The pathology of failed McKee-Farrar implants: correlation with modern metal-on-metal-implant failure

Size: px
Start display at page:

Download "The pathology of failed McKee-Farrar implants: correlation with modern metal-on-metal-implant failure"

Transcription

1 J Mater Sci: Mater Med (2017) 28:66 DOI /s y BIOMATERIAL-TISSUE INTERACTION IN HUMANS Original Research The pathology of failed McKee-Farrar implants: correlation with modern metal-on-metal-implant failure Mitsuru Munemoto 1,2 George Grammatopoulos 1,3 Yasuhito Tanaka 2 Max Gibbons 3 Nicholas A. Athanasou 3 Received: 8 December 2016 / Accepted: 8 March 2017 / Published online: 22 March 2017 The Author(s) 2017; This article is published with open access at Springerlink.com Abstract The McKee-Farrar (MF) prosthesis was the first widely used total hip replacement (THR) to employ a metalon-metal (MoM) articulation. These implants had a high rate of early aseptic loosening but a number achieved good long-term implant survival, stimulating the reintroduction of second and third generation implants of this type. In this study we analysed archival histopathology of periprosthetic tissues in twenty cases of MF aseptic implant failure to determine if there was evidence of an innate and adaptive immune response similar to that seen in modern MoM implants. The presence of macrophages, the extent of necrosis and the ALVAL response were graded semiquantitatively. Variable but in most cases extensive tissue necrosis was associated with a heavy macrophage response to Cobalt-Chrome (Co-Cr) wear particles in periprosthetic tissues; most cases also contained evidence of a predominantly lymphocyte response which in eight cases was moderate or heavy (Oxford Grade 2/3). Our findings show that inflammatory and necrotic changes to deposition of Co- Cr wear particles are found in periprosthetic tissues of failed MF implants, indicating that there is an innate and adaptive response similar to that noted in second/third generation MoM implants; they also suggest that the pathobiological * George Grammatopoulos george.grammatopoulos@ndorms.ox.ac.uk Nuffield Department Orthopaedics Rheumatology & Musculoskeletal Sciences (NDORMS), University of Oxford, Windmill Road, Oxford, UK Dept of Orthopaedic Surgery, Nara medical University, Kashihara- City Nara, Japan Nuffield Orthopaedic Centre, Windmill Road, Headington, Oxford OX3 7LD, UK response to metal wear particles is likely to have contributed to MF implant failure in these cases. Graphical Abstract 1 Introduction G.K. McKee and J Watson-Farrar developed a metal on metal (MoM) hip joint prosthesis for total hip replacement (THR) in the 1960s [1 4]. Large numbers of these metalon-metal (MoM) implants were implanted but they eventually fell out of favour when superior outcomes were reported with the low-friction Charnley metal on polyethylene (MoP) prosthesis. It was noted that McKee-Farrar (MF) hip replacements had a high rate of aseptic loosening in the first 10 years; McKee stopped using the all-metal system in 1972 [1]. Failure of MF implants was thought to be due mainly to poor implant design [1, 5]. The stem of the femoral component was curved with narrow medial and lateral borders and had multiple sharp corners [1, 6]. These

2 66 Page 2 of 7 J Mater Sci: Mater Med (2017) 28:66 features resulted in areas of high localised stress within the cement mantle; the broad neck of the MF femoral component was also prone to impingement against the metal rim of the acetabular component. The original surgical technique for implantation did not include reaming and routine deepening of the acetabulum, and, as a result, the components did not have adequate lateral bone cover [1, 6]. Despite these shortcomings, some MF hip replacements have shown impressive survival outcomes with several lasting more than 20 years [5 10]. The relative success of some MF and other firstgeneration MoM implant designs stimulated the drive to develop second- and third-generation MoM hip implants [11]. The modern MoM implants are composed of a cobaltchrome molybdenum alloy similar to that used in the MF implant. Although some second and third-generation MoM designs have performed reasonably, a number have been associated with implant failure [12 14]. This is considered to be due in large measure to the pathobiological response to deposition of cobalt chrome (Co Cr) wear particles in periprosthetic tissues in which extensive necrosis an innate and foreign body macrophage response to Co Cr wear is often seen as well as prominent lymphoid infiltrate which is characteristically perivascular in distribution; this has been termed ALVAL (aseptic lymphocyte-dominated vasculitisassociated lesion) and is believed to reflect the adaptive immune response to Co Cr wear in periprosthetic tissues [13, 15 21]. There are relatively few reports on the histopathology of failed MF hip implants; some of these noted pathological changes similar to those reported in second- and thirdgeneration MoM hip implants [22 27]. With one exception [22], these studies examined only a handful of cases. In this study, we have analysed histopathological findings in periprosthetic tissues around twenty failed MF hip implants with regard to the inflammatory and necrotic changes known to be associated with second- and third-generation MoM hip implants. Our aim has been to determine whether Co-Cr particles from MF implants induced a similar pathobiological response in periprosthetic tissues, particularly with regard to the necrotic and inflammatory changes that reflect the innate (non-specific) and adaptive (specific) host immune response to deposition of Co Cr wear particles. 2 Materials and methods Periprosthetic tissue histology from twenty cases of failed MF hip implants undergoing revision arthroplasty for aseptic loosening were obtained from the files of the Histopathology Department, Nuffield Orthopaedic Centre, Oxford. The cases studied included 15 females and 5 males; the age range of the patients was 55 to 84 years. The original hip arthroplasty procedure was carried out between 1968 and 1974 and revision surgery was carried out between 1975 and There was limited clinical information and radiology for most cases was not available. Implant duration was 2 to 41 years. Clinical information provided on the pathology forms noted that the specimens were associated with failure of an MF hip implant: eleven of the cases referred to loosening of an MF implant. Two cases of sepsis were excluded. Details of patient age and implant duration are shown in Table 1. The tissues analysed were the joint capsule in all cases and the femoral and/or acetabular membrane in 10 cases. The tissues were fixed in 10% formalin, embedded in paraffin wax, sectioned and stained with haematoxylin-eosin. Tissue necrosis and the extent of the inflammatory cell infiltrate in MoM periprosthetic tissues was assessed semiquantitatively by histology as previously described [16, 17]. The number of macrophages was scored as 0 (absent), 1+ (few), 2+ (many), 3+ (abundant); necrosis was scored as 0 (absent), 1+ (scattered small necrotic areas), 2+ (frequent small or large necrotic areas: up to 25% tissue involvement), 3+ (extensive necrosis: more than 25% tissue involvement). The perivascular lymphoid infiltrate was graded using the Oxford ALVAL scoring system [16], this has been shown to correlate well with the Campbell scoring system [28]. In the Oxford System, grade 0 = no evidence of perivascular lymphocyte infiltrate; grade 1 = little evidence of a perivascular lymphocytic infiltrate, with lymphocyte cuffing of vessels being fewer than five cells in thickness; grade 2 = several perivascular lymphoid aggregates, with lymphocytic cuffing of vessels being five to ten cells in thickness; and grade 3 = numerous large perivascular lymphoid aggregates, with lymphocyte cuffing around vessels being more than ten cells in thickness. Immunohistochemistry was carried out using an indirect immunoperoxidase technique to determine the antigenic phenotype of phagocytic macrophages and lymphoid cells in the inflammatory infiltrate of periprosthetic tissues. The list of monoclonal antibodies used in this study and their antigen specificity is shown in Table 2. 3 Results The most notable feature seen in the joint capsule and periprosthetic fibrous tissue membrane of all MF implant failures was a variable but generally extensive degree of surface tissue necrosis associated with an underlying heavy foreign body macrophage infiltrate (Fig. 1a). The tissues had a prominent zone of surface necrosis in which there were viable macrophages and occasional giant cells containing Co-Cr particles; some macrophages showed changes

3 J Mater Sci: Mater Med (2017) 28:66 Page 3 of 7 66 Table 1 Case details and pathological findings in 20 failed McKee Farrar hip implants Case details Pathological findings Case No. Sex Age at revision Implant duration (year) Necrosis Macrophage Oxford ALVAL score 1 F F M F F F F F F F M F M M F F F M F F Table 2 Antigen specificity of monoclonal antibodies used in this study Antibody (clone) Antigen specificity Source F CD3: T cells Dakopatts NCL-L-CD4-368 CD4: T helper cells Novocastra CR/144B CD8: T suppressor cells Dakopatts CD CD14: monocytes macrophages Novocastra L26 CD20: B cells Dakopatts KP1 CD68: monocytes macrophages Dakopatts CR3/43 HLA-DR Dakopatts 124 BCL-2 oncoprotein: apoptosis Dakopatts NCL-CD163 CD163: macrophages Novocastra of apoptosis such as nuclear pyknosis (Fig. 1b). Beneath this zone there was a prominent macrophage infiltrate. Viable macrophages had vesicular nuclei and discrete plump cytoplasm which contained granular brown or black particles, indicating Co Cr wear particle phagocytosis (Fig. 2a). On the surface, there was macrophage necrosis and apoptosis, evident as loss of cytoplasmic outline, nuclear pyknosis and release of wear particles into the degenerate tissue matrix (Fig. 2b). Histological findings regarding the extent of necrosis and macrophage infiltration are summarised in Table 1. A heavy eosinophil polymorph infiltrate was noted in two specimens. A few neutrophil polymorphs (less than 1 per high power field on average) were noted in two specimens. In all but two of the cases, the sampled periprosthetic tissues contained a scattered, focally heavy lymphocyte infiltrate (Table 1); plasma cells were seen in three cases. Eight cases (some of which were associated with relatively short implant duration) had a moderate or heavy lymphocyte infiltrate with Grade 2/3 Oxford ALVAL scores (Fig. 3). In nine cases the Oxford ALVAL score was Grade 1. There were three cases in which The Oxford ALVAL score was 0; in these cases, specimens of the capsule were composed entirely of necrotic tissue with only a few scattered macrophages and no obvious lymphocyte infiltrate. Immunohistochemistry showed that the mononuclear phagocytic cells containing metal wear particles strongly expressed the monocyte/macrophage markers CD14, CD68 (Fig. 4a) and CD163; these cells also strongly expressed HLA-DR. Lymphocytes in the perivascular lymphoid aggregates expressed the T-cell marker CD3 (Fig. 4b). A few cells also expressed the B-cell marker CD20. Most CD3 + T cells also expressed CD4, a marker of T-helper cells, with a minority of cells weakly expressing CD8, which is characteristic of T-cytotoxic/suppressor cells. Weak expression of the apoptosis marker bcl-2 was noted in macrophages and some lymphoid cells.

4 66 Page 4 of 7 J Mater Sci: Mater Med (2017) 28:66 Fig. 1 a Histology of periprosthetic fibrous tissue membrane from a failed MF implant showing necrosis on the surface with underlying foreign body macrophage infiltrate. b High power view of area of surface necrosis in 1a. showing a viable giant cell and macrophages with vesicular nuclei (some marked by thick arrows); phagocytosed aggregates of metal wear particles are evident as a black dot in the cytoplasm. There are also apoptotic macrophages with pyknotic nuclei (some marked by thin arrows) Fig. 2 Histology of periprosthetic tissue from a failed MF implant showing: a viable macrophages with phagocytosed (brown-black) Co Cr wear particles of variable size (some arrowed) b necrotic and apoptotic macrophages with released wear particles (some arrowed) (color figure online) Fig. 3 Histology of periprosthetic tissue membrane from a failed MF implant showing: a a zone of surface necrosis beneath which there are several perivascular lymphoid aggregates (some arrowed). b High power view of lymphoid aggregates in 3a 4 Discussion Although the MoM MF total hip prosthesis was characterised by a higher rate of early- and intermediate-term aseptic loosening than the MoP Charnley implant, a few studies reported similar rates of long-term implant loosening. Jacobsson et al. [9] reported a 20 year implant survival of 77% in 107 consecutive MF arthroplasties; this compared with 73% for a similar population with the Charnley prosthesis. Brown et al. [29] reported a 28 year MF implant survivorship of 74% and noted that the reported survival of MoM total hip arthroplasties was 53 89% at 10 to 15 years while that of the Charnley arthroplasty was 75 84% over 20 to 30 years. A review of short-term outcomes of the first 100 cases that received a MF hip implant in Oxford reported good or excellent results in 90% of cases [30]. A subsequent

5 J Mater Sci: Mater Med (2017) 28:66 Page 5 of 7 66 Fig. 4 Immunohistochemistry of periprosthetic tissue membrane from a failed MF implant showing a staining of foreign body macrophages for CD68 (pan-macrophage marker) and b staining of perivascular lymphocytes for CD3 (T cell marker) review comparing results of the MF and the Charnley total hip prosthesis in over 200 cases reported superior clinical outcomes and fewer complications with the Charnley implant [31]. Amstutz and Grigoris reviewed the clinical results of MF and other first- generation MoM hip implants and noted, as in the present study, some impressive outcomes [1]. The main cause of early MF hip implant failure was aseptic loosening. This complication was ascribed mainly to the MF implant being biomechanically at a disadvantage compared with the Charnley implant in design terms [1, 2, 5, 6]. A number of reports, however, noted pathological changes in periprosthetic tissues that were proposed as contributory factors to the development of implant loosening. Evans et al. [24] reported necrosis of bone and capsular tissue around MoM hip implants and noted a correlation with the severity and extent of obliterative changes in vessels; it was suggested that these changes occurred as a result of a sensitivity reaction to Co-Cr ions derived from metal wear particles. Winter [27] described necrosis in tissues around retrieved MoM hip replacements with particles deposited in acellular collagen and in phagocytic cells Jones et al. [26] also reported necrosis of bone and capsular tissue from retrieved first-generation MoM prostheses and attributed this to a toxic and hypersensitivity reaction to cobalt which he suggested caused an avascular phenomenon. The study of Brown et al. [22] noted significant tissue necrosis and a chronic inflammatory cell infiltrate in soft tissue and bone; metallic debris within cells and a prominent lymphocytic infiltrate with occasional plasma cells were also identified. Doorn et al. [23] and Howie et al. [25] also noted tissue necrosis and the presence of lymphocytes in retrieved periprosthetic tissues around MF and other first-generation MoM hip implants. Willert et al. [32], investigating (nonmf) failed first-generation MoM total hip arthroplasties, also noted a lymphocyte response in peri-implant tissues. Other studies allude to the presence of macrophages and giant cells, but do not mention tissue necrosis or a lymphoid infiltrate [7, 33 35]. The MF hip implant was composed of a similar Co Cr alloy to that used in second- and third-generation MoM implants. In this study we show that periprosthetic tissues around MF implants exhibit similar pathological changes to those reported around modern MoM implants. Cell and tissue necrosis, a heavy infiltrate of macrophages in response to Co Cr wear particles and ALVAL-like lymphoid aggregates, commonly seen in failed modern MoM hip implants [15 20, 36, 37], were also noted in failed MF hip implants with approximately 40% of cases exhibiting a Grade 2/3 ALVAL response. The MoM articulation tends to produce less volumetric wear than other implant combinations such as MoP; however, the number of wear particles generated is an order of magnitude higher, the particles being much smaller and therefore much more numerous per unit volume [38, 39]. The nanometre-size Co Cr particles induce a profound innate and adaptive immune response. Macrophage phagocytosis of metal wear particles is probably the major factor in causing the extensive necrosis in MoM periprosthetic tissues. Co Cr particles are subject to corrosion and produce high levels of metal ions which are cytotoxic, causing apoptosis and necrosis of macrophages that release lysosomal enzymes into the tissues. Metal ions can also act as haptens, combining with cell and tissue proteins to induce a specific cell-mediated delayed hypersensitivity immune response. This is evidenced histologically in MoM implants by the presence of a predominantly T lymphocyte infiltrate which is often perivascular in nature [17, 36, 37]. T cells predominated in the lymphoid aggregates seen in failed MF cases, suggesting that the adaptive immune response to metal wear is similar. The presence of profound necrotic and inflammatory changes in the MF periprosthetic tissues would indicate that Co Cr wear particle-induced hypersensitivity and cytotoxicity are likely to have contributed, at least in part, to implant failure. A major limitation of our study is that, due to the many years that have elapsed, we do not have full clinical and radiological information on the cases we have

6 66 Page 6 of 7 J Mater Sci: Mater Med (2017) 28:66 analysed. Pseudotumours have been strongly associated with failure of modern MoM hip implants [15, 16, 19, 20] and, though not seen in our cases, there are two reports of probable pseudotumour-like lesions associated with MF implant failure [23, 40]. Acknowledgements We would like to thank Sarah Turton for typing the manuscript, and Professor G Bentley and Professor J. Kenwright for helpful discussions on the cases included in this study. The authors acknowledge the support of the Rosetrees Trust and NIHR Biomedical Research Unit. Conflict of interest N.A.A. has provided expert medicolegal opinion in arthroplasty-related cases. The other authors declare that they have no competing interests. Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License ( creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. References 1. Amstutz HC, Grigoris P. Metal on metal bearings in hip arthroplasty. Clin Orthop Relat Res. 1996;329 Suppl:S Cuckler JM. The rationale for metal-on-metal total hip arthroplasty. Clin Orthop Relat Res. 2005;441: Knight SR, Aujla R, Biswas SP. Total hip arthroplasty - over 100 years of operative history. Orthop Rev (Pavia). 2011;3 2:e16 4. McKee GK, Watson-Farrar J. Replacement of arthritic hips by the McKee-Farrar prosthesis. J Bone Joint Surg Br. 1966;48 2: Schmalzried TP, Szuszczewicz ES, Akizuki KH, Petersen TD, Amstutz HC. Factors correlating with long term survival of McKee-Farrar total hip prostheses. Clin Orthop Relat Res. 1996;329 Suppl:S Walker PS, Salvati E, Hotzler RK. The wear on removed McKee- Farrar total hip prostheses. J Bone Joint Surg Am. 1974;56 1: Clarke MT, Darrah C, Stewart T, Ingham E, Fisher J, Nolan JF. Long-term clinical, radiological and histopathological follow-up of a well-fixed Mckee-Farrar metal-on-metal total hip arthroplasty. J Arthroplasty. 2005;20 4: Jacobsson SA, Djerf K, Wahlstrom O. A comparative study between McKee-Farrar and Charnley arthroplasty with long-term follow-up periods. J Arthroplasty. 1990;5 1: Jacobsson SA, Djerf K, Wahlstrom O. Twenty-year results of McKee-Farrar versus Charnley prosthesis. Clin Orthop Relat Res. 1996;329 Suppl:S Jantsch S, Schwagerl W, Zenz P, Semlitsch M, Fertschak W. Long-term results after implantation of McKee-Farrar total hip prostheses. Arch Orthop Trauma Surg. 1991;110 5: Amstutz HC, Campbell P, McKellop H, Schmalzreid TP, Gillespie WJ, Howie D, Jacobs J, Medley J, Merritt K. Metal on metal total hip replacement workshop consensus document. Clin Orthop Relat Res. 1996;329 Suppl:S de Steiger RN, Hang JR, Miller LN, Graves SE, Davidson DC. Five-year results of the ASR XL acetabular system and the ASR hip resurfacing system: an analysis from the Australian orthopaedic association national joint replacement registry. J Bone Joint Surg Am. 2011;93 24: Langton DJ, Jameson SS, Joyce TJ, Gandhi JN, Sidaginamale R, Mereddy P, Lord J, Nargol AV. Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br. 2011;93 8: Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br. 2010;92 1: Bauer T. What does the histology tell us? In: Jones LC, Greenwold AS, Haggard WO, editors. Metal-on-metal bearing. A clinical practicum. New York, Springer; P Grammatopoulos G, Pandit H, Kamali A, Maggiani F, Glyn-Jones S, Gill HS, Murray DW, Athanasou N. The correlation of wear with histological features after failed hip resurfacing arthroplasty. J Bone Joint Surg Am. 2013;95 12:e Mahendra G, Pandit H, Kliskey K, Murray D, Gill HS, Athanasou N. Necrotic and inflammatory changes in metal-on-metal resurfacing hip arthroplasties. Acta Orthop. 2009;80 6: Natu S, Sidaginamale RP, Gandhi J, Langton DJ, Nargol AV. Adverse reactions to metal debris: histopathological features of periprosthetic soft tissue reactions seen in association with failed metal on metal hip arthroplasties. J Clin Pathol. 2012;65 5: Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons CL, Ostlere S, Athanasou N, Gill HS, Murray DW. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90 7: Campbell P, Ebramzadeh E, Nelson S, Takamura K, De Smet K, Amstutz HC. Histological features of pseudotumour-like tissues from metal-on-metal hips. Clin Orthop Relat Res. 2010;468: Watters TS, Cardona DM, Menon KS, Vinson EN, Bolognesi MP, Dodd LG. Aseptic lymphocyte-dominated vasculitis-associated lesion: a clinicopathologic review of an underrecognized cause of prosthetic failure. Am J Clin Pathol. 2010;134 6: Brown GC, Lockshin MD, Salvati EA, Bullough PG. Sensitivity to metal as a possible cause of sterile loosening after cobaltchromium total hip-replacement arthroplasty. J Bone Joint Surg Am. 1977;59 2: Doorn PF, Mirra JM, Campbell PA, Amstutz HC. Tissue reaction to metal on metal total hip prostheses. Clin Orthop Relat Res. 1996;329 Suppl:S Evans EM, Freeman MA, Miller AJ, Vernon-Roberts B. Metal sensitivity as a cause of bone necrosis and loosening of the prosthesis in total joint replacement. J Bone Joint Surg Br. 1974;56-B-4: Howie DW. Tissue response in relation to type of wear particles around failed hip arthroplasties. J Arthroplasty. 1990;5-4: Jones DA, Lucas HK, O Driscoll M, Price CH, Wibberley B. Cobalt toxicity after McKee hip arthroplasty. J Bone Joint Surg Br. 1975;57 3: Winter G. Tissue recation to metallic wear and corrosion products in human patients. J Biomed Materials Res. 1974;8: Phillips EA, Klein GR, Cates HE, Kurtz SM, Steinbeck M. Histological characterization of periprosthetic tissue responses for metal-on-metal hip replacement. J Long Term Eff Med Implants. 2014;24: Brown SR, Davies WA, DeHeer DH, Swanson AB. Long-term survival of McKee-Farrar total hip prostheses. Clin Orthop Relat Res. 2002;402: Somerville EW, Bentley GJ. McKee-Farrar total hip replacement arthroplasty. a review of the first 100 cases at the nuffield orthopaedic centre. Orthopaedics Oxford 1970;3:73 7.

7 J Mater Sci: Mater Med (2017) 28:66 Page 7 of Bentley G, Duthie RB. A comparative review of the McKee- Farrar and Charnley total hip prostheses. Clin Orthop Relat Res. 1973;95: Willert HG, Buchhorn GH, Gobel D, Koster G, Schaffner S, Schenk R, Semlitsch M. Wear behavior and histopathology of classic cemented metal on metal hip endoprostheses. Clin Orthop Relat Res. 1996;329 Suppl:S Higuchi F, Inoue A, Semlitsch M. Metal-on-metal CoCrMo McKee- Farrar total hip arthroplasty: characteristics from a long-term followup study. Arch Orthop Trauma Surg. 1997;116 3: Schmalzried TP, Peters PC, Maurer BT, Bragdon CR, Harris WH. Long-duration metal-on-metal total hip arthroplasties with low wear of the articulating surfaces. J Arthroplasty. 1996;11 3: Zahiri CA, Schmalzried TP, Ebramzadeh E, Szuszczewicz ES, Salib D, Kim C, Amstutz HC. Lessons learned from loosening of the McKee-Farrar metal-on-metal total hip replacement. J Arthroplasty. 1999;14 3: Davies AP, Willert HG, Campbell PA, Learmonth ID, Case CP. An unusual lymphocytic perivascular infiltration in tissues around contemporary metal-on-metal joint replacements. J Bone Joint Surg Am. 2005;87 1: Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G, Lohmann CH. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study. J Bone Joint Surg Am. 2005;87 1: Hallab NJ, Jacobs JJ. Biological effects of implant debris. Bull NYU Hospit Jt Dis. 2009;67: Athanasou NA. The pathobiology and pathology of aseptic implant failure. Bone Joint Res. 2016;162: Madan S, Jowett RL, Goodwin MI. Recurrent intrapelvic cyst complicating metal-on-metal cemented total hip arthroplasty. Arch Orthop Trauma Surg. 2000;120 9:

Implant failure in bilateral metal-on-metal hip resurfacing arthroplasties: a clinical and pathological study

Implant failure in bilateral metal-on-metal hip resurfacing arthroplasties: a clinical and pathological study Journal of Materials Science: Materials in Medicine (2018) 29:28 https://doi.org/10.1007/s10856-018-6037-5 S.I.: BIOMATERIAL-TISSUE INTERACTION IN HUMANS Original Research Implant failure in bilateral

More information

Histological Features of Pseudotumor-like Tissues From Metal-on-Metal Hips

Histological Features of Pseudotumor-like Tissues From Metal-on-Metal Hips Clin Orthop Relat Res (2010) 468:2321 2327 DOI 10.1007/s11999-010-1372-y SYMPOSIUM: COMPLICATIONS OF HIP ARTHROPLASTY Histological Features of Pseudotumor-like Tissues From Metal-on-Metal Hips Pat Campbell

More information

CASE 56 Hip Pain. ? Diagnosis

CASE 56 Hip Pain. ? Diagnosis CSE 56 Hip Pain PIN IN THE HIP: 1. 50-year-old female who had undergone total hip arthroplasty [Metal on metal] of the right hip 3.4 years earlier. Shee had a tingling sensation in the trochanteric region

More information

Painful Metal-on-Metal Total Hip Arthroplasty

Painful Metal-on-Metal Total Hip Arthroplasty The Journal of Arthroplasty Vol. 21 No. 2 2006 Painful Metal-on-Metal Total Hip Arthroplasty Cambize Shahrdar, MD,* Pat Campbell, PhD,y Joseph Mirra, MD,y and Lawrence D. Dorr, MD* Abstract: Two patients

More information

THE CLINICAL PERFORMANCE OF METAL-ON-METAL AS AN ARTICULATION SURFACE IN TOTAL HIP REPLACEMENT

THE CLINICAL PERFORMANCE OF METAL-ON-METAL AS AN ARTICULATION SURFACE IN TOTAL HIP REPLACEMENT THE CLINICAL PERFORMANCE OF METAL-ON-METAL AS AN ARTICULATION SURFACE IN TOTAL HIP REPLACEMENT William T. Long, M.D.* ABSTRACT The metal-on-metal articulations in total hip arthroplasty (THA) were widely

More information

Lari Lehtovirta 1,2*, Aleksi Reito 2, Jyrki Parkkinen 3, Harry Hothi 4, Johann Henckel 4, Alister Hart 4 and Antti Eskelinen 2

Lari Lehtovirta 1,2*, Aleksi Reito 2, Jyrki Parkkinen 3, Harry Hothi 4, Johann Henckel 4, Alister Hart 4 and Antti Eskelinen 2 Lehtovirta et al. BMC Musculoskeletal Disorders (2017) 18:523 DOI 10.1186/s12891-017-1894-5 RESEARCH ARTICLE Open Access Analysis of bearing wear, whole blood and synovial fluid metal ion concentrations

More information

Adverse Local Tissue Reaction in Total Hip Replacement

Adverse Local Tissue Reaction in Total Hip Replacement Adverse Local Tissue Reaction in Total Hip Replacement Authors 1 David C Evans, 2 Ross A Miller, 3 Stephen J Incavo 1 MD Candidate, Baylor College of Medicine Class of 2018 2 Houston Methodist Hospital,

More information

Cover Page. The handle holds various files of this Leiden University dissertation

Cover Page. The handle   holds various files of this Leiden University dissertation Cover Page The handle http://hdl.handle.net/1887/25896 holds various files of this Leiden University dissertation Author: Weegen, Walter van der Title: Metal-on-metal hip arthroplasty : local tissue reactions

More information

Metal-on-metal articulations generate approximately to particles PSEUDOTUMORS FACTORS PATHOGENESIS

Metal-on-metal articulations generate approximately to particles PSEUDOTUMORS FACTORS PATHOGENESIS PSEUDOTUMORS Vasu Pai ARMD Adverse reaction to metal debris. Langton identified that there is no clear consensus in the literature defining the boundaries of the terms metallosis, aseptic lymphocyte-dominated

More information

Cover Page. The handle holds various files of this Leiden University dissertation

Cover Page. The handle   holds various files of this Leiden University dissertation Cover Page The handle http://hdl.handle.net/1887/25896 holds various files of this Leiden University dissertation Author: Weegen, Walter van der Title: Metal-on-metal hip arthroplasty : local tissue reactions

More information

Scandinavian Journal of Surgery 103: 54 59, 2013

Scandinavian Journal of Surgery 103: 54 59, 2013 345SJS103110.1177/1457496913495345Hip resurfacing arthroplasty vs. large headed metal-on-metal total hip arthroplastym. Junnila, et al. ORIGINAL ARTICLE Scandinavian Journal of Surgery 103: 54 59, 2013

More information

Diagnostic utility of joint fluid metal ion measurement for histopathological findings in metal-on-metal hip replacements

Diagnostic utility of joint fluid metal ion measurement for histopathological findings in metal-on-metal hip replacements Reito et al. BMC Musculoskeletal Disorders (2015) 16:393 DOI 10.1186/s12891-015-0851-4 RESEARCH ARTICLE Open Access Diagnostic utility of joint fluid metal ion measurement for histopathological findings

More information

Session 3: Controversies for Bearing Surfaces in THR

Session 3: Controversies for Bearing Surfaces in THR : Controversies for Bearing Surfaces in THR Learning Objectives Upon completion of this activity, participants should be able to: 1. Understand the indications and known methods for the use of each of

More information

Immunological response to metallic implants

Immunological response to metallic implants Immunological response to metallic implants Doc. dr. Peter Korošec Head of Laboratory for Clinical Immunology & Molecular Genetics Head of Research & Development Department University Clinic of Respiratory

More information

Pseudotumours associated with metal-onmetal hip resurfacings

Pseudotumours associated with metal-onmetal hip resurfacings Pseudotumours associated with metal-onmetal hip resurfacings H. Pandit, S. Glyn-Jones, P. McLardy-Smith, R. Gundle, D. Whitwell, C. L. M. Gibbons, S. Ostlere, N. Athanasou, H. S. Gill, D. W. Murray From

More information

Immune response in adverse reactions to metal debris following metal-on-metal total hip arthroplasty

Immune response in adverse reactions to metal debris following metal-on-metal total hip arthroplasty Hasegawa et al. BMC Musculoskeletal Disorders (2016) 17:221 DOI 10.1186/s12891-016-1069-9 RESEARCH ARTICLE Open Access Immune response in adverse reactions to metal debris following metal-on-metal total

More information

Adverse reaction metal debris (ARMD) after metal-on-metal (MOM) hip replacement: magnetic resonance imaging features with surgical feedback

Adverse reaction metal debris (ARMD) after metal-on-metal (MOM) hip replacement: magnetic resonance imaging features with surgical feedback Adverse reaction metal debris (ARMD) after metal-on-metal (MOM) hip replacement: magnetic resonance imaging features with surgical feedback Poster No.: C-0533 Congress: ECR 2012 Type: Scientific Paper

More information

Metal ion levels in patients with a lumbar metal-on-metal Total Disc Replacement:

Metal ion levels in patients with a lumbar metal-on-metal Total Disc Replacement: Metal ion levels in patients with a lumbar metal-on-metal Total Disc Replacement: Similar concerns compared to metal bearing Total Hip Arthroplasties? EuroSpine 2011, Milan Job L.C. van Susante, MD, PhD

More information

Case Report Inflammatory Pseudotumor Complicated by Recurrent Dislocations after Revision Total Hip Arthroplasty

Case Report Inflammatory Pseudotumor Complicated by Recurrent Dislocations after Revision Total Hip Arthroplasty Case Reports in Orthopedics, Article ID 792781, 5 pages http://dx.doi.org/10.1155/2014/792781 Case Report Inflammatory Pseudotumor Complicated by Recurrent Dislocations after Revision Total Hip Arthroplasty

More information

Examination of Porous-Coated Patellar Components and Analysis of the Reasons for Their Retrieval

Examination of Porous-Coated Patellar Components and Analysis of the Reasons for Their Retrieval Examination of Porous-Coated Patellar Components and Analysis of the Reasons for Their Retrieval J. P. Collier,* M. B. Mayor,+ V. A. Surprenant,* H. P. Surprenant,* and R.E. Jensen* *Dartmouth Biomedical

More information

Partial hip resurfacing is considered investigational/not medically necessary for all other indications not listed above.

Partial hip resurfacing is considered investigational/not medically necessary for all other indications not listed above. Subject: Policy #: Current Effective Date: 11/13/2006 Status: Revised Last Review Date: 09/14/2006 Description/Scope Hip resurfacing can be categorized in two ways: as a partial (hemi) hip resurfacing

More information

Avascular necrosis of the hip treated with Metal-on-Metal. surface hip arthroplasty

Avascular necrosis of the hip treated with Metal-on-Metal. surface hip arthroplasty Avascular necrosis of the hip treated with Metal-on-Metal surface hip arthroplasty H Sharma ( ), C Watson, AC Campbell Department of Trauma and Orthopaedics, Monklands Hospital, Airdrie, Lanarkshire, ML6

More information

Continuing the Tradition. VerSys Heritage Hip System

Continuing the Tradition. VerSys Heritage Hip System Continuing the Tradition VerSys Heritage Hip System Heritage Following the Tradition The low-friction hip prosthesis developed by Sir John Charnley has more than a 20-year history of outstanding results.

More information

Eleven-year follow-up of second-generation metal-on-metal total hip arthroplasty

Eleven-year follow-up of second-generation metal-on-metal total hip arthroplasty Journal of Orthopaedic Surgery 2010;18(1):15-21 Eleven-year follow-up of second-generation metal-on-metal total hip arthroplasty Myung-Sik Park, 1 Woo-Chul Chung, 1 Sun-Jung Yoon, 1 Hong-Man Cho, 2 Suk-Hyun

More information

OUTCOMES AFTER REVISION OF METAL-ON-METAL HIP RESURFACING ARTHROPLASTY THOMAS P. GROSS, MD FEI LIU, PHD MIDLANDS ORTHOPAEDICS, P.A.

OUTCOMES AFTER REVISION OF METAL-ON-METAL HIP RESURFACING ARTHROPLASTY THOMAS P. GROSS, MD FEI LIU, PHD MIDLANDS ORTHOPAEDICS, P.A. *Title Page (with full author details) OUTCOMES AFTER REVISION OF METAL-ON-METAL HIP RESURFACING ARTHROPLASTY THOMAS P. GROSS, MD FEI LIU, PHD MIDLANDS ORTHOPAEDICS, P.A. COLUMBIA, SC, USA Send Correspondence

More information

Imaging features of complications following hip replacement: A pictorial assay

Imaging features of complications following hip replacement: A pictorial assay Imaging features of complications following hip replacement: A pictorial assay Poster No.: C-2041 Congress: ECR 2014 Type: Educational Exhibit Authors: K. Pilania, B. Jankharia ; Mumbai, maharashtra/in,

More information

Metal Ion Levels Not Sufficient as a Screening Measure for Adverse Reactions in Metal-on- Metal Hip Arthroplasties

Metal Ion Levels Not Sufficient as a Screening Measure for Adverse Reactions in Metal-on- Metal Hip Arthroplasties The Journal of Arthroplasty Vol. 28 No. 1 2013 Metal Ion Levels Not Sufficient as a Screening Measure for Adverse Reactions in Metal-on- Metal Hip Arthroplasties Rory D. Macnair, MBBS, MRCS, MSc,* Henry

More information

Lesson Learned From Retrieval Analysis Of A Dislocating, Large Diameter MoM Revision THA A Case Report

Lesson Learned From Retrieval Analysis Of A Dislocating, Large Diameter MoM Revision THA A Case Report Chagrin Falls, Ohio, USA Lesson Learned From Retrieval Analysis Of A Dislocating, Large Diameter MoM Revision THA A Case Report Edward J. McPherson, MD, FACS º Ian Clarke, PhD*º Thomas K. Donaldson, MD

More information

Enhancing stability and increasing range of motion. Metasul LDH Large Diameter Head

Enhancing stability and increasing range of motion. Metasul LDH Large Diameter Head Enhancing stability and increasing range of motion Metasul LDH Large Diameter Head The Metasul large diameter head technology is the result of in-depth research, development and clinical experience that

More information

Imaging of Metallic Prostheses Using Novel Sequences: Early Experience

Imaging of Metallic Prostheses Using Novel Sequences: Early Experience Imaging of Metallic Prostheses Using Novel Sequences: Early Experience Leon D. Rybak, M.D. 1 ; Mary Bruno, B.S. 1 ; Mathias Nittka, Ph.D. 2 ; Christian Geppert, Ph.D. 3 ; Holly Delaney, M.D. 1 ; Park Jong,

More information

The development of alternate bearing surfaces in total

The development of alternate bearing surfaces in total S47 Adverse Soft-Tissue Reactions Around Non-Metal-on-Metal Total Hip Arthroplasty A Systematic Review of the Literature Alberto Carli, M.D., Avishai Reuven, M.D., David J. Zukor, M.D., F.R.C.S.(C), and

More information

Poor Survivorship and Frequent Complications at a Median of 10 Years After Metal-on-Metal Hip Resurfacing Revision

Poor Survivorship and Frequent Complications at a Median of 10 Years After Metal-on-Metal Hip Resurfacing Revision Clin Orthop Relat Res (2017) 475:304 314 DOI 10.1007/s11999-016-4882-4 Clinical Orthopaedics and Related Research A Publication of The Association of Bone and Joint Surgeons SYMPOSIUM: 2016 HIP SOCIETY

More information

The inflammatory process and immune reactions associated with implant use

The inflammatory process and immune reactions associated with implant use The inflammatory process and immune reactions associated with implant use Professor Peter A Revell Implant Biology Research Biomaterials and Tissue Engineering Division Eastman Dental Institute London

More information

Adverse reaction to metal debris following hip resurfacing

Adverse reaction to metal debris following hip resurfacing HIP Adverse reaction to metal debris following hip resurfacing THE INFLUENCE OF COMPONENT TYPE, ORIENTATION AND VOLUMETRIC WEAR D. J. Langton, T. J. Joyce, S. S. Jameson, J. Lord, M. Van Orsouw, J. P.

More information

Protocol. Hip Resurfacing

Protocol. Hip Resurfacing Protocol Hip Resurfacing (70180) Medical Benefit Effective Date: 10/01/14 Next Review Date: 11/19 Preauthorization No Review Dates: 07/12, 07/13, 07/14, 07/15, 11/15, 11/16, 11/17, 11/18 Preauthorization

More information

An assessment of the histological criteria used to diagnose infection in hip revision arthroplasty tissues

An assessment of the histological criteria used to diagnose infection in hip revision arthroplasty tissues 118 Department of Orthopaedic Surgery, NuYeld Orthopaedic Centre, Oxford OX3 7LD, UK R Pandey E Drakoulakis Department of Pathology, NuYeld Orthopaedic Centre N A Athanasou Correspondence to: Dr Athanasou.

More information

Surface Characterization of Retrieved Metal-on-Metal Total Hip Implants from Patients with Adverse Reaction to Metal Debris

Surface Characterization of Retrieved Metal-on-Metal Total Hip Implants from Patients with Adverse Reaction to Metal Debris Materials 2014, 7, 1866-1879; doi:10.3390/ma7031866 Article OPEN ACCESS materials ISSN 1996-1944 www.mdpi.com/journal/materials Surface Characterization of Retrieved Metal-on-Metal Total Hip Implants from

More information

Contamination of revision procedures in patients with adverse tissues reaction to metal on metal implant

Contamination of revision procedures in patients with adverse tissues reaction to metal on metal implant European Review for Medical and Pharmacological Sciences 019; 3( Suppl.): -93 Contamination of revision procedures in patients with adverse tissues reaction to metal on metal implant F. FAVETTI 1, G. MAZZOTTA,

More information

MRI Findings in Painful Metal-on- Metal Hip Arthroplasty

MRI Findings in Painful Metal-on- Metal Hip Arthroplasty Musculoskeletal Imaging Original Research Hayter et al. MRI of Painful Metal-on-Metal Hip Prostheses Musculoskeletal Imaging Original Research Catherine L. Hayter 1 Stephanie L. Gold 1 Matthew F. Koff

More information

Soft Tissue Imaging in. Total Hip Arthroplasty

Soft Tissue Imaging in. Total Hip Arthroplasty FDA Orthopaedic Rehabilitation Devices Panel Medical Devices Advisory Committee Meeting Thursday June 28th 2012 Soft Tissue Imaging in Metal-on on-metal Total Hip Arthroplasty Young-Min Kwon MD, PhD, FRCS,

More information

Cover Page. The handle holds various files of this Leiden University dissertation

Cover Page. The handle   holds various files of this Leiden University dissertation Cover Page The handle http://hdl.handle.net/1887/25896 holds various files of this Leiden University dissertation Author: Weegen, Walter van der Title: Metal-on-metal hip arthroplasty : local tissue reactions

More information

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Diagnosis: Ceramic head fracture In the 1970 s, Boutin implemented ceramic in modern total hip arthroplasty (THA).

More information

AGGRESSIVE GRANULOMATOUS LESIONS IN CEMENTLESS TOTAL HIP ARTHROPLASTY

AGGRESSIVE GRANULOMATOUS LESIONS IN CEMENTLESS TOTAL HIP ARTHROPLASTY AGGRESSIVE GRANULOMATOUS LESIONS IN CEMENTLESS TOTAL HIP ARTHROPLASTY SEPPO SANTAVIRTA, VEIJO HOIKKA, ANTTI ESKOLA, YRJO T. KONTTINEN, TIMO PAAVILAINEN, KAJ TALLROTH From the Orthopaedic Hospitalofthe

More information

Protocol. Hip Resurfacing

Protocol. Hip Resurfacing Protocol Hip Resurfacing (70180) Medical Benefit Effective Date: 10/01/14 Next Review Date: 11/18 Preauthorization No Review Dates: 07/12, 07/13, 07/14, 07/15, 11/15, 11/16, 11/17 Preauthorization is not

More information

Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty

Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty A. J. Price, A. Short, C. Kellett, D. Beard, H. Gill, H. Pandit, C. A. F. Dodd, D. W. Murray From

More information

Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement

Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement rrelation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement R. De Haan, C. Pattyn, H. S. Gill, D. W. Murray, P. A. Campbell, K. De Smet

More information

Pseudo tumor a cystic or solid mass (MRI or ultrasound) 1% of MOM pts develop pseudo tumors by 5 years 17 pts, all female, Pandit et al (JBJS, 2008)

Pseudo tumor a cystic or solid mass (MRI or ultrasound) 1% of MOM pts develop pseudo tumors by 5 years 17 pts, all female, Pandit et al (JBJS, 2008) Pseudo tumor a cystic or solid mass (MRI or ultrasound) 1% of MOM pts develop pseudo tumors by 5 years 17 pts, all female, Pandit et al (JBJS, 2008) 8% at 5yrs FU indicated by preliminary study from same

More information

A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY

A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY by Timothy McTighe, Exec. Dir., JISRF, Chagrin Falls, Ohio Ying Ko, Ph.D., Cincinnati, Ohio Russell B. Bennett, Ph.D., Cincinnati, Ohio James

More information

Bone Bangalore

Bone Bangalore Dr Suresh Annamalai MBBS, MRCS(Edn), FRCS( Tr & Orth)(Edn), FEBOT(European Board), Young Hip and Knee Fellowship(Harrogate, UK) Consultant Arthroplasty and Arthroscopic Surgeon Manipal Hospital, Whitefield,

More information

Imaging Metal-on-Metal Hip Replacements: the Norwich Experience

Imaging Metal-on-Metal Hip Replacements: the Norwich Experience HSSJ (2013) 9:247 256 DOI 10.1007/s11420-013-9357-5 REVIEW ARTICLE Imaging Metal-on-Metal Hip Replacements: the Norwich Experience Marianna S. Thomas, FRCR & James A. Wimhurst, FRCS(Orth) & John F. Nolan,

More information

Good sensitivity and specificity of ultrasound for detecting pseudotumors in 83 failed metal-on-metal hip replacements

Good sensitivity and specificity of ultrasound for detecting pseudotumors in 83 failed metal-on-metal hip replacements Acta Orthopaedica 2015; 86 (3): 339 344 339 Good sensitivity and specificity of ultrasound for detecting pseudotumors in 83 failed metal-on-metal hip replacements Olli Lainiala, Petra Elo, Aleksi Reito,

More information

Metal Ion Measurement as a Diagnostic Tool to Identify Problems with Metal-on-Metal Hip Resurfacing

Metal Ion Measurement as a Diagnostic Tool to Identify Problems with Metal-on-Metal Hip Resurfacing 202 COPYRIGHT 2008 BY THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED Metal Ion Measurement as a Diagnostic Tool to Identify Problems with Metal-on-Metal Hip Resurfacing By K. De Smet, MD, R. De Haan,

More information

Metal-on-metal bearings

Metal-on-metal bearings INSTRUCTIONAL REVIEW: HIP Metal-on-metal bearings THE EVIDENCE SO FAR F. S. Haddad, R. R. Thakrar, A. J. Hart, J. A. Skinner, A. V. F. Nargol, J. F. Nolan, H. S. Gill, D. W. Murray, A. W. Blom, C. P. Case

More information

BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code

BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code Patrick Chaboche, MD patrick.chaboche@wanadoo.fr Former Head of Department of Orthopaedic Surgery. Centre

More information

Metal-on-metal total hip arthroplasty: is there still a role in 2016?

Metal-on-metal total hip arthroplasty: is there still a role in 2016? Curr Rev Musculoskelet Med. 2016 Mar;; 9(1): 93 96. Published online 2016 Jan 20. doi: 10.1007/s12178-016-9323-1 PMCID: PMC4762806 Metal-on-metal total hip arthroplasty: is there still a role in 2016?

More information

Chapter 2 Bearing Surfaces for Joint Replacement: New Materials or New Problems

Chapter 2 Bearing Surfaces for Joint Replacement: New Materials or New Problems Chapter 2 Bearing Surfaces for Joint Replacement: New Materials or New Problems Stuart Goodman Total joint replacement (TJR) is one of the greatest technological advances in all of surgery. Hip, knee,

More information

PUBLISHED VERSION. Originally Published at: PERMISSIONS

PUBLISHED VERSION. Originally Published at:   PERMISSIONS PUBLISHED VERSION Marianne H Gillam, Nicole L Pratt, Maria C S Inacio, Elizabeth E Roughead, Sepehr Shakib, Stephen J Nicholls & Stephen E Graves Heart failure after conventional metal-on-metal hip replacements:

More information

Implant based differences in adverse local tissue reaction in failed total hip arthroplasties: a morphological and immunohistochemical study

Implant based differences in adverse local tissue reaction in failed total hip arthroplasties: a morphological and immunohistochemical study Perino et al. BMC Clinical Pathology 2014, 14:39 RESEARCH ARTICLE Open Access Implant based differences in adverse local tissue reaction in failed total hip arthroplasties: a morphological and immunohistochemical

More information

Poor mid-term results of total hip arthroplasty with use of a Hylamer liner

Poor mid-term results of total hip arthroplasty with use of a Hylamer liner Acta Orthop. Belg., 2008, 74, 337-342 ORIGINAL STUDY Poor mid-term results of total hip arthroplasty with use of a Hylamer liner Adrian SKWARA, Simone STRACKE, Carsten O. TIBESKU, Susanne FUCHS-WINKELMANN

More information

Results of Conversion Total Hip Prosthesis Performed Following Painful Hemiarthroplasty

Results of Conversion Total Hip Prosthesis Performed Following Painful Hemiarthroplasty M Nomura, S The Journal et al. of International Medical Research Endovascular 2000; 28: Embolization 307 312 of Unruptured Results of Conversion Total Hip Prosthesis Performed Following Painful Hemiarthroplasty

More information

Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs: Report of Two Cases

Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs: Report of Two Cases SAGE-Hindawi Access to Research Advances in Orthopedics Volume 2011, Article ID 519254, 4 pages doi:10.4061/2011/519254 Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs:

More information

ARTICLE IN PRESS. Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction. Effects on Polyethylene Wear

ARTICLE IN PRESS. Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction. Effects on Polyethylene Wear The Journal of Arthroplasty Vol. 00 No. 0 2008 Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction Effects on Polyethylene Wear William L. Jaffe, MD,* Eric J. Strauss,

More information

Hip Resurfacing Corporate Medical Policy

Hip Resurfacing Corporate Medical Policy Hip Resurfacing Corporate Medical Policy File name: Hip Resurfacing File code: UM.ORTHO.01 Last Review: 09/2015 Next Review: 09/2016 Effective Date: 04/01/16 Description/Summary Hip resurfacing is an alternative

More information

Table of content to which comment refers. Name of individual/ SCENIHRs Response. organisation

Table of content to which comment refers. Name of individual/ SCENIHRs Response. organisation Results of the public consultation on SCENIHR's preliminary opinion on Safety of Metal-on-Metal joint replacements with a particular focus on hip implants A public consultation on this opinion was opened

More information

The power of simplicity

The power of simplicity The power of simplicity The power of simplicity Consider the facts, and take a look at the future: Revision TKA procedures are projected to double between now and 2015. 1 Restoration of a normal joint

More information

Peachy Mae T. Piana, MD

Peachy Mae T. Piana, MD Peachy Mae T. Piana, MD OUTLINE History of Total Hip Arthroplasty MoM Prosthesis Production of Metal Particles Defining Terminology: Metallosis, ARMD, ALTR, ALVAL, pseudotumor, periprosthetic reactive

More information

The Challenge of Durability in Hip Replacement

The Challenge of Durability in Hip Replacement The Challenge of Durability in Hip Replacement Gregory A. Tocks, D.O. Orthopedic Associates of Lancaster Ltd. INTRODUCTION Hip replacement is a very common operation, and the number of long-term postoperative

More information

Plain Film CT. Principal Modality (2): Case Report # [] Date accepted: 15 March 2014

Plain Film CT. Principal Modality (2): Case Report # [] Date accepted: 15 March 2014 Radiological Category: Musculoskeletal Principal Modality (1): Principal Modality (2): Plain Film CT Case Report # [] Submitted by: Dr. Jason E. Lally, M.D. Faculty reviewer: Dr. Naga Ramesh Chinapuvvula,

More information

Wear 272 (2011) Contents lists available at ScienceDirect. Wear. jou rnal h om epage:

Wear 272 (2011) Contents lists available at ScienceDirect. Wear. jou rnal h om epage: Wear 272 (2011) 79 87 Contents lists available at ScienceDirect Wear jou rnal h om epage: www.elsevier.com/locate/wear Volumetric wear assessment of failed metal-on-metal hip resurfacing prostheses J.K.

More information

MEDICAL POLICY MEDICAL POLICY DETAILS POLICY STATEMENT. Page: 1 of 7

MEDICAL POLICY MEDICAL POLICY DETAILS POLICY STATEMENT. Page: 1 of 7 Page: 1 of 7 MEDICAL POLICY MEDICAL POLICY DETAILS Medical Policy Title METAL-ON-METAL TOTAL HIP RESURFACING Policy Number 7.01.74 Category Technology Assessment Effective Date 06/15/06 Revised Date 07/19/07,

More information

Hip dislocation after total hip arthroplasty can be classified

Hip dislocation after total hip arthroplasty can be classified Case Report Late Anterior Hip Dislocation after Metasul Metal-on-Metal Total Hip Arthroplasty for 13 Years: A Case Report Larbpaiboonpong V email : virojlarb@gmail.com Viroj Larbpaiboonpong 1 Suthorn Bavonratanavech

More information

Sex differences in the morphological failure patterns following hip resurfacing arthroplasty

Sex differences in the morphological failure patterns following hip resurfacing arthroplasty Sex differences in the morphological failure patterns following hip resurfacing arthroplasty Hinsch et al. Hinsch et al. BMC Medicine 2011, 9:113 (13 October 2011) RESEARCH ARTICLE Open Access Sex differences

More information

Thickness effects on maximum von-mises stress of a cement mantle in total hip replacement - a finite element study

Thickness effects on maximum von-mises stress of a cement mantle in total hip replacement - a finite element study Journal of Applied Biomaterials & Biomechanics 2009; Vol. 7 no. 2, pp. 111-115 Società Italiana Biomateriali Thickness effects on maximum von-mises stress of a cement mantle in total hip replacement -

More information

Assessment of radiolucent lines around the Oxford unicompartmental knee replacement

Assessment of radiolucent lines around the Oxford unicompartmental knee replacement KNEE Assessment of radiolucent lines around the Oxford unicompartmental knee replacement SENSITIVITY AND SPECIFICITY FOR LOOSENING S. Kalra, T. O. Smith, B. Berko, N. P. Walton From Norfolk and Norwich

More information

Histopathological characterization of corrosion product associated adverse local tissue reaction in hip implants: a study of 285 cases

Histopathological characterization of corrosion product associated adverse local tissue reaction in hip implants: a study of 285 cases Ricciardi et al. BMC Clinical Pathology (2016) 16:3 DOI 10.1186/s12907-016-0025-9 RESEARCH ARTICLE Open Access Histopathological characterization of corrosion product associated adverse local tissue reaction

More information

Cementless Oxford unicompartmental knee replacement shows reduced radiolucency at one year

Cementless Oxford unicompartmental knee replacement shows reduced radiolucency at one year Cementless Oxford unicompartmental knee replacement shows reduced radiolucency at one year H. Pandit, C. Jenkins, D. J. Beard, J. Gallagher, A. J. Price, C. A. F. Dodd, J. W. Goodfellow, D. W. Murray From

More information

2.79J/3.96J/BE.441/HST522J BIOMATERIALS FOR JOINT REPLACEMENT

2.79J/3.96J/BE.441/HST522J BIOMATERIALS FOR JOINT REPLACEMENT Massachusetts Institute of Technology Harvard Medical School Brigham and Women s/massachusetts General Hosp. VA Boston Healthcare System 2.79J/3.96J/BE.441/HST522J BIOMATERIALS FOR JOINT REPLACEMENT M.

More information

Metasul LDH Large Diameter Head

Metasul LDH Large Diameter Head Metasul LDH Large Diameter Head Surgical Technique Metasul LDH Large Diameter Head Surgical Technique Enhancing Stability and Increasing Range of Motion Metasul LDH Large Diameter Head Surgical Technique

More information

Ceramic-on-metal bearings in total hip replacement

Ceramic-on-metal bearings in total hip replacement Ceramic-on-metal bearings in total hip replacement WHOLE BLOOD METAL ION LEVELS AND ANALYSIS OF RETRIEVED COMPONENTS G. H. Isaac, C. Brockett, A. Breckon, D. van der Jagt, S. Williams, C. Hardaker, J.

More information

New data reinforces the proven safety and effectiveness of the BIRMINGHAM HIP Resurfacing System

New data reinforces the proven safety and effectiveness of the BIRMINGHAM HIP Resurfacing System New data reinforces the proven safety and effectiveness of the BIRMINGHAM HIP Resurfacing System Orthopaedics T 901.396.2121 Smith & Nephew, Inc. F 901.399.5187 1450 Brooks Road www.smith-nephew.com Memphis,

More information

Are There Long-term Benefits to Cementing the Metaphyseal Stem in Hip Resurfacing?

Are There Long-term Benefits to Cementing the Metaphyseal Stem in Hip Resurfacing? Clin Orthop Relat Res (2015) 473:3197 3203 DOI 10.1007/s11999-015-4402-y Clinical Orthopaedics and Related Research A Publication of The Association of Bone and Joint Surgeons CLINICAL RESEARCH Are There

More information

Stability without compromise. Epsilon Durasul Constrained Insert

Stability without compromise. Epsilon Durasul Constrained Insert Stability without compromise Epsilon Durasul Constrained Insert History of Dislocation and Constrained Acetabular Inserts The Need for a Better Solution The Problem Dislocation is the second most common

More information

Metal-on-metal hip surface replacement

Metal-on-metal hip surface replacement HIP TECHNOLOGIES Metal-on-metal hip surface replacement THE ROUTINE USE IS NOT JUSTIFIED M. J. Dunbar, V. Prasad, B. Weerts, G. Richardson From Halifax Infirmary Hospital, Halifax, Nova Scotia, Canada

More information

Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo?

Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo? Journal of Orthopaedic Surgery 2005;13(1):46-51 Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo? H Tsumura, N Kaku, T Torisu Department of Orthopedic Surgery, Oita University,

More information

The predictability of the results of THA has been shown to. Risk Factors Affecting Outcome of Metal-on-Metal Surface Arthroplasty of the Hip SYMPOSIUM

The predictability of the results of THA has been shown to. Risk Factors Affecting Outcome of Metal-on-Metal Surface Arthroplasty of the Hip SYMPOSIUM SYMPOSIUM Risk Factors Affecting Outcome of Metal-on-Metal Surface Arthroplasty of the Hip Paul E. Beaulé, MD, Frederick J. Dorey, PhD, Michel LeDuff, MA, Thomas Gruen, MS, and Harlan C. Amstutz, MD Abstract:

More information

Histopathology: Cell necrosis and cytoplasmic accumulations

Histopathology: Cell necrosis and cytoplasmic accumulations Histopathology: Cell necrosis and cytoplasmic accumulations These presentations are to help you identify basic histopathological features. They do not contain the additional factual information that you

More information

Do the cobalt and chromium released from metal-on-metal hip resurfacing cause cancer?

Do the cobalt and chromium released from metal-on-metal hip resurfacing cause cancer? ARTICLE: ORIGINAL RESEARCH Do the cobalt and chromium released from metal-on-metal hip resurfacing cause cancer? Mark Longman Trainee Intern Dunedin School of Medicine University of Otago ABSTRACT This

More information

SSSR HIP PROSTHESIS MODELS AND MATERIALS SURGICAL TECHNIQUES APPROACHES. Patrick Zingg Department of Orthopaedics University of Zurich, Balgrist

SSSR HIP PROSTHESIS MODELS AND MATERIALS SURGICAL TECHNIQUES APPROACHES. Patrick Zingg Department of Orthopaedics University of Zurich, Balgrist HIP PROSTHESIS MODELS AND MATERIALS SURGICAL TECHNIQUES APPROACHES Patrick Zingg Department of Orthopaedics University of Zurich, Balgrist Zürich, Switzerland www.balgrist.ch IMPLANT TO BONE INTERFACE

More information

Hip Arthroscopy after Resurfacing Arthroplasty: Surgical Technique, Clinical Examples and Literature Review

Hip Arthroscopy after Resurfacing Arthroplasty: Surgical Technique, Clinical Examples and Literature Review Griffith Research Online https://research-repository.griffith.edu.au Hip Arthroscopy after Resurfacing Arthroplasty: Surgical Technique, Clinical Examples and Literature Review Author Weinrauch, Patrick

More information

University of Groningen. Metal-on-metal total hip arthroplasty Zijlstra, Wierd Pieter

University of Groningen. Metal-on-metal total hip arthroplasty Zijlstra, Wierd Pieter University of Groningen Metal-on-metal total hip arthroplasty Zijlstra, Wierd Pieter IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please

More information

)501( COPYRIGHT 2018 BY THE ARCHIVES OF BONE AND JOINT SURGERY

)501( COPYRIGHT 2018 BY THE ARCHIVES OF BONE AND JOINT SURGERY )501( COPYRIGHT 2018 BY THE ARCHIVES OF BONE AND JOINT SURGERY RESEARCH ARTICLE Total Hip Replacement Revision in a Single Brand Small Cementless Stem Our Experience after the Findings of the National

More information

An Optimum Prosthesis Combination of Low Risk TKR Options in All Five Primary Categories of Design Results in a 60% Reduction in Revision Risk

An Optimum Prosthesis Combination of Low Risk TKR Options in All Five Primary Categories of Design Results in a 60% Reduction in Revision Risk An Optimum Prosthesis Combination of Low Risk TKR Options in All Five Primary Categories of Design Results in a 60% Reduction in Revision Risk Christopher Vertullo Andrea Peng Stephen Graves Peter Lewis

More information

Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces

Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces Timothy L. Tan, BS 1, Brenna C. Moeljadi, BS 2, Edward Ebramzadeh, PhD 3, Patricia Campbell, PhD 3, Sophia Nicole

More information

syngo WARP Metal Artifact Reduction Techniques in Magnetic Resonance Imaging

syngo WARP Metal Artifact Reduction Techniques in Magnetic Resonance Imaging Product News syngo WARP Metal Artifact Reduction Techniques in Magnetic Resonance Imaging Theresa Bachschmidt; Ferdinand Lipps, Ph.D.; Mathias Nittka, Ph.D. Siemens Healthcare, Erlangen, Germany The term

More information

Retrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report

Retrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report Journal of Long-Term Effects of Medical Implants, 18(1): 69 74 (2008) Retrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report W. Siebert, 1 S. Mai, 1 & S. Kurtz 2 1 Orthopädische Klinik

More information

AN ANALYSIS FROM THE NATIONAL JOINT REGISTRY FOR ENGLAND AND WALES

AN ANALYSIS FROM THE NATIONAL JOINT REGISTRY FOR ENGLAND AND WALES G. S. Matharu, A. Judge, H. G. Pandit, D. W. Murray From University of Oxford, Nuffield Orthopaedic Centre, Oxford, United Kingdom G. S. Matharu, BSc (Hons), MRCS, MRes, Clinical Research Fellow in Orthopaedics,

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Clinical consequences of pseudotumors in hip arthroplasty

Clinical consequences of pseudotumors in hip arthroplasty Journal of Research and Practice on the Musculoskeletal System JOURNAL OF RESEARCH AND PRACTICE Review Article Clinical consequences of pseudotumors in hip arthroplasty Georgios I. Karagiannis 4 th Department

More information

DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN USA Tel: +1 (800) Fax: +1 (574)

DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN USA Tel: +1 (800) Fax: +1 (574) References 1. Sanchez-Sotelo J, Haidukewych GJ, Boberg CJ. Hospital Cost of Dislocation After Primary Total Hip Arthroplasty. J Bone Joint Surg. 2006;88A:290-294. 2. Brodner W, Grübl A, Jankovsky R, Meisinger

More information