System Accuracy Evaluation LUX Chol HDL test
|
|
- Meghan Sabrina Stevens
- 5 years ago
- Views:
Transcription
1 System Accuracy Evaluation LUX Chol HDL test 1. Minimum Acceptable System Accuracy 1.1 System Accuracy Requirement The minimum acceptable accuracy for results produced by a Lipidmonitoring systemshall be as follows. Ninety-five percent (95%) of the individual HDL cholesterol results shall fall within ±15% at HDL cholesterol concentrations mmol/l ( 10mg/dL). 1.2 Data presentation For HDL cholesterol concentrations mmol/l ( 10mg/dL), results shall be expressed asthe percentage of values falling within the following intervals: ±5%, ±10% 2. Strip & Meter 1) Lots of HDL cholesterol teststrips; 1 strip lot per site Site # Strip lot number Quantity 1 L12BN1Z 150 strips from 15 vials 2 L12BN1S 150 strips from 15 vials 3 L12BD1H 150 strips from 15 vials 2) LUX Meters; 1 meter per site Site #1 Site #2 Site #3 meter # serial number meter # serial number meter # serial number 1 L L L )Cobas C111 The calibration accuracy of the Cobas C111was validated with Quality control solution calibration. 4)Instructions for use Lay users were provided with LUX METER-1 includinguser manual as draft labeled and packaged as it would for the US market in English. Lay users, who were able to read the user manual in English, were given sufficient time to read and understand the instructions and the contents of LUX METER 1. They did not receive additional training, other instruction, assistance or training material other than those. 5)Questionnaire Lay users were given the questionnaires to evaluate their understanding of the LUX METER andinstructions, and satisfaction of them. System Accuracy Evaluation report LUX Chol HDL test Page 1 of 10
2 3. Test date & environmental condition a)test Period: 14th February to 24th March 2014 b)temperature Range: 23.0 to Evaluation Protocol User performance study was performed using 150 participants from 3 sites - 50 participants at each site. Lay user performed his own fingerstick and tested him using LUX METER by himself. And then, healthcare gathered lay user s capillary whole blood by lay user s skin puncture and measured the HDL cholesterol using the same LUX METER. By additional skin puncture within 5 min, healthcare gathered capillary whole blood in the capillary anticoagulant tube. This capillary whole blood was measured by Cobas c111 After test, lay users filled the questionnaires. The questionnaires contained lay user information (sex, age, level of education), evaluation of their understanding of the system and instructions, and satisfaction of them. 4.1 Exclusion criteria a) Pregnancy or lactation period b) Severe acute disease c) Severe chronic disease with potential risk during the study d) Hematocrit value outside the range of 30~50% (based on manufacturer s instructions for use) System Accuracy Evaluation report LUX Chol HDL test Page 2 of 10
3 5.Test results System Accuracy Evaluation report LUX Chol HDL test Page 3 of 10
4 System Accuracy Evaluation report LUX Chol HDL test Page 4 of 10
5 System Accuracy Evaluation report LUX Chol HDL test Page 5 of 10
6 Lay user Lay user System Accuracy Evaluation report LUX Chol HDL test Page 6 of 10
7 Professional Professional System Accuracy Evaluation report LUX Chol HDL test Page 7 of 10
8 Human Factors and Questionnaire System Accuracy Evaluation report LUX Chol HDL test Page 8 of 10
9 System Accuracy Evaluation report LUX Chol HDL test Page 9 of 10
10 6.Conclusion [Acceptance criteria] Ninety-five percent (95%) of the individual HDLuser performance results shall fall within ±10% at HDL concentrations mmol/l ( 10mg/dL). [Result] A fine and direct correlation between Cobas c111 and LUX METER HDL strip for lay user vs professional were confirmed in the 150 blood samples with the correlation coefficient Lay User R2=0.977 and Professional R2= The results indicate that the use of the LUX METER HDL strip generate similar results as the Cobas c111, a laboratory instrument used in hospitals to test for HDL concentration. The data shown indicates that the results are within acceptance criteria. RESULT: Passed User Performance Evaluation Regression analysis results are summarized below: Arezzo, 24 th March 2014 BIOCHEMICAL SYSTEMS INTERNATIONAL SRL Dr.ssa Annalisa Imbriani QC Manager System Accuracy Evaluation report LUX Chol HDL test Page 10 of 10
3. System Accuracy Evaluation LUX Hgb test 1. Samples
3. System Accuracy Evaluation LUX Hgb test 1. Samples - Type of sample: Capillary Blood System accuracy evaluation was performed with 200 fresh capillary blood samples, each with sufficient volume to be
More informationClinical Evaluation for. Embrace No Code Blood Glucose Monitoring System
Embrace No Code Report Clinical Evaluation for Embrace No Code Blood Glucose Monitoring System Applicant: Apex Biotechnology Corp. Test Date: 2007/8/28~2007/9/21 Clinical Site: SCHMIDT Group Practice Clinic
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More informationAccuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method
Accuracy and Precision Performance of the Accu-Chek Mobile System I. ACCURACY The accuracy of the system was assessed according to ISO 15197. Introduction The purpose of this study was to determine the
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th
More informationEVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS
EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:23 ACCURACY AND USER EVALUATION REQUIREMENTS Douglas E. Bell, PhD Teri A. Sasse, RN, MS BACKGROUND: The
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch
More informationOGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking.
ITALIAN PRODUCT S E L F T E S T I N G Personal medical devices for metabolic disorders MONITORING and PREVENTION, taking advantage of reflectometric and electrochemical technologies. Blood glucose monitoring
More informationWhy do we need SD LipidoCare?
SD Biosensor, Inc. Innovative Global Leader of In Vitro Diagnostics Total supplier of Blood Glucose Monitoring System and Lipid Analyzer System in vitro diagnostics test kits for Point of Care Testing
More informationGB/T Translated English of Chinese Standard: GB/T
Translated English of Chinese Standard: GB/T 19634-2005 www.chinesestandard.net Sales@ChineseStandard.net GB ICS 11.100 C 44 National Standard of the People s Republic of China GB/T 19634-2005 In vitro
More information1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014
Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess
More informationWhite Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard
White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System
Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia
More informationWorksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide
LESSON 6-3 Laboratory Reagent Preparation and Calculations Worksheet LESSON 6-4 Chemistry Instrumentation in the Physician Office Laboratory Worksheet LESSON 6-6 Measuring Blood Glucose Worksheet LESSON
More information1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)
Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial
More informationAina Blood Monitoring System
Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System
Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,
More informationThe performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.
Chemical Composition Each Microdot strip contains the enzyme glucose dehydrogenase (Bacillus Sp.) = 1 IU and other ingredients (mediator, NAD, lysing agents etc.) = 200 µg. Test Strips When blood is added
More informationEquivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System
Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose
More informationClinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System
Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX AIR Blood Glucose Monitoring System, from Trividia
More informationOmnitest : 2013
ISO 15197: 2013 Omnitest 3 Blood Glucose Monitoring System Omnitest 3 Could it be easier? After strip insertion, Omnitest 3 is ready for the measurement. The code is set automatically. Only 0.3 µl sample
More information(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)
cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationSummary of Significant Changes at this Revision. Items Required
Update Approver/checker Summary of Significant Changes at this Revision Purpose and Scope Items Required 1. The Abbott FreeStyle Optium meter is a battery-powered device designed for the measurement of
More informationAlere INRatio 2 METHOD AND SAMPLE COLLECTION
Alere INRatio 2 METHOD AND SAMPLE COLLECTION 1. PURPOSE AND SCOPE The purpose of this document is to describe the procedure for performing an INR test using the Alere INRatio2 analyser. The INRatio2 meter
More informationReport. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013
Institute of Diabetes Gerhardt Katsch Karlsburg (IDK) Report Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Sponsor: The study was
More information1 PROTOCOL. Comparison Study Summary
Comparison Study Summary OnSite Health 210 S. Michigan Street, Suite 110 South Bend, IN 46601 574-243-5108 ext. 203 July 27, 2015 1 PROTOCOL This internal evaluation was conducted on July 17 & 20, 2015
More informationNICO /14 Nipro Diagnostics, Inc.
Clinical Performance of the TRUEresult twist Blood Glucose Monitoring System () Using ISO 15197:13 Criteria for Accuracy Utilizing GDH-FAD Enzyme NICO-8 1/1 Nipro Diagnostics, Inc. NICO_8_mmol.indd CLINICAL
More informationBern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.
Journal of Diabetes Science and Technology Volume 5, Issue 4, July 2011 Diabetes Technology Society TECHNOLOGY REPORTS Accuracy of the CONTOUR Blood Glucose Monitoring System Bern, B.A., Cheryl Leazenby,
More informationPROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:
PROCEDURE TITLE: Bedside Glucose Monitoring Issuing Department: Clinical Director Signature: Departments Involved: Laboratory Nursing Effective Date: 10/97 Review Dates: 09/01, 07/02, 05/13 Revision Dates:
More informationTitle: Abbott Optium Xceed Glucose Meter Effective date: 16/05/2011. Summary of Significant Changes at this Revision. SOP updated.
COPY Summary of Significant Changes at this Revision SOP updated Purpose and Scope Items Required 1. The Abbott Optium Xceed meter is a battery-powered device designed for the measurement of blood glucose
More informationINTRO SAFETY INFORMATION GENERAL INFORMATION TESTING PREANALYTICAL PHASE CHECKINGS RUNNING THE TEST ALTERNATIVE SITES MEMORY
04 12 30 60 66 72 INTRO SAFETY INFORMATION GENERAL INFORMATION THE LUX SYSTEM GETTING TO KNOW THE SYSTEM SETTING THE SYSTEM TESTING PREANALYTICAL PHASE CHECKINGS RUNNING THE TEST ALTERNATIVE SITES MEMORY
More informationItems in the package:
Intended Use: The EasyLife Hb Monitoring System is designed for in vitro diagnostic use only (external use only), and is suitable for self-testing. The system is for healthcare professionals and persons
More informationFreeStyle Lite A Blood Glucose Meter That Requires No Coding
Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationAll-in-One Portable Lipid Analyzer
All-in-One Portable Lipid Analyzer Small in Size! Big in Performance! Clinically proven accuracy & precision Evaluation study Extensive clinical studies proved the accuracy of the SD LipidoCare system
More informationManagement of Central Venous Access Devices. Blood Glucose Monitoring
Management of Central Venous Access Devices Blood Glucose Monitoring Purpose To provide education on the standard of care regarding the use and monitoring of the Accu- Chek Blood glucose machine, including
More informationClinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013
linical Performance of the TRUE2go lood Glucose Monitoring System Exceeds Minimum riteria for Accuracy Using ISO 197:213 LINIAL STUY SUMMARY EVALUATION OJETIVES This clinical evaluation: 1. Evaluates the
More informationTips for Training New Customers
Tips for Training New Customers Good Habits: Test Strips Recap vial immediately and tightly after removing a single test strip. Never remove the desiccant from the vial. Do not lay test strips out ahead
More informationI know my value CoaguChek XS Pro system
I know my value CoaguChek XS Pro system The smart way to manage your anticoagulation clinic CoaguChek XS Pro system A barcode scanner can make a difference There has been an extraordinary increase in the
More informationSummary of Significant Changes at this Revision
COPY Summary of Significant Changes at this Revision CR5890 Removal of Ketone guidelines - refer to Specialist Diabetes nurse (adult wards); Children's ward has their own protocol Remove all references
More informationMETHOD FOR THE MEASUREMENT OF BEDSIDE TROPONIN T (cobas h232)
Page 1 of 6 METHOD FOR THE MEASUREMENT OF BEDSIDE TROPONIN T (cobas h232) Edition No: 1.3.1 Author: Rosy Tirimacco / Paul Simpson Date: April 2012 Operative Date: April 2012 Review Date: November 2012
More informationAccu-Chek Inform II: Point of Care Glucose Testing. Sharp Healthcare 2014
Accu-Chek Inform II: Point of Care Glucose Testing Sharp Healthcare 2014 OBJECTIVES At the completion of this module the participant will be able to: Learn the proper technique of performing a finger stick
More informationBLOOD GLUCOSE METERS ACCURACY REQUIREMENTS
BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing
More informationCOBAS, ACCU-CHECK INFORM, MODULAR and LIFE NEEDS ANSWERS are trademarks of Roche Roche
Accu-Chek Inform II test strip Results of the evaluation study on newborns, conducted by the Flevo Hospital (FVZH) Almere and the Maxima Medical Center (MMC) Veldhoven in the Netherlands COBAS, ACCU-CHECK
More informationUrgent field safety notice
Accu-Chek Aviva Urgent field safety notice May 2018 Important information on selected lots of Accu-Chek Aviva (50s, 10s) potentially showing an increased number of strip errors prior to dosing or biased
More informationCoaguChek S System. Quick Reference Guide. This is a CLIA-waived System. Coagulation Testing with Fresh Whole Blood
CoaguChek S System This is a CLIA-waived System Quick Reference Guide Coagulation Testing with Fresh Whole Blood May not print or view at 100% All unmarked prints black. Coag Swoosh prints gradients of
More informationApolipoprotein A-1 ELISA
Apolipoprotein A-1 ELISA For the quantitative determination of apolipoprotein A1 in serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Please read carefully due to Critical
More informationFor Self-Testing Use Portable Whole Blood Test System. User Guide
For Self-Testing Use Portable Whole Blood Test System User Guide Table of Contents Materials Provided REF 1709 CardioChek analyzer system (1) Materials Needed but Not Provided PTS Panels self-test strips
More informationA meter designed for glucose self-measurement manufactured by LifeScan, Johnson & Johnson. Report from an evaluation organised by SKUP
SKUP Scandinavian evaluation of laboratory equipment for primary health care ONETOUCH Ultra A meter designed for glucose self-measurement manufactured by LifeScan, Johnson & Johnson Report from an evaluation
More informationEvaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)
Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November
More informationEvaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System
Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System Grace DeSantis, PhD Lisa Croner, PhD Michealle Havenhill, BA Shari Kipp, BSc (Honours) MBA Inverness Medical,
More informationIndividual Quality Control Plans (IQCP) Risk Assessment
Individual Quality Control Plans (IQCP) Risk Assessment James H. Nichols, PhD, DABCC, FACB Professor of Clinical Pathology, Microbiology and Immunology Medical Director of Clinical Chemistry and Point-of-Care
More informationA New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device: Accuracy and User Performance Evaluation
691301DSTXXX10.1177/1932296817691301Journal of Diabetes Science and TechnologyChristiansen et al research-article2017 Original Article A New, Wireless-enabled Blood Glucose Monitoring System That Links
More informationRat cholesterol ELISA Kit
Rat cholesterol ELISA Kit Catalog No. CSB-E11706r (96T) This immunoassay kit allows for the in vitro quantitative determination of rat Cholesterol concentrations in serum, plasma and other biological fluids.
More informationACE. Accurate, Compact & Economical Blood Glucose Meter. User Guide
ACE Accurate, Compact & Economical Blood Glucose Meter User Guide NOTICE: The essential features of the type described and illustrations herein remaining unaltered, Pulsatom Healthcare Pvt. Ltd. reserves
More informationUltra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC
VS UNISTRIP TECHNOLOGIES,LLC Charlotte, NC 28269 1-866-889-6229 V2-10/25/2018 UniStrip Generic Blood Glucose VS OneTouch Ultra Blue 1. Purpose This study is intended to evaluate the difference in values
More informationUser manual for healthcare professionals
Blood Glucose Management System Accu-Chek Performa User manual for healthcare professionals CONTENTS 3 Introduction 4 Staff Responsibilities 5 Protocol 6 About the Accu-Chek Performa 7 Coding/Calibrating
More informationLeadCare BLOOD LEAD ANALYZER. Quick Reference Guide
LeadCare II BLOOD LEAD ANALYZER Quick Reference Guide Precautions Precautions Caution The LeadCare II Blood Lead Analyzer is a CLIA-waived device. Facilities that perform tests with the LeadCare II System
More informationWHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: OraQuick HCV Rapid Antibody Test Kit WHO reference number: PQDx
WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: OraQuick HCV Rapid Antibody Test Kit WHO reference number: PQDx 0244-055-00 OraQuick HCV Rapid Antibody Test Kit with product codes 1001-0270
More informationCustomer Bulletin. Questions/Comments? Please call Customer Service at (Direct) or (Toll-free inside USA).
Customer Bulletin Assignment of Ranges for PTS Panels Multi-Chemistry Controls (Cat. No./REF 721) and PTS Panels Controls (Cat. No./REF 722) Background: PTS Panels Multi-Chemistry Controls and Controls
More informationA meter designed for glucose self-measurement manufactured by Roche Diagnostics. Report from an evaluation organised by SKUP
SKUP Scandinavian evaluation of laboratory equipment for primary health care ACCU-CHEK Compact Plus A meter designed for glucose self-measurement manufactured by Roche Diagnostics Report from an evaluation
More informationPoint of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter
Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter The Nova Stat Strip Blood Glucose meter is UHB s approved Glucometer
More informationPROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.
Department Of Pathology POC.514.06- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 6 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.05 LBH 1 OF 7 Printed copies are
More informationUser Guide. Test System. For Professional Use Portable Whole Blood Test System
User Guide Test System For Professional Use Portable Whole Blood Test System Table of Contents Materials Provided REF 1708 CardioChek PA analyzer (1) Materials Needed but not Provided Appropriate sterile,
More informationTitle/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised:
Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised: PURPOSE The Abbott FreeStyle Precision Pro (FSPP) Blood Glucose
More informationTechnical Bulletin. Cholesterol Reference Method Laboratory Network (CRMLN) Overview. TB Rev. 0
Technical Bulletin Reference Method Laboratory Network (CRMLN) Overview TB000020 Rev. 0 Reference Method Laboratory Network (CRMLN) - General Information With guidance from the US Centers for Disease Control
More informationGlucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers
Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,
More informationCareLink. software REPORT REFERENCE GUIDE. Management Software for Diabetes
CareLink Management Software for Diabetes software REPORT REFERENCE GUIDE How to use this guide Each type of CareLink report and its components are described in the following sections. Report data used
More information2/27/2017. Point of Care Testing- current and future opportunities for pharmacists in Virginia. Financial Disclosures. Pre-Assessment.
Point of Care Testing- current and future opportunities for pharmacists in Virginia Objectives Define the Pharmacy Practice Act in Virginia M AR G AR E T L AN D I S, P H AR M. D. K R O G E R P H AR M AC
More informationProcedure: CardioChek Plus Test System/PTS Panels Test Strips
Procedure: CardioChek Plus Test System/PTS Panels Test Strips Approved Director: Signature Date Prepared by Date: Adopted Date: Supersedes Procedure: Review Date: Reviewed By: Signature Date Distribution
More informationCurrent Status: Active PolicyStat ID: Implementation: 02/2012 Last Reviewed: 06/2017 Last Revised: 05/2017 Next Review: 06/2019
Current Status: Active PolicyStat ID: 3690447 Implementation: 02/2012 Last Reviewed: 06/2017 Last Revised: 05/2017 Next Review: 06/2019 Owner: Policy Area: References: Applicability: Christine Landerholm:
More informationSUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5
USER-EVALUATION ASCENSIA CONTOUR Blood Glucose Meter System Table of contents SUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5 MATERIALS AND METHODS... 6 ASCENSIA CONTOUR...
More informationBlood Glucose and ß -Ketone Monitoring System. Owner s Manual
Blood Glucose and ß -Ketone Monitoring System Owner s Manual English Dear Care Touch System Owner: Thank you for choosing the Care Touch Blood Glucose and ß-Ketone Monitoring System. This manual provides
More informationMission Hb accurately detects both
Mission Digital Hb-Testing System: Save Cost by testing 2- Parameters with Single Strip real time ie: i.e: Hemoglobin & Hematocrit-[PCV]. Presented By : Remedy Healthcare Private Limited, New Delhi, [INDIA].
More informationBlood Glucose Monitoring System. User Guide
Blood Glucose Monitoring System User Guide Table of Contents Introduction...2 Important Safety Instructions...2 About ipet PRO Blood Glucose Monitoring System...3 About ipet PRO Meter...4 About the ipet
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 15197 First edition 2003-05-01 In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus Systèmes d'essais
More informationChapter 14 Elderly, Home and Long-term Care Collections. Objectives:
EXERCISE 11: BEDSIDE GLUCOSE TESTING Textbook: Skills: Chapter 14 Elderly, Home and Long-term Care Collections 15 points Objectives: 1. Define diabetes mellitus. 2. Compare and contrast: Type 1 Diabetes,
More informationTitle: Glucose Testing: StatStrip Glucose Meter Page 1 of 6
Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Original: 5/02 Minor Revision: 9/15 Full Review: 1/15 Replaces: 1/15 Responsible Person: Point of Care Laboratory Specialist Approving Committee:
More informationAdiponectin, TG/HDL-cholesterol index and hs-crp. Predictors of insulin resistance.
ORIGINAL ARTICLE Adiponectin, TG/HDL-cholesterol index and hs-crp. Predictors of insulin resistance. Bonneau GA y Pedrozo WR 1 Ministry of Public Health, Province of Misiones, 2 School of Exact, Chemical
More informationWELL CARD HEALTH TOOL BOX
WELL CARD HEALTH TOOL BOX 1 ADVOCATE Brand Advantage Well Card Health Members Will Save 52-70% On Diabetic Supplies Free Talking Meter Or Dash Meter Programs To Well Card Health Members 3 to 5 day UPS
More informationBlood Glucose Monitoring System USER GUIDE. For Use With Ascensia CONTOUR Blood Glucose Meter and Ascensia MICROFILL Test Strips
Blood Glucose Monitoring System USER GUIDE For Use With Ascensia CONTOUR Blood Glucose Meter and Ascensia MICROFILL Test Strips Thank you for choosing the Ascensia CONTOUR Blood Glucose Monitoring System!
More informationEvaluation of a Portable Blood Lead Analyzer as an Alternative to Graphite Furnace Atomic Absorption Spectrophotometer
Evaluation of a Portable Blood Lead Analyzer as an Alternative to Graphite Furnace Atomic Absorption Spectrophotometer Tom M. Olewe 1*, Mutuku A. Mwanthi 2, Joseph K. Wang ombe 2 and Jeffrey K. Griffiths
More informationCoaguChek XS Plus System Whole Blood Protime/INR
1 CoaguChek XS Plus System Whole Blood Protime/INR A. PURPOSE The CoaguChek XS System is intended for use by professional healthcare providers for quantitative protime testing (PT) and calculated INR for
More informationSystem. User Manual for Self-Testing
CoaguChek System XS User Manual for Self-Testing Table of Contents CoaguChek XS System User Manual For investigational use only. Not for use in diagnostic procedures. Aspects of the CoaguChek XS System
More informationRoche CoaguChek XS Plus Error Codes For Firmware Version onwards
Roche CoaguChek XS Plus Error Codes For Firmware Version 03.01.00 onwards Description of Error The instrument will not turn on -The instrument has 'frozen' or not receiving power -Remove the power to the
More informationTITLE: Whole Blood Glucose Monitoring Using the Nova StatStrip Meter
Page 4 of 14 IV. Performance of Quality Control A. Purpose: To detect any errors due to system failures or operator performance B. Frequency: Levels 1 and 3 controls are run every 24 hours of use. C. Quality
More informationHuman LDL ELISA Kit. Innovative Research, Inc.
Human LDL ELISA Kit Catalog No: IRKTAH2582 Lot No: SAMPLE INTRODUCTION Human low-density lipoprotein (LDL) transports cholesterol from the liver to tissues where it is incorporated into cell membranes.
More informationPersonal medical devices for metabolic disorders MONITORING and PREVENTION, taking advantage of reflectometric and electrochemical technologies.
ITALIAN PRODUCT S E L F T E S T I N G Personal medical devices for metabolic disorders MONITORING and PREVENTION, taking advantage of reflectometric and electrochemical technologies. Altered blood parameters
More informationPrecision Xtra Training
Precision Xtra Training Objectives At the end of this module, the participant will be able to: List the steps to calibrate the machine Describe the steps for sample collection and blood glucose testing
More informationPRIMA Lab SA Via Antonio Monti Balerna Switzerland 0041 (0) primahometest.com
PRIMA Lab SA Via Antonio Monti 7 6828 Balerna Switzerland 0041 (0) 91 605 1030 info@primahometest.com primahometest.com prima self-testing kit user manual 2in1 / 3in1 Multiparameter diagnostic device cholesterol
More informationPrecision ceedpro. Online Training for. Glucometer & Capillary Collection
Online Training for Precision ceedpro Glucometer & Capillary Collection How long will it take? This module takes approximately 30 minutes to complete. What if I have to leave before I finish? Progress
More information5 BEST PRACTICES To Patient Safety. A Practical Guide to Blood Glucose Testing
5 BEST PRACTICES To Patient Safety A Practical Guide to Blood Glucose Testing In a seven-year hospital study, most errors (91% 97%) were operator-related rather than instrument-related. Nobels F, et al.,
More informationRETAIL BULK DIABETIC SUPPLY TOOL BOX
RETAIL BULK DIABETIC SUPPLY TOOL BOX 1 ADVOCATE Brand Advantage Employers Can Save 52-72% On Diabetic Supplies Free Talking Meter - No Coding Required One of the Most Accurate Blood Glucose Meters on the
More informationBlood Glucose Monitoring System
Blood Glucose Monitoring System Contents Chapter 1 OMRON Blood Glucose Monitoring System HEA-232 1.1 Introduction 1.2 List of Kit Contents 1.3 Product and Meter Display Chapter 2 Installing and Setting
More information(For National Authority Use Only) Name of Study Drug: to Part of Dossier:
2.0 Synopsis Abbott Laboratories Individual Study Table Referring to Part of Dossier: (For National Authority Use Only) Name of Study Drug: Volume: Niaspan Name of Active Ingredient: Page: Niacin extended-release
More information