1 593375JLAXXX / Journal of Laboratory AutomationBuoro et al. research-article2015 High-Throughput Imaging Technologies Reflex Testing Rules for Cell Count and Differentiation of Nucleated Elements in and Ascitic Fluids on Sysmex XE-5000 Journal of Laboratory Automation 2016, Vol. 21(2) Society for Laboratory Automation and Screening DOI: / jala.sagepub.com Sabrina Buoro 1, Sara Appassiti Esposito 1, Mauro Vavassori 1, Tommaso Mecca 1, Cosimo Ottomano 2, Paola Dominoni 1, Michela Seghezzi 1, Elisabetta Candiago 3, Claudio Farina 4, Andrea Gianatti 3, Alberto Crippa 1, and Giuseppe Lippi 2 Abstract Flow cytometry is widely used in many laboratories for automated nucleated cell counts and their differentiation in body fluids. The implementation of new reflex testing rules on these automated instruments could open new frontiers in laboratory workflow, improving characterization of body fluids and clinical diagnosis and decreasing costs. Ascitic (150) and pleural (33) fluids were collected and assessed by XE-5000 and optical microscopy. Cell counts performed with the methods showed a Pearson s correlation of 0.98, Passing Bablok regression y = 0.99x , and bias of In ascitic fluids, the best diagnostic performance was found for polymorphonuclear and neutrophil counts on XE-5000, which exhibited areas under the curve (AUCs) 0.98 and 0.99, respectively. In pleural fluids the best diagnostic performance was found for polymorphonuclear percent parameter, which displayed Specific reflex test rules based on these parameters were characterized by 92 diagnostic concordance, 1.00 sensitivity, and 0.84 specificity with optical microscopy. The application of a set of reflex testing rules may improve the diagnostic performance of XE-5000, increasing its reliability for routine automated cell count in body fluids. We acknowledge that further studies should be planned to validate our findings according to clinical data. Keywords body fluids, XE-5000, reflex testing rules, automated cell count Introduction The analysis of pleural (PF) and ascitic/peritoneal (AF) fluids by means of flow cytometry represents a valuable opportunity for diagnosis and categorization of body cavity effusions. A diagnosis of spontaneous bacterial peritonitis can be made when the neutrophil (NE) count in ascitic fluids exceeds 250 cells/μl, whereas the diagnosis of tuberculous peritonitis is very likely in the presence of a total nucleated cell (TNC) count of 1000 TNC/μL in combination with a prevalent population of lymphocytes (LYs). 1,2 A TNC count of 1000 TNC/μL in PFs is suggestive for a diagnosis of exudative pleural effusion. In particular, the presence of a prevalent population of NEs (i.e., >50) is indicative of acute inflammation or parapneumonic effusion, whereas the presence of a prevalent population of LYs (i.e., >50) is frequently observed in patients with tubercular infection, metastasis, lymphoproliferative disorder, or chylous effusion. 1 The presence of eosinophils (EOs) in 10 of TNCs may be indicative of pneumothorax, pulmonary embolism, parasitic diseases, or Churg Strauss syndrome. 1 Although optical microscopy (OM) remains the gold standard for identification and enumeration of TNCs in body fluids, 1,3 this technique has some obvious drawbacks, which include high imprecision and interobserved variability, 4 time-consuming sample preparation and examination, need for trained and qualified personnel for accurate morphological interpretation, and long turnaround time (TAT), along with costs that are greater than those of automated cell counting. 5 1 Clinical Chemistry Laboratory, A.O. Papa Giovanni XXIII, Bergamo, Italy 2 Clinical Chemistry and Hematology Laboratory, Azienda Ospedaliero Universitaria di Parma, Parma, Italy 3 Pathology Laboratory, A.O. Papa Giovanni XXIII, Bergamo, Italy 4 Microbiology Laboratory, A.O. Papa Giovanni XXIII, Bergamo, Italy Received March 30, Corresponding Author: Sabrina Buoro, PhD, Clinical Chemistry Laboratory, A.O. Papa Giovanni XXIII, Piazza OMS, 1, Bergamo, Italy. sbuoro@hpg23.it

2 298 Journal of Laboratory Automation 21(2) The new generation of fully automated cell counters, including hematological analyzers and flow cytofluorimeters coupled with image analysis software for blood cell analysis or automated urinalysis, has also been developed for cell counting and differentiation in body fluids It is noteworthy that the automated analysis of body fluid cellularity is mostly challenged by the presence of macrophages (MACROs), mesothelial cells (MESOs), and other abnormal or pathological cells (especially solid cancer cells) that are typically absent in blood or urine, so that the accuracy and differentiation of cellular elements may be impaired. 6,7,15,22,24 More specifically, some automated cell count analyzers provide a TNC count but not morphological differentiation, whereas specific information about the accuracy of cell recognition is unavailable for others. 8,23,24 Even more importantly, validation rules for automated cell counting and a definition of the reflex testing required for accurate result reporting are unavailable in the current scientific literature to the best of our knowledge. The hemocytometer XE-5000 (Sysmex Co., Kobe, Japan) is equipped with specific software for enumeration and identification of body fluid cellularity. The analytical performance (i.e., imprecision, functional sensitivity, linearity, and carryover) has been thoughtfully validated, 4 13 thus enabling its routine use for analysis of body fluids. Nevertheless, a set of rules and the associated reflex tests that may be applied for validation of cell counting and differentiation in body fluids have not have been defined or evaluated so far. 1 Therefore, the aims of this study were to compare the performances of Sysmex XE-5000 and OM for TNC counting and differentiation in different PF and AF specimens, to evaluate diagnostic agreement with cutoffs provided in the CLSI standard, 1 and to develop a new set of rules based on diagnostic parameters provided by XE-5000 that could be used for automated cell count validation in routine analysis. Materials and Methods Cell Counts A total of 184 consecutive fresh samples of body cavity fluids (151 AFs and 33 PFs) collected in 5.4 mg 2.0 ml K 3 EDTA tubes (Becton Dickinson, Franklin Lakes, NJ) and referred to the clinical laboratory for routine analysis from inpatient wards over approximately one month were simultaneously assessed by OM and Sysmex XE TNC and differential counts were performed with the XE-5000 body fluid (BF) mode, and results were compared to those obtained with OM after staining with Turk reagents and counting in a Nageotte chamber (dilution 1:20 or 1:200). Morphological classification was based on analysis of May Grunwald Giemsastained cytospin slides (Carlo Erba Reagents S.p.A., Milano, Italy) by two expert laboratory professionals, finally expressing results as the average of the two independent observations. The BF mode on Sysmex XE-5000 offers a number of default parameters, including total nucleated cell (TC-BF) and white blood cell (WBC-BF) counts and differential cell counts for mononuclear elements (MN-BF) and polymorphonuclear leukocytes (PMN-BF), both in percentage and absolute number (#). Additional research parameters include eosinophil (EO-BF) and high-fluorescence cell (HF-BF) counting, both in percentage and absolute number. The morphological differentiation by OM entailed cell classification in one of the following classes: neutrophils (NEs), lymphocytes (LYs), monocytes (MOs), eosinophils (EOs), basophils (BASOs), macrophages (MACROs), mesothelial cells (MESOs), and other cells (OTHER; also including blastlike cells). The cells were then clustered in a discrete number of homogeneous cell categories to enable direct comparison of enumeration and differentiation with the two methods (i.e., XE-5000 vs OM), as follows: TC-BF vs OM TNC # MN-BF vs OM-MN1 (= LY + MO) and # MN-BF vs OM-MN2 (= LY + MO + MACRO) and # PMN-BF vs OM-PMN (NE + EO + BASO) and # NE-XE-5000 (= PMN-BF-EO-BF) vs OM-NE and # HF-BF vs OM-NO-WBC (= MESO + OTHER) and # The agreement between XE-5000 and OM was assessed with Pearson s correlation coefficient (r), Passing Bablok regression, Bland Altman plot analysis, and Spearman s rank correlation. The diagnostic concordance of XE-5000 compared to OM was evaluated with receiver operating characteristic (ROC) curves, diagnostic concordance, sensitivity, and specificity. The analysis was performed by using either the diagnostic thresholds of cellularity in pleural and ascitic/peritoneal fluids defined by the CLSI (AFs, number of NE-XE-5000 or PMN-BF 250 cells/µl; PFs, percentage of NE-XE-5000 or PMN-BF or MN-BF 50) 1 or instrument-specific cutoffs identified by means of ROC curve analysis. Rules of Reflex Test The presence of morphological flags (i.e., WBC abnormal scattergram), along with DIFF scattergram analysis and data obtained for HF-BF and EO-BF, was then used to define a set of rules for assisting the validation of automated analysis of body fluids, activating reflex tests when needed (i.e., optical cell counting or differentiation using cytospin slides), and validating the use of this approach in routine

3 Buoro et al. 299 diagnostics of body fluids. The scattergrams were assessed using the same criteria as for complete blood cell (CBC) count. The following criteria were then used for establishing the positivity of the OM count and differentiation on cytospin slides: Presence of bacterial flora and/or mycetes Smudge cells 20 Abnormal cells count > 5 (i.e., morphologically aberrant elements and cells with specific peculiar characteristics such as lipophages, erythrophages, activated MESOs, etc.) Cancer cells count 1 (including blasts) Cell agglutination The performance of validation rules was evaluated against OM analysis by calculation of diagnostic concordance, sensitivity, specificity, Pearson s chi-squared test (χ 2 ), and Cohen s kappa coefficient of agreement. The statistical analysis was performed with Analyse-it for Microsoft Excel (Analyse-it Software, Ltd., Leeds, UK). Microbiological Testing Microscopic examination was performed on Gram-stained direct smears of pleural and peritoneal fluids according to standard procedures. The matrix-assisted laser desorption/ ionization (MALDI) time-of-flight (TOF) technique (VITEK MS, biomérieux, Marcy-l'Étoile, France) was used for the identification of positive cultures. Chemosensitivity testing was carried out using the automated microdilution method (VITEK2, biomérieux). Cytologic Examination Samples were fixed with Cytolyt hemolytic and preservative solution (Hologic, Inc., Marlborough, MA). Cells were spun at 1500 rotations/min and the sediment transferred in PreservCyt solution (Cytyc Corporation, Marlborough, MA) to be processed with the T5000 automated processor according to the manufacturer s recommendation (Hologic). The resulting slide was fixed in 95 ethanol and stained with the Papanicolaou method. The remaining material was stored in PreservCyt solution for possible later use in the preparation of additional slides for further investigations, including immunocytochemistry, molecular diagnosis, and flow cytometry. All specimens were classified as adequate or inadequate cytological material by a cytopathologist. Samples with adequate classification were further categorized as (1) negative, (2) atypical, (3) suspicious, or (4) malignant. The study was carried out in accordance with the Declaration of Helsinki, under the terms of all relevant local legislation. Figure 1. Comparison of cell counts with manual optical method (OM-TNC) and automated technique (TC-BF) in all samples. (A) Passing Bablok regression. (B) Bland Altman bias. Results Due to heavy cell deterioration, the TNC by OM count could not be performed in five AF samples. Therefore, the agreement of cell counting between XE-5000 and OM was studied in 179 samples, with a total cellularity ranging from 8 to 18,560 TNC/ µl (median, 1940 TNC/µL; 95 confidence interval [CI], TNC/µL). Eighteen samples were also excluded from morphological analysis for technical reasons (i.e., insufficient sample, unreadable cytospin slide, etc.). Therefore, the agreement of the cell differentiation between XE-5000 and OM was assessed in 161 samples (130 AFs and 31 PFs). The number of positive samples according to the CLSI standard was 60/161 (37.3) with OM analysis. Correlations The comparison of the TNC count between XE-5000 and OM (TC-BF vs OM-TNC) was as follows: Pearson s correlation r = 0.98, Passing Bablok regression y = 0.99x (95 CI of slope, ; 95 CI of intercept, 5.66 to 9.43), and Bland Altman bias 32.3 (95 CI, to 54.7) (Fig. 1A,B). The correlations between XE-5000 and OM for

4 300 Journal of Laboratory Automation 21(2) Table 1. Pearson Correlation, Passing Bablok Regression, and Bland Altman Bias for Different Cell Categories in Optical Microscopy (OM) vs XE-5000 Automated Count in Body Fluids (BF). a Pearson Correlation TC-BF vs OM-TNC 0.98 MN-BF vs OM-MN (LY + MO) MN-BF vs OM-MN1 # 0.95 (LY + MO) MN-BF vs OM-MN (LY + MO + MACRO) MN-BF vs OM-MN2 # 0.96 (LY + MO + MACRO) PMN-BF vs OM-PMN 0.93 (NE + EO + BASO) PMN-BF vs OM-PMN # 0.98 (NE + EO + BASO) NE-XE-5000 vs OM-NE 0.96 NE-XE-5000 vs OM-NE # 0.98 Passing Bablok Regression (95 CI Slope, Intercept) y = 0.99x ( , 5.66 to 9.43) y = 0.99x ( , ) y = 1.08x ( , ) y = 0.94x ( , ) y = 1.02x 5.43 ( , to 3.40) y = 0.94x ( , ) y = 0.91x ( , ) y = 0.94x ( , ) y = 0.92x ( , ) Bland Altman Bias (95 CI) 32.3 ( to 54.7) 18.4 ( ) 90.3 ( ) 7.0 ( ) ( ) 0.52 ( 2.3 to 1.3) 60 ( to 32.5) 0.79 ( 2.2 to 0.6) ( to 38.4) TC, total nucleated cell; WBC, white blood cell; MN, mononuclear element; PMN, polymorphonuclear leukocyte; NE, neutrophil; LY, lymphocyte; MO, monocyte; EO, eosinophil; BASO, basophil; MACRO, macrophage; CI, confidence interval. a Values in percentage and absolute number [#]. the different cell populations are listed in Table 1, with Pearson s correlation coefficients always comprised between 0.77 and 0.98 (all ). Diagnostic Concordance As regards the diagnostic concordance of XE-5000 in the 130 AFs, the area under the curve (AUC) was 0.98 (95 CI, ; p < ) for PMN-BF and 0.99 (95 CI, ; p < ) for NE-XE-5000, respectively. When a cutoff of 250 cells/µl for both PMN-BF # and NE-XE-5000 # was used, the diagnostic concordances compared to OM were 94 and 95, respectively. More specifically, 122 out of 130 samples were correctly classified using PMN-BF # (three false negative and six false positive results, 0.87 sensitivity and 0.96 specificity), whereas 123 out of 130 samples were correctly classified using the calculated parameter NE-XE-5000 (two false negative and six false positive results; 0.86 sensitivity and 0.97 specificity) (Table 2). An instrument-dependent threshold of 135 cells/μl could hence be identified with ROC curve analysis for both PMN-BF and NE-XE When this threshold was used, both parameters exhibited an identical diagnostic concordance of 94, with 1.00 sensitivity for both, 0.90 specificity for PMN-BF (nine false positive results), and 0.91 specificity for NE-XE-5000 (eight false positive results) (Table 2). As regards the diagnostic concordance of XE-5000 in the 31 PFs, the AUC was 0.75 (95 CI, ; p < ) for MN-BF and 0.97 (95 CI, ; p < ) for both PMN-BF and NE-XE-5000 (Table 2). When a cutoff of 50 cellular elements was used, the diagnostic concordance for MN-BF, PMN-BF, and NE-XE-5000 were 61, 97, and 97 compared to OM. In particular, 19 out of 31 samples were correctly classified with MN-BF (1 false negative and 11 false positive results; 0.93 sensitivity and 0.31 specificity), whereas 30 out of 31 samples were correctly classified with both PMN-BF and NE-XE-5000 (1 false positive result; 1.00 sensitivity and 0.96 specificity) (Table 2). Reflex Testing Rules The number of HF-BF was correlated with the number of MESO and/or cancer cells identified with OM analysis. The resulting Spearman s correlation was r = 0.70 for both absolute values and percentage. The ROC curve analysis of HF-BF compared to OM analysis of cytospin slides yielded AUCs of 0.75 (95 CI, ; p < ) for HF-BF # and 0.59 (95 CI, ; p < ) for HF-BF, respectively. Using a conventional cutoff of 50 cells/µl, the diagnostic concordance, sensitivity, and specificity of HF-BF # were 73, 0.49, and 0.90, respectively. According to these results, we developed the following validation rules for automated TNC enumeration and further cell differentiation:

5 Buoro et al. 301 Table 2. ROC Analysis of Different XE-5000 Parameters in Ascitic (AF) and (PF) Fluids. AUC (95 CI) Cellular Cutoff Sensitivity Specificity Diagnostic Concordance PMN-BF # on ascitic fluids NE-XE-5000 # on ascitic fluids PMN-BF # on ascitic fluids NE-XE-5000 # on ascitic fluids PMN-BF on pleural fluids NE-XE-5000 on pleural fluids MN-XE-5000 on pleural fluids 0.98 ( ) 0.99 ( ) 0.98 ( ) 0.99 ( ) 0.97 ( ) 0.97 ( ) 0.75 ( ) 250 cells/µl cells/µl cells/µl cells/µl TC, total nucleated cell; WBC, white blood cell; MN, mononuclear cell; PMF, polymorphonuclear cell; NE, neutrophil; CI, confidence interval; AUC, area under the curve. Rule 1: In the presence of WBC abnormal scattergram, perform OM analysis and further cell differentiation on stained cytospin slides. Rule 2: In the presence of HF-BF 50 cells/µl (the best cutoff with the ROC analysis evaluation of the sensitivity and specificity above described), perform OM analysis on stained cytospin slides. Rule 3: In the presence of EO > 10, perform OM analysis on stained cytospin slides. Rule 4: In the presence of abnormal WBC-DIFF scattergram, even without the WBC abnormal scattergram flag, perform OM analysis and further cell differentiation on stained cytospin slides. The application of these rules yielded the following diagnostic performance: diagnostic concordance of 92, sensitivity of 1.00, specificity of 0.84, a Pearson s chi-squared test of , and a Cohen s kappa coefficient of 0.84 (). Setting Reflex Testing Rules on XE-5000 and Their Application for Clinical Screening A full cytological examination was performed in 111 of the total number of samples (88 ascitic and 23 pleural samples), whereas microbiological testing was carried out in 132 samples (108 ascitic and 24 pleural samples). Overall, 38 out of 151 samples were obtained from patients with hepatocellular carcinoma (HCC), whereas the remaining 113 samples were collected from patients undergoing paracentesis drainage after surgery or cirrhosis-induced ascites. The cytologic examination performed on ascitic fluids did not reveal neoplastic cells (87 negative samples out of 88), with the notable exception of a single case in which atypia and cell necrosis was observed. This case also tested positive with both rule 1 (presence of WBC abnormal scattergram) and microbiological testing (presence of Escherichia coli). A complete bacterial culture test was available for 33 samples of ascitic fluids (out of 48) with PMN-BF > 250 cells/µl, and 16 of them (49) tested positive for the following pathogens: 9 samples with Enterococcus faecium, 2 with E. coli, 3 with mixed bacterial flora, and 2 with Candida albicans. On XE-5000, 63 samples of ascitic fluids were positive at the instrument-specific cutoff (PMN-BF > 135 cells/µl). The following bacterial culture test (available for 44 specimens) yielded the following results: 23 samples (53) were positive for E. faecium, 4 for E. coli, 2 for E. faecium and Klebsiella pneumonie, 3 for mixed bacterial flora, 2 for C. albicans, 1 for Enterecoccus faecalis, and 9 were negative. We evaluated 31 samples of pleural fluids and 22 of them showed a TC-BF of >1000 cell/µl; 6 of them were positive with cytologic examination and diagnosed as follows: metastasis from undifferentiated ovarian carcinoma, plasma cell infiltrates, localized chronic lymphoproliferative disorder, malignant mesothelioma, and two pulmonary

6 Table 3. Data of XE-5000, Optical Microscope Evaluation, and Reflex Testing Rules on Six Fluid (PF) Samples Positive by Cytologic Examination. XE-5000 Count OM Differential Count TC-BF cells/µl PMN-BF MN-BF HF-BF cells/µl EO-BF OM-NE OM-LY MACRO MESO OTHER Cells Positive Rules Results of Cytologic Examination Results of Microbiology Evaluation sample A sample B sample C sample D sample E sample F No No. 1 No No No No No. 2 Positive for malignant tumor cells Immunocytochemical profile consistent with ovarian origin (positive for cytokeratin [CK] 7, cancer antigen [Ca] 125, and Wilms tumor protein [WT-1] and negative for calretinin, CK20, and thyroid transcription factor [TTF] 1) Positive for malignant tumor cells (plasma cells) (Typing with flow cytometry cells is present with double positive CD138/CD38 for 87 of mononuclear cells and double positivity for CD138/CD56 and CD38/ CD56 for 85 of the cell population) Subject with multiple myeloma IgA monoclonal peak kappa and proteinuria Bence Jones (light-chain monoclonal kappa) Cytological atypia of undetermined significance for a mesothelial proliferation Profile immunocytochemical: calretinin positive, TTF-1 negative Positive for malignant tumor cells (small lymphocytes) (Typing with flow cytometric light-chain restriction confirmation in a subject with chronic lymphocytic leukemia [CLL]) Positive for malignant tumor cells Immunocytochemical profile consistent with pulmonary origin positive for TTF-1 +, CK7 +, A napsina Negative Negative Negative Negative None Positive for malignant tumor cells None 302

7 Buoro et al. 303 carcinomas. All samples were positive for the suggested set of reflex testing rules (see Table 3). In the remaining 16 of 22 pleural samples with positive testing, 6 of them showed a PMN-BF count of >50, with microbiological culture being positive in 3 of these (50; 1 E. coli, 1 Staphylococcus aureus, and 1 C. albicans). On XE-5000, nine samples of pleural fluids showed an MN-BF count of >50, which was confirmed by OM-LY in four of them. The last five samples were positive for rule 2. Interestingly, one sample exhibiting EO-BF of 5.6 and HF-BF of 212 cells/µl by XE-5000 was reclassified as positive 1 after microscopic evaluation since it displayed an EO count of 23 by OM. Another sample, which was positive using rule 3 (EO-BF 33.6), was reconfirmed as positive with OM. Discussion In agreement with previous data, 4 13 the results of this study confirm the strong correlation of the TNC counts between XE-5000 and OM in body cavity fluids. The evaluation of differential count also exhibited a satisfactory performance compared to OM, in particular for both PMN-BF and NE-XE-5000, as shown in Table 1. Interestingly, although the comparison of cell identification with XE-5000 and OM showed a lack of correlation between MN-BF and lymphocytes in OM (data not shown), this parameter was strongly associated with OM-MN1 (LY + MO), and an even better correlation was found with OM-MN2 (LY+MO+MACRO) (Table 1). The overall performance of the MN-BF parameter, as a percentage or an absolute, is apparently lower, exhibiting clear trends of nonlinearity and overestimation (Table 1). This is attributable to the fact that MN-BF include a variety of cells, namely, LYs, MOs, and MACROs. The parameter MN-BF is characterized by a modest diagnostic performance in PF samples, with an AUC that is much lower than those of both PMN-BF and NE-XE (i.e., 0.75 vs 0.97). Conversely, PMN-BF and NE-XE-5000 at the conventional 250 cells/μl threshold 1 showed much better diagnostic performance in AF samples, which could be further improved by introducing an instrumentspecific cutoff (i.e., 135 cells/μl). Using the same instrument-specific cutoff, the percentage of positive cultures increased to 53, in agreement with the practical guidelines of the American Association for the Study of Liver Diseases (AASLD). 27 The results of this study also confirm the well-known inclination of XE-5000 to classify as HF-BF both MESO and nonleukocyte cells of probable neoplastic origin, as earlier described. 5,6,10,13 This is valuable information for screening purposes, as well as for the definition of validation rules for automated cell counting and activating reflex testing. In fact, HF-BF, when expressed as absolute value, is characterized by an AUC of 0.75 compared with OM for pathological cell identification. Therefore, the inclusion of this parameter in validation rules may contribute to define a new set of acceptance rules for automated analysis of body fluids, exhibiting a high diagnostic agreement with OM. Taken together, these findings are in support of routine use of XE-5000 for automated cell counting and morphologic differentiation of nucleated elements in pleural and ascitic fluids according to the CSLI standard 1 and the most recent guidelines of the International Society for Laboratory Hematology (ICSH). 26 The WBC-DIFF scattergrams, with clearly defined clusters, allow a reliable quantitative analysis of leukocyte population, especially polymorphonuclear leukocytes and NEs, but not mononuclear cells. 1 Even more importantly, the definition of validation rules and related reflex tests for automated cell counting has contributed to further improve the diagnostic performance of this analyzer. This would provide an effective support to morphological analysis and diagnostic classification of cells in body fluids. Moreover, the low imprecision of XE-5000 further validates its use for follow-up of patients, as enabled by a suitable framework of validation rules. Finally, the positive impact of this analyzer on the optimization of human and economic resources should not be discounted, as previously emphasized by Zimmermann et al. 4 In this study we have also demonstrated that the excellent performance of XE-5000 in cell counting and differentiation, combined with a set of rules for reflex testing of easy implementation, enables a better clinical interpretation of data obtained, thus consistently lowering TAT and costs. A paradigmatic example is represented by the faster diagnosis of metastatic cancer in the six PFs testing positive with instrumental analysis. Declaration of Conflicting Interest The authors declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article. Funding The authors received no financial support for the research, authorship, and/or publication of this article. References 1. Clinical and Laboratory Standards Institute. Body Fluid Analysis for Cellular Composition: Approved Guideline. (accessed on Sept 18, 2013). 2. Lippi, G.; Danese, E.; Cervellin, G.; et al. Laboratory Diagnostics of Spontaneous Bacterial Peritonitis. Clin. Chim. Acta 2014, 430, Mahieu, S.; Vertessen, F.; Van der Planken, M. Evaluation of ADVIA 120 CSF Assay (Bayer) vs. Chamber Counting of Cerebrospinal Fluid Specimens. Clin. Lab. Haematol. 2004, 26 (3),

8 304 Journal of Laboratory Automation 21(2) 4. Zimmermann, M.; Ruprecht, K.; Kainzinger, F.; et al. Automated vs. Manual Cerebrospinal Fluid Cell Counts: A Work and Cost Analysis Comparing the Sysmex XE-5000 and the Fuchs-Rosenthal Manual Counting Chamber. Int. J. Lab. Hematol. 2011, 33 (6), Paris, A.; Nhan, T.; Cornet, E.; et al. Performance Evaluation of the Body Fluid Mode on the Platform Sysmex XE-5000 Series Automated Hematology Analyzer. Int. J. Lab. Hematol. 2010, 32 (5), de Jonge, R.; Brouwer, R.; de Graaf, M. T.; et al. Evaluation of the New Body Fluid Mode on the Sysmex XE-5000 for Counting Leukocytes and Erythrocytes in Cerebrospinal Fluid and Other Body Fluids. Clin. Chem. Lab. Med. 2010, 48 (5), Li, A.; Grönlund, E.; Brattsand, G. Automated White Blood Cell Counts in Cerebrospinal Fluid Using the Body Fluid Mode on the Platform Sysmex XE Scand. J. Clin. Lab. Invest. 2014, 74 (8), Lippi, G.; Cattabiani, C.; Benegiamo, A.; et al. Evaluation of White Blood Cell Count in Peritoneal Fluid with Five Different Hemocytometers. Clin. Biochem. 2013, 46 (1 2), Fleming, C.; Brouwer, R.; Lindemans, J.; et al. Improved Software on the Sysmex XE-5000 BF Mode for Counting Leukocytes in Cerebrospinal Fluid. Clin. Chem. Lab. Med. 2013, 51 (4), Perné, A.; Hainfellner, J. A.; Womastek, I.; et al. Performance Evaluation of the Sysmex XE-5000 Hematology Analyzer for White Blood Cell Analysis in Cerebrospinal Fluid. Arch. Pathol. Lab. Med. 2012, 136 (2), Kleine, T. O.; Nebe, C. T.; Löwer, C.; et al. Cell Analysis in Cerebrospinal Fluid (CSF) Using Sysmex Hematology Analyzers XT-4000i and XE-5000: Evaluation with CSF Controls of the Joint German Society for Clinical Chemistry and Laboratory Medicine (DGKL). Cytometry A 2012, 81 (3), Sandhaus, L. M.; Ciarlini, P.; Kidric, D.; et al. Automated Cerebrospinal Fluid Cell Counts Using the Sysmex XE-5000: Is It Time for New Reference Ranges? Am. J. Clin. Pathol. 2010, 134 (5), Boer, K.; Deufel, T.; Reinhoefer, M. Evaluation of the XE-5000 for the Automated Analysis of Blood Cells in Cerebrospinal Fluid. Clin Biochem. 2009, 42 (7 8), Aune, M. W.; Becker, J. L.; Brugnara, C.; et al. Automated Flow Cytometric Analysis of Blood Cells in Cerebrospinal Fluid: Analytic Performance. Am. J. Clin. Pathol. 2004, 121 (5), Buoro, S.; Ottomano, C.; Giustinetti, R.; et al. Evaluation of Sysmex XE-2100 for Enumaration and Differentiation of Cellular Elements in Peritoneal and Fluids. Clin. Biochem. 2013, 37 (4), Buoro, S.; Gustinetti, R.; Dominoni, P.; et al. Analytical Evaluation of Sysmex UF-1000i for Flow Cytometric Analysis of Peritoneal Fluid. Clin. Biochem. 2012, 45 (15), Buoro, S.; Ottomano, C.; Appassiti Esposito, S.; et al. Analytical and Clinical Evaluation of Sysmex UF1000i for Automated Screening of Cerebrospinal Fluids. J. Med. Biochem. 2013, 2, Walker, T. J.; Nelson, L. D.; Dunphy, B. W.; et al. Comparative Evaluation of the Iris iq200 Body Fluid Module with Manual Hemacytometer Count. Am. J. Clin. Pathol. 2009, 131 (3), Glasser, L.; Murphy, C. A.; Machan, J. T. The Clinical Reliability of Automated Cerebrospinal Fluid Cell Counts on the Beckman-Coulter LH750 and Iris iq200. Am. J. Clin. Pathol. 2009, 131 (1), Brown, W.; Keeney, M.; Chin-Yee, I.; et al. Validation of Body Fluid Analysis on the Coulter LH 750. Lab. Hematol. 2003, 9 (3), Lippi, G.; Cattabiani, C.; Benegiamo, A.; et al. Evaluation of the Fully Automated Hematological Analyzer Sysmex XE-5000 for Flow Cytometric Analysis of Peritoneal Fluid. J. Lab. Autom. 2013, 18 (3), Fleming, C.; Brouwer, R.; Lindemans, J.; et al. Validation of the Body Fluid Module on the New Sysmex XN-1000 for Counting Blood Cells in Cerebrospinal Fluid and Other Body Fluids. Clin. Chem. Lab. Med. 2012, 50 (10), Keuren, J. F.; Hoffmann, J. J.; Leers, M. P. Analysis of Serous Body Fluids Using the CELL-DYN Sapphire Hematology Analyzer. Clin. Chem. Lab. Med. 2013, 51 (6), De Smet, D.; Van Moer, G.; Martens, G. A.; et al. Use of the Cell-Dyn Sapphire Hematology Analyzer for Automated Counting of Blood Cells in Body Fluids. Am. J. Clin. Pathol. 2010, 133 (2), Danise, P.; Maconi, M.; Rovetti, A.; et al. Cell Counting of Body Fluids: Comparison between Three Automated Haematology Analysers and the Manual Microscope Method. Int. J. Lab. Hematol. 2013, 35 (6), Bourner, G.; De la Salle, B.; George, T.; et al. ICSH Guidelines for the Verification and Performance of Automated Cell Counters for Body Fluids. Int. J. Lab. Hematol. 2014, 36 (6), Runyon, B. A. AASLD Practice Guidelines Committee Management of Adult Patients with Ascites Due to Cirrhosis: An Update. Hepatology 2009, 49 (6),

Comparison between body fluid mode of Sysmex XN-9000 and optical microscopy for the counting of cells in ascitic fluid

Comparison between body fluid mode of Sysmex XN-9000 and optical microscopy for the counting of cells in ascitic fluid Original Article Page 1 of 5 Comparison between body fluid mode of Sysmex XN-9000 and optical microscopy for the counting of cells in ascitic fluid Vincenzo Roccaforte 1,2,3, Emanuela Zavaroni 1, Rosalba

More information

Literature List Body Fluids

Literature List Body Fluids Literature List Body Fluids Customer Information March 2018 Date: March 2018 Subject: Literature List body fluids Issued by: Scientific Customer Services Number: 180312 Note: Whether references are given

More information

A New Automated Technology for Cerebrospinal Fluid Cell Counts Comparison of Accuracy and Clinical Impact of GloCyte, Sysmex XN, and Manual Methods

A New Automated Technology for Cerebrospinal Fluid Cell Counts Comparison of Accuracy and Clinical Impact of GloCyte, Sysmex XN, and Manual Methods AJCP /ORIGINAL ARTICLE A New Automated Technology for Cerebrospinal Fluid Cell Counts Comparison of Accuracy and Clinical Impact of GloCyte, Sysmex XN, and Manual Methods Linda M. Sandhaus, MD, 1 Christine

More information

SIMTI Servizi Srl. Original Article

SIMTI Servizi Srl. Original Article Original Article Evaluation of nucleated red blood cell count by Sysmex XE-2100 in patients with thalassaemia or sickle cell anaemia and in neonates Sabrina Buoro 1, Mauro Vavassori 1, Silvia Pipitone

More information

Sysmex Journal International Vol.26 No.1 (2016)

Sysmex Journal International Vol.26 No.1 (2016) Sysmex Journal International Vol.26 No.1 (216) Performance Evaluation of the XN-55 Automated Hematology Analyzer Body Fluid Mode Considerations for Operational Conditions for Cell Counting with Cerebrospinal

More information

Performance evaluation of Celltac G: a new automated hematology analyzer

Performance evaluation of Celltac G: a new automated hematology analyzer Original Article Performance evaluation of Celltac G: a new automated hematology analyzer Masaaki Sugiyama 1, Tsukasa Kobayashi 1, Yuki Jisyage 2, Shigeko Yamamoto 2, Yutaka Nagai 2,3 and Hiroshi Kondo

More information

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer Hematopathology / EVALUATION OF THE ABBOTT CELL-DYN 4 HEMATOLOGY ANALYZER Evaluation of the Abbott CELL-DYN 4 Hematology Analyzer Ernesto Grimaldi, MD, and Francesco Scopacasa, PhD Key Words: Abbott CD

More information

The measurements of pleural fluid WBC count

The measurements of pleural fluid WBC count Variations in Pleural Fluid WBC Count and Differential Counts With Different Sample Containers and Different Methods* Barrett D. Conner, MD; Y. C. Gary Lee, MD; Paul Branca, MD, FCCP; Jeffrey T. Rogers,

More information

Customer Information Literature List White Blood Cells

Customer Information Literature List White Blood Cells Customer Information Literature List White Blood Cells Date: September 2015 Subject: Literature List White Blood Cells Issued by: Scientific Customer Services Number: 150901 Note: Whether references are

More information

10/30/2015. XN Case Studies: Every Picture Tells a Story

10/30/2015. XN Case Studies: Every Picture Tells a Story XN Case Studies: Every Picture Tells a Story Jill Crist MT(ASCP)Field Product Specialist OBJECTIVES Explain how scattergrams and histogram pictures can provide great insight into abnormal hematology samples

More information

Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. ABSTRACT

Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. ABSTRACT OriginalArticle Comparison of Leukocyte Differential Count by Beckman Coulter Unicel DxH800, Beckman Coulter LH780 and Sysmex XE-5000 Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. Department

More information

Lavender Top Management SUCCESS BEYOND FINANCES. Mission/ Vision

Lavender Top Management SUCCESS BEYOND FINANCES. Mission/ Vision Lavender Top Management Alfonso Ziccardi Lab.Operation Manager for AP/CP/Safety Officer New York Hospital Medical Center of Queens SUCCESS BEYOND FINANCES Mission/ Vision. VISION To successfully develop

More information

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes A. M. CENCI *1, M. MACONI *2, and B. CASOLARI *1 *1 Laboratory of

More information

Lavender Top Management SUCCESS BEYOND FINANCES

Lavender Top Management SUCCESS BEYOND FINANCES Lavender Top Management Alfonso Ziccardi Ass.Lab.Operation Manager for AP/CP/Safety Officer New York Hospital Medical Center of Queens SUCCESS BEYOND FINANCES 1 . VISION Mission/ Vision To successfully

More information

Screening for Urinary Tract Infection with the Sysmex UF-1000i Urine Flow Cytometer

Screening for Urinary Tract Infection with the Sysmex UF-1000i Urine Flow Cytometer JOURNAL OF CLINICAL MICROBIOLOGY, Mar. 2011, p. 1025 1029 Vol. 49, No. 3 0095-1137/11/$12.00 doi:10.1128/jcm.01669-10 Copyright 2011, American Society for Microbiology. All Rights Reserved. Screening for

More information

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation.

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Sysmex America White Paper One Nelson C. White Parkway,

More information

XN-SERIES. XN Technology and Case Studies

XN-SERIES. XN Technology and Case Studies XN Technology and Case Studies Karen Hoffman MT(ASCP) Clinical Applications Specialist OBJECTIVES Explain how scattergrams and histogram pictures can provide great insight into abnormal hematology samples

More information

Sysmex Educational Enhancement & Development

Sysmex Educational Enhancement & Development Sysmex Educational Enhancement & Development SEED-Africa Newsletter No 10 2011 Introducing automated body fluid analysis into the haematology laboratory Automated Body Fluid Analysis The purpose of this

More information

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa The Role of the FBC in clinical decision making History Examination Investigations Decision 70% FBC Laboratory

More information

BODY FLUID ANALYSIS. Synovial Fluid. Synovial Fluid Classification. CLS 426 Urinalysis and Body Fluid Analysis Body Fluid Lecture Session 1

BODY FLUID ANALYSIS. Synovial Fluid. Synovial Fluid Classification. CLS 426 Urinalysis and Body Fluid Analysis Body Fluid Lecture Session 1 BODY FLUID ANALYSIS Synovial Fluid Serous fluids the 3 P s Peritoneal Pleural Pericardial Cerebrospinal Fluid Karen Keller, MT(ASCP), SH Synovial Fluid Lubricant and sole nutrient source of joint. Normal

More information

A Look Into the Determination of Cell Morphology in Hematology in the 21 st Century. Ramon Simon-Lopez, MD Global Scientific Director Beckman Coulter

A Look Into the Determination of Cell Morphology in Hematology in the 21 st Century. Ramon Simon-Lopez, MD Global Scientific Director Beckman Coulter A Look Into the Determination of Cell Morphology in Hematology in the 21 st Century Ramon Simon-Lopez, MD Global Scientific Director Beckman Coulter Is cell morphology important? AML M7 CLL CD5 CD19 NHL

More information

MLAB 1211 Body Fluids BF Differentials - pretest

MLAB 1211 Body Fluids BF Differentials - pretest MLAB 1211 BF Differentials - pretest Instructor: Carolyn Ragland MLAB 1211 pre-test #2 Spring No Extra Materials Needed Name Date Instructions: Neatly write your answer in the space provided. Unless otherwise

More information

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags.

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags. WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags www.sysmex.com/us Introduction For many decades, physicians have relied on

More information

Sue Jung Kim, MD, MS; Yoonjung Kim, MD; Saeam Shin, MD; Jaewoo Song, MD, MS; Jong Rak Choi, MD, PhD

Sue Jung Kim, MD, MS; Yoonjung Kim, MD; Saeam Shin, MD; Jaewoo Song, MD, MS; Jong Rak Choi, MD, PhD Comparison Study of the Rates of Manual Peripheral Blood Smear Review From 3 Automated Hematology Analyzers, Unicel DxH 800, ADVIA 2120i, and XE 2100, Using International Consensus Group Guidelines Sue

More information

Analytic Performance of the iq200 Automated Urine Microscopy Analyzer and Comparison With Manual Counts Using Fuchs-Rosenthal Cell Chambers

Analytic Performance of the iq200 Automated Urine Microscopy Analyzer and Comparison With Manual Counts Using Fuchs-Rosenthal Cell Chambers Clinical Chemistry / PERFORMANCE OF THE IQ2 URINALYSIS SYSTEM Analytic Performance of the iq2 Automated Urine Microscopy Analyzer and Comparison With Manual Counts Using Fuchs-Rosenthal Cell Chambers David

More information

XN-SERIES. Hematology Case Studies: Every Picture Tells a Story

XN-SERIES. Hematology Case Studies: Every Picture Tells a Story Hematology Case Studies: Every Picture Tells a Story Jill Crist MT(ASCP)Field Product Specialist OBJECTIVES Explain how scattergrams and histogram pictures can provide great insight into abnormal hematology

More information

Applications of Flow Cytometry in Diagnostic Cytology of Body Cavity Fluids

Applications of Flow Cytometry in Diagnostic Cytology of Body Cavity Fluids Applications of Flow Cytometry in Diagnostic Cytology of Body Cavity Fluids Awtar Krishan, PhD. Professor, Department of Pathology University of Miami School of Medicine akrishan@med.miami.edu Beckman

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

Changes of hematological references depends on storage period and temperature conditions in rats and dogs

Changes of hematological references depends on storage period and temperature conditions in rats and dogs ISSN 1738-6055 (Print) ISSN 2233-7660 (Online) Lab Anim Res 2016: 32(4), 241-248 https://doi.org/10.5625/lar.2016.32.4.241 Changes of hematological references depends on storage period and temperature

More information

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S 111 Riv Med Lab - JLM, Vol. 4, N. 2, 2003 First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S H. Vandeputte a, D. Darre

More information

Successful flow cytometric immunophenotyping of body fluid specimens

Successful flow cytometric immunophenotyping of body fluid specimens Successful flow cytometric immunophenotyping of body fluid specimens Fiona E. Craig, MD Division of Hematopathology Mayo Clinic Arizona 2017 MFMER slide-1 Financial disclosure No conflicts 2017 MFMER slide-2

More information

Literature List White Blood Cells

Literature List White Blood Cells Literature List White Blood Cells Customer Information February 2018 Date: February 2018 Subject: Literature List White Blood Cells Issued by: Scientific Customer Services Number: 180202 Note: Whether

More information

Detail PRINCIPLE: Body fluids other than blood and urine will be analyzed according to their site of origin, and the providers specific orders.

Detail PRINCIPLE: Body fluids other than blood and urine will be analyzed according to their site of origin, and the providers specific orders. Subject Body Fluid Analysis - Affiliate Index Number Lab-8760 Section Laboratory Subsection Regional/Affiliates Category Departmental Contact Munson, Karen Last Revised 6/28/2017 References Required document

More information

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Thomas J. Spira, M.D. International Laboratory Branch Global AIDS Program Centers for Disease

More information

Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer

Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer Science Gene Gulati, PhD, Sean Cote, DO, Eric Behling, MD, William Kocher, MD

More information

Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis

Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis Hematopathology / LEFT SHIFT MODEL PREDICTS INFECTION Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis Michael J. Wile, MD, 1 Louis D. Homer, MD, PhD, 2 Stede Gaehler,

More information

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100 Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100 R. HERKLOTZ and A. R. HUBER Zentrum für Labormedizin, Kantonsspital Aarau, 5001 Aarau, Switzerland. The automated leukocyte(wbc)

More information

ACCME/Disclosures. Case 4 USCAP Pulmonary Panel Case 4 History

ACCME/Disclosures. Case 4 USCAP Pulmonary Panel Case 4 History Case 4 USCAP Pulmonary Panel 2016 Andrew Churg, MD Department of Pathology Vancouver General Hospital & University of British Columbia Vancouver, BC achurg@mail.ubc.ca. ACCME/Disclosures The USCAP requires

More information

Scoring System for Detecting Spurious Hemolysis in Anticoagulated Blood Specimens

Scoring System for Detecting Spurious Hemolysis in Anticoagulated Blood Specimens Original Article Laboratory Informatics Ann Lab Med 2015;35:341-347 http://dx.doi.org/10.3343/alm.2015.35.3.341 ISSN 2234-3806 eissn 2234-3814 Scoring System for Detecting Spurious Hemolysis in Anticoagulated

More information

Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation

Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation Laboratory Hematology 8:85-90 2002 Carden Jennings Publishing Co.. Ltd. Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation T. SCHISANO, L. VAN HOVE Abbott

More information

Cytological evaluation of effusion fluid with cell block technique and cytology smears among Sudanese patients

Cytological evaluation of effusion fluid with cell block technique and cytology smears among Sudanese patients EUROPEAN ACADEMIC RESEARCH Vol. IV, Issue 3/ June 2016 ISSN 2286-4822 www.euacademic.org Impact Factor: 3.4546 (UIF) DRJI Value: 5.9 (B+) Cytological evaluation of effusion fluid with cell block technique

More information

INFECTION/ INFLAMMATION

INFECTION/ INFLAMMATION HAEMATOLOGY OCTOBER 2017* WHITE PAPER INFECTION/ INFLAMMATION Novel haematological parameters for rapidly monitoring the immune system response Patients with inflammatory disease are common on hospital

More information

Immunohistochemistry on Fluid Specimens: Technical Considerations

Immunohistochemistry on Fluid Specimens: Technical Considerations Immunohistochemistry on Fluid Specimens: Technical Considerations Blake Gilks Dept of Pathology University of British Columbia, Vancouver, BC, Canada Disclosures None Learning Objectives At the end of

More information

3. Blood Cell Histograms:

3. Blood Cell Histograms: LECTURE MODULE 6c: ELECTRONIC CELL COUNTING PART III 3. Blood Cell Histograms: a. The Coulter cell counters today provides size distributions of the cellular content: 1) volume given in µm 3 or fl vs relative

More information

Automated Urinalysis Evaluation of the Sysmex UF-50

Automated Urinalysis Evaluation of the Sysmex UF-50 Clinical Chemistry / FLOW CYTOMETRY BASED URINALYSIS Automated Urinalysis Evaluation of the Sysmex UF-50 Hiroshi Okada, MD, 1 Yutaka Sakai, MD, 1 Gaku Kawabata, MD, 1 Masato Fujisawa, MD, 1 Soichi Arakawa,

More information

A HemoCue Novel Technology Product for 5-Part Differentiation of Leukocytes Point-of-Care

A HemoCue Novel Technology Product for 5-Part Differentiation of Leukocytes Point-of-Care A HemoCue Novel Technology Product for -Part Differentiation of Leukocytes Point-of-Care Stellan Lindberg, M Sc Medical Director HemoCue AB Kuvettgatan 1 SE-262 71 Ängelholm Sweden Stellan.Lindberg@hemocue.se

More information

Comparative Evaluation of Two Flowcytometric Analysers as Diagnostic Tools for the Automated Detection of Malaria

Comparative Evaluation of Two Flowcytometric Analysers as Diagnostic Tools for the Automated Detection of Malaria 82 Available online at www.annclinlabsci.org Comparative Evaluation of Two Flowcytometric Analysers as Diagnostic Tools for the Automated Detection of Malaria S. Mohapatra 1, J. C. Samantaray 1, S. Arulselvi

More information

Validation of the Efficacy of a Practical Method for Neutrophils Isolation from Peripheral Blood

Validation of the Efficacy of a Practical Method for Neutrophils Isolation from Peripheral Blood Validation of the Efficacy of a Practical Method for Neutrophils Isolation from Peripheral Blood JONATHAN DEGEL, MASIH SHOKRANI OBJECTIVE: The objectives of this study were to validate the Polymorphprep

More information

PointCare NOW TM Technical Note

PointCare NOW TM Technical Note PointCare NOW TM Technical Note Introduction to our Technology And Summary of Evaluation and Performance data 1 of 19 Introduction The PointCare NOW is a portable, and very fast, diagnostic instrument

More information

2008 CAP TODAY Q & A

2008 CAP TODAY Q & A 2008 CAP TODAY Q & A Q. How often should we document hematology competencies to ensure consistency of morphologic observations for manual differentials and fluids? Should we do this every six months or

More information

Pseudobasophilia in Pediatric Age Group

Pseudobasophilia in Pediatric Age Group Original Article DOI: 10.17354/ijss/2015/259 Pseudobasophilia in Pediatric Age Group Femela Muniraj 1, Vijay Amritraj 2 1 Assistant Professor, Department of Pathology, Chettinad Hospital & Research Institute,

More information

Sysmex Journal International Vol.9 No.2 (1999)

Sysmex Journal International Vol.9 No.2 (1999) Sysmex Journal International Vol.9 No. (999) Evaluation and Clinical Usefulness of the Automated Hematology Analyzer, Sysmex XE- TM Kazuto TSURUDA, Tomohiro TSUJI, Tetsuya USUI, Sachiko KITAJIMA, Atsuko

More information

Multiple studies are available concerning the use of

Multiple studies are available concerning the use of SPECIAL ARTICLE Evaluation of the Biosite Ò Quantitative Whole Blood D-dimer Assay and Comparison With the biomérieux VIDAS Ò D-dimer Exclusion Test Validation and Utility For Use in the Central Laboratory

More information

Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population

Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population Clinical immunology Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population EL BIETA GÓRSKA, URSZULA DEMKOW, ROMAN PIÑKOWSKI, BARBARA JAKUBCZAK,

More information

Evaluation of the QBC Star centrifugal three-part differential haematology system

Evaluation of the QBC Star centrifugal three-part differential haematology system ORIGINAL ARTICLE 67 Evaluation of the QBC Star centrifugal three-part differential haematology system O. ERHABOR, G. RICHARDSON, I. MOHAMMED, C. THORNTON, J. BARK, M. HURST, D. HAMER and P. KINSELLA Blood

More information

CAROL BRIGGS, DONNA GRANT, SAMUEL J. MACHIN INTRODUCTION. Department of Haematology, University College London Hospital, London, United Kingdom

CAROL BRIGGS, DONNA GRANT, SAMUEL J. MACHIN INTRODUCTION. Department of Haematology, University College London Hospital, London, United Kingdom ISLH Laboratory Hematology 7:75-80 2001 Carden Jennings Publishing Co., Ltd. Official Publication Comparison of the Automated Reticulocyte Counts and Immature Reticulocyte Fraction Measurements Obtained

More information

SEED HAEMATOLOGY. Medical statistics your support when interpreting results SYSMEX EDUCATIONAL ENHANCEMENT AND DEVELOPMENT APRIL 2015

SEED HAEMATOLOGY. Medical statistics your support when interpreting results SYSMEX EDUCATIONAL ENHANCEMENT AND DEVELOPMENT APRIL 2015 SYSMEX EDUCATIONAL ENHANCEMENT AND DEVELOPMENT APRIL 2015 SEED HAEMATOLOGY Medical statistics your support when interpreting results The importance of statistical investigations Modern medicine is often

More information

We noted a problem with the use of the hematology

We noted a problem with the use of the hematology Standard Article J Vet Intern Med 2017;31:1708 1716 Feline Differential Leukocyte Count with ProCyte Dx: Frequency and Severity of a Neutrophil-Lymphocyte Error and How to Avoid It H.W. Tvedten, V. Andersson,

More information

Detection of important abnormalities of the differential

Detection of important abnormalities of the differential J Clin Pathol 1989;42:772-776 Detection of important abnormalities of the differential count using the Coulter STKR blood counter D F BARNARD, S A BARNARD, A B CARTER, S J MACHN, K G PATTERSON, A YARDUMAN

More information

Brief Communication Diagnostic Hematology

Brief Communication Diagnostic Hematology Brief Communication Diagnostic Hematology Ann Lab Med 2016;36:244-249 http://dx.doi.org/10.3343/alm.2016.36.3.244 ISSN 2234-3806 eissn 2234-3814 Establishment of Age- and Gender-Specific Reference Ranges

More information

Serous Effusions. Spasenija Savic Prince, MD Pathology, University Hospital Basel, Switzerland

Serous Effusions. Spasenija Savic Prince, MD Pathology, University Hospital Basel, Switzerland Serous Effusions Spasenija Savic Prince, MD Pathology, University Hospital Basel, Switzerland Serous membrane Body cavities: Pleural Pericardial Peritoneal Effusion = Excess of fluid 80% Benign 20% Malignant

More information

HumaCount 5D. Outperforming 5-part Hematology System. CoreLab DX. Hematology

HumaCount 5D. Outperforming 5-part Hematology System. CoreLab DX. Hematology HumaCount 5D Outperforming 5-part Hematology System Direct Capillary Process by OptimalCount Technology Distinct 5-part Differentiation Definite Immature Cell Count (LIC, ALY) CoreLab DX Hematology 5-part

More information

Automated Hematology Analyzer. Differentiate with the XT-4000i

Automated Hematology Analyzer. Differentiate with the XT-4000i XT-4000i Automated Hematology Analyzer Differentiate with the XT-4000i Advanced Technology Solutions to Meet Your Lab s Needs Even with these challenges, the need for hematology testing has remained steady

More information

Flow cytometry leukocyte differential : a critical appraisal

Flow cytometry leukocyte differential : a critical appraisal Flow cytometry leukocyte differential : a critical appraisal Francis Lacombe Flow cytometry department University Hospital of Bordeaux, Pessac, France francis.lacombe@chu-bordeaux.fr 2008 HORIBA ABX, All

More information

ADx Bone Marrow Report. Patient Information Referring Physician Specimen Information

ADx Bone Marrow Report. Patient Information Referring Physician Specimen Information ADx Bone Marrow Report Patient Information Referring Physician Specimen Information Patient Name: Specimen: Bone Marrow Site: Left iliac Physician: Accession #: ID#: Reported: 08/19/2014 - CHRONIC MYELOGENOUS

More information

ACCESS hstni SCIENTIFIC LITERATURE

ACCESS hstni SCIENTIFIC LITERATURE ACCESS hstni SCIENTIFIC LITERATURE 2017 2018 Table of contents Performance Evaluation of Access hstni A critical evaluation of the Beckman Coulter Access hstni: Analytical performance, reference interval

More information

Investigating the recheck rules for urine analysis in children

Investigating the recheck rules for urine analysis in children Investigating the recheck rules for urine analysis in children Y.M. He*, S.W. Yao*, Y.J. Huang, B.S. Liang and H.Y. Liu Clinical Laboratory, Guangzhou Women and Children s Medical Center, Guangzhou Medical

More information

Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How?

Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How? Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How? Dennis B. DeNicola, DVM, PhD, DACVP Chief Veterinary Educator IDEXX Laboratories, Inc. Westbrook, ME USA Adjunct Professor of Veterinary

More information

Evaluation of the feasibility of the VACUETTE Urine CCM tube for microbial testing of urine samples

Evaluation of the feasibility of the VACUETTE Urine CCM tube for microbial testing of urine samples Evaluation of the feasibility of the VACUETTE Urine CCM tube for microbial testing of urine samples Background The VACUETTE Urine CCM tube is for the collection, transport and storage of urine samples

More information

The LaboratoryMatters

The LaboratoryMatters Laboratory Medicine Newsletter for clinicians, pathologists & clinical laboratory technologists. A Initiative. Complete Blood Count This issue highlights: CBC, while ubiquitous, is an excellent diagnostic

More information

BACKGROUND. Absolute neutrophil counts (ANCs) and absolute phagocyte counts

BACKGROUND. Absolute neutrophil counts (ANCs) and absolute phagocyte counts 2681 Utility of Automated Counting to Determine Absolute Neutrophil Counts and Absolute Phagocyte Counts for Pediatric Cancer Treatment Protocols Nobuko Hijiya, M.D. 1,2 Mihaela Onciu, M.D. 3 Scott C.

More information

HSCT MANAGEMENT WHITE PAPER. Managing stem cell apheresis effectively

HSCT MANAGEMENT WHITE PAPER. Managing stem cell apheresis effectively HAEMATOLOGY JANUARY 2017 WHITE PAPER HSCT MANAGEMENT Managing stem cell apheresis effectively Haematopoietic stem cell transplantation Haematopoietic stem cell transplantation (HSCT) is a treatment that

More information

Interpreting Hematology Scatter-Plots; One Cancer Center s Keys to Seeing the BIG Picture

Interpreting Hematology Scatter-Plots; One Cancer Center s Keys to Seeing the BIG Picture Interpreting Hematology Scatter-Plots; One Cancer Center s Keys to Seeing the BIG Picture Barbara L. Burch, MHA MT (ASCP) Laboratory Manager New York University Clinical Cancer Center Disclosure Ms Burch

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: BD FACSCount Instrument System with FACSCount Control Kit and BD FACSCount Reagent Kit (Absolute CD4+, CD8+, and CD3+ Counts) Number:

More information

XN series. Case interpretation. Gebruikersdag Vlaanderen- 6 oktober 2016

XN series. Case interpretation. Gebruikersdag Vlaanderen- 6 oktober 2016 XN series Case interpretation Gebruikersdag Vlaanderen- 6 oktober 2016 Fluorescence flow cytometry RET channel PLT-F channel WDF channel WPC channel WNR channel Case 1 Case 1: Initial measurement Patient

More information

FEAR. DEPENDS ON WHAT YOU DREAD THE MOST

FEAR. DEPENDS ON WHAT YOU DREAD THE MOST FEARLESS FLUIDS FEAR. DEPENDS ON WHAT YOU DREAD THE MOST YUCKY CELL COUNTS SCARY CELLS WHAT WE LL COVER. CSF, SEROUS, SYNOVIAL MANUAL CELL COUNTS A LITTLE, VERY LITTLE AUTOMATED CELL COUNTS AMR, CRR, METHOD

More information

Utilizing Sysmex RET He to Evaluate Anemia in Cancer Patients

Utilizing Sysmex RET He to Evaluate Anemia in Cancer Patients Utilizing Sysmex RET He to Evaluate Anemia in Cancer Patients Ellinor I. Peerschke, Ph.D., F.A.H.A. Vice Chair, Laboratory Medicine Chief, Hematology & Coagulation Laboratory Services Memorial Sloan Kettering

More information

Sample stability for complete blood cell count using the Sysmex XN haematological analyser

Sample stability for complete blood cell count using the Sysmex XN haematological analyser ORIGINAL ARTICLE Sample stability for complete blood cell count using the Sysmex XN haematological analyser Massimo Daves 1, Elmar M. Zagler 1, Roberto Cemin 2, Flora Gnech 1, Alexandra Joos 1, Stefan

More information

Value of antimesothelioma HBME 1 in the diagnosis of inflammatory and malignant pleural effusions

Value of antimesothelioma HBME 1 in the diagnosis of inflammatory and malignant pleural effusions Romanian Journal of Morphology and Embryology 2006, 47(4):351 355 ORIGINAL PAPER Value of antimesothelioma HBME 1 in the diagnosis of inflammatory and malignant pleural effusions LILIANA MOCANU 1), ANCA

More information

Update on Thyroid FNA The Bethesda System. Shikha Bose M.D. Associate Professor Cedars Sinai Medical Center

Update on Thyroid FNA The Bethesda System. Shikha Bose M.D. Associate Professor Cedars Sinai Medical Center Update on Thyroid FNA The Bethesda System Shikha Bose M.D. Associate Professor Cedars Sinai Medical Center Thyroid Nodules Frequent occurrence Palpable: 4-7% of adults Ultrasound: 10-31% Majority benign

More information

Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities

Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities JOANNE CORNBLEET, M.D. AND SALLY KESSINGER, MT(ASCP)SH The authors evaluated the Coulter 15 S-Plus

More information

LYDIA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE, M.D., SANDRA WINGRELD, MT(ASCP), AND EDWARD KEYSER, MT(ASCP)

LYDIA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE, M.D., SANDRA WINGRELD, MT(ASCP), AND EDWARD KEYSER, MT(ASCP) Laboratory Evaluation of Differential White Blood Cell Count nformation from the S-Plus V and H-1 in Patient Populations Requiring Rapid "Turnaround" Time LYDA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE,

More information

Flow Cytometric White Blood Cell Differential Using CytoDiff is Excellent for Counting Blasts

Flow Cytometric White Blood Cell Differential Using CytoDiff is Excellent for Counting Blasts Original Article Diagnostic Hematology Ann Lab Med 215;35:28-34 http://dx.doi.org/1.3343/alm.215.35.1.28 ISSN 2234-386 eissn 2234-3814 Flow Cytometric White Blood Cell Differential Using is Excellent for

More information

Andrea Tessari Microbiology Unit, Hospital of Rovigo, ULSS 18 Rovigo (Italy)

Andrea Tessari Microbiology Unit, Hospital of Rovigo, ULSS 18 Rovigo (Italy) URINARY SEDIMENT: STILL AN IMPORTANT DIAGNOSTIC TOOL International Symposium Rome, December 4 5 2014 The Application in Microbiology Andrea Tessari Microbiology Unit, Hospital of Rovigo, ULSS 18 Rovigo

More information

ADVANCED HAEMATOLOGY BATTLE OF THE BANDS. Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND

ADVANCED HAEMATOLOGY BATTLE OF THE BANDS. Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND ADVANCED HAEMATOLOGY BATTLE OF THE BANDS Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND The identification of immature neutrophils (bands, metamyelocytes,

More information

Combined efficacy of pleural fluid lymphocyte neutrophil ratio and pleural fluid adenosine deaminase for the diagnosis of tubercular pleural effusion

Combined efficacy of pleural fluid lymphocyte neutrophil ratio and pleural fluid adenosine deaminase for the diagnosis of tubercular pleural effusion Original article: Combined efficacy of pleural fluid lymphocyte neutrophil ratio and pleural fluid adenosine deaminase for the diagnosis of tubercular pleural effusion Kavita S Kore, Guruprasad Antin,

More information

Validation of Reagent Strips, Comparing with Automated and Manual Cell Counts for the Diagnosis of Spontaneous Bacterial Peritonitis ABSTRACT

Validation of Reagent Strips, Comparing with Automated and Manual Cell Counts for the Diagnosis of Spontaneous Bacterial Peritonitis ABSTRACT Original Article 91 Validation of Reagent Strips, Comparing with Automated and Manual Cell Counts for the Diagnosis of Spontaneous Bacterial Peritonitis Limmathurotsakul D 1 Bhokaisawan N 2 Thaimai P 1

More information

FEARLESS FLUIDS WHAT WE LL COVER. DEPENDS ON WHAT YOU DREAD THE MOST CSF, SEROUS, SYNOVIAL MANUAL CELL COUNTS A LITTLE, VERY LITTLE

FEARLESS FLUIDS WHAT WE LL COVER. DEPENDS ON WHAT YOU DREAD THE MOST CSF, SEROUS, SYNOVIAL MANUAL CELL COUNTS A LITTLE, VERY LITTLE FEARLESS FLUIDS DEPENDS ON WHAT YOU DREAD THE MOST YUCKY CELL COUNTS SCARY CELLS WHAT WE LL COVER. CSF, SEROUS, SYNOVIAL MANUAL CELL COUNTS A LITTLE, VERY LITTLE AUTOMATED CELL COUNTS WHAT S IMPORTANT

More information

Clinical Utility. Clinical Utility of the Automated IG-Count Xavier Troussard

Clinical Utility. Clinical Utility of the Automated IG-Count Xavier Troussard 5. Clinical Utility Clinical Utility of the Automated IG-Count Xavier Troussard Introduction Quantitative analysis of immature granulocytes (IG) is useful for the optimal clinical and biological management

More information

Do Your Flow Cytometric LDTs. Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine

Do Your Flow Cytometric LDTs. Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine Do Your Flow Cytometric LDTs Conform to the ICSH ICCS Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine How should LDTs be validated? Accuracy Specificity Sensitivity

More information

Introduction. 23 rd Annual Seminar in Pathology. FLUIDS, Part 1. Pittsburgh, PA Gladwyn Leiman UVMMC, VT

Introduction. 23 rd Annual Seminar in Pathology. FLUIDS, Part 1. Pittsburgh, PA Gladwyn Leiman UVMMC, VT 23 rd Annual Seminar in Pathology Pittsburgh, PA Gladwyn Leiman UVMMC, VT FLUIDS, Part 1 "Blue walls", Claudia Hansen, 2009 Introduction o Challenging to everyone o Almost any benign or malignant process

More information

SPECIMEN PREPARATION AND ADEQUACY OF THE MATERIAL

SPECIMEN PREPARATION AND ADEQUACY OF THE MATERIAL SPECIMEN PREPARATION AND ADEQUACY OF THE MATERIAL Guido FADDA, MD, MIAC Head, Cytopathology Section Department of Anatomic Pathology and Laboratory Medicine Agostino Gemelli School of Medicine and Hospital

More information

Efficient assessment of peripheral blood lymphocytosis in adults: developing new thresholds for blood smear review by pathologists

Efficient assessment of peripheral blood lymphocytosis in adults: developing new thresholds for blood smear review by pathologists Clin Chem Lab Med 24; aop Victor Tseng, Adam S. Morgan, Catherine P. Leith and David T. Yang * Efficient assessment of peripheral blood lymphocytosis in adults: developing new thresholds for blood smear

More information

Customer Information Literature List Platelets

Customer Information Literature List Platelets Customer Information Literature List Platelets Date: August 2015 Subject: Literature List Platelets Issued by: Scientific Customer Services Number: 150803 Note: Whether references are given in British

More information

Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana

Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana IS THERE STILL A ROLE FOR THE MANUAL BLOOD COUNT? Indication for examination

More information

M. Velizarova, T. Yacheva and K. Tzatchev Department of Clinical Laboratory and Clinical Immunology, MU Sofia

M. Velizarova, T. Yacheva and K. Tzatchev Department of Clinical Laboratory and Clinical Immunology, MU Sofia 50, 48, 2012, 3, ADVIA 2120.,.., ANALYSIS OF DIAGNOSTIC RED CELL CYTOGRAMS, GENERATED BY THE ADVIA 2120 HEMATOLOGY ANALYZER M. Velizarova, T. Yacheva and K. Tzatchev Department of Clinical Laboratory and

More information

Performance Evaluation of the CellaVision DM96 System WBC Differentials by Automated Digital Image Analysis Supported by an Artificial Neural Network

Performance Evaluation of the CellaVision DM96 System WBC Differentials by Automated Digital Image Analysis Supported by an Artificial Neural Network Hematopathology / DIGITAL IMAGE ANALYSIS IN HEMATOLOGY Performance Evaluation of the CellaVision DM96 System WBC Differentials by Automated Digital Image Analysis Supported by an Artificial Neural Network

More information

Hematology & Coagulation Practicum Objectives CLS - 647

Hematology & Coagulation Practicum Objectives CLS - 647 Hematology & Coagulation Practicum Objectives CLS - 647 The following objectives are to be completed by the student for successful completion of this clinical rotation. The objectives within the psychomotor

More information