HemoCue Glucose 201 RT Systems Performance Report

Size: px
Start display at page:

Download "HemoCue Glucose 201 RT Systems Performance Report"

Transcription

1 HemoCue Glucose 201 RT Systems Performance Report

2 Performance Report HemoCue Glucose 201 RT Systems This document summarizes the performance of the HemoCue Glucose 201 RT System and the HemoCue Glucose 201 DM RT System. Summary The HemoCue Glucose 201 RT Systems are based on a glucose dehydrogenase method and consist of a small dedicated analyzer and a unique disposable microcuvette. These systems combine the precision and accuracy of a central laboratory with the speed and convenience of obtaining results at the point-of-care. With just a fraction of a drop of whole blood and three simple steps, the HemoCue Glucose 201 RT Systems will produce immediate, accurate results for screening, diagnosing or monitoring diabetes, as well as monitoring neonatal blood glucose levels. The HemoCue 201 RT Analyzer stores patient and QC results together with date and time, the results can be downloaded to a PC. The analyzer is equipped with an outlet for printing the results. The HemoCue 201 DM RT Analyzer can forward measurements to host system using CLSI POCT1A. GPM

3 Content Performance Report HemoCue Glucose 201 RT Systems... 2 Summary... 2 Content... 3 HemoCue Glucose 201 RT System specifications... 4 HemoCue Glucose 201 DM RT System specifications... 5 System technology... 6 HemoCue Glucose 201 RT Microcuvette... 6 HemoCue Glucose 201 RT Analyzer... 8 HemoCue Glucose 201 DM RT Analyzer... 9 Traceability of calibration HemoCue Glucose 201 RT System and HemoCue Glucose 201 DM RT System Environmental Interferences Studies Precision studies with control and patient material Accuracy studies venous sample Comparative study for venous whole blood samples between HemoCue Glucose 201 RT System and ID GC-MS Accuracy study with capillary sample Capillary blood measured with HemoCue Glucose 201 RT Analyzer versus mean values for YSI 2300 STAT Plus Accuracy study with venous and capillary samples Comparative study between the HemoCue Glucose 201 RT System and YSI 2300 STAT Plus Accuracy study with arterial sample HemoCue Glucose 201 RT System compared with ABL 825 FLEX Accuracy and precision study with neonatal samples Clinical laboratory evaluation of neonatal samples Publications Comparison of the HemoCue Glucose 201 DM RT System with Cobas Modular (Roche) Comparison of the HemoCue Glucose 201 RT System with Architect ci8200 (Abbot) Comparison of the HemoCue Glucose 201 DM RT System with Vitros 5.1 (Ortho-Clinical Diagnostics) Comparison of the HemoCue Glucose 201 RT System with a hexokinase method Validation of a new generation POCT glucose device with emphasis on aspects important for glycemic control in the hospital care GPM

4 HemoCue Glucose 201 RT System specifications Intended purpose/intended use The HemoCue Glucose 201 RT System is used for quantitative determination of glucose in whole blood, supplementing the clinical evidence in the diagnosis and treatment of patients with diabetes as well as monitoring of neonatal blood glucose levels. The system should not be used on critically ill neonates in neonatal intensive care settings. The HemoCue Glucose 201 RT System is for In Vitro Diagnostic use only. The HemoCue Glucose 201 RT Analyzer is only to be used with HemoCue Glucose 201 RT Microcuvettes. For professional use only. Category Principle Calibration Sample Material Displayed range Limit of Detection Results Sample volume Shelf life Package Size Dimensions Weight Storage Temperature Operating temperature Operating humidity HemoCue Glucose 201 RT System specifications Modified glucose dehydrogenase in which the total amount of glucose is measured at the end point photometrically Factory calibrated and traceable to the ID GC-MS method, needs no further calibration and no coding Capillary, venous or arterial whole blood Plasma equivalent values 0-31 mmol/l (0-560 mg/dl) Whole blood values mmol/l (0-500 mg/dl) 0.56 mmol/l (10 mg/dl) Within 1 minute for normal glucose level <4 µl whole blood Hematocrit Range 20%-70% Altitude Memory function Power Interface Quality Control Data Management Each individually packed microcuvette: 12 months from manufacturing 25 single pack microcuvettes/box, 4 boxes/package 160 x 85 x 43 mm (6.3 x 3.35 x 1.69 inches) 350 g (0.77 pounds) including batteries Analyzer: 0-50 C ( F) Microcuvette: 0-30 C (32-86 F) C (59-80 F) Up to 90 % relative humidity Not affected by altitude 600 measurements AC adapter or 4 AA batteries Connectable to printer and HemoCue Basic Connect including barcode scanner Built-in self-test ; system can be verified using liquid control HemoCue Basic Connect software, store and view measurements and patient data GPM

5 HemoCue Glucose 201 DM RT System specifications Intended purpose/intended use The HemoCue Glucose 201 DM RT System is used for quantitative determination of glucose in whole blood, supplementing the clinical evidence in the diagnosis and treatment of patients with diabetes as well as monitoring of neonatal blood glucose levels. The system should not be used on critically ill neonates in neonatal intensive care settings. The HemoCue Glucose 201 DM RT System is for In Vitro Diagnostic use only. The HemoCue Glucose 201 DM RT Analyzer is only to be used with HemoCue Glucose 201 RT Microcuvettes. For professional use only. Category Principle Calibration Sample Material Displayed range Limit of Detection Results Sample volume Shelf life Package Size Dimensions Weight Storage Temperature Operating temperature Operating humidity HemoCue Glucose 201 DM RT System specifications Modified glucose dehydrogenase in which the total amount of glucose is measured at the end point photometrically Factory calibrated and traceable to the ID GC-MS method, needs no further calibration and no coding Capillary, venous or arterial whole blood Plasma equivalent values 0-31 mmol/l (0-560 mg/dl) Whole blood values mmol/l (0-500 mg/dl) 0.56 mmol/l (10 mg/dl) Within 1 minute for normal glucose level <4 µl whole blood Hematocrit Range 20%-70% Altitude Memory function Barcode scanner Power Interface Quality Control Each individually packed microcuvette: 12 months from manufacturing 25 single pack microcuvettes/box, 4 boxes/package Analyzer: 170 x 93 x 50 mm (6.7 x 3.66 x 1.97 inches). Docking Station: 206 x 135 x 61 mm (8.1 x 5.3 x 2.4 inches) Analyzer: 350 g (0.77 pounds) including batteries Docking Station: 566 g (1.24 pounds) Analyzer: 0-50 C ( F) Microcuvette: 0-30 C (32-86 F) C (59-80 F) Up to 90% relative humidity Not affected by altitude 4000 test results, 500 QC results and 500 Analyzer log entries Built in barcode scanner Internal rechargeable batteries or docking station with AC adapter USB/LAN POCT1-A (CLIS standard) Built-in self-test ; system can be verified using liquid control GPM

6 System technology Test method The HemoCue Glucose 201 RT System and the HemoCue Glucose 201 DM RT System use the Glucose dehydrogenase method and photometric detection. The glucose reaction is an enzymatic method, in which tetrazolium salt (MTT) is used to obtain a quantification of glucose in visible light. At first, erythrocytes are hemolyzed by saponin. Mutarotase catalyzes the conversion of α- D-glucose to β-d-glucose. β-d-glucose reacts with NAD + (Nicotinamide Adenine Dinucleotide) in the presence of the enzyme glucose dehydrogenase. The β-d-glucose is oxidized to D-gluconolactone and NAD + is reduced to NADH. The color reaction occurs when NADH, catalyzed by the enzyme diaphorase, reduces MTT (Methylthiazolylphenyltetrazolium), to MTTH, a colored formazan. The amount of colored formazan is spectrophotometric measured at 670 nm, with a second measurement at 840 nm for background compensation. The analysis time depends on the glucose concentration in the sample, and is an end-point measurement of the chemical reaction. Start measurement no later than 40 seconds after filling the microcuvette. The measurement will take approximately 1 min for normal glucose levels. HemoCue Glucose 201 RT Microcuvette Microcuvette Glucose 201 RT Microcuvette is specially designed and serves both as a pipette and as a measuring microcuvette. It is for single use and made of copolyester plastic, which turns into H 2 O and CO 2 when incinerated. The cavity of the microcuvette is filled by capillary force with less than 4 µl capillary, venous or arterial whole blood. The cavity contains an exact amount of dry reagents for the enzyme reaction to take place. Individually wrapped microcuvettes Each disposable microcuvette is individually wrapped to avoid contamination and to offer maximum shelf life. GPM

7 Microcuvette lots A lot to lot comparison was performed using five different cuvette lots and three different EDTA whole blood samples prepared to glucose levels 2.2, 6.1 and 22.2 mmol/l, whole blood values (Table 1). Each sample and level was analyzed with five Glucose 201 RT Analyzers and with four replicates per cuvette lot in alternated order. Table 1 Total Range of mean values, for whole blood samples between five different microcuvette lots. Glucose level mmol/l (mg/dl ) Max Mean mmol/l (mg/dl) Min Mean mmol/l (mg/dl) Range mmol/l (mg/dl) Level (40) 2.3 (40.9) 2.0 (36.8) 0.2 (4.1) Level (110) 6.1 (110.1) 5.9 (106.6) 0.2 (3.5) Level (400) 22.6 (406.6) 21.9 (394.0) 0.7 (12.6) Storage and expiry date The microcuvettes are to be stored at 0-30 C (32-86 F). Unused microcuvettes should be kept in the original package but once an individual package is opened, the microcuvette must be used immediately or discarded. Use the microcuvettes prior to the expiry date printed on the package. The expiry date is set 12 months from manufacturing date. GPM

8 HemoCue Glucose 201 RT Analyzer Calibration and quality control The HemoCue 201 RT Analyzer is factory calibrated and does not need any further calibration by the end user. It has an internal quality control, the selftest. At every start up the selftest will automatically verify the performance of the optronic unit. The selftest is performed every second hour when the analyzer is switched on. To ensure accuracy, the analyzer checks the stability of the blank value every fifth second. External quality control If an external quality control test is to be performed, GlucoTrol -NG from Eurotrol B.V. is recommended by HemoCue. GlucoTrol -NG is a glucose control that exhibits qualities similar to blood. It is available in five physiological levels and has 30-day open vial stability when refrigerated (2-8 C). Units The HemoCue Glucose 201 RT analyzer is available in 2 different units, mmol/l or mg/dl. This is factory set and can not be changed. Whole blood and plasma equivalent results Two versions of the HemoCue Glucose 201 RT Analyzer are available, displaying either whole blood results or plasma equivalent results. For plasma equivalent analyzers, the plasma conversion is made according to IFCC using the factor 1.11 to convert whole blood result to plasma equivalent result (Ref; D Orazio et al: Approved IFCC recommendation on Reporting Results for Blood Glucose, Clinical Chemistry 51: , 2005). Limit of detection The limit of detection is defined as the lowest amount of an analyte in a sample that can be detected with stated probability. (Ref; Protocols for determination of limits of detection and limits of quantification, CLSI EP17-A) The limit of detection for the HemoCue Glucose 201 RT in whole blood is 0.56 mmol/l (10 mg/dl). Memory function The HemoCue Glucose 201 RT Analyzer automatically stores up to 600 results. When the memory is full the analyzer will automatically overwrite the oldest result. If needed, single or all results can be deleted. Storage The HemoCue Glucose 201 RT Analyzer can be stored and transported at 0-50 C ( F). The operating temperature is C (59-80 F) and the analyzer needs to reach ambient temperature before use. The analyzer operates at a humidity <90% relative humidity. Connectivity The HemoCue Glucose 201 RT Analyzer can be connected to a serial printer for immediate printout of measurement. When there is a need for improved patient tracking, HemoCue Basic Connect Software provides analyzer with basic data management functions and connectivity to EMR (Electronic Medical Record) systems. The software makes it possible to connect up to 6 analyzers to one computer, electronically transfer test results to computer and forwarding test results from computer to EMR systems. The data management contains optional prompting for operator ID, site ID, lab ID, patient ID, patient name and microcuvette lot no. The software enables users to initiate tests, add comments to the results and mark results, falling outside a user-defined range, with a flag. GPM

9 HemoCue Glucose 201 DM RT Analyzer The HemoCue Glucose 201 DM RT is the data management version of the HemoCue Glucose 201 RT. Calibration The HemoCue 201 DM RT Analyzer is factory calibrated and does not need any further calibration by end user. It has an internal quality control, the selftest. At every start up the selftest will automatically verify the performance of the optronic unit. The selftest is performed every eight hour when the analyzer is switched on. To ensure accuracy, the analyzer checks the stability of the blank value every fifth second. External quality control If an external quality control test is to be performed, GlucoTrol -NG from Eurotrol B.V. is recommended by HemoCue. GlucoTrol -NG is a glucose control that exhibits qualities similar to blood. It is available in five physiological levels and has a 30-day open vial stability when refrigerated (2-8 C). Units The HemoCue Glucose 201 DM RT Analyzer is available in 2 different units, mmol/l or mg/dl. This is factory set and can not be changed. Whole blood and plasma equivalent results Two versions of the HemoCue Glucose 201 DM RT Analyzer are available, either displaying whole blood results or plasma equivalent results. For plasma equivalent analyzers, the plasma conversion is made according to IFCC using the factor 1.11 to convert whole blood result to plasma equivalent result (Ref; D Orazio et al: Approved IFCC recommendation on Reporting Results for Blood Glucose, Clinical Chemistry 51: , 2005). Limit of detection The limit of detection is defined as the lowest amount of an analyte in a sample that can be detected with stated probability (Ref; Protocols for determination of limits of detection and limits of quantification, CLSI EP17-A). The limit of detection for the HemoCue Glucose 201 DM RT in whole blood is 0.56 mmol/l (10 mg/dl). Memory function The HemoCue Glucose 201 DM RT Analyzer automatically stores up to results. When the memory is full there are two options, either the analyzer stores new test result by overwriting the oldest result or lockout. At lockout no patient tests can be performed until the stored results have been deleted. If needed, single or all results can be deleted. Storage The HemoCue Glucose 201 DM RT Analyzer can be stored and transported at 0-50 C ( F). The operating temperature is C (59-80 F) and the analyzer needs to reach ambient temperature before use. The analyzer operates at a humidity <90% relative humidity. HemoCue DM docking station Full data management is possible when the Analyzer is placed in the docking station which is connected via USB or LAN to a PC hosting the DMS-Software. The battery is recharged while in the docking station. Up to five docking stations can be connected to each other with only one of them connected to an AC adapter. GPM

10 Data management features The HemoCue Glucose 201 DM RT features to ensure patient safety: Barcode scanning of patient identification and comparison with current patient lists. Performing measurement can be set to only be accessible by certified operators/analyzer (up to operators/analyzers) Allows for Short-Turn-Around-Test in emergency situations. Storage of Patient measurements. Quality Control storage of 500 measurements. Quality Control management, QC scheduling and lock-out. The analyzer is designed to provide optimal workflow for the operator with built-in barcode scanner, touch screen and the possibility to enter comments for the results. The storage and possibility to transfer patient and QC results, ensures that no results are lost. The results are transferred automatically when the analyzer is docked in the docking-station. The workflow for the Point-Of-Care coordinator is supported by connectivity to the HemoCue 201 DM - DMS software and to all major data management server software (observation reviewers). These software applications will grant control of all analyzers, cuvette and QC lots used. The analyzer allows the software to perform QC scheduling centrally and in addition, the analyzers can be locked remotely by using supervisory lockout. The analyzer also has linearity and proficiency functions. The daisy-chain function of the docking-station makes it possible to connect up to five (5) docking-stations, only using one AC adapter and one network (LAN) connection. Training HemoCue provides a web-based interactive e-learning application that can be used for certification/re-certification of operators. The HemoCue e-learning can be integrated with the HemoCue 201 DM - DMS software and to selected middleware to enable an automatic workflow for certification. The workflow is as follows; the operator receives a mail with a link to e-learning - the operator takes the tailored certification course - the e-learning updates the operator list in the software - finally the software updates the operator list in the analyzer. GPM

11 Traceability of calibration HemoCue Glucose 201 RT System and HemoCue Glucose 201 DM RT System The calibration of the HemoCue Glucose 201 RT Analyzer and the HemoCue Glucose 201 DM RT Analyzer and HemoCue 201 RT Microcuvettes are fully traceable to ID GC-MS. HemoCue calibrates reference analyzers and reference lot of microcuvettes 4 to 5 times a year with the ID GC-MS. These analyzers and lots are used as reference for all manufactured analyzers and microcuvettes at HemoCue. The ID GC-MS system is calibrated with a NIST standard. The ID GC-MS method is controlled in an external IFCC control program for reference labs. HemoCue also controls the YSI used internally at HemoCue with ID GC-MS (Figure 1). Figure 1 Traceability of Calibration, HemoCue Glucose 201 RT Systems. Environmental HemoCue has an environmental management system that fulfills the requirements of SS- EN-ISO 14001:2004 with respect to design and development, production and distribution of test system for In-Vitro Diagnostics. Further information can be seen at According to REACH (EG 1907/2006), HemoCue is a downstream user of chemicals and a manufacturer of articles. HemoCue Microcuvettes is defined as a product where no Material Safety Data Sheet needs to be sent to customer. GPM

12 Interferences Interference studies have been performed for HemoCue Glucose 201 RT (Ref; Interference testing in Clinical Chemistry CLSI protocol EP7-A2, Vol.25 No. 27, Paired-difference testing, Worst Case, section 7.1). Potentially interfering substances were added to samples and bias relative to a control of the same samples was evaluated. The potential interfering substances were tested at relatively high concentrations to simulate worst case and were generally tested at 2-3 different glucose levels. If interference was found or known from earlier studies, dose-response experiments, (Ref; CLSI EP7-A2 section 7.2) were performed. In some cases the level of interference of a substance had to be determined with patient samples (Ref; CLSI EP7-A2, section 8). These studies were based on the analysis of two groups of patient specimen, i.e. a test group and a control group (Table 2). Table 2 Interferences Interferent Highest concentration tested without interference Anti-coagulants EDTA 5 x normal level Heparin 2.8 x normal level (8000U/L) Lithium-Heparin 5 x normal level Lithium-Iodo-Acetate 5 x normal level Sodium fluoride/ EDTA 5 x normal level Sodium fluoride/ Potassium oxalate 2 x normal level Sodium fluoride/ Heparin 3 x normal level Exogenous analyte Salicylic Acid 4.34 mmol/l (60 mg/dl ) Ibuprofen 2.19 mmol/l (50 mg/dl) Tetracycline 20 mg/dl Dopamine 13 mg/dl Ephedrine 1 mg/dl L-Dopa 50 mg/dl Methyldopa 2.5 mg/dl Tolazamide 25 mg/dl Tolbutamide 3.7 mmol/l (100 mg/dl) Acetaminophen 1.32 mmol/l (20 mg/dl) Ascorbic Acid 0.34 mmol/l (6 mg/dl) Level defined by EP7, 0.57 mmol/l (10 mg/dl) tested with no interference. Caffeine 0.31 mmol/l (6 mg/dl) Endogenous analyte Creatinine 2655 µmol/l (30 mg/dl) Uric acid 1050 µmol/l (20 mg/dl) Urea 83 mmol/l (500mg/dL) Unconjugated Bilirubin 342 µmol/l (20 mg/dl) Conjugated Bilirubin 475 µmol/l (40 mg/dl) ph / Leukocytes (WBC)** /L HHb (Reduced Hemoglobin) >75% HbCO* 30% Hematocrit* 70% Cholesterol** 9.6 mmol/l (350 mg/dl) *Data from Dose Response studies **Data from Patient studies GPM

13 Interferent Sugar Maltose Lactose Mannose Fructose Galactose Dextran* (Replacement Agent) Highest concentration tested without interference 15 mmol/l (540 mg/dl) 15 mmol/l (540 mg/dl) 15 mmol/l (270 mg/dl) 15 mmol/l (270 mg/dl) 15 mmol/l (270 mg/dl) 26 g/l Interfering substances with limitations in operating manual and package Insert Glucosamine is a natural monosaccharide used to treat osteoarthritis. Blood concentration >1.44 mmol/l (26 Glucosamine* mg/dl) will interfere with the glucose measurement. See limitations in operating manual and package insert. Orally administrated ascorbic acid will not affect the measurement. Intravenously administrated, ascorbic acid Ascorbic acid* resulting in blood concentration >0.57 mmol/l (10 mg/dl) will interfere with the measurement. See limitations in operating manual and package insert. During a xylose absorption test the concentration of xylose will likely exceed >1.44mmol/L (22 mg/dl) and Xylose* interfere with the glucose measurement. See limitations in operating manual and package insert. High levels of triglyceride can cause turbidity in the blood Triglycerides** sample and disturb the measurement. See limitations in operating manual and package insert. Intralipid (fat emulsion for intravenous use) that give >500 mg/l could interfere with the measurement. If a specimen is taken in this situation the firmware could Intralipid* detect it as a deviation from the normal pattern and an E70 or E71 information message will be displayed. See limitations in operating manual and package insert. The range of normal levels of Met Hemoglobin is 0-1%. Met Hemoglobin* Levels >7% will interfere with the analysis. See limitations in operating manual and package insert. *Data from Dose Response studies **Data from Patient studies Limitations in operating manual and package insert (extraction) The measured blood glucose value of, grossly lipemic specimens (intralipids >500 mg/l), turbid specimens, specimens containing >7 % methemoglobin and high concentrations of glucosamine (>1.44 mmol/l, >26 mg/dl) may give incorrect results and should be interpreted with caution. At normal oral administration of xylose the HemoCue Glucose 201 RT System does not have any interference (Xylose <1.44 mmol/l, <22 mg/dl). However the high dosage used during a xylose absorption test may give incorrect results and should be interpreted with caution. GPM

14 Studies The HemoCue Glucose 201 RT System and the HemoCue Glucose 201 DM RT System are based on the same platform and have identical performance. 1. Precision studies with control and patient material Precision study with control material The precision for Glucose 201 RT, repeatability and within laboratory precision, was determined according to EP5-A2 Ref; (CLSI Document EP5-A2, Vol. 24, No. 25, Evaluation of Precision Performance of Quantitative Measurement Methods, Approved Guideline Second Edition) and ISO Three different levels of Eurotrol, GlucoTrol-NG control solutions were analyzed in duplicate twice a day during 20 days on ten different analyzers. Two different lots of microcuvettes were used in the study. In total 800 samples/level (Table 3). Table 3 Precision study with control material. Repeatability Within-laboratory precision Control level N Mean mmol/l SD pooled mmol/l CV % SD pooled mmol/l (mg/dl) CV % (mg/dl) (mg/dl) (45.9) 0.06 (1.15) (1.28) (101.8) 0.08 (1.36) (1.62) (305.7) 0.18 (3.26) (3.72) 1.2 Precision study, with patient sample material Precision data is obtained from duplicate samples in validation studies at four different sites. Two different analyzers and three microcuvette lots were used. Standard deviation (SD), for duplicate samples, within each group was calculated (Table 4). Table 4 Precision study with patient sample material. SD mmol/l (mg/dl) Glucose range mmol/l (mg/dl) Capillary samples n=222 Venous samples n=180 All samples n= (0-200) 0.14 (2.60) 0.13 (2.32) 0.14 (2.49) ( ) 0.19 (3.50) 0.31 (5.60) 0.27.(4.81) ( ) 0.35 (6.05) 0.35 (6.36) 0.34 (6.22) GPM

15 2. Accuracy studies venous sample Comparative study for venous whole blood samples between HemoCue Glucose 201 RT System and ID GC-MS Patient samples, reference method ID GC-MS The study was performed according to CLSI EP9-A2. The evaluation time was 7 working days. Each sample was analyzed in duplicate with the HemoCue Glucose 201 RT and in parallel with the comparative reference method ID GC-MS. Human venous whole blood samples from 104 patients at Linköping University Hospital were used. Glucose solution was used to spike some samples to higher glucose levels. The user in this evaluation is one employee at the site, who is usually assigned to perform analysis of glucose samples with the ID GC-MS. The study shows very good agreement with ID GC-MS (Figure 2 and Figure 3). Figure 2 Venous whole blood measured at the HemoCue Glucose 201 RT compared to ID GC-MS. Difference (Glucose 201RT first replicate - ID GC-MS mean values), mg/dl Mean values ID GC-MS, mmol/l Figure 3 Mean values ID GC-MS, mg/dl Bland-Altman Difference plot for Glucose 201 RT System versus ID GC-MS, mg/dl and mmol/l Difference (Glucose 201RT first replicate - ID GC-MS mean values), mmol/l GPM

16 3. Accuracy study with capillary sample Capillary blood measured with HemoCue Glucose 201 RT Analyzer versus mean values for YSI 2300 STAT Plus A total of 222 human capillary blood samples were collected and analysed (some capillary samples collected in EDTA-microtainer tubes were spiked with glucose standards to obtain high levels of glucose). The test users in this evaluation were assistant nurses at three different sites. The test user analysed the samples by using two HemoCue Glucose 201 RT Analyzers and in parallel the samples were analysed with a comparative method, YSI 2300 STAT Plus, in duplicate, by a professional user (figure 4 and figure 5). YSI, mean values (mmol/l) Glucose 201RT, 1st repl. (mg/dl) Coefficients: y 0 = ±1.804 a = ±0.011 r 2 = n = 222 A Glucose 201RT, 1st repl. (mmol/l) YSI, mean values (mg/dl) 0 Figure 4 Capillary blood measured at the Glucose 201 RT System versus mean values for YSI 2300 Stat Plus, mg/dl and mmol/l. Differences (Glucose 201RT, 1st repl. - YSI, mean values), mg/dl Figure 5 Mean of Glucose 201RT and YSI, mmol/l Mean of Glucose 201RT and YSI, mg/dl Bland-Altman difference plot for Glucose 201 RT System versus YSI 2300 STAT Plus, mg/dl and mmol/l Differences (Glucose 201RT, 1st repl. - YSI, mean values), mmol/l GPM

17 4. Accuracy study with venous and capillary samples Comparative study between the HemoCue Glucose 201 RT System and YSI 2300 STAT Plus Patient samples, reference method YSI 2300 STAT Plus The study was performed according to CLIA Guidance. A total of 400 human capillary and venous whole blood samples were collected and analysed. Some venous samples and some capillary samples collected in EDTA-microtainer tubes were spiked with glucose standards to obtain high levels of glucose. In total 60 samples were spiked. The test users in this evaluation were assistant nurses at the 4 sites. Before performing analysis the test users read the Instructions for Use, Quick Reference Guide and the Package Insert. The samples were analysed by the test user with one sample/analyzer, in total two HemoCue Glucose 201 RT Analyzers and in parallel the samples were analysed with a comparative method, YSI 2300 STAT Plus, in duplicate, by a professional user (Fig 6). YSI, mean values (mmol/l) D B 400 E C A A Glucose 201RT, 1st repl. (mg/dl) C B Glucose 201RT, 1st repl. (mmol/l) D YSI, mean values (mg/dl) 0 Figure 6 Consensus grid for first replicate for Glucose 201 RT System versus mean values for YSI 2300 STAT Plus, mg/dl and mmol/l, venous and capillary samples (n = 400). The study was performed according to CLIA Guidance. The grid is divided into zones signifying the degree of risk posed by the incorrect measurement: Zone A no effect on clinical action Zone B altered clinical action little or no effect on clinical outcome Zone C altered clinical action likely to affect clinical outcome Zone D represents altered clinical action could have significant medical risk Zone E represents altered clinical action could have dangerous consequences GPM

18 5. Accuracy study with arterial sample HemoCue Glucose 201 RT System compared with ABL 825 FLEX Human arterial heparin blood samples, leftover samples from 140 patients at the intensive care unit (ICU) at Lund University Hospital. The study was performed according to CLSI EP9- A2. Each sample was analysed in single with the HemoCue Glucose 201 RT, in parallel with one single replicate with a comparative method, ABL 825 FLEX (Radiometer). The users in this evaluation were employees at the site, who are usually assigned to perform analysis of glucose samples with the ABL 825 FLEX (figure 7 and figure 8). ABL 825 FLEX, single replicate (mmol/l) Coefficients: y 0 = 3.62 a = r 2 = n = Glucose 201RT, 1st repl (mg/dl) Glucose 201RT, 1st repl (mmol/l) ABL 825 FLEX, single replicate (mg/dl) 0 Figure 7 First replicates for Glucose 201 RT System versus single replicate for ABL 825 FLEX, mg/dl and mmol/l. Bland-Altman difference plot for Glucose 201 RT compared with ABL 825 FLEX, arterial samples Mean values [ABL and Glucose 201 RT], mmol/l Difference (Glucose 201RT - ABL), mg/dl Difference (Glucose 201RT - ABL), mmol/l Mean values [ABL and Glucose 201 RT], mg/dl Figure 8 Bland-Altman difference plot for Glucose 201 RT System (first replicate) and ABL 825 FLEX versus mean of ABL and Glucose 201 RT, mg/dl and mmol/l. GPM

19 6. Accuracy and precision study with neonatal samples Clinical laboratory evaluation of neonatal samples Human venous EDTA whole blood or capillary heparin blood samples from 103 neonates in Linköping and Holbæk were collected and analyzed. The users in the study were employees at these sites, and usually assigned to perform glucose measurements with the comparative laboratory method. Since sufficient samples with glucose levels of less than or equal 55 mg/dl (plasma values) were not obtained from neonates, the study was complemented with 21 umbilical cord blood samples analyzed at HemoCue. The glucose levels in most of the samples were natural. However, some samples from the Linköping site were spiked with glucose solution in order to cover a larger glucose range. Each sample was analyzed in duplicate with the HemoCue Glucose 201 RT System, in parallel in single replicate with the comparative method ABL 835 FLEX/ID-GCMS/YSI 2300 STAT Plus. All values are recalculated and presented in plasma values (Table 5, Fig 9). The system should not be used on critically ill neonates in neonatal intensive care settings. Table 5 Glucose mmol/l (mg/dl) (0-200) Mean value, precision, SD and CV for Glucose 201 RT for neonatal samples from non-neonatal intensive care units. No of samples Mean value Glucose 201 RT mmol/l (mg/dl) SD duplicates mmol/l (mg/dl) CV duplicat e % Mean value comparative method mmol/l (mg/dl) (87.6) 0.24 (4.4) (85.5) Comparative method (YSI, ABL, ID GC-MS), single, mmol/l Glucose 201 RT, single, mg/dl ABL 835 FLEX (neonatal patient samples) ID GC-MS (neonatal patient samples) YS 2300 STAT Plus (Umbilical cord samples) x=y Glucose 201 RT, single, mmol/l Comparative method (YSI, ABL, ID GC-MS), single, mg/dl Figure 9 Neonatal samples from non-neonatal intensive care units. Scatterplot of Glucose 201 RT first replicates versus comp. method first replicate (Plasma equivalent values), mg/dl and mmol/l. GPM

20 Publications Comparison of the HemoCue Glucose 201 DM RT System with Cobas Modular (Roche) Ref: Poster by Jukka Saarimies 1, Anja Rautajoki 2 Charlotta Roos 3 1) Hospital District of Southwest Finland, TYSKLAB, Department of Clinical Chemistry, Turku Finland. 2) HemoCue Finland, Espoo, Finland. 3) HemoCue AB Ängelholm, Sweden This study compared the HemoCue Glucose 201 RT System performance against two different lab systems; Glucose 201 DM RT vs. blood gas Cobas b 221(Roche) and vs. plasma base Cobas Modular (Figure 10), a hexokinase method. The study shows good agreement between the HemoCue Glucose 201 DM RT and the two comparative methods, measuring glucose over a wide measuring range. This indicates that the methods can be used interchangeably for clinical purposes within the healthcare settings. Evaluating the variation of lot-to-lot and instrument-to-instrument confirmed that the system is robust and the precision is high (Table 6). Figure 10 Comparison of the HemoCue Glucose 201 DM RT System with Cobas Modular (Roche). Analysis between glucose measured with the HemoCue Glucose 201 DM RT System, individual replicates, and glucose measured with the Cobas Modular, mean values. No. of replicates = 100. Table 6 Evaluation of lot to lot variation and instrument to instrument variation regarding the precision. Pooled SD was calculated as well as 95% confidence level (CI) for these values. Type of comparison Pooled SD 95% Cl for SD Cuvette lot a Cuvette lot b Instrument Instrument GPM

21 Comparison of the HemoCue Glucose 201 RT System with Architect ci8200 (Abbot) Ref: Poster by Eva Olofsson 1, Lena Piscator 2 Charlotta Roos 2 1) Avdelningen för klinisk kemi/kliniklärarenheten, Falu lasarett, Falun, Sverige 2) HemoCue AB Ängelholm, Sweden The study compares the HemoCue Glucose 201 RT System to Architect ci8200 (Abbot) which uses a hexokinase method. The populations studied was diverse, some objects had pronounced hypo- and hyperglycemia as noted from the glucose values. There is a very good agreement between the HemoCue Glucose 201 RT and Architect ci8200 in a broad range of glycaemia. This indicates that this method can be used interchangeably for clinical purpose within the healthcare settings. (Figure 12) Figure 12 Scatter plot between glucose measured with the HemoCue Glucose 201 RT System and glucose measured with the Architect ci8200, y = 0.921x , R2 = , R = , N=100. GPM

22 Comparison of the HemoCue Glucose 201 DM RT System with Vitros 5.1 (Ortho-Clinical Diagnostics) Ref: Poster by Ingvor Kasselstrand 1, Eva Drott Heideman 2 Elin Johansson 2 1) Klinisk kemi, Västerviks sjukhus, Sverige. 2) HemoCue AB Ängelholm, Sweden The study compared the HemoCue Glucose 201 DM RT System Vitros 5.1 (Ortho-Clinical Diagnostics) which uses a glucose oxidase method. Samples were analyzed in duplicates on both systems. There is a very good agreement between the HemoCue Glucose 201 DM RT and Vitros 5.1 when measuring glucose over a wide measuring range. (Figure 13) By providing a direct transfer of the results to the hospitals laboratory information system or medical record, the HemoCue Glucose DM RT reduces risk of pre- and post-analytical errors and improves the patient treatment. Figure 13 Analysis between glucose measured with the HemoCue Glucose 201 DM RT System, individual replicates, and glucose measured with Vitros 5.1, mean values. No. of replicates = 114. GPM

23 Comparison of the HemoCue Glucose 201 RT System with a hexokinase method. Ref: Dr ir. R.J. Slingerland (PhD) at Isala klinieken, Zwolle, Netherlands. The evaluation of the HemoCue Glucose 201 RT plasma equivalent analyzers was performed to compare the HemoCue Glucose 201 RT to a hexokinase method, plasma (Figure 14). There is a very good agreement between the HemoCue Glucose 201 RT and the hexokinase method in a broad range of glycaemia. The evaluation was also made to confirm the hematocrit range (Figure 15) as well as the lack of dependency of maltose (Figure 16). Figure 14 Method comparison in capillary whole blood compared with calculated plasma reference values. 55 capillary samples were measured. Figure 15 Hematocrit dependence in capillary whole blood. GPM

24 Figure 16 Maltose dependence in capillary whole blood GPM

25 Validation of a new generation POCT glucose device with emphasis on aspects important for glycemic control in the hospital care Kos, Snježana 1, van Meerkerk, Arie 1, van der Linden, Joke 2, Stiphout, Theo 1, Wulkan, Remi 1. 1 Central Laboratory of the Hospital (Maasstad Lab), Department of Clinical Chemistry, Maasstad Hospital, Rotterdam, The Netherlands 2 Department of Internal Medicine, Maasstad Hospital, Rotterdam, The Netherlands 2012 Ref; Kos, S et al: Validation of a new generation POCT glucose device with emphasis on aspects important for glycemic control in the hospital care, Clin Chem Lab Med 2012; 50 (9): This article in full is available for download from our website In this study a new generation of POCT glucose device, the HemoCue Glucose 201 DM RT has been evaluated for its suitability for (tight) glycemic control. This study was performed according to the CLSI/STARD criteria. The HemoCue Glucose 201 DM RT was compare to the laboratory hexokinase glucose method, Siemens Dimension Vista (Fig 17). The variation among different POC devices and microcuvette lot numbers was examined. Additionally, the influence of the partial pressure of oxygen and hematocrit on glucose measurement was investigated. Figure 17 Comparison experiments (A) Comparison of the HemoCue Glucose 201 DM RT device with the laboratory reference method. (B) Comparison of the HemoCue Glucose 201 DM RT device with the HemoCue Glucose 201 DM analyzer. (C) Device to device comparison. (D) Cuvette lot-to-lot comparison. (A) and (B) Presented by means of Bland Altman plot, while (C) and (D) are presented by means of Bland-Altman absolute bias plot due to a very small deviation expressed as percentage of the mean. The dashed lines in (C) and (D) indicate 5% deviation from the mean concentration over the whole tested concentration range. GPM

26 Results The HemoCue Glucose 201 DM RT showed a good agreement with the laboratory reference method. This was examined using Deming regression analysis, percentage Bland-Altman plot and a modified Clarke-error grid. The total analytical error at the clinically critical glucose concentrations of 5.6, 7.0 and 11.1 mmol/l (101, 126 and 200 mg/dl) was 6.4%, 4.3% and 3.0%, respectively. The total error among the different POCT devices and among different microcuvette lot numbers was <6.5%. Glucose measurements on the HemoCue Glucose 201 DM RT were not affected by changes in partial pressure of oxygen; whereas changes in hematocrit had influence on the results (3.4% for every 0.10 L/L change in hematocrit). Conclusion (extraction) The HemoCue Glucose 201 DM RT shows a very good agreement with the reference hexokinase method. The total analytical error at clinically critical glucose concentrations was < 6.5%. The results were not influenced by changes in partial oxygen pressure, although influenced by changes in hematocrit in a predictable fashion. GPM

27 HemoCue is a global leader in a field of diagnostics known as near patient, or pointof-care, testing. In 1982, HemoCue AB, based in Ängelholm, Sweden, introduced the first system making accurate hemoglobin testing possible in near-patient settings. Since then, more than HemoCue Systems have been sold worldwide. The company also sells point-of-care tests for glucose, urine albumin, HbA1c, total and differential white blood cell count. HemoCue subsidiaries, affiliates, franchises, and third-party distributors generate revenue in more than 130 countries worldwide. HemoCue is owned by Radiometer, a Danaher company. Additional company information is available at GPM

28 GPM

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch

More information

HemoCue Hb METHOD AND SAMPLE COLLECTION

HemoCue Hb METHOD AND SAMPLE COLLECTION HemoCue Hb 201 + METHOD AND SAMPLE COLLECTION 1. PURPOSE AND SCOPE The purpose of this document is to describe the procedure for performing a Haemoglobin (Hb) test using the HemoCue Hb 201 + analyser.

More information

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing

More information

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical

More information

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Original: 5/02 Minor Revision: 9/15 Full Review: 1/15 Replaces: 1/15 Responsible Person: Point of Care Laboratory Specialist Approving Committee:

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th

More information

Aina Blood Monitoring System

Aina Blood Monitoring System Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale

More information

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000 v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination

More information

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626 Performance of the EKF Diagnostics, Stanbio β-hydroxybutyrate LiquiColor Assay on the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System. G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. GLUH Glucose REF B24985 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE

More information

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Polymer Technology Systems, Inc. CardioChek PA Comparison Study Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory

More information

5 BEST PRACTICES To Patient Safety. A Practical Guide to Blood Glucose Testing

5 BEST PRACTICES To Patient Safety. A Practical Guide to Blood Glucose Testing 5 BEST PRACTICES To Patient Safety A Practical Guide to Blood Glucose Testing In a seven-year hospital study, most errors (91% 97%) were operator-related rather than instrument-related. Nobels F, et al.,

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

bedside glucose testing systems

bedside glucose testing systems 42 CAP TODAY JANUARY 2017 Part 1 of 4 Abbott Diabetes Care Abbott Diabetes Care Arkray 1420 Harbor Bay Parkway 1420 Harbor Bay Parkway 5198 W. 76th St. Alameda, CA 94502 Alameda, CA 94502 Edina, MN 55439

More information

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose

More information

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D. Journal of Diabetes Science and Technology Volume 5, Issue 4, July 2011 Diabetes Technology Society TECHNOLOGY REPORTS Accuracy of the CONTOUR Blood Glucose Monitoring System Bern, B.A., Cheryl Leazenby,

More information

The performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.

The performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests. Chemical Composition Each Microdot strip contains the enzyme glucose dehydrogenase (Bacillus Sp.) = 1 IU and other ingredients (mediator, NAD, lysing agents etc.) = 200 µg. Test Strips When blood is added

More information

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff

More information

Gentian Canine CRP Immunoassay Application Note for scil VitroVet*

Gentian Canine CRP Immunoassay Application Note for scil VitroVet* v02-jan2014 Gentian Canine CRP Immunoassay Application Note for scil VitroVet* Intended Use The canine CRP immunoassay on scil VitroVet is an in vitro diagnostic test for quantitative determination of

More information

Summary of Significant Changes at this Revision

Summary of Significant Changes at this Revision COPY Summary of Significant Changes at this Revision CR5890 Removal of Ketone guidelines - refer to Specialist Diabetes nurse (adult wards); Children's ward has their own protocol Remove all references

More information

Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised:

Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised: Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised: PURPOSE The Abbott FreeStyle Precision Pro (FSPP) Blood Glucose

More information

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%) cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,

More information

WELCOME! HemoCue Presentation CPOCT San Diego, September 20 th Annika Eriksson Global Application Manager HemoCue AB

WELCOME! HemoCue Presentation CPOCT San Diego, September 20 th Annika Eriksson Global Application Manager HemoCue AB WELCOME! HemoCue Presentation CPOCT San Diego, September 20 th 2014 Annika Eriksson Global Application Manager HemoCue AB HemoCue Part of the Danaher Family 2 Pioneers in Point of care Testing When HemoCue

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess

More information

Summary of Significant Changes at this Revision. Items Required. Printed copies are uncontrolled unless there is an allocated Copy Number on page 1

Summary of Significant Changes at this Revision. Items Required. Printed copies are uncontrolled unless there is an allocated Copy Number on page 1 COPY Summary of Significant Changes at this Revision CR 5279 Update section 10 as per re-wording by NP Purpose and Scope 1. The Abbott Precision Xceed Pro/FreeStyle Precision Pro meter is a battery-powered

More information

Title: Abbott Optium Xceed Glucose Meter Effective date: 16/05/2011. Summary of Significant Changes at this Revision. SOP updated.

Title: Abbott Optium Xceed Glucose Meter Effective date: 16/05/2011. Summary of Significant Changes at this Revision. SOP updated. COPY Summary of Significant Changes at this Revision SOP updated Purpose and Scope Items Required 1. The Abbott Optium Xceed meter is a battery-powered device designed for the measurement of blood glucose

More information

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. 2013/07/30 A93A01315AUS A11A01641 60 ml 15 ml Pentra C400 Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. QUAL-QA-TEMP-0846

More information

Fraser Health Authority Accu-chek Inform II: Training Script

Fraser Health Authority Accu-chek Inform II: Training Script 1. Introduction Ensure that each participant Signs the attendance Sheet (please PRINT first & last name, and middle initial legibly) Refer to; For Reference Only: o Quick Reference Guide o What s new document

More information

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100) Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November

More information

Accu-Chek Inform II: Point of Care Glucose Testing. Sharp Healthcare 2014

Accu-Chek Inform II: Point of Care Glucose Testing. Sharp Healthcare 2014 Accu-Chek Inform II: Point of Care Glucose Testing Sharp Healthcare 2014 OBJECTIVES At the completion of this module the participant will be able to: Learn the proper technique of performing a finger stick

More information

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved: PROCEDURE TITLE: Bedside Glucose Monitoring Issuing Department: Clinical Director Signature: Departments Involved: Laboratory Nursing Effective Date: 10/97 Review Dates: 09/01, 07/02, 05/13 Revision Dates:

More information

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,

More information

Current Status: Active PolicyStat ID: Accu-Chek Whole Blood Glucose Testing Policy PURPOSE

Current Status: Active PolicyStat ID: Accu-Chek Whole Blood Glucose Testing Policy PURPOSE Current Status: Active PolicyStat ID: 4749849 Origination: 12/2013 Last Approved: 03/2018 Last Revised: 03/2018 Next Review: 03/2020 Owner: Area: References: Karla Gonzalez Anchondo: Point of Care Coordinator

More information

Charging the AccuChek Inform II on the Base Unit. Highlights of New AccuChek Inform II Whole Blood Glucose System

Charging the AccuChek Inform II on the Base Unit. Highlights of New AccuChek Inform II Whole Blood Glucose System Accu-Chek Inform II: Training Guide Introduction System Components Charging the AccuChek Inform II on the Base Unit AccuChek Inform II Meter tote Highlights of New AccuChek Inform II Whole Blood Glucose

More information

Summary of Significant Changes at this Revision. Items Required

Summary of Significant Changes at this Revision. Items Required Update Approver/checker Summary of Significant Changes at this Revision Purpose and Scope Items Required 1. The Abbott FreeStyle Optium meter is a battery-powered device designed for the measurement of

More information

Applicable To Employees of Gundersen Boscobel Area Hospital laboratory and Gundersen Palmer Lutheran Hospital and Clinic Laboratories.

Applicable To Employees of Gundersen Boscobel Area Hospital laboratory and Gundersen Palmer Lutheran Hospital and Clinic Laboratories. Subject Sodium Potassium Chloride ISE - C311 Index Number Lab-8828 Section Laboratory Subsection Regional/Affiliates Category Departmental Contact Betty Tilleraas Last Revised 4/12/2017 References Required

More information

Applicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory.

Applicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory. Subject Creatinine COBAS C311 Index Number Lab-8814 Section Laboratory Subsection Regional/Affiliates Location Category Departmental Contact Tilleraas, Betty Last Revised 1/17/2018 References Required

More information

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version. Department Of Pathology POC.514.06- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 6 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.05 LBH 1 OF 7 Printed copies are

More information

References Required document for Laboratory Accreditation by the College of American Pathologists.

References Required document for Laboratory Accreditation by the College of American Pathologists. Subject Hemoglobin A1c Cobas c501 Index Number Lab-4004 Section Laboratory Subsection Chemistry Category Departmental Contact Benjamin Michel Last Revised 9/12/2016 References Required document for Laboratory

More information

Multi User Model. Product Overview

Multi User Model. Product Overview Multi User Model Product Overview Features and Benefits NO CODING EASY TO USE o o o Large LCD display. Large buttons with simple functions Easy Setup. Pre-Setup before shipping. Easy strip insertion. SMALL

More information

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX AIR Blood Glucose Monitoring System, from Trividia

More information

Management of Central Venous Access Devices. Blood Glucose Monitoring

Management of Central Venous Access Devices. Blood Glucose Monitoring Management of Central Venous Access Devices Blood Glucose Monitoring Purpose To provide education on the standard of care regarding the use and monitoring of the Accu- Chek Blood glucose machine, including

More information

ASSAYED URINE CONTROL - LEVEL 2 (URN ASY CONTROL 2)

ASSAYED URINE CONTROL - LEVEL 2 (URN ASY CONTROL 2) ASSAYED URINE CONTROL - LEVEL 2 (URN ASY CONTROL 2) CAT. NO. AU 2352 LOT NO. 718UC SIZE: 12 x 10ml EXPIRY: 2017-10 INTENDED USE This product is intended for in vitro diagnostic use in the quality control

More information

Alere Technologies GmbH

Alere Technologies GmbH Alere Technologies GmbH Mobile Smart Instruments for Automated Image Analysis and Point of Care Diagnosis April 2011 Alere Technologies GmbH Alere Technologies GmbH Based in Jena, Thuringia, Germany ~300

More information

Blood Glucose Measurement. Omnitest 5 & Omnican Lance FAQ Frequently Asked Questions

Blood Glucose Measurement. Omnitest 5 & Omnican Lance FAQ Frequently Asked Questions Blood Glucose Measurement Omnitest 5 & Omnican Lance FAQ Frequently Asked Questions FAQ Omnitest 5 & Omnican Lance CONTENT Omnitest 5 Overview 3 Application Range 5 Target Groups 5 Test Strips 5 Settings

More information

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014 Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The

More information

Method Comparison Report Semi-Annual 1/5/2018

Method Comparison Report Semi-Annual 1/5/2018 Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community

More information

Smart Diagnostics for Point-of-Care!

Smart Diagnostics for Point-of-Care! Smart Diagnostics for Point-of-Care! The one special multi-parameter Point-of-Care (POC) instrument that covers the constantly expanding range of applications with the highest precision. Designed for HbA1c

More information

Ellis Jacobs, Ph.D., DABCC. New York University School of Medicine Bellevue Hospital Center New York, New York

Ellis Jacobs, Ph.D., DABCC. New York University School of Medicine Bellevue Hospital Center New York, New York Ellis Jacobs, Ph.D., DABCC New York University School of Medicine Bellevue Hospital Center New York, New York ADA Recommendation In terms of glucose in venous plasma WHO Recommendation In terms of glucose

More information

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date: BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET Revision date: 2017-02-24 BÜHLMANN LABORATORIES AG Baselstrasse 55 4124 Schönenbuch, Switzerland Tel.:

More information

The analytical phase

The analytical phase The analytical phase Result interpretation Test request Result Sampling Black box: the lab ANALYTICAL PHASE The CASE Uncle Pete, 67 years old Marked abdominal pain 8 pm, ED Acute abdomen? Assessment (+

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia

More information

GlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System

GlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System GlucCell TM SYSTEM USER S GUIDE Ver 2.1 Intended Use The GlucCell TM Glucose Monitoring System (The GlucCell TM System) is designed to quantitatively measure the concentration of glucose during cell culture.

More information

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide LESSON 6-3 Laboratory Reagent Preparation and Calculations Worksheet LESSON 6-4 Chemistry Instrumentation in the Physician Office Laboratory Worksheet LESSON 6-6 Measuring Blood Glucose Worksheet LESSON

More information

3. System Accuracy Evaluation LUX Hgb test 1. Samples

3. System Accuracy Evaluation LUX Hgb test 1. Samples 3. System Accuracy Evaluation LUX Hgb test 1. Samples - Type of sample: Capillary Blood System accuracy evaluation was performed with 200 fresh capillary blood samples, each with sufficient volume to be

More information

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009 Basic knowledge of POCT instuments The Tide Resort Bangsan Beach 18 June 2009 1 POCT instruments concepts Point of care testing Diagnostic medical devices used at or near the site of patient care Bed site,

More information

ASSAYED URINE CONTROL - LEVEL 3 (URN ASY CONTROL 3)

ASSAYED URINE CONTROL - LEVEL 3 (URN ASY CONTROL 3) SSAYED URINE CONTROL - LEVEL 3 (URN ASY CONTROL 3) CAT. NO. AU 2353 LOT NO. 797UC SIZE: 12 x 10 ml EXPIRY: 2019-03 INTENDED USE This product is intended for in vitro diagnostic use, in the quality control

More information

Creatinine Test Code: CRE

Creatinine Test Code: CRE Creatinine Test Code: CRE Intended Use For in vitro diagnostic use only CARESIDE cartridges are used with the CARESIDE Analyzer to quantitatively measure creatinine concentration in anti-coagulated whole

More information

Human Hemoglobin Colorimetric Detection Kit

Human Hemoglobin Colorimetric Detection Kit Human Hemoglobin Colorimetric Detection Kit CATALOG NO: IRAAKT2522 LOT NO: SAMPLE INTENDED USE The Hemoglobin detection kit is designed to quantitatively measure all forms of hemoglobin present in blood

More information

THE VALLEY HOSPITAL Ridgewood, New Jersey PATIENT CARE SERVICES (PCS) POLICIES AND PROCEDURES

THE VALLEY HOSPITAL Ridgewood, New Jersey PATIENT CARE SERVICES (PCS) POLICIES AND PROCEDURES 58:03 THE VALLEY HOSPITAL Ridgewood, New Jersey PATIENT CARE SERVICES (PCS) POLICIES AND PROCEDURES SUBJECT: Whole Blood Glucose PURPOSE: The Accu-chek Inform II is intended for use to quantitatively measure

More information

Traceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users. David Ducroq.

Traceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users. David Ducroq. Traceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users David Ducroq Weqas Unit 6, Parc Tŷ Glas Llanishen Cardiff UK www.weqas.com Programme

More information

Evaluation of VACUETTE Urine CCM tube for Clinical Chemistry

Evaluation of VACUETTE Urine CCM tube for Clinical Chemistry Evaluation of VACUETTE Urine CCM tube for Clinical Chemistry Background The VACUETTE Urine CCM tube is for the collection, transport and storage of urine samples for urine culture and urinalysis in the

More information

Dear Prodigy System Owner:

Dear Prodigy System Owner: User's Manual Dear Prodigy System Owner: You have chosen one of the best blood glucose monitoring systems available. This booklet has important information about the system. Please read it carefully. Your

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: BD FACSCount Instrument System with FACSCount Control Kit and BD FACSCount Reagent Kit (Absolute CD4+, CD8+, and CD3+ Counts) Number:

More information

GlucCell TM SYSTEM USER S GUIDE ver 2.3 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System

GlucCell TM SYSTEM USER S GUIDE ver 2.3 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System GlucCell TM SYSTEM USER S GUIDE ver 2.3 Intended Use The GlucCell TM Cell Culture Glucose Monitoring System (The GlucCell TM System) is designed to quantitatively measure the concentration of glucose during

More information

I know my value CoaguChek XS Pro system

I know my value CoaguChek XS Pro system I know my value CoaguChek XS Pro system The smart way to manage your anticoagulation clinic CoaguChek XS Pro system A barcode scanner can make a difference There has been an extraordinary increase in the

More information

POC Testing: The New Standard. InnovaStar From DiaSys.

POC Testing: The New Standard. InnovaStar From DiaSys. POC Testing: The New Standard. InnovaStar From DiaSys. Easy to use, superior precision, rapid results: InnovaStar from DiaSys brings POC testing to a whole new level. A fully automatic, compact analyzer

More information

Aspartate Aminotransferase Test Code: AST

Aspartate Aminotransferase Test Code: AST Aspartate Aminotransferase Test Code: AST Intended Use For in vitro diagnostic use only CARESIDE Aspartate Aminotransferase (AST) cartridges are used with the CARESIDE Analyzer to quantitatively measure

More information

CLINITEK Novus Automated Urine Chemistry Analyzer

CLINITEK Novus Automated Urine Chemistry Analyzer CLINITEK, Atlas, Novus, Status+, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks are properties of their respective owners. Product

More information

Why do we need SD LipidoCare?

Why do we need SD LipidoCare? SD Biosensor, Inc. Innovative Global Leader of In Vitro Diagnostics Total supplier of Blood Glucose Monitoring System and Lipid Analyzer System in vitro diagnostics test kits for Point of Care Testing

More information

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay TECHNICAL NOTE GSP Neonatal Phenylalanine kit 3308-0010/3308-001B GSP Neonatal Phenylalanine kit (3308-0010/3308-001B) is a fully Introduction automated enzymatic assay for the GSP system. Optimized for

More information

Individual Quality Control Plans (IQCP) Risk Assessment

Individual Quality Control Plans (IQCP) Risk Assessment Individual Quality Control Plans (IQCP) Risk Assessment James H. Nichols, PhD, DABCC, FACB Professor of Clinical Pathology, Microbiology and Immunology Medical Director of Clinical Chemistry and Point-of-Care

More information

Relationship between glucose meter error and glycemic control efficacy

Relationship between glucose meter error and glycemic control efficacy Relationship between glucose meter error and glycemic control efficacy Brad S. Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Department of Laboratory Medicine and Pathology Mayo Clinic

More information

METHOD VALIDATION: WHY, HOW AND WHEN?

METHOD VALIDATION: WHY, HOW AND WHEN? METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test

More information

SCIEX Vitamin D 200M Assay for the Topaz System

SCIEX Vitamin D 200M Assay for the Topaz System The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human

More information

Operator s Manual. Blood Glucose and β-ketone Monitoring System. Abbott Diabetes Care Inc South Loop Road Alameda, CA USA

Operator s Manual. Blood Glucose and β-ketone Monitoring System. Abbott Diabetes Care Inc South Loop Road Alameda, CA USA Blood Glucose and β-ketone Monitoring System Operator s Manual Precision Xceed Pro, TrueID, TrueMeasure, TrueAccess, PrecisionWeb, Precision, MediSense, Optium and Precision PCx are trademarks of the Abbott

More information

Bio-Rad Laboratories. Cardiac Assessment Controls. Value Assigned for Clinical Laboratory and Point of Care Test Systems

Bio-Rad Laboratories. Cardiac Assessment Controls. Value Assigned for Clinical Laboratory and Point of Care Test Systems Bio-Rad Laboratories Cardiac Assessment Controls Cardiac Assessment Controls Value Assigned for Clinical Laboratory and Point of Care Test Systems Solving the Challenges of Troponin I Testing for the Central

More information

State of the Art of HbA1c Measurement

State of the Art of HbA1c Measurement State of the Art of HbA1c Measurement XXI Congreso Latinoamericano de Patologia Clinica Y XLII Congreso Mexicano de Patologia Clinica, Cancun October 2012 Randie R. Little, Ph.D. NGSP Network Laboratory

More information

Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter

Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter The Nova Stat Strip Blood Glucose meter is UHB s approved Glucometer

More information

College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey

College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) The American Diabetes Association (ADA) recommends that laboratories use only HbA1c assay methods that have been NGSP certified and

More information

25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL

25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL 25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL I. PRINCIPLE The role of vitamin D in bone and mineral metabolism was recognized from its first identification as a factor that could cure rickets. However,

More information

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803) Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial

More information

Point of Care Capillary blood glucose/ketone monitoring with the NOVA meter. Clinical Guideline Register No: Status: Public

Point of Care Capillary blood glucose/ketone monitoring with the NOVA meter. Clinical Guideline Register No: Status: Public Point of Care Capillary blood glucose/ketone monitoring with the NOVA meter Clinical Guideline Register No: 08027 Status: Public Developed in response to: Best Practice Guidelines for clinicians Contributes

More information

Toxicology Blood Lead Comprehensive Testing

Toxicology Blood Lead Comprehensive Testing Clinical Laboratory s of Practice Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where

More information

Serodos and Serodos plus

Serodos and Serodos plus Design Verification Serodos and Serodos plus Contents 1 Value Adjustment... 2 2 Target Determination... 2 3 Stability... 2 Real-Time Stability... 3 Stability after Reconstitution... 4 Stability after Reconstitution

More information

Cardiac Assessment Controls A critical element of reliable cardiac testing

Cardiac Assessment Controls A critical element of reliable cardiac testing Cardiac Assessment Controls Cardiac Assessment Controls A critical element of reliable cardiac testing Choose the products best suited for your cardiac assessment needs Each year, 17.7 million people on

More information

Mission Hb accurately detects both

Mission Hb accurately detects both Mission Digital Hb-Testing System: Save Cost by testing 2- Parameters with Single Strip real time ie: i.e: Hemoglobin & Hematocrit-[PCV]. Presented By : Remedy Healthcare Private Limited, New Delhi, [INDIA].

More information

Precision ceedpro. Online Training for. Glucometer & Capillary Collection

Precision ceedpro. Online Training for. Glucometer & Capillary Collection Online Training for Precision ceedpro Glucometer & Capillary Collection How long will it take? This module takes approximately 30 minutes to complete. What if I have to leave before I finish? Progress

More information

Toxicology Blood Lead Comprehensive Testing

Toxicology Blood Lead Comprehensive Testing Clinical Laboratory s of Practice Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where

More information

TITLE: Whole Blood Glucose Monitoring Using the Nova StatStrip Meter

TITLE: Whole Blood Glucose Monitoring Using the Nova StatStrip Meter Page 4 of 14 IV. Performance of Quality Control A. Purpose: To detect any errors due to system failures or operator performance B. Frequency: Levels 1 and 3 controls are run every 24 hours of use. C. Quality

More information

Omnitest : 2013

Omnitest : 2013 ISO 15197: 2013 Omnitest 3 Blood Glucose Monitoring System Omnitest 3 Could it be easier? After strip insertion, Omnitest 3 is ready for the measurement. The code is set automatically. Only 0.3 µl sample

More information

Biocard. 1. Troponin I

Biocard.  1. Troponin I Biocard http://www.anibiotech.fi Biocard Quant Reader is a quantitative desktop reader for all Biocard rapid cardiac tests: Troponin I, Myoglobin, CRP, hscrp, MPO and CK-MB. It has a embedded printer and

More information

Glucose, glucose oxidase/peroxidase method, photometry, diabetes mellitus.

Glucose, glucose oxidase/peroxidase method, photometry, diabetes mellitus. Quantitative Determination TEAS Related topics Glucose, glucose oxidase/peroxidase method, photometry, diabetes mellitus. Principle Blood sugar is the key energy provider for our cells. The primary hormone

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Lactate REF A95550 For In Vitro Diagnostic Use ANNUAL REVIEW Refer to Review and Revision Coversheet at the front of each method. PRINCIPLE INTENDED USE reagent,

More information