1 CME JOURNAL OF MAGNETIC RESONANCE IMAGING 000: (2012) Original Research Quantification of Left Ventricular Indices From SSFP Cine Imaging: Impact of Real-World Variability in Analysis Methodology and Utility of Geometric Modeling Christopher A. Miller, MBChB, 1 3 * Peter Jordan, PhD, 1 Alex Borg, MD, 1 Rachel Argyle, MBChB, 1 David Clark, BSc, 4 Keith Pearce, BSc, 1 and Matthias Schmitt, PhD 1,3 Purpose: To assess the impact of real-world practice variation in the process of quantifying left ventricular (LV) mass, volume indices, and ejection fraction (EF) from steady-state free precession cardiovascular magnetic resonance (CMR) images. The utility of LV geometric modeling techniques was also assessed. Materials and Methods: The effect of short-axis- versus long-axis-derived LV base identification, simplified versus detailed endocardial contouring, and visual versus automated identification of end-systole were evaluated using CMR images from 50 consecutive, prospectively recruited patients. Additionally, the performance of six geometric models was assessed. Repeated measurements were performed on 25 scans (50%) in order to assess observer variability. Results: Simplified endocardial contouring significantly overestimated volumes and underestimated EF ( 6 6 4%, P < ) compared to detailed contouring. A mean difference of 34g (P < ) was observed between mass measurements made using short-axis- versus long-axisderived LV base positioning. A technique involving longaxis LV base identification, signal threshold-based detailed endocardial contouring, and automated identification of end-systole had significantly higher observer agreement. Geometric models showed poor agreement with conventional analysis and high variability. 1 Division of Cardiology and Cardiothoracic Surgery, University Hospital of South Manchester, Manchester, UK. 2 Cardiovascular Research Group, University of Manchester, UK. 3 Biomedical Imaging Institute, University of Manchester, UK. 4 Alliance Medical, Wythenshawe CMR Unit, Manchester, UK. Contract grant sponsor: National Institute for Health Research, UK; Contract grant number: Doctoral Research Fellowship NIHR-DRF (to C.A.M.); Contract grant sponsor: Greater Manchester Comprehensive Local Research Network funding (to M.S.). *Address reprint requests to: C.A.M., Division of Cardiology and Cardiothoracic Surgery, University Hospital of South Manchester, Southmoor Road, Wythenshawe, Manchester M23 9LT, U.K. chrismiller@doctors.org.uk Received July 8, 2011; Accepted September 17, DOI /jmri View this article online at wileyonlinelibrary.com. Conclusion: Real-world variability in CMR image analysis leads to significant differences in LV mass, volume and EF measurements, and observer variability. Appropriate reference ranges should be applied. Use of geometric models should be discouraged. Key Words: cardiovascular magnetic resonance imaging; left ventricle; analysis methodology; geometric modeling J. Magn. Reson. Imaging 2012;000: VC 2012 Wiley Periodicals, Inc. LEFT VENTRICULAR (LV) mass, volume indices, and ejection fraction (EF) remain some of the most evidenced-based indictors of prognosis (1 5). Both their absolute measurement and temporal change are used to guide pharmacotherapy, device implantation, and surgical intervention. Accurate and reproducible assessment of these parameters is an important strength of cardiovascular magnetic resonance imaging (CMR). The main components in the process of quantifying LV indices from CMR images are: LV base identification, myocardial border tracing, and identification of end-diastole and systole. While CMR is regarded as the gold standard technique for assessment of LV indices, in real-world CMR practice considerable variability exists in the methodology used for each of these components. First, rather than incorporating long-axis images into the analysis, many centers use only short-axis images for analysis, despite it often being difficult to determine the basal LV slice, and how much of it to include, from short-axis images alone. Arbitrary criteria are widely used, such as inclusion of a slice within the LV when at least 50% or 75% of the cavity is surrounded by ventricular myocardium (Fig. 1) (6 8). Alternatively, if a slice is thought to include both ventricular and atrial myocardium, others advocate tracing up to the apparent junction of atrium and ventricle before joining up the contours with a straight VC 2012 Wiley Periodicals, Inc. 1

2 2 Miller et al. The aim of this study was to assess the impact of these real-world methodological differences in CMR image analysis on LV mass, volume, and EF measurements in an unselected cohort of patients representative of clinical practice. Specifically, the effect of short-axis- versus long-axis-derived LV base identification, simplified versus detailed endocardial contouring, and manual versus automated identification of end-systole were evaluated. The effect on inter- and intraobserver variability was also assessed. In addition, the utility of LV geometric modeling techniques, advocated for analyzing CMR images and widely used by other imaging modalities (8,12 16), was assessed. MATERIALS AND METHODS Patients Fifty consecutive consenting patients undergoing clinically indicated CMR scanning at a single institution were enrolled. Patients were recruited prospectively, prior to undergoing imaging. The only exclusion criterion was known complex congenital heart disease. In order to reflect real-world clinical practice, patients were not excluded on the basis of image quality, arrhythmia, or breath-holding ability. The study was conducted according to the Helsinki Declaration. An Ethics Committee of the UK National Research Ethics Service gave ethical approval for the study (reference number 08/H1004/153) and written informed consent was obtained from all participants before entering the study. Figure 1. Methodological differences in LV base identification. Using the short-axis-derived method, basal slices were considered to be within the LV at end-diastole (a) and endsystole (b) if 50% or more of the cavity was surrounded by ventricular myocardium. If the basal slice contained both ventricular and atrial myocardium, contours were drawn up to the junction and joined by a curved line through the blood pool. With the long-axis-derived method, the mitral (and aortic) valve position was identified at end-diastole (c) and systole (d) in all three long-axis images (only horizontal longaxis image shown for illustration), which allowed the valve plane, and hence the LV base, to be tracked through the cardiac cycle. This was then integrated into the short-axis stack analysis (e, end-diastole; f, end-systole) so that short-axis data basal to the mitral valve plane were excluded from mass and volume calculations. or curved line through the blood pool, creating a semicircle or crescent-shaped cavity (9). Second, it is standard practice across many centers to only trace the compacted endocardial border rather than performing detailed tracing of papillary muscles and trabeculae (Fig. 2) (10,11). As a result, papillary muscles and trabeculae are included in cavity volumes rather than within mass measurements. Third, end-diastole and end-systole are usually identified visually by manually selecting the frame with the largest and smallest cavities, respectively (6). However, often this is not uniform across short-axis slices, particularly in dyssynchronous ventricles. Image Acquisition All patients underwent CMR imaging using a 1.5 T scanner (Avanto; Siemens Medical Imaging, Erlangen, Germany) with a 32-element phased-array coil. Steady-state free precession (SSFP) end-expiratory breath-hold cines were acquired in three long-axis planes (horizontal long-axis, vertical long-axis, and three-chamber long-axis). Using the horizontal and vertical long axis images to ensure perpendicularity to the ventricular septum, short-axis cines were then acquired from the atrioventricular ring to the apex (SA stack). Typical parameters included repetition time 2.9 msec, echo time 1.2 msec, flip angle 80, matrix , in-plane pixel size mm, slice thickness 8 mm (interslice gap 2 mm for the SA stack), temporal resolution msec depending on heart rate (25 acquired phases). The number of shortaxis cines acquired was patient-specific, depending on ventricular size (mean 11). Retrospective gating was used in 48 patients. Prospective gating was used in two patients due to arrhythmia. Image Analysis LV mass and volume indices were quantified for each scan using the three methods described below. Analysis was performed by a Level 2 accredited operator with over 3 years experience of CMR and of each analysis method.

3 CMR Methods for LV Assessment 3 Figure 2. Methodological differences in endocardial contouring. a: Simplified contouring; the compacted endocardial border was traced resulting in papillary muscles and trabeculae being included in volumetric and excluded from mass measurements. b: Detailed manual contouring; detailed tracing of the endocardium was performed manually such that papillary muscles and trabeculae were included in mass and excluded from volumetric measurements. c: Threshold-based detailed contouring; detailed tracing of the endocardium was performed using a signal intensity-based thresholding tool, which delineated individual voxels as blood or myocardium. The thresholding level was manually adjusted until endocardial appearances correlated with visual assessment. Method 1 The epicardial border was manually traced at end-diastole in successive short-axis slices (Argus Syngo MR software, Siemens Medical Imaging). The compacted endocardial border was manually traced on the shortaxis slices at end-diastole and end-systole such that papillary muscles and trabeculae were included in volumetric and excluded from mass measurements (Fig. 2A). At the base of the heart, slices were considered to be within the LV if 50% or more of the cavity was surrounded by ventricular myocardium. If the basal slice contained both ventricular and atrial myocardium, contours were drawn up to the junction and joined by a curved line through the blood pool (Fig. 1A,B), drawn appropriate to the part of the slice that appeared ventricular after observing the throughplane motion of the LV on the short-axis cine images. The end-diastolic and end-systolic frames were identified as the frames in which the LV cavity visually appeared largest and smallest, respectively. In cases where there was a discrepancy between slices, the frames in which the more basal slices had the largest cavity (for end-diastole) and smallest cavity (for endsystole) were used, as it was felt these would more closely represent the largest and smallest cavity volumes, respectively. Method 2 Analysis was identical to Method 1 except that manual detailed tracing of the endocardium was performed such that papillary muscles and trabeculae were included in mass and excluded from volumetric measurements (Fig. 2B). Method 3 The epicardial border was manually traced at end-diastole in successive short-axis slices (CMRtools, Cardiovascular Imaging Solutions, UK). A second contour was placed within the myocardium on the short-axis slices at end-diastole and systole allowing the signal intensity between the two contours (ie, the signal intensity of myocardium) to be automatically determined. The endocardial border was then determined using a signal intensity-based thresholding tool, which delineated individual voxels as blood or myocardium and as such papillary muscles and trabeculae were included in mass and excluded from volumetric measurements (Fig. 2C). The thresholding was adjusted manually until appearances correlated with visual assessment. The LV base was identified using the long-axis images; the mitral valve position was identified visually at end-diastole and systole on the long-axis images, which allowed the valve plane to be tracked through the cardiac cycle. This was then integrated into the SA stack analysis. As a result, shortaxis data basal to the mitral valve plane were excluded from mass and volume calculations (if the valve plane cut through a short axis slice, the part of the slice that was basal to the valve plane was excluded; Fig. 1C F). Finally, end-systole was automatically determined as being the frame with the smallest cavity volume by calculating cavity volume at each frame. For all methods, end-diastolic (EDV) and end-systolic (ESV) volumes were calculated as the sum of the endocardial areas in each short-axis slice at end-diastole and end-systole, respectively, multiplied by the interslice distance (ie, slice thickness plus interslice gap). Stroke volume (SV) and EF were derived from EDV and ESV. End-diastolic myocardial mass was determined as the sum of the areas between the endocardial and epicardial borders in each short-axis slice at end-diastole, multiplied by the interslice distance to give the myocardial volume, which was then multiplied by 1.05 g/cm 3, the specific density of myocardium, to give myocardial mass. Short axis slices were numbered with slice 1 being the most basal, and the most basal slice included in the analysis at end-diastole and end-systole was recorded for each technique, no matter how little of the slice was included (the number of short-axis slices included in the analysis at end-diastole and end-systole potentially differed due to the excursion of the mitral valve plane during

4 4 Miller et al. described. To assess intraobserver variability, 25 scans (50%) were reanalyzed by the first observer with a 1-month temporal separation between analyses. Statistics Figure 3. Geometric models and formulae for the determination of LV volume (LVV). All measurements are made at enddiastole and end-systole to give EDV and ESV. A M ¼ short axis area at the level of the base, A P ¼ short axis area at the level of the papillary muscles, A H ¼ long axis area in the horizontal plane, A V ¼ long axis area in the vertical plane, L ¼ length of the LV, D ¼ short axis diameter of LV measured in the long axis plane. [Color figure can be viewed in the online issue, which is available at wileyonlinelibrary.com.] the cardiac cycle). End-systolic frame used and the time taken for analysis was also recorded. Geometric Modeling LV volumetric analysis was performed on each scan using six geometric models (Fig. 3) (12,17 19). Appropriate contours were drawn for each model using opensource OsiriX Imaging Software (Pixmeo, Switzerland, v. 3.8). EDV, ESV, and EF were calculated using an OsiriX software plug-in, with the exception of the triplane model where contours were drawn using OsiriX but calculations were performed manually, as no appropriate plug-in was available. Papillary muscles and trabeculae were included in volumetric measurements. In the absence of a gold standard comparator, results obtained from each model were compared with the method found to have the highest observer agreement of the three described above, as prespecified. LV internal diameter, defined as the distance between the endocardial border of the septum and the inferolateral wall just distal to the mitral valve leaflet tips in the three-chamber long-axis view, was measured at end-diastole and end-systole and compared to EDV and ESV, respectively. Observer Variability To assess interobserver variability, 25 randomly selected scans (50%) were independently reanalyzed by a second experienced observer using the methods Values are expressed as mean 6 standard deviation (SD) unless stated otherwise. Agreement between each method was evaluated using Bland Altman testing by calculating mean difference (bias) and 95% limits of agreement (ie, mean difference 62 SD). The significance of the differences between each method was assessed using Wilcoxon rank testing. Correlation was assessed using Spearman s rank correlation coefficient (r s ). Inter- and intraobserver variability were evaluated using the repeatability coefficient (defined as 1.96 H[sum of the squares of the differences between observer measurements divided by n]) (20). The variability of each method was compared using a Wilcoxon rank comparison of the squared differences between observer measurements (an estimate of the within-subject variance for that method multiplied by two) for each technique (20,21). P < 0.05 was considered significant. Statistical analysis was performed using SPSS Statistics (v. 19, IBM). RESULTS Patient characteristics and scan indications are displayed in Table 1. Patients exhibited a variety of cardiovascular pathology and a wide range of mass, EDV, ESV, and EF (Table 2). Volumetric, EF, and Mass Measurement Compared to the other two techniques, Method 1 significantly overestimated both EDV and ESV, but ESV to a greater degree (Tables 2, 3; Fig. 4B,C,E,F). As a Table 1 Patient Characteristics, Scan Indications, and Main Findings Number (%) n ¼ 50 Age (range 19-84) Male 34 (76%) Arrhythmia 5 (10%) Frequent univentricular ectopics 2 Atrial fibrillation 2 Atrial flutter 1 Scan indication Myocardial perfusion assessment 36 (72%) Viability assessment* 10 (20%) Valvular assessment 2 (4%) Aortic assessment 2 (4%) Main scan diagnosis Ischemic heart disease 20 (40%) Cardiomyopathy 11 (22%) Normal 10 (20%) Valvular heart disease 4 (8%) Pericardial disease 2 (4%) Ascending aortic aneurysm 2 (4%) Intracardiac mass 1 (2%) *Includes assessment for cardiomyopathy, pericardial disease, intracardiac masses, and myocardial viability.

5 CMR Methods for LV Assessment 5 Table 2 Mean 6 SD (range) Volumetric, EF, and Mass Measurements Using Each Analysis Method Method 1 Method 2 Method 3 EDV (ml) (77-324) (69-315) (61-328) ESV (ml) (24-258) (19-245) (23-240) SV (ml) (21-142) (19-154) (13-150) EF (%) (19-72) (20-78) (21-81) Mass (g) (47-158) (52-176) (74-233) EDV, end-diastolic volume; ESV, end-systolic volume; SV, stroke volume; EF, ejection fraction. result, Method 1 significantly underestimated SV and EF (Tables 2, 3; Fig. 4H,I). As a consequence, six patients (12%) classified as having a normal EF by Method 3 (using a nominal normal EF cutoff of 55% (9)) were classified as having a reduced EF by Method 1. Furthermore, six patients (12%) classified as having mild to moderate LV impairment by Method 3 (EF <55% but >35%) were classified as having severe LV impairment (EF 35%) by Method 1. There was no significant difference between Method 2 and Method 3 for EDV, ESV, SV, and EF, but the 95% limits of agreement were wide for each parameter, suggesting these methods were not interchangeable (Tables 2, 3; Fig. 4A,D,G). While Method 1 significantly underestimated mass compared with to Method 2, both underestimated mass compared to Method 3 (Tables 2, 3; Fig. 4J L). Basal Slice Selection, End-Systolic Frame Selection, and Analysis Time Method 3 determined the base of the LV to be more basal than Methods 1 and 2 at both end-diastole and end-systole. At end-diastole, the most basal shortaxis slice included in Method 3 was (where short-axis slice number 1 is the most basal), compared to for Methods 1 and 2; P < Likewise at end-systole, the most basal short-axis slice included in Method 3 was , compared to for Methods 1 and 2; P < The frame used as end-systole was also significantly different between Method 3 (frame number ) and Methods 1 and 2 (frame number ; P ¼ 0.003). However, in the 21 patients where there was a difference in end-systolic frame selection, mean difference in ESV was only ml. Analysis time for Method 1 ( minutes) and Method 3 ( minutes) were both significantly shorter than for Method 2 ( minutes; P < for both). There was no significant difference between analysis time for Method 1 and Method 3. Observer Variability Method 3 had higher interobserver and intraobserver agreement than Methods 1 and 2 for measurement of EF (interobserver; P < compared to Method 1 and P ¼ compared to Method 2; intraobserver; P ¼ 0.04 compared to Method 1 and Method 2) but there was no difference in observer agreement between Methods 1 and 2 for measurement of EF (Fig. 5). Both Method 2 and Method 3 had significantly higher observer agreement than Method 1 for measurement of EDV and ESV. However, there was no Table 3 Comparison of Volumetric, EF, and Mass Measurements Using Each Analysis Method Mean difference 6SD P-value 95% limits of agreement Correlation coefficient (r s ) EDV (ml) Method 3 Method to Method 3 Method < to Method 2 Method < to ESV (ml) Method 3 Method to Method 3 Method < to Method 2 Method < to SV (ml) Method 3 Method to Method 3 Method < to Method 2 Method < to EF (%) Method 3 Method to Method 3 Method < to Method 2 Method < to Mass (g) Method 3 Method < to Method 3 Method < to Method 2 Method < to Mean difference (SD), significance of the difference, Bland-Altman 95% limits of agreement and correlation coefficient for the comparison of each analysis Method. Abbreviations as in Table 2.

6 6 Miller et al. Figure 4. Bland Altman comparisons of analysis methods for measurement of EDV (a c), ESV (d f), EF (g i), and mass (j l). Abbreviations as in Table 2. [Color figure can be viewed in the online issue, which is available at wileyonlinelibrary.com.] difference in observer agreement between the Method 2 and Method 3 for EDV and ESV measurements. Method 3 had higher interobserver agreement than Methods 1 and 2 for measurement of mass (P ¼ 0.02 compared to Method 1 and P < compared to Method 2), although intraobserver agreement was not significantly different. Geometric Modeling Geometric models were compared with Method 3. EDV, ESV, and EF varied substantially between geometric models (Table 4). The biplane model showed the highest agreement with Method 3, with no significant difference in EDV, ESV, and EF, although limits of agreement were wide for each parameter. For the majority of the other models, volume and EF measurements were significantly different from the semiautomated technique, with very wide limits of agreement. (Comparisons of geometric models with Methods 1 and 2 for measurement of EF are included in Table 5.) There was a moderate correlation between LV internal diameter in diastole and EDV (r S 0.65) and a good correlation between LV internal diameter in systole and ESV (r S 0.83). Interobserver and intraobserver agreement for measurement of EF was low for all models and was significantly lower than that of Method 3 (Fig. 6). Interobserver agreement for measurement of EDV using the Teicholz, triplane, modified Simpson s and biplane models was not significantly different from that of Method 3. However, all models had significantly lower interobserver agreement than Method 3 for measurement of ESV. All models also had significantly lower intraobserver agreement for measurement of EDV and ESV than Method 3. DISCUSSION This study assessed the impact of real-world practice variation in the process of quantifying LV indices from

7 CMR Methods for LV Assessment 7 Figure 5. Interobserver and intraobserver variability of each analysis method for measurement of volume indices, EF and mass, expressed as repeatability coefficients. The repeatability coefficient is the range within which measurements by two different observers (interobserver) and two measurements by the same observer (intraobserver) are expected to lie. *A significant difference, assessed using a Wilcoxon rank comparison of the squared differences (P < 0.05). Abbreviations as in Table 2. SSFP CMR images. Principally, different methods of LV base identification and endocardial contouring led to significant differences in LV mass, volume and EF measurements, and observer variability, findings that have direct implications for clinical and academic CMR practice. In addition, despite their widespread clinical use, particularly in the field of echocardiography, this study questions the value of geometric models. An increasing number of software packages are available for clinical assessment of LV volumes and mass from CMR images. All share the same fundamental methodological components, ie, LV base identification, myocardial border tracing, and identification of end-diastole and systole. However, common differences exist between analysis methods for each of these components. In particular, all involve either short-axis- or long-axis-derived LV base identification, simplified or detailed endocardial contouring, and manual versus automated identification of end-systole. The specific analysis software used in the current study were chosen because it is felt that they are representative of all commonly used analysis packages. As such, the findings of this study are relevant to the majority of CMR users. Recruitment was prospective and consecutive, without exclusion on grounds of image quality, arrhythmia, or breath-holding ability, with prospective gating required in 4% of patients. Patients exhibited a wide range of LV mass, volumes, and EF and a variety of pathology. It is therefore felt that the study is representative of real-world CMR practice. Methodological differences in endocardial border tracing had the greatest impact on LV volume and EF measurements. Simplified endocardial contouring, where papillary muscles and trabeculae are included within volumes and excluded from mass, led to a significant overestimation of volumes compared to detailed contouring (manual or threshold-derived). However, in keeping with other studies, ESV was overestimated to a greater degree (EDV overestimated by 5%, ESV by 19%, compared to detailed manual endocardial contouring) (22,23). Consequently, simplified contouring led to a significant underestimation of EF. The degree of EF underestimation in this study (6% difference between simplified and detailed manual contouring) is larger than that found by Weinsaft et al (23) (3%), although the current study includes patients with a wider range of volumes and EF. In a study assessing the impact of simplified versus detailed tracing of trabeculae alone, Papavassiliu et al (22) found the simplified technique underestimated EF by 2% (papillary muscles were included within mass measurements in all subjects). The greater underestimation of ESV relative to EDV by the simplified technique may result from it being more difficult to differentiate the compacted endocardial border from papillary muscles and trabeculae at end-systole. The technique of simplified endocardial contouring is widely practiced (10,11), despite detailed contouring (manual or threshold-based) having been used to establish normal reference ranges for SSFP imaging (6,9,24) and the greater accuracy of detailed contouring in ex vivo validatory studies (25,26). In this study, the use of simplified contouring had potential clinical implications, with nearly a quarter of patients being assigned to a different EF category compared to detailed contouring (ie, abnormal EF rather than normal, or severely impaired EF rather than mild-moderately impaired), thus highlighting the importance of using reference ranges appropriate to the analysis method used. Some software packages allow independent tracing of papillary muscles, which has the advantage of allowing volumetric (and mass) data to be presented both inclusive and exclusive of the papillary muscles. However, independent tracing of trabeculae is often more difficult, and the fundamental methodology of such software, ie, manual border tracing, is the same as that assessed in the current study. One of the reasons why simplified endocardial contouring is commonly used is in order to reduce analysis time; indeed, simplified contouring reduced analysis time by over a third compared to detailed manual contouring in the current study.

8 8 Miller et al. Table 4 Comparison of Volumetric and EF Measurements Using Each Geometric Model With Method 3 Mean difference 6SD P-value 95% limits of agreement Correlation coefficient (r s ) EDV (ml) Method 3 Teich < to Method 3 ModSimp to Method 3 Hemi < to Method 3 Mono to Method 3 Biplane to Method 3 Triplane < to ESV (ml) Method 3 Teich < to Method 3 ModSimp < to Method 3 Hemi < to Method 3 Mono to Method 3 Biplane to Method 3 Triplane < to EF (%) Method 3 Teich to Method 3 ModSimp 969 < to Method 3 Hemi 767 < to Method 3 Mono to Method 3 Biplane to Method 3 Triplane 366 < to Mean difference (SD), significance of the difference, Bland-Altman 95% limits of agreement and correlation coefficient for the comparison of each geometric model with Method 3. Teich Teicholz, ModSimp modified Simpson s, Hemi hemi-ellipse, Mono monoplane. Long-axis-derived LV base identification led to highly significantly more basal positioning of the LV base compared to the short-axis-derived method. This finding demonstrates that the LV cavity frequently extends beyond the slice determined to be the most basal using the widely used criteria whereby shortaxis slices are included within the LV if 50% or more of the cavity is surrounded by ventricular myocardium. Despite this finding, however, variability in base positioning did not have a significant impact on volumetric measurements, with no significant differences in volume and EF measurements seen between Methods 2 and 3. It is possible that the more basal positioning of the LV base using the long-axis method was offset by including a greater amount of cavity from the basal short axis slice using the short-axis method, due to the technique of joining up the endocardial contours with a curved line through the blood pool, as is widely practiced (27). Although there was a significant difference between automated and visually determined end-systole, the absolute difference in frame count was small, and the effect on ESV minimal (2 ml). Unlike with volumetric measurements, methodological differences in LV base identification appeared to be more important than method of endocardial tracing for measurement of mass, with a mean difference of 34g seen between Methods 2 and 3, which had comparable detailed endocardial contouring, but different base positioning. This large difference is in accordance with the difference seen between studies by Maceira et al (24) and Hudsmith at al (6) that defined normal values for LV mass (and volumes and EF) using different analysis techniques. Using Method 3, Table 5 Comparison of Ejection Fraction (EF) Measurements Using Each Geometric Model With Methods 1 and 2 Mean difference 6SD P-value 95% limits of agreement Correlation coefficient (r s ) EF (%) Method 1 Teich to Method 1 ModSimp to Method 1 Hemi to Method 1 Mono -967 < to Method 1 Biplane -767 < to Method 1 Triplane -566 < to EF (%) Method 2 Teich to Method 2 ModSimp 8611 < to Method 2 Hemi 767 < to Method 2 Mono to Method 2 Biplane to Method 2 Triplane to Mean difference (SD), significance of the difference, Bland-Altman 95% limits of agreement and correlation coefficient for the comparison of each geometric model with Method 3. Teich Teicholz, ModSimp modified Simpson s, Hemi hemi-ellipse, Mono monoplane.

9 CMR Methods for LV Assessment 9 Figure 6. Interobserver and intraobserver variability of each geometric model for measurement of volume indices and EF, expressed as repeatability coefficients. *A significant difference (P < 0.05) compared to conventional analysis (Method 3). Geometric models showed significantly greater variability than conventional analysis for all parameters, with the exception of the interobserver variability of triplane, biplane, modified Simpson s, and Teicholz model-derived measurements of EDV. Abbreviations as in Tables 2 and 4. Maceira et al found mean myocardial mass to be 156g in healthy males and 128g in healthy females but using Method 2, Hudsmith et al found mean mass to be 123g and 96g in healthy males and females, respectively. Of note, there were no discernable differences in epicardial contouring between Methods 2 and 3. This finding further highlights the importance of using normal reference ranges appropriate to analysis method used. The method of simplified endocardial border tracing did lead to a significant underestimation of mass compared to the detailed method; however, the difference was small in comparison (9g). Methodological differences in image analysis led to substantial disparity in observer agreement, a finding that has important implications for patient follow-up and for determining research study sample sizes (28). Method 3 had the highest observer agreement for measurement of EF and mass. This technique used the long-axis-derived method of identifying the LV base, which allows straightforward identification of the mitral valve plane from long-axis images, and hence LV base, with minimal scope for observer variability. Furthermore, end-systole was determined automatically. Therefore, the main variability for this technique stemmed from the manual adjustment required to set the thresholding level appropriate to the endocardial border. However, this involves substantially less user input than the manual endocardial tracing techniques, and our experience is that accustomed observers perform this consistently. Indeed, likely related to the minimal manual input required, these results demonstrate that the threshold-determined method of endocardial contouring is considerably more consistent than manual tracing. Epicardial border tracing was another potential source of variability, although this would only affect mass measurements, and due to its typically circular shape, epicardial contouring is generally straightforward. In light of these findings, it is suggested that the methods of long-axis-derived LV base identification, detailed endocardial contouring using a signal threshold (or similar)-based technique and automated determination of end-systole, which involve minimal manual input, should be incorporated into all analysis software. Measurement of mass showed greater variability than all other parameters, which is in keeping with the findings of other studies (28). A number of studies have advocated the use of geometric models as a timesaving alternate to conventional CMR image analysis (12 16). Indeed, these models are incorporated into state-of-the-art CMR software analysis packages, and are widely used by other imaging modalities, particularly echocardiography. The findings of this study, the largest to assess the utility of geometric modeling, show that the agreement between geometric modeling techniques and conventional analysis is poor, with wide limits of agreement for all models. In addition, the variability of all models was too high for clinical or academic use, with repeatability coefficients of 13% or greater for measurement of EF. These findings are in apparent contrast to a smaller study (n ¼ 25) by Thiele et al (12), which concluded that the level of agreement between the Triplane model and conventional analysis for measurement of EF was such that this model could be used clinically. However, agreement was assessed using mean relative difference, defined as the difference of two techniques divided by their mean value, and the 95% confidence interval of the mean difference, which represents the confidence interval for the mean difference derived from the sample of individuals and, importantly, does not represent the confidence interval for the difference for an individual, and by definition would result in substantially narrower limits of agreement than the method used in the present study (ie, absolute mean difference and 95% limits of agreement, calculated as the mean difference 6 2 SDs of the mean difference, as described by Bland and Altman (20)). Indeed, using the method described by Thiele et al, the mean relative difference in the present study between the Triplane model and Method 3 is 0.1%, with a 95% confidence interval of 61.6%, which is similar to that reported by Thiele ( % for patients with no regional dysfunction and %

10 10 Miller et al. for patients with regional dysfunction). Agreement data for measurement of EDV and ESV were not presented and patients with suboptimal image quality were excluded. It is suggested that geometric models devalue CMR and their use should be discouraged. Future methods for quantification of LV mass, volume, and EF may include deformable models, active shape and appearance modeling, and atlas-based methods, in which models are derived from anatomical or statistical properties of large databases of patients and then applied to individuals (29). Such techniques have the potential to allow fully automated, and hence highly accurate and reproducible analysis, although they are currently not sufficiently developed for clinical use. The main limitation of this study is that no comparison is made to an externally validated gold standard, so accuracy cannot be assessed. In addition, this is a single-site, single-vendor study. Interobserver variability was assessed using only two observers and intraobserver variability was assessed using only a single observer. Reproducibility of measurements, requiring two separate scans, was not assessed. In conclusion, real-world variability in the process of quantifying LV indices from SSFP CMR images leads to significant differences in LV mass, volume and EF measurements, and observer variability. Reference ranges appropriate to analysis method used must be applied. Use of geometric models for analyzing CMR images should be discouraged. REFERENCES 1. Ghali JK, Liao Y, Simmons B, Castaner A, Cao G, Cooper RS. The prognostic role of left ventricular hypertrophy in patients with or without coronary artery disease. Ann Intern Med 1992;117: White HD, Norris RM, Brown MA, Brandt PW, Whitlock RM, Wild CJ. Left ventricular end-systolic volume as the major determinant of survival after recovery from myocardial infarction. Circulation 1987;76: Pfeffer MA, Braunwald E, Moye LA, et al. Effect of captopril on mortality and morbidity in patients with left ventricular dysfunction after myocardial infarction. Results of the survival and ventricular enlargement trial. The SAVE Investigators. N Engl J Med 1992;327: Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005;352: Hwang MH, Hammermeister KE, Oprian C, et al. Preoperative identification of patients likely to have left ventricular dysfunction after aortic valve replacement. Participants in the Veterans Administration Cooperative Study on Valvular Heart Disease. Circulation 1989;80:I Hudsmith LE, Petersen SE, Francis JM, Robson MD, Neubauer S. Normal human left and right ventricular and left atrial dimensions using steady state free precession magnetic resonance imaging. J Cardiovasc Magn Reson 2005;7: Karamitsos TD, Hudsmith LE, Selvanayagam JB, Neubauer S, Francis JM. Operator induced variability in left ventricular measurements with cardiovascular magnetic resonance is improved after training. J Cardiovasc Magn Reson 2007;9: Bloomer TN, Plein S, Radjenovic A, et al. Cine MRI using steady state free precession in the radial long axis orientation is a fast accurate method for obtaining volumetric data of the left ventricle. J Magn Reson Imaging 2001;14: Alfakih K, Plein S, Thiele H, Jones T, Ridgway JP, Sivananthan MU. Normal human left and right ventricular dimensions for MRI as assessed by turbo gradient echo and steady-state free precession imaging sequences. J Magn Reson Imaging 2003;17: Natori S, Lai S, Finn JP, et al. Cardiovascular function in multiethnic study of atherosclerosis: normal values by age, sex, and ethnicity. AJR Am J Roentgenol 2006;186:S Salton CJ, Chuang ML, O Donnell CJ, et al. Gender differences and normal left ventricular anatomy in an adult population free of hypertension. A cardiovascular magnetic resonance study of the Framingham Heart Study Offspring cohort. J Am Coll Cardiol 2002;39: Thiele H, Paetsch I, Schnackenburg B, et al. Improved accuracy of quantitative assessment of left ventricular volume and ejection fraction by geometric models with steady-state free precession. J Cardiovasc Magn Reson 2002;4: Dulce MC, Mostbeck GH, Friese KK, Caputo GR, Higgins CB. Quantification of the left ventricular volumes and function with cine MR imaging: comparison of geometric models with threedimensional data. Radiology 1993;188: Hergan K, Schuster A, Fruhwald J, Mair M, Burger R, Topker M. Comparison of left and right ventricular volume measurement using the Simpson s method and the area length method. Eur J Radiol 2008;65: Sievers B, Brandts B, Franken U, Trappe HJ. Single and biplane TrueFISP cardiovascular magnetic resonance for rapid evaluation of left ventricular volumes and ejection fraction. J Cardiovasc Magn Reson 2004;6: Wassmuth R, Schulz-Menger J. Normal left and right ventricular volume and function. Kwong RY, editor. Totowa, NJ: Humana Press; Ambrose JA, Meller J, Herman MV, Teichholz LE. The ventricular A wave: a new echocardiographic index of late diastolic filling of the left ventricle. Am Heart J 1978;96: Wyatt HL, Heng MK, Meerbaum S, et al. Cross-sectional echocardiography. I. Analysis of mathematic models for quantifying mass of the left ventricle in dogs. Circulation 1979;60: Wyatt HL, Heng MK, Meerbaum S, et al. Cross-sectional echocardiography. II. Analysis of mathematic models for quantifying volume of the formalin-fixed left ventricle. Circulation 1980;61: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1: Moon JC, Lorenz CH, Francis JM, Smith GC, Pennell DJ. Breathhold FLASH and FISP cardiovascular MR imaging: left ventricular volume differences and reproducibility. Radiology 2002;223: Papavassiliu T, Kuhl HP, Schroder M, et al. Effect of endocardial trabeculae on left ventricular measurements and measurement reproducibility at cardiovascular MR imaging. Radiology 2005; 236: Weinsaft JW, Cham MD, Janik M, et al. Left ventricular papillary muscles and trabeculae are significant determinants of cardiac MRI volumetric measurements: effects on clinical standards in patients with advanced systolic dysfunction. Int J Cardiol 2008; 126: Maceira AM, Prasad SK, Khan M, Pennell DJ. Normalized left ventricular systolic and diastolic function by steady state free precession cardiovascular magnetic resonance. J Cardiovasc Magn Reson 2006;8: Fieno DS, Jaffe WC, Simonetti OP, Judd RM, Finn JP. TrueFISP: assessment of accuracy for measurement of left ventricular mass in an animal model. J Magn Reson Imaging 2002;15: Francois CJ, Fieno DS, Shors SM, Finn JP. Left ventricular mass: manual and automatic segmentation of true FISP and FLASH cine MR images in dogs and pigs. Radiology 2004;230: Bogaert J. Cardiac function. In: Bogaert J, Dymarkowski S, Taylor AM, Muthurangu V, editors. Clinical cardiac MRI, 2nd ed. New York: Springer; p Grothues F, Smith GC, Moon JC, et al. Comparison of interstudy reproducibility of cardiovascular magnetic resonance with twodimensional echocardiography in normal subjects and in patients with heart failure or left ventricular hypertrophy. Am J Cardiol 2002;90: Petitjean C, Dacher JN. A review of segmentation methods in short axis cardiac MR images. Med Image Anal 2011;15:

Analysis of left ventricular mass, volume indices and ejection fraction from. SSFP cine imaging: a comparison of semi-automated, simplified manual,

Analysis of left ventricular mass, volume indices and ejection fraction from. SSFP cine imaging: a comparison of semi-automated, simplified manual, Analysis of left ventricular mass, volume indices and ejection fraction from SSFP cine imaging: a comparison of semi-automated, simplified manual, detailed manual and geometric modelling techniques Christopher

More information

Impact of the ECG gating method on ventricular volumes and ejection fractions assessed by cardiovascular magnetic resonance imaging

Impact of the ECG gating method on ventricular volumes and ejection fractions assessed by cardiovascular magnetic resonance imaging Journal of Cardiovascular Magnetic Resonance (2005) 7, 441 446 Copyright D 2005 Taylor & Francis Inc. ISSN: 1097-6647 print / 1532-429X online DOI: 10.1081/JCMR-200053515 VENTRICULAR FUNCTION Impact of

More information

CME. Original Research

CME. Original Research CME JOURNAL OF MAGNETIC RESONANCE IMAGING 38:829 835 (2013) Original Research Interstudy Variability in Cardiac Magnetic Resonance Imaging Measurements of Ventricular Volume, Mass, and Ejection Fraction

More information

Measurement of Ventricular Volumes and Function: A Comparison of Gated PET and Cardiovascular Magnetic Resonance

Measurement of Ventricular Volumes and Function: A Comparison of Gated PET and Cardiovascular Magnetic Resonance BRIEF COMMUNICATION Measurement of Ventricular Volumes and Function: A Comparison of Gated PET and Cardiovascular Magnetic Resonance Kim Rajappan, MBBS 1,2 ; Lefteris Livieratos, MSc 2 ; Paolo G. Camici,

More information

CHAPTER. Quantification in cardiac MRI. This chapter was adapted from:

CHAPTER. Quantification in cardiac MRI. This chapter was adapted from: CHAPTER Quantification in cardiac MRI This chapter was adapted from: Quantification in cardiac MRI Rob J. van der Geest, Johan H.C. Reiber Journal of Magnetic Resonance Imaging 1999, Volume 10, Pages 602-608.

More information

How much are atrial volumes and ejection fractions assessed by cardiac magnetic resonance imaging influenced by the ECG gating method?

How much are atrial volumes and ejection fractions assessed by cardiac magnetic resonance imaging influenced by the ECG gating method? Journal of Cardiovascular Magnetic Resonance (2005) 7, 587 593 Copyright D 2005 Taylor & Francis Inc. ISSN: 1097-6647 print / 1532-429X online DOI: 10.1081/JCMR-200060635 VENTRICULAR FUNCTION How much

More information

Impact of Papillary Muscles in Ventricular Volume and Ejection Fraction Assessment by Cardiovascular Magnetic Resonance

Impact of Papillary Muscles in Ventricular Volume and Ejection Fraction Assessment by Cardiovascular Magnetic Resonance JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE 1 Vol. 6, No. 1, pp. 9 16, 2004 VENTRICULAR FUNCTION Impact of Papillary Muscles in Ventricular Volume and Ejection Fraction Assessment by Cardiovascular Magnetic

More information

Single and Biplane TrueFISP Cardiovascular Magnetic Resonance for Rapid Evaluation of Left Ventricular Volumes and Ejection Fraction

Single and Biplane TrueFISP Cardiovascular Magnetic Resonance for Rapid Evaluation of Left Ventricular Volumes and Ejection Fraction JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE 1 Vol. 6, No. 3, pp. 593 600, 2004 FUNCTION Single and Biplane TrueFISP Cardiovascular Magnetic Resonance for Rapid Evaluation of Left Ventricular Volumes and

More information

Impaired Regional Myocardial Function Detection Using the Standard Inter-Segmental Integration SINE Wave Curve On Magnetic Resonance Imaging

Impaired Regional Myocardial Function Detection Using the Standard Inter-Segmental Integration SINE Wave Curve On Magnetic Resonance Imaging Original Article Impaired Regional Myocardial Function Detection Using the Standard Inter-Segmental Integration Ngam-Maung B, RT email : chaothawee@yahoo.com Busakol Ngam-Maung, RT 1 Lertlak Chaothawee,

More information

Relationship of Number of Phases per Cardiac Cycle and Accuracy of Measurement of Left Ventricular Volumes, Ejection Fraction, and Mass

Relationship of Number of Phases per Cardiac Cycle and Accuracy of Measurement of Left Ventricular Volumes, Ejection Fraction, and Mass JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE 1 Vol. 6, No. 4, pp. 837 844, 2004 VENTRICULAR FUNCTION Relationship of Number of Phases per Cardiac Cycle and Accuracy of Measurement of Left Ventricular Volumes,

More information

Normal Human Left and Right Ventricular Dimensions for MRI as Assessed by Turbo Gradient Echo and Steady-State Free Precession Imaging Sequences

Normal Human Left and Right Ventricular Dimensions for MRI as Assessed by Turbo Gradient Echo and Steady-State Free Precession Imaging Sequences JOURNAL OF MAGNETIC RESONANCE IMAGING 17:323 329 (2003) Original Research Normal Human Left and Right Ventricular Dimensions for MRI as Assessed by Turbo Gradient Echo and Steady-State Free Precession

More information

Echocardiographic Assessment of the Left Ventricle

Echocardiographic Assessment of the Left Ventricle Echocardiographic Assessment of the Left Ventricle Theodora Zaglavara, MD, PhD, BSCI/BSCCT Department of Cardiovascular Imaging INTERBALKAN EUROPEAN MEDICAL CENTER 2015 The quantification of cardiac chamber

More information

Velocity Vector Imaging as a new approach for cardiac magnetic resonance: Comparison with echocardiography

Velocity Vector Imaging as a new approach for cardiac magnetic resonance: Comparison with echocardiography Velocity Vector Imaging as a new approach for cardiac magnetic resonance: Comparison with echocardiography Toshinari Onishi 1, Samir K. Saha 2, Daniel Ludwig 1, Erik B. Schelbert 1, David Schwartzman 1,

More information

Global left ventricular circumferential strain is a marker for both systolic and diastolic myocardial function

Global left ventricular circumferential strain is a marker for both systolic and diastolic myocardial function Global left ventricular circumferential strain is a marker for both systolic and diastolic myocardial function Toshinari Onishi 1, Samir K. Saha 2, Daniel Ludwig 1, Erik B. Schelbert 1, David Schwartzman

More information

Automated Image Analysis Techniques for Cardiovascular Magnetic Resonance Imaging

Automated Image Analysis Techniques for Cardiovascular Magnetic Resonance Imaging Automated Image Analysis Techniques for Cardiovascular Magnetic Resonance Imaging Robertus Jacobus van der Geest 2011 Printed by: Drukkerij Mostert & van Onderen, Leiden. ISBN 978-94-90858-04-9 2011, R.J.

More information

Chamber Quantitation Guidelines: What is New?

Chamber Quantitation Guidelines: What is New? Chamber Quantitation Guidelines: What is New? Roberto M Lang, MD J AM Soc Echocardiogr 2005; 18:1440-1463 1 Approximately 10,000 citations iase in itune Cardiac Chamber Quantification: What is New? Database

More information

LV FUNCTION ASSESSMENT: WHAT IS BEYOND EJECTION FRACTION

LV FUNCTION ASSESSMENT: WHAT IS BEYOND EJECTION FRACTION LV FUNCTION ASSESSMENT: WHAT IS BEYOND EJECTION FRACTION Jamilah S AlRahimi Assistant Professor, KSU-HS Consultant Noninvasive Cardiology KFCC, MNGHA-WR Introduction LV function assessment in Heart Failure:

More information

Full terms and conditions of use:

Full terms and conditions of use: This article was downloaded by:[uppsala Universitetsbibiotek] [Uppsala Universitetsbibiotek] On: 3 May 2007 Access Details: [subscription number 731962693] Publisher: Informa Healthcare Informa Ltd Registered

More information

Use of Cardiac Computed Tomography for Ventricular Volumetry in Late Postoperative Patients with Tetralogy of Fallot

Use of Cardiac Computed Tomography for Ventricular Volumetry in Late Postoperative Patients with Tetralogy of Fallot Korean J Thorac Cardiovasc Surg 2017;50:71-77 ISSN: 2233-601X (Print) ISSN: 2093-6516 (Online) CLINICAL RESEARCH https://doi.org/10.5090/kjtcs.2017.50.2.71 Use of Cardiac Computed Tomography for Ventricular

More information

INTRODUCTION ABSTRACT METHODS

INTRODUCTION ABSTRACT METHODS CHAPTER Assessment of left ventricular ejection fraction in patients eligible for ICD therapy: Discrepancy between cardiac magnetic resonance imaging and 2D echocardiography Stefan de Haan Karin de Boer

More information

Reproducibility and Accuracy of Echocardiographic Measurements of Left Ventricular Parameters Using Real-Time Three-Dimensional Echocardiography

Reproducibility and Accuracy of Echocardiographic Measurements of Left Ventricular Parameters Using Real-Time Three-Dimensional Echocardiography Journal of the American College of Cardiology Vol. 44, No. 4, 2004 2004 by the American College of Cardiology Foundation ISSN 0735-1097/04/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2004.05.050

More information

Automatic cardiac contour propagation in short axis cardiac MR images

Automatic cardiac contour propagation in short axis cardiac MR images International Congress Series 1281 (2005) 351 356 www.ics-elsevier.com Automatic cardiac contour propagation in short axis cardiac MR images G.L.T.F. Hautvast a,b, T, M. Breeuwer a, S. Lobregt a, A. Vilanova

More information

Quantification of Cardiac Chamber Size

Quantification of Cardiac Chamber Size 2017 KSE 2017-11-25 Quantification of Cardiac Chamber Size Division of Cardiology Keimyung University Dongsan Medical Center In-Cheol Kim M.D., Ph.D. LV size and function Internal linear dimensions PLX

More information

CARDIAC MRI. Cardiovascular Disease. Cardiovascular Disease. Cardiovascular Disease. Overview

CARDIAC MRI. Cardiovascular Disease. Cardiovascular Disease. Cardiovascular Disease. Overview CARDIAC MRI Dr Yang Faridah A. Aziz Department of Biomedical Imaging University of Malaya Medical Centre Cardiovascular Disease Diseases of the circulatory system, also called cardiovascular disease (CVD),

More information

Objectives. CMR Volumetric Analysis 8/25/11. CMR Volumetric Analysis Technique. Cardiac imaging plane acquisition. CMR Volumetric Analysis

Objectives. CMR Volumetric Analysis 8/25/11. CMR Volumetric Analysis Technique. Cardiac imaging plane acquisition. CMR Volumetric Analysis Objectives Cynthia K. Rigsby Children s Memorial Hospital Chicago, IL CMR volumetric analysis Techniques Normalized data Sources of error CMR phase contrast flow analysis Techniques What we can do with

More information

Normal values for cardiovascular magnetic resonance in adults and children

Normal values for cardiovascular magnetic resonance in adults and children Kawel-Boehm et al. Journal of Cardiovascular Magnetic Resonance (2015) 17:29 DOI 10.1186/s12968-015-0111-7 REVIEW Normal values for cardiovascular magnetic resonance in adults and children Nadine Kawel-Boehm

More information

General Cardiovascular Magnetic Resonance Imaging

General Cardiovascular Magnetic Resonance Imaging 2 General Cardiovascular Magnetic Resonance Imaging 19 Peter G. Danias, Cardiovascular MRI: 150 Multiple-Choice Questions and Answers Humana Press 2008 20 Cardiovascular MRI: 150 Multiple-Choice Questions

More information

Reference Normal Absolute and Indexed Values From ECG- Gated MDCT: Left Atrial Volume, Function, and Diameter

Reference Normal Absolute and Indexed Values From ECG- Gated MDCT: Left Atrial Volume, Function, and Diameter Cardiopulmonary Imaging Original Research Stojanovska et al. ECG-Gated MDCT Cardiopulmonary Imaging Original Research Jadranka Stojanovska Paul Cronin Smita Patel Barry H. Gross Hakan Oral Komal Chughtai

More information

2D Free-Breathing Dual Navigator-Gated Cardiac Function Validated Against the 2D Breath-hold Acquisition

2D Free-Breathing Dual Navigator-Gated Cardiac Function Validated Against the 2D Breath-hold Acquisition JOURNAL OF MAGNETIC RESONANCE IMAGING 28:773 777 (2008) Technical Note 2D Free-Breathing Dual Navigator-Gated Cardiac Function Validated Against the 2D Breath-hold Acquisition Dana C. Peters, PhD, 1 *

More information

LV function in ischemic heart failure - decreased correlation between Echo and CMR

LV function in ischemic heart failure - decreased correlation between Echo and CMR LV function in ischemic heart failure - decreased correlation between Echo and CMR Poster No.: C-0590 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Gruszczy#ska, L. Krzych, K. Golba, P. Ulbrych,

More information

Impact of the Revision of Arrhythmogenic Right Ventricular Cardiomyopathy/Dysplasia Task Force Criteria on Its Prevalence by CMR Criteria

Impact of the Revision of Arrhythmogenic Right Ventricular Cardiomyopathy/Dysplasia Task Force Criteria on Its Prevalence by CMR Criteria JACC: CARDIOVASCULAR IMAGING VOL. 4, NO. 3, 2011 2011 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION ISSN 1936-878X/$36.00 PUBLISHED BY ELSEVIER INC. DOI:10.1016/j.jcmg.2011.01.005 Impact of the Revision

More information

Automated Volumetric Cardiac Ultrasound Analysis

Automated Volumetric Cardiac Ultrasound Analysis Whitepaper Automated Volumetric Cardiac Ultrasound Analysis ACUSON SC2000 Volume Imaging Ultrasound System Bogdan Georgescu, Ph.D. Siemens Corporate Research Princeton, New Jersey USA Answers for life.

More information

좌심실수축기능평가 Cardiac Function

좌심실수축기능평가 Cardiac Function Basic Echo Review Course 좌심실수축기능평가 Cardiac Function Seonghoon Choi Cardiology Hallym university LV systolic function Systolic function 좌심실수축기능 - 심근의수축으로심실에서혈액을대동맥으로박출하는기능 실제임상에서 LV function 의의미 1Diagnosis

More information

Three-dimensional Wall Motion Tracking:

Three-dimensional Wall Motion Tracking: Three-dimensional Wall Motion Tracking: A Novel Echocardiographic Method for the Assessment of Ventricular Volumes, Strain and Dyssynchrony Jeffrey C. Hill, BS, RDCS, FASE Jennifer L. Kane, RCS Gerard

More information

Cardiac Chamber Quantification by Echocardiography

Cardiac Chamber Quantification by Echocardiography Cardiac Chamber Quantification by Echocardiography Maryam Bokhamseen, RCS, RCDS, EACVI Echotechnologist ǁ, Non invasive Cardiac Laboratory King Abdulaziz Cardiac Center. Outline: Introduction. Background

More information

Original Article Cardiovascular Imaging. Eun-Ah Park, MD 1, Whal Lee, MD 1, Hyung-Kwan Kim, MD 2, Jin Wook Chung, MD 1 INTRODUCTION

Original Article Cardiovascular Imaging. Eun-Ah Park, MD 1, Whal Lee, MD 1, Hyung-Kwan Kim, MD 2, Jin Wook Chung, MD 1 INTRODUCTION Original Article Cardiovascular Imaging http://dx.doi.org/1.3348/kjr.215.16.1.4 pissn 1229-6929 eissn 25-833 Korean J Radiol 215;16(1):4-12 Effect of Papillary Muscles and Trabeculae on Left Ventricular

More information

cardiac imaging planes planning basic cardiac & aortic views for MR

cardiac imaging planes planning basic cardiac & aortic views for MR cardiac imaging planes planning basic cardiac & aortic views for MR Dianna M. E. Bardo, M. D. Assistant Professor of Radiology & Cardiovascular Medicine Director of Cardiac Imaging cardiac imaging planes

More information

Disclosure of Interests. No financial relationships to disclose concerning the content of this presentation or session.

Disclosure of Interests. No financial relationships to disclose concerning the content of this presentation or session. Comparison of Free Breathing Cardiac MRI Radial Technique to the Standard Multi Breath-Hold Cine SSFP CMR Technique For the Assessment of LV Volumes and Function Shimon Kolker, Giora Weisz, Naama Bogot,

More information

Original Research. Girish Narayan, MD, 1 * Krishna Nayak, PhD, 2 John Pauly, PhD, 3 and Bob Hu, MD 2,4

Original Research. Girish Narayan, MD, 1 * Krishna Nayak, PhD, 2 John Pauly, PhD, 3 and Bob Hu, MD 2,4 JOURNAL OF MAGNETIC RESONANCE IMAGING 22:59 66 (2005) Original Research Single-Breathhold, Four-Dimensional, Quantitative Assessment of LV and RV Function Using Triggered, Real-Time, Steady-State Free

More information

10/7/2013. Systolic Function How to Measure, How Accurate is Echo, Role of Contrast. Thanks to our Course Director: Neil J.

10/7/2013. Systolic Function How to Measure, How Accurate is Echo, Role of Contrast. Thanks to our Course Director: Neil J. Systolic Function How to Measure, How Accurate is Echo, Role of Contrast Neil J. Weissman, MD MedStar Health Research Institute & Professor of Medicine Georgetown University Washington, D.C. No Disclosures

More information

Establishment and Performance of a Magnetic Resonance Cardiac Function Clinic

Establishment and Performance of a Magnetic Resonance Cardiac Function Clinic Journal of Cardiovascular Magnetic Resonance", 2( I), 15-22 (2000) Establishment and Performance of a Magnetic Resonance Cardiac Function Clinic Nicholas G. Bellenger, Jane M. Francis, Ceri L. Davies,

More information

Prof. JL Zamorano Hospital Universitario Ramón y Cajal

Prof. JL Zamorano Hospital Universitario Ramón y Cajal Prof. JL Zamorano Hospital Universitario Ramón y Cajal Fully Automated Quantification Software Adaptive analytical algorithm consists in knowledge-based identification of global shape and specific adaptation

More information

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor Cardiothoracic Radiology Disclosure I have no disclosure pertinent to this presentation.

More information

Noncoronary Cardiac MDCT

Noncoronary Cardiac MDCT Noncoronary Cardiac MDCT David A. Bluemke, M.D., Ph.D. Professor, of Radiology and Medicine Johns Hopkins University School of Medicine Baltimore, Maryland Toshiba Disclosures Grant support Noncoronary

More information

JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE 1 Vol. 6, No. 4, pp , 2004

JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE 1 Vol. 6, No. 4, pp , 2004 JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE 1 Vol. 6, No. 4, pp. 855 863, 2004 ATRIAL VOLUMES Assessment of Left Atrial Volumes in Sinus Rhythm and Atrial Fibrillation Using the Biplane Area-Length Method

More information

Planimetry of Mitral Valve Stenosis by Magnetic Resonance Imaging

Planimetry of Mitral Valve Stenosis by Magnetic Resonance Imaging Journal of the American College of Cardiology Vol. 45, No. 12, 2005 2005 by the American College of Cardiology Foundation ISSN 0735-1097/05/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2005.03.036

More information

Advanced Multi-Layer Speckle Strain Permits Transmural Myocardial Function Analysis in Health and Disease:

Advanced Multi-Layer Speckle Strain Permits Transmural Myocardial Function Analysis in Health and Disease: Advanced Multi-Layer Speckle Strain Permits Transmural Myocardial Function Analysis in Health and Disease: Clinical Case Examples Jeffrey C. Hill, BS, RDCS Echocardiography Laboratory, University of Massachusetts

More information

Contrast-enhanced echocardiography improves agreement on the assessment of ejection fraction and left ventricular function. A multicentre study

Contrast-enhanced echocardiography improves agreement on the assessment of ejection fraction and left ventricular function. A multicentre study Eur J Echocardiography 7 Suppl. 2 (2006) S16 S21 Contrast-enhanced echocardiography improves agreement on the assessment of ejection fraction and left ventricular function. A multicentre study Rainer Hoffmann*

More information

Martin G. Keane, MD, FASE Temple University School of Medicine

Martin G. Keane, MD, FASE Temple University School of Medicine Martin G. Keane, MD, FASE Temple University School of Medicine Measurement of end-diastolic LV internal diameter (LVIDd) made by properly-oriented M-Mode techniques in the Parasternal Long Axis View (PLAX):

More information

Left atrial function. Aliakbar Arvandi MD

Left atrial function. Aliakbar Arvandi MD In the clinic Left atrial function Abstract The left atrium (LA) is a left posterior cardiac chamber which is located adjacent to the esophagus. It is separated from the right atrium by the inter-atrial

More information

Ventricular function assessment using MRI: comparative study between cartesian and radial techniques of k-space filling

Ventricular function assessment using MRI: comparative study between cartesian and radial techniques of k-space filling Ventricular function assessment using MRI: comparative study between cartesian and radial techniques of k-space filling Poster No.: B-0681 Congress: ECR 014 Type: Scientific Paper Authors: C. Ferreira,

More information

MAYON VOLCANO: FAST FACTS

MAYON VOLCANO: FAST FACTS MAYON VOLCANO: FAST FACTS Type of Volcano: Stratovolcano Elevation: 2.46 km Base Diameter: 20 km Base Circumference: 62.8 km Area: 314.1 km 2 Reference: http://www.phivolcs.dost.gov.ph/html/update_vmepd/volcano/volcanolist/mayon.htm

More information

Normal LV Ejection Fraction Limits Using 4D-MSPECT: Comparisons of Gated Perfusion and Gated Blood Pool SPECT Data with Planar Blood Pool

Normal LV Ejection Fraction Limits Using 4D-MSPECT: Comparisons of Gated Perfusion and Gated Blood Pool SPECT Data with Planar Blood Pool Normal LV Ejection Fraction Limits Using 4D-MSPECT: Comparisons of Gated Perfusion and Gated Blood Pool SPECT Data with Planar Blood Pool EP Ficaro, JN Kritzman, JR Corbett University of Michigan Health

More information

Investigation of the Golden Ratio in Human Heart Anatomically by Cardiac Magnetic Resonance Imaging Method

Investigation of the Golden Ratio in Human Heart Anatomically by Cardiac Magnetic Resonance Imaging Method International Journal of Research Studies in Medical and Health Sciences Volume 3, Issue 11, 2018, PP 10-14 ISSN : 2456-6373 Investigation of the Golden Ratio in Human Heart Anatomically by Cardiac Magnetic

More information

Operator induced variability in cardiovascular MR: left ventricular measurements and their reproducibility

Operator induced variability in cardiovascular MR: left ventricular measurements and their reproducibility Journal of Cardiovascular Magnetic Resonance (2005) 7, 447 457 Copyright D 2005 Taylor & Francis Inc. ISSN: 1097-6647 print / 1532-429X online DOI: 10.1081/JCMR-200053578 VENTRICULAR FUNCTION Operator

More information

Assessment of cardiac function with 3D echocardiography. Đánh giá chức năng tim bằng siêu âm tim 3D

Assessment of cardiac function with 3D echocardiography. Đánh giá chức năng tim bằng siêu âm tim 3D Assessment of cardiac function with 3D echocardiography Đánh giá chức năng tim bằng siêu âm tim 3D TS. BS. Nguyễn Thị Thu Hoài Viện Tim Mạch Quốc Gia Việt Nam TỪ SIÊU ÂM M-mode ĐẾN SIÊU ÂM 3D TỪ SIÊU ÂM

More information

Reproducibility of Chronic and Acute Infarct Size Measurement by Delayed Enhancement-Magnetic Resonance Imaging

Reproducibility of Chronic and Acute Infarct Size Measurement by Delayed Enhancement-Magnetic Resonance Imaging Journal of the American College of Cardiology Vol. 47, No. 8, 2006 2006 by the American College of Cardiology Foundation ISSN 0735-1097/06/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2005.11.065

More information

Multiple Gated Acquisition (MUGA) Scanning

Multiple Gated Acquisition (MUGA) Scanning Multiple Gated Acquisition (MUGA) Scanning Dmitry Beyder MPA, CNMT Nuclear Medicine, Radiology Barnes-Jewish Hospital / Washington University St. Louis, MO Disclaimers/Relationships Standard of care research

More information

Normalized Left Ventricular Systolic and Diastolic Function by Steady State Free Precession Cardiovascular Magnetic Resonance

Normalized Left Ventricular Systolic and Diastolic Function by Steady State Free Precession Cardiovascular Magnetic Resonance Journal of Cardiovascular Magnetic Resonance (2006) 8, 4 426 Copyright c 2006 Taylor & Francis Group, LLC ISSN: 1097-6647 print / 32-429X online DOI: 10.1080/10976640600572889 VENTRICULAR FUNCTION Normalized

More information

Validation of performance of a free-of-charge plugin for the open-source platforms to perform cardiac segmentation in magnetic resonance imaging

Validation of performance of a free-of-charge plugin for the open-source platforms to perform cardiac segmentation in magnetic resonance imaging Free Free-of of-charge plugin to perform segmentation in CMR 83 Validation of performance of a free-of-charge plugin for the open-source platforms to perform cardiac segmentation in magnetic resonance

More information

4D Auto LAQ (Left Atrial Quantification)

4D Auto LAQ (Left Atrial Quantification) 4D Auto LAQ (Left Atrial Quantification) Introduction There has been an increased interest in quantification of the left atrium (LA) for various types of diseases; e.g. heart failure, heart valve diseases,

More information

Adult Echocardiography Examination Content Outline

Adult Echocardiography Examination Content Outline Adult Echocardiography Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 2 3 4 5 Anatomy and Physiology Pathology Clinical Care and Safety Measurement Techniques, Maneuvers,

More information

Strain-encoded cardiac magnetic resonance imaging: a new approach for fast estimation of left ventricular function

Strain-encoded cardiac magnetic resonance imaging: a new approach for fast estimation of left ventricular function Lapinskas et al. BMC Cardiovascular Disorders (2019) 19:52 https://doi.org/10.1186/s12872-019-1031-5 TECHNICAL ADVANCE Open Access Strain-encoded cardiac magnetic resonance imaging: a new approach for

More information

ORIGINAL ARTICLE. Keywords Real-time 3D echocardiography. Daily practice. Observer experience. CMR. Introduction

ORIGINAL ARTICLE. Keywords Real-time 3D echocardiography. Daily practice. Observer experience. CMR. Introduction Neth Heart J (2014) 22:383 390 DOI 10.1007/s12471-014-0577-1 ORIGINAL ARTICLE Assessment of LV ejection fraction using real-time 3D echocardiography in daily practice: direct comparison of the volumetric

More information

Objectives. Diastology: What the Radiologist Needs to Know. LV Diastolic Function: Introduction. LV Diastolic Function: Introduction

Objectives. Diastology: What the Radiologist Needs to Know. LV Diastolic Function: Introduction. LV Diastolic Function: Introduction Objectives Diastology: What the Radiologist Needs to Know. Jacobo Kirsch, MD Cardiopulmonary Imaging, Section Head Division of Radiology Cleveland Clinic Florida Weston, FL To review the physiology and

More information

Conflict of Interests

Conflict of Interests The Left Ventricle: How Should We Quantify Its Size and Function; Is It Time for 3D in Everyone? Roberto M Lang, MD Conflict of Interests Philips Medical Imaging Research Grants Speakers bureau Advisory

More information

Assessment of LV systolic function

Assessment of LV systolic function Tutorial 5 - Assessment of LV systolic function Assessment of LV systolic function A knowledge of the LV systolic function is crucial in the undertanding of and management of unstable hemodynamics or a

More information

Certificate in Clinician Performed Ultrasound (CCPU) Syllabus. Rapid Cardiac Echo (RCE)

Certificate in Clinician Performed Ultrasound (CCPU) Syllabus. Rapid Cardiac Echo (RCE) Certificate in Clinician Performed Ultrasound (CCPU) Syllabus Rapid Cardiac Echo (RCE) Purpose: Rapid Cardiac Echocardiography (RCE) This unit is designed to cover the theoretical and practical curriculum

More information

Right Ventricular Strain in Normal Healthy Adult Filipinos: A Retrospective, Cross- Sectional Pilot Study

Right Ventricular Strain in Normal Healthy Adult Filipinos: A Retrospective, Cross- Sectional Pilot Study Right Ventricular Strain in Normal Healthy Adult Filipinos: A Retrospective, Cross- Sectional Pilot Study By Julius Caesar D. de Vera, MD Jonnah Fatima B. Pelat, MD Introduction Right ventricle contributes

More information

Policy #: 222 Latest Review Date: March 2009

Policy #: 222 Latest Review Date: March 2009 Name of Policy: MRI Phase-Contrast Flow Measurement Policy #: 222 Latest Review Date: March 2009 Category: Radiology Policy Grade: Active Policy but no longer scheduled for regular literature reviews and

More information

Quantifying LV function how good are we?

Quantifying LV function how good are we? Quantifying LV function how good are we? Professor Alan G Fraser Wales Heart Research Institute Cardiff University, U.K. Support for research from Hitachi Aloka, & GE Ultrasound Visual assessment of synchronicity

More information

Myocardial Strain Imaging in Cardiac Diseases and Cardiomyopathies.

Myocardial Strain Imaging in Cardiac Diseases and Cardiomyopathies. Myocardial Strain Imaging in Cardiac Diseases and Cardiomyopathies. Session: Cardiomyopathy Tarun Pandey MD, FRCR. Associate Professor University of Arkansas for Medical Sciences Disclosures No relevant

More information

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre Pushing the limits of cardiac CT Steven Dymarkowski Radiology / Medical Imaging Research Centre 5 X 2013 Introduction Rapid technological advances and new clinical applications in cardiovascular imaging

More information

MEDVISO WHITE PAPER ON STRAIN ANALYSIS IN TAGGED MR IMAGES

MEDVISO WHITE PAPER ON STRAIN ANALYSIS IN TAGGED MR IMAGES Purpose of document The purpose of this document is to document validation of the Strain analysis module in Segment software packages. Intended audience The intended audiences of this document are: Engineering

More information

Comparison of Cardiac MDCT with MRI and Echocardiography in the Assessement of Left Ventricular Function

Comparison of Cardiac MDCT with MRI and Echocardiography in the Assessement of Left Ventricular Function Comparison of Cardiac MDCT with MRI and Echocardiography in the Assessement of Left Ventricular Function Poster No.: C-0969 Congress: ECR 2012 Type: Scientific Exhibit Authors: B. Kara, Y. Paksoy, C. Erol,

More information

Radiologic Assessment of Myocardial Viability

Radiologic Assessment of Myocardial Viability November 2001 Radiologic Assessment of Myocardial Viability Joshua Moss, Harvard Medical School Year III Patient EF 66yo female with a 3-year history of intermittent chest pain previously relieved by sublingual

More information

Coronary artery disease (CAD) risk factors

Coronary artery disease (CAD) risk factors Background Coronary artery disease (CAD) risk factors CAD Risk factors Hypertension Insulin resistance /diabetes Dyslipidemia Smoking /Obesity Male gender/ Old age Atherosclerosis Arterial stiffness precedes

More information

Assessment of right ventricular contraction by speckle tracking echocardiography in pulmonary hypertension patients.

Assessment of right ventricular contraction by speckle tracking echocardiography in pulmonary hypertension patients. Biomedical Research 2017; 28 (1): 173-177 ISSN 0970-938X www.biomedres.info Assessment of right ventricular contraction by speckle tracking echocardiography in pulmonary hypertension patients. Yudong Peng,

More information

UK Biobank. Imaging modality Cardiovascular Magnetic Resonance (CMR) Version th Oct 2015

UK Biobank. Imaging modality Cardiovascular Magnetic Resonance (CMR) Version th Oct 2015 Imaging modality Cardiovascular Magnetic Resonance (CMR) Version 1.0 http://www.ukbiobank.ac.uk/ 30 th Oct 2015 This document details the procedure for the CMR scan performed at an Imaging assessment centre

More information

MYOCARDIAL STRAIN MEASUREMENTS WITH MRI USING FEATURE TRACKING

MYOCARDIAL STRAIN MEASUREMENTS WITH MRI USING FEATURE TRACKING MYOCARDIAL STRAIN MEASUREMENTS WITH MRI USING FEATURE TRACKING Mercy Kataike Victoria MSc in Physics Submission date: May 2016 Supervisor: Pål Erik Goa, IFY Norwegian University of Science and Technology

More information

ORIGINAL ARTICLE-E-LEARNING. S. de Haan & K. de Boer & J. Commandeur & A. M. Beek & A. C. van Rossum & C. P. Allaart

ORIGINAL ARTICLE-E-LEARNING. S. de Haan & K. de Boer & J. Commandeur & A. M. Beek & A. C. van Rossum & C. P. Allaart Neth Heart J (2014) 22:449 455 DOI 10.1007/s12471-014-0594-0 ORIGINAL ARTICLE-E-LEARNING Assessment of left ventricular ejection fraction in patients eligible for ICD therapy: Discrepancy between cardiac

More information

What s New in Cardiac MRI

What s New in Cardiac MRI What s New in Cardiac MRI Katie M. Hawthorne, MD Director, Cardiac MRI Main Line Health Philadelphia Cardiovascular Summit November 18, 2017 Cardiac MRI: Disclosure 2 Disclosures No financial disclosures

More information

Cardiac Function Evaluation with Cine MRI of the Heart

Cardiac Function Evaluation with Cine MRI of the Heart Cardiac Function Evaluation with Cine MRI of the Heart UNIT A11.4 MRI is considered to be the gold standard for the calculation of hemodynamic parameters of cardiac function such as ejection fraction (EF),

More information

Cardiovascular magnetic resonance reference ranges for the heart and aorta in Chinese at 3T

Cardiovascular magnetic resonance reference ranges for the heart and aorta in Chinese at 3T Le et al. Journal of Cardiovascular Magnetic Resonance (2016) 18:21 DOI 10.1186/s12968-016-0236-3 RESEARCH Open Access Cardiovascular magnetic resonance reference ranges for the heart and aorta in Chinese

More information

Role of CMR in heart failure and cardiomyopathy

Role of CMR in heart failure and cardiomyopathy Role of CMR in heart failure and cardiomyopathy Hajime Sakuma Department of Radiology, Mie University Late gadolinium enhancement (LGE) LGE MRI can demonstrate site of necrosis, fibrosis or deposition

More information

A structured report for assessment of Tetralogy of Fallot by Cardiac MRI according to recent guidelines

A structured report for assessment of Tetralogy of Fallot by Cardiac MRI according to recent guidelines A structured report for assessment of Tetralogy of Fallot by Cardiac MRI according to recent guidelines Poster No.: C-0125 Congress: ECR 2016 Type: Educational Exhibit Authors: N. Stagnaro, G. Trocchio,

More information

3D-stress echocardiography Bernard Cosyns, MD, PhD

3D-stress echocardiography Bernard Cosyns, MD, PhD 3D-stress echocardiography Bernard Cosyns, MD, PhD No Disclosure The Pro-Technology bias Sicari et al. Cardiovascular Ultrasound 2006, 4:11 Overview 2D stress echocardiography: main limitations 3D echocardiography:

More information

EDITOR S PICK CURRENT STATUS OF FULLY AUTOMATED SOFTWARE WITH THREE-DIMENSIONAL ECHOCARDIOGRAPHY FOR THE QUANTIFICATION OF LEFT VENTRICULAR FUNCTION

EDITOR S PICK CURRENT STATUS OF FULLY AUTOMATED SOFTWARE WITH THREE-DIMENSIONAL ECHOCARDIOGRAPHY FOR THE QUANTIFICATION OF LEFT VENTRICULAR FUNCTION ITOR S PICK This paper, courtesy of Yang and Takeuchi, provides a timely and well-considered update on the current status of fully-automated software with three-dimensional echocardiography for quantifying

More information

When Does 3D Echo Make A Difference?

When Does 3D Echo Make A Difference? When Does 3D Echo Make A Difference? Wendy Tsang, MD, SM Assistant Professor, University of Toronto Toronto General Hospital, University Health Network 1 Practical Applications of 3D Echocardiography Recommended

More information

Xiaorong Chen 1,2, Hongjie Hu 2, Yue Qian 2, Jiner Shu 1. Introduction

Xiaorong Chen 1,2, Hongjie Hu 2, Yue Qian 2, Jiner Shu 1. Introduction Original Article Relation of late gadolinium enhancement in cardiac magnetic resonance on the diastolic volume recovery of left ventricle with hypertrophic cardiomyopathy Xiaorong Chen 1,2, Hongjie Hu

More information

Improved Detection of Subendocardial Hyperenhancement in Myocardial Infarction Using Dark Blood Pool Delayed Enhancement MRI

Improved Detection of Subendocardial Hyperenhancement in Myocardial Infarction Using Dark Blood Pool Delayed Enhancement MRI Cardiopulmonary Imaging Original Research Farrelly et al. Dark Blood Delayed Enhancement MRI in Myocardial Infarction Cardiopulmonary Imaging Original Research Cormac Farrelly 1 Wolfgang Rehwald 2 Michael

More information

Alicia Armour, MA, BS, RDCS

Alicia Armour, MA, BS, RDCS Alicia Armour, MA, BS, RDCS No disclosures Review 2D Speckle Strain (briefly) Discuss some various patient populations & disease pathways where Strain can be helpful Discuss how to acquire images for Strain

More information

Value of echocardiography in chronic dyspnea

Value of echocardiography in chronic dyspnea Value of echocardiography in chronic dyspnea Jahrestagung Schweizerische Gesellschaft für /Schweizerische Gesellschaft für Pneumologie B. Kaufmann 16.06.2016 Chronic dyspnea Shortness of breath lasting

More information

RIGHT VENTRICULAR SIZE AND FUNCTION

RIGHT VENTRICULAR SIZE AND FUNCTION RIGHT VENTRICULAR SIZE AND FUNCTION Edwin S. Tucay, MD, FPCC, FPCC, FPSE Philippine Society of Echocardiography Quezon City, Philippines Echo Mission, BRTTH, Legaspi City, July 1-2, 2016 NO DISCLOSURE

More information

Tissue Doppler Imaging in Congenital Heart Disease

Tissue Doppler Imaging in Congenital Heart Disease Tissue Doppler Imaging in Congenital Heart Disease L. Youngmin Eun, M.D. Department of Pediatrics, Division of Pediatric Cardiology, Kwandong University College of Medicine The potential advantage of ultrasound

More information

Valvular disease : Ο ρόλος του CMR. Sophie Mavrogeni MD FESC. Onassis Cardiac Surgery Center Athens Greece

Valvular disease : Ο ρόλος του CMR. Sophie Mavrogeni MD FESC. Onassis Cardiac Surgery Center Athens Greece Valvular disease : Ο ρόλος του CMR Sophie Mavrogeni MD FESC Onassis Cardiac Surgery Center Athens Greece Aortic Valve CMR EVALUATION OF AORTIC VALVE Phase-contrast CMR of the aorta to determine aortic

More information

Conflict of interest: none declared

Conflict of interest: none declared The value of left ventricular global longitudinal strain assessed by three-dimensional strain imaging in the early detection of anthracycline-mediated cardiotoxicity C. Mornoş, A. Ionac, D. Cozma, S. Pescariu,

More information

Restrictive Cardiomyopathy

Restrictive Cardiomyopathy ESC Congress 2011, Paris Imaging Unusual Causes of Cardiomyopathy Restrictive Cardiomyopathy Kazuaki Tanabe, MD, PhD Professor of Medicine Chair, Division of Cardiology Izumo, Japan I Have No Disclosures

More information

Automatic functional analysis of left ventricle in cardiac cine MRI

Automatic functional analysis of left ventricle in cardiac cine MRI Original Article Automatic functional analysis of left ventricle in cardiac cine MRI Ying-Li Lu 1, Kim A. Connelly 2,3, Alexander J. Dick 4, Graham A. Wright 1,5, Perry E. Radau 1 1 Imaging Research, Sunnybrook

More information

Basic Assessment of Left Ventricular Systolic Function

Basic Assessment of Left Ventricular Systolic Function WINFOCUS BASIC ECHO (WBE) Basic Assessment of Left Ventricular Systolic Function Ritesh Dhar, MD Director, Echocardiography Lab and Staff Cardiologist Intermountain Medical Center Murray, Utah Outline

More information