Performance of Dedicated Versus Integrated Bipolar Defibrillator Leads with CRT-Defibrillators: Results from a Prospective Multicenter Study

Size: px
Start display at page:

Download "Performance of Dedicated Versus Integrated Bipolar Defibrillator Leads with CRT-Defibrillators: Results from a Prospective Multicenter Study"

Transcription

1 Performance of Dedicated Versus Integrated Bipolar Defibrillator Leads with CRT-Defibrillators: Results from a Prospective Multicenter Study ROGER A. FREEDMAN, M.D.,* ALEX PETRAKIAN, M.D., KER BOYCE, M.D., CHARLES HAFFAJEE, M.D., JESUS E. VAL-MEJIAS, M.D., and ASHISH L. OZA, M.S.** From the *University of Utah Hospitals and Clinics, Salt Lake City, Utah; East Texas Medical Center, Tyler, Texas; First Health Moore Regional Hospital, Pinehurst, North Carolina; St. Elizabeth s Medical Center of Boston, Boston, Massachusetts; Galichia Heart Hospital, Wichita, Kansas; and **St. Jude Medical, Sylmar, California Introduction: Right ventricular (RV) anodal stimulation may occur in cardiac resynchronization therapy defibrillators (CRT-D) when left ventricular (LV) pacing is configured between the LV lead and an electrode on the RV defibrillator lead. RV defibrillator leads can have a dedicated proximal pacing ring electrode (dedicated bipolar) or utilize the distal shocking coil as the proximal pacing electrode (integrated bipolar). This study compares the performance of integrated versus dedicated leads with respect to anodal stimulation incidence, sensing, and inappropriate ventricular tachyarrhythmia detection in patients implanted with CRT-D. Methods: Two hundred ninety-two patients were randomly assigned to receive dedicated or integrated bipolar RV leads at the time of CRT-D implantation. Patients were followed for 6 months. Results: Patients with dedicated bipolar RV leads exhibited markedly higher rates of anodal stimulation than did patients with integrated leads. The incidence of anodal stimulation was 64% at implant for dedicated bipolar RV leads compared to 1% for integrated bipolar RV leads. The likelihood of anodal stimulation in patients with dedicated leads fell progressively during the 6-month follow-up (51.5%), but always exceeded the incidence of anodal stimulation in patients with integrated leads (5%). Clinically detectable undersensing and oversensing were very unusual and did not differ significantly between lead designs. There were no inappropriate ventricular tachyarrhythmia detections for either lead type. Conclusion: Integrated bipolar RV defibrillator leads had a significantly lower incidence of RV anodal stimulation when compared to dedicated bipolar RV defibrillation leads, with no clinically detectable oversensing or undersensing, and with no inappropriate ventricular tachyarrhythmia detections for either lead type. (PACE 2009; 32: ) integrated, dedicated, anodal stimulation, undersensing, oversensing, inappropriate therapies *Financial Support: This study was sponsored by St. Jude Medical. Conflict of Interest: (1) Dr. Roger Freedman, St. Jude Medical: research support, consulting, fellows program support; Boston Scientific: research support, consulting, fellows program support; Medtronic: research support, consulting, fellows program support; Sorin: consulting. (2) Dr. Alex Petrakian, St. Jude Medical: stock ownership; Boston Scientific: stock ownership; Medtronic: stock ownership. (3) Dr. Ker Boyce, St. Jude Medical: Research support, consulting; Boston Scientific research support; Spectranetics physician proctor. (4) Dr. Charles Haffajee, St. Jude Medical: Research support; Biotronik: research support; Boston Scientific: Research support, consulting; Medtronic: research support; ELA- Sorin: consulting, national PI. (5) Dr. Jesus E. Val-Mejias, St. Jude Medical: Research support, fellows program support. (6) Mr. Ashish Oza, St. Jude Medical CRMD: employee, stock ownership. Address for reprints: Roger A. Freedman, M.D., University of Utah Hospitals and Clinics, 30 North 1900 East, 4A100 UUHSC, Salt Lake City, UT Fax: ; roger.freedman@hsc.utah.edu Received July 22, 2008; revised September 14, 2008; accepted October 16, Introduction Cardiac resynchronization therapy (CRT) with biventricular pacing has been established as an effective treatment in selected patients with drug refractory heart failure. Various studies have demonstrated the clinical benefit of CRT with improvements in quality of life, heart failure symptoms, exercise capacity, mortality, and hospitalization rates. 1 3 In the United States, CRT is usually incorporated into devices that also have defibrillator capability given that this patient population has been shown to be at risk of sudden cardiac death. 4,5 An essential component of CRT defibrillators (CRT-D) is the right ventricular (RV) defibrillation lead. RV defibrillation leads are available with two configurations of sensing and pacing electrodes. (Fig. 1). Dedicated bipolar RV defibrillator leads have a ring electrode just proximal to the distal pacing electrode, similar to bipolar pacemaker leads, which is used as the anode for pacing and sensing. Integrated bipolar RV defibrillator leads do not have a proximal ring electrode; rather, the C 2009, The Authors. Journal compilation C 2009 Wiley Periodicals, Inc. PACE, Vol. 32 February

2 FREEDMAN, ET AL. Figure 1. Anodes for cross-chamber LV pacing using a dedicated bipolar RV lead (left) and an integrated bipolar RV lead (right). In both instances, the cathode for LV pacing is the LV pacing lead electrode. distal shocking coil is used as the anode for pacing and sensing. In CRT devices, left ventricular (LV) pacing is often done with the LV lead tip electrode serving as the cathode and the RV ring/coil as the anode. This configuration of pacing has been variously termed cross chamber pacing, extended bipolar pacing, and dual cathodal pacing. In most cases with this configuration, myocardial depolarization occurs only at the LV (cathode). However, in some patients myocardial depolarization may also occur from the RV ring/coil (anode) at sufficiently high output voltages; this is termed anodal stimulation. When anodal stimulation occurs, there is capture at both the RV and LV when only LV pacing was intended. If the CRT device is programmed to deliver the LV stimulus prior to the RV stimulus ( LV offset ) in order to improve resynchronization efficacy, anodal stimulation negates such offset since both the LV and RV will be depolarized simultaneously from the programmed LV stimulus. The subsequent RV stimulus, which typically would be delivered ms after the programmed LV stimulus, would encounter refractory myocardium and not result in RV cathodal capture. At very short V-V intervals (< 20 ms), cathodal stimulation from RV stimulus is still possible. 6 In either case, sequential V-V pacing is disrupted because of anodal stimulation resulting in unintended simultaneous biventricular stimulation. Anodal stimulation is also potentially arrhythmogenic. Ventricular refractoriness is shorter for anodal versus cathodal stimulation, particularly early after lead placement, and is associated with an increase in the vulnerable period for arrhythmias in the setting of anodal stimulation. 7 9 Additionally, anodal stimulation can make it difficult to determine LV pacing thresholds, occasionally resulting in an overestimation of the true LV pacing threshold and programming of unnecessarily high LV pacing outputs. On the other hand, anodal stimulation has been reported in some cases to have a beneficial effect on LV synchrony and LV systolic performance during biventricular pacing. 10 This prospective, randomized trial was designed to compare dedicated and integrated bipolar RV defibrillator leads in patients receiving CRT-D with respect to anodal RV stimulation, clinically detectable RV undersensing and oversensing, and inappropriate ventricular tachyarrhythmia detection due to oversensing. Methods Patient Population Nineteen centers enrolled and successfully implanted CRT-D devices in 292 patients. The centers and investigators are listed in the Appendix. The study was reviewed and approved by the appropriate Human Research Ethical Committees of each of the participating medical centers, and all patients gave written informed consent for participation in the study. Of the enrolled patients, 267 had not been previously implanted with a pacemaker or defibrillator. The other 25 patients were receiving a new RV defibrillator lead to upgrade or revise an existing pacemaker or defibrillator system. Patients were ineligible for the study if they had a mechanical valve in the tricuspid position, if their life expectancy was less than 6 months, if they were pregnant or planning pregnancy within 6 months, or if they were less than 18 years old. Lead System Implantation All patients were implanted with a Food and Drug Administration-approved CRT-D generator (Epic HF or Atlas HF models V-337, V-338, V-340, V343 or V355, St. Jude Medical, Sylmar, CA, USA). Patients were randomized and implanted with either a dedicated bipolar RV defibrillator lead (St. Jude Medical Riata or Riata ST models 1580, 1581, 1582, 7000, 7001, or 7002) or an integrated bipolar RV defibrillator lead (St. Jude Medical Riata I models 1590, 1591, 7010, or 7011) (Fig. 1). A variable randomization block scheme, within centers as the blocking factor, was 158 February 2009 PACE,Vol.32

3 INTEGRATED VS. DEDICATED DEFIBRILLATION LEADS used to provide the study center s randomization schedule. Both the dedicated and integrated RV leads were available in single- and dual-shocking coil designs and the choice between these was not controlled in this study. For both RV lead configurations, the surface area of the distal pacing electrode was 8 mm 2. The dedicated RV leads have a proximal pace/sense anodal ring electrode with surface area of 17 mm 2 and tip to ring spacing of 11 mm. The integrated RV leads do not have a proximal ring electrode and use the distal shocking coil as the pacing anode, with surface area of 367 mm 2 (models 7000, 7001, 7002, 7010, 7011) or 414 mm 2 (models 1580, 1581, 1582, 1590, 1591), all with a tip-to-coil spacing of 11 mm. Anodal Stimulation Threshold Testing and LV Capture Threshold Testing Anodal stimulation and LV capture threshold testing (Fig. 2) was performed by initiating LV pacing in the LV tip to RV ring/coil pacing configuration (cross chamber pacing) starting at 7.5 V and 0.5-ms pulse width and then progressively decrementing the output voltage. Surface electrocardiogram (ECG) and intracardiac electrograms were recorded during the testing. Anodal stimulation was identified by a local electrogram recorded on the RV lead at, or nearly at, the same time as LV pacing stimulus. Loss of anodal stimulation was identified by time displacement of the RV electrogram and a change in the surface QRS complex. Loss of LV capture was identified by the absence of a local LV electrogram and absence of a paced QRS complex on surface ECG. The anodal stimulation threshold was defined as the lowest LV output voltage resulting in anodal stimulation at 0.5-ms pulse width, and LV capture threshold was defined as the lowest LV output resulting in LV capture at 0.5-ms pulse width. Since pacing output is typically programmed to twice the capture threshold, patients who exhibited an anodal stimulation threshold less than or equal to twice the LV capture threshold were considered to have clinically relevant anodal stimulation. Device Programming Ventricular sensitivity was programmed using an automatic sensitivity control algorithm, with 93% of the patients programmed to the nominal maximum sensitivity value (0.3 mv). The ventricular tachyarrhythmia detection zones were programmed so that the lowest detection rate criterion was no higher than 150 beats/min. Stored electrogram recording was programmed to trigger on the diagnosis of an atrial or ventricular tachyarrhythmia. The duration of the stored electrogram prior to the trigger was set to 16 seconds (50%), 14 seconds (45%), or 12 seconds (5%). Follow-Up Patients were seen and their devices were interrogated and tested at intervals of 1, 3, and 6 months after implantation. Study follow-up was terminated 6 months postimplant. At the time of implant and at each follow-up, routine RV capture and sensing thresholds were measured. LV capture threshold and RV anodal stimulation threshold were measured with the cross-chamber pacing configuration as described above. At all follow-up visits, device-based stored electrograms were evaluated by each participating investigator to identify any episodes of at least one undersensed or oversensed beat or any inappropriate ventricular tachyarrhythmia detection. Inappropriate detection was defined as any detection interpreted by the device as a ventricular tachyarrhythmia Figure 2. LV capture threshold test in a patient with a dedicated bipolar RV lead, showing loss of anodal stimulation and loss of LV capture. Note the change in surface paced QRS morphology and the change in RV electrogram morphology when anodal stimulation is lost. PACE,Vol.32 February

4 FREEDMAN, ET AL. where the corresponding stored electrogram showed the presence of myopotential or other electromagnetic interference resulting in the detection. All episodes of undersensing or oversensing with documentation were reviewed by a single investigator (R.A.F.) for confirmation. Statistical Analysis A Fisher s exact test was used to compare the incidence of anodal stimulation, undersensing, and oversensing between the integrated and dedicated leads. A Cuzick s nonparametric test was used to check for trends in the incidence of clinically relevant anodal stimulation, undersensing, and oversensing over time across each group. A two-group t-test of equal means was used to compare RV pacing threshold, RV sensing threshold, LV pacing threshold, and LV pacing impedance between the dedicated and integrated leads. A p-value < 0.05 was considered statistically significant. All tests and fits were carried out in SAS (SAS Institute Inc., Cary, NC, USA) and Stata (Stata, Statistics/Data Analysis, College Station, TX, USA). Results Patient Demographics The study population of 292 patients was typical of those receiving CRT-D devices (Table I). Most of the patients were men, most had coronary artery disease as their underlying structural heart disease, and most were New York Heart Association (NYHA) functional class III. The typical LV ejection fraction was in the 20 30% range. Sixteen (5%) of the patients were in NYHA functional class II; patients with NYHA class II symptoms do not meet standard criteria for CRT implantation, and the decision to implant CRT-D systems in these patients was made by the individual investigator. Of the 292 implanted patients, 247 completed their 1-month follow-up visit, 230 completed their 3-month follow-up visit, and 214 patients completed their 6-month follow-up visit. There were a total of 11 deaths during the course of the study (seven with dedicated leads and four with integrated leads) that were reported to be unrelated to the device. There were no cases of RV lead dislodgement but there were two cases of RV lead perforation during the course of the study (one with dedicated lead and one with integrated lead). The remainder of the dropout was due to withdrawal from the study or noncompliance with follow-up. The drop-out rates between the two randomized groups were similar (26.5% with dedicated leads vs. 27.4% with integrated leads). Table I. Patient Population Dedicated Integrated Leads Leads (n = 147) (n = 145) Age (years) 69 ± ± 12 (Mean ± SD) Gender 72% male 76% male Ejection fraction (%) 25 ± 7 24 ± 7 (Mean ± SD) NYHA class (%) II 8 8 III IV 5 1 Unknown 3 3 Coronary artery disease (%) RV Lead Position (%) RV apex ICD Implant Indication (%) Primary Secondary Unknown 8 5 Anodal Stimulation The incidence of anodal stimulation at any LV output up to 7.5 V is shown in Figure 3. At implant, anodal stimulation was seen in 64% of patients with dedicated bipolar RV leads compared to only 1% with integrated bipolar RV leads (p < ). The incidence of anodal stimulation with the dedicated leads fell progressively to 51.5% at 6 months postimplant (p = 0.04), but always remained significantly higher than with the integrated leads. The incidence of clinically relevant anodal stimulation (anodal stimulation at a voltage less Figure 3. Prevalence of anodal stimulation, significantly lower with integrated bipolar RV lead systems as compared to dedicated bipolar RV lead systems. 160 February 2009 PACE,Vol.32

5 INTEGRATED VS. DEDICATED DEFIBRILLATION LEADS Figure 4. Prevalence of clinically relevant anodal stimulation (anodal stimulation at a voltage less than or equal to twice the LV capture threshold), significantly lower with integrated bipolar RV lead systems as compared to dedicated bipolar RV lead systems. than or equal to twice LV capture threshold) is shown in Figure 4. Again, at each follow-up, clinically relevant anodal stimulation was significantly more common with dedicated bipolar RV leads than with integrated bipolar RV leads (P < at implant; P = 0.02 at the 6-month follow-up). The incidence of clinically relevant anodal stimulation with dedicated bipolar RV leads was 22% at implant and fell progressively (P = 0.005) during follow-up to 10% at 6 months postimplant. By contrast, the incidence of clinically relevant anodal stimulation with integrated bipolar RV leads never exceeded 4% at any time of follow-up. There was no significant difference in LV pacing threshold during cross-chamber pacing between patients with dedicated versus inte- grated RV leads, but LV cross-chamber pacing impedances were lower in patients with integrated compared to those with dedicated RV leads (P < 0.05) (Table II). RV pacing thresholds and RV pacing impedances were lower with integrated compared to dedicated RV leads (Table II). Of the 75 patients with anodal stimulation at any LV output at 1-month postimplant, the LV output programmed at that time was available for analysis in 59 patients. A comparison of the anodal stimulation threshold, LV capture threshold, and actual programmed LV output in those patients is shown in Table III. The mean anodal stimulation was 3.89 ± 1.79 V, mean LV capture threshold was 1.58 ± 1.46 V, and mean programmed LV output was 3.50 ± 1.34 V. In 31 of the 59 patients (27 with dedicated RV leads and four with integrated RV leads), the programmed LV output exceeded the anodal stimulation threshold. In these patients, it is likely that there was actually anodal stimulation in the RV ongoing on an ambulatory basis. Ventricular Sensing There was no significant difference in RV sensing threshold between patients with dedicated versus integrated RV leads (Table II). There was no documentation of ventricular undersensing in any patient. Three episodes of oversensing due to electromagnetic interference (EMI) were documented; a single episode was documented in one patient with dedicated bipolar RV lead and two episodes documented in a second patient with an integrated bipolar RV lead. In all three episodes, the EMI was detected in both the atrial and ventricular channels (Fig. 5). No episodes of oversensing due to myopotentials Table II. Comparison of LV Capture Threshold, LV Pacing Impedance, RV Pacing Impedance, RV Pacing Threshold, and RV Sensing Thresshold between the Dedicated and Integrated Lead Systems at Different Follow-Up Points Implant 1 Month 3 Months 6 Months LV capture threshold (V at 0.5 ms) Dedicated 1.5 ± ± ± ± 1 Integrated 1.5 ± ± ± ± 1.4 LV pacing impedance (Ohms) Dedicated 713 ± 165* 690 ± 142* 705 ± 157* 696 ± 143* Integrated 609 ± ± ± ± 195 RV pacing impedance (Ohms) Dedicated 519 ± 91* 478 ± 68* 485 ± 75* 484 ± 96* Integrated 417 ± ± ± ± 84 RV pacing threshold (V at 0.5 ms) Dedicated 0.6 ± 0.2* 0.9 ± 0.8* 0.8 ± 0.6* 0.9 ± 0.7* Integrated 0.5 ± ± ± ± 0.4 RV sensing threshold (mv) Dedicated 10.8 ± ± ± ± 2.0 Integrated 10.3 ± ± ± ± 1.7 All values are mean ± standard deviation. * P < 0.05 compared to integrated leads. PACE,Vol.32 February

6 FREEDMAN, ET AL. Table III. Comparison of Anodal Stimulation Threshold, Left Ventricular (LV) Capture Threshold, and Programmed LV Output in 59 Patients with Anodal Stimulation at 1-Month Postimplant* Anodal LV Capture Anodal Stimulation Threshold LV Stimulation Patient Lead Threshold (V) at Programmed at Programmed Number Type (V) at 0.5 ms 0.5 ms Output (V) Output 2 Dedicated No 5 Dedicated No 7 Dedicated No 8 Dedicated Yes 10 Integrated Yes 13 Integrated Yes 18 Dedicated No 20 Dedicated Yes 22 Dedicated Yes 24 Dedicated Yes 25 Dedicated No 33 Dedicated Yes 37 Dedicated Yes 42 Dedicated No 63 Dedicated No 69 Dedicated No 74 Dedicated No 85 Dedicated Yes 86 Dedicated Yes 89 Dedicated No 95 Dedicated No 118 Dedicated Yes 119 Dedicated Yes 121 Dedicated Yes 122 Dedicated Yes 123 Dedicated No 126 Dedicated No 129 Dedicated No 132 Dedicated No 145 Dedicated No 153 Dedicated No 157 Dedicated Yes 162 Dedicated No 165 Dedicated No 166 Dedicated Yes 176 Dedicated No 181 Dedicated Yes 183 Dedicated Yes 185 Dedicated Yes 189 Dedicated Yes 197 Dedicated Yes 200 Dedicated No 213 Integrated Yes 216 Dedicated Yes 219 Integrated Yes 243 Dedicated Yes Continued. 162 February 2009 PACE,Vol.32

7 INTEGRATED VS. DEDICATED DEFIBRILLATION LEADS Table III. Continued Anodal LV Capture Anodal Stimulation Threshold LV Stimulation Patient Lead Threshold (V) at Programmed at Programmed Number Type (V) at 0.5 ms 0.5 ms Output (V) Output 248 Dedicated No 251 Dedicated No 252 Dedicated No 254 Dedicated Yes 270 Dedicated Yes 285 Dedicated No 301 Dedicated Yes 303 Dedicated Yes 322 Dedicated No 326 Dedicated No 327 Dedicated Yes 345 Dedicated No 371 Dedicated Yes Mean ± SD 3.89 ± ± ± 1.34 *An additional 16 patients had anodal stimulation at 1-month postimplant, but programmed LV output was not available for the analysis. were documented. There were no inappropriate ventricular tachyarrhythmia detections resulting from oversensing with either lead design. Far-Field R-Wave Sensing Twenty-five episodes of far-field sensing of R waves in the atrial channel were documented, occurring one or more times in 16 patients. Of these 16 patients, six had dedicated RV leads and 10 had integrated RV leads (P = NS). The far-field R-wave sensing invariably occurred during biventricular pacing and was detected as an inappropriate au- tomatic mode switch due to double counting of atrial rate. Discussion This prospective randomized trial compared integrated and dedicated bipolar RV defibrillator leads with respect to anodal stimulation and ventricular sensing performance. The principal results of this trial are (1) the incidence of anodal stimulation in integrated bipolar right ventricular leads was significantly lower than in dedicated bipolar right ventricular leads and (2) clinically Figure 5. Stored electrogram showing electromagnetic interference (EMI) of unknown origin detected as noise in a patient with a dedicated bipolar RV lead. Note that the EMI is detected in both the atrial and ventricular channels. Coincidentally, the EMI is detected shortly after the onset of ventricular tachycardia. RA = right atrial electrogram; RV = right ventricular electrogram. PACE,Vol.32 February

8 FREEDMAN, ET AL. detectable ventricular undersensing and oversensing were rare with both lead designs and did not differ between the two lead designs. The current study results are consistent with the findings of prior smaller studies. 11,12 In their study of 38 patients, Tamborero et al. found anodal stimulation during cross-chamber LV pacing in 74% of patients with dedicated bipolar RV leads versus none in patients with integrated bipolar RV leads. Thibault et al. found RV capture during cross-chamber LV stimulation in all 11 patients with CRT-D utilizing dedicated bipolar RV leads but none of 15 patients with integrated bipolar RV leads. Unlike the present study, neither Tamborero et al. nor Thibault et al. examined the likelihood of anodal stimulation at clinically relevant pacing outputs or the sensing characteristics of their leads. We found clinically relevant anodal stimulation in a significant proportion of patients with dedicated bipolar RV leads but only rarely in patients with integrated bipolar RV leads. We found no difference in RV sensing characteristics between the two lead designs. The differing incidence of anodal stimulation between dedicated and integrated bipolar RV leads is likely to be related to the differences in the surface area of the anodes. Due to the smaller surface area of an RV ring anode in a dedicated bipolar lead compared to the shocking coil anode in an integrated bipolar lead, current density in the area of the anode would be predicted to be higher with the dedicated bipolar lead than with the integrated bipolar lead. The higher current density would result in a greater likelihood of local myocardial depolarization at the anode with the dedicated bipolar leads. We found a progressive reduction in the incidence of anodal stimulation with cross-chamber LV pacing in dedicated leads following implantation. This trend is likely explained by the formation of a virtual electrode around the RV ring electrode with development of a conductive fibrotic capsule. The effective surface area of this virtual electrode is larger than the original electrode as it includes the original electrode and the newly formed fibrotic tissue. An increase in the virtual electrode size with time would be expected to progressively lower the current density around the anode and decrease the likelihood of anodal stimulation. RV pacing is often programmed to occur after a short delay following LV pacing in order to optimize the mechanical and hemodynamic benefit of CRT. If anodal RV stimulation occurs during cross-channel LV pacing, then this intended delay will not occur. In defibrillator systems incorporating a bipolar LV lead, anodal stimulation can be avoided by configuring LV pacing between the two LV electrodes rather than programming the system to cross-chamber LV pacing. This option is not available in patients where the LV lead is unipolar. Furthermore, even in patients with bipolar leads, cross-chamber LV pacing configuration is sometimes desired because of more favorable capture thresholds or better avoidance of phrenic nerve stimulation. Some newer resynchronizing defibrillator generators utilize the distal shocking coil as the anode for cross-chamber LV pacing even when the RV defibrillator lead is a dedicated bipolar design. In a study of 101 patients, Weretka et al. demonstrated that ventricular oversensing occurred more frequently in patients implanted with an integrated bipolar RV lead system (21/52) as compared to a dedicated bipolar RV lead system (4/49), with T-wave oversensing, P-wave oversensing, and diaphragmatic myopotential oversensing all demonstrated in patients with the integrated leads. 13 By contrast, in this study, ventricular oversensing was rarely seen, regardless of the lead design. Unlike the patients in this study, all the patients in the study by Weretka et al. underwent ambulatory monitoring, and this may partially explain the higher incidence of oversensing observed. However, there were inappropriate ventricular tachyarrhythmia therapies in 4% and repetitive inappropriate capacitor charges in 8% of the patients in that study, versus none in our study, and this difference cannot be attributed to difference in follow-up methodology. It is possible that the increased incidence of oversensing and inappropriate arrhythmia detections in the Weretka et al. study may be due to the difference in sense amplifier filters in the defibrillator generators used in that study (Boston Scientific, Inc., and Medtronic, Inc.) compared to the St. Jude Medical generators used in this study. Limitations Approximately 25% of the patients either withdrew from the study or were lost between implant and 6 months. It is conceivable that some bias was introduced into the study s results due to selective dropout of patients with a different likelihood of anodal stimulation or sensing performance. Only electrograms stored in the generator were available to examine the occurrence of ventricular undersensing or oversensing. Continuous ambulatory monitoring was not performed in these patients, and this might have detected a higher rate of sensing abnormalities. Clinical outcomes measured in this study were confined to measures of lead performance 164 February 2009 PACE,Vol.32

9 INTEGRATED VS. DEDICATED DEFIBRILLATION LEADS as well as oversensing, undersending, and inappropriate tachyarrhythmia detections, and followup in the study was terminated after 6 months. It is possible that longer follow-up would have revealed significant differences in clinical outcome in patients with one lead design versus the other. Also, the specific clinical impact of anodal stimulation and possible interference with LV-RV pacing offset were not examined in this study. The study utilized specific leads and pulse generators from a single manufacturer. Confining the study to defibrillator systems from a single manufacturer has the advantages of providing uniform lead designs, sensing amplifier filters, arrhythmia detection algorithms, and electrogram storage capability. However, the results cannot be necessarily extrapolated to defibrillator systems from other manufacturers. Conclusion We found that the incidence of RV anodal capture during LV cross-chamber pacing is significantly lower using an integrated lead system when compared to a dedicated lead system in patients receiving CRT-D systems. Oversensing and undersensing were uncommon with both dedicated and integrated RV lead systems. No patient had inappropriate ventricular arrhythmia detection or received inappropriate therapy due to oversensing of myopotentials or other electrical artifact. At the time of implantation, the physician s choice of a dedicated versus an integrated lead RV should take into account the clinical implications of the performance of the two lead designs. Following implantation, programming of LV pacing configuration and output should be done with the recognition of the possibility of anodal stimulation and its clinical implications. References 1. Salukhe TV, Francis DP, Sutton R. Comparison of medical therapy, pacing and defibrillation in heart failure (COMPANION) trial terminated early; combined biventricular pacemaker-defibrillators reduce all-cause mortality and hospitalization. Int J Cardiol 2003; 87: Abraham WT, Fisher WG, Smith AL, Delurgio DB, Leon AR, Loh E, Kocovic DZ, et al. Cardiac resynchronization in chronic heart failure. N Engl J Med 2002; 346: Cleland JG, Daubert JC, Erdmann E, Freemantle N, Gras D, Kappenberger L, Tavazzi L. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med 2005; 352: Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, Domanski M, et al. Amiodarone or an implantable cardioverter defibrillator for congestive heart failure. N Engl J Med 2005; 352: Hohnloser SH, Kuck KH, Dorian P, Roberts RS, Hampton JR, Hatala R, Fain E, et al. Prophylactic use of an implantable cardioverterdefibrillator after acute myocardial infarction. N Engl J Med 2004; 351: Van Gelder BM, Bracke FA, Meijer A. The effect of anodal stimulation on V-V timing at varying V-V intervals. Pacing Clin Electrophysiol 2005; 28: Mehra R, Furman S, Crump JF. Vulnerability of the mildly ischemic Appendix The following study centers and investigators participated in this trail: University of Utah Hospital and Clinics and Salt Lake VA Health Sciences Center, Salt Lake City, UT: Roger Freedman; East Texas Medical Center, Tyler, TX: Alex Petrakian; FirstHealth Moore Regional Hospital, Pinehurst, NC: Ker Boyce; St. Elizabeth s Medical Center of Boston, Boston, MA: Charles Haffajee; Galichia Heart Hospital, Wichita, KS: Jesus E. Val-Mejias; St. Luke s Methodist Hospital, Cedar Rapids, IA: Todd Langager; Lenox Hill Hospital, New York, NY: Noelle Langan; Fallon Clinic, Worcester, MA: Sanjeev Goyal; Montefiore Medical Center Moses ventricle to cathodal, anodal, and bipolar stimulation. Circ Res 1977; 41: Mehra R, McMullen M, Furman S. Time dependence of unipolar cathodal and anodal strength-interval curves. Pacing Clin Electrophysiol 1980; 3: Bennett JA, Roth BJ. Time dependence of anodal and cathodal refractory periods in cardiac tissue. Pacing Clin Electrophysiol 1999; 22: Bulava A, Ansalone G, Ricci R, Giannantoni P, Pignalberi C, Heinc P, Lukl J, et al. Triple-site pacing in patients with biventricular device Incidence of the phenomenon and cardiac resynchronization benefit. J Inter Card Electrophysiol 2004; 10: Tamborero D, Mont L, Alanis R, Berruezo A, Tolosana JM, Sitges M, Vidal B, et al. Anodal capture in cardiac resynchronization therapy implications for device programming. Pacing Clin Electrophysiol 2006; 29: Thibault B, Roy D, Guerra PG, Macle L, Dubuc M, Gagné P, Greiss I, et al. Anodal right ventricular capture during left ventricular stimulation in CRT-implantable cardioverter defibrillators. Pacing Clin Electrophysiol 2005; 28: Weretka S, Michaelsen J, Becker R, Karle CA, Voss F, Hilbel T, Osswald BR, et al. Ventricular oversensing: A study of 101 patients implanted with dual chamber defibrillators and two different lead systems. Pacing Clin Electrophysiol 2003; 26: Division, Bronx, NY: Jay Gross; Austin Heart, Austin, TX: David Kessler; St. Peters Hospital, Olympia, WA: Robert Wark; Hillcrest Hospital, Mayfield Heights, OH: Martin Wiseman; Roper Hospital, Charleston, SC: Brett Baker; Cardiology Associates of Marin and San Francisco, Larkspur, CA: Steven Hao; VA Medical Center Minneapolis, Minneapolis, MN: Venkat Tholakanahalli; Albany Medical Center, Albany, NY: James O Brien; St. Luke s Hospital Roosevelt, New York, NY: Jonathan Steinberg; Midwest Heart Foundation, Lombard, IL: Ray Kawasaki; Saint Barnabas Medical Center, Livingston, NJ: Marc Roelke. PACE,Vol.32 February

Anodal Capture in Cardiac Resynchronization Therapy Implications for Device Programming

Anodal Capture in Cardiac Resynchronization Therapy Implications for Device Programming Anodal Capture in Cardiac Resynchronization Therapy Implications for Device Programming DAVID TAMBORERO,* LLUIS MONT,* ROBERTO ALANIS, ANTONIO BERRUEZO,* JOSE MARIA TOLOSANA,* MARTA SITGES,* BARBARA VIDAL,*

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study ANEESH V. TOLAT, M.D.,* MELISSA WOICIECHOWSKI, M.S.N.,* ROSEMARIE KAHR, R.C.I.S.,* JOSEPH

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

EBR Systems, Inc. 686 W. Maude Ave., Suite 102 Sunnyvale, CA USA

EBR Systems, Inc. 686 W. Maude Ave., Suite 102 Sunnyvale, CA USA Over 200,000 patients worldwide are estimated to receive a CRT device each year. However, limitations prevent some patients from benefiting. CHALLENGING PROCEDURE 5% implanted patients fail to have coronary

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

Diagnostic capabilities of the implantable therapeutic systems

Diagnostic capabilities of the implantable therapeutic systems Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere

More information

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable

More information

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing

More information

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Pacing and Device Jargon Made Simple Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Disclosure Dr Jonathan Timperley Honorarium from Bayer Arne Larson (1915

More information

The Nuts and Bolts of ICD Therapy

The Nuts and Bolts of ICD Therapy Electrical Management of Cardiac Rhythm Disorders For Cardiology Fellows December 5-8 Austin, Texas The Nuts and Bolts of ICD Therapy 1 2 Action Potential Localized Differences in Conduction Conduction

More information

Implantable Cardioverter Defibrillator Therapy in MADIT II Patients with Signs and Symptoms of Heart Failure

Implantable Cardioverter Defibrillator Therapy in MADIT II Patients with Signs and Symptoms of Heart Failure Implantable Cardioverter Defibrillator Therapy in MADIT II Patients with Signs and Symptoms of Heart Failure Wojciech Zareba Postinfarction patients with left ventricular dysfunction are at increased risk

More information

Tachycardia Devices Indications and Basic Trouble Shooting

Tachycardia Devices Indications and Basic Trouble Shooting Tachycardia Devices Indications and Basic Trouble Shooting Peter A. Brady, MD., FRCP Cardiology Review Course London, March 6 th, 2014 2011 MFMER 3134946-1 Tachycardia Devices ICD Indications Primary and

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

LONGITUDINAL SURVEILLANCE REGISTRY OF ACUITY SPIRAL LEAD

LONGITUDINAL SURVEILLANCE REGISTRY OF ACUITY SPIRAL LEAD CLINICAL SUMMARY LONGITUDINAL SURVEILLANCE REGISTRY OF ACUITY SPIRAL LEAD CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and

More information

Cardiac resynchronization therapy for mild-to-moderate heart failure

Cardiac resynchronization therapy for mild-to-moderate heart failure For reprint orders, please contact reprints@expert-reviews.com Cardiac resynchronization therapy for mild-to-moderate heart failure Expert Rev. Med. Devices 8(3), 313 317 (2011) Haran Burri Electrophysiology

More information

GDT1000 SENSING ACUTE STUDY

GDT1000 SENSING ACUTE STUDY CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY GDT1000 SENSING ACUTE STUDY Boston

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France

How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France Presenter Disclosure Information Christophe Leclercq, MD,

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL

More information

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a

More information

Biventricular pacing (BVP) is effective in patients

Biventricular pacing (BVP) is effective in patients Case Report 178 Right Ventricular Lead Ring Capture in Sequential Biventricular Pacing with Pseudo-bipolar Left Ventricular Lead Configuration: an Unwanted Effect Oruganti Sai Satish, MD, DM; Kuan-Hung

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

G Lin, R F Rea, S C Hammill, D L Hayes, P A Brady

G Lin, R F Rea, S C Hammill, D L Hayes, P A Brady Division of Cardiovascular Diseases, Mayo Clinic, Rochester, MN, USA Correspondence to: Dr Peter A Brady, MD, FRCP, Mayo Clinic, 200 First Street SW, Rochester, MN, 55905, USA; brady.peter@mayo.edu Accepted

More information

Does Changing the Pacing Configuration Affect Phrenic Stimulation? - A substudy of the DETECT PS Trial -

Does Changing the Pacing Configuration Affect Phrenic Stimulation? - A substudy of the DETECT PS Trial - Does Changing the Pacing Configuration Affect Phrenic Stimulation? - A substudy of the DETECT PS Trial - Michael Glikson, MD 1, Vladmir Khalameizer, MD 2, Dominique Babuty, MD 3, Sunipa Saha 4, Holly Rockweiler

More information

58 Marc-Alexander Ohlow, MD 1 ; Bernward Lauer, PhD 1 ; Michele Brunelli, PhD 2, Yunis Daralammouri, MD 1 ; J Christoph Geller, PhD 2 ;

58 Marc-Alexander Ohlow, MD 1 ; Bernward Lauer, PhD 1 ; Michele Brunelli, PhD 2, Yunis Daralammouri, MD 1 ; J Christoph Geller, PhD 2 ; www.ipej.org 58 Original Article The Use of a Quadripolar Left Ventricular Lead Increases Successful Implantation Rates in Patients with Phrenic Nerve Stimulation and/or High Pacing Thresholds Undergoing

More information

Comparison of clinical trials evaluating cardiac resynchronization therapy in mild to moderate heart failure

Comparison of clinical trials evaluating cardiac resynchronization therapy in mild to moderate heart failure HOT TOPIC Cardiology Journal 2010, Vol. 17, No. 6, pp. 543 548 Copyright 2010 Via Medica ISSN 1897 5593 Comparison of clinical trials evaluating cardiac resynchronization therapy in mild to moderate heart

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Implantable cardioverter defibrillators for the treatment of arrhythmias and cardiac resynchronisation therapy for the treatment of heart failure (review

More information

A Square Peg in a Round Hole: CRT IN PAEDIATRICS AND CONGENITAL HEART DISEASE

A Square Peg in a Round Hole: CRT IN PAEDIATRICS AND CONGENITAL HEART DISEASE A Square Peg in a Round Hole: CRT IN PAEDIATRICS AND CONGENITAL HEART DISEASE Adele Greyling Dora Nginza Hospital, Port Elizabeth SA Heart November 2017 What are the guidelines based on? MADIT-II Size:

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Device Interrogation- Pacemakers, ICD and Loop Recorders Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Disclosures Consultant: Medtronic Speaker s Bureau: St. Jude Medical

More information

Advances in Arrhythmia and Electrophysiology

Advances in Arrhythmia and Electrophysiology Advances in Arrhythmia and Electrophysiology Troubleshooting Implanted Cardioverter Defibrillator Sensing Problems I Charles D. Swerdlow, MD; Samuel J. Asirvatham, MD; Kenneth A. Ellenbogen, MD; Paul A.

More information

Teaching Rounds in Cardiac Electrophysiology

Teaching Rounds in Cardiac Electrophysiology Teaching Rounds in Cardiac Electrophysiology Sustained Multiple Railroad Tracks on Implantable Cardiac Defibrillator Interval Plots Mechanisms and Management Alex Y. Tan, MD; Kenneth A. Ellenbogen, MD;

More information

Is This Thing Working?

Is This Thing Working? Is This *#@!* Thing Working? Pacemaker (and ICD) ECG and Telemetry Pitfalls Wayne O. Adkisson, MD adki0004@umn.edu Disclosures I currently receive research support from Medtronic, Inc. I have been compensated

More information

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds

More information

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART December 2015 This chart encompasses all Medtronic cardiac devices FDA-Approved as MR Conditional and included in the MRI SureScan portfolio. If a device

More information

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure Inappropriate Discharges of Intravenous Implantable Cardioverter Defibrillators Owing to Lead Failure Takashi WASHIZUKA, 1 MD, Masaomi CHINUSHI, 1 MD, Ryu KAZAMA, 1 MD, Takashi HIRONO, 1 MD, Hiroshi WATANABE,

More information

Cardiac resynchronisation therapy (biventricular pacing) for the treatment of heart failure

Cardiac resynchronisation therapy (biventricular pacing) for the treatment of heart failure NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Health Technology Appraisal for the treatment of heart failure Final scope Appraisal objective To appraise the clinical and cost effectiveness of cardiac

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Clinical Data Summary: Avoid FFS Study

Clinical Data Summary: Avoid FFS Study Atrial Pacing Lead with 1.1 mm Tip-to-Ring Spacing Clinical Data Summary: Avoid FFS Study A Multi-center, Randomized, Prospective Clinical Study Designed to Evaluate the 1699T Lead Clinical Data Summary:

More information

Dual-Chamber Implantable Cardioverter-Defibrillator

Dual-Chamber Implantable Cardioverter-Defibrillator February 1998 9 Dual-Chamber Implantable Cardioverter-Defibrillator A.SH. REVISHVILI A.N. Bakoulev Research Center for Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow, Russia Summary

More information

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Jason Collinson & Stuart Tan Essex Cardiothoracic Centre, Basildon and Thurrock University Hospital. Contact: jason.collinson@nhs.net

More information

Introduction. CLINICAL RESEARCH Clinical Trial Design. Mohammad Saeed 1 *, Mehdi Razavi 1, Curtis G. Neason 2, and Simona Petrutiu 2. Aims.

Introduction. CLINICAL RESEARCH Clinical Trial Design. Mohammad Saeed 1 *, Mehdi Razavi 1, Curtis G. Neason 2, and Simona Petrutiu 2. Aims. Europace (2011) 13, 1648 1652 doi:10.1093/europace/eur195 CLINICAL RESEARCH Clinical Trial Design Rationale and design for programming implantable cardioverter defibrillators in patients with primary prevention

More information

How to Approach the Patient with CRT and Recurrent Heart Failure

How to Approach the Patient with CRT and Recurrent Heart Failure How to Approach the Patient with CRT and Recurrent Heart Failure Byron K. Lee MD Associate Professor of Medicine Electrophysiology and Arrhythmia Section UCSF Update in Electrocardiography and Arrhythmias

More information

Programming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany

Programming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Programming of Bradycardic Parameters C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Programming of ICD Brady Parameters Conflict of Interest Biotronik

More information

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators 22 March 2002 Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators F. HINTRINGER, O. PACHINGER Division of Cardiology, Department for Internal Medicine, University

More information

Call Medtronic at 1 (800) to verify the patient s current implanted system

Call Medtronic at 1 (800) to verify the patient s current implanted system MRI SURESCAN SYSTEMS Patient Scanning Process Transvenous Implantable Cardiac Systems PATIENT PRESCREENING SureScan Pacing, Defibrillation, and CRT (CRT-D and CRT-P) Systems Verification Verify that patient

More information

CAPTIVATE SUMMARY CLINICAL SUMMARY. CAPTure Information Via Automatic Threshold Evaluation

CAPTIVATE SUMMARY CLINICAL SUMMARY. CAPTure Information Via Automatic Threshold Evaluation CLINICAL SUMMARY CAPTIVATE SUMMARY CAPTure Information Via Automatic Threshold Evaluation CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in

More information

Practice Questions.

Practice Questions. IBHRE Prep Practice Questions Question 1 The relative refractory yperiod of the ventricular myocardium corresponds to which of the following phases of the action potential? A. (0) B. (1) C. (2) D. (3)

More information

ICD leads survival and troubles in the last 10 years

ICD leads survival and troubles in the last 10 years XVI Symposium on Progress in Clinical Pacing, Rome, Italy - Dec 2 to 7, 2014 ICD leads survival and troubles in the last 10 years Antonio Raviele, MD, FESC, FHRS President ALFA Alliance to Fight Atrial

More information

Device based CRT optimization: is there a future? Lessons learned from published trials

Device based CRT optimization: is there a future? Lessons learned from published trials Device based CRT optimization: is there a future? Lessons learned from published trials C. Leclercq Department of Cardiology Centre Cardio-Pneumologique Rennes, France Presenter Disclosure Information

More information

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΖΩΡΤΖ ΔΑΔΟΥΣ ΕΠΙΚΟΥΡΟΣ ΚΑΘΗΓΗΤΗΣ Α.Π.Θ. ΜΑΡΙΑ ΚΑΡΑΛΙΟΛΙΟΥ ΕΙΔΙΚΕΥΟΜΕΝΗ ΙΑΤΡΟΣ ΚΑΡΔΙΟΛΟΓΟΣ Β ΚΑΡΔΙΟΛΟΓΙΚΗ ΚΛΙΝΙΚΗ Α.Π.Θ. ΙΠΠΟΚΡΑΤΕΙΟ Γ.Ν.Θ. 5 Why a Pacemaker is Implanted

More information

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia By Sandeep Joshi, MD and Jonathan S. Steinberg, MD Arrhythmia Service, Division of Cardiology

More information

INNOVATIONS IN DEVICE THERAPY:

INNOVATIONS IN DEVICE THERAPY: INNOVATIONS IN DEVICE THERAPY: Subcutaneous ICDs, Leadless Pacemakers, CRT Indications David J Wilber MD Loyola University Medical Center Disclosures: ACC Foundation: Consultant; Biosense / Webster: Consultant,

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Tseng ZH, Hayward RM, Clark NM, et al. Sudden death in patients with cardiac implantable electronic devices. JAMA Intern Med. Published online June 22, 2015. doi:10.1001/jamainternmed.2015.2641.

More information

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Identify patient with an ICD at POAC. Notify Cardiac Physiologist that patient is due to have surgery and

More information

CONSULTA CRT-P Model C4TR01

CONSULTA CRT-P Model C4TR01 CONSULTA CRT-P Model C4TR01 Physical characteristics Volume a 15 cm 3 Mass H x W x D Surface area of titanium device can Surface area of each titanium nitride LECG electrode Radiopaque ID Materials in

More information

Dialysis-Dependent Cardiomyopathy Patients Demonstrate Poor Survival Despite Reverse Remodeling With Cardiac Resynchronization Therapy

Dialysis-Dependent Cardiomyopathy Patients Demonstrate Poor Survival Despite Reverse Remodeling With Cardiac Resynchronization Therapy Dialysis-Dependent Cardiomyopathy Patients Demonstrate Poor Survival Despite Reverse Remodeling With Cardiac Resynchronization Therapy Evan Adelstein, MD, FHRS John Gorcsan III, MD Samir Saba, MD, FHRS

More information

Biatrial Stimulation and the Prevention of Atrial Fibrillation

Biatrial Stimulation and the Prevention of Atrial Fibrillation 374 September 2001 Biatrial Stimulation and the Prevention of Atrial Fibrillation L. MELCZER Heart Institute, Faculty of Medicine, University of Pécs, Pécs, Hungary I. LORINCZ 1 st Internal Department,

More information

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

A 'shocking' case rectified

A 'shocking' case rectified A 'shocking' case rectified Journal: Pacing and Clinical Electrophysiology Manuscript ID: PACE-13-00.R1 Wiley - Manuscript type: Device Rounds Date Submitted by the Author: n/a Complete List of Authors:

More information

8/8/2011. CARDIAC RESYCHRONIZATION THERAPY for Heart Failure. Case Presentation. Case Presentation

8/8/2011. CARDIAC RESYCHRONIZATION THERAPY for Heart Failure. Case Presentation. Case Presentation CARDIAC RESYCHRONIZATION THERAPY for Heart Failure James Taylor, DO, FACOS Cardiothoracic and Vascular surgery San Angelo Community Medical Center San Angelo, TX Case Presentation 64 year old female with

More information

Journal of the American College of Cardiology Vol. 50, No. 13, by the American College of Cardiology Foundation ISSN /07/$32.

Journal of the American College of Cardiology Vol. 50, No. 13, by the American College of Cardiology Foundation ISSN /07/$32. Journal of the American College of Cardiology Vol. 50, No. 13, 2007 2007 by the American College of Cardiology Foundation ISSN 0735-1097/07/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2007.04.096

More information

Implantation of a biventricular implantable cardioverter-defibrillator guided by an electroanatomic mapping system

Implantation of a biventricular implantable cardioverter-defibrillator guided by an electroanatomic mapping system Europace (2012) 14, 107 111 doi:10.1093/europace/eur250 CLINICAL RESEARCH Pacing and Resynchronization Therapy Implantation of a biventricular implantable cardioverter-defibrillator guided by an electroanatomic

More information

Biventricular Pacemakers Sensing

Biventricular Pacemakers Sensing Biventricular Pacemakers Sensing Sergio Dubner Director Arrhythmias and Electrophisiology Service, Clinica y Maternidad Suizo-Argentina, Buenos Aires, Argentina Left ventricular pacemaker lead technology

More information

Riata Lead Summit Jan 20, SJM Perspective Dr. Mark Carlson

Riata Lead Summit Jan 20, SJM Perspective Dr. Mark Carlson Riata Lead Summit Jan 20, 2012 SJM Perspective Dr. Mark Carlson 1 1 Presentation Outline Background, Characterization of Externalized Conductors Modeling and Testing Diagnostics and Therapeutics Recommendations

More information

Leadless pacemakers and subcutaneous ICD s: will we use them for most of our patients? K.-H. Kuck Asklepios Klinik St. Georg Hamburg, Germany

Leadless pacemakers and subcutaneous ICD s: will we use them for most of our patients? K.-H. Kuck Asklepios Klinik St. Georg Hamburg, Germany Leadless pacemakers and subcutaneous ICD s: will we use them for most of our patients? K.-H. Kuck Asklepios Klinik St. Georg Hamburg, Germany Disclosure Statement Research Grants Consultant / Advisory

More information

Primary prevention ICD recipients: the need for defibrillator back-up after an event-free first battery service-life

Primary prevention ICD recipients: the need for defibrillator back-up after an event-free first battery service-life Chapter 3 Primary prevention ICD recipients: the need for defibrillator back-up after an event-free first battery service-life Guido H. van Welsenes, MS, Johannes B. van Rees, MD, Joep Thijssen, MD, Serge

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 4,100 116,000 120M Open access books available International authors and editors Downloads Our

More information

Defibrillation threshold testing should no longer be performed: contra

Defibrillation threshold testing should no longer be performed: contra Defibrillation threshold testing should no longer be performed: contra Andreas Goette St. Vincenz-Hospital Paderborn Dept. of Cardiology and Intensive Care Medicine Germany No conflict of interest to disclose

More information

Biomedical Instrumentation

Biomedical Instrumentation Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Therapeutic and Prosthetic Devices 1 Figure 13.1 Block diagram of an asynchronous

More information

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Patient Scheduling and Screening Q: Is there any special (other than regular MRI consent) consent form that should be signed by the patient?

More information

His Bundle Pacing in Bundle Branch Block May 11, 2017

His Bundle Pacing in Bundle Branch Block May 11, 2017 His Bundle Pacing in Bundle Branch Block May 11, 2017 Gopi Dandamudi, MD FHRS System Medical Director, IUH Cardiac EP Program Director, IUH Atrial Fibrillation Center Assistant Professor of Clinical Medicine

More information

Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report

Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report Cardiol Ther (2017) 6:139 143 DOI 10.1007/s40119-017-0086-6 CASE REPORT Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report Abhishek Shenoy. Abhishek

More information

John H. Ip, M.D FACC Associate Professor of Medicine Michigan State University Medical Director, EP Service

John H. Ip, M.D FACC Associate Professor of Medicine Michigan State University Medical Director, EP Service ICD Trouble shooting: Case study John H. Ip, M.D FACC Associate Professor of Medicine Michigan State University Medical Director, EP Service Bullet proof evidence: ICD does prevent sudden death Disclosures

More information

Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention

Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention Dr Ngai-Yin Chan, MBBS, FRCP(Lond), FRCP(Edin), FRCP(Glasg), FACC, FHRS, Consultant Physician, Department of Medicine & Geriatrics, Princess

More information

Cardiac Devices CRT,ICD: Who is and is not a Candidate? Who Decides

Cardiac Devices CRT,ICD: Who is and is not a Candidate? Who Decides Cardiac Devices CRT,ICD: Who is and is not a Candidate? Who Decides Colette Seifer MB(Hons) FRCP(UK) Associate Professor, University of Manitoba, Cardiologist, Cardiac Sciences Program, St Boniface Hospital

More information

Apparent Failure of Automated Threshold Evaluation: What is the Mechanism? Vikas Kalra, M.B.B.S. Lynne Foreman, R.N. Deepak Bhakta, M.D.

Apparent Failure of Automated Threshold Evaluation: What is the Mechanism? Vikas Kalra, M.B.B.S. Lynne Foreman, R.N. Deepak Bhakta, M.D. Apparent Failure of Automated Threshold Evaluation: What is the Mechanism? Vikas Kalra, M.B.B.S. Lynne Foreman, R.N. Deepak Bhakta, M.D. Case Summary A 57-year-old man developed severe symptomatic tricuspid

More information

The Management of Heart Failure after Biventricular Pacing

The Management of Heart Failure after Biventricular Pacing The Management of Heart Failure after Biventricular Pacing Juan M. Aranda, Jr., MD University of Florida College of Medicine, Division of Cardiovascular Medicine, Gainesville, Florida Approximately 271,000

More information

Intermittent episodes of paced tachycardia: what is the cause?

Intermittent episodes of paced tachycardia: what is the cause? Intermittent episodes of paced tachycardia: what is the cause? Lindsey Parkinson and Lucy Broadhurst, Rotherham NHS Foundation Trust Contact: Lindsey.Parkinson@rothgen.nhs.uk Introduction A hospital in-patient

More information

CONTAK RENEWAL 3 AVT CLINICAL SUMMARY

CONTAK RENEWAL 3 AVT CLINICAL SUMMARY CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY CONTAK RENEWAL 3 AVT Boston Scientific

More information

Review of Pacemakers and ICD Therapy: Overview and Patient Management

Review of Pacemakers and ICD Therapy: Overview and Patient Management Review of Pacemakers and ICD Therapy: Overview and Patient Management Pacing Systems Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services OSU Division

More information

Device Diagnostics and Stroke Prevention: State of the Art

Device Diagnostics and Stroke Prevention: State of the Art Device Diagnostics and Stroke Prevention: State of the Art Ass. Prof. Dr. Carsten W. Israel Dept. of Cardiology Evangelical Hospital Bielefeld Germany C.W.Israel@em.uni-frankfurt.de Disclosure of Relationships

More information

A Rare Type of Pacemaker Mediated Tachycardia

A Rare Type of Pacemaker Mediated Tachycardia A Rare Type of Pacemaker Mediated Tachycardia Serkan ÇAY M.D., Ümit GÜRAY M.D., Cihan DEMİR M.D. Yuksek Ihtisas Heart-Education and Research Hospital, Department of Cardiology, Ankara, Turkey ABSTRACT

More information

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism?

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? CASE REPORTS Arrhythmia 2015;16(3):173-177 doi: http://dx.doi.org/10.18501/arrhythmia.2015.029 Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? Eun-Sun Jin, MD, PhD Cardiovascular

More information

Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm

Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm April 2000 107 Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm B. MERKELY Semmelweis University, Dept. of Cardiovascular Surgery,

More information

Devices and Other Non- Pharmacologic Therapy in CHF. Angel R. Leon, MD FACC Division of Cardiology Emory University School of Medicine

Devices and Other Non- Pharmacologic Therapy in CHF. Angel R. Leon, MD FACC Division of Cardiology Emory University School of Medicine Devices and Other Non- Pharmacologic Therapy in CHF Angel R. Leon, MD FACC Division of Cardiology Emory University School of Medicine Disclosure None University of Miami vs. OSU Renegade Miami football

More information

ESC Stockholm Arrhythmias & pacing

ESC Stockholm Arrhythmias & pacing ESC Stockholm 2010 Take Home Messages for Practitioners Arrhythmias & pacing Prof. Panos E. Vardas Professor of Cardiology Heraklion University Hospital Crete, Greece Disclosures Small teaching fees from

More information

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Case Report Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Taku Nishida MD, Tamio Nakajima MD, PhD, Yutaka Goryo MD, Ken-ichi Ishigami MD, PhD, Hiroyuki Kawata

More information

Inappropriate electrical shocks: Tackling the beast

Inappropriate electrical shocks: Tackling the beast ESC Paris 2011 Inappropriate electrical shocks: Tackling the beast Gerhard Hindricks University of Leipzig Heart Center Dept. of Electrophysiology ESC Paris 2011 Inappropriate electrical shocks: Tackling

More information

Preoperative Evaluation: Patients with Cardiac Disease

Preoperative Evaluation: Patients with Cardiac Disease Advances in Internal Medicine 2012 Preoperative Evaluation: Patients with Cardiac Disease Mary O. Gray, MD Professor of Medicine UC San Francisco Circulation 2007:100:e418-e500 (1) Cardiac Risk Assessment

More information

A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support

A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support 100 April 2000 A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support A. SH. REVISHVILI A.N. Bakoulev Institute of Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow,

More information

Journal of the American College of Cardiology Vol. 52, No. 23, by the American College of Cardiology Foundation ISSN /08/$34.

Journal of the American College of Cardiology Vol. 52, No. 23, by the American College of Cardiology Foundation ISSN /08/$34. Journal of the American College of Cardiology Vol. 52, No. 23, 2008 2008 by the American College of Cardiology Foundation ISSN 0735-1097/08/$34.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2008.08.027

More information

SARASOTA MEMORIAL HOSPITAL

SARASOTA MEMORIAL HOSPITAL SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE PACEMAKERS AND DETERMINING PACEMAKER DATE: REVIEWED: PAGES: 10/99 11/18 1 of 7 PS1094 ISSUED FOR: Nursing RESPONSIBILITY: *Qualified RN PURPOSE: To establish

More information

Continuous ECG telemonitoring with implantable devices: the expected clinical benefits

Continuous ECG telemonitoring with implantable devices: the expected clinical benefits Continuous ECG telemonitoring with implantable devices: the expected clinical benefits C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Declaration of

More information

Clinical and Electrocardiographic Characteristics of Patients with Brugada Syndrome: Report of Five Cases of Documented Ventricular Fibrillation

Clinical and Electrocardiographic Characteristics of Patients with Brugada Syndrome: Report of Five Cases of Documented Ventricular Fibrillation J Arrhythmia Vol 25 No 1 2009 Original Article Clinical and Electrocardiographic Characteristics of Patients with Brugada Syndrome: Report of Five Cases of Documented Ventricular Fibrillation Seiji Takashio

More information