X-Ray Guided Robotic Radiosurgery for Solid Tumors

Size: px
Start display at page:

Download "X-Ray Guided Robotic Radiosurgery for Solid Tumors"

Transcription

1 X-Ray Guided Robotic Radiosurgery for Solid Tumors Mohan Bodduluri Accuray Incorporated 570 Del Rey Avenue Sunnyvale, CA USA and J. M. McCarthy Department of Mechanical and Aerospace Engineering University of California Irvine, CA Abstract This paper presents an overview of the x-ray guided robotic radiosurgery system that has been developed for the ablation of solid tumors. A robot mounted linear accelerator is directed through a sequence of positions and orientations designed to deliver high radiation dosages focussed at a specific location. Patient movement during treatment is identified by stereo x-ray measurements and the robotic system adjusts the linear accelerator prior to the delivery of radiation at each location. The result is accurate delivery without rigid fixation of the tumor relative to the treatment system. Introduction For almost 20 years precisely administered large doses of radiation has provided a non-invasive surgical treatment for brain tumors, (Leksell 1951, Larsson et al. 1958). Careful measurements in three dimensions are required to locate a tumor relative to the reference frame of the treatment device. Magnetic resonance imaging (MRI) and computed tomography (CT) scans provide the ability to precisely locate a tumor relative to skeletal landmarks, or relative to implanted fiducial.markers. The challenge is to determine the location of these landmarks relative to the treatment system. One strategy for locating these landmarks is to attach a stereotactic frame to the skeleton of the patient which is then locked in place relative to the treatment device. This allows the tumor to be located at a center of a sphere of radiation sources. The intersection of the beams from these sources results in a high dose of radiation that treats the tumor directly with minimal impact on the surrounding tissue. The approach described here is called frameless radiosurgery (Adler 1993). It uses stereo x- ray imaging to regularly measure the location of landmarks relative to the treatment device. An

2 industrial robot is used to move a small linear accelerator through a preplanned path to focus an array of beams from many different directions. The robot and imaging system can work together to adjust the focus of these treatment beams in order to accommodate movement of the patient. See Chenery et al. (1998) for a description of initial clinical results, and Murphy et al. (2000) for a discussion of the feasibility of using this approach to treat tumors in the spine and pancreas. In this paper, we describe the overall system and future research directions. Figure 1. A robot mounted linac is guided by a stereo x-ray imaging system. Ceiling mounted x-ray sources illuminate two x-ray cameras mounted orthogonally. The CyberKnife Stereotactic Robotic-radiosurgery System consist of four major components: i) a radiation source mounted on an industrial manipulator; ii) a stereo x-ray imaging system; iii) a patient positioning system (or couch); and iv) the treatment planning and delivery software system. See Figure 1. The Radiation Source The radiation source consists of a pulsed electron gun combined with a waveguide that uses microwave radiation to accelerate the electrons into a tungsten target. The result is an x-ray beam that can be collimated into a diameter as small as 5mm. This radiation source is known as

3 a linear accelerator and is commonly called a linac. The weight of the linac is approximately 100kg and must be manipulated by the robot to accommodate tumor movement. For this reason, an industrial robot such as the Fanuc 420 or Kuka 210 is used for these systems. Figure 2. A fixed set of beam directions are identified on a sphere centered on the tumor. The trajectories are planned relative to the sphere which moves with the patient. The Stereo X-ray Imaging System The imaging system consists of two x-ray sources mounted to the ceiling of the treatment room and a pair of x-ray cameras mounted orthogonally in a v -shaped frame and fixed to the floor. Both CCD fluoroscopes and amorphous silicon detectors have been used to provide 20cm x 20cm x-ray images. The location of the sources, cameras and patient support system results in a spatial resolution of approximately 0.5mm of patient movement per pixel. The x-ray sources and cameras are precisely located within the room and relative to the robot to ensure the accuracy of the tracking algorithms.

4 Two cameras located perpendicular to each other provide two images of the skeletal structure surrounding the tumor. The tracking algorithm uses these images to locate the tumor in three dimensions relative to the robot with sub-millimeter accuracy. The Patient Positioning System The patient positioning system (couch) supports the patient relative to the imaging system. Appropriate location of the tumor and landmarks in the field of view of the cameras and patient positioning appropriate x-ray illumination enhances tracking performance. However, once in position the couch remains fixed throughout treatment. Patient movement during treatment is accommodated by adjusting the robot manipulator. The Planning and Delivery Software A Silicon Graphics Octane II is used to provide software integration to the system. It supports both the treatment planning and the treatment delivery software. The system is designed to have a fixed set of nodes on a sphere from which treatment beams can be delivered, Figure 2. A treatment plan consists of a dose rate and duration at each node, Figure 3.. The robot moves sequentially through each of the nodes and delivers the required dose. Figure 3. The treatment planning translates a desired dose distribution into robot path and linac beam plan.

5 Tracking Patient Movement The imaging system of CyberKnife consists of two diagnostic x-ray sources mounted in the ceiling, two amorphous silicon cameras mounted in the floor, Figure 4. The cameras are orthogonal to each other, the sources are aimed at the cameras directly. The distance between the camera and patient is approximately 65 cm, which eliminates most of the scatter radiation. The distance between the patient and the x-ray source is approximately 260 cm, which provides the robot ample access to the patient during the patient treatment. Digitally Reconstructed Radiographs Before the treatment begins, the system generates a set of digitally reconstructed radiographs (DRRs) for the patient. A DRR can be viewed as an x-ray image generated from a CT scan of the patient, using a computer model of the x-ray source and camera. By incrementing the alignment of the CT scan through the entire range of expected patient movements, an array of synthetic x-ray images is obtained for matching to a live image. In the present implementation, DRRs cover ±10mm range with step size of 2.5mm. Image Matching The patient is positioned so that real time live images match the nominal DRR defined as zero movement. Movement of the patient, and therefore of the tumor away from this nominal location is detected by matching the x-ray images of the two cameras to the precomputed array of DRRs. Standard image processing techniques are used to subtract the images and obtain a measure of the difference of the images, Figure 5. An interpolating surface is fit to the data to obtain a position estimate to less than 1mm error. Figure 4. Two x-ray sources and a cameras provide orthogonal views of the tumor. Tracking is achieved by matching the x-ray images to a set of synthetic images that correspond to various movements

6 Figure 5. Synthetic images and real x-ray images are compared from cameras A and B to compute patient position. System Calibration A system centerpoint called the isocenter is identified in order to calibrate the installation of the imaging system and the robot in the treatment facility. The system accuracy from treatment planning through delivery measured by exposing a film cube in a head phantom, and is found to be within 2mm (Murphy and Cox 1996). Imaging System Calibration The alignment of the cameras and location of the sources places the image of the isocenter within 5mm of the center of both cameras. A calibration process identifies variations in location over the face of the camera to account for small variations. The imaging system s accuracy is less than 1 mm. Robot Calibration The installation of the robot determines the orientation of the user frame which has its origin at the isocenter. The tool frame is identified by measuring the position and orientation of the line determined by a laser which is mounted coaxial to the therapeutic beam. Further calibration involves an automated process in which the robot illuminates an isocrystal and searches for the position and orientation that provides the maximum signal. A correction look-up table is maintained by the system. The residual error of the robot positioning system is less than 0.5 mm.

7 The Treatment Process A typical patient treatment process begins with a CT scan which is generally performed a day or two prior to the scheduled radiation delivery. The medical physicist and physician register the patient on the system, import the CT scans, and generate a treatment plan. On the day of the treatment, the patient is positioned and the treatment plan is executed. A normal time for radiation delivery is approximately one hour. Treatment Planning System The treatment planning system provides an interactive environment that is used to: 1. Display the CT scan in various formats, such as axial, sagittal, coronal, as well as three dimensional surface reconstructions; 2. Outline tumors and other critical structures, such as eyes, brain stem, optic chiasm for identification to the planning system; 3. Direct the beams at the tumor while avoiding critical structures. The following methods are available for directing the beams: a. All the beams are directed at the same point (isocentric), b. Subsets of beams are directed at various points (multi-center), c. The beams are directed so that they are uniformly distributed throughout the tumor conformal shape) d. Manual specification of beam directions. 4. Adjust the amount of radiation for each of the beams. The methods for specifying the radiation are: a. Forward planning the user selects the doses for each beam b. Inverse planning the user specifies dose requirements in the tumor, and maximum allowed dose in the critical structures, and the planning system calculates dose for each of the beams. 5. Calculate and display the resulting dose calculations in terms of isodose curves, color washes, three-dimensional isodose clouds, and dose volume histograms (DVHs). 6. Simulate the treatment. This allows verification of the dose distributions. Figure 6. shows an example of the results of a simulated treatment plan.

8 Figure 6. The radiation dose distributions are planned on the CT slices directly. Treatment Delivery On the day of the treatment the patient lies on the couch of the patient positioning system. The treatment plan is selected and the system assists the operator in aligning the patient in the center of the imaging system. This is done by acquiring x-ray images and evaluating offsets from the nominal position. Once the patient is aligned, the treatment begins. During the patient treatment, robot moves the linear accelerator through each beam position and orientation called a node. At each node, a pair of orthogonal x-ray images of the patient is obtained and the movement of the patient is calculated. This information is transmitted to the robot, which compensates for the patient movement before the therapeutic beam is turned on. This process is repeated through all the nodes to complete an entire treatment. Future Research Currently imaging accuracy and tracking algorithms are the primary focus of research. Imaging accuracy depends on the quality and performance of the x-ray sources, cameras, and room calibration. The tracking algorithms depend on quality of the CT scan, the DRR generation, and image processing techniques. Six Dimensional Tracking Current tracking algorithms focus on translational movement and assume that changes in orientation due to patient movement are small. However, in the initial alignment stage, larger rotational movements can provide the convenient treatment configurations. In addition, there are

9 tumor locations where small rotations result in large translations. For this reason, a tracking algorithm that identifies full six-degree of freedom movement has been implemented, and refinements are the focus of continuing research. Fiducial Tracking Recently, a capability has been released for tracking tumors that are in locations where the skeletal structure is not present, or not visible enough, to identify the movement of the patient. In this situation metallic landmarks called fiducials are implanted and CT scans generated. The x-ray system can compare the location of the fiducials to the synthesized DRRs and determine patient movement. The size, shape and composition of these fiducials are being refined to improve performance in certain anatomic locations. Compensating for Respiratory Motion The body and internal organs of the patient move with the breathing cycle. Therefore tumors in the lungs, liver, and other structures can move during treatment. One strategy is to have the patient hold a breadth while a beam of radiation is applied. Another is to track the patient movement during the application of the radiation and compensate for the respiratory movement. This places severe demands on the quality of the imaging, tracking and robot system. This motion-tracking feature is an important direction for research. Conclusion This paper presents an overview of a robotic system for frameless stereotactic radiosurgery. It relies on stereo x-ray imaging to identify the position of the patient and tumor relative to the treatment system. The accuracy of this process has been shown to be within the current specifications of radiosurgery devices. The elimination of rigid fixation framework to locate the tumor relative to the treatment device increases patient comfort and reduces the procedure to an outpatient treatment. This technology is increasingly finding a range of clinical applications. To date over 2000 patients have been treated with robotic radiosurgery around the world. See Figure 7. Acknowledgments The authors acknowledge the contributions of James Wang, John Dooley, Greg Glosser, Uisik Ro, Gopi Kuduvalli, Andy Irish, Shehrzad Qureshi, Bill Main, Chris Lee, and Eric Earnst to the successful performance of the CyberKnife system.

10 Spine 6% Prostate <1% Pancreas 1% Lung 4% Primary Intracranial Tumor 15% Benign/ Functional Tumor 43% Metastatic Brain Tumor 31% Figure 7 The distribution of patients treated by the CyberKnife in the US. References J. R. Adler, ``Frameless Radiosurgery,'' in, S. J. Goetsch and A. A. F. DeSalles (ed): Sterotactic Surgery and Radiosurgery, Medical Physics Publishing, Wisconsin, 17: , J. R. Adler, M. J. Murphy, S. D. Chang, and S. L. Hancock, ``Image-guided Robotic Radiosurgery,'' Neurosurgery, 44(6): , June, S. G. Chenery, H. H. Chehabi, D. M. Davis, J. R. Adler, ``The CyberKnife: Beta System Description and Initial Clinical Results,'' Journal of Radiosurgery, 1(4): , B. Larsson, L. Leksell, B. Rexed, ``The high energy proton beam as a neurosurgical tool,'' Nature, 182:1222-3, L. Leksell, ``The stereotaxic method and radiosurgery of the brain,'' Acta Chir Scand, 102:316-19, M. J. Murphy and R. S. Cox, ``The accuracy of dose localization for an image-guided frameless radiosugery system,'' Medical Physics, 23(12): , M. J. Murphy, J. R. Adler, M. Bodduluri, J. Dooley, K. Forster, J. Hai, Q. Le, G. Luxton, D. Martin, J. Poen, ``Image-Guided Radiosurgery for the Spine and Pancreas,'' Computer Aided Surgery, 5: , A. Schweikard, J. R. Adler, J. C. Latombe, ``Motion Planning in Stereotaxic Radiosurgery,'' Proc. International Conference on Robotics and Automation, 9: IEEE Press, 1993.

11 A. Schweikard, R. Z. Tombropoulos, J. R. Adler, J. C. Latombe, ``Treatment Planning for a Radiosurgical System with General Kinematics,'' Proc. International Conference on Robotics and Automation, 10: IEEE Press, S. B. Tatter, ``History of Stereotactic Radiosurgery,'' MGH Neurological Service, 1998.

Stereotactic Radiosurgery. Extracranial Stereotactic Radiosurgery. Linear accelerators. Basic technique. Indications of SRS

Stereotactic Radiosurgery. Extracranial Stereotactic Radiosurgery. Linear accelerators. Basic technique. Indications of SRS Stereotactic Radiosurgery Extracranial Stereotactic Radiosurgery Annette Quinn, MSN, RN Program Manager, University of Pittsburgh Medical Center Using stereotactic techniques, give a lethal dose of ionizing

More information

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017 CyberKnife Technology in Ablative Radiation Therapy Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017 Objectives Components and work flow of CyberKnife Motion management of CyberKnife

More information

Unrivaled, End-to-End

Unrivaled, End-to-End PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

SRS Uncertainty: Linac and CyberKnife Uncertainties

SRS Uncertainty: Linac and CyberKnife Uncertainties SRS Uncertainty: Linac and CyberKnife Uncertainties Sonja Dieterich, PhD Linac/CyberKnife Technological Uncertainties 1 Linac Mechanical/Radiation Isocenters Depuydt, Tom, et al. "Computer aided analysis

More information

A Patient s Guide to SRS

A Patient s Guide to SRS A Patient s Guide to SRS Stereotactic Radiosurgery 230 Nebraska St. Sioux City, IA 51101 NOTES 230 Nebraska St. Sioux City, IA 51101 Contents page Introduction 1 SRS and how it works 2 The technology involved

More information

02 CyberKnife: Treatment Delivery

02 CyberKnife: Treatment Delivery TREATMENT DELIVERY CyberKnife Treatment Delivery System The CyberKnife System is the first and only robotic radiosurgery system to offer highly precise and customizable, non-surgical treatment options

More information

The CyberKnife Robotic Radiosurgery System in 2010

The CyberKnife Robotic Radiosurgery System in 2010 Technology in Cancer Research and Treatment ISSN 1533-0346 Volume 9, Number 5, October 2010 Adenine Press (2010) The CyberKnife Robotic Radiosurgery System in 2010 www.tcrt.org This review provides a complete

More information

Cyberknife Stereotactic Treatment

Cyberknife Stereotactic Treatment Cyberknife Stereotactic Treatment Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA DISCLAIMER: I am not affiliated with any vendor and did not receive any financial support from any vendor.

More information

S Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M

S Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M s y n c h r o n y r e s p i r a t o r y t r a c k i n g s y s t e m S Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M The Synchrony System tracks respiration in real time and automatically

More information

ph fax

ph fax This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY

More information

Work partially supported by VisionRT

Work partially supported by VisionRT Work partially supported by VisionRT Background of frameless intracranial stereotactic radiosurgery UCSD SRS/SRT procedure Clinical Results Summary Total prescribed doses : order of 10 50 Gy Planning targets

More information

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

SBRT REQUIRES: STEREOTACTIC BODY RADIOTHERAPY STEREOTACTIC BODY RADIOTHERAPY (SBRT) (SBRT) What s s in a name? Stereotactic Body Radiotherapy

SBRT REQUIRES: STEREOTACTIC BODY RADIOTHERAPY STEREOTACTIC BODY RADIOTHERAPY (SBRT) (SBRT) What s s in a name? Stereotactic Body Radiotherapy INTRODUCTION TO STEREOTACTIC BODY RADIOTHERAPY: (I) Physics and Technology (II) Clinical Experience & (III) Radiobiological Considerations and Future Directions Stanley H. Benedict, Ph.D., Danny Song,

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 37-013 GARD Users Manual March 2005 Manual No. 37-013-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective

More information

Intensity-Modulated and Image- Guided Radiation Treatment. Outline. Conformal Radiation Treatment

Intensity-Modulated and Image- Guided Radiation Treatment. Outline. Conformal Radiation Treatment Intensity-Modulated and Image- Guided Radiation Treatment J. Daniel Bourland, PhD Professor Departments of Radiation Oncology, Physics, and Biomedical Engineering Wake Forest University School of Medicine

More information

02 CyberKnife: System Overview

02 CyberKnife: System Overview SYSTEM OVERVIEW Our hands push the technology. Your hands push the results. The CyberKnife System is the first and only robotic radiosurgery system to offer highly precise and customizable, non-surgical

More information

XSIGHT LUNG TRACKING SYSTEM

XSIGHT LUNG TRACKING SYSTEM XSIGHT LUNG TRACKING SYSTEM CYBERKNIFE SYSTEM HIGHLIGHTS Continual image guidance Without the need for staff intervention or treatment interruption, the CyberKnife s revolutionary image guidance technology

More information

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic

More information

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics

More information

Brain Tumor Radiosurgery. Gabor Fichtinger, PhD

Brain Tumor Radiosurgery. Gabor Fichtinger, PhD Brain Tumor Radiosurgery Gabor Fichtinger, PhD Surgical CAD/CAM Preoperative Computerassisted planning Surgical Intraoperative Update Model Update Plan Surgical CAD CAM Patient-specific Model Computer-

More information

Advanced tumor treatment. Precise. Comfortable. Convenient.

Advanced tumor treatment. Precise. Comfortable. Convenient. Advanced tumor treatment. Precise. Comfortable. Convenient. Precision CyberKnife is one of the most amazing breakthroughs in radiation delivery that I ve seen. - Dr. Martin Silverstein Radiation Oncologist,

More information

EXACTRAC HIGHLY ACCURATE PATIENT MONITORING

EXACTRAC HIGHLY ACCURATE PATIENT MONITORING EXACTRAC HIGHLY ACCURATE PATIENT MONITORING PATIENT POSITION MONITORING ExacTrac is an in-room based monitoring system that detects intrafractional motion during treatment delivery. Two kv X-Ray units

More information

Leksell Gamma Knife Icon A New User s Perspective

Leksell Gamma Knife Icon A New User s Perspective Leksell Gamma Knife Icon A New User s Perspective Steve Herchko, DMP 2017 MFMER slide-1 Disclosures None 2017 MFMER slide-2 Outline Icon Overview Mayo Clinic Experience Frame-Based System Mask-Based System

More information

Advanced tumor treatment. Precise. Comfortable. Convenient. Richard & Kathy Aquebogue

Advanced tumor treatment. Precise. Comfortable. Convenient. Richard & Kathy Aquebogue A PROGRAM OF MATHER HOSPITAL AT NEW YORK CANCER & BLOOD SPECIALISTS Advanced tumor treatment. Precise. Comfortable. Convenient. Richard & Kathy Aquebogue Precision CyberKnife is one of the most amazing

More information

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out

More information

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St. Quality Assurance of Ultrasound Imaging in Radiation Therapy Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St. Louis, Missouri Typical Applications of Ultrasound Imaging in Radiation Therapy

More information

Chapters from Clinical Oncology

Chapters from Clinical Oncology Chapters from Clinical Oncology Lecture notes University of Szeged Faculty of Medicine Department of Oncotherapy 2012. 1 RADIOTHERAPY Technical aspects Dr. Elemér Szil Introduction There are three possibilities

More information

PRECISE, ROBOTIC TREATMENT AS INDIVIDUAL AS EVERY PATIENT

PRECISE, ROBOTIC TREATMENT AS INDIVIDUAL AS EVERY PATIENT PRECISE, ROBOTIC TREATMENT AS INDIVIDUAL AS EVERY PATIENT BENEFITS OF THE CYBERKNIFE TREATMENT DELIVERY SYSTEM True robotic precision: Enable high definition radiotherapy anywhere in the body with the

More information

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,

More information

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague Radiosurgery by Leksell gamma knife Josef Novotny, Na Homolce Hospital, Prague Radiosurgery - Definition Professor Lars Leksell The tools used by the surgeon must be adapted to the task and where the human

More information

typical IMRT fraction time and expand high definition radiotherapy anywhere in the body with the widest range of motion of the

typical IMRT fraction time and expand high definition radiotherapy anywhere in the body with the widest range of motion of the Precisely maximize dose, Versatile, efficient and effective for the BENEFITS OFandTHE CYBERKNIFE TREATMENT minimize side effects range of radiation oncology patients maximize patient comfort DELIVERY SYSTEM

More information

Credentialing for the Use of IGRT in Clinical Trials

Credentialing for the Use of IGRT in Clinical Trials Credentialing for the Use of IGRT in Clinical Trials James M. Galvin, DSc Thomas Jefferson University Hospital Jefferson Medical College Philadelphia, PA and The Radiation Therapy Oncology Group RADIATION

More information

Radiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery

Radiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery Therapy SAM Symposium: WE-A-BRCD-1 Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery Kamil M. Yenice, PhD Associate Professor Chief of Clinical Physics

More information

Original Date: April 2016 Page 1 of 7 FOR CMS (MEDICARE) MEMBERS ONLY

Original Date: April 2016 Page 1 of 7 FOR CMS (MEDICARE) MEMBERS ONLY National Imaging Associates, Inc. Clinical guidelines STEREOTACTIC RADIATION THERAPY: STEREO RADIOSURGERY (SRS) AND STEREOTACTIC BODY RADIATION THERAPY (SBRT) CPT4 Codes: Please refer to pages 5-6 LCD

More information

Leksell Gamma Knife Icon. Treatment information

Leksell Gamma Knife Icon. Treatment information Leksell Gamma Knife Icon Treatment information You may be feeling frightened or overwhelmed by your recent diagnosis. It can be confusing trying to process a diagnosis, understand a new and challenging

More information

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system JBUON 2018; 23(1): 179-184 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Patient-specific quality assurance for intracranial cases in robotic

More information

4D Radiotherapy in early ca Lung. Prof. Manoj Gupta Dept of Radiotherapy & oncology I.G.Medical College Shimla

4D Radiotherapy in early ca Lung. Prof. Manoj Gupta Dept of Radiotherapy & oncology I.G.Medical College Shimla 4D Radiotherapy in early ca Lung Prof. Manoj Gupta Dept of Radiotherapy & oncology I.G.Medical College Shimla Presentation focus on ---- Limitation of Conventional RT Why Interest in early lung cancer

More information

Tumor Tracking Current & Future Developments. Josh Evans, Ph.D. Virginia Commonwealth University Richmond, VA

Tumor Tracking Current & Future Developments. Josh Evans, Ph.D. Virginia Commonwealth University Richmond, VA Tumor Tracking Current & Future Developments Josh Evans, Ph.D. Virginia Commonwealth University Richmond, VA None. Disclosures To understand: Learning Objectives the physiological characteristics of tumor

More information

Advanced Tumor Treatment

Advanced Tumor Treatment Advanced Tumor Treatment What is CyberKnife? The CyberKnife Stereotactic Radiosurgery System, offered by Southwest Washington Medical Center, is a noninvasive outpatient treatment without the risks and

More information

Technology Challenges of Commercial Medical Electron Accelerators

Technology Challenges of Commercial Medical Electron Accelerators Technology Challenges of Commercial Medical Electron Accelerators John Allen Chief Engineer Elekta Ltd Aims of this talk Medical Radiotherapy is already a large and profitable business Well established

More information

Current Concepts and Trends in Spinal Radiosurgery. Edward M. Marchan

Current Concepts and Trends in Spinal Radiosurgery. Edward M. Marchan Current Concepts and Trends in Spinal Radiosurgery Edward M. Marchan Spinal Neoplasia The spine is the most common site of skeletal metastatic disease. (70%) 40% of bony metastases involve the vertebrae

More information

Peer Reviewed Bibliography. (in English)

Peer Reviewed Bibliography. (in English) CyberKnife Stereotactic Radiosurgery System Peer Reviewed Bibliography (in English) Scientific Articles and Book Chapters H Adler, J.R., Murphy, M.J., Chang, S.D., Hancock, S.L.: Neurosurgery 44(6):1299-1307,

More information

What is the CyberKnife System?

What is the CyberKnife System? CYBERKNIFE Robotic Radiosurgery System PATIENT BROChURE What is the CyberKnife System? have stabilizing frames bolted to their head or limit their breathing during treatment to minimize movement of the

More information

Clinical Implementation of a New Ultrasound Guidance System. Vikren Sarkar Bill Salter Martin Szegedi

Clinical Implementation of a New Ultrasound Guidance System. Vikren Sarkar Bill Salter Martin Szegedi Clinical Implementation of a New Ultrasound Guidance System Vikren Sarkar Bill Salter Martin Szegedi Disclosure The University of Utah has research agreements with Elekta Agenda Historical Review Trans-Abdominal

More information

FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE. Geoffrey S. Ibbott, Ph.D. June 20, 2017

FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE. Geoffrey S. Ibbott, Ph.D. June 20, 2017 FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE Geoffrey S. Ibbott, Ph.D. June 20, 2017 1 DISCLOSURES My institution holds Strategic Partnership Research Agreements with Varian, Elekta, and Philips

More information

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER Bro. Dr. Collie Miller IARC/WHO Based on trends in the incidence of cancer, the International Agency for Research on Cancer (IARC) and WHO

More information

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System Varian Edge Experience Jinkoo Kim, Ph.D Henry Ford Health System Disclosures I participate in research funded by Varian Medical Systems. Outline of Presentation Review advanced imaging in Varian Edge Linear

More information

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed

More information

Subject: Image-Guided Radiation Therapy

Subject: Image-Guided Radiation Therapy 04-77260-19 Original Effective Date: 02/15/10 Reviewed: 01/25/18 Revised: 01/01/19 Subject: Image-Guided Radiation Therapy THIS MEDICAL COVERAGE GUIDELINE IS NOT AN AUTHORIZATION, CERTIFICATION, EXPLANATION

More information

Performance Evaluation of Calypso (R) 4D Localization and Kilovoltage Image Guidance Systems for Interfraction Motion Management of Prostate Patients

Performance Evaluation of Calypso (R) 4D Localization and Kilovoltage Image Guidance Systems for Interfraction Motion Management of Prostate Patients Performance Evaluation of Calypso (R) 4D Localization and Kilovoltage Image Guidance Systems for Interfraction Motion Management of Prostate Patients Tomi Ogunleye, Emory University Peter J Rossi, Emory

More information

Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife of-the-art and New Innovations

Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife of-the-art and New Innovations Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife State-of of-the-art and New Innovations Chad Lee, PhD CK Solutions, Inc. and CyberKnife Centers of San Diego Outline Basic overview

More information

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During

More information

Quality assurance program for prototype stereotactic system developed for Neptun 10 PC linac

Quality assurance program for prototype stereotactic system developed for Neptun 10 PC linac Iran. J. Radiat. Res., 2005; 3 (2): 73-78 Quality assurance program for prototype stereotactic system developed for Neptun 10 PC linac A.R. Khoshbin Khoshnazar 1*, M.T. Bahreyni Toossi 2, A.R. Hashemiyan

More information

IMAGE-GUIDED RADIATION THERAPY

IMAGE-GUIDED RADIATION THERAPY IMAGE-GUIDED RADIATION THERAPY Your Single Source Oncology Solutions Provider Plan. Target. Treat. At Best NOMOS, we design products and solutions that help medical professionals treat a variety of cancers.

More information

IGRT/Adaptive Gating

IGRT/Adaptive Gating IGRT/Adaptive Gating unlocking possibilities exceeding limits with confidence PIONEERING IGRT BrainLAB is committed to providing the best radiotherapy solutions for better treatment and patient care. BrainLAB

More information

Elekta - a partner and world-leading supplier

Elekta - a partner and world-leading supplier Experience Elekta Elekta - a partner and world-leading supplier of clinical solutions for image guided radiation therapy, stereotactic radiotherapy, radiosurgery and brachytherapy, as well as advanced

More information

Adaptive filtering to predict lung tumor motion during free breathing

Adaptive filtering to predict lung tumor motion during free breathing CARS 2002 - H.U Lemke, M W. Vannier; K. Inamura. A.G. Farman, K. Doi & J.H.c. Reiber (Editors) "CARS/Springer. All rights reserved. Adaptive filtering to predict lung tumor motion during free breathing

More information

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1 1 2 KV CBCT Imaging Part I Rabih Hammoud, MS, DABR Henry Ford Health System Detroit, Michigan Acknowledgements Indrin Chetty, PhD Teamour Nurushev, PhD Harrison Guan, PhD Jinkoo Kim, PhD JianYue Jin, PhD

More information

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT April 25 nd, 2015 Lung Cancer Lung cancer is the most lethal cancer: Over 224,000 new diagnoses in the U.S. predicted

More information

Dosimetry, see MAGIC; Polymer gel dosimetry. Fiducial tracking, see CyberKnife radiosurgery

Dosimetry, see MAGIC; Polymer gel dosimetry. Fiducial tracking, see CyberKnife radiosurgery Subject Index Acoustic neuroma, neurofibromatosis type 2 complications 103, 105 hearing outcomes 103, 105 outcome measures 101 patient selection 105 study design 101 tumor control 101 105 treatment options

More information

Delivery Quality Assurance with a High Resolution Liquid Filled Ion Chamber Array for Robotic Radiosurgery

Delivery Quality Assurance with a High Resolution Liquid Filled Ion Chamber Array for Robotic Radiosurgery Adamczyk S, Chan M, Stelljes T, Schmitt D, Skrobala A, Poppinga D, Adamczyk M, Pawalowski B, Hildebrandt G, Poppe B, Blanck O Dr. Oliver Blanck University Clinic Lübeck, Germany Phone: +491728033815 E-Mail:

More information

CYBERKNIFE M6 SERIES: Unmatched Precision and Patient Comfort

CYBERKNIFE M6 SERIES: Unmatched Precision and Patient Comfort CYBERKNIFE M6 SERIES: Unmatched Precision and Patient Comfort Precisely maximize dose, minimize side effects and maximize patient comfort The new CyberKnife M6 Series has the capabilities and efficiency

More information

Potential conflicts-of-interest. Respiratory Gated and Four-Dimensional Tumor Tracking Radiotherapy. Educational objectives. Overview.

Potential conflicts-of-interest. Respiratory Gated and Four-Dimensional Tumor Tracking Radiotherapy. Educational objectives. Overview. Respiratory Gated and Four-Dimensional Tumor Tracking Radiotherapy Potential conflicts-of-interest I am PI of a sponsored research agreement between Stanford University and Varian Medical Systems P Keall

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

Ultrasonographic guidance for spinal extracranial radiosurgery: technique and application for metastatic spinal lesions

Ultrasonographic guidance for spinal extracranial radiosurgery: technique and application for metastatic spinal lesions Neurosurg Focus 11 (6):Article 8, 2001, Click here to return to Table of Contents Ultrasonographic guidance for spinal extracranial radiosurgery: technique and application for metastatic spinal lesions

More information

Completion of Treatment Planning. Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA

Completion of Treatment Planning. Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA Completion of Treatment Planning Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA Outline Entering Prescription Plan Printout Print and Transfer DRR Segment BEV Export to R&V Physician approval

More information

Precisely Maximize Dose, side effects, Patient

Precisely Maximize Dose, side effects, Patient Precisely Maximize Dose, Minimize side effects, Maximize Patient Comfort Key Differentiators: The CyberKnife M6 Series Industry-leading precision and patient comfort Enables full-body robotic radiosurgery,

More information

I. Equipments for external beam radiotherapy

I. Equipments for external beam radiotherapy I. Equipments for external beam radiotherapy 5 linear accelerators (LINACs): Varian TrueBeam 6, 10 & 18 MV photons, 6-18 MeV electrons, image-guided (IGRT) and intensity modulated radiotherapy (IMRT),

More information

IGRT Protocol Design and Informed Margins. Conflict of Interest. Outline 7/7/2017. DJ Vile, PhD. I have no conflict of interest to disclose

IGRT Protocol Design and Informed Margins. Conflict of Interest. Outline 7/7/2017. DJ Vile, PhD. I have no conflict of interest to disclose IGRT Protocol Design and Informed Margins DJ Vile, PhD Conflict of Interest I have no conflict of interest to disclose Outline Overview and definitions Quantification of motion Influences on margin selection

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,

More information

The Revolution in Radiation Therapy

The Revolution in Radiation Therapy The Revolution in Radiation Therapy Calvin Huntzinger, MS Varian Medical Systems THE EMERGENCE OF IMRT... 1 THE IMRT PROCESS HOW IT IS PRACTICED TODAY... 5 PLANNING FOR IMRT - SECRETS OF SUCCESS... 11

More information

Institute of Oncology & Radiobiology. Havana, Cuba. INOR

Institute of Oncology & Radiobiology. Havana, Cuba. INOR Institute of Oncology & Radiobiology. Havana, Cuba. INOR 1 Transition from 2-D 2 D to 3-D 3 D conformal radiotherapy in high grade gliomas: : our experience in Cuba Chon. I, MD - Chi. D, MD - Alert.J,

More information

Stereotaxy. Outlines. Establishing SBRT Program: Physics & Dosimetry. SBRT - Simulation. Body Localizer. Sim. Sim. Sim. Stereotaxy?

Stereotaxy. Outlines. Establishing SBRT Program: Physics & Dosimetry. SBRT - Simulation. Body Localizer. Sim. Sim. Sim. Stereotaxy? Establishing SBRT Program: Physics & Dosimetry Lu Wang, Ph.D. Radiation Oncology Department Fox Chase Cancer Center Outlines Illustrate the difference between SBRT vs. CRT Introduce the major procedures

More information

Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies

Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies J. Y. C. Cheung and K. N. Yu a Department of Physics and Materials Science, City University of Hong Kong, Kowloon Tong,

More information

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of: Canadian Partnership for Quality Radiotherapy Technical Quality Control Guidelines for Gamma Knife Radiosurgery A guidance document on behalf of: Canadian Association of Radiation Oncology Canadian Organization

More information

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of Medicine Indianapolis, IN, USA Indra J. Das,

More information

AbstractID: 8073 Title: Quality Assurance, Planning and Clinical Results for Gamma Knife Radiosurgery S. Goetsch, Ph.D. Page 1 Seattle, WA May 2008

AbstractID: 8073 Title: Quality Assurance, Planning and Clinical Results for Gamma Knife Radiosurgery S. Goetsch, Ph.D. Page 1 Seattle, WA May 2008 Knife Radiosurgery S. Goetsch, Ph.D. Page 1 Quality Assurance, Planning and Clinical Results for Gamma Knife Radiosurgery Steven J. Goetsch, Ph.D., FAAPM ACMP 25 th Annual Meeting Seattle, WA May 2008

More information

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems Dental ConeBeam Computed Tomography (CBCT) X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL CONEBEAM COMPUTED TOMOGRAPHY (CBCT) X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability.

More information

Evaluation of accuracy in determination of stereotactical isocenter: from image acquisition to treatment.

Evaluation of accuracy in determination of stereotactical isocenter: from image acquisition to treatment. Evaluation of accuracy in determination of stereotactical isocenter: from image acquisition to treatment. ANNA SARDO 1, EUGENIA MADON 1, EDOARDO TREVISIOL 1, DANILO SANTUARI 1, VERONICA RICHETTO 1, ROBERTO

More information

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015 IROC Lung Phantom 3D CRT / IMRT Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015 The IROC requests that each institution keep the phantom for no more than 2 weeks. During

More information

Herlev radiation oncology team explains what MRI can bring

Herlev radiation oncology team explains what MRI can bring Publication for the Philips MRI Community Issue 46 2012/2 Herlev radiation oncology team explains what MRI can bring The radiotherapy unit at Herlev University Hospital investigates use of MRI for radiotherapy

More information

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES MELISSA C. MARTIN, M.S., FACR, FAAPM PRESIDENT AAPM - 2017 PRESIDENT - THERAPY PHYSICS INC., GARDENA, CA MELISSA@THERAPYPHYSICS.COM AAPM Spring

More information

Protura Robotic Patient Positioning System. for efficiency + performance

Protura Robotic Patient Positioning System. for efficiency + performance Protura Robotic Patient Positioning System for efficiency + performance Protura Robotic Patient Positioning System The Protura Robotic Patient Positioning System is the ultimate in robotic patient motion

More information

A Navigation System for Minimally Invasive CT-Guided Interventions

A Navigation System for Minimally Invasive CT-Guided Interventions A Navigation System for Minimally Invasive CT-Guided Interventions Markus Nagel 1,GerdSchmidt 2, Ralf Petzold 3, and Willi A. Kalender 1 1 Institute of Medical Physics, University of Erlangen-Nürnberg,

More information

On the road to Gamma Knife Icon workflow automation in radiation oncology. Young Lee Department of Medical Physics

On the road to Gamma Knife Icon workflow automation in radiation oncology. Young Lee Department of Medical Physics On the road to Gamma Knife Icon workflow automation in radiation oncology Young Lee Department of Medical Physics Outline - introduction to GK Icon and workflow - goals - processes for automation - on-going

More information

Strategies and Technologies for Cranial Radiosurgery Planning: Gamma Knife

Strategies and Technologies for Cranial Radiosurgery Planning: Gamma Knife Conflicts of Interest Strategies and Technologies for Cranial Radiosurgery Planning: Gamma Knife David Schlesinger, Ph.D. Research support: Elekta Instruments, AB Lars Leksell Gamma Knife Center University

More information

Results of Stereotactic radiotherapy for Stage I and II NSCLC Is There a Need for Image Guidance?

Results of Stereotactic radiotherapy for Stage I and II NSCLC Is There a Need for Image Guidance? Results of Stereotactic radiotherapy for Stage I and II NSCLC Is There a Need for Image Guidance? Frank Zimmermann Institute of Radiation Oncolgy University Clinic Basel Petersgraben 4 CH 4031 Basel radioonkologiebasel.ch

More information

ACR MRI Accreditation: Medical Physicist Role in the Application Process

ACR MRI Accreditation: Medical Physicist Role in the Application Process ACR MRI Accreditation: Medical Physicist Role in the Application Process Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics University of Texas M.D. Anderson Cancer Center Educational Objectives

More information

Application of Implanted Markers in Proton Therapy. Course Outline. McLaren Proton Therapy Center Karmanos Cancer Institute McLaren - Flint

Application of Implanted Markers in Proton Therapy. Course Outline. McLaren Proton Therapy Center Karmanos Cancer Institute McLaren - Flint Application of Implanted Markers in Proton Therapy Sung Yong Park, Ph.D. McLaren Proton Therapy Center Karmanos Cancer Institute McLaren - Flint AAPM 2016, SAM Therapy Educational Course, 2016.08.04. Course

More information

Motion gating and tracking techniques: overview and recent developments

Motion gating and tracking techniques: overview and recent developments Motion gating and tracking techniques: overview and recent developments Gig S Mageras, PhD, FAAPM Department of Medical Physics Memorial Sloan Kettering Cancer Center, New York MSK/gsm 15-Jun-2018 1 Disclosure

More information

Volumetric Modulated Arc Therapy - Patient Specific QA

Volumetric Modulated Arc Therapy - Patient Specific QA Volumetric Modulated Arc Therapy - Patient Specific QA Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA VMAT plan QA methods Composite dose measurement Film & ion chamber diode array Mapcheck

More information

Elekta Synergy Digital accelerator for advanced IGRT

Elekta Synergy Digital accelerator for advanced IGRT Elekta Synergy Digital accelerator for advanced IGRT Setting the standard for confident care The Field of Radiation Therapy is Constantly Changing Being able to take full advantage of the latest clinical

More information

RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy

RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy Guidelines for Planning and Irradiating the RPC Liver Phantom. Revised Dec 2005 Credentialing for this protocol requires four steps:

More information

FOR CMS (MEDICARE) MEMBERS ONLY NATIONAL COVERAGE DETERMINATION (NCD) FOR MAGNETIC RESONANCE IMAGING:

FOR CMS (MEDICARE) MEMBERS ONLY NATIONAL COVERAGE DETERMINATION (NCD) FOR MAGNETIC RESONANCE IMAGING: National Imaging Associates, Inc. Clinical guidelines BONE MARROW MRI Original Date: July 2008 Page 1 of 5 CPT Codes: 77084 Last Review Date: September 2014 NCD 220.2 MRI Last Effective Date: July 2011

More information

Robotically Assisted Nerve and Facet Blocks: A Cadaveric Study 1

Robotically Assisted Nerve and Facet Blocks: A Cadaveric Study 1 Technical Report Robotically Assisted Nerve and Facet Blocks: A Cadaveric Study 1 Kevin Cleary, PhD, Dan Stoianovici, PhD, Alexandru Patriciu, MS, Dumitru Mazilu, PhD David Lindisch, RT, Vance Watson,

More information