Repeatability and Reproducibility of Retinal Blood Flow Measurement Using a Doppler Optical Coherence Tomography Flowmeter in Healthy Subjects

Size: px
Start display at page:

Download "Repeatability and Reproducibility of Retinal Blood Flow Measurement Using a Doppler Optical Coherence Tomography Flowmeter in Healthy Subjects"

Transcription

1 Retina Repeatability and Reproducibility of Retinal Blood Flow Measurement Using a Doppler Optical Coherence Tomography Flowmeter in Healthy Subjects Tomofumi Tani, 1 Young-Seok Song, 1 Takafumi Yoshioka, 1 Tsuneaki Omae, 1 Akihiro Ishibazawa, 1 Masahiro Akiba, 2 and Akitoshi Yoshida 1 1 Department of Ophthalmology, Asahikawa Medical University, Asahikawa, Hokkaido, Japan 2 R&D Division, Topcon Corporation, Tokyo, Japan Correspondence: Young-Seok Song, Department of Ophthalmology, Asahikawa Medical University, Midorigaoka Higashi , Asahikawa, , Japan; ysong@asahikawa-med.ac.jp. Submitted: December 29, 2016 Accepted: April 10, 2017 Citation: Tani T, Song Y-S, Yoshioka T, et al. Repeatability and reproducibility of retinal blood flow measurement using a Doppler optical coherence tomography flowmeter in healthy subjects. Invest Ophthalmol Vis Sci. 2017;58: DOI: / iovs PURPOSE. To evaluate the repeatability and reproducibility of retinal blood flow (RBF) measurements in humans by using new auto-alignment and measurement software in a commercially available Doppler optical coherence tomography (DOCT) system. METHODS. The DOCT flowmeter assessed the intrasession repeatability and the intersession and interobserver reproducibility of the RBF measurements. For intrasession repeatability, the coefficients of variation (CVs) of five repeated RBF measurements were calculated at the retinal arteries and veins in 20 normal eyes of 20 healthy volunteers. For intersession reproducibility, two sets of three measurements obtained by one observer on 2 different days were compared. For interobserver reproducibility, two sets of three measurements obtained by two observers on the same day were compared. Intraclass correlation coefficients (ICCs) also were used to evaluate the repeatability and reproducibility. The relevance of the DOCT flowmeter and laser Doppler velocimetry (LDV) also was assessed. RESULTS. Regarding intrasession repeatability, the ICC of the RBF exceeded 0.90 in arterioles and venules (ICC: and 0.970, respectively). The CVs of the RBF in the arterioles and venules were 6.0% 6 3.4% and 8.8% 6 5.1%, respectively. The intersession and interobserver RBF values had high reproducibility in the arterioles (ICC: and 0.993, respectively) and venules (ICC: and 0.986, respectively). The RBF measured with the DOCT flowmeter was correlated strongly with LDV in the arterioles (r ¼ 0.76; P < 0.001). CONCLUSIONS. The DOCT flowmeter had good reproducibility in the arterioles and venules and precisely measured the RBF as compared to the LDV in the arterioles. Keywords: retinal blood flow, Doppler OCT flowmeter, laser Doppler velocimetry, retinal blood velocity profile Evaluation of the ocular circulation is highly important for gaining an understanding of the physiologic and pathologic features in a number of retinal diseases. 1 Although many devices are used to measure retinal ocular blood flow that is, the blue field entoptic technique, microsphere method, laser Doppler flowmeter, laser Doppler velocimetry (LDV), oxygen measurements, retinal vessel analyzer, color Doppler imaging, and laser speckle technique 2 we have shown that LDV is a reliable, noninvasive, and clinically useful tool for evaluating the retinal circulation in humans, based on the absolute vessel diameter and blood velocity. 3,4 Considering these advantages, we have measured the retinal blood flow (RBF) in several retinal diseases such as diabetic retinopathy (DR) in type 1 diabetes mellitus (DM), 5 DR in type 2 DM, 6 branch retinal vein occlusion, 7 and age-related macular degeneration. 8 However, the RBF measurements obtained with the LDV method have a number of disadvantages, that is, the required RBF should be considered as the parabolic flow distribution to calculate the absolute values of blood velocity, 9 fine alignment of the laser beam, and maintenance of strict eye fixation for measurements. Therefore, it is difficult to measure the RBF at the bifurcation or the margin of the optic nerve head because of uncertainty regarding the parabolic flow distribution. LDV also is limited to use in cooperative patients. A novel velocimetry technique using optical coherence tomography (OCT) technology, namely, Doppler OCT (DOCT), has been developed recently OCT can detect not only morphologic images but also a Doppler shift of reflected light, which provides information about flow and movement. 13,14 Although the technique facilitates measurements of the absolute RBF values in humans, with good repeatability, those DOCT methods have some shortcomings. In such methods, it is possible to measure the retinal venous blood flow velocity but not the arterial velocity. Another research group 15 has reported the results with the swept-source based OCT flowmeter in which both the arterial and venous total flow are assessed by using their prototype OCT. We have previously developed a DOCT instrument with novel software, referred to as a segmental-scanning method, 16 which enables simultaneous measurement of the RBF in the retinal arterioles and adjacent venules during one cardiac cycle. We have previously reported the accuracy of the measurements in in vitro glass capillaries. We then evaluated the reproducibility of the blood velocity measurements in the retinal arterioles and Copyright 2017 The Authors iovs.arvojournals.org j ISSN: This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.

2 Newly Developed Doppler OCT Flowmeter IOVS j June 2017 j Vol. 58 j No. 7 j 2892 venules in anesthetized cats. Recently, a new DOCT instrument has been developed that can be used in a clinical setting, namely, the DOCT flowmeter (Yoshida A, et al. IOVS 2016;57:ARVO E-Abstract 5922). The goal of the current study was to determine the repeatability and reproducibility of RBF measurements with the DOCT flowmeter in healthy subjects and the relevance of the arteriolar data when the DOCT flowmeter was compared with the LDV. MATERIALS AND METHODS Subjects and Ethics A total of 20 healthy volunteers (7 men, 13 women; age range, years) with no systemic or ophthalmic disease with mild refractive errors (1.00 and 3.75 diopters of spherical equivalent) participated in this study. This study adhered to the guidelines approved by the ethics committee of our institution and the tenets of the Declaration of Helsinki. Each subject provided written informed consent before enrollment after receiving a complete explanation of the study design and protocol. All volunteers underwent an ophthalmologic examination, including review of the medical history, slit-lamp biomicroscopy, intraocular pressure (IOP) measurement, and funduscopic examination using a 90-diopter lens. The axial length of each eye was measured with an A-mode ultrasound system (IOL- Master 500; Carl Zeiss Meditec, Jena, Germany). The mean arterial blood pressure (MABP) and heart rate (HR) were measured with an electronic sphygmomanometer (EP-88Si; Colin, Tokyo, Japan). The IOP was monitored by applanation tonometry (Haag Streit, Bern, Switzerland). The ocular perfusion pressure was calculated as 2/3MABP IOP. The pupil was dilated with one drop of 0.4% tropicamide (Mydrin M; Santen Pharmaceuticals, Osaka, Japan). The subjects were asked to abstain from drinking coffee and alcohol and smoking for 24 hours before the examination. Each subject rested for 10 minutes in a quiet room before the examination began. After the pupils were dilated, the subjects were required to rest in the sitting position for at least 5 minutes in a quiet dimly lit room in which the temperature was maintained at 258C, after which their BP and HR were measured in the left arm. The RBF measurements were assessed after the BP, HR, and IOP measurements. The relatively large arteries (>80 lm and sufficiently far from the bifurcations) chosen for measurement had relatively straight segments one disc diameter away from the optic disc. RBF Measurements An LDV system (Canon Laser Blood Flowmeter, model CLBF 100; Canon, Tokyo, Japan) was used to measure the blood flow in the superior branch of the first-order major temporal retinal vessels. The detailed system methodology has been described previously. 9 Briefly, the retinal LDV system allows noninvasive measurement of the absolute values of the red blood cells flowing in the centerline of the vessel, based on bidirectional LDV. 9 The mean retinal blood velocity (V mean ) was defined as the V of the averaged maximal speed during one cardiac cycle. Computer analysis of the signal produced by the arterial image on the array sensor automatically determined the retinal artery diameter (D). 9 The D measurements are corrected for the axial length (operator input) and refractive error of the eye, which is measured by the CLBF itself. The RBF was calculated as RBF ¼ V mean 3 Area, where V mean is calculated as V mean ¼ V of the averaged maximal speed/2, and area is the cross-sectional area of the retinal artery at the LDV measurement site. 9 Our DOCT flowmeter is based on a commercially available spectral-domain OCT system (3D OCT-1 Maestro; Topcon Corp., Tokyo, Japan) operated at an 800-nm wavelength range. 16 Only the software was modified for blood flow imaging, where extra image-processing software and quantification software were newly developed to measure blood flow. In DOCT, the flow velocity v(z) can be derived from the Doppler shift incurred by the moving blood: vz ¼ DUðz; sþ k 0 4p n s 1 cos h; ð1þ where z is the depth location, DU is the phase difference at the same depth location between the adjacent profiles after Fourier transform, k0 is the center wavelength, n is the refractive index of blood, s is the time interval between the adjacent profiles, and h is the Doppler angle between the flow vector and the incident probe beam. Besides measurement of the phase difference, the Doppler angle h must be known to calculate the velocity by using Equation 1. A detailed description has been published previously. 16 In the retina, most blood vessels run nearly parallel to the retinal surface except at the area around the optic nerve head. As h approaches 908, the measured velocity becomes very sensitive to the accuracy of h. To minimize this potential velocity measurement error, our Doppler blood flow measurements were performed where the Doppler angle h is considerably less than 908, that is, ~808. Automated alignment software was integrated to seek the proper Doppler angle. Such auto-alignment software traces the boundary of the internal limiting membrane in real time along the blood vessel to determine the Doppler angle. If the Doppler angle was out of appropriate range, the OCT beam was shifted farther away from the pupillary center. This alignment was repeated until the estimated Doppler angle was calculated in a certain range. The accurate Doppler angle estimation routine then was run where a pair of B-scan images was captured with a difference of 100 lm across the blood vessels. This measurement was repeated eight times to define the Doppler angle where it was chosen by using the median value among the eight repeated measurements. Since the time interval for capturing a pair of B- scan images is only 10 ms, most of the eye motion can be ignored. Obtaining the median value among eight repeated imaging makes our measurement more reliable and accurate for rejecting the unwanted value. DOCT imaging was performed for 2 seconds, followed by color fundus imaging. The blood vessels were detected automatically and identified from the OCT structured and phase images. The vessel diameter also was measured by using OCT phase images. Protocols To assess intrasession repeatability, 20 eyes of 20 subjects were recruited. Measurement of the right or left eye was determined by a random number table. One experienced observer (TY, observer 1) performed the DOCT flowmeter measurements for each subject. The RBF measurements of the arterioles and venules in either the superior or inferior temporal sites were repeated five times by using the follow-up mode in the software. The observer was masked to the results of the retinal circulatory parameters. The coefficients of variation (CVs) for the retinal circulatory parameters, retinal vessel D, maximal and averaged retinal blood velocity during one cardiac cycle (V maximum and V average ), and RBF were determined. The intersession reproducibility of the DOCT flowmeter measurements was evaluated. The same observer performed three measurements on 2 different days at each of the 20

3 Newly Developed Doppler OCT Flowmeter IOVS j June 2017 j Vol. 58 j No. 7 j 2893 arterioles and venules (days 1 and 2). Both measurements were performed in the morning under fasting conditions. The observer was masked to the results of the retinal circulatory parameters. The correlations for the different days of the D, V average, V maximum, and RBF were determined. The interobserver reproducibility of the DOCT flowmeter measurements was evaluated. Observer 1 and another inexperienced observer (TO, observer 2) measured the same sites of the same 20 eyes three times on the same day. A minimum of 10 minutes was allowed between measurements. The observers were masked to the results of the retinal circulatory parameters. The correlation between the two observers measurements was determined and the differences in the D, V average, V maximum, and RBF were determined. In the last part of this study, the correlations between the DOCT flowmeter and LDV were evaluated in the arterioles in the same 20 subjects. After an experienced observer (Y-SS, observer 3) performed the LDV measurements, observer 1 performed the DOCT flowmeter measurements at the same vessel sites of the LDV on the same day as those measured by experienced observer 4 (TT). The observers were masked to the results of the retinal circulatory parameters. A minimum of 10 minutes was allowed between the LDV and DOCT flowmeter measurements. The correlation of the measurements between the two different methods was determined in the RBF. Statistical Analysis All data are expressed as the mean 6 standard deviation (SD). Pearson s correlation analysis, determination of the CV, and Bland-Altman plotting were performed to assess the validity and reproducibility. 17 Intergroup differences were analyzed by using the paired t-test. P < 0.05 was considered significant. To evaluate the agreement in the repeatability of the intrasession measurements and the reproducibility of the intersession measurements and the interobserver measurements, the intraclass correlation coefficients (ICCs) one-way random model for the intrasession measurements and the two-way mixed model for the intersession and interobserver measurements with their 95% confidence intervals on absolute agreement were used to analyze measurement reliability. 18 The ICCs could range from 0 to 1, with a higher value indicating better reliability. An ICC above 0.9 indicated acceptable clinical repeatability. 19 IBM SPSS Statistics for Windows, version 24.0 (IBM Corp., Armonk, NY, USA) was used for statistical analyses. RESULTS Subjects and Images of DOCT Flowmeter Table 1 summarizes the systemic parameters of the population on the 2 study days. There were no significant differences in any systemic parameters between the 2 days. Figure 1 shows the imaging results from a healthy young subject, using our DOCT flowmeter. The OCT intensity and phase images are shown in Figures 1A and 1B, where the phase image was represented as a pseudocolor code. We confirmed that the blood velocity profiles of the superior temporal retinal arterioles and venules were obtained simultaneously by the DOCT flowmeter in a healthy young subject (Figs. 1C, 1D). Intrasession Repeatability Table 2 summarizes the results of intrasession repeatability of the arterioles and venules measured by DOCT flowmeter (n ¼ 20). In the arterioles, the Ds ranged from 80.9 to lm (mean, 94.5 lm). The V averages ranged from 9.2 to 21.3 mm/s (mean, 15.5 mm/s). The V maximums ranged from 18.1 to 44.4 TABLE 1. Systemic Parameters of the Population on the 2 Study Days Measurements Day 1, n ¼ 20 Day 2, n ¼ 20 P Value MABP, mm Hg IOP, mm Hg OPP, mm Hg HR, beats/min Data are expressed as the mean 6 SD.OPP,ocularperfusionpressure. mm/s (mean, 30.7 mm/s). The RBFs ranged from 3.2 to 14.2 ll/min (mean, 7.8 ll/min). The ICCs of the D, V average, V maximum, and RBF exceeded 0.9 (0.977, 0.957, 0.943, and 0.994, respectively). The CVs for the D, V average, V maximum, and RBF of the measured arterioles were 3.4% 6 2.1%, 8.8% 6 6.2%, 10.8% 6 3.9%, and 6.0% 6 3.4%, respectively. In the venules, the Ds ranged from 98.9 to lm (mean, lm). The V averages ranged from 5.6 to 21.2 mm/s (mean, 13.1 mm/s). The V maximums ranged from 7.2 to 26.3 mm/s (mean, 16.1 mm/s). The RBFs ranged from 4.9 to 19.4 ll/min (mean, 9.8 ll/min). The ICCs of the D, V average, V maximum, and RBF exceeded 0.9 (0.973, 0.977, 0.982, and 0.944, respectively). The CVs for the D, V average, V maximum, and RBF were 4.2% 6 2.7%, 8.8% 6 3.3%, 8.5% 6 3.6%, and 8.8% 6 5.1%, respectively. Intersession Reproducibility The DOCT flowmeter measurements obtained by the same examiner on 2 different days in the same 20 arterioles and venules were compared. The arteriolar correlations among the D, V average, V maximum, and RBF are shown in Figure 2A. Pearson s correlation analysis showed that the arteriolar circulatory parameters on day 1 were correlated strongly with those of day 2, that is, D (r ¼ 0.99; P < ), V average (r ¼ 0.90; P < ), V maximum (r ¼ 0.85; P < ), and RBF (r ¼ 0.92; P < ). In the Bland-Altman plots of intersession measurements, most values ranged within a mean 6 2 SDs of all retinal circulatory parameters (Fig. 2A). The ICCs of the D, V average, V maximum, and RBF exceeded 0.9 (0.986, 0.904, 0.919, and 0.980, respectively) (Table 3). The venular correlations among the D, V average, V maximum, and RBF measured by the DOCT flowmeter are shown in Figure 2B. Pearson s correlation analysis also showed a strong correlation in measurements between day 1 and day 2, that is, D(r ¼ 0.98; P < ), V average (r ¼ 0.98; P < ), V maximum (r ¼ 0.96; P < ), and RBF (r ¼ 0.98; P < ). In the Bland-Altman plots of intersession measurements, most values ranged within a mean 6 2 SDs of all retinal circulatory parameters (Fig. 2B). The ICCs of the D, V average, V maximum, and RBF exceeded 0.9 (0.975, 0.983, 0.962, and 0.982, respectively) (Table 3). Interobserver Reproducibility The DOCT flowmeter measurements performed by two observers on the same day at the same 20 arterioles and venules were compared. The arteriolar correlations among the D, V average, V maximum, and RBF are shown in Figure 3A. Pearson s correlation analysis showed that the arteriolar circulatory parameters of observer 1 were correlated strongly with those of observer 2, that is, D (r ¼ 0.97; P < ), V average (r ¼ 0.93; P < ), V maximum (r ¼ 0.96; P < ), and RBF (r ¼ 0.99; P < ). In the Bland-Altman plots of intersession measurements, most values ranged within a mean 6 2 SDs of all retinal circulatory parameters (Fig. 3A). The ICCs of the D, V average, V maximum, and RBF exceeded 0.9 (0.974, 0.929, 0.959, and 0.993, respectively) (Table 3).

4 Newly Developed Doppler OCT Flowmeter IOVS j June 2017 j Vol. 58 j No. 7 j 2894 FIGURE 1. Simultaneous measurement of the RBF in the retinal arterioles (A) and venules (V) in a healthy young subject. (A) A fundus image. The green bar indicates the measured location. (B) A phase image with color coding. (C) A velocity profile image of the red blood cells in the horizontal center of an artery at a given moment. (D) Blood velocity profiles of the retinal arterioles and venules measured simultaneously by DOCT flowmeter. TABLE 2. Results of Intrasession Repeatability of Arterioles and Venules Measured on DOCT Flowmeter Arterioles (n ¼ 20) Venules (n ¼ 20) Measurements Value ICC Value ICC CVs, % CVs, % Diameter, lm Averaged velocity, mm/s Maximum velocity, mm/s Blood flow, ll/min Data are expressed as the mean 6 SD.

5 Newly Developed Doppler OCT Flowmeter IOVS j June 2017 j Vol. 58 j No. 7 j 2895 FIGURE 2. (A) The left panels show the correlation of the D, V average, V maximum, and RBF between two occasionally different measurements by one observer. The right panels show the correlation of the D, V average, V maximum, and RBF in the arterioles in the Bland-Altman plots. (B) The left panels show the correlation of the D, V average, V maximum, and RBF between two occasionally different measurements by one observer; and the right panels show the correlation of the D, V average, V maximum, and RBF in the venules in the Bland-Altman plots. The venular correlations between the D, V average, V maximum, and RBF are shown in Figure 3B. Pearson s correlation analysis showed that the arteriolar circulatory parameters of observer 1 were correlated strongly with those of observer 2, that is, D (r ¼ 0.99; P < ), V average (r ¼ 0.98; P < ), V maximum (r ¼ 0.97; P < ), and RBF (r ¼ 0.99; P < ). In the Bland-Altman plots of intersession measurements, most values ranged within a mean 6 2 SDs of all retinal circulatory parameters (Fig. 3B). The ICCs of the D, V average, V maximum, and RBF exceeded 0.9 (0.986, 0.979, 0.973, and 0.986, respectively) (Table 3). Comparison of the DOCT Flowmeter and LDV Results in the Arterioles The RBF measured with the DOCT flowmeter ( ll/ min) was similar to that obtained by LDV ( ll/min) (paired t-test, P ¼ 0.35). The strong correlation (r ¼ 0.76; P <

6 Newly Developed Doppler OCT Flowmeter IOVS j June 2017 j Vol. 58 j No. 7 j 2896 TABLE 3. Reproducibility of Intersession and Interobserver of Arterioles and Venules Intersession Interobserver Measurements A (n ¼ 20) V (n ¼ 20) A (n ¼ 20) V (n ¼ 20) ICC ICC ICC ICC Diameter Averaged velocity Maximum velocity Blood flow A, arterioles; V, venules ) between the RBF measurements (n ¼ 20) is shown in Figure 4. DISCUSSION The current study showed the good repeatability of the intrasession measurements and the reproducibility of the intersession and interobserver measurements of the RBF, using the newly developed DOCT flowmeter in healthy subjects. This study also found that absolute RBF values measured with the DOCT flowmeter were similar to the RBF values of the same arterioles measured by LDV. The RBF measurements using the DOCT flowmeter and LDV also were correlated significantly, indicating that the performance of the DOCT flowmeter is equivalent to that of LDV (Fig. 4). In addition, in our previous LDV study, 4 the CVs of the RBF for intrasession repeatability reported are 13.8% in the arterioles and 12.7% in the venules. Regarding the current DOCT flowmeter, the CVs of the RBF were 6.0% in the arterioles and 8.8% in the venules (Table 2). The Bland-Altman plots showed that the deviations of the arterioles and venules in the RBF values were good in the plots of the intersession and interobserver measurements (Figs. 2, 3). These results also indicated that the DOCT flowmeter provides an accurate representation of the RBF. Until the advent of the DOCT technique, LDV was the only commercially available instrument that measured the absolute value of the RBF in the arterioles and venules. Werkmeister et al. 20 showed for the first time that the DOCT (dual-beam bidirectional Doppler Fourier-domain OCT) measurements of the human venules are well correlated with those obtained with LDV (LDV-5000; Oculix, Inc., Arbaz, Switzerland). No studies have investigated the correlation in the human retinal arterioles. Therefore, the current findings showed for the first time that the DOCT flowmeter measurements of the human arterioles are well correlated with those of the LDV (Fig. 4). No commercially available instrument has used the DOCT technique yet. Although the DOCT flowmeter is an experimental prototype, we are in the process of developing a commercially available product. The LDV method can measure only the maximal blood velocity. In this technique, the blood flow in the retinal vessels should be considered as laminar flow to calculate the absolute values of blood velocity. 9 It is difficult to measure the bifurcation or the margin of the optic nerve head where the flow is considered to be nonlaminar in those areas. In addition, since the LDV technique requires strict focusing on the target vessel, it must be performed by a well-trained examiner. Therefore, there are some difficult situations when measuring the velocity with LDV. However, since the DOCT technique can detect not only the maximal blood velocity but also the FIGURE 3. (A) The left panels show the correlation of the D, V average, V maximum, and RBF between two independent measurements by two observers; and the right panels show the correlation of the D, V average, V maximum, and RBF in the arterioles in the Bland-Altman plots. (B) The left panels show the correlation of the D, V average, V maximum, and RBF between two independent measurements by two observers; and the right panels show the correlation of the D, V average, V maximum, and RBF in the venules in the Bland-Altman plots. entire blood velocity, it does not require parabolic flow definition. There is no restriction on the measurement vessel site. Our DOCT flowmeter system required appropriate detection of the Doppler angle (<868 in target vessels). 16 It is difficult to manually find the proper Doppler angle. We overcame the angle problem by using an automatic setting of the proper Doppler angle using a combination of OCT intensity and phase images. 16 The DOCT flowmeter does not require any specialized technique for observation. After the simple preparations to set the head on the instrument (as with the standard OCT measurements) are completed, the observer selects the target vessels. The RBF is measured automatically. The simplicity of this method makes it suitable for screening large populations and for clinical use. These are the major advantages of the DOCT flowmeter. Another advantage of the DOCT flowmeter is the ability to analyze the blood flow pattern, that is, determination of the degree of laminar flow. Laminar flow is a type of blood flow in which the concentric layers of blood move in parallel through the vessel. The maximal velocity occurs in the center of the vessel and the minimal velocity is at the vessel wall. 21 The laminar flow is associated with shear stress, which is important for regulating blood flow. 22,23 The DOCT flowmeter can visualize the retinal blood velocity profile (RBVP), which is the velocity profile of the red blood cells in the horizontal center of a vessel at a given moment (Fig. 1C). Recently, we first showed an abnormal retinal arterial RBVP during the diastolic phase in a patient with Takayasu s arteritis and aortic insufficiency (Tani T, et al. IOVS 2015;56:ARVO E-Abstract 5892). These results indicate that the aortic valve function is related to the retinal arterial circulation. Therefore, not only absolute RBF but also RBVP analyses are expected to elucidate retinal circulatory disturbances. The DOCT flowmeter has some limitations. First, if the Doppler angle exceeds 868 at the measured vessels, even though the incident beam was shifted automatically, the RBF was not considered reliable in the DOCT flowmeter measurement. Second, in some clinical cases, that is, when a vessel is close to the optic disc or in patients with small pupils, it might be difficult to find the proper Doppler angle. Third, the number of vessels detected in a single scan is limited to a couple of vessels when using the 1-mm scan length. Even when the scan length was expanded to 4 mm, not all vessels

7 Newly Developed Doppler OCT Flowmeter IOVS j June 2017 j Vol. 58 j No. 7 j 2897 FIGURE 4. The correlation between the RBF as assessed with DOCT flowmeter and LDV, respectively, in the arterioles. could be covered in one scan. If the total blood flow measurement is required, multiple DOCT scans are necessary to cover all the vessels. Finally, scanning also might be difficult when there is an opacity in the eye, that is, a vitreous hemorrhage or hard cortical cataract. In conclusion, the current study confirmed the reproducibility of intraobserver, intersession, and interobserver measurements in a newly developed DOCT flowmeter and that the DOCT flowmeter can precisely measure the RBF, compared with the LDV. These data suggest that the DOCT flowmeter is acceptable for measuring the RBF measurements in humans. Thus, the DOCT flowmeter has potential for clinical use and screening RBF changes in a large population. Acknowledgments The authors thank Lynda Charters for manuscript review and also thank Jun Sakai and Shunsuke Nakamura for technical support. Supported by a Grant-in-Aid for Scientific Research from the Ministry of Education, Culture, Sports, Science, and Technology in Japan (15K20242 to TT and 16K20296 to TY). Disclosure: T. Tani, None; Y.-S. Song, None; T. Yoshioka, None; T. Omae, None; A. Ishibazawa, None; M. Akiba, None; A. Yoshida, None References 1. Pournaras CJ, Rungger-Brandle E, Riva CE, Hardarson SH, Stefansson E. Regulation of retinal blood flow in health and disease. Prog Retin Eye Res. 2008;27: Schmetterer L, Kiel J. Ocular Blood Flow. Berlin, Germany: Springer-Verlag Berlin Heidelberg; 2012: Nagaoka T, Ishii Y, Takeuchi T, Takahashi A, Sato E, Yoshida A. Relationship between the parameters of retinal circulation measured by laser Doppler velocimetry and a marker of early systemic atherosclerosis. Invest Ophthalmol Vis Sci. 2005;46: Yoshida A, Feke GT, Mori F, et al. Reproducibility and clinical application of a newly developed stabilized retinal laser Doppler instrument. Am J Ophthalmol. 2003;135: Konno S, Feke GT, Yoshida A, Fujio N, Goger DG, Buzney SM. Retinal blood flow changes in type I diabetes: a long-term follow-up study. Invest Ophthalmol Vis Sci. 1996;37: Nagaoka T, Sato E, Takahashi A, Yokota H, Sogawa K, Yoshida A. Impaired retinal circulation in patients with type 2 diabetes mellitus: retinal laser Doppler velocimetry study. Invest Ophthalmol Vis Sci. 2010;51: Nagaoka T, Sogawa K, Yoshida A. Changes in retinal blood flow in patients with macular edema secondary to branch retinal vein occlusion before and after intravitreal injection of bevacizumab. Retina. 2014;34: Sato E, Feke GT, Menke MN, Wallace McMeel J. Retinal haemodynamics in patients with age-related macular degeneration. Eye (Lond). 2006;20: Nagaoka T, Yoshida A. Noninvasive evaluation of wall shear stress on retinal microcirculation in humans. Invest Ophthalmol Vis Sci. 2006;47: Doblhoff-Dier V, Schmetterer L, Vilser W, et al. Measurement of the total retinal blood flow using dual beam Fourierdomain Doppler optical coherence tomography with orthogonal detection planes. Biomed Opt Express. 2014;5: Leitgeb RA, Werkmeister RM, Blatter C, Schmetterer L. Doppler optical coherence tomography. Prog Retin Eye Res. 2014;41: Wang Y, Bower BA, Izatt JA, Tan O, Huang D. Retinal blood flow measurement by circumpapillary Fourier domain Doppler optical coherence tomography. J Biomed Opt. 2008;13:

8 Newly Developed Doppler OCT Flowmeter IOVS j June 2017 j Vol. 58 j No. 7 j Wang XJ, Milner TE, Nelson JS. Characterization of fluid flow velocity by optical Doppler tomography. Opt Lett. 1995;20: Zhao Y, Chen Z, Saxer C, Xiang S, de Boer JF, Nelson JS. Phaseresolved optical coherence tomography and optical Doppler tomography for imaging blood flow in human skin with fast scanning speed and high velocity sensitivity. Opt Lett. 2000; 25: Baumann B, Potsaid B, Kraus MF, et al. Total retinal blood flow measurement with ultrahigh speed swept source/fourier domain OCT. Biomed Opt Express. 2011;2: Nagaoka T, Tani T, Song YS, et al. Evaluation of retinal circulation using segmental-scanning doppler optical coherence tomography in anesthetized cats. Invest Ophthalmol Vis Sci. 2016;57: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1: Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86: Kramer MS, Feinstein AR. Clinical biostatistics, LIV: the biostatistics of concordance. Clin Pharmacol Ther. 1981;29: Werkmeister RM, Palkovits S, Told R, et al. Response of retinal blood flow to systemic hyperoxia as measured with dualbeam bidirectional Doppler Fourier-domain optical coherence tomography. PLoS One. 2012;7:e Guyton AaJH. Textbook of Medical Physiology. Philadelphia, PA: W.B. Saunders Company; 2000: Ishibazawa A, Nagaoka T, Takahashi T, et al. Effects of shear stress on the gene expressions of endothelial nitric oxide synthase, endothelin-1, and thrombomodulin in human retinal microvascular endothelial cells. Invest Ophthalmol Vis Sci. 2011;52: Korenaga R, Ando J, Tsuboi H, et al. Laminar flow stimulates ATP- and shear stress-dependent nitric oxide production in cultured bovine endothelial cells. Biochem Biophys Res Commun. 1994;198:

Retinal blood flow reduction after panretinal photocoagulation in Type 2 diabetes mellitus: Doppler optical coherence tomography flowmeter pilot study

Retinal blood flow reduction after panretinal photocoagulation in Type 2 diabetes mellitus: Doppler optical coherence tomography flowmeter pilot study RESEARCH ARTICLE Retinal blood flow reduction after panretinal photocoagulation in Type 2 diabetes mellitus: Doppler optical coherence tomography flowmeter pilot study Youngseok Song ID 1 *, Tomofumi Tani

More information

Retinal tissue has an intrinsic ability to maintain a relatively

Retinal tissue has an intrinsic ability to maintain a relatively Special Issue Retinal Blood Flow Response to Hyperoxia Measured With En Face Doppler Optical Coherence Tomography Alex D. Pechauer, Ou Tan, Liang Liu, Yali Jia, Vivian Hou, William Hills, and David Huang

More information

CLINICAL SCIENCES. The Effect of Ocular Warming on Ocular Circulation in Healthy Humans 1-3 RECENTLY

CLINICAL SCIENCES. The Effect of Ocular Warming on Ocular Circulation in Healthy Humans 1-3 RECENTLY The Effect of Ocular Warming on Ocular Circulation in Healthy Humans Taiji Nagaoka, MD, PhD; Akitoshi Yoshida, MD, PhD CLINICAL SCIENCES Objective: To examine the effect of ocular warming on retinal blood

More information

Reproducibility of Scanning Laser Doppler Flowmetry in the Rat Retina and Optic Nervehead

Reproducibility of Scanning Laser Doppler Flowmetry in the Rat Retina and Optic Nervehead Reproducibility of Scanning Laser Doppler Flowmetry in the Rat Retina and Optic Nervehead Akitaka Tsujikawa,* Junichi Kiryu,* Atsushi Nonaka,* Kenji Yamashiro,* Yuichiro Ogura and Yoshihito Honda* *Department

More information

OCT Angiography: The Next Step in Retinal Imaging Jonathan Zelenak D.O.

OCT Angiography: The Next Step in Retinal Imaging Jonathan Zelenak D.O. OCT Angiography: The Next Step in Retinal Imaging Jonathan Zelenak D.O. Hillsdale Hospital Michigan State University Overview Evolution of OCT How does OCT angiography work? Clinical examples Potential

More information

Real-Time Blood Velocity Measurements in Human Retinal Vein Using the Laser Speckle Phenomenon

Real-Time Blood Velocity Measurements in Human Retinal Vein Using the Laser Speckle Phenomenon Real-Time Blood Velocity Measurements in Human Retinal Vein Using the Laser Speckle Phenomenon Miyuki Nagahara,* Yasuhiro Tamaki, Makoto Araie* and Hitoshi Fujii *Department of Ophthalmology, University

More information

Clinical Study Choroidal Thickness in Eyes with Unilateral Ocular Ischemic Syndrome

Clinical Study Choroidal Thickness in Eyes with Unilateral Ocular Ischemic Syndrome Hindawi Publishing Corporation Journal of Ophthalmology Volume 215, Article ID 62372, 5 pages http://dx.doi.org/1.1155/215/62372 Clinical Study Choroidal Thickness in Eyes with Unilateral Ocular Ischemic

More information

Ultrahigh Speed Imaging of the Rat Retina Using Ultrahigh Resolution Spectral/Fourier Domain OCT

Ultrahigh Speed Imaging of the Rat Retina Using Ultrahigh Resolution Spectral/Fourier Domain OCT Ultrahigh Speed Imaging of the Rat Retina Using Ultrahigh Resolution Spectral/Fourier Domain OCT The MIT Faculty has made this article openly available. Please share how this access benefits you. Your

More information

Corporate Medical Policy

Corporate Medical Policy Corporate Medical Policy Optical Coherence Tomography (OCT) Anterior Segment of the Eye File Name: Origination: Last CAP Review: Next CAP Review: Last Review: optical_coherence_tomography_(oct)_anterior_segment_of_the_eye

More information

NIH Public Access Author Manuscript Br J Ophthalmol. Author manuscript; available in PMC 2010 May 1.

NIH Public Access Author Manuscript Br J Ophthalmol. Author manuscript; available in PMC 2010 May 1. NIH Public Access Author Manuscript Published in final edited form as: Br J Ophthalmol. 2009 May ; 93(5): 634 637. doi:10.1136/bjo.2008.150276. Measurement of total blood flow in the normal human retina

More information

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Original article Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Dewang Angmo, 1 Shibal Bhartiya, 1 Sanjay K Mishra,

More information

Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis OCT

Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis OCT pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2011;25(3):166-173 DOI: 10.3341/kjo.2011.25.3.166 Original Article Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis

More information

CLINICAL SCIENCES. Repeatability and Reproducibility of Fast Macular Thickness Mapping With Stratus Optical Coherence Tomography

CLINICAL SCIENCES. Repeatability and Reproducibility of Fast Macular Thickness Mapping With Stratus Optical Coherence Tomography CLINICAL SCIENCES Repeatability and Reproducibility of Fast Macular Thickness Mapping With Stratus Optical Coherence Tomography Antonio Polito, MD; Michele Del Borrello, MD; Miriam Isola, MHS; Nicola Zemella,

More information

Swept-Source OCT Angiography: SS OCT Angio TM

Swept-Source OCT Angiography: SS OCT Angio TM Swept-Source OCT Angiography: SS OCT Angio TM Not available in all countries, please check with your distributor. 2015.09 Swept-Source OCT Angiography: SS OCT Angio TM Introduction Optical coherence tomography

More information

Measurement of Choroidal Thickness in Normal Eyes Using 3D OCT-1000 Spectral Domain Optical Coherence Tomography

Measurement of Choroidal Thickness in Normal Eyes Using 3D OCT-1000 Spectral Domain Optical Coherence Tomography pissn: 111-8942 eissn: 292-9382 Korean J Ophthalmol 212;26(4):255-259 http://dx.doi.org/1.3341/kjo.212.26.4.255 Original Article Measurement of Choroidal Thickness in Normal Eyes Using 3D OCT-1 Spectral

More information

Corporate Medical Policy

Corporate Medical Policy Corporate Medical Policy Glaucoma, Evaluation by Ophthalmologic Techniques File Name: Origination: Last CAP Review: Next CAP Review: Last Review: glaucoma_evaluation_by_ophthalmologic_techniques 3/2001

More information

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Original Article Philippine Journal of OPHTHALMOLOGY Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Dennis L. del Rosario, MD and Mario M. Yatco, MD University of Santo

More information

INTRODUCTION J. DAWCZYNSKI, E. KOENIGSDOERFFER, R. AUGSTEN, J. STROBEL. Department of Ophthalmology, University Hospital Jena, Jena - Germany

INTRODUCTION J. DAWCZYNSKI, E. KOENIGSDOERFFER, R. AUGSTEN, J. STROBEL. Department of Ophthalmology, University Hospital Jena, Jena - Germany European Journal of Ophthalmology / Vol. 17 no. 3, 2007 / pp. 363-367 Anterior segment optical coherence tomography for evaluation of changes in anterior chamber angle and depth after intraocular lens

More information

Effects of Oxygen and Carbon Dioxide on Human Retinal Circulation

Effects of Oxygen and Carbon Dioxide on Human Retinal Circulation Effects of ygen and on Dioxide on Human Retinal Circulation Stephen]. Pakola and Juan E. Grunwald Purpose. ogen, a gas mixture of 95% O 2 and 5% CO 2, is given to patients with retinal artery obstruction

More information

BLOOD VESSEL DIAMETER MEASUREMENT ON RETINAL IMAGE

BLOOD VESSEL DIAMETER MEASUREMENT ON RETINAL IMAGE Journal of Computer Science 10 (5): 879-883, 2014 ISSN: 1549-3636 2014 doi:10.3844/jcssp.2014.879.883 Published Online 10 (5) 2014 (http://www.thescipub.com/jcs.toc) BLOOD VESSEL DIAMETER MEASUREMENT ON

More information

Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography

Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography ORIGINAL STUDY Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography Huijuan Wu, MD, PhD,*w Johannes F. de Boer, PhD,z and Teresa C. Chen,

More information

Advances in OCT Murray Fingeret, OD

Advances in OCT Murray Fingeret, OD Disclosures Advances in OCT Murray Fingeret, OD Consultant Alcon, Allergan, Bausch & Lomb, Carl Zeiss Meditec, Diopsys, Heidelberg Engineering, Reichert, Topcon Currently Approved OCT Devices OCT Devices

More information

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma.

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma. Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma. Donald C. Hood 1,2,* and Ali S. Raza 1 1 Department of Psychology, Columbia

More information

Il contributo dell'angio-oct: valutazione integrata della componente nervosa e vascolare della malattia glaucomatosa

Il contributo dell'angio-oct: valutazione integrata della componente nervosa e vascolare della malattia glaucomatosa SIMPOSIO G.O.A.L. - LE NUOVE FRONTIERE DIAGNOSTICHE E LE LINEE DI INDIRIZZO AMBULATORIALI DEL GLAUCOMA Coordinatore e moderatore: D. Mazzacane Presidente: L. Rossetti Il contributo dell'angio-oct: valutazione

More information

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation C L I N I C A L S C I E N C E The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation Amjad Horani, MD; Shahar Frenkel, MD, PhD; Eytan Z. Blumenthal, MD BACKGROUND

More information

(Note these factors are not an issue in the current study as camera system, imaging field and resolution was consistent across images).

(Note these factors are not an issue in the current study as camera system, imaging field and resolution was consistent across images). Supplementary Discussion We identify seven potential sources of variation that will influence both software applications and may give rise to systematic differences between them. While not all factors

More information

doi: /s

doi: /s doi: 10.1186/s12886-015-0077-0 Miura et al. BMC Ophthalmology (2015) 15:79 DOI 10.1186/s12886-015-0077-0 RESEARCH ARTICLE Open Access Noninvasive vascular imaging of ruptured retinal arterial macroaneurysms

More information

Study of clinical significance of optical coherence tomography in diagnosis & management of diabetic macular edema

Study of clinical significance of optical coherence tomography in diagnosis & management of diabetic macular edema Original Research Article Study of clinical significance of optical coherence tomography in diagnosis & management of diabetic macular edema Neha Kantilal Desai 1,*, Somesh Vedprakash Aggarwal 2, Sonali

More information

Diffuse luminance flicker increases blood flow in major retinal arteries and veins

Diffuse luminance flicker increases blood flow in major retinal arteries and veins Vision Research 44 (2004) 833 838 www.elsevier.com/locate/visres Diffuse luminance flicker increases blood flow in major retinal arteries and veins G. Garh ofer a,b, C. Zawinka a, H. Resch a, K.H. Huemer

More information

A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT) Retinal Nerve Fiber Layer Measurements Based on Time Domain OCT Measurements

A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT) Retinal Nerve Fiber Layer Measurements Based on Time Domain OCT Measurements pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(5):369-377 http://dx.doi.org/10.3341/kjo.2012.26.5.369 Original Article A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT)

More information

PRIMUS 200 from ZEISS The essential OCT

PRIMUS 200 from ZEISS The essential OCT PRIMUS 200 from ZEISS The essential OCT Seeing beyond the surface. ZEISS PRIMUS 200 // INNOVATION MADE BY ZEISS Clear Visualization. Advanced Technology. Reliability. Essential elements of your first OCT.

More information

Choroidal Blood Flow Change in Eyes with High Myopia

Choroidal Blood Flow Change in Eyes with High Myopia pissn: 1011-892 eissn: 2092-9382 Korean J Ophthalmol 2015;29(5):309-31 http://dx.doi.org/10.331/kjo.2015.29.5.309 Choroidal Blood Flow Change in Eyes with High Myopia Original Article Young Seong Yang,

More information

Use of Scanning Laser Ophthalmoscope Microperimetry in Clinically Significant Macular Edema in Type 2 Diabetes Mellitus

Use of Scanning Laser Ophthalmoscope Microperimetry in Clinically Significant Macular Edema in Type 2 Diabetes Mellitus Use of Scanning Laser Ophthalmoscope Microperimetry in Clinically Significant Macular Edema in Type 2 Diabetes Mellitus Fumihiko Mori, Satoshi Ishiko, Norihiko Kitaya, Taiichi Hikichi, Eiichi Sato, Akira

More information

Laser Doppler Velocimefry Study of the Effect of Pure Oxygen Breathing on Retinal Blood Flow

Laser Doppler Velocimefry Study of the Effect of Pure Oxygen Breathing on Retinal Blood Flow Laser Doppler Velocimefry Study of the Effect of Pure Oxygen Breathing on Retinal Blood Flow C. E. Riva, J. E. Grunwald,* and S. H. Sinclair The noninvasive Laser Doppler Velocimetry (LDV) technique was

More information

This policy is applicable to Commercial Products only. For BlueCHiP for Medicare, see related policy section.

This policy is applicable to Commercial Products only. For BlueCHiP for Medicare, see related policy section. Medical Coverage Policy Ophthalmologic Techniques that Evaluate the Posterior Segment for Glaucoma EFFECTIVE DATE: 01 01 2017 POLICY LAST UPDATED: 09 18 2018 OVERVIEW Several techniques have been developed

More information

Functional OCT for Glaucoma Evaluation

Functional OCT for Glaucoma Evaluation Hawaiian Eye Meeting 18-24 January 2014 Functional OCT for Glaucoma Evaluation David Huang, MD, PhD Weeks Professor of Ophthalmic Research Professor of Ophthalmology & Biomedical Engineering Casey Eye

More information

Retinal Hemodynamics in Proliferative Diabetic Retinopathy A Laser Doppler Velocimetry Study

Retinal Hemodynamics in Proliferative Diabetic Retinopathy A Laser Doppler Velocimetry Study Retinal Hemodynamics in Proliferative Diabetic Retinopathy A Laser Doppler Velocimetry Study Juan E. Grunwald, Alexander J. Brucker, Sharon E. Grunwald, and Charles E. Riva Purpose. This study investigated

More information

Variability and Repeatability of Quantitative, Fourier- Domain Optical Coherence Tomography Doppler Blood Flow in Young and Elderly Healthy Subjects

Variability and Repeatability of Quantitative, Fourier- Domain Optical Coherence Tomography Doppler Blood Flow in Young and Elderly Healthy Subjects Retina Variability and Repeatability of Quantitative, Fourier- Domain Optical Coherence Tomography Doppler Blood Flow in Young and Elderly Healthy Subjects Faryan Tayyari, 1 Firdaus Yusof, 1,2 Michal Vymyslicky,

More information

What Is New in Optical Coherence Tomography. Ultrahigh-speed OCT makes 3D better

What Is New in Optical Coherence Tomography. Ultrahigh-speed OCT makes 3D better OOAA 66 th Meeting & CEI 20 th Birthday Casey Eye Institute, Portland, OR, 17 June 2011 What Is New in Optical Coherence Tomography David Huang, MD, PhD Weeks Professor of Ophthalmic Research Prof. of

More information

Structural examina.on: Imaging

Structural examina.on: Imaging ManaMa: Glaucoma Structural examina.on: Imaging Luís Abegão Pinto, MD, PhD Department of Ophthalmology CHLC Lisbon Faculty of Medicine, Lisbon University 1 11-10- 2013 Structural changes Qualitative changes

More information

* Şükrü Bayraktar, MD, Zerrin Bayraktar, MD, and Ömer Faruk Yilmaz, MD

* Şükrü Bayraktar, MD, Zerrin Bayraktar, MD, and Ömer Faruk Yilmaz, MD Journal of Glaucoma 10:163 169 2001 Lippincott Williams & Wilkins, Inc. Influence of Scan Radius Correction for Ocular Magnification and Relationship Between Scan Radius With Retinal Nerve Fiber Layer

More information

Effect of Timolol Maleate on the Retinal Circulation of Human Eyes with Ocular Hypertension

Effect of Timolol Maleate on the Retinal Circulation of Human Eyes with Ocular Hypertension Investigative Ophthalmology & Visual Science, Vol. 31, No. 3, March 1990 Copyright Association for Research in Vision and Ophthalmology Effect of Maleate on the Retinal Circulation of Human Eyes with Ocular

More information

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods Ophthalmology Volume 2011, Article ID 914250, 5 pages doi:10.1155/2011/914250 Clinical Study Macular Retinal Ganglion Cell Complex Thickness and Its Relationship to the Optic Nerve Head Topography in Glaucomatous

More information

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma Med. J. Cairo Univ., Vol. 83, No. 2, September: 67-72, 2015 www.medicaljournalofcairouniversity.net Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

More information

Diabetes Care 36: , 2013

Diabetes Care 36: , 2013 Pathophysiology/Complications O R I G I N A L A R T I C L E Relationship Between Retinal Blood Flow and Renal Function in Patients With Type 2 Diabetes and Chronic Kidney Disease TAIJI NAGAOKA, MD, PHD

More information

Structural changes of the anterior chamber following cataract surgery during infancy

Structural changes of the anterior chamber following cataract surgery during infancy Structural changes of the anterior chamber following cataract surgery during infancy Matthew Nguyen, Emory University Marla Shainberg, Emory University Allen Beck, Emory University Scott Lambert, Emory

More information

Optical Coherence Tomograpic Features in Idiopathic Retinitis, Vasculitis, Aneurysms and Neuroretinitis (IRVAN)

Optical Coherence Tomograpic Features in Idiopathic Retinitis, Vasculitis, Aneurysms and Neuroretinitis (IRVAN) Columbia International Publishing Journal of Ophthalmic Research (2014) Research Article Optical Coherence Tomograpic Features in Idiopathic Retinitis, Vasculitis, Aneurysms and Neuroretinitis (IRVAN)

More information

LEE EYE CENTRE. YOUR VISION, OUR PASSION LEC EyeNews

LEE EYE CENTRE. YOUR VISION, OUR PASSION LEC EyeNews LEE EYE CENTRE YOUR VISION, OUR PASSION LEC EyeNews FOR INTERNAL CIRCULATION ONLY www.lec.com.my ISSUE 51/003 SEPT OCT 2017 The American Society of Cataract and Refractive Surgery is one of the leading

More information

Yoshiro Minami 1*, Taiji Nagaoka 2, Akihiro Ishibazawa 1,2 and Akitoshi Yoshida 2

Yoshiro Minami 1*, Taiji Nagaoka 2, Akihiro Ishibazawa 1,2 and Akitoshi Yoshida 2 Minami et al. BMC Ophthalmology (2017) 17:90 DOI 10.1186/s12886-017-0485-4 RESEARCH ARTICLE Open Access Correlation between short- and long-term effects of intravitreal ranibizumab therapy on macular edema

More information

Cirrus TM HD-OCT. Details define your decisions

Cirrus TM HD-OCT. Details define your decisions Cirrus TM HD-OCT Details define your decisions 2 With high-definition OCT Carl Zeiss Meditec takes you beyond standard spectral domain Built on 10 years experience at the vanguard of innovation, Carl Zeiss

More information

Predictability and accuracy of IOL formulas in high myopia

Predictability and accuracy of IOL formulas in high myopia ARTICLE Predictability and accuracy of IOL formulas in high myopia Mohamed Yasser Sayed Saif, MD 1 ; Mohamed Othman Abdel Khalek, MD 1 ; Ahmed Tamer Sayed Saif, MD 2 ; Passant Sayed Saif, MD 3 ; Sherif

More information

PRIMUS 200 from ZEISS The essential OCT

PRIMUS 200 from ZEISS The essential OCT EN 00_00I The contents of the brochure may differ from the current status of approval of the product in your country. Please contact your regional representative for more information. Subject to change

More information

Populations Interventions Comparators Outcomes Individuals: Who are being evaluated for angleclosure

Populations Interventions Comparators Outcomes Individuals: Who are being evaluated for angleclosure Protocol Optical Coherence Tomography of the Anterior Eye Segment (90318) Medical Benefit Effective Date: 07/01/14 Next Review Date: 05/18 Preauthorization No Review Dates: 07/11, 07/12, 07/13, 05/14,

More information

Clinically Significant Macular Edema (CSME)

Clinically Significant Macular Edema (CSME) Clinically Significant Macular Edema (CSME) 1 Clinically Significant Macular Edema (CSME) Sadrina T. Shaw OMT I Student July 26, 2014 Advisor: Dr. Uwaydat Clinically Significant Macular Edema (CSME) 2

More information

R&M Solutions

R&M Solutions Mohamed Hosny El-Bradey, MD., Assistant Professor of Ophthalmology, Tanta University. Wael El Haig, MD., Professor of Ophthalmology. Zagazeeg University. 1 Myopic CNV is considered the most common vision

More information

Retinal blood flow in type 1 diabetes patients with no or mild diabetic retinopathy during euglycemic clamp

Retinal blood flow in type 1 diabetes patients with no or mild diabetic retinopathy during euglycemic clamp Diabetes Care Publish Ahead of Print, published online June 28, 2010 Retinal blood flow in type 1 diabetes patients with no or mild diabetic retinopathy during euglycemic clamp Berthold Pemp MD 1, Elżbieta

More information

Accuracy of Biometry for Intraocular Lens Implantation Using the New Partial Coherence Interferometer, AL-scan

Accuracy of Biometry for Intraocular Lens Implantation Using the New Partial Coherence Interferometer, AL-scan pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2014;28(6):444-450 http://dx.doi.org/10.3341/kjo.2014.28.6.444 Original Article Accuracy of Biometry for Intraocular Lens Implantation Using the New

More information

Pilot Study of Optical Coherence Tomography Measurement of Retinal Blood Flow in Retinal and Optic Nerve Diseases METHODS

Pilot Study of Optical Coherence Tomography Measurement of Retinal Blood Flow in Retinal and Optic Nerve Diseases METHODS Retina Pilot Study of Optical Coherence Tomography Measurement of Retinal Blood Flow in Retinal and Optic Nerve Diseases Yimin Wang, 1 Amani A. Fawzi, 1 Rohit Varma, 1 Alfredo A. Sadun, 1 Xinbo Zhang,

More information

NIH Public Access Author Manuscript Cornea. Author manuscript; available in PMC 2014 May 29.

NIH Public Access Author Manuscript Cornea. Author manuscript; available in PMC 2014 May 29. NIH Public Access Author Manuscript Published in final edited form as: Cornea. 2013 December ; 32(12): 1544 1548. doi:10.1097/ico.0b013e3182a7f39d. Repeatability of corneal epithelial thickness measurements

More information

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Eytan Z. Blumenthal, MD, 1 Julia M. Williams, BS, 1 Robert N. Weinreb, MD, 1 Christopher A. Girkin, MD,

More information

Retinal blood flow. 1.Retinal circulation. 1. Retinal circulation Ø Supplies the inner retinal layers Ø Non fenestrated capillaries

Retinal blood flow. 1.Retinal circulation. 1. Retinal circulation Ø Supplies the inner retinal layers Ø Non fenestrated capillaries www.ophtalmique.ch Glucose Retinal blood flow Pournaras Jean-Antoine C Vitreoretinal Surgery Unit Constantly regulated hemodynamics assuming delivery of and metabolic substrates, as well as intact blood-retinal

More information

The leading causes of blindness. ocular circulation: Macular Degeneration

The leading causes of blindness. ocular circulation: Macular Degeneration Taiwan Academy of Ophthalmology Taipei, March 31, 2012 Measurement of Blood Flow in the Retina and Optic Disc with OCT David Huang, MD, Weeks Professor of Ophthalmic Research Professor of Ophthalmology

More information

Diabetic Retinopathy. Barry Emara MD FRCS(C) Giovanni Caboto Club October 3, 2012

Diabetic Retinopathy. Barry Emara MD FRCS(C) Giovanni Caboto Club October 3, 2012 Diabetic Retinopathy Barry Emara MD FRCS(C) Giovanni Caboto Club October 3, 2012 Outline Statistics Anatomy Categories Assessment Management Risk factors What do you need to do? Objectives Summarize the

More information

Research Article Repeatability of Perimacular Ganglion Cell Complex Analysis with Spectral-Domain Optical Coherence Tomography

Research Article Repeatability of Perimacular Ganglion Cell Complex Analysis with Spectral-Domain Optical Coherence Tomography Ophthalmology Volume 2015, Article ID 605940, 5 pages http://dx.doi.org/10.1155/2015/605940 Research Article Repeatability of Perimacular Ganglion Cell Complex Analysis with Spectral-Domain Optical Coherence

More information

Five Things You re Missing with Your Fundus Camera

Five Things You re Missing with Your Fundus Camera ebook Five Things You re Missing with Your Fundus Camera By Donald J. Siegel, OD, Sun City West Eye Care Sponsored by: Before I began incorporating EIDON true-color imaging into my practice, my retinal

More information

NIH Public Access Author Manuscript Br J Ophthalmol. Author manuscript; available in PMC 2010 April 29.

NIH Public Access Author Manuscript Br J Ophthalmol. Author manuscript; available in PMC 2010 April 29. NIH Public Access Author Manuscript Published in final edited form as: Br J Ophthalmol. 2009 August ; 93(8): 1057 1063. doi:10.1136/bjo.2009.157875. Retinal nerve fibre layer thickness measurement reproducibility

More information

Case Report: Indocyanine Green Dye Leakage from Retinal Artery in Branch Retinal Vein Occlusion

Case Report: Indocyanine Green Dye Leakage from Retinal Artery in Branch Retinal Vein Occlusion Case Report: Indocyanine Green Dye Leakage from Retinal Artery in Branch Retinal Vein Occlusion Hiroki Fujita, Kyoko Ohno-Matsui, Soh Futagami and Takashi Tokoro Department of Visual Science, Tokyo Medical

More information

INTRODUCTION MIKLOS SCHNEIDER 1, GABOR BORGULYA 2, ANDRAS SERES 1, ZOLTAN Z. NAGY 1, JANOS NEMETH 1

INTRODUCTION MIKLOS SCHNEIDER 1, GABOR BORGULYA 2, ANDRAS SERES 1, ZOLTAN Z. NAGY 1, JANOS NEMETH 1 European Journal of Ophthalmology / Vol. 19 no. 2, 2009 / pp. 180-187 Central corneal thickness measurements with optical coherence tomography and ultrasound pachymetry in healthy subjects and in patients

More information

OCT angiography of ONH blood flow in glaucoma

OCT angiography of ONH blood flow in glaucoma Hawaiian Eye Meeting 19-25 January 2013 OCT angiography of ONH blood flow in glaucoma David Huang, MD, Weeks Professor of Ophthalmic Research Professor of Ophthalmology & Biomedical Engineering Casey Eye

More information

Ganglion cell complex scan in the early prediction of glaucoma

Ganglion cell complex scan in the early prediction of glaucoma Original article in the early prediction of glaucoma Ganekal S Nayana Super Specialty Eye Hospital and Research Center, Davangere, Karnataka, India Abstract Objective: To compare the macular ganglion cell

More information

Learn Connect Succeed. JCAHPO Regional Meetings 2017

Learn Connect Succeed. JCAHPO Regional Meetings 2017 Learn Connect Succeed JCAHPO Regional Meetings 2017 Faculty Biometry and IOL Calculations ASCRS and ASOA Symposium and Congress Los Angeles, CA Daniel H. Chang, M.D. - Empire Eye and Laser Center Bakersfield,

More information

RECENT ADVANCES IN ANTERIOR SEGMENT IMAGING OF THE EYE. by Eszter Szalai, M.D. Supervisor: László Módis, M.D., Ph.D.

RECENT ADVANCES IN ANTERIOR SEGMENT IMAGING OF THE EYE. by Eszter Szalai, M.D. Supervisor: László Módis, M.D., Ph.D. SHORT THESIS FOR THE DEGREE OF DOCTOR OF PHILOSOPHY (Ph.D.) RECENT ADVANCES IN ANTERIOR SEGMENT IMAGING OF THE EYE by Eszter Szalai, M.D. Supervisor: László Módis, M.D., Ph.D. UNIVERSITY OF DEBRECEN DOCTORAL

More information

Dehiscence of detached internal limiting membrane in eyes with myopic traction maculopathy with spontaneous resolution

Dehiscence of detached internal limiting membrane in eyes with myopic traction maculopathy with spontaneous resolution Hirota et al. BMC Ophthalmology 2014, 14:39 RESEARCH ARTICLE Open Access Dehiscence of detached internal limiting membrane in eyes with myopic traction maculopathy with spontaneous resolution Kazunari

More information

The pathogenesis of type 2 diabetes mellitus (DM) is

The pathogenesis of type 2 diabetes mellitus (DM) is Retina Relationship Between Retinal Blood Flow and Serum Adiponectin Concentrations in Patients With Type 2 Diabetes Mellitus Tsuneaki Omae, Taiji Nagaoka, and Akitoshi Yoshida Department of Ophthalmology,

More information

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(1):32-38 http://dx.doi.org/10.3341/kjo.2012.26.1.32 Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence

More information

Optical Coherence Tomography Fast versus Regular Macular Thickness Mapping in Diabetic Retinopathy

Optical Coherence Tomography Fast versus Regular Macular Thickness Mapping in Diabetic Retinopathy Original Paper DOI: 10.1159/000127830 Received: March 18, 2007 Accepted after revision: July 9, 2007 Published online: April 25, 2008 Optical Coherence Tomography Fast versus Regular Macular Thickness

More information

OCCLUSIVE VASCULAR DISORDERS OF THE RETINA

OCCLUSIVE VASCULAR DISORDERS OF THE RETINA OCCLUSIVE VASCULAR DISORDERS OF THE RETINA Learning outcomes By the end of this lecture the students would be able to Classify occlusive vascular disorders (OVD) of the retina. Correlate the clinical features

More information

Reproducibility of Choroidal Thickness Measurements Across Three Spectral Domain Optical Coherence Tomography Systems

Reproducibility of Choroidal Thickness Measurements Across Three Spectral Domain Optical Coherence Tomography Systems Reproducibility of Choroidal Thickness Measurements Across Three Spectral Domain Optical Coherence Tomography Systems The MIT Faculty has made this article openly available. Please share how this access

More information

Ophthalmology Department, Lozano Blesa University Hospital, c/ San Juan Bosco 15, Zaragoza, Spain 2

Ophthalmology Department, Lozano Blesa University Hospital, c/ San Juan Bosco 15, Zaragoza, Spain 2 Ophthalmology Volume 2012, Article ID 107053, 6 pages doi:10.1155/2012/107053 Clinical Study Comparison of Retinal Nerve Fiber Layer Thickness Measurements in Healthy Subjects Using Fourier and Time Domain

More information

Comparison of cross sectional optical coherence tomography images of elevated optic nerve heads across acquisition devices and scan protocols

Comparison of cross sectional optical coherence tomography images of elevated optic nerve heads across acquisition devices and scan protocols Patel et al. Eye and Vision (2018) 5:17 https://doi.org/10.1186/s40662-018-0112-3 RESEARCH Open Access Comparison of cross sectional optical coherence tomography images of elevated optic nerve heads across

More information

Influence of cataract on image quality and macular thickness measured using spectral domain optical coherence tomography: a prospective cohort study

Influence of cataract on image quality and macular thickness measured using spectral domain optical coherence tomography: a prospective cohort study International Journal of Advances in Medicine Lathika VK et al. Int J Adv Med. 2017Apr;4(2):546-550 http://www.ijmedicine.com pissn2349-3925 eissn 2349-3933 Original Research Article DOI: http://dx.doi.org/10.18203/2349-3933.ijam20171058

More information

Automated Detection of Vascular Abnormalities in Diabetic Retinopathy using Morphological Entropic Thresholding with Preprocessing Median Fitter

Automated Detection of Vascular Abnormalities in Diabetic Retinopathy using Morphological Entropic Thresholding with Preprocessing Median Fitter IJSTE - International Journal of Science Technology & Engineering Volume 1 Issue 3 September 2014 ISSN(online) : 2349-784X Automated Detection of Vascular Abnormalities in Diabetic Retinopathy using Morphological

More information

International Journal of Ophthalmic Research

International Journal of Ophthalmic Research International Journal of Ophthalmic Research Online Submissions: http://www.ghrnet.org/index./ijor/ doi:10.17554/j.issn.2409-5680.2017.03.55 Int. J. Ophthalmic Res 2017 June; 3(2): 226-230 ISSN 2409-5680

More information

10/17/2017. FDA Approved. Zeiss AngioPlex TM Optovue AngioVue TM

10/17/2017. FDA Approved. Zeiss AngioPlex TM Optovue AngioVue TM Images retinal microvasculature without dye injection Displays structure and function from a single imaging system Standard of Care-2011 DFE, Fundus Photos, VF 10-2, SD-OCT, FAF, or mferg 2016-AAO Baseline

More information

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography Original articles in Indian eyes with optical coherence tomography Malik A, Singh M, Arya SK, Sood S, Ichhpujani P Department of Ophthalmology Government Medical College and Hospital, Sector 32, Chandigarh,

More information

Overview. Macular OCT Artifact Study

Overview. Macular OCT Artifact Study Imaging Artifacts Sarah Moyer, CRA, OCT-C Director, Ophthalmic Imaging Kittner Eye Center University of North Carolina Chapel Hill, NC Disclose financial interest now Overview Sarah s Thoughts on Artifacts

More information

Comparative Efficacy of the New Optical Biometer on Intraocular Lens Power Calculation (AL-Scan versus IOLMaster)

Comparative Efficacy of the New Optical Biometer on Intraocular Lens Power Calculation (AL-Scan versus IOLMaster) pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2018;32(3):241-248 https://doi.org/10.3341/kjo.2017.0063 Original Article Comparative Efficacy of the New Optical Biometer on Intraocular Lens Power

More information

measure of your overall performance. An isolated glucose test is helpful to let you know what your sugar level is at one moment, but it doesn t tell you whether or not your diabetes is under adequate control

More information

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD Original Article Philippine Journal of OPHTHALMOLOGY Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 1: Categorical Outcomes Based on a Normative Database

More information

INTRODUCTION S. HERDENER, D. HAFIZOVIC, M. PACHE, S. LAUTEBACH, J. FUNK. University Eye Hospital, Freiburg - Germany

INTRODUCTION S. HERDENER, D. HAFIZOVIC, M. PACHE, S. LAUTEBACH, J. FUNK. University Eye Hospital, Freiburg - Germany European Journal of Ophthalmology / Vol. 18 no. 1, 2008 / pp. 39-43 Is the PASCAL -Tonometer suitable for measuring intraocular pressure in clinical routine? Long- and short-term reproducibility of dynamic

More information

Introduction. Zohreh Hosseinaee 1,2, *, Bingyao Tan 2, *, Adam Martinez 2, and Kostadinka K. Bizheva 1,2,3. Article

Introduction. Zohreh Hosseinaee 1,2, *, Bingyao Tan 2, *, Adam Martinez 2, and Kostadinka K. Bizheva 1,2,3. Article Article https://doi.org/10.1167/tvst.7.4.18 Comparative Study of Optical Coherence Tomography Angiography and Phase-Resolved Doppler Optical Coherence Tomography for Measurement of Retinal Blood Vessels

More information

Structural & Functional Optical Coherence Tomography for Glaucoma Diagnosis

Structural & Functional Optical Coherence Tomography for Glaucoma Diagnosis Congreso Argentino de Oftalmologia, May 2009, Buenos Aires Structural & Functional Optical Coherence Tomography for Glaucoma Diagnosis David Huang, MD, PhD Assoc. Prof. of Ophthalmology & Biomedical Engineering

More information

CAPILLARY NETWORK ANOMALIES IN BRANCH RETINAL VEIN OCCLUSION ON OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY

CAPILLARY NETWORK ANOMALIES IN BRANCH RETINAL VEIN OCCLUSION ON OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY CAPILLARY NETWORK ANOMALIES IN BRANCH RETINAL VEIN OCCLUSION ON OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY MARCO RISPOLI, MD, MARIA CRISTINA SAVASTANO, MD, PHD, BRUNO LUMBROSO, MD Purpose: To analyze the

More information

Comparison of ultrasound and optic biometry with respect to eye refractive errors after phacoemulsification

Comparison of ultrasound and optic biometry with respect to eye refractive errors after phacoemulsification Received: 5.5.2007 Accepted: 22.1.2008 Comparison of ultrasound and optic biometry with respect to eye refractive errors after phacoemulsification Heidarali Moeini*, Fateme Eslami**, Akram Rismanchian*,

More information

Cirrus TM HD-OCT. Details defi ne your decisions

Cirrus TM HD-OCT. Details defi ne your decisions Cirrus TM HD-OCT Details defi ne your decisions 2 With high-defi nition OCT Carl Zeiss Meditec takes you beyond standard spectral domain Built on 10 years experience at the vanguard of innovation, Carl

More information

Retinal vessel dilatation and elongation precedes diabetic macular oedema

Retinal vessel dilatation and elongation precedes diabetic macular oedema 274 Department of Ophthalmology, University of Iceland, Landakotsspítali, Reykjavík, Iceland K Kristinsson M S Gottfredsdóttir E Stefánsson Correspondence to: Einar Stefánsson, MD, PhD, University of Iceland,

More information

Katsuhiko, Fukui ; Akitoshi, Yoshida ; Hiromasa, Igarashi ; Hong-Ming, Cheng ; Hironori, Isobe

Katsuhiko, Fukui ; Akitoshi, Yoshida ; Hiromasa, Igarashi ; Hong-Ming, Cheng ; Hironori, Isobe The Journal of ophthalmic photography (1995) 17(2):54-60. Anterior Segment Fluorescein Angiography Using a Modified Photo Slit- Lamp Katsuhiko, Fukui ; Akitoshi, Yoshida ; Hiromasa, Igarashi ; Hong-Ming,

More information

Neovascular Glaucoma Associated with Cilioretinal Artery Occlusion Combined with Perfused Central Retinal Vein Occlusion

Neovascular Glaucoma Associated with Cilioretinal Artery Occlusion Combined with Perfused Central Retinal Vein Occlusion Neovascular Glaucoma Associated with Cilioretinal Artery Occlusion Combined with Perfused Central Retinal Vein Occlusion Man-Seong Seo,* Jae-Moon Woo* and Jeong-Jin Seo *Department of Ophthalmology, Chonnam

More information

1. Background. Correspondence should be addressed to Atsushi Miki;

1. Background. Correspondence should be addressed to Atsushi Miki; Hindawi Ophthalmology Volume 2017, Article ID 3596587, 5 pages https://doi.org/10.1155/2017/3596587 Research Article Quantitative Analysis of Macular Inner Retinal Layer Using Swept-Source Optical Coherence

More information

Characteristics of Retinal Neovascularization in Proliferative Diabetic Retinopathy Imaged by Optical Coherence Tomography Angiography

Characteristics of Retinal Neovascularization in Proliferative Diabetic Retinopathy Imaged by Optical Coherence Tomography Angiography Retina Characteristics of Retinal Neovascularization in Proliferative Diabetic Retinopathy Imaged by Optical Coherence Tomography Angiography Akihiro Ishibazawa, 1 Taiji Nagaoka, 1 Harumasa Yokota, 1 Atsushi

More information