

More information

Functional anatomy of the aortic root. ΔΡΟΣΟΣ ΓΕΩΡΓΙΟΣ Διεσθσνηής Καρδιοθωρακοτειροσργικής Κλινικής Γ.Ν. «Γ. Παπανικολάοσ» Θεζζαλονίκη

Functional anatomy of the aortic root. ΔΡΟΣΟΣ ΓΕΩΡΓΙΟΣ Διεσθσνηής Καρδιοθωρακοτειροσργικής Κλινικής Γ.Ν. «Γ. Παπανικολάοσ» Θεζζαλονίκη Functional anatomy of the aortic root ΔΡΟΣΟΣ ΓΕΩΡΓΙΟΣ Διεσθσνηής Καρδιοθωρακοτειροσργικής Κλινικής Γ.Ν. «Γ. Παπανικολάοσ» Θεζζαλονίκη What is the aortic root? represents the outflow tract from the LV provides

More information

Vascular Mechanobiology: growth and remodeling in the aorta in health and disease

Vascular Mechanobiology: growth and remodeling in the aorta in health and disease Vascular Mechanobiology: growth and remodeling in the aorta in health and disease Dr.-Ing. Christian J. Cyron Technical University of Munich funded by the German Research Foundation (Emmy-Noether Grant

More information

PPM: How to fit a big valve in a small heart

PPM: How to fit a big valve in a small heart PPM: How to fit a big valve in a small heart Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC King Abdulaziz Cardiac Centre National Guard Health Affairs Riyadh, Saudi Arabia GHA meeting Muscat

More information

WHAT CAN WE LEARN FROM COMPUTER SIMULATION?

WHAT CAN WE LEARN FROM COMPUTER SIMULATION? WHAT CAN WE LEARN FROM COMPUTER SIMULATION? Ehud Raanani, MD Gil Marom, Phd Cardiothoracic Surgery, Sheba Medical Center Sackler School of Medicine, Biomechanical Engineering, Tel Aviv University Homburg,

More information

HEART MURMURS: DECIPHERING THEIR CAUSE AND SIGNIFICANCE

HEART MURMURS: DECIPHERING THEIR CAUSE AND SIGNIFICANCE Vet Times The website for the veterinary profession https://www.vettimes.co.uk HEART MURMURS: DECIPHERING THEIR CAUSE AND SIGNIFICANCE Author : Pedro Oliveira Categories : Vets Date : May 27, 2013 PEDRO

More information

The aortic valve has been extensively studied over the

The aortic valve has been extensively studied over the Dynamic Analysis of the Aortic Valve Using a Finite Element Model Ramakrishna Gnyaneshwar, BTech, Ramarathnam Krishna Kumar, PhD, and Komarakshi R. Balakrishnan, MBBS, MS Department of Mechanical Engineering,

More information

Oil Transmission Pipelines Condition Monitoring Using Wavelet Analysis and Ultrasonic Techniques

Oil Transmission Pipelines Condition Monitoring Using Wavelet Analysis and Ultrasonic Techniques Engineering, 2013, 5, 551-555 doi:10.4236/eng.2013.56066 Published Online June 2013 (http://www.scirp.org/journal/eng) Oil Transmission Pipelines Condition Monitoring Using Wavelet Analysis and Ultrasonic

More information

A Two-layered Model for the Analysis of Arterial Rheology

A Two-layered Model for the Analysis of Arterial Rheology IJCSIT International Journal of Computer Science and Information Technology, Vol., o., June, pp. 37- A Two-layered Model for the Analysis of Arterial Rheology Sapna Singh Department of Mathematics, Harcourt

More information

EFFECTS OF SPATIAL VARIATION OF SEISMIC INPUTS ON BRIDGE LONGITUDINAL RESPONSE

EFFECTS OF SPATIAL VARIATION OF SEISMIC INPUTS ON BRIDGE LONGITUDINAL RESPONSE 13 th World Conference on Earthquake Engineering Vancouver, B.C., Canada August 1-6, 24 Paper No. 64 EFFECTS OF SPATIAL VARIATION OF SEISMIC INPUTS ON BRIDGE LONGITUDINAL RESPONSE Jiachen WANG 1, Athol

More information

Possible link with isolated aortic stenosis

Possible link with isolated aortic stenosis British Heart Journal, 1977, 39, 1006-1011 Minor congenital variations of cusp size in tricuspid aortic valves Possible link with isolated aortic stenosis F. E. M. G. VOLLEBERGH AND A. E. BECKER From the

More information

Numerical analysis of the embedded abutments of integral bridges

Numerical analysis of the embedded abutments of integral bridges Numerical analysis of the embedded abutments of integral bridges Ming XU Researcher, Dept. of Civil & Env. Engrg., University of Southampton, UK Summary Alan G. BLOODWORTH Lecturer, Dept. of Civil & Env.

More information

The Bicuspid AV Surgical Considerations

The Bicuspid AV Surgical Considerations The Bicuspid AV Surgical Considerations Ehud Raanani, MD Cardiothoracic Surgery, Sheba Medical Center Sackler School of Medicine, Tel Aviv University September 12, 2014 Homburg BAV Repair Congenital variations

More information

Oscillating Pressure Experiments on Porcine Aorta

Oscillating Pressure Experiments on Porcine Aorta Oscillating Pressure Experiments on Porcine Aorta V. V. Romanov, S. Assari, and K. Darvish Tissue Biomechanics Lab, College of Engineering, Temple University ABSTRACT This paper addresses the problem of

More information

Computational Simulation of Penetrating Trauma in Biological Soft Tissues using the Material Point Method

Computational Simulation of Penetrating Trauma in Biological Soft Tissues using the Material Point Method Computational Simulation of Penetrating Trauma in Biological Soft Tissues using the Material Point Method Irina IONESCU* +, James GUILKEY* ++, Martin BERZINS*, Robert M. KIRBY*, Jeffrey WEISS* + *Scientific

More information

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor Cardiothoracic Radiology Disclosure I have no disclosure pertinent to this presentation.

More information

Image Analysis and Cytometry in Three-Dimensional Digital Reconstruction of Porcine Native Aortic Valve Leaflets

Image Analysis and Cytometry in Three-Dimensional Digital Reconstruction of Porcine Native Aortic Valve Leaflets Image Analysis and Cytometry in Three-Dimensional Digital Reconstruction of Porcine Native Aortic Valve Leaflets Introduction Chi Zheng, M1 - University of Pittsburgh; BSE - University of Michigan In association

More information

Simulation of Aortic Valve Replacement Surgery

Simulation of Aortic Valve Replacement Surgery SMGr up Simulation of Aortic Valve Replacement Surgery Hadi Mohammadi 1,2,3 *and Guy Fradet 1,2 1 The Heart Valve Performance Laboratory, University of British Columbia, Canada 2 Department of Surgery,

More information

The Cardiac Cycle Clive M. Baumgarten, Ph.D.

The Cardiac Cycle Clive M. Baumgarten, Ph.D. The Cardiac Cycle Clive M. Baumgarten, Ph.D. OBJECTIVES: 1. Describe periods comprising cardiac cycle and events within each period 2. Describe the temporal relationships between pressure, blood flow,

More information

Mechanisms of heart failure with normal EF Arterial stiffness and ventricular-arterial coupling. What is the pathophysiology at presentation?

Mechanisms of heart failure with normal EF Arterial stiffness and ventricular-arterial coupling. What is the pathophysiology at presentation? Mechanisms of heart failure with normal EF Arterial stiffness and ventricular-arterial coupling What is the pathophysiology at presentation? Ventricular-arterial coupling elastance Central arterial pressure

More information

Mechanical Properties and Active Remodeling of Blood Vessels. Systemic Arterial Tree. Elastic Artery Structure

Mechanical Properties and Active Remodeling of Blood Vessels. Systemic Arterial Tree. Elastic Artery Structure Mechanical Properties and Active Remodeling of Blood Vessels Gross anatomy of systemic and pulmonary circulation Microscopic structure Mechanical properties and testing Residual stress Remodeling Systemic

More information

P have been used for mitral and aortic valve replacement

P have been used for mitral and aortic valve replacement A -Year Comparison of Mitral Valve Replacement With Carpentier-Edwards and Hancock Porcine Bioprostheses P. Perier, MD, A. Deloche, MD, S. Chauvaud, MD, J. C. Chachques, MD, J. Relland, MD, J. N. Fabiani,

More information

By: Stephanie Bendtsen, Joseph Calderan, and Celeste Dupont Team 17 Client: Dr. Sun

By: Stephanie Bendtsen, Joseph Calderan, and Celeste Dupont Team 17 Client: Dr. Sun By: Stephanie Bendtsen, Joseph Calderan, and Celeste Dupont Team 17 Client: Dr. Sun Flow Loop Design Introduction Purpose: to build a pulsatile flow loop that allows for implementation and testing of mechanical

More information

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 Presented at the COMSOL Conference 2010 India ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 1 DEPARTMENT OF MECHANICAL ENGINEERING & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING

More information

Mechanical Properties and Active Remodeling of Blood Vessels. Blood Vessels

Mechanical Properties and Active Remodeling of Blood Vessels. Blood Vessels Mechanical Properties and Active Remodeling of Blood Vessels Gross anatomy of systemic and pulmonary circulation Microscopic structure Mechanical properties and testing Residual stress Remodeling Blood

More information

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries Copyright 2015 Tech Science Press MCB, vol.12, no.1, pp.37-47, 2015 CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries P. Jhunjhunwala,, P.M. Padole, and S.B. Thombre, Abstract:

More information

Anatomy of aortic valve and root

Anatomy of aortic valve and root Anatomy of aortic valve and root Emmanuel Lansac, Isabelle Di Centa Cardiac Surgery Institut Mutualiste Montsouris, Paris, France The aortic valve : a passive or dynamic structure? Leonardo da Vinci 1508

More information

4D model of hemodynamics in the abdominal aorta

4D model of hemodynamics in the abdominal aorta Bio-Medical Materials and Engineering 26 (2015) S257 S264 DOI 10.3233/BME-151312 IOS Press S257 4D model of hemodynamics in the abdominal aorta Ireneusz Zbicinski a,*, Natalia Veshkina a and Ludomir Stefa

More information

Finite Element Analysis of the Human Middle Ear and an Application for Clinics for Tympanoplasty (Static and Harmonic Vibration Analysis)

Finite Element Analysis of the Human Middle Ear and an Application for Clinics for Tympanoplasty (Static and Harmonic Vibration Analysis) 2015 American Transactions on Engineering & Applied Sciences. American Transactions on Engineering & Applied Sciences http://tuengr.com/ateas Finite Element Analysis of the Human Middle Ear and an Application

More information

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery Proceedings of the 3 rd International Conference on Fluid Flow, Heat and Mass Transfer (FFHMT 16) Ottawa, Canada May 2 3, 2016 Paper No. 170 Numerical Simulation of Blood Flow through Asymmetric and Symmetric

More information

Dynamic Role of the Cardiac Jelly

Dynamic Role of the Cardiac Jelly 61 Chapter 6 Dynamic Role of the Cardiac Jelly Before looping, when the embryonic heart is still a straight tube, the cardiac jelly occupies the bulk of the heart tube walls. Despite its preeminence in

More information

Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A Three-Dimensional Finite Element Analysis

Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A Three-Dimensional Finite Element Analysis J Indian Prosthodont Soc (Oct-Dec 2013) 13(4):555 559 DOI 10.1007/s13191-012-0189-7 ORIGINAL ARTICLE Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A

More information

Analysis of the effects of plaque deposits on the blood flow through human artery

Analysis of the effects of plaque deposits on the blood flow through human artery ISSN 2395-1621 Analysis of the effects of plaque deposits on the blood flow through human artery #1 Sajid S. Mulani, #2 Pankaj I. Jagad 1 sajidsmulani21@gmail.com 2 pjjagad.scoe@sinhgad.edu #12 Department

More information

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics.

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics. In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics. Chandrasenan.P 1, Vishnu.G 2, Akshay K Nair 3 1M Tech student, Department of Mechanical

More information

ICE: Echo Core Lab-CRF

ICE: Echo Core Lab-CRF APPENDIX 1 ICE: Echo Core Lab-CRF Study #: - Pt Initials: 1. Date of study: / / D D M M M Y Y Y Y 2. Type of Study: TTE TEE 3. Quality of Study: Poor Moderate Excellent Ejection Fraction 4. Ejection Fraction

More information

Experimental Prediction of Contact Area in Hip Replacement and Hemi- Arthroplasty

Experimental Prediction of Contact Area in Hip Replacement and Hemi- Arthroplasty Experimental Prediction of Contact Area in Hip Replacement and Hemi- Arthroplasty Qianqian Wang, John Fisher, Sophie Williams. Institute of Medical and Biological Engineering, School of Mechanical Engineering,

More information

Patient-Specific Computational Simulation of the Mitral Valve Function Using Three-Dimensional Echocardiography

Patient-Specific Computational Simulation of the Mitral Valve Function Using Three-Dimensional Echocardiography Patient-Specific Computational Simulation of the Mitral Valve Function Using Three-Dimensional Echocardiography Y. Rim 1, S. T. Laing 1, P. Kee 1, K. B. Chandran 2, D. D. McPherson 1 and H. Kim 1 1 Department

More information

Modelling of stress-strain states in arteries as a pre-requisite for damage prediction

Modelling of stress-strain states in arteries as a pre-requisite for damage prediction 2002 WT Press, Ashurst Lodge, Southampton, SO40 7AA, UK. All rights reserved. Paper from: Damage and Fracture Mechanics V, CA Brebbia, & S Nishida (Editors). SBN 1-85312-926-7 Modelling of stress-strain

More information

Anatomy of aortic valve and root Emmanuel Lansac MD PhD

Anatomy of aortic valve and root Emmanuel Lansac MD PhD Anatomy of aortic valve and root Emmanuel Lansac MD PhD Cardiac Surgery Institut Mutualiste Montsouris, Paris, France The aortic valve : a passive or dynamic structure? Leonardo da Vinci 1508 Quadr Anat

More information

PREDICTION OF BLOOD FLOW VELOCITY AND LEAFLET DEFORMATION VIA 2D MITRAL VALVE MODEL

PREDICTION OF BLOOD FLOW VELOCITY AND LEAFLET DEFORMATION VIA 2D MITRAL VALVE MODEL Journal of Mechanical Engineering and Sciences (JMES) e-issn: 2231-8380; Volume 2, pp. 217-225, June 2012 FKM, Universiti Malaysia Pahang PREDICTION OF BLOOD FLOW VELOCITY AND LEAFLET DEFORMATION VIA 2D

More information

Numerical Analysis of Coronary Stent for Diverse Materials

Numerical Analysis of Coronary Stent for Diverse Materials Numerical Analysis of Coronary Stent for Diverse Materials Vasantha Kumar Assistant Professor Department of Mechanical Engineering Bearys Institute of Technology Mangalore, Karnataka, India Abstract Coronary

More information

Update on Seismic Behavior and Design of

Update on Seismic Behavior and Design of Update on Seismic Behavior and Design of Steel Plate Girder Bid Bridges Ahmad M. Itani, Ph.D., P.E., S.E. Professor Department of Civil and Environmental Engineering University of Nevada, Reno Background

More information

FSI within Aortic Arch Model over Cardiac Cycle and Influence of Wall Stiffness on Wall Stress in Layered Wall

FSI within Aortic Arch Model over Cardiac Cycle and Influence of Wall Stiffness on Wall Stress in Layered Wall Engineering Letters, 13:2, EL_13_2_15 (Advance online publication: 4 August 2006) FSI within Aortic Arch Model over Cardiac Cycle and Influence of Wall Stiffness on Wall Stress in Layered Wall F. Gao and

More information

The stentless bioprosthesis has many salient features that

The stentless bioprosthesis has many salient features that Aortic Valve Replacement with the Medtronic Freestyle Xenograft Using the Subcoronary Implantation Technique D. Michael Deeb, MD The stentless bioprosthesis has many salient features that make it an attractive

More information

CYCLIC TESTING OF REINFORCED CONCRETE COLUMNS WITH UNBONDED REINFORCEMENT

CYCLIC TESTING OF REINFORCED CONCRETE COLUMNS WITH UNBONDED REINFORCEMENT CYCLIC TESTING OF REINFORCED CONCRETE COLUMNS WITH UNBONDED REINFORCEMENT Keerthi RANASINGHE* 1, Hiroshi MUTSUYOSHI* 2 and Hiroki UCHIBORI* 3 ABSTRACT: This paper describes the seismic behavior of RC columns

More information

PATIENT BOOKLET MEDTRONIC SURGICAL VALVE REPLACEMENT. Tissue Valve for Aortic and Mitral Valve Replacement

PATIENT BOOKLET MEDTRONIC SURGICAL VALVE REPLACEMENT. Tissue Valve for Aortic and Mitral Valve Replacement PATIENT BOOKLET MEDTRONIC SURGICAL VALVE REPLACEMENT Tissue Valve for Aortic and Mitral Valve Replacement ARE MEDTRONIC SURGICAL TISSUE HEART VALVES RIGHT FOR YOU? Medtronic surgical heart valves are for

More information

Echo in Asymptomatic Mitral and Aortic Regurgitation

Echo in Asymptomatic Mitral and Aortic Regurgitation 2017 ASE Florida Orlando, FL October 9, 2017 10:40 11:00 PM 20 min Grand Harbor Ballroom South Echo in Asymptomatic Mitral and Aortic Regurgitation Muhamed Sarić MD, PhD, MPA Director of Noninvasive Cardiology

More information

Comparative study of the mechanical behavior of the superior thoracic artery and abdominal arteries using the finite elements method

Comparative study of the mechanical behavior of the superior thoracic artery and abdominal arteries using the finite elements method J. Biomedical Science and Engineering, 2012, 5, 52-57 JBiSE http://dx.doi.org/10.4236/jbise.2012.52008 Published Online February 2012 (http://www.scirp.org/journal/jbise/) Comparative study of the mechanical

More information

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm I. Husain, C. Langdon and J. Schwark Department of Mathematics Luther College University of Regina Regina, Saskatchewan

More information

FRACTION IN THE HUMAN LEFT VENTRICLE

FRACTION IN THE HUMAN LEFT VENTRICLE FIBER ORIENTATION AND EJECTION FRACTION IN THE HUMAN LEFT VENTRICLE EDWARD A. SALLIN From the Department of Bionmathematics, Birmingham, Alabama 35233 University of Alabama Medical Center, ABSTRACT Fiber

More information

Mechanical Aspects of an Interference Screw Placement in ACL Reconstruction

Mechanical Aspects of an Interference Screw Placement in ACL Reconstruction Mechanical Aspects of an Interference Screw Placement in ACL Reconstruction Mahmoud Chizari 1, Mohammad Alrashidi 2, Khaled Alrashdan 2, Ibrahim Yildiz 3, Jamaluddin Mahmud 4 1 School of Engineering and

More information

A practical set-up to create an adjustable insufficiency in a porcine aortic root.

A practical set-up to create an adjustable insufficiency in a porcine aortic root. A practical set-up to create an adjustable insufficiency in a porcine aortic root. Internship Hemolab BMTE07.13 T.L.Plantenga Supervision: P.H.M. Bovendeerd S.v.Tuijl M.Stijnen TU/e HemoLab HemoLab Eindhoven,

More information

T sors in the following aspects: the porcine aortic valve

T sors in the following aspects: the porcine aortic valve Clinical and Hemodynamic Assessment of the Hancock I1 Bioprosthesis Tirone E. David, MD, Susan Armstrong, MSc, and Zhao Sun, MA Division of Cardiovascular Surgery, The Toronto Hospital and University of

More information

Verification and Validation Methodologies for Prosthetic Heart Valves: Review and Considerations. ASME V&V40 Subcommittee Heart Valve Subgroup

Verification and Validation Methodologies for Prosthetic Heart Valves: Review and Considerations. ASME V&V40 Subcommittee Heart Valve Subgroup Verification and Validation Methodologies for Prosthetic Heart Valves: Review and Considerations. ASME V&V40 Subcommittee Heart Valve Subgroup ASME V&V 40 Heart Valve Subgroup Members Andrew Rau (Exponent

More information

Posterior leaflet prolapse is the most common lesion seen

Posterior leaflet prolapse is the most common lesion seen Techniques for Repairing Posterior Leaflet Prolapse of the Mitral Valve Robin Varghese, MD, MS, and David H. Adams, MD Posterior leaflet prolapse is the most common lesion seen in degenerative mitral valve

More information

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION Jayme Pinto Ortiz University of São Paulo - Avenida Prof. Luciano Gualberto, travessa3 nº 380 - CEP - 05508-900

More information

15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses

15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses ORIGINAL CONTRIBUTION 15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses WR Eric Jamieson, MD, Eva Germann, MSc, Michel R Aupart, MD 1, Paul H Neville, MD 1, Michel A Marchand,

More information

Soft tissue biomechanics

Soft tissue biomechanics Soft tissue biomechanics Caroline Öhman Pula, 22/06-08 TABLE OF CONTENTS Introduction to soft tissues Tendon and ligaments Introduction Composition Function and structure In vitro testing Stress-strain

More information

Effect of leaflet-to-chordae contact interaction on computational mitral valve evaluation

Effect of leaflet-to-chordae contact interaction on computational mitral valve evaluation Rim et al. BioMedical Engineering OnLine 2014, 13:31 RESEARCH Open Access Effect of leaflet-to-chordae contact interaction on computational mitral valve evaluation Yonghoon Rim, David D McPherson and Hyunggun

More information

The Phvsical Properties of Bovine Pericardium: A Stud? of the Elffects of Stretching during Chemical Treatment in Glutaraldehyde

The Phvsical Properties of Bovine Pericardium: A Stud? of the Elffects of Stretching during Chemical Treatment in Glutaraldehyde The Phvsical Properties of Bovine Pericardium: A Stud? of the Elffects of Stretching during Chemical Treatment in Glutaraldehyde I. J. Reece, M.B., F.R.C.S., R. van Noort, B.Sc., D.Phil., T. R. P. Martin,

More information

Hemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics

Hemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics Hemodynamic Assessment Matt M. Umland, RDCS, FASE Aurora Medical Group Milwaukee, WI Assessment of Systolic Function Doppler Hemodynamics Stroke Volume Cardiac Output Cardiac Index Tei Index/Index of myocardial

More information

TAVR or SAVR: Beyond the STS Score

TAVR or SAVR: Beyond the STS Score TAVR or SAVR: Beyond the STS Score Juan Crestanello, MD Division of Cardiac Surgery Lakshmi Prasad Dasi, PhD Department of Biomedical Engineering The Ohio State University Wexner Medical Center 1 Disclosures

More information

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM H. GIRIJA BAI 1 and K.B. NAIDU 2 Department of Mathematics, Sathyabama University, Chennai-600 119, Tamil Nadu, India 1 girijanameprakash@gmail.com

More information

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt Washington University in St. Louis Washington University Open Scholarship Mechanical Engineering and Materials Science Independent Study Mechanical Engineering & Materials Science 12-23-2017 Computational

More information

The Edge-to-Edge Technique f For Barlow's Disease

The Edge-to-Edge Technique f For Barlow's Disease The Edge-to-Edge Technique f For Barlow's Disease Ottavio Alfieri, Michele De Bonis, Elisabetta Lapenna, Francesco Maisano, Lucia Torracca, Giovanni La Canna. Department of Cardiac Surgery, San Raffaele

More information

Medical Biophysics 3970Z Six Week Project Determining the Mechanical Properties of the Eardrum through Modelling

Medical Biophysics 3970Z Six Week Project Determining the Mechanical Properties of the Eardrum through Modelling 1 Medical Biophysics 3970Z Six Week Project Determining the Mechanical Properties of the Eardrum through Modelling INTRODUCTION The mechanism of hearing involves a series of coordinated transfers, which

More information

Aortic valve repair: Techniques and Pitfalls. Allan Stewart, MD Columbia University Medical Center New York, NY

Aortic valve repair: Techniques and Pitfalls. Allan Stewart, MD Columbia University Medical Center New York, NY Aortic valve repair: Techniques and Pitfalls Allan Stewart, MD Columbia University Medical Center New York, NY Take Away Points 1. Valve anatomy is essential to assess repair 2. Unique Decisions with Aneurysm/AI

More information

Annular Stabilization Techniques in the Context of Aortic Valve Repair

Annular Stabilization Techniques in the Context of Aortic Valve Repair Annular Stabilization Techniques in the Context of Aortic Valve Repair Prashanth Vallabhajosyula, MD MS University of Pennsylvania, Philadelphia, Pennsylvania 2 nd North American Aortic Valve Repair Symposium

More information

Aortic Valve Repair a Modular and Geometric Approach. H.-J. Schäfers Dept. of Thoracic and Cardiovascular Surgery University Hospital of Saarland

Aortic Valve Repair a Modular and Geometric Approach. H.-J. Schäfers Dept. of Thoracic and Cardiovascular Surgery University Hospital of Saarland Aortic Valve Repair a Modular and Geometric Approach H.-J. Schäfers Dept. of Thoracic and Cardiovascular Surgery University Hospital of Saarland 12.09.2018 Limitations: Purely echocardiographic, does not

More information