1 Parker et al. Breast Radiation During CT Pulmonary Angiography Chest Imaging Original Research Mark S. Parker 1 Ferdinand K. Hui Marc A. Camacho Jiyearn K. Chung Dean W. Broga Narinder N. Sethi Parker MS, Hui FK, Camacho MA, Chung JK, Broga DW, Sethi NN DOI: /AJR Received May 13, 2004; accepted after revision November 29, Presented at the 2004 annual meeting of the American Roentgen Ray Society, Miami Beach, FL. 1 All authors: Department of Radiology, Thoracic Imaging, Medical College of Virginia Hospitals, Virginia Commonwealth University Health System, 1250 E Marshall St., Main Bldg., 3rd Fl., PO Box , Richmond, VA Address correspondence to M. S. Parker (msparker@mail1.vcu.edu). AJR 2005; 185: X/05/ American Roentgen Ray Society Female Breast Radiation Exposure During CT Pulmonary Angiography OBJECTIVE. The objective of our study was to estimate the effective radiation dose to the female breast during CT pulmonary angiography compared with other routine diagnostic imaging techniques. MATERIALS AND METHODS. We retrospectively reviewed the demographic data of patients who underwent CT pulmonary angiography between May 2000 and December 2002, the diagnostic yield of those studies, and the estimated effective radiation dose to the breast incurred during CT. The estimated effective radiation dose was calculated using the ImPACT CT (Impact Performance Assessment of CT) dosimetry calculator and the CT dose index (CTDI) and was compared with the average glandular dose for two-view screening mammography. RESULTS. During the study period, 1,325 CT pulmonary angiograms were obtained. Sixty percent (797) of the scans were obtained on female patients. The mean age of scanned females was 52.5 years (range, years). Of the studies performed in females, 401 (50.31%) were negative, 151 (18.95%) were nondiagnostic, and 245 (30.74%) were positive for pulmonary thromboembolism. The calculated effective minimum dose to the breast of an average 60-kg woman during CT was 2.0 rad (20 mgy) per breast compared with an average glandular dose of rad (3 mgy) for standard two-view screening mammography. CONCLUSION. CT pulmonary angiography delivers a minimum radiation dose of 2.0 rad (20 mgy) to the breasts of an average-sized woman. This greatly exceeds the American College of Radiology recommendation of rad (3 mgy) or less for standard two-view mammography. The potential latent carcinogenic effects of such radiation exposure at this time remain unknown. We encourage the judicious use of CT pulmonary angiography and lower doses and nonionizing radiation alternatives when appropriate. adiologic imaging studies have become an integral diagnostic tool in R the practice of clinical medicine. The role of body CT in particular has grown exponentially since its introduction in the mid 1970s. Today, more than 35 million CT scans are obtained annually in the United States alone. It is estimated that CT is now responsible for approximately 13% of all radiologic procedures performed in the United States and contributes 30% of the medical diagnostic radiation dosage to patients [1 3]. As the use of CT has markedly increased over the past two decades, so have its technical capabilities. CT now plays a critical role in the diagnosis and management of most patients with complex thoracic and cardiopulmonary disease, including pulmonary thromboembolism. With the advent of multidetector scanners, CT has become the de facto gold standard for imaging pulmonary emboli [4]. Although the morbidity and mortality of unrecognized and untreated acute pulmonary embolism are well known, the potential carcinogenic effects of ionizing radiation on radiosensitive tissues such as the eye, thyroid gland, and female breast in particular are often not considered or deemed relevant to acute patient care. Because a large percentage of patients evaluated with CT pulmonary angiography are women of reproductive age, our objective was to estimate the effective radiation dose to the female breast incurred during these studies and compare that value with the average glandular breast dose for routine mammography. This information will allow referring physicians and radiologists to describe the effective radiation dose and relative risk of CT in terms patients understand. We also encourage the judicious use of CT pulmonary angiography and lower doses and nonionizing radiation alternatives when appropriate AJR:185, November 2005

2 Breast Radiation During CT Pulmonary Angiography Materials and Methods Our study involved four phases. First, we retrospectively reviewed the demographic data of all patients who underwent CT pulmonary angiography studies at our institution between May 2000 and December Second, the diagnostic outcome of each of these CT studies was reviewed and categorized as negative for pulmonary thromboembolism, nondiagnostic for establishing or excluding the diagnosis of pulmonary thromboembolism, and positive for pulmonary thromboembolism. Third, we divided the clinical status of studied patients into one of three categories: inpatient, outpatient, and emergency department. We did not include data on incidental pulmonary thromboembolism detected on scans acquired for other clinical concerns or diagnoses. Fourth, on the basis of our CT pulmonary angiography protocol and imaging parameters, our physicists estimated the effective radiation dose to the female breast of an average 60-kg female using the ImPACT CT (Impact Performance Assessment of CT, Bence Jones Offices, St. George s Hospital, London) dosimetry calculator and the CT dose index (CTDI) listed for Siemens Plus 4 scanners and compared that value with the average glandular breast dose for a typical two-view screening mammogram reported in the literature. All CT pulmonary angiography studies were performed on Plus 4 or Volume Zoom units (Siemens Medical Solutions): 5-mm noncontrast images were acquired first from the thoracic inlet to the adrenal glands using 140 kv, 120 mas, and a 1.5 pitch. The scanning time delay was determined by administering a 30-mL timing bolus of iohexol iodinated contrast material (Omnipaque 300, Amersham Health Products) with sequential stationary images acquired through the main pulmonary artery. Subsequently, 2-mm contrast-enhanced images reconstructed every 1.5 mm were acquired in the caudocranial plane after the IV administration of 170 ml of iohexol using 140 kv, 150 mas, and a 2.0 pitch. Because our study straddled the transition to PACS, approximately 20% of the studies were reviewed both on hard copy and at workstations. The remaining 80% of studies were interpreted directly on the PACS monitors in the reading room for the thoracic imaging division. Fellowship-trained board-certified thoracic radiologists interpreted 90% of the CT studies. Board-certified nonthoracic radiologists interpreted the remaining studies. Results During the defined study period, 1,325 CT pulmonary angiography studies were performed. As suspected, the majority of studies, 60% (797), were on female patients. Forty percent (528) of the studies were on male patients. Most of the scanned females were of reproductive age with active and radiosensitive fibroglandular breast tissue. The mean age of scanned females was 52.5 years (range, years) (Fig. 1). One hundred ninetynine CT studies (25%) were performed on females less than 40 years old. One hundred forty-one examinations (17.7%) were on women years old, 46 examinations (6%) on women who ranged in age from 20 years to less than 25 years, and 24 studies (3%) on those less than 20 years old. The CT pulmonary angiography diagnostic outcomes in the female cohort included 401 studies (50.31%) negative for pulmonary thromboembolism, 151 studies (18.95%) nondiagnostic or inconclusive for pulmonary thromboembolism, and 245 studies (30.74%) positive for pulmonary thromboembolism (Fig. 2). Forty-one (29.1%) of the 141 CT studies performed on women 35 years old or younger were positive, 65 (46.1%) were negative, and 35 (24.8%) were nondiagnostic. Two hundred four (31.1%) of the 656 CT examinations performed on women older than 35 years were positive, 329 (50.2%) were negative, and 119 (18.14%) were nondiagnostic. Quantum mottle related to the patient s body habitus, respiratory motion degradation, and suboptimal vascular opacification was responsible for most nondiagnostic (4%) or inconclusive (15%) studies. The majority of these patients underwent no further diagnostic imaging and were managed empirically on the basis of the level of clinical suspicion. The positive yield for CT pulmonary angiography studies was greatest for female inpatient and emergency department patients (Fig. 3). Sixty-four percent (157) of positive CT studies were in inpatient females, and 27% (66) of positive studies were in emergency department females. Only 7% (16) of positive studies were identified in outpatient females. Between May 2000 and December 2002, 314 CT pulmonary angiography studies were performed in the emergency department. Two hundred thirteen scans were obtained on women, and 101 scans on men (female male ratio, 2:1). The mean age of the female patients was 49 years (range, years). A total of 14 patients had more than one scan, including one individual who was scanned eight times, each time for suspected pulmonary thromboembolism disease. Ten (71%) of these 14 individuals were women. The mean age of this subgroup was 48.5 years (range, years). Of the 314 scans, a D-dimer level was obtained in 89 patients (28.3%). Using a D-dimer level of 500 ng/ml or less to exclude acute pulmonary thromboembolism, we found that 34 patients had negative D-dimer results. Twenty-four of these patients had a negative CT pulmonary angiogram and 10 patients had a positive CT examination. Fifty-five patients had a positive D- dimer (> 500 ng/ml). Thirty-six of these patients had a negative CT examination, and 11 had a positive CT examination. Eight of these scans (14.5%) were interpreted as inconclusive for acute pulmonary thromboembolism. The D-dimer level showed a sensitivity of 52.4%, specificity of 40%, a positive predictive value of 23.4%, and a negative predictive value of 70.6%. Basic dosimetric quantities are absorbed dose, expressed in grays (1 Gy = 100 rad), and equivalent dose, expressed in sieverts (1 Sv = 100 rem). These quantities are numerically equal for diagnostic radiographs. Organ dose is expressed in milligrays or rads. The principal CT radiation dose descriptor, which integrates the radiation dose delivered both within and beyond the scanned volume, is the CT dose index (CTDI). Although the CTDI does not represent an actual radiation dose to a specific patient, it can be used as a standardized index of the average dose delivered from a CT scan and can be used to calculate the effective radiation dose [2, 3, 5]. The CTDI is calculated as follows: CTDI 1 +L 2 T L 2 where T is the slice thickness, L is the length of the CT ionization chamber, D(z) is the dose as a function of the position along the z-axis of rotation, and dz is integration distance. The CTDI is determined for a centered singlescan dose profile and depends on the peak kilovoltage and milliampere-second settings and the distance from the focus to the center of rotation [3, 6]. Our physicists calculated the estimated female breast effective radiation dose using the ImPACT CT dosimetry calculator and the CTDI listed for Siemens Plus 4 scanners for the contrast-enhanced phase of our chest CT examinations. The calculated dose to the average 60-kg woman during CT pulmonary angiography was 2.0 rad (20 mgy) per breast. This dose compares with an average glandular breast dose of rad (300 mrad or 3 mgy) for standard two-view screening mammography [7]. Our physicists did not calculate the radiation dose delivered to the breasts during the unenhanced image acquisition. Because unen- = Dz ( ) dz AJR:185, November

3 Parker et al. Fig. 1 Bar graph shows age breakdown of female patients at time of CT pulmonary angiography. Fig. 2 Pie chart shows diagnostic outcomes negative (light gray), positive (white), or nondiagnostic (dark gray) for female patients who underwent CT pulmonary angiography. Fig. 3 Chart shows CT pulmonary angiography outcome negative (light gray), positive (white), or nondiagnostic (dark gray) in female patients based on clinical setting: No. of Studies No. of Patients hanced image acquisition is part of the scanning protocol and likewise contributes to the total breast dose, inclusion of this additional piece of data would have been insightful Emergency Department Age (yr) Inpatient Outpatient Discussion Our study confirmed our hypothesis that a large percentage of patients evaluated with CT pulmonary angiography are indeed women, 60% in our cohort, and that a significant percentage of scanned women (26.7%) are young, under 40 years old, with radiosensitive fibroglandular breast tissue. We also confirmed that the breast tissue is exposed to a much higher radiation dose, a minimum of 2.0 rad, than radiologists, medical physicists, and the American College of Radiology, according to their guidelines, would desire [2, 3, 7, 8]. The clinical diagnosis of pulmonary thromboembolic disease is both problematic and unreliable. It is difficult to triage patients for CT pulmonary angiography on the basis of clinical findings alone. Patient age is likewise unreliable for triage purposes. One might assume that the rate of pulmonary thromboembolism disease would be considerably lower in younger patients and that a patient s age might be used to triage the CT evaluation of patients with suspected pulmonary thromboembolism. Our study does not support that assumption: 31% of our female cohort who were younger than 35 years old had a positive CT pulmonary angiogram. Similarly, 31% of our female patients over 35 years had a positive CT pulmonary angiogram. Although this may be peculiar to our particular institution, we do not believe age alone is a reliable tool for triage, and we believe that women of any age with the clinical suspicion of pulmonary thromboembolism will continue to be evaluated with CT pulmonary angiography. In fact, the number of CT examinations is likely to continue to increase as the clinical threshold for scanning patients continues to decrease. Thus, it is imperative to reduce the exposure of breast tissue to ionizing radiation. The issue of female breast radiation exposure during chest CT is relevant for many thoracic applications of CT, including CT pulmonary angiography. Exposures ranging from 2.0 rad, as in our study, to 5.0 rad, at the body surface, are not rare for conventional chest CT. CT studies that overlap scanned regions or rescan the same anatomic region of interest (e.g., unenhanced and enhanced scans) have 2 3 times the radiation dose of nonoverlapped scans. This would include the protocol currently used at our institution. Our physicists did not calculate this additional radiation dose delivered to the breasts during the noncontrast phase of CT pulmonary angiography, which also contributes to the total breast dose. Theoretically, this would have further increased our minimum dose estimate of 2.0 rad (20 mgy) by two- or threefold. Dose rates for CT angiography are typically in the same range as conventional chest CT, rad (20 40 mgy) [9]. This compares with an effective radiation dose equivalent of rad ( mgy) for twoview chest radiography and an average glandular breast dose of rad (300 mrad or 3 mgy) for standard two-view screening mammography [8, 10]. Few physicians are aware that conventional diagnostic chest CT imparts a radiation dose of rad (20 50 mgy) to the breasts of an average-sized woman. This dose is roughly equivalent to twoview mammograms and up to as many as chest radiographs [9 12]. The technical capabilities of CT have been further enhanced with MDCT. State-of-theart MDCT provides exquisite anatomic detail, reduced examination time, and the ability to perform complex multiplanar vascular and 3D reconstructions that have had a dramatic impact on vascular imaging in particular. However, the radiation dose is 30 50% greater with 4-MDCT compared with singleslice helical scanners as a result of scan over AJR:185, November 2005

4 Breast Radiation During CT Pulmonary Angiography lap, positioning of the X-ray tube closer to the patient, and increased scatter created by the wider X-ray beam [3, 8, 12, 13]. The adverse biologic effects associated with radiation exposure are classified as either deterministic or stochastic. Deterministic effects result from cell death and are quantified by radiation dose to a particular region that has a threshold level. If this threshold is exceeded, cell death usually occurs. Such effects do not occur with the doses typically used for CT. The potential radiation risks in patients undergoing CT are due to stochastic effects. These effects are believed to result in carcinomas in the patient and genetic mutations and alterations in offspring of affected patients. The probability of stochastic effects depends on the amount of radiation absorbed. According to the International Commission on Radiation Protection (ICRP) Special Task Force Report 2000, the radiation doses used in CT often approach or exceed those levels known to increase the probability of nonfatal and fatal cancers [2, 14]. To date, no scientific study has reported a direct link between cancer and CT radiation. Most authors have drawn inferences of an increased risk of cancer on the basis of the outcomes of victims surviving the nuclear fallout in Hiroshima and Nagasaki. However, the increasing radiation exposure from CT and its potential adverse effects on both young patients and radiosensitive tissues is a real concern. Available data for radiation-induced cancers suggest that 1 msv (0.1 rad) of radiation exposure may lead to five additional cancers in 100,000 exposed patients [15, 16]. Assuming a linear relationship between increasing radiation dose exposure and the stochastic effects of ionizing radiation on biologic tissue, one can extrapolate a possible additional 100 cancers per 100,000 exposed individuals from CT angiography for pulmonary thromboembolism evaluation alone. This falsely assumes all fibroglandular breast tissue has the same radiosensitivity, regardless of patient age, and breast volume. The risk of diagnostic imaging procedures inducing breast carcinoma is not hypothetical. It has already been observed. A study of 1,030 women with scoliosis who routinely underwent multiple thoracic spine radiography examinations as young girls revealed a nearly twofold statistically significant increased risk for incident breast carcinoma [17]. Furthermore, radiologic technologists, exposed to low levels of occupational ionizing radiation, have elevated risks of female breast cancer, thyroid cancer, and melanoma [18]. Most physicians would agree that CT is an excellent imaging technique and that the benefit of any one given CT examination outweighs the risk of the study. Collectively however, the risk of several thousand or possibly million CT scans and not infrequently several in the same patient could become a public health issue. Recall, in our study, that 14 patients had more than one CT pulmonary angiogram, including one individual who was scanned eight times. Ten of these patients were young women with a mean age of 48.5 years. This same phenomenon is likely encountered at many other institutions across the country as well. Efforts should be made to reduce the number of repeat scans and the CT radiation dose without impacting the clinical usefulness of the study. It is possible to reduce the radiation dose in performing chest CT examinations by modifying scanning protocols and technical settings. Radiologists typically do this by reducing the tube current (normally between 80 and 300 mas, or ma/sec), increasing the table increment (pitch), reducing the exposure time and therefore the exposure level, and reducing the tube voltage [18, 19]. However, each manipulation is associated with a compromise in image quality and potentially in diagnostic information. For example, diagnostic images of the lung may be obtained with tube currents as low as ma. However, this reduction in milliamperes is associated with an increase in noise and quantum mottle and may compromise the evaluation of mediastinal structures, lung findings, and the detection of distal segmental and subsegmental emboli and pulmonary nodules less than or equal to 5 mm in diameter [9, 19, 20]. On some MDCT scanners, increases in pitch are associated with a proportional increase in tube current to maintain similar noise. In such cases, the increase in pitch does not affect the radiation dose, but rather compromises image quality and the value of multiplanar reconstructions [21]. Reducing tube voltage below kv leads to an unfavorable ratio of absorbed radiation and signal-to-noise ratio in areas of the body with high radiation absorption (e.g., shoulder girdle) and is not useful [9, 20, 21]. Radiologists and clinicians must be willing to sacrifice image quality and potentially diagnostic information if such manipulations in scanning parameters are to be the sole means of reducing patient radiation exposure. Recently, manufacturers have responded to the need for radiation dose minimization and have implemented significant improvements in the newer, now widely available, 8- and 16-MDCT scanners [19]. One such innovation, similar to automatic exposure control in conventional diagnostic radiologic imaging, involves dynamic real-time tube current modulation based on body geometry and attenuation. Although the resultant radiation doses are lower than those incurred with the 4-MDCT scanners, the dose is still proportionally greater compared with the doses of other routine diagnostic imaging studies. It is possible to reduce the radiation dose to superficial radiosensitive organs such as the female breast, thyroid gland, and eye during chest CT examinations without adversely affecting imaging quality or diagnostic information. Thin-layered bismuth radioprotective garments have been designed for this purpose (AttenuRad, Dyna Medical). Although available, these devices are not typically used outside of pediatric imaging centers. Bismuth garments are far more malleable to the body s surface, fitting better than traditional lead aprons or shields. Proper placement of these bismuth garments can reduce breast radiation exposure by 57%, thyroid gland dose by 60%, and that to the eye by 40% [22, 23]. The reusable protective breast shields cost approximately $ The shields for the thyroid and eye are approximately $5.00 [22]. These shields could be used routinely, especially for female patients, whenever the head, chest, or abdomen is scanned. We have recently begun using the AttenuRad breast shield at our institution during CT pulmonary angiograms. We are also in the preliminary stages of developing our own custom-designed breast shield. We are interested in conducting a study with various-sized breast phantoms to evaluate the effectiveness of this breast shield on breast radiation dose reduction on our newly installed 16-MDCT scanners. Alternative examinations, other than chest CT, are another means of reducing radiation exposure to radiosensitive organ systems. This may be feasible particularly in the female outpatient population who, according to our study, had positive CT pulmonary angiograms in only 7% of the cases. The poor diagnostic performance of the radionuclide ventilation perfusion (V/Q) scans among patients with abnormal chest radiographs, significant cardiopulmonary disease, and emphysema is well known to clinicians and radiologists alike. However, in the setting of a normal chest radiograph and no history of significant cardiopulmonary disease, the V/Q scan can AJR:185, November

5 Parker et al. be effectively used to triage patients with suspected acute pulmonary thromboembolism [24]. The female breast receives less than 1 rad (< 1 cgy) of exposure during V/Q lung scanning [8]. Although CT is currently the preferred study, MR angiography is another alternative imaging technique that provides information similar to that obtained with CT and should be considered in those patients in whom iodinated contrast agents or cumulative ionizing radiation exposure is of concern (e.g., young patients, reproductive-age women, and especially women who have undergone multiple prior CT examinations). Contrast-enhanced 3D MR angiography for the evaluation of acute pulmonary thromboembolism has sensitivities ranging between 75% and 100% and specificities of %. The positive predictive value and negative predictive value are 87% and 100%, respectively, for central emboli. MR angiography does have limitations in its ability to detect subsegmental emboli and is also not appropriate for critically ill patients requiring close monitoring [24, 25]. The plasma D-dimer enzyme-linked immunosorbent assay (ELISA) has become recognized as a sensitive screening test for excluding acute pulmonary thromboembolism [26 29]. Numerous other qualitative and quantitative D- dimer assays have been introduced. One such quantitative assay, the immunoturbidimetric assay, has been shown to be equivalent to the ELISA [27]. D-dimer is a specific degradation product of cross-linked fibrin. The pairing of the D domains of fibrin monomers occurs only with full cross-linking of the monomers. D- dimer is thus elevated in the setting of deep venous thrombosis and pulmonary embolism. Unfortunately, disseminated intravascular coagulation, pregnancy, sepsis, liver disease, trauma, surgery, and neoplastic disease are also associated with elevated D-dimer levels; thus, it cannot be used to triage such patients. A positive D-dimer result is of limited value, especially in hospitalized or critically ill patients. However, a negative D-dimer result is potentially useful in excluding acute pulmonary thromboembolism, with reported negative predictive values of % far exceeding that encountered in our study. Some advocate incorporating the D-dimer assay into the triage of emergency department and outpatient populations [26 29]. Our experience with the D-dimer assay has been unrewarding. Even though many of our patients had underlying neoplastic, rheumatic, and surgical diseases, which may explain the lower sensitivity (52.4%), specificity (40%), and negative predictive value (70.6%) than reported in the literature, we can understand the reluctance of our clinical colleagues to routinely integrate the D-dimer assay into their clinical algorithm. We are currently revisiting this issue at our institution. The latency period for potential cancer induction is estimated to be years in the dose ranges used in CT pulmonary angiography [8]. It is an accepted concept among radiation biologists and public health officials that the younger the patient is at the time of exposure, the greater their lifetime risk of developing nonfatal and fatal cancers. The greater lifetime risk is compounded by the increased biologic susceptibility to radiationinduced cancer. Thus, CT pulmonary angiography may not always be the best diagnostic option in young patients or reproductive-age and perimenopausal women. If CT is indeed justified in this patient population, every effort should be made to reduce the radiation dose, shield the patient, and limit the number of CT examinations performed. In conclusion, the technologic advances in CT have had a dramatic impact on the diagnosis and management of almost all patients with complex thoracic and cardiopulmonary disease, including pulmonary thromboembolism. The role of CT will only continue to grow with the availability of MDCT and its multiplanar vascular and 3D reconstruction capabilities. However, the need for any given CT examination should always be justified on sound medical grounds. Both radiologists and referring clinicians need to work together to develop better selection criteria of patients for CT; reduce the radiation exposure of patients; and shield the eyes, thyroid gland, and female breasts whenever possible. In those situations in which the diagnostic yield of CT is expected to be low, alternative examinations should be considered. References 1. Gamsu G, Held BT, Czum JM. Reduced radiation for adult thoracic CT: a practical approach. J Thorac Imaging 2004; 19: Kalra MK, Maher MM, Saini S. CT radiation exposure: rationale for concern and strategies for dose reduction. Proceedings from the SCBT/MR. Applied Radiol 2003; 7: Mayo JR, Aldrich J, Muller NL. Radiation exposure at chest CT: a statement of the Fleischner Society. Radiology 2003; 228: Abella H. CT gains favor for imaging pulmonary embolism: high-resolution scans can spot abnormalities beyond PE and may influence treatment. Diagn Imaging 2003; Shope TB, Gagne RM, Johnson GC. A method for describing the doses delivered by transmission x-ray computed tomography. Med Phys 1981; 8: van der Bruggen-Bogaarts BAHA, Broerse JJ, Lammers JWJ, van Waes PFGM, Geleijns J. Radiation exposure in standard and high-resolution chest CT scans. Chest 1995; 107: American College of Radiology (ACR). ACR practice guideline for the performance of screening mammography. Reston, VA: ACR, 1999: Nickoloff EL, Alderson PO. Radiation exposures to patients from CT: reality, public perception, and policy. AJR 2001; 177: Diederich S, Lenzen H. Radiation exposure associated with imaging of the chest: comparison of different radiographic and computed tomography techniques. Cancer 2000; 89[suppl 11]: McCollough CH, Liu HH. Breast dose during electron-beam CT: measurement with film dosimetry. Radiology 1995; 196: Trigaux JP, Lacrosse M. Radiation exposure and computed tomography. Rev Mal Respir 1999; 16: McCollough CH, Zink FE. Performance evaluation of a multi-slice CT system. Med Phys 1999; 26: Hidajat N, Maurer J, Schroder RJ, Wolf M, Vogel T, Felix R. Radiation exposure in spiral computed tomography: dose distribution and dose reduction. Invest Radiol 1999; 34: Rehani MM, Bongartz G, Kalender W, et al. Managing X-ray dose in computed tomography: ICRP Special Task Force report. Ann ICRP 2000; 30: International Commission on Radiation Protection (ICRP). Recommendations of the International Commission on Radiological Protection. Oxford, UK: Pergamon Press, 1991: publication no Maher MM, Kalra MK, Toth TL, Wittram C, Saini S, Shepard J. Application of rational practice and technical advances for optimizing radiation dose for chest CT. J Thorac Imaging 2004; 19: Morin DM, Lonstein JE, Stovall M, Hacker DG, Luckyanov N, Land CE. Breast cancer morality after diagnostic radiography: findings from the U.S. scoliosis cohort study. Spine 2000; 25: Sigurdson AJ, Doody MM, Rao RS, et al. Cancer incidence in the US radiologic technologists health study, Cancer 2003; 97: Mayo JR, Kim K, MacDonald SLS, et al. Reduced radiation dose helical chest CT: effect on reader evaluation of structures and lung findings. Radiology 2004; 232: Diederich S, Lenzen H, Windmann R, et al. Pulmonary nodule: experimental and clinical studies at 1232 AJR:185, November 2005

6 Breast Radiation During CT Pulmonary Angiography low-dose CT. Radiology 1999; 213: Mahesh M, Scatarige JC, Cooper J, Fishman EK. Dose and pitch relationship for a particular multislice CT scanner. AJR 2001; 177: Hopper KD, King SH, Lobell ME, et al. The breast: in-plane x-ray protection during diagnostic thoracic CT shielding with bismuth radioprotective garments. Radiology 1997; 205: Fricke BL, Donnelly LF, Frush DP, et al. In-plane bismuth breast shields for pediatric CT: effects on radiation dose and image quality using experimental and clinical data. AJR 2003; 180: Hatabu H, Uematsu H, Nguyen B, Miller WT Jr, Hasegawa I, Gefter WB. CT and MR in pulmonary embolism: a changing role for nuclear medicine in diagnostic strategy. Semin Nucl Med 2002; 32: Meaney JF, Weg JG, Chenevert TL, et al. Diagnosis of pulmonary embolism with magnetic resonance angiography. N Engl J Med 1997; 336: Knecht MF, Heinrich F. Clinical evaluation of an immunoturbidimetric D-dimer assay in the diagnostic procedure of deep vein thrombosis and pulmonary embolism. Thromb Res 1997; 88: Dunn KL, Wolf JP, Dorfman DM, Fitzpatrick P, Baker JL, Goldhaber SZ. Normal D-dimer levels in emergency department patients with suspected acute pulmonary embolism. J Am Coll Cardiol 2002; 40: Wells PS, Anderson DR, Rodger M, et al. Excluding pulmonary embolism at the bedside without diagnostic imaging: management of patients with suspected pulmonary embolism presenting to the emergency department by using a simple clinical model and D-dimer. Ann Intern Med 2001; 135: Quinn DA, Fogel RB, Smith CD, et al. D-dimers in the diagnosis of pulmonary embolism. Am J Respir Crit Care Med 1999; 159: AJR:185, November

Why is CT Dose of Interest?

Why is CT Dose of Interest? Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase

More information

CT Radiation Risks and Dose Reduction

CT Radiation Risks and Dose Reduction CT Radiation Risks and Dose Reduction Walter L. Robinson, M.S. D.A.B.S.N.M., D.A.B.M.P., D.A.B.R. Consultant Certified Medical Radiation Health & Diagnostic Imaging Physicist Medical Radiation and Children

More information

Paediatric Dose Reduction and Image Quality

Paediatric Dose Reduction and Image Quality Paediatric Dose Reduction and Image Quality Alan Whiteside The majority of this work was undertaken as part of MSc Thesis of Helen Dixon. Introduction Paediatric CT protocols result in a higher effective

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

To Shield or Not to Shield? Lincoln L. Berland, M.D.

To Shield or Not to Shield? Lincoln L. Berland, M.D. To Shield or Not to Shield? Lincoln L. Berland, M.D. Disclosures Consultant to: Nuance, Inc. Page 2 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation.

More information

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Lincoln L. Berland, M.D., F.A.C.R. Michael V. Yester, Ph.D. University of Alabama at Birmingham Breast Radiation on CT Use of chest CT

More information

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Minimizing CT Radiation Dose CT examinations improve health care and are an essential part

More information

How to Develop CT Protocols for Children

How to Develop CT Protocols for Children How to Develop CT Protocols for Children Introduction Prior to 2001 the vast majority of CT imaging of children was conducted using the same or similar techniques used for adult imaging. In 2001, several

More information

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan Radiation Dose To Pediatric Patients in Computed Tomography in Sudan Omer Osman,Saeed Medical Physics Department, ALNeelain University, Sudan Presentation outlines Introduction Objectives Materials and

More information

Radiation Exposure in Pregnancy. John R. Mayo UNIVERSITY OF BRITISH COLUMBIA

Radiation Exposure in Pregnancy. John R. Mayo UNIVERSITY OF BRITISH COLUMBIA Radiation Exposure in Pregnancy John R. Mayo UNIVERSITY OF BRITISH COLUMBIA Illustrative Clinical Scenario 32 year old female 34 weeks pregnant with recent onset shortness of breath and central chest pain

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Debra Pennington, MD Director of Imaging Dell Children s Medical Center Debra Pennington, MD Director of Imaging Dell Children s Medical Center 1 Gray (Gy) is 1 J of radiation energy/ 1 kg matter (physical quantity absorbed dose) Diagnostic imaging doses in mgy (.001 Gy)

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation

Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation Medical Physics and Informatics Original Research Angel et al. Radiation Dose During Chest CT Medical Physics and Informatics Original Research Erin Angel 1,2 Nazanin Yaghmai 1 Cecilia Matilda Jude 1 John

More information

Page 1 of 5 Patient Safety: Radiation Dose in X-Ray and CT Exams What are x-rays and what do they do? X-rays are forms of radiant energy, like light or radio waves. Unlike light, x-rays can penetrate the

More information

Outline. NCRP Scientific Committee 6-2

Outline. NCRP Scientific Committee 6-2 Magnitude of Medical Radiation Exposures to US population Mahadevappa Mahesh, MS, PhD, FAAPM. Assistant Professor of Radiology & Cardiology Chief Physicist - Johns Hopkins Hospital The Russell H. Morgan

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

Managing Patient Dose in Computed Tomography (CT)

Managing Patient Dose in Computed Tomography (CT) Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

Comparative Analysis of Radiation Dose and Image Quality Between Thyroid Shielding and Unshielding During CT Examination of the Neck

Comparative Analysis of Radiation Dose and Image Quality Between Thyroid Shielding and Unshielding During CT Examination of the Neck Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Young Hen Lee 1 Eun-tae Park 1 Pyong Kon Cho 1 Hyung Suk Seo 1 Bo-Kyung Je 1 Sang-il Suh 2 Kyung-Sook

More information

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

A more accurate method to estimate patient dose during body CT examinations with tube current modulation A more accurate method to estimate patient dose during body CT examinations with tube current modulation Poster No.: C-0738 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Kawaguchi 1, Y. Matsunaga

More information

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center CT Dose Estimation John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology 1 University of California Davis Medical Center CT Dose Estimation Introduction The CTDI Family of Metrics

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

Skyscan 1076 in vivo scanning: X-ray dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in

More information

Pulmonary Embolism. Thoracic radiologist Helena Lauri

Pulmonary Embolism. Thoracic radiologist Helena Lauri Pulmonary Embolism Thoracic radiologist Helena Lauri 8.5.2017 Statistics 1-2 out of 1000 adults annually are diagnosed with deep vein thrombosis (DVT) and/or pulmonary embolism (PE) About half of patients

More information

Translating Protocols Across Patient Size: Babies to Bariatric

Translating Protocols Across Patient Size: Babies to Bariatric Translating Protocols Across Patient Size: Babies to Bariatric Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

Radiation Dose in Pediatric Imaging

Radiation Dose in Pediatric Imaging Radiation Dose in Pediatric Imaging A Brief History of Radiology Dose: Why Does It Matter? Measuring Exposure and Dose Deterministic Effects Stochastic Effects Common Exams: What is the Risk? Reducing

More information

Radiation Dose Reduction Strategies in Coronary CT Angiography

Radiation Dose Reduction Strategies in Coronary CT Angiography Radiation Dose Reduction Strategies in Coronary CT Angiography Noor Diyana Osman, PhD noordiyana@usm.my Contents: Introduction Radiation dosimetry in CT Radiation risk associated with coronary CT angiography

More information

People Exposed to More Radiation from Medical Exams

People Exposed to More Radiation from Medical Exams People Exposed to More Radiation from Medical Exams With its release of a new report, titled Ionizing Radiation Exposure of the Population of the United States (Report No. 160, 2009), the National Council

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT

Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT Mannudeep K. Kalra, MD, DNB Michael M. Maher, MD, FRCR, FFR (RCSI) Srinivasa R.

More information

CT Dose Optimization for Whole- Body PET/CT Examinations

CT Dose Optimization for Whole- Body PET/CT Examinations Nuclear Medicine and Molecular Imaging Original Research Tonkopi et al. CT Dose Optimization for PET/CT Nuclear Medicine and Molecular Imaging Original Research FOCUS ON: Elena Tonkopi 1,2 Andrew A. Ross

More information

To Shield or Not to Shield: Application of Bismuth Breast Shields

To Shield or Not to Shield: Application of Bismuth Breast Shields Cardiopulmonary Imaging Opinion Colletti et al. Bismuth Breast Shields Cardiopulmonary Imaging Opinion FOCUS ON: Patrick M. Colletti 1 Orlando A. Micheli Kai H. Lee Colletti PM, Micheli OA, Lee KH Keywords:

More information

Dose Estimates for Nuclear Medicine Procedures: What are they? Where do they come from?

Dose Estimates for Nuclear Medicine Procedures: What are they? Where do they come from? Dose Estimates for Nuclear Medicine Procedures: What are they? Where do they come from? SNM Continuing Education Lecture Salt Lake City, UT -- June 6, 2010 Darrell R. Fisher Pacific Northwest National

More information

The Increasing Use of CT and Its Risks

The Increasing Use of CT and Its Risks STUDENT SCOPE The Increasing Use of CT and Its Risks Matthew Voress is a radiography student at Owens Community College in Toledo, Ohio. This article was awarded first prize in the Ohio Society of Radiologic

More information

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

Estimation of the Risk of Cancer Associated with Pediatric Cranial Computed Tomography

Estimation of the Risk of Cancer Associated with Pediatric Cranial Computed Tomography British Journal of Medicine & Medical Research 9(10): 1-7, 2015, Article no.bjmmr.18079 ISSN: 2231-0614 SCIENCEDOMAIN international www.sciencedomain.org Estimation of the Risk of Cancer Associated with

More information

Ionizing Radiation Exposure from Radiologic Imaging: The Issue and What Can We Do?

Ionizing Radiation Exposure from Radiologic Imaging: The Issue and What Can We Do? Ionizing Radiation Exposure from Radiologic Imaging: The Issue and What Can We Do? Background, The increased use of diagnostic imaging requiring the use of ionizing radiation, the rapidly expanding use

More information

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy What is radiation? Radiation is energy emitted from a source, that travels through space and can penetrate matter. Listed below are two types that we are exposed to and contribute to our overall radiation

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Proper Diagnosis of Venous Thromboembolism (VTE)

Proper Diagnosis of Venous Thromboembolism (VTE) Proper Diagnosis of Venous Thromboembolism (VTE) Whal Lee, M.D. Seoul National University Hospital Department of Radiology 2 nd EFORT Asia Symposium, 3 rd November 2010, Taipei DVT - Risk Factors Previous

More information

CT Dose Reduction in Pediatric Patients

CT Dose Reduction in Pediatric Patients CT Dose Reduction in Pediatric Patients By Kelly Firestine, RT(R)(CT)(M) Executive Summary CT is an incredibly valuable imaging tool, but there are unique concerns with pediatric patients, including the

More information

Pulmonary Embolism. Pulmonary Embolism. Pulmonary Embolism. PE - Clinical

Pulmonary Embolism. Pulmonary Embolism. Pulmonary Embolism. PE - Clinical Pulmonary embolus - a practical approach to investigation and treatment Sam Janes Wellcome Senior Fellow and Respiratory Physician, University College London Background Diagnosis Treatment Common: 50 cases

More information

Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography

Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography The British Journal of Radiology, 81 (2008), 499 503 SHORT COMMUNICATION Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography 1 G R IBALL, MSc, BSc, 2

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

New Criteria for Ventilation-Perfusion Lung Scan Interpretation: A Basis for Optimal Interaction with Helical CT Angiography 1

New Criteria for Ventilation-Perfusion Lung Scan Interpretation: A Basis for Optimal Interaction with Helical CT Angiography 1 1206 July-August 2000 RG Volume 20 Number 4 New Criteria for Ventilation-Perfusion Lung Scan Interpretation: A Basis for Optimal Interaction with Helical CT Angiography 1 Alexander Gottschalk, MD Introduction

More information

Radiation Dose in X-Ray and CT Exams

Radiation Dose in X-Ray and CT Exams Scan for mobile link. Patient Safety: Radiation Dose in X-Ray and CT Exams What are x-rays and what do they do? X-rays are forms of radiant energy, like light or radio waves. Unlike light, x-rays can penetrate

More information

Ask EuroSafe Imaging Tips & Tricks. CT Working Group

Ask EuroSafe Imaging Tips & Tricks. CT Working Group Ask EuroSafe Imaging Tips & Tricks CT Working Group Organ Based Tube Current Modulation to Reduce Radiation Dose to Superficial Radiosensitive Organs Eileen Kelly (Galway University Hospitals, IE) Matthias

More information

State of the art and future development for standardized estimation of organ doses in CT

State of the art and future development for standardized estimation of organ doses in CT State of the art and future development for standardized estimation of organ doses in CT March 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Agenda Introduction Duke Florida

More information

Canadian Association of Nuclear Medicine Association Canadienne de Médecine Nucléaire

Canadian Association of Nuclear Medicine Association Canadienne de Médecine Nucléaire CHOOSING WISELY Canadian Association of Nuclear Medicine Association Canadienne de Médecine Nucléaire Five Things Physicians and Patients Should Question 1- Don t perform stress cardiac imaging or coronary

More information

Imaging of acute pulmonary thromboembolism*

Imaging of acute pulmonary thromboembolism* Silva, Isabela et al. Imaging of acute pulmonary thromboembolism Imaging of acute pulmonary thromboembolism* C. ISABELA S. SILVA, NESTOR L. MÜLLER The diagnosis of acute pulmonary thromboembolism is based

More information

Dose Reduction Options in Cardiac CT

Dose Reduction Options in Cardiac CT Dose Reduction Options in Cardiac CT Doyle P, Ball P*, Donnelly P # Radiological Sciences & Imaging, Forster Green Hospital *Department of Radiology, Ulster Hospital # Department of Cardiology, Ulster

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

RADIATION SAFETY. Junior Radiology Course

RADIATION SAFETY. Junior Radiology Course RADIATION SAFETY Junior Radiology Course Expectations for the Junior Radiology Course Medical School wants students to learn basic principles, factual knowledge, safety info, etc. Medical Students want

More information

Radiation Safety in the Catheterization Lab

Radiation Safety in the Catheterization Lab SCAI FALL FELLOWS COURSE - 2015 Radiation Safety in the Catheterization Lab V. Vivian Dimas, MD, FSCAI Associate Professor Pediatrics, Cardiology UT Southwestern Medical Center Dallas TX None Disclosures

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

Organ-Based Dose Current Modulation and Thyroid Shields: Techniques of Radiation Dose Reduction for Neck CT

Organ-Based Dose Current Modulation and Thyroid Shields: Techniques of Radiation Dose Reduction for Neck CT Medical Physics and Informatics Original Research Hoang et al. Radiation Dose Reduction for Neck CT Medical Physics and Informatics Original Research Jenny K. Hoang 1,2 Terry T. Yoshizumi 1 Kingshuk Roy

More information

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial Medical Physics and Informatics Original Research Larke et al. Estimated Radiation Dose for Low-Dose Chest CT Medical Physics and Informatics Original Research Frederick J. Larke 1 Randell L. Kruger 2

More information

Patient Dose in the Diagnosis of PE

Patient Dose in the Diagnosis of PE Patient Dose in the Diagnosis of PE IZAAZ BADSHAH 2018 CANM-CAMRT Joint Annual Meeting March 22-24, 2018 Disclosure I do not have a financial interest, arrangement or affiliation including receipt of honoraria

More information

Epidermiology Early pulmonary embolism

Epidermiology Early pulmonary embolism Epidermiology Early pulmonary embolism Sitang Nirattisaikul Faculty of Medicine, Prince of Songkla University 3 rd most common cause of cardiovascular death in the United States, following ischemic heart

More information

Assessment of effective dose in paediatric CT examinations

Assessment of effective dose in paediatric CT examinations Assessment of effective dose in paediatric CT examinations E. Dougeni 1,2 CL. Chapple 1, J. Willis 1, G. Panayiotakis 2 1 Regional Medical Physics Department, Freeman Hospital, Freeman Road, Newcastle

More information

Medical Physics and Informatics Original Research

Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Christner et al. Estimating Effective Dose for CT Medical Physics and Informatics Original Research FOCUS ON: Jodie A. Christner 1 James M. Kofler Cynthia

More information

Low-Dose CT: Clinical Studies & the Radiologist Perspective

Low-Dose CT: Clinical Studies & the Radiologist Perspective Low-Dose CT: Clinical Studies & the Radiologist Perspective RD-ASiR RD-MBIR SD-FBP RD=0.35 msv (80% dose reduction) Perry J. Pickhardt, MD UW School of Medicine & Public Health Low-Dose CT: Clinical Overview

More information

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Department of Medical Physics and Bioengineering, Royal United Hospital, Bath Overview Varian Acuity

More information

Radiation Exposure 1980 to 2006

Radiation Exposure 1980 to 2006 Radiation Exposure 1980 to 2006 Background 3-6 msv/yr Natural (85% 45%) Radon Cosmic Rays Air travel Living at Altitude Man-made (15% 55%) Medical Imaging** mgy Radiation Therapy cgy Radiation Whole Body

More information

Seattle Children s Hospital Radiology Department. Statement regarding radiation exposure related to computed. tomography (CT) exams

Seattle Children s Hospital Radiology Department. Statement regarding radiation exposure related to computed. tomography (CT) exams Seattle Children s Hospital Radiology Department Statement regarding radiation exposure related to computed tomography (CT) exams Computed tomography (CT) scanners use radiation in the form of X- rays

More information

8/1/2017. Financial Disclosures. Dose Tracking at MGH. How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center

8/1/2017. Financial Disclosures. Dose Tracking at MGH. How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center Mannudeep K. Kalra, MD Webster Center for Quality and Safety Massachusetts General Hospital Harvard Medical School

More information

Ionizing Radiation Exposure from Radiologic Imaging

Ionizing Radiation Exposure from Radiologic Imaging Ionizing Radiation Exposure from Radiologic Imaging Background The increased use of diagnostic imaging requiring the use of ionizing radiation, the rapidly expanding use of computed tomography in the emergency

More information

Dose-equivalent equivalent = absorbed

Dose-equivalent equivalent = absorbed UCSF General Surgery 2010 Radiation Risks of Diagnostic Radiology in Trauma Robert A. Izenberg, M.D., FACS University of California, San Francisco San Francisco General Hospital Context Increasingly liberal

More information

PULMONARY EMBOLISM ANGIOCT (CTA) ASSESSMENT OF VASCULAR OCCLUSION EXTENT AND LOCALIZATION OF EMBOLI 1. BACKGROUND

PULMONARY EMBOLISM ANGIOCT (CTA) ASSESSMENT OF VASCULAR OCCLUSION EXTENT AND LOCALIZATION OF EMBOLI 1. BACKGROUND JOURNAL OF MEDICAL INFORMATICS & TECHNOLOGIES Vol. 11/2007, ISSN 1642-6037 Damian PTAK * pulmonary embolism, AngioCT, postprocessing techniques, Mastora score PULMONARY EMBOLISM ANGIOCT (CTA) ASSESSMENT

More information

Missed Pulmonary Embolism on Abdominal CT

Missed Pulmonary Embolism on Abdominal CT Cardiopulmonary Imaging Original Research Lim et al. Pulmonary Emboli Missed on CT of the Abdomen Cardiopulmonary Imaging Original Research Kun Young Lim 1 Seth J. Kligerman 2 Cheng Ting Lin 2 Charles

More information

Spine MRI and Spine CT Test Request Tip Sheet

Spine MRI and Spine CT Test Request Tip Sheet Spine MRI and Spine CT MRI is almost always preferred over CT scan; if ordering CT, CLEARLY document why MRI is not appropriate. In cases of back pain without red flags, six weeks of multimodality supervised

More information

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA Full Length Research Article 1 Ibrahim, U, 3 Daniel, I.H., 3 Ayaninola, O., 4 Ibrahim,

More information

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433 - 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.

More information

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Conventional and spiral CT dose indices in Yazd general hospitals, Iran Iran. J. Radiat. Res., 2006; 3 (4): 183-189 Conventional and spiral CT dose indices in Yazd general hospitals, Iran F. Bouzarjomehri 1*,M.H.Zare 2, D. Shahbazi 2 1 Department of Medical Physics, Shahid

More information

The print quality of this copy is not an accurate representation of the original.

The print quality of this copy is not an accurate representation of the original. Cardiovasc Intervent Radiol DOI 10.1007/s00270-010-9945-4 TECHNICAL NOTE Staff Radiation Doses in a Real-Time Display Inside the Angiography Room Roberto Sanchez E. Vano J. M. Fernandez J. J. Gallego Received:

More information

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result. Minimizing Dose Understand the importance and methods of pre-procedure patient assessment including a review of previous radiologic exams, disease processes and anatomical considerations that may increase

More information

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose Ask EuroSafe Imaging Tips & Tricks CT Working Group Optimization of scan length to reduce CT radiation dose Alban Gervaise (Centre Hospitalier Universitaire Nancy, FR) Mika Kortesniemi (HUS Medical Imaging

More information

Radiation Safety for New Medical Physics Graduate Students

Radiation Safety for New Medical Physics Graduate Students Radiation Safety for New Medical Physics Graduate Students John Vetter, PhD Medical Physics Department UW School of Medicine & Public Health Background and Purpose of This Training This is intended as

More information

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Poster No.: C-2584 Congress: ECR 2015 Type: Scientific

More information

Dual Energy CT Aortography: Can We Reduce Iodine Dose??

Dual Energy CT Aortography: Can We Reduce Iodine Dose?? Dual Energy CT Aortography: Can We Reduce Iodine Dose?? William P. Shuman MD, FACR FSCBTMR Department of Radiology University of Washington SCBTMR Annual Course Boston, October 10, 2012 Conflict of Interest

More information

BrightSpeed Elite CT with ASiR: Comparing Dose & Image Quality Rule Out Pulmonary Embolism on Initial & Follow-Up Exam

BrightSpeed Elite CT with ASiR: Comparing Dose & Image Quality Rule Out Pulmonary Embolism on Initial & Follow-Up Exam GE Healthcare BrightSpeed Elite CT with ASiR: Comparing Dose & Image Quality Rule Out Pulmonary Embolism on Initial & Follow-Up Exam Michael Swack, MD Diagnostic Radiologist Irvington Radiologists, PC

More information

Improved image quality of low-dose thoracic CT examinations with a new postprocessing software*

Improved image quality of low-dose thoracic CT examinations with a new postprocessing software* JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 3, Summer 2010 Improved image quality of low-dose thoracic CT examinations with a new postprocessing software* Anne Catrine Traegde Martinsen,

More information

Spine MRI and Spine CT Test Request Tip Sheet

Spine MRI and Spine CT Test Request Tip Sheet Spine MRI and Spine CT With/Without Contrast CT, MRI Studies should NOT be ordered simultaneously as dual studies (i.e., with and without contrast). Radiation exposure is doubled and both views are rarely

More information

Current status of diagnostic imaging in dental university hospitals in Japan

Current status of diagnostic imaging in dental university hospitals in Japan Oral Radiol (2004) 20:15 21 Japanese Society for Oral and Maxillofacial Radiology and Springer-Verlag Tokyo 2004 DOI 10.1007/s11282-004-0010-3 ORIGINAL ARTICLE Takehito Sasaki Minoru Fujita Tsuguhisa Katoh

More information

After the Chest X-Ray:

After the Chest X-Ray: After the Chest X-Ray: What To Do Next Alan S. Brody Professor of Radiology and Pediatrics Chief of Thoracic Imaging Cincinnati Children s Hospital Cincinnati, Ohio USA What Should We Do Next? CT scan?

More information

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION Managing Pediatric CT Patient Doses Keith J. Strauss, MSc, FAAPM, FACR President X-Ray Computations, Inc. Boston, Massachusetts Acknowledgements Marilyn Goske, MD John Boone, PhD Cynthia McCollough, PhD

More information

Steven Aaron Ross, M.D. Pediatric Radiologist El Paso Imaging Consultants El Paso Children s Hospital

Steven Aaron Ross, M.D. Pediatric Radiologist El Paso Imaging Consultants El Paso Children s Hospital Steven Aaron Ross, M.D. Pediatric Radiologist El Paso Imaging Consultants El Paso Children s Hospital I will prescribe regimens for the good of my patients according to my ability and my judgment and never

More information

Usefulness of Clinical Pre-test Scores for a Correct Diagnostic Pathway in Patients with Suspected Pulmonary Embolism in Emergency Room

Usefulness of Clinical Pre-test Scores for a Correct Diagnostic Pathway in Patients with Suspected Pulmonary Embolism in Emergency Room Send Orders for Reprints to reprints@benthamscience.net The Open Emergency Medicine Journal, 2013, 5, (Suppl 1: M-4) 19-24 19 Open Access Usefulness of Clinical Pre-test Scores for a Correct Diagnostic

More information

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS ABBY DUROJAYE,M.D CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS BOTH HAVE BEEN SHOWN TO BE EFFECTIVE TOOLS EARLY DETECTION OF BREAST

More information

FDG-18 PET/CT - radiation dose and dose-reduction strategy

FDG-18 PET/CT - radiation dose and dose-reduction strategy FDG-18 PET/CT - radiation dose and dose-reduction strategy Poster No.: C-1856 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit P. Nicholson, S. McSweeney, K. O'Regan; Cork/IE Radiation

More information

Computed Tomography (CT) - Chest

Computed Tomography (CT) - Chest Scan for mobile link. Computed Tomography (CT) - Chest What is CT Scanning of the Chest? Computed tomography, more commonly known as a CT or CAT scan, is a diagnostic medical test that, like traditional

More information

Ask EuroSafe Imaging. Tips & Tricks. Paediatric Imaging Working Group. Shielding in pediatric CT

Ask EuroSafe Imaging. Tips & Tricks. Paediatric Imaging Working Group. Shielding in pediatric CT Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Shielding in pediatric CT Claudio Granata (IRCCS Istituto Giannina Gaslini, IT) Joana Santos (ESTeSC-Coimbra Health School, PT) Elina

More information

Cover Page. The handle holds various files of this Leiden University dissertation.

Cover Page. The handle  holds various files of this Leiden University dissertation. Cover Page The handle http://hdl.handle.net/1887/21764 holds various files of this Leiden University dissertation. Author: Mos, Inge Christina Maria Title: A more granular view on pulmonary embolism Issue

More information

Introduction and Background

Introduction and Background CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section

More information

Cardiovascular Imaging in Pregnancy. Diana Litmanovich, MD

Cardiovascular Imaging in Pregnancy. Diana Litmanovich, MD Cardiovascular Imaging in Pregnancy Diana Litmanovich, MD Cardiovascular disorders 4% of all pregnancies PE DVT Aortic dissection Peripartum cardiomyopathy Acute coronary syndrome Hypertension Pneumonia

More information

Detectability of subsegmental pulmonary vessels in 64 MDCT-pulmonary angiography.

Detectability of subsegmental pulmonary vessels in 64 MDCT-pulmonary angiography. ISPUB.COM The Internet Journal of Radiology Volume 11 Number 2 Detectability of subsegmental pulmonary vessels in 64 MDCT-pulmonary angiography. T Niemann, G Bongartz Citation T Niemann, G Bongartz. Detectability

More information