Numerical Study on Effects of Drug-coating Position of Drug-eluting Stents on Drug Concentration

Size: px
Start display at page:

Download "Numerical Study on Effects of Drug-coating Position of Drug-eluting Stents on Drug Concentration"

Transcription

1 Journal of Medical and Biological Engineering, 34(5): Numerical Study on Effects of Drug-coating Position of Drug-eluting Stents on Drug Concentration Yu Chen 1 Fei Yan 1 Wen-Tao Jiang 1,* Qing-Yuan Wang 1 Yu-Bo Fan 2 1 Laboratory of Biomechanical Engineering, Department of Applied Mechanics, Sichuan University, Chengdu , China 2 School of Biological Science and Medical Engineering, Beihang University, Beijing , China Received 22 May 2013; Accepted 26 Sep 2013; doi: /jmbe.1570 Abstract As consequences of the implantation of drug-eluting stents (DESs), hemodynamic changes and drug diffusion have significant effects on in-stent restenosis. In particular, the drug-coating position on DESs can affect drug concentration distribution in blood and the vascular wall. Therefore, investigating the effects of the drug-coating position on drug distribution is important for the optimal design of DESs, in terms of both geometry and drug coatings. In this paper, three-dimensional virtual stent models with different drug-coating positions were built to numerically investigate the drug concentration distribution. The results were compared with those from two-dimensional simulation models. It was demonstrated that an increase of the contacting surface increases drug concentration in the vascular wall, and constrains the increase of the low wall shear stress (WSS) region (which normally increases with the height of DESs). The results show that the drug coated on the contacting surface effectively enters the vascular wall without interference from blood flow. Increasing the drug on the contacting surface and decreasing the height of a DES are optimal for reducing the risk of restenosis. Keywords: Drug-eluting stent (DES), Drug concentration, Drug-coating position, Hemodynamics, Contacting surface 1. Introduction Coronary arteries are the most susceptible arteries to atherosclerosis, which is a common occurrence with cardiovascular disease. Drug-eluting stent (DES) therapy is considered as a milestone in the evolution of percutaneous coronary intervention therapy [1]. However, studies have found that DESs can increase the risks of cardiac death and myocardial infarction, which are caused by coronary thrombosis in non-cardiac death [2]. Many researchers have investigated the effects of geometry, drug-coating/loading scheme, and drug release patterns of stents. A previous study [3] analyzed the influence of Reynolds number and the width/ height ratio of the stent on drug concentration distribution; they found that coronary DESs should be considered where luminal flow, strut design and pulsatility have direct effects on tissue drug uptake after local delivery. Researchers [4] studied the contribution of the drug released into the blood flow with respect to the efficacy of drug deposition and penetration into the arterial walls using three-dimensional (3D) models. Other schemes [5] simulated the concentration distribution of heparin and taxol in blood. An experimental approach [6] confirmed * Corresponding author: Wen-Tao Jiang Tel: ; Fax: scubme@aliyun.com that drug-coating position significantly affects drug distribution. Balakrishnan [7,8] and Dong [9] applied numerical methods to investigate the effects of drug-coating position on drug distribution. Balakrishnan studied two-dimensional (2D) straight artery stents and found that increasing the stent height alone increases drug concentration. Dong studied 2D curved artery stents and found that the drug-coating position and stent spacing have significant effects on drug deposition in a curved artery. The present study uses 3D models to further study the influence of drug-coating position in DESs on the drug distribution in blood and the vascular wall. Computational fluid dynamics (CFD) is used to investigate the effects of drug-coating position on the drug concentration in 3D straight/curved DESs. The relationships between drug-coating position, drug concentration, and in-stent restenosis are studied. The results provide theoretical guidance for the optimization of DES design. 2. Methods 2.1 Models To compare the effects of drug-coating position on drug concentration, two stents with different configurations were built and respectively implanted into a straight vessel and a curved vessel. Below, the stent in the straight vessel is referred to as the straight stent and that in the curved vessel is referred

2 488 J. Med. Biol. Eng., Vol. 34 No to as the curved stent. The configuration of the straight stent, with four stent struts, is shown in Fig. 1. The coronary artery diameter is 3 mm [10], the arterial wall thickness is 0.5 mm [11], and the stent diameter and length are 3 and 9 mm, respectively [7]. The cross-section of a stent strut is a 0.1 mm 0.1 mm square. The configuration of the curved stent, with six stent struts, is shown in Fig. 1. The coronary artery diameter is 3 mm, the arterial wall thickness is 0.5 mm, and the stent diameter and length are 3 and 15 mm, respectively. The curvature radius is 10 mm. The cross-section of a stent strut is a 0.1 mm 0.1 mm square. For both stent configurations, the stent struts and links are identical in size, thickness, and surface area. The stent surface was assumed to perfectly fit with the vascular wall. grid independence, which was achieved at 3,000,000 cells for all models. 2.2 Boundary conditions Blood was assumed to be a steady, isotropic, and incompressible Newtonian fluid with constant density and viscosity. The vessel walls were assumed to be impermeable and rigid; although to some extent this assumption is not accurate, the elasticity of the walls is less important than the gross features of the flow. Previous studies have shown that the effect of non-newtonian fluid behavior is negligible [12,13]. Consequently, the governing equations can be written as: u j 0 x j u jui P u u i j ( ) x j xi x j x j xi ( i, j 1,2,3) ( 1 ) where and are the density and dynamic viscosity of blood, respectively. = 1055 kg/m 3 and = kg/m s [14]. u i is the velocity vector of blood and P is the pressure. The parabolic profile with a centerline velocity of 0.46 m/s [7] was applied at the inlet: Figure 1. 3D drug-eluting stent models in straight and curved arteries. According to different drug-coating positions on the stents, there are four models for each stent configuration (straight and curved): Model A: all-coated stents, i.e., drug coating on all surfaces (I, II, III, and IV), as shown in Fig. 2; Model B: drug coating on the contacting surface (I); Model C: drug coating on the non-contacting proximal and distal surfaces (II and III); Model D: drug coating on the non-contacting top surface (IV). V z inlet 2 2 x y 0.46 (1 ) 2 r V 0 r inlet where V z and V r are the blood velocities in the axial and radial directions, respectively, with radius r = m. A zero-pressure boundary condition was applied at the outlet: P 0 (3) outlet Luminal drug transport was described by the steady-state convection-diffusion equation and drug transport through the vascular wall was modeled by the steady-state diffusion equation: (2) f ( uif kf ) 0 xi xi t ( kt ) 0 xi xi ( i 1, 2,3) (4) where f and t are the drug concentrations in the blood and vascular wall, respectively. The diffusion coefficients of the drug in blood ( kf ) and tissue ( kt ) are 10-7 and m 2 /s, respectively [7]. Figure 2. Coating positions of drug-coated section. Gambit (version 2.3, ANSYS, Inc., USA) software was used to mesh these models. Unstructured meshes were selected for all eight models. Grids were refined at the stents and the vascular surfaces to improve computational accuracy. The size-function in Gambit and the grid adaptation in Fluent (version 6.3, ANSYS, Inc.) were applied to refine the meshes, particularly in the areas near the stents. The refinement on the meshes was stopped when the computational results reached f inlet t r f kf r f 0, 0 z perivascular wall blood outlet 0 t kt r tissue (5) (6) (Tissue-blood interface) (7) t 0 (Upstream/downstream boundaries of the tissue) (8) z

3 Numerical Analysis of Drug Coating of DES 489 1(1 indicates the stent surfaces assigned for drug-coatin)(9) The drug transport in tissues was modeled as a simple diffusion process, with an impermeable boundary condition on the perivascular wall (shown in Eq. 6). The continuity of flux at the tissue-blood interface is shown in Eq. 7. The boundary conditions of the upstream and downstream boundaries of the tissue (shown in Eq. 8) allowed drug transport to distal tissue segments. As shown in Eq. 9, the drug release of the stents was simulated with a Dirichlet boundary condition. For the drug coated, unit 1 was defined as 100% full drug coated and 0 means no drug on the stent surface, thus, the drug concentration values of the drug-coated surfaces in Models A, B, C and D were set as 1. All the values of drug concentration mentioned in this paper, which represent the range of the drug concentration from 0 to 1, are relative value, and the maximum value (near 1) indicates relatively high drug concentration. The computations were conducted by using the commercial CFD package, FLUENT 6.3(ANSUS Inc.), in which a finite volume method was used to discretize the governing equations. A 3D single-precision format and a segregated solver were used, and the SIMPLEC algorithm was applied for the velocity-pressure correction. The standard format was chosen for the pressure discretization and a second-order upwind format was selected for the momentum equations. The convergence threshold of residual error was set as Results To clearly observe the hemodynamic behavior and drug concentration for the two stent configurations, three segments and four lines were chosen for the investigation (as shown in Fig. 3). The stented arteries in the models were divided into proximal, middle, and distal segments. Line A is the reference line in blood with 1.25-mm offset from the centre line of blood vessel, and line B, C and D are the reference lines with a 1.45-mm, 1.55-mm and 1.75 mm offset from the centre line of blood vessel, respectively. Figure 3. Illustrations of reference lines and segments. The drug concentration curves of lines A and B (reference lines in blood) are shown in Figs. 4 and 4, respectively. It can be seen that the drug concentration on line A in Model C is lower than that in Model D, and that recirculation zones formed near the stents, causing drug deposition and thus a high drug concentration. As shown in the circle in Fig. 4, the drug concentration on line B in Model C is much higher than that in Model D. This indicates the effect of the stents on drug deposition near the vascular wall: the closer to the vascular wall, the higher is the level of deposition [15]. There was nearly no drug released into the blood in Model B. Figure 4. Drug concentration curves in straight blood vessel. Lines A and B. The drug concentration curves on lines C and D (in the vascular wall) are shown in Figs. 5 and 5, respectively. The drug concentration peak on line C appears on the stent struts and links in the four models. Because the drug in Model B directly entered the vascular wall, the peak value reached 70% that in Model A. In Models C and D, recirculation zones were formed by the stent struts and links in blood, which caused drug accumulation and high-drug-concentration regions, eventually increasing the amount of drug entering the vascular wall. In contrast, the drug concentration of the regions without stent struts and links is very low, as shown in Model B, because the drug entry into the vascular wall in Model B is different from those in Models C and D. The drug concentration curves in Models C and D are similar for a given flow field, and the drug concentration in the distal region is higher than that in the proximal region. Because the drug diffused in both the circumferential and radial directions, all drug concentration curves are flat on line D. Due to the diffusion effect on the regions without stent struts and links, the drug concentrations in Models B and C also show a significant increase. As the near-stent drug concentration in Model D is relatively low, the effect of the drug spread from the stent struts and links to the regions far from the stent struts and links is not obvious. The above analysis demonstrates that: (1) as the drug diffused along the radial direction into the vascular wall, the drug concentration in Model B increased and became uniform; (2) the drug concentration in Model A was the sum of those in Models B, C, and D; (3) there was no interaction between the different drug-coated surfaces in the vascular walls.

4 490 J. Med. Biol. Eng., Vol. 34. No Figure 5. Drug concentration curves on straight vascular wall. Lines C and D. To better observe the drug diffusion in the circumferential direction in the vascular walls, a cross section was selected in the middle of the stented blood vessel. In the cross section, the drug concentration at the mid-wall position of the vascular walls (as shown in Fig. 6) was unfolded in the circumferential direction. The drug concentration curves are shown in Fig. 6. It can be clearly seen that the drug concentration peak near the stent struts and links appeared in Models A, B, and C, but not in Model D. The peak-trough difference is 0.2 in Model A, 0.12 in Model B, and 0.09 in Model C. This implies that the increase of drug concentration near the stent struts and links was mostly caused by the drug coating on the contacting surface and the non-contacting proximal/distal surfaces, with the drug coating on the noncontacting top surface having little effect. Figure 6. Drug distribution in circumferential direction in vascular wall. Reference location and drug concentration curves. The drug concentration curves of inner/outer bends along line A in the curved vessel are shown in Figs. 7 and 7, respectively. The drug concentration trends in Models C and D are similar, showing high drug concentrations near the stent struts and links. Due to blood flow, the drug concentration is near zero in the proximal region. There is a ten-fold difference in the drug concentration between the inner and outer bends of the stents. Blood flow greatly influences the drug concentration at the outer bend of the stents, and the curve shows significant fluctuation. On line A, the drug concentration in Model D is higher than that in Model C, and the drug concentration in Model B is nearly zero. The drug concentration curves on line B of the inner/outer bends of the curved stents are shown in Figs. 7(c) and 7(d), respectively. It can be seen that the drug concentration difference between the inner and outer bends is reduced; the drug concentration of the inner bend is 3 to 4 times greater than that of the outer bend; the drug concentration peaks of the inner and outer bends are nearly the same, but the drug concentration differences in regions far from the stent struts and links are obvious. For the drug concentration at the inner bend, the first stent struts are consistent, but the remaining stent struts showed the proximal high and distal low phenomena, that is high concentrations at proximal segments and low concentrations at distal segments. The drug concentration curves of the inner/outer bends of the curved vascular wall on line C are shown in Figs. 8 and, respectively. There are still drug concentration differences between the inner and outer bends, but the differences have gotten smaller. Even though the drug in Model B is free from blood flow and theoretically the drug concentration of the inner and outer bends should be consistent, in reality there are differences. This is because under bending deformation, the outer bend is stretched and the inner bend is compressed. It means that if the same amount of drug enters the different volumes, drug concentration in the inner bend and outer bend of the vascular wall will be different. With the influence of blood flow and stent bends, in the regions far from the stent struts and links, the drug concentration of the inner bend is higher than that of the outer bend. The drug concentration curves of the inner and outer bends of the curved vessel on line D are shown in Figs. 8(c) and (d), respectively. There are still drug concentration differences for the inner bend and outer bend, especially in Models C and D, for which the drug concentration of the inner bend is nearly twice as much as that of the outer bend. In Model B, the peak value of drug concentration near the struts and links in the vascular wall decreased from around 75% to around 50%; however, the drug concentration between the struts and links significantly increased. This implies that in high-drug-concentration regions, the drug diffused in the radial and circumferential directions, and the drug concentration gradually became uniform. The biggest difference between the drug concentrations of the inner bend and the outer bend is the significant decrease between the stent struts and the links. In particular, Models C and D show significant decline of the outer-bend drug concentration between the second and third struts; the drug concentration in the distal region was smaller than that in the proximal region.

5 Numerical Analysis of Drug Coating of DES 491 (c) (c) (d) Figure 7. Drug concentration curves in Lines A and B in the blood. Inner and outer bends of Line A and (c) inner and (d) outer bends of Line B. This implies that for the curved stents, the effects of blood flow and stent bend on the drug concentration should be taken into account. In Models C and D, the drug concentrations in the middle and proximal regions were half that of the inner bend stents. The comparisons of mean values of the drug concentrations in the proximal, middle, and distal regions of the stents are shown in Fig. 9. For Model B, the drug concentration had few changes in the proximal, middle, and distal regions. Due to blood flow, Models C and D show gradual increases of drug concentration. In particular, for Model D, the drug concentration in the proximal region was half that in the distal region. In Model B, the drug directly entered the vascular wall with little influence from blood flow; therefore, the drug concentrations are nearly same in the proximal, middle, and distal regions of stents in the straight vessel. The drug concentrations in Models C and D show the same trend, namely (d) Figure 8. Drug concentration curves in Lines C and D on vascular wall. Inner and outer bends of Line C and (c) inner and (d) outer bends of Line D. increasing distribution curves for both straight and curved vessels. Because the non-contacting top surface is closer to blood, the drug concentration in Model D shows an obvious phenomenon- proximal low and distal high, that is low concentrations at proximal segments and high concentrations at distal segments, especially in the curved vessel. Based on the above analysis, the different drug concentrations in the proximal and distal regions are caused by the drug coating on the non-contacting surfaces of the stents. The drug concentration in the curved vessel does not only show the phenomena of proximal low and distal high, which is caused by blood flow, but also show the phenomena of inner high and outer low, that is high concentrations in inner bend and low concentrations in outer bend, which is caused by different curves (as shown in Fig. 10). In terms of the overall drug concentration, the drug concentration of the outer bend in the curved vessel is 70.4% that of the inner bend; more precisely, it is 86%, 67.5%, and 64% those for the proximal,

6 492 J. Med. Biol. Eng., Vol. 34 No Figure 9. Comparison of drug concentration for the models in proximal, middle, and distal regions of stents in straight and curved vessels. Figure 11. Comparison of percentages of Model B, C, and D for allcoated Model A in 2D and 3D models. 4. Discussion Figure 10. Comparison of drug concentration at inner and outer bends in curved artery. middle, and distal regions, respectively. This implies that the differences in drug concentration between the proximal and distal regions, as well as between the inner and outer bends, are caused by blood flow and vessel curve. The volume-weighted average concentrations in the straight and curved models were calculated. The results are shown in Fig. 11. In the straight vessel, the drug amounts from the contacting surface and the non-contacting top surface respectively account for 27% and 27% of the total drug deposition, compared to 11% and 30% in 2D models [7]. In the curved vessel, the drug amounts from contacting surface and the non-contacting top surface respectively account for 25.4% and 27.2% of the total drug deposition, compared to 12% and 27% in 2D models [9]. The percentages of Model B, C and D are the proportion of drug concentration in the vascular wall for all-coated surfaces Model A (Model A is 100% in this situation), so the sum of Model B, C, and D is 100%. In numerical simulations, the vascular wall is commonly simplified as a rigid wall [16]. Studies have demonstrated that this assumption has little impact on the calculation results [17]. Even though the conditions do not perfectly reflect reality, the surface elasticity of the wall is not very critical under the flow conditions considered here [12]. Moreover, the arterial elasticity is gradually reduced during the development of atherosclerosis, and after stents are implanted, the arterial rigidity gets strengthened [13]. Therefore, all in all, the assumption of rigid wall in the models is reasonable. For large and medium arteries, it is appropriate to simplify the blood as an incompressible Newtonian fluid with constant density and viscosity. Because red blood cells are smaller than the blood vessel diameter in large and medium arteries, blood can be considered a continuous fluid. Because the diameter of blood vessels is larger than 0.5 mm, the blood flow shear rate is very high in the vessel, and consequently blood viscosity becomes a constant and has no relationship with the shear rate [18]. Finally, calculation error should be less than 2% [19]. Therefore for blood in large and medium arteries, it is appropriate to simplify the blood as an incompressible Newtonian fluid of constant density and viscosity. From the drug concentration curves in the blood and vascular wall, it can be seen that the drug concentration in the vascular wall consists of two parts. Some of the drug entered the vascular wall directly from the contacting surface and some entered the blood from the non-contacting surfaces, and finally the vascular wall. The total drug quantity is the sum of these two parts.

7 Numerical Analysis of Drug Coating of DES 493 Intuitively, if the drug entered the vascular wall directly from the contacting surface, the advantages are no influence of blood flow and less drug loss, and the disadvantage are uneven drug distribution in the vascular wall (the drug concentration near the stents could be much higher than that of regions far from the stents). In contrast, if the drug entered the blood from the non-contacting surfaces, the drug concentration in the vascular wall is more even, but the influence of blood flow is large, and thus the drug concentration in the distal region is much higher than that in the proximal region and the drug loss is large. Based on the above analysis, the following can be concluded: (1) Along with the diffusion of the drug in the radial direction of the vascular wall, the drug concentration curves tended to be flat because the drug in the circumferential direction can spread to the region far from the stents, and consequently the drug concentration in this region increased gradually. (2) The drug deposition from both the contacting and non-contacting proximal/distal surfaces can cause uneven distribution. Therefore, increasing the amount of drug on the contacting surfaces cannot always make the uniformity of drug distribution worse. (3) For the stents in a curved artery, as shown in Fig. 10, the drug concentration in the distal region is higher than that in the proximal region and that of the inner bend is higher than that of the outer bend. Blood flow in the curved artery thus leads to uneven drug concentration at inner/outer bends. (4) When the stents are bent, the outer bend links get stretched and the spaces between the outer bends of the struts become bigger, whereas the inner bend links get squeezed and the spaces between the inner bends of struts become smaller. This leads to different drug concentrations on the inner and outer bends. (5) Regarding drug deposition from the various surfaces, the trends from 2D models are consistent with those from 3D models, but the values are different. The results of 2D simulation may be completely different from those of 3D simulation. Therefore, even though a considerable amount of the drug in the vascular wall comes from the drug on the non-contacting top and proximal/distal surfaces, when all stent surfaces were drug-coated, the phenomenon- proximal low and distal high of drug distribution can appear in the blood and vascular walls. These uneven distributions are mainly caused by blood flow. In order to improve the uniformity of distribution, the drug coating on the surface of the stents should be optimized, such as using uneven drug coating (a heavier coating in the proximal region than that in the distal region). The previous analysis showed that because the drug coating on the non-contacting surface is affected by blood flow, which can affect the drug release into the vascular wall, an optimized drug coating for one blood flow state may not be suitable for another. However, optimizing the drug distribution for various states of blood flow is quite complicated and impractical. Drug coating on the contacting surface can provide effective drug release into the vascular wall that is unaffected by blood flow. The uniformity of drug distribution in the vascular wall can thus be improved by optimizing the drug loading on the contacting surface. Using 2D simulation to simplify the model analysis can reduce computation time and resources at the cost of some inaccuracy. When accuracy is important, 3D analysis should be used. As shown in Section 3, the amounts of the drug from the contacting surface calculated by the 3D method (27%, 25.4%) are much bigger than those obtained from 2D simulation (11%, 12%). The contribution of the contacting surface to the drug concentration in the vascular wall was underestimated in the 2D numerical simulation. This is due to the following reasons: (1) in 2D models, the non-contacting proximal/distal surfaces of stents are perpendicular to the direction of blood flow, and the drug is prone to gather in the recirculation region, whereas in 3D models, the blood flow vector and most of the proximal/ distal surfaces are at acute angles (in the same direction at some positions). (2) In 2D models, the drug only diffuses in the axial and radial directions in the vascular wall, whereas in 3D simulation models, the drug diffuses in the axial, radial, and circumferential directions, which is consistent with reality. Due to the limitations of 2D models, Balakrishnan [7] found that an increase of the stents height was the only factor that improved the drug concentration in the vascular wall. However, based on the results of 3D simulation, the contacting surfaces enhance drug release and absorption; therefore, increasing the area of the contacting surface improves the drug concentration in the vascular wall and avoids the increase of the low WSS region caused by an increase in DES height. Studies on the drug release and absorption of DES should thus be conducted using 3D models. 5. Conclusion In this paper, the effects of drug-coating position on the drug concentration were analyzed using 3D DES models. It was found that: In traditional 2D models, the blood flow directions are perpendicular to most of the non-contacting proximal/distal surfaces, but in 3D models, in most situations, the blood flow directions are not perpendicular to most of the non-contacting proximal/distal surfaces and the nonperpendicular cannot effectively block the blood flow and caused drug accumulation. Therefore, it is very necessary to study DES drug release and absorption using 3D models. Increasing the area of the contacting surface increases the drug concentration in the vascular wall and avoids the possible increase of the low WSS region caused by an increased DES height. Drug coating on the contacting surface can provide effective drug release in the vascular wall without interference from blood flow. Therefore, it is possible to improve the uniformity of drug concentration in the vascular wall by optimizing drug loading on the contacting surface. DES design can be optimized 1) by increasing the drug on the contacting surface (to improve drug concentration in the vascular wall) and 2) by reducing the stent height (to reduce the low WSS distribution). Such a design can make the distribution of drug concentration more uniform and inhibit restenosis.

8 494 J. Med. Biol. Eng., Vol. 34 No Acknowledgements This work is supported by Grants-in-Aid from the National Natural Science Research Foundation of China (No , ) and Applied Basic Research Programs of science & technology department of Sichuan Province (No. 2014JY0148). The authors would like to thank Dr. Jian Chen at the School of Mechanical and Manufacturing Engineering University of New South Wales, Australia, for his help during the paper writing. References [1] C. Yang and H. M. Burt, Drug-eluting stents: factors governing local pharmacokinetics, Adv. Drug Deliv. Rev., 58: , [2] A. V. Finn, G. Nakazawa, M. Joner, F. D. Kolodgie, E. K. Mont, H. K. Gold and R. Virmani, Vascular responses to drug eluting stents importance of delayed healing, Arterioscler. Thromb. Vasc. Biol., 27: , [3] V. B. Kolachalama, A. R. Tzafriri, D. Y. Arifin and E. R. Edelman, Luminal flow patterns dictate arterial drug deposition in stent-based delivery, J. Control. Release, 133: 24-30, [4] P. Zunino, C. D Angelo, L. Petrini, C. Vergara, C. Capelli and F. Migliavacca, Numerical simulation of drug eluting coronary stents: mechanics, fluid dynamics and drug release, Comput. Meth. Appl. Mech. Eng., 198: , [5] C. Hwang, D. Wu and E. R. Edelman, Physiological transport forces govern drug distribution for stent-based delivery, Circulation, 104: , [6] A. Seidlitz, S. Nagel, B. Semmling, N. Grabow, H. Martin, V. Senz, C. Harder, K. Sternberg, K. P. Schmitz, H. K. Kroemer and W. Weitschies, Examination of drug release and distribution from drug-eluting stents with a vessel-simulating flow-through cell, Eur. J. Pharm. Biopharm., 78: 36-48, [7] B. Balakrishnan, A. R. Tzafriri, P. Seifert, A. Groothuis, C. Rogers and E. R. Edelman, Strut position, blood flow, and drug deposition implications for single and overlapping Drug-Eluting Stents, Circulation, 111: , [8] B. Balakrishnan, J. F. Dooley, G Kopia and E. R. Edelman, Intravascular drug release kinetics dictate arterial drug deposition, retention, and distribution, J. Control. Release, 123: , [9] R. Dong, W. Jiang and T. Zheng, Numerical analysis on drug deposition from drug-eluting stents in curved artery, J. Med. Biomech., 26: 13-17, [10] J. F. LaDisa, L. E. Olson, R. C. Molthen, D. A. Hettrick, P. F. Pratt, M. D. Hardel, J. R. Kersten, D. C. Warltier and P. S. Pagel, Alterations in wall shear stress predict sites of neointimal hyperplasia after stent implantation in rabbit iliac arteries, Am. J. Physiol.-Heart Circul. Physiol., 288: H2465-H2475, [11] A. Kastrati, J. Mehilli, J. Dirschinger, J. Pache, K.Ulm, H. Schühlen, M. Seyfarth, C. Schmitt, R. Blasini, F. J. Neumann and A. Schömig, Restenosis after coronary placement of various stent types, Am. J. Cardiol., 87: 34-39, [12] T. Zheng, W. Wang, W. Jiang, X. Deng and Y. Fan, Assessing hemodynamic performances of small diameter helical grafts: transient simulation, J. Mech. Med. Biol., 12: , [13] D. Massai, G. Soloperto, D. Gallo, X. Xu and U. Morbiducci, Shear-induced platelet activation and its relationship with blood flow topology in a numerical model of stenosed carotid bifurcation, Eur. J. Mech. B-Fluids, 35: , [14] J. J. Wentzel, R. Krams, J. C. Schuurbiers, J. A. Oomen, J. Kloet, W. J. van der Giessen, P. W. Serruys and C. J. Slager, Relationship between neointimal thickness and shear stress after Wallstent implantation in human coronary arteries, Circulation, 103: , [15] X. Chen, W. Jiang, Y. Fan, Y. Chen, T. Zheng, M. Guo and J. Chen, Numerical investigation of effects of drug eluting stent link on flow field and concentration field of Hemoynamics, Space Med. Med. Eng., 23: , [16] Z. Chen, Y. Fan, X. Deng and Z. Xu, A new way to reduce flow disturbance in endovascular stents: a numerical study, Artif. Organs, 35: , [17] U. Morbiducci, R. Ponzini, M. Grigioni and A. Redaelli, Helical flow as fluid dynamic signature for atherogenesis risk in aortocoronary bypass. A numeric study, J. Biomech., 40: , [18] W. Jiang, T. Zheng, J. Chen, X. Deng and Y. Fan, Numerical investigation of pulsatile flow in an S-type bypass graft, J. Mech. Med. Biol., 12: , [19] C. Tu, M. Deville, L. Dheur and L. Vanderschuren, Finite element simulation of pulsatile flow through arterial stenosis, J. Biomech., 25: , 1992.

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents Dublin Institute of Technology ARROW@DIT Conference Papers School of Mechanical and Design Engineering 2008-09-01 Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational

More information

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm I. Husain, C. Langdon and J. Schwark Department of Mathematics Luther College University of Regina Regina, Saskatchewan

More information

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches Korea-Australia Rheology Journal Vol. 16, No. 2, June 2004 pp. 75-83 Numerical simulations of fluid mechanical interactions between two abdominal aortic branches Taedong Kim, Taewon Seo* 1,2 and Abdul.I.

More information

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery Proceedings of the 3 rd International Conference on Fluid Flow, Heat and Mass Transfer (FFHMT 16) Ottawa, Canada May 2 3, 2016 Paper No. 170 Numerical Simulation of Blood Flow through Asymmetric and Symmetric

More information

Developing Pulsatile Flow in a Deployed Coronary Stent

Developing Pulsatile Flow in a Deployed Coronary Stent Divakar Rajamohan Department of Mechanical Engineering, University of Cincinnati, Cincinnati, OH 45221 Rupak K. Banerjee 1 Department of Mechanical Engineering, and Department of Biomedical Engineering,

More information

Developed pulsatile flow in a deployed coronary stent

Developed pulsatile flow in a deployed coronary stent Biorheology 44 (2007) 91 102 91 IOS Press Developed pulsatile flow in a deployed coronary stent Rupak K. Banerjee a,b,, Surendra B. Devarakonda a, Divakar Rajamohan a and Lloyd H. Back c a Mechanical Engineering

More information

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries Copyright 2015 Tech Science Press MCB, vol.12, no.1, pp.37-47, 2015 CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries P. Jhunjhunwala,, P.M. Padole, and S.B. Thombre, Abstract:

More information

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY LABORATORY OF BIOLOGICAL STRUCTURE MECHANICS www.labsmech.polimi.it FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY Rossella Balossino, Francesca Gervaso, Francesco Migliavacca,

More information

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 Presented at the COMSOL Conference 2010 India ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 1 DEPARTMENT OF MECHANICAL ENGINEERING & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING

More information

Comparison of Stent Designs using Computational Fluid Dynamics

Comparison of Stent Designs using Computational Fluid Dynamics Dublin Institute of Technology ARROW@DIT Conference Papers School of Mechanical and Design Engineering 2007-03-28 Comparison of Stent Designs using Computational Fluid Dynamics Jonathan Murphy Dublin Institute

More information

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis Biomedical & Pharmacology Journal Vol. 8(1), 123-131 (2015) Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis M. JAHANGIRI 1 *, M.

More information

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System Contents 1 Computational Haemodynamics An Introduction... 1 1.1 What is Computational Haemodynamics (CHD)... 1 1.2 Advantages of CHD... 3 1.3 Applications in the Cardiovascular System... 4 1.3.1 CHD as

More information

Arteriovenous Graft Modeling and Hemodynamic Interpretation

Arteriovenous Graft Modeling and Hemodynamic Interpretation Open Journal of Fluid Dynamics, 2012, 2, 324-330 http://dx.doi.org/10.4236/ojfd.2012.24a040 Published Online December 2012 (http://www.scirp.org/journal/ojfd) Arteriovenous Graft Modeling and Hemodynamic

More information

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Excerpt from the Proceedings of the COMSOL Conference 2010 Paris (COMSOL Conference) Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Vaidehi

More information

The influence of strut-connectors in stented vessels: A comparison of pulsatile flow through

The influence of strut-connectors in stented vessels: A comparison of pulsatile flow through Manuscript Click here to download Manuscript: paper_pant_et_al_revised_.ps 1 1 1 1 1 1 1 1 0 1 0 1 0 1 The influence of strut-connectors in stented vessels: A comparison of pulsatile flow through five

More information

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture RAMON BERGUER, a,b JOSEPH L. BULL, a,b AND KHALIL KHANAFER a a Vascular Mechanics Laboratory, Department of Biomedical Engineering,

More information

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION Jayme Pinto Ortiz University of São Paulo - Avenida Prof. Luciano Gualberto, travessa3 nº 380 - CEP - 05508-900

More information

Blood Flow Simulation toward Actual Application at Hospital

Blood Flow Simulation toward Actual Application at Hospital THE 5 TH ASIAN COMPUTAITIONAL FLUID DYNAMICS BUSAN, KOREA, OCTOBER 27 ~ OCTOBER 30, 2003 Blood Flow Simulation toward Actual Application at Hospital Abstract R. Himeno 1 1. Advanced Center for Computing

More information

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement Interactive CardioVascular and Thoracic Surgery 19 (2014) 16 20 doi:10.1093/icvts/ivu034 Advance Access publication 22 April 2014 ORIGINAL ARTICLE ADULTCARDIAC A computational fluid dynamics simulation

More information

FOR many decades, cardiovascular disease has been one of

FOR many decades, cardiovascular disease has been one of Vol:1, No:2, 27 Effect of Non-Newtonian Behaviour of Blood on Pulsatile Flows in Stenotic Arteries Somkid Amornsamankul, Benchawan Wiwatanapataphee, Yong Hong Wu, Yongwimon Lenbury International Science

More information

Analysis of the effects of plaque deposits on the blood flow through human artery

Analysis of the effects of plaque deposits on the blood flow through human artery ISSN 2395-1621 Analysis of the effects of plaque deposits on the blood flow through human artery #1 Sajid S. Mulani, #2 Pankaj I. Jagad 1 sajidsmulani21@gmail.com 2 pjjagad.scoe@sinhgad.edu #12 Department

More information

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics 54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics Andžela Šešok 1, Donatas Lukšys 2 Vilnius Gediminas Technical University,

More information

Blood flow in vessels with artificial or pathological geometrical changes

Blood flow in vessels with artificial or pathological geometrical changes Blood flow in vessels with artificial or pathological geometrical changes P. Tibaut 1, B. Wiesler 1, M. Mayer 2 & R. Wegenkittel 3 1 AVL LIST GmbH, Graz, Austria 2 VRVIs, Vienna, Austria 3 Tiani Medgraph

More information

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM H. GIRIJA BAI 1 and K.B. NAIDU 2 Department of Mathematics, Sathyabama University, Chennai-600 119, Tamil Nadu, India 1 girijanameprakash@gmail.com

More information

Luminal Flow Amplifies Stent-based Drug Deposition in Arterial Bifurcations

Luminal Flow Amplifies Stent-based Drug Deposition in Arterial Bifurcations Luminal Flow Amplifies Stent-based Drug Deposition in Arterial Bifurcations The Harvard community has made this article openly available. Please share how this access benefits you. Your story matters Citation

More information

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt Washington University in St. Louis Washington University Open Scholarship Mechanical Engineering and Materials Science Independent Study Mechanical Engineering & Materials Science 12-23-2017 Computational

More information

Simulations of the blood flow in the arterio-venous fistula for haemodialysis

Simulations of the blood flow in the arterio-venous fistula for haemodialysis Acta of Bioengineering and Biomechanics Vol. 16, No. 1, 2014 Original paper DOI: 10.5277/abb140109 Simulations of the blood flow in the arterio-venous fistula for haemodialysis DANIEL JODKO*, DAMIAN OBIDOWSKI,

More information

The relative effects of arterial curvature and lumen diameter on wall shear stress distributions in human right coronary arteries

The relative effects of arterial curvature and lumen diameter on wall shear stress distributions in human right coronary arteries The relative effects of arterial curvature and lumen diameter on wall shear stress distributions in human right coronary arteries Author Johnston, Barbara, Johnston, Peter Published 2007 Journal Title

More information

restenosis; neointimal hyperplasia; scaffolding; prolapse; three-dimensional

restenosis; neointimal hyperplasia; scaffolding; prolapse; three-dimensional J Appl Physiol 98: 947 957, 005. First published November 5, 004; doi:10.115/japplphysiol.0087.004. Circumferential vascular deformation after stent implantation alters wall shear stress evaluated with

More information

Flow Analysis with Stent Placement in the Cerebral Aneurysm

Flow Analysis with Stent Placement in the Cerebral Aneurysm Proceedings of the ECCOMAS Thematic International Conference on Simulation and Modeling of Biological Flows (SIMBIO 2011) September 21-23, 2011, VUB, Brussels, Belgium Flow Analysis with Stent Placement

More information

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation EasyChair Preprint 259 Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation Tianyi Xia, Matthew Doyle, Thomas Forbes and Cristina H. Amon EasyChair preprints

More information

Treatment of lesions at bifurcations: General perspective

Treatment of lesions at bifurcations: General perspective ESC Congress 2011 27-31 August, Paris, France Treatment of lesions at bifurcations: General perspective Goran Stankovic MD, PhD, FESC Division of Cardiology Clinical Center of Serbia, Belgrade Serbia Potential

More information

1992 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 59, NO. 7, JULY 2012

1992 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 59, NO. 7, JULY 2012 1992 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 59, NO. 7, JULY 2012 Identification of Hemodynamically Optimal Coronary Stent Designs Based on Vessel Caliber Timothy J. Gundert, Alison L. Marsden,

More information

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm Andrzej Polańczyk, MSc Ireneusz Zbiciński, PhD, DSc Abstract The aim of this study was to estimate whether

More information

NUMERICAL SIMULATION OF EFFECTS OF REYNOLDS NUMBER ON NON-NEWTONIAN BLOOD FLOW WITH SPIRAL COMPONENT THROUGH A REGULAR STENOSED ARTERY

NUMERICAL SIMULATION OF EFFECTS OF REYNOLDS NUMBER ON NON-NEWTONIAN BLOOD FLOW WITH SPIRAL COMPONENT THROUGH A REGULAR STENOSED ARTERY Proceedings of the International Conference on Mechanical Engineering and Renewable Energy 2017 (ICMERE2017) 18 20 December, 2017, Chittagong, Bangladesh ICMERE2017-PI-325 NUMERICAL SIMULATION OF EFFECTS

More information

Experimental Flow Studies on Carotid Artery Models with and without Stents

Experimental Flow Studies on Carotid Artery Models with and without Stents Experimental Flow Studies on Carotid Artery Models with and without Stents Liepsch, D 1, Schmid, Th. 1, Klügel, G. 1, Sakurai, A. 3, Berger, H. 2, Greil, O. 2 1 FB 05, Fachhochschule München, Lothstr.

More information

Post-conditioning. P a g e 1. To my Thesis Committee,

Post-conditioning. P a g e 1. To my Thesis Committee, P a g e 1 To my Thesis Committee, This document seeks to clarify my research project. After describing what post-conditioning (PC) is, I will explain differences between my research and the recent peristaltic

More information

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties Lin Chen, Shen Jingfeng & Chen Bing School of mechanical engineering, University of Shanghai for Science & Technology,

More information

Identification of Atherosclerotic Lesion-Prone Sites through Patient-Specific Simulation of Low-Density Lipoprotein Accumulation

Identification of Atherosclerotic Lesion-Prone Sites through Patient-Specific Simulation of Low-Density Lipoprotein Accumulation Identification of Atherosclerotic Lesion-Prone Sites through Patient-Specific Simulation of Low-Density Lipoprotein Accumulation Ufuk Olgac 1, Vartan Kurtcuoglu 1, Stefan C. Saur 2, and Dimos Poulikakos

More information

Three-Dimensional Computational Fluid Dynamics Modeling of Alterations in Coronary Wall Shear Stress Produced by Stent Implantation

Three-Dimensional Computational Fluid Dynamics Modeling of Alterations in Coronary Wall Shear Stress Produced by Stent Implantation Annals of Biomedical Engineering, Vol. 31, pp. 972 980, 2003 Printed in the USA. All rights reserved. 0090-6964/2003/31 8 /972/9/$20.00 Copyright 2003 Biomedical Engineering Society Three-Dimensional Computational

More information

Medical device design using Computational Fluid Dynamics (CFD)

Medical device design using Computational Fluid Dynamics (CFD) Medical device design using Computational Fluid Dynamics (CFD) Session: Winter 2016 IMPORTANT NOTE: This project has 8 deliverables, for each one timely work is expected. 1. General Design Specifications

More information

Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass

Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass BioImpacts, 2011, 1(2), 99-104 http://bi.tbzmed.ac.ir/ Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass Seyed Esmail Razavi *, Ramin Zanbouri, Omid Arjmandi-Tash

More information

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery 1 Sajid S. Mulani, 2 P. I. Jagad 1,2 Department of Mechanical Engineering, SCoE, Pune 411041, India Email: 1 sajidsmulani21@gmail.com,

More information

Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics

Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics Sinjae Hyun, PhD, a Clement Kleinstreuer, PhD, b and Joseph P. Archie, Jr, PhD,

More information

MULTIVESSEL PCI. IN DRUG-ELUTING STENT RESTENOSIS DUE TO STENT FRACTURE, TREATED WITH REPEAT DES IMPLANTATION

MULTIVESSEL PCI. IN DRUG-ELUTING STENT RESTENOSIS DUE TO STENT FRACTURE, TREATED WITH REPEAT DES IMPLANTATION MULTIVESSEL PCI. IN DRUG-ELUTING STENT RESTENOSIS DUE TO STENT FRACTURE, TREATED WITH REPEAT DES IMPLANTATION C. Graidis, D. Dimitriadis, A. Ntatsios, V. Karasavvides Euromedica Kyanous Stavros, Thessaloniki.

More information

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique International Journal of Engineering Research and Technology. ISSN 0974-3154 Volume 6, Number 3 (2013), pp. 399-406 International Research Publication House http://www.irphouse.com Study of Newtonian and

More information

Computational design of Intracranial Stent using 3D visualization system

Computational design of Intracranial Stent using 3D visualization system Computational design of Intracranial Stent using 3D visualization system Institute of Fluid Science, Tohoku University Makoto OHTA Graduate school of Engineering Hitomi Anzai Graduate school of Biomedical

More information

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents Augusto F. Sanches and Eva Gutheil Interdisciplinary Center for Scientific Computing, Heidelberg University Flow diverters

More information

Flow in re-stenosed artery after angioplasty

Flow in re-stenosed artery after angioplasty Data Management and Security 209 Flow in re-stenosed artery after angioplasty S. I. Bernad 1, A. Totorean 2, E. S. Bernad 3 & R. Susan-Resiga 2 1 Romanian Academy, Timisoara Branch, Romania 2 Politehnica

More information

Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent

Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent Kuang-Huei Lee, Hao-Ming Hsiao, Ying-Chih Liao, Yi-Hsiang Chiu, Yu-Seng Tee Dept. of

More information

Investigation of the Hemodynamics of Coronary Arteries - Effect of Stenting. Naresh Kumar Coimbatore Selvarasu

Investigation of the Hemodynamics of Coronary Arteries - Effect of Stenting. Naresh Kumar Coimbatore Selvarasu Investigation of the Hemodynamics of Coronary Arteries - Effect of Stenting Naresh Kumar Coimbatore Selvarasu Dissertation submitted to the faculty of the Virginia Polytechnic Institute and State University

More information

Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries

Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries Author Johnston, Barbara, Johnston, Peter Published 2009 Conference Title 18th IMACS World Congress MODSIM09

More information

A numerical study on the effect of hematocrit on hemodynamic characteristics in arteriovenous graft

A numerical study on the effect of hematocrit on hemodynamic characteristics in arteriovenous graft Korea-Australia Rheology Journal, Vol.26, No.3, pp.327-334 (August 2014) DOI: 10.1007/s13367-014-0037-x www.springer.com/13367 A numerical study on the effect of hematocrit on hemodynamic characteristics

More information

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases Proceedings of the 215 2nd International Symposium on Physics and Technology of Sensors, 8-1th March, 215, Pune, India Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

More information

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft BURASKORN NUNTADILOK 1, BENCHAWAN WIWATANAPATAPHEE 1 MEECHOKE CHUEDOUNG 1, THANONGCHAI SIRIAPISITH

More information

Evaluation of Stent Performances using FEA considering a Realistic Balloon Expansion

Evaluation of Stent Performances using FEA considering a Realistic Balloon Expansion Evaluation of Stent Performances using FEA considering a Realistic Balloon Expansion Won-Pil Park, Seung-Kwan Cho, Jai-Young Ko, Anders Kristensson, S.T.S. Al-Hassani, Han-Sung Kim, and Dohyung Lim Abstract

More information

The study on hemodynamic effect of series type LVAD on aortic blood flow pattern: a primary numerical study

The study on hemodynamic effect of series type LVAD on aortic blood flow pattern: a primary numerical study DOI 10.1186/s12938-016-0252-4 BioMedical Engineering OnLine RESEARCH Open Access The study on hemodynamic effect of series type LVAD on aortic blood flow pattern: a primary numerical study Qi Zhang, Bin

More information

4D model of hemodynamics in the abdominal aorta

4D model of hemodynamics in the abdominal aorta Bio-Medical Materials and Engineering 26 (2015) S257 S264 DOI 10.3233/BME-151312 IOS Press S257 4D model of hemodynamics in the abdominal aorta Ireneusz Zbicinski a,*, Natalia Veshkina a and Ludomir Stefa

More information

Mathematical and Computational study of blood flow through diseased artery

Mathematical and Computational study of blood flow through diseased artery Mathematical and Computational study of blood flow through diseased artery Abstract This paper presents the study of blood flow through a tapered stenosed artery. The fluid (blood) medium is assumed to

More information

FFR Fundamentals and Measurements

FFR Fundamentals and Measurements FFR Fundamentals and Measurements Ghassan S. Kassab Thomas Linnemeier Chair Professor Biomedical Engineering, Indiana University Purdue University Indianapolis Principle of FFR Q S ( P P ) / R P max d

More information

Numerical analysis of blood flow in human abdominal aorta

Numerical analysis of blood flow in human abdominal aorta Advances in Fluid Mechanics VI 603 Numerical analysis of blood flow in human abdominal aorta M. Podyma 1, I. Zbicinski 1, J. Walecki 2, M. L. Nowicki 2, P. Andziak 3, P. Makowski 4 & L. Stefanczyk 5 1

More information

FSI Analysis of Diseased Coronary using Patient Specific Data

FSI Analysis of Diseased Coronary using Patient Specific Data FSI Analysis of Diseased Coronary using Patient Specific Data Mingchao Cai, * Haofei Liu, a Chun Yang, a,c Jie Zheng, b Richard Bach, b Mehmet H. Kural, d Kristen L. Billiar, d David Muccigrosso, b Dongsi

More information

Estimation and Comparison of T Graft Versus Conventional Graft for Coronary Arteries

Estimation and Comparison of T Graft Versus Conventional Graft for Coronary Arteries World Applied Sciences Journal 27 (10): 1336-1344, 2013 ISSN 1818-4952 IDOSI Publications, 2013 DOI: 10.5829/idosi.wasj.2013.27.10.2908 Estimation and Comparison of T Graft Versus Conventional Graft for

More information

Citation. As Published Publisher. Version

Citation. As Published Publisher. Version Modelling intravascular delivery from drug-eluting stents with biodurable coating: investigation of anisotropic vascular drug diffusivity and arterial drug distribution The MIT Faculty has made this article

More information

Investigation of metallic surface area of coronary stents

Investigation of metallic surface area of coronary stents Investigation of metallic surface area of coronary stents Dóra KÁROLY 1, Miksa KOVÁCS 1, Andrew Attila TERDIK 1, Eszter BOGNÁR 1,2 1 Department of Materials Science and Engineering, Faculty of Mechanical

More information

Hemodynamics in Idealized Stented Coronary Arteries: Important Stent Design Considerations

Hemodynamics in Idealized Stented Coronary Arteries: Important Stent Design Considerations Annals of Biomedical Engineering, Vol. 44, No. 2, February 2016 ( 2015) pp. 315 329 DOI: 10.1007/s10439-015-1387-3 Medical Stents: State of the Art and Future Directions Hemodynamics in Idealized Stented

More information

Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes

Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes Titus PETRILA 1, Balazs ALBERT,2 Corresponding author 1 Vasile Goldis Western University,

More information

ISSUES ON COMPUTATIONAL MODELING FOR COMPUTATION-AIDED DIAGNOSIS 臨床診断支援ツールのための計算力学モデリング

ISSUES ON COMPUTATIONAL MODELING FOR COMPUTATION-AIDED DIAGNOSIS 臨床診断支援ツールのための計算力学モデリング ISSUES ON COMPUTATIONAL MODELING FOR COMPUTATION-AIDED DIAGNOSIS 臨床診断支援ツールのための計算力学モデリング Hao LIU Advanced Computer and Information Division, RIKEN 2-1, Hirosawa, Wako-shi, Saitama 351-0198 JAPAN e-mail:

More information

NIH Public Access Author Manuscript J Control Release. Author manuscript; available in PMC 2014 June 10.

NIH Public Access Author Manuscript J Control Release. Author manuscript; available in PMC 2014 June 10. NIH Public Access Author Manuscript Published in final edited form as: J Control Release. 2013 June 10; 168(2): 115 124. doi:10.1016/j.jconrel.2013.03.014. Impact of flow pulsatility on arterial drug distribution

More information

An effort is made to analyse the stresses experienced by the human femur. In order

An effort is made to analyse the stresses experienced by the human femur. In order Finite Element Analysis of Human Femur Bone U N Mughal 1, H A Khawaja 2*, M Moatamedi 1, M Souli 3 1. Narvik University College, Norway 2. UiT-The Arctic University of Norway, Norway 3. University of Lille,

More information

Left main coronary artery (LMCA): The proximal segment

Left main coronary artery (LMCA): The proximal segment Anatomy and Pathology of Left main coronary artery G Nakazawa Tokai Univ. Kanagawa, Japan 1 Anatomy Difinition Left main coronary artery (LMCA): The proximal segment RCA AV LAD LM LCX of the left coronary

More information

Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation

Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation American Journal of Applied Sciences 7 (9): 1237-1242, 2010 ISSN 1546-9239 2010 Science Publications Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation 1 Vahab Dehlaghi and

More information

There has been extensive speculation in the literature about

There has been extensive speculation in the literature about TECHNICAL NOTE A. Mantha C. Karmonik G. Benndorf C. Strother R. Metcalfe Hemodynamics in a Cerebral Artery before and after the Formation of an Aneurysm SUMMARY: Using data obtained from 3D digital subtraction

More information

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images Applied Mathematics Volume, Article ID 77, pages http://dx.doi.org/.//77 Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images Mahsa Dabagh,, Wakako

More information

Catch-up Phenomenon: Insights from Pathology

Catch-up Phenomenon: Insights from Pathology Catch-up Phenomenon: Insights from Pathology Michael Joner, MD CVPath Institute Inc. Gaithersburg, MD USA Path Lessons learned from the BMS and DES (1 st Gen) era Neointimal Thickness [mm] In Stent Re

More information

BioMedical Engineering OnLine

BioMedical Engineering OnLine BioMedical Engineering OnLine BioMed Central Research Evaluation of the effect of stent strut profile on shear stress distribution using statistical moments Juan Mejia 1,4, Bilal Ruzzeh 1, Rosaire Mongrain*

More information

BLOOD FLOW VISUALISATION THROUGH CAROTID BIFURCATION USING ANSYS CFD

BLOOD FLOW VISUALISATION THROUGH CAROTID BIFURCATION USING ANSYS CFD BLOOD FLOW VISUALISATION THROUGH CAROTID BIFURCATION USING ANSYS CFD Roopa.V.Chanashetty 1, Dr.Channappa Bhyri 2 and Vijaykumar Chanashetty 3 1 Department of Electronics and Communication Engineering,

More information

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents.

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents. Blucher Mechanical Engineering Proceedings May 2014, vol. 1, num. 1 www.proceedings.blucher.com.br/evento/10wccm AN ANALYSIS OF THE CONTACT BETWEEN THE STENT AND THE ARTERY USING TUBE HIDROFORMING SIMULATION

More information

Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery

Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery Korea-Australia Rheology Journal Vol. 21, No. 2, June 2009 pp. 119-126 Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery Kun Hyuk Sung, Kyoung

More information

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta No Author Given No Institute Given Abstract. Abdominal Aortic Aneurysm is a potentially life-threatening disease if not treated adequately.

More information

Dynamic simulation of stent deployment effects of design, material and coating

Dynamic simulation of stent deployment effects of design, material and coating Journal of Physics: Conference Series OPEN ACCESS Dynamic simulation of stent deployment effects of design, material and coating To cite this article: A Schiavone et al 2013 J. Phys.: Conf. Ser. 451 012032

More information

Noninvasive Fractional Flow Reserve from Coronary CT Angiography

Noninvasive Fractional Flow Reserve from Coronary CT Angiography 2016 KSC Annual Spring Scientific Conference Noninvasive Fractional Flow Reserve from Coronary CT Angiography Bon-Kwon Koo, MD, PhD, Seoul, Korea Why the hemodynamics for coronary artery disease? Twinlifemarketing.com.au

More information

A Two-layered Model for the Analysis of Arterial Rheology

A Two-layered Model for the Analysis of Arterial Rheology IJCSIT International Journal of Computer Science and Information Technology, Vol., o., June, pp. 37- A Two-layered Model for the Analysis of Arterial Rheology Sapna Singh Department of Mathematics, Harcourt

More information

Modeling of the Impact of Blood Vessel Flow on the Temperature Distribution during Focused Ultrasound Treatments

Modeling of the Impact of Blood Vessel Flow on the Temperature Distribution during Focused Ultrasound Treatments Presented at the COMSOL Conference 2010 Boston Modeling of the Impact of Blood Vessel Flow on the Temperature Distribution during Focused Ultrasound Treatments Elisabetta Sassaroli, King C. P. Li, Brian

More information

Hematocrit Level on Blood flow through a Stenosed Artery with Permeable Wall: A Theoretical Study

Hematocrit Level on Blood flow through a Stenosed Artery with Permeable Wall: A Theoretical Study Available at http://pvamu.edu/aam Appl. Appl. Math. ISSN: 1932-9466 Vol. 12, Issue 1 (June 2017), pp. 291-304 Applications and Applied Mathematics: An International Journal (AAM) Hematocrit Level on Blood

More information

Non-Newtonian blood flow in human right coronary arteries: Transient simulations

Non-Newtonian blood flow in human right coronary arteries: Transient simulations Non-Newtonian blood flow in human right coronary arteries: Transient simulations Author Johnston, Barbara, Johnston, Peter, Corney, Stuart, Kilpatrick, David Published 2006 Journal Title Journal of Biomechanics

More information

Extraction and FSI modeling of Left coronary artery structure from patient s CTA images

Extraction and FSI modeling of Left coronary artery structure from patient s CTA images Extraction and FSI modeling of Left coronary artery structure from patient s CTA images Safia Salim Dept. of Electronics & Communication College of Engineering, Kerala University Trivandrum. 695016 Email:

More information

STUDY OF FLOW PHENOMENA IN AORTIC DISSECTION

STUDY OF FLOW PHENOMENA IN AORTIC DISSECTION STUDY OF FLOW PHENOMENA IN AORTIC DISSECTION Maurizio Bordone *, Eugenio Oñate *, Paula Rudenick, Bart Bijnens and Eduardo Soudah * * International Centre for Numerical Methods in Engineering (CIMNE) Campus

More information

Coronary Artery Bifurcation Biomechanics and Implications for Interventional Strategies

Coronary Artery Bifurcation Biomechanics and Implications for Interventional Strategies Catheterization and Cardiovascular Interventions 76:836 843 (2010) Basic Science Coronary Artery Bifurcation Biomechanics and Implications for Interventional Strategies James E. Moore, Jr., 1 * PhD, Lucas

More information

Three Dimensional Large Eddy Simulation of Blood Flow and Deformation in an Elastic Constricted Artery

Three Dimensional Large Eddy Simulation of Blood Flow and Deformation in an Elastic Constricted Artery Three Dimensional Large Eddy Simulation of Blood Flow and Deformation in an Elastic Constricted Artery Xi Gu, Guan Heng Yeoh, Victoria Timchenko Abstract In the current work, a three-dimensional geometry

More information

Is a Flat Inlet Profile Sufficient for WSS Estimation in the Aortic Arch?

Is a Flat Inlet Profile Sufficient for WSS Estimation in the Aortic Arch? Is a Flat Inlet Profile Sufficient for WSS Estimation in the Aortic Arch? J. Renner 1, D. Loyd 1, T. Länne 2, M. Karlsson 1 1 Department of Management and Engineering 2 Department of Medical and Health

More information

Visualization and Quantification of Blood Flow in the Human Aorta. From in vivo 4D Phase Contrast MRI to Subject-Specific Computational Hemodynamics

Visualization and Quantification of Blood Flow in the Human Aorta. From in vivo 4D Phase Contrast MRI to Subject-Specific Computational Hemodynamics Visualization and Quantification of Blood Flow in the Human Aorta. From in vivo 4D Phase Contrast MRI to Subject-Specific Computational Hemodynamics Industrial Bioengineering Group Department of Mechanics,

More information

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction Eann A. Patterson Department of Mechanical Engineering, The University of Sheffield Mappin Street, Sheffield, S1 3JD

More information

Numerical analysis of urine flow through the side holes of a double J stent in a ureteral stenosis

Numerical analysis of urine flow through the side holes of a double J stent in a ureteral stenosis Technology and Health Care 25 (2017) S63 S72 DOI 10.3233/THC-171307 IOS Press S63 Numerical analysis of urine flow through the side holes of a double J stent in a ureteral stenosis Hyoung-Ho Kim a, Young

More information

Prediction of stent endflare, arterial stresses and flow patterns in a stenotic artery

Prediction of stent endflare, arterial stresses and flow patterns in a stenotic artery Prediction of stent endflare, arterial stresses and flow patterns in a stenotic artery M. R. Hyre 1, R. M. Pulliam 2 & J. C. Squire 1 1 Department of Mechanical Engineering, Virginia Military Institute,

More information

Computational Analysis on Commercially Available Stent Designs

Computational Analysis on Commercially Available Stent Designs Computational Analysis on Commercially Available Stent Designs Abhijit Chanda 1, Shuvrangsu Das 2, Sounak Bhattacharjee 2, Pranab Ghosh 2, K. Basu 3 1 School of Bio Science and Engineering, Jadavpur University,

More information

Marquette University Sung Kwon Marquette University Recommended Citation

Marquette University Sung Kwon Marquette University Recommended Citation Marquette University e-publications@marquette Master's Theses (2009 -) Dissertations, Theses, and Professional Projects Quantification of Local Hemodynamic Alterations Caused by Virtual Implantation of

More information

Changes in Carotid Artery Flow Velocities After Stent Implantation: A Fluid Dynamics Study With Laser Doppler Anemometry

Changes in Carotid Artery Flow Velocities After Stent Implantation: A Fluid Dynamics Study With Laser Doppler Anemometry J ENDOVASC THER 275 EXPERIMENTAL INVESTIGATION Changes in Carotid Artery Flow Velocities After Stent Implantation: A Fluid Dynamics Study With Laser Doppler Anemometry Oliver Greil, MD; Gottlieb Pflugbeil,

More information

FFR CT : Beyond FFR. Bon-Kwon Koo, MD, PhD. Seoul National University Hospital, Seoul, Korea. Seoul National University Hospital Cardiovascular Center

FFR CT : Beyond FFR. Bon-Kwon Koo, MD, PhD. Seoul National University Hospital, Seoul, Korea. Seoul National University Hospital Cardiovascular Center FFR CT : Beyond FFR Bon-Kwon Koo, MD, PhD, Seoul, Korea Patient-specific non-invasive coronary hemodynamic assessment Non-invasive, Pt-specific Hemodynamics Pressure Pressure difference Pressure gradient

More information