1 Accu-Chek Inform II test strip Results of the evaluation study on newborns, conducted by the Flevo Hospital (FVZH) Almere and the Maxima Medical Center (MMC) Veldhoven in the Netherlands COBAS, ACCU-CHECK INFORM, MODULAR and LIFE NEEDS ANSWERS are trademarks of Roche. 212 Roche Roche Diagnostics International Ltd. CH-6343 Rotkreuz Switzerland

2 Evaluation of the Accu-Chek II test strip for neonatal patients The current Accu-Chek Inform II test strip is the successor of the Performa test strip for point-of-care glucose testing. The current test strip is an optimized version of the previous Performa test strip and is used in combination with the Accu-Chek Inform II meter. In November 21, this test strip was evaluated by two Dutch laboratories and approved for use on adults.* The results of neonatal samples were also recently tested with this test strip at the clinical chemistry laboratory for comparison with glucose tests performed with the Modular P analyzer. This method comparison was carried out as described below by the laboratory of the Flevo Hospital (FVZH) in Almere and the Maxima Medical Center (MMC) in Veldhoven. * Accu-Chek Inform II test strip, report from November 21, data from Roche Diagnostics Netherlands. Method comparison Samples from neonatal patients; only samples with glucose concentrations below 7 mmol/l were included in the method comparison. Sample collection; lithium-heparin whole blood taken from the heel. Measurements; MMC: glucose from blood taken from the heel with the Accu-Chek Inform II test strip was measured directly on the Accu-Chek Inform II. After the whole blood had been centrifuged, the glucose was also tested with the Modular P analyzer. During this process, the time difference between the two measurements was kept at a minimum (less than 2 minutes). FVZH: the lithium-heparin sample was taken from the heel of the newborn. Then the same sample was used to measure the glucose with the Accu-Chek Inform II test strip and the Modular P analyzer. Measurements were performed on different weekdays within a time span of several weeks.

3 Results of the method comparison: Fig Hospital n Method X Lot number of test strip Method Y Lot number of test strip Slope () Intercept () 1 FVZH 22 Modular p ACI II* ( ).19 (.54.84) 2 FVZH 21 Modular p ACI II ( ).1 (.52.67) 3 FVZH 21 ACI II 4711 ACI II ( ).1 (.58.18) 4 MMC 13 Modular p ACI II ( ).22 ( ) 5 MMC 13 Modular p ACI II ( ).2 ( ) 6 MMC 13 ACI II ACI II ( ).37 ( ) r: correlation coefficient; : 95 % confidence interval; all results are indicated in mmol/l. *ACI II: Accu-Chek Inform II The table above shows the results of the method comparison in neonate samples with glucose concentrations below 7 mmol/l at the Flevo Hospital (FVZH) and the Maxima Medical Center (MMC). A regression analysis was conducted in accordance with the Passing-Bablok method. Conclusion The validation shows that the results from the Accu-Chek Inform II test strip are highly comparable to glucose measurements performed with the Modular P analyzer at the central clinical chemistry laboratory. This demonstrates that the Accu-Chek Inform II test strip is also reliable for use on newborns. Fig 1: Results of the method comparison: Modular P versus Accu-Chek Inform II lot 4711 n 22 Modular P 1 Range ACI II (4711) 1 ACI II lot 4711 Passing-Bablok fit ( x) Constant Proportional H: Proportional bias = 1. H1: Proportional bias 1. plot

4 Fig 2: Results of the method comparison: versus Accu-Chek Inform II lot 4733 Fig 3: Results of the method comparison: Accu-Chek Inform II lot 4711 versus Accu-Chek Inform II lot 4733 n 21* 1 Constant n 21* ACI II (4711) 1 Constant Range ACI II (4733) 1 Proportional Range ACI II (4733) 1 Proportional *cases excluded: 1 due to missing values. H: Proportional bias = 1. H1: Proportional bias 1. *cases excluded: 1 due to missing values. H: Proportional bias = 1. H1: Proportional bias 1. ACI II lot 4733 Passing-Bablok fit ( x) ACI II lot 4733 Passing-Bablok fit ( x) ACI II lot 4711 plot plot

5 Fig 4: Results of the method comparison: versus Accu-Chek Inform II lot 4711 Fig 5: Results of the method comparison: versus Accu-Chek Inform II lot n 13 1 Constant n 13 1 Constant Range ACI II (4711) 1 Proportional Range ACI II (37242) 1 Proportional H: Proportional bias = 1. H1: Proportional bias 1. H: Proportional bias = 1. H1: Proportional bias 1. ACI II lot 4711 Passing-Bablok fit ( x) ACI II lot Passing-Bablok fit ( x) plot plot

6 Fig 6: Results of the method comparison: Accu-Chek Inform II lot versus Accu-Chek Inform II lot 4711 n 13 ACI II (37242) 1 Range 5.4 ACI II (4711) 1 Constant Proportional H: Proportional bias = 1. H1: Proportional bias 1. ACI II lot 4711 Passing-Bablok fit ( x) ACI II lot plot

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%) cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care

More information

Accuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method

Accuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method Accuracy and Precision Performance of the Accu-Chek Mobile System I. ACCURACY The accuracy of the system was assessed according to ISO 15197. Introduction The purpose of this study was to determine the

More information

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014 Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The

More information

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803) Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess

More information

1 PROTOCOL. Comparison Study Summary

1 PROTOCOL. Comparison Study Summary Comparison Study Summary OnSite Health 210 S. Michigan Street, Suite 110 South Bend, IN 46601 574-243-5108 ext. 203 July 27, 2015 1 PROTOCOL This internal evaluation was conducted on July 17 & 20, 2015

More information

Performance of three portable blood glucose monitoring devices used in a veterinary application

Performance of three portable blood glucose monitoring devices used in a veterinary application Performance of three portable blood glucose monitoring devices used in a veterinary application Prepared by Michael I. Lindinger, PhD, Vice-President of Research, The Nutraceutical Alliance, Campbellville,

More information

Introduction. Study Objective. Methodology. Results. Data Analysis. Interpretation

Introduction. Study Objective. Methodology. Results. Data Analysis. Interpretation WHITE PAPER COMPARISON STUDY BETWEEN THE HETTICH EBA 200S (HP) BLOOD TUBE PACKAGE 8 (FORMALLY HELMER QUICKSPIN PLUS) AND A TRADITIONAL 3 LITER CENTRIFUGE IN BLOOD CHEMISTRY DETERMINATIONS Introduction

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose

More information

Urgent field safety notice

Urgent field safety notice Accu-Chek Aviva, Accu-Chek Performa Urgent field safety notice May 2018 Important information on selected lots of Accu-Chek Aviva (50s, 10s) and Accu-Chek Performa (10s) strips potentially showing an increased

More information

Introduction. Study Objective. Methodology. Results. Data Analysis. Interpretation

Introduction. Study Objective. Methodology. Results. Data Analysis. Interpretation WHITE PAPER COMPARISON STUDY BETWEEN THE HETTICH EBA200S (HP) BLOOD TUBE PACKAGE 8 (FORMERLY QUICKSPIN PLUS) AND A SMALL CLINICAL CENTRIFUGE IN BLOOD CHEMISTRY DETERMINATIONS Introduction The Hettich EBA

More information

Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase

Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase Washington University School of Medicine Digital Commons@Becker Open Access Publications 2013 Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase Dennis J. Dietzen

More information

3. System Accuracy Evaluation LUX Hgb test 1. Samples

3. System Accuracy Evaluation LUX Hgb test 1. Samples 3. System Accuracy Evaluation LUX Hgb test 1. Samples - Type of sample: Capillary Blood System accuracy evaluation was performed with 200 fresh capillary blood samples, each with sufficient volume to be

More information

Urgent field safety notice

Urgent field safety notice Accu-Chek Aviva Urgent field safety notice May 2018 Important information on selected lots of Accu-Chek Aviva (50s, 10s) potentially showing an increased number of strip errors prior to dosing or biased

More information

Evaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers

Evaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers Evaluation Report Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers Location: Operators: Eurolyser Diagnostica GmbH Simone Wieser, Bettina Schönauer, Dr. Jürgen Berlanda Report created on 17

More information

Aina Blood Monitoring System

Aina Blood Monitoring System Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th

More information

Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES

Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES Accuracy of Roche Accu-Chek Inform Whole Blood Capillary, Arterial, and Venous Glucose Values in Patients Receiving Intensive Intravenous

More information

cobas u 601 urine analyzer Providing precise and secure test results

cobas u 601 urine analyzer Providing precise and secure test results Providing precise and secure test results 1 of 6 Providing precise and secure test results cobas u 601 urine analyzer is a modular, analytical system platform for urine strip testing and features a fully

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,

More information

Elecsys bone marker panel. Optimal patient management starts in the laboratory

Elecsys bone marker panel. Optimal patient management starts in the laboratory bone marker panel Optimal patient management starts in the laboratory Complete solution for osteoporosis The most complete bone metabolism panel on a single platform bone marker assays are important diagnostic

More information

I know my value CoaguChek XS Pro system

I know my value CoaguChek XS Pro system I know my value CoaguChek XS Pro system The smart way to manage your anticoagulation clinic CoaguChek XS Pro system A barcode scanner can make a difference There has been an extraordinary increase in the

More information

Management of Central Venous Access Devices. Blood Glucose Monitoring

Management of Central Venous Access Devices. Blood Glucose Monitoring Management of Central Venous Access Devices Blood Glucose Monitoring Purpose To provide education on the standard of care regarding the use and monitoring of the Accu- Chek Blood glucose machine, including

More information

Essential advice for people with diabetes from Accu-Chek. Time to test. G1812 Time to Test leaflet v5.indd 1 25/02/ :43

Essential advice for people with diabetes from Accu-Chek. Time to test. G1812 Time to Test leaflet v5.indd 1 25/02/ :43 Essential advice for people with diabetes from Accu-Chek Time to test G1812 Time to Test leaflet v5.indd 1 25/02/2016 16:43 Why do I need to check m Going for a jog, drinks after work, eating out or a

More information

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX AIR Blood Glucose Monitoring System, from Trividia

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

NICE to. know... The facts you need to make the most of your diabetes care.

NICE to. know... The facts you need to make the most of your diabetes care. NICE to know... 3 The facts you need to make the most of your diabetes care. NICE to know... The National Institute for Health and Care Excellence (NICE) provides advice on the care and support that should

More information

Multicenter Evaluation of the Glucometer Elite XL Meter, an Instrument Specifically Designed for Use With Neonates

Multicenter Evaluation of the Glucometer Elite XL Meter, an Instrument Specifically Designed for Use With Neonates Emerging Treatments and O R I G I N A L A R T I C L E Technologies Multicenter Evaluation of the Glucometer Elite XL Meter, an Instrument Specifically Designed for Use With Neonates JOËL GIROUARD, MD JEAN-CLAUDE

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia

More information

Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube

Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube Coagulation and Transfusion Medicine / Discard Tube Unnecessary for Coagulation Assays Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube Maarten

More information

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Polymer Technology Systems, Inc. CardioChek PA Comparison Study Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory

More information

Current Status: Active PolicyStat ID: Accu-Chek Whole Blood Glucose Testing Policy PURPOSE

Current Status: Active PolicyStat ID: Accu-Chek Whole Blood Glucose Testing Policy PURPOSE Current Status: Active PolicyStat ID: 4749849 Origination: 12/2013 Last Approved: 03/2018 Last Revised: 03/2018 Next Review: 03/2020 Owner: Area: References: Karla Gonzalez Anchondo: Point of Care Coordinator

More information

Time to test. Do you do it enough? Essential advice for people with diabetes from Accu-Chek

Time to test. Do you do it enough? Essential advice for people with diabetes from Accu-Chek Do you do it enough? Essential advice for people with diabetes from Accu-Chek Regular tests provide valuable information for both you and your diabetes team Find out what level of testing is recommended

More information

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff

More information

Guidance for the choice of blood glucose testing meters, strips and lancets

Guidance for the choice of blood glucose testing meters, strips and lancets Guidance for the choice of blood glucose testing meters, strips and lancets This guidance is intended to assist healthcare professionals in their selection of appropriate blood glucose meters and testing

More information

13. POINT-OF-CARE CREATININE TESTING IN HIGH- RISK PATIENTS

13. POINT-OF-CARE CREATININE TESTING IN HIGH- RISK PATIENTS How to Cite this article: Point-of-Care Creatinine Testing in High-Risk Patients - ejifcc 20/01 2009 http://www.ifcc.org 13. POINT-OF-CARE CREATININE TESTING IN HIGH- RISK PATIENTS Yolanda B. de Rijke

More information

MQ Comparison of glucometers using whole blood

MQ Comparison of glucometers using whole blood MQ 2017-1 Comparison of glucometers using whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing devices

More information

Validity and Reliability of a Glucometer Against Industry Reference Standards

Validity and Reliability of a Glucometer Against Industry Reference Standards 514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference

More information

SKUP Scandinavian evaluation of laboratory equipment for primary health care

SKUP Scandinavian evaluation of laboratory equipment for primary health care S K U P i n D e n m a r k, D E K S, R i g s h o s p i t a l e t - G l o s t r u p, DK- 2 6 0 0 G l o s t r u p, P h o n e + 4 5 3 8 6 3 4 4 0 0, w w w. s k u p. n u S K U P i n N o r w a y, N o k l u s,

More information

MQ Comparison of glucometers using whole blood

MQ Comparison of glucometers using whole blood MQ 2018-2 Comparison of glucometers using whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing devices

More information

Better connected for better decisions

Better connected for better decisions Introducing the Accu-Chek Connect diabetes management system Better connected for better decisions Help your patients use the power of their mobile devices and the web to manage their diabetes How many

More information

cobas h 232 POC system The power to make decisions in cardiac emergencies

cobas h 232 POC system The power to make decisions in cardiac emergencies cobas h 232 POC system The power to make decisions in cardiac emergencies cobas h 232 POC system Rapid results at the Point of Care (POC) Wherever and whenever you face an emergency situation, measurement

More information

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,

More information

Diabetes Care Publish Ahead of Print, published online February 2, 2011

Diabetes Care Publish Ahead of Print, published online February 2, 2011 Clinical Care/Education/Nutrition/Psychosocial Research O R I G I N A L A R T I C L E Diabetes Care Publish Ahead of Print, published online February 2, 2011 Self-Monitoring of Blood Glucose: The Use of

More information

ACCESS hstni SCIENTIFIC LITERATURE

ACCESS hstni SCIENTIFIC LITERATURE ACCESS hstni SCIENTIFIC LITERATURE 2017 2018 Table of contents Performance Evaluation of Access hstni A critical evaluation of the Beckman Coulter Access hstni: Analytical performance, reference interval

More information

System Accuracy Evaluation LUX Chol HDL test

System Accuracy Evaluation LUX Chol HDL test System Accuracy Evaluation LUX Chol HDL test 1. Minimum Acceptable System Accuracy 1.1 System Accuracy Requirement The minimum acceptable accuracy for results produced by a Lipidmonitoring systemshall

More information

MQ Comparison of glucometers using heparin whole blood

MQ Comparison of glucometers using heparin whole blood MQ 2014-1 Comparison of glucometers using heparin whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing

More information

Plug-and-play goes online

Plug-and-play goes online The next-generation Accu-Chek T Smart Pix software Plug-and-play goes online Proven tools for efficient and effective diabetes management 1 even for patients who are not in your office How many of these

More information

Variation in Glucose and A1c Measurement: Which Diabetes Test is Best?

Variation in Glucose and A1c Measurement: Which Diabetes Test is Best? Variation in Glucose and A1c Measurement: Which Diabetes Test is Best? Martha E Lyon, PhD, DABCC, FAACC Division of Clinical Biochemistry Department of Pathology & Laboratory Medicine Saskatchewan Health

More information

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff

More information

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide LESSON 6-3 Laboratory Reagent Preparation and Calculations Worksheet LESSON 6-4 Chemistry Instrumentation in the Physician Office Laboratory Worksheet LESSON 6-6 Measuring Blood Glucose Worksheet LESSON

More information

SCIEX Vitamin D 200M Assay for the Topaz System

SCIEX Vitamin D 200M Assay for the Topaz System The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human

More information

In Vitro Diagnostic Glucose Test System

In Vitro Diagnostic Glucose Test System CDRH Final Guidance 2-Page Cover Sheet Guidance for Industry In Vitro Diagnostic Glucose Test System Document issued on: July 6, 1998 U.S. Department Of Health and Human Services Food and Drug Administration

More information

Non-competitive Internal Control Concept for PCR-based Qualitative Assays

Non-competitive Internal Control Concept for PCR-based Qualitative Assays Christian O. Simon Roche Molecular Diagnostics Rotkreuz, CH Non-competitive Internal Control Concept for PCR-based Qualitative Assays 23 rd SoGAT Meeting, 16 th 17 th April, 2012, Vilnius, Lithuania Design

More information

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System Embrace No Code Report Clinical Evaluation for Embrace No Code Blood Glucose Monitoring System Applicant: Apex Biotechnology Corp. Test Date: 2007/8/28~2007/9/21 Clinical Site: SCHMIDT Group Practice Clinic

More information

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version. Department Of Pathology POC.514.06- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 6 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.05 LBH 1 OF 7 Printed copies are

More information

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,

More information

Clinical Study Thrombin-Accelerated Quick Clotting Serum Tubes: An Evaluation with 22 Common Biochemical Analytes

Clinical Study Thrombin-Accelerated Quick Clotting Serum Tubes: An Evaluation with 22 Common Biochemical Analytes Advances in Hematology Volume 13, Article ID 7979, 8 pages http://dx.doi.org/1.11/13/7979 Clinical Study Thrombin-Accelerated Quick Clotting Serum Tubes: An Evaluation with Common Biochemical Analytes

More information

No influence of heparin plasma and other (pre)analytic variables on D-dimer determination

No influence of heparin plasma and other (pre)analytic variables on D-dimer determination CHAPTER 2 No influence of heparin plasma and other (pre)analytic variables on D-dimer determination R.E.G. Schutgens 1, F.J.L.M. Haas 2, H.J.T. Ruven 2, M. Spannagl 3, K. Horn 3, D.H. Biesma 1 Departments

More information

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE Brooke Ann Spaeth, Anne K Shephard, Mark DS Shephard, Timothy H Mathew ABSTRACT

More information

Fraser Health Authority Accu-chek Inform II: Training Script

Fraser Health Authority Accu-chek Inform II: Training Script 1. Introduction Ensure that each participant Signs the attendance Sheet (please PRINT first & last name, and middle initial legibly) Refer to; For Reference Only: o Quick Reference Guide o What s new document

More information

Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration

Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration Clinical Chemistry / Whole Blood Creatinine for egfr Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration Nichole

More information

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole

More information

There has been increasing interest in glucose measurement

There has been increasing interest in glucose measurement J Vet Intern Med 28;22:1189 1195 Comparison of the Accu-Chek Aviva Point-of-Care Glucometer with Blood Gas and Laboratory Methods of Analysis of Glucose Measurement in Equine Emergency Patients A.R. Hollis,

More information

THE VALLEY HOSPITAL Ridgewood, New Jersey PATIENT CARE SERVICES (PCS) POLICIES AND PROCEDURES

THE VALLEY HOSPITAL Ridgewood, New Jersey PATIENT CARE SERVICES (PCS) POLICIES AND PROCEDURES 58:03 THE VALLEY HOSPITAL Ridgewood, New Jersey PATIENT CARE SERVICES (PCS) POLICIES AND PROCEDURES SUBJECT: Whole Blood Glucose PURPOSE: The Accu-chek Inform II is intended for use to quantitatively measure

More information

ABSTRACT ORIGINAL RESEARCH. Guido Freckmann. Nina Jendrike. Annette Baumstark. Stefan Pleus. Christina Liebing. Cornelia Haug

ABSTRACT ORIGINAL RESEARCH. Guido Freckmann. Nina Jendrike. Annette Baumstark. Stefan Pleus. Christina Liebing. Cornelia Haug Diabetes Ther (2018) 9:683 697 https://doi.org/10.1007/s13300-018-0392-6 ORIGINAL RESEARCH User Performance Evaluation of Four Blood Glucose Monitoring Systems Applying ISO 15197:2013 Accuracy Criteria

More information

A meter designed for glucose self-measurement manufactured by Roche Diagnostics. Report from an evaluation organised by SKUP

A meter designed for glucose self-measurement manufactured by Roche Diagnostics. Report from an evaluation organised by SKUP SKUP Scandinavian evaluation of laboratory equipment for primary health care ACCU-CHEK Compact Plus A meter designed for glucose self-measurement manufactured by Roche Diagnostics Report from an evaluation

More information

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical

More information

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats IDEXX Catalyst SDMA Test for in-house measurement of SDMA in serum from dogs and cats By Graham Bilbrough, Barbara Evert, Karen Hathaway, Gina Panagakos, Jane Robertson, and Maha Yerramilli Introduction

More information

Human epididymal protein 4 The role of HE4 in the management of patients presenting with pelvic mass Publication abstracts

Human epididymal protein 4 The role of HE4 in the management of patients presenting with pelvic mass Publication abstracts Human epididymal protein 4 The role of HE4 in the management of patients presenting with pelvic mass Publication abstracts Ovarian cancer is diagnosed annually in more than 200,000 women worldwide, with

More information

Traceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users. David Ducroq.

Traceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users. David Ducroq. Traceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users David Ducroq Weqas Unit 6, Parc Tŷ Glas Llanishen Cardiff UK www.weqas.com Programme

More information

Simple Linear Regression

Simple Linear Regression Simple Linear Regression Assoc. Prof Dr Sarimah Abdullah Unit of Biostatistics & Research Methodology School of Medical Sciences, Health Campus Universiti Sains Malaysia Regression Regression analysis

More information

Multi User Model. Product Overview

Multi User Model. Product Overview Multi User Model Product Overview Features and Benefits NO CODING EASY TO USE o o o Large LCD display. Large buttons with simple functions Easy Setup. Pre-Setup before shipping. Easy strip insertion. SMALL

More information

This fi le may not print or view at 100%. Die lines and color breaks do not print.

This fi le may not print or view at 100%. Die lines and color breaks do not print. This fi le may not print or view at 100%. Die lines and color breaks do not print. Roche USA 43149 V6/1 05212685002 Schwarz Roche USA 43149 V6/2 05212685002 Cyan Roche USA 43149 V6/3 05212685002 Magenta

More information

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing

More information

Performance Characteristics of the Daktari CD4 System

Performance Characteristics of the Daktari CD4 System Daktari Diagnostics, Inc. 85 Bolton Street Cambridge, MA 02140 USA Performance Characteristics of the Daktari CD4 System 2 February 2015 Scope The Daktari CD4 system has been evaluated through studies

More information

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood

More information

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Institute of Diabetes Gerhardt Katsch Karlsburg (IDK) Report Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Sponsor: The study was

More information

Evidence Based Commutability: Bias 2 Study. Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA

Evidence Based Commutability: Bias 2 Study. Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA Evidence Based Commutability: Bias 2 Study Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA Australian Bias Studies conducted by Gus Koerbin, ACT Pathology on behalf of AACB Harmonisation Committee

More information

College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey

College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) The American Diabetes Association (ADA) recommends that laboratories use only HbA1c assay methods that have been NGSP certified and

More information

Performa BLOOD GLUCOSE METER. Operator s Manual

Performa BLOOD GLUCOSE METER. Operator s Manual Performa BLOOD GLUCOSE METER Operator s Manual The Accu Chek Performa System Intended Use The Accu Chek Performa Meter is designed to be used with the Accu Chek Inform II Test Strip to quantitatively

More information

Accu-Chek Insight. Travelling & Holiday Guide. 3583_HolidayGuideA5booklet_Final.indd 1

Accu-Chek Insight. Travelling & Holiday Guide. 3583_HolidayGuideA5booklet_Final.indd 1 Accu-Chek Insight Travelling & Holiday Guide 3583_HolidayGuideA5booklet_Final.indd 1 4/16/15 10:58 AM Preparation Carry your supplies in your hand luggage and split with a travel companion if possible.

More information

Current Status: Active PolicyStat ID: Implementation: 02/2012 Last Reviewed: 06/2017 Last Revised: 05/2017 Next Review: 06/2019

Current Status: Active PolicyStat ID: Implementation: 02/2012 Last Reviewed: 06/2017 Last Revised: 05/2017 Next Review: 06/2019 Current Status: Active PolicyStat ID: 3690447 Implementation: 02/2012 Last Reviewed: 06/2017 Last Revised: 05/2017 Next Review: 06/2019 Owner: Policy Area: References: Applicability: Christine Landerholm:

More information

Accuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems

Accuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems Original Article Accuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems DOI: 10.7860/JCDR/2018/34055.11671 Internal Medicine Section Anastasios Tentolouris 1,

More information

bedside glucose testing systems

bedside glucose testing systems 42 CAP TODAY JANUARY 2017 Part 1 of 4 Abbott Diabetes Care Abbott Diabetes Care Arkray 1420 Harbor Bay Parkway 1420 Harbor Bay Parkway 5198 W. 76th St. Alameda, CA 94502 Alameda, CA 94502 Edina, MN 55439

More information

Performance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose

Performance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose ORIGINL RTILE https://doi.org/10.15263/jlmqa.2016.38.4.234 Performance Evaluation of. raun Omnitest 5 lood Glucose Monitoring System for Self-Monitoring of lood Glucose Kyungso Jeon 1 and Miseon Shin 2

More information

Method Precision and Frequent Causes of Errors Observed in Point-of-Care Glucose Testing. A Proficiency Testing Program Perspective

Method Precision and Frequent Causes of Errors Observed in Point-of-Care Glucose Testing. A Proficiency Testing Program Perspective Method Precision and Frequent Causes of Errors Observed in Point-of-Care Testing A Proficiency Testing Program Perspective Berna Aslan, MD, 1 Julia Stemp, ART, 1 Paul Yip, PhD, 2 and Jane Gun-Munro, MLT,

More information

Application Note 201

Application Note 201 201 Application note: Determination of the of Milk Application note no.: 201 By: E. de Jong Date: January 2001 Copyright Delta Instruments 2005 www.deltainstruments.com Table of contents Table of contents

More information

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS)

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Type 2 Diabetes DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Clinical Guideline Self-monitoring of blood glucose (SMBG) should be considered but not be routinely offered for adults with type 2 diabetes

More information

Walsall CCG Supported Blood Glucose Meters April 2018

Walsall CCG Supported Blood Glucose Meters April 2018 Meter Name Picture Type Strips Lancets Information Company PRIMARY CARE INITIATION TEE2+ Uncomplicated Type 2 diabetes patient on TEE2 test CareSens Lancets Easy to use. Large clear display Bluetooth functionality

More information

Urine/CSF Albumin Reagent. Urine/CSF Albumin Calibrator

Urine/CSF Albumin Reagent. Urine/CSF Albumin Calibrator December 2014 PRODUCT ANNOUNCEMENT AU Clinical Chemistry Systems AU400/400 e, AU480, AU640/640 e, AU680, AU2700, AU5400, and AU5800 Product Urine/CSF Albumin Reagent Urine/CSF Albumin Reagent Urine/CSF

More information

GlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System

GlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System GlucCell TM SYSTEM USER S GUIDE Ver 2.1 Intended Use The GlucCell TM Glucose Monitoring System (The GlucCell TM System) is designed to quantitatively measure the concentration of glucose during cell culture.

More information

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS)

DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Type 2 Diabetes Clinical Guideline Self-monitoring of blood glucose (SMBG) should be considered but not routinely offered for adults with type 2 diabetes

More information

Urgent Field Safety Notice

Urgent Field Safety Notice Urgent Field Safety Notice Location, Date 2014 Important notice: Enhanced instructions for the proper handling of the Accu-Chek Mobile system to avoid the potential of falsely elevated

More information

University of Virginia Health System AccuChek Inform II Glucose Meters POC C2.21

University of Virginia Health System AccuChek Inform II Glucose Meters POC C2.21 PURPOSE Whole blood glucose testing with Accu-Chek Inform II monitors is used to monitor blood glucose concentrations for adjustments of insulin dosage in diabetic patients. It is not used for definitive

More information

User manual for healthcare professionals

User manual for healthcare professionals Blood Glucose Management System Accu-Chek Performa User manual for healthcare professionals CONTENTS 3 Introduction 4 Staff Responsibilities 5 Protocol 6 About the Accu-Chek Performa 7 Coding/Calibrating

More information

The performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.

The performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests. Chemical Composition Each Microdot strip contains the enzyme glucose dehydrogenase (Bacillus Sp.) = 1 IU and other ingredients (mediator, NAD, lysing agents etc.) = 200 µg. Test Strips When blood is added

More information

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Original: 5/02 Minor Revision: 9/15 Full Review: 1/15 Replaces: 1/15 Responsible Person: Point of Care Laboratory Specialist Approving Committee:

More information