1 Accu-Chek Inform II test strip Results of the evaluation study on newborns, conducted by the Flevo Hospital (FVZH) Almere and the Maxima Medical Center (MMC) Veldhoven in the Netherlands COBAS, ACCU-CHECK INFORM, MODULAR and LIFE NEEDS ANSWERS are trademarks of Roche. 212 Roche Roche Diagnostics International Ltd. CH-6343 Rotkreuz Switzerland
2 Evaluation of the Accu-Chek II test strip for neonatal patients The current Accu-Chek Inform II test strip is the successor of the Performa test strip for point-of-care glucose testing. The current test strip is an optimized version of the previous Performa test strip and is used in combination with the Accu-Chek Inform II meter. In November 21, this test strip was evaluated by two Dutch laboratories and approved for use on adults.* The results of neonatal samples were also recently tested with this test strip at the clinical chemistry laboratory for comparison with glucose tests performed with the Modular P analyzer. This method comparison was carried out as described below by the laboratory of the Flevo Hospital (FVZH) in Almere and the Maxima Medical Center (MMC) in Veldhoven. * Accu-Chek Inform II test strip, report from November 21, data from Roche Diagnostics Netherlands. Method comparison Samples from neonatal patients; only samples with glucose concentrations below 7 mmol/l were included in the method comparison. Sample collection; lithium-heparin whole blood taken from the heel. Measurements; MMC: glucose from blood taken from the heel with the Accu-Chek Inform II test strip was measured directly on the Accu-Chek Inform II. After the whole blood had been centrifuged, the glucose was also tested with the Modular P analyzer. During this process, the time difference between the two measurements was kept at a minimum (less than 2 minutes). FVZH: the lithium-heparin sample was taken from the heel of the newborn. Then the same sample was used to measure the glucose with the Accu-Chek Inform II test strip and the Modular P analyzer. Measurements were performed on different weekdays within a time span of several weeks.
3 Results of the method comparison: Fig Hospital n Method X Lot number of test strip Method Y Lot number of test strip Slope () Intercept () 1 FVZH 22 Modular p ACI II* ( ).19 (.54.84) 2 FVZH 21 Modular p ACI II ( ).1 (.52.67) 3 FVZH 21 ACI II 4711 ACI II ( ).1 (.58.18) 4 MMC 13 Modular p ACI II ( ).22 ( ) 5 MMC 13 Modular p ACI II ( ).2 ( ) 6 MMC 13 ACI II ACI II ( ).37 ( ) r: correlation coefficient; : 95 % confidence interval; all results are indicated in mmol/l. *ACI II: Accu-Chek Inform II The table above shows the results of the method comparison in neonate samples with glucose concentrations below 7 mmol/l at the Flevo Hospital (FVZH) and the Maxima Medical Center (MMC). A regression analysis was conducted in accordance with the Passing-Bablok method. Conclusion The validation shows that the results from the Accu-Chek Inform II test strip are highly comparable to glucose measurements performed with the Modular P analyzer at the central clinical chemistry laboratory. This demonstrates that the Accu-Chek Inform II test strip is also reliable for use on newborns. Fig 1: Results of the method comparison: Modular P versus Accu-Chek Inform II lot 4711 n 22 Modular P 1 Range ACI II (4711) 1 ACI II lot 4711 Passing-Bablok fit ( x) Constant Proportional H: Proportional bias = 1. H1: Proportional bias 1. plot
4 Fig 2: Results of the method comparison: versus Accu-Chek Inform II lot 4733 Fig 3: Results of the method comparison: Accu-Chek Inform II lot 4711 versus Accu-Chek Inform II lot 4733 n 21* 1 Constant n 21* ACI II (4711) 1 Constant Range ACI II (4733) 1 Proportional Range ACI II (4733) 1 Proportional *cases excluded: 1 due to missing values. H: Proportional bias = 1. H1: Proportional bias 1. *cases excluded: 1 due to missing values. H: Proportional bias = 1. H1: Proportional bias 1. ACI II lot 4733 Passing-Bablok fit ( x) ACI II lot 4733 Passing-Bablok fit ( x) ACI II lot 4711 plot plot
5 Fig 4: Results of the method comparison: versus Accu-Chek Inform II lot 4711 Fig 5: Results of the method comparison: versus Accu-Chek Inform II lot n 13 1 Constant n 13 1 Constant Range ACI II (4711) 1 Proportional Range ACI II (37242) 1 Proportional H: Proportional bias = 1. H1: Proportional bias 1. H: Proportional bias = 1. H1: Proportional bias 1. ACI II lot 4711 Passing-Bablok fit ( x) ACI II lot Passing-Bablok fit ( x) plot plot
6 Fig 6: Results of the method comparison: Accu-Chek Inform II lot versus Accu-Chek Inform II lot 4711 n 13 ACI II (37242) 1 Range 5.4 ACI II (4711) 1 Constant Proportional H: Proportional bias = 1. H1: Proportional bias 1. ACI II lot 4711 Passing-Bablok fit ( x) ACI II lot plot
(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)
cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care
More informationAccuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method
Accuracy and Precision Performance of the Accu-Chek Mobile System I. ACCURACY The accuracy of the system was assessed according to ISO 15197. Introduction The purpose of this study was to determine the
More information1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014
Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The
More information1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)
Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess
More information1 PROTOCOL. Comparison Study Summary
Comparison Study Summary OnSite Health 210 S. Michigan Street, Suite 110 South Bend, IN 46601 574-243-5108 ext. 203 July 27, 2015 1 PROTOCOL This internal evaluation was conducted on July 17 & 20, 2015
More informationPerformance of three portable blood glucose monitoring devices used in a veterinary application
Performance of three portable blood glucose monitoring devices used in a veterinary application Prepared by Michael I. Lindinger, PhD, Vice-President of Research, The Nutraceutical Alliance, Campbellville,
More informationIntroduction. Study Objective. Methodology. Results. Data Analysis. Interpretation
WHITE PAPER COMPARISON STUDY BETWEEN THE HETTICH EBA 200S (HP) BLOOD TUBE PACKAGE 8 (FORMALLY HELMER QUICKSPIN PLUS) AND A TRADITIONAL 3 LITER CENTRIFUGE IN BLOOD CHEMISTRY DETERMINATIONS Introduction
More informationUSING THE ACCESS AMH ASSAY IN YOUR LABORATORY
INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
More informationEquivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System
Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose
More informationUrgent field safety notice
Accu-Chek Aviva, Accu-Chek Performa Urgent field safety notice May 2018 Important information on selected lots of Accu-Chek Aviva (50s, 10s) and Accu-Chek Performa (10s) strips potentially showing an increased
More informationIntroduction. Study Objective. Methodology. Results. Data Analysis. Interpretation
WHITE PAPER COMPARISON STUDY BETWEEN THE HETTICH EBA200S (HP) BLOOD TUBE PACKAGE 8 (FORMERLY QUICKSPIN PLUS) AND A SMALL CLINICAL CENTRIFUGE IN BLOOD CHEMISTRY DETERMINATIONS Introduction The Hettich EBA
More informationPoint-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase
Washington University School of Medicine Digital Commons@Becker Open Access Publications 2013 Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase Dennis J. Dietzen
More information3. System Accuracy Evaluation LUX Hgb test 1. Samples
3. System Accuracy Evaluation LUX Hgb test 1. Samples - Type of sample: Capillary Blood System accuracy evaluation was performed with 200 fresh capillary blood samples, each with sufficient volume to be
More informationUrgent field safety notice
Accu-Chek Aviva Urgent field safety notice May 2018 Important information on selected lots of Accu-Chek Aviva (50s, 10s) potentially showing an increased number of strip errors prior to dosing or biased
More informationEvaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers
Evaluation Report Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers Location: Operators: Eurolyser Diagnostica GmbH Simone Wieser, Bettina Schönauer, Dr. Jürgen Berlanda Report created on 17
More informationAina Blood Monitoring System
Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th
More informationClinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES
Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES Accuracy of Roche Accu-Chek Inform Whole Blood Capillary, Arterial, and Venous Glucose Values in Patients Receiving Intensive Intravenous
More informationcobas u 601 urine analyzer Providing precise and secure test results
Providing precise and secure test results 1 of 6 Providing precise and secure test results cobas u 601 urine analyzer is a modular, analytical system platform for urine strip testing and features a fully
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System
Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,
More informationElecsys bone marker panel. Optimal patient management starts in the laboratory
bone marker panel Optimal patient management starts in the laboratory Complete solution for osteoporosis The most complete bone metabolism panel on a single platform bone marker assays are important diagnostic
More informationI know my value CoaguChek XS Pro system
I know my value CoaguChek XS Pro system The smart way to manage your anticoagulation clinic CoaguChek XS Pro system A barcode scanner can make a difference There has been an extraordinary increase in the
More informationManagement of Central Venous Access Devices. Blood Glucose Monitoring
Management of Central Venous Access Devices Blood Glucose Monitoring Purpose To provide education on the standard of care regarding the use and monitoring of the Accu- Chek Blood glucose machine, including
More informationEssential advice for people with diabetes from Accu-Chek. Time to test. G1812 Time to Test leaflet v5.indd 1 25/02/ :43
Essential advice for people with diabetes from Accu-Chek Time to test G1812 Time to Test leaflet v5.indd 1 25/02/2016 16:43 Why do I need to check m Going for a jog, drinks after work, eating out or a
More informationClinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System
Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX AIR Blood Glucose Monitoring System, from Trividia
More informationFreeStyle Lite A Blood Glucose Meter That Requires No Coding
Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood
More informationNICE to. know... The facts you need to make the most of your diabetes care.
NICE to know... 3 The facts you need to make the most of your diabetes care. NICE to know... The National Institute for Health and Care Excellence (NICE) provides advice on the care and support that should
More informationMulticenter Evaluation of the Glucometer Elite XL Meter, an Instrument Specifically Designed for Use With Neonates
Emerging Treatments and O R I G I N A L A R T I C L E Technologies Multicenter Evaluation of the Glucometer Elite XL Meter, an Instrument Specifically Designed for Use With Neonates JOËL GIROUARD, MD JEAN-CLAUDE
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System
Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia
More informationCollection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube
Coagulation and Transfusion Medicine / Discard Tube Unnecessary for Coagulation Assays Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube Maarten
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More informationCurrent Status: Active PolicyStat ID: Accu-Chek Whole Blood Glucose Testing Policy PURPOSE
Current Status: Active PolicyStat ID: 4749849 Origination: 12/2013 Last Approved: 03/2018 Last Revised: 03/2018 Next Review: 03/2020 Owner: Area: References: Karla Gonzalez Anchondo: Point of Care Coordinator
More informationTime to test. Do you do it enough? Essential advice for people with diabetes from Accu-Chek
Do you do it enough? Essential advice for people with diabetes from Accu-Chek Regular tests provide valuable information for both you and your diabetes team Find out what level of testing is recommended
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationGuidance for the choice of blood glucose testing meters, strips and lancets
Guidance for the choice of blood glucose testing meters, strips and lancets This guidance is intended to assist healthcare professionals in their selection of appropriate blood glucose meters and testing
More information13. POINT-OF-CARE CREATININE TESTING IN HIGH- RISK PATIENTS
How to Cite this article: Point-of-Care Creatinine Testing in High-Risk Patients - ejifcc 20/01 2009 http://www.ifcc.org 13. POINT-OF-CARE CREATININE TESTING IN HIGH- RISK PATIENTS Yolanda B. de Rijke
More informationMQ Comparison of glucometers using whole blood
MQ 2017-1 Comparison of glucometers using whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing devices
More informationValidity and Reliability of a Glucometer Against Industry Reference Standards
514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference
More informationSKUP Scandinavian evaluation of laboratory equipment for primary health care
S K U P i n D e n m a r k, D E K S, R i g s h o s p i t a l e t - G l o s t r u p, DK- 2 6 0 0 G l o s t r u p, P h o n e + 4 5 3 8 6 3 4 4 0 0, w w w. s k u p. n u S K U P i n N o r w a y, N o k l u s,
More informationMQ Comparison of glucometers using whole blood
MQ 2018-2 Comparison of glucometers using whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing devices
More informationBetter connected for better decisions
Introducing the Accu-Chek Connect diabetes management system Better connected for better decisions Help your patients use the power of their mobile devices and the web to manage their diabetes How many
More informationcobas h 232 POC system The power to make decisions in cardiac emergencies
cobas h 232 POC system The power to make decisions in cardiac emergencies cobas h 232 POC system Rapid results at the Point of Care (POC) Wherever and whenever you face an emergency situation, measurement
More informationGlucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers
Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,
More informationDiabetes Care Publish Ahead of Print, published online February 2, 2011
Clinical Care/Education/Nutrition/Psychosocial Research O R I G I N A L A R T I C L E Diabetes Care Publish Ahead of Print, published online February 2, 2011 Self-Monitoring of Blood Glucose: The Use of
More informationACCESS hstni SCIENTIFIC LITERATURE
ACCESS hstni SCIENTIFIC LITERATURE 2017 2018 Table of contents Performance Evaluation of Access hstni A critical evaluation of the Beckman Coulter Access hstni: Analytical performance, reference interval
More informationSystem Accuracy Evaluation LUX Chol HDL test
System Accuracy Evaluation LUX Chol HDL test 1. Minimum Acceptable System Accuracy 1.1 System Accuracy Requirement The minimum acceptable accuracy for results produced by a Lipidmonitoring systemshall
More informationMQ Comparison of glucometers using heparin whole blood
MQ 2014-1 Comparison of glucometers using heparin whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing
More informationPlug-and-play goes online
The next-generation Accu-Chek T Smart Pix software Plug-and-play goes online Proven tools for efficient and effective diabetes management 1 even for patients who are not in your office How many of these
More informationVariation in Glucose and A1c Measurement: Which Diabetes Test is Best?
Variation in Glucose and A1c Measurement: Which Diabetes Test is Best? Martha E Lyon, PhD, DABCC, FAACC Division of Clinical Biochemistry Department of Pathology & Laboratory Medicine Saskatchewan Health
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationWorksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide
LESSON 6-3 Laboratory Reagent Preparation and Calculations Worksheet LESSON 6-4 Chemistry Instrumentation in the Physician Office Laboratory Worksheet LESSON 6-6 Measuring Blood Glucose Worksheet LESSON
More informationSCIEX Vitamin D 200M Assay for the Topaz System
The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human
More informationIn Vitro Diagnostic Glucose Test System
CDRH Final Guidance 2-Page Cover Sheet Guidance for Industry In Vitro Diagnostic Glucose Test System Document issued on: July 6, 1998 U.S. Department Of Health and Human Services Food and Drug Administration
More informationNon-competitive Internal Control Concept for PCR-based Qualitative Assays
Christian O. Simon Roche Molecular Diagnostics Rotkreuz, CH Non-competitive Internal Control Concept for PCR-based Qualitative Assays 23 rd SoGAT Meeting, 16 th 17 th April, 2012, Vilnius, Lithuania Design
More informationClinical Evaluation for. Embrace No Code Blood Glucose Monitoring System
Embrace No Code Report Clinical Evaluation for Embrace No Code Blood Glucose Monitoring System Applicant: Apex Biotechnology Corp. Test Date: 2007/8/28~2007/9/21 Clinical Site: SCHMIDT Group Practice Clinic
More informationPROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.
Department Of Pathology POC.514.06- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 6 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.05 LBH 1 OF 7 Printed copies are
More informationIDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats
IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,
More informationClinical Study Thrombin-Accelerated Quick Clotting Serum Tubes: An Evaluation with 22 Common Biochemical Analytes
Advances in Hematology Volume 13, Article ID 7979, 8 pages http://dx.doi.org/1.11/13/7979 Clinical Study Thrombin-Accelerated Quick Clotting Serum Tubes: An Evaluation with Common Biochemical Analytes
More informationNo influence of heparin plasma and other (pre)analytic variables on D-dimer determination
CHAPTER 2 No influence of heparin plasma and other (pre)analytic variables on D-dimer determination R.E.G. Schutgens 1, F.J.L.M. Haas 2, H.J.T. Ruven 2, M. Spannagl 3, K. Horn 3, D.H. Biesma 1 Departments
More informationASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE
ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE Brooke Ann Spaeth, Anne K Shephard, Mark DS Shephard, Timothy H Mathew ABSTRACT
More informationFraser Health Authority Accu-chek Inform II: Training Script
1. Introduction Ensure that each participant Signs the attendance Sheet (please PRINT first & last name, and middle initial legibly) Refer to; For Reference Only: o Quick Reference Guide o What s new document
More informationComparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration
Clinical Chemistry / Whole Blood Creatinine for egfr Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration Nichole
More informationWhite Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard
White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More informationThere has been increasing interest in glucose measurement
J Vet Intern Med 28;22:1189 1195 Comparison of the Accu-Chek Aviva Point-of-Care Glucometer with Blood Gas and Laboratory Methods of Analysis of Glucose Measurement in Equine Emergency Patients A.R. Hollis,
More informationTHE VALLEY HOSPITAL Ridgewood, New Jersey PATIENT CARE SERVICES (PCS) POLICIES AND PROCEDURES
58:03 THE VALLEY HOSPITAL Ridgewood, New Jersey PATIENT CARE SERVICES (PCS) POLICIES AND PROCEDURES SUBJECT: Whole Blood Glucose PURPOSE: The Accu-chek Inform II is intended for use to quantitatively measure
More informationABSTRACT ORIGINAL RESEARCH. Guido Freckmann. Nina Jendrike. Annette Baumstark. Stefan Pleus. Christina Liebing. Cornelia Haug
Diabetes Ther (2018) 9:683 697 https://doi.org/10.1007/s13300-018-0392-6 ORIGINAL RESEARCH User Performance Evaluation of Four Blood Glucose Monitoring Systems Applying ISO 15197:2013 Accuracy Criteria
More informationA meter designed for glucose self-measurement manufactured by Roche Diagnostics. Report from an evaluation organised by SKUP
SKUP Scandinavian evaluation of laboratory equipment for primary health care ACCU-CHEK Compact Plus A meter designed for glucose self-measurement manufactured by Roche Diagnostics Report from an evaluation
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More informationIDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats
IDEXX Catalyst SDMA Test for in-house measurement of SDMA in serum from dogs and cats By Graham Bilbrough, Barbara Evert, Karen Hathaway, Gina Panagakos, Jane Robertson, and Maha Yerramilli Introduction
More informationHuman epididymal protein 4 The role of HE4 in the management of patients presenting with pelvic mass Publication abstracts
Human epididymal protein 4 The role of HE4 in the management of patients presenting with pelvic mass Publication abstracts Ovarian cancer is diagnosed annually in more than 200,000 women worldwide, with
More informationTraceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users. David Ducroq.
Traceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users David Ducroq Weqas Unit 6, Parc Tŷ Glas Llanishen Cardiff UK www.weqas.com Programme
More informationSimple Linear Regression
Simple Linear Regression Assoc. Prof Dr Sarimah Abdullah Unit of Biostatistics & Research Methodology School of Medical Sciences, Health Campus Universiti Sains Malaysia Regression Regression analysis
More informationMulti User Model. Product Overview
Multi User Model Product Overview Features and Benefits NO CODING EASY TO USE o o o Large LCD display. Large buttons with simple functions Easy Setup. Pre-Setup before shipping. Easy strip insertion. SMALL
More informationThis fi le may not print or view at 100%. Die lines and color breaks do not print.
This fi le may not print or view at 100%. Die lines and color breaks do not print. Roche USA 43149 V6/1 05212685002 Schwarz Roche USA 43149 V6/2 05212685002 Cyan Roche USA 43149 V6/3 05212685002 Magenta
More informationBLOOD GLUCOSE METERS ACCURACY REQUIREMENTS
BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing
More informationPerformance Characteristics of the Daktari CD4 System
Daktari Diagnostics, Inc. 85 Bolton Street Cambridge, MA 02140 USA Performance Characteristics of the Daktari CD4 System 2 February 2015 Scope The Daktari CD4 system has been evaluated through studies
More informationEvaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter
Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood
More informationReport. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013
Institute of Diabetes Gerhardt Katsch Karlsburg (IDK) Report Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Sponsor: The study was
More informationEvidence Based Commutability: Bias 2 Study. Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA
Evidence Based Commutability: Bias 2 Study Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA Australian Bias Studies conducted by Gus Koerbin, ACT Pathology on behalf of AACB Harmonisation Committee
More informationCollege of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey
College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) The American Diabetes Association (ADA) recommends that laboratories use only HbA1c assay methods that have been NGSP certified and
More informationPerforma BLOOD GLUCOSE METER. Operator s Manual
Performa BLOOD GLUCOSE METER Operator s Manual The Accu Chek Performa System Intended Use The Accu Chek Performa Meter is designed to be used with the Accu Chek Inform II Test Strip to quantitatively
More informationAccu-Chek Insight. Travelling & Holiday Guide. 3583_HolidayGuideA5booklet_Final.indd 1
Accu-Chek Insight Travelling & Holiday Guide 3583_HolidayGuideA5booklet_Final.indd 1 4/16/15 10:58 AM Preparation Carry your supplies in your hand luggage and split with a travel companion if possible.
More informationCurrent Status: Active PolicyStat ID: Implementation: 02/2012 Last Reviewed: 06/2017 Last Revised: 05/2017 Next Review: 06/2019
Current Status: Active PolicyStat ID: 3690447 Implementation: 02/2012 Last Reviewed: 06/2017 Last Revised: 05/2017 Next Review: 06/2019 Owner: Policy Area: References: Applicability: Christine Landerholm:
More informationAccuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems
Original Article Accuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems DOI: 10.7860/JCDR/2018/34055.11671 Internal Medicine Section Anastasios Tentolouris 1,
More informationbedside glucose testing systems
42 CAP TODAY JANUARY 2017 Part 1 of 4 Abbott Diabetes Care Abbott Diabetes Care Arkray 1420 Harbor Bay Parkway 1420 Harbor Bay Parkway 5198 W. 76th St. Alameda, CA 94502 Alameda, CA 94502 Edina, MN 55439
More informationPerformance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose
ORIGINL RTILE https://doi.org/10.15263/jlmqa.2016.38.4.234 Performance Evaluation of. raun Omnitest 5 lood Glucose Monitoring System for Self-Monitoring of lood Glucose Kyungso Jeon 1 and Miseon Shin 2
More informationMethod Precision and Frequent Causes of Errors Observed in Point-of-Care Glucose Testing. A Proficiency Testing Program Perspective
Method Precision and Frequent Causes of Errors Observed in Point-of-Care Testing A Proficiency Testing Program Perspective Berna Aslan, MD, 1 Julia Stemp, ART, 1 Paul Yip, PhD, 2 and Jane Gun-Munro, MLT,
More informationApplication Note 201
201 Application note: Determination of the of Milk Application note no.: 201 By: E. de Jong Date: January 2001 Copyright Delta Instruments 2005 www.deltainstruments.com Table of contents Table of contents
More informationDIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS)
Type 2 Diabetes DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Clinical Guideline Self-monitoring of blood glucose (SMBG) should be considered but not be routinely offered for adults with type 2 diabetes
More informationWalsall CCG Supported Blood Glucose Meters April 2018
Meter Name Picture Type Strips Lancets Information Company PRIMARY CARE INITIATION TEE2+ Uncomplicated Type 2 diabetes patient on TEE2 test CareSens Lancets Easy to use. Large clear display Bluetooth functionality
More informationUrine/CSF Albumin Reagent. Urine/CSF Albumin Calibrator
December 2014 PRODUCT ANNOUNCEMENT AU Clinical Chemistry Systems AU400/400 e, AU480, AU640/640 e, AU680, AU2700, AU5400, and AU5800 Product Urine/CSF Albumin Reagent Urine/CSF Albumin Reagent Urine/CSF
More informationGlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System
GlucCell TM SYSTEM USER S GUIDE Ver 2.1 Intended Use The GlucCell TM Glucose Monitoring System (The GlucCell TM System) is designed to quantitatively measure the concentration of glucose during cell culture.
More informationDIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS)
DIABETES BLOOD GLUCOSE MONITORING AT HOME (ADULTS) Type 2 Diabetes Clinical Guideline Self-monitoring of blood glucose (SMBG) should be considered but not routinely offered for adults with type 2 diabetes
More informationUrgent Field Safety Notice
Urgent Field Safety Notice Location, Date 2014 Important notice: Enhanced instructions for the proper handling of the Accu-Chek Mobile system to avoid the potential of falsely elevated
More informationUniversity of Virginia Health System AccuChek Inform II Glucose Meters POC C2.21
PURPOSE Whole blood glucose testing with Accu-Chek Inform II monitors is used to monitor blood glucose concentrations for adjustments of insulin dosage in diabetic patients. It is not used for definitive
More informationUser manual for healthcare professionals
Blood Glucose Management System Accu-Chek Performa User manual for healthcare professionals CONTENTS 3 Introduction 4 Staff Responsibilities 5 Protocol 6 About the Accu-Chek Performa 7 Coding/Calibrating
More informationThe performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.
Chemical Composition Each Microdot strip contains the enzyme glucose dehydrogenase (Bacillus Sp.) = 1 IU and other ingredients (mediator, NAD, lysing agents etc.) = 200 µg. Test Strips When blood is added
More informationTitle: Glucose Testing: StatStrip Glucose Meter Page 1 of 6
Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Original: 5/02 Minor Revision: 9/15 Full Review: 1/15 Replaces: 1/15 Responsible Person: Point of Care Laboratory Specialist Approving Committee:
More information