Update on Device Innovation (S-ICD, Wearable, Leadless)
|
|
- Lorin Wilkerson
- 6 years ago
- Views:
Transcription
1 Update on Device Innovation (S-ICD, Wearable, Leadless) C. W. Israel Dept. of Medicine - Cardiology Evangelical Hospital Bielefeld J. W. Goethe University Frankfurt Carsten.Israel@em.uni-frankfurt.de
2 Conflicts of Interest Honoraria (speakers bureau, etc.): Biotronik, Boston-Scientific, Medtronic, St. Jude Medical, Sorin, Zoll Advisory Board: Medtronic, St. Jude Medical, Zoll Participation in sponsored clinical trials: Biotronik, Boston-Scientific, Medtronic, Sorin, St. Jude Medical
3 Subcutaneous ICD (S-ICD)
4 Subcutaneous ICD: Size, Longevity new device thinner
5 S-ICD: EFFORTLESS Median FU 498 days Czech Republic, Denmark, Germany, Italy, Netherlands, New Zealand, UK Lambiase, Eur Heart J 2014;35:
6 S-ICD: EFFORTLESS 1 pt died: asystole followed by VF, 11 ineffective shocks (successful DFT at implant) Explants in 17 pts (3.7%): Infection (n=8), recurrent VT (n=2), inappropriate shocks (n=2), erosion, DFT failure at implant, failure to convert spontaneous VT/VF, inappropriate sensing, inappropriate shocks, pain (all n=1) 9 deaths Lambiase, Eur Heart J 2014;35:
7 S-ICD: EFFORTLESS Lambiase, Eur Heart J 2014;35:
8 S-ICD: Study/Registry Overiew McLeod, Eur Heart J 2015 (advance access, Oct 29 th, 2015)
9 Lead Selection to Prevent T Wave Oversensing McLeod, Eur Heart J 2015 (advance access, Oct 29 th, 2015)
10 S-ICD: Pooled Data Burke, J Am Coll Cardiol 2015;65:
11 Achilles Heel: Implanter, not Lead! 1. Ligaments compress the lead subclavian crush 2. Calcified ligaments crush the lead when sheath is retracted 3. Leads trapped in muscle are rubbed off by bone movements
12 ICD Leads: No Subclavian Puncture! 1. Ligaments compress the lead subclavian crush 2. Calcified ligaments crush the lead when sheath is retracted 3. Leads trapped in muscle are rubbed off by bone movements
13 Transvene : Subclavia vs. Cephalica 767 Medtronic Transvene ICD leads Lead fracture: 7/523 leads via subclavian vein 0/221 leads via cephalic vein All lead failures occurred in patients implanted via subclavian vein, no lead failure occurred in patients implanted via cephalic vein! Gallik, PACE 1996;19:
14 Sprint Fidelis: Subclavia vs. Cephalica Canadian Heart Rhythm Society Device Committee Birnie, Circulation 2012;125:
15 Riata : Subclavia vs. Cephalica Erkapic, J Carciovasc Electrophysiol 2011;22:
16 ESC Guidelines on SCD 2015 Priori, Eur Heart J 2015 (epub)
17 Subcutaneous ICD: Candidates patients with (recurrent) device infection, septicemia and TV ICD explantation patients with ICD lead fracture due to anatomic reasons (not only bad implanter ) patients with lack/obstruction of SVC patients on dialysis problem: risk of bradycardia patients with TV replacement
18 Subcutaneous ICD: No Candidates patients who require antibradycardia pacing patients who require CRT patients with VT (require antitachycardia pacing) patients who require AADs for VT or SVT ( patients with β blockers for CAD or HF) patients with HCM if T wave oversensing cannot be resolved thin patients young, sportive patients (device lifetime 5-6 years)
19 Wearable ICD
20 Wearable ICD: WEARIT II Registry Kutyifa, Circulation 2015;132:1613-9
21 WCD: Use in Hours per Day Kutyifa, Circulation 2015;132:1613-9
22 Wearable ICD: WEARIT II Registry Kutyifa, Circulation 2015;132:1613-9
23 Wearable ICD: WEARIT II Registry Kutyifa, Circulation 2015;132:1613-9
24 Wearable ICD: WEARIT II Registry Kutyifa, Circulation 2015;132:1613-9
25 Wearable ICD Post MI 1.4% of pts successfully treated for VT/VF in the first 3 months post MI Epstein, J Am Coll Cardiol 2013;62:2000-7
26 Wearable ICD Post MI Epstein, J Am Coll Cardiol 2013;62: % resuscitation survival rate
27 Wearable ICD: SCD Guidelines Prochnau, Successful use of a wearable cardioverter-defibrillator in myocarditis with normal ejection fraction. Clin Res Cardiol 2010;99: Chung MK. The role of the wearable cardioverter defibrillator in clinical practice. Cardiol Clin 2014;32: Priori et al., Eur Heart J 2015
28 Wearable ICD: SCD Guidelines 2015 Priori et al., Eur Heart J 2015
29 Wearable ICD: SCD Guidelines 2015 Priori et al., Eur Heart J 2015
30 Leadless Pacing
31 Nanostim (St. Jude Medical) Length 42 mm, maximum diameter 5.99 mm, 18F sheath nonretractable screw-in steroid-eluting helix (max. penetration 1.3 mm) Bipolar (tip: centre of the fixation helix, ring: titanium pacemaker case), electrode spacing >10 mm feature for docking delivery and retrieval catheters estimated battery life 8.4 years with 100% pacing (12.4 yrs with 50%) RV blood temperature-responsive VVIR Reddy, Circulation 2014;129:
32 Nanostim: Implantation Reddy, New Engl J Med 2015;129:
33 Innovationen Schrittmacher/ICD LEADLESS Trial Reddy, Circulation 2014;129:
34 Innovationen Schrittmacher/ICD Nanostim (St. Jude Medical) Reddy, Circulation 2014;129:
35 Nanostim: LEADLESS II successfully implantation in 504 of 526 patients (95.8%) Reddy, New Engl J Med 2015;129:
36 Nanostim: LEADLESS II Reddy, New Engl J Med 2015;129:
37 LEADLESS II: Complications Reddy, New Engl J Med 2015; 129:
38 Micra (Medtronic) Size 0.8 cm³, 2.0 g Length: 25.9 mm Outer diameter: 6.7 mm (27F sheath) Bipolar sensing/pacing Fixation mechanism: Nitinol tines Battery Lithium silver vanadium oxide/ Carbon monoflouride Longevity: 7.4/10.1 year (2.0/1.5V output, 60 bpm, 100% paced) Capabilities VVIR RV capture management Sensing assurance Diagnostics Deactivated to OOO at EoL
39 Micra: Implantation Reynolds, New Engl J Med 2015 (advance access)
40 MICRA Study successfully implantation in 719 of 725 patients (99.2%) Reynolds, New Engl J Med 2015 (advance access)
41 MICRA Study Reynolds, New Engl J Med 2015
42 What should be remembered? S-ICD: Alternative to TV-ICD for patients with (recurrent) lead problems (infection, fracture) or lack of venous access; performance compared to TV-ICD unclear (shock efficacy, VF undersensing, T wave oversensing, infection) WCD: Interesting for selected patients during waiting periods (compliance tool!), trials pending Leadless/transcatheter VVIR pacemakers: Promising performance, 4-6% complications (tamponade, vascular access, device dislodgement)
43 Thank you very much for your attention!
New generations pacemakers and ICDs: an update
Advances in Cardiac Arrhythmias and Great Innovations in Cardiology XXVII Giornate Cardiologiche Torinesi New generations pacemakers and ICDs: an update Prof. Fiorenzo Gaita, MD Division of Cardiology
More informationINNOVATIONS IN DEVICE THERAPY:
INNOVATIONS IN DEVICE THERAPY: Subcutaneous ICDs, Leadless Pacemakers, CRT Indications David J Wilber MD Loyola University Medical Center Disclosures: ACC Foundation: Consultant; Biosense / Webster: Consultant,
More informationContinuous ECG telemonitoring with implantable devices: the expected clinical benefits
Continuous ECG telemonitoring with implantable devices: the expected clinical benefits C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Declaration of
More informationSingle- versus Dual-Coil ICD Leads: Does it Matter?
Single- versus Dual-Coil ICD Leads: Does it Matter? C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Conflict of Interest Biotronik Boston-Scientific
More informationNOVEL DEVICE TECHNOLOGIES
NOVEL DEVICE TECHNOLOGIES Leadless Pacemakers and Subcutaneous ICDs Do the Benefits Outweigh MRI Incompatibility? Disclosures None Background PPMs and ICDs are very effective therapy for treating bradyarrhythmias
More informationLeadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt
Leadless Pacing Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt The weakest link in Pacemaker system the lead. The more the leads the more the complications Dislodgement Fracture Insulation
More informationNews der kardialen Device - Therapie! Christian Binggeli! Rhythmologie Klinik St. Anna, Luzern! HerzGefässZentrum Zürich, KlinikImPark, Zürich!
News der kardialen Device - Therapie! Christian Binggeli! Rhythmologie Klinik St. Anna, Luzern! HerzGefässZentrum Zürich, KlinikImPark, Zürich! Complications after Implantation / Replacement / Upgrade
More informationLeadless pacemakers and subcutaneous ICD s: will we use them for most of our patients? K.-H. Kuck Asklepios Klinik St. Georg Hamburg, Germany
Leadless pacemakers and subcutaneous ICD s: will we use them for most of our patients? K.-H. Kuck Asklepios Klinik St. Georg Hamburg, Germany Disclosure Statement Research Grants Consultant / Advisory
More informationProgramming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany
Programming of Bradycardic Parameters C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Programming of ICD Brady Parameters Conflict of Interest Biotronik
More informationThe EFFORTLESS SICD Registry. The EFFORTLESS SICD Registry. SICD Implant 11/14/2016
11/14/2016 VuMedi Webinar: Safety and Efficacy of the Subcutaneous ICD Primary Results of The EFFORTLESS SICD Registry November 21, 2016 Philip Chang, MD, FACC, FHRS Assistant Professor of Medicine and
More informationEssentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH
Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing
More informationTachycardia Devices Indications and Basic Trouble Shooting
Tachycardia Devices Indications and Basic Trouble Shooting Peter A. Brady, MD., FRCP Cardiology Review Course London, March 6 th, 2014 2011 MFMER 3134946-1 Tachycardia Devices ICD Indications Primary and
More informationSubcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention
Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention Dr Ngai-Yin Chan, MBBS, FRCP(Lond), FRCP(Edin), FRCP(Glasg), FACC, FHRS, Consultant Physician, Department of Medicine & Geriatrics, Princess
More informationMEET MICRA. Micra TM ACTUAL SIZE. Transcatheter Pacing System
MEET MICRA ACTUAL SIZE Micra TM Transcatheter Pacing System MEET MICRA The world s smallest pacemaker 1 MINIATURIZED. 93% smaller than modern-day pacemakers 7 Completely self contained within the heart,
More informationDevice Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI
Device Interrogation- Pacemakers, ICD and Loop Recorders Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Disclosures Consultant: Medtronic Speaker s Bureau: St. Jude Medical
More informationHow to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France
How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France Presenter Disclosure Information Christophe Leclercq, MD,
More informationICD: Basics, Programming and Trouble-shooting
ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications
More informationMEET MICRA. Micra TM ACTUAL SIZE. Transcatheter Pacing System
MEET MICRA ACTUAL SIZE Micra TM Transcatheter Pacing System MEET MICRA The world s smallest pacemaker 1 MINIATURIZED. 93% smaller than modern-day pacemakers 6 Completely self contained within the heart,
More informationInappropriate electrical shocks: Tackling the beast
ESC Paris 2011 Inappropriate electrical shocks: Tackling the beast Gerhard Hindricks University of Leipzig Heart Center Dept. of Electrophysiology ESC Paris 2011 Inappropriate electrical shocks: Tackling
More informationSubcutaneous Implantable Cardioverter Defibrillator (S-ICD)
Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) D. D. MANOLATOS, MD, PhD, FESC Electrophysiology and Device Lab General Hospital Evangelismos, Athens The Problem: 300,000 people die each year
More informationLifeVest. Sven Reek, Aarau. Conflict of interest: none
LifeVest Sven Reek, Aarau Conflict of interest: none WCD (LifeVest, ZOLL, Pittsburg,PA) Technical aspects Wearable garment with 4 dry ECG electrodes (detection) 3 defibrillation pads External defibrillator
More informationDefibrillation threshold testing should no longer be performed: contra
Defibrillation threshold testing should no longer be performed: contra Andreas Goette St. Vincenz-Hospital Paderborn Dept. of Cardiology and Intensive Care Medicine Germany No conflict of interest to disclose
More informationDevice Diagnostics and Stroke Prevention: State of the Art
Device Diagnostics and Stroke Prevention: State of the Art Ass. Prof. Dr. Carsten W. Israel Dept. of Cardiology Evangelical Hospital Bielefeld Germany C.W.Israel@em.uni-frankfurt.de Disclosure of Relationships
More informationWhat You Should Know About Subcutaneous and Transvenous ICD
What You Should Know About Subcutaneous and Transvenous ICD N. A. Mark Estes III MD Professor of Medicine Tufts University School of Medicine Director, New England Cardiac Arrhythmia Center Tufts Medical
More informationREVIEW ARTICLE. Leadless Cardiac Pacemaker Therapy. An Overview for the Hospitalist Richard Weachter 1
Leadless Cardiac Pacemaker Therapy. An Overview for the Hospitalist Richard Weachter 1 1 Division of Cardiovascular Medicine, Department of Medicine, University of Missouri, Columbia, MO Correspondence:
More informationWhat s New in CIEDs. Keimyung University Dongsan Medical Center Hyoung-Seob Park
What s New in CIEDs Keimyung University Dongsan Medical Center Hyoung-Seob Park Transvenous CIEDs Technology: - Highly mature & reliable - Still includes generator, connectors and leads Procedure : -
More informationRiata lead extraction- a single centre experience
Riata lead extraction- a single centre experience Rebecca L Noad, Keith W Morrice, Vivek N Kodoth, Carol M Wilson, Michael JD Roberts. Royal Victoria Hospital, Belfast, United Kingdom Background of previous
More informationLeadless pacemakers a panacea for bradyarrhythmias?
Leadless pacemakers a panacea for bradyarrhythmias? Peysh A Patel Take Home Messages Why may leadless systems be required? Where the cessation of vital action is very complete, and continues long, we ought
More informationNATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE
NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE INTERVENTIONAL PROCEDURES PROGRAMME Interventional procedure overview of subcutaneous implantable cardioverter defibrillator insertion for preventing sudden
More informationWe are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors
We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 4,000 116,000 120M Open access books available International authors and editors Downloads Our
More informationWhere Does the Wearable Cardioverter Defibrillator (WCD) Fit In?
Where Does the Wearable Cardioverter Defibrillator (WCD) Fit In? 24 th Annual San Diego Heart Failure Symposium June 1-2, 2018 La Jolla, CA Barry Greenberg, MD Distinguished Professor of Medicine Director,
More informationRecurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm
Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm Guy Amit, MD, MPH Soroka University Medical Center Ben-Gurion University of the Negev Beer-Sheva, Israel Disclosures Consultant:
More informationICD leads survival and troubles in the last 10 years
XVI Symposium on Progress in Clinical Pacing, Rome, Italy - Dec 2 to 7, 2014 ICD leads survival and troubles in the last 10 years Antonio Raviele, MD, FESC, FHRS President ALFA Alliance to Fight Atrial
More informationPrimary Therapy for High Risk LQT Patients Should Be an ICD
Primary Therapy for High Risk LQT Patients Should Be an ICD Raul Weiss MD, FAHA, FACC, FHRS, CCDS Director, Electrophysiology Fellowship Program Associate Professor of Medicine The Ohio State University
More informationSelecting the Best ICD for your Patient-SICD v. TV. Martin C. Burke DO, FACOI CorVita Science Foundation & Academic Medical Center, Amsterdam
Selecting the Best ICD for your Patient-SICD v. TV Martin C. Burke DO, FACOI CorVita Science Foundation & Academic Medical Center, Amsterdam COI DISCLOSURES I have received lecture and proctoring honoraria
More informationMultisite Left Ventricular Pacing for Cardiac Resynchronization
Multisite Left Ventricular Pacing for Cardiac Resynchronization C. W. Israel, M.D. Ass. Prof. Dept. of Medicine Cardiology Evangelical Hospital Bielefeld Germany J. W. Goethe University Frankfurt C.W.Israel@em.uni-frankfurt.de
More informationTroubleshooting ICD. NASPE Training Lancashire & South Cumbria Cardiac Network
Troubleshooting ICD NASPE Training Lancashire & South Cumbria Cardiac Network Bradycardia Pacing by ICD Isolated ICD discharge Multiple discharges Appropriate Inappropriate No ICD Therapy or delay in ICD
More informationLeadless pacing: Technical challenges and concerns. Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital
Leadless pacing: Technical challenges and concerns Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital Modest consultancy fees/honoraria from Bayer Boehringer Ingelheim
More informationLead Selection and Subcutaneous ICD Considerations. The Future of CIEDs
Lead Selection and Subcutaneous ICD Considerations The Future of CIEDs February, 2013 Steven P. Kutalek, MD Director, Cardiac Electrophysiology Drexel University College of Medicine New Technologies 2012
More informationESC Stockholm Arrhythmias & pacing
ESC Stockholm 2010 Take Home Messages for Practitioners Arrhythmias & pacing Prof. Panos E. Vardas Professor of Cardiology Heraklion University Hospital Crete, Greece Disclosures Small teaching fees from
More informationDanish Pacemaker and ICD Register Annual report 2015
Danish Pacemaker and ICD Register Annual report 2015 Preface The Danish Pacemaker Register was founded in 1982 by physicians from all Danish hospitals where pacemakers were implanted. When the first implantable
More information2017 AHA/ACC/HRS Ventricular Arrhythmias and Sudden Cardiac Death Guideline. Top Ten Messages. Eleftherios M Kallergis, MD, PhD, FESC
2017 AHA/ACC/HRS Ventricular Arrhythmias and Sudden Cardiac Death Guideline Top Ten Messages Eleftherios M Kallergis, MD, PhD, FESC Cadiology Department - Heraklion University Hospital No actual or potential
More informationDevice Implantation: Long-Term Commitments. Fayez Bokhari, MD, FRCPC Director of Cardiology Military Hospital, Jeddah
Device Implantation: Long-Term Commitments Fayez Bokhari, MD, FRCPC Director of Cardiology Military Hospital, Jeddah Who should Implant ICD? 1. Electrophysiologist 2. Surgeons 3. General Cardiologist with
More informationEBR Systems, Inc. 686 W. Maude Ave., Suite 102 Sunnyvale, CA USA
Over 200,000 patients worldwide are estimated to receive a CRT device each year. However, limitations prevent some patients from benefiting. CHALLENGING PROCEDURE 5% implanted patients fail to have coronary
More informationTreatment of VT of Purkinje fiber origin: ablation targets and outcome
Treatment of VT of Purkinje fiber origin: ablation targets and outcome Ch. Piorkowski University Leipzig - Heart Center - Dept. of Electrophysiology Leipzig, Germany Presenter Disclosure Information Gerhard
More informationImplantable cardioverter defibrillator, Inappropriate shock, Lead failure
Inappropriate Discharges of Intravenous Implantable Cardioverter Defibrillators Owing to Lead Failure Takashi WASHIZUKA, 1 MD, Masaomi CHINUSHI, 1 MD, Ryu KAZAMA, 1 MD, Takashi HIRONO, 1 MD, Hiroshi WATANABE,
More informationEHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs
EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for
More informationAssessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis
Assessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis Investigator: Keiko Saito, MD Mentor: Yuji Saito, MD, PhD, FACP, FACC Department
More informationClinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm
April 2000 107 Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm B. MERKELY Semmelweis University, Dept. of Cardiovascular Surgery,
More informationThe subcutaneous implantable cardioverter defibrillator review of the recent data
Journal of Geriatric Cardiology (2018) 15: 222 228 2018 JGC All rights reserved; www.jgc301.com Review Open Access The subcutaneous implantable cardioverter defibrillator review of the recent data Stacy
More informationShock Reduction Strategies Michael Geist E. Wolfson MC
Shock Reduction Strategies Michael Geist E. Wolfson MC Shock Therapy Thanks, I needed that! Why Do We Need To Reduce Shocks Long-term outcome after ICD and CRT implantation and influence of remote device
More informationNATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE
NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE INTERVENTIONAL PROCEDURES PROGRAMME Interventional procedure overview of a leadless cardiac pacemaker implantation for bradyarrhythmias Bradyarrhythmias
More informationΕκθορηίζεις απινιδωηή και θνηηόηηηα: μέθοδοι μείωζης ηων θεραπειών απινίδωζης
Εκθορηίζεις απινιδωηή και θνηηόηηηα: μέθοδοι μείωζης ηων θεραπειών απινίδωζης Εμμ. Μ. Κανοσπάκης Καρδιολογική Κλινική Πανεπιζηημίοσ Κρήηης Lessons from large trials Conditioning Rhythm and Electrical Therapy
More informationDanish Pacemaker and ICD Register Annual Report 2016
Danish Pacemaker and ICD Register Annual Report 2016 Preface The Danish Pacemaker Register was founded in 1982 by physicians from all Danish hospitals where pacemakers were implanted. When the first implantable
More informationPerformance of a Miniaturized Transcatheter Pacing System: First-in-human experience
Performance of a Miniaturized Transcatheter Pacing System: First-in-human experience Philippe Ritter, MD Gabor Z Duray, MD, PhD, FESC; Clemens Steinwender, MD, FESC; Kyoko Soejima, MD; Razali Omar, MD;
More informationNATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE
NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Implantable cardioverter defibrillators for the treatment of arrhythmias and cardiac resynchronisation therapy for the treatment of heart failure (review
More informationCRF procedure PROCEDURE FLOW. PATIENT ASSESSMENT Symptoms and indication. Pacemaker & ICD registration. Procedures. Procedure ICD.
Pacemaker & ICD registration Procedures CRF procedure PROCEDURE FLOW Procedure ICD ICD PM ICM lead only PATIENT ASSESSMENT Symptoms and indication Symptoms and events (multiple possibilities) Asymptomatic
More informationNATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE
NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE INTERVENTIONAL PROCEDURES PROGRAMME Interventional procedure overview of leadless cardiac pacemaker implantation for bradyarrhythmias Bradyarrhythmias
More informationSUBCUTANEOUS ICD ITS JOURNEY FROM PROOF OF CONCEPT TO CURRENT EVOLUTION
SUBCUTANEOUS ICD ITS JOURNEY FROM PROOF OF CONCEPT TO CURRENT EVOLUTION Giovanni Luca Botto MD, FEHRA, FESC Electrophysiology Unit ASST Lariana, Como, Italy Presenter Disclosure Information Research support:
More informationThe Nuts and Bolts of ICD Therapy
Electrical Management of Cardiac Rhythm Disorders For Cardiology Fellows December 5-8 Austin, Texas The Nuts and Bolts of ICD Therapy 1 2 Action Potential Localized Differences in Conduction Conduction
More informationLeft Atrial Appendage Closure: The Good, The Bad and The Ugly
Left Atrial Appendage Closure: The Good, The Bad and The Ugly John D. Hummel, MD Director of Electrophysiology Research Ross Heart Hospital, The Ohio State University Columbus, Ohio USA Disclosures Modest
More informationSUBCUTANEOUS IMPLANTABLE CARDIOVERTER DEFIBRILLATORS
SUBCUTANEOUS IMPLANTABLE CARDIOVERTER DEFIBRILLATORS Protocol: CAR044 Effective Date: December 1, 2017 Table of Contents Page COMMERCIAL & MEDICAID COVERAGE RATIONALE... 1 MEDICARE COVERAGE RATIONALE...
More informationAdvanced ICD Concepts
1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation
More informationAutomatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms
Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable
More informationDevice Update Implantable Cardioverter Defibrillator (ICD) 박상원
2012 년춘계학술대회 Device Update Implantable Cardioverter Defibrillator (ICD) 박상원 Arrhythmia Center, KUMC www.korea-heartrhythm.com Korea University Medical Center Seoul, Korea The Development of ICD by a team
More informationIntegration of diagnostic device information to improve patient management
Integration of diagnostic device information to improve patient management Haran Burri Associate Professor Cardiology Service University Hospital of Geneva Disclosures Biotronik: fellowship support, speaker
More informationCardiovascular Implantable Electronic Device Leads & Arteriovenous Hemodialysis Access
Cardiovascular Implantable Electronic Device Leads & Arteriovenous Hemodialysis Access Theodore F. Saad, MD Nephrology Associates, PA Christiana Care Health System Newark, Delaware Cardiovascular Implantable
More informationFollow-up of CRT patients ESC Munich Clinical and biological follow-up of CRT patients
Follow-up of CRT patients ESC Munich 12-08-26 Clinical and biological follow-up of CRT patients Frieder Braunschweig MD PhD FESC Associate Professor of Cardiology Karolinska University Hospital Stockholm,
More informationCIEDs Infection: Lead Extraction, First or Last option?
CIEDs Infection: Lead Extraction, First or Last option? More CRM Systems & Leads 600,000 new devices annually 1.2 million new leads annually Sources: 1) Medtronic CRDM Product Performance Report, Jan 2009.
More informationPediatric Pacemaker Implantation Endocardial or Epicardial
Pediatric Pacemaker Implantation Endocardial or Epicardial HAITHAM BADRAN, MD, FEHRA CONSULTANT OF INTERVENTIONAL CARDIOLOGY CONSULTANT OF CARDIAC PACING AND ELECTROPHYSIOLOGY LECTURER OF CARDIOLOGY AIN
More informationWearable Cardioverter-Defibrillators
Medical Policy Manual Durable Medical Equipment, Policy No. 61 Wearable Cardioverter-Defibrillators Next Review: June 2019 Last Review: June 2018 Effective: August 1, 2018 IMPORTANT REMINDER Medical Policies
More informationPolicy #: 168 Latest Review Date: October 2014
Name of Policy: Cardioverter Defibrillator: Implantable Policy #: 168 Latest Review Date: October 2014 Category: Surgery Policy Grade: A Background/Definitions: As a general rule, benefits are payable
More informationLeadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece
Leadless pacing P.E. Vardas Professor of Cardiology University of Crete, Greece DISCLOSURES Small teaching fees from: Bayer/J&J, Boeringer, Menarini Small consulting fees from: Medtronic 1957: First External
More informationRecent Advances in Pacing and Defibrillation Harish Doppalapudi, MD
Recent Advances in Pacing and Defibrillation Harish Doppalapudi, MD Harish Doppalapudi, MD Assistant Professor of Medicine Director, Clinical Cardiac Electrophysiology Training Program University of Alabama
More informationPacing Without Wires: Leadless Cardiac Pacing
REVIEWS AND CONTEMPORARY UPDATES Ochsner Journal 16:238 242, 2016 Ó Academic Division of Ochsner Clinic Foundation Pacing Without Wires: Leadless Cardiac Pacing Michael L. Bernard, MD, PhD Department of
More informationCase presentation: A 14-yearold
CLINICIAN UPDATE Prevention of Sudden Death in Hypertrophic Cardiomyopathy But Which Defibrillator for Which Patient? Giuseppe Boriani, MD, PhD; Barry J. Maron, MD; Win-Kuang Shen, MD; Paolo Spirito, MD
More informationWearable Cardioverter-Defibrillators
Wearable Cardioverter-Defibrillators Policy Number: 2.02.15 Last Review: 12/2018 Origination: 10/1988 Next Review: 12/2019 Policy Blue Cross and Blue Shield of Kansas City (Blue KC) will provide coverage
More informationCARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART
CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART December 2015 This chart encompasses all Medtronic cardiac devices FDA-Approved as MR Conditional and included in the MRI SureScan portfolio. If a device
More informationPreventing Sudden Death Current & Future Role of ICD Therapy
Preventing Sudden Death Current & Future Role of ICD Therapy Derek V Exner, MD, MPH, FRCPC, FACC, FAHA, FHRS Professor, Libin Cardiovascular Institute of Alberta Canada Research Chair, Cardiovascular Clinical
More informationInitial Clinical Experience With A Novel Left Ventricular Quadripolar Lead. Johannes Sperzel Kerckhoff Heart Center, Bad Nauheim, Germany
Initial Clinical Experience With A Novel Left Ventricular Quadripolar Lead Johannes Sperzel Kerckhoff Heart Center, Bad Nauheim, Germany Conflict of Interest - Disclosure Within the past 12 months, I have
More informationWho does not need a primary preventive ICD?
Who does not need a primary preventive ICD? Hildegard Tanner, Bern Universitätsklinik für Kardiologie Disclosure of potential conflicts of interest Travel grants for educational purposes from: Biosense
More informationNORTH OF ENGLAND CARDIOVASCULAR NETWORK (NECVN)
NORTH OF ENGLAND CARDIOVASCULAR NETWORK (NECVN) Operational policy for optimal management of: Cardiac pacemakers, Implantable cardioverter defibrillators (ICD), Cardiac resynchronisation devices, (CRT-D,
More informationEndpoints When Treating VT/VF in Patients with ICDs Programming Wojciech Zareba, MD, PhD
Endpoints When Treating VT/VF in Patients with ICDs Programming Wojciech Zareba, MD, PhD Professor of Cardiology/Medicine Director of the Heart Research Follow Up Program, University of Rochester, Rochester,
More informationBSH Annual Autumn Meeting 2017
BSH Annual Autumn Meeting 2017 Presentation title: The Development of CRT Speaker: John GF Cleland Conflicts of interest: I have received research support and honoraria from Biotronik, Boston Scientific,
More informationNew Developments in Cardiac Intervention. Dr Elaine Chau HK Sanatorium & Hospital, Hong Kong 17 January 2016
New Developments in Cardiac Intervention Dr Elaine Chau HK Sanatorium & Hospital, Hong Kong 17 January 2016 Cardiac Intervention Coronary intervention BVS Valvular heart disease TAVI, Mitraclips Ventricular
More informationThe Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia
The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia By Sandeep Joshi, MD and Jonathan S. Steinberg, MD Arrhythmia Service, Division of Cardiology
More information12 YEARS PROVEN PERFORMANCE. Sprint Quattro Defibrillation Leads
12 YEARS PROVEN PERFORMANCE Sprint Quattro Defibrillation Leads PROVEN PERFORMANCE. PROVEN RELIABLE. We made intentional choices and took proactive steps from lead design to active product monitoring to
More informationCall Medtronic at 1 (800) to verify the patient s current implanted system
MRI SURESCAN SYSTEMS Patient Scanning Process Transvenous Implantable Cardiac Systems PATIENT PRESCREENING SureScan Pacing, Defibrillation, and CRT (CRT-D and CRT-P) Systems Verification Verify that patient
More informationDECREASE-HF CLINICAL SUMMARY
CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY DECREASE-HF Boston Scientific Corporation
More informationLeadless Pacemakers A Review
Leadless Pacemakers A Review Mikhael El-Chami, MD, FACC Emory University Atlanta, Ga Consultant: Medtronic Boston Scientific Disclosure One of the leadless pacemakers discussed in this presentation is
More informationSubcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death
NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE Interventional procedure consultation document Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death A subcutaneous
More informationPediatric pacemakers & ICDs:
Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components
More informationMEDICAL POLICY Cardioverter Defibrillators
POLICY........ PG-0224 EFFECTIVE......06/01/09 LAST REVIEW... 01/27/17 MEDICAL POLICY Cardioverter Defibrillators GUIDELINES This policy does not certify benefits or authorization of benefits, which is
More informationRemote monotoring of cardiac rhythm devices: present and future Pacemaker and ICD
Remote monotoring of cardiac rhythm devices: present and future Pacemaker and ICD Philippe Mabo University Hospital, Rennes, France ESC congress, Paris 30 August 2011 U642 Disclosures Biotronik: research
More informationPatient Safety: the optimal lead body design
Patient Safety: the optimal lead body design E. Soldati U.O. Malattie Cardiovascolari II Azienda Ospedaliero Universitaria Pisana Advances in Cardiac Arrhythmias Torino, 25-27 Ottobre 2012 Lead Malfunction
More informationDual-Chamber Implantable Cardioverter-Defibrillator
February 1998 9 Dual-Chamber Implantable Cardioverter-Defibrillator A.SH. REVISHVILI A.N. Bakoulev Research Center for Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow, Russia Summary
More informationSilvia G Priori MD PhD
The approach to the cardiac arrest survivor Silvia G Priori MD PhD Molecular Cardiology, IRCCS Fondazione Salvatore Maugeri Pavia, Italy AND Leon Charney Division of Cardiology, Cardiovascular Genetics
More informationHF and CRT: CRT-P versus CRT-D
HF and CRT: CRT-P versus CRT-D Andrew E. Epstein, MD Professor of Medicine, Cardiovascular Division University of Pennsylvania Chief, Cardiology Section Philadelphia VA Medical Center Philadelphia, PA
More informationAdvances in Arrhythmia and Electrophysiology
Advances in Arrhythmia and Electrophysiology Troubleshooting Implanted Cardioverter Defibrillator Sensing Problems I Charles D. Swerdlow, MD; Samuel J. Asirvatham, MD; Kenneth A. Ellenbogen, MD; Paul A.
More informationRiata Lead Extraction. Thomas D. Callahan, MD, FACC, FHRS
Riata Lead Extraction Thomas D. Callahan, MD, FACC, FHRS Outline Riata lead history Lead design Lead Failure Extraction outcomes Techniques Special considerations Riata Lead Background 8F introduced in
More information