Patient dose values during interventional cardiology examinations in Yazd hospital, Iran

Size: px
Start display at page:

Download "Patient dose values during interventional cardiology examinations in Yazd hospital, Iran"

Transcription

1 Patient dose values during interventional cardiology examinations in Yazd hospital, Iran F. Bouzarjomehri 1* and V. Tsapaki 2 Iran. J. Radiat. Res., 2009; 6 (4): Department of Medical Physics, Shahid Sadooghi University of Medical Sciences, Yazd, Iran 2 Department of Medical Physics, Konstantopoulio Hospital, Nea Ionia, Athens, Greece Background: The number of interventional cardiology (IC) procedures has increased rapidly. coronary angiography (CA) and percutaneous transluminal coronary angioplasty (PTCA) are now widely performed as a matter of routine, and they are considered safe procedures for experienced cardiologists. However, it is also known that these procedures are associated with high radiation doses due to long fluoroscopy time (T), and large number of cineradiography frames (F). These levels of radiation may even lead to radiation skin injuries under certain conditions. Materials and Methods: A detailed study of radiation doses received by 168 patients who underwent coronary angiography (CA), and 84 patients who underwent percutaneous transluminal coronary angioplasty (PTCA) using 3 angiography X-ray systems in two hospitals of Yazd-Iran is presented. An air kerma-area product (KAP) meter was used for patient dosimetry. KAP, fluoroscopy time and total number cine frames for CA and PTCA procedures were recorded for each patient. Results: Mean ± SD of KAP in CA and PTCA were 33 Gy.cm 2 ± 18.8 Gy.cm 2 and 80.3 Gy.cm 2 ± 65.6 Gy.cm 2 respectively. The comparison showed that CA KAP (33Gy.cm 2 ), fluoroscopy time (2.7±2.4min), and cine frames number (571±149) except of on case, were lower than (P<0.001) the results of other studies and mean KAP due to PTCA procedures, except for three cases, were not significatly different from the other references results. Conclusion: The high level expert cardiologists couldn't have a significant effect on the decrease of patient dose since they should also teach angiography examinations to medicine students. With increasing patient BMI the value of KAP increased, but the fluoroscopy time and cineframes number did not change significantly. In addition, the results showed that the use of flat panel detector was not sufficient for decreasing patient dose, and system's adjustment was more important. Iran. J. Radiat. Res., 2008; 6 (4): Keywords: IC, patient dose. KAP, CA, PTCA, Yazd. INTRODUCTION The number of interventional cardiology (IC) procedures has increased rapidly (1 3). Coronary Angiography (CA) and percutaneous transluminal coronary angioplasty (PTCA) procedures are now widely performed as a matter of routine and are considered safe procedures in the hand of experienced cardiologists (4). However, it is also known that these procedures are associated with high radiation doses due to long fluoroscopy time (T), and large numbers of cineradiography frames (F). These levels of radiation may even lead to skin injuries under certain conditions. Numerous incidents of radiation-induced skin injuries have recently been reported (5-8). Doses from the prolonged use of fluoroscopy can be very high and place the skin at risk for injury. United States Food and Drug Administration (FDA) has published recommendations on how to avoid these injuries (9). A number of studies (10 16) have investigated patient radiation doses in IC procedures, revealing variability not only in the methods of radiation measurement, but also in the level of radiation dose received by the patient. The complexity of the procedure (17), the experience of the operator (18), the level of training in radiation protection (3), and the type of X-ray equipment (19) are some of the factors responsible for various results. In Iran, such patient dose surveys in IC examinations are rarely performed. Very recently (in 2008), one patient dose study in Mashhad and another phantom study in Tabriz were performed (20, 21). Yazd hospitals treat cardiac diseases in more than one *Corresponding author: Dr. Fatollah Bouzarjomehri, Medical physics department, Shahid Sadooghi University of Medical Sciences, Yazd, Iran Fax: bouzarj_44@ssu.ac.ir

2 F. Bouzarjomehri and V. Tsapaki million Iranian people (Iran population is people). The current study has been the first patient dose monitoring of IC procedures in this city. As the need for continuously monitoring radiation dose in IC procedures has already been recognized in many European countries (22), the authors felt that similar studies have also been necessary in Middle East countries such as Iran. The main objectives of this work were therefore: (1) to investigate the level of knowledge of cardiologists on radiation protection in such techniques, and (2) to optimize practice so that radiation doses would be the lowest practically achievable, consistent with the clinical needs. More specifically, the subjects of this investigation included: Air kerma-area product (KAP) measurement of CA and PTCA procedures in 2 Yazd hospitals and comparison with the other studies in the recent literature. KAP comparison between IC procedures performed by flat-panel digital detector and conventional image intensifier machines. Investigation of the effects of patient body mass index (BMI) and physician skill on KAP. MATERIALS AND METHODS This study was performed in two general hospitals of Yazd, Iran. The first hospital had two angiography rooms: A-room with a General Electric angiography system (Advantx LC model, GE, USA) having an overcouch image intensifier (II) detector which was installed 15 years ago, and B-room with a Siemens system (AXIOM Artis model, Germany) having an overcouch flat panel detector (FD) that was installed 6 months ago. There was one cine mode in each machine, 25 frames/s, routinely used for adult patients. The second general hospital had one angiography room (C-room) with a General Electric angiography system (Advantx LC+DLX model) having an overcouch image intensifier detector, that was installed 7 years ago. All cardiologists used 25 cm field of view for all patients and magnification was seldom used. The detector (image intensifier or flat panel detector) was always placed as close to patient as possible. Patient radiation dose was measured by a calibrated KAP meter (PTW, Diamentor, Freiberg Germany) attached on the head of each X-ray tube. The device consisted of a large area ionization chamber (this was placed on the X-ray tube) and a control box for KAP measurement display. Data were for CA and PTCA. Patient clinical information was recorded manually in rooms A and C. Data included patient s sex, age, height, weight, type of procedure, KAP reading, fluoroscopy time, total number of cine frames, kv and name of cardiologist. Patient data in B-room were recorded automatically by machine. The cardiologists were divided into three groups depending on their experience. Level I included cardiologists with more than 10 years experience, level II cardiologists with 5 to 10 years experience and level III cardiologists with 1 to 5 years experience. RESULTS During the study, 252 IC procedures were performed by 12 cardiologists, 168 cases of CA (66.6%), and 84 cases of PTCA (33.4%). 55.7% of the patients sample were male and 44.3% were female patients. As may be deduced from figure 1, the highest percentage of IC procedures (40%) was performed in patients' age group of 50 to 60 years, followed by the 40 to 50 age group (30%). Radiation dose measurements in terms of KAP, fluoroscopy time (T) and total cine frame number (F) for CA and PTCA procedures in the three angiography rooms are given in table 1. As expected, PTCA presents higher values of KAP, T and F relative to CA. As shown in figure 2, KAP 168 Iran. J. Radiat. Res., Vol. 6, No. 4, Spring 2009

3 Patient dose during interventional cardiology values did not exhibit normal distribution, (KAP distribution slant to right hand), therefore apart from mean and standard deviation, median and 3 rd quartile values were also calculated for T, F and KAP both for the CA and PTCA as shown in table 1. Figure 1. The age distribution of patients whom referred to the Yazd hospitals for coronary angiography (CA) and percutaneous transluminal coronary angioplasty (PTCA) procedures is presented. Figure 2. Air Kerma-area Product (KAP) distribution in CA procedures (168 patients) is presented. Table 2, summarizes KAP, T and F in CA separately for the three angiography systems in A-room (GE Advantx LC machine), B-room (Siemens AXIOM Artis), and C-room (GE Adventx LC+DLX). For quality comparison of these X- ray systems, CA examination was used. As mentioned in materials and methods section, the two GE Table1. KAP results in Gy.cm 2, fluoroscopy time (T) in minutes and total number of frames (F) in coronary Angiography (CA) and percutaneous transluminal coronary angioplasty (PTCA) procedures are shown. Parameter Type of procedure Range Mean ± SD Median 3 rd quart KAP (Gy.cm 2 ) CA ( 168 cases) ± PTCA (84 cases) ± T (minutes) CA ± PTCA ± F CA ± PTCA ± Table 2. Mean and standard deviations of KAP, T and F are presented for the three X-ray machines (both General Electric systems had an image intensifier (II) detector and the Siemens machine had a flat panel detector (FD)) in CA. Cardiology Room *n KAP (Gy.cm 2 ) T (min) F A (GE Advantx LC) ± ± ± 150 B (Siemens AXIOM Artis) ± ± ± 138 C (GE Adventx LC+DLX ) ± ± ± 176 P values of C and B Room Vs A Room P (C&A) <0.001 P (B&A) <0.001 A&B: ns** P (A&C) <0.01 P ( B&C) <0.04 ns *n: number of CA examinations, **ns: not significant Iran. J. Radiat. Res., Vol. 6, No. 4, Spring

4 F. Bouzarjomehri and V. Tsapaki Adventx systems had an II detector, whereas the Siemens AXIOM Artis machine had a FD. The results showed that there were no statistically significant differences between KAP, F and T in II and FD machines. Table 3 shows the relationship of KAP with BMI. The results indicated that an increase of BMI increases patient KAP (P<0.01), whereas T and F values of level III and level I BMI did not significantly changed [P(I vs III)<0.7 and P(I vs III)<0.1 respectively]. The patient frequency of three groups consist of I<25 kg.m -2, 25<II<30 kg.m -2 and 30<III<35 kg.m -2 are 81, 112 and 39, respectively. Regression coefficients of KAP to body mass index (BMI) provided prediction of increasing radiation exposure by BMI (figure 3). (KAP) = (BMI) was linear regression function and coefficient correlation was r 2 =0.042, which is very weak. In table 4, mean ± SD of KAP, T and F for CA procedures are presented according to cardiologists' experience from C-room's data. The results showed that KAP, T and F values were not significantly different between I, II and III levels of cardiologists skill. Table 3. Mean ± SD of KAP, T and F values in CA procedures in three levels of mass index (BMI, Weight per tall 2 ). Level I: <25, Level II: 25-35, Level III: BMI (kg.m -2 ) Mean (BMI) N KAP(Gy.cm 2 ) T (min) F I: < ± ± ±150 25< II < ± ± ± < III < ± ± ± 181 P value BMI (III vs I) <0.01 <0.8 <0.1 BMI, body mass index; KAP: kerma area product, F: total cineframess, T: fluoroscopic time; N: number of patients Figure 3. Straight line regression of KAP (Gy.cm 2 ) to body mass index (BMI, kg/m 2 ). Regression equation: KAP= BMI, coefficient correlation: r 2 =0.042 Table 4. Mean and SD of KAP, T and F in CA procedures are presented according to cardiologist s experience only in B-Room. Level (I) indicate more than 10 years experience, level (II) between 5 to10 years experience and level (III) between 1 to5 years experience, N: the number of patients involved in this study for each experience level. Cardiologist N KAP (Gy.cm 2 ) T (min) F Level I ± ± ± 134 Level II ± ± ± 174 Level III ± ± ± Iran. J. Radiat. Res., Vol. 6, No. 4, Spring 2009

5 Patient dose during interventional cardiology DISCUSSION Although many patients derive great diagnostic and therapeutic benefit from IC procedures, the use of ionizing X-ray constitutes an associated hazard which must be justified by the procedure s benefits (23). Cardiac catheterizations are the highest patient radiation dose among the radiological X-ray procedures. In Yazd province, IC procedures have begun 15 years ago, and in the recent years an increase in the number of CA and PTCA techniques has been observed (4870 procedures in 2008 compared with 2505 procedures in 2007). In general, the justification of these procedures is evident, because complicated invasive surgery is usually avoided. However, the complexity of the procedures results in higher radiation exposures caused by longer irradiation times. The high patient doses and the introduction of new types of interventional procedures stress the need for an inventory of doses delivered to patients who undergo these high-dose X-ray examinations (24). Unfortunately in Yazd city, IC patient dose monitoring was not performed until now. It is evident that determination of patient dose and its effective parameters help to optimize IC techniques. Unlike the increasing KAP observed among overweight patients (table 3), the other parameters such as T and F almost remained constant. It could be concluded, therefore, that exposure parameters such as kv and ma for overweight patients were increased without any effect on T and F values. A comparison of this study s results with others found in the literature is shown in tables 5 and 6. The comparison showed that mean KAP in CA found in this study (except the result of Bahreyni et al. Iranian survey (20) ) was substantially lower than the other studies presented in table 5 (18, 4, 11-13). These results show that Yazd hemodynamic departments appear to be acceptable regarding radiation protection principals. A European survey was conducted by SENTINEL consortium to investigate doses in selected interventional cardiac procedures, and to establish updated reference levels (RLs). The survey involved nine European partners and near 2000 procedures were examined (22). RLs for T, F and KAP in CA are 6.5 min, 700 cine frames and 45 Gy.cm 2, respectively. Corresponding RLs in PTCA are 15.5 min, 1000 cine frames and 85 Gy.cm 2. Our results show that mean KAP (33Gy.cm 2 ), T (2.7min) and F (570) in CA procedure are lower than SENTINEL RLs. In PTCA, our values of KAP (90 Gy.cm 2 ), T (10.1min) and F (1057) are similar to SENTINEL RLs. KAP differences in CA procedures performed by three different skill level cardiologists were not statistically significant. This result may be expected in an educational hospital. In these hospitals, the most experienced cardiologists must teach fellows and consume more time during of CA or PTCA examinations than cardiologists in other general hospitals. The result of the present study also showed that KAP differences between II and FD machines were not statistically significant. It cleared that the regulation of imaging system was more important than the type of detector, so using flat panel detector was not certainly equal to decrease of patient dose. At last, there should be a concern about the decrease of cardiac patient age in Yazd; one reason may be the frequency of diabetic patients in Yazd. CONCLUSION The results of this study showed that patient doses in terms of KAP in CA examinations in 3 Yazd hospitals were lower than European RL and values found in other studies in the recent literature. Corresponding doses in PTCA procedures were similar to analogous studies. Proper quality control of the imaging system seems to be more important than the type of detector as initially mentioned by the manufacturers. Iran. J. Radiat. Res., Vol. 6, No. 4, Spring

6 F. Bouzarjomehri and V. Tsapaki Therefore, the use of FD isn't sufficient for decrease patient dose. The values of KAP by increase of BMI will be increased. ACKNOWLEDGEMENT The authors would like to acknowledge the cooperation of the all Afshar hospital cardiologists and the efforts of Hossein Karbalaeian and Manochehr Shareghzadeh in data acquisition. RREFERENCES 1. Sources and effects of ionizing radiation (2000) United Nations Scientific Committee on the Effects of Atomic Radiation. UNSCEAR 2000 Report to the General Assembly with Scientific Annexes. New York: United Nations. 2. Balter S and Shope TB (1995) Syllabus: A categorical course in physics: physical and technical aspects of angiography and interventional radiology. 81st Scientific Assembly, Dec RSNA 95 Scientific Program World Health Organization (2000) Efficacy and radiation safety in interventional radiology, Geneva: World Health Organization. 4. Padovani R, Novario R, Bernardim G (1998) Optimization in coronary angiography and percutaneous transluminal coronary angioplasty. Radiat Prot Dosim, 80: Kovoor P, Ricciardello M, Collins L, Uther JB, Ross DL (1998) Risk to patients from radiation associated with radiofrequency ablation for supraventricular tachycardia. Circulation, 15: Nahass GT (1997) acute radiodermatitis after radiofrequency catheter ablation. J Am Acad Dermatol, 5: Shope TB (1996) Radiation-induced skin injuries from fluoroscopy. Radiographics, 5: Knautz MA, Abele DC, Reynolds TL (1997) Radiodermatitis after transjugular intrahepatic portosystemic shunt. South Med J, 3: United States Food and Drug Administration (FDA) (1994) Avoidance of serious X-ray induced skin injuries to patients during fluoroscopically guided procedures. Medical Bulletin, 24: Kuon E, Glaser C, Dahm JB (2003) Effective techniques for reduction of radiation dosage to patient undergoing invasive cardiac procedures. The British J Radiology, 76: Vano E, Gonzalez L, Fernandez JM, Guibelalde E (1995) Patient dose values in interventional radiology. Br J Radiol, 68: Broadhead DA, Chapple CL, Faulkner K, Davies ML, McCallum H (1997) The impact of cardiology on the collective effective dose in the North of England. Br J Radiol, 70: Zorzetto M, Bernardi G, Morocutti G, Fontanelli A (1997) Radiation exposure to patients and operators during diagnostic catheterization and coronary angioplasty. Cathet Cardiovasc Diagn, 40: Widmark A, Fosmark H, Einarsson G, Gundtoft P, Hjardemaal O, Leitz W, Pukkila E (2001) Guidance levels in the Nordic Countries: A preliminary report for selected interventional procedures. Radiat Prot Dosim, 94: Cusma JT, Bell MR, Wondrow MA, Taubel JP, Holmes DR (1999) Real-time measurement of radiation exposure to patients during diagnostic coronary angiography and percutaneous interventional procedures. J Am Coll Cardiol, 33: Clark AL, Brennan AG, Robertson LJ, McArthur JD (2000) Factors affecting patient radiation exposure during routine coronary angiography in the tertiary referral centre. Br J Radiol, 73: Larrazet F, dibie A, Philippe F, palau R, Klauszand R, labored F (2003) Factors influencing fluoroscopy time and dose area product values during vessel percutaneous coronary angioplasty. British Journal Radiology, 76: Tsapaki V, Kottou S, Vano E, Faulkuer K, Giannouleas J, Padovni R, Kyrozi E, Koutelou M, Vardalaki E, Neofotistou V (2003) Patient dose values in a dedicated Greek cardiac centre. British Journal Radiology, 76: Suzuki S, Furui S, Kobayashi I, Yamauchi T, Kohtake T, Takeshita K, Takada K, Yamagishi M (2005) Radiation Dose to Patients and Radiologists During Transcatheter Arterial Embolization: Comparison of a Digital Flat- Panel System and Conventional Unit. Amer J Rontgen, 185: Bahreyni Toossi M T, Zare H, Bayani S, Esmaili S (2008) Organ and effective doses of patients arising from coronary angiography and percutaneous transluminal coronary angioplasty at two hospitals in Mashhad -Iran. Radiation Protection Dosimetry, 128: Mesbahi A, Mehnati P, Keshtkar A, Aslanabadi N (2008) Comparison of radiation dose to patient and staff for two interventional cardiology units: a phantom study. Radiation Protection Dosimetry, 11: Padovani R, Vano E, Trianni A, Bokou C, Bosmans H, Bor D, Jankowski J, Torbica P, Kepler K, Dowling A, Milu C, Tsapaki V, Salat D, Vassileva J, Faulkner K (2008) Reference levels at European level for cardiac interventional procedures. Radiation Protection Dosimetry, 129: Hirshfeld JW, Balter S, Brinker JA, Kern MJ, Klein LW, Lindsay BD, Tommaso CL, Tracy CM, Wagner LK (2005) Clinical Competence Statement on Physician Knowledge to Optimize Patient Safety and Image Quality in Fluoroscopically Guided Invasive Cardiovascular Procedures. Circulation, 111: Bacher K, Bogaert E Lapere R, Wolf D, Thierens H (2005) Patient-Specific Dose and Radiation Risk Estimation in Pediatric Cardiac Catheterization. Circulation, 111: Iran. J. Radiat. Res., Vol. 6, No. 4, Spring 2009

Research Journal of Pharmaceutical, Biological and Chemical Sciences

Research Journal of Pharmaceutical, Biological and Chemical Sciences Research Journal of Pharmaceutical, Biological and Chemical Sciences Evaluation of Patient Dose in Interventional Cardiology. Ayoub Momivand 1, Reza Zohdiaghdam 2*, Zhaleh Behrouzkia 1, and Ebrahim Khayati

More information

Radiation Protection Dosimetry Submitted Manuscript

Radiation Protection Dosimetry Submitted Manuscript Patient radiation doses in the most common interventional cardiology procedures in Croatia: first results Journal: Radiation Protection Dosimetry Manuscript ID: RPD-09-0121 Manuscript Type: Scientific

More information

A pilot programme on patient dosimetry on paediatric interventional cardiology in Chile

A pilot programme on patient dosimetry on paediatric interventional cardiology in Chile A pilot programme on patient dosimetry on paediatric interventional cardiology in Chile Patricia Miranda *1, Fernando Leyton 2, Eliseo Vano 3, Alfonso Espinoza 2, Raul Ramirez 4, Otto Delgado 2, Pedro

More information

Tyne NE4 6BE, UK. Corresponding author:

Tyne NE4 6BE, UK. Corresponding author: Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 54 58 doi:10.1093/rpd/nci719 Advance Access published on February 3, 2006 REFERENCE LEVELS IN PTCA AS A FUNCTION OF PROCEDURE COMPLEXITY A.

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Assessment of Patient Radiation Dose in Interventional Procedures at Shahid Madani Heart Center in Khorramabad, Iran Mehrdad Gholami 1*, Somayeh Gharloghi

More information

MEASUREMENT OF EFFECTIVE DOSE TO PATIENT DURING INTERVENTIONAL CARDIAC PROCEDURE

MEASUREMENT OF EFFECTIVE DOSE TO PATIENT DURING INTERVENTIONAL CARDIAC PROCEDURE J. Asiat. Soc. Bangladesh, Sci. 40(1): 1-7, June 2014 MEASUREMENT OF EFFECTIVE DOSE TO PATIENT DURING INTERVENTIONAL CARDIAC PROCEDURE M.M.M. SIRAZ 1, ALEYA BEGUM, R.K.KHAN, A.HOQUE AND A.BEGUM 2 Health

More information

Establishing the Greek National Reference Levels for Interventional Cardiology procedures

Establishing the Greek National Reference Levels for Interventional Cardiology procedures Establishing the Greek National Reference Levels for Interventional Cardiology procedures George Simantirakis*, Christina Koukorava, Maria Kalathaki, Christos Pa

More information

Radiation Protection Dosimetry Advance Access published February 17, Radiation Protection Dosimetry (2011), pp. 1 9

Radiation Protection Dosimetry Advance Access published February 17, Radiation Protection Dosimetry (2011), pp. 1 9 Radiation Protection Dosimetry Advance Access published February 17, 2011 Radiation Protection Dosimetry (2011), pp. 1 9 doi:10.1093/rpd/ncq600 AN AUDIT OF DIAGNOSTIC REFERENCE LEVELS IN INTERVENTIONAL

More information

Radiation Dose Reduction in the Invasive Cardiovascular Laboratory

Radiation Dose Reduction in the Invasive Cardiovascular Laboratory JACC: CARDIOVASCULAR INTERVENTIONS VOL. 5, NO. 8, 2012 2012 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION ISSN 1936-8798/$36.00 PUBLISHED BY ELSEVIER INC. http://dx.doi.org/10.1016/j.jcin.2012.05.003

More information

Clinical Determinants of Radiation Dose in Percutaneous Coronary Interventional Procedures

Clinical Determinants of Radiation Dose in Percutaneous Coronary Interventional Procedures JACC: CARDIOVASCULAR INTERVENTIONS VOL. 4, NO. 3, 2011 2011 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION ISSN 1936-8798/$36.00 PUBLISHED BY ELSEVIER INC. DOI: 10.1016/j.jcin.2010.10.014 Clinical Determinants

More information

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Radiation exposure of the Yazd population from medical conventional X-ray examinations Iran. J. Radiat. Res., 2007; 4 (4): 195-200 Radiation exposure of the Yazd population from medical conventional X-ray examinations F. Bouzarjomehri 1*, M.H. Dashti 2, M.H. Zare 1 1 Department of Medical

More information

Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures

Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures The British Journal of Radiology, 84 (2011), 70 77 Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures 1 E P EFSTATHOPOULOS, PhD, 1 I PANTOS,

More information

Relationship Between Fluoroscopic Time, Dose Area Product, Body Weight, and Maximum Radiation Skin Dose in Cardiac Interventional Procedures

Relationship Between Fluoroscopic Time, Dose Area Product, Body Weight, and Maximum Radiation Skin Dose in Cardiac Interventional Procedures Radiation Dose in Cardiac Intervention al Procedures Interventional Radiology Original Research A C M E D E N T U R I C A L I M A G I N G AJR 26; 186:774 778 361 83X/6/1863 774 American Roentgen Ray Society

More information

Evaluation of occupational radiation exposure of cardiologists in interventional radiography in Mashhad CATHLABs. Mohammad Taghi Bahreyni Toossi*

Evaluation of occupational radiation exposure of cardiologists in interventional radiography in Mashhad CATHLABs. Mohammad Taghi Bahreyni Toossi* 60 Int. J. Low Radiation, Vol. 9, No. 2, 203 Evaluation of occupational radiation exposure of cardiologists in interventional radiography in Mashhad CATHLABs Reza Fardid Department of Radiology, School

More information

Radiation Dose to the Pediatric Cardiac Catheterization and Intervention Patient

Radiation Dose to the Pediatric Cardiac Catheterization and Intervention Patient Medical Physics and Informatics Original Research Chida et al. Radiation Dose to Pediatric Patients Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 37.44.26.14 on

More information

Estimation of Organ and Effective Doses for Neonate and Infant Diagnostic Cardiac Catheterizations

Estimation of Organ and Effective Doses for Neonate and Infant Diagnostic Cardiac Catheterizations Medical Physics and Informatics Original Research Kawasaki et al. Dose Estimates for Neonate and Infant Diagnostic Cardiac Catheterization Medical Physics and Informatics Original Research Toshio Kawasaki

More information

Comparison of Five Developed Algorithms to Estimate Staff Effective Dose in Interventional Cardiology: Are They Interchangeable?

Comparison of Five Developed Algorithms to Estimate Staff Effective Dose in Interventional Cardiology: Are They Interchangeable? RESEARCH ARTICLE Copyright 2015 American Scientific Publishers All rights reserved Printed in the United States of America Journal of Medical Imaging and Health Informatics Vol. 5, 647 651, 2015 Comparison

More information

Does Ad Hoc Coronary Intervention Reduce Radiation Exposure? Analysis of 568 Patients

Does Ad Hoc Coronary Intervention Reduce Radiation Exposure? Analysis of 568 Patients Does Ad Hoc Coronary Intervention Reduce Radiation Exposure? Analysis of 568 Patients Márcio A. M. Truffa, Gustavo M.P. Alves, Fernando Bernardi, Antonio Esteves Filho, Expedito Ribeiro, Micheli Z. Galon,

More information

RADIATION INJURY OF THE SKIN FOLLOWING DIAGNOSTIC AND INTERVENTIONAL FLUOROSCOPIC PROCEDURES

RADIATION INJURY OF THE SKIN FOLLOWING DIAGNOSTIC AND INTERVENTIONAL FLUOROSCOPIC PROCEDURES XA0101657 RADIATION INJURY OF THE SKIN FOLLOWING DIAGNOSTIC AND INTERVENTIONAL FLUOROSCOPIC PROCEDURES T.R. Koenig, L.K. Wagner Department of Radiology, The University of Texas - Houston Medical School,

More information

OBJECTIVES BACKGROUND METHODS RESULTS CONCLUSIONS

OBJECTIVES BACKGROUND METHODS RESULTS CONCLUSIONS Journal of the American College of Cardiology Vol. 33, No. 2, 1999 1999 by the American College of Cardiology ISSN 0735-1097/99/$20.00 Published by Elsevier Science Inc. PII S0735-1097(98)00591-9 Real-time

More information

Radiation Protection Dosimetry Advance Access published April 22, 2008

Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 5 doi:10.1093/rpd/ncn144 EVALUATION OF EXPOSURE PARAMETERS IN PLAIN RADIOGRAPHY: A COMPARATIVE

More information

Radiation protection in the cardiac catheterization laboratory: special focus on the role of the operator

Radiation protection in the cardiac catheterization laboratory: special focus on the role of the operator special report Radiation protection in the cardiac catheterization laboratory: special focus on the role of the operator The use of ionizing radiation is associated with a risk of inducing malignant disease

More information

Patient Radiation Doses in Interventional Cardiology Procedures

Patient Radiation Doses in Interventional Cardiology Procedures Current Cardiology Reviews, 2009, 5, 111 1 Patient Radiation Doses in Interventional Cardiology Procedures Ioannis Pantos 1,2, Georgios Patatoukas 2, Demosthenes G. Katritsis 1 and Efstathios Efstathopoulos

More information

Why radiation protection matters?

Why radiation protection matters? Why radiation protection matters? Elias Brountzos Professor of Radiology 2 nd Department of Radiology Medical School, University of Athens Athens, Greece A definition for radiation protection Radiation

More information

Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis

Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis Europace (2006) 8, 443 448 doi:10.1093/europace/eul041 Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis Efstathios

More information

Article Excerpts: Radiation protection concerns among staff performing Fluoroscopic procedures.

Article Excerpts: Radiation protection concerns among staff performing Fluoroscopic procedures. Article Excerpts: Radiation protection concerns among staff performing Fluoroscopic procedures. E.P./Cath Lab Pain Management Radiology Contents: Interventional Radiology Carries Occupational Risk for

More information

A systematic review of the efficiency of radiation protection training in raising. University, Perth, Western Australia 6845, Australia.

A systematic review of the efficiency of radiation protection training in raising. University, Perth, Western Australia 6845, Australia. A systematic review of the efficiency of radiation protection training in raising awareness of medical staff working in catheterisation laboratory Mohammed Ali S Alahmari 1, Zhonghua Sun 1 1. Discipline

More information

International Journal of Radiation Research, January 2017

International Journal of Radiation Research, January 2017 Downloaded from ijrr.com at : +0330 on Tuesday November th Volume 5, No International Journal of Radiation Research, January 7 Radiation dose to the thyroid, eyes and parotid glands of patients undergoing

More information

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Conventional and spiral CT dose indices in Yazd general hospitals, Iran Iran. J. Radiat. Res., 2006; 3 (4): 183-189 Conventional and spiral CT dose indices in Yazd general hospitals, Iran F. Bouzarjomehri 1*,M.H.Zare 2, D. Shahbazi 2 1 Department of Medical Physics, Shahid

More information

A Survey of Radiation Exposure Techniques Factors Used for Common Diagnostics X- Ray Examinations in Akwa Ibom State, Nigeria

A Survey of Radiation Exposure Techniques Factors Used for Common Diagnostics X- Ray Examinations in Akwa Ibom State, Nigeria International Journal of Medical Imaging 2015; 3(4): 69-74 Published online June 6, 2015 (http://www.sciencepublishinggroup.com/j/ijmi) doi: 10.11648/j.ijmi.20150304.11 ISSN: 2330-8303 (Print); ISSN: 2330-832X

More information

Assessing the standard dose to standard patients for x-ray investigations

Assessing the standard dose to standard patients for x-ray investigations Assessing the standard dose to standard patients for x-ray investigations A. Almén and W. Leitz Swedish Radiation Protection Authority, SE-171 16 Stockholm, Sweden E-mail: anja.almen@ssi.se Abstract. The

More information

RADIATION DOSES TO PATIENTS UNDERGOING BARIUM MEAL AND BARIUM ENEMA EXAMINATIONS

RADIATION DOSES TO PATIENTS UNDERGOING BARIUM MEAL AND BARIUM ENEMA EXAMINATIONS Radiation Protection Dosimetry (2004), Vol. 109, No. 3, pp. 243±247 DOI: 10.1093/rpd/nch047 RADIATION DOSES TO PATIENTS UNDERGOING BARIUM MEAL AND BARIUM ENEMA EXAMINATIONS M. G. Delichas 1, K. Hatziioannou

More information

Ask EuroSafe Imaging Tips & Tricks. Interventional Radiology Working Group. How to control or reduce staff doses during IR procedures

Ask EuroSafe Imaging Tips & Tricks. Interventional Radiology Working Group. How to control or reduce staff doses during IR procedures Ask EuroSafe Imaging Tips & Tricks Interventional Radiology Working Group How to control or reduce staff doses during IR procedures Tommy Berglund (St. Olavs University Hospital, NO) Philipp Wiggermann

More information

Page 1 of Higashiibarakigun Oaraimachi/JP

Page 1 of Higashiibarakigun Oaraimachi/JP Study on accurate dose mapping system with radiophotoluminescence glass dosimeter (RPLD) measuring the direct patient entrance dose in cardiac interventional procedures. Poster No.: C-0687 Congress: ECR

More information

Patient s Radiation Exposure in Coronary Angiography and Angioplasty: The Impact of Different Projections

Patient s Radiation Exposure in Coronary Angiography and Angioplasty: The Impact of Different Projections doi:.15171/jcvtr.14. http://journals.tbzmed.ac.ir/jcvtr TUOMS Publishing Group Original Article Patient s Radiation Exposure in Coronary Angiography and Angioplasty: The Impact of Different Projections

More information

The print quality of this copy is not an accurate representation of the original.

The print quality of this copy is not an accurate representation of the original. Cardiovasc Intervent Radiol DOI 10.1007/s00270-010-9945-4 TECHNICAL NOTE Staff Radiation Doses in a Real-Time Display Inside the Angiography Room Roberto Sanchez E. Vano J. M. Fernandez J. J. Gallego Received:

More information

ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED *

ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED * ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED * Olivera Ciraj-Bjelac 1**, Danijela Arandjic 1, Predrag Bozovic 1, Sandra Ceklic 2, Jelena Stankovic

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

Survey of patients CT radiation dose in Jiangsu Province

Survey of patients CT radiation dose in Jiangsu Province Original Article Page 1 of 6 Survey of patients CT radiation dose in Jiangsu Province Yuanyuan Zhou 1, Chunyong Yang 1, Xingjiang Cao 1, Xiang Du 1, Ningle Yu 1, Xianfeng Zhou 2, Baoli Zhu 1, Jin Wang

More information

Coaching Reduced the Radiation Dose of Pain Physicians by Half during Interventional Procedures

Coaching Reduced the Radiation Dose of Pain Physicians by Half during Interventional Procedures REVIEW ARTICLE Coaching Reduced the Radiation Dose of Pain Physicians by Half during Interventional Procedures A. S. Slegers, MSc* ; I. G ultuna, MD ; J. A. Aukes, MD ; E. J. J. A. A. van Gorp, MD ; F.

More information

Application of international standards to diagnostic radiology dosimetry

Application of international standards to diagnostic radiology dosimetry Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-8 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry in Interventional Radiology Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

National diagnostic reference levels for fluoroscopy-guided procedures in Gastroenterology, Urology and Pediatric Cardiology

National diagnostic reference levels for fluoroscopy-guided procedures in Gastroenterology, Urology and Pediatric Cardiology National diagnostic reference levels for fluoroscopy-guided procedures in Gastroenterology, Urology and Pediatric Cardiology Eleni-Theano Samara, Abbas Aroua, Barbara Ott, Philipp R. Trueb and Francis

More information

Radiation Protection- Cath lab

Radiation Protection- Cath lab Radiation Protection- Cath lab Dr. Mawya A Khafaji Associate Prof. Medical Physics, Faculty of Medicine, KAU Head of Medical Physics Unit Dept. of Radiology -KAUH Head, Volunteer Office -KAUH Outline:

More information

The interdependence of staff and patient doses in interventional radiology

The interdependence of staff and patient doses in interventional radiology T he British Journal of Radiology, 70 (1997), 498 503 1997 The British Institute of Radiology The interdependence of staff and patient doses in interventional radiology J R WILLIAMS, MSc, FIPEM Department

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers Patient effective dose evaluation for chest X-ray examination in three digital radiography centers R. Paydar 1,3*, A. Takavar 1, M.R. Kardan 2,3, A. Babakhani 3,4, M.R. Deevband 3, S. Saber 5 1Medical

More information

Radiation Safety in the Catheterization Lab

Radiation Safety in the Catheterization Lab SCAI FALL FELLOWS COURSE - 2015 Radiation Safety in the Catheterization Lab V. Vivian Dimas, MD, FSCAI Associate Professor Pediatrics, Cardiology UT Southwestern Medical Center Dallas TX None Disclosures

More information

Quality Control Status of Radiology Centers of Hospitals Associated with Mashhad University of Medical Sciences

Quality Control Status of Radiology Centers of Hospitals Associated with Mashhad University of Medical Sciences Iranian Journal of Medical Physics Vol. 10, No. 4, Autumn 2013 & Vol. 11, No. 1, Winter 2014, 182-187 Received: December 2, 2013; Accepted: May 25, 2013 Original Article Quality Control Status of Radiology

More information

Beam projections and radiation exposure in transradial and transfemoral approaches during coronary angiography

Beam projections and radiation exposure in transradial and transfemoral approaches during coronary angiography 298 Original Investigation Beam projections and radiation exposure in transradial and transfemoral approaches during coronary angiography Ali Tarighatnia *,**,1, Amirhossien Mohammadalian 1, Morteza Ghojazade

More information

Patient dose in routine X-ray examinations in Yazd state

Patient dose in routine X-ray examinations in Yazd state Iran. J. Radiat. Res., 2004; 1(4): 199-204 Patient dose in routine X-ray examinations in Yazd state F. Bouzarjomehri Health Physics Dept. Shahid Sadoghi Univ. of Medical Sciences, Yazd, Iran ABSTRACT Background:

More information

Evaluation of radiation dose to patients undergoing interventional radiology procedures at Ramathibodi Hospital, Thailand

Evaluation of radiation dose to patients undergoing interventional radiology procedures at Ramathibodi Hospital, Thailand Available online at http://www.biij.org/2011/3/e22 doi: 10.2349/biij.7.3.e22 biij Biomedical Imaging and Intervention Journal ORIGINAL ARTICLE Evaluation of radiation dose to patients undergoing interventional

More information

Patient doses from X-ray computed tomography examinations by a single-array detector unit: Axial versus spiral mode

Patient doses from X-ray computed tomography examinations by a single-array detector unit: Axial versus spiral mode Patient doses from X-ray computed tomography examinations by a single-array detector unit: Axial versus spiral mode S.M. Ghavami 1, A. Mesbahi 2,3*, I. Pesianian 1 Iran. J. Radiat. Res., 2012; 10(2): 89-94

More information

Patient Dosimetry in Interventional Procedures: what to measure and estimate, why and how?

Patient Dosimetry in Interventional Procedures: what to measure and estimate, why and how? ICTP/IAEA Training Course on Radiation Protection of Patients 16-27 September 2013, Trieste, Italy Patient Dosimetry in Interventional Procedures: what to measure and estimate, why and how? Renato Padovani

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran Global Journal of Health Science; Vol. 7, No. 5; 2015 ISSN 1916-9736 E-ISSN 1916-9744 Published by Canadian Center of Science and Education Measurement of Entrance Skin Dose and Calculation of Effective

More information

The Use of Transoesophageal Echocardiography for Guiding Left Atrial Appendage Occlusion: Impact on Radiation Exposure

The Use of Transoesophageal Echocardiography for Guiding Left Atrial Appendage Occlusion: Impact on Radiation Exposure The Use of Transoesophageal Echocardiography for Guiding Left Atrial Appendage Occlusion: Impact on Radiation Exposure Abstract The aim of this study was primarily to determine if left atrial appendage

More information

Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks

Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks Poster No.: C-2993 Congress: ECR 2010 Type: Topic: Scientific Exhibit

More information

Effective dose conversion factors in paediatric interventional cardiology

Effective dose conversion factors in paediatric interventional cardiology The British Journal of Radiology, 82 (2009), 748 755 Effective dose conversion factors in paediatric interventional cardiology 1 A KARAMBATSAKIDOU, MSc, 2 B SAHLGREN, MD, PhD, 3 B HANSSON, MSc, 4 M LIDEGRAN,

More information

Radiation Protection Dosimetry Advance Access published September 13, 2006

Radiation Protection Dosimetry Advance Access published September 13, 2006 Radiation Protection Dosimetry Advance Access published September 13, 26 Radiation Protection Dosimetry (26), 1 of 6 doi:1.193/rpd/ncl113 TECHNICAL NOTE RADIATION EXPOSURE AND DOSE EVALUATION IN INTRAORAL

More information

The interventional cardiology team who gets the most exposure in percutaneous recanalisation of chronic total occlusion (CTO) procedures?

The interventional cardiology team who gets the most exposure in percutaneous recanalisation of chronic total occlusion (CTO) procedures? The interventional cardiology team who gets the most exposure in percutaneous recanalisation of chronic total occlusion (CTO) procedures? L. Price, A. Pascoal Proteção Radiológica na Saúde, 18-20 Sep 2013,

More information

Chronic Radiodermatitis Following Percutaneous Coronary Interventions

Chronic Radiodermatitis Following Percutaneous Coronary Interventions Chronic Radiodermatitis Following Percutaneous Coronary Interventions Jung-Yi Chan Chia-Yu Chu The incidence of radiodermatitis after percutaneous coronary interventions is rising with the increasing number

More information

Procedural radiation exposure of interventional cardiologists and radiologists Kuipers, G.

Procedural radiation exposure of interventional cardiologists and radiologists Kuipers, G. UvA-DARE (Digital Academic Repository) Procedural radiation exposure of interventional cardiologists and radiologists Kuipers, G. Link to publication Citation for published version (APA): Kuipers, G. (2011).

More information

Interventional radiology, by virtue of its low invasiveness, is

Interventional radiology, by virtue of its low invasiveness, is ORIGINAL RESEARCH T. Moritake Y. Matsumaru T. Takigawa K. Nishizawa A. Matsumura K. Tsuboi Dose Measurement on Both Patients and Operators during Neurointerventional Procedures Using Photoluminescence

More information

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology Iran. J. Radiat. Res., 2006; 4 (1): 21-27 Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology M.T. Bahreyni

More information

U/S Guided Interventions for Complex Lower Extremity Disease

U/S Guided Interventions for Complex Lower Extremity Disease U/S Guided Interventions for Complex Lower Extremity Disease A Vascular Surgeons Effort to Significantly Reduce Radiation Exposure Bryan T. Fisher Sr., MD January 23, 2017 Chief of Vascular Surgery and

More information

available at journal homepage:

available at  journal homepage: Physica Medica (2012) 28, 61e70 available at www.sciencedirect.com journal homepage: http://intl.elsevierhealth.com/journals/ejmp TECHNICAL NOTE Proposed local diagnostic reference levels in angiography

More information

Asymmetric breast dose in coronary angiography

Asymmetric breast dose in coronary angiography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Asymmetric breast dose in coronary angiography Kam Shan Mong, 1,3a Anthony B. Wallace, 2 and Rick D. Franich 3 Medical Physics Department,

More information

Research Article. Corresponding author H. Osman

Research Article. Corresponding author H. Osman Scholars Journal of Applied Medical Sciences (SJAMS) ISSN 2320-6691 Sch. J. App. Med. Sci., 2013; 1(5):511-515 Scholars Academic and Scientific Publisher (An International Publisher for Academic and Scientific

More information

STONY BROOK UNIVERSITY HOSPITAL VASCULAR CENTER CREDENTIALING POLICY BROCHURE

STONY BROOK UNIVERSITY HOSPITAL VASCULAR CENTER CREDENTIALING POLICY BROCHURE STONY BROOK UNIVERSITY HOSPITAL VASCULAR CENTER CREDENTIALING POLICY BROCHURE Per Medical Board decision March 18, 2008: These credentialing standards do NOT apply to peripheral angiography performed in

More information

CT Quality Control Manual FAQs

CT Quality Control Manual FAQs CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues

More information

Medical Fluoroscopic Exposures and Incidents

Medical Fluoroscopic Exposures and Incidents Medical Fluoroscopic Exposures and Incidents FCHPS Spring 2014 Meeting Orlando, FL. April 11, 2014 Kevin Nelson, Ph.D., CHP Radiation Safety Officer Mayo Clinic, Jacksonville, FL nelson.kevin2@mayo.edu

More information

THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES

THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES IJRRAS 9 (1) October 11 www.arpapress.com/volumes/vol9issue1/ijrras_9_1_16.pdf THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES Bamidele

More information

Interventional Fluoroscopy Patient and Staff Safety. Procedures. Safety objectives. Medical physics tasks. Comparative patient risks.

Interventional Fluoroscopy Patient and Staff Safety. Procedures. Safety objectives. Medical physics tasks. Comparative patient risks. Interventional Fluoroscopy Patient and Staff Safety Stephen Balter, Ph.D. Columbia University Medical Center Presented at AAPM Annual Meeting Minneapolis - 2007 Procedures Numbers of procedures will continue

More information

JOURNAL OF VASCULAR SURGERY Volume 49, Number 6 Bannazadeh et al 1521 Table I. Units of radiation Quantity Unit Definition Exposure R The total charge

JOURNAL OF VASCULAR SURGERY Volume 49, Number 6 Bannazadeh et al 1521 Table I. Units of radiation Quantity Unit Definition Exposure R The total charge Patterns of procedure-specific radiation exposure in the endovascular era: Impetus for further innovation Mohsen Bannazadeh, MD, Ozcan Altinel, MD, Vikram S. Kashyap, MD, Zhiyuan Sun, MS, Daniel Clair,

More information

Diagnostic reference levels in intraoral dental radiography in Korea

Diagnostic reference levels in intraoral dental radiography in Korea Imaging Science in Dentistry 2012; 42 : 237-42 http://dx.doi.org/10.5624/isd.2012.42.4.237 Diagnostic reference levels in intraoral dental radiography in Korea Eun-Kyung Kim, Won-Jeong Han, Jin-Woo Choi,

More information

Interprovincial Billing Out-Patient Rates Effective for Visits on or After April 1, 2018

Interprovincial Billing Out-Patient Rates Effective for Visits on or After April 1, 2018 Service Code Interprovincial Billing Out-Patient Rates Effective for Visits on or After April 1, 2018 Description Rate ($) 01 Standard Out-patient Visit, including select discrete high cost diagnostic

More information

Chronic radiodermatitis following cardiac catheterisation: a report of two cases and a brief review of the literature

Chronic radiodermatitis following cardiac catheterisation: a report of two cases and a brief review of the literature 308 Heart 1999;81:308 312 CASE STUDY Skin Research Institute and the Department of Pathology, Saint Louis University Hospital, Paris, France L Dehen C Vilmer C Humilière L Ollivaud D Chatelain L Dubertret

More information

Influence of the exposure time in the area monitors at radiodiagnostic

Influence of the exposure time in the area monitors at radiodiagnostic Journal of Physics: Conference Series PAPER OPEN ACCESS Influence of the exposure time in the area monitors at radiodiagnostic To cite this article: F Leyton et al 2016 J. Phys.: Conf. Ser. 733 012087

More information

Document Control Report

Document Control Report Document Control Report Title Recording Clinical Evaluation and Dose - IR(ME)R 2000 Standard Operating Procedure Author Gill Kite Author s job title Senior Radiographer Directorate Sub-directorate Department

More information

Learning Curve for Transradial and Transfemoral Coronary Angiography amongst Cardiology Trainees

Learning Curve for Transradial and Transfemoral Coronary Angiography amongst Cardiology Trainees Research Article imedpub Journals www.imedpub.com Interventional Cardiology Journal ISSN 2471-8157 DOI: 10.21767/2471-8157.100071 Abstract Learning Curve for Transradial and Transfemoral Coronary Angiography

More information

Coronary angiography and PCI

Coronary angiography and PCI Coronary arteries Coronary angiography and PCI Samo Granda, Franjo Naji Department of Cardiology Clinical department of internal medicine University clinical centre Maribor Coronary arteries Atherosclerosis

More information

Doses Received by Patients during Thorax X-Ray Examinations

Doses Received by Patients during Thorax X-Ray Examinations Iranian Journal of Medical Physics Vol. 9, No. 4, Autumn 01, 45-51 Received: September 16, 01; Accepted: February 0, 013 Original Article Doses Received by Patients during Thorax X-Ray Examinations Nsikan

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Dose Evaluation for Common Digital Radiographic s in Selected Hospitals in Pahang Malaysia Soo-Foon Moey 1*, Zubir Ahmad Shazli 1, Inayatullah Shah Sayed

More information

Quality Control and Patient Dosimetry on line for Computed Tomography

Quality Control and Patient Dosimetry on line for Computed Tomography Quality Control and Patient Dosimetry on line for Computed Tomography Jose I. Ten 1,2, Eliseo Vano 2,3, Jose M. Fernandez-Soto 2,3, Roberto Sanchez 3, Juan Arrazola 1,2 1 Diagnostic Radiology Service and

More information

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan *M.T.Bahreyni Toossi, **S.Mohandes Dastgherdi Medical Physics Dep., Faculty

More information

2010 Canadian Cardiovascular Society/ Canadian Heart Rhythm Society Training and Maintenance of Competency in Adult Clinical Cardiac

2010 Canadian Cardiovascular Society/ Canadian Heart Rhythm Society Training and Maintenance of Competency in Adult Clinical Cardiac 2010 Canadian Cardiovascular Society/ Canadian Heart Rhythm Society Training and Maintenance of Competency in Adult Clinical Cardiac Electrophysiology Martin S. Green, Chair, CHRS Education Committee Peter

More information

Reduction of Radiation Exposure in Percutaneous Pedicle Screw Placement by Using a New Extension Instrument

Reduction of Radiation Exposure in Percutaneous Pedicle Screw Placement by Using a New Extension Instrument Reduction of Radiation Exposure in Percutaneous Pedicle Screw Placement by Using a New Extension Instrument Franquiz M. Juan, Myers J.Behnam, Thomas A. Michael, Greenhalgh J. Travis, Shoup J. Andrew Study

More information

Managing Patient Dose in Computed Tomography (CT)

Managing Patient Dose in Computed Tomography (CT) Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

Interprovincial Billing Out-Patient Rates Effective for Visits on or After April 1, 2017

Interprovincial Billing Out-Patient Rates Effective for Visits on or After April 1, 2017 Service Code Interprovincial Billing Out-Patient Rates Effective for Visits on or After April 1, 2017 Description Rate ($) 01 Standard Out-patient Visit, including select discrete high cost diagnostic

More information

Chapter 2 How Radiation Protection Influences Quality in Radiology

Chapter 2 How Radiation Protection Influences Quality in Radiology Chapter 2 How Radiation Protection Influences Quality in Radiology Eliseo Vano, Kwan-Hoong Ng, and Lawrence Lau Abstract Radiation safety is a key quality element in medical imaging and interventional

More information

Joint Working Group to produce guidance on delivering an Endovascular Aneurysm Repair (EVAR) Service.

Joint Working Group to produce guidance on delivering an Endovascular Aneurysm Repair (EVAR) Service. Joint Working Group to produce guidance on delivering an Endovascular Aneurysm Repair (EVAR) Service. Royal College of Radiologists British Society of Interventional Radiology The Vascular Society of Great

More information

Assessment of Patient Dose from CT Examinations in Khorasan, Iran

Assessment of Patient Dose from CT Examinations in Khorasan, Iran Iranian Journal of Medical Physics Vol. 9, No. 4, Autumn 2012, 233-238 Received: July 18, 2012; Accepted: November 17, 2012 Original Article Assessment of Patient Dose from CT Examinations in Khorasan,

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

2017 Course of the Nordic Association for Clinical Physics on occupational dosimetry in hospitals

2017 Course of the Nordic Association for Clinical Physics on occupational dosimetry in hospitals 2017 Course of the Nordic Association for Clinical Physics on occupational dosimetry in hospitals 27 September 2017 Dr. Pedro Ortiz López Retired member of the ICRP and the IAEA 2 Pedro Ortiz López (Chair)

More information

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: Rob Loader Oliver Gosling Introduction Approached by Dr Oliver Gosling (Research

More information

Horizon Scanning Technology Summary. Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease

Horizon Scanning Technology Summary. Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease Horizon Scanning Technology Summary National Horizon Scanning Centre Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease April 2007 This technology summary is based

More information

Radiographic techniques in screen-film mammography

Radiographic techniques in screen-film mammography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 3, SUMMER 2002 Radiographic techniques in screen-film mammography Thomas R. LaVoy* 4314 Kelsey Drive, Syracuse, New York 13215 Walter Huda

More information

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS RADIOPROTECTION AND DOSIMETRY MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS DANIELA ADAM 1, ANA STOCHIOIU

More information

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques

More information