TWIDDLER S syndrome is characterized by coiling of the pacemaker lead due

Size: px
Start display at page:

Download "TWIDDLER S syndrome is characterized by coiling of the pacemaker lead due"

Transcription

1 Reel Syndrome and Pulsatile Liver in a Patient With a Two-chamber Pacemaker Ahmet VURAL, 1 MD, Aysen AGACDIKEN, 1 MD, Dilek URAL, 1 MD, and Baki KOMSUOGLU, 1 MD, FESC SUMMARY Twiddler s syndrome is characterized by coiling of the pacemaker lead due to the rotation of the pacemaker generator on its long axis. Reel syndrome is another form of Twiddler's syndrome. It occurs due to the rotation of the pacemaker generator on its transverse axis with subsequent coiling of the pacemaker leads around the pulse generator. In this article we describe a patient with a two-chamber pacemaker who presented with sudden onset of abdominal pulsation and was subsequently diagnosed as Reel syndrome. To the best of our knowledge, this case is the first case of Reel syndrome that developed in a patient with a two-chamber pacemaker. (Jpn Heart J 2004; 45: ) Key words: Twiddler s syndrome, Reel syndrome, Pulsatile liver TWIDDLER S syndrome is characterized by coiling of the pacemaker lead due to the rotation of the pacemaker generator on its long axis. Although a loose subcutaneous pocket is the main reason for the rotation of the pacemaker generator, its size and weight may also play an important role in the development of this syndrome. Since its first description by Bayliss, et al in 1968, many interesting cases with Twiddler's syndrome have been published. 1) It is frequently observed in overweight female patients with loose, fatty subcutaneous tissue. 2) Rotation of the generator may occur by the patient s pathologic manipulations or spontaneously. 3) Lead coiling is the manifestation of the rotation of the generator on its long axis. It is usually observed close to the generator and sometimes intracardially. 4,5) It may cause lead dislocation or fraction and these in turn may cause life-threatening symptoms in patients with pacemaker dependence. 6,7) Reel syndrome is another form of Twiddler s syndrome. It occurs due to the rotation of the pacemaker generator on its transverse axis with subsequent coiling of the pacemaker leads around the pulse generator. It was described in a patient with a single-chamber pacemaker, which had been implanted because of the diag- From the 1 Department of Cardiology, Kocaeli University Medical Faculty, Kocaeli, Turkey. Address for correspondence: Ahmet Vural, MD, Department of Cardiology, Kocaeli University Medical Faculty, Fatih Mahallesi, Seymen Doktorlar sit. B-blok, D-7, Kurucesme, Kocaeli, Turkey. Received for publication November 5, Revised and accepted April 8,

2 1038 VURAL, ET AL Jpn Heart J November 2004 nosis of presyncope, a low ventricular rate, atrial fibrillation, and previous mitral valve replacement due to rheumatic heart valve disease. 8) Similar to Twiddler's syndrome, Reel syndrome may also lead to electrode dislocation or fracture followed by several clinical symptoms such as presyncope, syncope, or even death in patients with absolute pacemaker dependence. In this report, we describe a patient with a two-chamber pacemaker who presented with sudden onset abdominal pulsation and was subsequently diagnosed as having Reel syndrome. CASE REPORT A 68-year-old man was admitted to our emergency department because of sudden onset of abdominal pulsation. He had been complaining of fatigue and lightheadness for 10 years. After his evaluation at another hospital he had been diagnosed with sick sinus syndrome and a DDDR pacemaker [Medtronic 5594 atrial lead, 5092 ventricular lead, KDR 701VPP pacemaker generator (Medtronic, Inc., Minneapolis, MN, USA)] was implanted. He did well until 1 month after implantation when an abdominal pulsation developed suddenly. On physical examination his arterial blood pressure was 120/70 mmhg and his pulse rate 75/min. Cardiac auscultation findings were normal. A pulsation in his abdomen was observed on inspection and a regular pulsation in the liver was palpated. On ECG, pacing spikes at the beginning of the P waves and in the QRS complexes and rare ventricular ectopic beats were observed. P waves were thought to have been developed by pacing and QRS morphology by fusion pacemaker beats or normal ventricular activation-ineffective pacing spikes. Pacemaker interrogation showed effective atrial and ineffective ventricular pacing. Atrial pacing beats were regularly conducted to the ventricle (Figure 1). The pacemaker mode was changed to AAI mode and the patient s symptoms disappeared. In the evaluation of bedside chest radiography, which had been obtained just after implantation at the previous hospital, the atrial and ventricular leads and pulse generator were in normal positions (Figure 2). On chest X-rays taken at our university hospital, the atrial lead was straightened, the ventricular lead was in the vena cava superior, and leads were coiled around the pacemaker generator (Figure 3). A new surgical procedure was performed. The pacemaker pocket was opened; leads were uncoiled, repositioned, and connected to the same generator. The leads and pulse generator were fixed tightly to the fascia (Figure 4).

3 Vol 45 No 6 REEL SYNDROME AND PULSATILE LIVER 1039 Figure 1. Pacing spikes at the beginning of the P waves and in the QRS complexes and rare ventricular ectopic beats on electrocardiographic tracing. Pacemaker interrogation shows effective atrial and ineffective ventricular pacing. Atrial pacing beats are conducted to the ventricle. Figure 2. Bedside antero-posterior chest radiography in previous hospital after implantation. a: Atrial lead, b: Ventricular lead, c: Pulse generator.

4 1040 VURAL, ET AL Jpn Heart J November 2004 Figure 3. On postero-anterior chest radiograph, atrial lead is straightened, ventricular lead is in the vena cava superior, and leads are coiled around the pacemaker generator one month after implantation. a. Ventricular lead, b. Atrial lead, c. Pulse generator. Figure 4. On postero-anterior chest radiograph, leads are repositioned in the right atrium and in the right ventricular apex. a: Atrial lead; b: Ventricular lead; c: Pulse generator. DISCUSSION To the best of our knowledge, the present case is the first case of Reel syndrome that developed in a patient with a two-chamber pacemaker. The patient

5 Vol 45 No 6 REEL SYNDROME AND PULSATILE LIVER 1041 had liver pulsations synchronous with heartbeats. The pulsations were thought to be due to stimulation of the right phrenic nerve by the dislocated ventricular lead. We suggest that Reel syndrome, which is another form of Twiddler s syndrome, should be included among the etiologies of liver pulsation in patients with pacemakers. It is easily diagnosed by chest X-rays with the observation of dislocated leads which are coiled around the pulse generator. In our case we were not able to find any etiological factor for the pacemaker rotation. The patient was male, not overweight, had thick subcutaneous tissue, and no history of willing or unwilling pathological manipulation. The size of the generator pocket was not large. We suggest that implantation of the atrial lead loosely during the first procedure could be the reason for the straightened and functional lead. The same leads were replaced and connected to the same pacemaker because their functions were normal. Development of new complications was prevented by creating a new generator pocket instead of reducing the pocket s size and tight fixation of the leads and generator to the fascia. Because of the patient's thick and strong subcutaneous tissue, we thought subpectoral placement was not necessary. During follow-up we have not observed any further complications and pacemaker function was normal after 10 months. Although Twiddler s syndrome and its other forms are rare conditions, they may result in serious clinical complaints and life-threatening events. To prevent Twiddler s syndrome, the pulse generator should be implanted between the fascia of the musculus pectoralis major and minor, and fixation to a loose subcutaneous tissue should be avoided. 9,10) Although fixation of electrodes with a butterfly and unresorbable suture is not a gold standard, fixation of the generator in at least two points increases the safety of the procedure and decreases the risk of rotation. Other preventive measures include the creation of a small surgical pocket, using a compression band around the upper chest and shoulder, and tightening of the patient s arm for at least five to seven days. REFERENCES 1. Bayliss CE, Beanlands DS, Baird RJ. The pacemaker - Twiddler s syndrome: a new complication of implantable transvenous pacemakers. Can Med Assoc J 1968; 99: Graf R. Twiddler's Syndrome Revisited. Tex Heart Inst J 1983; 10: Roberts JS, Wenger NK. Pacemaker twiddler s syndrome. Am J Cardiol 1989; 63: Kouba C, Liu D, Everett G, Kerber RE. Extreme intracardiac coiling of a transvenous pacemaker: a case report. Pacing Clin Electrophysiol 1985; 8: Chernilas JZ, Bilfinger TV, Vlay SC. Pacemaker Twiddler s syndrome. Pacing Clin Electrophysiol 1996; 19: Tegtmeyer CJ, Deignan JM. The cardiac pacemaker: a different twist. Am J Roentgenol 1976; 126:

6 1042 VURAL, ET AL Jpn Heart J November Jaschke W. Pacemaker - Twiddler s syndrome as the cause of the pacing-lead break. Chirurg 1983; 54: (German) 8. Carnero-Varo A, Perez-Paredes M, Ruiz-Ros JA, et al. Reel Syndrome : a new form of Twiddler's syndrome? Circulation 1999; 100: e Klotz S, Wollmann C, Asfour B, Scheld HH. Twiddler s syndrome - not a disease of the old patient. A case report. Herzschr Elektrophys 2003; 14: Khalilullah M, Khanna SK, Gupta U, Padmavati S. Pacemaker Twiddler s syndrome: a note on its mechanism. J Cardiovasc Surg 1979; 20:

The pacemaker twiddler s syndrome: an infrequent cause of pacemaker failure

The pacemaker twiddler s syndrome: an infrequent cause of pacemaker failure DOI 10.1186/s13104-015-1818-0 BMC Research Notes CASE REPORT Open Access The pacemaker twiddler s syndrome: an infrequent cause of pacemaker failure Mohammad Salahuddin, Fathima Aaysha Cader *, Sahela

More information

Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement

Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement 217 Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement Partha Prateem Choudhury, Vivek Chaturvedi, Saibal Mukhopadhyay, Jamal Yusuf Department

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Implantable cardioverter defibrillators for the treatment of arrhythmias and cardiac resynchronisation therapy for the treatment of heart failure (review

More information

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable

More information

Review Article Pacemaker Lead Displacement: Mechanisms And Management

Review Article Pacemaker Lead Displacement: Mechanisms And Management www.ipej.org 231 Review Article Pacemaker Lead Displacement: Mechanisms And Management Beatriz Fuertes, MD, Jorge Toquero, MD, Ramón Arroyo-Espliguero, MD, Ignacio F. Lozano, MD. Department of Cardiology,

More information

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program Electrophysiology Implant Classification Table The table below contains the codes that apply to our UnitedHealthcare Medicare Advantage cardiology prior Description Includes Generator Placement Includes

More information

Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement

Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement www.ipej.org 49 Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement Naga VA Kommuri, MD, MRCPCH, Sri Lakshmi S Kollepara, MD,

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity. Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded

More information

Pacing Codes and Modes Concepts

Pacing Codes and Modes Concepts Pacing Codes and Modes Concepts Pacing codes and modes concepts Objectives Upon completion of this program the participant will be able to: State what the first four positions of the NBG code represent.

More information

Pediatric Pacemaker Implantation Endocardial or Epicardial

Pediatric Pacemaker Implantation Endocardial or Epicardial Pediatric Pacemaker Implantation Endocardial or Epicardial HAITHAM BADRAN, MD, FEHRA CONSULTANT OF INTERVENTIONAL CARDIOLOGY CONSULTANT OF CARDIAC PACING AND ELECTROPHYSIOLOGY LECTURER OF CARDIOLOGY AIN

More information

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution Francesco Zanon, MD, FESC, FHRS Arrhythmia and Electrophysiology Unit, Cardiology Dept. Santa Maria della Misericordia General

More information

Implantation of Transvenous Pacemakers in Infants and Small Children

Implantation of Transvenous Pacemakers in Infants and Small Children Implantation of Transvenous Pacemakers in Infants and Small Children J. Ernesto Molina, MD, PhD, Ann C. Dunnigan, MD, and Jane E. Crosson, MD Divisions of Cardiovascular and Thoracic Surgery and Pediatric

More information

Transvenous Pacemaker Implantation 22 years after the Mustard Procedure

Transvenous Pacemaker Implantation 22 years after the Mustard Procedure Case Report Transvenous Pacemaker Implantation 22 years after the Mustard Procedure Masato Sakamoto MD, Yoshie Ochiai MD, Yutaka Imoto MD, Akira Sese MD, Mamie Watanabe MD, Kunitaka Joo MD Department of

More information

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism?

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? CASE REPORTS Arrhythmia 2015;16(3):173-177 doi: http://dx.doi.org/10.18501/arrhythmia.2015.029 Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? Eun-Sun Jin, MD, PhD Cardiovascular

More information

Pre-participation Screening for the Prevention of sudden Cardiac Death in Young Athletes. Thomas W. Allen, DO, MPH

Pre-participation Screening for the Prevention of sudden Cardiac Death in Young Athletes. Thomas W. Allen, DO, MPH Pre-participation Screening for the Prevention of sudden Cardiac Death in Young Athletes Thomas W. Allen, DO, MPH Cardiovascular disorders are the leading cause of sudden death in young athletes accounting

More information

Model 5392 EPG Temporary Pacer

Model 5392 EPG Temporary Pacer Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short

More information

IB TOPIC 6.2 THE BLOOD SYSTEM

IB TOPIC 6.2 THE BLOOD SYSTEM IB TOPIC 6.2 THE BLOOD SYSTEM TERMS TO KNOW circulation ventricle artery vein THE BLOOD SYSTEM 6.2.U1 - Arteries convey blood at high pressure from the ventricles to the tissues of the body Circulation

More information

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a

More information

THE HEART THE CIRCULATORY SYSTEM

THE HEART THE CIRCULATORY SYSTEM THE HEART THE CIRCULATORY SYSTEM There are three primary closed cycles: 1) Cardiac circulation pathway of blood within the heart 2) Pulmonary circulation blood from the heart to lungs and back 3) Systemic

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

Repetitive narrow QRS tachycardia in a 61-year-old female patient with recent palpitations

Repetitive narrow QRS tachycardia in a 61-year-old female patient with recent palpitations Journal of Geriatric Cardiology (2018) 15: 193 198 2018 JGC All rights reserved; www.jgc301.com Case Report Open Access Repetitive narrow QRS tachycardia in a 61-year-old female patient with recent palpitations

More information

Epicardial Pacemaker Implantation for Complete Heart Block

Epicardial Pacemaker Implantation for Complete Heart Block for Complete Heart Block Donald G. Mulder, M.D., and C. Gordon Frank, M.D. S udden death or progressive cardiac failure threatens life in the patient with complete heart block. The slow heart rate and

More information

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants EKG Pop Quiz Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants 1 Disclosures No financial relationships to disclose. EKG #1 75 y/o woman with a dual-chamber pacemaker

More information

Implantation of a biventricular implantable cardioverter-defibrillator guided by an electroanatomic mapping system

Implantation of a biventricular implantable cardioverter-defibrillator guided by an electroanatomic mapping system Europace (2012) 14, 107 111 doi:10.1093/europace/eur250 CLINICAL RESEARCH Pacing and Resynchronization Therapy Implantation of a biventricular implantable cardioverter-defibrillator guided by an electroanatomic

More information

Project Title Temporary Pacemaker Training Simulator

Project Title Temporary Pacemaker Training Simulator Project Title Temporary Pacemaker Training Simulator Project Description Problem: There is no available training device for temporary pacemakers (pacemakers). A training device will have to essentially

More information

Reduction of Mitral Regurgitation by Endocardial Right Ventricular Bifocal Pacing in Cases of Dilated Cardiomyopathy

Reduction of Mitral Regurgitation by Endocardial Right Ventricular Bifocal Pacing in Cases of Dilated Cardiomyopathy June 2000 233 Reduction of Mitral Regurgitation by Endocardial Right Ventricular Bifocal Pacing in Cases of Dilated Cardiomyopathy J. C. PACHON M., R. N. ALBORNOZ, E. I. PACHON M., V. M. GIMENES, J. PACHON

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

Lab #3: Electrocardiogram (ECG / EKG)

Lab #3: Electrocardiogram (ECG / EKG) Lab #3: Electrocardiogram (ECG / EKG) An introduction to the recording and analysis of cardiac activity Introduction The beating of the heart is triggered by an electrical signal from the pacemaker. The

More information

Atrioventricular (AV) Nodal Reentry Associated with 2:1 Infra-His Conduction Block during Tachycardia in a Patient with AV Nodal Triple Pathways

Atrioventricular (AV) Nodal Reentry Associated with 2:1 Infra-His Conduction Block during Tachycardia in a Patient with AV Nodal Triple Pathways Atrioventricular (AV) Nodal Reentry Associated with 2:1 Infra-His Conduction Block during Tachycardia in a Patient with AV Nodal Triple Pathways Haruhiko ABE, M.D., Takashi OHKITA, M.D., Masasuke FUJITA,

More information

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul,

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul, Mayo Clin Proc, August 2002, Vol 77 Unintentional Deactivation of ICDs 855 Case Report Unintentional Deactivation of Implantable Cardioverter-Defibrillators in Health Care Settings MARY JANE RASMUSSEN,

More information

Chapter 80 Implantable Cardiac Devices

Chapter 80 Implantable Cardiac Devices Chapter 80 Implantable Cardiac Devices Episode Overview: 1. List 5 Indications for permanent pacing 2. Pacemaker nomenclature - what do the 5 letters mean a. What are common Pacemaker types? 3. List the

More information

10. Thick deposits of lipids on the walls of blood vessels, called, can lead to serious circulatory issues. A. aneurysm B. atherosclerosis C.

10. Thick deposits of lipids on the walls of blood vessels, called, can lead to serious circulatory issues. A. aneurysm B. atherosclerosis C. Heart Student: 1. carry blood away from the heart. A. Arteries B. Veins C. Capillaries 2. What is the leading cause of heart attack and stroke in North America? A. alcohol B. smoking C. arteriosclerosis

More information

IB TOPIC 6.2 THE BLOOD SYSTEM

IB TOPIC 6.2 THE BLOOD SYSTEM IB TOPIC 6.2 THE BLOOD SYSTEM THE BLOOD SYSTEM TERMS TO KNOW circulation ventricle artery vein 6.2.U1 - Arteries convey blood at high pressure from the ventricles to the tissues of the body Circulation

More information

Case #4 Diagnosis: Pregnancy at 39 weeks with fetal distress Procedure: This is a 28-year-old patient who had a previous cesarean section for

Case #4 Diagnosis: Pregnancy at 39 weeks with fetal distress Procedure: This is a 28-year-old patient who had a previous cesarean section for Case #4 Diagnosis: Pregnancy at 39 weeks with fetal distress Procedure: This is a 28-year-old patient who had a previous cesarean section for delivery due to fetal distress. During this pregnancy, she

More information

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Case Report Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Taku Nishida MD, Tamio Nakajima MD, PhD, Yutaka Goryo MD, Ken-ichi Ishigami MD, PhD, Hiroyuki Kawata

More information

Arrhythmias. Pulmonary Artery

Arrhythmias. Pulmonary Artery Arrhythmias Introduction Cardiac arrhythmia is an irregularity of the heart beat that causes the heart to beat too slowly, too fast, or irregularly. There are different types of arrhythmias. Most arrhythmias

More information

Different indications for pacemaker implantation are the following:

Different indications for pacemaker implantation are the following: Patient Resources: ICD/Pacemaker Overview ICD/Pacemaker Overview What is a pacemaker? A pacemaker is a device that uses low energy electrical pulses to prompt the heart to beat whenever a pause in the

More information

Lab 16. The Cardiovascular System Heart and Blood Vessels. Laboratory Objectives

Lab 16. The Cardiovascular System Heart and Blood Vessels. Laboratory Objectives Lab 16 The Cardiovascular System Heart and Blood Vessels Laboratory Objectives Describe the anatomical structures of the heart to include the pericardium, chambers, valves, and major vessels. Describe

More information

Cardiovascular System Notes: Heart Disease & Disorders

Cardiovascular System Notes: Heart Disease & Disorders Cardiovascular System Notes: Heart Disease & Disorders Interesting Heart Facts The Electrocardiograph (ECG) was invented in 1902 by Willem Einthoven Dutch Physiologist. This test is still used to evaluate

More information

37 1 The Circulatory System

37 1 The Circulatory System H T H E E A R T 37 1 The Circulatory System The circulatory system and respiratory system work together to supply cells with the nutrients and oxygen they need to stay alive. a) The respiratory system:

More information

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure Inappropriate Discharges of Intravenous Implantable Cardioverter Defibrillators Owing to Lead Failure Takashi WASHIZUKA, 1 MD, Masaomi CHINUSHI, 1 MD, Ryu KAZAMA, 1 MD, Takashi HIRONO, 1 MD, Hiroshi WATANABE,

More information

The Mammalian Circulatory System

The Mammalian Circulatory System The Mammalian Heart The Mammalian Circulatory System Recall: What are the 3 cycles of the mammalian circulatory system? What are their functions? What are the three main vessel types in the mammalian circulatory

More information

Biology Unit 3 The Human Heart P

Biology Unit 3 The Human Heart P Biology 2201 Unit 3 The Human Heart P 314-321 Structure and Function of the Human Heart Structure of the Human Heart Has four Chambers (2 Atria and 2 Ventricles) Made of Cardiac Muscle Found in Chest Cavity

More information

Electrocardiography for Healthcare Professionals

Electrocardiography for Healthcare Professionals Electrocardiography for Healthcare Professionals Kathryn A. Booth Thomas O Brien Chapter 10: Pacemaker Rhythms and Bundle Branch Block Learning Outcomes 10.1 Describe the various pacemaker rhythms. 10.2

More information

4. The two inferior chambers of the heart are known as the atria. the superior and inferior vena cava, which empty into the left atrium.

4. The two inferior chambers of the heart are known as the atria. the superior and inferior vena cava, which empty into the left atrium. Answer each statement true or false. If the statement is false, change the underlined word to make it true. 1. The heart is located approximately between the second and fifth ribs and posterior to the

More information

CASE 10. What would the ST segment of this ECG look like? On which leads would you see this ST segment change? What does the T wave represent?

CASE 10. What would the ST segment of this ECG look like? On which leads would you see this ST segment change? What does the T wave represent? CASE 10 A 57-year-old man presents to the emergency center with complaints of chest pain with radiation to the left arm and jaw. He reports feeling anxious, diaphoretic, and short of breath. His past history

More information

IBHRExam Prep Implanted CRM Device System Radiography

IBHRExam Prep Implanted CRM Device System Radiography Implanted CRM Device System Radiography IBHRExam Prep www.pacericd.com 2 Where do they go? 3 Anatomy diagram 4 Anatomy 5 Pulse generator configurations www.pacericd.com 6 Guidant pacemaker pulse generator

More information

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program:

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: Notification and Prior Authorization Program: Electrophysiology Implant Classification Table The following chart contains the codes that require notification or prior authorization as part of UnitedHealthcare

More information

Management strategies for atrial fibrillation Thursday, 20 October :27

Management strategies for atrial fibrillation Thursday, 20 October :27 ALTHOUGH anyone who has had to run up a flight of steps or has had a frightening experience is quite familiar with a racing heartbeat, for the more than 2 million Americans who suffer from atrial fibrillation

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

How to Approach the Patient with CRT and Recurrent Heart Failure

How to Approach the Patient with CRT and Recurrent Heart Failure How to Approach the Patient with CRT and Recurrent Heart Failure Byron K. Lee MD Associate Professor of Medicine Electrophysiology and Arrhythmia Section UCSF Update in Electrocardiography and Arrhythmias

More information

HISTORY. Question: What category of heart disease is suggested by this history? CHIEF COMPLAINT: Heart murmur present since early infancy.

HISTORY. Question: What category of heart disease is suggested by this history? CHIEF COMPLAINT: Heart murmur present since early infancy. HISTORY 18-year-old man. CHIEF COMPLAINT: Heart murmur present since early infancy. PRESENT ILLNESS: Although normal at birth, a heart murmur was heard at the six week check-up and has persisted since

More information

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II). Cardiac Pacemaker Premature Beats When one of ectopic foci becomes irritable, it may spontaneously fire, leading to one or more premature beats. Atrial and junctional foci may become irritable from excess

More information

Self-Conversion of Paroxysmal Atrial Tachycardia by Radio-frequency-Triggered Pacemaker

Self-Conversion of Paroxysmal Atrial Tachycardia by Radio-frequency-Triggered Pacemaker Self-Conversion of Paroxysmal Atrial Tachycardia by Radio-frequency-Triggered Pacemaker Parashos Geleris, M.D., Harisios Boudoulas, M.D., Stephen F. Schaal, M.D., and Charles Meckstroth, M.D. ABSTRACT

More information

CRITICAL CARE OF THE CARDIAC PATIENT WEBINAR VET 2017

CRITICAL CARE OF THE CARDIAC PATIENT WEBINAR VET 2017 CRITICAL CARE OF THE CARDIAC PATIENT WEBINAR VET 2017 The Heart The heart is undoubtedly the most important organ in the body. Unfortunately, when it is not functioning properly, it can have dire consequences.

More information

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs).

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs). Management of Children with Cardiac Devices Guideline originally developed by Leann Miles, APRN; Lindsey Pumphrey, RN; Srikant Das, MD, and the ANGELS Team. Last reviewed by Lindsey Pumphrey, RN, Srikant

More information

Congestive Heart Failure or Heart Failure

Congestive Heart Failure or Heart Failure Congestive Heart Failure or Heart Failure Dr Hitesh Patel Ascot Cardiology Group Heart Failure Workshop April, 2014 Question One What is the difference between congestive heart failure and heart failure?

More information

COMMONLY BILLED CODES AND ASSOCIATED 2018 MEDICARE RATES

COMMONLY BILLED CODES AND ASSOCIATED 2018 MEDICARE RATES CRHF REIMBURSEMENT & HEALTH POLICY Pacemaker Therapy COMMONLY BILLED CODES AND ASSOCIATED 2018 MEDICARE RATES This document reflects commonly billed codes for Pacemaker Therapy and their associated National

More information

Ch 19: Cardiovascular System - The Heart -

Ch 19: Cardiovascular System - The Heart - Ch 19: Cardiovascular System - The Heart - Give a detailed description of the superficial and internal anatomy of the heart, including the pericardium, the myocardium, and the cardiac muscle. Trace the

More information

Lab 2. The Intrinsic Cardiac Conduction System. 1/23/2016 MDufilho 1

Lab 2. The Intrinsic Cardiac Conduction System. 1/23/2016 MDufilho 1 Lab 2 he Intrinsic Cardiac Conduction System 1/23/2016 MDufilho 1 Figure 18.13 Intrinsic cardiac conduction system and action potential succession during one heartbeat. Superior vena cava ight atrium 1

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. The Miracle of Living February 21, 2018 Matthew Ostrom MD,FACC,FHRS Division of

More information

Circulation. Circulation = is a process used for the transport of oxygen, carbon! dioxide, nutrients and wastes through-out the body

Circulation. Circulation = is a process used for the transport of oxygen, carbon! dioxide, nutrients and wastes through-out the body Circulation Circulation = is a process used for the transport of oxygen, carbon! dioxide, nutrients and wastes through-out the body Heart = muscular organ about the size of your fist which pumps blood.

More information

Heart Pacemakers On Chest X Ray Revisited: What the radiologist has to know

Heart Pacemakers On Chest X Ray Revisited: What the radiologist has to know Heart Pacemakers On Chest X Ray Revisited: What the radiologist has to know Poster No.: C-2314 Congress: ECR 2015 Type: Educational Exhibit Authors: J. Peña Conde, J. De Luis - Yanes, M. Martinez-Perez,

More information

The Cardiovascular System. Chapter 15. Cardiovascular System FYI. Cardiology Closed systemof the heart & blood vessels. Functions

The Cardiovascular System. Chapter 15. Cardiovascular System FYI. Cardiology Closed systemof the heart & blood vessels. Functions Chapter 15 Cardiovascular System FYI The heart pumps 7,000 liters (4000 gallons) of blood through the body each day The heart contracts 2.5 billion times in an avg. lifetime The heart & all blood vessels

More information

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt Leadless Pacing Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt The weakest link in Pacemaker system the lead. The more the leads the more the complications Dislodgement Fracture Insulation

More information

The HEART. What is it???? Pericardium. Heart Facts. This muscle never stops working It works when you are asleep

The HEART. What is it???? Pericardium. Heart Facts. This muscle never stops working It works when you are asleep This muscle never stops working It works when you are asleep The HEART It works when you eat It really works when you exercise. What is it???? Located between the lungs in the mid thoracic region Apex

More information

Diseases of the Conduction System

Diseases of the Conduction System 4 CHAPTER 4 Diseases of the Conduction System Diseases of the conduction system are numerous and varied. The authors have selected a few representative entities for this section: complete heart block as

More information

Interesting ECG Right bundle branch block pattern during right ventricular permanent pacing: Is it safe or not?

Interesting ECG Right bundle branch block pattern during right ventricular permanent pacing: Is it safe or not? www.ipej.org 187 Interesting ECG Right bundle branch block pattern during right ventricular permanent pacing: Is it safe or not? Okan Erdogan, MD, Feyza Aksu, MD Trakya Universitesi Tıp Fakultesi, Kardiyoloji

More information

Differentiation of Ventricular Tachycardia from Junctional Tachycardia with Aberrant Conduction

Differentiation of Ventricular Tachycardia from Junctional Tachycardia with Aberrant Conduction Differentiation of Ventricular Tachycardia from Junctional Tachycardia with Aberrant Conduction The Use of Competitive Atrial Pacing By ROBERT M. EASLEY, JR., M.D., AND SIDNEY GOLDSTEIN, M.D. SUMMARY A

More information

Chapter 20 (2) The Heart

Chapter 20 (2) The Heart Chapter 20 (2) The Heart ----------------------------------------------------------------------------------------------------------------------------------------- Describe the component and function of

More information

BHRS Prep course Pub style Quiz NOT A

BHRS Prep course Pub style Quiz NOT A BHRS Prep course Pub style Quiz NOT A Round 1 Quiz questions Common ECG s in CRM Colin Cunnington Over the past 3 months, a 24-yearold man who works as a truck driver has had frequent episodes of pre-syncope

More information

The Electrocardiogram

The Electrocardiogram The Electrocardiogram Chapters 11 and 13 AUTUMN WEDAN AND NATASHA MCDOUGAL The Normal Electrocardiogram P-wave Generated when the atria depolarizes QRS-Complex Ventricles depolarizing before a contraction

More information

Hatim Al Lawati. MD, FRCPC, DABIM(CV), FACC

Hatim Al Lawati. MD, FRCPC, DABIM(CV), FACC Hatim Al Lawati. MD, FRCPC, DABIM(CV), FACC Consultant Interventional Cardiology & Structural Heart Disease Department of Medicine Sultan Qaboos University Hospital hatim.al.lawati@gmail.com April 2017

More information

Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children

Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children ERALD A. SERWER, M.D., MACDONALD DICK 11, M.D., KAREN UZARK, R.N., Ph.D., WILLIAM A. SCOTT, M.D., and EDWARD L.

More information

Matters of the Heart: Comprehensive Cardiology SARAH BEANLANDS RN BSCN MSC

Matters of the Heart: Comprehensive Cardiology SARAH BEANLANDS RN BSCN MSC Matters of the Heart: Comprehensive Cardiology SARAH BEANLANDS RN BSCN MSC Who am I? Class Outline Gross anatomy of the heart Trip around the heart Micro anatomy: cellular and tissue level Introduction

More information

Large Arteries of Heart

Large Arteries of Heart Cardiovascular System (Part A-2) Module 5 -Chapter 8 Overview Arteries Capillaries Veins Heart Anatomy Conduction System Blood pressure Fetal circulation Susie Turner, M.D. 1/5/13 Large Arteries of Heart

More information

Cardiac Cycle. Each heartbeat is called a cardiac cycle. First the two atria contract at the same time.

Cardiac Cycle. Each heartbeat is called a cardiac cycle. First the two atria contract at the same time. The Heartbeat Cardiac Cycle Each heartbeat is called a cardiac cycle. First the two atria contract at the same time. Next the two ventricles contract at the same time. Then all the chambers relax. http://www.youtube.com/watch?v=frd3k6lkhws

More information

Permanent Pacemaker Implantation via the Iliac Vein: An Alternative in 4 Cases with Contraindications to the Pectoral Approach

Permanent Pacemaker Implantation via the Iliac Vein: An Alternative in 4 Cases with Contraindications to the Pectoral Approach Case Report Permanent Pacemaker Implantation via the Iliac Vein: An Alternative in 4 Cases with Contraindications to the Pectoral Approach Koji Tsutsumi MD, Kenichi Hashizume MD, Naritaka Kimura MD, Shinichi

More information

Implantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect

Implantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect The Ochsner Journal 10:27 31, 2010 f Academic Division of Ochsner Clinic Foundation Implantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect Anas Bitar, MD, Maria Malaya

More information

Patient Resources: Arrhythmias and Congenital Heart Disease

Patient Resources: Arrhythmias and Congenital Heart Disease Patient Resources: Arrhythmias and Congenital Heart Disease Overview Arrhythmias (abnormal heart rhythms) can develop in patients with congenital heart disease (CHD) due to thickening/weakening of their

More information

MR Advance Techniques. Cardiac Imaging. Class IV

MR Advance Techniques. Cardiac Imaging. Class IV MR Advance Techniques Cardiac Imaging Class IV Heart The heart is a muscular organ responsible for pumping blood through the blood vessels by repeated, rhythmic contractions. Layers of the heart Endocardium

More information

Cigna - Prior Authorization Procedure List Cardiology

Cigna - Prior Authorization Procedure List Cardiology Cigna - Prior Authorization Procedure List Cardiology Category CPT Code CPT Code Description 33206 Insertion of new or replacement of permanent pacemaker with transvenous electrode(s); atrial 33207 Insertion

More information

Lab Activity 23. Cardiac Anatomy. Portland Community College BI 232

Lab Activity 23. Cardiac Anatomy. Portland Community College BI 232 Lab Activity 23 Cardiac Anatomy Portland Community College BI 232 Cardiac Muscle Histology Branching cells Intercalated disc: contains many gap junctions connecting the adjacent cell cytoplasm, creates

More information

The Heart and Heart Disease

The Heart and Heart Disease The Heart and Heart Disease Illustration of the heart by Leonardo DaVinci heart-surgeon.com/ history.html 2/14/2010 1 I. Location, Size and Position of the Heart A. Triangular organ located 1. of mass

More information

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program:

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: Electrophysiology Implant Code Classification Table The

More information

Pacing in patients with congenital heart disease: part 1

Pacing in patients with congenital heart disease: part 1 Pacing in patients with congenital heart disease: part 1 September 2013 Br J Cardiol 2013;20:117 20 doi: 10.5837/bjc/2013.028 Authors: Khaled Albouaini, Archana Rao, David Ramsdale View details Only a

More information

PRACTICAL GUIDE TO FETAL ECHOCARDIOGRAPHY IC Huggon and LD Allan

PRACTICAL GUIDE TO FETAL ECHOCARDIOGRAPHY IC Huggon and LD Allan PRACTICAL GUIDE TO FETAL ECHOCARDIOGRAPHY IC Huggon and LD Allan Fetal Cardiology Unit, Harris Birthright Research Centre for Fetal Medicine, King's College Hospital, London, UK IMPORTANCE OF PRENATAL

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

ECG Tutorial Series. Case 1. Case 2. Case 3. Prof. Wang, Tzong-Luen

ECG Tutorial Series. Case 1. Case 2. Case 3. Prof. Wang, Tzong-Luen ECG Tutorial Series Prof. Wang, Tzong-Luen MD, PhD, JM, FESC, FACC, FCAPSC Chief, ED, Shin-Kong Wu Ho-Su Memorial Hospital Case 1 A 38-year-old male is visiting a family member in the hospital, but collapes

More information

Interpreting Electrocardiograms (ECG) Physiology Name: Per:

Interpreting Electrocardiograms (ECG) Physiology Name: Per: Interpreting Electrocardiograms (ECG) Physiology Name: Per: Introduction The heart has its own system in place to create nerve impulses and does not actually require the brain to make it beat. This electrical

More information

as the cause of recurrent syncope 3 allows appropriate management aimed

as the cause of recurrent syncope 3 allows appropriate management aimed Case Report Hellenic J Cardiol 2009; 50: 155-159 The Role of the Implantable Loop Recorder in the Investigation of Recurrent Syncope SKEVOS K. SIDERIS 1, TERESA A. MOUSIAMA 1, PAVLOS N. STOUGIANNOS 1,

More information

X-Plain Atrial Fibrillation Reference Summary

X-Plain Atrial Fibrillation Reference Summary X-Plain Atrial Fibrillation Reference Summary Introduction Atrial fibrillation is a common heart condition that affects approximately 2.5 million Americans every year. Atrial fibrillation requires immediate

More information

EKG Abnormalities. Adapted from:

EKG Abnormalities. Adapted from: EKG Abnormalities Adapted from: http://www.bem.fi/book/19/19.htm Some key terms: Arrhythmia-an abnormal rhythm or sequence of events in the EKG Flutter-rapid depolarizations (and therefore contractions)

More information

EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology

EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important

More information

The production of murmurs is due to 3 main factors:

The production of murmurs is due to 3 main factors: Heart murmurs The production of murmurs is due to 3 main factors: high blood flow rate through normal or abnormal orifices forward flow through a narrowed or irregular orifice into a dilated vessel or

More information