1 Biorheology 44 (2007) IOS Press Developed pulsatile flow in a deployed coronary stent Rupak K. Banerjee a,b,, Surendra B. Devarakonda a, Divakar Rajamohan a and Lloyd H. Back c a Mechanical Engineering Department, University of Cincinnati, Cincinnati, OH 45221, USA b Biomedical Engineering Department, University of Cincinnati, Cincinnati, OH 45221, USA c Jet Propulsion Laboratory, California Institute of Technology, Pasadena, CA 91109, USA Received 19 December 2005 Accepted in revised form 1 February 2007 Abstract. The patho-physiologic process of restenosis and tissue growth may not be completely eliminated and is the primary concern of clinicians performing angioplasty and stent implantation procedures. Recent evidence suggests that the restenosis process is influenced by several factors: (1) geometry and size of vessel; (2) stent design; and (3) it s location that alter hemodynamic parameters, including local wall shear stress (WSS) distributions. The present three-dimensional (3D) analysis of pulsatile flow in a deployed coronary stent: (1) shows complex 3D variation of hemodynamic parameters; and (2) quantifies the changes in local WSS distributions for developed flow and compares with recently published WSS data for developing flow. Higher order of magnitude of WSS of 290 dyn/cm 2 is observed on the surface of cross-link intersections at the entrance of the stent for developed flow, which is about half of that for developing flow. Low WSS of 0.8 dyn/cm 2 and negative WSS of 8 dyn/cm 2 are seen at the immediate upstream and downstream regions of strut intersections. Persistent recirculation is observed at the downstream region of each strut cross-link and the regions of low and negative WSS may lead to patho-physiologic conditions near the stented region. The key finding of this study is that the location of stent in the coronary artery determines the developing or developed nature of the flow, which in turn, results in varied level of WSS. Keywords: Coronary artery, stent, pulsatile flow, stenosis, computational fluid dynamics, wall shear stress. 1. Introduction A major limiting factor to the effectiveness of the stent implantation procedure has been the phenomenon of restenosis, or occurrence of re-narrowing in the treated initially narrowed artery segment. From the context of hemodynamic perspective, our previous study (Rajamohan et al. [21]) has evaluated the impact of developing pulsatile flow in a deployed coronary stent on local wall shear stress (WSS) distributions in addition to other hemodynamic parameters. The present study examines the developed pulsatile flow in a deployed coronary stent. The authors believe that the literature would remain incomplete without the comparison of results between the developed and the developing [21] flow. * Address for correspondence: R.K. Banerjee, Dept. of Mechanical, Industrial and Nuclear Engineering, 688 Rhodes Hall, PO Box , Cincinnati, OH , USA. Tel.: ; Fax: ; Rupak.Banerjee@UC. Edu X/07/$ IOS Press and the authors. All rights reserved

2 92 R.K. Banerjee et al. / Developed pulsatile flow in a deployed coronary stent There are several factors affecting restenosis like geometry and size of vessel, stent design and its location. These factors impact areas of flow recirculation, flow separation and WSS. Studies have shown that there are critical interactions between stents and vessel walls [15,16,23,26] and stent design has played a vital role for tissue damage in neointimal hyperplasia. Researchers [5,12,18] have also shown the dependence of flow recirculation and flow reattachment on strut spacing. They observed minimum platelet deposition and continual flow stagnation at low strut spacing, maximum platelet deposition and partial flow reattachment at intermediate strut spacing, and high platelet deposition and complete flow reattachment at higher strut spacing. The study by Henry [14] concluded that the occurrence of flow separation near the strut led to large WSS gradients in the immediate vicinity of the flow separation and reattachment points, and this might have an impact on the rate of endothelial repair following injury induced by stent placement. Several other researchers have studied and indicated the effect of wall shear stress on restenosis [4,10,17,22,24,25,27]. These studies have been discussed in detail in our previous publication [21]. Due to the space limitation, the authors have refrained from repeating the literature review in the present manuscript. The primary focus of the present study is to perform a detailed calculation on WSS distribution, flow recirculation and flow stagnation for developed flow and compare that with our previous study [21] for developing flow, where the stent is deployed at the origin of a branched coronary artery. This study improves upon previously published literatures as a 3D comparative analysis has been conducted to quantify the stent induced changes on the flow patterns for developed pulsatile flow in relation to developing flow, having non-newtonian viscosity of blood. 2. Methodology 2.1. Geometry A pulsatile flow analysis is performed for a freshly deployed coronary stent with helical strut matrix, in a human coronary artery having a 3D geometry with an axial length of 10 mm and diameter of 3 mm. The design and the geometry of the stent strut used in the present study (Fig. 1A) are based on the model used in the past studies [5,18,21]. The meshing scheme, governing equations, definition of modified oscillatory shear index (MOSI) and the finite volume method are discussed in our previous study [21] Boundary conditions To simulate the best case scenario of a developed flow in stented artery, a time varying (pulsatile) parabolic spatial variation of velocity boundary condition is imposed at the inlet [21]. Thus, the present model represents a best case scenario and evaluates the condition of the stent being deployed at a downstream region of a coronary artery, where a developed flow is expected. For developing flow [21] the stent was deployed at the origin of a branched coronary artery. To model this worst case flow scenario at the entrance region of the artery, a time-varying pulse with a uniform spatial variation of velocity boundary condition was imposed at the inlet with a user-defined subroutine. No-slip boundary condition is specified on the wall which makes all flow velocity components equal to zero. Considering the diseased condition of the coronary artery, the wall is specified to be rigid [20]. At the outlet, zero gauge pressure boundary condition is applied.

3 R.K. Banerjee et al. / Developed pulsatile flow in a deployed coronary stent 93 Fig. 1. (A) Geometry and stent struts; (B) validation of wall shear stress Validation WSS data of the present study (coronary stent: d = 0.3 cm,u peak = 22.8 cm/s,re peak = 208) of developed flow as well as developing [21] flow are compared with the past study by LaDisa et al. [17] (coronary stent: d = cm, u avg = 22.5 cm/s,re avg = 177). Figure 1B shows the comparison of WSS at the four strut intersections of the stent. For developed flow, WSS near the entrance region of stented artery i.e. at strut intersection 1 is 53 dyn/cm 2 ; this value is within 0.7% of the result reported by LaDisa et al. [17]. As the flow proceeds from strut 1 to strut 2, WSS decreases to 48.3 dyn/cm 2.This value further decreases to 46 dyn/cm 2 as the flow reaches the strut intersection 4. This trend of reduction in WSS has also been shown by LaDisa et al. [17]. The parabolic velocity profile at inlet is disturbed when the flow encounters the struts. The disturbance is higher at the location of strut 1 resulting in higher WSS compared to downstream strut locations 2, 3 and 4. The flow is relatively more developed at downstream strut locations. The variation in the WSS values between the present study of developed flow and LaDisa et al. [17] is around 7% at strut intersections 2 to 4. This variation is primarily due to some differences in strut spacing and height between the present study and the geometry studied by LaDisa et al. [17]. WSS value at the entrance for the developed flow is about half of the value that was obtained for developing flow [21]. As the flow develops, the WSS value decreases rapidly and

4 94 R.K. Banerjee et al. / Developed pulsatile flow in a deployed coronary stent the variation in the WSS values between developed and developing flow decreases from 20.6% at strut intersection 2 to 9.8% at strut intersection 3, and then to 7.5% at strut intersection 4. It may be noted that a comparative study of developed and developing pulsatile flows has not been reported in the past. The additional validation of pressure data of the present study is shown in Fig Results and discussion To study the effect of stent implantation on the blood flow, velocity profiles, WSS, recirculation length and height and pressure drop data are plotted at various locations inside the stent. Figure 1A shows a pictorial representation of the locations of strut intersections, midpoints and vertices. The results have been obtained for both basal and hyperemic flow rates, but due to space constraints only the data for hyperemic flow rates are shown in the figures. Results for basal flow have been discussed wherever applicable Velocity profiles To analyze the developed pulsatile flow, axial velocity profiles are plotted along the radius of the artery at strut intersections and midpoints of the stent. The velocity profiles are plotted for hyperemic flow rate at different time points: 1.86 s, the early acceleration phase (u in = 0.5u inmax ); 2.00 s, the peak flow (u in = u inmax ); 2.16 s, the early deceleration phase of the cycle (u in = 0.75u inmax ) and 2.30 s, the late deceleration phase of the cycle (u in = 0.25u inmax ). Due to space limitation velocity profiles near the entrance (Intersection 1 & Midpoint 1) and near the exit (Intersection 4 & Midpoint 3) of the stented artery are reported (Fig. 1A). Figures 2A and 2B show the axial velocity profiles at strut intersections 1 (near the entrance) and 4 (near the exit) for developed flow. For comparison, these figures are superimposed with a representative velocity profile for developing flow [21] at t = 2.0 s, the time at which the inlet flow attains its peak. The velocity profiles are plotted along the radius of the artery i.e. from the center of the artery to the vessel wall where cm represents the point of strut intersection. For developed flow, a near parabolic velocity profile is observed where the peak velocity during the entire cardiac cycle occurs at the center of the artery for all flow rates. Peak velocity at t = 2.0 s is 40.2 cm/s (not shown) for basal flow and cm/s for hyperemic flow near the entrance (strut intersection 1). Near the exit of the artery (strut intersection 4), the peak velocity is 42.1 cm/s (not shown) and cm/s for basal and hyperemic flows, respectively. There is no significant variation in the velocity from the entrance to the exit of the artery for developed flow. For developing flow [21], the peak velocity at the entrance and exit occurs near the strut intersections 1 and 4, respectively, i.e. near the stented artery wall. There is a decrease of 83% in the velocity gradient near the entrance and 37% near the exit of the stented artery for developed flow compared to developing flow. Figures 2C and 2D show the axial velocity profiles at midpoints 1 and 3 for developed flow. These figures are also superimposed with a representative velocity profile for developing flow [21] for comparison at t = 2.0 s. Similar to the velocity profiles obtained for developed flow at strut intersections, near parabolic velocity profiles are observed at midpoints close to the entrance and exit of the artery. For basal flow rate, the peak velocity remains constant at 40 cm/s whereas, for hyperemic flow rate, peak velocity is 157 cm/s and 161 cm/s near the entrance and exit of the artery, respectively. For developing flow [21], peak velocity occurs near the stent wall (Figs 2C and 2D) for all flows except at the exit region for basal flow(not plotted). There is a decrease of 55% in the velocity gradient near midpoint 1 and 34%

5 R.K. Banerjee et al. / Developed pulsatile flow in a deployed coronary stent 95 t = 1.85 s t = 2.00s t = 2.20 s t = 2.00 s; [21] t = 2.30 s Fig. 2. Variation of axial velocity along the radius of the artery at strut intersections 1 (A) and 4 (B) at times of accelerating (t = 1.85 s), peak (t = 2.00 s) and decelerating flows (t = 2.20 s and 2.30 s); Variation of axial velocity along the radius of artery at midpoints 1 (C) and 3 (D) at times of accelerating (t = 1.85 s), peak (t = 2.00 s) and decelerating flows (t = 2.20 s and 2.30 s). near midpoint 3 for developed flow compared to developing flow. It can be concluded that the developed flow does not have any significant variation in the velocity profile from the entrance to the exit of the artery and has lower velocity gradient at the wall, thereby leading to lower WSS value compared to developing flow Wall shear stress (WSS) distribution Figures 3A and 3B show the phasic variation of axial WSS for hyperemic flow rates at strut intersections 1 to 4 and midpoints 1 to 3, respectively for developed flow. For comparison, these figures are superimposed with a representative WSS plot for developing flow [21] at t = 2.0 s. For the pulse cycle,

6 96 R.K. Banerjee et al. / Developed pulsatile flow in a deployed coronary stent Fig. 3. Temporal variation of wall shear stress (WSS) at strut intersections [I] (A) and midpoints [M] (B); (C) Temporal variation of WSS at vertices. I upstream of strut intersections: V1, V3, V5, and V7; II downstream of strut intersection: V2 (values for V4, V6 and V8 are similar to V2); (D) Axial variation of WSS along arterial wall at times of accelerating, peak and decelerating flows. the WSS attains its maxima at peak diastole (t = 2.0 s). WSS values at strut intersections are generally higher than that at midpoints due to high velocity gradient at the strut intersections. Maximum peak WSS values of 252 dyn/cm 2 and 76 dyn/cm 2 are observed for hyperemic flow rates near the entrance at strut intersection 1 and midpoint 1, respectively. For basal flow, peak WSS values are between 20 to 25% of that of hyperemic flow. There is an average decrease of 30% in peak WSS values at strut intersections and 20% at midpoints near the entrance for developed flow compared to developing flow. As the flow develops along the stent from strut intersection 1 to strut intersection 4, the reduction in peak WSS value is 20% and 41% for developed and developing flows, respectively. Thus, the developed flow has less impact on changes in local WSS distributions than the developing flow. Figure 3C shows the phasic variation of axial WSS at the vertices 1 to 8. Vertices 1, 3, 5 and 7 represent the upstream and vertices 2, 4, 6 and 8 represent the downstream of strut intersections 1, 2, 3 and 4, respectively. These figures are also superimposed with a representative pulsatile WSS profile for developing flow [21] for comparison at t = 2.0 s. Overall positive WSS occurs at the upstream and

7 R.K. Banerjee et al. / Developed pulsatile flow in a deployed coronary stent 97 negative WSS occurs at the downstream of strut intersections during the entire cardiac cycle for all flow rates, except for the basal flow (not shown) rate where negative shear stress occurs even at the upstream location of strut intersection 1 for both developed and developing flows. For developed flow, peak values of WSS of 11 dyn/cm 2 and 5.5 dyn/cm 2, occur near the entrance at vertices 1 and 2, respectively for hyperemic flow. These results show that the peak positive WSS at the immediate upstream and the peak negative WSS at the immediate downstream of each strut intersection for developed flow are about 20% and 75% of that for developing flow, respectively. Figure 3D shows the axial variation of WSS along the artery wall for hyperemic flow rates at times of accelerating, peak and decelerating flows. For developed flow, there is an increase of 5 to 6 times in WSS at strut intersections for hyperemic flow (Fig. 3D) as compared to the basal flow (not shown) for which 52 dyn/cm 2 is calculated near the entrance at strut intersection 1 and 47 dyn/cm 2 is seen near the exit at strut intersection 4. For hyperemic flow, WSS of 290 dyn/cm 2 occurs just upstream of strut cross-link intersection 1 (x = cm). As the flow proceeds in the axial direction, there is a sharp decrease of WSS to a negative value of 8 dyn/cm 2 at the immediate downstream of strut intersection 1 (x = cm) showing recirculation. Due to the flow reattachment between strut intersections 1 and 2, the WSS increases from negative value gradually to 80 dyn/cm 2 and then decreases to a low value of 0.8 dyn/cm 2 at the immediate upstream of strut intersection 2 (x = 0.3 cm), where there is flow stagnation. Axially downstream the WSS again increases to 250 dyn/cm 2 at the strut intersection 2. The percentage reduction in WSS from strut intersection 1 to 4 is higher for hyperemic flow (24%) than for basal flow (9%) because at hyperemic flow the flow disturbance is higher at the upstream location of strut 1. The comparison of axial variation of WSS for the developed flow at peak inlet velocity (at t = 2.0 s) with developing flow data [21] is also presented in Fig. 3D for hyperemic flows. From Fig. 3D, it can be seen that for developed flow WSS values are half of that for developing flow near the entrance and nearly coincide at the exit Modified oscillatory shear index (MOSI) Figure 4 shows the variation of MOSI with flow rate. To analyze the flow, MOSI is plotted at upstream and downstream of each strut intersection (vertices 1 to 8) from basal to hyperemic flow rates. For developed flow, MOSI value of nearly +1 and 1 are observed at the upstream and downstream of each strut intersection, respectively for all flow rates. This shows that there is negative WSS at the Fig. 4. Variation of modified oscillatory shear index (MOSI) with mean flow rate at vertices (V) near stent intersections.

8 98 R.K. Banerjee et al. / Developed pulsatile flow in a deployed coronary stent downstream of each strut intersection and positive WSS at the upstream of each strut intersection during the entire cardiac cycle for all flow rates except near the entrance (vertex 1), where there is a negative WSS (negative MOSI value) for 50, 100 and 150 ml/min flow rates. Similar observation is seen for developing flow [21] at vertex 1 at basal flow; however, as the flow rate increases, positive WSS is seen at vertex 1. This happens due to the suppressed recirculation region at the upstream of vertex 1 because of high momentum of the fluid at hyperemic flow rate. In general, there is a persistent recirculation at the downstream of each strut intersection. It is evident from the Figs 3C, 3D and 4 that zones of flow stagnation and recirculation zones are formed at the immediate upstream and downstream of each strut intersection, respectively. In line with these results, previous studies have show that large amounts of dye accumulated between adjacent struts [5] Recirculation length and height Figure 5 shows the variation of recirculation length and height for basal to hyperemic flow rates at times of accelerating, peak and decelerating flows. Recirculation length and height [21] are calculated as the maximum length of recirculation zone formed at the immediate downstream of each strut intersection in axial and radial directions, respectively. For developed flow, there is no significant variation in size Fig. 5. Variation of recirculation length along the struts for basal & hyperemic flow rates at times of accelerating (A), peak (B) and decelerating (C, D) flows.

9 R.K. Banerjee et al. / Developed pulsatile flow in a deployed coronary stent 99 of the recirculation zone formed at the downstream of each strut intersection except for hyperemic flow. For hyperemic flow, at time of peak diastole (t = 2.0 s) and during deceleration (t = 2.20 s), the length and height of recirculation zone formed near the entrance i.e. strut intersection 1 is smaller than those formed near strut intersections 2, 3 and 4. Maximum recirculation length of 0.22 mm (Fig. 5B) and height of 0.04 mm (not shown) are observed near strut intersection 2 at peak diastole for developed hyperemic flow. It can be seen that the developed flow recirculation length and height are either less than or equal to the developing flow values for the basal flow rate. Near strut intersection 2, at peak diastole, the maximum recirculation length and recirculation height for developing hyperemic flow are 100% and 50% higher, respectively, than that for developed flow. Thus, the nature of the flow is critical in determining the size of recirculation zones [2,3] formed inside the stented artery Pressure drop Figure 6A shows the pressure drop ( p = p out p in ) between outlet and inlet of the artery for a pulse cycle for developed and developing [21] flows at basal and hyperemic conditions. As expected, the p profile follows the inlet velocity profile. The existence of the stent causes the minimum p to be Fig. 6. Pressure drop along the stent for a pulse cycle (A) and a comparison between developing and developed time-averaged mean pressure drops for a deployed stent (B).

10 100 R.K. Banerjee et al. / Developed pulsatile flow in a deployed coronary stent 0.35 mmhg for basal flow rate and 1.78 mmhg for hyperemic flow rate at peak diastole (t = 2.0 s) for developed flow. When compared with the developed flow, the minimum p for developing flow is more than 50% higher for basal flow ( 0.55 mmhg) and more than twice for hyperemic flow ( 4 mmhg). There is a positive p occurring at the end (t = 2.4 s) of the pulse cycle because of the flow reversal occurring near the arterial and stent wall regions. Integration of p(t) over the pulse cycle provides the time averaged (mean) pressure drop, p. The p Q (mean flow rate) relation is shown in Fig. 6B. As the flow rate increases from basal to hyperemic, the p increases by a factor of 4.5 for developed flow and 6.4 for developing flow. Also, as the Q increases, there is a sharp increase in mean p for developing flow reaching a maximum of 1.8 mmhg at hyperemic condition, which is more than twice that of developed flow (0.8 mmhg). 4. Conclusions The present study quantifies and compares the effect of developed and developing [21] pulsatile flows on hemodynamic parameters in a stented artery. Past studies report that the WSS reaching a value of 20 dyn/cm 2 over a short span depict a rapid increase in the activation of platelets [6,26,27]. The present results show peak WSS value of 290 dyn/cm 2 at the entrance of the stent for developed flow, which is half of that for developing flow. This implies that the developing flow might have a larger impact on the arterial wall such as higher level of platelet activation compared to developed flow. Negative WSS, signifying recirculation region, is observed at the immediate downstream region of each strut cross-link for developed flow. When compared to developed flow, maximum recirculation length and height for developing flow are 100% higher and 50% higher, respectively. Caro et al. [7] and Nerem and Levesque [19] have shown that recirculation zones of low shear stresses are susceptible to the deposition of macromolecules and other lipid derivatives. The present results show that the location of stent in the coronary artery determines the developing or developed nature of the flow, which in turn, results in varied level of WSS. Developing flow may be subjected to increased patho-physiologic conditions. The limitation of the present study is that it is carried out for a straight rigid artery. However, the coronary arteries are curved and, hence, there would be alterations in WSS distributions due to centrifugal acceleration. As reported in several previous studies, diseased coronary arteries behave as rigid tubes and, thus, the impact of either compliant arterial wall or compliance mismatch has been ignored. To delineate the effects of compliance and compliance mismatch, further investigations are needed. Flow field downstream of stent needs to be studied to obtain a better understanding of the flow disturbances induced by stent. Also, stent implantation resulting in residual stenosis requires investigation. Currently, clinicians use: (1) stents of improved designs with complex structures than that used in the present study; and (2) drug coated stents. Flow through the stents of improved design needs to be evaluated in conjunction with pharmacokinetics associated with drug coated stents. Nomenclature d vessel diameter, cm p static pressure, mmhg p pressure drop (p out p in ), mmhg Q mean flow rate, ml/min Re mean-flow Reynolds number (4 Q/πνd) t time, s

11 R.K. Banerjee et al. / Developed pulsatile flow in a deployed coronary stent 101 u axial velocity, cm/s x axial distance, cm α frequency parameter (d/2)(ω/ν) 0.5 γ shear rate, s 1 γ ij rate of strain tensor, s 1 η blood viscosity, poise ν kinematic viscosity (η/ρ), cm 2 /s ρ density, g/cm 3 τ wall shear stress [WSS], dyn/cm 2 T period of cardiac cycle, s ω circular frequency (2π/T), s 1 Subscript avg spatial average in inlet of artery out outlet of artery peak peak of cardiac cycle w wall condition Superscript time-average (mean over cardiac cycle) References [1] R.K. Banerjee, L.H. Back and Y.I. Cho, Computational fluid dynamics modeling techniques, using finite element methods to predict arterial blood flow in: Biomechanical Systems Techniques and Applications: Biofluid Methods in Vascular and Pulmonary Systems, Vol. 4, CRC Press, 2001, Chapter 8. [2] A.I. Barakat and E.T. Cheng, Numerical simulation of fluid mechanical disturbance induced by intravascular stents, Proceedings of the 11th International Conference on Mechanics in Medicine and Biology, Maui, Hawaii, April 2 5, [3] A.I. Barakat and L.G. Schachter, Computational study of arterial flow disturbance induced by intravascular stents, in: Proceedings of the Summer Bioengineering Conference, BED 50 (2001), [4] J.L. Berry, J.E. Moore, E. Manoach and P.H. Rolland, Hemodynamics and wall mechanics of a compliance matching stent: in vitro and in vivo analysis, J. Vasc. Int. Rad. 13 (2002), [5] J.L. Berry, A. Santamarina, J.E. Moore, S. Roychowdhury and W.D. Routh, Experimental and computational flow evaluation of coronary stents, Ann. Biomed. Eng. 28 (2000), [6] D.Y. Bluestein, Y.M. Li and I.B. Krukenkamp, Free emboli formation in the wake of bi-leaflet mechanical heart valves and the effects of implantation techniques, J. Biomech. 35 (2002), [7] C.G. Caro, J.M. Fitzgerald and R.C. Schroter, Arterial wall shear and distribution of early atheroma in man, Nature 223 (1969), [8] Y.I. Cho, L.H. Back, D.W. Crawford and R.F. Cuffel, Experimental study of pulsatile and steady flow through a smooth tube and an atherosclerotic coronary artery casting of man, J. Biomech. 16 (1983), [9] Y.I. Cho and K.R. Kensey, Effects of the non-newtonian viscosity of blood on flows in a diseased arterial vessel: part I, steady flows, Biorheology 28 (1991), [10] E.R. Edelman and C. Rogers, Pathobiologic responses to stenting, Am. J. Cardiol. 81 (1998), 4E 6E. [11] Fluent User Manual, ver , Fluent Inc., Lebanon, NH, [12] A.O. Frank, P.W. Walsh and J.E. Moore, Computational fluid dynamics and stent design, Artif. Organs 26 (2002), [13] Heart and Stroke Statistical Update, American Heart Association, 2001, pp [14] F.S. Henry, Simulation of flow through model stented arteries, Proceedings of the Summer Bioengineering Conference, BED 50 (2001), [15] A. Kastrati, J. Mehilli, J. Dirschinger, F. Dotzer, H. Schuhlen, F.J. Neumann, M. Fleckenstein, C. Pfafferott, M. Seyfarth

12 102 R.K. Banerjee et al. / Developed pulsatile flow in a deployed coronary stent and A. Schomig, Intracoronary stenting and angiographic results Strut thickness effect on restenosis outcome (ISAR- STEREO) trial, Circulation 103 (2001), [16] A. Kastrati, J. Mehilli, J. Dirschinger, J. Pache, K. Ulm, H. Schuhlen, M. Seyfarth, C. Schmitt, R. Blasini, F.J. Neumann and A. Schomig, Restenosis after coronary placement of various stent types, Am. J. Cardiol. 87 (2001), [17] J.F. LaDisa, I. Guler, L.E. Olson, D.A. Hettrick, J.R. Kersten, D.C. Warltier and P.S. Pagel, Three-dimensional computational fluid dynamics modeling of alterations in coronary wall shear stress produced by stent implantation, Ann. Biomed. Eng. 31 (2003), [18] J.E. Moore and J.L. Berry, Fluid and solid mechanical implications of vascular stenting, Ann. Biomed. Eng. 30 (2002), [19] R.M. Nerem and M.J. Levesque, Fluid mechanics in atherosclerosis, in: Handbook of Bioengineering, R. Skalak and S. Chien, eds, 1987, Chapter 21, [20] W.W. Nicholas and M.F. O Rourke, McDonald s Blood Flow in Arteries Theoretical, Experimental and Clinical Principles, Oxford University Press, New York, 1998, pp [21] D. Rajamohan, R.K. Banerjee, A.L. Ibrahim, M.A. Jog and L.H. Back, Developing pulsatile flow in a deployed coronary stent, ASME J. Biomech. Eng. 128 (2006), [22] C. Rogers and E.R. Edelman, Endovascular stent design dictates experimental restenosis and thrombosis, Circulation 91 (1995), [23] T. Seo, L.G. Schachter and A.I. Barakat, Computational study of fluid mechanical disturbance induced by endovascular stents, Ann. Biomed. Eng. 33(4) (2005), [24] J.J. Wentzel, R. Krams, C.H. Schuurbiers, J.A. Oomen, J. Kloet, W.J. van der Giesen, P.W. Serruys and C.J. Slager, Relationship between neointimal thickness and shear stress after Wallstent implantation in human coronary arteries, Circulation 103 (2001), [25] J.J. Wentzel, D.M. Whelan,W.J. van dergiessen, H.M.M. van Beusekom, I. Andhyiswara, P.W. Serruys, C.J. Slager and R. Krams, Coronary stent implantation changes 3-D vessel geometry and 3-D shear stress distribution, J. Biomech. 33 (2000), [26] N. Li, A.H. Goodall and P. Hjemdahl, Efficient flow cytometric assay for platelet leukocyte aggregates in whole blood using fluorescence signal triggering, Cytometry 35 (1999), [27] K.T. Nguyen, S.-H. Su, A. Sheng, D. Wawro, N.D. Schwade, C.F. Brouse, P.E. Greilich, L. Tang and R.C. Eberhart, In vitro hemocompatibility studies of drug-loaded poly-(l-lactic acid) fibers, Biomaterials 24 (2003),

Developing Pulsatile Flow in a Deployed Coronary Stent

Developing Pulsatile Flow in a Deployed Coronary Stent Divakar Rajamohan Department of Mechanical Engineering, University of Cincinnati, Cincinnati, OH 45221 Rupak K. Banerjee 1 Department of Mechanical Engineering, and Department of Biomedical Engineering,

More information

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches Korea-Australia Rheology Journal Vol. 16, No. 2, June 2004 pp. 75-83 Numerical simulations of fluid mechanical interactions between two abdominal aortic branches Taedong Kim, Taewon Seo* 1,2 and Abdul.I.

More information

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents Dublin Institute of Technology ARROW@DIT Conference Papers School of Mechanical and Design Engineering 2008-09-01 Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational

More information

Numerical Study on Effects of Drug-coating Position of Drug-eluting Stents on Drug Concentration

Numerical Study on Effects of Drug-coating Position of Drug-eluting Stents on Drug Concentration Journal of Medical and Biological Engineering, 34(5): 487-494 487 Numerical Study on Effects of Drug-coating Position of Drug-eluting Stents on Drug Concentration Yu Chen 1 Fei Yan 1 Wen-Tao Jiang 1,*

More information

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY LABORATORY OF BIOLOGICAL STRUCTURE MECHANICS www.labsmech.polimi.it FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY Rossella Balossino, Francesca Gervaso, Francesco Migliavacca,

More information

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm I. Husain, C. Langdon and J. Schwark Department of Mathematics Luther College University of Regina Regina, Saskatchewan

More information

Comparison of Stent Designs using Computational Fluid Dynamics

Comparison of Stent Designs using Computational Fluid Dynamics Dublin Institute of Technology ARROW@DIT Conference Papers School of Mechanical and Design Engineering 2007-03-28 Comparison of Stent Designs using Computational Fluid Dynamics Jonathan Murphy Dublin Institute

More information

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery Proceedings of the 3 rd International Conference on Fluid Flow, Heat and Mass Transfer (FFHMT 16) Ottawa, Canada May 2 3, 2016 Paper No. 170 Numerical Simulation of Blood Flow through Asymmetric and Symmetric

More information

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture RAMON BERGUER, a,b JOSEPH L. BULL, a,b AND KHALIL KHANAFER a a Vascular Mechanics Laboratory, Department of Biomedical Engineering,

More information

The influence of strut-connectors in stented vessels: A comparison of pulsatile flow through

The influence of strut-connectors in stented vessels: A comparison of pulsatile flow through Manuscript Click here to download Manuscript: paper_pant_et_al_revised_.ps 1 1 1 1 1 1 1 1 0 1 0 1 0 1 The influence of strut-connectors in stented vessels: A comparison of pulsatile flow through five

More information

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries Copyright 2015 Tech Science Press MCB, vol.12, no.1, pp.37-47, 2015 CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries P. Jhunjhunwala,, P.M. Padole, and S.B. Thombre, Abstract:

More information

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION Jayme Pinto Ortiz University of São Paulo - Avenida Prof. Luciano Gualberto, travessa3 nº 380 - CEP - 05508-900

More information

Physiological flow analysis in significant human coronary artery stenoses

Physiological flow analysis in significant human coronary artery stenoses Biorheology 40 (2003) 451 476 451 IOS Press Physiological flow analysis in significant human coronary artery stenoses Rupak K. Banerjee a,, Lloyd H. Back b, Martin R. Back c and Young I. Cho d a Department

More information

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Excerpt from the Proceedings of the COMSOL Conference 2010 Paris (COMSOL Conference) Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Vaidehi

More information

Arteriovenous Graft Modeling and Hemodynamic Interpretation

Arteriovenous Graft Modeling and Hemodynamic Interpretation Open Journal of Fluid Dynamics, 2012, 2, 324-330 http://dx.doi.org/10.4236/ojfd.2012.24a040 Published Online December 2012 (http://www.scirp.org/journal/ojfd) Arteriovenous Graft Modeling and Hemodynamic

More information

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 Presented at the COMSOL Conference 2010 India ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 1 DEPARTMENT OF MECHANICAL ENGINEERING & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING

More information

The relative effects of arterial curvature and lumen diameter on wall shear stress distributions in human right coronary arteries

The relative effects of arterial curvature and lumen diameter on wall shear stress distributions in human right coronary arteries The relative effects of arterial curvature and lumen diameter on wall shear stress distributions in human right coronary arteries Author Johnston, Barbara, Johnston, Peter Published 2007 Journal Title

More information

Simulations of the blood flow in the arterio-venous fistula for haemodialysis

Simulations of the blood flow in the arterio-venous fistula for haemodialysis Acta of Bioengineering and Biomechanics Vol. 16, No. 1, 2014 Original paper DOI: 10.5277/abb140109 Simulations of the blood flow in the arterio-venous fistula for haemodialysis DANIEL JODKO*, DAMIAN OBIDOWSKI,

More information

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis Biomedical & Pharmacology Journal Vol. 8(1), 123-131 (2015) Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis M. JAHANGIRI 1 *, M.

More information

Three-Dimensional Computational Fluid Dynamics Modeling of Alterations in Coronary Wall Shear Stress Produced by Stent Implantation

Three-Dimensional Computational Fluid Dynamics Modeling of Alterations in Coronary Wall Shear Stress Produced by Stent Implantation Annals of Biomedical Engineering, Vol. 31, pp. 972 980, 2003 Printed in the USA. All rights reserved. 0090-6964/2003/31 8 /972/9/$20.00 Copyright 2003 Biomedical Engineering Society Three-Dimensional Computational

More information

Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent

Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent Kuang-Huei Lee, Hao-Ming Hsiao, Ying-Chih Liao, Yi-Hsiang Chiu, Yu-Seng Tee Dept. of

More information

Medical device design using Computational Fluid Dynamics (CFD)

Medical device design using Computational Fluid Dynamics (CFD) Medical device design using Computational Fluid Dynamics (CFD) Session: Winter 2016 IMPORTANT NOTE: This project has 8 deliverables, for each one timely work is expected. 1. General Design Specifications

More information

Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery

Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery Korea-Australia Rheology Journal Vol. 21, No. 2, June 2009 pp. 119-126 Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery Kun Hyuk Sung, Kyoung

More information

Post-conditioning. P a g e 1. To my Thesis Committee,

Post-conditioning. P a g e 1. To my Thesis Committee, P a g e 1 To my Thesis Committee, This document seeks to clarify my research project. After describing what post-conditioning (PC) is, I will explain differences between my research and the recent peristaltic

More information

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt Washington University in St. Louis Washington University Open Scholarship Mechanical Engineering and Materials Science Independent Study Mechanical Engineering & Materials Science 12-23-2017 Computational

More information

Physiological Flow Simulation in Residual Human Stenoses After Coronary Angioplasty

Physiological Flow Simulation in Residual Human Stenoses After Coronary Angioplasty Rupak K. Banerjee Staff Scientist, Bioengineering and Physical Science Program, Bldg. 13, Rm. 3N17, National Institute of Health (NIH), Bethesda, MD 20892 Lloyd H. Back Jet Propulsion Laboratory, California

More information

Experimental Flow Studies on Carotid Artery Models with and without Stents

Experimental Flow Studies on Carotid Artery Models with and without Stents Experimental Flow Studies on Carotid Artery Models with and without Stents Liepsch, D 1, Schmid, Th. 1, Klügel, G. 1, Sakurai, A. 3, Berger, H. 2, Greil, O. 2 1 FB 05, Fachhochschule München, Lothstr.

More information

Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries

Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries Author Johnston, Barbara, Johnston, Peter Published 2009 Conference Title 18th IMACS World Congress MODSIM09

More information

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement Interactive CardioVascular and Thoracic Surgery 19 (2014) 16 20 doi:10.1093/icvts/ivu034 Advance Access publication 22 April 2014 ORIGINAL ARTICLE ADULTCARDIAC A computational fluid dynamics simulation

More information

Non-Newtonian blood flow in human right coronary arteries: Transient simulations

Non-Newtonian blood flow in human right coronary arteries: Transient simulations Non-Newtonian blood flow in human right coronary arteries: Transient simulations Author Johnston, Barbara, Johnston, Peter, Corney, Stuart, Kilpatrick, David Published 2006 Journal Title Journal of Biomechanics

More information

Effects of surface geometry and non-newtonian viscosity on the flow field in arterial stenoses

Effects of surface geometry and non-newtonian viscosity on the flow field in arterial stenoses Journal of Mechanical Science and Technology 23 (2009) 2424~2433 Journal of Mechanical Science and Technology www.springerlink.com/content/1738-494x DOI 10.1007/s12206-009-0627-6 Effects of surface geometry

More information

restenosis; neointimal hyperplasia; scaffolding; prolapse; three-dimensional

restenosis; neointimal hyperplasia; scaffolding; prolapse; three-dimensional J Appl Physiol 98: 947 957, 005. First published November 5, 004; doi:10.115/japplphysiol.0087.004. Circumferential vascular deformation after stent implantation alters wall shear stress evaluated with

More information

FFR CT : Beyond FFR. Bon-Kwon Koo, MD, PhD. Seoul National University Hospital, Seoul, Korea. Seoul National University Hospital Cardiovascular Center

FFR CT : Beyond FFR. Bon-Kwon Koo, MD, PhD. Seoul National University Hospital, Seoul, Korea. Seoul National University Hospital Cardiovascular Center FFR CT : Beyond FFR Bon-Kwon Koo, MD, PhD, Seoul, Korea Patient-specific non-invasive coronary hemodynamic assessment Non-invasive, Pt-specific Hemodynamics Pressure Pressure difference Pressure gradient

More information

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images Applied Mathematics Volume, Article ID 77, pages http://dx.doi.org/.//77 Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images Mahsa Dabagh,, Wakako

More information

Mathematical and Computational study of blood flow through diseased artery

Mathematical and Computational study of blood flow through diseased artery Mathematical and Computational study of blood flow through diseased artery Abstract This paper presents the study of blood flow through a tapered stenosed artery. The fluid (blood) medium is assumed to

More information

Computational Analysis on Commercially Available Stent Designs

Computational Analysis on Commercially Available Stent Designs Computational Analysis on Commercially Available Stent Designs Abhijit Chanda 1, Shuvrangsu Das 2, Sounak Bhattacharjee 2, Pranab Ghosh 2, K. Basu 3 1 School of Bio Science and Engineering, Jadavpur University,

More information

A Two-layered Model for the Analysis of Arterial Rheology

A Two-layered Model for the Analysis of Arterial Rheology IJCSIT International Journal of Computer Science and Information Technology, Vol., o., June, pp. 37- A Two-layered Model for the Analysis of Arterial Rheology Sapna Singh Department of Mathematics, Harcourt

More information

BioMedical Engineering OnLine

BioMedical Engineering OnLine BioMedical Engineering OnLine BioMed Central Research Evaluation of the effect of stent strut profile on shear stress distribution using statistical moments Juan Mejia 1,4, Bilal Ruzzeh 1, Rosaire Mongrain*

More information

FOR many decades, cardiovascular disease has been one of

FOR many decades, cardiovascular disease has been one of Vol:1, No:2, 27 Effect of Non-Newtonian Behaviour of Blood on Pulsatile Flows in Stenotic Arteries Somkid Amornsamankul, Benchawan Wiwatanapataphee, Yong Hong Wu, Yongwimon Lenbury International Science

More information

Roles of Flow Mechanics in Vascular Cell Biology in Health and Disease

Roles of Flow Mechanics in Vascular Cell Biology in Health and Disease Roles of Flow Mechanics in Vascular Cell Biology in Health and Disease Shu Chien Dept. of Bioengineering & Medicine UC, San Diego Presented by Ming-Shaung Ju Dept. of Mech. Eng., NCKU, Tainan Background

More information

Hematocrit Level on Blood flow through a Stenosed Artery with Permeable Wall: A Theoretical Study

Hematocrit Level on Blood flow through a Stenosed Artery with Permeable Wall: A Theoretical Study Available at http://pvamu.edu/aam Appl. Appl. Math. ISSN: 1932-9466 Vol. 12, Issue 1 (June 2017), pp. 291-304 Applications and Applied Mathematics: An International Journal (AAM) Hematocrit Level on Blood

More information

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM H. GIRIJA BAI 1 and K.B. NAIDU 2 Department of Mathematics, Sathyabama University, Chennai-600 119, Tamil Nadu, India 1 girijanameprakash@gmail.com

More information

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation EasyChair Preprint 259 Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation Tianyi Xia, Matthew Doyle, Thomas Forbes and Cristina H. Amon EasyChair preprints

More information

Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics

Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics Sinjae Hyun, PhD, a Clement Kleinstreuer, PhD, b and Joseph P. Archie, Jr, PhD,

More information

Computational design of Intracranial Stent using 3D visualization system

Computational design of Intracranial Stent using 3D visualization system Computational design of Intracranial Stent using 3D visualization system Institute of Fluid Science, Tohoku University Makoto OHTA Graduate school of Engineering Hitomi Anzai Graduate school of Biomedical

More information

Particle Hemodynamics Analysis after Coronary Angioplasty

Particle Hemodynamics Analysis after Coronary Angioplasty , June 30 - July 2, 2010, London, U.K. Particle Hemodynamics Analysis after Coronary Angioplasty S.I. Bernad, A.F. Totorean, V.F. Vinatu, R.F. Susan-Resiga Abstract The resistance to flow through a stenosis

More information

Blood Flow Simulation toward Actual Application at Hospital

Blood Flow Simulation toward Actual Application at Hospital THE 5 TH ASIAN COMPUTAITIONAL FLUID DYNAMICS BUSAN, KOREA, OCTOBER 27 ~ OCTOBER 30, 2003 Blood Flow Simulation toward Actual Application at Hospital Abstract R. Himeno 1 1. Advanced Center for Computing

More information

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft BURASKORN NUNTADILOK 1, BENCHAWAN WIWATANAPATAPHEE 1 MEECHOKE CHUEDOUNG 1, THANONGCHAI SIRIAPISITH

More information

Treatment of lesions at bifurcations: General perspective

Treatment of lesions at bifurcations: General perspective ESC Congress 2011 27-31 August, Paris, France Treatment of lesions at bifurcations: General perspective Goran Stankovic MD, PhD, FESC Division of Cardiology Clinical Center of Serbia, Belgrade Serbia Potential

More information

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases Proceedings of the 215 2nd International Symposium on Physics and Technology of Sensors, 8-1th March, 215, Pune, India Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

More information

Investigation of the Hemodynamics of Coronary Arteries - Effect of Stenting. Naresh Kumar Coimbatore Selvarasu

Investigation of the Hemodynamics of Coronary Arteries - Effect of Stenting. Naresh Kumar Coimbatore Selvarasu Investigation of the Hemodynamics of Coronary Arteries - Effect of Stenting Naresh Kumar Coimbatore Selvarasu Dissertation submitted to the faculty of the Virginia Polytechnic Institute and State University

More information

NUMERICAL SIMULATION OF EFFECTS OF REYNOLDS NUMBER ON NON-NEWTONIAN BLOOD FLOW WITH SPIRAL COMPONENT THROUGH A REGULAR STENOSED ARTERY

NUMERICAL SIMULATION OF EFFECTS OF REYNOLDS NUMBER ON NON-NEWTONIAN BLOOD FLOW WITH SPIRAL COMPONENT THROUGH A REGULAR STENOSED ARTERY Proceedings of the International Conference on Mechanical Engineering and Renewable Energy 2017 (ICMERE2017) 18 20 December, 2017, Chittagong, Bangladesh ICMERE2017-PI-325 NUMERICAL SIMULATION OF EFFECTS

More information

A numerical study on the effect of hematocrit on hemodynamic characteristics in arteriovenous graft

A numerical study on the effect of hematocrit on hemodynamic characteristics in arteriovenous graft Korea-Australia Rheology Journal, Vol.26, No.3, pp.327-334 (August 2014) DOI: 10.1007/s13367-014-0037-x www.springer.com/13367 A numerical study on the effect of hematocrit on hemodynamic characteristics

More information

Blood flow in S-shaped in-plane and out-of-plane coronary arteries

Blood flow in S-shaped in-plane and out-of-plane coronary arteries Blood flow in S-shaped in-plane and out-of-plane coronary arteries Author Johnston, Peter, Johnston, Barbara Published 2008 Journal Title The A N Z I A M Journal DOI https://doi.org/10.21914/anziamj.v49i0.330

More information

Flow in re-stenosed artery after angioplasty

Flow in re-stenosed artery after angioplasty Data Management and Security 209 Flow in re-stenosed artery after angioplasty S. I. Bernad 1, A. Totorean 2, E. S. Bernad 3 & R. Susan-Resiga 2 1 Romanian Academy, Timisoara Branch, Romania 2 Politehnica

More information

4D model of hemodynamics in the abdominal aorta

4D model of hemodynamics in the abdominal aorta Bio-Medical Materials and Engineering 26 (2015) S257 S264 DOI 10.3233/BME-151312 IOS Press S257 4D model of hemodynamics in the abdominal aorta Ireneusz Zbicinski a,*, Natalia Veshkina a and Ludomir Stefa

More information

BLOOD FLOW THROUGH A COMPOSITE STENOSIS IN CATHETERIZED ARTERIES INTRODUCTION

BLOOD FLOW THROUGH A COMPOSITE STENOSIS IN CATHETERIZED ARTERIES INTRODUCTION 55 BLOOD FLOW THROUGH A COMPOSITE STENOSIS IN CATHETERIZED ARTERIES V P Srivastava, Rochana Vishnoi, Shailesh Mishra, Poonam Sinha * Department of Mathematics, Krishna Girls Engineering College, Kanpur-917,

More information

Hemodynamics in Idealized Stented Coronary Arteries: Important Stent Design Considerations

Hemodynamics in Idealized Stented Coronary Arteries: Important Stent Design Considerations Annals of Biomedical Engineering, Vol. 44, No. 2, February 2016 ( 2015) pp. 315 329 DOI: 10.1007/s10439-015-1387-3 Medical Stents: State of the Art and Future Directions Hemodynamics in Idealized Stented

More information

CVS Hemodynamics. Change in blood pressure:

CVS Hemodynamics. Change in blood pressure: CVS Hemodynamics -The distribution of blood inside the circulation: The major part of blood volume is found in the venous system 60% (2/3), that s why veins are called the capacitance vessels. -Arteries

More information

CPM Specifications Document Aortofemoral Normal:

CPM Specifications Document Aortofemoral Normal: CPM Specifications Document Aortofemoral Normal: OSMSC 0003_0000 0006_0000 May 24, 2013 Version 1 Open Source Medical Software Corporation 2013 Open Source Medical Software Corporation. All Rights Reserved.

More information

Analysis of the effects of plaque deposits on the blood flow through human artery

Analysis of the effects of plaque deposits on the blood flow through human artery ISSN 2395-1621 Analysis of the effects of plaque deposits on the blood flow through human artery #1 Sajid S. Mulani, #2 Pankaj I. Jagad 1 sajidsmulani21@gmail.com 2 pjjagad.scoe@sinhgad.edu #12 Department

More information

CPM Specifications Document Aortic Coarctation: Exercise

CPM Specifications Document Aortic Coarctation: Exercise CPM Specifications Document Aortic Coarctation: Exercise OSMSC 0091_2000 0102_2000 0107_0000 0111_0000 May 29, 2013 Version 1 Open Source Medical Software Corporation 2013 Open Source Medical Software

More information

Blood Flow At Arterial Branches: Complexities To Resolve For The Angioplasty Suite

Blood Flow At Arterial Branches: Complexities To Resolve For The Angioplasty Suite Blood Flow At Arterial Branches: Complexities To Resolve For The Angioplasty Suite P.D. Richardson 1, I.V. Pivkin 2, G.E. Karniadakis 2, D.H. Laidlaw 3 1 Division of Engineering 2 Division of Applied Mathematics

More information

Blood flow in vessels with artificial or pathological geometrical changes

Blood flow in vessels with artificial or pathological geometrical changes Blood flow in vessels with artificial or pathological geometrical changes P. Tibaut 1, B. Wiesler 1, M. Mayer 2 & R. Wegenkittel 3 1 AVL LIST GmbH, Graz, Austria 2 VRVIs, Vienna, Austria 3 Tiani Medgraph

More information

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction Eann A. Patterson Department of Mechanical Engineering, The University of Sheffield Mappin Street, Sheffield, S1 3JD

More information

Flow Analysis with Stent Placement in the Cerebral Aneurysm

Flow Analysis with Stent Placement in the Cerebral Aneurysm Proceedings of the ECCOMAS Thematic International Conference on Simulation and Modeling of Biological Flows (SIMBIO 2011) September 21-23, 2011, VUB, Brussels, Belgium Flow Analysis with Stent Placement

More information

Mathematical Modeling of Bingham Plastic Model of Blood Flow Through Stenotic Vessel

Mathematical Modeling of Bingham Plastic Model of Blood Flow Through Stenotic Vessel RESEARCH ARTICLE OPEN ACCESS Mathematical Modeling of Bingham Plastic Model of Blood Flow Through Stenotic Vessel S.R. Verma Department of Mathematics, D.A-V. (P.G.) College, Kanpur-208001, India Abstract

More information

Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid Artery with Stenosis

Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid Artery with Stenosis See discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/323417296 Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid

More information

Influence of Oscillating Flow on LDL Transport and Wall Shear Stress in the Normal Aortic Arch

Influence of Oscillating Flow on LDL Transport and Wall Shear Stress in the Normal Aortic Arch 128 The Open Cardiovascular Medicine Journal, 2009, 3, 128-142 Open Access Influence of Oscillating Flow on LDL Transport and Wall Shear Stress in the Normal Aortic Arch J. Soulis 1,*, G. Giannoglou 2,

More information

CPM Specifications Document Aortofemoral Normal:

CPM Specifications Document Aortofemoral Normal: CPM Specifications Document Aortofemoral Normal: OSMSC 0110_0000 May 27, 2013 Version 1 Open Source Medical Software Corporation 2013 Open Source Medical Software Corporation. All Rights Reserved. 1. Clinical

More information

1992 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 59, NO. 7, JULY 2012

1992 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 59, NO. 7, JULY 2012 1992 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 59, NO. 7, JULY 2012 Identification of Hemodynamically Optimal Coronary Stent Designs Based on Vessel Caliber Timothy J. Gundert, Alison L. Marsden,

More information

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System Contents 1 Computational Haemodynamics An Introduction... 1 1.1 What is Computational Haemodynamics (CHD)... 1 1.2 Advantages of CHD... 3 1.3 Applications in the Cardiovascular System... 4 1.3.1 CHD as

More information

Vascular reconstruction: CFD predictions of bypass graft haemodynamics

Vascular reconstruction: CFD predictions of bypass graft haemodynamics Vascular reconstruction: CFD predictions of bypass graft haemodynamics J.S. Cole 1, J.K. Watterson 1 & M.J.G. O Reilly 2 1 School of Mechanical and Aerospace Engineering, The Queen s University of Belfast,

More information

FSI Analysis of Diseased Coronary using Patient Specific Data

FSI Analysis of Diseased Coronary using Patient Specific Data FSI Analysis of Diseased Coronary using Patient Specific Data Mingchao Cai, * Haofei Liu, a Chun Yang, a,c Jie Zheng, b Richard Bach, b Mehmet H. Kural, d Kristen L. Billiar, d David Muccigrosso, b Dongsi

More information

Intracoronary Stenting and Angiographic Results: Strut Thickness Effect on Restenosis Outcome (ISAR-STEREO-2) Trial

Intracoronary Stenting and Angiographic Results: Strut Thickness Effect on Restenosis Outcome (ISAR-STEREO-2) Trial Journal of the American College of Cardiology Vol. 41, No. 8, 2003 2003 by the American College of Cardiology Foundation ISSN 0735-1097/03/$30.00 Published by Elsevier Science Inc. doi:10.1016/s0735-1097(03)00119-0

More information

E. Schuit July 2007 BMTE 08.15

E. Schuit July 2007 BMTE 08.15 Internship Pulsatile flow in a stented and non-stented 2D cerebral aneurysm model E. Schuit July 2007 BMTE 08.15 ir. G. Mulder prof. dr. ir. F.N. v.d. Vosse Eindhoven University of Technology Department

More information

On the Existence of an Optimum End-to-side Junctional Geometry in Peripheral Bypass Surgery A Computer Generated Study

On the Existence of an Optimum End-to-side Junctional Geometry in Peripheral Bypass Surgery A Computer Generated Study Eur J Vasc Endovasc Surg 26, 649 656 (2003) doi: 10.1016/j.ejvs.2003.08.004, available online at http://www.sciencedirect.com on On the Existence of an Optimum End-to-side Junctional Geometry in Peripheral

More information

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery 1 Sajid S. Mulani, 2 P. I. Jagad 1,2 Department of Mechanical Engineering, SCoE, Pune 411041, India Email: 1 sajidsmulani21@gmail.com,

More information

Coronary stenting is an established form of treatment for

Coronary stenting is an established form of treatment for Intracoronary Stenting and Angiographic Results Strut Thickness Effect on Restenosis Outcome (ISAR-STEREO) Trial Adnan Kastrati, MD; Julinda Mehilli, MD; Josef Dirschinger, MD; Franz Dotzer, MD; Helmut

More information

A Numerical Methodology to Fully Elucidate the Altered Wall Shear Stress in a Stented Coronary Artery

A Numerical Methodology to Fully Elucidate the Altered Wall Shear Stress in a Stented Coronary Artery Dublin Institute of Technology ARROW@DIT Articles School of Mechanical and Design Engineering 2010-11-11 A Numerical Methodology to Fully Elucidate the Altered Wall Shear Stress in a Stented Coronary Artery

More information

time-dependent solution of an incompressible. non-newtonian fluid residual neck after incomplete surgical clipping are found in the MEJHQPS

time-dependent solution of an incompressible. non-newtonian fluid residual neck after incomplete surgical clipping are found in the MEJHQPS FED-Vol. 216, Unsteady Flows -1995 ASME 1995 RECURRENCE OF BALLOON TREATED INTRACRANIAL TERMINAL ANEURYSM: A STUDY WITH PULSATILE BLOOD FLOW Rupak K. Banerjee, PhD. Young I. Cho, PhD. Department of Mechanical

More information

Asymmetric flows of non-newtonian fluids in symmetric stenosed artery

Asymmetric flows of non-newtonian fluids in symmetric stenosed artery Korea-Australia Rheology Journal Vol. 16, No. 2, June 2004 pp. 101-108 Asymmetric flows of non-newtonian fluids in symmetric stenosed artery Hun Jung*, Jong Wook Choi 1 and Chan Guk Park 2 Graduate School,

More information

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents Augusto F. Sanches and Eva Gutheil Interdisciplinary Center for Scientific Computing, Heidelberg University Flow diverters

More information

History of Vascular Modelling. William Harvey discovery of the circulation 1628

History of Vascular Modelling. William Harvey discovery of the circulation 1628 History of Vascular Modelling William Harvey discovery of the circulation 1628 William Harvey (1578-1657) Since all things, both argument and ocular demonstration, show that the blood passes through the

More information

Timothy J. Gundert Department of Biomedical Engineering, Marquette University, 1515 West Wisconsin Avenue, Milwaukee, WI 53233

Timothy J. Gundert Department of Biomedical Engineering, Marquette University, 1515 West Wisconsin Avenue, Milwaukee, WI 53233 Timothy J. Gundert Department of Biomedical Engineering, Marquette University, 1515 West Wisconsin Avenue, Milwaukee, WI 53233 Ronak J. Dholakia Department of Neurological Surgery, Stony Brook University

More information

BEE 453. May 4, Ravi Sood, David Nahlik, Weston Nichols, and Evan Graham

BEE 453. May 4, Ravi Sood, David Nahlik, Weston Nichols, and Evan Graham Laser Irradiation of Tumors for the Treatment of Cancer: An Analysis of Blood Flow, Temperature and Oxygen Transport BEE 453 May 4, 2007 Ravi Sood, David Nahlik, Weston Nichols, and Evan Graham 2 Executive

More information

3D Modeling of Plaque Progression in the Human Coronary Artery

3D Modeling of Plaque Progression in the Human Coronary Artery Proceedings 3D Modeling of Plaque Progression in the Human Coronary Artery Igor Saveljic 1,2, *, Dalibor Nikolic 1,2, Zarko Milosevic 1,2, Velibor Isailovic 1,2, Milica Nikolic 1,2, Oberdan Parodi 3 and

More information

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics 54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics Andžela Šešok 1, Donatas Lukšys 2 Vilnius Gediminas Technical University,

More information

Blood Flow at Arterial Branches: Complexities to Resolve for the Angioplasty Suite

Blood Flow at Arterial Branches: Complexities to Resolve for the Angioplasty Suite Blood Flow at Arterial Branches: Complexities to Resolve for the Angioplasty Suite P.D. Richardson 1,I.V.Pivkin 2, G.E. Karniadakis 2, and D.H. Laidlaw 3 1 Division of Engineering 2 Division of Applied

More information

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm Andrzej Polańczyk, MSc Ireneusz Zbiciński, PhD, DSc Abstract The aim of this study was to estimate whether

More information

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with Stenosis

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with Stenosis Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with Stenosis PEARANAT CHUCHARD 1, BENCHAWAN WIWATANAPATAPHEE 1, THITIKOM PUAPANSAWAT 1, THANONGCHAI SIRIAPISITH 2 1 Department

More information

There has been extensive speculation in the literature about

There has been extensive speculation in the literature about TECHNICAL NOTE A. Mantha C. Karmonik G. Benndorf C. Strother R. Metcalfe Hemodynamics in a Cerebral Artery before and after the Formation of an Aneurysm SUMMARY: Using data obtained from 3D digital subtraction

More information

Senior Honors Thesis Prospectus

Senior Honors Thesis Prospectus Senior Honors Thesis Prospectus [Author s Name] [Author s Contact Info] [Thesis Director s Name] University of Houston Biomedical Engineering Background Coronary Atherosclerosis -Stents -Oscillatory Wall

More information

Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation

Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation American Journal of Applied Sciences 7 (9): 1237-1242, 2010 ISSN 1546-9239 2010 Science Publications Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation 1 Vahab Dehlaghi and

More information

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique International Journal of Engineering Research and Technology. ISSN 0974-3154 Volume 6, Number 3 (2013), pp. 399-406 International Research Publication House http://www.irphouse.com Study of Newtonian and

More information

Investigation of metallic surface area of coronary stents

Investigation of metallic surface area of coronary stents Investigation of metallic surface area of coronary stents Dóra KÁROLY 1, Miksa KOVÁCS 1, Andrew Attila TERDIK 1, Eszter BOGNÁR 1,2 1 Department of Materials Science and Engineering, Faculty of Mechanical

More information

FFR Fundamentals and Measurements

FFR Fundamentals and Measurements FFR Fundamentals and Measurements Ghassan S. Kassab Thomas Linnemeier Chair Professor Biomedical Engineering, Indiana University Purdue University Indianapolis Principle of FFR Q S ( P P ) / R P max d

More information

Saccular Aneurysm Formation in Curved and Bifurcating Arteries

Saccular Aneurysm Formation in Curved and Bifurcating Arteries AJNR Am J Neuroradiol 20:1309 1317, August 1999 Saccular Aneurysm Formation in Curved and Bifurcating Arteries George N. Foutrakis, Howard Yonas, and Robert J. Sclabassi BACKGROUND AND PURPOSE: Distinguishing

More information

Resolute Integrity. Independent Conformability Assessment

Resolute Integrity. Independent Conformability Assessment Resolute Integrity Independent Conformability Assessment Stent Conformability and its Clinical Importance Immediate post-deployment stent geometry determines neointimal thickness independently of arterial

More information