Urgent field safety notice

Size: px
Start display at page:

Download "Urgent field safety notice"

Transcription

1 Accu-Chek Aviva, Accu-Chek Performa Urgent field safety notice May 2018 Important information on selected lots of Accu-Chek Aviva (50s, 10s) and Accu-Chek Performa (10s) strips potentially showing an increased number of strip errors prior to dosing or biased results Dear Healthcare Professional, At Roche Diabetes Care we hold our products to the highest standards of quality and are committed to communicating any issues impacting the operation of our products. This is why we would like to inform you today about an issue that might occur in certain lots of the Accu-Chek Aviva and Accu-Chek Performa test strips. As part of our ongoing quality monitoring and market surveillance processes, we have identified certain test strip lots that potentially show an increase in strip errors prior to dosing. Due to the designed fail-safe in the blood glucose meter, the issue can be identified by an error message displayed on the meter upon strip insertion or through the device not recognising the test strip. However, in a very limited number of cases the test strip can produce a biased result i.e. a falsely too high or too low value, which your patients might not be able to detect easily and which could lead to erroneous therapy adaptations. The affected UK products are: Accu-Chek Aviva 50s Accu-Chek Aviva 10s (Sold in retail only) Accu-Chek Performa 10s (Sold in retail only) Accu-Chek Performa 10s contained with Accu-Chek Performa Nano meter kits As patient safety is our first priority, we would therefore like to ask you to advise your patients to: check the lot numbers of their test strip supplies against the complete list of lot numbers in Appendix A. You can also do this online by visiting They can find the lot number on the top flap of the vial carton packaging as well as on the label of each test strip vial as shown in the picture below. Internal

2 Consumer Letter (Accu-Chek Aviva, Accu-Chek Performa) Urgent field safety notice May, 2018 Important information on selected lots of Accu-Chek Aviva and Accu-Chek Performa test strips potentially showing an increased number of strip errors prior to dosing or biased results Dear Customer, At Roche Diabetes Care we hold our products to the highest standards of quality and are committed to communicating any issues impacting the operation of our products. This is why we would like to inform you today about an issue that might occur in certain lots of the Accu-Chek Aviva, Accu-Chek Performa test strips, and Accu-Chek Performa strips that have been included in Accu-Chek Performa Nano meter kits. As part of our ongoing quality monitoring and market surveillance processes, we have identified certain test strip lots that potentially show an increase in strip errors prior to dosing. Due to the designed fail-safe in the blood glucose meter, the issue can be identified by an error message displayed on the meter upon strip insertion or through the device not recognising the test strip. However, in a very limited number of cases the test strip can produce a biased result i.e. a falsely too high or too low value, which you might not be able to detect easily and which could lead to incorrect therapy adaptations. As patient safety is our first priority, we would therefore like to ask you to: Check the lot numbers of your test strip supplies against the complete list of lot numbers in Appendix A. You can find the lot number on the top flap of the vial carton packaging as well as on the label of each test strip vial as shown in the picture below. discontinue using strips from the affected lots immediately and please return to the pharmacy or store where you obtained the strips for a replacement pack Please complete the attached acknowledgement form and return it via freepost within 7 days of receipt of this notification

3 If your test strip supply is not from the affected lots, you can continue using these test strips and no additional action is required. We have thoroughly investigated this issue to identify the root cause of this potential error and have already started to implement the appropriate corrective measures. Please be assured that this issue only affects specific lots of the Accu-Chek Aviva and Accu-Chek Performa test strips. Other Accu-Chek blood glucose test strips available in your market are not impacted by this issue. The MHRA, users of the affected blood glucose monitoring systems, healthcare professionals, hospitals, distributors and retailers will all be informed about this field action. If you need any additional advice on the operation of your Accu-Chek blood glucose meter and test strips or have any further questions or concerns, please call our Accu-Chek Customer Careline on or visit the website. If you are a pump user please contact the Accu-Chek Pump Careline on It is through the careful monitoring of customer reports that we are able to identify issues and implement solutions. We appreciate your time and attention to this important notification. Kind regards, Roche Diabetes Care Appendix A: List of affected lot numbers 2/4

4 APPENDIX A: LOT NUMBERS UK ACCU-CHEK AVIVA STRIPS (50s) ACCU-CHEK AVIVA STRIPS (10s) ACCU-CHEK PERFORMA STRIPS (10s) Accu-Chek Performa Nano mmol Kit. [Please note that the FSN only applies to the pack of Performa 10 test strips contained within the kit] /4

5 Pharmacy Letter (Accu-Chek Aviva, Accu-Chek Performa) Urgent field safety notice May, 2018 Important information on selected lots of Accu-Chek Aviva and Accu-Chek Performa test strips potentially showing an increased number of strip errors prior to dosing or biased results Dear Pharmacy, At Roche Diabetes Care we hold our products to the highest standards of quality and are committed to communicating any issues impacting the operation of our products. This is why we would like to inform you today about an issue that might occur in certain lots of the Accu-Chek Aviva test strips, Accu-Chek Performa test strips, and Accu-Chek Performa strips that have been included in Accu-Chek Performa Nano meter kits. As part of our ongoing quality monitoring and market surveillance processes, we have identified certain test strip lots that potentially show an increase in strip errors prior to dosing. Due to the designed fail-safe in the blood glucose meter, the issue can be identified by an error message displayed on the meter upon strip insertion or through the device not recognising the test strip. However, in a very limited number of cases the test strip can produce a biased result i.e. a falsely too high or too low value, which you might not be able to detect easily and which could lead to incorrect therapy adaptations. As patient safety is our first priority, we would therefore like to ask you to: Check the lot numbers of your test strip stock in your pharmacy against the complete list of lot numbers in appendix A. You can find the lot number on the top flap of the vial carton packaging as well as on the label of each test strip vial as shown in the picture below. Discontinue dispensing/selling strips from the affected lots immediately

6 Accept any product from the affected lots that that is returned by a patient/customer and issue a replacement from another lot. Return all affected product to your wholesaler We have thoroughly investigated this issue to identify the root cause of this potential error and have already started to implement the appropriate corrective measures. Please be assured that this issue only affects the Accu-Chek Aviva and Accu-Chek Performa test strips, as well as strips contained within the Accu-Chek Performa Nano kits in the United Kingdom. The MHRA, users of the affected blood glucose monitoring systems, healthcare professionals and hospitals will all be informed about this field action. Please contact us on if you have any further questions or concerns. Kind regards, Roche Diabetes Care Appendix A: List of affected lot numbers 2/3

7 APPENDIX A: Affected lot numbers UK ACCU-CHEK AVIVA STRIPS (50s) PIP CODE: EAN CODE: ACCU-CHEK AVIVA STRIPS (10s) PIP CODE: EAN CODE: ACCU-CHEK PERFORMA STRIPS (10s) PIP CODE: EAN CODE: Accu-Chek Performa Nano mmol Kit. [Please note that the FSN only applies to the pack of Performa 10 test strips contained within the kit] EAN CODE: /3

8 discontinue using strips from the affected lots immediately and please return them to the pharmacy or store where they obtained the strips for a replacement pack at no charge. Please check for any affected strip lots within your centers: Strips within the Accu-Chek Aviva Expert and Accu-Chek Performa Nano meter kits Accu-Chek Aviva 50s and 10s Accu-Chek Performa 10s Quarantine any affected lots and notify your local sales representative for collection and to arrange for replacement. Please select the right option when completing the acknowledgement form and return it on behalf of your practice within 7 days of receipt of this notification. We have thoroughly investigated this issue to identify the root cause of this potential error and have already started to implement the appropriate corrective measures. Please be assured that this issue only affects specific lots of the Accu-Chek Aviva and Accu-Chek Performa test strips. Other Accu-Chek blood glucose test strips available in your market are not impacted by this issue. The MHRA, users of the affected blood glucose monitoring systems, distributors and retailers will all be informed about this field action. Please call your local sales representative if you need any additional advice on the operation of Accu- Chek blood glucose meters and test strips, or have any further questions or concerns. It is through the careful monitoring of customer reports that we are able to identify issues and implement solutions. We appreciate your time and attention to this important notification. Kind regards, Roche Diabetes Care Appendix A: List of affected lot numbers Internal 2/3

9 APPENDIX A: AFFECTED LOTS UK ACCU-CHEK AVIVA STRIPS (50s) ACCU-CHEK AVIVA STRIPS (10s) ACCU-CHEK PERFORMA STRIPS (10s) Accu-Chek Performa Nano mmol Kit. [Please note that the FSN only applies to the pack of Performa 10 test strips contained within the kit] *Our records indicate that you may have received stock from these specific affected lots. Internal 3/3

10 URGENT FIELD SAFETY NOTICE PLEASE COMPLETE AND RETURN BY FAX/ Roche Diabetes Care Reference: Medical Device: Manufacturer: Issue: SB_RDC_2018_02 Accu-Chek Aviva and Accu-Chek Performa test strips Roche Diabetes Care Important information on selected lots of Accu-Chek Aviva and Accu-Chek Performa test strips potentially showing an increased number of strip errors prior to dosing or biased results Acknowledgement Form / Faxback Please submit this acknowledgement form to Roche Diabetes Care: via or postal address: burgesshill.dcsafety@roche.com Local Regulatory and Safety Officer Roche Diabetes Care Ltd Charles Avenue Burgess Hill West Sussex RH15 9RY I confirm to have received, read and understood the information in this Field Safety Notice of May 2018 for the above mentioned medical devices and intend to perform the requested actions. Form completed and returned by: Name (CAPITAL LETTERS): Position/Title of Person completing this Form: Company Name: Phone Number: Address: Date (dd-mmm-yyyy): Signature Internal 1/1

Urgent field safety notice

Urgent field safety notice Accu-Chek Aviva Urgent field safety notice May 2018 Important information on selected lots of Accu-Chek Aviva (50s, 10s) potentially showing an increased number of strip errors prior to dosing or biased

More information

Important information on the Accu-Chek Insight insulin pump system: Update of the handling instructions for changing an insulin cartridge

Important information on the Accu-Chek Insight insulin pump system: Update of the handling instructions for changing an insulin cartridge Consumer Letter (Accu-Chek Insight PFC) Urgent field safety notice Location, Date Important information on the Accu-Chek Insight insulin pump system: Update of the handling instructions

More information

Urgent Field Safety Notice

Urgent Field Safety Notice Urgent Field Safety Notice Location, Date 2014 Important notice: Enhanced instructions for the proper handling of the Accu-Chek Mobile system to avoid the potential of falsely elevated

More information

URGENT: VOLUNTARY DRUG RECALL 07/05/2016

URGENT: VOLUNTARY DRUG RECALL 07/05/2016 9075 Centre Pointe Drive, Suite 140 West Chester, OH 45069 Phone: 800.543.2111 URGENT: VOLUNTARY DRUG RECALL 07/05/2016 GLUCOSE TEST STRIPS, TRUETRACK, 50/BOX NDC# 56151-0850-50 Besse# 31330 TRUETRACK

More information

Accu-Chek Insight. Travelling & Holiday Guide. 3583_HolidayGuideA5booklet_Final.indd 1

Accu-Chek Insight. Travelling & Holiday Guide. 3583_HolidayGuideA5booklet_Final.indd 1 Accu-Chek Insight Travelling & Holiday Guide 3583_HolidayGuideA5booklet_Final.indd 1 4/16/15 10:58 AM Preparation Carry your supplies in your hand luggage and split with a travel companion if possible.

More information

URGENT FIELD SAFETY NOTICE

URGENT FIELD SAFETY NOTICE AMO Ireland Block B, Liffey Valley Office Campus, Quarryvale, Dublin D22 XOY3 Ireland URGENT FIELD SAFETY NOTICE April X, 2017 Dear AMO Customer: RE: Voluntary Recall of Select Lots of AMO OVDs Abbott

More information

Essential advice for people with diabetes from Accu-Chek. Time to test. G1812 Time to Test leaflet v5.indd 1 25/02/ :43

Essential advice for people with diabetes from Accu-Chek. Time to test. G1812 Time to Test leaflet v5.indd 1 25/02/ :43 Essential advice for people with diabetes from Accu-Chek Time to test G1812 Time to Test leaflet v5.indd 1 25/02/2016 16:43 Why do I need to check m Going for a jog, drinks after work, eating out or a

More information

Time to test. Do you do it enough? Essential advice for people with diabetes from Accu-Chek

Time to test. Do you do it enough? Essential advice for people with diabetes from Accu-Chek Do you do it enough? Essential advice for people with diabetes from Accu-Chek Regular tests provide valuable information for both you and your diabetes team Find out what level of testing is recommended

More information

If you have any questions, please contact Jerry Webb at ext or

If you have any questions, please contact Jerry Webb at ext or 1. On May 3, 2013, Health & Life, Co., LTD, initiated a nationwide recall of 242,892 EZ Breathe Atomizers, Model #EZ-100, after becoming aware of an increase in the number of complaints. The Atomizers

More information

URGENT MEDICAL DEVICE CORRECTION. Important Information about Animas 2020, IR 1250 or IR 1200 Insulin Pumps

URGENT MEDICAL DEVICE CORRECTION. Important Information about Animas 2020, IR 1250 or IR 1200 Insulin Pumps URGENT MEDICAL DEVICE CORRECTION May 26, 2015 Important Information about Animas 2020, IR 1250 or IR 1200 Insulin Pumps Dear Healthcare Professional: At Animas, we hold our products to the highest standards

More information

URGENT FIELD SAFETY NOTICE Overconsumption following ElectroStatic Discharge or MRI scan

URGENT FIELD SAFETY NOTICE Overconsumption following ElectroStatic Discharge or MRI scan URGENT FIELD SAFETY NOTICE Overconsumption following ElectroStatic Discharge or MRI scan FSCA identifier: CRM-SAL-2017-002 Affected Devices: Platinium Implantable Cardiac Defibrillators (ICDs) and Cardiac

More information

Essential advice for people with diabetes from Accu-Chek. Get the low-down on hypos

Essential advice for people with diabetes from Accu-Chek. Get the low-down on hypos Essential advice for people with diabetes from Accu-Chek Get the low-down on hypos The low-down on hypos If you have diabetes, a hypo is one of those things you have to deal with from time to time. FIRST

More information

Essential advice for people with diabetes from Accu-Chek. The inside story on diabetes

Essential advice for people with diabetes from Accu-Chek. The inside story on diabetes Essential advice for people with diabetes from Accu-Chek The inside story on diabetes What is diabetes? Glucose is a form of sugar that is found in food you eat. It is a vital energy source for your body

More information

Urgent Patient Notification

Urgent Patient Notification Urgent Patient Notification Comfort, Comfort Short, and Contact Detach Infusion Sets - Tubing May Detach From the Connect/Disconnect Location November 24, 2014 Dear Valued Customer: At Animas, we hold

More information

Essential advice for men with diabetes from Accu-Chek. Man talk

Essential advice for men with diabetes from Accu-Chek. Man talk Essential advice for men with diabetes from Accu-Chek Man talk Diabetes and ere More than 50% of m problems either obtai an erection. Whether you have diabetes or not, erectile dysfunction is a really

More information

Plug-and-play goes online

Plug-and-play goes online The next-generation Accu-Chek T Smart Pix software Plug-and-play goes online Proven tools for efficient and effective diabetes management 1 even for patients who are not in your office How many of these

More information

NovoPen Echo URGENT MEDICAL DEVICE RECALL

NovoPen Echo URGENT MEDICAL DEVICE RECALL NovoPen Echo URGENT MEDICAL DEVICE RECALL ISSUE Novo Nordisk A/S has detected that the insulin cartridge holder used in a small number of batches may crack or break if exposed to certain chemicals, for

More information

Manometer Tray, Jamshidi Needle Biopsy, Illinois (TJ) Needle Aspiration, Jamshidi (TJ) Needle Bone Marrow, Thoracentesis/Paracentesis Kit

Manometer Tray, Jamshidi Needle Biopsy, Illinois (TJ) Needle Aspiration, Jamshidi (TJ) Needle Bone Marrow, Thoracentesis/Paracentesis Kit URGENT FIELD SAFETY NOTICE Affected product: MANOMETER TRAY JAMSHIDI NEEDLE BIOPSY ILLINOIS (TJ) NEEDLE ASPIRATION JAMSHIDI (TJ) NEEDLE BONE MARROW THORACENTESIS/PARACENTESIS KIT FSCA reference number:

More information

Attachment 12 URGENT FIELD SAFETY REMOVAL FSCA3193. ETEST Ceftriaxone TXL32 (Ref , ) FOAM packaging

Attachment 12 URGENT FIELD SAFETY REMOVAL FSCA3193. ETEST Ceftriaxone TXL32 (Ref , ) FOAM packaging 16 January 2017 Attachment 12 URGENT FIELD SAFETY REMOVAL FSCA3193 ETEST Ceftriaxone TXL32 (Ref. 507058, 507018) FOAM packaging Dear This letter is intended for all ETEST Ceftriaxone TXL32 (Ref. 507058,

More information

URGENT MEDICINE RECALL

URGENT MEDICINE RECALL Wednesday 7 th September 2016 URGENT MEDICINE RECALL GlucaGen HypoKit (glucagon (rys) as HCl 1 mg (1 IU) powder vial with 1 ml, prefilled solvent syringe); AUST R 47105 Batch numbers: FS6X465, FS6X536,

More information

DATE RECEIVED 09/10/2017 SUBJECT Formulary PASSED TO Mary DATE PASSED 09/10/2017 RESPOND BY 27/10/2017 CATEGORY Business FoI NUMBER

DATE RECEIVED 09/10/2017 SUBJECT Formulary PASSED TO Mary DATE PASSED 09/10/2017 RESPOND BY 27/10/2017 CATEGORY Business FoI NUMBER DATE RECEIVED 09/10/2017 SUBJECT Formulary PASSED TO Mary DATE PASSED 09/10/2017 RESPOND BY 27/10/2017 CATEGORY Business FoI NUMBER 2017-404 Question/s to be Answered Can you please provide the following

More information

The inside story on diabetes

The inside story on diabetes Did you know? Essential advice for people with diabetes from Accu-Chek There are two main types of diabetes type 1 and type 2 Signs are tiredness, excess thirst, urinating a lot, blurry vision and weight

More information

Important Information about your 2.0 ml Animas Insulin Pump Cartridges

Important Information about your 2.0 ml Animas Insulin Pump Cartridges URGENT: INSULIN PUMP 2.0 ml CARTRIDGE RECALL February 24, 2011 Dear Animas Pumper: Important Information about your 2.0 ml Animas Insulin Pump Cartridges Animas Corporation is dedicated to supporting pumpers

More information

Essential advice for people with diabetes from Accu-Chek Alcohol

Essential advice for people with diabetes from Accu-Chek Alcohol Essential advice for people with diabetes from Accu-Chek Alcohol and diabetes The impact of alcohol on blood glucose If I have diabetes does that mean I can t have a drink? Enjoying the occasional drink

More information

NICE to. know... The facts you need to make the most of your diabetes care.

NICE to. know... The facts you need to make the most of your diabetes care. NICE to know... 3 The facts you need to make the most of your diabetes care. NICE to know... The National Institute for Health and Care Excellence (NICE) provides advice on the care and support that should

More information

Contacts Consumers:

Contacts Consumers: Prinston Pharmaceutical Inc. issues Voluntary Nationwide Recall of Irbesartan and Irbesartan HCTZ Tablets Due to detection of a Trace Amount of Unexpected Impurity, N- nitrosodiethylamine (NDEA) in the

More information

Get the low-down on hypos

Get the low-down on hypos Are you prepared? Essential advice for people with diabetes from Accu-Chek 07299508001 Date of prep: May 2014 Having a hypo means your blood glucose is too low Test your blood glucose regularly and look

More information

Better connected for better decisions

Better connected for better decisions Introducing the Accu-Chek Connect diabetes management system Better connected for better decisions Help your patients use the power of their mobile devices and the web to manage their diabetes How many

More information

URGENT FIELD SAFETY NOTICE FSCA3193. Etest Ceftriaxone TXL32 (Ref , ) SPB Packaging

URGENT FIELD SAFETY NOTICE FSCA3193. Etest Ceftriaxone TXL32 (Ref , ) SPB Packaging 16 January 2017 Attachment 3 URGENT FIELD SAFETY NOTICE FSCA3193 Etest Ceftriaxone TXL32 (Ref. 412302, 412303) SPB Packaging Dear This letter is intended for all ETEST Ceftriaxone TXL32 (Ref. 412302, 412303)

More information

URGENT FIELD SAFETY NOTICE/ DEVICE RECALL

URGENT FIELD SAFETY NOTICE/ DEVICE RECALL URGENT FIELD SAFETY NOTICE/ DEVICE RECALL COMMERCIAL NAME: NC Trek RX Coronary Dilatation Catheter NC Traveler RX Coronary Dilatation Catheter NC Tenku RX PTCA Balloon Catheter FSCA-Identifier: Type of

More information

Guidance on the Self-Monitoring of Blood Glucose in Adults with Diabetes

Guidance on the Self-Monitoring of Blood Glucose in Adults with Diabetes Introduction Guidance on the Self-Monitoring of Blood Glucose in Adults with Diabetes This guideline is designed to offer guidance for primary and secondary care on the use of selfmonitoring of blood glucose

More information

Product NDC Code Lot Number Expiry Dates Distribution Date VALSARTAN TABLETS 40MG 30CT

Product NDC Code Lot Number Expiry Dates Distribution Date VALSARTAN TABLETS 40MG 30CT Prinston Pharmaceutical Inc Issues Voluntary Nationwide Recall of Valsartan and Valsartan HCTZ Tablets Due to detection of a Trace Amount of Unexpected Impurity, N- nitrosodimethylamine (NDMA) in the Products

More information

Guidance for the choice of blood glucose testing meters, strips and lancets

Guidance for the choice of blood glucose testing meters, strips and lancets Guidance for the choice of blood glucose testing meters, strips and lancets This guidance is intended to assist healthcare professionals in their selection of appropriate blood glucose meters and testing

More information

Government Gazette REPUBLIC OF SOUTH AFRICA

Government Gazette REPUBLIC OF SOUTH AFRICA Government Gazette REPUBLIC OF SOUTH AFRICA Vol. 523 Cape Town 9 January 2009 No. 31790 THE PRESIDENCY No. 23 9 January 2009 It is hereby notified that the President has assented to the following Act,

More information

FIELD SAFETY NOTICE (FSN) (Please refer to Annex 5 of the Guideline: MEDDEV REV.8 Guideline Vigilance)

FIELD SAFETY NOTICE (FSN) (Please refer to Annex 5 of the Guideline: MEDDEV REV.8 Guideline Vigilance) FIELD SAFETY NOTICE (FSN) (Please refer to Annex 5 of the Guideline: MEDDEV 2.12-1 REV.8 Guideline Vigilance) In Vitro Diagnostics Medical Devices : Type: Instruments for professional use. Attention :

More information

Guidance on the Self-Monitoring of Blood Glucose in Adults with Diabetes

Guidance on the Self-Monitoring of Blood Glucose in Adults with Diabetes Introduction Guidance on the Self-Monitoring of Blood Glucose in Adults with Diabetes This guideline is designed to offer guidance for primary and secondary care on the use of selfmonitoring of blood glucose

More information

Company Announcement. For Immediate Release. Contact. Announcement

Company Announcement. For Immediate Release. Contact. Announcement Company Announcement When a company announces a recall, market withdrawal, or safety alert, the FDA posts the company's announcement as a public service. FDA does not endorse either the product or the

More information

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS)

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Type 2 Diabetes DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Clinical Guideline Self-monitoring of blood glucose (SMBG) should be considered but not be routinely offered for adults with type 2 diabetes

More information

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS)

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Type 2 Diabetes Clinical Guideline Self-monitoring of blood glucose (SMBG) should be considered but not routinely offered for adults with type 2 diabetes

More information

Please see the attached drug alert for onward transmission as below.

Please see the attached drug alert for onward transmission as below. Chief Medical Officer Directorate Pharmacy and Medicines Division T: 0131-244 2528 E: irene.fazakerley@gov.scot IMMEDIATE MESSAGE TO: 1. Directors of Pharmacy 2. Medical Directors NHS Boards 18 April 2018

More information

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS)

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Type 2 Diabetes DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Clinical Guideline Self-monitoring of blood glucose should be offered to all newly diagnosed type 2 diabetic patients as an integral part

More information

Blood glucose Meter Roche USA V6/ PMS 287 Proof Print Number 209 Advanced Owner s Booklet

Blood glucose Meter Roche USA V6/ PMS 287 Proof Print Number 209 Advanced Owner s Booklet This file may not print or view at 100%. Die lines and color breaks do not print. Roche USA 50841 V6/1 06434258002 Black Proof Print Number 209 06434258002-0512 Roche USA 50841 V6/2 06434258002 Cyan Proof

More information

URGENT FIELD SAFETY NOTICE - EXPANSION

URGENT FIELD SAFETY NOTICE - EXPANSION URGENT FIELD SAFETY NOTICE - EXPANSION August 18, 2017 Product Name Catalog Number BD MAX TM Vaginal Panel 443710 BD MAX UVE Specimen Collection Kit 443376 BD MAX CT/GC/TV 442970 BD MAX CT/GC 442969 Dear

More information

Guidelines for Product Recall or Withdrawal

Guidelines for Product Recall or Withdrawal REPUBLIC OF KENYA PHARMACY AND POISONS BOARD Guidelines for Product Recall or Withdrawal Edition 1 Date of release for publication June 2006 Date of implementation June 2006 This document has been prepared

More information

ORDER of the Minister of Health No. 269/2017 of 14 March 2017 on mandatory provision of medicinal product adequate and continuous supplies

ORDER of the Minister of Health No. 269/2017 of 14 March 2017 on mandatory provision of medicinal product adequate and continuous supplies ORDER of the Minister of Health No. 269/2017 of 14 March 2017 on mandatory provision of medicinal product adequate and continuous supplies ISSUING BODY: The Ministry of Health published in: the Official

More information

Approval of Multiple Diabetes Management Products and Mesalazine proposals

Approval of Multiple Diabetes Management Products and Mesalazine proposals 8 June Approval of Multiple Diabetes Management Products and Mesalazine proposals PHARMAC is pleased to announce the approval of five agreements in relation to funding of blood glucose, blood glucose diagnostic

More information

Important Safety Notice!!! March 2018 Dräger Jaundice Meter JM-103 Out of Range Indication

Important Safety Notice!!! March 2018 Dräger Jaundice Meter JM-103 Out of Range Indication To all customers of the JM-103 Jaundice Meters Important Safety Notice!!! March 2018 Dräger Jaundice Meter JM-103 Out of Range Indication Dear Ladies and Gentlemen, Our continuous post market surveillance

More information

Changes to the National Diabetes Services Scheme (NDSS)

Changes to the National Diabetes Services Scheme (NDSS) Changes to the National Diabetes Services Scheme (NDSS) June 2016 FREQUENTLY ASKED QUESTIONS 1. What are the changes from 1 July 2016? 2. Why are the supply changes happening? 3. What is the role of Diabetes

More information

URGENT: Important Safety Information

URGENT: Important Safety Information October 2018 URGENT: Important Safety Information Subject: Notice of New Special Handling Instructions due to Potential for Crystallization, Cracked Needle Hubs and Particulate in Diazepam Injection, USP,

More information

Ref: PFSN18 Deviations of high (>4.5) CoaguChek INR values due to calibration with WHO reference standard rtf/16 SBN-CPS

Ref: PFSN18 Deviations of high (>4.5) CoaguChek INR values due to calibration with WHO reference standard rtf/16 SBN-CPS Consumer Letter (CoaguChek XS PT Test PST, CoaguChek XS PT Test, CoaguChek PT Test) Urgent Field Safety Notice Limited, Ref: PFSN18 Deviations of high (>4.5) CoaguChek INR values due

More information

Essential advice for people with diabetes from Accu-Chek Alcohol

Essential advice for people with diabetes from Accu-Chek Alcohol Essential advice for people with diabetes from Accu-Chek Alcohol and diabetes The impact of alcohol on blood glucose If I have diabetes does that mean I can t have a drink? Enjoying the occasional drink

More information

We are writing to inform you about an issue concerning your CoaguChek system and test strips. Please read this information carefully.

We are writing to inform you about an issue concerning your CoaguChek system and test strips. Please read this information carefully. Consumer Letter (CoaguChek XS PT Test PST, CoaguChek XS PT Test, CoaguChek PT Test) Urgent Field Safety Notice Location, Date We are writing to inform you about an issue concerning

More information

URGENT FIELD SAFETY NOTICE: INSTRUCTIONS TO END USER

URGENT FIELD SAFETY NOTICE: INSTRUCTIONS TO END USER URGENT FIELD SAFETY NOTICE: INSTRUCTIONS TO END USER Commercial Name: Unomedical Sterile Urine Drainage Bags LOT No: Please see attached sheet for reference numbers & affected LOT numbers FSCA Id: 2010/11

More information

URGENT FIELD SAFETY NOTICE MEDICAL DEVICE FIELD CORRECTION Maquet/Datascope CARDIOSAVE Intra-Aortic Balloon Pump (IABP)

URGENT FIELD SAFETY NOTICE MEDICAL DEVICE FIELD CORRECTION Maquet/Datascope CARDIOSAVE Intra-Aortic Balloon Pump (IABP) May 4, 2018 via [INSERT METHOD] URGENT FIELD SAFETY NOTICE MEDICAL DEVICE FIELD CORRECTION Maquet/Datascope CARDIOSAVE Intra-Aortic Balloon Pump (IABP) AFFECTED PRODUCT PART NUMBER DISTRIBUTION DATE Cardiosave

More information

Magnesium Sulfate in Water for Injection is indicated for the prevention and control of seizures in preeclampsia and eclampsia, respectively.

Magnesium Sulfate in Water for Injection is indicated for the prevention and control of seizures in preeclampsia and eclampsia, respectively. a HOSPIRA ISSUES A VOLUNTARY NATIONWIDE RECALL OF ONE LOT OF MAGNESIUM SULFATE IN WATER FOR INJECTION DUE TO INCORRECT BARCODE LABELING ON THE PRIMARY CONTAINER Consumers: 1-888-345-4680 Media: 610-329-1340

More information

URGENT FIELD SAFETY NOTICE

URGENT FIELD SAFETY NOTICE August X, 2018 Olympus reference: QIL 151-005 URGENT FIELD SAFETY NOTICE IMPORTANT UPDATED LABELING INFORMATION NEW REPROCESSING INSTRUCTIONS FOR THE OLYMPUS PJF-160 DUODENOSCOPE ATTENTION: Endoscopy Department,

More information

Walsall CCG Supported Blood Glucose Meters April 2018

Walsall CCG Supported Blood Glucose Meters April 2018 Meter Name Picture Type Strips Lancets Information Company PRIMARY CARE INITIATION TEE2+ Uncomplicated Type 2 diabetes patient on TEE2 test CareSens Lancets Easy to use. Large clear display Bluetooth functionality

More information

In the consumer advisory we are instructing consumers as follows:

In the consumer advisory we are instructing consumers as follows: Urgent Field Safety Notice Affected products: Poligrip Ultra and Poligrip Total Care Type of Action: Cessation of Manufacture and Distribution of our zinc-containing denture adhesives Date: 18 th February

More information

Urgent Notice September 16, 2014

Urgent Notice September 16, 2014 Urgent Notice September 16, 2014 Voluntary Recall Notification: EDS 3 CSF External Drainage System PLEASE DISTRIBUTE THIS INFORMATION TO CLINICIANS WHO USE THIS DEVICE Dear OR Manager, Materials Manager,

More information

New York State Department of Health. Notice of Voluntary Recall of Influenza A (H1N1) 2009 Monovalent Vaccine

New York State Department of Health. Notice of Voluntary Recall of Influenza A (H1N1) 2009 Monovalent Vaccine New York State Department of Health Notice of Voluntary Recall of Influenza A (H1N1) 2009 Monovalent Vaccine The U.S. Centers for Disease Control and Prevention have announced a Non safety Related Voluntary

More information

Guidelines on the Sourcing of Medicinal Products for Sale or Supply by a Retail Pharmacy Business

Guidelines on the Sourcing of Medicinal Products for Sale or Supply by a Retail Pharmacy Business Guidelines on the Sourcing of Medicinal Products for Sale or Supply by a Retail Pharmacy Business to Facilitate Compliance with Regulations 5(1)(g), 6, 8 and 11 of the Regulation of Retail Pharmacy Businesses

More information

Guidelines for the Recall or Withdrawal of a Medical Product

Guidelines for the Recall or Withdrawal of a Medical Product National Drug Authority Rumee Towers Plot 19, Lumumba Avenue P. O. Box 23096 Kampala, Uganda. Tel: +256-0414 - 255665/347391/2 E-mail: ndaug@nda.or.ug Website: http://www.nda.or.ug TABLE OF CONTENTS 0.2

More information

URGENT FIELD SAFETY NOTICE IMMEDIATE ACTION REQUIRED

URGENT FIELD SAFETY NOTICE IMMEDIATE ACTION REQUIRED Ref No: Date: 22.08.2018 Type of Action: PFSN18 Deviations of high (>4.5) CoaguChek INR values due to calibration with WHO reference standard rtf/16 SBN-CPS-2018-014 Field Safety Corrective Action (FSCA)

More information

For more comprehensive information, please refer to the t:connect Application User Guide available online at: Getting Started Guide.

For more comprehensive information, please refer to the t:connect Application User Guide available online at: Getting Started Guide. Congratulations on the purchase of your new insulin pump from Tandem Diabetes Care. Your decision to use insulin pump therapy is a sign of your commitment to actively manage your diabetes. This guide provides

More information

Multi User Model. Product Overview

Multi User Model. Product Overview Multi User Model Product Overview Features and Benefits NO CODING EASY TO USE o o o Large LCD display. Large buttons with simple functions Easy Setup. Pre-Setup before shipping. Easy strip insertion. SMALL

More information

Recall Guidelines. for Chinese Medicine Products

Recall Guidelines. for Chinese Medicine Products Recall Guidelines for Chinese Medicine Products April 2018 Recall Guidelines for Chinese Medicine Products Chinese Medicines Board Chinese Medicine Council of Hong Kong Compiled in September 2005 1 st

More information

COBAS, ACCU-CHECK INFORM, MODULAR and LIFE NEEDS ANSWERS are trademarks of Roche Roche

COBAS, ACCU-CHECK INFORM, MODULAR and LIFE NEEDS ANSWERS are trademarks of Roche Roche Accu-Chek Inform II test strip Results of the evaluation study on newborns, conducted by the Flevo Hospital (FVZH) Almere and the Maxima Medical Center (MMC) Veldhoven in the Netherlands COBAS, ACCU-CHECK

More information

Guideline on the Regulation of Therapeutic Products in New Zealand

Guideline on the Regulation of Therapeutic Products in New Zealand Guideline on the Regulation of Therapeutic Products in New Zealand Part 10: Requirements for information for prescribers and consumers Edition 7.0 January 2016 Section 1: Legislation Section summary This

More information

URGENT: Important Safety Information

URGENT: Important Safety Information July 2018 URGENT: Important Safety Information Subject: Notice of New Special Handling Instructions due to Potential for Cracked Needle Hubs and Particulate in Multiple Carpuject Luer Lock Glass Products-Corrected

More information

User manual for healthcare professionals

User manual for healthcare professionals Blood Glucose Management System Accu-Chek Performa User manual for healthcare professionals CONTENTS 3 Introduction 4 Staff Responsibilities 5 Protocol 6 About the Accu-Chek Performa 7 Coding/Calibrating

More information

Performa BLOOD GLUCOSE METER. Operator s Manual

Performa BLOOD GLUCOSE METER. Operator s Manual Performa BLOOD GLUCOSE METER Operator s Manual The Accu Chek Performa System Intended Use The Accu Chek Performa Meter is designed to be used with the Accu Chek Inform II Test Strip to quantitatively

More information

NEWCASTLE CLINICAL TRIALS UNIT STANDARD OPERATING PROCEDURES

NEWCASTLE CLINICAL TRIALS UNIT STANDARD OPERATING PROCEDURES SOP details SOP title: SOP number: SOP category: Version number: 03 Version date: 19 December 2016 Effective date: 19 January 2017 Revision due date: 19 January 2020 NEWCASTLE CLINICAL TRIALS UNIT STANDARD

More information

I trust you are well, it is pleasing to see that your campaign is providing positive results.

I trust you are well, it is pleasing to see that your campaign is providing positive results. Travelers on insulin pump therapy should always remove their insulin pump, sensor, transmitter, and meter before entering a room that has x-ray, magnetic resonance imaging (MRI), diathermy, or CT scan

More information

Access Points: Frequently Asked Questions 15 June 2016

Access Points: Frequently Asked Questions 15 June 2016 Access Points: Frequently Asked Questions 15 June 2016 Access Point Interim Agreement Q. Why have I received an Interim Agreement? The Commonwealth and Diabetes Australia have entered into a new NDSS Agreement

More information

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff

More information

March 21, Subject: Availability of Blister-Packaged XELODA (capecitabine) Tablets

March 21, Subject: Availability of Blister-Packaged XELODA (capecitabine) Tablets March 21, 2011 Subject: Availability of Blister-Packaged XELODA (capecitabine) Tablets Dear Healthcare Professional: The purpose of this letter is to inform you of the temporary availability of an alternative

More information

Experience the ultimate test strip for use with OneTouch Ultra, Ultra 2, and UltraMini meters

Experience the ultimate test strip for use with OneTouch Ultra, Ultra 2, and UltraMini meters Experience the ultimate test strip for use with OneTouch Ultra, Ultra 2, and UltraMini meters An estimated 29.1 million people in the US are living with diabetes The population of people living with diabetes

More information

Urgent Medical Device Recall

Urgent Medical Device Recall PENTAX Medical 3 Paragon Drive Montvale, New Jersey 07645 Toll-free: 800-431-5880 Tel: 201-571-2300 Fax: 201-799-4032 June 12, 2014 Urgent Medical Device Recall Field Safety Corrective Action IFU Addendum

More information

How to Access A Free Blood Glucose Meter

How to Access A Free Blood Glucose Meter How to Access A Free Abbott FreeStyle Lite FreeStyle Lite FreeStyle Freedom Lite FreeStyle InsuLinx How to Access A Free contacting their health care professionals at diabetes clinics; or calling the FreeStyle

More information

BLOOD GLUCOSE METER TEST STRIP STEP THERAPY CRITERIA

BLOOD GLUCOSE METER TEST STRIP STEP THERAPY CRITERIA BLOOD GLUCOSE METER TEST STRIP STEP THERAPY CRITERIA Coverage for services, procedures, medical devices and drugs are dependent upon benefit eligibility as outlined in the member's specific benefit plan.

More information

Flash Glucose Monitoring (Flash GM) Frequently asked questions (November 2018)

Flash Glucose Monitoring (Flash GM) Frequently asked questions (November 2018) Flash Glucose Monitoring (Flash GM) Frequently asked questions (November 2018) What is Flash Glucose Monitoring (Flash GM)? Flash glucose monitoring is a small sensor that you wear on your skin (usually

More information

HOME DIAGNOSTIC TESTS FOR PERSONAL HEALTH PRODUCT CATALOGUE

HOME DIAGNOSTIC TESTS FOR PERSONAL HEALTH PRODUCT CATALOGUE HOME DIAGNOSTIC TESTS FOR PERSONAL HEALTH PRODUCT CATALOGUE 2 WELCOME TO CIGA HEALTHCARE AND Take Charge of Your Own Health With consumers all over the world becoming increasingly more concerned about

More information

Urgent Field Safety Notice

Urgent Field Safety Notice ADVIA Centaur Systems Dimension Vista Systems IMMULITE Systems Urgent Field Safety Notice CC 17-06.A.OUS Elevated Results in Patient s Due to Cross of with Our records indicate that your facility may have

More information

LABELLING OF FOOD ADDITIVES REGULATIONS (under section 13(1)) (11th April, 2003) ARRANGEMENT OF REGULATIONS

LABELLING OF FOOD ADDITIVES REGULATIONS (under section 13(1)) (11th April, 2003) ARRANGEMENT OF REGULATIONS REGULATION LABELLING OF FOOD ADDITIVES REGULATIONS (under section 13(1)) (11th April, 2003) ARRANGEMENT OF REGULATIONS 1. Citation 2. Interpretation 3. Application 4. General Principles PART I Preliminary

More information

A new model for prescribing varenicline

A new model for prescribing varenicline Pharmacist Independent Prescribers in partnership with A new model for prescribing varenicline Dear Stop Smoking Advisor You will be aware of the stop smoking drug varenicline that goes under the brand

More information

DRAFT FOR CONSULTATION

DRAFT FOR CONSULTATION DRAFT FOR CONSULTATION Member s Bill Explanatory note General policy statement Smoking rates and tobacco consumption have declined in recent decades. However, 4 500 to 5 000 New Zealanders still die prematurely

More information

Important Update to Field Safety Notice Nellix EndoVascular Aneurysm Sealing System Updated Instructions for Use (IFU)

Important Update to Field Safety Notice Nellix EndoVascular Aneurysm Sealing System Updated Instructions for Use (IFU) October 6, 2017 Important Update to Field Safety Notice Nellix EndoVascular Aneurysm Sealing System Updated Instructions for Use (IFU) Dear Physician, This notification is to provide you with further information

More information

THIS PROJECT CAN STILL BE COMPLETED IF YOUR SHELF HANGER DID NOT ARRIVED PRIOR TO THE START OF THIS PROJECT.

THIS PROJECT CAN STILL BE COMPLETED IF YOUR SHELF HANGER DID NOT ARRIVED PRIOR TO THE START OF THIS PROJECT. Job Name: Walgreens ML Rx Diabetes 3/12-3/18 Revision Project: 170220000635 Date: 3/12/2017-3/18/2017 SERVICE OBJECTIVE OVERVIEW In this project you will be updating the POG for the Rx DIABETES & DIAGNOSTICS

More information

Allergan s Blephamide (sulfacetamide/prednisolone ophthalmic ointment) 10%/0.2%, 3.5Gm tube

Allergan s Blephamide (sulfacetamide/prednisolone ophthalmic ointment) 10%/0.2%, 3.5Gm tube Allergan s Blephamide (sulfacetamide/prednisolone ophthalmic ointment) 10%/0.2%, 3.5Gm tube On Aug. 24, 2015, Allergan recalled three lots (86258 expires September 2017, 87189 December 2017 and 87514 February

More information

M E M O R A N D U M. Please share this information as applicable within your organizations/practices.

M E M O R A N D U M. Please share this information as applicable within your organizations/practices. ESRD NETWORK 18 M E M O R A N D U M To: From: Subject: Facility Manager Lana Kacherova, QI Director Lisle Mukai, QI Coordinator FDA Recalls & Alerts Date: September 8, 2009 The Network is required to distribute

More information

Aviva Combo BLOOD GLUCOSE METER. Getting Started Guide for Self-Testing Only

Aviva Combo BLOOD GLUCOSE METER. Getting Started Guide for Self-Testing Only Aviva Combo BLOOD GLUCOSE METER Getting Started Guide for Self-Testing Only Before You Start Testing About the meter and test strips Carefully read and follow the instructions in the Getting Started Guide,

More information

Calcium Folinate Solution for Injection

Calcium Folinate Solution for Injection 1 of 7 6/10/2016 4:14 PM Dosage Form: injection, solution Calcium Folinate Solution for Injection URGENT: LEUCOVORIN UPDATE June 2011 Dear Healthcare Professional: Due to the current critical shortage

More information

Self-regulation program for the sale of tobacco products

Self-regulation program for the sale of tobacco products Self-regulation program for the sale of tobacco products Section 12 c of the Tobacco Act (1993:581) ENGELSKA Good to know about the self-regulation program Persons intending to sell tobacco products to

More information

*Newly Released Vials consist of lot numbers: 183aG117, 183aG217, 184aG117, 184aG aG117 and 185aG217

*Newly Released Vials consist of lot numbers: 183aG117, 183aG217, 184aG117, 184aG aG117 and 185aG217 *Newly Released Vials consist of lot numbers: 183aG117, 183aG217, 184aG117, 184aG217 185aG117 and 185aG217 SPECIAL HANDLING INSTRUCTIONS VIALS of ERWINASE FROM BATCHes 183a, 184a And 185a* should be used

More information

MEDICINES CONTROL COUNCIL

MEDICINES CONTROL COUNCIL MEDICINES CONTROL COUNCIL ACCESS to and CONTROL of MEDICAL DEVICES and IVDs This guideline is intended to provide recommendations to Manufacturers, Importers, Exporters, Distributors, Wholesalers and Holders

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE. Health Technology Appraisal

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE. Health Technology Appraisal NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Health Technology Appraisal Continuous subcutaneous insulin infusion for the treatment of diabetes (review) Final scope Appraisal objective To review

More information

Health Warning 6. Unofficial Translation OFFICIAL PUBLICATION No. 207 FROM THE REPUBLIC OF SURINAME

Health Warning 6. Unofficial Translation OFFICIAL PUBLICATION No. 207 FROM THE REPUBLIC OF SURINAME 2013-12- No. 207 Health Warning 6 2013 OFFICIAL PUBLICATION No. 207 FROM THE REPUBLIC OF SURINAME DECISION of the Minister of Health of December 9, 2013 No. 4061/13, on detailed rules on the packaging

More information

Feidhmeannacht na Seirbhíse Sláinte

Feidhmeannacht na Seirbhíse Sláinte Feidhmeannacht na Seirbhíse Sláinte Health Service Executive Feidhmeannacht na Seirbhíse Sláinte Seirbhís Aisíocaíochta Cúraim Phríomhúil Plás J5 Lárionad Gnó na Páirce Thuaidh Bealach Amach 5, M50 An

More information

PRODUCT RECALL NOTICE URGENT

PRODUCT RECALL NOTICE URGENT PRODUCT RECALL NOTICE URGENT Date: July 21, 2018 To: Re: Mondelēz Global LLC Customers Recall Notice Certain Ritz Cracker Sandwiches and Ritz Bits product in the U.S., including Puerto Rico & U.S. Virgin

More information

STANDARDISED PACKAGING AND TOBACCO PRODUCTS DIRECTIVE

STANDARDISED PACKAGING AND TOBACCO PRODUCTS DIRECTIVE This guidance document contains Assured Advice under the ACS Primary Authority Scheme. For more details visit www.acs.org.uk/assured-advice STANDARDISED PACKAGING AND TOBACCO PRODUCTS DIRECTIVE Laws regulating

More information