Effective dose conversion factors in paediatric interventional cardiology

Size: px
Start display at page:

Download "Effective dose conversion factors in paediatric interventional cardiology"

Transcription

1 The British Journal of Radiology, 82 (2009), Effective dose conversion factors in paediatric interventional cardiology 1 A KARAMBATSAKIDOU, MSc, 2 B SAHLGREN, MD, PhD, 3 B HANSSON, MSc, 4 M LIDEGRAN, MD, PhD and 1 A FRANSSON, PhD Departments of 1 Medical Physics and 2 Paediatric Cardiology, Karolinska University Hospital, S Stockholm, 3 Department of Hospital Physics, Danderyd Hospital, S Stockholm, and 4 Department of Paediatric Radiology, Karolinska University Hospital, S Stockholm, Sweden ABSTRACT. Conversion factors for effective dose (CF E 5 effective dose/dose area product (msv (Gy cm 2 ) 21 ) in paediatric interventional cardiology were estimated retrospectively for 249 patients using the dose area product (DAP), irradiation geometry, exposure parameters and tissue-weighting factors (TWFs) from the International Commission on Radiological Protection (ICRP) 60. Two methods for estimating the conversion factors, which differed in the description of the irradiation geometry, were evaluated. The effective doses obtained with the two methods were almost identical. The results showed that irradiation geometry had no significant impact on the CF E, and a single factor was defined for both diagnostic and interventional examinations. In addition, the effect of the new tissue-weighting factor for breast tissue (TWF b ) given in ICRP 103 on the effective dose was assessed. The CF E was msv (Gy cm 2 ) 21 (neonate), msv (Gy cm 2 ) 21 (1 year), msv (Gy cm 2 ) 21 (5 years), msv (Gy cm 2 ) 21 (10 years) and msv (Gy cm 2 ) 21 (15 years). Applying these CFs to the individual DAP values of each patient yielded mean effective doses of 13.0 msv (neonate), 8.6 msv (1 year), 6.4 msv (5 years), 8.6 msv (10 years) and 12.7 msv (15 years). The maximum estimated skin dose (15 patients) did not exceed 60 mgy. With the new ICRP value for TWF b, increases in the CFs in the order of 10 30%, and in the effective dose of 10 20%, were indicated. The results indicated that the effective dose in paediatric interventional cardiology is of much greater concern than the skin dose. Furthermore, age-dependent CF E values are required so as not to underestimate the doses to very young patients. Received 19 May 2008 Revised 21 October 2008 Accepted 24 October 2008 DOI: /bjr/ The British Institute of Radiology As the number of interventional radiological procedures increases, the need for accurate methods to determine organ and effective doses also increases [1]. It is well known that adult patients undergoing cardiac interventional radiological procedures could suffer skin injuries as a consequence of the intervention [2 4]. In children, the main concern is instead the increased risk of cancer induction [5]. As the radiation sensitivity in children is considered significantly higher than in adults, radiological interventions in children should always be performed using techniques that minimise the radiation exposure. It is well known that the maximum skin dose is highly dependent on radiation geometry and on the total dose area product (DAP) [6 8]. The influence of these and other factors on the effective dose in children undergoing intervention is, however, not yet well established. This is most likely related to a lack of standardised techniques reflecting a multitude of paediatric congenital heart diseases, as well as large variations in body shape with age in this patient group. Address correspondence to: A Karambatsakidou, Department of Medical Physics, Karolinska University Hospital, S Stockholm, Sweden. angeliki.karambatsakidou@karolinska. se Despite these difficulties, several studies have been performed to estimate the effective dose in children undergoing radiological interventions. The techniques used in these studies include thermoluminescent dosimeter (TLD) measurements on patients [9], phantom studies with TLDs [10], calculations applying conversion factors (effective dose/dap) from the National Radiological Protection Board [11] and Monte Carlo simulations [12, 13]. Common to all these methods is their limited applicability to dose estimations in routine work, owing to the wide range of patient age and size and irradiation geometry encountered in clinical practice. The aim of this study was to establish conversion factors for effective dose (CF E 5 effective dose/dap (msv (Gy cm 2 ) 21 )) in paediatric interventional cardiology, and to evaluate the influence of radiation geometry and age on these factors. The results reported are based on a total of 249 paediatric patient examinations performed on the same interventional X-ray equipment during a 6-year period. In addition, the maximum estimated skin dose (MESD) and typical effective doses from such procedures are reported. The effective doses are based on tissue-weighting factors (TWFs) specified in International Commission on Radiological Protection (ICRP) Report 60 [14]. Owing to the recent update in TWFs by ICRP in 2007 [15], an 748 The British Journal of Radiology, September 2009

2 Effective doses in paediatric interventional cardiology attempt to evaluate the impact of the new set of factors on the effective dose is also reported. Methods and materials X-ray equipment At the Karolinska University Hospital, Solna, paediatric cardiac angiographic and interventional procedures are performed on a biplane Philips Integris H 5000C X- ray unit (Philips, Eindhoven, the Netherlands). This digital cardiac imaging system has an integrated DAP meter (Diamentor; PTW-Freiburg, Germany) mounted on each X-ray tube housing. The DAP value that is reported corresponds to the sum of the exposures using the frontal and lateral planes. The integrated DAP meters are checked once a year against a reference DAP meter (Doseguard 100 (VacuTec 70157); RTI Electronics AB, Molndal, Sweden) calibrated at the secondary standard dosimetry laboratory in Sweden. The image intensifiers (IIs) have three magnification modes 23 cm, 17 cm and 14 cm and are equipped with antiscatter grids with a ratio of 10:1. For the three different II modes (14 cm, 17 cm and 23 cm), the air kerma rates at the entrance window of the II (without grid) are 0.40 mgy per frame, 0.28 mgy per frame and 0.14 mgy per frame, with digital image acquisition, and 0.39 mgy s 1, 0.30 mgy s 1 and 0.22 mgy s 1, respectively, in fluoroscopy mode. The unit uses automatic exposure control, which includes programmes that automatically select beam quality parameters based on the type of examination and patient size. Typical values for the tube peak voltage in paediatric examinations are between 60 kv and 85 kv. The angiographic X-ray unit is equipped with a spectral filter on the tube side, consisting of a sandwich of copper (Cu) and aluminium (Al). The total filtration of the tube for the majority of cardiology examinations is in digital image acquisition mode: 6 mm Al mm Cu (,4 years) or 6 mm Al mm Cu (.4 years and for slow procedures). In fluoroscopy mode, the total filtration was 6 mm Al mm Cu. The programme chosen for cardiology procedures uses a digital image acquisition of 25 or 50 frames per second (,4 years), 25 frames per second (.4 years) and 12.5 frames per second (slow procedures), and a matrix size of Estimation of conversion factors for effective dose Effective dose CFs for 249 paediatric patients between 0 years and 18 years old who underwent cardiac radiological examinations between 1999 and 2004 at the Karolinska University Hospital, Solna, were determined retrospectively. The cardiac procedures were divided into interventional (including a diagnostic part; 110 patients) and pure diagnostic procedures (139 patients). The interventional procedures were transcatheter closure of the patent ductus arteriosus (PDA) or atrial septal defect (ASD) with a device, balloon dilatation of pulmonary valvar stenosis, peripheral pulmonary stenosis or aortic coarctation and, in a few cases, closure of collateral arteries or fistulas with coils. The diagnostic examinations were almost exclusively pre-operative evaluation of complex congenital cardiac anomalies such as tetralogy of Fallot or pulmonary atresia with ventricular septal defects (VSDs) and systemic to pulmonary collaterals (major aortopulmonary collateral arteries; MAPCAs) or univentricular heart before palliative surgery with bidirectional or total cavopulmonary connection (TCPC). Examination reports containing information on cine and fluoroscopy data acquisitions were retrieved for all patients. The patients were divided into five age groups (neonate ( 0 ), 1 year, 5 years, 10 years and 15 years old). The entrance radiation field size used during these procedures varies with age. For the purpose of this study, the field size was assumed to be constant for all patients in each age group (see Table 1). Two different methods (A and B) were used to estimate effective dose CFs. The fundamental difference between the methods is in the definition of the irradiation geometry. Method A Conversion factors for all 249 paediatric patients were estimated using data published by Schmidt et al [12]. They reported CFs (effective dose/dap value) for a wide range of projection angles and for children of age 0, 1 year, 5 years, 10 years and 15 years, assuming a fixed isocentre of 70 cm and a radiation quality of 65 kvp and 3 mm Al beam filtration. For radiation qualities differing from 65 kvp and 3 mm Al filtration, a set of correction factors for tube peak voltage and total beam filtration were reported. The starting point for using their method in our study was to calculate a single CF for effective dose for each acquisition plane (frontal, lateral). This was done by calculating the mean value of the CFs for the most frequent projection angles used at our hospital, and for a radiation quality of 65 kvp and 3 mm Al filtration (Table 2). The projections included were for the frontal plane right anterior oblique (45, 30, 15 ), anteroposterior, left anterior oblique (15, 30, 45 ) (tilt: cranial 15, 0 ; caudal 15, 45 ) and for the lateral plane lateral (tilt 0 ). As the tube voltage and total filtration vary between patients, the technique provided by Schmidt et al [12] to adjust the CFs for tube voltage and total beam filtration was applied. Firstly, the CF (Table 2) for the Table 1. Entrance radiation field size for each age interval Age group (years) Age interval (years) Entrance radiation field (cm 6 cm) The British Journal of Radiology, September

3 A Karambatsakidou, B Sahlgren, B Hansson et al Table 2. Mean value of DAP to effective dose conversion factor (msv (Gy cm 2 ) 21 ; Method A) for various ages and most frequent radiation angles (frontal X-ray tube: RAO (45, 30, 15 ), AP, LAO (15, 30, 45 ) (tilt: cranial 15, 0, caudal (15, 45 )); and for lateral X-ray tube: LAT (tilt 0 ) Age (years) Conversion factor for lateral X-ray tube (msv (Gy cm 2 ) 21 ) Conversion factor for frontal X-ray tube (msv (Gy cm 2 ) 21 ) Adult AP, anteroposterior; DAP, dose area product; LAO, left anterior oblique; LAT, lateral; RAO, right anterior oblique. frontal plane was divided into fluoroscopy and cine fluorography as, with the equipment used at our hospital, different filtrations are used with the different modes. It was further assumed that all fluoroscopy had been performed using only the frontal plane. The median tube voltage value using this plane was retrieved from the examination report for each patient. With this information, a correction for the tube voltage was performed using the CFs published by Schmidt et al (see Tables 9 and 10 in [12]). The correction for total filtration was performed in a similar manner using the technique reported by Schmidt et al (see Table 12 in [12]) and extending the data to include the total filtration values used at our hospital. These calculations were performed using the software PCXMC (version 1.5; Radiation and Nuclear Safety Authority, Helsinki, Finland) [16], and were cross-checked with the filtration correction data provided by Schmidt et al [12]. The two data sets (from Schmidt et al [12] and ours) were identical for all radiation qualities included in the work by Schmidt et al [12]. The CF for the lateral plane was retrieved using the same procedure. In the next step, the effective doses from the different modes (fluoroscopy and cine fluorography) using the frontal plane were calculated by multiplying the CFs from fluoroscopy and cine fluorography with the corresponding DAP values provided in the examination reports. The effective dose from the lateral plane (only cine fluorography) was estimated in the same way. Finally, a single CF for the entire procedure (frontal and lateral planes) was estimated by adding the effective doses from each plane, and dividing by the total DAP value of the examination. Method B For 52 of the 249 patients, examination reports containing comprehensive information on exposure parameters for the cine fluorography part (beam projection and cranio/caudal angles, tube voltage, tube filtration) could be retrieved. Combined with the information on patient age and DAP values, these data were fed into the X-ray dose calculation software PCXMC [16], and the effective dose corresponding to each projection angle of the cine fluorography procedure was calculated. For simplicity, for any given projection angle and for each 4 kv interval, a single tube voltage value equal to the mean tube voltage was used. In clinical practice, an isocentric irradiation geometry is used, such that the focus-to-patient distance will vary with patient size. In a previous study, and using anthropomorphic paediatric phantoms, focus-to-patient distances were typically between 50 cm and 60 cm [10]. As in this work, no data on patient size were available and a focus-to-skin distance of 60 cm was assumed for all projection angles and patients. The field sizes used with Method B are reported in Table 1. The fluoroscopy mode data were incomplete in that they did not contain information on projection angle and tube voltage. Based on observations of patient examinations as they are performed today and on statements from staff, a frontal projection angle was assigned to the complete fluoroscopy part of the examination. The tube voltage used in fluoroscopy was set to the median tube voltage used in frontal projections in the cine procedure. Finally, the CF for the complete cardiac procedure was determined by adding the effective doses from all projections, and dividing by the total DAP value. In order to evaluate the impact on the CF of the radiation geometry during fluoroscopy, two different situations were assessed: one using only the frontal plane for fluoroscopy (100% frontal) and the other assuming that fluoroscopy was performed using both planes (50% frontal, 50% lateral). CFs using either geometry were estimated with Method A. Effective dose For all 249 patients and for both Methods (A and B), the effective doses were estimated by using TWFs from ICRP Report 60 [14]. With the recent update of such values by the ICRP, the impact on the effective dose from irradiation of breast tissue has increased significantly (weighting factor for breast tissue has increased from 0.05 to 0.12). In addition, other TWFs have changed, but these changes can be expected to have only a minor effect on the effective dose from cardiac procedures. To evaluate the importance of the new value for breast tissue on the effective dose, effective doses for the subset of 52 patients were calculated replacing the weighting factor for breast tissue from ICRP Report 60 with the new value in ICRP Report 103 [15]. In addition, an estimate of the effective dose by applying the complete set of new TWFs was attempted for five patients. These calculations were performed using Method B. Skin dose For the evaluation of maximum skin dose in paediatric cardiac procedures, 15 patients undergoing PDA (5) or 750 The British Journal of Radiology, September 2009

4 Effective doses in paediatric interventional cardiology Figure 1. Mean dose area product (DAP) values from diagnostic and interventional procedures, and the corresponding number of patients in the different age groups. Results for adult patients [6] are included for comparison. ASD closure (4), pre-operative evaluation of pulmonary atresia and VSD with MAPCA (4) and univentricular heart before TCPC (2) were selected for film dosimetry. These procedures are all high-dose procedures and are also the most common interventional and diagnostic paediatric heart examinations performed at Karolinska University Hospital. In these patients, the MESD was estimated using slow radiographic film (EDR2, 35 cm 6 43 cm; Eastman Kodak Co, Rochester, NY). This method has been applied by our group in a previous study on adults [7], and the same methodology as that used for the adult patients was applied in this work. Based on observations of paediatric examinations performed at this unit, together with information from examination reports, it was concluded that the maximum skin dose was located on the patient s back. The skin dose from the frontal plane (includes the region of the MESD on the patient s back) was documented on a film positioned between the patient and the table, and the corresponding MESD was obtained from the maximum film density (Figure 3 in [7]). The film processor used was a Kodak RP X-OMAT Processor, Model M8, and the transmission densitometer was an X-rite 331 (X-rite Co., Grand Rapids, MI). Statistical analysis The two-tailed Mann Whitney test was applied to evaluate if there was a statistically significant difference between the results using Method A and those using Method B, between diagnostic and interventional procedures, and between using one (frontal) or two (frontal plus lateral) projections during fluoroscopy. To investigate correlations in scatter plots displaying linear or exponential behaviour, linear and exponential regressions were used, respectively. Linear regression was applied on effective dose vs DAP value for the different age groups. CF vs age for Method A and B and mean CF vs age for diagnostic and interventional procedures both showed exponential behaviour, and exponential regression was used in these cases. Mean lifetime mortality risk The multiplicative model of ICRP 60 [14] was used to calculate the lifetime probability of death following irradiation. The mortality risk in this model is age- and gender-dependent, and the risk factors vary with age (0 9 years 5 16%/Sv (girls) and 13%/Sv (boys); years 5 9.5%/Sv (girls) and 7.5%/Sv (boys)). Figure 2. Comparison of the two methods for estimating mean conversion factors (CFs) for effective dose in 52 paediatric patients. Results The mean DAP values, including all procedures (diagnostic and interventional) in the different age groups, together with data on the number of patients in each group, are shown in Figure 1. For comparison, data for adult cardiac procedures (diagnostic, 73 Gy cm 2 ; interventional, 120 Gy cm 2 ) taken from a previous study are included [6]. The paediatric DAP values include fluoroscopy and cine fluorography from both the frontal and lateral planes. There is a clear trend for an increase in total DAP with increasing age. There is also a tendency for a slightly higher DAP value for diagnostic procedures ( Gy cm 2 (neonate; 1 SD (standard deviation)), Gy cm 2 (1 year; 1 SD), Gy cm 2 (5 years; 1 SD), Gy cm 2 (10 years; 1 SD), Gy cm 2 (15 years; 1 SD)) compared The British Journal of Radiology, September

5 A Karambatsakidou, B Sahlgren, B Hansson et al Table 3. Contribution from fluoroscopy to the total DAP in diagnostic and interventional procedures for the various age groups Age (years) (DAP F /DAP T ) diagnostic (%) (DAP F /DAP T ) interventional (%) DAP F, dose area product from fluoroscopy; DAP T, dose area product from the complete procedure. with interventional procedures ( Gy cm 2 (neonate; 1 SD), Gy cm 2 (1 year; 1 SD), Gy cm 2 (5 years; 1 SD), Gy cm 2 (10 years; 1 SD), Gy cm 2 (15 years; 1 SD)) in the paediatric age groups. The contribution to the total DAP from fluoroscopy is displayed in Table 3. The percentage share of fluoroscopy for diagnostic and interventional procedures is lowest for neonates (31% and 49%), reaches a peak at the age of 5 years (64% and 79%), and finally decreases to values of 51% and 59%, respectively, at the age of 15 years. CFs for effective dose using Methods A and B for the subset of 52 patients are shown in Figure 2. There was no significant difference (a>0.05; Mann Whitney test) in the results using the two methods, showing a mean deviation of 5%. The difference in CFs using Method A compared with Method B was always less than 10%, except in four patients (216%, 15%, 20% and 25% difference). Detailed inspection of the corresponding patient data revealed that, for three of these patients, a program containing a lower tube filtration than normal had been applied whereas, in the fourth patient, a radiation quality using a 25% higher tube voltage than for patients in the same age group had been used. Recalculation of the CFs taking these differences into account re-established the limit of 10% difference between the two methods. The data could be fitted with an exponential function: (CF E 5 2.7e 20.15*age ; age 0 15 years; r50.97 for both Methods A and B). Based on the very good agreement of the CFs obtained using either of the two methods (A and B) in the subset of 52 patients, the simplest technique (Method A) was applied to the whole patient sample (249 patients) and the results used to evaluate the extent to which CFs from diagnostic and interventional procedures differed. The results showed almost identical factors for both diagnostic and interventional procedures within each age group (a>0.05; Mann Whitney test; Figure 3), and could be fitted with exponential equations (diagnostic, CF E 5 2.6e 20.14*age ; interventional, CF E 5 2.6e 20.15*age ; r50.96, age 0 15 years). The corresponding data for adult patients, as reported in a previous study [6], are included for comparison. Table 4 shows the influence of radiation geometry during the fluoroscopy part of the examination on the CF. There was no significant difference using the two geometries for neonates (a>0.05; Mann Whitney test), and 1-year-old (a,0.05; Mann Whitney test) and 15- year-old patients (a>0.05; Mann Whitney test). In the 5 year and 10 year age groups, there was a minor yet significant difference (a,0.01 and a,0.001, respectively; Mann Whitney test). The MESD for the subset of 15 patients did not exceed 60 mgv. The mean value of MESD for patients undergoing ASD closure (monoplane, 4 patients) was 12 mgy whereas, in patients undergoing PDA closure, TCPC or MAPCA (biplane; 11 patients), the mean MESD was 20 mgv. The relationship between age and effective doses as a function of DAP values is illustrated in Figure 4, showing a linear correlation for all age groups (R ). Table 5 shows the impact of the new TWF b (tissueweighting factor for breast tissue) from ICRP Report 103 [15] on the effective dose. This recalculation (Method B) resulted in an upward adjustment of the mean effective dose by 13% (newborn), 16% (1 year), 14% (5 years), 10% (10 years) and 11% (15 years). In five patients, the effective dose was estimated taking into account all new TWFs from ICRP [15]. There was only a small difference (,5%) compared with results obtained applying only the new TWF b. The effect of the updated TWF b on the contribution to effective dose from the different projections (posteroanterior and lateral) is also reported in Table 5. The effective dose from posteroanterior and lateral projections in the different age groups increased by up to 12% and 32%, respectively. As can be observed from these data, there is a significant increase in the effective dose from the Table 4. Conversion factors obtained using two different radiation geometries for the fluoroscopy part of the procedures (Method A) Figure 3. Mean conversion factors (CFs) (Method A) for 249 paediatric patients undergoing diagnostic and interventional procedures. Age (years) CF E 1 (msv (Gy cm 2 ) 21 ) ( 1 SD) CF E 2 (msv (Gy cm 2 ) 21 ) ( 1 SD) CF 1 E, conversion factor from 100% fluoroscopy using only frontal plane; CF 2 E, conversion factor obtained from using both planes during fluoroscopy (50% frontal and 50% lateral plane); SD, standard deviation. 752 The British Journal of Radiology, September 2009

6 Effective doses in paediatric interventional cardiology Figure 4. Effective dose estimated using conversion factors from Method A and individual dose area product (DAP) values for all 249 patients. A linear fit of the patient data within each age group yielded: 0 year, effective dose DAP 0.19 (R ); 1 year, effective dose DAP 0.40 (R ); 5 years, effective dose DAP+0.01 (R ); 10 years, effective dose DAP+0.85 (R ); 15 years, effective dose DAP+1.60 (R ). lateral projection using the new TWF b. As the contribution to effective dose from the lateral projection is relatively high, the change in the TWF b therefore also changes the relative impact on effective dose from the posteroanterior vs lateral projections. The estimates of effective dose (Method A, 249 patients), together with the mean lifetime risk factors of death following irradiation as defined by ICRP Report 60 [14], yielded increases in mortality risk of % for girls and % for boys (using ICRP Report 60 TWFs), with the highest risk estimates corresponding to neonates. The risk factors have not been changed in ICRP Report 103 [15]. Taking into account the change in TWF for breast tissue in the new report, the resulting increase in effective dose is reflected by a corresponding increase in the mortality risk. Discussion Patients undergoing cardiac radiological procedures can receive radiation doses that exceed threshold levels Table 5. Mean effective doses (Method B; 52 patients) using tissue-weighting factors from ICRP Report 60 (E 60 ), and including the new tissue-weighting factor for breast from ICRP 103 (E 103 ). Ratios (E 103 /E 60 ) PA,LAT for the contribution to effective dose from the posteroanterior (PA) and lateral (LAT) projections, respectively, are included Age (years) E 60 (msv) E 103 (msv) (E 103 / E 60 ) PA (E 103 /E 60 ) LAT ICRP, International Commission on Radiological Protection. for deterministic effects (skin erythema, >2 Sv). However, in paediatric cardiac radiological examinations, the skin dose is usually well below the 2 Sv limit [13, 17, 18], and in the present study a MESD of 60 msv was registered for an MAPCA (high-dose diagnostic examination). In the younger patients, the concern is that the effective dose, owing to the smaller patient size, can be significantly greater than in adults. This is clearly demonstrated in Figure 4, which shows clinical data on the dependence of age on the effective dose. Comparing the mean lifetime risk of death related to irradiation [14] (ICRP Report 60, multiplicative model) of newborns (using data from the present study and TWFs from ICRP Report 60) and adults (data taken from the article by Hansson and Karambatsakidou [6]), the increased mortality risk for adults from cardiac procedures was 0.16% (intervention) and 0.08% (diagnostic), whereas the corresponding figures for newborns were 0.21% (girls) and 0.17% (boys) for all procedures. The results presented in this work are based on clinical data, and could be useful as reference data for sites performing radiological procedures in children with congenital heart disease in order to estimate the effective dose from such interventions. As the younger children are expected to have an increased risk for radiationinduced cancer [5, 14, 19], it is of the utmost importance to monitor the doses to these patients. This becomes especially important in patients who undergo multiple radiological procedures. Paediatric cardiac patients represent a relatively inhomogeneous group in the sense that there exist many different types of congenital heart disease in children. Consequently, the radiological interventions do not, in general, follow a standardised scheme. This contributes to the difficulty in comparing DAP values from this study with values published by other groups. An attempt to do so indicates that the total DAP values reported for paediatric cardiac patients by other groups are almost double those reported in this work [20, 21]. The significance of this fact is, however, doubtful in the light of the variations in age and in the type and severity of disease that these different data represent. In addition, differences in radiation quality between different radiological equipment will influence the total DAP (or rather the DAP rate) required to produce images of sufficient quality. The mean percentage share of fluoroscopy for all 249 paediatric patients in this study was close to 60%, which is similar to the results presented by Schueler et al [20]. For comparison, a 20 70% contribution to the mean DAP value from fluoroscopy has been reported in cardiac procedures on adult patients [6, 7, 22 24]. The CFs for effective dose were almost identical using Methods A and B, as evaluated for a subset of 52 patients (Figure 2). The two methods differ in the description of the irradiation geometry, with Method A being less exact than Method B. The results show that the dependence of the irradiation geometry (number of projections) on the effective dose is of minor importance, and that a relatively rough estimation of the geometry could allow for a calculation of effective dose from these types of procedures on children. This conclusion is furthermore supported by the results presented in Table 4 on the effect of irradiation geometry during fluoroscopy on CF E The British Journal of Radiology, September

7 A Karambatsakidou, B Sahlgren, B Hansson et al Table 6. Comparison of conversion factors (msv (Gy cm 2 ) 21 ) obtained in this study with previously published values Study CF E 0 (msv (Gy cm 2 ) 21 ) CF E 1 (msv (Gy cm 2 ) 21 ) CF E 5 (msv (Gy cm 2 ) 21 ) CF E 10 (msv (Gy cm 2 ) 21 ) CF E 15 (msv (Gy cm 2 ) 21 ) This study PA: 3.65 LAT: 3.74 PA: 1.80 LAT: 1.97 PA: 0.94 LAT: 0.98 PA: 0.62 LAT: 0.66 PA: 0.33 LAT: 0.34 Rassow et al [11] PA: LAT: PA: 0.47 LAT: 1.02 PA: 0.35 LAT: 0.71 PA: 0.26 LAT: 0.53 PA: 0.16 LAT: Axelsson et al [10] PA: 1.8 LAT: 1.4 PA: 0.9 LAT: 0.7 Schmidt et al [12] PA: PA: PA: PA: PA: LAT: LAT: LAT: LAT: LAT: Bacher et al [13] a 1.59 b /1.31 c 1.59 b /1.31 c 1.59 b /1.31 c 1.59 b /1.31 c Schultz et al [25] CF 0 E, conversion factor for age group 0; CF 1 E, conversion factor for age group 1; CF 5 E, conversion factor for age group 5; CF 10 E, conversion factor for age group 10; CF 15 E, conversion factor for age group 15; PA, posteroanterior; LAT, lateral. a Children,10 years. b Tube filtration mm Al mm Cu. c Tube filtration mm Al mm Cu. and by results from phantom studies reported by Axelsson et al [10]. Statistical evaluation (Mann Whitney U test) of the significance of the irradiation geometry during fluoroscopy (data as in Table 4) was negative for age groups 0, 1 year and 15 years, whereas there was a weak correlation for the 5 year and 10 year age groups. There is no clear explanation for this, although weak irradiation geometry dependence of the fluoroscopy part on the effective dose has been found for these two age groups. In addition, there was no significant difference in the CF between interventional (including a diagnostic part) and purely diagnostic procedures for a given age group. In conclusion, these results indicate that a single CF for effective dose can be applied to each age group, independent of the type of procedure. The CFs estimated in the present work have been compared with factors published by other groups (see Table 6). Except for the results given by Bacher et al [13] (no age grouping), all studies conclude that CFs for effective dose decrease with age [10, 12, 20, 25]. It should be noted that the CF E values from the present study are the highest of all. This is believed to reflect differences in radiation quality used at different hospitals/equipments, indicating that the higher tube filtration and tube voltage used at our hospital yields a higher effective dose per DAP. Conversely, owing to the lower total DAP values resulting from the use of a more penetrating beam, the effective doses in this study are lower than those reported by Schultz et al [25], except in the newborn for which the values in our study are approximately 20% higher. A similar trend with relatively higher effective dose estimates for newborns was given by Rassow et al [11] (approximately 30% higher). The high value of CF E for neonates most likely reflects the difficulties in collimating properly the radiation field, with the result that larger body regions will be irradiated in very young children. This has also been shown by Rassow et al [11] and Li et al [26]. The TWFs from ICRP Report 60 were used throughout the whole work. In addition, the impact of the new TWF b on the effective dose from paediatric cardiac procedures was assessed. These recalculations showed, as expected, a relative increase in the contribution to effective dose from the lateral plane compared with the frontal plane. There are most likely to be several factors contributing to such a change, with the most important possibly being a larger contribution to effective dose from irradiation of breast tissue in the lateral projection than in the posteroanterior projection. The results show that CFs based on previous ICRP TWFs need to be adjusted upwards, and initial calculations indicate an adjustment up to +12% for posteroanterior projections and +32% for lateral projections (see Table 5). A detailed analysis of the impact of the new set of TWF values on effective dose for cardiac procedures is ongoing and will be presented in a future publication. Conclusions A technique to define CFs for the effective dose to paediatric cardiac patients has been presented. Using ICRP Report 60 TWF, it has been shown that a single CF can be used within a given age group, independent of the type of procedure. The technique was applied to 249 paediatric examinations, yielding effective doses in the range of msv. Initial results applying updated TWFs from ICRP Report 103 indicate an increase in the CFs of approximately 10% for posteroanterior projections and 30% for lateral projections, and an increase in the effective dose of 10 20%, depending on age. In sites in which similar beam qualities as those presented in our study are used, the CFs reported here could be applied directly to radiological procedures in children with congenital heart disease in order to estimate the effective dose from such interventions. References 1. ICRU. Patient dosimetry for X rays used in medical imaging. Report 74. Oxford, UK: Oxford University Press, Barnea Y, Amir A, Sharif R, Weiss J, Gur E. Chronic radiodermatitis injury after cardiac catheterization. Ann Plast Surg 2002;49: Dehen L, Vilmer C, Humiliere C, Cordos T, Pentousis D, Ollivaud L, et al. Chronic radiodermatitis following cardiac catheterisation: a report of two cases and a brief review of the literature. Heart 1999;81: Aerts A, Decraene T, van den Oord JJ, Dens J, Janssens S, Guelinckx P, et al. Chronic radiodermatitis following percutaneous coronary interventions: a report of two cases. J Eur Acad Dermatol Venereol 2003;17: The British Journal of Radiology, September 2009

8 Effective doses in paediatric interventional cardiology 5. ICRP 85. Avoidance of Radiation Injuries from Medical Interventional Procedures. Oxford, UK: Pergamon Press, Hansson B, Karambatsakidou A. Relationships between entrance skin dose, effective dose and dose area product for patients in diagnostic and interventional cardiac procedures. Radiat Prot Dosim 2000;90: Karambatsakidou A, Tornvall P, Saleh N, Chouliaras T, Löfberg P-O, Fransson A. Skin dose alarm levels in cardiac angiography procedures: is a single DAP value sufficient? Br J Radiol 2005;78: Van de Putte S, Verhaegen F, Taeymans Y, Thierens H. Correlation of patient skin doses in cardiac interventional radiology with dose-area product. Br J Radiol 2000;73: Faulkner K, Love HG, Sweeney JK, Bardsley RA. Radiation doses and somatic risk to patients during cardiac radiological procedures. Br J Radiol 1986;59: Axelsson B, Khalil C, Lidegran M, Schuwert P, Mortensson W. Estimating the effective dose to children undergoing heart investigations-a phantom study. Br J Radiol 1999;72: Rassow J, Schmaltz AA, Hentrich F, Streffer C. Effective doses to patients from pediatric cardiac catheterization. Br J Radiol 2000;73: Schmidt PWE, Dance DR, Skinner CL, Castellano Smith IA, McNeill JG. Conversion factors for the estimation of effective dose in paediatric cardiac angiography. Phys Med Biol 2000;45: Bacher K, Bogaert E, Lapere R, De Wolf D, Thierens H. Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization. Circulation 2005;111: ICRP. Recommendations of the International Commission on Radiological Protection. Publication 60. Oxford, UK: Pergamon Press; ICRP 103. The 2007 Recommendations of the International Commission on Radiological Protection. Elsevier Ltd, PCXMC v1.5, user s guide. Available from: stuk.fi/sateilyn_kaytto/ohjelmat/pcxmc/en_gb/pcxmc/ _files/ /default/pcxmc15_usersguide.pdf 17. Boothroyd A, McDonald E, Moores BM, Sluming V, Carty H. Radiation exposure to children during cardiac catheterization. Br J Radiol 1997;70: Moore JD, Shim D, Sweet JL, Arheart KH. Beekman III R. Radiation exposure to children during coil occlusion of the patent ductus arteriosus. Catheter Cardiovasc Interv 1999;47: Darby SC. Epidemiological evaluation of radiation risk using population exposed at high doses. Health Phys 1986;51: Schueler BA, Jubrud PR, Gray JE, Stears JG, Wu KY. Radiation exposure and efficacy of exposure-reduction techniques during cardiac catheterisation in children. AJR Am J Roentgenol 1994;162: Leibovic SJ, Fellows KE. Patient radiation exposure during pediatric cardiac catheterization. Cardiovasc Intervent Radiol 1983;6: Tsapaki V, Patsilinakos S, Voudris V, Magginas A, Pavlidis S, Maounis T, et al. Level of patient and operator dose in the largest cardiac centre in Greece. Radiat Prot Dosimetry 2008;129: Kuon E, Glaser C, Dahm JB. Effective techniques for reduction of radiation dosage to patients undergoing invasive cardiac procedures. Br J Radiol 2003;76: Livingstone RS, Chandy S, Peace TB, George PV, John B, Pati P. Audit of radiation dose to patients during coronary angiography. Indian J Med Sci 2007;61: Schultz F W, Geleijns J, Spoelstra FM, Zoetelief J. Monte Carlo calculations for assessment of radiation dose to patients with congenital heart defect and to staff during cardiac catheterizations. Br J Radiol 2003;76: Li LB, Kai M, Kusama T. Radiation exposure to patients during paediatric cardiac catheterization. Radiat Prot Dosimetry 2001;94: The British Journal of Radiology, September

Estimation of Organ and Effective Doses for Neonate and Infant Diagnostic Cardiac Catheterizations

Estimation of Organ and Effective Doses for Neonate and Infant Diagnostic Cardiac Catheterizations Medical Physics and Informatics Original Research Kawasaki et al. Dose Estimates for Neonate and Infant Diagnostic Cardiac Catheterization Medical Physics and Informatics Original Research Toshio Kawasaki

More information

A pilot programme on patient dosimetry on paediatric interventional cardiology in Chile

A pilot programme on patient dosimetry on paediatric interventional cardiology in Chile A pilot programme on patient dosimetry on paediatric interventional cardiology in Chile Patricia Miranda *1, Fernando Leyton 2, Eliseo Vano 3, Alfonso Espinoza 2, Raul Ramirez 4, Otto Delgado 2, Pedro

More information

Radiation Dose to the Pediatric Cardiac Catheterization and Intervention Patient

Radiation Dose to the Pediatric Cardiac Catheterization and Intervention Patient Medical Physics and Informatics Original Research Chida et al. Radiation Dose to Pediatric Patients Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 37.44.26.14 on

More information

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS RADIOPROTECTION AND DOSIMETRY MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS DANIELA ADAM 1, ANA STOCHIOIU

More information

RADIATION DOSES TO PATIENTS UNDERGOING BARIUM MEAL AND BARIUM ENEMA EXAMINATIONS

RADIATION DOSES TO PATIENTS UNDERGOING BARIUM MEAL AND BARIUM ENEMA EXAMINATIONS Radiation Protection Dosimetry (2004), Vol. 109, No. 3, pp. 243±247 DOI: 10.1093/rpd/nch047 RADIATION DOSES TO PATIENTS UNDERGOING BARIUM MEAL AND BARIUM ENEMA EXAMINATIONS M. G. Delichas 1, K. Hatziioannou

More information

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals Iran. J. Radiat. Res., 2006; 4 (1): 29-33 Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals D. Shahbazi-Gahrouei * Department of Medical Physics and

More information

Asymmetric breast dose in coronary angiography

Asymmetric breast dose in coronary angiography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Asymmetric breast dose in coronary angiography Kam Shan Mong, 1,3a Anthony B. Wallace, 2 and Rick D. Franich 3 Medical Physics Department,

More information

Measurement of Absorbed Doses in Anatomical Phantoms

Measurement of Absorbed Doses in Anatomical Phantoms International Journal of Pure and Applied Physics. ISSN 0973-1776 Volume 8, Number 1 (2012), pp. 53-58 Research India Publications http://www.ripublication.com/ijpap.htm Measurement of Absorbed Doses in

More information

Radiation Safety in the Catheterization Lab

Radiation Safety in the Catheterization Lab SCAI FALL FELLOWS COURSE - 2015 Radiation Safety in the Catheterization Lab V. Vivian Dimas, MD, FSCAI Associate Professor Pediatrics, Cardiology UT Southwestern Medical Center Dallas TX None Disclosures

More information

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: Rob Loader Oliver Gosling Introduction Approached by Dr Oliver Gosling (Research

More information

Research Article. Corresponding author H. Osman

Research Article. Corresponding author H. Osman Scholars Journal of Applied Medical Sciences (SJAMS) ISSN 2320-6691 Sch. J. App. Med. Sci., 2013; 1(5):511-515 Scholars Academic and Scientific Publisher (An International Publisher for Academic and Scientific

More information

Relationship Between Fluoroscopic Time, Dose Area Product, Body Weight, and Maximum Radiation Skin Dose in Cardiac Interventional Procedures

Relationship Between Fluoroscopic Time, Dose Area Product, Body Weight, and Maximum Radiation Skin Dose in Cardiac Interventional Procedures Radiation Dose in Cardiac Intervention al Procedures Interventional Radiology Original Research A C M E D E N T U R I C A L I M A G I N G AJR 26; 186:774 778 361 83X/6/1863 774 American Roentgen Ray Society

More information

Pilot study of patient and phantom breast dose measurements in Bulgaria

Pilot study of patient and phantom breast dose measurements in Bulgaria Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast

More information

Low-dose prospective ECG-triggering dual-source CT angiography in infants and children with complex congenital heart disease: first experience

Low-dose prospective ECG-triggering dual-source CT angiography in infants and children with complex congenital heart disease: first experience Low-dose prospective ECG-triggering dual-source CT angiography in infants and children with complex congenital heart disease: first experience Ximing Wang, M.D., Zhaoping Cheng, M.D., Dawei Wu, M.D., Lebin

More information

Determination of the conversion factor for the estimation of effective dose in lungs, urography and cardiac procedures

Determination of the conversion factor for the estimation of effective dose in lungs, urography and cardiac procedures Department of Medical Radiation Physics KAROLINSKA INSTITUTET STOCKHOLM UNIVERSITY Determination of the conversion factor for the estimation of effective dose in lungs, urography and cardiac procedures

More information

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers Patient effective dose evaluation for chest X-ray examination in three digital radiography centers R. Paydar 1,3*, A. Takavar 1, M.R. Kardan 2,3, A. Babakhani 3,4, M.R. Deevband 3, S. Saber 5 1Medical

More information

The print quality of this copy is not an accurate representation of the original.

The print quality of this copy is not an accurate representation of the original. Cardiovasc Intervent Radiol DOI 10.1007/s00270-010-9945-4 TECHNICAL NOTE Staff Radiation Doses in a Real-Time Display Inside the Angiography Room Roberto Sanchez E. Vano J. M. Fernandez J. J. Gallego Received:

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Assessment of Patient Radiation Dose in Interventional Procedures at Shahid Madani Heart Center in Khorramabad, Iran Mehrdad Gholami 1*, Somayeh Gharloghi

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3 Radiation Exposure by Routine Radiographic Examinations: Multicenter Study in Japan with Thermoluminescence Dosimetry and Estimation from the Radiographic Data T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1,

More information

Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis

Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis Europace (2006) 8, 443 448 doi:10.1093/europace/eul041 Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis Efstathios

More information

Evaluation of an On-line Patient Exposure Meter in Neuroradiology

Evaluation of an On-line Patient Exposure Meter in Neuroradiology Medical Physics Nikolaos A. Gkanatsios, Walter Huda, PhD Keith R. Peters, MD James A. Freeman, RT Evaluation of an On-line Patient Exposure Meter in Neuroradiology PURPOSE: To assess the clinical performance

More information

Radiation Protection Dosimetry Advance Access published February 17, Radiation Protection Dosimetry (2011), pp. 1 9

Radiation Protection Dosimetry Advance Access published February 17, Radiation Protection Dosimetry (2011), pp. 1 9 Radiation Protection Dosimetry Advance Access published February 17, 2011 Radiation Protection Dosimetry (2011), pp. 1 9 doi:10.1093/rpd/ncq600 AN AUDIT OF DIAGNOSTIC REFERENCE LEVELS IN INTERVENTIONAL

More information

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Debra Pennington, MD Director of Imaging Dell Children s Medical Center Debra Pennington, MD Director of Imaging Dell Children s Medical Center 1 Gray (Gy) is 1 J of radiation energy/ 1 kg matter (physical quantity absorbed dose) Diagnostic imaging doses in mgy (.001 Gy)

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

Application of international standards to diagnostic radiology dosimetry

Application of international standards to diagnostic radiology dosimetry Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Radiation exposure of the Yazd population from medical conventional X-ray examinations Iran. J. Radiat. Res., 2007; 4 (4): 195-200 Radiation exposure of the Yazd population from medical conventional X-ray examinations F. Bouzarjomehri 1*, M.H. Dashti 2, M.H. Zare 1 1 Department of Medical

More information

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology G K Korir, BSc, MSc (Nuclear Science), PhD (Radiological Science) Department of Physics

More information

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS Bayero Journal of Pure and Applied Sciences, 2(1): 1-5 Received: October, 2008 Accepted: February, 2009 PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS *Sharifat,

More information

Patient dose assessment of CT perfusion scanning at the RSCH

Patient dose assessment of CT perfusion scanning at the RSCH Patient dose assessment of CT perfusion scanning at the RSCH Lesley Leavesley, Emma Whitehead, Matthew Pryor, Debbie Peet Regional Radiation Protection Service Royal Surrey County Hospital, Guildford Overview

More information

Evaluation of radiation dose and image quality following changes to tube potential (kvp) in conventional paediatric chest radiography

Evaluation of radiation dose and image quality following changes to tube potential (kvp) in conventional paediatric chest radiography Available online at http://www.biij.org/2006/3/e35 doi: 10.2349/biij.2.3.e35 biij Biomedical Imaging and Intervention Journal ORIGINAL PAPER Evaluation of radiation dose and image quality following changes

More information

Procedural radiation exposure of interventional cardiologists and radiologists Kuipers, G.

Procedural radiation exposure of interventional cardiologists and radiologists Kuipers, G. UvA-DARE (Digital Academic Repository) Procedural radiation exposure of interventional cardiologists and radiologists Kuipers, G. Link to publication Citation for published version (APA): Kuipers, G. (2011).

More information

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography Wilson Otto Gomes Batista Instituto Federal da Bahia Rua Emidio dos Santos s/n. Barbalho 40301-015

More information

Minimizing radiation exposure to the vascular surgeon

Minimizing radiation exposure to the vascular surgeon From the Society for Vascular Surgery Minimizing radiation exposure to the vascular surgeon Omar P. Haqqani, MD, Prakhar K. Agarwal, BA, Neil M. Halin, DO, and Mark D. Iafrati, MD, Boston, Mass Objectives:

More information

UNMH Pediatric Cardiology Clinical Privileges. Name: Effective Dates: From To

UNMH Pediatric Cardiology Clinical Privileges. Name: Effective Dates: From To All new applicants must meet the following requirements as approved by the UNMH Board of Trustees, effective August 18, 2017: Initial Privileges (initial appointment) Renewal of Privileges (reappointment)

More information

Assessing the standard dose to standard patients for x-ray investigations

Assessing the standard dose to standard patients for x-ray investigations Assessing the standard dose to standard patients for x-ray investigations A. Almén and W. Leitz Swedish Radiation Protection Authority, SE-171 16 Stockholm, Sweden E-mail: anja.almen@ssi.se Abstract. The

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Evaluation of the doses delivered to premature babies in the Belgian Neonatal Intensive Care Units

Evaluation of the doses delivered to premature babies in the Belgian Neonatal Intensive Care Units Evaluation of the doses delivered to premature babies in the Belgian Neonatal Intensive Care Units PreDos Project Jérémie Dabin Lara Struelens Filip Vanhavere Table of Contents Executive summary... Introduction...

More information

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals B.Aghahadi 1,2*,Z.Zhang 1,S.Zareh 3, S. Sarkar 3, P.S.Tayebi 4 1 Department of Physics, Harbin Institute of Technology, Harbin 150080. P.R. China 2 Department of National Radiation Protection, Atomic Energy

More information

Radiation Protection Dosimetry Advance Access published April 22, 2008

Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 5 doi:10.1093/rpd/ncn144 EVALUATION OF EXPOSURE PARAMETERS IN PLAIN RADIOGRAPHY: A COMPARATIVE

More information

A Standardised Approach to Optimisation

A Standardised Approach to Optimisation The Radiographer vol. 51: 105-110 A Standardised Approach to Optimisation Helen M. Warren-Forward 1 & Ronald Beckhaus 2 ABSTRACT Purpose: Optimisation of radiographic images is said to have been obtained

More information

Patient dose in routine X-ray examinations in Yazd state

Patient dose in routine X-ray examinations in Yazd state Iran. J. Radiat. Res., 2004; 1(4): 199-204 Patient dose in routine X-ray examinations in Yazd state F. Bouzarjomehri Health Physics Dept. Shahid Sadoghi Univ. of Medical Sciences, Yazd, Iran ABSTRACT Background:

More information

Paediatrica Indonesiana

Paediatrica Indonesiana Paediatrica Indonesiana VOLUME 53 July NUMBER 4 Original Article Transcatheter vs. surgical closure of patent ductus arteriosus: outcomes and cost analysis Mulyadi M Djer, Mochammading, Mardjanis Said

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Dose Evaluation for Common Digital Radiographic s in Selected Hospitals in Pahang Malaysia Soo-Foon Moey 1*, Zubir Ahmad Shazli 1, Inayatullah Shah Sayed

More information

The interdependence of staff and patient doses in interventional radiology

The interdependence of staff and patient doses in interventional radiology T he British Journal of Radiology, 70 (1997), 498 503 1997 The British Institute of Radiology The interdependence of staff and patient doses in interventional radiology J R WILLIAMS, MSc, FIPEM Department

More information

Patient s Radiation Exposure in Coronary Angiography and Angioplasty: The Impact of Different Projections

Patient s Radiation Exposure in Coronary Angiography and Angioplasty: The Impact of Different Projections doi:.15171/jcvtr.14. http://journals.tbzmed.ac.ir/jcvtr TUOMS Publishing Group Original Article Patient s Radiation Exposure in Coronary Angiography and Angioplasty: The Impact of Different Projections

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

Introduction and Background

Introduction and Background CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section

More information

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science PATIENT DOSIMETRY IN DIAGNOSTIC RADIOLOGY MODALITIES Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science Ankara University Institute of Nuclear Science USE OF RADIATION! INCREASING? Natural

More information

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system Original article NOWOTWORY Journal of Oncology 2017, volume 67, number 2, 115 120 DOI: 10.5603/NJO.2017.0018 Polskie Towarzystwo Onkologiczne ISSN 0029 540X www.nowotwory.edu.pl Mean glandular dose values

More information

Assessment of radiation risk to pediatric patients undergoing conventional X-ray examinations

Assessment of radiation risk to pediatric patients undergoing conventional X-ray examinations Radioprotection 50(1), 19-25 (2015) c EDP Sciences 2015 DOI: 10.1051/radiopro/2014023 Available online at: www.radioprotection.org Article Assessment of radiation risk to pediatric patients undergoing

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2

More information

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Investigation of the clinical performance of a novel solid-state diagnostic dosimeter Jason Tse, a Donald McLean Medical Physics and

More information

The impact of SID and collimation on backscatter in radiography

The impact of SID and collimation on backscatter in radiography The impact of SID and collimation on backscatter in radiography Poster No.: C-2931 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: M. Joyce 1, J. T. Ryan 2, M. F. Mc Entee

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria

Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria Bidemi I.

More information

Effective radiation dosage of three-dimensional rotational angiography in children

Effective radiation dosage of three-dimensional rotational angiography in children Europace (2015) 17, 611 616 doi:10.1093/europace/euu207 CLINICAL RESEARCH Cardiac electrophysiology Effective radiation dosage of three-dimensional rotational angiography in children Marjolein Peters 1,

More information

Radiation Protection Dosimetry Advance Access published September 13, 2006

Radiation Protection Dosimetry Advance Access published September 13, 2006 Radiation Protection Dosimetry Advance Access published September 13, 26 Radiation Protection Dosimetry (26), 1 of 6 doi:1.193/rpd/ncl113 TECHNICAL NOTE RADIATION EXPOSURE AND DOSE EVALUATION IN INTRAORAL

More information

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Clarita Saldarriaga Vargas, Corinne Amalberto, Lara Struelens,

More information

Staff Exposure Monitoring in Interventional Radiology

Staff Exposure Monitoring in Interventional Radiology Conference on Physics in Medicine: From Diagnosis to Treatment "Enhancing safety and quality in radiation medicine". KFMC, Riyadh 7-9 November 2017 Staff Exposure Monitoring in Interventional Radiology

More information

Establishing the Greek National Reference Levels for Interventional Cardiology procedures

Establishing the Greek National Reference Levels for Interventional Cardiology procedures Establishing the Greek National Reference Levels for Interventional Cardiology procedures George Simantirakis*, Christina Koukorava, Maria Kalathaki, Christos Pa

More information

MEASUREMENT OF EFFECTIVE DOSE TO PATIENT DURING INTERVENTIONAL CARDIAC PROCEDURE

MEASUREMENT OF EFFECTIVE DOSE TO PATIENT DURING INTERVENTIONAL CARDIAC PROCEDURE J. Asiat. Soc. Bangladesh, Sci. 40(1): 1-7, June 2014 MEASUREMENT OF EFFECTIVE DOSE TO PATIENT DURING INTERVENTIONAL CARDIAC PROCEDURE M.M.M. SIRAZ 1, ALEYA BEGUM, R.K.KHAN, A.HOQUE AND A.BEGUM 2 Health

More information

Dosimetry in digital mammography

Dosimetry in digital mammography Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol

More information

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result. Minimizing Dose Understand the importance and methods of pre-procedure patient assessment including a review of previous radiologic exams, disease processes and anatomical considerations that may increase

More information

Assessment of effective dose in paediatric CT examinations

Assessment of effective dose in paediatric CT examinations Assessment of effective dose in paediatric CT examinations E. Dougeni 1,2 CL. Chapple 1, J. Willis 1, G. Panayiotakis 2 1 Regional Medical Physics Department, Freeman Hospital, Freeman Road, Newcastle

More information

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio Quality Control in Mammography: Results from a study in Recife, Brazil H. J. Khoury, V. S. de Barros, M. Sampaio Grupo de Dosimetria e Instrumentação, Universidade Federal de Pernambuco, Rua Prof. Luiz

More information

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Poster No.: C-2078 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Exhibit F. Salmani Rezaei,

More information

Kerma Area Product Method for Effective Dose Estimation During Lumbar Epidural Steroid Injection Procedures: Phantom Study

Kerma Area Product Method for Effective Dose Estimation During Lumbar Epidural Steroid Injection Procedures: Phantom Study Vascular and Interventional Radiology Original Research Kim et al. Dose for Lumbar Steroid Injection Vascular and Interventional Radiology Original Research Sangroh Kim 1 Greta Toncheva 2,3 Colin Anderson-Evans

More information

Interventional radiology, by virtue of its low invasiveness, is

Interventional radiology, by virtue of its low invasiveness, is ORIGINAL RESEARCH T. Moritake Y. Matsumaru T. Takigawa K. Nishizawa A. Matsumura K. Tsuboi Dose Measurement on Both Patients and Operators during Neurointerventional Procedures Using Photoluminescence

More information

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA Full Length Research Article 1 Ibrahim, U, 3 Daniel, I.H., 3 Ayaninola, O., 4 Ibrahim,

More information

RADIATION INJURY OF THE SKIN FOLLOWING DIAGNOSTIC AND INTERVENTIONAL FLUOROSCOPIC PROCEDURES

RADIATION INJURY OF THE SKIN FOLLOWING DIAGNOSTIC AND INTERVENTIONAL FLUOROSCOPIC PROCEDURES XA0101657 RADIATION INJURY OF THE SKIN FOLLOWING DIAGNOSTIC AND INTERVENTIONAL FLUOROSCOPIC PROCEDURES T.R. Koenig, L.K. Wagner Department of Radiology, The University of Texas - Houston Medical School,

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

International Journal of Radiation Research, January 2017

International Journal of Radiation Research, January 2017 Downloaded from ijrr.com at : +0330 on Tuesday November th Volume 5, No International Journal of Radiation Research, January 7 Radiation dose to the thyroid, eyes and parotid glands of patients undergoing

More information

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology Iran. J. Radiat. Res., 2006; 4 (1): 21-27 Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology M.T. Bahreyni

More information

Index. cardiology.theclinics.com. Note: Page numbers of article titles are in boldface type.

Index. cardiology.theclinics.com. Note: Page numbers of article titles are in boldface type. Index Note: Page numbers of article titles are in boldface type. A ACHD. See Adult congenital heart disease (ACHD) Adult congenital heart disease (ACHD), 503 512 across life span prevalence of, 504 506

More information

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.93 Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

More information

Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures

Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures The British Journal of Radiology, 84 (2011), 70 77 Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures 1 E P EFSTATHOPOULOS, PhD, 1 I PANTOS,

More information

Patient Dosimetry in Mammography and Tomosynthesis:

Patient Dosimetry in Mammography and Tomosynthesis: 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor

More information

EVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY L. Lança 1, * and A. Silva 2

EVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY L. Lança 1, * and A. Silva 2 Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 7 doi:10.1093/rpd/ncn143 EVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran Global Journal of Health Science; Vol. 7, No. 5; 2015 ISSN 1916-9736 E-ISSN 1916-9744 Published by Canadian Center of Science and Education Measurement of Entrance Skin Dose and Calculation of Effective

More information

Cardiac Computed Tomography

Cardiac Computed Tomography Cardiac Computed Tomography Authored and approved by Koen Nieman Stephan Achenbach Francesca Pugliese Bernard Cosyns Patrizio Lancellotti Anastasia Kitsiou Contents CARDIAC COMPUTED TOMOGRAPHY Page 1.

More information

Radiologic Units: What You Need to Know

Radiologic Units: What You Need to Know Radiologic Units: What You Need to Know TODD VAN AUKEN M.ED. RT (R)(MR) Agenda Greys, Sieverts, Coulombs per kg, & Becquerel's Conventional Units Other Concepts (LET, Q-Factor, Effective Dose, NCRP Report

More information

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Conventional and spiral CT dose indices in Yazd general hospitals, Iran Iran. J. Radiat. Res., 2006; 3 (4): 183-189 Conventional and spiral CT dose indices in Yazd general hospitals, Iran F. Bouzarjomehri 1*,M.H.Zare 2, D. Shahbazi 2 1 Department of Medical Physics, Shahid

More information

OBJECTIVES BACKGROUND METHODS RESULTS CONCLUSIONS

OBJECTIVES BACKGROUND METHODS RESULTS CONCLUSIONS Journal of the American College of Cardiology Vol. 33, No. 2, 1999 1999 by the American College of Cardiology ISSN 0735-1097/99/$20.00 Published by Elsevier Science Inc. PII S0735-1097(98)00591-9 Real-time

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

Radiation protection in the cardiac catheterization laboratory: special focus on the role of the operator

Radiation protection in the cardiac catheterization laboratory: special focus on the role of the operator special report Radiation protection in the cardiac catheterization laboratory: special focus on the role of the operator The use of ionizing radiation is associated with a risk of inducing malignant disease

More information

Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks

Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks Poster No.: C-2993 Congress: ECR 2010 Type: Topic: Scientific Exhibit

More information

Radiation protection for patient and staff during routine EVAR and TEVAR procedures Miltos Matsagkas, MD, PhD, FEBVS

Radiation protection for patient and staff during routine EVAR and TEVAR procedures Miltos Matsagkas, MD, PhD, FEBVS Radiation protection for patient and staff during routine EVAR and TEVAR procedures Miltos Matsagkas, MD, PhD, FEBVS Professor of Vascular Surgery University of Thessaly, Larissa, Greece Disclosure of

More information

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433 - 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.

More information

Tyne NE4 6BE, UK. Corresponding author:

Tyne NE4 6BE, UK. Corresponding author: Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 54 58 doi:10.1093/rpd/nci719 Advance Access published on February 3, 2006 REFERENCE LEVELS IN PTCA AS A FUNCTION OF PROCEDURE COMPLEXITY A.

More information

The Heart Center. Quality Counts: Cardiothoracic Surgery and Interventional Cardiology

The Heart Center. Quality Counts: Cardiothoracic Surgery and Interventional Cardiology The Heart Center Quality Counts: Cardiothoracic Surgery and Interventional Cardiology The Cardiothoracic Surgery Program at Nationwide Children s Hospital is dedicated to the treatment of all patients,

More information

The effect of breast shielding during lumbar spine radiography

The effect of breast shielding during lumbar spine radiography 26 research article The effect of breast shielding during lumbar spine radiography Nejc Mekis 1, Dejan Zontar 2, Damijan Skrk 2 1 University of Ljubljana, Faculty of Health Sciences, Medical Imaging and

More information

Aortic Coarctation: Evaluation with Computed Tomography Angiography in Pediatric Patients

Aortic Coarctation: Evaluation with Computed Tomography Angiography in Pediatric Patients Med. J. Cairo Univ., Vol. 83, No. 2, June: 63-70, 2015 www.medicaljournalofcairouniversity.net Aortic Coarctation: Evaluation with Computed Tomography Angiography in Pediatric Patients MOHAMED ZAKI, M.D.

More information

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME Number of mammography X-ray sets Radiation Protection Dosimetry (year), Vol. 0, No. 0, pp. 0 0 DOI: 10.1093/rpd/nc0000 SECTION TITLE HERE PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY

More information

Recent technical advances and increasing experience

Recent technical advances and increasing experience Pediatric Open Heart Operations Without Diagnostic Cardiac Catheterization Jean-Pierre Pfammatter, MD, Pascal A. Berdat, MD, Thierry P. Carrel, MD, and Franco P. Stocker, MD Division of Pediatric Cardiology,

More information