Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions

Size: px
Start display at page:

Download "Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions"

Transcription

1 Europace (2010) 12, doi: /europace/euq092 CLINICAL RESEARCH Pacing and CRT Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions Michael Mollerus*, Glenn Albin, Margaret Lipinski, and Jill Lucca St. Mary s/duluth Clinic, 407 E Third Street, Duluth, MN 55805, USA Received 18 October 2009; accepted after revision 4 March 2010; online publish-ahead-of-print 30 March 2010 Aims The purpose of the current study is to evaluate the safety profile of patients with pacemakers or implantable cardioverter-defibrillators (ICDs) undergoing a medically necessary magnetic resonance imaging (MRI) scan without limitation on peak specific absorption rate (SAR). Recent series suggest that MRI scanning can be performed safely in select patients with pacemakers or ICDs. These studies, though, limited peak SAR.... Methods One-hundred and three patients with a total of 240 leads underwent a total of 127 scans of any body landmark using and results usual protocols with standard peak SAR settings for the scan. No patient was pacemaker dependent. Thresholds were obtained immediately before and after the scan. For all scans, the median (25th and 75th percentiles) peak SAR was 2.5 (1.3, 3.2) W/kg whereas the median scan time was 1650 (1236, 2099) s. Pre- and post-scan pacing thresholds were unchanged [0.7 (0.5, 0.8) vs. 0.6 (0.5, 0.8) V at 0.5 ms, P ¼ NS], though the sensed amplitudes [6.7 (2.9, 11.5) vs. 6.1 (2.9, 11.2) mv, P, ] and pacing impedances [500 (440, 609) vs. 491 (437, 593) V, P, ] both decreased significantly.... Conclusion The current series suggests that MRI scans may be performed safely in appropriately selected patients up to a peak SAR of 3.2 W/kg. Furthermore, peak SAR level poorly predicts the safety profile of patients with pacemakers or ICDS who are exposed to an MRI environment Keywords Pacemaker Specific absorption rate Magnetic resonance imaging Introduction Magnetic resonance imaging (MRI) is a diagnostic technique that has become the imaging modality of choice for many neurological and musculoskeletal disorders. MRI makes use of both static and gradient magnetic fields as well as pulsed radiofrequency fields. Pacemakers have been relatively contraindicated in MRI scans because of potential adverse effects including tissue heating, failure of capture, runaway pacemaker function, unpredictable reed switch behaviour, asynchronous pacing, or damage to pacemaker circuitry. 1 To date, no known deaths have been reported of patients with pacemakers or implantable cardioverter-defibrillators (ICD) undergoing an MRI scan under the supervision of a physician while on telemetry. Irnich et al., 2 though, did raise the possibility of six deaths of patients with pacemakers while undergoing MRI scans, but the decision criteria by which the coroner made the diagnosis were not listed. Several large, prospective series have been published demonstrating the relative safety of scanning patients with pacemakers in a controlled environment. 3 5 The peak specific absorption rate (SAR) in these studies, though, was 2.0 W/kg in nearly all of the scans. In order to limit peak SAR, techniques including increasing repetition time, adjusting flip angles, and changing matrix size are employed but may result in less than optimal studies. The purpose of the current study is to evaluate the safety profile of patients with pacemakers or ICDs undergoing a medically necessary MRI scan without limitation on peak SAR. * Corresponding author. Tel: ; fax: , mike@mollerus.org Published on behalf of the European Society of Cardiology. All rights reserved. & The Author For permissions please journals.permissions@oxfordjournals.org.

2 948 M. Mollerus et al. Methods Study subjects Eligible subjects consisted of patients with implanted permanent pacemakers or ICDs referred by their health-care providers for medically necessary MRI scans. Devices must have been in place for at least 6 weeks at the time of the scan and have a battery status that was beginning of life (BOL). Patients who had a native ventricular rate of,40 bpm, had an epicardial pacing lead, had a known or suspected fractured lead, had a generator with battery status that was at elective replacement indicator (ERI) or end of life (EOL), or had a device with known increased risk to exposure to an MRI scan 6 were excluded. Study subjects were enrolled between July, 2005 and September Study protocol The study was a prospective, non-randomized, unblended cohort study of patients undergoing a medically necessary MRI scan. The study population included those patients reported earlier in two smaller studies. 7,8 Prior to the scan, each patient received counselling regarding potential risks of an MRI scan and gave written consent to participate. Underlying rhythm was confirmed by temporarily programming the device to VVI with a lower rate of 40 bpm. Baseline sensing, impedance, and pacing threshold measurements at a fixed pulse width of 0.5 ms were obtained. Sensing and impedance measurements were made using the automated features of the manufacturers programmers. The devices were programmed to either DDI or VVI with a lower rate of 40 bpm for the first 97 scans. Subsequently, devices were programmed either to ODO/OVO mode or to subthreshold pacing. 8 No changes were made to the sensing parameters. If possible, magnet mode was disabled. If an ICD was present, therapy features were disabled. Histograms and event logs were saved. Each patient underwent an MRI scan in a Siemens Symphony 1.5 T scanner using an AS39T gradient coil and NUMARIS/4 (ver. syngo MR A30 DHHS) software. No limitations were placed on peak SAR level or body landmark. Scans were performed using usual protocols with standard peak SAR settings for the scan. Turbo spin echo, FLAIR, and diffusion sequences were permitted. Continuous monitoring was performed with an Invivo 3155MVS non-invasive monitoring system including telemetry and continuous pulse oximetry with plethysmographic waveform. Blood pressure measurements were obtained every 3 min. A cardiac electrophysiologist was present for the entire scan and resuscitation equipment was available in the MRI suite. Following the scan, the device was reinterrogated for any changes to pre-scan settings. Histograms and event logs were reinterrogated. Thresholds were repeated using the same protocol as pre-scan. Prior to discharge from the MRI suite, each device was reprogrammed to its pre-scan settings. Patients were followed for at least 3 months following the scan for any adverse events including loss of capture, unexpected generator malfunction or failure, device resetting, early battery depletion, or death. Significant changes in battery status were defined as a change from BOL to ERI or EOL status immediately following the scan. A threshold change or difference was defined as the pre-scan threshold value minus the post-scan threshold value. A significant increase in pacing threshold was defined as an increase the pacing amplitude by 1 V or more at a pulse width of 0.5 ms following the scan. The study complied with the Declaration of Helsinki. All participants gave written, informed consent to participate in the trial. The study was approved by the local Institutional Review Board. Statistics Pacing, sensing, and impedance thresholds were evaluated for right atrial, right ventricular, and coronary sinus leads both individually and as an aggregate. Scans were categorized as either truncal (cervical spine, thoracic spine, lumber spine, pelvis, abdomen, or chest) or non-truncal (head and extremities), and as high SAR (peak SAR.2.0 W/kg) or low SAR (peak SAR 2.0 W/kg). If a scan included both truncal and non-truncal landmarks, that scan was recorded as a truncal scan. Analysis was performed of all leads in aggregate, as well as subdivided into lead locations of right atrium, right ventricle, and coronary sinus. Because of the small number of coronary sinus leads, thresholds from these leads were not evaluated separately during subanalysis of high- vs. low-sar scans. Statistical analysis was performed using R software, version (R Foundation for Statistical Computing, Vienna, Austria). The Shapiro Wilk test was used to assess normality. Because the data were not normally distributed, results are expressed as median (25th, 75th percentiles). Analysis was performed using the paired Wilcoxon rank sum test with continuity correction for continuous variables, and the Kruskal Wallis test for categorical data. Comparison between pre- and post-scan samples was performed using the paired data. A two-sided P-value of,0.05 was considered significant. Results One hundred and three patients with a total of 240 leads underwent a total of 127 scans. One hundred and four right atrial, 127 right ventricular, and 9 coronary sinus leads were scanned. Twenty-two ICDs and 105 pacemakers were exposed to the MRI scanner. All patients had a documented, native ventricular rhythm.40 bpm. For all scans, the median peak SAR was 2.5 (1.3, 3.2) W/kg, whereas the median scan time was 1650 (1236, 2099) s. Sixty-two scans included at least one truncal landmark. The median peak SAR was significantly higher for truncal than nontruncal scans [3.2 (2.9, 3.3) vs. 1.3 (1.1, 1.7) W/kg, P, ]. The median scan time was similar between truncal and non-truncal scans [1686 (1268, 2248) vs (1236, 2032) s, P ¼ NS]. One patient experienced the onset of atrial fibrillation during a scan. 8 One pacemaker had a device reset which required reprogramming. One ICD had its arrhythmia log erased during a scan. No significant changes in battery status were seen immediately following a scan. No significant study-related events were seen at the 3-month follow-up. When evaluated as an aggregate, pacing thresholds were unchanged pre- from post-scan, though sensed amplitudes and pacing impedances both decreased significantly (see Table 1). When analysed by lead location, the right atrial leads showed only a decrease in pacing impedances, whereas the right ventricular leads showed a decrease in both sensed amplitudes and pacing impedances. There were no differences in coronary sinus lead thresholds, though the numbers may have been too small to detect a difference. The largest increase in pacing thresholds immediately following a scan was 0.5 V at a pulse width of 0.5 ms of an atrial lead in a patient undergoing a brain MRI. The peak SAR for that scan was 1.0 W/kg. Sixty-nine scans of 60 right atrial, 69 right ventricular, and 7 coronary sinus leads were high-sar scans whereas 58 scans of 44 right atrial, 58 right ventricular, and 2 coronary sinus leads

3 MRI and pacemakers without SAR restriction 949 Table 1 Thresholds pre- and post-scan Pre-scan Post-scan P-value All leads (240) Pacing (V) 0.7 (0.5, 0.8) 0.6 (0.5, 0.8) NS Sensing (mv) 6.7 (2.9, 11.5) 6.1 (2.9, 11.2), Impedance (V) 500 (440, 609) 491 (437, 593), RA Leads (104) Sensing (mv) 2.7 (1.8, 3.6) 2.6 (1.8, 3.5) NS Impedance (V) 479 (430, 540) 466 (427, 542), RV leads (127) Sensing (mv) 11.2 (8.0, 15.7) 10.8 (7.8, 14.9), Impedance (V) 530 (462, 620) 522 (440, 610), CS leads 9 Pacing (V) 1.0 (0.8, 2.0) 1.0 (1.0, 2.0) NS Sensing (mv) 10.5 (9.5, 17.7) 10.2 (9.5, 14.9) NS Impedance (V) 688 (496, 989) 676 (483, 966) NS NS, not significant. Comparison of pacing, sensing, and impedance thresholds preand post-scan. Pacing thresholds were obtained at a fixed pulse width of 0.5 ms. Data are expressed as median (25th, 75th percentiles). Lead totals are placed in parentheses. were low-sar scans. The median peak SAR [3.2 (2.9, 3.2) vs. 1.3 (1.0, 1.5) W/kg, P, ] was larger and the median scan time [1678 (1346, 2329) vs (1187, 1947) s, P ¼ 0.008] was longer for the high-sar group than the low-sar group. The pacing thresholds of all leads evaluated as an aggregate were unchanged between pre- and post-scans for both high- and low-sar groups, whereas sensed amplitudes and pacing impedances decreased significantly (see Table 2). During low-sar scans, the right atrial leads showed only a decrease in pacing impedances whereas the right ventricular leads showed a decrease in pacing impedances but a slight increase in median sensed amplitude. Though the median RV sensed amplitude was smaller pre-scan, there was no difference in the median change in sensed amplitude from pre- to post-scan (see Table 3). During high-sar scans, right atrial leads showed only a decrease in pacing impedances whereas right ventricular leads showed a decrease in both sensed amplitudes and pacing impedances. When high-sar threshold changes were compared with low-sar threshold changes, no differences between SAR level groups were seen in pacing, sensing, or impedances changes when the leads were analysed by lead location. When the leads were analysed as an aggregate, the changes in pacing thresholds were greater in the low-sar group than in the high-sar group (see Table 3). Discussion Several series have been published demonstrating the relative safety of MRI scanning of patients with pacemakers and ICDs in a controlled environment. In the first large series published, Martin et al. 3 reported on 54 patients undergoing 62 scans using Table 2 Pre- and post-scan thresholds by SAR level Pre-scan Post-scan P-value Low SAR RA leads (44) Sensing (mv) 2.5 (1.7, 3.3) 2.4 (1.6, 3.1) NS Impedance (V) 457 (415, 532) 443 (415, 521) RV leads (58) Pacing (V) 0.7 (0.5, 0.9) 0.7 (0.5, 1.0) NS Sensing (mv) 9.7 (7.7, 14.1) 9.9 (7.4, 14.1) Impedance (V) 539 (440, 659) 530 (440, 610), All leads (104) Sensing (mv) 6.2 (2.7, 11.2) 5.9 (2.8, 10.4) Impedance (V) 500 (430, 603) 489 (430, 590), High SAR RA leads (60) Pacing (V) 0.6 (0.5, 0.9) 0.6 (0.5, 0.8) NS Sensing (mv) 2.8 (1.9, 4.4) 2.9 (2.0, 4.0) NS Impedance (V) 490 (438, 563) 475 (430, 560), RV leads (69) Sensing (mv) 11.2 (8.6, 15.7) 11.2 (8.1, 15.0) Impedance (V) 523 (470, 612) 520 (460, 607), All leads (136) Sensing (mv) 6.7 (3.2, 11.9) 6.4 (3.1, 11.2) Impedance (V) 503 (449, 610) 500 (440, 602), NS, not significant. Comparison of pacing, sensing, and impedance thresholds preand post-scan by peak SAR level. Pacing thresholds were obtained at a fixed pulse width of 0.5 ms. Data are expressed as median (25th, 75th percentiles). Lead totals are placed in parentheses. a GE Signa CV/i 1.5 T scanner. In their series, 9.4% of patients had significant pacing threshold changes, though only 1.9% required device reprogramming. The peak SAR attained during any scan was 2.0 W/kg. Nazarian et al. 4 evaluated 55 patients undergoing 68 scans in a GE Signa CV/i 1.5 T scanner. No significant adverse events were seen in their series, though peak SAR was kept to,2.0 W/kg during nearly all the scans. Sommer et al. 5 reported a series of 82 patients with Medtronic devices undergoing 115 scans in a Philips Intera 1.5 T scanner. In their series, significant increases in pacing thresholds were seen in 3.1% of leads. Peak SAR was limited to 1.5 W/kg and chest scans were avoided. To our knowledge, the current series is the first study to date with a large number of devices exposed to a high-sar environment.2.0 W/kg. In this series, no significant decreases in pacing thresholds were seen either in a high- or low-sar environment. With the exception of coronary sinus leads, a decrease in impedances was seen for right atrial and right ventricular leads both in a high- or low-sar environment. The small coronary sinus lead sample size was most likely underpowered to detect any differences. A decrease in sensed amplitude was inconsistently

4 950 M. Mollerus et al. Table 3 Comparison of high- and low-sar threshold changes High SAR Low SAR P-value RA leads D Pacing (V) 0.00 (0.00, 0.03) 0.00 (20.1, 0.00) NS D Sensing (mv) 0.0 (20.3, 0.2) 0.0 (0.0, 0.2) NS D Impedance (V) 9 (0, 12) 3 (0, 10) NS RV leads D Pacing (V) 0.00 (0.00, 0.00) 0.00 (0.00, 0.00) NS D Sensing (mv) 0.2 (0.0, 0.8) 0.0 (0.0, 0.6) NS D Impedance (V) 10 (0, 20) 10 (0, 20) NS All leads D Pacing (V) 0.00 (0.00, 0.01) 0.0 (20.04, 0.00) D Sensing (mv) 0.00 (20.1, 0.6) 0.0 (0.0, 0.4) NS D Impedance (V) 10 (0, 19) 7 (0, 17) NS NS, not significant. Comparison of high- and low-sar threshold changes. A positive value signifies that the pre-scan value was greater than the post-scan value. Pacing thresholds were obtained at a fixed pulse width of 0.5 ms. Data are expressed as median (25th, 75th percentiles). seen for right atrial and right ventricular leads in high- or low-sar environments. When the leads, though, were evaluated as an aggregate, a significant decrease in sensed amplitude was seen in both high- and low-sar environments. The larger sample size of the aggregate data most likely increased the power to detect a significant decrease in sensed amplitude. The only difference seen in threshold changes between high- and low-sar group was the pacing threshold change when leads were evaluated as an aggregate, where the pacing threshold change was greater in the low-sar group than in the high-sar group. Radiofrequency fields during an MRI scan can lead to resistive heating at the lead tip-tissue interface. 9 Phantom gel and animal models have detected local tissue heating from 5.7 to 23.58C during exposure of a pacing system to the MRI environment. 6,10 12 Luechinger et al. 9 evaluated local tissue heating in a swine model with peak SAR set to 3.8 W/kg, while Roguin et al. 6, evaluated the effect of worse-case scenario MRI protocols on thresholds in a canine model. Neither study found evidence of tissue necrosis at post-scan necropsy, though Roguin et al. reported on a failure to capture for 12 h following the scan in one animal. The decrease in both lead impedance and sensed amplitudes seen in the current series suggests the possibility of some local lead tip-tissue interface changes such as local tissue heating or oedema or of changes within the device circuitry. These changes occurred irrespective of SAR levels and may have no clinical significance. The current scientific statement from the American Heart Association does not comment on SAR limitations, 13 though the European Society of Cardiology recommends a SAR limit of 2 W/kg. 14 The current series would suggest that a 2 W/kg SAR limit would be unduly restrictive, and that the safety profile of an MRI scan is poorly predicted by peak SAR level. The MRI environment is dynamic with multiple interrelated factors that may influence risk in the MRI scanner including not only peak SAR level and body landmark, but also MRI manufacturer, generator and lead system manufacturer and model, 6 proximity of the generator to the end of the bore, proximity of the generator to the surface of the bore, body habitus, lead geometry and loops, 15,16 and lead system integrity. The present study represents a real-world situation where patients may be exposed to uncontrolled MRI environments and where attempts to limit SAR levels may lead to suboptimal scans. Adjusting flip angles may adversely affect the signal-to-noise ratio. Reducing matrix size may decrease scan resolution. Increasing repetition time may prolong scan time and so increase the possibility of scan artifact from patient motion. This series suggests that MRI scans can be performed safely in patients up to a peak SAR of 3.2 W/kg and may include truncal landmarks. Nevertheless, consistent decreases in lead impedances indicative of possible local lead tip-tissue changes, as well as the results of Luechinger et al. and Roguin et al. suggest that a level of caution and vigilance must be maintained when submitting a patient with a pacemaker or ICD to an MRI scan irrespective of peak SAR level, and that preparations must be in place to address potential device failure during or following a scan. Conclusions The current series suggests that MRI scans may be performed safely in appropriately selected patients up to a peak SAR of 3.2 W/kg. Furthermore, peak SAR level poorly predicts the safety profile of patients with pacemakers or ICDs who are exposed to an MRI environment. Limitations Because of the sample size, the current study may not have enough power to detect uncommon complications related to MRI scanning. This study is underpowered to determine the safety profile of coronary sinus leads. Measurements reported by the MRI console may be imprecise and may vary between manufacturers and models. 11,17 The current study cannot be extrapolated to 3 T MRI scanners or to other models of 1.5 T MRI scanners. Because of limited battery data acquired during the study, no conclusions about the affect of MRI scanning on battery performance or longevity can be drawn. The absence of clinically significant pacing threshold changes does not predict long-term device function or battery longevity. Furthermore, because pre- and post-scan defibrillation threshold testing was not performed, no conclusions about defibrillation thresholds can be made. The current study cannot be used to risk stratify recently implanted systems or to generators at ERI or EOL. Caution should be observed in applying this study to pacemaker-dependent patients. Conflict of interest: M.M. received research grants from Boston Scientific and Medtronic. Funding This study was funded by the Duluth Clinic Foundation and by the Department of Cardiology, St. Mary s Duluth Clinic.

5 MRI and pacemakers without SAR restriction 951 References 1. Goldschlager N, Epstein A, Friedman P, Gang E, Krol R, Olshansky B. Environmental and drug effects on patients with pacemakers and implantable cardioverter/defibrillators: a practical guide to patient treatment. Arch Intern Med 2001;161: Irnich W, Irnich B, Bartsch C, Stertmann WA, Gufler H, Weiler G. Do we need pacemakers resistant to magnetic resonance imaging? Europace 2005;7: Martin ET, Coman JA, Shellock FG, Pulling CC, Fair R, Jenkins K. Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla. JAmCollCardiol2004;43: Nazarian S, Roguin A, Zviman MM, Lardo AC, Dickfeld TL, Calkins H et al. Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 Tesla. Circulation 2006;114: Sommer T, Naehle CP, Yang A, Zeijlemaker V, Hackenbroch M, Schmiedel A et al. Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 Tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: a prospective study with 115 examinations. Circulation 2006;114: Roguin A, Zviman MM, Meininger GR, Rodrigues ER, Dickfeld TM, Bluemke DA et al. Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic resonance imaging safe: in vitro and in vivo assessment of safety and function at 1.5 T. Circulation 2004;110: Mollerus M, Albin G, Lipinski M, Lucca J. Cardiac biomarkers in patients with permanent pacemakers and implantable cardioverter-defibrillators undergoing an MRI scan. Pacing Clin Electrophysiol 2008;31: Mollerus M, Albin G, Lipinski M, Lucca J. Ectopy in patients with permanent pacemakers and implantable cardioverter-defibrillators undergoing an MRI scan. Pacing Clin Electrophysiol 2009;32: Luechinger R, Zeijlemaker VA, Pedersen EM, Mortensen P, Falk E, Duru F et al. In vivo heating of pacemaker leads during magnetic resonance imaging. Eur Heart J 2005;26: Sommer T, Vahlhaus C, Lauck G, von Smekal A, Reinke M, Hofer U et al. MR imaging and cardiac pacemakers: in vitro evaluation and in vivo studies in 51 patients at 0.5 T. Radiology 2000;215: Baker KB, Tkach JA, Nyenhuis JA, Phillips M, Shellock FG, Gonzalez-Martinez J et al. Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating. J Magn Reson Imaging 2004;20: Achenbach S, Moshage W, Diem B, Dieberle T, Schibgilla V, Bachmann K. Effects of magnetic resonance imaging on cardiac pacemakers and electrodes. Am Heart J 1997;134: Levine GN, Gomes AS, Arai AE, Bluemke DA, Flamm SD, Kanal E et al. Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association Scientific Statement from the Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology, and the Council on Cardiovascular Radiology and Intervention: Endorsed by the American College of Cardiology Foundation, the North American Society for Cardiac Imaging, and the Society for Cardiovascular Magnetic Resonance. Circulation 2007;116: Roguin A, Schwitter J, Vahlhaus C, Lombardi M, Brugada J, Vardas P et al. Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. Europace 2008;10: Mattei E, Calcagnini G, Triventi M, Censi F, Bartolini P, Kainz W et al. MRI induced heating of pacemaker leads: effect of temperature probe positioning and pacemaker placement on lead tip heating and local SAR. Conf Proc IEEE Eng Med Biol Soc 2006;1: Tandri H, Zviman MM, Wedan SR, Lloyd T, Berger RD, Halperin H. Determinants of gradient field-induced current in a pacemaker lead system in a magnetic resonance imaging environment. Heart Rhythm 2008;5: Faris OP, Shein MJ. Government viewpoint: U.S. Food & Drug Administration: Pacemakers, ICDs and MRI. Pacing Clin Electrophysiol 2005;28:268 9.

Safe magnetic resonance image scanning of the pacemaker patient: current technologies and future directions

Safe magnetic resonance image scanning of the pacemaker patient: current technologies and future directions Europace (2012) 14, 631 637 doi:10.1093/europace/eur391 REVIEW Safe magnetic resonance image scanning of the pacemaker patient: current technologies and future directions Werner Jung*, Vlada Zvereva, Bajram

More information

Cardiac magnetic resonance imaging at 1.5 T in patients with cardiac rhythm devices

Cardiac magnetic resonance imaging at 1.5 T in patients with cardiac rhythm devices Europace (2011) 13, 533 538 doi:10.1093/europace/euq501 CLINICAL RESEARCH Pacing and Resynchronization Therapy Cardiac magnetic resonance imaging at 1.5 T in patients with cardiac rhythm devices Francisco

More information

Safety and Imaging Quality of MRI in Pediatric and Adult Congenital Heart Disease Patients with Pacemakers

Safety and Imaging Quality of MRI in Pediatric and Adult Congenital Heart Disease Patients with Pacemakers Safety and Imaging Quality of MRI in Pediatric and Adult Congenital Heart Disease Patients with Pacemakers AARON F. PULVER, M.D.,* MICHAEL D. PUCHALSKI, M.D.,* DAVID J. BRADLEY, M.D., L. LUANN MINICH,

More information

It is estimated that after implantation, up to 75% of patients. Original Research

It is estimated that after implantation, up to 75% of patients. Original Research Annals of Internal Medicine Original Research A Prospective Evaluation of a Protocol for Magnetic Resonance Imaging of Patients With Implanted Cardiac Devices Saman Nazarian, MD; Rozann Hansford, RN, MPH;

More information

Safety of MRI at 1.5 Tesla in Patients with Implanted Sacral Nerve Neurostimulator

Safety of MRI at 1.5 Tesla in Patients with Implanted Sacral Nerve Neurostimulator european urology 50 (2006) 311 316 available at www.sciencedirect.com journal homepage: www.europeanurology.com Neuro-urology Safety of MRI at 1.5 Tesla in Patients with Implanted Sacral Nerve Neurostimulator

More information

MRI for Patients with MRI-Conditional Pacing System: Radiographers Role. Lawrance Yip DM, DR, QMH

MRI for Patients with MRI-Conditional Pacing System: Radiographers Role. Lawrance Yip DM, DR, QMH MRI for Patients with MRI-Conditional Pacing System: Radiographers Role Lawrance Yip DM, DR, QMH MRI & Pacemaker MRI is a crucial and growing imaging modality Plays an important role in clinical management

More information

MRI and CIEDs. Disclosures

MRI and CIEDs. Disclosures Date Clinical Title MRI and CIEDs C. p. Lao Raul Weiss MD, FAHA, FACC, FHRS and CCDS Director, Electrophysiology Fellowship Program Professor of Medicine The Ohio State University Wexner Medical Center

More information

Cardiovascular magnetic resonance with an MR compatible pacemaker

Cardiovascular magnetic resonance with an MR compatible pacemaker Washington University School of Medicine Digital Commons@Becker Open Access Publications 2013 Cardiovascular magnetic resonance with an MR compatible pacemaker Anita R. Bhandiwad Kristopher W. Cummings

More information

MR in Patients with Pacemakers and ICDs: Defining the Issues

MR in Patients with Pacemakers and ICDs: Defining the Issues Journal of Cardiovascular Magnetic Resonance (2007) 9, 5 13 Copyright c 2007 Informa Healthcare ISSN: 1097-6647 print / 1532-429X online DOI: 10.1080/10976640601117056 MR in Patients with Pacemakers and

More information

Magnetic resonance imaging uses high-strength magnetic

Magnetic resonance imaging uses high-strength magnetic Advances in Arrhythmia and Electrophysiology Magnetic Resonance Imaging and Implantable Devices Saman Nazarian, MD, PhD; Roy Beinart, MD; Henry R. Halperin, MD, MA Magnetic resonance imaging uses high-strength

More information

Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices

Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices The new england journal of medicine Original Article Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices Saman Nazarian, M.D., Ph.D., Rozann Hansford, R.N., M.P.H., Amir A. Rahsepar,

More information

Micra MC1VR01. MRI procedural information for Micra Model MC1VR01 MR Conditional single chamber transcatheter pacing device (VVIR)

Micra MC1VR01. MRI procedural information for Micra Model MC1VR01 MR Conditional single chamber transcatheter pacing device (VVIR) Micra MC1VR01 MRI procedural information for Micra Model MC1VR01 MR Conditional single chamber transcatheter pacing device (VVIR) MRI Technical Manual The following list includes trademarks or registered

More information

Magnetic resonance imaging (MRI) is now the modality. Imaging

Magnetic resonance imaging (MRI) is now the modality. Imaging Imaging Clinical Utility and Safety of a Protocol for Noncardiac and Cardiac Magnetic Resonance Imaging of Patients With Permanent Pacemakers and Implantable-Cardioverter Defibrillators at 1.5 Tesla Saman

More information

Lead Management for MRI

Lead Management for MRI Lead Management for MRI Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services The Ohio State University Medical Center Columbus, OH USA President, International

More information

Nuclear Magnetic Resonance Imaging in Patients With Cardiac Pacing Devices

Nuclear Magnetic Resonance Imaging in Patients With Cardiac Pacing Devices BRIEF REPORT Nuclear Magnetic Resonance Imaging in Patients With Cardiac Pacing Devices Francisco Buendía, Juan M. Sánchez-Gómez, María J. Sancho-Tello, José Olagüe, Joaquín Osca, Óscar Cano, Miguel A.

More information

Revo MRI Pacing System

Revo MRI Pacing System Brief Statement The Revo MRI SureScan pacing system is MR Conditional and as such is designed to allow patients to undergo MRI under the specified conditions for use. Indications: The Revo MRI SureScan

More information

MRI SureScan pacemakers & ICDs what you need to know

MRI SureScan pacemakers & ICDs what you need to know MRI SureScan pacemakers & ICDs what you need to know Many variables can affect the device patient s likelihood of risks w/ an MRI Type of imaging sequence Type of scanner MRI scan duration Strength of

More information

STANDARD OPERATING PROCEDURE (SOP)

STANDARD OPERATING PROCEDURE (SOP) STANDARD OPERATING PROCEDURE (SOP) MRI Scanning for Patients with Cardiac Implantable Electronic Devices (Pacemakers and Implantable Cardiac Defibrillators) Effective date 05/04/2018 Version number V1.0

More information

MRI Exams in Patients with Non MR Conditional Pacemakers and with MR Conditional Pacemakers

MRI Exams in Patients with Non MR Conditional Pacemakers and with MR Conditional Pacemakers MRI Exams in Patients with Non MR Conditional Pacemakers and with MR Conditional Pacemakers Frank G. Shellock, Ph.D., FACC Adjunct Clinical Professor of Radiology and Medicine Keck School of Medicine,

More information

Cardiac Rhythm Therapy // Advanced Features // ProMRI. ProMRI. Checklist and Quick Reference Guide

Cardiac Rhythm Therapy // Advanced Features // ProMRI. ProMRI. Checklist and Quick Reference Guide Cardiac Rhythm Therapy // Advanced Features // ProMRI ProMRI Checklist and Quick Reference Guide PAGE 2 Conditions for MR scans with a BIOTRONIK ProMRI system Usage of an MR scan on a patient having an

More information

MRI conditional devices. A luxury or a real clinical need? Sophie Mavrogeni, MD, FESC Onassis Cardiac Surgery Center, Athens, Greece

MRI conditional devices. A luxury or a real clinical need? Sophie Mavrogeni, MD, FESC Onassis Cardiac Surgery Center, Athens, Greece MRI conditional devices. A luxury or a real clinical need? Sophie Mavrogeni, MD, FESC Onassis Cardiac Surgery Center, Athens, Greece Increase of MRI scans in recent decades In the USA, the number of procedures

More information

Cardiac Pacemakers and Implantable Cardioverter Defibrillators: In Vitro Magnetic Resonance Imaging Evaluation at 1.5-Tesla

Cardiac Pacemakers and Implantable Cardioverter Defibrillators: In Vitro Magnetic Resonance Imaging Evaluation at 1.5-Tesla Journal of Cardiovascular Magnetic Resonance (2007) 9, 21 31 Copyright c 2007 Informa Healthcare ISSN: 1097-6647 print / 1532-429X online DOI: 10.1080/10976640600897237 Cardiac Pacemakers and Implantable

More information

MRI imaging for patients with cardiac implantable electronic devices (CIEDs)

MRI imaging for patients with cardiac implantable electronic devices (CIEDs) MRI imaging for patients with cardiac implantable electronic devices (CIEDs) 13 th annual International Winter Arrhythmia School Collingwood, Ontario, Canada February 6, 2016 Andrew C.T. Ha, MD, MSc, FRCPC

More information

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only.

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only. SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000 1 APPROVED MRI SCAN CONDITIONS 1.5 TESLA FULL BODY 2 SETUP SJM MRI ACTIVATOR HANDHELD DEVICE The SJM MRI Activator handheld device must be enabled

More information

Protocol. This trial protocol has been provided by the authors to give readers additional information about their work.

Protocol. This trial protocol has been provided by the authors to give readers additional information about their work. Protocol This trial protocol has been provided by the authors to give readers additional information about their work. Protocol for: Russo RJ, Costa HS, Silva PD, et al. Assessing the risks associated

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System MRI-Ready Systems Manual MRI Procedure Information for the St. Jude Medical MR Conditional System CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. WARNING: This

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

Safety of Pacemakers and Defibrillators in Electromagnetic Navigation Bronchoscopy

Safety of Pacemakers and Defibrillators in Electromagnetic Navigation Bronchoscopy CHEST Original Research INTERVENTIONAL PULMONOLOGY Safety of Pacemakers and Defibrillators in Electromagnetic Navigation Bronchoscopy Ahmed Y. Khan, MD ; David Berkowitz, MD ; William S. Krimsky, MD, FCCP;

More information

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Patient Scheduling and Screening Q: Is there any special (other than regular MRI consent) consent form that should be signed by the patient?

More information

Advisa MRI and Ensura MRI SureScan pacing systems

Advisa MRI and Ensura MRI SureScan pacing systems Advisa MRI and Ensura MRI SureScan pacing systems MR Conditional dual chamber and single chamber SureScan pacing systems with SureScan leads MRI Technical Manual The following list includes trademarks

More information

Managing Metallic Artifacts in MRI

Managing Metallic Artifacts in MRI Managing Metallic Artifacts in MRI William Faulkner, BS, RT (R) (MR) (CT), FSMRT Background Magnetism is a fundamental property of matter. As such, all substances have some type of magnetic property. Magnetic

More information

VNS Therapy System Physician s Manual

VNS Therapy System Physician s Manual VNS Therapy System Physician s Manual Pulse Generator Model 102 Pulse Duo Generator Model 102R Demipulse Generator Model 103 Demipulse Duo Generator Model 104 AspireHC Generator Model 105 AspireSR Generator

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

EXAMPLE PROTOCOL FOR PATIENT SCANNING PROCESS

EXAMPLE PROTOCOL FOR PATIENT SCANNING PROCESS EXAMPLE PROTOCOL FOR PATIENT SCANNING PROCESS MRI SureScan Pacing, ICD, and CRT (CRT-D and CRT-P) Transvenous Implantable Cardiac Systems Timothy S. E. Albert, M.D. President CVMR Solutions, Inc. Carmel

More information

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500) 1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500). ONLY NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro

More information

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine

More information

MRI-conditional pacemakers: current perspectives

MRI-conditional pacemakers: current perspectives Medical Devices: Evidence and Research open access to scientific and medical research Open Access Full Text Article Review MRI-conditional pacemakers: current perspectives António M Ferreira 1,2 Francisco

More information

Call Medtronic at 1 (800) to verify the patient s current implanted system

Call Medtronic at 1 (800) to verify the patient s current implanted system MRI SURESCAN SYSTEMS Patient Scanning Process Transvenous Implantable Cardiac Systems PATIENT PRESCREENING SureScan Pacing, Defibrillation, and CRT (CRT-D and CRT-P) Systems Verification Verify that patient

More information

Automatic assessment of atrial pacing threshold in current medical practice

Automatic assessment of atrial pacing threshold in current medical practice Europace (212) 14, 1615 1619 doi:1.193/europace/eus76 CLINICAL RESEARCH Leads and lead extraction Automatic assessment of atrial pacing threshold in current medical practice Jean Luc Rey 1, Serge Quenum

More information

MRI in Patients with Pacemakers and Defibrillators. Paul A. Friedman, MD Win K Shen, MD Robert E. Watson, Jr. MD, Phd October 2016

MRI in Patients with Pacemakers and Defibrillators. Paul A. Friedman, MD Win K Shen, MD Robert E. Watson, Jr. MD, Phd October 2016 MRI in Patients with Pacemakers and Defibrillators Paul A. Friedman, MD Win K Shen, MD Robert E. Watson, Jr. MD, Phd October 2016 Case 71 y/o woman, adriamycin CM since 2001, presents Mar 2013 with AF/MI,

More information

Review Article Magnetic Resonance Imaging in patients with ICDs and Pacemakers

Review Article Magnetic Resonance Imaging in patients with ICDs and Pacemakers www.ipej.org 197 Review Article Magnetic Resonance Imaging in patients with ICDs and Pacemakers Prashant Nair MD, DM, Ariel Roguin MD, PhD Department of Cardiology, Rambam Medical Center, B. Rappaport

More information

2014 AAPM 56 th Annual Meeting

2014 AAPM 56 th Annual Meeting 2014 AAPM 56 th Annual Meeting SAM Diagnostic Radiology MR Safety - Deep Brain Stimulator and Other Neurostimulators Yunhong Shu, Ph.D. Mayo Clinic, Rochester, MN Outline Background of Neurostimulator

More information

Devices and EP- Safety in MRI

Devices and EP- Safety in MRI Devices and EP- Safety in MRI Mohammad Shenasa M.D. FACC, FHRS, FESC O Connor Hospital San Jose, CA Biblioteca-Alexandria Alexandria-Egypt Cardio Alex June 10-13, 2014 Effect of External Electrical and

More information

ENABLE MRI CLINICAL STUDY SUMMARY

ENABLE MRI CLINICAL STUDY SUMMARY CLINICAL STUDY SUMMARY ENABLE MRI CAUTION: Federal Law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. TABLE OF CONTENTS

More information

MRI ACCESS. Evera MRI SureScan ICD Systems

MRI ACCESS. Evera MRI SureScan ICD Systems ACCESS Evera SureScan ICD Systems ACCESS. 36% of ICD patients are likely to have an ordered over 4 years 3 Introducing Evera, featuring SureScan Technology. ICD Patients are not receiving s 1.4% 36% CONTOURED.

More information

MRI Guidelines for the Axonics Sacral Neuromodulation System

MRI Guidelines for the Axonics Sacral Neuromodulation System MRI Guidelines for the Axonics Sacral Neuromodulation System For Healthcare Professionals Rx only Note: This document is a supplement to the physician manuals provided with the Axonics SNM System. This

More information

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Case Report Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Taku Nishida MD, Tamio Nakajima MD, PhD, Yutaka Goryo MD, Ken-ichi Ishigami MD, PhD, Hiroyuki Kawata

More information

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul,

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul, Mayo Clin Proc, August 2002, Vol 77 Unintentional Deactivation of ICDs 855 Case Report Unintentional Deactivation of Implantable Cardioverter-Defibrillators in Health Care Settings MARY JANE RASMUSSEN,

More information

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500) 1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500) ONLY NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro

More information

CVIA. Cardiac Implantable Electronic Devices and Magnetic Resonance Imaging INTRODUCTION

CVIA. Cardiac Implantable Electronic Devices and Magnetic Resonance Imaging INTRODUCTION CVIA REVIEW ARTICLE pissn 2508-707X / eissn 2508-7088 https://doi.org/10.22468/cvia.2016.00101 CVIA 2017;1(1):60-66 Cardiac Implantable Electronic Devices and Magnetic Resonance Imaging Sung Ho Hwang 1,

More information

Peri-operative management of pacemakers and implantable cardiac defibrillators

Peri-operative management of pacemakers and implantable cardiac defibrillators Guideline Title: Peri-operative management of patients fitted with Permanent Pacemakers (PPMs) and Implantable Cardioverter Defibrillators (ICDs). Author(s): Jessica Osman (Chief Pacing Physiologist),

More information

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds

More information

IMAGEREADY MR CONDITIONAL PACING SYSTEM

IMAGEREADY MR CONDITIONAL PACING SYSTEM CAUTION: Federal law (USA) restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. MRI TECHNICAL GUIDE IMAGEREADY MR CONDITIONAL

More information

Warning: Patients should not be exposed to MRI environments until the surgical site following pump implantation is fully healed.

Warning: Patients should not be exposed to MRI environments until the surgical site following pump implantation is fully healed. Prometra and Prometra II Programmable Pumps Magnetic Resonance Imaging (MRI) Safety Information GENERAL MR Conditional WARNING: FAILURE TO EMPTY THE PUMP PRIOR TO EXPOSURE TO MRI ENVIRONMENT COULD RESULT

More information

Lack of interference of electromagnetic navigation bronchoscopy to implanted cardioverter-defibrillator: in-vivo study

Lack of interference of electromagnetic navigation bronchoscopy to implanted cardioverter-defibrillator: in-vivo study Europace (2014) 16, 1767 1771 doi:10.1093/europace/euu156 CLINICAL RESEARCH Sudden death and ICDs Lack of interference of electromagnetic navigation bronchoscopy to implanted cardioverter-defibrillator:

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator

Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator The new england journal of medicine Original Article Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator Robert J. Russo, M.D., Ph.D., Heather S. Costa, Ph.D., Patricia

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program Electrophysiology Implant Classification Table The table below contains the codes that apply to our UnitedHealthcare Medicare Advantage cardiology prior Description Includes Generator Placement Includes

More information

Clinical Data Summary: Avoid FFS Study

Clinical Data Summary: Avoid FFS Study Atrial Pacing Lead with 1.1 mm Tip-to-Ring Spacing Clinical Data Summary: Avoid FFS Study A Multi-center, Randomized, Prospective Clinical Study Designed to Evaluate the 1699T Lead Clinical Data Summary:

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART December 2015 This chart encompasses all Medtronic cardiac devices FDA-Approved as MR Conditional and included in the MRI SureScan portfolio. If a device

More information

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a

More information

Magnetic resonance imaging in patients with cardiac implantable electronic devices

Magnetic resonance imaging in patients with cardiac implantable electronic devices Neth Heart J (2018) 26:584 590 https://doi.org/10.1007/s12471-018-1192-3 POINT OF VIEW Magnetic resonance imaging in patients with cardiac implantable electronic devices A. H. Maass 1 M.E.W.Hemels 2,3

More information

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System 1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System ONLY 11096 Rev B 1 NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro Corp.

More information

MRI SURESCAN SYSTEMS. Patient Scanning Process PATIENT PRESCREENING. SureScan Pacing, Defibrillation, and CRT-D Systems Verification

MRI SURESCAN SYSTEMS. Patient Scanning Process PATIENT PRESCREENING. SureScan Pacing, Defibrillation, and CRT-D Systems Verification MRI SURESCAN SYSTEMS Patient Scanning Process PATIENT PRESCREENING SureScan Pacing, Defibrillation, and CRT-D Systems Verification Verify that patient has a complete SureScan pacing, defibrillation, or

More information

Preoperative Evaluation: Patients with Cardiac Disease

Preoperative Evaluation: Patients with Cardiac Disease Advances in Internal Medicine 2012 Preoperative Evaluation: Patients with Cardiac Disease Mary O. Gray, MD Professor of Medicine UC San Francisco Circulation 2007:100:e418-e500 (1) Cardiac Risk Assessment

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

In vivo heating of pacemaker leads during magnetic resonance imaging

In vivo heating of pacemaker leads during magnetic resonance imaging European Heart Journal (2005) 26, 376 383 doi:10.1093/eurheartj/ehi009 Preclinical research In vivo heating of pacemaker leads during magnetic resonance imaging Roger Luechinger 1, Volkert A. Zeijlemaker

More information

Safety Issue of Static Magnetic Field (1)

Safety Issue of Static Magnetic Field (1) MRI Safety Safety Issue of Static Magnetic Field (1) Currently, typical MRI utilizes static field with strengths of 0.5 to 3 Tesla (FDA guideline is max 4 T for infants < 1 month and 8 T for others) 1

More information

Procedure for exceptional magnetic resonance scanning in patients with standard (non-mr-conditional) pacemakers

Procedure for exceptional magnetic resonance scanning in patients with standard (non-mr-conditional) pacemakers Procedure for exceptional magnetic resonance scanning in patients Category: Summary: Standard Operating Procedure This procedure outlines the process for assessing the safety and suitability of, and for

More information

12 th International Winter Arrhythmia School Collingwood, Ontario February 7, Dr. Derek Yung. MD, FRCPC

12 th International Winter Arrhythmia School Collingwood, Ontario February 7, Dr. Derek Yung. MD, FRCPC 12 th International Winter Arrhythmia School Collingwood, Ontario February 7, 2015 Dr. Derek Yung. MD, FRCPC Presenter: Dr. Derek Yung Relationships with industry: Educational material provided by Biotronik,

More information

How to set up a device clinic for heart failure patients Dr C Sridevi

How to set up a device clinic for heart failure patients Dr C Sridevi How to set up a device clinic for heart failure patients Dr C Sridevi Consultant Cardiologist /Electro physiologist CARE Hospitals, CARE Foundation Need to set up a device clinic for heart failure patients

More information

Key words: Decision-making processes; case study; MRI with pacemaker; A&E. ASEAN J. Sci. Technol. Dev., 32(2):

Key words: Decision-making processes; case study; MRI with pacemaker; A&E. ASEAN J. Sci. Technol. Dev., 32(2): ASEAN J. Sci. Technol. Dev., 32(2): 133 144 Decision Making Processes for a Patient with Cardiac Pacemaker Admitted to the Accident and Emergency Undergoing Magnetic Resonance Imaging A Case Study F. P.

More information

JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE w Vol. 5, No. 2, pp , 2003

JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE w Vol. 5, No. 2, pp , 2003 JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE w Vol. 5, No. 2, pp. 387 397, 2003 SAFETY Cardiac s, ICDs, and Loop Recorder: Evaluation of Translational Attraction Using Conventional ( Long-Bore ) and Short-Bore

More information

We are IntechOpen, the first native scientific publisher of Open Access books. International authors and editors. Our authors are among the TOP 1%

We are IntechOpen, the first native scientific publisher of Open Access books. International authors and editors. Our authors are among the TOP 1% We are IntechOpen, the first native scientific publisher of Open Access books 3,350 108,000 1.7 M Open access books available International authors and editors Downloads Our authors are among the 151 Countries

More information

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants EKG Pop Quiz Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants 1 Disclosures No financial relationships to disclose. EKG #1 75 y/o woman with a dual-chamber pacemaker

More information

Model 5392 EPG Temporary Pacer

Model 5392 EPG Temporary Pacer Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short

More information

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt Leadless Pacing Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt The weakest link in Pacemaker system the lead. The more the leads the more the complications Dislodgement Fracture Insulation

More information

Leadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece

Leadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece Leadless pacing P.E. Vardas Professor of Cardiology University of Crete, Greece DISCLOSURES Small teaching fees from: Bayer/J&J, Boeringer, Menarini Small consulting fees from: Medtronic 1957: First External

More information

Review of Pacemakers and ICD Therapy: Overview and Patient Management

Review of Pacemakers and ICD Therapy: Overview and Patient Management Review of Pacemakers and ICD Therapy: Overview and Patient Management Pacing Systems Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services OSU Division

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs).

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs). Management of Children with Cardiac Devices Guideline originally developed by Leann Miles, APRN; Lindsey Pumphrey, RN; Srikant Das, MD, and the ANGELS Team. Last reviewed by Lindsey Pumphrey, RN, Srikant

More information

You have a what, inside you?

You have a what, inside you? Costal Emergency Medicine Conference You have a what, inside you? Less than mainstream medical devices encountered in the ED. Eric Ossmann, MD, FACEP Associate Professor Duke University Medical Center

More information

New generations pacemakers and ICDs: an update

New generations pacemakers and ICDs: an update Advances in Cardiac Arrhythmias and Great Innovations in Cardiology XXVII Giornate Cardiologiche Torinesi New generations pacemakers and ICDs: an update Prof. Fiorenzo Gaita, MD Division of Cardiology

More information

Cardiac Rhythm Management // Manual // ProMRI. ProMRI. MR conditional device systems

Cardiac Rhythm Management // Manual // ProMRI. ProMRI. MR conditional device systems Cardiac Rhythm Management // Manual // ProMRI ProMRI MR conditional device systems Table of Contents Table of Contents Version Comparison............................................ 3 Change History.....................................

More information

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie 268 Case Report Cardiac Resynchronization Therapy in a Patient with Persistent Left Superior Vena Cava Draining into the Coronary Sinus and Absent Innominate Vein: A Case Report and Review of Literature

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Device Interrogation- Pacemakers, ICD and Loop Recorders Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Disclosures Consultant: Medtronic Speaker s Bureau: St. Jude Medical

More information

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Identify patient with an ICD at POAC. Notify Cardiac Physiologist that patient is due to have surgery and

More information

Magnetic Resonance Imaging and Cardiac Pacemaker Safety at 1.5-Tesla

Magnetic Resonance Imaging and Cardiac Pacemaker Safety at 1.5-Tesla Journal of the American College of Cardiology Vol. 43, No. 7, 2004 2004 by the American College of Cardiology Foundation ISSN 0735-1097/04/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2003.12.016

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

You have a what, inside you?

You have a what, inside you? Costal Emergency Medicine Conference You have a what, inside you? Less than mainstream medical devices encountered in the ED. Eric Ossmann, MD, FACEP Associate Professor Duke University Medical Center

More information

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Pacing and Device Jargon Made Simple Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Disclosure Dr Jonathan Timperley Honorarium from Bayer Arne Larson (1915

More information

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity. Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded

More information

Cochlear Implants. Medical Procedures. for MED EL CI Systems. AW33290_5.0 (English US)

Cochlear Implants. Medical Procedures. for MED EL CI Systems. AW33290_5.0 (English US) Cochlear Implants Medical Procedures for MED EL CI Systems AW33290_5.0 (English US) This manual provides important instructions and safety information for MED EL CI System users who have to undergo a medical

More information

Perioperative Management of Cardiac Implantable Devices

Perioperative Management of Cardiac Implantable Devices Financial Disclosures Breandan Sullivan MD Assistant Professor Co Director Cardiothoracic and Vascular Surgery ICU Department of Anesthesiology and Critical Care Medicine University of Colorado None Perioperative

More information

Management of RT patients with implanted cardiac devices: From recommendation to implementation

Management of RT patients with implanted cardiac devices: From recommendation to implementation Management of RT patients with implanted cardiac devices: From recommendation to implementation Coen Hurkmans Clinical Physicist Catharina Hospital Eindhoven The Netherlands 1/22 Outline CIED relocation.

More information

Remote monotoring of cardiac rhythm devices: present and future Pacemaker and ICD

Remote monotoring of cardiac rhythm devices: present and future Pacemaker and ICD Remote monotoring of cardiac rhythm devices: present and future Pacemaker and ICD Philippe Mabo University Hospital, Rennes, France ESC congress, Paris 30 August 2011 U642 Disclosures Biotronik: research

More information