MANAGING STERILIZATION PROCESS MONITOR FAILURES
|
|
- Jonah Allen
- 5 years ago
- Views:
Transcription
1 MANAGING STERILIZATION PROCESS MONITOR FAILURES Joseph F. LeBouef, RST, CST, CRCST, CHL Regional Sterile Processing Educator, Kaiser Foundation Health Plan of the Northwest President, IAHCSMM Cascade Chapter
2 Why Monitor the Sterilization Process? Critical items contact normally sterile tissues of the body-items must be sterile You can t see sterility You can t reasonably verify or test sterility Sterility is a probability function, affected by: Human factors Equipment and utility issues 2
3 Types of Sterilization Monitors Physical Mechanical charts, printouts or graphs Chemical Indicate exposure to sterilizing agent Biological Measure lethality of cycle 3
4 Information provided: Physical Monitors Whether or not the sterilizer performed as instructed Parametric only Only measures at one location in chamber No indication of whether sterilant reached device or if lethality was achieved 4
5 Chemical Monitors (Indicators) Information provided: Whether adequate sterilant concentration and/or other critical variables required for sterilization occurred at specific location inside sterilizer chamber or processed package No measure of cycle lethality Both external and internal for each package 6 Classifications not hierarchical 5
6 Classification of Chemical Indicators Class I: Process Indicators React to 1 critical variable Used to differentiate items that were exposed from those that were not Class II: Indicators for Use in Specific Tests Special tests (Bowie-Dick/DART - used to assess dynamic air removal sterilizer s ability to remove air) 6
7 Classification of Chemical Indicators Class III: Single Variable Indicators React to 1 critical variable Indicates performance in relation to at least 1 stated value Not commonly used Class IV: Multi-variable Indicators React to at least 2 critical variables May be used as internal pack monitors 7
8 Classification of Chemical Indicators Class V: Integrating Indicators React to ALL critical variables Used as internal pack monitors Performance must mimic biological performance Class VI: Emulating Indicators React to ALL critical variables Used as internal pack monitors Performance does not mimic biological performance Cycle specific 8
9 Placement of Chemical Indicators AAMI: Placed in area least accessible to sterilant penetration For containers, refer to manufacturer s written instructions AORN: For wrapped items: 1 in geographical center of package, on each level if multiple levels For containerized items: placed in 2 opposing corners, on each level 9
10 Biological Monitors (Indicators) Only monitor that directly measures cycle lethality Steam, H2O2 and ozone processes use known quantities of highly resistant Geobacillus stearothermophilus spores Specifically designed for particular modality employed Sterilizers monitored on routine basis and with ALL implantable devices Steam: at least weekly, preferably daily H2O2: at least daily Ozone: every cycle 10
11 Application of Biological Indicators Placed directly over drain, on lowest shelf of sterilization cart/rack Indicator must be specifically approved for use in the modality employed: 250⁰ F gravity displacement air removal (example: 3M 1292 BROWN top BI) 270⁰ F gravity displacement air removal (example: 3M 1291 BLUE top BI) 270⁰ F dynamic (pre-vac) air removal (example: 3M 1292 BROWN top BI, possibly in a process challenge device) All modalities employed must be tested 11
12 Application of BIs (Continued) Unprocessed BI used as control test to ensure viability of spores Daily, and With each new lot number of BI Implantable devices should not be released until BI results are known If emergency necessitates early release, an exception form should be used to communicate risk to provider 12
13 Exception Form for Premature Release of Implantable Device/Tray NOTE In a documented emergency situation, implantable devices will be released from quarantine without the biological monitor result. This form should accompany the implant to the Operating Room. Operating Room personnel should complete this form and return it to Central Service within 24 hours. PLEASE COMPLETE ALL INFORMATION: DATE: SHIFT: TIME: AM PM PERSON COMPLETING THIS REPORT IN CENTRAL SERVICE: The following implantable devices/trays were prematurely released to the Operating Room: NAME OF OR PERSON REQUESTING PREMATURE RELEASE OF DEVICES: OPERATING ROOM REPORT: PATIENT NAME: SURGEON NAME: TIME OF PROCEDURE: AM PM DATE: REASON PREMATURE RELEASE WAS NEEDED: WHAT COULD HAVE PREVENTED PREMATURE RELEASE OF THIS DEVICE/TRAY? NAME OF OR PERSON COMPLETING THIS REPORT: DATE REPORT COMPLETED: FORM RETURNED TO CENTRAL SERVICE ON: 13
14 When Monitors Indicate Failure Quarantine items/remove sterilizer from service If reason is apparent (incorrect cycle selected, sterilizer loaded improperly, etc.) Correct issue, repackage and reprocess items If reason is not apparent Recall all items processed since last negative biological test Any items used must be communicated to Infection Prevention & Control and to provider responsible for care of patient 14
15 Minor repair If Failure Results in Repair of a Sterilizer Routine monitors performed, then sterilizer released into service Major repair Repairs that effect the basic performance of the sterilizer (chamber welding, major refitting of piping, software overhaul, etc.) Sterilizer must be requalified 15
16 Steam Sterilizer Qualification Performed when a new sterilizer is installed, a sterilizer s location is changed or it undergoes major repair 3 back-to-back biological indicators performed in an empty chamber 3 Bowie-Dick/DART tests performed backto-back in an empty chamber Sterilizers quarantined until test results are obtained 16
17 Questions? Thank you! 17
18 References ANSI/AAMI ST79:2010 & Amendment 1/Amendment 2 AORN Perioperative Standards and Recommended Practices,
ST79:2006 Section 10: Quality Control Key Changes
ST79:2006 Section 10: Quality Control Key Changes Background: The Association for the Advancement of Medical Instrumentation (AAMI) newest recommended practice, Comprehensive guide to steam sterilization
More informationSterile U Network. Quality Control for Table-top Steam Sterilizers. Introduction: T U T O R I A L S
3M Attest Products TM TM Sterile U Network T U T O R I A L S Quality Control for Table-top Steam Sterilizers Introduction: Many health care professionals in office-based settings are not aware that they
More informationLarry! 28/06/2016. House Keeping. Welcome! From the GoToWebinar page:
A Review of Sterilization Methods and Recommended Practices for Healthcare Facilities PART I June 30, 2016 Larry Talapa Welcome! Topic: A Review of Sterilization Methods and Recommended Practices for Healthcare
More informationLoad Control Monitoring Using Process Challenge Devices
Load Control Monitoring Using Process Challenge Devices ANSI/AAMI ST79: 2010 update by Martha Young, BS, MS, CSPDT Objectives After completion of this self-study activity, the learner will be able to:
More informationDid You Know? Education & Training. Sterilization process monitoring Part One by Martha Young, BS, MS, CSPDT
Did You Know? Sterilization process monitoring Part One by Martha Young, BS, MS, CSPDT Objectives After completion of this self-study activity, the learner will be able to: 1. Identify when a Bowie-Dick
More informationAST Standards of Practice for Monitoring Sterility
AST Standards of Practice for Monitoring Sterility Introduction The following Standards of Practice were researched and authored by the AST Education and Professional Standards Committee and have been
More informationAST Guidelines for Best Practices in Monitoring Sterility
1 AST Guidelines for Best Practices in Monitoring Sterility Approved October 2008 Revised February 2015 Revised April 14, 2017 Introduction The following Guidelines for Best Practices were researched and
More informationApproval Signature: Original signed by B. Postl. June 2007
POLICY REGIONAL Applicable to all WRHA governed sites and facilities (including hospitals and personal care homes), and all funded hospitals and personal care homes. All other funded entities are excluded
More informationCondensation of the AAMI Steam Sterilization Recommended Practices Quality Control (Section 10), Part II
Condensation of the AAMI Steam Sterilization Recommended Practices Quality Control (Section 10), Part II Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities (ANSI/AAMI
More informationMonitoring The Sterilization Process. Are We Over The Top?
Monitoring The Sterilization Process. Are We Over The Top? Dr Brian Kirk, Senior Technical Services Specialist, Sterilization 3M Health Care Are We Over The Top? NO! Get it wrong and you kill the patient!
More informationThe AAMI Impact: What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing
The AAMI Impact: What Dentistry Can Learn from AAMI ST79 Best Practices for device and instrument reprocessing Learning goals: Purpose and goals of AAMI organization & overview of ST79. Review regulatory
More informationTo view an archived recording of this presentation please click the following link: Please scroll down this
To view an archived recording of this presentation please click the following link: http://pho.adobeconnect.com/p7f3sm3cbe4i/ Please scroll down this file to view a copy of slides from the session. Helpful
More informationReprocessing Guide. Autoclavable Arthroscope and Hardware Sterilization Tray
Reprocessing Guide Autoclavable Arthroscope and Hardware Sterilization Tray 0233032116 Contents Introduction...2 Intended Use of Sterilization Trays...4 Warnings...5 Cautions...5 Instructions...6 Sterilization
More informationComplex Medical Devices and Steam Sterilization. By Andrew Gay Director - Sterilizer Validation Australia
Complex Medical Devices and Steam Sterilization By Andrew Gay Director - Sterilizer Validation Australia About 10 years ago, during a routine Performance Qualification test, an observation was made that
More informationBecause You Can t See Sterile.
Sterilization Because You Can t See Sterile. Sterilization EVERY LOAD. Sure-Check Sterilization Pouches EVERY DAY. STEAMPlus Type 5 Sterilization Integrators EVERY WEEK. ConFirm Mail-in Sterilizer Monitoring
More informationJoint Commission In Dental Settings
Joint Commission In Dental Settings Jay Afrow DMD, MHA Surveyor, Ambulatory Health Care The Joint Commission The Joint Commission Disclaimer These slides are current as of 1/19/2018. The Joint Commission
More informationM Manufacturer by: Guide for Cleaning, Sterilization and Storage of Instruments for Caldera Medical Desara Product Family of Implants
Guide for Cleaning, Sterilization and Storage of Instruments for Caldera Medical Desara Product Family of Implants M Manufacturer by: Caldera Medical, Inc. 5171 Clareton Drive Agoura Hills, CA 91301 U.S.
More informationHOGAN & IFbuxrso~ LLP.
HOGAN & IFbuxrso~ LLP. C. STEPHEN LAWRENCE PARTNER (948) 940-0400 CSLAWRENCEBHHLAW. COM April 30, 1999 4675 MA- cow S V m 670 NEWPORT B UM, CA 92660 TEL (949) 250450 FAX (949) 83f0976 Janice Rullo President
More informationSterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects
AAMI/FDS-1 11137-03 2006-01-12 (Revision of ANSI/AAMI/ISO 11137:1994; AAMI TIR27:2001; AAMI/ISO TIR13409:1996) Final draft American National Standard AAMI/ American National Standard NOTE - This document
More informationCATHETER ACCESS KIT. For use with Prometra Programmable Infusion Systems
CATHETER ACCESS KIT Caution: Federal (USA) Law restricts this device to sale by or on the order of a physician. Table of Contents Contents... 3 Description... 3 Indications... 3 Contraindications... 3
More informationCustomized Cranial and Craniofacial Implants. Instructions For Use. 7000revD_Instructions_For_Use. Page 1 of 8
Customized Cranial and Craniofacial Implants Instructions For Use Page 1 of 8 Table of Contents INDICATIONS... 3 DESCRIPTION OF DEVICE... 3 CONDITIONS OF USE... 3 CONTRAINDICATIONS... 3 KELYNIAM GLOBAL
More informationAMENDATORY SECTION (Amending WSR , filed 10/10/95, effective 11/10/95)
AMENDATORY SECTION (Amending WSR 95-21-041, filed 10/10/95, effective 11/10/95) WAC 246-817-601 Purpose. The purpose of WAC 246-817-601 through ((246-817-630)) 246-817-660 is to establish requirements
More informationTABLE OF CONTENTS. PAL-650 Instrument System Instruction Manual. Applicable Part Numbers Introduction General Warnings...
1 NOTES 19 TABLE OF CONTENTS PAL-650 Instrument System Instruction Manual Applicable Part Numbers................................................................... 1-4 Introduction...............................................................................
More informationTable of Contents. Introduction Indications For Use Contraindications Warnings Precautions...5
User Manual 3 Table of Contents Introduction....4 1. Indications For Use...4 2. Contraindications...4 3. Warnings...5 4. Precautions...5 5. Adverse Reactions...5 6. Step-By-Step Instructions...6 A. Contents...6
More informationD. The following factors are of extreme importance to the eventual success of the procedure.
Reprocessed by Stryker Sustainability Solutions 1810 W. Drake Drive Tempe, AZ 85283 sustainability.stryker.com phone: 888.888.3433 English REPROCESSED EXTERNAL FIXATION DEVICES ATTENTION OPERATING SURGEON
More informationProduct Sterility Testing... To Test or Not to Test? That Is the Question
Product Sterility Testing... To Test or Not to Test? That Is the Question Elaine Daniell, Trabue Bryans, Kimbrell Darnell, Joyce Hansen, Victoria M. Hitchins, and Manuel Saavedra Abstract The applications
More informationProper Handling Techniques for Sterilization Wrap and Wrapped OR Trays
CE ONLINE Proper Handling Techniques for Sterilization Wrap and Wrapped OR Trays Continuing Nursing & Allied Health Education Provider Funds Provided By Proper Handling Techniques for Sterilization Wrap
More informationEyewash & Emergency Shower Equipment (Plumbed) Academic Affairs Laboratories of Ferris State University
EYEWASH & EMERGENCY SHOWER SECURED DOCUMENT Academic Affairs Laboratory Safety Eyewash & Emergency Shower Equipment (Plumbed) Academic Affairs Laboratories of Ferris State University Number: AALSD-2-50-1001
More informationPRE-STERILIZED SPINE IMPLANTS
PRE-STERILIZED SPINE IMPLANTS Surgical Technical Technique Report Hip Knee Spine Navigation PRE-STERILIZED SPINE IMPLANTS Technical Report Hip Knee Spine Navigation 1PRE-STERILIZED INTRODUCTION SPINE IMPLANTS
More informationRecommendations for cleaning, disinfection and sterilization of orthopaedic instruments and devices from Swemac.
Recommendations for cleaning, disinfection and sterilization of orthopaedic instruments and devices from Swemac. This document is intended as guidance for decontamination and sterilization for medical
More informationDIRECTIVES. (Text with EEA relevance)
L 238/44 DIRECTIVES COMMISSION DIRECTIVE (EU) 2017/1572 of 15 September 2017 supplementing Directive 2001/83/EC of the European Parliament and of the Council as regards the principles and guidelines of
More informationM6-C Artificial Cervical Disc Surgical Instruments
CARE AND HANDLING INSTRUCTIONS Orthofix Inc. 3451 Plano Parkway Lewisville, Texas 75056-9453 U.S.A. 1-214-937-3199 1-888-298-5700 www.orthofix.com OSI-CustomerService@Orthofix.com Spinal Kinetics LLC,
More informationS5 Rotary Files. only. Manufacturer. Packing unit. Batch code. For professional use only (0)
EN S5 Rotary Files AR Manufacturer 100 Packing unit www.sendoline.com +46 (0)8 445 88 30 Batch code only For professional use only Catalogue number Consult instructions for use Sendoline AB Box 7037, Tillverkarvägen
More informationTechnique Manual Technique Manual Rev. A 01/11/2011
Technique Manual Technique Manual Table of Contents Introduction 4 Indications 5 Instrumentation 6 Frame Assembly 7 Position Guide 8-9 Tips and Tricks 10-13 Cleaning 14 Sterilization 14 Storage Instructions
More informationPeninsula Dental Social Enterprise (PDSE)
Peninsula Dental Social Enterprise (PDSE) Decontamination - Cleaning and Disinfection Version 3.0 Date approved: November 2016 Approved by: The Board Review due: November 2019 Policy will be updated as
More informationISO INTERNATIONAL STANDARD. Sterilization of health care products Radiation Part 2: Establishing the sterilization dose
INTERNATIONAL STANDARD ISO 11137-2 First edition 2006-04-15 Sterilization of health care products Radiation Part 2: Establishing the sterilization dose Stérilisation des produits de santé Irradiation Partie
More informationINFECTION CONTROL IN PRACTICE. Continuing Education
INFECTION CONTROL IN PRACTICE Continuing Education CONTINUING EDUCATION: INFECTION CONTROL IN PRACTICE DESCRIPTION This seminar introduces infection control principals and best practices pertaining to
More informationExperimental Design Process. Things you can change or vary: Things you can measure or observe:
Experimental Design Process Things you can change or vary: Things you can measure or observe: Choosing Variables I will change (independent variable): I will measure (dependent variable): I will not change,
More informationTEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED P18L TYPE USER S MANUAL. from a CURRENT LOOP
TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED from a CURRENT LOOP P18L TYPE USER S MANUAL 1 2 Contents 1. Application... 2 2. Basic requirements, operational safety... 2 3. Installation... 2 3.1. Assembly...
More informationCHAPTER 34 LINN COUNTY DRUG PARAPHERNALIA ORDINANCE
CHAPTER 34 LINN COUNTY DRUG PARAPHERNALIA ORDINANCE TABLE OF CONTENTS 34.1 Definitions 34.2 Possession of Drug Paraphernalia 34.3 Manufacture or Delivery of Drug Paraphernalia 34.4 Effective Date LINN
More informationSterilization of health care products - Radiation - Part 2: Establishing the sterilization dose
AAMI/FDS-1 11137-02 2006-01-12 (Revision of ANSI/AAMI/ISO 11137:1994; AAMI/ISO TIR13409:1996; AAMI/ISO TIR15844:1998) Final draft American National Standard AAMI/ American National Standard NOTE - This
More informationHow Do You Know If a Sterilized Package Remains Sterile?
How Do You Know If a Sterilized Package Remains Sterile? 1 Acknowledgments Bioaerosol and Applied Microbiology Team Brian Heimbuch, MS Group Lead Del Harnish, MS Lead Technical Mike McDonald, MS Lead Engineer
More informationScope This policy applies to all personnel and departments that clean, prepare and/or sterilize items intended for patient care use.
Dental Sterilization Procedures Policy Number VIM4(4)-10 Purpose The purpose of this policy is to ensure patient and employee safety when using instruments with potential for exposure to bloodborne pathogens
More informationDesara TV and Desara Blue TV
Desara TV and Desara Blue TV Sling for Female Stress Urinary Incontinence Instructions For Use D I Prescription Use only Do not reuse Sterilized using ethylene oxide Available Electronically M Manufactured
More informationDisinfection and Decontamination Core Subject. The Process of Instrument Decontamination
Disinfection and Decontamination Core Subject The Process of Instrument Decontamination Aims: To give an overview of the processes involved in dental instrument decontamination using essential quality
More informationVelocity Orthopedics Instrument Cleaning and Sterilization
Velocity Orthopedics Instrument Cleaning and Sterilization It is important to read the Instructions For Use in its entirety prior to using the product. Caution: Federal law (USA) restricts this device
More informationRDHA710FR10A1NK. Neutron and Total Ionizing Dose Test Report. October 2009
Neutron and Total Ionizing Dose Test Report International Rectifier currently does not have a DSCC approved Radiation Hardness Assurance Program for MIL-PRF-38534. 205 Crawford St, Leominster, Ma 01453
More informationSterilization of health care products Radiation. Part 2: Establishing the sterilization dose
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 11137-2 Third edition 2013-06-01 Sterilization of health care products Radiation Part 2: Establishing the sterilization dose Stérilisation des
More informationIN THE HEALTHCARE SETTING, GENERAL INSTRUMENTS REFER TO
CIS Self-Study Lesson Plan Lesson No. CIS 268 (Instrument Continuing Education - ICE) by Kristen Bursey, CRCST Sponsored by: General Instrumentation: An Overview LEARNING OBJECTIVES 1. Discuss the most
More informationPRATICAL GUIDE. Techniques for using, cleaning and sterilizing Riellens products including magnetic resonance imaging MRI
PRATICAL GUIDE Techniques for using, cleaning and sterilizing Riellens products including magnetic resonance imaging MRI Welcome! Here you will find the use, cleaning and sterilization Techniques Guide,
More informationREVISED SEPTEMBER 2017 Commercial Processing Example: Hot Smoked Salmon, Reduced-Oxygen Packed
National Seafood HACCP Alliance for Training and Education REVISED SEPTEMBER 2017 Commercial Processing Example: Hot Smoked Salmon, Reduced-Oxygen Packed Example: Narrative This is a Special Training Model
More informationFor use by an Accredited Orthopaedic Surgeon only
Page 1 of 5 For use by an Accredited Orthopaedic Surgeon only 1. Purpose: External fixators are intended to aid in surgical stabilization following operative procedures to treat fractures, enable correction
More informationTOTAL JOINT ARTHROPLASTY, OR TOTAL JOINT REPLACEMENT, IS
Lesson No. CIS 267 (Instrument Continuing Education - ICE) Sponsored by: by Tyler Sather, CRCST, CIS, CHL Traveling SPD Technician and Casey Czarnowski, BA, CRCST, CIS CS Educator Essentia Health Hospital
More informationMedical Device Conservation
Medical Device Conservation The Science and Safety of Medical Device Reprocessing Think Green help save our planet! Session Description The session describes the history, science and safety of medical
More informationReprocessing of Implants: What are the Issues?
Reprocessing of Implants: What are the Issues? Dr. Michelle J. Alfa, Ph.D., FCCM Medical Director, Clinical Microbiology, Diagnostic Services of Manitoba, Winnipeg, Canada Surgical Instrument Sets: When
More informationWHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: INSTI HIV-1/HIV-2 Antibody Test Number: PQDx
WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: INSTI HIV-1/HIV-2 Antibody Test Number: PQDx 0002-002-00 Abstract INSTI HIV-1/HIV-2 Antibody Test with product codes, 90-1013, 90-1010,
More informationHealthcare Sterilisation: Challenging Practices Volume 2
Contents 1 Traditional Chemical Sterilisation Methods... 1 1.1 Ethylene Oxide Sterilisation... 3 1.1.1 Green sterilisation (Benefits versus Risks)... 4 1.1.2 Ethylene Oxide Residuals... 8 1.1.3 Advantages
More informationInstruction Manual Aerogen, Inc. Part No. AG-AL1010 Rev. C
Instruction Manual 2004 Aerogen, Inc. Part No. AG-AL1010 Rev. C Table of Contents Introduction... 3 System description... 4 Warnings... 5 Cautions... 5 Electromagnetic Susceptibility... 5 Symbols... 6
More informationHow Do You Know If Your Sterilized Instruments Remain Sterile?
How Do You Know If Your Sterilized Instruments Remain Sterile? Disclosure Harry Shaffer is the first author of the sterility maintenance study covered in this presentation and has been compensated by Halyard
More informationcgmp (21 CFR 111) Regulation and Compliance Overview
cgmp (21 CFR 111) Regulation and Compliance Overview Neogen Effective Compliance Seminar September 23, 2014 Michael McGuffin President, American Herbal Products Association mmcguffin@ahpa.org Regulation
More informationHealthcare Sterilisation: Introduction and Standard Practices, Volume 1
Contents 1. Overview of Sterilisation and Background... 1 1.1 Magic Wand... 1 1.1.1 Definition of Sterilisation... 2 1.1.2 Detectives or Forensic Scientists... 3 1.1.3 Idealised Process... 6 1.1.4 Lesser
More informationEffectiveness of Decontamination of Laboratory Room Surfaces with Low Concentrations of Hydrogen Peroxide using Atmospheric Cold Plasma Activation
Effectiveness of Decontamination of Laboratory Room Surfaces with Low Concentrations of Hydrogen Peroxide using Atmospheric Cold Plasma Activation Miguel A. Grimaldo, M.Eng. Assistant Professor Department
More informationPractical Application of Basic Sterilization Principles
Practical Application of Basic Sterilization Principles By DORIS C. MacCLELLAND, RN, MS Commander, United States Navy Nurse Corps The opinions or assertions contained herein are those of the author and
More informationAMENDATORY SECTION (Amending WSR , filed 10/10/95, effective. WAC Purpose. The purpose of WAC through
AMENDATORY SECTION (Amending WSR 95-21-041, filed 10/10/95, effective 11/10/95) WAC 246-817-601 Purpose. The purpose of WAC 246-817-601 through ((246-817-630)) 246-817-660 is to establish requirements
More informationThis is the author s final accepted version.
Winter, S., Smith, A., Lappin, D., McDonagh, G. and Kirk, B. (2017) Failure of non-vacuum steam sterilization processes for dental handpieces. Journal of Hospital Infection, 97(4), pp. 343-347. (doi:10.1016/j.jhin.2017.09.004)
More informationROBOTIC-ASSISTED SURGERY REPRESENTS THE USE OF TECHNOLOGY
CIS Self-Study Lesson Plan Lesson No. CIS 265 (Instrument Continuing Education - ICE) Sponsored by: by Casey Czarnowski, BA, CRCST, CIS CS Educator Essentia Health Hospital Fargo, N.D. Safe Reprocessing
More informationSample Enhanced Visual Inspection of medical devices. SUBJECT: Enhanced Visual Inspection of medical devices
Sample Enhanced Visual Inspection of medical devices SUBJECT: Enhanced Visual Inspection of medical devices DEPARTMENT: Sterile Processing Area** APPROVED BY: EFFECTIVE: REVISED: 9/2017 PURPOSE: The purpose
More informationEX-PRESS Glaucoma Filtration Device Surgical Procedure
EX-PRESS Glaucoma Filtration Device Surgical Procedure Surgical Recommendations EX-PRESS Glaucoma Filtration Device preloaded device (plus 1 back-up) Sterile caliper Lid speculum Corneal forceps 0.12 forceps
More informationMetallic Internal Fixation Devices The following languages are included in this packet:
Metallic Internal Fixation Devices 150848-0 The following languages are included in this packet: English (en) Deutsch (de) Nederlands (nl) Français (fr) Español (es) Italiano (it) Português (pt) 中文 - Chinese
More informationParametric Release Regulatory Requirements. Terry E. Munson Technical Vice President PAREXEL International, LLC
Parametric Release Regulatory Requirements Terry E. Munson Technical Vice President History First application approved for parametric release in 1985 Formed the basis for the FDA Compliance Policy Guide
More informationTEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED P18L TYPE USER S MANUAL. from a CURRENT LOOP
TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED from a CURRENT LOOP P18L TYPE USER S MANUAL Contents 1. Application... 2 2. Basic requirements, operational safety... 2 3. Installation... 2 3.1. Assembly...
More informationIssue : Infection Control in the Dental Practice. Part 2
Issue 4 2018: Infection Control in the Dental Practice. Part 2 Compiled by Mrs Carin Brent Dipl OH (US) EF(UWC) 3 Clinical CEU s Sterilization and Disinfection of Patient-Care Items Patient-care items
More informationMedidée Services SA. Nano-Tera.ch. 05 February 2015 part 4. Intro ISO GMP - GLP Pierre-Alain Sommer
Nano-Tera.ch 05 February 2015 part 4 Intro ISO GMP - GLP Pierre-Alain Sommer Pierre-alain.sommer@medidee.com www.medidee.com Nano-Tera 2015 05.02.2015 Introduction to ISO 13485, cgmp s and GLP s Context
More informationMedical/Surgical Asepsis. Presented by: Cynthia Bartlau, RN, PHN, MSN
Medical/Surgical Asepsis Presented by: Cynthia Bartlau, RN, PHN, MSN Objectives At the completion of this lesson the student will be able to Discuss concepts related to the chain of infection Differentiate
More informationTo standardize wound care and prevent infection in compromised patients who have a Berlin Heart Ventricular Assist Device (VAD).
PURPOSE To standardize wound care and prevent infection in compromised patients who have a Berlin Heart Ventricular Assist Device (VAD). POLICY STATEMENTS Dressing change should be done no sooner than
More informationMedGyn Gellhorn Pessary. MedGyn Cup Pessary. Products, Inc. Products, Inc.
MedGyn Ring Pessary Device Description: MedGyn Ring Pessary is ring-shaped with a solid core and an indentation on each side. The Ring Pessary size ranges from 51mm to 102mm and is available with support
More informationCHAPTER 1 GENERAL INFORMATION
BUMEDINST 6600.10 DENTAL INFECTION CONTROL PROGRAM CHAPTER 1 GENERAL INFORMATION Job Classifications with occupational exposure 1. Housekeeping 2. Laundry 3. Maintenance ( Radiology, Pros Lab, DTR, sterilization
More informationAesculap Spine Instructions for use
Aesculap Spine S 4 Spondylolisthesis Reduction Instrument (SRI) Instructions for use S 4 Spondylolisthesis Reduction Instrument (SRI) 5 4 3 2 Fig. 1 1 9 10 11 8 6 7 Fig. 2 Aesculap Spine S 4 Spondylolisthesis
More informationProducts, Inc. w/ & w/o Support R#1 R#2 R#3 R#4 R#5. inch mm R#6 R#7 R#8 R#9. inch
EVALUATING THE FIT: Patient should be able to stand and bear down slightly without dislodging Pessary. Expulsion or turning of Pessary may indicate that Pessary size is too small. Therefore, the next larger
More informationToronto Public Health HCV Outbreak Investigation Ontario Endoscopy Clinic 1315 Finch Avenue West, Suite 302, Toronto Final Report April 24, 2015
Toronto Public Health HCV Outbreak Investigation Ontario Endoscopy Clinic 1315 Finch Avenue West, Suite 302, Toronto Final Report April 24, 2015 Initial Case Identification On May 24, 2013, a 60 year old
More informationCHARLOTTE. 7.0 Multi-Use Compression Screw System SURGICAL TECHNIQUE
CHARLOTTE 7.0 Multi-Use Compression Screw System SURGICAL TECHNIQUE Contents Chapter 1 1 Chapter 2 2 2 3 4 Appendix A 5 6 Appendix B 7 8 Preoperative Planning Surgical Technique Subtalar Arthrodesis Guide
More informationPreparation Instructions for the CAMLOG /CONELOG Implant System
J8000.0032 Rev.6 EN 06/2015 Preparation Instructions for the CAMLOG /CONELOG Implant System ENGLISH The following descriptions contain detailed instructions on cleaning, disinfection and sterilization
More informationICD-10-CM & ICD-10-PCS CODING ROUNDTABLES. Emergency Room Scenarios
ICD-10-CM & ICD-10-PCS CODING ROUNDTABLES Emergency Room Scenarios THANK YOU TO OUR CODING ROUNDTABLE SPONSOR CHANNEL PUBLISHING! ICD-10-CM: Diagnosis Coding Chapter 19 Injury, poisoning and certain other
More informationSTERILISATION PACKAGING CATALOGUE 2017 AUSTRALIA & NEW ZEALAND
STERILISATION PACKAGING CATALOGUE 2017 AUSTRALIA & NEW ZEALAND QUICK LINKS The HALYARD* Sterilization Wrap Advantage page 3 Why Double-Wrapping is Imperative for Optimal Barrier Protection page 6 SEQUENTIAL
More informationISO INTERNATIONAL STANDARD. Male condoms Requirements and test methods for condoms made from synthetic materials
INTERNATIONAL STANDARD ISO 23409 First edition 2011-02-15 Male condoms Requirements and test methods for condoms made from synthetic materials Préservatifs masculins Exigences et méthodes d'essai pour
More informationUnidirectional Airflow Smoke Test Protocol
Unidirectional Airflow Smoke Test Protocol Persons signing as approvers shall have examined the Unidirectional Airflow Protocol. This approval shall signify that testing of operational parameters is adequate
More informationMETAL HEMI SYSTEM
METAL HEMI SYSTEM 152206-0 The following languages are included in this packet: English (en) Deutsch (de) Nederlands (nl) Français (fr) Español (es) Italiano (it) Português (pt) 中文 - Chinese (sch) Türkçe
More informationOBSOLETE. Visit for the latest version. Ankle Plating System 3 PKGI-79-B EFFECTIVE
Ankle Plating System 3 MediMark Europe Sarl. 11 rue Emile ZOLA. BP 2332 38033 GRENOBLE CEDEX 2 FRANCE +33.4.76.86.43.22 Acumed LLC 5885 NW Cornelius Pass Road Hillsboro, OR 97124-9432 +1.503.627.9957 acumed.net
More informationA Summary of Knee Recalls Consumers Union Safe Patient Project September 9, 2013
A Summary of Knee Recalls Consumers Union Safe Patient Project September 9, 213 An estimated 4.4 million Americans are living with knee implants. In 211, over 711, 1 knee replacements were performed primarily
More informationI will tell you that when we left, we were just physically kind of sick. Susan Rogers, executive director of the Oklahoma Board of Dentistry
1 2 3 Infection Control Conundrums Moving Towards New Solutions Michael Joseph, DDS MSD VA Dental Infection Control Consultant Michael.Joseph2@va.gov Disclosure I receive no money and have no financial
More informationAKA Good Manufacturing Practice (GMP) Certification Program
AKA Good Manufacturing Practice (GMP) Certification Program Preamble The American Kratom Association (AKA) is establishing this program to assure the safety and integrity of kratom dietary supplements
More informationBody Contouring Implants - Calf
Body Contouring Implants - Calf The body you were born with may or may not have the musculature you desire and it may be difficult to improve upon certain areas with exercise alone. In some cases, damage
More informationAre we ready for the switch over?
Are we ready for the switch over? More info about this article: http://www.ndt.net/?id=21194 Sadasivan Narayana Materials Evaluation Services (Mumbai) Pvt. Ltd. sadasivan.n24@gmail.com Abstract: The increasing
More informationORIFICE FLOW METERS WITH CHAMBERS MEASURING RINGS
Bulletin E79 rev0 02/03/03 ORIFICE FLOW METERS WITH CHAMBERS MEASURING RINGS POP-U-S SERIES FEATURES UNI PN6 flanges: iron Chamber measuring rings: iron Disc: AISI30 Max. operating temperature: 00 C Max.
More informationChapter 11 How to state your conditions clearly to the doctor
Chapter 11 How to state your conditions clearly to the doctor Chapter 11 How to state your conditions clearly to the doctor While seeing doctor, details of your statement should include the following:
More informationPlease note: Forms with PXXXX designations can be found at All other forms can be found at
ADULT SAFETY NET (ASN) PROGRAM 2018 COMPLIANCE SITE VISIT FOLLOW-UP PLAN This document is used to track required follow-up actions at the completion of an ASN compliance site visit. When all follow-up
More informationCryptosporidium Proficiency Test Program for Laboratories using US EPA Method 1623: Cryptosporidium and Giardia
Cryptosporidium Proficiency Test Program for Laboratories using US EPA Method 1623: Cryptosporidium and Giardia in Water by Filtration/IMS/FA, December 2005 version Preferable Specifications to Establish
More informationPUBLIC HEALTH AND SAFETY
CHAPTER 41 PUBLIC HEALTH AND SAFETY 41.01 Drug Paraphernalia 41.04 Throwing and Shooting 41.02 Fireworks 41.05-41.14 (Reserved) 41.03 Discharge of Weapons 41.15 Urinating and Defecating 41.01 DRUG PARAPHERNALIA.
More informationDISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY
0086 Instructions for Use RCS Anterior Buttress Plate System Caution: Federal (USA) law restricts this device to sale by or on the order of a physician DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY There
More informationCTO Establishment Inspections. an Atlantic update - Tissues
Transplant Atlantic CTO Establishment Inspections an Atlantic update - Tissues October 14, 2010 Compliance & Enforcement Deliver a national compliance and enforcement program, pursuant of the Food & Drugs
More information