DataSim Operators Manual. Patient Simulator

Size: px
Start display at page:

Download "DataSim Operators Manual. Patient Simulator"

Transcription

1 DataSim 6100 Patient Simulator Operators Manual PN February Fluke Corporation, All rights reserved. Printed in USA. Specifications subject to change without notice. All product names are trademarks of their respective companies.

2 Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of original purchase. During the warranty period, we will repair or at our option replace, at no charge, a product that proves to be defective, provided you return the product, shipping prepaid, to Fluke Biomedical. This warranty covers the original purchaser only and is not transferable. The warranty does not apply if the product has been damaged by accident or misuse or has been serviced or modified by anyone other than an authorized Fluke Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, IN- DIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THE- ORY. This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration of instruments is not covered under the warranty This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since some jurisdictions do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you. If any provision of this warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision. 07/07

3 Notices All Rights Reserved Copyright 2008, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language without the written permission of Fluke Biomedical. Copyright Release Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke Biomedical. Unpacking and Inspection Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking the instrument. Notify the carrier and ask for an agent to be present while the instrument is unpacked. There are no special unpacking instructions, but be careful not to damage the instrument when unpacking it. Inspect the instrument for physical damage such as bent or broken parts, dents, or scratches. Technical Support For application support or answers to technical questions, either techservices@flukebiomedical.com or call or Claims Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all packing materials in their original condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not operate within specifications, or if there are any other problems not caused by shipping damage, please contact Fluke Biomedical or your local sales representative. Standard Terms and Conditions Refunds and Credits Please note that only serialized products and their accessory items (i.e., products and items bearing a distinct serial number tag) are eligible for partial refund and/or credit. Nonserialized parts and accessory items (e.g., cables, carrying cases, auxiliary modules, etc.) are not eligible for return or refund. Only products returned within 90 days from the date of original purchase are eligible for refund/credit. In order to receive a partial refund/credit of a product purchase price on a serialized product, the product must not have been damaged by the customer or by the carrier chosen by the customer to return the goods, and the product must be returned complete (meaning with all manuals, cables, accessories, etc.) and in as new and resalable condition. Products not returned within 90 days of purchase, or products which are not in as new and resalable condition, are not eligible for credit return and will be returned to the customer. The Return Procedure (see below) must be followed to assure prompt refund/credit. Restocking Charges Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15 %. Products returned in excess of 30 days after purchase, but prior to 90 days, are subject to a minimum restocking fee of 20 %. Additional charges for damage and/or missing parts and accessories will be applied to all returns.

4 Return Procedure All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factory location. When you return an instrument to Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper packaging or handling. Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide for repackaging: Use a double-walled carton of sufficient strength for the weight being shipped. Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive material around all projecting parts. Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the instrument. Returns for partial refund/credit: Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obtained from our Order Entry Group at or Repair and calibration: To find the nearest service center, go to or In the U.S.A.: Cleveland Calibration Lab Tel: globalcal@flukebiomedical.com Everett Calibration Lab Tel: service.status@fluke.com In Europe, Middle East, and Africa: Eindhoven Calibration Lab Tel: ServiceDesk@fluke.com In Asia: Everett Calibration Lab Tel: service.international@fluke.com Certification This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical s manufacturing specifications when it was shipped from the factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibration standards are measured against in-house performance standards using accepted test procedures. WARNING Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. Fluke Biomedical will not be responsible for any injuries sustained due to unauthorized equipment modifications.

5 Restrictions and Liabilities Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the information in this document will be incorporated in new editions of the publication. No responsibility is assumed by Fluke Biomedical for the use or reliability of software or equipment that is not supplied by Fluke Biomedical, or by its affiliated dealers. Manufacturing Location The DataSim 6100 Patient Simulator is manufactured in Everett, Washington by Fluke Biomedical, 6920 Seaway Blvd., Everett, WA, U.S.A.

6

7 Table of Contents Chapter Title Page 1 Introduction and Specifications Introduction Features General Safety Considerations Symbols Warnings and Cautions Instrument Familiarity Front Panel Pendant Keypad Controller Direct Functions Special Functions PAC and PVC Insertion Modifier/Cal Mode Specifications Accessories Operation Introduction Size Adjustments Calibration Heart Rate Adjustment Artifact Simulation Respiration Defibrillation Training Blood Pressure Simulation Personality Modules (Optional Special Functions) Interactive IABP Personality Modules Cardiac Output Personality Modules Blood Pressure Cables Connector Interface Pin Definitions Maintenance, Service, and Calibration Maintenance Avoiding Damage Cleaning i

8 DataSim 6100 Operators Manual Battery Charging and Replacement Service and Calibration ii

9 List of Tables Table Title Page 1-1. Symbols Simulator Front Panel Controls and Indicators Lead ECG Amplitude Pendant Keypad Controls and Indicators Pendant Keypad Codes and Functions Functions Used in the Modifier/Cal Mode Standard Accessories Optional Accessories (Series 90) Blood Pressure Cables Waveform Amplitudes Effects of Size Adjustments on Parameters Calibration Functions Heart Rates Obtained via Rate Adjust Special Function Artifact Modifier Effects Personality Modules Appropriate Pressure Simulator Interface Cables ECG Connectors Pressure Connectors iii

10 DataSim 6100 Operators Manual iv

11 List of Figures Figure Title Page 1-1. Simulator Front Panel Controls and Indicators Pendant Keypad Controls and Indicators Pressure Cable Wiring for Monitor with 5 µv/v mmhg Sensitivity Pressure Cable Wiring for Monitor with 40 µv/v mmhg Sensitivity Connector Pin Assignments v

12 DataSim 6100 Operators Manual vi

13 Chapter 1 Introduction and Specifications Contents Page Introduction Features General Safety Considerations Symbols Warnings and Cautions Instrument Familiarity Front Panel Pendant Keypad Controller Direct Functions Special Functions PAC and PVC Insertion Modifier/Cal Mode Specifications Accessories

14 DataSim 6100 Operators Manual 1-2

15 Introduction and Specifications Introduction 1 Introduction The DataSim 6100 Patient Simulator, hereafter referred to as the Simulator, is a battery-operated, six-channel patient simulator and defibrillation trainer. Critical care nurses, ACLS program instructors, and other clinical educators can generate an extensive range of simulations, from a simple normal sinus rhythm to a complex Swan-Ganz catheter insertion, as well as a wide range of real world ECG, pressure, and respiration artifacts. In the standard defibrillation training mode, the instructor can interface the Simulator with patient manikins and teach students the correct way to defibrillate a patient, plus the desired effect on ECG and BP activity. All control keys and display prompts are on a hand-held pendant keypad connected to the unit with a flexible, 20-foot, coiled telephone-style cord that gives a wide range of motion. Standard waveforms and sequences, selected via the keypad, are stored in microcomputer memory within the Simulator. A variety of optional Personality Modules expands the number of available waveform selections and sequences and includes functions for special purpose ECG, Cardiac Output Dilution Curves, Capnography, and IABP. Each personality module plugs into the Simulator front panel. A video adapter allows display of generated ECG and BP waveforms on any standard TV set. The enlarged format increases visibility and comfort level for a class of students. Features The Simulator has the following features: Training capabilities Six channels for generating ECG, arrhythmias, blood pressure, and respiration Synchronization of hemodynamic waveforms Manual PAC and PVC insertions Swan-Ganz procedure Interface with Resusci-Anne, Arrhythmia Anne, and Chris Clean manikins Optional Personality Modules Video adapter interface 1-3

16 DataSim 6100 Operators Manual General Safety Considerations Read the Users Manual before operating the Analyzer. Note If calibration or other service of the Simulator is required, only qualified service personnel should be permitted to remove the front panel of the Simulator. Symbols Table 1-1 describes the symbols associated with the Analyzer. Table 1-1. Symbols Symbol X W P ~ Hazardous voltage Description Important information; refer to manual. Conforms to European Union directives Do not dispose of this product as unsorted municipal waste. Go to Fluke s website for recycling information. Warnings and Cautions A Warning identifies hazardous conditions and actions that could cause bodily harm or death. A Caution identifies conditions and actions that could damage the Analyzer, the equipment under test, or cause permanent loss of data. Levels of signals available via the output connectors and the Personality Module connector do not exceed 12 volts, and as such, do not constitute a potential danger to the operator. However, observe the following: 1-4

17 Introduction and Specifications Instrument Familiarity 1 XW Warning To avoid possible electrical shock or personal injury, follow these guidelines: Use this Simulator only in the manner specified by the manufacturer or the protection provided may be impaired. Read the Users Manual before operating the Simulator. Do not connect the Simulator to a patient or equipment connected to a patient. The Simulator is intended for equipment evaluation only and should never be used in diagnostics, treatment or in any other capacity where the Simulator would come in contact with a patient Instrument Familiarity The functions of the Simulator are controlled from the front panel and the pendant keypad. Front Panel Figure 1-1 shows the front panel controls and indicators of the Simulator. Table 1-2 lists these components with accompanying descriptions. 1-5

18 DataSim 6100 Operators Manual 1 2 DataSim 6100 Patient Simulator LL (+) 3 LA (-) On NSR TACH BRAD ASYS Off POWER RA (-) 0 A-FLTR 1 A-FIB 2 2AVB1 3 2AVB2 ECG RL 4 3AVB 8 5 PAC 9 6 PVC1 7 PVC2 ART V1 10 CPLT V-TAC V-FIB SPEC STD FUNC PA V2 KEYPAD RA V3 V PERSONALITY MODULE AUX MANKIN V5 V6 SIMULATE DEFIB Figure 1-1. Simulator Front Panel Controls and Indicators fdg001.eps 1-6

19 Introduction and Specifications Instrument Familiarity 1 Table 1-2. Simulator Front Panel Controls and Indicators Label Component Description A Power Indicator Illuminates when power is ON B Power Switch Switches power on and off C D ECG Snap Connectors Pressure Output Connectors This battery operated unit is rechargeable with the supplied charger. Outputs low level 12 lead ECG Attach standard patient ECG cable lead wires to these snap connectors to display the simulated ECG waveforms on the monitor screen. To minimize 60-cycle artifact, use a reference electrode connection. For example, most five-lead ECG diagnostic recorders require that a Simulator REF signal is connected to the RL input of the patient cable. The Simulator can also output the ECG in a modified chest lead configuration (MCL2) used for bedside monitoring. Simply attach the LL (+) red, RA (-) white, and LA (ref) patient cable snaps to the patient Simulator. Refer to the operators manual of the ECG monitor for hook-up instructions. Output for arterial, PA, and RA pressures This output simulates the electrical output of the BP transducer that would be used with the patient monitor. The Simulator generates signals that are compatible with either 5 or 40-microvolt transducers. 1-7

20 DataSim 6100 Operators Manual Table 1-2. Simulator Front Panel Controls and Indicators (cont.) Label Component Description E Aux Input/Output Connector Cable interconnects for optional waveform, such as capnography and IABP F Manual Defib Switch Manually simulates defibrillation G Defib Connector Input for Arrhythmia Anne or Chris Clean manikins equipped with defibrillation option H Personality Module Connector Input for optional modules Used to add numerous functions and waveforms to the standard set I Keypad Connector Input for hand-help keypad J Keypad Rest Storage area for the keypad Inner Well Back Battery Charger Input Connector Jack used to recharge battery The unit can be operated while recharging. Table 1-3. Lead ECG Amplitude Lead # ECG Amplitude I +.25 mv (+/- 5%) II +1.0 III +.75 AVR.60 AVL -.25 AVF +.90 V1.40 V2.90 V V V V

21 Pendant Keypad Controller Introduction and Specifications Instrument Familiarity 1 The pendant keypad controller has two general classes of functions, direct and special, that are accessed by pressing individual keys or combinations of keys. The following pages describe functions and codes available in the standard Simulator unit. Figure 1-2 shows the keypad controls and indicators, and Table 1-4 lists and describes these components. DataSim 6100 Patient Simulator 1 2 NSR TACH BRAD ASYS A-FLTR 4 A-FIB 5 2AVB1 6 2AVB AVB PAC 8 9 PVC1 PVC2 3 CPLT V-TAC V-FIB SPEC STD FUNC 5 Figure 1-2. Pendant Keypad Controls and Indicators fdg002.eps 1-9

22 DataSim 6100 Operators Manual Table 1-4. Pendant Keypad Controls and Indicators Label Element Description A LCD Display Screen B Direct Function Keys C Special Function Key D Palpitation Keys Eight-character display shows prompts, messages, confirmations, results Fifteen of the 16 keys; excluding the SPEC FUNC key Key at the lower right corner (fourth row) of the keypad; labeled SPEC FUNC The second, third, and fourth keys in the third row of the keypad; labeled PAC, PVC1, and PVC2 Allow user to add a palpitation to any other selected rhythm E Mode/Cal Keys Dual function; the first, second, and third keys in the fourth row of the keypad; labeled CPLT/], V-TAC/[, and V-FIB/STD Allow user to select a mode or, when used with Special Functions, to adjust the value of the selected parameter Direct Functions To perform Direct Functions, the user selects the desired rhythm by pressing one of the 15 direct function keys (0-9, PVC1, PVC2, CPLT, V-TAC, and V- FIB). For example, to initiate a bradycardia rhythm, the user presses and releases the 2 key that is labeled BRAD. The Simulator automatically determines the proper time in the cardiac cycle to begin a new rhythm. Special Functions To perform Special Functions, the user presses and releases the SPEC FUNC key, followed by the appropriate two-digit code, found in the printed menus attached to either side of the keypad. The Simulator displays the * symbol in response to the SPEC FUNC key selection. The two-digit function code must be entered when the * prompt is displayed. For example, to select a bigeminy rhythm, the user presses the SPEC FUNC key followed by the code 1-10

23 Introduction and Specifications Instrument Familiarity Functions can be selected in any order to simulate virtually any patient condition sequence. After the code is input, the following prompt displays: ], [ or STD At this prompt, the user can adjust the value of the parameter as desired: ] - Increases parameter value; momentarily displays at max when maximum value is reached [ - Decreases parameter value; momentarily displays at min when minimum value is reached STD - Resets parameter value to the standard value Some of the Special Functions are actually programmed sequences of rhythms or events. Table 1-5 describes the codes and functions. Table 1-5. Pendant Keypad Codes and Functions Code Function Description 00 NORMAL SINUS Provides a convenient way to return to normal sinus rhythm at a rate of 70 BPM 01 SINUS TACHYCARDIA 02 SINUS BRADYCARDIA Rapid regular rhythm with normal P wave and heart rate of 140 BPM Slow regular rhythm with normal P wave and heart rate of 35 BPM 03 ASYSTOLE No beats but very small irregular signal remains 04 ATRIAL FLUTTER 05 ATRIAL FIBRILLATION 06 2nd DEGREE AV BLOCK I Regular ventricular rhythm; atrial flutter with 4:1 conduction Rapid irregular atrial signal with no real P waves; irregular ventricular rate Wenckebach; irregular rhythm with normal P waves; PR interval lengthens progressively until a dropped beat occurs and cycle repeats. 1-11

24 DataSim 6100 Operators Manual Table 1-5. Pendant Keypad Codes and Functions (cont.) Code Function Description 07 2nd DEGREE AV BLOCK II 08 3rd DEGREE AV BLOCK Slow regular rhythm with normal P wave; PR interval in conduced beats is normal; 3:1 block Slow regular ventricular rhythm; regular atrial rhythm; independent atrial and ventricular rhythms 09 PAC INSERT Premature atrial contraction; a single PAC is inserted into the current rhythm. 10 PVC1 INSERT Premature ventricular contraction; a Single PVC1 is inserted into the current. 11 PVC2 INSERT Premature ventricular contraction; a single PVC2 is inserted into the current. 12 COUPLET Irregular rhythms with frequent couplets and 1 pair of ventricular beats; PVCs are premature; compensatory pause follows couplet. 13 VENTRICULAR TACHYCARDIA 14 VENTRICULAR FIBRILLATION 15 ATRIAL TACHYCARDIA 16 ATRIAL TACHYCARDIA/ ABERRANCY 17 FREQUENT PACs 18 PEDIATRIC TACHYCARDIA 19 1st DEGREE AV BLOCK Irregular, rapid rhythm with irregular, successive PVCs; no P waves Irregular, rapid, chaotic waveforms with no QRS ATRIAL Rapid, regular rhythm with biphasic P Rapid, regular rhythms with biphasic P wave and notched wide QRS. Irregular rhythm with frequent premature PACs and a biphasic P wave Rapid regular rhythm with tall R wave and narrow QRS BLOCKS Normal beats except with long PR interval of.25 sec 1-12

25 Introduction and Specifications Instrument Familiarity 1 Table 1-5. Pendant Keypad Codes and Functions (cont.) Code Function Description 20 BUNDLE BRANCH BLOCK 22 ACCELERATED JUNCTIONAL Wide QRS complexes; regular rhythm with a PR interval of.16 sec JUNCTIONAL Regular rhythm with inverted P wave and short PR interval 23 JUNCTIONAL Slow regular rhythm with inverted P wave and short PR interval 24 FREQUENT PJCs Irregular rhythm with frequent, premature PJCs, followed by a pause; inverted P wave and short PR interval VENTRICULAR 26 IDIOVENTRICULAR Slow regular rhythm, no P waves, wide QRS 27 BIGEMINY Irregular rhythm alternates between PVC and normal beat. 28 UNIFOCAL PVCs Frequent, premature unifocal PVCs followed by pause; PVC has wide QRS. 29 MULTIFOCAL PVCs Regular rhythms with frequent PVCs that alternate between type 1 and type 2; pause follows PVCs 30 TRIPLET Frequent runs of 3 consecutive PVCs PACEMAKER 31 VENTRICULAR Pacer artifact precedes QRS. Regular rhythm, no P waves, wide QRS 32 ATRIAL Normal rhythm, pacer artifact precedes P wave. 33 AV SEQUENTIAL Prolonged PR interval; pacemaker artifact precedes P and QRS waves. Wide QRS. 1-13

26 DataSim 6100 Operators Manual Table 1-5. Pendant Keypad Codes and Functions (cont.) Code Function Description 34 SENSE and CAPTURE FAIL 35 FAILURE TO CAPTURE 38 CARDIAC FAILURE Frequent loss of capture, low amplitude escape beat, sensing failure of escape beat No capture; pacer artifact amplitude is 8 mv with 1 msec duration. MISC ECG Intermittent PVCs and couplet followed by ventricular tachycardia, followed by ventricular fibrillation, followed by asystole 39 ST ELEVATION Regular rhythm with ST Elevation at 2 mm 80 msec after J point 40 ST DEPRESSION Regular rhythm with ST Depression at 3 mm 80 msec after J point 41 AGONAL No P waves; wide complexes with periods of asystole 42 VENTRICULAR ASYSTOLE No QRS, regular atrial rhythm 43 CONVERSION Vfib with CPR followed by Vfib without CPR for 2 seconds prior to defib: Defib artifact, Brady rhythm with CPR, Brady rhythm without CPR, followed by NSR ECG ARTIFACT 46 DEFIBRILLATION Positive and negative baseline saturation; artifact is automatically followed by pre-selected rhythm (see #59 below) CYCLE ARTIFACT 48 ECG/RESP ARTIFACT Artifact amplitudes 0.1, 0.4, and 0.7 mv p-p; use ] and [ keys or eliminate with STD key. Artifact amplitudes 0.4, 0.6 and 1.0 mv peak to peak; use ] and [ keys or eliminate with STD key. 1-14

27 Introduction and Specifications Instrument Familiarity 1 Table 1-5. Pendant Keypad Codes and Functions (cont.) Code Function Description 49 MUSCLE ARTIFACT Artifact amplitudes 0.4, 0.6 and 1.0 mv peak to peak; use ] and [ keys or eliminate with STD key. 50 CPR CPR artifact a 70/min 53 ECG RATE ADJUST ADJUST Increases or decreases rate in 10% increments; STD returns to standard rate. 54 SIZE ADJUST Increases and decreases ECG, BP, and respiration waveforms or returns them to standard size; use after selecting desired channel. 55 AUTO TREND Automatically varies NSR heart and respiration rates and blood pressures; Vtach episode 4 every 1 hour; trend repeats every 2 hours. 59 RHYTHM SELECT DEFIB CONVERSATION Permits selection of post defib rhythm, which is then automatically inserted after defib artifact. 60 DEFIB DISABLE Disables auto defib sequence RESPIRATION 71 APNEA Thoracic impedance is simulated via ECG output snaps. 72 RESP AT 10 BPM Respiration waveform at 10 BPM 73 RESP AT 20 BPM Respiration waveform at 20 BPM 74 RESP AT 40 BPM Respiration waveform at 40 BPM 75 RESP AT 80 BPM Respiration waveform at 80 BPM 76 CVA COINCIDENCE ECG and respiration are synchronized and timed appropriately for cardiovascular artifact. Impedance drops immediately after each R wave. 1-15

28 DataSim 6100 Operators Manual Table 1-5. Pendant Keypad Codes and Functions (cont.) Code Function Description 77 PEDIATRIC APNEA/BRADY 78 HEART RATE CALIBRATION Intermittent apnea periods; heart rate drops from 140 BPM to 70 BPM each apnea period. CALIBRATION Calibrated heart rates from 30 to 300 BPM 81 LINEARITY/SPEED 2.5 Hz triangular ECG waveform 82 ECG SIZE CALIBRATION 83 ARRHYTHMIA SEQUENCE 84 RESPIRATION SIZE CAL 62 ZERO PRESSURE Cal pulse amplitudes from.25 mv to 2.5 mv Automatic sequence of arrhythmias over 3 minute period of time: NSR, Dropped Beat, AV Pace, V Pace, Atrial Tach, Couplet, RBBB, Triplet, Bigeminy (3 Foci), Vtach Respiratory depth ranging from 0.25 to 2.5 ohm; calibrated for RA-LA lead BLOOD PRESSURE Zeros all three-pressure channels as well as auxiliary CO2 channel 63 PATENT LINE Outputs a 50 mmhg flush waveform on PA channel PA waveform follows flush. Trailing edge of flush and PA pressure waveforms demonstrate ideal frequency characteristics of pressure line. 64 RESONANT LINE Outputs a 50 mmhg flush waveform on PA channel PA waveform follows flush. Trailing edge of flush and PA pressure waveforms demonstrate resonant frequency characteristics of pressure line. 1-16

29 Introduction and Specifications Instrument Familiarity 1 Table 1-5. Pendant Keypad Codes and Functions (cont.) Code Function Description 65 DAMPED LINE Outputs a 50 mmhg flush waveform on PA channel. PA waveform follows flush. Trailing edge of flush and PA pressure waveforms demonstrate damped frequency characteristics of pressure line. 66 CATHETER WHIP 67 PRES/RESP ARTIFACT Pressure artifact caused by motion of catheter; on PA channel Varies level of respiration artifact on pressures PROCEDURES 68 IABP Arterial pressure with cardiac assists pressure pulse, 2:1 augmentation. Simulation only; not interactive with the IABP; output on arterial channel 69 PA WEDGE Outputs a PAW waveform 70 SWAN-GANZ INSERTION PAC and PVC Insertion Output on PA channel: right atrium to right ventricle; right ventricle to pulmonary artery; pulmonary artery to wedged pulmonary artery; wedged pulmonary artery to pulmonary artery The third row of the keypad has three palpitation simulation keys: 9/PAC PVC1 PVC2 A single PAC or PVC can be inserted into any rhythm by pressing one of these keys. The respective PAC or PVC is inserted into the next cardiac cycle. Only one PAC, PVC1, or PVC2 can be inserted at any one time. The eight-character LCD display responds to key entries by displaying the abbreviated function name. 1-17

30 DataSim 6100 Operators Manual Modifier/Cal Mode The bottom row of the keypad has three dual function keys: CPLT/up arrow V-TAC/down arrow V-FIB/STD Pressing one of these keys without first pressing the SPEC FUNC key puts the Simulator into the Modifier/Cal mode. This mode can also be used in conjunction with the functions listed in Table 1-6. Table 1-6. Functions Used in the Modifier/Cal Mode Code Function 53 Rate Adjust 54 Size 78 Heart Rate Cal 70 BPM 82 ECG Size Cal 84 Resp Size Cal Cycle Artifact 48 ECG/Resp Artifact 49 ECG/Muscle Artifact After the Modifier/Cal mode is invoked and one of the functions is selected, the user can press one of these dual function keys to increase, decrease, or reset the standard value for the selected parameter. Pressing any of the remaining keys exits the Modifier/Cal function, retains the parameter level, and simulates the rhythm selected (NSR, Bradycardia, etc.) 1-18

31 Introduction and Specifications Specifications 1 Specifications Dimensions in H x 10 in W x 4.7 in D (33 cm H x 25.4 cm W x cm D) Weight lb (3.2 kg) Power AC power V/240 V, 1.6 A max, Hz; output: +20V, 2.5 A; output power: 50 W max Battery Type V, 1.9 AH sealed lead acid rechargeable Battery Capacity hr Battery Charge Time... 5 to 95 % of complete charge in 14 hr Temperature ºC to 40 ºC Humidity... 80% maximum ECG Output signals... High Level: 1 V/1 mv Heart Rate Range to 300 BPM Accuracy... ±1 % Output Connectors Low Level Lead electrode snaps High Level... Switchcraft 15GM7F Pacemaker Artifact mv, 1 ms Sample Rate (Maximum) sample/s Performance Testing Linearity Hz triangular wave HR Cal Check to 300 BPM Chart Speed Hz square wave Amplitude , 0.50, 1.00, 1.50, 2.00, and 2.50 mv Blood Pressure Output signals High Level... 1 V/100 mmhg Transducer... 5 and 40 µv/v mmhg Exciter Voltage V ac or dc max Static Range... 0 to 250 mmhg, adjustable in 5 % increments Accuracy... ±5 % or 1 mmhg, whichever is greater 1-19

32 DataSim 6100 Operators Manual Waveforms...Static, square wave, and physiological dynamic Output connector...switchcraft 15GM7F Performance Testing Static...0 to 250 adjustable in 5 % increments Square Wave...0 to 250 mmhg adjustable in 5 % increments Respiration Output signals Delta Impedance to 2.50 Ω Lead I Base Impedance Ω Lead I, 750 Ω Lead II Rate Range...0 to 80 BPM Lead...All leads Performance Testing Rate...0, 10, 20, 40, and 80 BPM Delta Impedance , 0.50, 1.00, 1.50, 2.00, and 2.50 Ω Coincidence check General Information Channels...ECG/arrhythmia, respiration, arterial pressure, RA pressure, and auxiliary channel for optional parameters. Fundamental Rhythms and Sequences...Normal sinus rhythm Sinus tachycardia and bradycardia Ventricular and sinus asystole Atrial tachycardia Atrial flutter Fibrillation AV blocks (1st degree, 2nd degree Mobitz I and II, and 3rd degree) Unifocal and multifocal PVCs Ventricular tachycardia Ventricular fibrillation PVCs at 1 to 35/min for any rhythm Couplet Triplet Bigeminy Junctional Accelerated junctional PACs and PJCs Atrial tachycardia with aberrant conduction Idioventricular Agonal ST-segment elevation and depression Pacemaker (atrial, ventricular, and AV sequential) Failure to capture and sense 1-20

33 Introduction and Specifications Accessories 1 Bundle-branch block Cardiac-failure sequence Conversion sequence Accessories The following are accessories for the Simulator. Table 1-7 lists standard accessories shipped with the tester. Table 1-8 lists optional accessories that must be ordered separately. To order, contact your Fluke Biomedical equipment dealer and use the Fluke Biomedical part numbers provided Table 1-7. Standard Accessories Description Part Number Operators Manual LCD Pendant (Keyboard) Controller VAC Battery Charger (Wall Mount) VAC Battery Charger B Cardiac Output Injectate B Marquette CO module B Cardiac Output Injectate B Interactive IABP Datascope Personality Module B Normal/Diseased Heart Personality Module

34 DataSim 6100 Operators Manual Table 1-8. Optional Accessories (Series 90) Blood Pressure Cables Description Part Number BCI International (6 M) Criticare Systems (1100) (6 M) Critikon (Dinamap Plus) (6 M) Datascope (800 series) (6 F) Fakuda Denshi (DS 3300) (12 M) GE Marquette Medical (7000/Early Tram - AR series only) (8 M) GE Marquette Medical (Dash, Eagle, Solar, Tram, and MacLab) (rectangular - 11 M) GE Marquette Medical (PDS 3100) (6 F) GE Marquette Medical (PPG/E for M: DR, IR, IM4, VR series) (6 F) Hewlett Packard 40 µv (78300, - 500, - 800, and Merlin/Viridia/Omnicare) (12 M) Hewlett Packard 5 µv (78300, - 500, - 800) Invivo Research (Omni-Trak) (6 M) Ivy Biomedical (6 M) Kontron (Mini-, Super-, Color-Mon) (6 M) MDE (Escort series) (6 M) Mennen Medical (All) (6 M) North American Drager (Vitalert 2000) (6 M)

35 Introduction and Specifications Accessories 1 Table 1-8. Optional Accessories (Series 90) Blood Pressure Cables (cont.) Description Part Number Ohmeda (Modulus CD-CV) (6 M) Physio Control (All) (6 M) Protocol Systems (Propaq Series 100) (6 M) Quinton (Q-Cath) (6 M) Siemens Mingo (Cath System) b (7 F) SpaceLabs (1050, 1700, PCMS series) (for use with SpaceLabs adapters and when testing the multiparameter Ultraview Command Module) (6 M) SpaceLabs (Alpha 9, Alpha 14, 703R) (5 M) Unterminated BP cable (7-pin DIN, one end only) Witt Biomedical (6 F) BCI International (6 M)

36 DataSim 6100 Operators Manual 1-24

37 Chapter 2 Operation Contents Page Introduction Size Adjustments Calibration Heart Rate Adjustment Artifact Simulation Respiration Defibrillation Training Blood Pressure Simulation Personality Modules (Optional Special Functions) Interactive IABP Personality Modules Cardiac Output Personality Modules Blood Pressure Cables Connector Interface Pin Definitions

38 DataSim 6100 Operators Manual 2-2

39 Introduction Operation Introduction 2 The following describes procedures for using the Simulator to demonstrate and train in situations that healthcare professionals experience on a daily basis. Such situations include normal and abnormal blood pressures, arrhythmias, and defibrillation. Size Adjustments The Size Adjust function, Special Function 54, lets the user adjust the waveform size up or down from the STD level. This function is intended for in-service and demonstration applications when size adjustments may be helpful in demonstrating a clinical situation, such as increasing the PA wedge pressure to simulate LV heart failure. Table 2-1 lists waveform amplitudes for each channel. Table 2-1. Waveform Amplitudes Channel Parameter 1 ECG 2 ART 3 PA 4 RA 5 AUX 6 RESP To adjust the size of a waveform channel: 1. Using the keypad, press SPEC FUNC, followed by the numbers 54. The keypad briefly displays: SIZE 2-3

40 DataSim 6100 Operators Manual followed by the prompt: CHAN? 2. Enter the desired channel number by pressing the numbers 1-6. The following prompt displays: ], [ or STD 3. Press ] to increase or [ to decrease the value, or press STD to reset to the standard level. Table 2-2 shows the effect of incremental waveform amplitude size adjustments on the parameters. Table 2-2. Effects of Size Adjustments on Parameters Channel/ Parameter (1) ECG R Wave Amplitude (mv) STD mv (2) ART SYS/DIAS Mean (mmhg) 116/ / / / / / / / (3) PA SYS/DIAS Mean (mmhg) 14/6 9 21/ / / /19 27/\ 48/ / /28 42 (4) RA Mean (mmhg) (5) AUX (CO) (6) RESP (RA-LA) (Ω) Not Adjustable

41 Operation Calibration 2 Note Pressure and R wave values are defined for normal sinus beats only. To obtain non-varying pressure valves, select SPEC FUNC 67 and decrease minimum level. For calibration of performance check applications, use SPEC FUNC 78 (Heart Rate Cal), SPEC FUNC 81 (Linearity/Speed), SPEC FUNC 82 (ECG Size Cal), and SPEC FUNC 84 (Resp Size Cal). Calibration The Calibration functions, Special Functions 78, 82, and 84, let the user adjust calibration levels for heart rate, linearity/speed, ECG size, and respiration size, as illustrated in Table 2-3. Table 2-3. Calibration Functions Adjusment Levels (78) HR CAL (BPM) Function (82) ECG/SIZE/CAL (mv) (84) RESP SIZE CAL (Ω) STD

42 DataSim 6100 Operators Manual Table 2-3. Calibration Functions (cont.) Adjusment Levels (78) HR CAL (BPM) Function (82) ECG/SIZE/CAL (mv) (84) RESP SIZE CAL (Ω) To adjust calibration levels: 1. Using the keypad, press SPEC FUNC, followed by the numbers 78, 82, or 84. For example, using the number 78, the keypad briefly displays: followed by the prompt: ECG RATE ], [ or STD 2. Press ] to increase or [ to decrease the value, or press STD to reset to the standard level. Heart Rate Adjustment The Rate Adjust function, Special Function 53, offers the user the option of adjusting rates up or down in 10% increments. Table 2-4. lists the heart rates available for the various rhythms. On power-up, each rhythm defaults to the standard heart rate (STD). 2-6

43 Operation Artifact Simulation 2 Table 2-4. Heart Rates Obtained via Rate Adjust Special Function RHYTHM -40% -30% -20% -10% STD +10% +20% +30% Bradycardia Ventricular Junctional Normal Sinus Sinus Tach Atrial Tach Atrial Fib Note For performance check applications, the Heart Rate Calibration, Special Function 78 (ECG RATE), provides calibrated heart rates from 30 to 300 BPM, See Table 2-3. To adjust the heart rate: 1. Using the keypad, press SPEC FUNC, followed by the number 53. The keypad briefly displays: followed by the prompt: RATE ], [ or STD 2. Press ] to increase or [ to decrease the value, or press STD to reset to the standard level. Artifact Simulation Various types of artifact can be superimposed on the ECG. The type and level of artifact is selected via the Special Functions 47, 48, and

44 DataSim 6100 Operators Manual To enable an artifact: 1. Using the keypad, press SPEC FUNC, followed by the numbers 47, 48, or 49. For example, using the number 47, the keypad briefly displays: followed by the prompt: 60 CYCLE ], [ or STD 2. Press ] to increase or STD to reset to standard level. The [ capability is not available. Table 2-5 shows the effect of artifact modification on the parameters. Table 2-5. Artifact Modifier Effects FUNC TYPE STD ECG 60 Hz mv 48 ECG RESP mv 49 ECG MUSCLE mv Respiration The Simulator provides a varying thoracic impedance signal via all 10 ECG electrode output snaps. Several respiration rates from apnea to 80 BPM can be selected. After a respiration rate has been selected, the respiration rate remains fixed during subsequent rhythm selections or until a new respiration rate is selected. To vary the amplitude of the thoracic impedance signal (R): 1. Using the keypad, press SPEC FUNC, followed by the numbers 54. The keypad briefly displays: SIZE 2-8

45 Operation Defibrillation Training 2 followed by the prompt: CHAN? 2. Enter number 6. The following prompt displays: ], [ or STD Press ] to increase, [ to decrease, or STD to reset to standard level. Respiration amplitudes differ, depending upon which lead is monitored. Tables 2-2 and 2-3 list amplitudes for the RA-LA lead. Note For calibration applications, use the Respiration Size Cal function, Special Function 84, to display impedance amplitudes. Pressure waveforms are modulated by the respiration waveforms to simulate respiration artifact. The respiration artifact level can be varied using Pressure/Respiration Artifact, Special Function 67. Cardiovascular artifact or ECG/Respiration coincidence can be simulated by selecting the CVA Coincidence, Special Function 76. The impedance signal is synchronized with the ECG. Impedance decreases after each R wave, simulating a decrease in impedance due to an increase in pulmonary blood volume during systole. Defibrillation Training The Defibrillation Training mode lets the user interface the Simulator with the Laerdal Arrhythmia Anne and the Armstrong Chris Clean manikins for realistic defibrillation and ECG paddle monitoring training. To train in defibrillation using a manikin: 1. Follow the guidelines provided by the training manikin for proper cable connection. 2. Plug the torso skin cable into the DEFIB MANIKIN jack. 3. Power on the Simulator. 4. On the keypad, select a life-threatening arrhythmia, such as V-FIB 2-9

46 DataSim 6100 Operators Manual 5. To enable the automatic defib conversion feature and select the post defib rhythm: a. Press and release the SPEC FUNC key b. Enter code number 59 to initiate the DEFIB CONVERSION Rhythm Select function. The keypad displays: SEL RTHM c. Select the desired post-defib rhythm by pressing a Direct Function key, or by pressing the SPEC FUNC key followed by the two-digit code number of the desired rhythm. Note If a post defib rhythm is not programmed, the defib artifact is followed by asystole until another rhythm is manually selected. 6. To initiate the defibrillation sequence: Defibs equipped with apex/sternum paddles: Firmly place the paddles on the electrodes located on the manikin chest skin or torso cover. Defibs equipped with disposable electrodes: Plug in adapters and connect cables from defib to adapter snaps. 7. Charge the defib to 25 joules for Arrhythmia Anne or up to 400 joules for Chris Clean. 8. Press the SIMULATE DEFIB button. W Caution To avoid damage to the Simulator or adverse affects on its performance, refer to the Arrhythmia Anne and Chris Clean instruction manuals for the proper procedures and use of the defibrillator on the manikin. Under no circumstances should a defibrillator be discharged directly into the Simulator. After defibrillation of the manikin, the Simulator simulates defibrillation artifact, followed by the pre-selected rhythm. The programmed post defibrillation rhythm can be changed at any time. 2-10

47 Operation Blood Pressure Simulation 2 9. To repeat a sequence, select another life-threatening rhythm and defibrillate. The Defib Conversion feature can be disabled at any time by selecting SPEC FUNC 60 (DEFIB DISABLE). Manually Initiating Defib Sequence To initiate the defib sequence manually, press the SIMULATE DEFIB switch on the Simulator front panel or select SPEC FUNC 46 (DEFIB). When this function is selected, defib artifact is simulated. If a post defib rhythm has been programmed, the defib artifact is automatically followed by the programmed rhythm. Blood Pressure Simulation The Simulator can assist in checking the calibration of an invasive pressure monitor. It is important to zero each pressure channel of the monitor. To check the calibration of the monitor: 1. Connect BP cables from the Simulator to the monitor. 2. Zero the monitor before powering on the Simulator. When the Simulator is powered on, it automatically outputs an NSR ECG waveform and the normal BP waveforms. Alternatively, power on the Simulator and select SPEC FUNC 62 (0mmHg). Up to three pressure channels can be zeroed simultaneously via channel 2-4. Use the pressure monitor manufacturer suggested procedure to obtain a 0 mmhg pressure baseline of each pressure channel. 3. After successfully adjusting the monitor zero point, select SPEC FUNC 79 to check calibration of the pressure monitor. 4. Select the desired pressure waveform via Special Function or Direct Function keys to output dynamic BP waveforms on all three channels. Pressure cables are available to connect the Simulator to most pressure monitors. Table 2-2 defines systolic, diastolic, and mean values of the dynamic pressure waveforms for normal sinus beats. If the monitor or recorder does not closely agree with these pressure values, any of the following may apply. The monitor or recorder is not calibrated or properly zeroed. 2-11

48 DataSim 6100 Operators Manual The inter-connect cable is not supplying an excitation reference voltage required by the Simulator. The inter-connect cable is wired to the wrong signal output pin. Respiration artifact on the pressures modulates the pressure values in sync with the respiration cycle. This artifact significantly alters the systolic/diastolic values. 5. To remove respiration artifact, select SPEC FUNC 67 (PRES RESP) and decrease to the minimum level. Personality Modules (Optional Special Functions) Optional Personality Modules increase the number of Special Functions available in the Simulator. These can be plugged into the front panel connector and provide the following scenarios: Interactive IABP helps students practice setting inflation/deflation timings. The Simulator can be interfaced with all Kontron, Datascope, and Aries balloon pumps. The Simulator responds to the inflation/deflation timing of the balloon pump by changing the timing and shape of augmented arterial pressures. Cardiac Output simulates normal and abnormal thermal dilution curves by connecting the Simulator to any cardiac output computer that accepts the Edwards 3-pin catheters. PALS Training an optional Pediatric ECG Personality Module used with the DEFIB training feature during Pediatric (PALS) training workshops. Note Power off the Simulator before inserting or removing a Personality Module. Each module has a function menu listing the functions and code numbers within the module s memory. These functions are assigned code numbers between 89 and 99. Available Personality Modules are listed in Table

49 Operation Personality Modules (Optional Special Functions) 2 Table 2-6. Personality Modules MDE Ref # Part # Description Additional Detail B Intra-Aortic Balloon Assist (manual IABP waveform selection) Early Inflation Late Inflation Early Deflation Late Deflation Proper Timing B Pediatric ECG Sinus Arrhythmia Junctional Junctional Escape Wandering Pacer Hyperkalemia Enlarged Atrium CPR Artifact Sinus Tachycardia B Intra-Cranial Pressures (ICP) Normal ICP Normal with Resp Artifact Damped with Resp Artifact Cough Artifact Hypercapnia B Wave and A Wave Jugular Comp B Advanced Pacer Undersensing Oversensing Muscle Artifact Oversensing tall T Waves Fusion Pseudo Fusion Runaway Pacer PAC with DVI Pacer Retrograde VA Conduction 2-13

50 DataSim 6100 Operators Manual Table 2-6. Personality Modules (cont.) MDE Ref # Part # Description Additional Detail B MCL1 Atrials Normal Sinus Rhythm Bradycardia Sinus Arrest Atrial Tach Atrial Flutter Atrial Fib PAC PAT B MCL1 Blocks Normal Sinus Rhythm 1 st degree AV block 2 nd degree AV block Type I 2 nd degree AV block Type II 3 rd degree AV block RBBB LBBB B MCL1 Ectopy/ Aberrancy B Left Heart Pressures Normal Sinus Rhythm Right PVC Left PVC Multifocal PVC s R on T Bigeminy Atrial Fib with Aberrancy Cardiac Failure Sequence Pressure Zero Normal AO, LV, LA AO to LV Insertion LV to AO Pullback AO, LV, and LA with PVCs B Valve Disease Mitral Stenosis Mitral Regurgitation Aortic Stenosis Aortic Regurgitation LV/AO Pullback with AS LV/AO Pullback with AR B Lead Set Includes three previous modules B Lead Normal ECG 2-14

51 Operation Personality Modules (Optional Special Functions) 2 Table 2-6. Personality Modules (cont.) MDE Ref # Part # Description Additional Detail B Lead Anterior infarct B Lead Inferior infarct Leads I, II, III, AVR, AVL, AVF, V B ST Segments 2.5 mm Horizontal ST Depression 2.5 mm Horizontal ST Elevation ST Downslope Depression ST Upslope Elevation ST Upslope Depression 45 ST Upslope Depression 30 ST Upslope Depression B MCL1 Set (07, 08, 09) B Cardiac Output (CO)(Injectate Temp = 0 C) B Marquette CO Module B Cardiac Output (Injectate Temp = 25 C) B Defib Training Module B Interactive IABP/Kontron (K2000 and KAAT) B Interactive IABP/Datasco pe Includes , and Injectate temp for use with Injectate temp for use with (7, 10, K2000, KAAT ) System

52 DataSim 6100 Operators Manual Table 2-6. Personality Modules (cont.) MDE Ref # Part # Description Additional Detail B Interactive IABP/Datascope B Interactive IABP/Aries B Normal/Disease Heart System 90 Early Inflation Late Inflation Early Deflation Late Deflation Proper Timing Interactive IABP Personality Modules The Interactive Intra-Aortic Balloon Pump Personality Modules are used with the Simulator and Kontron, Datascope System 80, Datascope System 90, and Aries monitors. Invasive arterial BP and sync timing cables are included with each module. To generate the special waveforms: 1. Determine the need for a balloon according to the monitor used. Datascope 90: It is essential that a balloon (or suitable replacement) be connected to the Balloon Output connector for proper operation of the Datascope 90 pump. Datascope 80, Kontron Aries: It may be necessary to defeat the alarm by connecting a balloon or suitable replacement to the Balloon Output connector. 2. Connect Datascope/Kontron/Aries ECG cable to the Simulator ECG snap connectors. 3. Insert the Interactive IABP module into the Simulator Personality Module connector. 4. Power on the Simulator. 5. Connect the following cables to their respective monitors: 2-16

ProSim 6/8. Users Manual. Vital Signs Simulator

ProSim 6/8. Users Manual. Vital Signs Simulator ProSim 6/8 Vital Signs Simulator Users Manual January 2011, Rev. 2, 9/11 2011 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks

More information

ProSim 4 Vital Signs Simulator

ProSim 4 Vital Signs Simulator ProSim 4 Vital Signs Simulator Technical Data ProSim 4 Vital Signs Simulator with breakthrough touchscreen technology offers quick and simple one-tap testing for patient monitor performance checks and

More information

ProSim 3 and 2 Vital Signs Simulators

ProSim 3 and 2 Vital Signs Simulators ProSim 3 and 2 Vital Signs Simulators Technical Data Don t need a comprehensive patient monitor tester? The 6-in-1 ProSim 3 and 4-in-1 ProSim 2 Vital Signs Simulators are clear choices for biomedical engineers

More information

ProSim 3 and 2. Vital Signs Simulators

ProSim 3 and 2. Vital Signs Simulators ProSim 3 and 2 Vital Signs Simulators Technical Data Don t need a comprehensive patient monitor tester? The 6-in-1 ProSim 3 and 4-in-1 ProSim 2 Vital Signs Simulators are clear choices for biomedical engineers

More information

ProSim 8. Users Manual. Vital Signs Simulator

ProSim 8. Users Manual. Vital Signs Simulator ProSim 8 Vital Signs Simulator Users Manual January 2011, Rev. 3, 3/16 2011-2016 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks

More information

medsim 300B Patient Simulator

medsim 300B Patient Simulator medsim 300B Patient Simulator Technical Data The full-featured medsim 300B is the most versatile patient simulator on the market today. It is designed to meet the most complex inspection requirements and

More information

Rigel UNI-SiM Lite. The most cost-effective patient simulator on the market. Key Features n Compact and cost-effective

Rigel UNI-SiM Lite. The most cost-effective patient simulator on the market. Key Features n Compact and cost-effective Rigel UNI-SiM Lite The most cost-effective patient simulator on the market The UNI-SiM Lite is a handheld and battery-operated vital signs simulator, designed to be a cost effective and portable solution

More information

Rigel UNI-SiM. The world s smallest integrated NiBP, SpO2 and patient simulator. Key Features n Compact and cost-effective

Rigel UNI-SiM. The world s smallest integrated NiBP, SpO2 and patient simulator. Key Features n Compact and cost-effective Rigel UNI-SiM The world s smallest integrated NiBP, SpO2 and patient simulator. The UNI-SiM is a handheld and battery-operated vital signs simulator capable of undertaking six synchronized vital signs

More information

MPS450. Operators Manual. Multiparameter Simulator

MPS450. Operators Manual. Multiparameter Simulator MPS450 Multiparameter Simulator Operators Manual PN 2243350 June 2010, Rev. 2, 3/11 2010-2011 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice.

More information

CORONARY ARTERIES. LAD Anterior wall of the left vent Lateral wall of left vent Anterior 2/3 of interventricluar septum R & L bundle branches

CORONARY ARTERIES. LAD Anterior wall of the left vent Lateral wall of left vent Anterior 2/3 of interventricluar septum R & L bundle branches CORONARY ARTERIES RCA Right atrium Right ventricle SA node 55% AV node 90% Posterior wall of left ventricle in 90% Posterior third of interventricular septum 90% LAD Anterior wall of the left vent Lateral

More information

medsim 300B Operators Manual Patient Simulator

medsim 300B Operators Manual Patient Simulator medsim 300B Patient Simulator Operators Manual PN 2243039 December 2006 2006 Fluke Corporation, All rights reserved. Printed in USA All product names are trademarks of their respective companies. LIMITED

More information

-RHYTHM PRACTICE- By Dr.moanes Msc.cardiology Assistant Lecturer of Cardiology Al Azhar University. OBHG Education Subcommittee

-RHYTHM PRACTICE- By Dr.moanes Msc.cardiology Assistant Lecturer of Cardiology Al Azhar University. OBHG Education Subcommittee -RHYTHM PRACTICE- By Dr.moanes Msc.cardiology Assistant Lecturer of Cardiology Al Azhar University The Normal Conduction System Sinus Node Normal Sinus Rhythm (NSR) Sinus Bradycardia Sinus Tachycardia

More information

Basic Dysrhythmia Interpretation

Basic Dysrhythmia Interpretation Basic Dysrhythmia Interpretation Objectives 2 To understand the Basic ECG To understand the meaning of Dysrhythmia To describe the normal heart conduction system. To describe the normal impulse pathways.

More information

MULTI-PARAMETER PATIENT SIMULATOR

MULTI-PARAMETER PATIENT SIMULATOR MULTI-PARAMETER PATIENT SIMULATOR PS-2200 SERIES USER MANUAL BC BIOMEDICAL PS-2200 SERIES TABLE OF CONTENTS WARNINGS, CAUTIONS, NOTICES... ii DESCRIPTION... 1 LAYOUT... 7 ECG--NORMAL SINUS RHYTHM... 14

More information

ProSim 4. Users Manual. Vital Signs Simulator

ProSim 4. Users Manual. Vital Signs Simulator ProSim 4 Vital Signs Simulator Users Manual PN FBC-0007 January 2011, Rev. 3, 2/15 2011-2015 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names

More information

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use DEFIBRILLATOR OPTION General Information Introduction M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for defibrillation and cardioversion.

More information

Electrocardiography for Healthcare Professionals

Electrocardiography for Healthcare Professionals Electrocardiography for Healthcare Professionals Chapter 9: Ventricular Dysrhythmias 2012 The Companies, Inc. All rights reserved. Learning Outcomes 9.1 Describe the various ventricular dysrhythmias 9.2

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing D. J. McMahon 141001 cewood rev 2017-10-04 Key Points Defibrillators: - know the definition & electrical value

More information

CORONARY ARTERIES HEART

CORONARY ARTERIES HEART CARDIAC/ECG MODULE THE HEART CORONARY ARTERIES FIBRILLATING HEART CORONARY ARTERIES HEART PRACTICE RHYTHMS PRACTICE RHYTHMS ELECTRICAL CONDUCTION SA Node (60 100) Primary pacemaker AV Node (40 60) ***Creates

More information

Cardiac Arrhythmias in Sleep

Cardiac Arrhythmias in Sleep Cardiac Arrhythmias in Sleep Only 53 Slides! Gauresh H Kashyap, MD, FACP, FCCP, FAASM 1 2 Cardiac Arrhythmias in Sleep Out of 400 Patients with OSA, 48% had some Arrhythmias 20% - 2 PVCs/min 7% - Bradycardia

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Key Points Defibrillators: - know the definition & electrical value of a joule - monophasic vs biphasic types:

More information

UNDERSTANDING YOUR ECG: A REVIEW

UNDERSTANDING YOUR ECG: A REVIEW UNDERSTANDING YOUR ECG: A REVIEW Health professionals use the electrocardiograph (ECG) rhythm strip to systematically analyse the cardiac rhythm. Before the systematic process of ECG analysis is described

More information

Course Objectives. Proper Lead Placements. Review the ECG print paper. Review the mechanics of the Myocardium. Review basics of ECG Rhythms

Course Objectives. Proper Lead Placements. Review the ECG print paper. Review the mechanics of the Myocardium. Review basics of ECG Rhythms ECG Interpretations Course Objectives Proper Lead Placements Review the ECG print paper Review the mechanics of the Myocardium Review basics of ECG Rhythms How Leads Work The ECG Leads we use are Bipolar

More information

2017 BDKA Review. Regularity Rate P waves PRI QRS Interpretation. Regularity Rate P waves PRI QRS Interpretation 1/1/2017

2017 BDKA Review. Regularity Rate P waves PRI QRS Interpretation. Regularity Rate P waves PRI QRS Interpretation 1/1/2017 1. 2017 BDKA Review 2. 3. 4. Interpretation 5. QT 6. 7. 8. 9. 10. QT 11. 12. 13. 14. 15. 16. 17. 18. QT 19. 20. QT 21. 22. QT 23. 24. Where are pacer spikes? Before the P wave or before the QRS complex?

More information

CARESCAPE Patient Data Module

CARESCAPE Patient Data Module GE Healthcare CARESCAPE Patient Data Module High-acuity mobile patient monitoring The CARESCAPE Patient Data Module helps you transport patients to the right place at the right time efficiently and safely.

More information

QUICK SETUP GUIDE SECULIFE DFBASE

QUICK SETUP GUIDE SECULIFE DFBASE QUICK SETUP GUIDE SECULIFE DFBASE SECULIFE DFBASE is a microprocessor-based instrument that is used in the testing of defibrillators. It measures the energy output and provides information about the defibrillator

More information

Cardiology Flash Cards

Cardiology Flash Cards Cardiology Flash Cards EKG in a nut shell www.brain101.info Conduction System www.brain101.info 2 Analyzing EKG Step by step Steps in Analyzing ECG'S 1. Rhythm: - Regular _ Sinus, Junctional or Ventricular.

More information

Study methodology for screening candidates to athletes risk

Study methodology for screening candidates to athletes risk 1. Periodical Evaluations: each 2 years. Study methodology for screening candidates to athletes risk 2. Personal history: Personal history of murmur in childhood; dizziness, syncope, palpitations, intolerance

More information

Defibrillator. BiomedGuy

Defibrillator. BiomedGuy Defibrillator BiomedGuy Medtronic Physiocontrol LifePak 10 Introduction This life-support system is used by paramedic, hospital staff, and other trained authorized healthcare providers. Provides, ECG,

More information

TEST BANK FOR ECGS MADE EASY 5TH EDITION BY AEHLERT

TEST BANK FOR ECGS MADE EASY 5TH EDITION BY AEHLERT Link download full: http://testbankair.com/download/test-bank-for-ecgs-made-easy-5thedition-by-aehlert/ TEST BANK FOR ECGS MADE EASY 5TH EDITION BY AEHLERT Chapter 5 TRUE/FALSE 1. The AV junction consists

More information

ECG Interpretation Cat Williams, DVM DACVIM (Cardiology)

ECG Interpretation Cat Williams, DVM DACVIM (Cardiology) ECG Interpretation Cat Williams, DVM DACVIM (Cardiology) Providing the best quality care and service for the patient, the client, and the referring veterinarian. GOAL: Reduce Anxiety about ECGs Back to

More information

Miscellaneous Stuff Keep reading the Outline

Miscellaneous Stuff Keep reading the Outline Miscellaneous Stuff Keep reading the Outline Welcome to the 5-Step Method ECG #: Mearurements: Rhythm (s): Conduction: Waveform: Interpretation: A= V= PR= QRS= QT= Axis= 1. Compute the 5 basic measurements:

More information

Panorama. Arrhythmia Analysis Frequently Asked Questions

Panorama. Arrhythmia Analysis Frequently Asked Questions Panorama Arrhythmia Analysis Frequently Asked Questions What ECG vectors are used for Beat Detection? 3-wire lead set 5-wire lead set and 12 lead What ECG vectors are used for Beat Typing? 3-wire lead

More information

Blocks & Dissociations. Reading Assignment (p47-52 in Outline )

Blocks & Dissociations. Reading Assignment (p47-52 in Outline ) Blocks & Dissociations Reading Assignment (p47-52 in Outline ) Objectives Who are Wenckebach and Mobitz? Review SA and AV Blocks AV Dissociations: learning who s the boss and why 2 nd degree SA Block:

More information

ECG Monitoring, Defibrillation and Synchronized Cardioversion with Hands-Free Electrodes

ECG Monitoring, Defibrillation and Synchronized Cardioversion with Hands-Free Electrodes Suggestions for Performance Evaluation with the LIFEPAK 20 Defibrillator/Monitor ECG Monitoring, Defibrillation and Synchronized Cardioversion with Hands-Free Electrodes Name: Unit: Reviewer: Date: This

More information

AMPS-1. Advanced Modular Patient Simulator. Operating Manual

AMPS-1. Advanced Modular Patient Simulator. Operating Manual AMPS-1 Advanced Modular Patient Simulator Operating Manual AMPS-1 Advanced Modular Patient Simulator Operating Manual 2008 Datrend Systems Inc. Unit #1-3531 Jacombs Road Richmond, BC CANADA. V6V 1Z8 Tel:

More information

Electrocardiography Biomedical Engineering Kaj-Åge Henneberg

Electrocardiography Biomedical Engineering Kaj-Åge Henneberg Electrocardiography 31650 Biomedical Engineering Kaj-Åge Henneberg Electrocardiography Plan Function of cardiovascular system Electrical activation of the heart Recording the ECG Arrhythmia Heart Rate

More information

REtrive. REpeat. RElearn Design by. Test-Enhanced Learning based ECG practice E-book

REtrive. REpeat. RElearn Design by. Test-Enhanced Learning based ECG practice E-book Test-Enhanced Learning Test-Enhanced Learning Test-Enhanced Learning Test-Enhanced Learning based ECG practice E-book REtrive REpeat RElearn Design by S I T T I N U N T H A N G J U I P E E R I Y A W A

More information

FLB s What Are Those Funny-Looking Beats?

FLB s What Are Those Funny-Looking Beats? FLB s What Are Those Funny-Looking Beats? Reading Assignment (pages 27-45 in Outline ) The 5-Step Method ECG #: Mearurements: Rhythm (s): Conduction: Waveform: Interpretation: A= V= PR= QRS= QT= Axis=

More information

COPYRIGHT Ramsey Medical, Inc.

COPYRIGHT Ramsey Medical, Inc. Now your petmap+ii can monitor ECG 1 This manual applies to the operation of the ECG Module (REF# 9026), available as an option with the petmap+ii. The ECG Module requires a petmap+ii to be useful. Read

More information

ECG ABNORMALITIES D R. T AM A R A AL Q U D AH

ECG ABNORMALITIES D R. T AM A R A AL Q U D AH ECG ABNORMALITIES D R. T AM A R A AL Q U D AH When we interpret an ECG we compare it instantaneously with the normal ECG and normal variants stored in our memory; these memories are stored visually in

More information

CARESCAPE Patient Data Module

CARESCAPE Patient Data Module CARESCAPE Patient Data Module High-acuity mobile patient monitoring The CARESCAPE Patient Data Module helps you to transport patients to the right place at the right time so you can deliver a consistent

More information

The Nuts and Bolts of ICD Therapy

The Nuts and Bolts of ICD Therapy Electrical Management of Cardiac Rhythm Disorders For Cardiology Fellows December 5-8 Austin, Texas The Nuts and Bolts of ICD Therapy 1 2 Action Potential Localized Differences in Conduction Conduction

More information

Please check your answers with correct statements in answer pages after the ECG cases.

Please check your answers with correct statements in answer pages after the ECG cases. ECG Cases ECG Case 1 Springer International Publishing AG, part of Springer Nature 2018 S. Okutucu, A. Oto, Interpreting ECGs in Clinical Practice, In Clinical Practice, https://doi.org/10.1007/978-3-319-90557-0

More information

PHM-100 PORTABLE HANDHELD TRANSDUCER INDICATOR OPERATORS MANUAL. Transducer Techniques R

PHM-100 PORTABLE HANDHELD TRANSDUCER INDICATOR OPERATORS MANUAL. Transducer Techniques R PHM-100 PORTABLE HANDHELD TRANSDUCER INDICATOR OPERATORS MANUAL Transducer Techniques R 42480 RIO NEDO, TEMECULA, CA 92590 (909) 719-3965 FAX (909) 719-3900 URL: http://www.ttloadcells.com E-mail: tti@ttloadcells.com

More information

VitalSim Directions for Use

VitalSim Directions for Use VitalSim Directions for Use Overview and Intended use. The VitalSim Base Unit and Remote Control are intended to be used in conjunction with an approved manikin to simulate vital signs on the manikin.

More information

BEDSIDE ECG INTERPRETATION

BEDSIDE ECG INTERPRETATION BEDSIDE ECG INTERPRETATION Presented by: Ryan Dean, RN, MSN, CCRN, CCNS, CFRN Flight Nurse 2017 Based on presentations originally by Gennifer DePaoli, RN Objectives Hospital policies Electrical conduction

More information

CRC 431 ECG Basics. Bill Pruitt, MBA, RRT, CPFT, AE-C

CRC 431 ECG Basics. Bill Pruitt, MBA, RRT, CPFT, AE-C CRC 431 ECG Basics Bill Pruitt, MBA, RRT, CPFT, AE-C Resources White s 5 th ed. Ch 6 Electrocardiography Einthoven s Triangle Chest leads and limb leads Egan s 10 th ed. Ch 17 Interpreting the Electrocardiogram

More information

Junctional Premature Contraction (JPC)

Junctional Premature Contraction (JPC) Where s the PAC? Junctional Premature Contraction (JPC) A junctional premature contraction (JPC) is a beat that originates prematurely in the AV node. It can occur sporadically or in a grouped pattern.

More information

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. Defibrillators OBJECTIVE 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. 4. Types and classes of defibrillator 5. Describe

More information

PATIENT WITH ARRHYTHMIA IN DENTIST S OFFICE. Małgorzata Kurpesa, MD., PhD. Chair&Department of Cardiology

PATIENT WITH ARRHYTHMIA IN DENTIST S OFFICE. Małgorzata Kurpesa, MD., PhD. Chair&Department of Cardiology PATIENT WITH ARRHYTHMIA IN DENTIST S OFFICE Małgorzata Kurpesa, MD., PhD. Chair&Department of Cardiology Medical University of Łódź The heart is made up of four chambers Left Atrium Right Atrium Left Ventricle

More information

Activated Transponder & Activator Instruction Manual

Activated Transponder & Activator Instruction Manual Activated Transponder & Activator Instruction Manual Westhold Corporation General Warranty Modules and other equipment ("Goods") purchased from Westhold Corporation are warranted against defects in materials

More information

STANDARD INDOOR SIGNS INSTRUCTION MANUAL REVISION DATE: PART#:

STANDARD INDOOR SIGNS INSTRUCTION MANUAL REVISION DATE: PART#: STANDARD INDOOR SIGNS INSTRUCTION MANUAL REVISION DATE: 10-02-2009 PART#: 98-0016-01 SERVICE & CUSTOMER INFORMATION CUSTOMER MUST HAVE PART NUMBER WHEN ORDERING ITEMS THROUGH THE SERVICE DEPARTMENT. IF

More information

Cardioversion / Defibrillation / Transcutaneous Pacer Helpful Hints [ZOLL] 2018

Cardioversion / Defibrillation / Transcutaneous Pacer Helpful Hints [ZOLL] 2018 Cardioversion / Defibrillation / Transcutaneous Pacer Helpful Hints [ZOLL] 08 4 Universal cable may be inserted to the side Defib Short Block to perform 0 J check Universal Cable must be connected to APEX

More information

Patient Simulator. Instruction Manual. 04/2007, Peter Kron GmbH Subject to Technical Changes

Patient Simulator. Instruction Manual. 04/2007, Peter Kron GmbH Subject to Technical Changes Patient Simulator Instruction Manual Subject to Technical Changes Zossener Straße 41, D-10961 Berlin, Germany Tel.: +49 (0)30 / 69 81 88-40 Fax: +49 (0)30 69 81 88 49 Contents 1. Purpose and Application...

More information

Rhythm ECG Characteristics Example. Normal Sinus Rhythm (NSR)

Rhythm ECG Characteristics Example. Normal Sinus Rhythm (NSR) Normal Sinus Rhythm (NSR) Rate: 60-100 per minute Rhythm: R- R = P waves: Upright, similar P-R: 0.12-0.20 second & consistent P:qRs: 1P:1qRs Sinus Tachycardia Exercise Hypovolemia Medications Fever Substances

More information

2) Heart Arrhythmias 2 - Dr. Abdullah Sharif

2) Heart Arrhythmias 2 - Dr. Abdullah Sharif 2) Heart Arrhythmias 2 - Dr. Abdullah Sharif Rhythms from the Sinus Node Sinus Tachycardia: HR > 100 b/m Causes: o Withdrawal of vagal tone & Sympathetic stimulation (exercise, fight or flight) o Fever

More information

Appendix D Output Code and Interpretation of Analysis

Appendix D Output Code and Interpretation of Analysis Appendix D Output Code and Interpretation of Analysis 8 Arrhythmia Code No. Description 8002 Marked rhythm irregularity 8110 Sinus rhythm 8102 Sinus arrhythmia 8108 Marked sinus arrhythmia 8120 Sinus tachycardia

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Reading Assignment (p1-91 in Outline ) Objectives What s in an ECG?

Reading Assignment (p1-91 in Outline ) Objectives What s in an ECG? Reading Assignment (p1-91 in Outline ) Objectives What s in an ECG? The 5-Step Method ECG #: Mearurements: Rhythm (s): Conduction: Waveform: Interpretation: A= V= PR= QRS= QT= Axis= 1. Compute the 5 basic

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Arrhythmia Study Guide 3 Junctional and Ventricular Rhythms

Arrhythmia Study Guide 3 Junctional and Ventricular Rhythms Arrhythmia Study Guide 3 Junctional and Ventricular Rhythms JUNCTIONAL RHYTHMS The AV Junction (Bundle of His and surrounding cells) only acts as pacemaker of the heart when the SA Node is not firing normally

More information

ABCs of ECGs. Shelby L. Durler

ABCs of ECGs. Shelby L. Durler ABCs of ECGs Shelby L. Durler Objectives Review the A&P of the cardiac conduction system Placement and obtaining 4-lead and 12-lead ECGs Overview of the basics of ECG rhythm interpretation Intrinsic

More information

COURSE DESCRIPTION. Rev 2.0 7/2013. Page 1 of 26

COURSE DESCRIPTION. Rev 2.0 7/2013. Page 1 of 26 COURSE DESCRIPTION Ventricular dysrhythmias (arrhythmias) are unique and potentially dangerous cardiac rhythms. They are often associated with Code Blue calls, and life and death situations. The only two

More information

Electrocardiography for Healthcare Professionals

Electrocardiography for Healthcare Professionals Electrocardiography for Healthcare Professionals Kathryn A. Booth Thomas O Brien Chapter 10: Pacemaker Rhythms and Bundle Branch Block Learning Outcomes 10.1 Describe the various pacemaker rhythms. 10.2

More information

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah DEFIBRILLATORS Prof. Yasser Mostafa Kadah Basics Defibrillation is definitive treatment for life-threatening cardiac arrhythmias such as ventricular fibrillation Defibrillation consists of delivering therapeutic

More information

Cardiac Arrhythmia How to approach นพ.พ น จ แกวส วรรณะ หน วยโรคห วใจและหลอดเล อด

Cardiac Arrhythmia How to approach นพ.พ น จ แกวส วรรณะ หน วยโรคห วใจและหลอดเล อด Cardiac Arrhythmia How to approach นพ.พ น จ แกวส วรรณะ หน วยโรคห วใจและหลอดเล อด EKG paper is a grid where time is measured along the horizontal axis. Each small square is 1 mm in length and represents

More information

INSTRUCTION MANUAL. phantom 320 ECG Simulator 1.4

INSTRUCTION MANUAL. phantom 320 ECG Simulator 1.4 VERSION 1.4 INSTRUCTION MANUAL phantom 320 ECG Simulator This instruction manual is an essential part of the system, in accordance with EN61010-1. Comply with it and keep it. MedTec-Science GmbH July,

More information

Multiparameter Simulation

Multiparameter Simulation MedSim 400 Multiparameter Simulation Operation Manual Dynatech CBET 2525 Industrial Street Leesburg, FL 34748 Dynatech-CBET.com I content Introduction... 1 Safety Considerations... 1 Abbreviation and Symbols...

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

Project Title Temporary Pacemaker Training Simulator

Project Title Temporary Pacemaker Training Simulator Project Title Temporary Pacemaker Training Simulator Project Description Problem: There is no available training device for temporary pacemakers (pacemakers). A training device will have to essentially

More information

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work Electrical Activity of the Human Heart Oguz Poroy, PhD Assistant Professor Department of Biomedical Engineering The University of Iowa Outline Basic Facts about the Heart Heart Chambers and Heart s The

More information

HR: 50 bpm (Sinus) PR: 280 ms QRS: 120 ms QT: 490 ms Axis: -70. Sinus bradycardia with one ventricular escape (*)

HR: 50 bpm (Sinus) PR: 280 ms QRS: 120 ms QT: 490 ms Axis: -70. Sinus bradycardia with one ventricular escape (*) 1? HR: 50 bpm (Sinus) PR: 280 ms QRS: 120 ms QT: 490 ms Axis: -70 1 Sinus P waves? 2 sinus cycles The pause (2 sinus cycles) suggests that the sinus fired (?) but did not conduct to the atria (i.e., missing

More information

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa INTENSIVE CARE POLICY & PROCEDURES Policy Number: R-25 Subject: Policy: Purpose: Cardioversion - Synchronized An RN may perform a synchronized

More information

SENSATION and CS300 Smaller meets faster. IABP System

SENSATION and CS300 Smaller meets faster. IABP System SENSATION and CS300 Smaller meets faster IABP System Cardiac Assist SENSATION and CS300 3 Cardiac Assist SENSATION and CS300 5 Smaller The SENSATION 7Fr. Fiber-Optic IAB Catheter Faster The CS300 with

More information

EKG Competency for Agency

EKG Competency for Agency EKG Competency for Agency Name: Date: Agency: 1. The upper chambers of the heart are known as the: a. Atria b. Ventricles c. Mitral Valve d. Aortic Valve 2. The lower chambers of the heart are known as

More information

Supraventricular Arrhythmias. Reading Assignment. Chapter 5 (p17-30)

Supraventricular Arrhythmias. Reading Assignment. Chapter 5 (p17-30) Supraventricular Arrhythmias Reading Assignment Chapter 5 (p17-30) The Supraventricular Rhythms In Our Lives Site of Origin Single Events Slow Rates Intermediate Rates Fast Rates (>100 bpm) Sinus Sinus

More information

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Testing the Accuracy of ECG Captured by through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Data Analysis a) R-wave Comparison: The mean and standard deviation of R-wave amplitudes

More information

12-Lead ECG Interpretation. Kathy Kuznar, RN, ANP

12-Lead ECG Interpretation. Kathy Kuznar, RN, ANP 12-Lead ECG Interpretation Kathy Kuznar, RN, ANP The 12-Lead ECG Objectives Identify the normal morphology and features of the 12- lead ECG. Perform systematic analysis of the 12-lead ECG. Recognize abnormalities

More information

Pennsylvania Academy of Family Physicians Foundation & UPMC 43rd Refresher Course in Family Medicine CME Conference March 10-13, 2016

Pennsylvania Academy of Family Physicians Foundation & UPMC 43rd Refresher Course in Family Medicine CME Conference March 10-13, 2016 Pennsylvania Academy of Family Physicians Foundation & UPMC 43rd Refresher Course in Family Medicine CME Conference March 10-13, 2016 Disclosures: EKG Workshop Louis Mancano, MD Speaker has no disclosures

More information

ECG Interpretation and Clinical Significance

ECG Interpretation and Clinical Significance Confirming Pages ECG Interpretation and Clinical Significance Chapter Outline 5.1 5.2 5.3 5.4 5.5 5.6 5.7 5.8 5.9 5.10 Learning Outcomes Key Terms Introduction (p. 114) Identifying the Components of the

More information

Basic ECG Interpretation Module Notebook

Basic ECG Interpretation Module Notebook Basic ECG Interpretation Module Notebook ECG_Notebook_04.27.05 Page 1 of 142 Basic ECG Interpretation Table of Contents Module Objectives... 3 Module Outline... 6 Lesson I... 6 Lesson II... 8 Lesson III...

More information

Module 1: Introduction to ECG & Normal ECG

Module 1: Introduction to ECG & Normal ECG Module 1: Introduction to ECG & Normal ECG Importance of Correct anatomical positions Measurements & Morphologies ONLY accurate if Precise anatomical positions adhered to Standardised techniques are used

More information

SARASOTA MEMORIAL HOSPITAL

SARASOTA MEMORIAL HOSPITAL SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE PACEMAKERS AND DETERMINING PACEMAKER DATE: REVIEWED: PAGES: 10/99 11/18 1 of 7 PS1094 ISSUED FOR: Nursing RESPONSIBILITY: *Qualified RN PURPOSE: To establish

More information

Diploma in Electrocardiography

Diploma in Electrocardiography The Society for Cardiological Science and Technology Diploma in Electrocardiography The Society makes this award to candidates who can demonstrate the ability to accurately record a resting 12-lead electrocardiogram

More information

TIP CARD. Model Temporary External Pacemaker

TIP CARD. Model Temporary External Pacemaker TIP CARD Model 5392 Temporary External Pacemaker PRE-USE What to do... What NOT to do... Battery Replace the battery for each new patient. Do NOT reuse battery. Physical Check case for cracks/damage.**

More information

ER75 Electro-Acoustic Ear Simulator. Operating Manual

ER75 Electro-Acoustic Ear Simulator. Operating Manual ER75 Electro-Acoustic Ear Simulator Operating Manual ABOUT THIS MANUAL READ THIS OPERATING MANUAL BEFORE ATTEMPTING TO USE THE INSTRUMENT. Amplivox Ltd. 6 Oasis Park, Eynsham Oxfordshire, OX29 4TP United

More information

Electrocardiography for Healthcare Professionals

Electrocardiography for Healthcare Professionals Electrocardiography for Healthcare Professionals Chapter 7: Junctional Dysrhythmias 2012 The Companies, Inc. All rights reserved. Learning Outcomes 7.1 Describe the various junctional dysrhythmias 7.2

More information

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD External Pacemaker Set-up and Revised Date: 11/13/2014 Page 1 of 5 POLICY: The transcutaneous external pacemaker will be used as a noninvasive method of delivering electrical stimuli to the myocardium

More information

8/20/2012. Learning Outcomes (Cont d)

8/20/2012. Learning Outcomes (Cont d) 1 2 3 4 Electrocardiography for Healthcare Professionals Chapter 7: Junctional Dysrhythmias Learning Outcomes 7.1 Describe the various junctional dysrhythmias 7.2 Identify premature junctional complexes

More information

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION 1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,

More information

PACIFIC MEDICAL TRAINING Arrhythmia Interpretation

PACIFIC MEDICAL TRAINING Arrhythmia Interpretation PACIFIC MEDICAL TRAINING Arrhythmia Interpretation Introduction Activity Summary Target Audience Educational Objectives Nursing Educational Objective Faculty Physician Continuing Medical Education Nursing

More information

Cardiac Telemetry Self Study: Part One Cardiovascular Review 2017 THINGS TO REMEMBER

Cardiac Telemetry Self Study: Part One Cardiovascular Review 2017 THINGS TO REMEMBER Please review the above anatomy of the heart. THINGS TO REMEMBER There are 3 electrolytes that affect cardiac function o Sodium, Potassium, and Calcium When any of these electrolytes are out of the normal

More information

This presentation will deal with the basics of ECG description as well as the physiological basics of

This presentation will deal with the basics of ECG description as well as the physiological basics of Snímka 1 Electrocardiography basics This presentation will deal with the basics of ECG description as well as the physiological basics of Snímka 2 Lecture overview 1. Cardiac conduction system functional

More information

TABLE OF CONTENTS 1.1. WARRANTY DAMAGED SHIPMENTS SERVICE INTRODUCTION UNPACKING AND INSTALLATION

TABLE OF CONTENTS 1.1. WARRANTY DAMAGED SHIPMENTS SERVICE INTRODUCTION UNPACKING AND INSTALLATION TABLE OF CONTENTS EDITION 5 Section 1 - WARRANTY & SERVICE 1.1. WARRANTY 2 1.2. DAMAGED SHIPMENTS 2 1.3. SERVICE 2 Section 2 - INTRODUCTION 2.1. INTRODUCTION 3 Section 3 - GENERAL 3.1. UNPACKING AND INSTALLATION

More information

ProSim 8 Vital Signs Simulator

ProSim 8 Vital Signs Simulator ProSim 8 Vital Signs Simulator Technical Data The 8-in-1 ProSim 8 Vital Signs Simulator offers fast and comprehensive preventative maintenance (PM) testing for your entire patient monitor fleet. Designed

More information

Model 5392 EPG Temporary Pacer

Model 5392 EPG Temporary Pacer Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short

More information

Rate: The atrial and ventricular rates are equal; heart rate is greater than 100 bpm (usually between bpm).

Rate: The atrial and ventricular rates are equal; heart rate is greater than 100 bpm (usually between bpm). Sinus Bradycardia Regularity: The R-R intervals are constant; the rhythm is regular. Rate: The atrial and ventricular rates are equal; heart rate is less than 60 bpm. P wave: There is a uniform P wave

More information

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II). Cardiac Pacemaker Premature Beats When one of ectopic foci becomes irritable, it may spontaneously fire, leading to one or more premature beats. Atrial and junctional foci may become irritable from excess

More information