Organ-Based Dose Current Modulation and Thyroid Shields: Techniques of Radiation Dose Reduction for Neck CT

Size: px
Start display at page:

Download "Organ-Based Dose Current Modulation and Thyroid Shields: Techniques of Radiation Dose Reduction for Neck CT"

Transcription

1 Medical Physics and Informatics Original Research Hoang et al. Radiation Dose Reduction for Neck CT Medical Physics and Informatics Original Research Jenny K. Hoang 1,2 Terry T. Yoshizumi 1 Kingshuk Roy Choudhury 1 Giao B. Nguyen 1 Greta Toncheva 1 Andreia R. Gafton 1 James D. Eastwood 1 Carolyn Lowry 1 Lynne M. Hurwitz 1 Hoang JK, Yoshizumi TT, Roy Choudhury K, et al. Keywords: CT, neck CT, organ-based dose modulation, radiation dose reduction, tube current modulation DOI: /AJR Received June 25, 2011; accepted after revision September 5, J.K. Hoang is a GE-AUR Fellow for Department of Radiology, Duke University Medical Center, Box 3808, Erwin Rd, Durham, NC Address correspondence to J. K. Hoang (jennykh@gmail.com). 2 Department of Radiation Oncology, Duke University Medical Center, Durham, NC. AJR 2012; 198: X/12/ American Roentgen Ray Society Organ-Based Dose Current Modulation and Thyroid Shields: Techniques of Radiation Dose Reduction for Neck CT OBJECTIVE. The purpose of this study was to assess the difference in absorbed organ dose and image quality for MDCT neck protocols using automatic tube current modulation alone compared with organ-based dose modulation and in-plane thyroid bismuth shielding. MATERIALS AND METHODS. An anthropomorphic female phantom with metal oxide semiconductor field effect transistor (MOSFET) detectors was scanned on a 64-MDCT scanner. The protocols included a reference neck CT protocol using automatic tube current modulation and three modified protocols: organ-based dose modulation, automatic tube current modulation with thyroid shield, and organ-based dose modulation with thyroid shield. Image noise was evaluated quantitatively with the SD of the attenuation value, and subjectively by two neuroradiologists. RESULTS. Organ-based dose modulation, automatic tube current modulation with thyroid shield, and organ-based dose modulation with thyroid shield protocols reduced the thyroid dose by 28%, 33%, and 45%, respectively, compared with the use of automatic tube current modulation alone (p 0.005). Organ-based dose modulation also reduced the radiation dose to the ocular lens (33 47%) compared with the use of automatic tube current modulation (p 0.04). There was no significant difference in measured noise and subjective image quality between the protocols. CONCLUSION. Both organ-based dose modulation and thyroid shields significantly reduce the thyroid organ dose without degradation of subjective image quality compared with automatic tube current modulation. Organ-based dose modulation has the additional benefit of dose reduction to the ocular lens. O ptimization of CT parameters considers the balance between image quality needed for accurate diagnosis and radiation dose exposure. A neck CT examination extends from the lower portion of the brain to the upper portion of the chest, irradiating several radiosensitive organs, including the eyes, thyroid, and bone marrow. These organs are regarded to have greater stochastic risk for injury and future malignancy, especially with cumulative doses and in younger patients [1, 2]. Prior studies assessing dose reduction techniques for neck CT have focused on thyroid shielding and automatic tube current modulation [3 5]. These forms of dose reduction have been shown to reduce radiation dose to the thyroid by 47% and 78%, respectively, compared with fixed tube current [3, 6]. Organ-based dose modulation is a new technique of radiation dose reduction that modulates tube current and can be used for neck imaging. Current MDCT scanners use tube current modulation software to modulate the tube current in the x-, y-, and z-axes on the basis of body habitus. This is known as automatic tube current modulation or real-time tube current modulation. Organ-based dose modulation similarly modulates along the z- axis on the basis of body habitus, but in contrast to automatic tube current modulation, it modulates the beam in the x- and y-planes by reducing the tube current over a prescribed 120 radial arc over the anterior, lateral, or posterior aspect of the body (operator preference). The overall radiation dose is kept constant with organ-based dose modulation as the tube current is increased in the remaining 240 radial arc. As a result, centrally located organs have a constant radiation exposure, whereas superficially located organs have a reduced exposure within the 120 radial arc and increased exposure over the 240 arc. Organ-based dose modulation has 1132 AJR:198, May 2012

2 Radiation Dose Reduction for Neck CT been proposed for use in chest and brain imaging to reduce breast and ocular lens dose, respectively [7]. Its use during neck imaging to reduce dose to the thyroid gland has not been evaluated. The purpose of this study was to assess the difference in absorbed organ dose and image quality for MDCT neck protocols by comparing the use of organ-based dose modulation and in-plane thyroid bismuth shielding to a reference protocol with only automatic tube current modulation. Materials and Methods Anthropomorphic Phantom and Dosimetry Institutional review board approval and HIPAA compliance were not required because this was a phantom study. A commercially available anthropomorphic female phantom (model 702-D, CIRS) (Fig. 1) was used for measurement of absorbed organ doses. The phantom is equivalent to a human measuring 160 cm in height and 55 kg in weight. Twenty metal oxide semiconductor field effect transistor (MOSFET) dosimeters (model TABLE 1: MDCT Scanning Parameters A 1002RD, Best Medical) with active detector areas of µm (total dimensions, 2.5 mm width 1.3 mm thickness 8 mm length) were placed in specific anatomic locations in the neck, chest, abdomen, and pelvis to allow determination of whole-body effective dose and absorbed organ doses. The detectors were in the following organs: breast, skin over the mid anterior neck, cerebellum, ocular lens, mandible, cervical spine, thyroid, esophagus, upper lung, mid lung, lower lung, sternum, thymus, thoracolumbar spine, heart, liver, kidney, spleen-adrenals, and stomach. Each MOSFET detector was calibrated for the appropriate beam energy, and individual calibration factors for all 20 detectors were stored in a laptop computer. Detailed calibration methods and validation of MOSFET methods have been described previously by Yoshizumi et al. [8]. The lower limit of detection of absorbed dose for the MOSFETs with the Autosense patient dose verification system (TN-RD-60, Best Medical) was 1.50 mgy. The bismuth thyroid shield was 15 cm long, 9 cm wide, and 1 mm thick (40 ml thickness or 4-ply) (AttenuRad, JRT Associates). We attached B Fig. 1 Anthropomorphic phantom. A, Photograph shows anthropomorphic phantom in supine position with leads attached to metal oxide semiconductor field effect transistor (MOSFET) dosimeters. Thyroid bismuth shield has been placed on phantom after obtaining scout image. Note 1-cm foam standoff pad attached to shield. B, Scout image of phantom shows z-axis coverage of neck CT protocols (double-headed arrow). Dashed lines indicate upper and lower boundaries. it to a 1-cm-thick foam standoff pad (Fig. 1). The standoff foam was used to reduce streak artifact over the thyroid [6, 9]. CT Protocols The MDCT neck protocols are summarized in Table 1. A dual source 64-MDCT scanner (Definition, Siemens Healthcare) was used for all examinations. The reference clinical protocol was helically acquired using real-time dose modulation, CARE Dose4D (Siemens Healthcare). CARE Dose4D combines two types of tube current modulation: axial tube current modulation based on the topogram image attenuation profile in the z-axis and angular tube current modulation based on an axial image during the examination in the x- and y-axis [10]. The parameters for this protocol were 120 kvp; effective tube current-seconds, 192 mas; rotation time, 1 second; pitch, 0.9; and beam collimation, mm (single tube). There were three modified protocols. The organ-based dose modulation protocol used organ-based dose modulation (X-CARE, Siemens Healthcare, adapted thoracic version) in combination with dose modula- CT Protocol ATCM OBDM ATCM With TS OBDM With TS Tube current modulation a On On On On OBDM Off On Off On Bismuth shielding No No Yes Yes Peak beam energy (kvp) Reference tube current (mas) Scan length (cm) Effective tube current (mas) CTDI vol (mgy) DLP (mgy cm) Note ATCM = automatic tube current modulation, OBDM = organ-based dose modulation, TS = thyroid shield, CTDI vol = volume CT dose index, DLP = dose-length product. a CARE Dose4D (GE Healthcare) with tube current modulation in the x-, y-, and z-axis. When used in combination with OBDM, tube current modulation only along the z-axis was used. AJR:198, May

3 Hoang et al. tion, CARE Dose4D. When using CARE Dose4D with organ-based dose modulation, only the z-axis dose modulation of the CARE Dose4D software was used in combination with the 120 reduction in tube output along the radial arc in the x- and y- planes. For this protocol, the parameters were 120 kvp; effective tube current-seconds, 186 mas; rotation time, 1 second; pitch 0.6; and beam collimation, mm (single tube). For the organbased dose modulation protocol, the dose reduction arc was selected for the anterior portion of the neck to reduce the dose to the thyroid gland and ocular lens. The reference (automatic tube current modulation) protocol and organ-based dose modulation protocols were repeated using a bismuth thyroid shield placed over the lower neck to cover the thyroid gland after the scout images were obtained. These were labeled as the automatic tube current modulation with thyroid shield and organ-based dose modulation with thyroid shield protocols, respectively. The latter protocol was performed to evaluate the additional dose reduction possible with combined techniques. For all protocols, the scan length was constant at 24 cm, covering from sella turcica to the bottom of the aortic arch, and the FOV was constant at 22 cm (Fig. 1). The reconstruction algorithm was 2.5-mm thickness with 2.5-mm increments using a soft-tissue algorithm filter of B31s. The volume CT dose index (CTDI) and dose-length product were matched for each protocol to assess the difference in absorbed organ doses with a constant overall radiation exposure. The phantom was scanned three times for each of the four A Fig. 2 Noise measurements. A C, Noise measurements were obtained from three axial levels at upper neck (just below mandible) (A), thyroid gland (B), and upper mediastinum (C). ROI = region of interest. protocols to obtain the mean and SD of absorbed organ doses. Radiation Dose The mean absorbed organ doses were obtained from the MOSFET dosimeters. The mean organ doses for the modified protocols were compared with the automatic tube current modulation (ATCM) protocol (reference protocol). Comparison was made as absolute values and as percentage organ dose reduction. The equation for percentage dose reduction was [(ATCM protocol) OBDM (modified protocol)] 100 ATCM protocol where ACTM is automatic tube current modulation and OBDM is organ-based dose modulation. The effective dose (ED), an equivalent uniform dose to the entire body, was calculated from the measured organ dose (D) by applying tissueweighting factors (W T ) using publication number 103 of the International Commission on Radiologic Protection (ICRP) [11], and by assuming a radiation-weighting factor (W R ) of 1.0 for x-rays. The effective dose is computed by the following equation: ED = i W T i H i = W R i W T i D i where H i is the equivalent dose for organ i. The effective dose for the organ-based dose modulation, automatic tube current modulation with thyroid shield, and organ-based dose modulation with thyroid shield protocols were compared with the automatic tube current modulation protocol as absolute values and as percentage organ dose reduction. B Image Quality Assessment Image quality was evaluated by quantitative and subjective analysis using an Advantage workstation (GE Healthcare). For quantitative analysis, a circular 1-cm 2 region of interest (ROI) was placed in the soft tissues of the anterior and posterior neck on three selected axial images at the level of the upper neck (just below the mandible), mid neck (at the level of the thyroid gland), and lower neck-upper mediastinum as shown in Figure 2. Noise measurements were made on each of the three CT scans for each protocol by recording the SD of the attenuation value (in Hounsfield units) for the ROIs. The anterior and posterior aspects of the neck were selected because organ-based dose modulation and thyroid shields could result in increased image noise in the region of the anterior neck. For qualitative assessment, images were reviewed independently by two fellowship-trained neuroradiologists with 8 and 18 years of experience. The images were reviewed in soft-tissue window settings (window level, 40 HU; window width, 300 HU). Image quality was assessed on the same three selected axial levels used for quantitative assessment and were chosen to be in a region separated from the MOSFET detectors to exclude any change in image quality due to the presence of the detectors. Readers compared image quality for noise and streak artifact for the modified protocols with images from the automatic tube current modulation protocol and graded images on a 5-point scale. Assessment was performed by placing two images side by side on the same screen, with the anterior neck and chest masked for all images to C 1134 AJR:198, May 2012

4 Radiation Dose Reduction for Neck CT remove identification of the presence or absence of the bismuth shields. The automatic tube current modulation protocol was identified, but the readers were blinded to the modified protocol. On a 5-point scale, in comparison with the automatic tube current modulation protocol, a score of 5 represented much better image quality with markedly less noise and streak artifact, a score of 4 represented slightly better image quality with slightly less noise and streak artifact, a score of 3 represented no significant difference in image quality with similar noise and artifact, a score of 2 represented slightly inferior image quality with slightly more noise and streak artifact, and a score of 1 represented markedly inferior image quality with markedly more noise and streak artifact. Statistical Analysis Data were entered into a Microsoft Excel, version , spreadsheet. Statistical analyses were performed using Enterprise (version 4.2, SAS Institute) and the R program (R Development Core Team). By scanning the phantom with each protocol three times, we were able to obtain dose data for calculation of the mean organ dose and mean effective dose measurements for each protocol. The mean organ doses for the organ-based dose modulation, organ-based dose modulation with thyroid shield, and automatic tube current modulation with thyroid shield protocols were compared with the organ doses for the reference automatic tube current modulation protocol using a two-factor analysis of variance with all possible pairwise interactions. The cervical spine was used as a reference organ for treatment-type interaction contrasts. The noise values for the modified protocols were also compared with the reference automatic tube current modulation protocol using a three-factor analysis of variance with all possible interactions. The three factors were the type of protocol, side (anterior or posterior), and location (above the thyroid shield, thyroid, mediastinum). A two-tailed p value of less than 0.05 was considered statistically significant. Results Radiation Dose Comparison For the automatic tube current modulation protocol, the organ with the highest dose was the thyroid gland followed by the ocular lens, the bone marrow of the sternum, and the skin over the anterior neck (Table 2). These are all anteriorly located organs. The thyroid gland dose was reduced by 29% with organ-based dose modulation, by 33% with automatic tube current modulation with thyroid shield, and by 45% with organ-based dose modulation with thyroid shield (p 0.005) (Table 2). Using automatic tube current modulation as the base protocol, significant reductions in dose were observed for only one organ (thyroid, p < 0.001) when using the automatic TABLE 2: Mean Organ Dose and Effective Dose (ED) Measured by MOSFET Dosimeters Organ ATCM (mgy) Dose (mgy) tube current modulation with thyroid shield protocol. For the organ-based dose modulation protocol, significant reductions were observed for four organs (both ocular lenses, skin, and thyroid). For the organ-based dose modulation with thyroid shield protocol, significant reductions were observed for the same four organs as organ-based dose modulation plus the sternum (p = ). There was no statistically significant change in the dose to cervical spine bone marrow with organ-based dose modulation compared with the automatic tube current modulation protocol. However, use of organ-based dose modulation resulted in an increased dose to some posterior structures: the upper lungs (p = 0.01) and the brain (p = 0.02). The calculated effective dose did not change when scanning the neck with the organ-based dose modulation protocol compared with automatic tube current modulation. The automatic tube current modulation with thyroid shield and organ-based dose modulation with thyroid shield protocols resulted in 15% and 21% reduction in the calculated effective dose, respectively. Image Quality Assessment Table 3 shows the noise measurements for the different protocols. By analysis of variance analysis, 91% of the total variability in noise OBDM ATCM With TS OBDM With TS p Reduction (%) Dose (mgy) p Reduction (%) Dose (mgy) p Reduction (%) Anterior structures Anterior skin (overlying the chin) (1.51) 9.72 (1.99) (0.17) (0.56) Thyroid (1.70) (3.00) (2.44) < (2.05) < Right ocular lens (3.80) (2.91) (2.59) (0.56) Left ocular lens (2.96) (1.39) (1.75) (1.42) Bone marrow mandible (4.10) (0.68) (1.80) (1.76) Bone marrow sternum (1.29) (2.35) (2.86) (2.43) Posterior structures Brain (cerebellum) (2.41) (1.21) (1.75) (3.63) Lung upper (1.96) (0.50) (2.52) (1.01) Central structures Bone marrow cervical spine (4.17) (1.80) (3.10) (1.15) Esophagus (4.00) (2.08) (3.53) (0.69) Thymus (2.19) (0.91) (1.92) (1.90) Effective dose 3.69 (0.14) 3.51 (0.26) (0.16) (0.15) Note Reported p values correspond to the significance of estimated effects in the two-way analysis of variance model. The corresponding analysis of variance table and estimated effect values are shown in Table 4. Data in parentheses are SD. Only the 10 organs with the highest radiation doses (mgy) are shown; all other organs measured had doses of less than 8 mgy. Changes with dose reduction techniques of organ-based dose modulation (OBDM) and thyroid shield (TS) are shown as percentage dose reduction with positive values indicating dose reduction compared with the automated tube current modulation (ATCM) protocol. AJR:198, May

5 Hoang et al. was attributed to the anterior or posterior location (6%), level in the neck (57%), or the interaction between these two factors (28%). By contrast, only 5% of the variability was due to the type of protocol or its interaction with these factors. There was no statistically significant difference in overall noise between the modified protocols compared with the automatic tube current modulation reference protocol (p, ). The detailed three-way analysis of variance table and estimated effect values are shown in Table 4. By subjective assessment, both neuroradiologists graded all images as grade 3 on the 5-point scale, indicating no appreciable difference in image quality with the use of organ-based dose modulation or thyroid shields or both compared with the automatic tube current modulation protocol (Fig. 3). Discussion Organ-based dose modulation is a new radiation dose reduction technique for MDCT scanners that was originally intended for thoracic and brain imaging for dose reduction to the breast and ocular lens, respectively [7]. To the best of our knowledge, organ-based dose modulation has not been evaluated for radiation dose reduction for MDCT neck examinations. This study shows that both organ-based dose modulation and thyroid shielding result in significant dose reductions to the thyroid A C D Fig. 3 Images at thyroid level for automatic tube current modulation (A), organ-based dose modulation (B), automatic tube current modulation with thyroid shield (C), and organ-based dose modulation with thyroid shield (D) protocols. Two readers found no difference in subjective image quality. B gland compared with automatic tube current modulation. Additionally, organ-based dose modulation (in contrast to the use of thyroid bismuth shields) reduces dose to the ocular lens without impairing subjective image quality or increasing overall image noise. To date, dose reduction for MDCT neck imaging has been reported with the use of automatic tube current modulation and thyroid bismuth shields. Organ-based dose modulation is a variant of tube current modulation, but there is a paucity of literature evaluating the effect of organ-based dose modulation. The only other work in the literature is by Vollmar and Kalender [7], who evaluated a preclinical version of this software (known as partial scanning) during which the tube current is turned off completely during part of the image acquisition. Those authors compared the partial scanning technique for thoracic imaging to the full scanning technique using bismuth breast shields and showed similar reductions in breast dose for the partial scanning technique (48%) and breast bismuth shielding (47%) [7]. Our study for neck CT shows similar dose savings to the thyroid gland for organ-based dose modulation and bismuth shields: 28% for organ-based dose modulation compared with 33% for the automatic tube current modulation with thyroid shield protocol. An advantage of organ-based dose modulation compared with thyroid shields is additional dose reduction to other anterior structures, such as the ocular lens and sternum. In adult and pediatric patients, the dose reduction to the bone marrow and the ocular lens is important because of the risk of radiationinduced cataracts and leukemia, respectively [12 14]. Furthermore, from a practical perspective, organ-based dose modulation may be easier to implement than thyroid shields in a busy radiology department because the use of thyroid shields requires the technologist to enter the scanning room to place the shield over the patient after the scout images are acquired [15]. Combining both of these TABLE 3: Quantitative Mean Noise Measurements in the Anterior and Posterior Neck Mean Noise at Neck Site ATCM OBDM ATCM With TS OBDM With TS Upper neck Anterior 4.65 (0.32) 5.08 (0.40) 4.86 (0.21) 4.99 (0.48) Posterior 5.91 (0.44) 6.57 (0.48) 5.56 (0.91) 6.09 (0.55) Mid neck Anterior 4.48 (0.23) 4.66 (0.49) 5.68 (0.77) 6.00 (0.51) Posterior 5.24 (0.18) 5.14 (0.35) 5.73 (0.25) 5.46 (0.24) Lower neck Anterior (1.37) (1.87) (1.47) (1.01) Posterior 7.43 (0.63) 6.40 (0.31) 7.39 (0.50) 7.13 (0.26) Mean of three levels Anterior 7.02 (0.43) 7.22 (0.71) 8.45 (0.66) 7.75 (0.48) Posterior 6.19 (0.18) 6.04 (0.13) 6.23 (0.31) 6.23 (0.26) Note Data are mean (SD). ATCM = automatic tube current modulation, OBDM = organ-based dose modulation, TS = thyroid shield, CTDI vol = volume CT dose index, DLP = dose-length product AJR:198, May 2012

6 Radiation Dose Reduction for Neck CT TABLE 4: Three-Way Analysis of Variance of Image Noise Parameter Degrees of Freedom Sum of Squares dose reduction techniques (organ-based dose modulation and thyroid shields) may be beneficial in pediatric patients because of their greater stochastic risk from radiation exposure [16]. Our results show that the use of thyroid shields with organ-based dose modulation can almost halve the thyroid organ dose compared with automatic tube current modulation, as well as lower the absorbed dose to the lens of the eye and the bone marrow. An important consideration for CT dose reduction techniques is the effect on image quality. In our study, readers found no difference in subjective image quality assessment for organ-based dose modulation compared with the automatic tube current modulation protocol. Furthermore, organ-based dose modulation did not increase the image noise, even when noise was evaluated separately for the anterior and posterior neck. These findings agree with the study by Vollmar and Kalender [7] of partial scanning in the thorax and other studies for neck CT examination using automatic tube current modulation [3 5, 7]. For protocols that used a bismuth shield, there were no significant differences in subjective evaluation of image quality and only slight increases in image noise (Fig. 3). Prior studies have criticized thyroid bismuth shields for producing significant streak artifact; this effect is most pronounced when the shields are folded or undulate across the patient s body [3, 17]. Our study did not show any increase in streak artifacts from thyroid shields, and we attribute this to the use of a 1-cm-thick foam standoff pad with the manufacturer s bismuth shield. A foam pad has also been used with breast shields and acts to elevate the shield from the patient, thereby displacing streak artifact from the anterior thorax [18, 19]. A potential disadvantage of organ-based dose modulation is the increase in organ dose Mean of Squares F p Variability (%) Protocol Side Location Protocol:side Protocol:location Side:location Protocol:side:location Residuals Total to posterior neck structures because the output of the primary beam may be increased over the posterior 240 arc. Vollmar and Kalender [7] found that partial scanning in the thorax increased the dose to the posteriorly located spinal bone marrow by 15 20%. Although the bone marrow located anteriorly may have a decreased dose, the dose to the bone marrow in total may be increased because a large proportion of the bone marrow lies in the spine. Our study found no increase in dose to the cervical spine, possibly because the detector was placed in the vertebral body, making it more centrally located. The absorbed dose did increase in other posteriorly located organs, such as the upper lung (29%) and cerebellum (20%). Additional sampling of bone marrow along the entire spine may allow better evaluation of this effect. The effective dose remained similar with organ-based dose modulation compared with the automatic tube current modulation protocol, reflecting the balance of change in organ dose with change in tube current along the arc of the patient (i.e., tube current decreases in the anterior aspect and increases in the posterolateral aspect) with a constant central CTDI. The protocols using thyroid shields resulted in moderate decreases in effective dose. There are several limitations to our study. First, only one phantom size was used. Evaluation of this technique should also be studied in larger phantoms or pediatric phantoms because the effect of organ-based dose modulation on radiation dose may differ significantly depending on body size. The posterior beam tube current may vary by different magnitudes in large versus small patients to maintain constant image quality. Second, although we did not find differences in subjective image quality, there are limitations to evaluating image quality in phantoms because iodinated contrast media cannot be administered and the small neck structures cannot be individually evaluated. Third, the current version of organ-based dose modulation is prescribed as a 120 angle and was selected because it was the preset angle for organ-based dose modulation used for thoracic imaging. In imaging the chest, the objective for dose reduction was similar to the neck: to reduce dose to the anterior structures. Future research may be able to determine the ideal radial arc of dose reduction for neck structures. Furthermore, we used only a single scanner for our experiments because at the time of the study organ-based dose modulation was only available on one of our CT scanners. We recognize that there are other techniques to reduce dose, such as lowering the peak kilovoltage or reference tube current, but we elected to keep these factors constant to isolate the effect of organ-based dose modulation on dose reduction. Finally, individual organ doses obtained in our study are point-location specific and may not represent the overall absorbed dose for the entire organs; this is due to the point-dose measurement method used with the MOSFET detectors, i.e., an inherent limitation in the maximal number of detectors that can be placed in the organs. In conclusion, both organ-based dose modulation and thyroid shields reduce the thyroid organ dose without degradation of image quality, but organ-based dose modulation has additional benefits of dose reduction to the ocular lens. Combining thyroid shields with organ-based dose modulation further reduces thyroid organ dose and effective dose and could be considered for pediatric patients who are more sensitive to the long-term risks of radiation exposure. References 1. Brenner DJ, Hall EJ. Computed tomography: an increasing source of radiation exposure. N Engl J Med 2007; 357: Brenner D, Elliston C, Hall E, Berdon W. Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR 2001; 176: Leswick DA, Hunt MM, Webster ST, Fladeland DA. Thyroid shields versus z-axis automatic tube current modulation for dose reduction at neck CT. Radiology 2008; 249: Lee EJ, Lee SK, Agid R, Howard P, Bae JM, ter- Brugge K. Comparison of image quality and radiation dose between fixed tube current and combined automatic tube current modulation in craniocervical CT angiography. AJNR 2009; 30: AJR:198, May

7 Hoang et al. 5. Russell MT, Fink JR, Rebeles F, Kanal K, Ramos M, Anzai Y. Balancing radiation dose and image quality: clinical applications of neck volume CT. AJNR 2008; 29: Hohl C, Wildberger JE, Suss C, et al. Radiation dose reduction to breast and thyroid during MDCT: effectiveness of an in-plane bismuth shield. Acta Radiol 2006; 47: Vollmar SV, Kalender WA. Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations. Eur Radiol 2008; 18: Yoshizumi TT, Goodman PC, Frush DP, et al. Validation of metal oxide semiconductor field effect transistor technology for organ dose assessment during CT: comparison with thermoluminescent dosimetry. AJR 2007; 188: Lee YH, Park ET, Cho PK, et al. Comparative analysis of radiation dose and image quality between thyroid shielding and unshielding during ct examination of the neck. AJR 2011; 196: Lee CH, Goo JM, Ye HJ, et al. Radiation dose modulation techniques in the multidetector CT era: from basics to practice. RadioGraphics 2008; 28: International Commission on Radiological Protection recommendations of the International Commission on Radiological Protection (publication no. 103). Ann ICRP 2007; 37: Berrington de González A, Mahesh M, Kim KP, et al. Projected cancer risks from computed tomographic scans performed in the United States in Arch Intern Med 2009; 169: National Research Council; Committee to Assess Health Risks from Exposure to Low Level of Ionizing Radiation. Health risks from exposure to low levels of ionizing radiation: BEIR VII phase 2. Washington, DC: National Academies Press, Hein E, Rogalla P, Klingebiel R, Hamm B. Lowdose CT of the paranasal sinuses with eye lens protection: effect on image quality and radiation dose. Eur Radiol 2002; 12: Coursey C, Frush DP, Yoshizumi T, Toncheva G, Nguyen G, Greenberg SB. Pediatric chest MDCT using tube current modulation: effect on radiation dose with breast shielding. AJR 2008; 190:244; [web]w54 W Mazonakis M, Tzedakis A, Damilakis J, Gourtsoyiannis N. Thyroid dose from common head and neck CT examinations in children: is there an excess risk for thyroid cancer induction? Eur Radiol 2007; 17: Gunn ML, Kanal KM, Kolokythas O, Anzai Y. Radiation dose to the thyroid gland and breast from multidetector computed tomography of the cervical spine: does bismuth shielding with and without a cervical collar reduce dose? J Comput Assist Tomogr 2009; 33: Hurwitz LM, Yoshizumi TT, Goodman PC, et al. Radiation dose savings for adult pulmonary embolus 64-MDCT using bismuth breast shields, lower peak kilovoltage, and automatic tube current modulation. AJR 2009; 192: Fricke BL, Donnelly LF, Frush DP, et al. In-plane bismuth breast shields for pediatric CT: effects on radiation dose and image quality using experimental and clinical data. AJR 2003; 180: FOR YOUR INFORMATION The AJR has made getting the articles you really want really easy with an online tool, Really Simple Syndication, available at It s simple. Click the RSS button located in the menu on the right side of the page. You ll be on your way to syndicating your AJR content in no time AJR:198, May 2012

Radiation Dose Estimations to the Thorax Using Organ-Based Dose Modulation

Radiation Dose Estimations to the Thorax Using Organ-Based Dose Modulation Cardiopulmonary Imaging Original Research Lungren et al. Organ-Based Dose Modulation Cardiopulmonary Imaging Original Research Matthew P. Lungren 1 Terry T. Yoshizumi 1 Samuel M. Brady 1 Greta Toncheva

More information

Paediatric Dose Reduction and Image Quality

Paediatric Dose Reduction and Image Quality Paediatric Dose Reduction and Image Quality Alan Whiteside The majority of this work was undertaken as part of MSc Thesis of Helen Dixon. Introduction Paediatric CT protocols result in a higher effective

More information

Comparative Analysis of Radiation Dose and Image Quality Between Thyroid Shielding and Unshielding During CT Examination of the Neck

Comparative Analysis of Radiation Dose and Image Quality Between Thyroid Shielding and Unshielding During CT Examination of the Neck Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Young Hen Lee 1 Eun-tae Park 1 Pyong Kon Cho 1 Hyung Suk Seo 1 Bo-Kyung Je 1 Sang-il Suh 2 Kyung-Sook

More information

Ask EuroSafe Imaging Tips & Tricks. CT Working Group

Ask EuroSafe Imaging Tips & Tricks. CT Working Group Ask EuroSafe Imaging Tips & Tricks CT Working Group Organ Based Tube Current Modulation to Reduce Radiation Dose to Superficial Radiosensitive Organs Eileen Kelly (Galway University Hospitals, IE) Matthias

More information

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Lincoln L. Berland, M.D., F.A.C.R. Michael V. Yester, Ph.D. University of Alabama at Birmingham Breast Radiation on CT Use of chest CT

More information

To Shield or Not to Shield? Lincoln L. Berland, M.D.

To Shield or Not to Shield? Lincoln L. Berland, M.D. To Shield or Not to Shield? Lincoln L. Berland, M.D. Disclosures Consultant to: Nuance, Inc. Page 2 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation.

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

Translating Protocols Across Patient Size: Babies to Bariatric

Translating Protocols Across Patient Size: Babies to Bariatric Translating Protocols Across Patient Size: Babies to Bariatric Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose Ask EuroSafe Imaging Tips & Tricks CT Working Group Optimization of scan length to reduce CT radiation dose Alban Gervaise (Centre Hospitalier Universitaire Nancy, FR) Mika Kortesniemi (HUS Medical Imaging

More information

To Shield or Not to Shield: Application of Bismuth Breast Shields

To Shield or Not to Shield: Application of Bismuth Breast Shields Cardiopulmonary Imaging Opinion Colletti et al. Bismuth Breast Shields Cardiopulmonary Imaging Opinion FOCUS ON: Patrick M. Colletti 1 Orlando A. Micheli Kai H. Lee Colletti PM, Micheli OA, Lee KH Keywords:

More information

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Poster No.: C-2584 Congress: ECR 2015 Type: Scientific

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

CT examination is a high-radiation-dose imaging technique

CT examination is a high-radiation-dose imaging technique ORIGINAL RESEARCH J.S.P. Tan K.-L. Tan J.C.L. Lee C.-M. Wan J.-L. Leong L.-L. Chan Comparison of Eye Lens Dose on Neuroimaging Protocols between 16- and 64-Section Multidetector CT: Achieving the Lowest

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Lens Dose in Routine Head CT: Comparison of Different Optimization Methods With Anthropomorphic Phantoms

Lens Dose in Routine Head CT: Comparison of Different Optimization Methods With Anthropomorphic Phantoms Medical Physics and Informatics Original Research Nikupaavo et al. Lens Dose in Head CT Medical Physics and Informatics Original Research Ulla Nikupaavo 1,2 Touko Kaasalainen 1,3 Vappu Reijonen 1 Sanna-Mari

More information

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial Medical Physics and Informatics Original Research Larke et al. Estimated Radiation Dose for Low-Dose Chest CT Medical Physics and Informatics Original Research Frederick J. Larke 1 Randell L. Kruger 2

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School OUTLINE 1 st Presentation: Dose optimization strategies

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Debra Pennington, MD Director of Imaging Dell Children s Medical Center Debra Pennington, MD Director of Imaging Dell Children s Medical Center 1 Gray (Gy) is 1 J of radiation energy/ 1 kg matter (physical quantity absorbed dose) Diagnostic imaging doses in mgy (.001 Gy)

More information

Pediatric Imaging Original Research

Pediatric Imaging Original Research Pediatric Imaging Original Research Podberesky et al. Dose Estimates of Pediatric Cardiac CT Angiography Using 320-MDCT Scanner Pediatric Imaging Original Research Daniel J. Podberesky 1 Erin Angel 2 Terry

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Assessment of effective dose in paediatric CT examinations

Assessment of effective dose in paediatric CT examinations Assessment of effective dose in paediatric CT examinations E. Dougeni 1,2 CL. Chapple 1, J. Willis 1, G. Panayiotakis 2 1 Regional Medical Physics Department, Freeman Hospital, Freeman Road, Newcastle

More information

CT Radiation Risks and Dose Reduction

CT Radiation Risks and Dose Reduction CT Radiation Risks and Dose Reduction Walter L. Robinson, M.S. D.A.B.S.N.M., D.A.B.M.P., D.A.B.R. Consultant Certified Medical Radiation Health & Diagnostic Imaging Physicist Medical Radiation and Children

More information

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

A more accurate method to estimate patient dose during body CT examinations with tube current modulation A more accurate method to estimate patient dose during body CT examinations with tube current modulation Poster No.: C-0738 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Kawaguchi 1, Y. Matsunaga

More information

How to Develop CT Protocols for Children

How to Develop CT Protocols for Children How to Develop CT Protocols for Children Introduction Prior to 2001 the vast majority of CT imaging of children was conducted using the same or similar techniques used for adult imaging. In 2001, several

More information

Ask EuroSafe Imaging. Tips & Tricks. Paediatric Imaging Working Group. Shielding in pediatric CT

Ask EuroSafe Imaging. Tips & Tricks. Paediatric Imaging Working Group. Shielding in pediatric CT Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Shielding in pediatric CT Claudio Granata (IRCCS Istituto Giannina Gaslini, IT) Joana Santos (ESTeSC-Coimbra Health School, PT) Elina

More information

Radiation Dose Reduction Strategies in Coronary CT Angiography

Radiation Dose Reduction Strategies in Coronary CT Angiography Radiation Dose Reduction Strategies in Coronary CT Angiography Noor Diyana Osman, PhD noordiyana@usm.my Contents: Introduction Radiation dosimetry in CT Radiation risk associated with coronary CT angiography

More information

CT Radiation Safety in Adults: Where are we now? What can be done?

CT Radiation Safety in Adults: Where are we now? What can be done? CT Radiation Safety in Adults: Where are we now? What can be done? David A Leswick MD FRCPC University of Saskatchewan Canadian Association of Radiologists Annual General Meeting Friday April 26, 2013

More information

Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation

Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation Medical Physics and Informatics Original Research Angel et al. Radiation Dose During Chest CT Medical Physics and Informatics Original Research Erin Angel 1,2 Nazanin Yaghmai 1 Cecilia Matilda Jude 1 John

More information

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report www.siemens.com/healthcare SOMATOM Drive System Owner Manual Dosimetry and imaging performance report Table of contents 1 Dosimetry and imaging performance report 5 1.1 Dose information 5 1.1.1 General

More information

Breast Dose Reduction Options During Thoracic CT: Influence of Breast Thickness

Breast Dose Reduction Options During Thoracic CT: Influence of Breast Thickness Cardiopulmonary Imaging Original Research Revel et al. Breast Dose Reduction in Thoracic CT Cardiopulmonary Imaging Original Research Marie-Pierre Revel 1 Isabelle Fitton 2 Etienne Audureau 3 Joseph Benzakoun

More information

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell CT Optimisation for Paediatric SPECT/CT Examinations Sarah Bell Sarah.bell14@nhs.net Outline 1. Introduction 2. Aims and Objectives 3. Methods 4. Results 5. Discussion 6. Conclusions 7. References Introduction

More information

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Alice B. Smith, MD

More information

Skyscan 1076 in vivo scanning: X-ray dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in

More information

Radiation Dosimetry for CT Protocols

Radiation Dosimetry for CT Protocols Radiation Dosimetry for CT Protocols This document contains radiation dosimetry information from CT scans and can be used by investigators to estimate the dosimetry information required by the JRSC or

More information

Optimizing radiation dose by varying age at pediatric temporal bone CT

Optimizing radiation dose by varying age at pediatric temporal bone CT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Optimizing radiation dose by varying age at pediatric temporal bone CT Daichi Noto, 1 Yoshinori Funama, 2a Mika Kitajima, 3 Daisuke

More information

Austin Radiological Association Ga-68 NETSPOT (Ga-68 dotatate)

Austin Radiological Association Ga-68 NETSPOT (Ga-68 dotatate) Austin Radiological Association Ga-68 NETSPOT (Ga-68 dotatate) Overview Ga-68 dotatate binds to somatostatin receptors, with highest affinity for subtype 2 receptors (sstr2). It binds to cells that express

More information

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols Personalization of Pediatric Imaging in Terms of Needed Indication-Based Quality Per Dose Acknowledgments Duke University Medical Center Ehsan Samei, PhD Donald Frush, MD Xiang Li PhD DABR Cleveland Clinic

More information

STRUCTURED EDUCATION REQUIREMENTS EFFECTIVE: JANUARY 1, 2016

STRUCTURED EDUCATION REQUIREMENTS EFFECTIVE: JANUARY 1, 2016 Computed Tomography The purpose of structured education is to provide the opportunity for individuals to develop mastery of discipline-specific knowledge that, when coupled with selected clinical experiences,

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

The Computed Tomography Examination

The Computed Tomography Examination CONTENT SPECIFICATIONS The Computed Tomography Examination The purpose of The American Registry of Radiologic Technologists (ARRT ) Computed Tomography Examination is to assess the knowledge and cognitive

More information

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

B. CT protocols for the spine

B. CT protocols for the spine B. CT protocols for the spine Poster No.: A-003 Congress: ECR 2010 Type: Invited Speaker Topic: Neuro Authors: B. Tins; Oswestry/UK Keywords: CT, spine, diagnostic imaging protocol DOI: 10.1594/ecr2010/A-003

More information

Dose Reduction of the Female Breast in Chest CT

Dose Reduction of the Female Breast in Chest CT Cardiopulmonary Imaging Original Research Seidenfuss et al. Reduction of CT Radiation Dose to Female Breast Cardiopulmonary Imaging Original Research Downloaded from www.ajronline.org by 46.3.197.9 on

More information

CT Dose Optimization for Whole- Body PET/CT Examinations

CT Dose Optimization for Whole- Body PET/CT Examinations Nuclear Medicine and Molecular Imaging Original Research Tonkopi et al. CT Dose Optimization for PET/CT Nuclear Medicine and Molecular Imaging Original Research FOCUS ON: Elena Tonkopi 1,2 Andrew A. Ross

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

Austin Radiological Association Nuclear Medicine Procedure PET SODIUM FLUORIDE BONE SCAN (F-18 NaF)

Austin Radiological Association Nuclear Medicine Procedure PET SODIUM FLUORIDE BONE SCAN (F-18 NaF) Austin Radiological Association Nuclear Medicine Procedure PET SODIUM FLUORIDE BONE SCAN (F-18 NaF) Overview Indication Sodium Fluoride F18 injection is a radioactive diagnostic agent for positron emission

More information

CT Dosimetry in the Clinical Environment: Methods and Analysis

CT Dosimetry in the Clinical Environment: Methods and Analysis CT Dosimetry in the Clinical Environment: Methods and Analysis Manuel Arreola, Ph.D. DABR Associate Chair of Radiology Director, Medical Physics Graduate Program Department of Radiology University of Florida

More information

Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose)

Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose) Eur Radiol (2002) 12:1571 1576 DOI 10.1007/s00330-001-1255-4 PHYSICS Holger Greess Anton Nömayr Heiko Wolf Ulrich Baum Michael Lell Bernhard Böwing Willi Kalender Werner A. Bautz Dose reduction in CT examination

More information

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center CT Dose Estimation John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology 1 University of California Davis Medical Center CT Dose Estimation Introduction The CTDI Family of Metrics

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION Managing Pediatric CT Patient Doses Keith J. Strauss, MSc, FAAPM, FACR President X-Ray Computations, Inc. Boston, Massachusetts Acknowledgements Marilyn Goske, MD John Boone, PhD Cynthia McCollough, PhD

More information

Kerma Area Product Method for Effective Dose Estimation During Lumbar Epidural Steroid Injection Procedures: Phantom Study

Kerma Area Product Method for Effective Dose Estimation During Lumbar Epidural Steroid Injection Procedures: Phantom Study Vascular and Interventional Radiology Original Research Kim et al. Dose for Lumbar Steroid Injection Vascular and Interventional Radiology Original Research Sangroh Kim 1 Greta Toncheva 2,3 Colin Anderson-Evans

More information

Bone Densitometry Radiation dose: what you need to know

Bone Densitometry Radiation dose: what you need to know Bone Densitometry Radiation dose: what you need to know John Damilakis, PhD Associate Professor and Chairman University of Crete, Iraklion, Crete, GREECE Estimation of bone status using X-rays Assessment

More information

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Department of Medical Physics and Bioengineering, Royal United Hospital, Bath Overview Varian Acuity

More information

Pediatric CT: Strategies to Lower Radiation Dose

Pediatric CT: Strategies to Lower Radiation Dose Pediatric Imaging Review Zacharias et al. Strategies to Lower Pediatric CT Radiation Dose Pediatric Imaging Review FOCUS ON: Claudia Zacharias 1 Adam M. Alessio 2 Randolph K. Otto 3 Ramesh S. Iyer 3 Grace

More information

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017 EXAMINATION CONTENT SPECIFICATIONS Computed Tomography The purpose of the computed tomography examination is to assess the knowledge and cognitive skills underlying the intelligent performance of the tasks

More information

Absorbed Radiation Dose in Radiosensitive Organs Using 64- and 320-Row Multidetector Computed Tomography: A Comparative Study

Absorbed Radiation Dose in Radiosensitive Organs Using 64- and 320-Row Multidetector Computed Tomography: A Comparative Study Absorbed Radiation Dose in Radiosensitive Organs Using 64- and 320-Row Multidetector Computed Tomography: A Comparative Study The Harvard community has made this article openly available. Please share

More information

Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography

Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography The British Journal of Radiology, 81 (2008), 499 503 SHORT COMMUNICATION Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography 1 G R IBALL, MSc, BSc, 2

More information

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan *M.T.Bahreyni Toossi, **S.Mohandes Dastgherdi Medical Physics Dep., Faculty

More information

Dose Reduction Options in Cardiac CT

Dose Reduction Options in Cardiac CT Dose Reduction Options in Cardiac CT Doyle P, Ball P*, Donnelly P # Radiological Sciences & Imaging, Forster Green Hospital *Department of Radiology, Ulster Hospital # Department of Cardiology, Ulster

More information

Reducing Radiation Dose in Body CT: A Primer on Dose Metrics and Key CT Technical Parameters

Reducing Radiation Dose in Body CT: A Primer on Dose Metrics and Key CT Technical Parameters Medical Physics and Informatics Review Maldjian and Goldman Reducing Radiation Dose in Body CT Medical Physics and Informatics Review FOCUS ON: Pierre D. Maldjian 1 Alice R. Goldman Maldjian PD, Goldman

More information

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction?

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Poster No.: C-0096 Congress: ECR 2012 Type: Scientific Paper Authors: M. Mazonakis,

More information

Radiation Exposure 1980 to 2006

Radiation Exposure 1980 to 2006 Radiation Exposure 1980 to 2006 Background 3-6 msv/yr Natural (85% 45%) Radon Cosmic Rays Air travel Living at Altitude Man-made (15% 55%) Medical Imaging** mgy Radiation Therapy cgy Radiation Whole Body

More information

Patient / Organ Dose in CT

Patient / Organ Dose in CT Patient / Organ Dose in CT Patient specific and organ dose estimation H.D. Nagel Dr. HD Nagel, Science & Technology for Radiology Buchholz / Germany www.sascrad.com 1 Topics CTDI & patient dose SSDE Organ

More information

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017 STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017 Computed Tomography The purpose of structured education is to provide the opportunity for individuals to develop mastery of discipline-specific

More information

IMAGE GENTLY HOW CAN YOU HELP?

IMAGE GENTLY HOW CAN YOU HELP? IMAGE GENTLY HOW CAN YOU HELP? Keith J. Strauss, MSc, FAAPM, FACR Director, Radiology Physics & Engineering Children s s Hospital Boston Harvard Medical School Acknowledgment Marilyn J. Goske,, MD Robert

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Rebecca Milman Marsh, Ph.D. University of Colorado Department of Radiology Who in the Facility Works

More information

CT Dose Reduction in Pediatric Patients

CT Dose Reduction in Pediatric Patients CT Dose Reduction in Pediatric Patients By Kelly Firestine, RT(R)(CT)(M) Executive Summary CT is an incredibly valuable imaging tool, but there are unique concerns with pediatric patients, including the

More information

Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT

Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT protocols Tips Leukaemia / lymphoma: ~ 35% acute lymphoblastic

More information

Optimization of kvp and mas for Pediatric Low-Dose Simulated Abdominal CT: Is It Best to Base Parameter Selection on Object Circumference?

Optimization of kvp and mas for Pediatric Low-Dose Simulated Abdominal CT: Is It Best to Base Parameter Selection on Object Circumference? Pediatric Imaging Original Research Reid et al. Parameter Selection for Pediatric Abdominal CT Downloaded from www.ajronline.org by 46.3.3.24 on 2/3/18 from IP address 46.3.3.24. Copyright ARRS. For personal

More information

CT Radiation Dosimetry Study Using Monte Carlo Simulation and. Computational Anthropomorphic Phantoms

CT Radiation Dosimetry Study Using Monte Carlo Simulation and. Computational Anthropomorphic Phantoms CT Radiation Dosimetry Study Using Monte Carlo Simulation and Computational Anthropomorphic Phantoms by Yakun Zhang Graduate Program in Medical Physics Duke University Date: Approved: Ehsan Samei, Supervisor

More information

Recent technical advances have greatly increased the clinical

Recent technical advances have greatly increased the clinical ORIGINAL RESEARCH E.J. Lee S.K. Lee R. Agid P. Howard J.M. Bae K. terbrugge Comparison of Image Quality and Radiation Dose between Fixed Tube Current and Combined Automatic Tube Current Modulation in Craniocervical

More information

Effects of shielding the radiosensitive superficial organs of ORNL pediatric phantoms on dose reduction in computed tomography

Effects of shielding the radiosensitive superficial organs of ORNL pediatric phantoms on dose reduction in computed tomography Original Article Effects of shielding the radiosensitive superficial organs of ORNL pediatric phantoms on dose reduction in computed tomography Parisa Akhlaghi, Hashem Miri Hakimabad, Laleh Rafat Motavalli

More information

Medical Physics and Informatics Original Research

Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Christner et al. Estimating Effective Dose for CT Medical Physics and Informatics Original Research FOCUS ON: Jodie A. Christner 1 James M. Kofler Cynthia

More information

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean Doses from Cervical Spine Computed Tomography (CT) examinations in the UK John Holroyd and Sue Edyvean Why a new dose survey? Number of enquires received concerning the current NDRL Concern that could

More information

Introduction and Background

Introduction and Background CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section

More information

Pediatric CT Protocols (18 years old or less)

Pediatric CT Protocols (18 years old or less) Pediatric CT Protocols (18 years old or less) Ped1: Head CT Ped2: Cervical spine CT Ped3: Sinus CT Ped4: Neck CT Ped5: Chest CT Ped6: Abdomen and pelvis CT Ped7: Thoracic or lumbar spine CT Ped8: Extremity

More information

Low Dose Era in Cardiac CT

Low Dose Era in Cardiac CT Low Dose Era in Cardiac CT DIANA E. LITMANOVICH, MD Department of Radiology Beth Israel Deaconess Medical Center Harvard Medical School Disclosures Neither I nor my immediate family members have a financial

More information

Seattle Children s Hospital Radiology Department. Statement regarding radiation exposure related to computed. tomography (CT) exams

Seattle Children s Hospital Radiology Department. Statement regarding radiation exposure related to computed. tomography (CT) exams Seattle Children s Hospital Radiology Department Statement regarding radiation exposure related to computed tomography (CT) exams Computed tomography (CT) scanners use radiation in the form of X- rays

More information

FDG-18 PET/CT - radiation dose and dose-reduction strategy

FDG-18 PET/CT - radiation dose and dose-reduction strategy FDG-18 PET/CT - radiation dose and dose-reduction strategy Poster No.: C-1856 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit P. Nicholson, S. McSweeney, K. O'Regan; Cork/IE Radiation

More information

Patient Radiation Doses from Adult and Pediatric CT

Patient Radiation Doses from Adult and Pediatric CT Radiation Doses from CT CT Imaging Original Research Walter Huda 1 Awais Vance 1,2 Huda W, Vance A Keywords: MDCT, pediatric CT, physics of radiology, radiation dose DOI:.2214/AJR.06.01 Received January

More information

Why is CT Dose of Interest?

Why is CT Dose of Interest? Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase

More information

Chief Radiographer TEI Clinical Associate 2016

Chief Radiographer TEI Clinical Associate 2016 MDCT Principles i and Applications Ε ΑGADAKOS MSc Ε. ΑGADAKOS MSc Chief Radiographer TEI Clinical Associate 2016 Aim To understand d recent technological advances in MSCT and how they can be effectively

More information

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Learning Objectives: Limitations for estimating patient dose for CT Methods

More information

Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT

Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT Mannudeep K. Kalra, MD, DNB Michael M. Maher, MD, FRCR, FFR (RCSI) Srinivasa R.

More information

Estimating Testicular Cancer specific Mortality by Using the Surveillance Epidemiology and End Results Registry

Estimating Testicular Cancer specific Mortality by Using the Surveillance Epidemiology and End Results Registry Appendix E1 Estimating Testicular Cancer specific Mortality by Using the Surveillance Epidemiology and End Results Registry To estimate cancer-specific mortality for 33-year-old men with stage I testicular

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

Radiation Dose in Pediatric Imaging

Radiation Dose in Pediatric Imaging Radiation Dose in Pediatric Imaging A Brief History of Radiology Dose: Why Does It Matter? Measuring Exposure and Dose Deterministic Effects Stochastic Effects Common Exams: What is the Risk? Reducing

More information

Ultralow Dose Chest CT with MBIR

Ultralow Dose Chest CT with MBIR Ultralow Dose Chest CT with MBIR Ella A. Kazerooni, M.D. Professor & Director Cardiothoracic Radiology Associate Chair for Clinical Affairs University of Michigan Disclosures Consultant: GE Healthcare

More information

Estimating Patient Radiation Dose from Computed Tomography

Estimating Patient Radiation Dose from Computed Tomography Estimating Patient Radiation Dose from Computed Tomography C. Cagnon, J. DeMarco, E. Angel, M. McNitt-Gray UCLA David Geffen School of Medicine 1 Patient Dose from CT Advances in Technology... Helical

More information

CT SCAN PROTOCOL. Shoulder

CT SCAN PROTOCOL. Shoulder CT SCAN PROTOCOL Shoulder Purpose and Summary CT images made with this protocol are used to provide the orthopedic surgeon with a detailed 3D anatomical reconstruction of the patient s scapula and proximal

More information

Combined Anatomical and Functional Imaging with Revolution * CT

Combined Anatomical and Functional Imaging with Revolution * CT GE Healthcare Case studies Combined Anatomical and Functional Imaging with Revolution * CT Jean-Louis Sablayrolles, M.D. Centre Cardiologique du Nord, Saint-Denis, France Case 1 Whole Brain Perfusion and

More information

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: Rob Loader Oliver Gosling Introduction Approached by Dr Oliver Gosling (Research

More information