1 Methods of Implantable Cardioverter-Defibrillator- Pacemaker Insertion to Avoid Interactions Henry M. Spotnitz, MD, Gary Y. Ott, MD, J. Thomas Bigger, Jr, MD, Jonathan S. Steinberg, MD, and Frank Livelli, Jr, MD Departments of Surgery and Medicine, Columbia-Presbyterian Medical Center, New York, New York Experience with combined transvenous pacemaker and implantable cardioverter-defibrillator insertion in 21 patients is described. Special techniques are needed to avoid potentially lethal pacemaker-implantable cardioverter-defibrillator interaction. Separation between leads for the two devices should be maximized. The electrophysiologic criteria for successful device function can be met, even when some leads for both devices must be placed by a transvenous approach. (Ann Thorac Surg 1992;53:253-7) mplantable cardioverter-defibrillators (ICDs) markedly I reduce arrhythmic death in survivors of cardiac arrest [l-31. Because the majority of patients who are candidates for ICD implantation suffer from advanced heart disease [l, 21, some patients who undergo ICD implantation will also require permanent pacemakers for sinus bradycardia, sinus arrest, or atrioventricular block [4,5]. The indication for an ICD or pacemaker may also be discovered after one or the other has already been implanted. Maximum safety is promoted in combined implantation by making pacemaker output invisible to the ICD, thereby avoiding inappropriate ICD discharges and failure of the ICD to recognize and respond to ventricular fibrillation. We describe methods that we have found useful in achieving this goal. Material and Methods We have previously described our approach to ICD implantation [6]. These patients are generally, but not always [2, 31, survivors of cardiac arrest in whom antiarrhythmic drugs are ineffective, poorly tolerated, or untestable. The need for transvenous pacing may be discovered in the course of electrophysiologic testing for ICD implantation. Some patients who are candidates for an ICD will have previously undergone pacemaker insertion. Other patients will be discovered to need a pacemaker after ICD insertion. The technique described here is best suited for patients undergoing ICD insertion through a left thoracotomy without simultaneous coronary bypass grafting. General anesthesia, maintenance of heart rate with drugs or temporary pacing, excellent fluoroscopy, and Accepted for publication Aug 1, Presented in part at the annual meeting of the New York Society for Thoracic Surgery, Nov Dr Ott s present address is University of Oregon Health Sciences Center, Portland, OR. Dr Steinberg s present address is St. Luke s-roosevelt Hospital, New York, NY. Address reprint requests to Dr Spotnitz, Columbia-Presbyterian Medical Center, 630 West 168th St, New York, NY electrocardiographic monitoring are essential. Adhesive external defibrillation electrodes are applied to the anterior right chest and posterolateral left chest. A fluoroscopy table must be used, and careful positioning is required to allow successful fluoroscopy and left thoracotomy in the presence of adhesive external defibrillation electrodes. We begin with a standard approach to transvenous pacing, exposing a cephalic or other small venous branch through a 3- to 4-cm incision beneath the left clavicle. Pacing wires are introduced directly through a small venotomy or through a 7F strip-away percutaneous catheter introducer (No ; Daig Corp, Minnetonka, MN) passed over a small guidewire. Fluoroscopy is then used to insert ventricular or atrial pacing wires as needed. The pacemaker system must be bipolar [v. For ventricular pacing, we prefer two unipolar, positive fixation, screw-in permanent pacing electrodes (Model ; Intermedics, Inc, Freeport, TX) positioned close together in the outflow tract of the right ventricle (RV) or in the anterior RV, just across the tricuspid valve (Figs 1, 2). Adequacy of sensing and pacing thresholds is confirmed before installing the pacemaker generator. A pacemaker system analyzer is used to determine thresholds for each unipolar lead, seeking pacing thresholds of less than 0.7 V, an R wave greater than 5 mv, and impedance of approximately 500 R. The leads are then tested in the bipolar configuration, usually providing pacing thresholds and R-wave amplitude similar to the unipolar configuration, except that bipolar impedance is about 1,000 R. Both possible bipolar polarities are tested, and the configuration giving the lower pacing threshold is identified with a marking tie on the negative lead. The leads are connected to a bipolar pacing generator. An adapter is usually required for these connections, because most current pacemakers use in-line bipolar connectors. Y adapters (5 mm to vs-1 or is-1) are available from Medtronic, CPI, Telectronics, and other manufacturers. Similar techniques of insertion, testing, and connection of paired atrial leads are employed when atrial or dual by The Society of Thoracic Surgeons /92/$5.00

2 254 SPOTNITZ ET AL Ann Thorac Surg 1992;53:25>7 A C B Fig 1. Posteroanterior (A) and lateral (C) roentgenograms reveal transvenous bipolar VVI pacemaker and epicardial ZCD inserted by left thoracotomy in a 56-year-old woman. Schematic diagrams (B, D) identify two positive-fixation pacemaker leads in the right ventricle, two myocardial ICD rate-sensing leads, and two large extrapericardial patch leads. D chamber pacing is indicated. Leads are placed in the atrial appendage or lateral right atrium as required. A left anterolateral thoracotomy incision is then performed, usually excising a small (7 cm) segment of the fifth rib. IntrapericardiaYmyocardial rate-sensing electrodes and extrapericardial large patch electrodes for the ICD are installed. Alternatively, the rate-sensing leads and a spring-coil defibrillator lead may be inserted through the subclavian vein (see Fig 2). During transvenous pacing at maximum pacemaker output and pulse width, the electrograms in both the rate-sensing and morphology leads are examined to confirm that pacemaker spikes are less than 0.2 mv in amplitude or less than 5% of the QRS amplitude in the rate-sensing leads. If these criteria are not met, either the pacemaker or ICD leads are repositioned (Fig 3). After lead positioning, cardioversion thresholds are measured by our electrophysiologists (J.T.B., J.S.S., F.L.) in ventricular tachycardia (when inducible) and in ventricular fibrillation, with transesophageal or epicardial echocardiographic monitoring of wall motion and ejection fraction when possible, as we have described previously [4]. Defibrillation thresholds are measured with an external cardioverter defibrillator (CPI) and must be 20 J or less. External defibrillator backup and availability of internal paddles can be critical if thresholds are unexpectedly high. Implantable cardioverter-defibrillator lead configurations are adjusted until defibrillation thresholds are adequate. A hemostatic, abdominal subcutaneous pocket is fashioned through a left upper quadrant incision. The ICD leads are tunneled through the diaphragm and attached

3 Ann Thorac Surg 1992;53:25>7 SPOTNITZ ET AL 255 A C SprlnQ COII Defibrlllator Lead, SVC Blmlar Rate Sensing Bipolar Rate Sensing Lead, RV Apex B Large Patch Electrode. Leads, RV%utflow Tract Fig 2. Posteroanterior (A) and lateral (C) roentgenograms reveal transvenous bipolar VVZ pacemaker and ICD in a 60-year-old man with previous coronary artery bypass grafting. Schematic diagrams (B, D) identify leads. Two positive-fixation pacemaker leads are visible in the right ventricular outflow tract. Technical considerations led also to transvenous insertion of an endocardial, bipolar rate-sensing lead and a spring-coil defibrillator lead in the superior vena cava. One large patch was inserted in an extrapericardial location through a left thoracotomy. Successful function of this lead configuration is illustrated in Figure 4. (RV = right ventricular; SVC = superior vena cava.) D - to the ICD, which is activated and tested. Our experience encompasses AICD models AID-B (Intek) and Ventak models 1510, 1520, 1550, and 1600 (CPI). The most stringent test is successful ICD detection and cardioversion from ventricular fibrillation during magnet-forced asynchronous ventricular or asynchronous atrioventricular pacing at maximum pacemaker output and pulse width (Fig 4). The absence of double counting of pacing spikes and QRS complex during asynchronous ventricular or asynchronous atrioventricular pacing should also be confirmed by listening to the R-wave synchronous tones emitted when a magnet is applied to an activated ICD. Finally, the ICD is temporarily inactivated and all incisions are closed, incorporating a small left chest tube. The patient is generally awakened and extubated in the operating room. Epidural fentanyl analgesia is administered in the intensive care unit for 24 to 48 hours. All devices are again tested in the electrophysiology laboratory 1 week postoperatively before discharge. Results Our clinical experience includes 114 ICD system insertions since 1984, with one infarction-related perioperative

4 256 SPOTNITZ ET AL Ann Thorac Surg 1!392;5325>7 Fig 3. The pacing spikes illustrated in the left panel were judged unacceptably large in the rate-sensing leads. After repositioning, as shown on the right, the pacemaker spikes decreased in size, the amplitude of the native QRS increased, and the lead position was considered adequate. LA Initial Recordings - Output 8.lv PM Leads Repositioned - 8.lv mortality. At mean follow-up of 2.9 years, only one arrhythmic death may have occurred. Twenty-one of these patients have required ICDs in combination with pacemakers (19) or pacing leads (2). Pacemakers implanted consisted of 12 WI, 6 DDD, and one antitachycardia pacemaker. In 3 patients, the need for transvenous pacing was established and pacemaker insertion occurred weeks or months after ICD insertion. In 12 patients, ICD implantation was combined with pacemaker insertion (10) or lead insertion only (2) using the methods described. Five additional patients were referred for ICD insertion with previously inserted pacemakers; in 2 of these patients (DDD) satisfactory function of both devices was achieved by moving the ICD rate-sensing leads as far as possible to the posterolateral aspect of the left ventricle. The remaining 3 patients required new transvenous leads and testing, as described in the Material and Methods section. Fig 4. Successful test of leads illustrated in Figure 2. Continuous right ventricular bipolar pacing is forced by a magnet applied to the pacemaker. The pacing spikes do not prevent detection and cardioversion by the ICD. The pacing spikes are invisible except in the bipolar morphology lead (patch and superior vena cava electrodes), where they measure less than 0.4 mv. (AICD = automatic implantable cardioverter-defibrillator; VF = ventricular fibrillation; VOO = asynchronous ventricular.) - AlCD CardioverSiOn VF i Overall, we found it necessary to reposition or replace ventricular pacemaker leads four times-twice from the RV apex to the outflow tract and twice from outflow tract to the anterior RV, close to the tricuspid valve. In 1 patient with DDD pacing and chronically implanted unipolar leads in the right atrial appendage and RV apex, successful conversion to bipolar pacing was accomplished by placing a second unipolar lead close to the chronic leads and proceeding as described. We have found that either RV or right atrial pacing wires can cause interactions with ICDs. Accordingly, both lead systems must be individually tested. Comment The methods and testing described here were not used in one patient, our first pacemaker recipient, who received a bipolar DDD pacemaker. That patient is the only one in Rate Sensing Lead I f I I I I I I I I I I I I I Ill11 II I I I I I I I I I I I I I I I I I I I I I I I1 I I I I\ I I I I I I1 I I

5 Ann Thorac Surg 1992;5325>7 SPOTNlTZ ET AL 257 our series of 114 ICD recipients who may have suffered arrhythmic death. The circumstances of that event are not fully known, as the patient died while sleeping alone, at home. Nevertheless, that experience leads us to recommend that all steps to avoid pacemaker-icd interaction be followed stringently. Pacemaker-ICD interaction can cause either unnecessary ICD discharges or inappropriate ICD inhibition permitting sudden death [3]. Unnecessary ICD discharges can occur if the highly sensitive circuits of the ICD detect both the pacing spike and the QRS depolarization in W I pacing (double counting). Similarly, triple counting can occur (atrial spike, ventricular spike, and QRS) in DDD or DVI pacing. In both cases, a paced rate of 90 beats/min will exceed the common ICD rate cutoff of 173 and lead to ICD firing [ 81. More ominous is the potential problem in ventricular fibrillation. Ventricular fibrillation may not inhibit ventricular pacing [4]. The ICD may not detect ventricular fibrillation in the presence of unipolar ventricular pacing spikes if the amplitude of the pacing spikes substantially exceeds the low amplitude of the waves of ventricular fibrillation. The ICD thus may interpret pacemaker spikes as a normal ventricular rhythm, a phenomenon that has been observed in patients [4]. Other adverse ICD-pacemaker experiences described in the literature include potentially lethal reciprocal triggering of an ICD and antitachycardia pacemaker [8], inadvertent induction of ventricular tachycardia triggered by an "R on T" phenomenon during a pacemaker magnet test [8] or burst pacing [9], and transient sensing and pacing dysfunction of transvenous pacemakers after ICD discharges [4]. Many of these problems will be eliminated when devices integrating ICD and pacemaker functions in a single unit are released, hopefully within 2 to 3 years. Many difficulties of ICD-pacemaker interaction can be avoided with present, independent devices by making the pacemaker signals invisible to the ICD. Unipolar pacing is contraindicated in this setting because of the large pacing spikes generated [;7. Recently, it has been recommended that pacemakers implanted in ICD recipients should be capable of function only in the bipolar mode, because pacemakers capable of function in the unipolar mode could be inadvertently programmed to the unipolar mode or may use that mode as an indicator of battery depletion. The optimal lead configuration consists of bipolar RV endocardial pacing electrodes and bipolar left ventricular epicardial ICD electrodes. In-line bipolar pacemaker leads can be useful but are stiff in comparison with unipolar leads and, as we have observed, may not provide good R-wave amplitude in the outflow tract. The increased flexibility of unipolar positive-fixation leads promotes stability of pacemaker leads in locations such as the RV outflow tract and anterior RV, which we prefer. We have observed no displacement of more than 30 leads implanted in these locations with the technique and lead described. Others have also described successful experiences with outflow tract pacing [lo]. We have found moving the epicardial rate-sensing leads, described by others [8], useful to minimize ICD-pacemaker interaction. But we find the ability to move twin unipolar leads widely within the RV to be even more useful. This has even allowed endocardial ICD rate-sensing leads and pacemaker leads to be combined (see Fig 2). Successful operation of this lead configuration is illustrated in Figure 3. In summary, the need for simultaneous insertion of both an ICD and a permanent pacemaker is a problem created in part by delayed development of a single device capable of both functions. The techniques must be well understood, because bradycardic death in ICD recipients has been reported [5]. However, even when dual-function devices become available, special situations will arise that require familiarity with the techniques described. Detection of pacemaker spikes by the ICD can lead to either inappropriate ICD discharges or failure of the ICD to detect and fire in ventricular fibrillation. These problems can be avoided by use of bipolar pacemakers and lead configurations physically separating the lead systems. Supported in part by US Public Health Service grant HL References 1. Winkle RA, Mead RH, Ruder MA, et al. Long-term outcome with the automatic implantable cardioverter-defibrillator. J Am Coll Cardiol 1989;13: Lehman MH, Saksena S. NASPE policy statement. Implantable cardioverter defibrillators in cardiovascular practice: report of the policy conference of the North American Society of Pacing and Electrophysiology. PACE 1991; Kral MA, Spotnitz HM, Hordof A, Bigger JT Jr, Steinberg JS, Livelli FD Jr. Automatic implantable cardioverter defibrillator implantation for malignant ventricular arrhythmias associated with congenital heart disease. Am J Cardiol 1989;63: Cohen AI, Wish MH, Fletcher RD, et al. The use and interaction of permanent pacemakers and the automatic implantable cardioverter defibrillator. PACE 1988;11: Khastgir T, Aarons D, Veltri E. Sudden bradyarrhythmic death in patients with the implantable cardioverterdefibrillator: report of two cases. PACE 1991; Antunes ML, Spotnitz HM, Steinberg JS, Bigger JT Jr, Livelli FD Jr. Measurement of ejection fraction during intraoperative electrophysiologic testing of the automatic implantable cardioverter-defibrillator. Arrhythmia Clin 1988;5:1% Kim SG, Furman S, Waspe LE, Brodman R, Fisher JD. Unipolar pacer artifacts induced failure of an automatic implantable cardioverter/defibrillator to detect ventricular fibrillation. Am J Cardiol 1986; Epstein AE, Kay GN, Plumb VJ, Shepard RB, Kirklin JK. Combined automatic implantable cardioverter-defibrillator and pacemaker systems: implantation techniques and followup. J Am Coll Cardiol 1989;13: Ahem TS, Nydegger C, McCormick DJ, et al. Device interaction-antitachycardia pacemakers and defibrillators for sustained ventricular tachycardia. PACE 1991;14: Barin SE, Jones SM, Ward DE, Camm AJ, Nathan AW. The right ventricular outflow tract as an alternative permanent pacing site: long-term follow-up. PACE 1991;14:34.

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm

Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm April 2000 107 Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm B. MERKELY Semmelweis University, Dept. of Cardiovascular Surgery,

More information

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul,

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul, Mayo Clin Proc, August 2002, Vol 77 Unintentional Deactivation of ICDs 855 Case Report Unintentional Deactivation of Implantable Cardioverter-Defibrillators in Health Care Settings MARY JANE RASMUSSEN,

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

You have a what, inside you?

You have a what, inside you? Costal Emergency Medicine Conference You have a what, inside you? Less than mainstream medical devices encountered in the ED. Eric Ossmann, MD, FACEP Associate Professor Duke University Medical Center

More information

You have a what, inside you?

You have a what, inside you? Costal Emergency Medicine Conference You have a what, inside you? Less than mainstream medical devices encountered in the ED. Eric Ossmann, MD, FACEP Associate Professor Duke University Medical Center

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

Different indications for pacemaker implantation are the following:

Different indications for pacemaker implantation are the following: Patient Resources: ICD/Pacemaker Overview ICD/Pacemaker Overview What is a pacemaker? A pacemaker is a device that uses low energy electrical pulses to prompt the heart to beat whenever a pause in the

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

SUPPLEMENTAL MATERIAL

SUPPLEMENTAL MATERIAL SUPPLEMENTAL MATERIAL Table S1. Patient Selection Codes, CIED Generator Procedures Code Type Code Description ICD9 Proc 00.51 Implantation of cardiac resynchronization defibrillator, total system [CRT-D]

More information

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Jason Collinson & Stuart Tan Essex Cardiothoracic Centre, Basildon and Thurrock University Hospital. Contact: jason.collinson@nhs.net

More information

Transvenous Pacemaker Procedures

Transvenous Pacemaker Procedures Cardiology: Pacemaker and Defibrillator Coding Presented By: Moderate Sedation 2017 99151 : under age 5, initial 15 minutes by MD performing intervention 99152: age 5 or older, initial 15 minutes by MD

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases

Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases Case Report Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases Koji Tsutsumi MD, Tatsuru Niibori MD, Keiichiro Katsumoto MD Department

More information

Cigna - Prior Authorization Procedure List Cardiology

Cigna - Prior Authorization Procedure List Cardiology Cigna - Prior Authorization Procedure List Cardiology Category CPT Code CPT Code Description 33206 Insertion of new or replacement of permanent pacemaker with transvenous electrode(s); atrial 33207 Insertion

More information

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular

More information

Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS

Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS NBG Codes 1st Letter 2nd Letter 3rd Letter A V D Chamber(s) Paced = atrium = ventricle = dual (both atrium and ventricle) Chamber(s) Sensed A = atrium

More information

Review of Pacemakers and ICD Therapy: Overview and Patient Management

Review of Pacemakers and ICD Therapy: Overview and Patient Management Review of Pacemakers and ICD Therapy: Overview and Patient Management Pacing Systems Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services OSU Division

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

Model 5392 EPG Temporary Pacer

Model 5392 EPG Temporary Pacer Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program Electrophysiology Implant Classification Table The table below contains the codes that apply to our UnitedHealthcare Medicare Advantage cardiology prior Description Includes Generator Placement Includes

More information

Dual-Chamber Implantable Cardioverter-Defibrillator

Dual-Chamber Implantable Cardioverter-Defibrillator February 1998 9 Dual-Chamber Implantable Cardioverter-Defibrillator A.SH. REVISHVILI A.N. Bakoulev Research Center for Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow, Russia Summary

More information

Implantation of the Automatic Defibrillator: The Subxiphoid Approach

Implantation of the Automatic Defibrillator: The Subxiphoid Approach Implantation of the Automatic Defibrillator: The Subxiphoid Approach Levi Watkins, Jr., M.D., M. Mirowski, M.D., Morton M. Mower, M.D., Philip R. Reid, M.D., Paul Freund, P.A., Andra Thomas, R.N., Myron

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL

More information

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment Karen L. Booth, MD, Lucile Packard Children s Hospital Arrhythmias are common after congenital heart surgery [1]. Postoperative electrolyte

More information

Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement

Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement 217 Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement Partha Prateem Choudhury, Vivek Chaturvedi, Saibal Mukhopadhyay, Jamal Yusuf Department

More information

MEDICAL POLICY Cardioverter Defibrillators

MEDICAL POLICY Cardioverter Defibrillators POLICY........ PG-0224 EFFECTIVE......06/01/09 LAST REVIEW... 01/27/17 MEDICAL POLICY Cardioverter Defibrillators GUIDELINES This policy does not certify benefits or authorization of benefits, which is

More information

evicore cardiology procedures and services requiring prior authorization

evicore cardiology procedures and services requiring prior authorization evicore cardiology procedures and services requiring prior authorization Moda Health Commercial Group and Individual Members* *Check EBT to verify member enrollment in evicore program Radiology Advanced

More information

Cardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand:

Cardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand: Cardiac Pacing CHAPTER 10 Learning outcomes To understand: The indications for cardiac pacing in the peri-arrest setting How to perform percussion pacing How to apply non-invasive, transcutaneous electrical

More information

Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children

Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children ERALD A. SERWER, M.D., MACDONALD DICK 11, M.D., KAREN UZARK, R.N., Ph.D., WILLIAM A. SCOTT, M.D., and EDWARD L.

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Implantable cardioverter defibrillators for the treatment of arrhythmias and cardiac resynchronisation therapy for the treatment of heart failure (review

More information

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! What is a

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART II

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART II 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART II 2012 Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 ADDITIVES IN DEVICE THERAPY Key Choice / CNEA 1 Adaptive Rate

More information

Need to Know: Implantable Devices. Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia

Need to Know: Implantable Devices. Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia Need to Know: Implantable Devices Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia Disclosure Statement I have no relationships to disclose. Objectives Discuss the most

More information

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. The Miracle of Living February 21, 2018 Matthew Ostrom MD,FACC,FHRS Division of

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program:

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: Notification and Prior Authorization Program: Electrophysiology Implant Classification Table The following chart contains the codes that require notification or prior authorization as part of UnitedHealthcare

More information

Pacing in patients with congenital heart disease: part 1

Pacing in patients with congenital heart disease: part 1 Pacing in patients with congenital heart disease: part 1 September 2013 Br J Cardiol 2013;20:117 20 doi: 10.5837/bjc/2013.028 Authors: Khaled Albouaini, Archana Rao, David Ramsdale View details Only a

More information

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure Inappropriate Discharges of Intravenous Implantable Cardioverter Defibrillators Owing to Lead Failure Takashi WASHIZUKA, 1 MD, Masaomi CHINUSHI, 1 MD, Ryu KAZAMA, 1 MD, Takashi HIRONO, 1 MD, Hiroshi WATANABE,

More information

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies

More information

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie 268 Case Report Cardiac Resynchronization Therapy in a Patient with Persistent Left Superior Vena Cava Draining into the Coronary Sinus and Absent Innominate Vein: A Case Report and Review of Literature

More information

RN-BC, MS, CCRN, FAHA

RN-BC, MS, CCRN, FAHA Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! 1 A. Atropine

More information

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds

More information

With the advent of implantable cardioverter defibrillators

With the advent of implantable cardioverter defibrillators One-Incision Approach for Insertion of Implantable Cardioverter Defibrillators Andreas Markewitz, MD, Hans Kaulbach, MD, Soren Mattke, MD, Uwe Dorwarth, MD, Christian Weinhold, MD, Ellen Hoffmann, MD,

More information

Pacemaker-ICD/Drug Interaction

Pacemaker-ICD/Drug Interaction Review Article Pacemaker-ICD/Drug Interaction Tsu-Juey Wu When a drug is prescribed for a patient with a permanent pacemaker or an implantable cardioverter defibrillator (ICD), consideration must be given

More information

Pediatric Pacemaker Implantation Endocardial or Epicardial

Pediatric Pacemaker Implantation Endocardial or Epicardial Pediatric Pacemaker Implantation Endocardial or Epicardial HAITHAM BADRAN, MD, FEHRA CONSULTANT OF INTERVENTIONAL CARDIOLOGY CONSULTANT OF CARDIAC PACING AND ELECTROPHYSIOLOGY LECTURER OF CARDIOLOGY AIN

More information

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Case Report Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Taku Nishida MD, Tamio Nakajima MD, PhD, Yutaka Goryo MD, Ken-ichi Ishigami MD, PhD, Hiroyuki Kawata

More information

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Pacing and Device Jargon Made Simple Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Disclosure Dr Jonathan Timperley Honorarium from Bayer Arne Larson (1915

More information

Implantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect

Implantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect The Ochsner Journal 10:27 31, 2010 f Academic Division of Ochsner Clinic Foundation Implantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect Anas Bitar, MD, Maria Malaya

More information

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt Leadless Pacing Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt The weakest link in Pacemaker system the lead. The more the leads the more the complications Dislodgement Fracture Insulation

More information

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program:

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: Electrophysiology Implant Code Classification Table The

More information

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing

More information

The Nuts and Bolts of ICD Therapy

The Nuts and Bolts of ICD Therapy Electrical Management of Cardiac Rhythm Disorders For Cardiology Fellows December 5-8 Austin, Texas The Nuts and Bolts of ICD Therapy 1 2 Action Potential Localized Differences in Conduction Conduction

More information

PERFORMANCE MEASURE TECHNICAL SPECIFICATIONS. HRS-3: Implantable Cardioverter-Defibrillator (ICD) Complications Rate

PERFORMANCE MEASURE TECHNICAL SPECIFICATIONS. HRS-3: Implantable Cardioverter-Defibrillator (ICD) Complications Rate PERFORMANCE MEASURE TECHNICAL SPECIFICATIONS HRS-3: Implantable Cardioverter-Defibrillator (ICD) Complications Rate Measure Title Description Measure Type Data Source Level of Analysis Numerator HRS-3:

More information

Epicardial Pacemaker Implantation for Complete Heart Block

Epicardial Pacemaker Implantation for Complete Heart Block for Complete Heart Block Donald G. Mulder, M.D., and C. Gordon Frank, M.D. S udden death or progressive cardiac failure threatens life in the patient with complete heart block. The slow heart rate and

More information

Biventricular Pacing - Hemodynamic Benefit for Patients with Congestive Heart Failure

Biventricular Pacing - Hemodynamic Benefit for Patients with Congestive Heart Failure 428 December 2000 Biventricular Pacing - Hemodynamic Benefit for Patients with Congestive Heart Failure K. MALINOWSKI Helios Clinics, Aue, Germany Summary Congestive heart failure afflicts a large and

More information

Repetitive narrow QRS tachycardia in a 61-year-old female patient with recent palpitations

Repetitive narrow QRS tachycardia in a 61-year-old female patient with recent palpitations Journal of Geriatric Cardiology (2018) 15: 193 198 2018 JGC All rights reserved; www.jgc301.com Case Report Open Access Repetitive narrow QRS tachycardia in a 61-year-old female patient with recent palpitations

More information

Experience with pectoral versus abdominal implantation of a small defibrillator

Experience with pectoral versus abdominal implantation of a small defibrillator European Heart Journal (1998) 19, 185 198 Article No. hj9816 Experience with pectoral versus abdominal implantation of a small defibrillator A multicenter comparison in 778 patients E. Hoffmann, G. Steinbeck,

More information

Permanent Pacemaker Implantation via the Iliac Vein: An Alternative in 4 Cases with Contraindications to the Pectoral Approach

Permanent Pacemaker Implantation via the Iliac Vein: An Alternative in 4 Cases with Contraindications to the Pectoral Approach Case Report Permanent Pacemaker Implantation via the Iliac Vein: An Alternative in 4 Cases with Contraindications to the Pectoral Approach Koji Tsutsumi MD, Kenichi Hashizume MD, Naritaka Kimura MD, Shinichi

More information

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΖΩΡΤΖ ΔΑΔΟΥΣ ΕΠΙΚΟΥΡΟΣ ΚΑΘΗΓΗΤΗΣ Α.Π.Θ. ΜΑΡΙΑ ΚΑΡΑΛΙΟΛΙΟΥ ΕΙΔΙΚΕΥΟΜΕΝΗ ΙΑΤΡΟΣ ΚΑΡΔΙΟΛΟΓΟΣ Β ΚΑΡΔΙΟΛΟΓΙΚΗ ΚΛΙΝΙΚΗ Α.Π.Θ. ΙΠΠΟΚΡΑΤΕΙΟ Γ.Ν.Θ. 5 Why a Pacemaker is Implanted

More information

IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT JANUARY 24, 2012

IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT JANUARY 24, 2012 IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT201203 JANUARY 24, 2012 The IHCP to reimburse implantable cardioverter defibrillators separately from outpatient implantation Effective March 1, 2012, the

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing D. J. McMahon 141001 cewood rev 2017-10-04 Key Points Defibrillators: - know the definition & electrical value

More information

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study ANEESH V. TOLAT, M.D.,* MELISSA WOICIECHOWSKI, M.S.N.,* ROSEMARIE KAHR, R.C.I.S.,* JOSEPH

More information

Finally, a pacemaker may be either permanent or temporary, which will also factor into your code selections.

Finally, a pacemaker may be either permanent or temporary, which will also factor into your code selections. 2015 Cardiology Survival Guide Chapter 15: Pacemakers With more than 30 codes and many variables to choose from, you must weigh your options carefully when reporting insertion, revision, or removal of

More information

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity. Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded

More information

Call Medtronic at 1 (800) to verify the patient s current implanted system

Call Medtronic at 1 (800) to verify the patient s current implanted system MRI SURESCAN SYSTEMS Patient Scanning Process Transvenous Implantable Cardiac Systems PATIENT PRESCREENING SureScan Pacing, Defibrillation, and CRT (CRT-D and CRT-P) Systems Verification Verify that patient

More information

EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology

EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important

More information

Technical option of surgical approach for trouble-shooting

Technical option of surgical approach for trouble-shooting JHRS Corner Device and lead trouble-shooting - standard strategy and technical option - Technical option of surgical approach for trouble-shooting Katsuhiko IMAI Department of Cardiovascular surgery, Hiroshima

More information

Three case studies have reported that patients have received

Three case studies have reported that patients have received Interactions Between Electronic Article Surveillance Systems and Implantable Cardioverter-Defibrillators William J. Groh, MD; Scott A. Boschee, BS; Erica D. Engelstein, MD; William M. Miles, MD; M. Erick

More information

Chapter 80 Implantable Cardiac Devices

Chapter 80 Implantable Cardiac Devices Chapter 80 Implantable Cardiac Devices Episode Overview: 1. List 5 Indications for permanent pacing 2. Pacemaker nomenclature - what do the 5 letters mean a. What are common Pacemaker types? 3. List the

More information

EXAMPLE PROTOCOL FOR PATIENT SCANNING PROCESS

EXAMPLE PROTOCOL FOR PATIENT SCANNING PROCESS EXAMPLE PROTOCOL FOR PATIENT SCANNING PROCESS MRI SureScan Pacing, ICD, and CRT (CRT-D and CRT-P) Transvenous Implantable Cardiac Systems Timothy S. E. Albert, M.D. President CVMR Solutions, Inc. Carmel

More information

Pacemakers, Defibrillators, and Electrophysiology in 2012

Pacemakers, Defibrillators, and Electrophysiology in 2012 Pacemakers, Defibrillators, and Electrophysiology in 2012 Presented by: David B. Dunn, MD, FACS, CIRCC, CPC-H, CCC National Coding Standards Sources of information Centers for Medicare & Medicaid (CMS)

More information

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable

More information

Transvenous Pacemaker Implantation 22 years after the Mustard Procedure

Transvenous Pacemaker Implantation 22 years after the Mustard Procedure Case Report Transvenous Pacemaker Implantation 22 years after the Mustard Procedure Masato Sakamoto MD, Yoshie Ochiai MD, Yutaka Imoto MD, Akira Sese MD, Mamie Watanabe MD, Kunitaka Joo MD Department of

More information

Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement

Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement www.ipej.org 49 Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement Naga VA Kommuri, MD, MRCPCH, Sri Lakshmi S Kollepara, MD,

More information

Rise in Chronic Defibrillation Energy Requirements Necessitating Implantable Defibrillator Lead System Revision

Rise in Chronic Defibrillation Energy Requirements Necessitating Implantable Defibrillator Lead System Revision Rise in Chronic Defibrillation Energy Requirements Necessitating Implantable Defibrillator Lead System Revision EMILE G. DAOUD, K. CHING MAN, ERED MORADY, and S. ADAM STRIGKBERGER From the Division of

More information

Ventricular Tachycardia in Structurally Normal Hearts (Idiopathic VT) Patient Information

Ventricular Tachycardia in Structurally Normal Hearts (Idiopathic VT) Patient Information Melbourne Heart Rhythm Ventricular Tachycardia in Structurally Normal Hearts (Idiopathic VT) Patient Information What is Ventricular Tachycardia? Ventricular tachycardia (VT) is an abnormal rapid heart

More information

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia By Sandeep Joshi, MD and Jonathan S. Steinberg, MD Arrhythmia Service, Division of Cardiology

More information

Leadless pacing: Technical challenges and concerns. Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital

Leadless pacing: Technical challenges and concerns. Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital Leadless pacing: Technical challenges and concerns Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital Modest consultancy fees/honoraria from Bayer Boehringer Ingelheim

More information

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use DEFIBRILLATOR OPTION General Information Introduction M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for defibrillation and cardioversion.

More information

Temporary Pacemakers

Temporary Pacemakers 07 Part I Anaesthesia Refresher Course 2018 University of Cape Town Dr Owen Porrill Private Practice Honorary lecturer- University of Cape Town Temporary pacemakers are usually inserted as part of the

More information

Practice Questions.

Practice Questions. IBHRE Prep Practice Questions Question 1 The relative refractory yperiod of the ventricular myocardium corresponds to which of the following phases of the action potential? A. (0) B. (1) C. (2) D. (3)

More information

When to implant an ICD in systemic right ventricle?

When to implant an ICD in systemic right ventricle? When to implant an ICD in systemic right ventricle? Département de rythmologie et de stimulation cardiaque Nicolas Combes n.combes@clinique-pasteur.com Pôle de cardiologie pédiatrique et congénitale Risk

More information

Leadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece

Leadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece Leadless pacing P.E. Vardas Professor of Cardiology University of Crete, Greece DISCLOSURES Small teaching fees from: Bayer/J&J, Boeringer, Menarini Small consulting fees from: Medtronic 1957: First External

More information

T lator (ICD) in the nonpharmacological treatment of

T lator (ICD) in the nonpharmacological treatment of Implantation of Cardioverter Defibrillators in the Post-S ternotomy Patient Ralph J. Damiano, Jr, MD, Andrew H. Foster, MD, Kenneth A. Ellenbogen, MD, Mark A. Wood, MD, Bruce S. Stambler, MD, William J.

More information

Ass. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital

Ass. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital Treatment Bradyarrhythmia Ass. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital 1 What is the rhythm? Sinus Bradycardia What s rhythm Sinus Bradycardia Treatment Asymptomatic No treatment

More information

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. Defibrillators OBJECTIVE 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. 4. Types and classes of defibrillator 5. Describe

More information

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

The heart's natural pacemaker is called the sinoatrial (SA) node or sinus node. PACEMAKER Natural pacemaker: The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node. Artificial pacemaker: It is a small, battery-operated device that helps the heart beat in

More information

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II). Cardiac Pacemaker Premature Beats When one of ectopic foci becomes irritable, it may spontaneously fire, leading to one or more premature beats. Atrial and junctional foci may become irritable from excess

More information

A Mechanism for "False" Inhibition of Demand Pacemakers

A Mechanism for False Inhibition of Demand Pacemakers A Mechanism for "False" Inhibition of Demand Pacemakers By KENNETH C. LASSETER, M.D., JACK W. BUCHANAN, JR., M.S.E.E., AND KARL F. YOSHONIS, M.D. SUMMARY Certain variations in discharge rate of demand

More information

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION 1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,

More information

Summary of the Final 2013 Policy, Payment Changes For Hospital Outpatient Departments

Summary of the Final 2013 Policy, Payment Changes For Hospital Outpatient Departments Summary of the Final 2013 Policy, Payment Changes For Hospital Outpatient Departments A. Introduction On November 1, 2012, the Centers for Medicare & Medicaid Services (CMS) issued a final rule that will

More information

Heart Pacemakers On Chest X Ray Revisited: What the radiologist has to know

Heart Pacemakers On Chest X Ray Revisited: What the radiologist has to know Heart Pacemakers On Chest X Ray Revisited: What the radiologist has to know Poster No.: C-2314 Congress: ECR 2015 Type: Educational Exhibit Authors: J. Peña Conde, J. De Luis - Yanes, M. Martinez-Perez,

More information

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE Interventional procedure consultation document Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death A subcutaneous

More information

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs).

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs). Management of Children with Cardiac Devices Guideline originally developed by Leann Miles, APRN; Lindsey Pumphrey, RN; Srikant Das, MD, and the ANGELS Team. Last reviewed by Lindsey Pumphrey, RN, Srikant

More information