Establishing the Greek National Reference Levels for Interventional Cardiology procedures

Size: px
Start display at page:

Download "Establishing the Greek National Reference Levels for Interventional Cardiology procedures"

Transcription

1 Establishing the Greek National Reference Levels for Interventional Cardiology procedures George Simantirakis*, Christina Koukorava, Maria Kalathaki, Christos Pa<ilis, Ioannis Kaisas, Sotirios Economides, Costas J. Hourdakis and Vassiliki Kamenopoulou Greek Atomic Energy Commission * T: , F: , E: gsim@eeae.gr

2 Contents of the presentation Introduction Methods Measurements related to the performance of the X- ray systems Reference Levels related to the clinical procedures Results Discussion 2

3 Introduction The frequency of <luoroscopically guided Interventional Cardiology (IC) procedures has seen signi<icant growth over the past decade. The extended use of these procedures raises various radiation protection issues, as they result in high radiation dose to the patient as well as to the medical personnel. Speci<ically the dose to the patients skin can, in some cases, even exceed the threshold of tissue reactions. As a consequence, optimisation of radiation protection on IC procedures is of great importance. Vliestra et al. (2004) 3

4 Introduction Directive 97/43/EURATOM introduced the concept of Diagnostic Reference Levels as a tool for optimisation for typical examinations. They are de<ined as dose levels in medical radiodiagnostic practices to patients of standard- sized groups or standard phantoms. Although IC procedures can hardly be standardized because of the variations of patients anatomy and pathology, the adoption of Reference Levels (RLs) for these procedures seems feasible and has been investigated in the literature. The establishment of RLs in image guided interventional procedures at a national level is required in the recent revision of the BSS*. * ΙΑΕΑ (2011) Radiation Protection and Safety of Radiation Sources: International Basic Safety Standards Interim Edition, Safety Series No. GSR Part 3 (Interim), IAEA, Vienna 4

5 Introduction The purpose of this speech is to present the results of a national survey that was performed in Greece for the establishment of national RLs for the most common IC procedures. 5

6 Methods Two types of Reference Levels were set, as suggested in the literature*: RLs related to the performance of the X- ray systems used in IC, RLs related to the clinical procedures. For this purpose, the study included collection and analysis of clinical data as well as measurements related to the performance of the X- ray systems used in IC. * Vano E, Gonzalez L (2001) Approaches to establishing reference levels in interventional radiology. Radiat Prot Dosimetry 94: * Padovani R et al. (2008) Reference levels at European level for cardiac interventional procedures. Radiat Prot Dosimetry 129:

7 Measurements related to the performance of the X- ray systems The performance of 30 systems that are used in IC was evaluated in respect to the following parameters: Phantom entrance dose rate during <luoroscopy Phantom entrance dose rate and dose per frame during image acquisition (CINE) Air kerma- area product (P KA ) meter accuracy Image quality [Low Contrast Sensitivity (LCS) and High Contrast Resolution (HCR)] 7

8 Measurements related to the performance of the X- ray systems Dose measurements* 20 cm of water 60 cm 90 cm Typical PA projection Measurements were performed without magni<ication(20-25 cm FoV) and with magni<ication to cm FoV. All exposure parameters (<luoroscopy and CINE mode, pulse and frame rate, <iltration etc.) during the measurements were those that are used in typical clinical cases on each system. The accuracy of the P KA meters was evaluated using the beam area method (i.e. by measuring the air kerma and the <ield area at the plane of measurement) without the presence of the patient table (free- in- air). * International Atomic Energy Agency (2007) Dosimetry in Diagnostic Radiology: An International Code of Practice. Technical Reports Series 457, IAEA, Vienna 8

9 RLs related to the clinical procedures The total <luoroscopy time and the P KA of a total of 5250 IC procedures from 26 hospitals were recorded and analyzed. Classi<ication of procedures: i) diagnostic Coronary Angiography (CA), ii) Percutaneous Coronary Intervention (PCI): any therapeutic procedure (with or without a diagnostic component)* iii) Pacemaker Implantation (PMI) iv) Radiofrequency Cardiac Ablation (RFCA). *ΙΑΕΑ (2009) Establishing Guidance Levels in X Ray Guided Medical Interventional Procedures: A Pilot Study. Safety Reports Series No. 59, IAEA, Vienna. 9

10 RLs related to the clinical procedures The P KA meter accuracy that was assessed during the systems performance evaluation, was used in order to investigate the need for applying corrective factors to the collected data. The minimum sample size for each procedure was 30 patients per hospital. For each type of procedure, the mean <luoroscopy time and P KA values were calculated for each hospital. The rounded 3 rd quartile of the mean values was set as the RL for each procedure.* *Hart D, Hillier MC, Wall BF (2002), Doses to patients from medical X-ray Examinations in the UK 2000 Review, NRPB-W14 edition, National Radiation Protection Board, Chilton. 10

11 Results Systems measurements 120,00 100,00 Fluoroscopy - Phantom entrance dose rate FoV Signi<icant differences 1/ FoV on the dose rates and dose per frame values were observed, even for systems of the same type: 80,00 mgy/min 60,00 40,00 20,00 0,00 A1 A2 B1 B2 C1 C2 C3 C4 C5 C6 C7 C8 D1 D2 E1 F1 G1 H1 H2 H3 H4 I1 I2 I3 K1 L1 L2 L3 L4 M1 Each letter on the horizontal axis represents a specific system model 11

12 Results Systems measurements CINE - Phantom entrance dose per frame 1/ FoV FoV μgy/fr A1 A2 B1 B2 C1 C2 C3 C4 C5 C6 C7 C8 D1 D2 E1 F1 G1 H1 H2 H3 H4 I1 I2 I3 K1 L1 L2 L3 L4 M1 Each letter on the horizontal axis represents a specific system model 12

13 Results Systems measurements The accuracy of all P KA meters lied within the ± 35% tolerance* Eight systems presented inaccuracies greater than the estimated uncertainty of the P KA measurement (6.4%). A respective correction factor was applied to the clinical P KA data that was collected from these systems. Flat Detector systems presented better high contrast resolution against those with Image Intensi<iers (mean values were 2.4 lp/mm and 1.5 lp/mm respectively). No signi<icant differences were observed for the rest of the measured parameters. * International Electrotechnical Commission (2010) Medical electrical equipment - Part 2 43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures, IEC Report No Ed. 2.0, Geneva. 13

14 Results Clinical data PCI - Mean DAP values per center PCI and RFCA were the procedures with the highest P KA mean values (112.0 mgy cm 2 and 98.3 mgy cm 2 respectively). 200 Gy cm

15 Results Clinical data 250 RFCA - Mean DAP values per center 200 Gy cm pps fluoroscopy (provides adequate image quality for RFCA) 15

16 Suggested national RLs Other studies 1.CA: 50 Gycm 2, PCI: 125 Gycm CA: Gycm 2 PTCA: Gycm PMI: 38 Gycm 2, 27 Gycm 2 1. ΙΑΕΑ (2009) Establishing Guidance Levels in X Ray Guided Medical Interventional Procedures: A Pilot Study. Safety Reports Series No. 59, IAEA, Vienna. 2. Neofotistou V et al. (2003) Preliminary reference levels in interventional cardiology. Eur Radiol 13: Padovani R et al. (2008) Reference levels at European level for cardiac interventional procedures. Radiat Prot Dosimetry 129: Aroua A et al. (2004) Adult reference levels in diagnostic and interventional radiology for temporary use in Switzerland. Radiat Prot Dosimetry 111: Hart D et al. (2002), Doses to patients from medical X- ray Examinations in the UK 2000 Review, NRPB- W14 edition, National Radiation Protection Board, Chilton. 16

17 Discussion The wide range of the dose rate and dose per frame values of the systems that were evaluated, represent differences not only in the systems dosimetric performance, but also in the protocols used by the operating staff during the clinical work. Optimisation by changing the systems settings used during the clinical practice seems feasible in some cases. Scaling of the RLs according to the complexity of the procedures was not feasible during the present study (no such information was available in the collected data). Each individual facility should compare its own values against the suggested RLs, after applying an adjustment for complexity to the local data. 17

18 Procedure complexity PTCA Complexity Index (CI)* CI = No.ves 1 + No.LesType No.Occl3m No.SevTort No.BifSt 0.58 (a) Simple PTCA procedures: CI = 1; (b) Medium complex PTCA procedures: 1 < CI < 2; (c) Complex PTCA procedures: CI > 2. *ΙΑΕΑ (2009) Establishing Guidance Levels in X Ray Guided Medical Interventional Procedures: A Pilot Study. Safety Reports Series No. 59, IAEA, Vienna. PMI: type of implanted device (conventional or biventricular) RFCA: type of treated arrhythmia Padovani R et al. (2008) Reference levels at European level for cardiac interventional procedures. Radiat Prot Dosimetry 129:

19 Discussion Future re- evaluation of the suggested RLs should take into account The complexity of the procedures The number of acquired frames per procedure The patient s morphometric characteristics (e.g. Body Mass Index) More information: Simantirakis G., Koukorava C., Kalathaki M., Pa<ilis C., Kaisas I., Economides S., Hourdakis C. J., Kamenopoulou V., Georgiou E. "Reference levels and patient doses in interventional cardiology procedures in Greece", Eur Radiol (2013) 23:

20 Thank you Greek Atomic Energy Commission P.O Box Agia Paraskevi 15310, Athens T: F: E: facebook.com/eeaegr twitter.com/#eeaegr 20

Radiation Protection Dosimetry Advance Access published February 17, Radiation Protection Dosimetry (2011), pp. 1 9

Radiation Protection Dosimetry Advance Access published February 17, Radiation Protection Dosimetry (2011), pp. 1 9 Radiation Protection Dosimetry Advance Access published February 17, 2011 Radiation Protection Dosimetry (2011), pp. 1 9 doi:10.1093/rpd/ncq600 AN AUDIT OF DIAGNOSTIC REFERENCE LEVELS IN INTERVENTIONAL

More information

A pilot programme on patient dosimetry on paediatric interventional cardiology in Chile

A pilot programme on patient dosimetry on paediatric interventional cardiology in Chile A pilot programme on patient dosimetry on paediatric interventional cardiology in Chile Patricia Miranda *1, Fernando Leyton 2, Eliseo Vano 3, Alfonso Espinoza 2, Raul Ramirez 4, Otto Delgado 2, Pedro

More information

Application of international standards to diagnostic radiology dosimetry

Application of international standards to diagnostic radiology dosimetry Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-8 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry in Interventional Radiology Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS RADIOPROTECTION AND DOSIMETRY MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS DANIELA ADAM 1, ANA STOCHIOIU

More information

Why radiation protection matters?

Why radiation protection matters? Why radiation protection matters? Elias Brountzos Professor of Radiology 2 nd Department of Radiology Medical School, University of Athens Athens, Greece A definition for radiation protection Radiation

More information

Assessing the standard dose to standard patients for x-ray investigations

Assessing the standard dose to standard patients for x-ray investigations Assessing the standard dose to standard patients for x-ray investigations A. Almén and W. Leitz Swedish Radiation Protection Authority, SE-171 16 Stockholm, Sweden E-mail: anja.almen@ssi.se Abstract. The

More information

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Investigation of the clinical performance of a novel solid-state diagnostic dosimeter Jason Tse, a Donald McLean Medical Physics and

More information

Patient dose values during interventional cardiology examinations in Yazd hospital, Iran

Patient dose values during interventional cardiology examinations in Yazd hospital, Iran Patient dose values during interventional cardiology examinations in Yazd hospital, Iran F. Bouzarjomehri 1* and V. Tsapaki 2 Iran. J. Radiat. Res., 2009; 6 (4): 167-172 1 Department of Medical Physics,

More information

National diagnostic reference levels for fluoroscopy-guided procedures in Gastroenterology, Urology and Pediatric Cardiology

National diagnostic reference levels for fluoroscopy-guided procedures in Gastroenterology, Urology and Pediatric Cardiology National diagnostic reference levels for fluoroscopy-guided procedures in Gastroenterology, Urology and Pediatric Cardiology Eleni-Theano Samara, Abbas Aroua, Barbara Ott, Philipp R. Trueb and Francis

More information

Radiation Protection- Cath lab

Radiation Protection- Cath lab Radiation Protection- Cath lab Dr. Mawya A Khafaji Associate Prof. Medical Physics, Faculty of Medicine, KAU Head of Medical Physics Unit Dept. of Radiology -KAUH Head, Volunteer Office -KAUH Outline:

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Assessment of Patient Radiation Dose in Interventional Procedures at Shahid Madani Heart Center in Khorramabad, Iran Mehrdad Gholami 1*, Somayeh Gharloghi

More information

Ask EuroSafe Imaging Tips & Tricks. Interventional Radiology Working Group. How to control or reduce staff doses during IR procedures

Ask EuroSafe Imaging Tips & Tricks. Interventional Radiology Working Group. How to control or reduce staff doses during IR procedures Ask EuroSafe Imaging Tips & Tricks Interventional Radiology Working Group How to control or reduce staff doses during IR procedures Tommy Berglund (St. Olavs University Hospital, NO) Philipp Wiggermann

More information

Radiation Protection Dosimetry Advance Access published September 13, 2006

Radiation Protection Dosimetry Advance Access published September 13, 2006 Radiation Protection Dosimetry Advance Access published September 13, 26 Radiation Protection Dosimetry (26), 1 of 6 doi:1.193/rpd/ncl113 TECHNICAL NOTE RADIATION EXPOSURE AND DOSE EVALUATION IN INTRAORAL

More information

Research Journal of Pharmaceutical, Biological and Chemical Sciences

Research Journal of Pharmaceutical, Biological and Chemical Sciences Research Journal of Pharmaceutical, Biological and Chemical Sciences Evaluation of Patient Dose in Interventional Cardiology. Ayoub Momivand 1, Reza Zohdiaghdam 2*, Zhaleh Behrouzkia 1, and Ebrahim Khayati

More information

The Use of Transoesophageal Echocardiography for Guiding Left Atrial Appendage Occlusion: Impact on Radiation Exposure

The Use of Transoesophageal Echocardiography for Guiding Left Atrial Appendage Occlusion: Impact on Radiation Exposure The Use of Transoesophageal Echocardiography for Guiding Left Atrial Appendage Occlusion: Impact on Radiation Exposure Abstract The aim of this study was primarily to determine if left atrial appendage

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures

Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures The British Journal of Radiology, 84 (2011), 70 77 Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures 1 E P EFSTATHOPOULOS, PhD, 1 I PANTOS,

More information

ISEMIR An IAEA Tool for Radiation Protection Optimization in Interventional Cardiology and Industrial Radiography

ISEMIR An IAEA Tool for Radiation Protection Optimization in Interventional Cardiology and Industrial Radiography An IAEA Tool for Radiation Protection Optimization in Interventional Cardiology and Industrial Radiography Information System on Occupational Exposure in Medicine, Industry and Research What is? is an

More information

MEASUREMENT OF EFFECTIVE DOSE TO PATIENT DURING INTERVENTIONAL CARDIAC PROCEDURE

MEASUREMENT OF EFFECTIVE DOSE TO PATIENT DURING INTERVENTIONAL CARDIAC PROCEDURE J. Asiat. Soc. Bangladesh, Sci. 40(1): 1-7, June 2014 MEASUREMENT OF EFFECTIVE DOSE TO PATIENT DURING INTERVENTIONAL CARDIAC PROCEDURE M.M.M. SIRAZ 1, ALEYA BEGUM, R.K.KHAN, A.HOQUE AND A.BEGUM 2 Health

More information

International Journal of Radiation Research, January 2017

International Journal of Radiation Research, January 2017 Downloaded from ijrr.com at : +0330 on Tuesday November th Volume 5, No International Journal of Radiation Research, January 7 Radiation dose to the thyroid, eyes and parotid glands of patients undergoing

More information

Diagnostic reference levels in intraoral dental radiography in Korea

Diagnostic reference levels in intraoral dental radiography in Korea Imaging Science in Dentistry 2012; 42 : 237-42 http://dx.doi.org/10.5624/isd.2012.42.4.237 Diagnostic reference levels in intraoral dental radiography in Korea Eun-Kyung Kim, Won-Jeong Han, Jin-Woo Choi,

More information

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review Receptor dose and Patient Dose in Radiographic exposures - a 15 year review D J Peet*, N Tyler, M Pryor, P Hollaway, C Strudley and L Leavesley Regional Radiation Protection Service, Royal Surrey County

More information

Tyne NE4 6BE, UK. Corresponding author:

Tyne NE4 6BE, UK. Corresponding author: Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 54 58 doi:10.1093/rpd/nci719 Advance Access published on February 3, 2006 REFERENCE LEVELS IN PTCA AS A FUNCTION OF PROCEDURE COMPLEXITY A.

More information

Patient Dosimetry in Interventional Procedures: what to measure and estimate, why and how?

Patient Dosimetry in Interventional Procedures: what to measure and estimate, why and how? ICTP/IAEA Training Course on Radiation Protection of Patients 16-27 September 2013, Trieste, Italy Patient Dosimetry in Interventional Procedures: what to measure and estimate, why and how? Renato Padovani

More information

A Standardised Approach to Optimisation

A Standardised Approach to Optimisation The Radiographer vol. 51: 105-110 A Standardised Approach to Optimisation Helen M. Warren-Forward 1 & Ronald Beckhaus 2 ABSTRACT Purpose: Optimisation of radiographic images is said to have been obtained

More information

Radiation Dose Reduction in the Invasive Cardiovascular Laboratory

Radiation Dose Reduction in the Invasive Cardiovascular Laboratory JACC: CARDIOVASCULAR INTERVENTIONS VOL. 5, NO. 8, 2012 2012 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION ISSN 1936-8798/$36.00 PUBLISHED BY ELSEVIER INC. http://dx.doi.org/10.1016/j.jcin.2012.05.003

More information

Quality Control and Patient Dosimetry on line for Computed Tomography

Quality Control and Patient Dosimetry on line for Computed Tomography Quality Control and Patient Dosimetry on line for Computed Tomography Jose I. Ten 1,2, Eliseo Vano 2,3, Jose M. Fernandez-Soto 2,3, Roberto Sanchez 3, Juan Arrazola 1,2 1 Diagnostic Radiology Service and

More information

RADIATION DOSES TO PATIENTS UNDERGOING BARIUM MEAL AND BARIUM ENEMA EXAMINATIONS

RADIATION DOSES TO PATIENTS UNDERGOING BARIUM MEAL AND BARIUM ENEMA EXAMINATIONS Radiation Protection Dosimetry (2004), Vol. 109, No. 3, pp. 243±247 DOI: 10.1093/rpd/nch047 RADIATION DOSES TO PATIENTS UNDERGOING BARIUM MEAL AND BARIUM ENEMA EXAMINATIONS M. G. Delichas 1, K. Hatziioannou

More information

Diagnostic reference levels as a quality assurance tool

Diagnostic reference levels as a quality assurance tool Australian Institute of Radiography The Radiographer 2009; 56 (3): 32 37 Literature review Diagnostic reference levels as a quality assurance tool KD Edmonds Medical Physics Section, Medical Radiation

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 233-3 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 29 Dosimetry Framework, Formalism including Uncertainties Costas Hourdakis Greek Atomic

More information

Influence of the exposure time in the area monitors at radiodiagnostic

Influence of the exposure time in the area monitors at radiodiagnostic Journal of Physics: Conference Series PAPER OPEN ACCESS Influence of the exposure time in the area monitors at radiodiagnostic To cite this article: F Leyton et al 2016 J. Phys.: Conf. Ser. 733 012087

More information

National collection of local diagnostic reference levels in Norway and their role in optimization of X-ray examinations

National collection of local diagnostic reference levels in Norway and their role in optimization of X-ray examinations National collection of local diagnostic reference levels in Norway and their role in optimization of X-ray examinations Eva Godske Friberg *, Anders Widmark and Ingrid Helen Ryste Hauge Norwegian Radiation

More information

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals Iran. J. Radiat. Res., 2006; 4 (1): 29-33 Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals D. Shahbazi-Gahrouei * Department of Medical Physics and

More information

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system Original article NOWOTWORY Journal of Oncology 2017, volume 67, number 2, 115 120 DOI: 10.5603/NJO.2017.0018 Polskie Towarzystwo Onkologiczne ISSN 0029 540X www.nowotwory.edu.pl Mean glandular dose values

More information

Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran

Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran Soraya Emamgholizadeh Minaei 1, Fariba Firouzi 1 2, 3,*, Hamid Reza Khosravi 1 Department of radiology, Shahid

More information

Radiation Protection Dosimetry Submitted Manuscript

Radiation Protection Dosimetry Submitted Manuscript Patient radiation doses in the most common interventional cardiology procedures in Croatia: first results Journal: Radiation Protection Dosimetry Manuscript ID: RPD-09-0121 Manuscript Type: Scientific

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Radiation Protection Dosimetry Advance Access published April 22, 2008

Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 5 doi:10.1093/rpd/ncn144 EVALUATION OF EXPOSURE PARAMETERS IN PLAIN RADIOGRAPHY: A COMPARATIVE

More information

Patient dose assessment of CT perfusion scanning at the RSCH

Patient dose assessment of CT perfusion scanning at the RSCH Patient dose assessment of CT perfusion scanning at the RSCH Lesley Leavesley, Emma Whitehead, Matthew Pryor, Debbie Peet Regional Radiation Protection Service Royal Surrey County Hospital, Guildford Overview

More information

Estimation of Organ and Effective Doses for Neonate and Infant Diagnostic Cardiac Catheterizations

Estimation of Organ and Effective Doses for Neonate and Infant Diagnostic Cardiac Catheterizations Medical Physics and Informatics Original Research Kawasaki et al. Dose Estimates for Neonate and Infant Diagnostic Cardiac Catheterization Medical Physics and Informatics Original Research Toshio Kawasaki

More information

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report www.siemens.com/healthcare SOMATOM Drive System Owner Manual Dosimetry and imaging performance report Table of contents 1 Dosimetry and imaging performance report 5 1.1 Dose information 5 1.1.1 General

More information

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Rebecca Milman Marsh, Ph.D. University of Colorado Department of Radiology Who in the Facility Works

More information

Radiation Safety in the Catheterization Lab

Radiation Safety in the Catheterization Lab SCAI FALL FELLOWS COURSE - 2015 Radiation Safety in the Catheterization Lab V. Vivian Dimas, MD, FSCAI Associate Professor Pediatrics, Cardiology UT Southwestern Medical Center Dallas TX None Disclosures

More information

Survey of patients CT radiation dose in Jiangsu Province

Survey of patients CT radiation dose in Jiangsu Province Original Article Page 1 of 6 Survey of patients CT radiation dose in Jiangsu Province Yuanyuan Zhou 1, Chunyong Yang 1, Xingjiang Cao 1, Xiang Du 1, Ningle Yu 1, Xianfeng Zhou 2, Baoli Zhu 1, Jin Wang

More information

Radiation Protection Dosimetry (2008), Vol. 131, No. 1, pp Advance Access publication 23 August 2008

Radiation Protection Dosimetry (2008), Vol. 131, No. 1, pp Advance Access publication 23 August 2008 Radiation Protection Dosimetry (2008), Vol. 131, No. 1, pp. 62 66 Advance Access publication 23 August 2008 doi:10.1093/rpd/ncn229 AN OVERVIEW OF THE USE OF EXTREMITY DOSEMETERS IN SOME EUROPEAN COUNTRIES

More information

Evaluation of radiation dose to patients undergoing interventional radiology procedures at Ramathibodi Hospital, Thailand

Evaluation of radiation dose to patients undergoing interventional radiology procedures at Ramathibodi Hospital, Thailand Available online at http://www.biij.org/2011/3/e22 doi: 10.2349/biij.7.3.e22 biij Biomedical Imaging and Intervention Journal ORIGINAL ARTICLE Evaluation of radiation dose to patients undergoing interventional

More information

OPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY

OPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY March 2010 EAST AFRICAN MEDICAL JOURNAL 127 East African Medical Journal Vol. 87 No. 3 March 2010 OPTIMIsATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY G. K. Korir, BSc, MSc, PhD

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements Standards Revisions for Organizations Providing Fluoroscopy Services The Joint Commission has approved the following revisions for prepublication. While revised requirements

More information

The ORAMED project: Optimization of Radiation Protection for Medical Staff in Interventional Radiology, Cardiology and Nuclear Medicine

The ORAMED project: Optimization of Radiation Protection for Medical Staff in Interventional Radiology, Cardiology and Nuclear Medicine The ORAMED project: Optimization of Radiation Protection for Medical Staff in Interventional Radiology, Cardiology and Nuclear Medicine N. Ruiz Lopez 1, M. Sans Merce 1, I. Barth 2, E. Carinou 3, A. Carnicer

More information

Interventional Fluoroscopy Patient and Staff Safety. Procedures. Safety objectives. Medical physics tasks. Comparative patient risks.

Interventional Fluoroscopy Patient and Staff Safety. Procedures. Safety objectives. Medical physics tasks. Comparative patient risks. Interventional Fluoroscopy Patient and Staff Safety Stephen Balter, Ph.D. Columbia University Medical Center Presented at AAPM Annual Meeting Minneapolis - 2007 Procedures Numbers of procedures will continue

More information

available at journal homepage:

available at  journal homepage: Physica Medica (2012) 28, 61e70 available at www.sciencedirect.com journal homepage: http://intl.elsevierhealth.com/journals/ejmp TECHNICAL NOTE Proposed local diagnostic reference levels in angiography

More information

Radiation Burden in Cardiology. A quick tour! David Sutton

Radiation Burden in Cardiology. A quick tour! David Sutton Radiation Burden in Cardiology A quick tour! David Sutton Contents Radiation Doses to Staff Doses to the eye Radiation Doses to Patients Nuclear Cardiology CCTA Cardiac x-ray procedures Guiding Principles

More information

A systematic review of the efficiency of radiation protection training in raising. University, Perth, Western Australia 6845, Australia.

A systematic review of the efficiency of radiation protection training in raising. University, Perth, Western Australia 6845, Australia. A systematic review of the efficiency of radiation protection training in raising awareness of medical staff working in catheterisation laboratory Mohammed Ali S Alahmari 1, Zhonghua Sun 1 1. Discipline

More information

European Guidelines on DRLs for Paediatric Imaging

European Guidelines on DRLs for Paediatric Imaging European Guidelines on DRLs for Paediatric Imaging Final complete draft 1 1 1 1 1 1 1 1 0 1 0 1 Contents PREFACE... EXECUTIVE SUMMARY... 1. Background.... Introduction.... Purpose and scope.... Definitions...

More information

Managing the imaging dose during image-guided radiation therapy

Managing the imaging dose during image-guided radiation therapy Managing the imaging dose during image-guided radiation therapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Richmond VA Imaging during radiotherapy Radiographic

More information

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology G K Korir, BSc, MSc (Nuclear Science), PhD (Radiological Science) Department of Physics

More information

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS Bayero Journal of Pure and Applied Sciences, 2(1): 1-5 Received: October, 2008 Accepted: February, 2009 PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS *Sharifat,

More information

Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis

Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis Europace (2006) 8, 443 448 doi:10.1093/europace/eul041 Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis Efstathios

More information

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration Establishment of the new IEC 61267 mammography qualities in a clinical system used for instruments calibration 1 Corrêa, E.L.; 2 Vivolo, V.; 3 Potiens, M.P.A. Abstract Instituto de Pesquisas Energéticas

More information

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals B.Aghahadi 1,2*,Z.Zhang 1,S.Zareh 3, S. Sarkar 3, P.S.Tayebi 4 1 Department of Physics, Harbin Institute of Technology, Harbin 150080. P.R. China 2 Department of National Radiation Protection, Atomic Energy

More information

Overview of ICRP Committee 3 Protection in Medicine

Overview of ICRP Committee 3 Protection in Medicine Overview of ICRP Committee 3 Protection in Medicine E. Van o a, D.L. Miller b, M.M. Rehani c a Radiology Department, Complutense University, Madrid, 28040 Spain; e-mail: eliseov@med.ucm.es b Center for

More information

CT Quality Control Manual FAQs

CT Quality Control Manual FAQs CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues

More information

Relationship Between Fluoroscopic Time, Dose Area Product, Body Weight, and Maximum Radiation Skin Dose in Cardiac Interventional Procedures

Relationship Between Fluoroscopic Time, Dose Area Product, Body Weight, and Maximum Radiation Skin Dose in Cardiac Interventional Procedures Radiation Dose in Cardiac Intervention al Procedures Interventional Radiology Original Research A C M E D E N T U R I C A L I M A G I N G AJR 26; 186:774 778 361 83X/6/1863 774 American Roentgen Ray Society

More information

Chapter 2 How Radiation Protection Influences Quality in Radiology

Chapter 2 How Radiation Protection Influences Quality in Radiology Chapter 2 How Radiation Protection Influences Quality in Radiology Eliseo Vano, Kwan-Hoong Ng, and Lawrence Lau Abstract Radiation safety is a key quality element in medical imaging and interventional

More information

2017 Course of the Nordic Association for Clinical Physics on occupational dosimetry in hospitals

2017 Course of the Nordic Association for Clinical Physics on occupational dosimetry in hospitals 2017 Course of the Nordic Association for Clinical Physics on occupational dosimetry in hospitals 27 September 2017 Dr. Pedro Ortiz López Retired member of the ICRP and the IAEA 2 Pedro Ortiz López (Chair)

More information

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3 Radiation Exposure by Routine Radiographic Examinations: Multicenter Study in Japan with Thermoluminescence Dosimetry and Estimation from the Radiographic Data T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1,

More information

Patient dose in routine X-ray examinations in Yazd state

Patient dose in routine X-ray examinations in Yazd state Iran. J. Radiat. Res., 2004; 1(4): 199-204 Patient dose in routine X-ray examinations in Yazd state F. Bouzarjomehri Health Physics Dept. Shahid Sadoghi Univ. of Medical Sciences, Yazd, Iran ABSTRACT Background:

More information

Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks

Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks Poster No.: C-2993 Congress: ECR 2010 Type: Topic: Scientific Exhibit

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

Doses Received by Patients during Thorax X-Ray Examinations

Doses Received by Patients during Thorax X-Ray Examinations Iranian Journal of Medical Physics Vol. 9, No. 4, Autumn 01, 45-51 Received: September 16, 01; Accepted: February 0, 013 Original Article Doses Received by Patients during Thorax X-Ray Examinations Nsikan

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Cardiac-Interventional Radiography

Cardiac-Interventional Radiography STRUCTURED EDUCATION REQUIREMENTS Cardiac-Interventional Radiography The purpose of structured education is to provide the opportunity for individuals to develop mastery of discipline-specific knowledge

More information

Diagnostic Reference Levels (DRLs)

Diagnostic Reference Levels (DRLs) Section 7 Diagnostic Reference Levels (DRLs) Establishment and Use of Diagnostic Reference Levels (DRLs) The National Radiation Safety Committee (NRSC), established in November 2007, is a statutory committee

More information

Interventional radiology, by virtue of its low invasiveness, is

Interventional radiology, by virtue of its low invasiveness, is ORIGINAL RESEARCH T. Moritake Y. Matsumaru T. Takigawa K. Nishizawa A. Matsumura K. Tsuboi Dose Measurement on Both Patients and Operators during Neurointerventional Procedures Using Photoluminescence

More information

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography Wilson Otto Gomes Batista Instituto Federal da Bahia Rua Emidio dos Santos s/n. Barbalho 40301-015

More information

Radiation Dose to the Pediatric Cardiac Catheterization and Intervention Patient

Radiation Dose to the Pediatric Cardiac Catheterization and Intervention Patient Medical Physics and Informatics Original Research Chida et al. Radiation Dose to Pediatric Patients Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 37.44.26.14 on

More information

The print quality of this copy is not an accurate representation of the original.

The print quality of this copy is not an accurate representation of the original. Cardiovasc Intervent Radiol DOI 10.1007/s00270-010-9945-4 TECHNICAL NOTE Staff Radiation Doses in a Real-Time Display Inside the Angiography Room Roberto Sanchez E. Vano J. M. Fernandez J. J. Gallego Received:

More information

Individual Dosimetry Monitoring: What Is Available In The Context Of A Retrospective Evaluation Of Potential Exposure To The Lens?

Individual Dosimetry Monitoring: What Is Available In The Context Of A Retrospective Evaluation Of Potential Exposure To The Lens? Individual Dosimetry Monitoring: What Is Available In The Context Of A Retrospective Evaluation Of Potential Exposure To The Lens? ORAMED 2011 Barcelona Thursday 20 January Sophie Jacob, PhD sophie.jacob@irsn.fr

More information

Performance of several active personal dosemeters in interventional radiology and cardiology

Performance of several active personal dosemeters in interventional radiology and cardiology Performance of several active personal dosemeters in interventional radiology and cardiology S. Chiriotti 1,*, M. Ginjaume 1, E. Vano 2,3, R. Sanchez 3, J.M. Fernandez 2,3, M.A. Duch 1, J. Sempau 1 1.

More information

Pilot study of patient and phantom breast dose measurements in Bulgaria

Pilot study of patient and phantom breast dose measurements in Bulgaria Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast

More information

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Radiation exposure of the Yazd population from medical conventional X-ray examinations Iran. J. Radiat. Res., 2007; 4 (4): 195-200 Radiation exposure of the Yazd population from medical conventional X-ray examinations F. Bouzarjomehri 1*, M.H. Dashti 2, M.H. Zare 1 1 Department of Medical

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

Clinical Determinants of Radiation Dose in Percutaneous Coronary Interventional Procedures

Clinical Determinants of Radiation Dose in Percutaneous Coronary Interventional Procedures JACC: CARDIOVASCULAR INTERVENTIONS VOL. 4, NO. 3, 2011 2011 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION ISSN 1936-8798/$36.00 PUBLISHED BY ELSEVIER INC. DOI: 10.1016/j.jcin.2010.10.014 Clinical Determinants

More information

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute

More information

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS I. Villanueva (1), I. Amor (1), A. Brosed (), R. Villanueva () M. Ginjaume (3), X. Ortega (3), M. Roig (3) (1)

More information

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis 60 Journal of Radiology and Diagnostic Imaging, 2013, 1, 60-64 Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

More information

Doses to Patients arising from Dental X-ray Examinations in the UK, A Review of Dental X-ray Protection Service Data

Doses to Patients arising from Dental X-ray Examinations in the UK, A Review of Dental X-ray Protection Service Data HPA-RPD-22 Doses to Patients arising from Dental X-ray Examinations in the UK, 22 24 A Review of Dental X-ray Protection Service Data A D Gulson, T A Knapp and P G Ramsden ABSTRACT The Dental X-ray Protection

More information

ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED *

ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED * ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED * Olivera Ciraj-Bjelac 1**, Danijela Arandjic 1, Predrag Bozovic 1, Sandra Ceklic 2, Jelena Stankovic

More information

ABSTRACT I. INTRODUCTION

ABSTRACT I. INTRODUCTION 2017 IJSRSET Volume 3 Issue 6 Print ISSN: 2395-1990 Online ISSN : 2394-4099 Themed Section: Engineering and Technology Estimation of Entrance Surface Dose of Adult Patients undergoing Computed Radiography

More information

Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria

Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria Bidemi I.

More information

Medical Fluoroscopic Exposures and Incidents

Medical Fluoroscopic Exposures and Incidents Medical Fluoroscopic Exposures and Incidents FCHPS Spring 2014 Meeting Orlando, FL. April 11, 2014 Kevin Nelson, Ph.D., CHP Radiation Safety Officer Mayo Clinic, Jacksonville, FL nelson.kevin2@mayo.edu

More information

Introduction to RP154 methology for population dose estimates

Introduction to RP154 methology for population dose estimates Introduction to RP154 methology for population dose estimates Hilde M. Olerud (NRPA, NO) DDM 1 & DDM2 WP 3 Contract ENER/2010/NUCL/SI2.581237 Study on European Population Doses from Medical Exposure (Dose

More information

Radiation protection in the cardiac catheterization laboratory: special focus on the role of the operator

Radiation protection in the cardiac catheterization laboratory: special focus on the role of the operator special report Radiation protection in the cardiac catheterization laboratory: special focus on the role of the operator The use of ionizing radiation is associated with a risk of inducing malignant disease

More information

Combined Anatomical and Functional Imaging with Revolution * CT

Combined Anatomical and Functional Imaging with Revolution * CT GE Healthcare Case studies Combined Anatomical and Functional Imaging with Revolution * CT Jean-Louis Sablayrolles, M.D. Centre Cardiologique du Nord, Saint-Denis, France Case 1 Whole Brain Perfusion and

More information

Krueger-Gilbert Health Physics, Inc.

Krueger-Gilbert Health Physics, Inc. Krueger-Gilbert Health Physics, Inc. 1 Educational Objectives Radiation bioeffects Sources of radiation for the US population Typical radiation doses in diagnostic imaging Maryland, FDA and JCAHO guidelines

More information

Asymmetric breast dose in coronary angiography

Asymmetric breast dose in coronary angiography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Asymmetric breast dose in coronary angiography Kam Shan Mong, 1,3a Anthony B. Wallace, 2 and Rick D. Franich 3 Medical Physics Department,

More information