1 CET Gonioscopy Part 2 The first part of the two-part article reviewed the indications and contraindications to performing gonioscopy, the optics of the procedure, and the types of goniolens. In this second part, Dr Michael E Johnson adopts a practical how to approach to the technique and discusses the recording of observations. Common variations of normal anterior chamber angles and angle abnormalities that are seen in primary eye care will also be described. Module C14116, one specialist CET point for AS, SP and IP optometrists, one general CET point for optometrists Becoming competent in gonioscopy requires the ability to identify angle structures that even in normal eyes demonstrate considerable variability. It also demands an appreciation of mirrored imagery and good dexterity. In my opinion it is the most challenging ocular diagnostic technique. Difficulty in reliably obtaining a good view and uncertainty as to what is being seen is normal for clinicians learning gonioscopy. There are no short-cuts to gaining proficiency and so when starting it is necessary to practice gonioscopy with as many eyes as possible, rather than reserving the technique for just those eyes in which it is especially important. Contact lens patients are often good for improving familiarity with the technique because these individuals are used to having something in the eye, and no-fluid goniolenses can even be applied to the surface of a soft lens without the instillation of an anaesthetic. Images are better than any textual description and so I recommend viewing the excellent video clips at If possible it is also extremely helpful to arrange to observe at a local glaucoma clinic. Preparation Instruction Patients need to be told why it is necessary to put a lens on their eye. As a suggestion of what to say: I want to look in the corners of the eye where its internal fluid drains. This is because problems with the drainage of fluid from the eye can cause glaucoma. The drain of the eye is hidden from view and to see it I need to put a lens on to the tear film, on the eye s surface. Patients should be reassured that the eye will be numbed and so the procedure will not be painful, although they may have an awareness of gentle pressure or coldness. Some clinicians suggest that warming the goniolens improves Figure 1 Note the mirrored image is not laterally reversed tolerance. If a Goldmann-type lens is to be used then patients should also be warned that during the procedure coupling solution may dribble onto the face, and that this is not harmful. Set up A topical anaesthetic is instilled before gonioscopy. Alternatively, with no-fluid goniolenses it is possible to not anaesthetise the eye by the prior insertion of a soft contact goniolens. Anaesthetic instillation does not need to be repeated if it has already been done for applanation tonometry, although in patients with twitchy eyelids or where a longer than normal examination is anticipated I prefer to instil a second drop. One of the keys to successful gonioscopy is that the clinician and patient be in a correct and comfortable position. Some clinicians find it helpful to support the elbow to enable steady holding of the goniolens on the eye. Technique Goldmann-type lens specifics The goniolens should be positioned with its concave surface upwards to receive a viscous coupling solution, such as carboxymethylcellulose (eg Celluvisc) or carbomer gel (eg Viscotears). The lower the viscosity of the coupling solution the greater the chance of the goniolens lifting from the eye during the examination and so introducing air bubbles that disturb viewing, particularly when turning the lens on the eye. However, advantages of lower viscosity coupling solutions are that they are less likely to interfere with subsequent viewing of the fundus or leave the patient with sticky eyes and blurred vision. It is useful to store the bottle of the coupling solution upside-down to avoid air bubbles. With very viscous solutions it may also be of value to start a stream of fluid on a tissue before transferring the flow to the lens. If air bubbles do gather beneath the goniolens when on the eye that obstruct viewing the only remedy is to remove the lens, wipe away the coupling solution and start again. To apply the goniolens the patient should be positioned at the slit lamp and asked to look upwards. While looking around the side of the biomicroscope and holding the lens (left hand for right eye and vice versa), corneal side up and tilted at about 45 towards the globe, the little finger of the hand holding the goniolens is used to retract the lower eyelid of the patient. Then the lower edge of the lens engages and fully retracts the lid downward. Finally, the goniolens is touched onto the eye and rapidly tilted into an upright position as the patient is asked to slowly look straight ahead. This should expel any trapped air upwards. When viewing the angle, most clinicians apply the lens so that the mirror is at the top of the eye. This is to allow the inferior angle to be examined first. The inferior portion of the angle is typically the widest and is where the trabecular meshwork has the most pigment, and so is the easiest to identify structures and become familiar with the 20 Optician
2 Continuing education CET appearance of a patient s anatomy. After the inferior angle has been assessed, the goniolens is rotated to view another portion of the angle. When the angle has been examined in its entirety the goniolens is sometimes suckered on the eye and resists removal. The air-tight seal between the goniolens and the eye can usually be broken by asking the patient to blink hard. If this is not successful, the lens can be slid towards the sclera across the change in surface curvatures. Rarely it is necessary to apply pressure through the edge of the lens to indent the globe to break suction. Zeiss-type lens specifics Coupling solution is generally not needed with these flatter goniolenses. However, a small drop of artificial tears to the concavity of the lens improves optical contiguity in patients with dry eye or marked corneal irregularity, and may minimise epithelial trauma. To apply the goniolens the patient is positioned at the slit lamp and asked to look straight ahead. The handle of the goniolens, or the lens itself, is held between thumb and forefinger, with the remaining three fingers braced against the patient s face (left hand for right eye and vice versa). I personally use the index finger to support the upper eyelid against the upper bony rim of the orbit in all but the most cooperative of patients. While looking around the side of the biomicroscope the goniolens is applied directly onto the centre of the cornea. A delicate touch is needed to apply the correct level of pressure with the lens. Too much pressure causes Descemet s membrane to wrinkle, which prevents a quality view of the angle and, of more concern, forces aqueous into the recesses of the angle and artifactually deepens the angle. Too little pressure allows air bubbles to gather beneath the goniolens, which prevents viewing of the angle. The optimal level of pressure has very narrow tolerance limits, and so air should constantly be seen darting in and out under the goniolens when viewing through the slit lamp. The mirrors of the goniolens are initially put on in a square position, rather than a diamond configuration, because it is more comfortable if the patient blinks. The inferior angle is usually examined first using the top mirror. Other portions of the angle are then examined, but without the rotation required for most Goldmann-type goniolenses. When all mirrors have been inspected, if it is considered necessary to view the small portions of the angle not visible when the goniolens is held squarely, the goniolens can be minimally turned on Figure 2 The corneal wedge the eye, or, to allow the patient to blink and reduce epithelial damage, removed and applied again. Mirrored imagery The portion of the angle viewed is that opposite the mirror. A mirrored image of the angle is seen, which differs to an inverted image as is formed with indirect ophthalmoscopy. For example, when viewing the superior angle in the bottom mirror of the goniolens the image is upside-down but not laterally reversed (Figure 1). Relating what is seen in the mirrors of a goniolens to a location in the angle can be confusing due to mirrored imagery. A helpful strategy when unsure is to turn the goniolens so that the lesion is seen in the centre of the mirror, because at this point the location of the lesion in the angle is exactly opposite. Viewing methods Gonioscopy is a dynamic technique. The illumination characteristics are varied and the goniolens is manipulated to help appreciate subtle findings. Regarding lighting, the room lights can be on or dimmed, and the brightness, size and orientation of the slit beam can be adjusted. The attention of the clinician may be focused on the area directly illuminated or just off to the side. When assessing if an angle is open or closed, and if open the likelihood of closure, the testing conditions are critical. The room lights are dimmed, the illumination and the height of the slit beam are decreased so that it does not impinge on the pupil and cause pupillary constriction with attendant artifactual opening of the angle. In this regard, there is a conflict with ease of viewing. Having assessed the occludability of the angle, the next task is to ascertain other features of interest in the angle: pigment, pseudoexfoliative (PXF) material, anterior synechiae, recession, neovascularisation, etc. For this purpose it is permissible, and often helpful, to increase the room and slit lamp illumination. Indeed, the view of angle structures is improved when the pupil is constricted, particularly in narrow angles. Some clinicians like to rotate the slit beam so that it is always perpendicular to the angle, whereas others feel that this additional manipulation is not necessary. A specific technique that is valuable in identifying structures in eyes with very little or excessive trabecular pigmentation, or other causes of confusing anatomy, is the corneal wedge. To set up the corneal wedge the slit beam is narrowed and the illumination system is displaced to create an optical section of the cornea (Figure 2). When correctly focused, the corneal light reflex from the inner aspect of the cornea will be sharp, and that from the outer corneal surface will be slightly hazy. The latter curves as it approaches the angle to meet the reflex from the inner surface, meeting at Schwalbe s line, where after they continue as one line down the angle structures and onto the iris. By pointing to Schwalbe s line the corneal wedge locates the top of the trabecular meshwork. Viewing of the corneal wedge is challenging. Its identification is assisted by turning room lights down and ensuring adequate brightness of the slit beam. The beam must be narrow, and the observation and viewing systems of the slit lamp must be offset, but not so much that the two no longer point down the barrel of the goniolens. The corneal wedge is easier to see with Goldmann-type goniolenses because in these the mirror has a relatively low angle of inclination and so gives a better cross-sectional view of the cornea, which in turn can be followed down to locate the wedge. With Zeiss-type goniolenses the tilt of the mirror may be reduced by sliding the goniolens in the direction of the mirror being used. It should be noted that the corneal wedge never appears as obvious as it does in non-photographic representations in textbooks. Owing to the design of slit lamps, it is usually only possible to view the corneal wedge in the superior and inferior angle because decoupling of the illumination and observation system is only possible horizontally. However, this is normally sufficient to familiarise oneself with the appearance of the angle structures in an individual. The examination of narrow angles is often difficult because a convex iris obstructs the view of the angle from a mirror located close to the opposite limbus. Owing to their hidden recesses, these angles are susceptible to being misdiagnosed as closed. Disparity or Optician 21
3 CET Continuing education parallax of a slit beam on the inner wall of the eye and the iris, using the set up described above for a corneal wedge, indicate that the depths of the angle are not being seen. The view in these situations can be improved by asking the patient to look towards the mirror being used (Figure 3a). Only a small change in viewing direction is needed. It is my experience that, regardless of the completeness of instruction, most patients move their eyes too much without a visual target, often causing the goniolens to eject. To improve control I find it helpful to ask the patient to follow my finger of the non-goniolens holding hand and slowly move it from the straight ahead position until the desired view of the angle through the slit lamp is obtained. This procedure alters the tilt of the mirror relative to the microscope, such that the line-ofsight of the slit lamp viewing system now strikes higher up the mirror and has a steeper angle of approach into the angle. An alternative technique that does not require the patient to alter their direction of gaze is to slide the mirror of the goniolens centrally, towards the angle being examined, and thus away from the mirror being used (Figure 3b). Note that this is the opposite direction to that recommended to identify the corneal wedge. This second technique is quicker because it avoids an explanation to obtain patient cooperation, but is only possible with goniolenses that have a relatively small area of contact and so allow for mobility on the eye. An inability to see the trabecular meshwork or the apex of the corneal wedge when the angle is viewed optimally indicates angle-closure. Increasing the light level, which induces miosis and centripetal movement of iris tissue, or applying pressure to the eye through the goniolens, can be used to deliberately try and open a closed portion of the angle. The latter technique is known as indentation gonioscopy and is more dependable, but can only be performed with relatively flat, small contact diameter goniolenses. If structures previously hidden become visible with these methods the angleclosure was appositional (touching) rather than synechial (adherent). This indicates that an iridotomy or cataract extraction would likely be successful in opening up the angle, whereas, in synechial closure the angle would remain closed. Indentation gonioscopy can also be helpful in differentiating relative pupil block from the more rare plateau iris. In relative pupil block the iris indents smoothly, whereas in plateau iris assumes a sine wave Figure 3 Moving gaze direction influences the angle view (a) (b) configuration as it indents maximally centrally and drapes peripherally over anteriorly located ciliary processes. Angle grading systems There are numerous systems for grading the anterior chamber angle, each with their virtues and proponents. It is, however, important to appreciate the limitations of reducing complex anatomical relationships to a single figure. It should not be assumed that angles with the same grade are equivalent, or that they have an equal probability of closure. The most commonly employed grading systems in use today are based on those developed by Shaffer and Spaeth. The Shaffer method simply estimates the angle that the iris makes with the trabecular meshwork and posterior corneal surface in four quadrants. Practically, the angle width is usually determined by the ease at which the various angle structures can be seen, or conversely, by the structures obscured by the iris (Table 1). The Spaeth system expands on this method and formally assesses other relevant angle characteristics. It sequentially records the level at which the iris apparently inserts (A-E), the geometric angle of iris contact in degrees, the peripheral iris contour (r, s, or q), and the degree of trabecular pigmentation (TMP 1 minimal to 4 very heavy) (Table 2). For example, a Spaeth grading of B 20.r.TMP-2 refers to an iris that inserts just behind Schwalbe s line at an angle of 20, with a regular contour and moderate trabecular pigmentation. Such a grading would be typical for a hypermetropic eye with a narrow angle. Angle anatomy The structures visible in a wide open anterior chamber angle from posterior to anterior are the iris, ciliary body, scleral spur, trabecular meshwork, Schwalbe s line, and cornea. Iris When performing gonioscopy, the contour of the iris plane and its surface should be assessed. A slight convexity of the iris is common. This occurs due to relative pupil block and tends to be greater in hypermetropic eyes. This forward bowing is grossly exaggerated when pupil block is absolute, caused either by primary or secondary mechanisms. Conversely, flat or minimally concave irides may occur in myopia, pseudophakia, and aphakia. Pronounced concavity is often encountered in pigment dispersion syndrome (PDS), where in anatomically predisposed eyes, reverse pupillary block is created when aqueous humour is forced from the posterior chamber into the anterior chamber by the action of blinking. After surveying the general iris contour, attention should be directed to its angle of approach and the position at which it appears to insert into the ciliary body or other angle structures. A narrow anterior chamber angle is more at risk of angle-closure than one that is widely open. Normally, the angle is narrowest superiorly and widest inferiorly, possibly due to a combination of eyelid forces and gravity. As a general rule, anterior chamber depth, iris convexity, and anterior chamber angle width are correlated. However, exceptions do occur, exemplified by iris plateau configuration/syndrome, which is characterised by a normal 24 Optician
4 Continuing education CET anterior chamber depth and an iris that remains relatively flat until its periphery, whereupon it sharply turns, draping over anteriorly displaced ciliary processes to create a narrow angle recess. Blood vessels on the surface of the iris are rarely seen in brown irides, but are relatively common in fair iris colours. Normal iris blood vessels tend to have a relatively thick calibre, and a radial or circular orientation. In contrast, abnormal new blood vessels are finer, lacy, tortuous, and have a random orientation. It is extremely rare to see normal blood vessels crossing the scleral spur to encroach on the trabecular meshwork. Early angle neovascularisation may only be observable as a reddish hue. As neovascular disease progresses, the fibrovascular membrane contracts and leads to the development of posterior anterior synechiae (PAS). The fine, wandering abnormal blood vessels in Fuch s iridocyclitis only rarely cause PAS. Ciliary body The ciliary body is most commonly seen as a dull-brown band, although it may appear slate-grey or pink in lighter eyes. Its visibility and width depends on the position of iris insertion. An unusually wide area of ciliary body exposure, particularly when it is irregular or asymmetric, indicates angle-recession. This occurs due to a tear between the longitudinal and circular muscle fibres of the ciliary body following blunt trauma. Although it does not directly cause glaucoma, angle-recession is a marker for concomitant trabecular damage and, by a degree proportional to its circumferential extent, is a significant risk factor for elevated IOP and glaucoma, although this may not occur until years after the original trauma. Table 1 Shaffer grading system for angle width Shaffer Grade Angle of approach Most posterior structure that can be visualised Scleral spur The scleral spur is an internal projection of scleral tissue, identified as a white opaque line that yellows with age. This structure is difficult to see when the angle is narrow, but when visible is of considerable value as a landmark in gonioscopy. Strands of superficial iris tissue often reflect in the angle to extend over the ciliary body to insert into the scleral spur. In light-coloured eyes these iris processes are pale and subtle, and are often missed with diffuse illumination. In brown eyes they are darker and stand out Interpretation IV > 35 Broad band of ciliary body Incapable of closure III 20 < Angle 35 Scleral spur Incapable of closure Often narrow band of ciliary body II 10 < Angle 20 Posterior trabeculum Closure possible I 0 < Angle 10 Schwalbe s line High risk of closure Possibly anterior trabeculum 0 0 No structures visible. Apex of corneal wedge cannot be visualised Closed Table 2 Spaeth classification of iris insertion and iris contour Iris insertion A Anterior to the trabecular meshwork B Behind Schwalbe s line C At the scleral spur D Deeply ciliary body visible E Extremely deeply Peripheral iris contour s Steep and convex r Regular flat to mildly convex q Queer concave against the scleral spur. Iris processes are benign and inconsequential, unless they extend over the trabecular meshwork to connect to a posterior embryotoxon where they constitute Axenfeld s anomaly. Iris processes need to be differentiated from PAS, which tend to have a broad base and extend beyond the scleral spur, and functionally in that they impede aqueous outflow. PAS with fine points of adhesion can occur following laser trabeculoplasty if the laser is directed too far posteriorly, its energy is too high, or the angle is very narrow Optician 25
5 CET Continuing education Trabecular meshwork The trabecular meshwork extends anteriorly to Schwalbe s line. Unlike other structures in the angle, scrutiny at high magnification reveals this tissue to have texture and depth. In youth the trabecular meshwork has a translucent blue-grey appearance, but with increasing age it darkens in colour. There is little variation in the degree of pigmentation with race and iris colour, but it is increased in PDS and PXF syndrome, and secondary to inflammation, trauma, ocular surgery, iris cysts and melanomas. Of these, PDS is notable in that its dense pigmentation of the trabecular meshwork is relatively uniform in its distribution around the angle s circumference, whereas in other aetiologies it is concentrated inferiorly. Keratic precipitates are occasionally seen on the trabecular meshwork in cases of trabeculitis, most commonly related to Herpes infections or glaucomatocyclitic crisis. Internal to the trabecular meshwork is the canal of Schlemm, which can only be seen when filled with blood. This does not usually happen because IOP prevents reflux from the venous system. However, reversal of the usual pressure gradient may occur in the setting of ocular hypotony or raised episcleral venous pressure, from either disease or excessive pressure with the edge of a goniolens. It is important that this sign is not confused with early angle neovascularisation, which can have a similar appearance when its blood vessels are too fine to discern individually and only visible as a reddish flush. Schwalbe s line Schwalbe s line consists of a condensation of connective tissue fibres that represents the termination of Descemet s membrane and marks the transition from transparent cornea to opaque scleral tissue. On gonioscopy it is seen as a fine protuberance, encircling the anterior chamber. Its smooth surface may glisten somewhat in comparison with the adjacent trabeculum, and display a smattering of pigment, most prominent in the inferior angle. When its location is uncertain the corneal wedge can help identify this structure. An unusually prominent Schwalbe s line in known as a posterior embryotoxon. Owing to an anterior displacement it is often visible with a normal slit lamp examination as a thin white line at the limbus, typically most prominent temporally. A posterior embryotoxon as an isolated finding is benign, but is linked with glaucoma when accompanied by attached strands of peripheral iris, where it represents Axenfeld-Rieger syndrome. In this anterior segment dysgenesis the prominence of Schwalbe s line is often marked and it may even be detached from the wall of the eye and hang, suspended, in the anterior chamber. Occasionally pigment extends anterior to Schwalbe s line, patchy and scalloped in appearance, termed a Sampaolesi s line. This may occur from previous episodes of angleclosure, or any condition that results in pigment release, but most commonly in association with PXF syndrome. Goniolens disinfection The Medical Devices Agency advise that wherever practicable a device that comes into contact with the ocular surface should not be used on more than one patient, as to do so may expose patients to unnecessary risk through the transmission of disease. 1 This is not practicable with goniolenses and so they should be cleaned and disinfected. In this regard, an interesting development is that Haag-Streit have developed a single-use disposable cap (Stery Cup) for some of its lenses. Guidance on the disinfection of ophthalmic devices has been published by the Department Health s Advisory Committee on Dangerous Pathogens (ACDP). This advice has been strongly influenced by the theoretical risk of transmitting prions from one patient to another, although there are no known cases of transmission of variant Creutzfeldt-Jakob Disease (vcjd) by ophthalmic devices. Until recently the recommended method for decontaminating ophthalmic devices was soaking in sodium chlorite solution providing 20,000ppm of available chlorine (2 per cent solution) for 1 hour. This is was more aggressive than disinfection procedures recommended by the leading goniolens manufacturers (Haag-Streit, Ocular-Instruments and Volk), which for sodium hypochlorite suggests either a 0.5 per cent solution for 25 minutes or a 1 per cent solution for 10 minutes. My experience is that Goldmann-type goniolenses with mirrors enclosed in a plastic casing can withstand the more aggressive cleaning regime, but some Zeiss-type goniolenses that have mirrors protected externally by only a thin coating cannot and quickly develop mirror peeling. As an aside, alcohol wipes are even more corrosive, and may fix prions to the surface of a goniolens, and so should not be used. Advice from the College of Optometrists on the matter is that clinicians should to use their discretion 26 Optician
6 Continuing education CET on whether the clinical benefits of an instrument outweigh the theoretical risks of prion transmission. Recently the advice for the decontamination of ophthalmic devices by the ACDP ophthalmology sub-group has been updated. It now allows for a lower concentration of sodium hypochlorite of 10,000ppm of available chlorine, and immersion for the shorter time of 10 minutes. 2 This is combined with prior cleaning with soap/detergent and a series of rinses with water for irrigation (not tap water), and then finally storing the device dry. Soap/ detergent is recommended because sodium hypochlorite is not effective against spores and cysts of certain microorganisms. Multiple rinsing is recommended to avoid the accidental damage to subsequent patients with caustic sodium hypochlorite. It is recommended that tap water be avoided to avoid the risk of Acanthamoeba spp. Puzzlingly, this new guidance appears to be inconsistent with another even more recent publication by the ACDP, which continues to state that sodium hypochlorite is considered to be effective at reducing infectivity but only at concentrations of 20,000ppm of available chlorine and when used for at least one hour. 3 Summary Gonioscopy is a critical part of the examination of patients with suspected glaucoma and with established disease, and on all individuals identified as having possibly closable angles with screening techniques. In addition to the diagnosis and treatment of glaucoma, gonioscopy is often necessary in the diagnosis and management of ocular Multiple-choice questions take part at Which of the following is not an 1 advantage of Zeiss-type goniolenses? A No coupling fluid B Minimal rotation on eye to view all of angle C Can perform indentation gonioscopy D Resistant to ejection from blinking What does the corneal wedge help 2 identify? A Schwalbe s line B Scleral spur C Ciliary body D Trabecular pigmentation Which of the following is true about an 3 angle graded as B 10.r.TMP-1? A The angle has almost no risk of closure B The trabecular meshwork is heavily pigmented C The iris inserts at the level of the scleral spur D The iris has a slightly convex contour trauma, inflammation and neoplasia. Besides its functional importance at the business end of aqueous dynamics, no doubt in part due to difficulty in its visualisation, the angle of the anterior chamber is an immensely rewarding and sometimes beautiful part of ocular anatomy to study. References 1 Medical Devices Agency. MDA AN1994 (04). Single patient use of ophthalmic medical devices: Implications for clinical practice, October Where is the angle normally widest? 4 A Inferior B Superior C Nasal D Temporal Where is the angle normally most 5 pigmented? A Inferior B Superior C Nasal D Temporal In what circumstances can the canal of 6 Schlemm be identified with gonioscopy? A Pigment dispersion syndrome B Pseudoexfoliation syndrome C Raised episcleral pressure D Raised IOP Successful participation in this module counts as one credit towards the GOC CET scheme administered by Vantage and one towards the Association of Optometrists Ireland s scheme. The deadline for responses is July Guidance from the ACDP TSE Working Group: Annex L Managing CJDvCJD risk in ophthalmology. TSEguidance/index.htm, September Guidance from the ACDP TSE Working Group: Annex C - General principles of decontamination and waste disposal. www. dh.gov.uk/ab/acdp/tseguidance/index.ht, November Dr Michael Johnson is an optometrist at Bristol Eye Hospital and a research fellow at Bristol University. He has no commercial interest in any product mentioned Optician 27
5/18/2014. Fundamentals of Gonioscopy Workshop Aaron McNulty, OD, FAAO Walt Whitley, OD, MBA, FAAO
1 Fundamentals of Gonioscopy Workshop Aaron McNulty, OD, FAAO Walt Whitley, OD, MBA, FAAO 2 3 4 5 6 Optometry s Meeting 2014 The Most Valuable Glaucoma Tool Glaucoma Diagnosis Gonioscopy Central corneal
More informationSystems for Anterior Chamber Angle Evaluation 長庚紀念醫院青光眼科吳秀琛
Systems for Anterior Chamber Angle Evaluation 長庚紀念醫院青光眼科吳秀琛 Clinical Techniques for Assessing Angle Width A light from the side showing physiological iris bombe Slit lamp-grading of peripheral AC depth
More informationAre traditional assessments a waste of time? NZAO 2015
Are traditional assessments a waste of time? NZAO 2015 Disclosures No financial interests other than Optometry Practice owner Full time optometrist Not a glaucoma prescriber ODOB Board Chair Previously
More information_ Assessment of the anterior chamber. Review of anatomy of the angle
Assessment of the anterior chamber Dr Simon Barnard PhD BSc FCOptom FAAO DCLP Department of Optometry & Visual Science City University London, UK Review of anatomy of the angle Figure 1. Anatomical section
More informationGonioscopy and Slit Lamp Exam for the Glaucoma Suspect. Disclosure GONIOSCOPY: Gonioscopy Why?? What should I look for? GONIOSCOPY
Gonioscopy and Slit Lamp Exam for the Glaucoma Suspect Disclosure Michael Chaglasian has the following disclosures:» 1. Advisory Board: Alcon, Allergan, Bausch+Lomb, Carl Zeiss Meditec, Merck, Sucampo»
More informationWritten by Administrator Wednesday, 13 January :27 - Last Updated Thursday, 21 January :34
angle closure glaucoma A type of glaucoma caused by a sudden and severe rise in eye pressure. Occurs when the pupil enlarges too much or too quickly, and the outer edge of the iris blocks the eye s drainage
More informationAdvanced Examination of the Retina: Scleral Indentation & Retinal 3-Mirror
Advanced Examination of the Retina: Scleral Indentation & Retinal 3-Mirror Meredith Whiteside, OD, FAAO Nimesh Patel, OD, FAAO John Shan, OD, FAAO Please silence all mobile devices. Unauthorized recording
More information03/04/2015. LOC Talk Anterior Chamber & Gonioscopy 1st April Methods of Assessing Anterior Chamber Depth (and angle width) Outline
LOC Talk Anterior & 1st April 2015 Mr. Areeb Moosavi MBBS BSc FRCOphth Glaucoma Consultant Milton Keynes University Hospital NHS Foundation Trust Methods of Assessing Anterior Open Versus Closed angle
More informationANTERIOR SEGMENT EXAMINATION TECHNIQUES
ANTERIOR SEGMENT EXAMINATION TECHNIQUES GERS October 2017 Amanda Harding - Principal Optometrist, MREH MSc., MCOptom, Dip Glauc., Dip TP (IP). Anterior Segment Examination Depth of anterior chamber Central
More informationDownloaded from:
Philippin, H; Shah, P; Burton, M (2012) The next step: Detailed assessment of an adult glaucoma patient. Community eye health / International Centre for Eye Health, 25 (79-80). pp. 50-53. ISSN 0953-6833
More information3/16/2018. Ultrasound Biomicroscopy in Glaucoma By Ahmed Salah Abdel Rehim. Prof. of Ophthalmology Al-Azhar University
Ultrasound Biomicroscopy in Glaucoma By Ahmed Salah Abdel Rehim Prof. of Ophthalmology Al-Azhar University 1 Ultrasound biomicroscopy (UBM) is a recent technique to visualize anterior segment with the
More informationRelationship between limbal incisions. angle. and the structures of the anterior chamber
Brit. _7. Ophthal. (I 973) 57, 722 Relationship between limbal incisions and the structures of the anterior chamber angle MOHAMED I. AYOUB AND AHMED H. SAID Department of Ophthalmology, Faculty of Medicine,
More informationPrimary Angle Closure Glaucoma
www.eyesurgeonlondon.co.uk Primary Angle Closure Glaucoma What is Glaucoma? Glaucoma is a condition in which there is damage to the optic nerve. This nerve carries visual signals from the eye to the brain.
More informationUnderstanding Angle Closure
Case Understanding Angle Closure Dominick L. Opitz, OD, FAAO Associate Professor Illinois College of Optometry 56 year old Caucasian Male Primary Eye Exam BCVA: 20/25 OD with+1.25 DS 20/25 OS with +1.75
More informationGLAUCOMA. An Overview
GLAUCOMA An Overview Compiled by Campbell M Gold (2004) CMG Archives http://campbellmgold.com --()-- IMPORTANT The health information contained herein is not meant as a substitute for advice from your
More informationGonioscopy and 3-Mirror Retinal Evaluation Workshop Edeline Lu, O.D., FAAO Benedicte Gonzalez, O.D., MPH, FAAO Tina Zheng, O.D.
Gonioscopy and 3-Mirror Retinal Evaluation Workshop Edeline Lu, O.D., FAAO Benedicte Gonzalez, O.D., MPH, FAAO Tina Zheng, O.D., FAAO Please silence all mobile devices and remove items from chairs so others
More informationTHE CHRONIC GLAUCOMAS
THE CHRONIC GLAUCOMAS WHAT IS GLAUCOMA People with glaucoma have lost some of their field of all round vision. It is often the edge or periphery that is lost. That is why the condition can be missed until
More informationTHE CHRONIC GLAUCOMAS
THE CHRONIC GLAUCOMAS WHAT IS GLAUCOMA? People with glaucoma have lost some of their field of all round vision. It is often the edge or periphery that is lost. That is why the condition can be missed until
More informationClosed Angle Glaucoma Or Narrow Angle Glaucoma. What s is a closed angle type of glaucoma,
Closed Angle Glaucoma Or Narrow Angle Glaucoma What s is a closed angle type of glaucoma, This is where the iris is found to be blocking the drainage of the eye through the trabecular meshwork. The eye
More informationYear 1 MBChB Clinical Skills Session Ophthalmoscopy
Year 1 MBChB Clinical Skills Session Ophthalmoscopy Reviewed & ratified by: Dr V Taylor-Jones, Mr M Batterbury Consultant Ophthalmologist Learning objectives o To understand the anatomy and physiology
More informationLEUKAEMIA*t INFILTRATION OF THE IRIS IN CHRONIC LYMPHATIC. pattemn * Received for pubiication November io, i967.
Brit. J. Ophthal. (1968) 52, 781 INFILTRATION OF THE IRIS IN CHRONIC LYMPHATIC LEUKAEMIA*t BY BRIAN MARTIN The General Infirmary, Leeds OCULAR involvement is common in the leukaemias though the anterior
More informationMeet Libby. Corneal Dysgenesis, Degeneration, and Dystrophies Definitions. Dr. Victor Malinovsky
Meet Libby Corneal Dysgenesis, Degeneration, and Dystrophies 2006 Dr. Victor Malinovsky Definitions Dysgenesis: (congenital anomalies) A development disorder that results in a congenital malformation of
More informationnon-perforating injury
Brit. J. Ophthal. (I 972) 56, 418 Anterior chamber angle tears after non-perforating injury DAVID MOONEY Croydon Eye Unit, Croydon, Surrey Recession of the anterior chamber angle is a common finding after
More informationSecondary open-angle glaucoma
Secondary open-angle glaucoma Kathy Hondeghem ZNA Middelheim MaNaMa 12/10/13 Definition Open anterior chamber angle (at least 270 ) Trabecular meshwork (and thus aqueous humor outflow) is occluded by a
More informationDifferential diagnosis of the red eye. Carol Slight Nurse Practitioner Ophthalmology
Differential diagnosis of the red eye Carol Slight Nurse Practitioner Ophthalmology The red eye Conjunctivitis HSV Keratitis Acute angle closure glaucoma Anterior Uveitis Red eye Scleritis Subconjunctival
More informationYear 2 MBChB Clinical Skills Session Ophthalmoscopy. Reviewed & ratified by: Mr M Batterbury Consultant Ophthalmologist
Year 2 MBChB Clinical Skills Session Ophthalmoscopy Reviewed & ratified by: o Mr M Batterbury Consultant Ophthalmologist Learning objectives o To understand the anatomy and physiology of the external and
More informationFundamentals of Gonioscopy
Fundamentals of Gonioscopy Denise Goodwin, OD, FAAO Introduction: Evaluating the anterior chamber angle with gonioscopy is an essential part of evaluating patients at risk for glaucoma and neovascularization.
More information02/03/2014. Average Length: 23mm (Infant ~16mm) Approximately the size of a quarter Volume: ~5mL
Identify the anatomy of the eye. Explain the basic physiology of the parts of the eye. Briefly discuss various surgeries related to different parts of the anatomy. Average Length: 23mm (Infant ~16mm) Approximately
More informationCorporate Medical Policy
Corporate Medical Policy Optical Coherence Tomography (OCT) Anterior Segment of the Eye File Name: Origination: Last CAP Review: Next CAP Review: Last Review: optical_coherence_tomography_(oct)_anterior_segment_of_the_eye
More informationTraining Checking Vision Tonometry
Training 101-2 Checking Vision Tonometry Checking Vision The classic example of an eye chart is the Snellen eye chart, in general they show 11 rows of capital letters. The top row contains one letter (usually
More informationAround The Globe in 60 Minutes
Around The Globe in 60 Minutes Around the GLOBE in Sixty Minutes Basic Ocular Anatomy, Examination, and Diagnostic Techniques Introduction Focusing on canine and feline ocular anatomy and basic examination
More informationChronicity. Narrow Minded. Course Outline. Acute angle closure. Subacute angle closure. Classification of Angle Closure 5/19/2014
Chronicity Narrow Minded The management of narrow angles in the optometric practice Acute Subacute Chronic Aaron McNulty, OD, FAAO Course Outline Classification of Angle Closure Evaluation of narrow angles
More informationFACING YOUR FUNDIC FEARS: EXAMINATION OF THE OCULAR FUNDUS J. Seth Eaton, VMD, DACVO Cornell University Veterinary Specialists
FACING YOUR FUNDIC FEARS: EXAMINATION OF THE OCULAR FUNDUS J. Seth Eaton, VMD, DACVO Cornell University Veterinary Specialists The goal of a thorough fundus examination is to clinically evaluate the structures
More informationMaximizing your Retinal Exam with Slit Lamp Fundus Lenses Caroline B. Pate, OD, FAAO and Elizabeth A. Steele, OD, FAAO
Maximizing your Retinal Exam with Slit Lamp Fundus Lenses (2 Hour Workshop) Caroline B. Pate, OD, FAAO and Elizabeth The University of Alabama at Birmingham School of Optometry 1716 University Boulevard
More informationThe Anterior Segment & Glaucoma Visual Recognition & Interpretation of Clinical Signs
The Anterior Segment & Glaucoma Visual Recognition & Interpretation of Clinical Signs Quiz created by Jane Macnaughton MCOptom & Peter Chapman BSc MCOptom FBDO CET Accreditation C19095 2 CET Points (General)
More informationIntraoperative Gonioscopy: A Key to Angle Surgery
Intraoperative Gonioscopy: A Key to Angle Surgery Shakeel Shareef, MD Associate Professor Flaum Eye Institute Univ. of Rochester School of Med. Rochester, NY Phaco vs. MIGS PHACO MIGS Viewing Full corneal
More information4/2/2016. Bond, Brent "We Should Be Doing More. Hpw?" We Should Be Doing More Gonioscopy and Iridoplasty: GONIOSCOPY GONIOSCOPY
We Should Be Doing More Gonioscopy and Iridoplasty: Why and How Brent Bond, MD Associate Professor Wake Forest University Department of Ophthalmology We Should Be Doing More Gonioscopy and Iridoplasty:
More informationAnterior segment imaging
Article Date: 11/1/2016 Anterior segment imaging AS OCT vs. UBM vs. endoscope; case based approaches BY BENJAMIN BERT, MD, FACS AND BRIAN FRANCIS, MD, MS Currently, numerous imaging modalities are available
More informationFrequently Asked Questions about General Ophthalmology:
1. Normal Eye Structure The eye is a slightly asymmetrical globe, about an inch in diameter. The parts of the eye include: Cornea (a clear dome over the iris), Iris (the pigmented part); Pupil (the black
More informationAngle recession and secondary glaucoma
Brit. j. Ophthal. (I973) 57, 6o8 Angle recession and secondary glaucoma D. MOONEY Kent County Ophthalmic and Aural Hospital, Maidstone In a previous article (Mooney, I972), the incidence and extent of
More informationGlaucoma Glaucoma is a complication which has only recently been confirmed as a feature of
1.2.4 OPHTHALMOLOGICAL ABNORMALITIES Ocular abnormalities are well documented in patients with NPS 6 62 81 95. 1.2.4.1 Glaucoma Glaucoma is a complication which has only recently been confirmed as a feature
More informationEntropion uveae: Early sphincter atrophy, signposting primary angle closure glaucoma?
European Journal of Ophthalmology / Vol. 14 no. 4, 2004 / pp. 290-297 Entropion uveae: Early sphincter atrophy, signposting primary angle closure glaucoma? R. SIHOTA, R. SAXENA, H.C. AGARWAL Glaucoma Service,
More informationAnatomy: There are 6 muscles that move your eye.
Thyroid Eye Disease Your doctor thinks you have thyroid orbitopathy. This is an autoimmune condition where your body's immune system is producing factors that stimulate enlargement of the muscles that
More informationPatient & Family Guide. Glaucoma Management.
Patient & Family Guide Glaucoma Management 2018 www.nshealth.ca Glaucoma Management What is glaucoma? Your eyes are filled with a fluid made by the ciliary body (1). This fluid flows through the pupil
More informationA LITTLE ANATOMY. three layers of eye: 1. outer: corneosclera. 2. middle - uvea. anterior - iris,ciliary body. posterior - choroid
GLAUCOMA A LITTLE ANATOMY three layers of eye: 1. outer: corneosclera 2. middle - uvea anterior - iris,ciliary body posterior - choroid connection at the pars plana between post and ant uvea 3. retina
More information1/31/2018. Course Objectives. Diagnostic Testing. Optic Nerve Damage ANATOMY AND PHYSIOLOGY OF A GLAUCOMA WORK-UP/TONOMETRY TECHNICIAN: -SDP
ANATOMY AND PHYSIOLOGY OF A GLAUCOMA WORK-UP/TONOMETRY KNOW THE DISEASE PROCESS TECHNICIAN: EXPLAIN PROCESS OF EXAMINATION Presenters: Dana McMahan, COA Nicole Smith, COA Engage with patient s, help alleviate
More information4/22/16. Eye. External Anatomy of Eye. Accessory Structures. Bio 40B Dr. Kandula
Eye Bio 40B Dr. Kandula External Anatomy of Eye Accessory Structures l Eyebrows l Levator Palpebrae Superioris - opens eye l Eyelashes l Ciliary glands modified sweat glands l Small sebaceous glands l
More informationUBM and glaucoma: diagnosis and follow-up of plateau iris. M. Puech
ophtalmologiques UBM and glaucoma: diagnosis and follow-up of plateau iris M. Puech Extrait du mensuel Réalités Ophtalmologiques N 204 Juin 2013 Cahier 1 Editeur : Performances Médicales 91, avenue de
More informationSurgical management of Duane's
Brit. J. Ophthal. (I974) 58, 30 I Surgical management of Duane's syndrome M. H. GOBIN ljniversity Eye Clinic, Leyden, IHolland Ten years ago I introduced a surgical technique for the correction of Duane's
More informationICO-Ophthalmology Surgical Competence Assessment Rubric Vitrectomy (ICO-OSCAR:VIT)
ICO-Ophthalmology Surgical Competence Assessment Rubric Vitrectomy (ICO-OSCAR:VIT) Date Resident Evaluator Novice (score = 2) Beginner (score = 3) Advanced Beginner (score = 4) Competent (score = 5) Not
More informationGoals. Glaucoma PARA PEARL TO DO. Vision Loss with Glaucoma
Glaucoma Janet R. Fett, OD Drs. Kincaid, Fett and Tharp So Sioux City, NE eyewear21@hotmail.com Goals Understand Glaucoma Disease process Understand how your data (objective and subjective) assists in
More informationMIXED GLAUCOMA* glaucoma may be the first episode. DURING the past 20 years it has become widely accepted that primary
Brit. J. Ophthal. (1961) 45, 503. MIXED GLAUCOMA* BY J. D. ABRAMS London DURING the past 20 years it has become widely accepted that primary glaucoma consists of two distinct conditions: chronic simple
More informationEars. Mouth. Jowls 6 Major Bones of the Face Nasal bone Two
1 2 3 4 5 Chapter 25 Injuries to the Face, Neck, and Eyes Injuries to the Face and Neck Face and neck are to injury Relatively unprotected positions on body Some injuries are life-threatening. trauma to
More informationSTAB INCISION GLAUCOMA SURGERY (SIGS)
STAB INCISION GLAUCOMA SURGERY (SIGS) Dr. Soosan Jacob, MS, FRCS, DNB Senior Consultant Ophthalmologist, Dr. Agarwal's Eye Hospital, Chennai, India dr_soosanj@hotmail.com Videos available in Youtube channel:
More informationAssisting in Ophthalmology. Copyright 2011, 2007, 2003, 1999 by Saunders, an imprint of Elsevier Inc. All rights reserved.
Assisting in Ophthalmology Learning Objectives Define, spell, and pronounce the terms listed in the vocabulary. Apply critical thinking skills in performing patient assessment and care. Explain the differences
More informationANGLE VESSELS IN NORMAL EYES
Brit. J. Ophthal. (1964) 48, 551. ANGLE VESSELS IN NORMAL EYES A GONIOSCOPIC EVALUATION AND ANATOMIC. CORRELATION*t BY PAUL HENKIND From the Department of Ophthalmology, New York University College of
More informationManagement of Angle Closure Glaucoma Hospital Authority Convention 18 May 2015
Management of Angle Closure Glaucoma Hospital Authority Convention 18 May 2015 Jimmy Lai Clinical Professor Department of Ophthalmology The University of Hong Kong 1 Primary Angle Closure Glaucoma PACG
More informationGlaucoma. How is Glaucoma Diagnosed? Glaucoma Testing
Glaucoma How is Glaucoma Diagnosed? Glaucoma Testing There is no single test for glaucoma. The diagnosis is made by evaluating the patient from a number of perspectives, using specialized instruments.
More informationPedsCases Podcast Scripts
PedsCases Podcast Scripts This is a text version of a podcast from Pedscases.com on Approach to Childhood Glaucoma. These podcasts are designed to give medical students an overview of key topics in pediatrics.
More informationThe Sense Organs 10/13/2016. The Human Eye. 1. Sclera 2. Choroid 3. Retina. The eye is made up of three layers:
The human body gathers information from the outside world by using the five senses of: The Sense Organs 12.3 Sight Hearing Taste Smell Touch This information is essential in helping the body maintain homeostasis.
More informationA Guide to Administering
A Guide to Administering INDICATIONS AND USAGE YUTIQ (fluocinolone acetonide intravitreal implant) 0.18 mg is indicated for the treatment of chronic non-infectious uveitis affecting the posterior segment
More informationNate Lighthizer, O.D., F.A.A.O. Assistant Professor, NSUOCO Assistant Dean, Clinical Care Services Director of CE Chief of Specialty Care Clinics
Nate Lighthizer, O.D., F.A.A.O. Assistant Professor, NSUOCO Assistant Dean, Clinical Care Services Director of CE Chief of Specialty Care Clinics Chief of Electrodiagnostics Clinic lighthiz@nsuok.edu YAG
More informationGLAUCOMA SUMMARY BENCHMARKS FOR PREFERRED PRACTICE PATTERN GUIDELINES
SUMMARY BENCHMARKS FOR PREFERRED PRACTICE PATTERN GUIDELINES Introduction These are summary benchmarks for the Academy s Preferred Practice Pattern (PPP) guidelines. The Preferred Practice Pattern series
More informationAn Injector s Guide to OZURDEX (dexamethasone intravitreal implant) 0.7 mg
An Injector s Guide to OZURDEX (dexamethasone intravitreal implant) 0.7 mg This guide is intended to provide injectors with information on the recommended injection technique and the important risks related
More informationAnterior Chamber Angle: Assessment and Anomalies
Anterior Chamber Angle: Assessment and Anomalies Dave Hicks, OD, FAAO Please silence all mobile devices and remove items from chairs so others can sit. Unauthorized recording of this session is prohibited.
More informationGlaucoma. Glaucoma. Optic Disc Cupping
Glaucoma What is Glaucoma? Bruce James A group of diseases in which damage to the optic nerve occurs as a result of intraocualar pressure being above the physiological norm for that eye Stoke Mandeville
More informationrelative s privacy, do not identify your relative by full name in any assignment.
Overview Do you or a family member have glaucoma? Do you wonder what this diagnosis means? Glaucoma affects tens of millions of people worldwide. Despite its prevalence, many people lack accurate information
More informationBy Darlene Jones, Nurse. May 2017
By Darlene Jones, Nurse May 2017 Disclosure of potential conflict of interest Darlene Jones, Nurse I have no conflict of interest Course objectives Become familiar with the different pathologies in ophthalmology
More informationGENERAL INFORMATION GLAUCOMA GLAUCOMA
GENERAL INFORMATION GLAUCOMA GLAUCOMA WHAT IS GLAUCOMA? Glaucoma is commonly known as the sneak thief of sight because it can cause irreversible vision loss without any obvious symptoms. The term glaucoma
More informationTrabeculectomy. Draining the aqueous humour reduces the pressure on the optic nerve that causes loss of vision in glaucoma.
Trabeculectomy Other formats If you need this information in another format such as audio tape or computer disk, Braille, large print, high contrast, British Sign Language or translated into another language,
More informationDr Jo-Anne Pon. Dr Sean Every. 8:30-9:25 WS #70: Eye Essentials for GPs 9:35-10:30 WS #80: Eye Essentials for GPs (Repeated)
Dr Sean Every Ophthalmologist Southern Eye Specialists Christchurch Dr Jo-Anne Pon Ophthalmologist Southern Eye Specialists, Christchurch Hospital, Christchurch 8:30-9:25 WS #70: Eye Essentials for GPs
More informationTISSUE* ABERRANT INTRA-OCULAR LACRIMAL GLAND. of the left eye of a full-term infant of 2 weeks; the tumour was globular and
Brit. J. Ophthal. (1960) 44, 619. ABERRANT INTRA-OCULAR LACRIMAL GLAND TISSUE* BY W. S. HUNTERt Department of Pathology, Institute of Ophthalmology, University oflondon THE occurrence of aberrant lacrimal
More informationStep 4: Ask permission to turn off lights or draw the curtains
STEPS OF EYE EXAMINATION - FUNDUS Step 1: Approach the patient Read the instructions carefully for clues Shake hands, introduce yourself Ask permission to examine him I would like to examine your eyes,
More informationAcknowledgement goes to WHO for use of their hand washing technique
How to administer third party insulin injections safely with a pen device, safety pen needle and syringe 1 Hand washing It takes at least 15 seconds to wash your hands. This is about how long it takes
More informationAnterior segment imaging
CET CONTINUING Sponsored by 1 CET POINT Anterior segment imaging Sundeep Vaswani, BSc (Hons), MCOptom 39 The anterior segment of the eye encompasses all structures from the front surface of the cornea
More informationOcular Hypertension. A Guide
Ocular Hypertension A Guide Printed: July 2015 Review: July 2018 This free booklet is brought to you by the International Glaucoma Association (IGA), the charity for people with glaucoma. We haven t charged
More informationSinus trabeculectomy. Preliminary results of IOO operations
Brit. J. Ophthal. (I 972) 56, 833 Sinus trabeculectomy Preliminary results of IOO operations A. P. NESTEROV, N. V. FEDEROVA, AND Y. E. BATMANOV Department of Ophthalmology, Kazan Medical Institute, Kazan,
More informationarthritis "Contact lens" cornea in rheumatoid (opposite). Brit. J. Ophthal. (I970) 54, 410 Peterborough District Hospital
Brit. J. Ophthal. (I970) 54, 410 "Contact lens" cornea in rheumatoid arthritis A. J. LYNE Peterborough District Hospital It has been noted that patients suffering from long-standing rheumatoid arthritis
More informationEye conditions in Samoyeds
Eye conditions in Samoyeds Information for breeders and pet owners Melbourne EyeVet Mulgrave Essendon Bundoora Frankston Geelong Bendigo Wodonga Traralgon Darwin Extra eyelashes/distichiasis Distichia
More informationADVANCED DIAGNOSTIC TECHNIQUES
DIVISION OF VISION SCIENCES SESSION: 2008/2009 DIET: 1ST ADVANCED DIAGNOSTIC TECHNIQUES VISP216 LEVEL:2 MODULE LEADER: DR GUNTER LOFFLER B.Sc/B.Sc. (HONS) OPTOMETRY MAY 2009 DURATION: 2 HOURS CANDIDATES
More informationThis module introduces students to the basic concepts of human anatomy and physiology, and correlating structures and functions.
DIPLOMA IN OPTOMETRY (PT) DOP 101 General Anatomy, Physiology and Pathology This module introduces students to the basic concepts of human anatomy and physiology, and correlating structures and functions.
More informationAcute Eyes for ED. Enis Kocak. The Alfred Ophthalmology
Acute Eyes for ED Enis Kocak The Alfred Ophthalmology The problem with eyes Things to cover Ocular anatomy Basic assessment Common presentations Eye first aid and procedures Ophthalmic emergencies What
More informationCataract Surgery: Information for patients. Back of eye. Vitreous. Retina. Lens
Patient information Cataract Surgery: Information for patients Front of eye Cornea Pupil Iris Back of eye Vitreous Retina Lens The anatomy of the eye is illustrated above. Your cataract is a clouding of
More informationophthalmoscopy seeing the eye from greek: ophthalmos - the eye skopeos to see
ophthalmoscopy seeing the eye from greek: ophthalmos - the eye skopeos to see structure principles of fundus examination direct ophthalmoscopy indirect ophthalmoscopy magnification in ophthalmoscopy field
More informationProbe Selection A high frequency (7-12 MHz) linear array transducer should be used to visualize superficial structures (Image 1).
! Teresa S. Wu, MD, FACEP Director, Emergency Ultrasound Program & Fellowships Co-Director, Women s Imaging Fellowship Maricopa Medical Center Associate Professor, Emergency Medicine Director, Simulation
More informationTest Bank for Medical Surgical Nursing An Integrated Approach 3rd Edition by White
Test Bank for Medical Surgical Nursing An Integrated Approach 3rd Edition by White Link full download : http://testbankair.com/download/test-bank-for-medical-surgical-nursing-anintegrated-approach-3rd-edition-by-white/
More information3/23/2016. Phaco vs. Angle Surgery. Car Analogy. 10 Essential Steps of Perioperative Gonioscopy. Intraoperative Gonioscopy: A Key to Angle Surgery
Intraoperative Gonioscopy: A Key to Angle Surgery Shakeel Shareef, MD Wallace Alward, MD Alan Crandall, MD Steven Vold, MD Ike Ahmed, MD No financial conflicts of interest 10 Essential Steps of Perioperative
More informationPollen Slide Mounting Protocol
Pollen Slide Mounting Protocol Materials: Syn-Matrix mounting medium Microcentrifuge Microscope slides Slide coverslips (18mm x 18mm) Coverslip podium (see Figure 1) Capillary tubes Dissecting microscope
More informationPREVALENCE OF GLAUCOMA AMONG FISHERMEN COMMUNITY OF MUNDRA TALUKA OF KUTCH DISTRICT- A CROSS- SECTIONAL STUDY
ORIGINAL RESEARCH PREVALENCE OF GLAUCOMA AMONG FISHERMEN COMMUNITY OF MUNDRA TALUKA OF KUTCH DISTRICT- A CROSS- SECTIONAL STUDY Sanjay Upadhyay 1, Jayantilal Shah 2 1 Assistant Professor, 2 Associate Professor,
More informationistent Implantation Procedure
istent Implantation Procedure Design Anatomy Lumen 120 μm Snorkel Retention Arches 1 mm Rail Self-Trephining Tip istent Indications for Use The istent Trabecular Micro-Bypass Stent Model GTS100R/L is indicated
More informationWORLD GLAUCOMA WEEK What is Glaucoma? Can I develop glaucoma if I have increased eye pressure?
WORLD GLAUCOMA WEEK What is Glaucoma? Glaucoma is a group of diseases that damage the optic nerve and can result in gradual vision loss and blindness. However, with early detection and treatment, you can
More informationHistology of the Eye
Histology of the Eye Objectives By the end of this lecture, the student should be able to describe: The general structure of the eye. The microscopic structure of:»cornea.»retina. EYE BULB Three coats
More informationPENETRATING EYE INJUIRES
PENETRATING EYE INJUIRES King Harold receives a mortal penetrating injury to the eye at the Battle of Hastings 1066, Detail Bayeux Tapestry, Eleventh century. Then Earl William came from Normandy into
More informationOPHTHALMOLOGY AND ULTRASOUND
Vet Times The website for the veterinary profession https://www.vettimes.co.uk OPHTHALMOLOGY AND ULTRASOUND Author : JAMES OLIVER Categories : Vets Date : April 28, 2008 JAMES OLIVER discusses why ultrasound
More informationGlaucoma What You Should Know
Glaucoma What You Should Know U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES National Institutes of Health National Eye Institute The National Eye Institute (NEI) conducts and supports research that leads
More informationINTRODUCTION: ****************************************************************************************************
BIOLOGY 211: HUMAN ANATOMY & PHYSIOLOGY **************************************************************************************************** EYES AND VISION ****************************************************************************************************
More informationScleral buckling. Surgical Treatment
Dr. Ayman M. Khattab MD, FRCS professor of Ophthalmology Cairo University Surgical Treatment Pneumatic retinopexy. Primary pars plana vitrectomy. 1 Indications for scleral buckling. SB is used to treat
More informationSecondary Glaucomas. Mr Nick Strouthidis MBBS MD PhD FRCS FRCOphth FRANZCO Consultant Ophthalmologist, Glaucoma Service, Moorfields Eye Hospital
Secondary Glaucomas Mr Nick Strouthidis MBBS MD PhD FRCS FRCOphth FRANZCO Consultant Ophthalmologist, Glaucoma Service, Moorfields Eye Hospital Introduction: What is glaucoma? Glaucoma is the name given
More informationIntegrating the LenSx Laser Laser into your Clinic: Our advice for those considering it
Integrating the LenSx Laser Laser into your Clinic: Our advice for those considering it (DAY) Facilitator: (Name) LenSx Laser Important Safety Information Caution United States Federal Law restricts this
More information