Safe magnetic resonance image scanning of the pacemaker patient: current technologies and future directions

Size: px
Start display at page:

Download "Safe magnetic resonance image scanning of the pacemaker patient: current technologies and future directions"

Transcription

1 Europace (2012) 14, doi: /europace/eur391 REVIEW Safe magnetic resonance image scanning of the pacemaker patient: current technologies and future directions Werner Jung*, Vlada Zvereva, Bajram Hajredini, and Sebastian Jäckle Department of Cardiology, Academic Teaching Hospital of the University of Freiburg, Schwarzwald-Baar Klinikum, Vöhrenbacher Street 23, Villingen-Schwenningen, Germany Received 24 July 2011; accepted after revision 23 vember 2011; online publish-ahead-of-print 10 January 2012 Magnetic resonance imaging (MRI) is the imaging modality of choice in many clinical situations, and its use is likely to grow due to expanding indications and an ageing population. Many patients with implantable devices are denied MRI except in cases of urgent need, and when scans must be performed they are complicated by the need for burdensome and costly personnel and monitoring requirements that have the net effect of restricting access to scans. Several small studies, enrolling a total of 344 patients, suggest that some patients with conventional systems may undergo MR examinations without clinically overt adverse events. However, a number of potential interactions exist between implantable cardiac devices and the static and gradient magnetic fields and modulated radio frequency (RF) fields generated during MR scans; nearly all studies have reported pacing capture threshold changes, troponin elevations, ectopy, unpredictable reed switch behaviour, and other subclinical issues with pacemakers and implantable cardioverter-defibrillators (ICDs) in patients who have undergone MRI. Attention has turned to devices that are specifically designed to be safe in the MRI environment. A clinical study of one such device documented its ability to be exposed to MRI in a 1.5 T scanner without adverse impact on patient outcomes or pacemaker system function. Such new technologies may enable scanning of pacemaker and ICD patients with reduced concerns regarding the shortand long-term effects of MRI. As importantly, these devices may increase the number of centres that are able to safely perform MRI and, thus, expand access to scans for patients with these devices Keywords Pacemakers MRI Safety Implantable defibrillators Introduction Magnetic resonance imaging (MRI) has rapidly become the imaging modality of choice in many therapeutic areas as a result of its ability to provide superior soft tissue contrast without exposing patients to the risks associated with invasive procedures, ionizing radiation, or contrast agents. For these reasons, the use of MRI is likely to grow due to the confluence of expanding indications for its use and an ageing population with a corresponding increase in disease states that could benefit from MRI. 1 3 Worldwide, 5 million patients are implanted with a pacemaker or implantable cardioverter-defibrillator (ICD). As these patients are generally older and sicker than the general population, estimates suggest that these patients have between a 50 and 75% likelihood of having a clinical indication for MRI over the lifetime of the device. 1 In total, it has been estimated that patients with devices had an indication for an MRI scan in As the majority of current implantable cardiac devices is not approved for use in the MR environment, many patients with implantable devices are excluded from MRI except in cases of urgent need. Current US guidelines for safe MRI scanning strongly discourage MR examination of pacemaker and ICD patients, except in cases of urgent need. 4 Similarly, current European guidelines for safe MRI indicate that patients should have life-threatening or severely quality-of-life- limiting conditions. 5 In both guidelines, exceptional scans can be performed under stringent safety conditions with documentation of informed consent and careful analysis of the risks and benefits of treatment, provision of specialized personnel and emergency medical equipment, monitoring during the procedure, and extensive follow-up. The conditions imposed by current guidelines are burdensome and costly, and have the net effect of dramatically restricting access to a potentially beneficial diagnostic modality in a clinically important patient population. Several questions must be asked: first, what do * Corresponding author. Tel: ; fax: , werner.jung@sbk-vs.de Published on behalf of the European Society of Cardiology. All rights reserved. & The Author For permissions please journals.permissions@oup.com.

2 632 W. Jung et al. Table 1 Indications for MRI in patients with pacemakers/implantable cardiac defibrillators a Patient group European recommendation US recommendation... Pacemaker-dependent patients (very high risk) ICD patients (non-dependent) (high risk) Pacemaker patient (non-dependent) (low risk) If underlying rhythm is too slow, reconsider indication. The threshold for imaging and safety requirements are higher, but no absolute contraindication The patient must have a documented, extremely serious, life-threatening, or severely quality-of-life- limiting condition The patient must have a documented, very serious, life-threatening, or severely quality-of-life-limiting condition MRI should not be performed unless there are highly compelling circumstances and when the benefits clearly outweigh the risks MRI should not be performed unless there are highly compelling circumstances and when the benefits clearly outweigh the risks MRI is discouraged and should only be considered in cases in which there is a strong clinical indication and in which the benefits clearly outweigh the risks ICD, Implantable cardiac defibrillators; MRI, magnetic resonance imaging. a Data from Levine et al. 4 and Rougin et al. 5 current guidelines say about safe MRI in the pacemaker patient? Second, do these guidelines represent an overreaction based on current data? Third, what can we do to ensure safe scans in patients implanted with current devices? Fourth and finally what is the role of new MRI-conditional technologies in the context of the current clinical environment? This review will explore each of these issues. What do current guidelines say about safemagnetic resonance imaging? Recommendations exist to guide MRI scanning in patients with pacemakers and ICDs. In general, both US and European guidelines do not impose absolute contraindications against MRI. The European guidelines strongly recommend against MRI in pacemaker patients, except in situations where the patients have a documented extremely serious, life-threatening, or severely quality-of-lifelimiting condition (Table 1). 5 Similarly, the American Heart Association (AHA) guidelines indicate that MRI examination of non-pacemaker-dependent patients is discouraged, except in cases with a strong clinical indication and in which the benefits clearly outweigh the risks. 4 Among pacemaker-dependent patients and those with ICDs, MRI examination is generally not recommended except under highly compelling circumstances. Both sets of guidelines impose stringent monitoring requirements. In the European guidelines an advanced cardiac life supportcertified health-care provider must be present during the entire exam to monitor the patient and perform cardiac life support if needed. 5 A cardiologist and a pacemaker/icd programmer should be present during the scan. Monitoring requirements include electrocardiogram (ECG), pulse oximetry, and in some cases non-invasive blood pressure measurement and breathing sensors. A crash cart with an external defibrillator-pacemaker must be present during the scan. Following examination of pacemaker patients, a physician with electrophysiology experience should interrogate the pacemaker and reprogram as needed. 4 In patients with ICDs, the physician should perform post-scan device reprogramming and defibrillation threshold testing. Are we overreacting to potential hazards? Questions remain regarding whether safe scans can be conducted in patients implanted with current devices. Several small studies have been published concerning newer conventional pacemaker systems; these reports suggest that some patients may undergo MR examinations without clinically overt adverse events and, potentially, without burdensome monitoring requirements. In a report by Martin et al., 6 54 non-pacemaker-dependent patients underwent a total of 62 MRI examinations at 1.5 T. Restrictions were not imposed on the type of pacemaker present in the patient, nor were limitations placed on the type of MRI examination. Scans included cardiac, vascular, and general MRI studies using a range of whole-body averaged specific absorption rates (SARs). Overall, a total of 107 leads and 61 pulse generators were evaluated. overt adverse events occurred in this study. In a second study conducted by Sommer et al., 7 82 pacemaker patients underwent a total of 115 MRI examinations at 1.5 T; radiofrequency (RF)-induced lead heating was minimized by limiting the specific absorption rate to 1.5 W/kg. As in the Martin study, all MR examinations were completed safely. Similarly, in a third study conducted by Nazarian et al. 8 in 55 patients (of whom 31 had permanent pacemakers and 24 had ICDs), again, there were no overt clinically relevant adverse events. In a recent study conducted by Mollerus et al., 9 scans were successfully completed in 52 non-pacemaker-dependent patients with no immediate adverse effects. Several additional small studies have evaluated safety at low field strength (0.5 T) with no serious clinical repercussions. 7,10,11 Potential adverse events associated with magnetic resonance imaging It is worthwhile to note that only about 1500 scans of patients with pacemakers or ICDs have been reported in the literature in any form, 12 far too few to completely allay concerns about MR scans in patients implanted with conventional devices. Many of these

3 Safe MRI scanning of the pacemaker patient 633 reports consist of very small series and case reports. At the time of this writing, well-designed clinical studies reported in reputable venues have included a grand total of 344 patients, of whom 81 were studied at a field strength of 0.5 T, 7,10,11 13 at a field strength of 2.0 T, 13 and 250 at a field strength of 1.5 T. 6,8,9,14,15 A number of potential interactions exist between implantable cardiac devices and the static and gradient magnetic fields and modulated RF fields generated during MR scans, including mechanical forces on ferromagnetic components, RF-induced heating of leads, unintended cardiac stimulation, interference with pacemaker function, and electrical reset. Force and torque on the implanted device is a concern that has been reduced due to the lower ferromagnetic content of current pacemakers. Studies conducted at 1.5 T found that the force in pacemakers and ICDs at this field strength ranges from 0.05 to 3.6 N and 1.0 to 5.9 N. 16 tably, the amount of force generated at 1.5 N was dependent upon not only on the amount of ferromagnetic material employed in the design of the device, but also the year the device was manufactured. These data illustrate the considerable advances that have been made in conventional pacemaker design over the last decade. Increases in pacing capture thresholds attributable to heating of pacemaker leads during MRI are common enough to be potentially clinically significant. In vitro temperature increases of up to 63.18C have been observed at a field strength of 1.5 T; animal studies have demonstrated increases of up to 20.88C at 1.5 T. 17,18 In the study conducted by Martin et al., 6 40 (37%) leads had pacing threshold changes, and 10 (9.4%) leads underwent significant change (defined as a change.1 voltage or pulse width increment or decrement). Two leads required a change in programmed output. In the Sommer study, scans were conducted with significant precautions against RF-induced heating, including limiting the specific absorption rate to,1.5 W/kg, and excluding anatomic regions with full coverage of the pacemaker lead loop, including the thoracic spine, heart, and breasts. 15 The total active scan time was also limited to 30 min. Despite these precautions, cardiac troponins were increased in 4 of 114 examinations; in one case these increases were associated with a significant increase in pacingcapture threshold. 15 Pacing-capture threshold changes 1.0 V were identified in six cases, and statistical analysis showed that there was a significant increase in pacing-capture threshold from pre- to post-mri (P ¼ 0.017). Possible long-term effects of scanning were observed in two cases, in which increased pacingcapture threshold was not detected until the 3 month follow-up. These changes were considered likely to be related to scar tissue development around the lead tips as a result of RF-related thermal injury. The behaviour of reed switches (open or closed) and the pacemaker (synchronous or asynchronous) in conventional devices are often not predictable in the strong static magnetic field of the MR device. This has potentially important implications, particularly in the pacemaker-dependent patient. The state of the reed switch in various orientations and positions in the magnetic fields of 0.5, 1.5, and 3.0 T was evaluated by Luechinger et al. 19 When oriented parallel to magnetic fields, reed switches closed at mt and opened at mt. In low magnetic fields (,50 mt), reed switches were closed and in high magnetic fields (.200 mt), reed switches opened in 50% of orientations. These data are supported by the results of the Sommer study, in which the reed switch remained inactivated in 44.7% of patients. 15 Electrical reset, an emergency safety feature that guarantees minimal pacemaker function in the case of battery voltage dips, has been demonstrated to occur in patients with implanted devices subjected to MRI. Electrical reset presents an important safety hazard for two reasons. First, if reset occurs concomitant with an open reed switch, bradycardia/asystole may occur in patients with low intrinsic heart rates as a result of inhibition of pacemaker output by time-varying gradient fields. Second, the default pacing mode may not provide adequate functionality for some patients. 5 In the study conducted by Sommer et al., 15 post-mri interrogation showed that electrical pacemaker reset occurred in 7 (6.1%) examinations. In all cases, however, the pacemaker could be reprogrammed to the parameters used prior to the MRI scan. Ectopy has been reported in patients with pacemakers and ICDs. In a study conducted by Mollerus et al., 9 52 nonpacemaker-dependent patients with a total of 119 leads underwent 59 MRI scans of any landmark. Scans were conducted using usual protocols with standard peak-specific absorption rates for the scan. Both telemetry and pulse oximetry plethysmographic waveform were observed throughout the scans for ectopy. Seven patients had ectopy observed either on telemetry or on observation of the oxygen saturation plethysmographic waveform. According to the investigators, ectopy was likely due to normal device noise-rejection behaviour. Solution: a protocol for performing safe magnetic resonance imaging in patients with conventional devices Nazarian and colleagues recently published the safety protocol used in their institution to improve safety in patients with pacemakers and ICDs (Figure 1). The authors recommend that device generators prone to electromagnetic interference (generally older devices), as well as patients with leads that are prone to movement (e.g. patients with,6 weeks time since device implant and those with no fixation), be excluded. Scanning should also be avoided in patients with device leads that are prone to healing (e.g. non-transvenous epicardial and capped leads), and pacemaker-dependent patients with ICDs. The authors suggest programming to an asynchronous, dedicated pacing mode in pacemaker-dependent patients to reduce the risk of inappropriate inhibition of pacing due to detection of radiofrequency pulses. To avoid inappropriate activation of pacing in non-pacemaker-dependent patients, the authors suggest programming to a non-tracking ventricular or dual-chamber inhibited pacing mode and deactivation of rate response, premature ventricular contraction response, ventricular sense response, and conducted atrial fibrillation response to ensure that sensing of vibrations or radiofrequency pulses does not lead to unwarranted

4 634 W. Jung et al. Device with Satisfactory Previous MRI Testing? Pacemakers with Satisfactory Previous MRI Testing Device Implant >6 Weeks Prior to MRI? Alternative Tests St. Jude Pacesetter AFP (262) Trilogy (2360) Entity (5326) Affinity (5130, 5330) Integrity (5142, 5342, 5346) Identity (5172, 5370, 5376, 5380, 5386) Absense of n- Transvenous Epicardial, Abandoned, and - Fixation Leads? Guidant Vigor (1232) Discovery (1272) Insignia (1194, 1290) Medtronic EnPulse (AT-500, E2SRO1, E2DRO1) Kappa (701, 901) Prodigy (7860) InSync BiV (8040, 8042) Check Device Parameters: Lead Impedance and Threshold, P/R Wave Amplitude, and Battery Voltage ICD Monitoring and Tachy-Arrhythmia Therapies (Anti-Tachycardia Pacing / Defibrillation) - Off ICDs with Satisfactory Previous MRI Testing Pacemaker St. Jude Photon (V-194, V-230, V-232) Atlas (V-240) Epic (V-197, V-235, V-239) Pacemaker Dependent? Pacemaker Dependent? Guidant Prizm (1850, 1851, 1852, 1860, 1861) Contak (1823, H119, H170, H175) Vitality (T125, T135) VOO / DOO VVI / DDI Medtronic Maximo (7232) Gem-II (7273) Gem-III (7275) Marquis (7274) InSync (7272) Magnet, Rate, PVC, ise, Ventricular Sense, Conducted Atrial Fibrillation Response - Off Estimated SAR of Sequence >2.0 W/kg Monitor Blood Pressure, ECG, Oxygen Saturation and Symptoms During MRI Limit SAR by Increasing Repetition Time or Field of View, and/or Reducing Flip Angle or the Acquisition Bandwidth Re-Check Device Parameters: Lead Impedence and Threshold, P/R Wave Amplitude, and Battery Voltage Figure 1 A safety protocol for performing MRI in patients implanted with conventional devices. Reproduced from Nazarian and Halperin. 20 ECG, electrocardiography; MRI, magnetic resonance imaging; PVC, premature ventricular complex; SAR, specific absorption rate. pacing. When possible, asynchronous pacing is minimized in non-pacemaker-dependent patients through deactivation of the magnet mode. The authors also usually deactivate tachyarrhythmia monitoring to avoid battery drainage due to recording of multiple RF pulse sequences as arrhythmic episodes. Because reed-switch function can be unpredictable during MRI, therapies should be

5 Safe MRI scanning of the pacemaker patient 635 disabled to avoid unwarranted antitachycardia pacing or shocks. Finally, to reduce the risk of thermal injury and changes in lead threshold and impedance, the authors recommend limiting the estimated whole-body, averaged specific absorption rate of MRI sequences to,2.0 W/kg. The protocol also has provisions for extensive monitoring during MRI. Blood pressure, ECG, pulse oximetry, and symptoms should be monitored for the duration of the examination. A radiologist and cardiac electrophysiologist (or an individual trained in advanced cardiac life support familiar with device programming and troubleshooting) are present at all scans. At the end of the examination, all device parameters should be checked and programming should be restored to pre-mri settings. Another solution: can we be magnetic resonance imaging safe by design? While it appears that MRI scans can be performed in some device patients without immediate clinical consequences, a sufficient number of patients have not yet been evaluated using these protocols, nor has there been sufficient long-term follow-up to determine if there are long-term effects on outcomes. Moreover, these protocols limit scans to centres with substantial experience in this patient population, effectively limiting access to MRI for a substantial percentage of the population. Recently, attention has turned to devices that are safe by design. These devices have extensive design modifications to improve MRI compatibility. One such system includes a heavily modified pulse generator (EnRhythm MRI SureScan) and leads (CapSureFix MRI leads) that together are designed specifically to mitigate the potential hazards associated with MR scans. Design modifications to the device and leads include (i) modification of the leads to reduce the potential for radiofrequency lead-tip heating; (ii) revision of the internal circuitry to reduce the potential for cardiac stimulation; (iii) reductions in the amount of ferromagnetic materials in the device; (iv) changes to internal circuit protection to prevent disruption of the internal power supply; and (v) replacement of the reed switch with a Hall sensor to improve predictability in static magnetic fields. 21 In addition, a dedicated programming care pathway was developed to facilitate the choice between asynchronous vs. non-stimulation modes, increase the pacing output to 5.0 V/1.0 ms during MRI scanning, prevent programming to the MRI mode in the absence of positive systemintegrity checks, and simplify restoration of pre-scan program states and values. Initial clinical experience with the SureScan system compared with conventional dual-chamber devices was reported in 2010 in a small, single-centre feasibility study. 22 In this study, 107 consecutive patients were implanted with either the MRI-compatible device or a conventional dual-chamber, active-fixation lead, non-mri system. Implantation success was 100% for both groups; lead cephalic and subclavian access were 63.0 and 37.0%, respectively, for MRI patients and 70.2 and 29.8%, respectively, for those with the non-mri system; subclavian vein puncture was required to place at least one lead in 40% of MRI patients and 31.6% of conventional-system patents. There was a non-significant trend towards shorter procedural times with the conventional system. complications were observed at a median follow-up of 6.8 months. The EnRhythm device was recently evaluated in a large-scale, prospective, multicentre, randomized study published in In this study, patients were enrolled who met Class I or II dualchamber, pacemaker-implant indications according to current American College of Cardiology (ACC)/AHA/Heart Rhythm Society (HRS) guidelines; patients could be pacemaker-dependent or non-pacemaker-dependent. Following successful implant, patients (N ¼ 464) were randomized to undergo an MRI scan between 9 and 12 weeks post-implant or not to undergo MRI. Clinical evaluation of pacemaker-system function was conducted immediately before and after the scan and at 1 week and 1 month post-mri. The primary study end points were the MRI procedure complication-free rate and capture and sensing performance between the MRI and control groups. A total of 464 patients were randomized after successful implantation; of these, 211 patients underwent MRI and completed the 1 month post-mri visit. The complication-free rate was 100% (e.g. no patient experienced MRI-related complications). inhibition of pacemaker output, asystole, sustained ventricular arrhythmias, unexpected changes of heart rate, or electrical resets occurred during MRI. At 1 month following MRI scanning, the expected rate of pacemaker system-related complications was expected to be 80% and was measured to be 91.7% (P, vs. the expected rate). System-related complications at 1 month included lead dislodgement (n ¼ 17), elevated capture thresholds (n ¼ 9), pericardial effusion (n ¼ 3), and failure to capture (n ¼ 3). ne of these complications were judged related to the scan. tably, there were no differences between the MRI and control groups in the proportions of patients who experienced an increase in pacing capture threshold, the proportion of patients who maintained the sensed electrogram amplitudes above 1.5 mv (atrial) or 5 mv (ventricle), and there were no differences between the two groups in impedance results. It is important to note that this study examined an MRIconditional scenario in that the static magnetic field strength was limited to 1.5 T, with a maximum specific absorption rate value of 2 W/kg for each sequence and a maximum gradient slew rate of 200 T/m/s. The position of the isocentre of the RF transmitter coil was required to be above the superior surface of the C1 vertebra or below the inferior surface of the body of T12. conclusions can be drawn from this study about the safety of scans conducted between these landmarks; however, these restrictions still permit scanning of the complete cervical region and most of the thoracic region by widening the field of view. As noted previously, conventional pacemakers are associated with cardiac myocardial injury even when used under carefully controlled conditions as indicated by increases in troponin I levels and increases in pacing capture thresholds. This study is notable in that only one increase in pacing capture threshold was observed (0.2%), compared with 3.1% in the study by Sommer et al. 15 and 9.4% in the report published by Martin et al. 6 Although neither of these trials demonstrated immediate clinical implications associated

6 636 W. Jung et al. with this damage, it is reasonable to assume that such damage may influence intermediate- and long-term outcomes. These data suggest that the design of the MRI-conditional system has the potential to dramatically reduce the incidence of subclinical myocardial injury, although the longer-term implications of this effect remain to be described. Conclusions The current literature suggests that MRI scans can be conducted safely in patients with conventional ICDs. However, it is important to note that only about 1500 scans have been reported in the medical literature; 12 of these, prospective clinical studies designed specifically to assess the safety of MRI in this patient population have included a total of 344 patients, of whom a significant percentage were evaluated at 0.5 T. To consider a drug, device, or procedure safe after successful scans in 344 patients conducted in centres with considerable expertise is a considerably lower standard than is usual in medicine and is unlikely to reflect the rates of complications in real-world clinical practice. Moreover, while all scans were completed without overt clinical adverse events occurring during the scan or during follow-up periods of up to 6 months, potentially clinically relevant changes in pacemaker/icd and lead function including changes in pacing capture threshold, increases in cardiac troponin levels, electrical reset, changes in battery voltage, and unpredictable reed switch behaviour were observed in all studies. As a result of these concerns, labelling of conventional devices is conservative and cautions physicians against the use of MRI. An editorial, written on behalf of the US Food and Drug Administration (FDA), indicates that while the FDA recognizes that there are pacemaker and ICD patients for whom, on a case-by-case basis, the diagnostic benefit from MRI outweighs the presumed risks...those risks have not yet been characterized and mitigated sufficiently to justify the routine use of MRI examination in those populations. 23 New technologies may enable scanning of pacemaker and ICD patients with reduced concerns regarding the short- and long-term effects of MRI. These devices may increase the number of centres that are able to safely perform MRI and thus expand access to scans for patients with these devices. Furthermore, burdensome monitoring requirements may be reduced after sufficient realworld clinical expertise has been gained with MRI-conditional devices, expanding access and reducing the costs of scanning these patients. It must, however, be emphasized that these devices have been adequately evaluated only at field strengths of up to 1.5 T and as 3.0 T MR scanners come into broader use there is an urgent need to evaluate the safety of scanning these devices at higher field strengths. Given the acknowledged value of MRI in a broad range of therapeutic areas as well as the increasing number of pacemaker/icd patients who could benefit from MRI it is clear that new technologies that can improve the safety of scans are to be welcomed. Acknowledgement The authors acknowledge the editorial assistance of John R. Ferguson and Tara Fish, whose services were funded by Medtronic, Inc. The authors meet the criteria for authorship as recommended by the International Committee of Medical Journal Editors (ICMJE) and were fully responsible for the content and editorial decisions. Conflict of interest: none declared. References 1. Kalin R, Stanton MS. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. Pacing Clin Electrophysiol 2005;28: Harris CH. MRI Procedure Growth Rate Slows, Study Shows. Business & Industry Digital Community Web site. sub=bai&pag=dis&itemid=81355 (15 September 2010, date last accessed). 3. Sutton R, Kanal E, Wilkoff BL, Bellow D, Luechinger R, Jenniskens I et al. Safety of magnetic resonance imaging of patients with a new Medtronic EnRhythm MRI SureScan pacing system: clinical study design. Trials 2008;9: Levine GN, Gomes AS, Arai AE, Bluemke DA, Flamm SD, Kanal E et al. Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology, and the Council on Cardiovascular Radiology and Intervention: endorsed by the American College of Cardiology Foundation, the rth American Society for Cardiac Imaging, and the Society for Cardiovascular Magnetic Resonance. Circulation 2007;116: Roguin A, Schwitter J, Vahlhaus C, Lombardi M, Brugada J, Vardas P et al. Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. Europace 2008;10: Martin ET, Coman JA, Shellock FG, Pulling CC, Fair R, Jenkins K. Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla. J Am Coll Cardiol 2004;43: Sommer T, Vahlhaus C, Lauck G, Smekal Av, Reinke M, Hofer U et al. MR imaging and cardiac pacemakers: in-vitro evaluation and in-vivo studies in 51 patients at 0.5 T. Radiology 2000;215: Nazarian S, Roguin A, Zviman MM, Lardo AC, Dickfeld TL, Calkins H et al. Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 tesla. Circulation 2006;114: Mollerus M, Albin G, Lipinski M, Lucca J. Ectopy in patients with permanent pacemakers and implantable cardioverter-defibrillators undergoing an MRI scan. Pacing Clin Electrophysiol 2009;32: Gimbel JR, Johnson D, Levine PA, Wilkoff BL. Safe performance of magnetic resonance imaging on five patients with permanent cardiac pacemakers. Pacing Clin Electrophysiol 1996;19: Vahlhaus C, Sommer T, Lewalter T, Schimpf R, Schumacher B, Jung W et al. Interference with cardiac pacemakers by magnetic resonance imaging: are there irreversible changes at 0.5 Tesla? Pacing Clin Electrophysiol 2001;24: Gimbel JR. The safety of MRI scanning of pacemakers and ICDs: what are the critical elements of safe scanning? Ask me again at Europace 2010;12: Del Ojo JL, Moya F, Villalba J, Sanz O, Garcia PD, Pastor L. Is magnetic resonance imaging safe in cardiac pacemaker recipients? Pacing Clin Electrophysiol 2005;28: Gimbel JR, Kanal E, Schwartz KM, Wilkoff BL. Outcome of magnetic resonance imaging (MRI) in selected patients with implantable cardioverter defibrillators (ICDs). Pacing Clin Electrophysiol 2005;28: Sommer T, Naehle CP, Yang A, Zeijlemaker V, Hackenbroch M, Schmiedel A et al. Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: a prospective study with 115 examinations. Circulation 2006;114: Luechinger R, Duru F, Scheidegger MB, Boesiger P, Candinas R. Force and torque effects of a 1.5-Tesla MRI scanner on cardiac pacemakers and ICDs. Pacing Clin Electrophysiol 2001;24: Achenbach S, Moshage W, Diem B, Bieberle T, Schibgilla V, Bachmann K. Effects of magnetic resonance imaging on cardiac pacemakers and electrodes. Am Heart J 1997;134: Luechinger R, Zeijlemaker VA, Pedersen EM, Mortensen P, Falk E, Duru F et al. In vivo heating of pacemaker leads during magnetic resonance imaging. Eur Heart J 2005;26: ; discussion Luechinger R, Duru F, Zeijlemaker VA, Scheidegger MB, Boesiger P, Candinas R. Pacemaker reed switch behavior in 0.5, 1.5, and 3.0 Tesla magnetic resonance

7 Safe MRI scanning of the pacemaker patient 637 imaging units: are reed switches always closed in strong magnetic fields? Pacing Clin Electrophysiol 2002;25: Nazarian S, Halperin HR. How to perform magnetic resonance imaging on patients with implantable cardiac arrhythmia devices. Heart Rhythm 2009;6: Wilkoff BL, Bello D, Taborsky M, Vymazal J, Kanal E, Heuer H et al. Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environment. Heart Rhythm 2010;8: Forleo GB, Santini L, Della Rocca DG, Romano V, Papavasilieiou LP, Magliano G et al. Safety and efficacy of a new magnetic resonance imaging-compatible pacing system: early results of a prospective comparison with conventional dualchamber implant outcomes. Heart Rhythm 2010;7: Faris OP, Shein M. Food and Drug Administration perspective: magnetic resonance imaging of pacemaker and implantable cardioverter-defibrillator patients. Circulation 2006;114:

Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions

Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions Europace (2010) 12, 947 951 doi:10.1093/europace/euq092 CLINICAL RESEARCH Pacing and CRT Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption

More information

MRI conditional devices. A luxury or a real clinical need? Sophie Mavrogeni, MD, FESC Onassis Cardiac Surgery Center, Athens, Greece

MRI conditional devices. A luxury or a real clinical need? Sophie Mavrogeni, MD, FESC Onassis Cardiac Surgery Center, Athens, Greece MRI conditional devices. A luxury or a real clinical need? Sophie Mavrogeni, MD, FESC Onassis Cardiac Surgery Center, Athens, Greece Increase of MRI scans in recent decades In the USA, the number of procedures

More information

MRI and CIEDs. Disclosures

MRI and CIEDs. Disclosures Date Clinical Title MRI and CIEDs C. p. Lao Raul Weiss MD, FAHA, FACC, FHRS and CCDS Director, Electrophysiology Fellowship Program Professor of Medicine The Ohio State University Wexner Medical Center

More information

STANDARD OPERATING PROCEDURE (SOP)

STANDARD OPERATING PROCEDURE (SOP) STANDARD OPERATING PROCEDURE (SOP) MRI Scanning for Patients with Cardiac Implantable Electronic Devices (Pacemakers and Implantable Cardiac Defibrillators) Effective date 05/04/2018 Version number V1.0

More information

It is estimated that after implantation, up to 75% of patients. Original Research

It is estimated that after implantation, up to 75% of patients. Original Research Annals of Internal Medicine Original Research A Prospective Evaluation of a Protocol for Magnetic Resonance Imaging of Patients With Implanted Cardiac Devices Saman Nazarian, MD; Rozann Hansford, RN, MPH;

More information

MRI for Patients with MRI-Conditional Pacing System: Radiographers Role. Lawrance Yip DM, DR, QMH

MRI for Patients with MRI-Conditional Pacing System: Radiographers Role. Lawrance Yip DM, DR, QMH MRI for Patients with MRI-Conditional Pacing System: Radiographers Role Lawrance Yip DM, DR, QMH MRI & Pacemaker MRI is a crucial and growing imaging modality Plays an important role in clinical management

More information

Safety and Imaging Quality of MRI in Pediatric and Adult Congenital Heart Disease Patients with Pacemakers

Safety and Imaging Quality of MRI in Pediatric and Adult Congenital Heart Disease Patients with Pacemakers Safety and Imaging Quality of MRI in Pediatric and Adult Congenital Heart Disease Patients with Pacemakers AARON F. PULVER, M.D.,* MICHAEL D. PUCHALSKI, M.D.,* DAVID J. BRADLEY, M.D., L. LUANN MINICH,

More information

Lead Management for MRI

Lead Management for MRI Lead Management for MRI Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services The Ohio State University Medical Center Columbus, OH USA President, International

More information

Devices and EP- Safety in MRI

Devices and EP- Safety in MRI Devices and EP- Safety in MRI Mohammad Shenasa M.D. FACC, FHRS, FESC O Connor Hospital San Jose, CA Biblioteca-Alexandria Alexandria-Egypt Cardio Alex June 10-13, 2014 Effect of External Electrical and

More information

Cardiovascular magnetic resonance with an MR compatible pacemaker

Cardiovascular magnetic resonance with an MR compatible pacemaker Washington University School of Medicine Digital Commons@Becker Open Access Publications 2013 Cardiovascular magnetic resonance with an MR compatible pacemaker Anita R. Bhandiwad Kristopher W. Cummings

More information

MRI ACCESS. Evera MRI SureScan ICD Systems

MRI ACCESS. Evera MRI SureScan ICD Systems ACCESS Evera SureScan ICD Systems ACCESS. 36% of ICD patients are likely to have an ordered over 4 years 3 Introducing Evera, featuring SureScan Technology. ICD Patients are not receiving s 1.4% 36% CONTOURED.

More information

MRI SureScan pacemakers & ICDs what you need to know

MRI SureScan pacemakers & ICDs what you need to know MRI SureScan pacemakers & ICDs what you need to know Many variables can affect the device patient s likelihood of risks w/ an MRI Type of imaging sequence Type of scanner MRI scan duration Strength of

More information

12 th International Winter Arrhythmia School Collingwood, Ontario February 7, Dr. Derek Yung. MD, FRCPC

12 th International Winter Arrhythmia School Collingwood, Ontario February 7, Dr. Derek Yung. MD, FRCPC 12 th International Winter Arrhythmia School Collingwood, Ontario February 7, 2015 Dr. Derek Yung. MD, FRCPC Presenter: Dr. Derek Yung Relationships with industry: Educational material provided by Biotronik,

More information

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Patient Scheduling and Screening Q: Is there any special (other than regular MRI consent) consent form that should be signed by the patient?

More information

Cardiac magnetic resonance imaging at 1.5 T in patients with cardiac rhythm devices

Cardiac magnetic resonance imaging at 1.5 T in patients with cardiac rhythm devices Europace (2011) 13, 533 538 doi:10.1093/europace/euq501 CLINICAL RESEARCH Pacing and Resynchronization Therapy Cardiac magnetic resonance imaging at 1.5 T in patients with cardiac rhythm devices Francisco

More information

Advisa MRI and Ensura MRI SureScan pacing systems

Advisa MRI and Ensura MRI SureScan pacing systems Advisa MRI and Ensura MRI SureScan pacing systems MR Conditional dual chamber and single chamber SureScan pacing systems with SureScan leads MRI Technical Manual The following list includes trademarks

More information

MRI Exams in Patients with Non MR Conditional Pacemakers and with MR Conditional Pacemakers

MRI Exams in Patients with Non MR Conditional Pacemakers and with MR Conditional Pacemakers MRI Exams in Patients with Non MR Conditional Pacemakers and with MR Conditional Pacemakers Frank G. Shellock, Ph.D., FACC Adjunct Clinical Professor of Radiology and Medicine Keck School of Medicine,

More information

EXAMPLE PROTOCOL FOR PATIENT SCANNING PROCESS

EXAMPLE PROTOCOL FOR PATIENT SCANNING PROCESS EXAMPLE PROTOCOL FOR PATIENT SCANNING PROCESS MRI SureScan Pacing, ICD, and CRT (CRT-D and CRT-P) Transvenous Implantable Cardiac Systems Timothy S. E. Albert, M.D. President CVMR Solutions, Inc. Carmel

More information

MRI imaging for patients with cardiac implantable electronic devices (CIEDs)

MRI imaging for patients with cardiac implantable electronic devices (CIEDs) MRI imaging for patients with cardiac implantable electronic devices (CIEDs) 13 th annual International Winter Arrhythmia School Collingwood, Ontario, Canada February 6, 2016 Andrew C.T. Ha, MD, MSc, FRCPC

More information

Micra MC1VR01. MRI procedural information for Micra Model MC1VR01 MR Conditional single chamber transcatheter pacing device (VVIR)

Micra MC1VR01. MRI procedural information for Micra Model MC1VR01 MR Conditional single chamber transcatheter pacing device (VVIR) Micra MC1VR01 MRI procedural information for Micra Model MC1VR01 MR Conditional single chamber transcatheter pacing device (VVIR) MRI Technical Manual The following list includes trademarks or registered

More information

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System MRI-Ready Systems Manual MRI Procedure Information for the St. Jude Medical MR Conditional System CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. WARNING: This

More information

Call Medtronic at 1 (800) to verify the patient s current implanted system

Call Medtronic at 1 (800) to verify the patient s current implanted system MRI SURESCAN SYSTEMS Patient Scanning Process Transvenous Implantable Cardiac Systems PATIENT PRESCREENING SureScan Pacing, Defibrillation, and CRT (CRT-D and CRT-P) Systems Verification Verify that patient

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

Nuclear Magnetic Resonance Imaging in Patients With Cardiac Pacing Devices

Nuclear Magnetic Resonance Imaging in Patients With Cardiac Pacing Devices BRIEF REPORT Nuclear Magnetic Resonance Imaging in Patients With Cardiac Pacing Devices Francisco Buendía, Juan M. Sánchez-Gómez, María J. Sancho-Tello, José Olagüe, Joaquín Osca, Óscar Cano, Miguel A.

More information

Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices

Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices The new england journal of medicine Original Article Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices Saman Nazarian, M.D., Ph.D., Rozann Hansford, R.N., M.P.H., Amir A. Rahsepar,

More information

MR in Patients with Pacemakers and ICDs: Defining the Issues

MR in Patients with Pacemakers and ICDs: Defining the Issues Journal of Cardiovascular Magnetic Resonance (2007) 9, 5 13 Copyright c 2007 Informa Healthcare ISSN: 1097-6647 print / 1532-429X online DOI: 10.1080/10976640601117056 MR in Patients with Pacemakers and

More information

Protocol. This trial protocol has been provided by the authors to give readers additional information about their work.

Protocol. This trial protocol has been provided by the authors to give readers additional information about their work. Protocol This trial protocol has been provided by the authors to give readers additional information about their work. Protocol for: Russo RJ, Costa HS, Silva PD, et al. Assessing the risks associated

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

Revo MRI Pacing System

Revo MRI Pacing System Brief Statement The Revo MRI SureScan pacing system is MR Conditional and as such is designed to allow patients to undergo MRI under the specified conditions for use. Indications: The Revo MRI SureScan

More information

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine

More information

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul,

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul, Mayo Clin Proc, August 2002, Vol 77 Unintentional Deactivation of ICDs 855 Case Report Unintentional Deactivation of Implantable Cardioverter-Defibrillators in Health Care Settings MARY JANE RASMUSSEN,

More information

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds

More information

Magnetic resonance imaging (MRI) is now the modality. Imaging

Magnetic resonance imaging (MRI) is now the modality. Imaging Imaging Clinical Utility and Safety of a Protocol for Noncardiac and Cardiac Magnetic Resonance Imaging of Patients With Permanent Pacemakers and Implantable-Cardioverter Defibrillators at 1.5 Tesla Saman

More information

Magnetic resonance imaging uses high-strength magnetic

Magnetic resonance imaging uses high-strength magnetic Advances in Arrhythmia and Electrophysiology Magnetic Resonance Imaging and Implantable Devices Saman Nazarian, MD, PhD; Roy Beinart, MD; Henry R. Halperin, MD, MA Magnetic resonance imaging uses high-strength

More information

Management of RT patients with implanted cardiac devices: From recommendation to implementation

Management of RT patients with implanted cardiac devices: From recommendation to implementation Management of RT patients with implanted cardiac devices: From recommendation to implementation Coen Hurkmans Clinical Physicist Catharina Hospital Eindhoven The Netherlands 1/22 Outline CIED relocation.

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

Managing Metallic Artifacts in MRI

Managing Metallic Artifacts in MRI Managing Metallic Artifacts in MRI William Faulkner, BS, RT (R) (MR) (CT), FSMRT Background Magnetism is a fundamental property of matter. As such, all substances have some type of magnetic property. Magnetic

More information

Safety of MRI at 1.5 Tesla in Patients with Implanted Sacral Nerve Neurostimulator

Safety of MRI at 1.5 Tesla in Patients with Implanted Sacral Nerve Neurostimulator european urology 50 (2006) 311 316 available at www.sciencedirect.com journal homepage: www.europeanurology.com Neuro-urology Safety of MRI at 1.5 Tesla in Patients with Implanted Sacral Nerve Neurostimulator

More information

Model 5392 EPG Temporary Pacer

Model 5392 EPG Temporary Pacer Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short

More information

09/05/2018. True/False: All pacemakers have defibrillation capabilities.

09/05/2018. True/False: All pacemakers have defibrillation capabilities. Paul G. Tarasi M.D. Staff Anesthesiologist Allegheny Health Network Adjunct Clinical Assistant Professor of Anesthesiology Temple University School of Medicine Describe different types of implanted cardiac

More information

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System MRI-Ready Systems Manual MRI Procedure Information for the St. Jude Medical MR Conditional System Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical or

More information

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable

More information

MRI in Patients with Pacemakers and Defibrillators. Paul A. Friedman, MD Win K Shen, MD Robert E. Watson, Jr. MD, Phd October 2016

MRI in Patients with Pacemakers and Defibrillators. Paul A. Friedman, MD Win K Shen, MD Robert E. Watson, Jr. MD, Phd October 2016 MRI in Patients with Pacemakers and Defibrillators Paul A. Friedman, MD Win K Shen, MD Robert E. Watson, Jr. MD, Phd October 2016 Case 71 y/o woman, adriamycin CM since 2001, presents Mar 2013 with AF/MI,

More information

IMAGEREADY MR CONDITIONAL PACING SYSTEM

IMAGEREADY MR CONDITIONAL PACING SYSTEM CAUTION: Federal law (USA) restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. MRI TECHNICAL GUIDE IMAGEREADY MR CONDITIONAL

More information

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only.

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only. SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000 1 APPROVED MRI SCAN CONDITIONS 1.5 TESLA FULL BODY 2 SETUP SJM MRI ACTIVATOR HANDHELD DEVICE The SJM MRI Activator handheld device must be enabled

More information

MRI SURESCAN SYSTEMS. Patient Scanning Process PATIENT PRESCREENING. SureScan Pacing, Defibrillation, and CRT-D Systems Verification

MRI SURESCAN SYSTEMS. Patient Scanning Process PATIENT PRESCREENING. SureScan Pacing, Defibrillation, and CRT-D Systems Verification MRI SURESCAN SYSTEMS Patient Scanning Process PATIENT PRESCREENING SureScan Pacing, Defibrillation, and CRT-D Systems Verification Verify that patient has a complete SureScan pacing, defibrillation, or

More information

Review of Pacemakers and ICD Therapy: Overview and Patient Management

Review of Pacemakers and ICD Therapy: Overview and Patient Management Review of Pacemakers and ICD Therapy: Overview and Patient Management Pacing Systems Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services OSU Division

More information

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART December 2015 This chart encompasses all Medtronic cardiac devices FDA-Approved as MR Conditional and included in the MRI SureScan portfolio. If a device

More information

Procedure for exceptional magnetic resonance scanning in patients with standard (non-mr-conditional) pacemakers

Procedure for exceptional magnetic resonance scanning in patients with standard (non-mr-conditional) pacemakers Procedure for exceptional magnetic resonance scanning in patients Category: Summary: Standard Operating Procedure This procedure outlines the process for assessing the safety and suitability of, and for

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

MRI-conditional pacemakers: current perspectives

MRI-conditional pacemakers: current perspectives Medical Devices: Evidence and Research open access to scientific and medical research Open Access Full Text Article Review MRI-conditional pacemakers: current perspectives António M Ferreira 1,2 Francisco

More information

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs).

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs). Management of Children with Cardiac Devices Guideline originally developed by Leann Miles, APRN; Lindsey Pumphrey, RN; Srikant Das, MD, and the ANGELS Team. Last reviewed by Lindsey Pumphrey, RN, Srikant

More information

Review Article Magnetic Resonance Imaging in patients with ICDs and Pacemakers

Review Article Magnetic Resonance Imaging in patients with ICDs and Pacemakers www.ipej.org 197 Review Article Magnetic Resonance Imaging in patients with ICDs and Pacemakers Prashant Nair MD, DM, Ariel Roguin MD, PhD Department of Cardiology, Rambam Medical Center, B. Rappaport

More information

JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE w Vol. 5, No. 2, pp , 2003

JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE w Vol. 5, No. 2, pp , 2003 JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE w Vol. 5, No. 2, pp. 387 397, 2003 SAFETY Cardiac s, ICDs, and Loop Recorder: Evaluation of Translational Attraction Using Conventional ( Long-Bore ) and Short-Bore

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System MRI-Ready Systems Manual MRI Procedure Information for the St. Jude Medical MR Conditional System Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical or

More information

Percepta CRT-P MRI SureScan W1TR01

Percepta CRT-P MRI SureScan W1TR01 Percepta CRT-P MRI SureScan W1TR01 MR Conditional pacemaker with cardiac resynchronization therapy, SureScan technology, and Bluetooth wireless telemetry (OAE-DDDR) Device Manual Caution: Federal law (USA)

More information

New generations pacemakers and ICDs: an update

New generations pacemakers and ICDs: an update Advances in Cardiac Arrhythmias and Great Innovations in Cardiology XXVII Giornate Cardiologiche Torinesi New generations pacemakers and ICDs: an update Prof. Fiorenzo Gaita, MD Division of Cardiology

More information

PACING SYSTEMS.

PACING SYSTEMS. References 1 Ensura SR MRI SureScan EN1SR01 Clinician Manual. 2 Medtronic Adapta ADSR01/03/06 Implant Manual, 2005. 3 Advisa MRI SureScan pacing system included on the ARTG 22/8/2009, with full body MRI

More information

Magnetic resonance imaging in patients with cardiac implantable electronic devices

Magnetic resonance imaging in patients with cardiac implantable electronic devices Neth Heart J (2018) 26:584 590 https://doi.org/10.1007/s12471-018-1192-3 POINT OF VIEW Magnetic resonance imaging in patients with cardiac implantable electronic devices A. H. Maass 1 M.E.W.Hemels 2,3

More information

Cardiac Rhythm Management // Manual // ProMRI. ProMRI. MR conditional device systems

Cardiac Rhythm Management // Manual // ProMRI. ProMRI. MR conditional device systems Cardiac Rhythm Management // Manual // ProMRI ProMRI MR conditional device systems Table of Contents Table of Contents Version Comparison............................................ 3 Change History.....................................

More information

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a

More information

ICD Diagnostics: Making the most of it

ICD Diagnostics: Making the most of it ICD Basics and Beyond ICD Diagnostics: Making the most of it Dulce Obias-Manno RN, BSN, MHSA, FHRS, CCDS, CEPS Cardiac Arrhythmia Center, Device Clinic Washington Hospital Center, Washington DC Objectives

More information

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt Leadless Pacing Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt The weakest link in Pacemaker system the lead. The more the leads the more the complications Dislodgement Fracture Insulation

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

Perioperative Management of Cardiac Implantable Devices

Perioperative Management of Cardiac Implantable Devices Financial Disclosures Breandan Sullivan MD Assistant Professor Co Director Cardiothoracic and Vascular Surgery ICU Department of Anesthesiology and Critical Care Medicine University of Colorado None Perioperative

More information

Safety Issue of Static Magnetic Field (1)

Safety Issue of Static Magnetic Field (1) MRI Safety Safety Issue of Static Magnetic Field (1) Currently, typical MRI utilizes static field with strengths of 0.5 to 3 Tesla (FDA guideline is max 4 T for infants < 1 month and 8 T for others) 1

More information

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500) 1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500). ONLY NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro

More information

Peri-operative management of pacemakers and implantable cardiac defibrillators

Peri-operative management of pacemakers and implantable cardiac defibrillators Guideline Title: Peri-operative management of patients fitted with Permanent Pacemakers (PPMs) and Implantable Cardioverter Defibrillators (ICDs). Author(s): Jessica Osman (Chief Pacing Physiologist),

More information

Cardiac Rhythm Therapy // Advanced Features // ProMRI. ProMRI. Checklist and Quick Reference Guide

Cardiac Rhythm Therapy // Advanced Features // ProMRI. ProMRI. Checklist and Quick Reference Guide Cardiac Rhythm Therapy // Advanced Features // ProMRI ProMRI Checklist and Quick Reference Guide PAGE 2 Conditions for MR scans with a BIOTRONIK ProMRI system Usage of an MR scan on a patient having an

More information

How to set up a device clinic for heart failure patients Dr C Sridevi

How to set up a device clinic for heart failure patients Dr C Sridevi How to set up a device clinic for heart failure patients Dr C Sridevi Consultant Cardiologist /Electro physiologist CARE Hospitals, CARE Foundation Need to set up a device clinic for heart failure patients

More information

Clinical Data Summary: Avoid FFS Study

Clinical Data Summary: Avoid FFS Study Atrial Pacing Lead with 1.1 mm Tip-to-Ring Spacing Clinical Data Summary: Avoid FFS Study A Multi-center, Randomized, Prospective Clinical Study Designed to Evaluate the 1699T Lead Clinical Data Summary:

More information

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500) 1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500) ONLY NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro

More information

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Identify patient with an ICD at POAC. Notify Cardiac Physiologist that patient is due to have surgery and

More information

Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator

Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator The new england journal of medicine Original Article Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator Robert J. Russo, M.D., Ph.D., Heather S. Costa, Ph.D., Patricia

More information

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism?

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? CASE REPORTS Arrhythmia 2015;16(3):173-177 doi: http://dx.doi.org/10.18501/arrhythmia.2015.029 Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? Eun-Sun Jin, MD, PhD Cardiovascular

More information

We are IntechOpen, the first native scientific publisher of Open Access books. International authors and editors. Our authors are among the TOP 1%

We are IntechOpen, the first native scientific publisher of Open Access books. International authors and editors. Our authors are among the TOP 1% We are IntechOpen, the first native scientific publisher of Open Access books 3,350 108,000 1.7 M Open access books available International authors and editors Downloads Our authors are among the 151 Countries

More information

Automatic assessment of atrial pacing threshold in current medical practice

Automatic assessment of atrial pacing threshold in current medical practice Europace (212) 14, 1615 1619 doi:1.193/europace/eus76 CLINICAL RESEARCH Leads and lead extraction Automatic assessment of atrial pacing threshold in current medical practice Jean Luc Rey 1, Serge Quenum

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators 22 March 2002 Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators F. HINTRINGER, O. PACHINGER Division of Cardiology, Department for Internal Medicine, University

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

BRITISH HEART RHYTHM SOCIETY GUIDELINES FOR THE MANAGEMENT OF PATIENTS WITH CARDIAC IMPLANTABLE ELECTRONIC DEVICES (CIEDs) AROUND THE TIME OF SURGERY

BRITISH HEART RHYTHM SOCIETY GUIDELINES FOR THE MANAGEMENT OF PATIENTS WITH CARDIAC IMPLANTABLE ELECTRONIC DEVICES (CIEDs) AROUND THE TIME OF SURGERY BRITISH HEART RHYTHM SOCIETY GUIDELINES FOR THE MANAGEMENT OF PATIENTS WITH CARDIAC IMPLANTABLE ELECTRONIC DEVICES (CIEDs) AROUND THE TIME OF SURGERY Honey Thomas, Andy Turley and Chris Plummer on behalf

More information

Diagnostic capabilities of the implantable therapeutic systems

Diagnostic capabilities of the implantable therapeutic systems Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Device Interrogation- Pacemakers, ICD and Loop Recorders Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Disclosures Consultant: Medtronic Speaker s Bureau: St. Jude Medical

More information

Safety of Pacemakers and Defibrillators in Electromagnetic Navigation Bronchoscopy

Safety of Pacemakers and Defibrillators in Electromagnetic Navigation Bronchoscopy CHEST Original Research INTERVENTIONAL PULMONOLOGY Safety of Pacemakers and Defibrillators in Electromagnetic Navigation Bronchoscopy Ahmed Y. Khan, MD ; David Berkowitz, MD ; William S. Krimsky, MD, FCCP;

More information

Cardiac Pacemakers and Implantable Cardioverter Defibrillators: In Vitro Magnetic Resonance Imaging Evaluation at 1.5-Tesla

Cardiac Pacemakers and Implantable Cardioverter Defibrillators: In Vitro Magnetic Resonance Imaging Evaluation at 1.5-Tesla Journal of Cardiovascular Magnetic Resonance (2007) 9, 21 31 Copyright c 2007 Informa Healthcare ISSN: 1097-6647 print / 1532-429X online DOI: 10.1080/10976640600897237 Cardiac Pacemakers and Implantable

More information

Leadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece

Leadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece Leadless pacing P.E. Vardas Professor of Cardiology University of Crete, Greece DISCLOSURES Small teaching fees from: Bayer/J&J, Boeringer, Menarini Small consulting fees from: Medtronic 1957: First External

More information

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Pacing and Device Jargon Made Simple Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Disclosure Dr Jonathan Timperley Honorarium from Bayer Arne Larson (1915

More information

Azure XT DR MRI SureScan W2DR01

Azure XT DR MRI SureScan W2DR01 Azure XT DR MRI SureScan W2DR01 MR Conditional dual chamber pacemaker with SureScan technology and Bluetooth wireless telemetry (OAE-DDDR) Device Manual The following list includes trademarks or registered

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE GUIDANCE EXECUTIVE (GE)

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE GUIDANCE EXECUTIVE (GE) NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE GUIDANCE EXECUTIVE (GE) Review of TA88; Dual-chamber pacemakers for symptomatic bradycardia due to sick sinus syndrome and/or atrioventricular block

More information

Solara CRT-P MRI SureScan W1TR06

Solara CRT-P MRI SureScan W1TR06 Solara CRT-P MRI SureScan W1TR06 MR Conditional pacemaker with cardiac resynchronization therapy, SureScan technology, and Bluetooth wireless telemetry (OAE-DDDR) Device Manual The following list includes

More information

Size (mm) (H W D) Mass (g) Volume (cc)

Size (mm) (H W D) Mass (g) Volume (cc) The FORMIO pacing system from Boston Scientific provides an ImageReady * MR conditional pacing system. FORMIO offers RightRate the newest generation of MV sensor technology, and the only MV sensor clinically

More information

CVIA. Cardiac Implantable Electronic Devices and Magnetic Resonance Imaging INTRODUCTION

CVIA. Cardiac Implantable Electronic Devices and Magnetic Resonance Imaging INTRODUCTION CVIA REVIEW ARTICLE pissn 2508-707X / eissn 2508-7088 https://doi.org/10.22468/cvia.2016.00101 CVIA 2017;1(1):60-66 Cardiac Implantable Electronic Devices and Magnetic Resonance Imaging Sung Ho Hwang 1,

More information

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE Interventional procedure consultation document Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death A subcutaneous

More information

8/2/2017. Acknowledgement. Disclaimer. How to manage radiotherapy patients with CIED from initial consult to treatment: TG203 recommendations

8/2/2017. Acknowledgement. Disclaimer. How to manage radiotherapy patients with CIED from initial consult to treatment: TG203 recommendations How to manage radiotherapy patients with CIED from initial consult to treatment: TG203 recommendations Moyed Miften, Ph.D. Department of Radiation Oncology University of Colorado School of Medicine Disclaimer

More information

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System 1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System ONLY 11096 Rev B 1 NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro Corp.

More information