Clinical experience with a new thoracic stent graft system (Ankura TM ). Procedural analysis, 30 days and 6-month results in a single center study
|
|
- May Wilcox
- 5 years ago
- Views:
Transcription
1 Clinical experience with a new thoracic stent graft system (Ankura TM ). Procedural analysis, 30 days and 6-month results in a single center study Theodoros Kratimenos, MD Consultant, Interventional Radiologist Interventional Radiology Unit EVANGELISMOS General Hospital of Athens, Greece
2 Disclosure Speaker name: Theodoros Kratimenos I have the following potential conflicts of interest to report: Consulting (Bolton, Lifetech, Vascutek) Employment in industry Stockholder of a healthcare company Owner of a healthcare company Other(s) I do not have any potential conflict of interest
3 ANKURA TM THORACIC STENT GRAFT FEATURES The Ankura Thoracic Stent Graft is designed for endovascular treatment of descending thoracic aorta diseases such as: type B dissection, Aneurysm, Dissecting aneurysm, penetrating aortic ulcer, intramural hematoma, traumatic rupture, etc. It reinforces the vascular wall of thoracic aorta from rupture. 1. e-ptfe Dual Membrane (Excellent biocompatibility and durability) 2. No suture on the main body (Avoid pinhole leakage -Type Ⅳ endoleak) 3. Sinusoidal shape Self-expanding nitinol springs -(provide flexibility and conformability exert active radial force to enhance seal and fixation) 4. Kink-resistant delivery system with Hydrophilic Coating - (facilitates stent graft delivery) 5. The first nitinol spring strut has smaller diameter -(in order to provide enhanced apposition to the aortic wall, decreased endoleak risk and prevent migration). 6. Proximal Bare Stents which are embedded in the graft- (less traumatic)
4 ANKURA TM THORACIC STENT GRAFT FEATURES Tip capture system (provides controlled deployment and precise placement in the thoracic aorta). Proximal releaser Front handle Safe buckle Tip capture release in 2 movements: remove the safe buckle (green), pull back the tip releaser (black), use left hand to fix the front handle 2 radiopaque markers at the prox. Part of the graft (figure 8 marker at the greater curvature and figure 0 marker at the lesser curvature), Longitudinal supporting bar on the greater curvature (helps to avoid stent shortening, providing axial support to the nitinol springs. No bar at the first and last nitinol spring strut). (21-24) F OD
5 ANKURA TM THORACIC STENT GRAFT FEATURES Easy and operator friendly deployment process: Deploy Slowly for maximum control keeping fixed with the left hand the first handle and rotating with the right hand the second handle. Straight grafts: 20-46mm diameter mm length Tapered grafts: 4,6, and 8 mm diameter difference, mm length IFU Adequate femoral/iliac access Fast deployment: Pull back the blue safety valve and withdraw fast the right handle keeping fixed the left one Proximal landing Zone: 15mm length, 18-42mm width Distal landing Zone: 15mm, 18-42mm width
6 METHODS: From February 2014 till June 2017, 83 patients, 62 male, 21 female (average age 64,7 years, max 85-min 32) received endovascular treatment with the Ankura Stent Graft device at a single center (Evangelismos Hospital) in Athens, Greece, for diseases of thoracic aorta. Informed consent from each patient and Institutional Review Board permission were obtained. A chest and abdominal Computed Tomography Angiography (CTA) was performed in every patient before the endovascular procedure. All post-cta images were transferred into a dedicated software. A multidisciplinary team, including interventional radiologists, cardiac and vascular surgeons discussed and planned each case on an individual basis. All TEVAR interventions were performed in the Angio Suite A femoral artery cut-down was performed on the device access site We did not routinely used spinal drainage before TEVAR (however drainage was readily available in cases of previous abdominal aortic aneurysm repair, when coverage of LT subclavian artery was planned, or when extensive coverage of thoracic aorta by the stent graft was required).
7 Table I Number of patients Average n o of devices used Pathology treated 83 1,36 Degenerative aneurysms 30 1,5 Type B dissection 22 1,22 Dissecting aneurysms 13 1,5 Traumatic pseudoaneurysm 7 1 Penetrating Aortic Ulcer (PAU)+Intramural Heamatoma (IMH) 5 1,2 PAU 3 1,3 IMH 2 1 Isthmic Stenosis 1 1
8 RESULTS: INTRAPROCEDURAL DATA During operation there was no major complication or death. Successful Implantation in 81/83 patients (97,6%) 2 cases demanded the implantation of a surgical conduit in the common iliac arteries through which finally passed the device. In one single case the device presented mechanical damage of the delivery sheath, due to extreme tortuosity of descending thoracic aorta, where a different company s device also did not succeded to be released. In one case (under hemodialysis) was not possible to advance the delivery system of the device through the iliac art. (rigid and tortuous). In all other cases the stent graft was easily and precisely released with no migration.
9 Clinical case: 77y/o male with an 8 cm diameter descending thoracic aorta aneurysm, Pre Operation CT: Aortic Arch with a normal caliber, Descending Thoracic Aorta Aneurysm with mural thrombus starting a few centimeters after the origin of Lt Subcl. Art., Atelectasis of the lt pulmonary lower lobe and pleural effusion 6 months post op. CT: Exclusion of the aneurysm, no proximal or distal Type Endoleak, No Atelectasis, No Pleural Effusion
10 Table 2. Operation-related characteristics Operation-related characteristics Clinical symptomatology elective repair (n, %) 42 50,6% symptomatic (n, %) 22 27% emergency (n, %) 19 23% Type of prior aortic treatment none (n, %) 77 92,8% surgical repair of type A aortic dissection (n, %) 4 4,8% surgical repair for AAA (n, %) NONE EVAR (n, %) 2 2,4% Type of anaesthesia used local /IV sedation (n, %) 81 97,6% general (n, %) 2-2,4% regional (n, %) NONE Coverage of LSA (n, %) Need for carotid-subcl. Bypass (n, %) 9 10,8% 1 1,2% Fluoroscopy time (min) (mean±sd) 13 ± 6,2 DAP (dgy cm2) (mean±sd) 313,08±228,2 Contrast administered (ml) (mean±sd) 120 ± 55,44 Blood loss (ml) (mean±sd) 172 ±149,4 Need for Transfusion (n, %) 1 1,2 % Length of hospital stay (days) (mean±sd) 4,25 ± 2,93
11 Clinical case: Male, 76 years old with acute thoracic pain CT scan images: PAU+IMH in descending Thor. Ao 1 MONTH POST TEVAR CT Clinical case : Female 34 years old, bilateral lower limb bone fractures and pelvis fracture, post car accident, Pre op. MRA, DSA, CT: 6-Months f-up
12 Table day and 6 months outcomes 30-day Outcomes 6 months Outcomes Access site complications (n, %) 10 9,6% Endoleak (n, %) type I prox. 2 2,4% 1 1,2% Type I distal 2 2,4% 1 1,2% type II 3 3,6% 5 6% Type III NONE 1 1,2% Type IV NONE NONE Stroke / TIA (cerebral complications) (n, %) 1 1,2% / 2 2,4% 1 1,2% MI (cardiac complications) (n, %) 1 1,2% 2 2,4% Paralysis / Paraparesis (n, %) 1 1,2% / 3 3,6%* NONE/ 1 1,2% Pulmonary complications (n, %) 2 2,4% NONE Renal failure (n, %) NONE NONE Retro type A dissection (n, %) NONE NONE Re-intervention 1 1,2% (Type Ia endoleak) Aorta related Death (n, %) NONE NONE All cause death (n, %) 2 2,4 % 5 6% In-hospital Death (n, %) 1 1,2 % 4 4,8% Graft migration 0 1 1,2% Graft thrombosis 0 0 Graft infection 0 0 Conversion to open surgery 0 0 *All cases treated with lumbar drainage and medications Paraparesis cases did not leave permanent sequela The case with paralysis became permanent paraparesis 3 3,6% (2 Type Ib, and 1 III endoleak)
13 Clinical case: 73yo male, TEVAR 18 months ago, now symptomatic aneurysm 7,5cm of diameter in the distal aortic arch, because of type Ia endoleak, with no prox. landing zonesurgical debranching of aortic arch vessels was performed in order to expand the prox. landing zone
14 CONCLUSION In this single center clinical study the Ankura Thoracic Stent Graft system, has demonstrated: high intra- procedural technical success, very satisfactory 30 days and 6 months clinical outcome, showing a promising capacity to treat a variety of thoracic aortic diseases. Further studies and longer follow ups are needed.
15 1884 Evaggelismos General Hospital of Athens/Greece 1200 beds >65 standard TEVAR cases/ year >60 standard EVAR cases / year 1934 In addition on 2017: 4 scalloped TEVAR cases (prox. or distal) 3 BR-EVAR+TEVAR (thoraco-abdominal aneurysm cases) 3 F-EVAR (3-4 fenestrations, Pararenal aneurysm cases) 2 IBD EVAR cases (1 single 1 double) 3 Ch-EVAR cases (1 single and 2 double chimneys) new laboratories building THE COMPANY AKTOR in Brief History / Milestones Strategy Management Locations AKTOR's Main Subsidiaries Health & Safety Total Quality Management ACTIVITIES Construction Solar Power Mining Quarrying Facility Management Project Management PROJECTS Building Projects Infrastructure Projects M.E.P. Projects Industrial Projects NEWS Newsletters FINANCIAL DATA Key Financial Figures Summary Financial Data Financial Statements - Figures and Information according to IFRS Investors Update (till Dec. 2005) SUSTAINABILITY Corporate Social Responsibility Environment Economic Responsibility Relationships with Interested Parties Strategic Objective Sustainability Report & Environmental Statement per year Other Reports Events CAREERS Working at AKTOR Our People Job Search 2016 new OR building Search OK Home Page Links Access / Contact us ΕΛΛΗΝΙΚΑ Search OK THE COMPANY AKTOR in Brief History / Milestones Strategy Management Locations AKTOR's Main Subsidiaries Health & Safety Total Quality Management
16 Clinical experience with a new thoracic stent graft system (Ankura TM ). Procedural analysis, 30 days and 6-month results in a single center study Theodoros Kratimenos, MD Consultant, Interventional Radiologist Interventional Radiology Unit EVANGELISMOS General Hospital of Athens, Greece
Challenges. 1. Sizing. 2. Proximal landing zone 3. Distal landing zone 4. Access vessels 5. Spinal cord ischemia 6. Endoleak
Disclosure I have the following potential conflicts of interest to report: Consulting: Medtronic, Gore Employment in industry Stockholder of a healthcare company Owner of a healthcare company Other(s)
More informationSanti Trimarchi, MD, PhD Vascular Surgeon Thoracic Aortic Research Center, Director IRCCS Policlinico San Donato University of Milan
The Gore GREAT Registry: Update about a real life data collection Santi Trimarchi, MD, PhD Vascular Surgeon Thoracic Aortic Research Center, Director IRCCS Policlinico San Donato University of Milan Disclosures
More informationPerformance of the conformable GORE TAG device in Type B aortic dissection from the GORE GREAT real world registry
University of Milan Thoracic Aortic Research Center Performance of the conformable GORE TAG device in Type B aortic dissection from the GORE GREAT real world registry Santi Trimarchi, MD, PhD Associate
More informationEarly Clinical Results with the Valiant Mona LSA Branch Stent-Graft
Early Clinical Results with the Valiant Mona LSA Branch Stent-Graft Frank R. Arko III, MD Professor of Cardiovascular Surgery Director, Endovascular Surgery Co-Director, Aortic Institute Carolinas Medical
More informationIntroducing the GORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System
Introducing the GORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System K. Oikonomou Department of Vascular Surgery Regensburg University Disclosures W.L. Gore & Associates Sponsored Presentation
More informationDevelopment of a Branched LSA Endograft & Ascending Aorta Endograft
Development of a Branched LSA Endograft & Ascending Aorta Endograft Frank R. Arko III, MD Sanger Heart & Vascular Institute Carolinas Medical Center Charlotte, North Carolina, USA Disclosures Proximal
More informationAortic stents, types, selection, tricks in deployment.
Aortic stents, types, selection, tricks in deployment. Hamdy Soliman.M.D,FSCAI Consultant of Cardiology&Head of Endovascular Unit National Heart Institute Endovascular Treatment of Thoracic Aortic Aneurysms
More informationPercutaneous Approaches to Aortic Disease in 2018
Percutaneous Approaches to Aortic Disease in 2018 Wendy Tsang, MD, SM Assistant Professor, University of Toronto Toronto General Hospital, University Health Network Case 78 year old F Lower CP and upper
More informationArch Repair with the Bolton Medical RelayBranch Thoracic Stent-graft system: Multicenter experience
Arch Repair with the Bolton Medical RelayBranch Thoracic Stent-graft system: Multicenter experience Joost van Herwaarden, MD, PhD University Medical Center, Utrecht Disclosure I have the following potential
More informationAbdominal and thoracic aneurysm repair
Abdominal and thoracic aneurysm repair William A. Gray MD Director, Endovascular Intervention Cardiovascular Research Foundation Columbia University Medical Center Abdominal Aortic Aneurysm Endografts
More informationRole of Gender in TEVAR and EVAR results from the GREAT registry
Role of Gender in TEVAR and EVAR results from the GREAT registry Mauro Gargiulo Vascular Surgery University of Bologna - DIMES Policlinico S.Orsola-Malpighi Bologna, Italy mauro.gargiulo2@unibo.it Disclosure
More informationRobert F. Cuff, MD FACS SHMG Vascular Surgery
Robert F. Cuff, MD FACS SHMG Vascular Surgery Objectives To become familiar with the commercially available fenestrated EVAR graft Discuss techniques to increase success Review available data to determine
More informationNeurological Complications of TEVAR. Frank J Criado, MD. Union Memorial-MedStar Health Baltimore, MD USA
ISES Online Neurological Complications of Frank J Criado, MD TEVAR Union Memorial-MedStar Health Baltimore, MD USA frank.criado@medstar.net Paraplegia Incidence is 0-4% after surgical Rx of TAAs confined
More informationDurability of The Endurant Stent-Graft through 5 Years
Durability of The Endurant Stent-Graft through 5 Years Michel S. Makaroun MD Co-Director, UPMC Heart and Vascular Institute Professor and Chair, Division of Vascular Surgery University of Pittsburgh School
More informationEVAR and TEVAR: Extending Their Use for Rupture and Traumatic Injury. Conflict of Interest. Hypotensive shock 5/5/2014. none
EVAR and TEVAR: Extending Their Use for Rupture and Traumatic Injury Bruce H. Gray, DO MSVM FSCAI Professor of Surgery/Vascular Medicine USC SOM-Greenville Greenville, South Carolina none Conflict of Interest
More informationMy personal experience with INCRAFT in standard and challenging cases
My personal experience with INCRAFT in standard and challenging cases G Pratesi, MD Vascular Surgery University of Rome Tor Vergata giovanni.pratesi@uniroma2.it Disclosure Speaker name: Giovanni Pratesi,
More informationCUSTOM-MADE SCALLOPED THORACIC ENDOGRAFTS IN DIFFERENT HOSTILE AORTIC ANATOMIES
CUSTOM-MADE SCALLOPED THORACIC ENDOGRAFTS IN DIFFERENT HOSTILE AORTIC ANATOMIES A SERIES OF THREE CASE REPORTS Joel Sousa Department of Department of Angiology and Vascular Surgery Hospital S. João, Porto,
More informationTalent Abdominal Stent Graft
Talent Abdominal with THE Xcelerant Hydro Delivery System Expanding the Indications for EVAR Treat More Patients Short Necks The Talent Abdominal is the only FDA-approved device for proximal aortic neck
More informationTitle. Different arch branched devices are available, is morphology the. main criteria of choice? Ciro Ferrer, MD
Different arch branched devices are available, is morphology the Title main criteria of choice? Ciro Ferrer, MD Vascular Surgery Unit Sapienza University of Rome Disclosure Speaker name: Ciro Ferrer Proctoring/speaking
More informationOptimizing Accuracy of Aortic Stent Grafts in Short Necks
Optimizing Accuracy of Aortic Stent Grafts in Short Necks Venkatesh Ramaiah, MD, FACS Medical Director Arizona Heart Hospital Director Peripheral Vascular and Endovascular Research Arizona Heart Institute
More informationGORE EXCLUDER AAA Endoprosthesis demonstrates long-term durability. Michel Reijnen Rijnstate Hospital Arnhem, The Netherlands
GORE EXCLUDER AAA Endoprosthesis demonstrates long-term durability Michel Reijnen Rijnstate Hospital Arnhem, The Netherlands Disclosure Speaker name: Michel Reijnen I have the following potential conflicts
More informationCOMPLICATIONS OF TEVAR
COMPLICATIONS OF TEVAR P. Bergeron, A.Petrosyan, F.Markatis, T.Abdulamit, J.-C. Trastour IMAD CONGRESS 2010 Liège Belgium BACKGROUND Stentgrafting is a recognized treatment for TAA & TAD and has been proposed
More informationHow to achieve a successful proximal sealing in TEVAR? Pr L Canaud
How to achieve a successful proximal sealing in TEVAR? Pr L Canaud CHU de Montpellier France Disclosure I have the following potential conflicts of interest to report: Consulting: Medtronic. Proximal neck
More informationAnatomical challenges in EVAR
Anatomical challenges in EVAR M.H. EL DESSOKI, MD,FRCS PROFESSOR OF VASCULAR SURGERY CAIRO UNIVERSITY Disclosure Speaker name:... I have the following potential conflicts of interest to report: Consulting
More informationJean M Panneton, MD Professor of Surgery Program Director Vascular Surgery Chief EVMS. Arch Pathology: The Endovascular Era is here
Jean M Panneton, MD Professor of Surgery Program Director Vascular Surgery Chief EVMS Arch Pathology: The Endovascular Era is here Disclosures Consultant: Cook Medical, Bolton Medical, Medtronic Inc, Volcano,
More informationEndovascular Management of Thoracic Aortic Pathology Stéphan Haulon, J Sobocinski, B Maurel, T Martin-Gonzalez, R Spear, A Hertault, R Azzaoui
Endovascular Management of Thoracic Aortic Pathology Stéphan Haulon, J Sobocinski, B Maurel, T Martin-Gonzalez, R Spear, A Hertault, R Azzaoui Aortic Center, Lille University Hospital, France Disclosures
More informationMODERN METHODS FOR TREATING ABDOMINAL ANEURYSMS AND THORACIC AORTIC DISEASE
MODERN METHODS FOR TREATING ABDOMINAL ANEURYSMS AND THORACIC AORTIC DISEASE AAA FACTS 200,000 New Cases Each Year Ruptured AAA = 15,000 Deaths per Year in U.S. 13th Leading Cause of Death 80% Chance of
More informationAcute dissections of the descending thoracic aorta (Debakey
Endovascular Treatment of Acute Descending Thoracic Aortic Dissections Nimesh D. Desai, MD, PhD, and Joseph E. Bavaria, MD Acute dissections of the descending thoracic aorta (Debakey type III or Stanford
More informationZenith Alpha T HORACIC ENDOVASCULAR GRAFT
Device description Zenith Alpha T HORACIC ENDOVASCULAR GRAFT www.cookmedical.com AI-D21181-EN-F Modular design The two-piece modular system allows the physician to customize a graft system to fit each
More informationHome-made Fenestrations for Various Pathologies of Abdominal Aorta
Home-made Fenestrations for Various Pathologies of Abdominal Aorta Chen, Jer-Shen Director of Vascular Surgery, Far Eastern Memorial Hospital Clinical Assistant Professor, National Yang-Ming University
More informationTHE THE MORE MORE NATURAL APPROACH TO OPTIMAL FIT
THE THE MORE MORE NATURAL APPROACH Natural Approach TO Optimal Fit TO OPTIMAL FIT Conformability without Compromise THE STANDARD IN Conformability AND Designed for flexibility and conformability in tortuous
More informationYoung-Guk Ko, M.D. Severance Cardiovascular Hospital, Yonsei University Health System,
Young-Guk Ko, M.D., Dangas G, J Am Coll Cardiol Intv 2012;5:1071 All-cause Mortality Dangas Severance G, J Am Coll Cardiovascular Cardiol Intv Hospital, 2012;5:1071 Yonsei University Health System Aneurysm-related
More informationLow profile TEVAR: is it an added value? Michel Bosiers, G. Torsello Münster
Low profile TEVAR: is it an added value? Michel Bosiers, G. Torsello Münster Disclosure Speaker name:...michel Bosiers... I have the following potential conflicts of interest to report: Consulting Employment
More informationFROM THE EVERYDAY TO THE EXTRAORDINARY
FROM THE EVERYDAY TO THE EXTRAORDINARY Created with the collaboration of more than 250 physicians around the world, ENDURANT empowers you to create stronger outcomes for more patients, including those
More informationOptimal repair of acute aortic dissection
Optimal repair of acute aortic dissection Dept. of Vascular Surgery, The 2nd Xiang-Yale Hospital, Central-South University, China Hunan Major Vessels Diseases Clinical Center Chang Shu Email:changshu01@yahoo.com
More informationAccess More Patients. Customize Each Seal.
Access More. Customize Each Seal. The Least Invasive Path Towards Proven Patency ULTRA LOW PROFILE TO EASE ADVANCEMENT The flexible, ultra-low 12F ID Ovation ix delivery system enables you to navigate
More informationThoracic Endovascular Aortic Repair (TEVAR) Indications and Basic Procedure
Thoracic Endovascular Aortic Repair (TEVAR) Indications and Basic Procedure Tilo Kölbel, MD, PhD University Heart Center Hamburg University Hospital Eppendorf Disclosure Speaker name: Tilo Kölbel, MD I
More informationENCORE, a Study to Investigate the Durability of Polymer EVAR with Ovation A Contemporary Review of 1296 Patients
ENCORE, a Study to Investigate the Durability of Polymer EVAR with Ovation A Contemporary Review of 1296 Patients The Ovation System is approved to treat infrarenal abdominal aortic aneurysms and is not
More informationGoing the distance: Conformable Gore TAG device demonstrates durable outcomes over the long. D.Böckler University Hospital Heidelberg, Germany
Going the distance: Conformable Gore TAG device demonstrates durable outcomes over the long D.Böckler University Hospital Heidelberg, Germany Disclosures Speaker name: Dittmar Böckler I have the following
More informationA New EVAR Device for Infrarenal AAAs
A New EVAR Device for Infrarenal AAAs Peter Nelson, MD, MS Assistant Professor of Surgery MM0203 Rev. 01 Current U.S. EVAR Devices Anatomical Fixation Proximal Fixation Powerlink - Endologix Excluder WL
More informationBifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully
Physician Training Bifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully supported by self-expanding z-stents H&L-B
More informationExperience of endovascular procedures on abdominal and thoracic aorta in CA region
Experience of endovascular procedures on abdominal and thoracic aorta in CA region May 14-15, 2015, Dubai Dr. Viktor Zemlyanskiy National Research Center of Emergency Care Astana, Kazakhstan Region Characteristics
More informationManagement of Acute Aortic Syndromes. M. Grabenwoger, MD Dept. of Cardiovascular Surgery Hospital Hietzing, Vienna, Austria
Management of Acute Aortic Syndromes M. Grabenwoger, MD Dept. of Cardiovascular Surgery Hospital Hietzing, Vienna, Austria I have nothing to disclose. Acute Aortic Syndromes Acute Aortic Dissection Type
More informationIntravascular Ultrasound in the Treatment of Complex Aortic Pathologies. Naixin Kang, M.D. Vascular Surgery Fellow April 26 th, 2018
Intravascular Ultrasound in the Treatment of Complex Aortic Pathologies Naixin Kang, M.D. Vascular Surgery Fellow April 26 th, 2018 DISCLOSURES Nothing To Disclose 2 ENDOVASCULAR AORTIC INTERVENTION Improved
More informationReimbursement Guide Zenith Fenestrated AAA Endovascular Graft
MEDICAL Reimbursement Guide Zenith Fenestrated AAA Endovascular Graft Disclaimer: The information provided herein reflects Cook s understanding of the procedure(s) and/or device(s) from sources that may
More informationChimney endovascular aneurysm sealing (ch-evas) for ruptured abdominal aortic aneurysms (AAA) due to type Ia endoleak following failed EVAS
Chimney endovascular aneurysm sealing (ch-evas) for ruptured abdominal aortic aneurysms (AAA) due to type Ia endoleak following failed EVAS Saritphat Orrapin MD FRCS (Thailand), Thoetphum Benyakorn, Tunyarat
More informationInnovating for life. CardioVascular LifeLine Customer Support Tel: Tel:
Indications The Valiant Thoracic with the Captivia Delivery System is intended for the endovascular repair of fusiform aneurysms and saccular aneurysms/penetrating ulcers of the descending thoracic aorta
More informationPercutaneous Axillary Artery Access For Branch Grafting for complex TAAAs and pararenal AAAs: How to do it safely
Percutaneous Axillary Artery Access For Branch Grafting for complex TAAAs and pararenal AAAs: How to do it safely Daniela Branzan, MD, Department of Vascular Surgery University Hospital Leipzig Disclosure
More informationTaming The Aorta. David Minion, MD Program Director, Vascular Surgery University of Kentucky Medical Center Lexington, Kentucky, USA
Taming The Aorta David Minion, MD Program Director, Vascular Surgery University of Kentucky Medical Center Lexington, Kentucky, USA Faculty Disclosure Consulting: Endologix, Cook 1 Objectives Review the
More informationRight Choice for Right Angles
Right Choice for Right Angles The Anatomy of Technology Aorfix gives you technology that conforms to patient anatomy, optimising both procedure and post-operative performance. Fishmouth for optimum neck
More informationZenith Renu AAA Converter Graft. Device Description Planning and Sizing Deployment Sequence Patient Follow-Up
Zenith Renu AAA Converter Graft Device Description Planning and Sizing Deployment Sequence Patient Follow-Up Device description: Device indications The Zenith Renu AAA Converter Graft with Z-Trak Introduction
More informationChallenging anatomies demand versatility.
Challenging anatomies demand versatility. The Distinct Advantages of Separating Seal and Fixation ANATOMICAL FIXATION Unlike proximal fixation designs, the AFX bifurcated unibody endograft allows for natural
More informationCurrent State of Thoracic Branch Devices and Ongoing Clinical Trials
Current State of Thoracic Branch Devices and Ongoing Clinical Trials Hiroo Takayama, MD, PhD Associate Professor of Surgery Director of Cardiovascular Institute Co-Director of Aortic Center NY Presbyterian/Columbia
More informationRadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved.
Interventional Radiology Coding Case Studies Prepared by Stacie L. Buck, RHIA, CCS-P, RCC, CIRCC, AAPC Fellow President & Senior Consultant INDICATION: Abdominal aortic aneurysm. INTERVENTIONAL RADIOLOGIST:
More informationEndovascular Repair of Aortic Arch/Thoracic Aneurysms: Bolton RelayBranch Device
Endovascular Repair of Aortic Arch/Thoracic Aneurysms: Bolton RelayBranch Device Luis A. Sanchez MD Gregorio A. Sicard Distinguished Professor of Surgery & Radiology Chief, Section of Vascular Surgery
More informationIndications for use. Contraindications within the United States
Indications for use Indications within the United States The GORE TAG Thoracic Endoprosthesis is intended for endovascular repair of all lesions of the descending thoracic aorta, including: Isolated lesions
More informationHostile Neck During EVAR, The Role Of Endoanchores
Hostile Neck During EVAR, The Role Of Endoanchores Samer Koussayer, MD, FACS, RVT Prof, Al Faisal University Section Head and consultant Vascular & Endovascular Surgery Division King Faisal Specialist
More informationChallenges with Complex Anatomies Advancing Care in Endovascular Aortic Treatment
Challenges with Complex Anatomies Advancing Care in Endovascular Aortic Treatment Robert Y. Rhee, MD Chief, Vascular and Endovascular Surgery Director, Aortic Center Maimonides Medical Center Brooklyn,
More informationFour-year Surgical Results for Traumatic Aortic Injury in China Medical University Hospital, Mid-Taiwan
Four-year Surgical Results for Traumatic Aortic Injury in China Medical University Hospital, Mid-Taiwan Yi-Chun Lin 林怡均 (5 th grade medical student), MingLi Li 李明禮, Chih-Hsiang Hsu, Ching-Feng Wu, Hui-Han
More informationTreatment of complex thoracic cases Focus on the new Gore Active Control TAG device
Treatment of complex thoracic cases Focus on the new Gore Active Control TAG device Professor Christoph A. Nienaber The Royal Brompton and Harefield NHS Trust Cardiology and Aortic Centre C.Nienaber@rbht.nhs.uk
More informationINCRAFT system: Update from the Pivotal INSPIRATION Study
INCRAFT system: Update from the Pivotal INSPIRATION Study Michel S. Makaroun MD Co-Director, UPMC Heart and Vascular Institute Professor and Chair, Division of Vascular Surgery University of Pittsburgh
More informationWhat's on the Horizon for AAA: Unilateral & Percutaneous, "UP-EVAR" System Zoran Rancic M.D., Ph.D.
What's on the Horizon for AAA: Unilateral & Percutaneous, "UP-EVAR" System Zoran Rancic M.D., Ph.D. Clinic for Cardiovascular Surgery University Hospital Zurich DISCLOSURES COMMON SITUATIONS FOR UNILATERAL
More informationI-Hui Wu, M.D. Ph.D. Clinical Assistant Professor Cardiovascular Surgical Department National Taiwan University Hospital
Comparisons of Aortic Remodeling and Outcomes after Endovascular Repair of Acute and Chronic Complicated Type B Aortic Dissections I-Hui Wu, M.D. Ph.D. Clinical Assistant Professor Cardiovascular Surgical
More informationHow to select FEVAR versus EVAR + endoanchors in short-necked AAAs
How to select FEVAR versus EVAR + endoanchors in short-necked AAAs Jean-Paul P.M. de Vries, Richte C.L. Schuurmann St. Antonius Hospital Nieuwegein, The Netherlands 21st Critical Issues Congress Nuernberg,
More informationEndo-Bentall: Fact or Fiction?
Endo-Bentall: Fact or Fiction? Eric E. Roselli, MD Chief, Adult Cardiac Surgery Surgical Director, Aorta Center Heart and Vascular Institute, Cleveland Clinic Disclosures Bolton Cook Cryolife Edwards Gore
More informationOvation. Sean Lyden, MD Department Chair, Vascular Surgery Cleveland Clinic
Ovation Sean Lyden, MD Department Chair, Vascular Surgery Cleveland Clinic Disclosure Statement of Financial Interest Within the past 12 months, I or my spouse/partner have had a financial interest/arrangement
More informationGORE TAG Thoracic Endoprosthesis ANNUAL CLINICAL UPDATE SEPTEMBER 2018 Abstract. Section I GORE TAG Device Clinical experience. Section II Conformable
GORE TAG Thoracic Endoprosthesis ANNUAL CLINICAL UPDATE SEPTEMBER 08 Abstract This annual clinical update provides a review of the ongoing experience with the GORE TAG Thoracic Endoprosthesis There have
More informationThe Auckland Experience with the Nellix EVAS System. Andrew Holden, MBChB, FRANZCR
The Auckland Experience with the Nellix EVAS System Andrew Holden, MBChB, FRANZCR Disclosure Speaker name: Associate Professor Andrew Holden I have the following potential conflicts of interest to report:
More informationEVAR Revision Setting - How can Heli-FX EndoAnchors improve the outcomes?
D. Böckler Department of Vascular and Endovascular Surgery University Hospital Heidelberg EVAR Revision Setting - How can Heli-FX EndoAnchors improve the outcomes? Disclosures Speaker name: Dittmar Böckler
More informationLOWERING THE PROFILE RAISING THE BAR
LOWERING THE PROFILE RAISING THE BAR INNOVATIVE LOW PROFILE. ADVANCED CLINICAL PERFORMANCE. The AFX TM Endovascular AAA System integrates anatomical fixation with an advanced delivery system and graft
More informationSubclavian Artery Plug Embolization (SAPE study): a real experience about endovascular subclavian occlusion prior to thoracic vascular repair
Subclavian Artery Plug Embolization (SAPE study): a real experience about endovascular subclavian occlusion prior to thoracic vascular repair Simone Salvati, Luca Bertoglio, Alessandra Fittipaldi, Andrea
More informationEndovascular therapy for Ischemic versus Nonischemic complicated acute type B aortic dissection (catbad).
Endovascular therapy for Ischemic versus Nonischemic complicated acute type B aortic dissection (catbad). AS. Eleshra, MD 1, T. Kölbel, MD, PhD 1, F. Rohlffs, MD 1, N. Tsilimparis, MD, PhD 1,2 Ahmed Eleshra
More informationHistory of the Powerlink System Design and Clinical Results. Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ
History of the Powerlink System Design and Clinical Results Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ Powerlink System: Unibody-Bifurcated Design Long Main Body Low-Porosity Proprietary eptfe
More informationthoracic stent graft system Smallest delivery profile as from 20 F
thoracic stent graft system Smallest delivery profile as from 20 F Third-generation thoracic stent graft system effective, precise and safe JOTEC develops stent graft systems in close collaboration with
More informationEndovascular Repair o Abdominal. Aortic Aneurysms. Cesar E. Mendoza, M.D. Jackson Memorial Hospital Miami, Florida
Endovascular Repair o Abdominal Aortic Aneurysms Cesar E. Mendoza, M.D. Jackson Memorial Hospital Miami, Florida Disclosure Nothing to disclose. 2 Mr. X AAA Mr. X. Is a 70 year old male who presented to
More informationAnatomical applicability of current off-the-shelf branched endografts in thoracoabdominal aortic aneurysms managed by open surgery.
Leipzig Interventional Course Update on clinical trials Tuesday, January 30 th, 2018 Anatomical applicability of current off-the-shelf branched endografts in thoracoabdominal aortic aneurysms managed by
More informationCombined Endovascular and Surgical Repair of Thoracoabdominal Aortic Pathology: Hybrid TEVAR
Combined Endovascular and Surgical Repair of Thoracoabdominal Aortic Pathology: Hybrid TEVAR William J. Quinones-Baldrich MD Professor of Surgery Director UCLA Aortic Center UCLA Medical Center Los Angeles,
More informationThe evolution. AORFIX AAA Stent Graft now with the new AORFLEX Delivery System
The evolution of AORFIX AORFIX AAA Stent Graft now with the new AORFLEX Delivery System The confident choice for the right angles AORFIX the only AAA stent graft approved for all angulations from 0 90,
More informationEVAR replaced standard repair in most cases. Why?
EVAR replaced standard repair in most cases. Why? Initial major steps in endograft evolution Papazoglou O. Konstantinos M.D. The story of a major breakthrough in vascular surgery 1991 Parodi introduces
More informationAbdominal Aortic Aneurysms. A Surgeons Perspective Dr. Derek D. Muehrcke
Abdominal Aortic Aneurysms A Surgeons Perspective Dr. Derek D. Muehrcke Aneurysm Definition The abnormal enlargement or bulging of an artery caused by an injury or weakness in the blood vessel wall A localized
More informationManagement of intramural hematoma and penetrating ulcers - what is different? D.Böckler University Hospital Heidelberg, Germany
Management of intramural hematoma and penetrating ulcers - what is different? D.Böckler University Hospital Heidelberg, Germany Disclosure Speaker name: Dittmar Böckler I have the following potential conflicts
More informationFLEXIBLE, BALOON EXPANDABLE
EARLY RESULTS OF A CLINICAL TRIAL OF FLEXIBLE, BALOON EXPANDABLE COVERED STENT GRAFT IN ILIAC OCCLUSIVE DISEASE Chris LeCroy Coastal Vascular and Interventional Pensacola, Florida Clinical Trial WL GORE
More informationAortic Center of Excellence at Sentara
Gordon K. Stokes, MD, FACS Eastern Virginia Medical School Sentara Vascular Specialists Norfolk, VA Aortic Center of Excellence at Sentara April 25, 2014 Disclosures I have no disclosures relevant to this
More informationEXPERIENCE AFTER 500 ENDOLUMINAL STENT GRAFTS. DEVICES SPECIFIC OUTCOME AND LESSONS LEARNED.
EXPERIENCE AFTER 500 ENDOLUMINAL STENT GRAFTS. DEVICES SPECIFIC OUTCOME AND LESSONS LEARNED. A/Prof. B. Patrice Mwipatayi MD, MMed (Surg), MClinEd (Melb), FCS, FRACS Department of Vascular Surgery, Royal
More informationΓιατί είναι σημαντικό για τον θεράποντα χειρουργό να γνωρίζει πως να μετρήσει ένα ανεύρυσμα και να επιλέξει το κατάλληλο μόσχευμα?
Γιατί είναι σημαντικό για τον θεράποντα χειρουργό να γνωρίζει πως να μετρήσει ένα ανεύρυσμα και να επιλέξει το κατάλληλο μόσχευμα? Μ. Ματσάγκας MD, PhD, FEBVS Καθηγητής Αγγειοχειρουργικής Πανεπιστημίου
More informationPreliminary Data from the Gore EXCLUDER Thoracoabdominal Branch Endoprosthesis Early Feasibility Study
Preliminary Data from the Gore EXCLUDER Thoracoabdominal Branch Endoprosthesis Early Feasibility Study Mark A Farber, MD Professor of Surgery and Radiology Director, UNC Aortic Network University of North
More informationPioneering EVAR techniques in aortic dissection
Pioneering EVAR techniques in aortic dissection Jianing Yue, Weiguo Fu Department of Vascular Surgery Zhongshan Hospital Fudan University Shanghai, China LINC Asia-Pacific 2016 March 8-10, 2016 Hong Kong
More informationVALIANT THORACIC STENT GRAFT WITH THE CAPTIVIA DELIVERY SYSTEM 2016 ANNUAL PHYSICIAN CLINICAL UPDATE
VALIANT THORACIC STENT GRAFT WITH THE CAPTIVIA DELIVERY SYSTEM 2016 ANNUAL PHYSICIAN CLINICAL UPDATE MARCH 27, 2016 2016 ANNUAL PHYSICIAN CLINICAL UPDATE VALIANT THORACIC STENT GRAFT WITH THE CAPTIVIA
More informationPatient selection in Hostile Necks and how. to prevent endoleaks a word of caution
Patient selection in Hostile Necks and how to prevent endoleaks a word of caution P M Kasprzak, K. Pfister Department of Vascular Surgery Endovascular Surgery University Hospital Regensburg, Germany Conflict
More informationObjective assessment of current stent grafts: which graft for which lesion. Ludovic Canaud, MD, PhD Pierre Alric, MD, PhD Montpellier, France
Objective assessment of current stent grafts: which graft for which lesion Ludovic Canaud, MD, PhD Pierre Alric, MD, PhD Montpellier, France Conflict of interest: none 1 Introduction Different diseases
More informationTreatment options of late failures of EVAS. Michel Reijnen Rijnstate Arnhem The Netherlands
Treatment options of late failures of EVAS Michel Reijnen Rijnstate Arnhem The Netherlands Disclosure Speaker name: Michel Reijnen I have the following potential conflicts of interest to report: Consulting
More informationEndovascular aortic stent grafts have forever
Identifying the Appropriate Thoracic Device Size How to correctly size a thoracic device in diameter, landing zone, and length. BY MICHAEL E. BARFIELD, MD, AND THOMAS S. MALDONADO, MD Endovascular aortic
More informationDurable outcomes. Proven performance.
Durable outcomes. Proven performance. GORE EXCLUDER AAA Endoprosthesis GORE EXCLUDER Iliac Branch Endoprosthesis GORE EXCLUDER AAA Endoprosthesis The most-studied* EVAR stent graft designed for durable
More informationCase Report 1. CTA head. (c) Tele3D Advantage, LLC
Case Report 1 CTA head 1 History 82 YEAR OLD woman with signs and symptoms of increased intra cranial pressure in setting of SAH. CT Brain was performed followed by CT Angiography of head. 2 CT brain Extensive
More informationIMH/Penetrating Aortic Ulcers/ Saccular Aneurysms: How to manage and when to intervene
IMH/Penetrating Aortic Ulcers/ Saccular Aneurysms: How to manage and when to intervene UCSF Vascular Surgery Symposium 2018 Sukgu M Han, MD, MS Assistant Professor of Clinical Surgery Co-director, Comprehensive
More informationStandardization of the CHEVAR procedure: How a standard approach has improved outcomes. Prof Peter Holt St George s, London
Standardization of the CHEVAR procedure: How a standard approach has improved outcomes Prof Peter Holt St George s, London Disclosure Speaker name: Prof Peter Holt I have the following potential conflicts
More informationTechniques For Endovascular Aortic Arch Repair. Mario Lachat, MD University Hospital of Zurich Matteo Orrico, MD San Camillo-Forlanini Hospital, Rome
Techniques For Endovascular Aortic Arch Repair Mario Lachat, MD University Hospital of Zurich Matteo Orrico, MD San Camillo-Forlanini Hospital, Rome Disclosure Speaker name: Mario Lachat Endospan, Jotec,
More informationEuropean Experience with a New Thoracic Device. D.Böckler University Hospital Heidelberg Germany
European Experience with a New Thoracic Device D.Böckler University Hospital Heidelberg Germany Disclosures Speaker name: Dittmar Böckler I have the following potential conflicts of interest to report:
More informationTriVascular Ovation Prime Abdominal Stent Graft System
TriVascular Ovation Prime Abdominal Stent Graft System Science of the Seal O-Ring Sealing Technology O-Ring Sealing in Proven Engineering Solutions O-rings are designed to seal by blocking the flow of
More informationAnatomy-Driven Endograft Selection for Abdominal Aortic Aneurysm Repair S. Jay Mathews, MD, MS, FACC
Anatomy-Driven Endograft Selection for Abdominal Aortic Aneurysm Repair S. Jay Mathews, MD, MS, FACC Interventional Cardiologist/Endovascular Specialist Bradenton Cardiology Center Bradenton, FL, USA Disclosures
More information