Rates of Abnormal Retinal Nerve Fiber Layer and Ganglion Cell Layer OCT Scans in Healthy Myopic Eyes: Cirrus Versus RTVue

Size: px
Start display at page:

Download "Rates of Abnormal Retinal Nerve Fiber Layer and Ganglion Cell Layer OCT Scans in Healthy Myopic Eyes: Cirrus Versus RTVue"

Transcription

1 CLINICAL SCIENCE Rates of Abnormal Retinal Nerve Fiber Layer and Ganglion Cell Layer OCT Scans in Healthy Myopic Eyes: Cirrus Versus RTVue Jean-Claude Mwanza, MD, MPH, PhD; Fouad E. Sayyad, MD; Ahmad A. Aref, MD; Donald L. Budenz, MD, MPH n BACKGROUND AND OBJECTIVE: To compare rates of abnormal peripapillary retinal nerve fiber layer (RNFL) and macular ganglion cell layer scans acquired with Cirrus HD-OCT (Carl Zeiss Meditec, Inc., Dublin, CA) and RTVue-100 (Optovue Inc., Fremont, CA) in healthy myopic eyes. n PATIENTS AND METHODS: Forty-one non-glaucomatous myopic eyes (41 individuals) were scanned with Cirrus to measure RNFL and ganglion cell-inner plexiform layer (GCIPL) and with RTVue to measure peripapillary RNFL and ganglion cell complex (GCC) thicknesses. Rates of abnormal scans were calculated and compared between devices. Inter-device agreement in falsely classifying scans as abnormal was also assessed. n RESULTS: The rate of abnormal average and fourquadrant RNFL was 4.8% to 7.3% on Cirrus and 2.4% to 9.7% on RTVue (P >.05). The overall rate of abnormal scans was 19.2% on Cirrus and 29.3% on RTVue (P =.3). Rates of abnormal Cirrus average and segmental GCIPL (12.2% to 17%) were similar to those of RTVue average and segmental GCC (9.7% to 14.6%) (P >.05). The overall rate of abnormal GCIPL (36.6%) was higher than that of GCC (14.6%) (P =.023). The inter-device agreement was poor for average RNFL (k = -0.09), very good for average ganglion cell (k = 0.81), and fair for overall RNFL (k = 0.35) and overall ganglion cell (k = 0.34). n CONCLUSION: The high rates of abnormal RNFL and ganglion cell layer scans on both devices call for caution, particularly when attempting to diagnose glaucoma in myopic eyes using these devices. The RNFL and ganglion cell layer analyses may not be interchangeable on either of these devices. These two devices are not interchangeable for classifying healthy myopic eyes based on RNFL or ganglion cell layer analysis. [Ophthalmic Surg Lasers Imaging 2012;43:S67- S74.] From the Bascom Palmer Eye Institute, University of Miami School of Medicine, Miami, Florida. Drs. Mwanza and Budenz are currently affiliated with the Department of Ophthalmology, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina. Dr. Aref is currently affiliated with the Illinois Eye & Ear Infirmary, University of Illinois at Chicago School of Medicine, Chicago, Illinois. Originally submitted June 9, Accepted for publication October 2, Presented at the Association for Research in Vision and Ophthalmology annual meeting, May 6-10, 2012, Fort Lauderdale, Florida. Supported by an unrestricted grant from Research to Prevent Blindness, Inc., New York, New York, and unrestricted research support from Carl Zeiss Meditec, Inc., Dublin, California. The authors have no financial or proprietary interest in the materials presented herein. Address correspondence to Donald L. Budenz, MD, MPH, Department of Ophthalmology, UNC-Chapel Hill, 130 Mason Farm Road, 5151 Bioinformatics Bldg., CB# 7040, Chapel Hill, NC donald_budenz@med.unc.edu doi: / Ophthalmic Surgery, Lasers & Imaging Vol. 43, No. 6 (Suppl), 2012 S67

2 INTRODUCTION Major population-based studies 1-8 have found a significant association between myopia and glaucoma. A recent review based on a meta-analysis of data from 11 population-based studies concluded that both low and high myopia increase the risk of developing open-angle glaucoma. 9 Millions of individuals are diagnosed as having glaucoma worldwide each year. However, the risk of falsely classifying a normal individual as glaucomatous or a glaucomatous individual as normal may be high, particularly in early stages of the disease. Falsely diagnosing a truly normal individual as having glaucoma, with potential subsequent lifelong treatment, may have tremendous psychological, economic, and physical consequences. It is also well established that glaucomatous cupping is difficult to assess in myopic discs due to shallow cup, thin pre-laminar tissue, oblique implantation of the optic nerve resulting in tilted disc, and areas of peripapillary atrophy In addition, non-glaucomatous myopic individuals may exhibit visual field defects similar to those seen in glaucomatous individuals. 14,15 Following recent advances in ocular imaging, optical coherence tomography (OCT) has developed rapidly to become an important clinical diagnostic and monitoring tool for glaucoma, mainly through quantitative analysis of peripapillary retinal nerve fiber layer (RNFL) thickness. In addition, studies are available to indicate that analysis of ganglion cell complex (GCC) 16,17 and ganglion cell-inner plexiform layer (GCIPL) 18 can successfully discriminate normal individuals and individuals with glaucoma. There is also indication that a substantial proportion of non-glaucomatous myopic eyes have thinner RNFL and GCC, and their thicknesses are falsely classified as outside normal limits on OCT. 13,19-22 The current study was designed to compare the rate of abnormal peripapillary RNFL and macular ganglion cell scans obtained with two different spectral-domain OCT devices in healthy myopic eyes. The study was also designed to determine the agreement between devices and between peripapillary RNFL and macular ganglion cell scans in classifying individuals as abnormal. PATIENTS AND METHODS Patients This prospective study was conducted on nonglaucomatous myopic volunteers aged 18 years and older recruited among University of Miami medical students, ophthalmology and optometry trainees, and employees. After individuals provided written informed consent to participate in the study, they underwent a screening eye examination that included visual acuity, automated refraction, slit-lamp examination, intraocular pressure (IOP) measurement, ophthalmoscopy, and visual field testing using the screening mode of frequency-doubling technology (software version Welch Allyn Humphrey Systems; Carl Zeiss Meditec, Inc., Dublin, CA). Individuals were excluded for the following reasons: best-corrected visual acuity worse than 20/40, contraindication to dilation or intolerance to topical anesthetics; IOP 22 mm Hg or greater, history of raised IOP, or any type of glaucoma in either eye, history of intraocular surgery in the study eye (except cataract or refractive surgery performed more than 1 year prior to enrollment), cup-to-disc area ratio interocular asymmetry 0.2 or greater, evidence or history of non-glaucomatous neuropathy or other optic nerve abnormalities in either eye, or retinal diseases including diabetic retinopathy, macular edema, or other vitreoretinal disease. Abnormal frequency-doubling technology test results, defined as one or more locations identified as abnormal at a P value of.05 or less at the same location(s) on repeated testing, was also an exclusion criterion. OCT Instrumentation and Rates of Abnormal Scans Following the screening eye examination, individuals underwent OCT peripapillary and macular scanning sequentially with Cirrus HD-OCT (Carl Zeiss Meditec, Inc.) and RTVue-100 (Optovue Inc., Fremont, CA). Only one eye randomly selected was scanned per individual. Six scans including three peripapillary scans centered on the optic disc were obtained with Cirrus OCT (Optic Disc axial protocol) and with RTVue (Optic Nerve Head protocol). In addition, the same number of macular scans centered on the fovea was acquired using the Macular Cube cube protocol for Cirrus OCT and GCC protocol for RTVue. Macular scans obtained with Cirrus OCT were processed with the ganglion cell analysis algorithm, 23,24 which was a pre-release version of software version 6.0. Only good quality images, defined by a signal strength of 6 or greater for Cirrus OCT and a signal strength index of 30 or greater for RTVue were used for S68 Copyright SLACK Incorporated

3 TABLE 1 Comparison and Correlation of Average and Hemispheric Ganglion Cell-Inner Plexiform Layer (GCIPL) and Ganglion Cell Complex (GCC) Thickness Parameter GCIPL GCC P Pearson r (P) Average 81.9 (6.1) 94.8 (7.4) < (<.001) Superior hemisphere 82.4 (6.0) 94.7 (7.5) < (<.001) Inferior hemisphere 81.5 (6.4) 94.9 (7.6) < (<.001) analysis. For peripapillary RNFL, the overall average and quadrant thicknesses were used in the analysis. For macular ganglion cell analysis, the overall average and hemispheric (superior and inferior) GCIPL for Cirrus OCT and GCC for RTVue were considered. In addition, the minimum GCIPL was analyzed for Cirrus OCT and compared with focal loss volume measured with RTVue. To determine the frequency of abnormal scans, the average and quadrant significance maps for RNFL and the average and hemispheric maps for GCIPL for each participant were examined. Any yellow or red color on any of the maps was considered abnormal. An individual was considered positive when at least two of the three scans consistently showed a thickness value lower than the given percentile of the instrument intrinsic normative database it was compared to, specifically, less than 5% indicated in yellow or less than 1% level in red. Statistical Analysis All analyses were performed using SPSS version 19.0 (SPSS, Inc., Chicago, IL). The actual percent agreement, without the adjustment for agreement that would have occurred by chance, was calculated as follows using only the actual observations: % agreement = (number of agreements / number of cases compared) 3 100%. The statistical significance of the difference between rates of abnormal scans was calculated using the chi-square test. P values of less than.05 were considered significant. The degree of agreement between instruments in classifying individuals as normal or abnormal was assessed with kappa (k) statistic. The classification proposed by Altman was used to determine the degree of agreement between instruments and between peripapillary and macular scans 25 : poor (k < 0.2), fair (k = 0.21 to 0.40), moderate (k = 0.41 to 0.60), good (k = 0.61 to 0.80), and very good (k = 0.81 to 1.00). RESULTS Demographic and Ocular Characteristics Forty-one non-glaucomatous myopic eyes of 41 individuals were included in the study. Mean age was 31.7 ± 8.3 years (range: 24 to 64 years), axial length was 25.2 ± 1.2 mm (range: 23.1 to 28.4 mm), spherical equivalent was -4.6 ± 1.9 diopters (range: -2.0 to diopters), and IOP was 13.4 ± 2.2 mm Hg (range: 9 to 17 mm Hg). The mean signal strength was 8.3 ± 1.1 (range: 7 to 9) for RNFL scans and 8.8 ± 0.9 (range: 8 to 10) for macular scans on Cirrus OCT; the mean signal strength index was 74.0 ± 8.6 (range: 55.4 to 89.8) for RNFL and 65.3 ± 10.8 (range: 48.5 to 83.9) for macular scans on RTVue. The average and hemispheric ganglion cell thickness for both OCT devices are displayed in Table 1. All GCIPL measurements were significantly thinner than GCC measurements (all P <.001), but correlated well with each other (Pearson r = 0.84 to 0.87, all P <.001). Table 2 shows the proportions of individuals classified as having abnormal RNFL and ganglion cell thickness scans based on each parameter that was analyzed. For RNFL, the rate of abnormal scans was 4.8% on Cirrus OCT versus 9.7% on RTVue for average, and ranged from 4.8% to 7.3% on Cirrus OCT and from 2.4% to 9.7% on RTVue for quadrants. The rate of abnormal scans calculated based on the presence of any abnormal quadrant was comparable on the two instruments. The overall rate based on any abnormal parameter was 19.2% on Cirrus OCT and 29.3% on RTVue. The differences in rates of abnormal RNFL parameters between devices were all not statistically significant (all P >.05). The ganglion cell layer analysis showed that Cirrus OCT average, minimum, and hemispheric GCIPL were abnormal in 12.2% to 17% of the individuals, whereas the rate of abnormal scans ranged from 2.4% to 14.6% for RTVue average, focal loss vol- Ophthalmic Surgery, Lasers & Imaging Vol. 43, No. 6 (Suppl), 2012 S69

4 TABLE 2 Rate of Abnormal Average and Sectoral Measurements of Peripapillary RNFL and Macular Layers and Their Agreement on Cirrus OCT and RTVue Parameter Cirrus OCT RTVue P Kappa RNFL Average 2 (4.8%) 4 (9.7%) Temporal quadrant 2 (4.8%) 2 (2.4%) Superior quadrant 2 (4.8%) 1 (2.4%) Nasal quadrant 3 (7.3%) 5 (9.7%) Inferior quadrant 2 (4.8%) 3 (7.3%) Any quadrant 7 (17.0%) 9 (21.9%) Overall 8 (19.2%) 12 (29.3%) Ganglion cell a Average 6 (14.6%) 6 (14.6%) Superior hemisphere 5 (12.2%) 5 (12.2%) Inferior hemisphere 7 (17.0%) 4 (9.7%) Any hemisphere 7 (17.0%) 6 (14.6%) Minimum/focal loss volume b 5 (12.2%) 1 (2.4%) Overall 15 (36.6%) 6 (14.6%) RNFL = retinal nerve fiber layer; OCT = optical coherence tomography. a Denotes ganglion cell-inner plexiform layer and ganglion cell complex for Cirrus OCT and RTVue, respectively. b Measured with Cirrus OCT and RTVue, respectively. Cirrus OCT is manufactured by Carl Zeiss Meditec, Inc., Dublin, CA, and RTVue is manufactured by Optovue Inc., Fremont, CA. Figure 1. Venn diagrams comparing (A) the number of eyes with overall abnormal retinal nerve fiber layer scans on Cirrus OCT (Carl Zeiss Meditec, Inc., Dublin, CA) and RTVue (Optovue Inc., Fremont, CA) and (B) the number of eyes with overall abnormal ganglion cell-inner plexiform layer scans on Cirrus OCT and abnormal ganglion cell complex scans on RTVue. ume, and hemispheric GCC, with no significant differences compared to Cirrus OCT GCIPL rates (P =.09 to 1). However, the ganglion cell analysis produced a higher overall abnormal rate on Cirrus OCT than on RTVue (P =.023). The actual percent agreement between the two devices (not shown in Table 1), without the adjustment for agreement that would have occurred by chance, in classifying individuals as abnormal was 20% for average, 50% for superior quadrant, 12.5% for nasal quadrant, 0% for temporal and inferior quadrants, and 33.3% for any abnormal parameter based on RNFL analysis. For ganglion cell analysis, these values were 71.4% for average, 66.7% for superior hemisphere, 57.1% for inferior hemisphere, 62.5% for any hemisphere, 20% for minimum GCIPL/GCC focal loss volume, and 31.3% for any abnormal parameter. On both instruments, 33.3% of eyes with abnormal average ganglion cell layer thickness had abnormal average RNFL thickness. Kappa statistic for agreement between instruments was fair (k = 0.34) for any abnormal ganglion cell S70 Copyright SLACK Incorporated

5 Figure 2. Illustrative examples of optical coherence tomography scans acquired in a myopic individual (age: 27 years, spherical equivalent: diopters, intraocular pressure: 18 mm Hg) showing (A) abnormal Cirrus OCT (Carl Zeiss Meditec, Inc., Dublin, CA) peripapillary retinal nerve fiber layer (RNFL) and (B) macular ganglion cell-inner plexiform layer (GCIPL). Scans acquired in the same eye with RTVue-100 (Optovue Inc., Fremont, CA) showed (C) a normal peripapillary RNFL but (D) an abnormal ganglion cell complex (GCC). Panels E through H are scans from a 31-year-old individual (spherical equivalent: diopters, intraocular pressure: 12 mm Hg) showing abnormal RNFL (E and G) and ganglion cell (F and H) on Cirrus OCT and RTVue, respectively. The last set of panels depicts normal RNFL scans on both instruments (I and K), and normal GCIPL scan on Cirrus OCT (J) but abnormal GCC scan on RTVue (L) in a 63-year-old individual (spherical equivalent: diopters, intraocular pressure: 13 mm Hg). parameter (Figure 1A) and for minimum GCIPL/ GCC focal loss volume (k = 0.31), good (k = 0.69 to 0.77) for GCIPL/GCC hemispheres, and very good (k = 0.81) for average thickness. The agreement between instruments on classifying individuals was poor for average RNFL (k = -0.09), fair to moderate for quadrants (k = 0.36 to 0.56), and fair for any abnormal RNFL parameter (k = 0.35) (Figure 1B). Average RNFL and average ganglion cell analyses showed moderate agreement on Cirrus OCT (k = 0.46) and fair agreement on RTVue (k = 0.32). When the analysis was based on any abnormal parameter, the agreement between RNFL and ganglion cell scans was poor on both devices. Examples of agreement between RNFL and ganglion cell scans on both instruments are illustrated in Figure 2. DISCUSSION Technological advances in ocular imaging in recent years have led to the development of OCT, which provides the clinician with information that has been shown to improve the diagnostic performance for early diagnosis and assist in the management of diseases affecting the optic nerve, retina, and choroid. Consequently, OCT has been introduced successfully for routine assessment of glaucoma. In recent evaluations of the glaucoma diagnostic accuracy of OCT, sensitivity values of 78% to 92% for Cirrus OCT and 80% to 88% for RTVue with specificities of 77% to 95% and 90% to 98%, respectively, have been reported However, information about rates of abnormal OCT in healthy myopic eyes and normal OCT in glaucomatous myopic eyes Ophthalmic Surgery, Lasers & Imaging Vol. 43, No. 6 (Suppl), 2012 S71

6 is largely lacking. We chose to focus on the rate of abnormal OCT RNFL and ganglion cell analysis in nonglaucomatous myopic eyes because they have thinner RNFL 13,31-33 and ganglion cell layers 24,32,34 and therefore may fall outside normal limits because current OCT normative databases do not take refractive error into account. Although sensitivity and specificity are the most commonly used measures of test performance, the rate of abnormal OCT results is important to know, particularly in healthy myopic individuals because of the likelihood of them being falsely classified as glaucomatous by clinical examination. 12,35 It is important to note that the current study is not about the subtle discordance that can commonly occur between glaucoma diagnostic devices used for assessing structural changes, which are in this case the thicknesses of the RNFL and ganglion cell layer. Our main concerns are rather substantial inter-device differences and equivocal or abnormal OCT results in glaucoma-free individuals using devices approved by the U.S. Food and Drug Administration. Although it can be argued that some discordance is acceptable in myopic individuals when different OCT devices are used, it is potentially problematic if a substantial proportion of otherwise normal myopic individuals is classified as having abnormal RNFL and ganglion cells scans by these devices, which could result in initiation of treatment inappropriately. Although a few studies are available that have compared RNFL thickness between Cirrus OCT and RTVue, this is the first report to establish a comparison between GCIPL and GCC in both thickness and rates of abnormal scans in myopic individuals. As expected, average and hemispheric GCIPLs were significantly thinner compared with GCC measurements. This finding is understandable because, unlike the GCC algorithm, the Cirrus OCT ganglion cell analysis algorithm leaves the RNFL out to segment and measure only the GCIPL as previously described. 23 The only available study that compared the ability of macular ganglion cells and peripapillary RNFL scans acquired with RTVue for glaucoma diagnosis in highly myopic eyes is by Kim et al. 32 In that study, falsepositive rates were not reported, but the data suggest that they ranged between 33.3% and 45.8% in highly myopic eyes and between 16.3% and 24.5% in nonhighly myopic eyes for average, hemispheric, and focal loss volume of GCC. False-positive rates based on average RNFL thickness were 25% and 10.2% in highly and non-highly myopic eyes, respectively. Direct comparison of these rates with ours is difficult because we did not separate our study population into high and non-high myopic eyes. Our results showed that approximately one in every five myopic individuals had a false-positive RNFL on Cirrus OCT. This rate is far lower than the 52% found on time-domain OCT by Rauscher et al. 13 Vernon et al., 20 Leung et al., 19 and Qiu et al. 22 assessed the proportion of abnormal peripapillary RNFL scans in myopic eyes using time-domain OCT. None of those studies reported the overall proportion of abnormal RNFL, but a careful look at their results tends to suggest abnormal time-domain peripapillary RNFL scans in more than 40%, approximately 45%, and more than 20%, respectively. In addition, more than 20% of the individuals had abnormal RNFL on Cirrus OCT in the study by Qiu et al. 22 Overall, results of the current study and those of previous studies indicate high rates of abnormal RNFL scans in healthy myopic eyes. The lack of difference in rates of abnormal RNFL scans between Cirrus OCT and RTVue may be in line with earlier findings that the two devices have a similar ability to distinguish normal eyes and eyes with glaucoma using peripapillary RNFL, 36 although RTVue yields thicker measurements than Cirrus OCT, as reported in the current study and elsewhere. 37,38 It is important to emphasize that the high rate of abnormal results obtained with both instruments does not necessarily lessen the usefulness of OCT in myopic individuals. In addition, the overall lower rate of abnormal results based on ganglion cell analysis on RTVue compared to Cirrus OCT does not indicate with certainty that measurements obtained with RTVue in myopes are correct or that those obtained with Cirrus OCT are erroneous. Another important aspect to keep in mind is that although a high specificity is desirable because it influences the significance of a finding, a test with high specificity such as RTVue GCC may be prone to missing true glaucomatous change in comparison with a test that has a low specificity, such as Cirrus OCT GCIPL in this study. Overall, the findings call for caution when interpreting OCT results in this group of individuals based on both RNFL and ganglion cell layer thickness measurements. One interesting observation in the current study is the level of agreement between Cirrus OCT and RTVue for classifying individuals as having abnor- S72 Copyright SLACK Incorporated

7 mal scans based on RNFL and ganglion cell analysis, and the agreement between RNFL and ganglion cell analysis on Cirrus OCT and RTVue. Because of the anatomical relationship between retinal ganglion cell bodies and their axons, one would expect a good agreement between RNFL and ganglion cell layer in classifying individuals. If we consider average thicknesses, the agreement was moderate on Cirrus OCT, suggesting that to some extent the performance of RNFL reflected that of GCIPL. In contrast, agreement between GCC and RNFL parameters in classification was poor with RTVue. It is difficult to explain this discrepancy, but it is possible that combining RNFL and GCIPL the way that is done with RTVue, and assuming that the thickness of this complex represents just the ganglion cell layer, may account for the observed lack of agreement. Another factor that may have contributed to low agreement between macular ganglion cell and peripapillary RNFL analysis is that the former is only based on a sampling containing approximately 50% of the retinal ganglion cell population, whereas the latter results from all retinal ganglion cell axons as axons emanating from different parts of the retina converge onto the optic disc. Interestingly, the two instruments agreed well in classifying individuals based on most parameters of macular ganglion cell analysis. This agreement likely reflects the similarity in diagnostic efficacy of ganglion cell analysis performed with these two devices, although further studies are needed to confirm or refute this assumption. The reason for this similarity may be that the RNFL in the macular region is thin and should not have a significant impact on the ganglion cell layer thickness measurement and diagnostic accuracy despite the difference in the shape and size of macular regions used to determine the thickness of GCIPL and GCC. 16,23,39 In contrast, there was a lack of agreement based on RNFL analysis, which is probably related to differences in segmentation algorithms. We are not aware of any other study examining the agreement between Cirrus OCT and RTVue or other spectral-domain OCT devices in classifying RNFL and ganglion cell scans as positive based on RNFL or ganglion cell analysis. Abnormal peripapillary RNFL and macular ganglion cell layer scans occur frequently on Cirrus OCT and RTVue in healthy myopic eyes. From a practical standpoint, clinicians should rely more on clinical judgment than on current imaging techniques. The RNFL and ganglion cell layer analyses may not be interchangeable on either of the two devices. Similarly, these two devices are not interchangeable for classifying healthy myopic eyes based on RNFL or ganglion cell layer analysis. REFERENCES 1. Wu SY, Nemesure B, Leske MC. Refractive errors in a black adult population: the Barbados Eye Study. Invest Ophthalmol Vis Sci. 1999;40: Mitchell P, Hourihan F, Sandbach J, et al. The relationship between glaucoma and myopia: the Blue Mountains Eye Study. Ophthalmology. 1999;106: Suzuki Y, Iwase A, Araie M, et al. Risk factors for open-angle glaucoma in a Japanese population: the Tajimi Study. Ophthalmology. 2006;113: Perera SA, Wong TY, Tay WT, et al. Refractive error, axial dimensions, and primary open-angle glaucoma: the Singapore Malay Eye Study. Arch Ophthalmol. 2010;128: Czudowska MA, Ramdas WD, Wolfs RC, et al. Incidence of glaucomatous visual field loss: a ten-year follow-up from the Rotterdam Study. Ophthalmology. 2010;117: Wong TY, Klein BE, Klein R, et al. Refractive errors, intraocular pressure, and glaucoma in a white population. Ophthalmology. 2003;110: Casson RJ, Gupta A, Newland HS, et al. Risk factors for primary open-angle glaucoma in a Burmese population: the Meiktila Eye Study. Clin Experiment Ophthalmol. 2007;35: Sia DI, Edussuriya K, Sennanayake S, et al. Prevalence of and risk factors for primary open-angle glaucoma in central Sri Lanka: the Kandy Eye Study. Ophthalmic Epidemiol. 2010;17: Marcus MW, de Vries MM, Montolio FG, et al. Myopia as a risk factor for open-angle glaucoma: a systematic review and meta-analysis. Ophthalmology. 2011;118: Greve EL, Furuno F. Myopia and glaucoma. Graefes Arch Clin Exp Ophthalmol. 1980;213: Jonas JB, Gusek GC, Naumann GO. Optic disk morphometry in high myopia. Graefes Arch Clin Exp Ophthalmol. 1988;226: Tay E, Seah SK, Chan SP, et al. Optic disk ovality as an index of tilt and its relationship to myopia and perimetry. Am J Ophthalmol. 2005;139: Rauscher FM, Sekhon N, Feuer WJ, et al. Myopia affects retinal nerve fiber layer measurements as determined by optical coherence tomography. J Glaucoma. 2009;18: Doshi A, Kreidl KO, Lombardi L, et al. Nonprogressive glaucomatous cupping and visual field abnormalities in young Chinese males. Ophthalmology. 2007;114: Kumar RS, Baskaran M, Singh K, et al. Clinical Characterization of young Chinese myopes with optic nerve and visual field changes resembling glaucoma. J Glaucoma. In press. 16. Tan O, Chopra V, Lu AT, et al. Detection of macular ganglion cell loss in glaucoma by Fourier-domain optical coherence tomography. Ophthalmology. 2009;116: Wang M, Hood DC, Cho JS, et al. Measurement of local retinal ganglion cell layer thickness in patients with glaucoma using frequency-domain optical coherence tomography. Arch Ophthalmol. 2009;127: Mwanza JC, Durbin MK, Budenz DL, et al. Glaucoma diagnostic accuracy of ganglion cell-inner plexiform layer thickness: comparison with nerve fiber layer and optic nerve head. Ophthalmology. In press. 19. Leung CK, Mohamed S, Leung KS, et al. Retinal nerve fiber layer measurements in myopia: an optical coherence tomography study. Ophthalmic Surgery, Lasers & Imaging Vol. 43, No. 6 (Suppl), 2012 S73

8 Invest Ophthalmol Vis Sci. 2006;47: Vernon SA, Rotchford AP, Negi A, et al. Peripapillary retinal nerve fibre layer thickness in highly myopic Caucasians as measured by Stratus optical coherence tomography. Br J Ophthalmol. 2008;92: Kim NR, Lim H, Kim JH, et al. Factors associated with false positives in retinal nerve fiber layer color codes from spectral-domain optical coherence tomography. Ophthalmology. 2011;118: Qiu KL, Zhang MZ, Leung CK, et al. Diagnostic classification of retinal nerve fiber layer measurement in myopic eyes: a comparison between time-domain and spectral-domain optical coherence tomography. Am J Ophthalmol. 2011;152: Mwanza JC, Oakley JD, Budenz DL, et al. Macular ganglion cell-inner plexiform layer: automated detection and thickness reproducibility with spectral domain-optical coherence tomography in glaucoma. Invest Ophthalmol Vis Sci. 2011;52: Mwanza JC, Durbin MK, Budenz DL, et al. Profile and predictors of normal ganglion cell-inner plexiform layer thickness measured with frequency-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2011;52: Altman DG. Practical Statistics for Medical Research. Boca Raton, FL: Chapman and Hall; 1997: Bengtsson B, Andersson S, Heijl A. Performance of time-domain and spectral-domain optical coherence tomography for glaucoma screening. Acta Ophthalmol. 2012;90: Chang RT, Knight OJ, Feuer WJ, et al. Sensitivity and specificity of time-domain versus spectral-domain optical coherence tomography in diagnosing early to moderate glaucoma. Ophthalmology. 2009;116: Mwanza JC, Oakley JD, Budenz DL, et al. Ability of Cirrus HD- OCT optic nerve head parameters to discriminate normal from glaucomatous eyes. Ophthalmology. 2011;118: Rao HL, Zangwill LM, Weinreb RN, et al. Comparison of different spectral domain optical coherence tomography scanning areas for glaucoma diagnosis. Ophthalmology. 2010;117: Sehi M, Grewal DS, Sheets CW, et al. Diagnostic ability of Fourierdomain vs time-domain optical coherence tomography for glaucoma detection. Am J Ophthalmol. 2009;148: Budenz DL, Anderson DR, Varma R, et al. Determinants of normal retinal nerve fiber layer thickness measured by Stratus OCT. Ophthalmology. 2007;114: Kim NR, Lee ES, Seong GJ, et al. Comparing the ganglion cell complex and retinal nerve fibre layer measurements by Fourier domain OCT to detect glaucoma in high myopia. Br J Ophthalmol. 2011;95: Schweitzer KD, Ehmann D, Garcia R. Nerve fibre layer changes in highly myopic eyes by optical coherence tomography. Can J Ophthalmol. 2009;44: Kim NR, Kim JH, Lee J, et al. Determinants of perimacular inner retinal layer thickness in normal eyes measured by Fourier-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2011;52: Loyo-Berrios NI, Nlustein JN. Primary open-angle glaucoma and myopia: a narrative review. The Wisconsin Medical Journal. 2007;106: Leite MT, Rao HL, Zangwill LM, et al. Comparison of the diagnostic accuracies of the Spectralis, Cirrus, and RTVue optical coherence tomography devices in glaucoma. Ophthalmology. 2011;118: Leite MT, Rao HL, Weinreb RN, et al. Agreement among spectraldomain optical coherence tomography instruments for assessing retinal nerve fiber layer thickness. Am J Ophthalmol. 2011;151: Savini G, Carbonelli M, Barboni P. Retinal nerve fiber layer thickness measurement by Fourier-domain optical coherence tomography: a comparison between cirrus-hd OCT and RTVue in healthy eyes. J Glaucoma. 2010;19: Bagci AM, Shadidi M, Ansari R, et al. Thickness profiles of retinal layers by optical coherence tomography image segmentation. Am J Ophthalmol. 2008;146: S74 Copyright SLACK Incorporated

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Original Article Philippine Journal of OPHTHALMOLOGY Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Dennis L. del Rosario, MD and Mario M. Yatco, MD University of Santo

More information

Assessment of Retinal Nerve Fiber Layer Changes by Cirrus High-definition Optical Coherence Tomography in Myopia

Assessment of Retinal Nerve Fiber Layer Changes by Cirrus High-definition Optical Coherence Tomography in Myopia Divya Singh et al Original REASEARCH 10.5005/jp-journals-10028-1223 Assessment of Retinal Nerve Fiber Layer Changes by Cirrus High-definition Optical Coherence Tomography in Myopia 1 Divya Singh, 2 Sanjay

More information

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma Med. J. Cairo Univ., Vol. 83, No. 2, September: 67-72, 2015 www.medicaljournalofcairouniversity.net Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

More information

OPTOMETRY RESEARCH PAPER

OPTOMETRY RESEARCH PAPER C L I N I C A L A N D E X P E R I M E N T A L OPTOMETRY RESEARCH PAPER Interocular symmetry of retinal nerve fibre layer thickness in healthy eyes: a spectral-domain optical coherence tomographic study

More information

Available online at Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6):

Available online at   Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6): Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(6):201-206 ISSN: 0976-8610 CODEN (USA): AASRFC Comparison of glaucoma diagnostic ability of retinal nerve

More information

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Original article Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Dewang Angmo, 1 Shibal Bhartiya, 1 Sanjay K Mishra,

More information

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD Original Article Philippine Journal of OPHTHALMOLOGY Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 1: Categorical Outcomes Based on a Normative Database

More information

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography American Journal of Engineering Research (AJER) e-issn : 2320-0847 p-issn : 2320-0936 Volume-02, Issue-11, pp-104-110 www.ajer.org Research Paper Open Access To assess the glaucoma diagnostic ability of

More information

Advances in OCT Murray Fingeret, OD

Advances in OCT Murray Fingeret, OD Disclosures Advances in OCT Murray Fingeret, OD Consultant Alcon, Allergan, Bausch & Lomb, Carl Zeiss Meditec, Diopsys, Heidelberg Engineering, Reichert, Topcon Currently Approved OCT Devices OCT Devices

More information

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography Original articles in Indian eyes with optical coherence tomography Malik A, Singh M, Arya SK, Sood S, Ichhpujani P Department of Ophthalmology Government Medical College and Hospital, Sector 32, Chandigarh,

More information

International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December ISSN

International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December ISSN International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December-2013 108 Name of Chief and Corresponding Author : Dr Chandrima Paul TITLE : Comparison of glaucoma diagnostic ability

More information

Ganglion cell complex scan in the early prediction of glaucoma

Ganglion cell complex scan in the early prediction of glaucoma Original article in the early prediction of glaucoma Ganekal S Nayana Super Specialty Eye Hospital and Research Center, Davangere, Karnataka, India Abstract Objective: To compare the macular ganglion cell

More information

Glaucoma Diagnosis. Definition of Glaucoma. Diagnosing Glaucoma. Vision Institute Annual Fall Conference

Glaucoma Diagnosis. Definition of Glaucoma. Diagnosing Glaucoma. Vision Institute Annual Fall Conference Glaucoma Diagnosis Vision Institute Annual Fall Conference Mitchell W. Dul, OD, MS, FAAO mdul@sunyopt.edu Richard J. Madonna, MA, OD, FAAO rmadonna@sunyopt.edu Definition of Glaucoma Glaucoma can be regarded

More information

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma.

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma. Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma. Donald C. Hood 1,2,* and Ali S. Raza 1 1 Department of Psychology, Columbia

More information

A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT) Retinal Nerve Fiber Layer Measurements Based on Time Domain OCT Measurements

A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT) Retinal Nerve Fiber Layer Measurements Based on Time Domain OCT Measurements pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(5):369-377 http://dx.doi.org/10.3341/kjo.2012.26.5.369 Original Article A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT)

More information

Research Article Repeatability of Perimacular Ganglion Cell Complex Analysis with Spectral-Domain Optical Coherence Tomography

Research Article Repeatability of Perimacular Ganglion Cell Complex Analysis with Spectral-Domain Optical Coherence Tomography Ophthalmology Volume 2015, Article ID 605940, 5 pages http://dx.doi.org/10.1155/2015/605940 Research Article Repeatability of Perimacular Ganglion Cell Complex Analysis with Spectral-Domain Optical Coherence

More information

Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis OCT

Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis OCT pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2011;25(3):166-173 DOI: 10.3341/kjo.2011.25.3.166 Original Article Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis

More information

Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC)

Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC) Dada T et al Original article Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC) Dada T1, Aggarwal A1, Bali SJ2, Sharma A1, Shah

More information

Jong Chul Han, Da Ye Choi, and Changwon Kee. 1. Introduction

Jong Chul Han, Da Ye Choi, and Changwon Kee. 1. Introduction Ophthalmology Volume 2015, Article ID 641204, 7 pages http://dx.doi.org/10.1155/2015/641204 Clinical Study The Different Characteristics of Cirrus Optical Coherence Tomography between Superior Segmental

More information

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma Module 3a GDx Educational Slide Deck Carl Zeiss Meditec, Inc. November 2005 1 Structure & Function Modules Module

More information

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods Ophthalmology Volume 2011, Article ID 914250, 5 pages doi:10.1155/2011/914250 Clinical Study Macular Retinal Ganglion Cell Complex Thickness and Its Relationship to the Optic Nerve Head Topography in Glaucomatous

More information

OtticaFisiopatologica

OtticaFisiopatologica Anno quindicesimo dicembre 2010 How to assess the retinal nerve fiber layer thickness Antonio Ferreras Miguel Servet University Hospital, Zaragoza. Aragón Health Sciences Institute University of Zaragoza

More information

CLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography

CLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography CLINICAL SCIENCES Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography Seong Bae Park, MD; Kyung Rim Sung, MD, PhD; Sung Yong Kang, MD; Kyung Ri Kim, BS;

More information

Evaluation of myopia on retinal nerve fiber layer thickness measured by Spectralis optical coherence tomography

Evaluation of myopia on retinal nerve fiber layer thickness measured by Spectralis optical coherence tomography 2716 Evaluation of myopia on retinal nerve fiber layer thickness measured by Spectralis optical coherence tomography YI ZHA, JINFEI ZHUANG, DA LIN, WANGQIANG FENG, HAIHUA ZHENG and JIANQIU CAI Ophthalmology

More information

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects Romanian Journal of Ophthalmology, Volume 60, Issue 3, July-September 2016. pp:158-164 GENERAL ARTICLE Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013

EXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013 268 Comparison of optic nerve morphology in eyes with glaucoma and eyes with non-arteritic anterior ischemic optic neuropathy by Fourier domain optical coherence tomography YUXIN YANG 1, HAITAO ZHANG 1,

More information

RETINAL NERVE FIBER LAYER

RETINAL NERVE FIBER LAYER CLINICAL SCIENCES The Effect of Scan Diameter on Retinal Nerve Fiber Layer Thickness Measurement Using Stratus Optic Coherence Tomography Giacomo Savini, MD; Piero Barboni, MD; Michele Carbonelli, MD;

More information

Mark Dunbar: Disclosure

Mark Dunbar: Disclosure Important Things to Understand About OCT Mark T. Dunbar, O.D., F.A.A.O. Bascom Palmer Eye Institute University of Miami, School of Medicine Mark Dunbar: Disclosure Optometry Advisory Board for: Allergan

More information

Generally, myopia is a condition of the eye that makes it

Generally, myopia is a condition of the eye that makes it Glaucoma Effect of Myopia on the Progression of Primary Open- Angle Glaucoma Jin Young Lee, 1 Kyung Rim Sung, 1 Seungbong Han, 2 and Jung Hwa Na 1 1 Department of Ophthalmology, College of Medicine, University

More information

Validation of the UNC OCT Index for the Diagnosis of Early Glaucoma

Validation of the UNC OCT Index for the Diagnosis of Early Glaucoma Article https://doi.org/10.1167/tvst.7.2.16 Validation of the UNC OCT Index for the Diagnosis of Early Glaucoma Jean-Claude Mwanza 1, Gary Lee 2, Donald L. Budenz 1, Joshua L. Warren 3, Michael Wall 4,

More information

Differences between Non-arteritic Anterior Ischemic Optic Neuropathy and Open Angle Glaucoma with Altitudinal Visual Field Defect

Differences between Non-arteritic Anterior Ischemic Optic Neuropathy and Open Angle Glaucoma with Altitudinal Visual Field Defect pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2015;29(6):418-423 http://dx.doi.org/10.3341/kjo.2015.29.6.418 Original Article Differences between Non-arteritic Anterior Ischemic Optic Neuropathy

More information

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(1):32-38 http://dx.doi.org/10.3341/kjo.2012.26.1.32 Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence

More information

CLINICAL SCIENCES. optic neuropathy characterized

CLINICAL SCIENCES. optic neuropathy characterized CLINICAL SCIENCES Spectral-Domain Optical Coherence Tomography for Detection of Localized Retinal Nerve Fiber Layer Defects in Patients With Open-Angle Glaucoma Na Rae Kim, MD; Eun Suk Lee, MD, PhD; Gong

More information

Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography

Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography ORIGINAL STUDY Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography Huijuan Wu, MD, PhD,*w Johannes F. de Boer, PhD,z and Teresa C. Chen,

More information

Research Article The Pattern of Retinal Nerve Fiber Layer and Macular Ganglion Cell-Inner Plexiform Layer Thickness Changes in Glaucoma

Research Article The Pattern of Retinal Nerve Fiber Layer and Macular Ganglion Cell-Inner Plexiform Layer Thickness Changes in Glaucoma Hindawi Ophthalmology Volume 2017, Article ID 78365, 8 pages https://doi.org/10.1155/2017/78365 Research Article The Pattern of Retinal Nerve Fiber Layer and Macular Ganglion Cell-Inner Plexiform Layer

More information

Clinical Use of OCT in Assessing Glaucoma Progression

Clinical Use of OCT in Assessing Glaucoma Progression r e v i e w Clinical Use of OCT in Assessing Glaucoma Progression Jacek Kotowski, MD; Gadi Wollstein, MD; Lindsey S. Folio, BS; Hiroshi Ishikawa, MD; Joel S. Schuman, MD ABSTRACT Detection of disease progression

More information

International Journal of Health Sciences and Research ISSN:

International Journal of Health Sciences and Research  ISSN: International Journal of Health Sciences and Research www.ijhsr.org ISSN: 2249-9571 Original Research Article Conversion of Ocular Hypertensives into Glaucoma: A Retrospective Study Aditi Singh 1, Shibi

More information

Structural examina.on: Imaging

Structural examina.on: Imaging ManaMa: Glaucoma Structural examina.on: Imaging Luís Abegão Pinto, MD, PhD Department of Ophthalmology CHLC Lisbon Faculty of Medicine, Lisbon University 1 11-10- 2013 Structural changes Qualitative changes

More information

The Role of the RNFL in the Diagnosis of Glaucoma

The Role of the RNFL in the Diagnosis of Glaucoma Chapter 1. The Role of the RNFL in the Diagnosis of Glaucoma Introduction Glaucoma is an optic neuropathy characterized by a loss of of retinal ganglion cells and their axons, the Retinal Nerve Fiber Layer

More information

Science & Technologies

Science & Technologies STANDARD COMPUTERIZED PERIMETRY IN FUNCTION OF DIAGNOSTIC GLAUCOMA Iljaz Ismaili, 1 Gazepov Strahil, 2, Goshevska Dashtevska Emilija 1 1 University Eye Clinic,Skopje 2 Clinical Hospital, Shtip Abstract

More information

S Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science

S Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science Department of Ophthalmology and Visual Sciences, Kyoto University Graduate School of Medicine, Kyoto, Japan Correspondence to: Dr T Tanabe, Department of Ophthalmology, The Tazuke Kofukai Medical Institute,

More information

Advances in the Structural Evaluation of Glaucoma with Optical Coherence Tomography

Advances in the Structural Evaluation of Glaucoma with Optical Coherence Tomography Curr Ophthalmol Rep (2013) 1:98 105 DOI 10.1007/s40135-013-0014-4 DIAGNOSIS AND MONITORING OF GLAUCOMA (S SMITH, SECTION EDITOR) Advances in the Structural Evaluation of Glaucoma with Optical Coherence

More information

Position of retinal blood vessels correlates with retinal nerve fibre layer thickness profiles as measured with GDx VCC and ECC

Position of retinal blood vessels correlates with retinal nerve fibre layer thickness profiles as measured with GDx VCC and ECC Department of Ophthalmology, Medical University of Vienna, Austria Correspondence to Clemens Vass, Department of Ophthalmology and Optometry, Medical University of Vienna, General Hospital, Währinger Gürtel

More information

Clinical and OCT features of different types and stages of diabetic optic neuropathy

Clinical and OCT features of different types and stages of diabetic optic neuropathy Clinical and OCT features of different types and stages of diabetic optic neuropathy P.A. Bezditko 1, Dr Sc (Med), Prof.; M.A. Karliychuk 2, Cand Sc (Med) 1 Kharkiv National Medical University ; Kharkiv

More information

Comparison of macular GCIPL and peripapillary RNFL deviation maps for detection of glaucomatous eye with localized RNFL defect

Comparison of macular GCIPL and peripapillary RNFL deviation maps for detection of glaucomatous eye with localized RNFL defect Comparison of macular GCIPL and peripapillary RNFL deviation maps for detection of glaucomatous eye with localized RNFL defect Mi Jeung Kim, 1,2 Ki Ho Park, 1,2 Beong Wook Yoo, 3 Jin Wook Jeoung, 1,2 Hee

More information

Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma

Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma Reza Zarei, MD 1 Mohammad Soleimani, MD 2 Sasan Moghimi, MD 3 Mohammad Yaser Kiarudi, MD 2 Mahmoud Jabbarvand, MD 1 Yadollah

More information

New Normative Database of Inner Macular Layer Thickness Measured by Spectralis OCT Used as Reference Standard for Glaucoma Detection

New Normative Database of Inner Macular Layer Thickness Measured by Spectralis OCT Used as Reference Standard for Glaucoma Detection Article https://doi.org/10.1167/tvst.7.1.20 New Normative Database of Inner Macular Layer Thickness Measured by Spectralis OCT Used as Reference Standard for Glaucoma Detection María Nieves-Moreno 1,2,

More information

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS Glaucoma Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study Jin Wook Jeoung, 1,2 Seok Hwan Kim, 1,3 Ki Ho Park, 1,2

More information

Thickness Changes in the Fovea and Peripapillary Retinal Nerve Fiber Layer Depend on the Degree of Myopia

Thickness Changes in the Fovea and Peripapillary Retinal Nerve Fiber Layer Depend on the Degree of Myopia Thickness Changes in the Fovea and Peripapillary Retinal Nerve Fiber Layer Depend on the Degree of Myopia Sung-Won Choi, MD, Seok-Joon Lee, MD Department of Ophthalmology, Wonju Christian Hospital. Yonsei

More information

Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve Fiber Layer Defects

Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve Fiber Layer Defects 접수번호 : 2008-105 Korean Journal of Ophthalmology 2009;23:169-175 ISSN : 1011-8942 DOI : 10.3341/kjo.2009.23.3.169 Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve

More information

Relationship Between Structure

Relationship Between Structure Original Article Relationship Between Structure and Function of the Optic Nerve Head-Glaucoma versus Normal Dr Savita Bhat, Dr Anna Elias, Dr Siddharth Pawar, Dr S.J. Saikumar, Dr Alpesh Rajput, superior,

More information

Ganglion cell analysis by optical coherence tomography (OCT) Jonathan A. Micieli, MD Valérie Biousse, MD

Ganglion cell analysis by optical coherence tomography (OCT) Jonathan A. Micieli, MD Valérie Biousse, MD Ganglion cell analysis by optical coherence tomography (OCT) Jonathan A. Micieli, MD Valérie Biousse, MD Figure 1. Normal OCT of the macula (cross section through the line indicated on the fundus photo)

More information

Corporate Medical Policy

Corporate Medical Policy Corporate Medical Policy Glaucoma, Evaluation by Ophthalmologic Techniques File Name: Origination: Last CAP Review: Next CAP Review: Last Review: glaucoma_evaluation_by_ophthalmologic_techniques 3/2001

More information

Although glaucoma is classified as an optic nerve disease,

Although glaucoma is classified as an optic nerve disease, Glaucoma Macular Ganglion Cell Inner Plexiform Layer: Automated Detection and Thickness Reproducibility with Spectral Domain Optical Coherence Tomography in Glaucoma Jean-Claude Mwanza, 1 Jonathan D. Oakley,

More information

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Chiharu Matsumoto, Shiroaki Shirato, Mai Haneda, Hiroko Yamashiro

More information

Optical coherence tomography (OCT) is a noninvasive,

Optical coherence tomography (OCT) is a noninvasive, Ability of Stratus OCT to Detect Progressive Retinal Nerve Fiber Layer Atrophy in Glaucoma Eun Ji Lee, 1,2 Tae-Woo Kim, 1,2 Ki Ho Park, 2 Mincheol Seong, 3 Hyunjoong Kim, 4 and Dong Myung Kim 2 PURPOSE.

More information

OCT in the Diagnosis and Follow-up of Glaucoma

OCT in the Diagnosis and Follow-up of Glaucoma OCT in the Diagnosis and Follow-up of Glaucoma Karim A Raafat MD. Professor Of Ophthalmology Cairo University Hmmmm! Do I have Glaucoma or not?! 1 Visual Function 100% - N Gl Structure : - 5000 axon /

More information

Il contributo dell'angio-oct: valutazione integrata della componente nervosa e vascolare della malattia glaucomatosa

Il contributo dell'angio-oct: valutazione integrata della componente nervosa e vascolare della malattia glaucomatosa SIMPOSIO G.O.A.L. - LE NUOVE FRONTIERE DIAGNOSTICHE E LE LINEE DI INDIRIZZO AMBULATORIALI DEL GLAUCOMA Coordinatore e moderatore: D. Mazzacane Presidente: L. Rossetti Il contributo dell'angio-oct: valutazione

More information

R&M Solutions

R&M Solutions Mohamed Hosny El-Bradey, MD., Assistant Professor of Ophthalmology, Tanta University. Wael El Haig, MD., Professor of Ophthalmology. Zagazeeg University. 1 Myopic CNV is considered the most common vision

More information

Nobuko Enomoto, Ayako Anraku, Kyoko Ishida, Asuka Takeyama, Fumihiko Yagi, and Goji Tomita

Nobuko Enomoto, Ayako Anraku, Kyoko Ishida, Asuka Takeyama, Fumihiko Yagi, and Goji Tomita Ophthalmology Volume 2015, Article ID 186249, 6 pages http://dx.doi.org/10.1155/2015/186249 Research Article Size of the Optic Nerve Head and Its Relationship with the Thickness of the Macular Ganglion

More information

Effect of refractive errors/axial length on peripapillary retinal nerve fibre layer thickness (RNFL) measured by Topcon SD-OCT

Effect of refractive errors/axial length on peripapillary retinal nerve fibre layer thickness (RNFL) measured by Topcon SD-OCT 1054 RESEARCH ARTICLE Effect of refractive errors/axial length on peripapillary retinal nerve fibre layer thickness (RNFL) measured by Topcon SD-OCT Ayisha Kausar, 1 Nadia Akhtar, 2 Farooq Afzal, 3 Khalid

More information

Since the introduction of optical coherence tomography

Since the introduction of optical coherence tomography Multidisciplinary Ophthalmic Imaging Repeatability of Peripapillary Retinal Nerve Fiber Layer and Inner Retinal Thickness Among Two Spectral Domain Optical Coherence Tomography Devices Juliane Matlach,

More information

Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes

Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes Hiroshi Nakamura,* Toshine Maeda,* Yasuyuki Suzuki and Yoichi Inoue* *Eye Division of Olympia Medical Clinic, Tokyo, Japan; Department of

More information

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note.

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note. Case 1 Normal Tension Glaucoma 53 year old woman attends your practice for routine exam. She has no past medical history or family history of note. Table 1. Right Eye Left Eye Visual acuity 6/6 6/6 Ishihara

More information

Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach

Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach ORIGINAL STUDY Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach Joong Won Shin, MD, Kyung Rim Sung, MD, PhD, Jiyun

More information

Reproducibility of Bruch Membrane Opening-Minimum Rim Width Measurements With Spectral Domain Optical Coherence Tomography

Reproducibility of Bruch Membrane Opening-Minimum Rim Width Measurements With Spectral Domain Optical Coherence Tomography ORIGINAL STUDY Reproducibility of Bruch Membrane Opening-Minimum Rim Width Measurements With Spectral Domain Optical Coherence Tomography Keunheung Park, MD,* Jinmi Kim, PhD, and Jiwoong Lee, MD, PhD*

More information

Optical Coherence Tomography-Measured Nerve Fiber Layer and Macular Thickness in Emmetropic, High-Myopic and High-Hyperopic Eyes

Optical Coherence Tomography-Measured Nerve Fiber Layer and Macular Thickness in Emmetropic, High-Myopic and High-Hyperopic Eyes Optical Coherence Tomography-Measured Nerve Fiber Layer and Macular Thickness in Emmetropic, High-Myopic and High-Hyperopic Eyes Mohammad-Mehdi Parvaresh, MD 1 Marjan Imani, MD 2 Mohsen Bahmani-Kashkouli,

More information

Evaluating Optic Nerve Damage: Pearls and Pitfalls

Evaluating Optic Nerve Damage: Pearls and Pitfalls 54 The Open Ophthalmology Journal, 9, 3, 54-58 Evaluating Optic Nerve Damage: Pearls and Pitfalls Open Access Paul J. Mackenzie * and Frederick S. Mikelberg Division of Glaucoma, Department of Ophthalmology

More information

Performance of time-domain and spectral-domain Optical Coherence Tomography for glaucoma screening

Performance of time-domain and spectral-domain Optical Coherence Tomography for glaucoma screening Performance of time-domain and spectral-domain Optical Coherence Tomography for glaucoma screening Boel Bengtsson, Sabina Andersson and Anders Heijl Department of Clinical Sciences, Ophthalmology in Malmo,

More information

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS Quantitative Assessment of Atypical Birefringence Images Using Scanning Laser Polarimetry With Variable Corneal Compensation HARMOHINA BAGGA, MD, DAVID S. GREENFIELD, MD, AND WILLIAM J. FEUER, MS PURPOSE:

More information

EFFECT OF OPTIC DISK FOVEA DISTANCE ON MEASUREMENTS OF INDIVIDUAL MACULAR INTRARETINAL LAYERS IN NORMAL SUBJECTS

EFFECT OF OPTIC DISK FOVEA DISTANCE ON MEASUREMENTS OF INDIVIDUAL MACULAR INTRARETINAL LAYERS IN NORMAL SUBJECTS EFFECT OF OPTIC DISK FOVEA DISTANCE ON MEASUREMENTS OF INDIVIDUAL MACULAR INTRARETINAL LAYERS IN NORMAL SUBJECTS KUNLIANG QIU, MD,* BINYAO CHEN, MD,* HAOYU CHEN, MD,* ENTING GAO, PHD, JIANLING YANG, MD,*

More information

Glaucoma Evaluation. OCT Pearls for Glaucoma. OCT: Retinal Nerve Fiber Layer. Financial Disclosures. OCT: Macula. Case Example

Glaucoma Evaluation. OCT Pearls for Glaucoma. OCT: Retinal Nerve Fiber Layer. Financial Disclosures. OCT: Macula. Case Example OCT Pearls for Glaucoma using OCT of the macula for glaucoma Glaucoma Evaluation Right eye Visual Acuity 20/25 20/25 IOP 13 13 Central corneal 530 530 thickness Anterior exam Normal with PCIOL Normal with

More information

MATERIALS AND METHODS

MATERIALS AND METHODS Glaucoma Analysis of Peripapillary Retinal Nerve Fiber Distribution in Normal Young Adults Seung Woo Hong, 1,2 Myung Douk Ahn, 2 Shin Hee Kang, 1,3 and Seong Kyu Im 1,4 PURPOSE. To determine the anatomic

More information

Evolving glaucoma management True diagnostic integration for the preservation of vision

Evolving glaucoma management True diagnostic integration for the preservation of vision Evolving glaucoma management True diagnostic integration for the preservation of vision // GLAUCOMA MANAGEMENT MADE BY ZEISS The moment you are certain it is glaucoma. This is the moment we work for. There

More information

Optic Nerve: Clinical Examination

Optic Nerve: Clinical Examination Optic Nerve: Clinical Examination Marcelo T. Nicolela 2 Core Messages Optic disc evaluation is of fundamental importance in the management of glaucoma. Clinical examination of the optic disc is best performed

More information

Diagnostic Accuracy of Scanning Laser Polarimetry with Enhanced versus Variable Corneal Compensation

Diagnostic Accuracy of Scanning Laser Polarimetry with Enhanced versus Variable Corneal Compensation Diagnostic Accuracy of Scanning Laser olarimetry with Enhanced versus Variable Corneal T. A. Mai, MD, N. J. Reus, MD, hd, H. G. Lemij, MD, hd urpose: To compare the diagnostic accuracy of scanning laser

More information

Glaucoma: a disease of the macula?

Glaucoma: a disease of the macula? Glaucoma: a disease of the macula? Derek MacDonald, OD, FAAO Waterloo, Ontario, Canada Please silence all mobile devices and remove items from chairs so others can sit. Unauthorized recording of this session

More information

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7 Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7 61 Abstract Purpose The GDx is a scanning laser polarimeter that assesses

More information

21st Century Visual Field Testing

21st Century Visual Field Testing Supplement to Supported by an educational grant from Carl Zeiss Meditec, Inc. Winter 2011 21st Century Visual Field Testing the Evolution Continues 21st Century Visual Field Testing 21st Century Visual

More information

Spectral- and time-domain optical coherence tomography measurements of macular thickness in young myopic eyes

Spectral- and time-domain optical coherence tomography measurements of macular thickness in young myopic eyes Liu et al. Diagnostic Pathology 2014, 9:38 RESEARCH Open Access Spectral- and time-domain optical coherence tomography measurements of macular thickness in young myopic eyes Lin Liu, Jun Zou *, Lili Jia,

More information

Eye Movements, Strabismus, Amblyopia, and Neuro-Ophthalmology

Eye Movements, Strabismus, Amblyopia, and Neuro-Ophthalmology Eye Movements, Strabismus, Amblyopia, and Neuro-Ophthaology Scanning Laser Polarimetry, but Not Optical Coherence Tomography Predicts Permanent Visual Field Loss in Acute Nonarteritic Anterior Ischemic

More information

How to Be Efficient and Effective. Disclosure. Topics CASE CM. Case JF 2007 OHTN / POAG? How to Be Efficient and Effective with. with New Technology

How to Be Efficient and Effective. Disclosure. Topics CASE CM. Case JF 2007 OHTN / POAG? How to Be Efficient and Effective with. with New Technology How to Be Efficient and Effective with Disclosure COPE Course ID: 40750 GL Michael Chaglasian has the following disclosures: 1. Advisory Board: Allergan, Inc., Alcon Labs, B+L Carl Zeiss Meditec 2. Research:

More information

A Comparative Analysis of Ganglion Cell Complex Parameters in Nigerian Negroes with Glaucoma and Macular Disease

A Comparative Analysis of Ganglion Cell Complex Parameters in Nigerian Negroes with Glaucoma and Macular Disease Nigerian Journal of Ophthalmology 2012; 20(1): 14-20 A Comparative Analysis of Ganglion Cell Complex Parameters in Nigerian Negroes with Glaucoma and Macular Disease O Oderinlo, AO Ogunro FWACS, AO Hassan

More information

Ganglion Cell-Inner Plexiform Layer Assesment with Cirrus Hd 5000 in Early Glaucoma

Ganglion Cell-Inner Plexiform Layer Assesment with Cirrus Hd 5000 in Early Glaucoma AASCIT Journal of Medicine 2017; 3(6): 40-45 http://www.aascit.org/journal/medicine ISSN: 2381-1420 (Print); ISSN: 2381-1447 (Online) Ganglion Cell-Inner Plexiform Layer Assesment with Cirrus Hd 5000 in

More information

Glaucoma Diagnosis & Tracking with Optical Coherence Tomography

Glaucoma Diagnosis & Tracking with Optical Coherence Tomography Glaucoma Diagnosis & Tracking with Optical Coherence Tomography David Huang, MD, PhD Charles C. Manger III, MD Chair of Corneal Laser Surgery Assoc. Prof. of Ophthalmology & Biomedical Engineering Doheny

More information

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Yamanashi Med. J. 18(1), 15~ 20, 2003 Original Article Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Satoshi KOGURE, Yoshiki TODA, Hiroyuki IIJIMA and Shigeo TSUKAHARA Department of

More information

STANDARD AUTOMATED PERIMETRY IS A GENERALLY

STANDARD AUTOMATED PERIMETRY IS A GENERALLY Comparison of Long-term Variability for Standard and Short-wavelength Automated Perimetry in Stable Glaucoma Patients EYTAN Z. BLUMENTHAL, MD, PAMELA A. SAMPLE, PHD, LINDA ZANGWILL, PHD, ALEXANDER C. LEE,

More information

Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients

Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients Yoshio Yamazaki, Kenji Mizuki, Fukuko Hayamizu and Chizuru Tanaka Department of Ophthalmology, Nihon

More information

Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 2: Optic Nerve Head and Retinal

Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 2: Optic Nerve Head and Retinal Original Article Philippine Journal of OPHTHALMOLOGY Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 2: Optic Nerve Head and Retinal Nerve Fiber Layer

More information

C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e. The GDxVCC Early answers and ongoing assessment for glaucoma

C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e. The GDxVCC Early answers and ongoing assessment for glaucoma C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e The GDxVCC Early answers and ongoing assessment for glaucoma The quantifiable approach to quality care Only Humphrey GPA software Early insight

More information

Measurement of Choroidal Thickness in Normal Eyes Using 3D OCT-1000 Spectral Domain Optical Coherence Tomography

Measurement of Choroidal Thickness in Normal Eyes Using 3D OCT-1000 Spectral Domain Optical Coherence Tomography pissn: 111-8942 eissn: 292-9382 Korean J Ophthalmol 212;26(4):255-259 http://dx.doi.org/1.3341/kjo.212.26.4.255 Original Article Measurement of Choroidal Thickness in Normal Eyes Using 3D OCT-1 Spectral

More information

Cirrus High-Definition Optical Coherence Tomography Compared with Stratus Optical Coherence Tomography in Glaucoma Diagnosis

Cirrus High-Definition Optical Coherence Tomography Compared with Stratus Optical Coherence Tomography in Glaucoma Diagnosis Glaucoma Cirrus High-Definition Optical Coherence Tomography Compared with Stratus Optical Coherence Tomography in Glaucoma Diagnosis Javier Moreno-Montañés, Natalia Olmo, Aurora Alvarez, Noelia García,

More information

OCT Angiography: The Next Step in Retinal Imaging Jonathan Zelenak D.O.

OCT Angiography: The Next Step in Retinal Imaging Jonathan Zelenak D.O. OCT Angiography: The Next Step in Retinal Imaging Jonathan Zelenak D.O. Hillsdale Hospital Michigan State University Overview Evolution of OCT How does OCT angiography work? Clinical examples Potential

More information

Changes in Peripapillary Retinal Nerve Fiber Layer Thickness after Pattern Scanning Laser Photocoagulation in Patients with Diabetic Retinopathy

Changes in Peripapillary Retinal Nerve Fiber Layer Thickness after Pattern Scanning Laser Photocoagulation in Patients with Diabetic Retinopathy pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2014;28(3):220-225 http://dx.doi.org/10.3341/kjo.2014.28.3.220 Original Article Changes in Peripapillary Retinal Nerve Fiber Layer Thickness after

More information

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Eytan Z. Blumenthal, MD, 1 Julia M. Williams, BS, 1 Robert N. Weinreb, MD, 1 Christopher A. Girkin, MD,

More information

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2010 November 18.

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2010 November 18. NIH Public Access Author Manuscript Published in final edited form as: Arch Ophthalmol. 2009 July ; 127(7): 875 881. doi:10.1001/archophthalmol.2009.145. Measurement of Local Retinal Ganglion Cell Layer

More information

Cirrus TM HD-OCT. Details defi ne your decisions

Cirrus TM HD-OCT. Details defi ne your decisions Cirrus TM HD-OCT Details defi ne your decisions 2 With high-defi nition OCT Carl Zeiss Meditec takes you beyond standard spectral domain Built on 10 years experience at the vanguard of innovation, Carl

More information

Clinical Study Choroidal Thickness in Eyes with Unilateral Ocular Ischemic Syndrome

Clinical Study Choroidal Thickness in Eyes with Unilateral Ocular Ischemic Syndrome Hindawi Publishing Corporation Journal of Ophthalmology Volume 215, Article ID 62372, 5 pages http://dx.doi.org/1.1155/215/62372 Clinical Study Choroidal Thickness in Eyes with Unilateral Ocular Ischemic

More information

Clinical Ophthalmology

Clinical Ophthalmology UDC 616.7.-007.681-085.27:577.112.385:617.735/.736-018.82-07 The thickness of the retinal ganglion cells complex (gcc) and nerve fibers in the macular area in patients with primary open-angle glaucoma

More information

Ethambutol (EMB), the first-line drug used to treat mycobacterium

Ethambutol (EMB), the first-line drug used to treat mycobacterium Eye Movements, Strabismus, Amblyopia and Neuro-Ophthalmology Ganglion Cell Layer and Inner Plexiform Layer as Predictors of Vision Recovery in Ethambutol-Induced Optic Neuropathy: A Longitudinal OCT Analysis

More information