Effect of alumina femoral heads on polyethylene wear in cemented total hip arthroplasty

Size: px
Start display at page:

Download "Effect of alumina femoral heads on polyethylene wear in cemented total hip arthroplasty"

Transcription

1 Effect of alumina femoral heads on polyethylene wear in cemented total hip arthroplasty OLD VERSUS CURRENT ALUMINA K. Tanaka, J. Tamura, K. Kawanabe, M. Shimizu, T. Nakamura From Kyoto University, Japan W e examined the behaviour of alumina ceramic heads in 156 cemented total hip arthroplasties, at a minimum follow-up of eight years. They were divided into three groups according to the size of the femoral head; 22, 26, and 28 mm. We measured polyethylene wear radiologically using a computer-aided technique. The linear wear rate of polyethylene sockets for the 28 mm heads was high (0.156 mm/year), whereas those for the 22 and 26 mm heads were relatively low (0.090 and mm/year, respectively). Moreover, the surface roughness data of retrieved femoral heads clearly showed maintenance of an excellent surface finish of the current alumina. We conclude that the alumina ceramic femoral heads currently used are associated with a reduced rate of polyethylene wear. J Bone Joint Surg [Br] 2003;85-B: Received 14 October 2002; Accepted after revision 18 February 2003 Polyethylene wear has been a major cause of periprosthetic osteolysis, leading to aseptic loosening after total hip arthroplasty (THA). 1-3 In order to reduce such wear, for more than 20 years, ceramic femoral heads have been used with polyethylene cups. 4-9 Alumina ceramics, which previously had been used clinically as a ceramic-on-ceramic combination, gave great promise because of their excellent wear characteristics and high resistance to scratching. 10 In laboratory tests of wear, many investigators have shown that the wear rate of polyethylene against alumina is lower than that against metal Although early clinical results of the alumina-on-polyethylene bearing combination were excellent, K. Tanaka, MD, Orthopaedic Surgeon J. Tamura, MD, PhD, Instructor K. Kawanabe, MD, PhD, Associate Professor M. Shimizu, MD, PhD, Instructor T. Nakamura, MD, PhD, Chairman and Professor of Orthopaedics Department of Orthopaedic Surgery, Faculty of Medicine, Kyoto University, 54 Shogoin, Kawahara-cho, Sakyo-ku, Kyoto , Japan. Correspondence should be sent to Professor T. Nakamura British Editorial Society of Bone and Joint Surgery doi: / x.85b $2.00 mid- and long-term wear of this combination has been controversial and some authors have questioned its use. 6,8,9 There are, however, differences between the old alumina, which was used in the early 1980s, and the newer material, which is currently in use. 16 Newer alumina ceramics seem better in terms of purity of composition, grain size, mechanical properties and surface quality. It is still premature to draw conclusions concerning the effect of alumina femoral heads on reducing polyethylene wear. In this study, therefore, our aim was to evaluate the effect of femoral heads made of new alumina on polyethylene wear. Patients and Methods Between March 1981 and December 1992, we carried out 225 consecutive primary cemented THAs in 181 patients using Kyocera hip prostheses (Bioceram series and KC series: Kyocera, Kyoto, Japan), in which alumina femoral heads articulated with all-polyethylene acetabular components. Of these, we selected only the 171 hips with a minimum follow-up of eight years, in order to assess the rate of polyethylene wear. We excluded 14 hips because of missing or incomplete radiological records. Four hips had been revised within eight years of the primary operation, three for aseptic loosening and these were included, and one for infection and this was excluded. The remaining 156 hips (124 patients) were divided into three groups for measurement of wear based on the size of the femoral head (Table I). The 28 mm heads were used between 1981 and 1988; the 26 mm heads between 1988 and 1991, and the 22 mm heads between 1991 and The preoperative diagnosis included osteoarthritis (primary or secondary) in 149 hips, avascular necrosis in four and rheumatoid arthritis in three. Since 1988, the manufacturer has made improvements in the properties of alumina ceramics, enabling the femoral head size to be reduced from 28 mm diameter (old alumina) to 26 or 22 mm (current alumina). The old alumina contained 7% of yttria, while the current material is high purity alumina containing >99.8% Al 2 O 3. Table II compares the material properties of the old and current alumina. Current alumina has a greater strength and a much smaller grain size as seen on SEM (Fig. 1). With regard to the femoral components, the Bioceram stem was designed for patients with dysplastic hips, which VOL. 85-B, No. 5, JULY

2 656 K. TANAKA, J. TAMURA, K. KAWANABE, M. SHIMIZU, T. NAKAMURA Table I. Details of the 124 patients and the implants used in cemented THA between 1981 and 1992 Alumina head (mm) Type of alumina Containing 7% yttria High purity High purity Implantation period 1981 to to to 1992 Number of hips Gender Male Female Mean age in years (range) 50.1 (37 to 61) 57.8 (29 to 77) 58.3 (42 to 74) Mean follow-up in years (range) 13.6 (6.4 to 17.8) 10.8 (8.5 to 12.5) 9.2 (8.0 to 10.9) Diagnosis (number of hips) Osteoarthritis Rheumatoid arthritis Avascular necrosis Table II. Material properties of old, containing 7% yttria, and current alumina, high purity, used in the implants Old alumina Current alumina Density (g/cm 3 ) Bending strength (MPa) Vickers hardness (HV) Grain size (µm) are frequently seen in Japan, and it has been described in detail elsewhere. 8,9,17 The KC stem (Kyocera) was similar in design to the Charnley prosthesis, but was made of stainless steel or its alloy, while the Charnley prosthesis was made of titanium alloy. The Bioceram stem had a 28 mm modular alumina head, and the KC stem had a 26 or 22 mm modular alumina head. All the acetabular polyethylene components in both series were machined from compression-moulded GUR 412 sheets and were sterilised with ethylene oxide gas. Radiological analysis. We measured polyethylene wear radiologically by determining the migration of the centre of the femoral head relative to the centre of the cup, based on the computer-aided technique of Sychterz et al. 18 For each patient, the initial postoperative and the most recent anteroposterior radiographs of the pelvis were selected. First, the radiographs were scanned using a scanner (GT9500, EPSON, Nagano, Japan) to generate 600 dpi TIFF images. Using the custom software Image-Pro Plus version 4.0 (Media Cybernetics, Maryland), an observer digitised about ten points around the periphery of the femoral head on the radiograph and also around the periphery of the polyethylene cup at the cement-cup interface. This software provided the best-fit circles and their centres for the head and cup from the digitised points, by which the two-dimensional coordinate for each centre was obtained on the radiograph (Fig. 2). By comparing the co-ordinates of the two centres on the initial postoperative and most recent radiographs, the Fig. 1 SEM (a) of old and (b) of current alumina ( 2500). THE JOURNAL OF BONE AND JOINT SURGERY

3 EFFECT OF ALUMINA FEMORAL HEADS ON POLYETHYLENE WEAR IN CEMENTED TOTAL HIP ARTHROPLASTY 657 Fig. 2 Radiological measurement of polyethylene wear. BP1 to 10, digitised points around the periphery of the polyethylene socket; BP11 to 20, digitised points around the periphery of the femoral head; BC1, centre of the socket; BC2, centre of the femoral head; D1, distance between two centres. amount and the direction of penetration of the femoral head into the polyethylene were determined after correction for pelvic tilting and magnification. Although the migration of the femoral head thus determined may consist not only of true wear, but also creep deformation of polyethylene, we defined it as linear wear for the purpose of this study. The direction of wear was also defined relative to a vertical line drawn through the centre of the femoral head and perpendicular to an inter-teardrop line, as described by Livermore, Ilstrup and Morrey. 19 The wear angle was defined as positive if the direction was medial to this line, and negative if lateral to it. In addition, volumetric wear was calculated from the linear wear data using the same equation as Hashimoto et al, 20 Maruyama et al 21 and Sychterz et al. 22 We determined the annual wear rate by dividing the total wear by the duration of time after implantation. In order to eliminate interobserver errors, all measurements were undertaken by the same observer (KT). To validate this technique, we compared the radiological measurements with direct measurements using ten cups retrieved from nine patients at revision surgery. These were chosen because their wear measurements had not been disturbed by excessive osteolysis or significant migration of the cup on the radiographs taken just before revision. The linear wear was measured directly from the retrieved polyethylene cups using a three-dimensional co-ordinate measuring machine (BHN305; Mitutoyo Corporation, Tokyo, Japan). For each patient, we also measured the polyethylene wear radiologically on five occasions using the method described above. Analysis of retrieved femoral heads. We analysed 11 femoral heads retrieved at revision surgery from ten patients with regard to surface roughness; seven were of old and four of current alumina. As controls, eight retrieved stainlesssteel heads from Charnley prostheses were also analysed. A contacting two-dimensional profilometer (S-405; Kosaka Laboratory Ltd, Tokyo, Japan) determined the surface roughness (Ra and Rmax) of the retrieved heads. The adopted cut-off and evaluation lengths were 0.8 mm for both. Ra is defined as the arithmetic mean roughness, while Rmax is the largest peak-to-valley height. Measurements of roughness were made at the apex and at the equator of the femoral head and on two lines perpendicular to each other at each point. Statistical analysis. The data were statistically analysed using a one-way ANOVA, with Scheffe s method as a posthoc test. Differences with p < 0.05 were considered to be statistically significant. VOL. 85-B, No. 5, JULY 2003

4 658 K. TANAKA, J. TAMURA, K. KAWANABE, M. SHIMIZU, T. NAKAMURA Table III. Comparison between the direct and radiological measurement of polyethylene wear in ten implants obtained from ten women at revision THA Wear (mm) Case Age Time in situ (yrs) Type of prosthesis Direct (A) Radiological (B)* Difference (B - A) Charnley ± Bioceram ± Bioceram ± Bioceram ± Bioceram ± Bioceram ± Bioceram ± KC (26) ± KC (22) ± KC (22) ± Mean ± 0.27 *the values are expressed as the mean and the SD of five measurements positive values indicate overestimation and negative values indicate underestimation Table IV. Wear data, at the last follow-up, expressed as mean ±SD, for 156 hips (124 patients) after THA with 28, 26 or 22 mm ceramic heads Alumina head (mm) Number of hips Socket angle (degrees) 39.0 ± ± ± 6.6 Polyethylene thickness (mm) 7.9 ± ± ± 1.1 Linear wear (mm) ± ± ± Linear wear rate (mm/year) ± ± ± Volumetric wear (mm 3 ) ± ± ± Volumetric wear rate (mm 3 /year) 91.1 ± 47.2* 50.1 ± ± 28.4 Wear angle (degrees) 4.7 ± ± ± 16.1 *significantly higher than for 26 and 22 mm heads (p < ) Results Validation testing of the measuring technique revealed a good correlation between the radiological and direct measurements of polyethylene wear (R 2 = 0.95), and that the radiological measurement underestimated the direct measurement by a mean of 0.11 mm (Table III). Moreover, measurements by the same observer were reproducible; the SD of five measurements ranged from 0.04 to 0.13 mm. The mean age at operation for the 28 mm group (50.1 years) was significantly lower than that for the 26 mm and 22 mm (57.8 and 58.3 years, respectively). This is because alumina ceramic heads were selectively used in younger patients and metal heads were used in older patients during the same period as the 28 mm group. There were significant differences in the mean lengths of follow-up among the three groups: 13.6 years for 28 mm, 10.8 for 26 mm and 9.2 for 22 mm. Of the 156 hips, 23 had been revised at the latest followup; 20 in the 28 mm group, one in the 26 mm group, and two in the 22 mm group. The wear data are summarised in Table IV. Although there were no significant differences in the cup abduction angle among the three groups, there were significant differences in the thickness of polyethylene. The mean linear wear rates ranged from to mm/year for the 28 mm group, from to mm/year for the 26 mm group, and from to mm/year for the 22 mm group. The mean volumetric wear rates ranged from 14.7 to mm 3 /year for the 28 mm group, from 2.7 to mm 3 /year for the 26 mm group, and from 2.5 to mm 3 /year for the 22 mm group. The volumetric wear rate for the 28 mm group of patients was significantly higher than that for the other two groups (p < ), whereas there were no significant differences between the 22 mm and 26 mm groups (p = 0.07). Interestingly, the mean wear angle was negative for the 22 mm group of patients and positive for the other two groups (p < 0.05). The surface roughness data for the retrieved femoral heads clearly showed the maintenance of an excellent surface finish of the current alumina (Table V). We did not compare the three groups statistically, because the number for the current alumina group was so small (n = 4) and the follow-up time for these patients was short. Both Ra and Rmax values on the apex for the current alumina heads (0.013 ± µm Ra and ± µm Rmax) were lower than those for both the old alumina (0.028 ± µm Ra and ± µm Rmax) and the stainless-steel heads (0.026 ± µm Ra and ± µm Rmax). It should be noted that the values for the roughness for the stainless steel heads were similar to those for the old alumina heads. THE JOURNAL OF BONE AND JOINT SURGERY

5 EFFECT OF ALUMINA FEMORAL HEADS ON POLYETHYLENE WEAR IN CEMENTED TOTAL HIP ARTHROPLASTY 659 Table V. Surface roughness data of 11 retrieved femoral heads, unused heads and eight stainless-steel controls, expressed as the mean ± SD (range) Surface roughness (µm) Apex Equator Number Time in vivo (years) Ra Rmax Ra Rmax Unused heads 22 mm alumina 0.02 to to to to mm alumina 0.01 to to to to mm alumina 0.02 to to to to 0.15 Retrieved heads Current alumina (22- or 26 mm) ± 1.5 (8.0 to 11.4) ± (0.01 to 0.02) ± (0.09 to 0.13) ± (0.01 to 0.02) ± (0.08 to 0.17) Old alumina (28 mm) ± 1.1 (13.0 to 16.3) ± ± ± ± (0.02 to 0.04) Stainless steel controls (22 mm) ± 1.9 (11.2 to 16.0) ± (0.01 to 0.05) (0.22 to 1.15) ± (0.07 to 0.47) (0.01 to 0.04) ± (0.01 to 0.10) (0.11 to 0.60) ± (0.06 to 0.86) Discussion The measurement of polyethylene wear is very important when predicting the longevity of a hip arthroplasty. The technique for measuring it, as described by Livermore et al, 19 has been used in many studies. 7,8,23-29 In cemented THA using all-polyethylene sockets, however, the measurements are subjective as they depend on the manual determination of the centre of the femoral head and the direction of maximum wear. Since our technique is based on the concept of vector wear analysis as described by Sychterz et al 18 or Martell and Berdia, 30 the amount and direction of wear are objectively determined from the co-ordinates of the centres of the head and the cup on the initial and latest radiographs. Additionally, even if the identification of the cement-polyethylene interface may be difficult when determining the direction of the maximum wear, the choice of well-recognisable points on the rest of the interface should provide the best-fit circle and its centre. We believe that computer-aided methods for measuring the wear of all-polyethylene sockets are preferable as they reduce the subjectivity of measurement. In recent years, several authors have questioned the use of alumina femoral heads in THA. Sychterz et al 18 reported no significant difference in the wear rate of polyethylene between alumina and cobalt-chrome femoral heads at a mean of seven years in cementless THA. Hasegawa et al 8 and Haraguchi et al 9 have recently investigated the longterm results of cemented THA with Bioceram prostheses and reported mean linear wear rates of 0.10 mm/year at 13 years and 0.15 mm/year at 13.8 years of follow-up, respectively. The mean linear wear rate of mm/year for the 28 mm group (Bioceram) in our study was disappointing, compared with the reported wear rates of all-polyethylene cups articulating with 28 mm metal heads which ranged from 0.05 to 0.12 mm/year. 19,23,24,31 By contrast, the wear rate of 0.09 mm/year for the 22 mm group was satisfactory compared with the recently reported values for the wear rate of polyethylene with 22 mm metal heads which ranged from 0.08 to 0.15 mm/year. 19,23,26,28,32-34 Wang and Essner 15 found that ceramic heads were effective in reducing polyethylene wear because of the inability of third-body particles to adhere to their surfaces rather than due to their resistance to scratching. On the other hand, Hasegawa et al 8 reported that the retrieved alumina heads (28 mm Bioceram) showed smooth surfaces without scratches, but there was debris stuck to the gaps on the surfaces. Therefore, we infer from these findings and our data that the increased surface roughness, in particular Rmax values, and the high wear rate for the old alumina heads may be due to the adherence of such third-body particles and the resultant abrasive wear process, while we believe that the better wear achieved in the current alumina groups is mainly attributable to the improved quality of the surface. Our study is limited because it is retrospective and there were significant differences in the age of the patients and the thickness of the polyethylene among the three groups. We are aware that young age and male gender, with higher activity, are risk factors for increased wear. We consider that the differences in the ages of the patients had little influence on the wear results, because Japanese patients after THA tend to be relatively inactive regardless of age and most of our patients are women. Although whether polyethylene thickness is correlated to increased wear remains controversial, we think that our results may be affected, partly, by this factor. 9,25,35 The wear data for the 22 mm group are, however, comparable, but not superior, to the excellent wear rates reported for 22 mm metal heads in some studies, with the preservation of excellent surface finish for the current alumina in vivo. 33,34 Metal heads with improved surface quality may show excellent wear performance. 36,37 These data should not be interpreted to mean that the current alumina heads are disappointing, because the possible advantageous effects of the current alumina heads in this study may have been masked by other factors including implant design, the quality of the polyethylene and the method of sterilisation. 25,27,29,38-41 In particular, since sterilisation by gamma irradiation may affect the cross-linking of polyethylene, VOL. 85-B, No. 5, JULY 2003

6 660 K. TANAKA, J. TAMURA, K. KAWANABE, M. SHIMIZU, T. NAKAMURA even if it is done in air, sterilisation with ethylene oxide gas in our study may have been disadvatageous In summary, our study has shown that alumina heads are not always associated with reduced polyethylene wear if the initial surface quality is poor, but that there may be reduced wear with the current alumina heads which have excellent surface quality. No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article. References 1. Amstutz HC, Campbell P, Kossovksy N, Clarke IC. Mechanism and clinical significance of wear debris-induced osteolysis. Clin Orthop 1992;276: Schamlzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: polyethylene wear debris and the concept of the effective joint space. J Bone Joint Surg [Am] 1992;74-A: Jiranek WA, Machado M, Jasty M, et al. Production of cytokines around loosened cemented acetabular components: analysis with immunohistochemical techniques and in situ hybridization. J Bone Joint Surg [Am] 1993;75-A: Schuller HM, Marti RK. Ten-year socket wear in 66 hip arthroplasties: ceramic versus metal heads. Acta Orthop Scand 1990;61: Zichner LP, Willert HG. Comparison of alumina-polyethylene and metal-polyethylene in clinical trials. Clin Orthop 1992;282: Sychterz CJ, Engh CA Jr, Young AM, Hopper RH Jr, Engh CA. Comparison of in vivo wear between polyethylene liners articulating with ceramic and cobalt-chrome femoral heads. J Bone Joint Surg [Br] 2000;82-B: Urban JA, Garvin KL, Boese CK, et al. Ceramic-on-polyethylene bearing surfaces in total hip arthroplasty: seventeen to twenty-one-year results. J Bone Joint Surg [Am] 2001;83-A: Hasegawa M, Ohashi T, Tani T. Poor outcome of 44 cemented total hip arthroplasties with alumina ceramic heads: clinical evaluation and retrieval analysis after years. Acta Orthop Scand 2001;72: Haraguchi K, Sugano N, Nishii T, et al. Analysis of survivorship after total hip arthroplasty using a ceramic head. Clin Orthop 2001;391: Clarke IC, Gustafson A. Clinical and hip simulator comparisons of ceramic-on-polyethylene and metal-on-polyethylene wear. Clin Orthop 2000;379: Semlitsch M, Lehmann M, Weber H, Doerre E, Willert HG. New prospects for a prolonged functional life-span of artificial hip joints by using the material combination polyethylene/aluminium oxide ceramic/ metal. J Biomed Mater Res 1977;11: Kumar P, Oka M, Ikeuchi K, et al. Low wear rate of UHMWPE against zirconia ceramic (Y-PSZ) in comparison to alumina ceramic and SUS 316L alloy. J Biomed Mater Res 1991;25: Weightman B, Light D. The effect of the surface finish of alumina and stainless steel on the wear rate of UHMW polyethylene. Biomaterials 1986;7: Saikko VO, Paavolainen PO, Slatis P. Wear of the polyethylene acetabular cup: metallic and ceramic heads compared in a hip simulator. Acta Orthop Scand 1993;64: Wang A, Essner A. Three-body wear of UHMWPE acetabular cups by PMMA particles against CoCr, alumina and zirconia heads in a hip joint simulator. Wear 2001;250: Willmann G. Ceramic femoral head retrieval data. Clin Orthop 2000;379: Iida H, Matsusue Y, Kawanabe K, et al. Cemented total hip arthroplasty with acetabular bone graft for developmental dysplasia: longterm results and survivorship analysis. J Bone Joint Surg [Br] 2000;82- B: Sychterz CJ, Engh CA Jr, Shah N, Engh CA Sr. Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time. J Bone Joint Surg [Am] 1997;79-A: Livermore J, Ilstrup L, Morrey B. Effect of femoral head size on wear in wear of the polyethylene acetabular component. J Bone Joint Surg [Am] 1990;72-A: Hashimoto Y, Bauer TW, Jiang M, Stulberg BN. Polyethylene wear in total hip arthroplasty: volumetric wear measurement of retrieved acetabular components. Trans Orthop Res Soc 1995;20: Maruyama M, Capello WN, D Antonio JA, Jaffe WL, Bierbaum BE. Effect of low friction ion treated femoral heads on polyethylene wear rates. Clin Orthop 2000;370: Sychterz CJ, Moon KH, Hashimoto Y, et al. Wear of polyethylene cups in total hip arthroplasty: a study of specimens retrieved post mortem. J Bone Joint Surg [Am] 1996;78-A: Callaghan JJ, Pedersen DR, Olejniczak JP, Goetz DD, Johnston RC. Radiographic measurement of wear in 5 cohorts of patients observed for 5 to 22 years. Clin Orthop 1995;317: Bankston AB, Cates H, Ritter MA, Keating EM, Faris PM. Polyethylene wear in total hip arthroplasty. Clin Orthop 1995;317: Oonishi H, Wakitani S, Murata N, et al. Clinical experience with ceramics in total hip replacement. Clin Orthop 2000;379: Hirakawa K, Mitsugi N, Koshino T, et al. Effect of acetabular cup position and orientation in cemented total hip arthroplasty. Clin Orthop 2001;388: Sakalkale DP, Sharkey PF, Eng K, Hozack WJ, Rothman RH. Effect of femoral component offset on polyethylene wear in total hip arthroplasty. Clin Orthop 2001;388: Kesteris U, Ilchamnn T, Wingstrand H, Onnerfalt R. Polyethylene wear in Scanhip arthroplasty with a 22 or 32 mm head: 62 matched patients followed for 7-9 years. Acta Orthop Scand 1996;67: Urquhart AG, D Lima DD, Venn-Watson E, Colwell CW Jr, Walker RH. Polyethylene wear after total hip arthroplasty: the effect of a modular femoral head with an extended flange-reinforced neck. J Bone Joint Surg [Am] 1998;80-A: Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg [Am] 1997;79-A: Cates HE, Faris PM, Keating EM, Ritter MA. Polyethylene wear in cemented metal-backed acetabular cups. J Bone Joint Surg [Br] 1993;75-B: Eggli S, z Brun S, Gerber C, Ganz R. Comparison of polyethylene wear with femoral heads of 22 mm and 32 mm: a prospective, randomised study. J Bone Joint Surg [Br] 2002;84-B: Kobayashi S, Eftekhar NS, Terayama K, Iorio R. Risk factors affecting radiological failure of the socket in primary Charnley low friction arthroplasty: a 10- to 20-year followup study. Clin Orthop 1994;306: Wroblewski BM, Fleming PA, Siney PD. Charnley low-frictional torque arthroplasty of the hip: 20- to 30-year results. J Bone Joint Surg [Br] 1999;81-B: Nashed RS, Becker DA, Gustilo RB. Are cementless acetabular components the cause of excess wear and osteolysis in total hip arthroplasty? Clin Orthop 1995;317: Lancaster JG, Dowson D, Isaac GH, Fisher J. The wear of ultra-high molecular weight polyethylene sliding on metallic and ceramic counterfaces representative of current femoral surfaces in joint replacement. Proc Inst Mech Eng [H] 1997;211: Kobayashi S, Saito N, Horiuchi H, Iorio R, Takaoka K. Poor bone quality or hip structure as risk factors affecting survival of total hip arthroplasty. Lancet 2000;355: Bankston AB, Keating EM, Ranawat C, Faris PM, Ritter MA. Comparison of polyethylene wear in machined versus molded polyethylene. Clin Orthop 1995;317: Affatato S, Bordini B, Fagnano C, et al. Effects of the sterilisation method on the wear of UHMWPE acetabular cups tested in a hip joint simulator. Biomaterials 2002;23: Karrholm J, Digas G, Nivbrant B, Rohrl S, Thanner J. Increased femoral head penetration using polyethylene sterilised with ethylene oxide. Trans Orthop Res Soc 2001;47: McKellop HA, Shen FW, Campbell P, Ota T. Effect of molecular weight, calcium stearate, and sterilisation methods on the wear of ultra high molecular weight polyethylene acetabular cups in a hip joint simulator. J Orthop Res 1999;17: THE JOURNAL OF BONE AND JOINT SURGERY

Surface Roughness of Femoral Head Prostheses After Dislocation

Surface Roughness of Femoral Head Prostheses After Dislocation An Original Study Surface Roughness of Femoral Head Prostheses After Dislocation Kenny T. Mai, MD, Christopher Verioti, DO, Darryl D Lima, MD, PhD, Clifford W. Colwell, Jr., MD, and Kace A. Ezzet, MD Abstract

More information

ORIGINAL PAPER. Department of Orthopedic Surgery, Hamamatsu Medical Center ABSTRACT

ORIGINAL PAPER. Department of Orthopedic Surgery, Hamamatsu Medical Center ABSTRACT Nagoya J. Med. Sci. 71. 145 ~ 150, 2009 ORIGINAL PAPER AUTOGENOUS BULK STRUCTURAL BONE GRAFTING FOR RECONSTRUCTION OF THE ACETABLUM IN PRIMARY TOTAL HIP ARTHROPLASTY: AVERAGE 12-YEAR FOLLOW-UP TETSUO MASUI,

More information

Poor mid-term results of total hip arthroplasty with use of a Hylamer liner

Poor mid-term results of total hip arthroplasty with use of a Hylamer liner Acta Orthop. Belg., 2008, 74, 337-342 ORIGINAL STUDY Poor mid-term results of total hip arthroplasty with use of a Hylamer liner Adrian SKWARA, Simone STRACKE, Carsten O. TIBESKU, Susanne FUCHS-WINKELMANN

More information

Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene Liner

Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene Liner ORIGINAL ARTICLE Hip Pelvis 27(4): 216-222, 2015 http://dx.doi.org/10.5371/hp.2015.27.4.216 Print ISSN 2287-3260 Online ISSN 2287-3279 Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene

More information

Polyethylene wear in total hip prostheses: the influence of direction of linear wear on volumetric wear determined from radiographic data

Polyethylene wear in total hip prostheses: the influence of direction of linear wear on volumetric wear determined from radiographic data Skeletal Radiol (2003) 32:679 686 DOI 10.1007/s00256-003-0685-2 ARTICLE R. Košak V. Antolič V. Pavlovčič V. Kralj-Iglič I. Milošev G. Vidmar A. Iglič Polyethylene wear in total hip prostheses: the influence

More information

Radiographic Wear Measurements in a Cementless Metal-Backed Modular Cobalt-Chromium Acetabular Component

Radiographic Wear Measurements in a Cementless Metal-Backed Modular Cobalt-Chromium Acetabular Component The Journal of Arthroplasty Vol. 16 No. 7 2001 Radiographic Wear Measurements in a Cementless Metal-Backed Modular Cobalt-Chromium Acetabular Component Robert L. Barrack, MD,* Carlos Lavernia, MD, Edward

More information

A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY

A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY by Timothy McTighe, Exec. Dir., JISRF, Chagrin Falls, Ohio Ying Ko, Ph.D., Cincinnati, Ohio Russell B. Bennett, Ph.D., Cincinnati, Ohio James

More information

Investigation performed at the Anderson Orthopaedic Research Institute, Alexandria, Virginia

Investigation performed at the Anderson Orthopaedic Research Institute, Alexandria, Virginia 464 COPYRIGHT 2003 BY THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED Effect of Terminal Sterilization with Gas Plasma or Gamma Radiation on Wear of Polyethylene Liners BY ROBERT H. HOPPER JR., PHD,

More information

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Diagnosis: Ceramic head fracture In the 1970 s, Boutin implemented ceramic in modern total hip arthroplasty (THA).

More information

Total hip replacement with a zirconium oxide ceramic femoral head

Total hip replacement with a zirconium oxide ceramic femoral head Total hip replacement with a zirconium oxide ceramic femoral head A RANDOMISED ROENTGEN STEREOPHOTOGRAMMETRIC STUDY T. von Schewelov, L. Sanzén, I. Önsten, Å. Carlsson, J. Besjakov From Lund University

More information

Department of Orthopaedic Surgery, Graduate School of Medical Sciences, Kyushu University, Maidashi, Higashi-ku, Fukuoka , Japan

Department of Orthopaedic Surgery, Graduate School of Medical Sciences, Kyushu University, Maidashi, Higashi-ku, Fukuoka , Japan Wear Resistant Performance of Highly Cross-Linked and Annealed Ultra-High Molecular Weight Polyethylene against Ceramic Heads in Total Hip Arthroplasty Taishi Sato, Yasuharu Nakashima, Mio Akiyama, Takuaki

More information

Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists

Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists : 31 35 doi 10.1308/1478708051450 Audit Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists Jac Ciampolini, Matthew JW Hubble Princess Elizabeth Orthopaedic Centre,

More information

Clinical Study Comparison of Conventional Polyethylene Wear and Signs of Cup Failure in Two Similar Total Hip Designs

Clinical Study Comparison of Conventional Polyethylene Wear and Signs of Cup Failure in Two Similar Total Hip Designs Advances in Orthopedics Volume 2013, Article ID 710621, 7 pages http://dx.doi.org/10.1155/2013/710621 Clinical Study Comparison of Conventional Polyethylene Wear and Signs of Cup Failure in Two Similar

More information

Enhancing stability and increasing range of motion. Metasul LDH Large Diameter Head

Enhancing stability and increasing range of motion. Metasul LDH Large Diameter Head Enhancing stability and increasing range of motion Metasul LDH Large Diameter Head The Metasul large diameter head technology is the result of in-depth research, development and clinical experience that

More information

Stability without compromise. Epsilon Durasul Constrained Insert

Stability without compromise. Epsilon Durasul Constrained Insert Stability without compromise Epsilon Durasul Constrained Insert History of Dislocation and Constrained Acetabular Inserts The Need for a Better Solution The Problem Dislocation is the second most common

More information

Wear Analysis of Second-generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty

Wear Analysis of Second-generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty Wear Analysis of Second-generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty Christopher Samujh, MD; Samrath Bhimani, MS; Langan Smith, BS; Arthur L. Malkani, MD abstract A major

More information

ARTICLE IN PRESS. Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction. Effects on Polyethylene Wear

ARTICLE IN PRESS. Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction. Effects on Polyethylene Wear The Journal of Arthroplasty Vol. 00 No. 0 2008 Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction Effects on Polyethylene Wear William L. Jaffe, MD,* Eric J. Strauss,

More information

The Spatial Location of Impingement in Total Hip Arthroplasty

The Spatial Location of Impingement in Total Hip Arthroplasty The Journal of Arthroplasty Vol. 15 No. 3 2000 The Spatial Location of Impingement in Total Hip Arthroplasty Motoi Yamaguchi, MD, PhD,* Toshihiro Akisue, MD, PhD,* Thomas W. Bauer, MD, PhD,* and Yasushi

More information

Wear and Osteolysis of Highly Crosslinked Polyethylene at 10 to 14 Years: The Effect of Femoral Head Size

Wear and Osteolysis of Highly Crosslinked Polyethylene at 10 to 14 Years: The Effect of Femoral Head Size Clin Orthop Relat Res (2016) 474:365 371 DOI 10.1007/s11999-015-4319-5 Clinical Orthopaedics and Related Research A Publication of The Association of Bone and Joint Surgeons SYMPOSIUM: 2015 HIP SOCIETY

More information

111. ACETABULAR FIXATION

111. ACETABULAR FIXATION CLINICAL ORTHOPAEDICS AND RELATED RESEARCH Number 369, pp. 157-164 0 1999 Lippincott Williams & Wilkins, Inc. 111. ACETABULAR FIXATION IN PRIMARY TOTAL HIP ARTHROPLASTY Fixation, Polyethylene Wear, and

More information

UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD

UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS 66th Annual Meeting February 4-9, 1999 Anaheim, California COMMITTEE ON BIOMEDICAL ENGINEERING COMMITTEE ON

More information

Wear 268 (2010) Contents lists available at ScienceDirect. Wear. journal homepage:

Wear 268 (2010) Contents lists available at ScienceDirect. Wear. journal homepage: Wear 268 (2010) 617 621 Contents lists available at ScienceDirect Wear journal homepage: www.elsevier.com/locate/wear Wear comparison between a dual mobility total hip prosthesis and a typical modular

More information

Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty

Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty The Journal of Arthroplasty Vol. 23 No. 2 2008 Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty 7135 Primary Total Hip Arthroplasties after Developmental Dysplasia of the

More information

Femoral Osteolysis Around the Unrevised Stem During Isolated Acetabular Revision

Femoral Osteolysis Around the Unrevised Stem During Isolated Acetabular Revision Clin Orthop Relat Res (2009) 467:1501 1506 DOI 10.1007/s11999-008-0499-6 ORIGINAL ARTICLE Femoral Osteolysis Around the Unrevised Stem During Isolated Acetabular Revision Byung-Woo Min MD, Kwang-Soon Song

More information

Bone Bangalore

Bone Bangalore Dr Suresh Annamalai MBBS, MRCS(Edn), FRCS( Tr & Orth)(Edn), FEBOT(European Board), Young Hip and Knee Fellowship(Harrogate, UK) Consultant Arthroplasty and Arthroscopic Surgeon Manipal Hospital, Whitefield,

More information

Zimmer TM. Continuum Acetabular System. Bringing together proven technologies.

Zimmer TM. Continuum Acetabular System. Bringing together proven technologies. Zimmer TM Continuum Acetabular System Bringing together proven technologies. The power to meet individual patient needs. Zimmer Continuum Acetabular System provides highly flexible solutions for orthopaedic

More information

Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital)

Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital) Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital) 1. Introduction Total hip replacement arthroplasty (THRA) using a metal-on-polyethylene articulation has become one of the most

More information

FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY

FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY Fixation First FIXATION FIRST 1958 Charnley Hip Over 1,000,000 implanted 1 96.2 percent survivorship at 32 years

More information

Twenty-five-Year Results After Charnley Total Hip Arthroplasty in Patients Less than Fifty Years Old A CONCISE FOLLOW-UP OF A PREVIOUS REPORT*

Twenty-five-Year Results After Charnley Total Hip Arthroplasty in Patients Less than Fifty Years Old A CONCISE FOLLOW-UP OF A PREVIOUS REPORT* 1066 COPYRIGHT 2003 BY THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED Twenty-five-Year Results After Charnley Total Hip Arthroplasty in Patients Less than Fifty Years Old A CONCISE FOLLOW-UP OF A

More information

Total Hip Arthroplasty around the Inception of the Interface Bioactive Bone Cement Technique

Total Hip Arthroplasty around the Inception of the Interface Bioactive Bone Cement Technique Review Article Clinics in Orthopedic Surgery 2016;8:237-242 http://dx.doi.org/10.4055/cios.2016.8.3.237 Total Hip Arthroplasty around the Inception of the Interface Bioactive Bone Cement Technique Hiroyuki

More information

Continuing the Tradition. VerSys Heritage Hip System

Continuing the Tradition. VerSys Heritage Hip System Continuing the Tradition VerSys Heritage Hip System Heritage Following the Tradition The low-friction hip prosthesis developed by Sir John Charnley has more than a 20-year history of outstanding results.

More information

High wear rates and extensive osteolysis in 3 types of uncemented total hip arthroplasty

High wear rates and extensive osteolysis in 3 types of uncemented total hip arthroplasty Acta Orthopaedica 6; 77 (4): 575 584 575 High wear rates and extensive osteolysis in 3 types of uncemented total hip arthroplasty A review of the PCA, the Harris Galante and the Profile/Tri-Lock Plus arthroplasties

More information

MARATHON Cross-Linked Polyethylene

MARATHON Cross-Linked Polyethylene MARATHON Cross-Linked Polyethylene Fixation First... 1 Wear Reduction... 2 Oxidative Stability... 3 Mechanical Integrity... 4 Proven Technology...5-6 References... 7 Fixation First FIXATION FIRST Prodigy

More information

Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty

Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty A. J. Price, A. Short, C. Kellett, D. Beard, H. Gill, H. Pandit, C. A. F. Dodd, D. W. Murray From

More information

Tri-Lock Bone Preservation Stem

Tri-Lock Bone Preservation Stem Tri-Lock Bone Preservation Stem Clinical Results of the Tri-lock BPS Femoral Stem at One Year Follow-Up William L. Healy, M.D. Lahey Clinic Medical Center Burlington, Massachusetts Introduction Cementless

More information

Developing techniques for the dimensional assessment of true clinical wear in retrieved total joint replacements.

Developing techniques for the dimensional assessment of true clinical wear in retrieved total joint replacements. Developing techniques for the dimensional assessment of true clinical wear in retrieved total joint replacements. P. Bills*, L. Blunt*, G. Chakrabarty** *University of Huddersfield, Huddersfield, UK **West

More information

CLINICAL PAPER / ORTHOPEDIC

CLINICAL PAPER / ORTHOPEDIC HIP LEG LENGTH AND OFFSET Kelley T.C. and Swank M.L. (2009) Using CAS leads to more accurate positioning within the safe zone (inclination between 30 and 50, anteversion between 5 and 25 ) CAS improves

More information

Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces

Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces Timothy L. Tan, BS 1, Brenna C. Moeljadi, BS 2, Edward Ebramzadeh, PhD 3, Patricia Campbell, PhD 3, Sophia Nicole

More information

Exeter. Technical Guide

Exeter. Technical Guide Exeter Technical Guide Exeter Hip System Anatomic Restoration The Exeter hip system allows interoperative changes in offset, leg length and version, all independent of one another. This flexibility facilitates

More information

Hip Joint Prostheses

Hip Joint Prostheses Translated English of Chinese Standard: YY0118-2005 www.chinesestandard.net Sales@ChineseStandard.net PHARMACEUTICAL INDUSTRY STANDARD YY ICS 11.040.40 OF THE PEOPLE S REPUBLIC OF CHINA C 35 YY 0118-2005

More information

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 6, Issue 12 (May 2013), PP. 119-123 A Mathematical Model to Calculate Contact Stresses

More information

Outcomes of Charnley total hip arthroplasty using improved cementing with so-called second- and third-generation techniques

Outcomes of Charnley total hip arthroplasty using improved cementing with so-called second- and third-generation techniques J Orthop Sci (212) 17:118 DOI 1.17/s776-11-18-x ORIGINAL ARTICLE Outcomes of Charnley total hip arthroplasty using improved cementing with so-called second- and third-generation techniques Shiro Hirose

More information

Product Information. *smith&nephew POLARCUP Dual Mobility System

Product Information. *smith&nephew POLARCUP Dual Mobility System APPLYING SCIENCE TO MOBILITY Combined with VERILAST Technology for unmatched performance Product Information *smith&nephew POLARCUP Dual Mobility System APPLYING SCIENCE TO MOBILITY Combined with VERILAST

More information

Acetabular Cup System. Clinical Summary

Acetabular Cup System. Clinical Summary Acetabular Cup System Clinical Summary A Prospective, Randomized Study of Cross-Linked and Non-Cross-Linked Polyethylene for Total Hip Arthroplasty at 10-Year Follow-Up Engh CA Jr., Hopper RH Jr., Huynh

More information

A modular femoral neck and head system works well in cementless total hip replacement for patients with developmental dysplasia of the hip

A modular femoral neck and head system works well in cementless total hip replacement for patients with developmental dysplasia of the hip A modular femoral neck and head system works well in cementless total hip replacement for patients with developmental dysplasia of the hip T. Sakai, K. Ohzono, T. Nishii, H. Miki, M. Takao, N. Sugano From

More information

Painful Metal-on-Metal Total Hip Arthroplasty

Painful Metal-on-Metal Total Hip Arthroplasty The Journal of Arthroplasty Vol. 21 No. 2 2006 Painful Metal-on-Metal Total Hip Arthroplasty Cambize Shahrdar, MD,* Pat Campbell, PhD,y Joseph Mirra, MD,y and Lawrence D. Dorr, MD* Abstract: Two patients

More information

6.1 Hip Joint Simulators: State of the Art

6.1 Hip Joint Simulators: State of the Art Tribology 171 6.1 Hip Joint Simulators: State of the Art S. Affatato, W. Leardini and M. Zavalloni Abstract Wear and simulator testing are complicated tasks. Controlled wear testing should not be routinely

More information

Featuring. Technology. Product Rationale

Featuring. Technology. Product Rationale Featuring Technology Product Rationale 2 Optimum implant geometry Extending proven TRI-LOCK heritage The original TRI-LOCK Stem was introduced in 1981. This implant was the first proximally coated tapered-wedge

More information

Are Portable Imaging Intraoperative Radiographs Helpful for Assessing Adequate Acetabular Cup Positioning in Total Hip Arthroplasty?

Are Portable Imaging Intraoperative Radiographs Helpful for Assessing Adequate Acetabular Cup Positioning in Total Hip Arthroplasty? J Korean Med Sci 2009; 24: 315-9 ISSN 1011-8934 DOI: 10.3346/jkms.2009.24.2.315 Copyright The Korean Academy of Medical Sciences Are Portable Imaging Intraoperative Radiographs Helpful for Assessing Adequate

More information

Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo?

Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo? Journal of Orthopaedic Surgery 2005;13(1):46-51 Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo? H Tsumura, N Kaku, T Torisu Department of Orthopedic Surgery, Oita University,

More information

Polyethylene wear rate and osteolysis: critical threshold versus continuous dose-response relationship

Polyethylene wear rate and osteolysis: critical threshold versus continuous dose-response relationship ELSEVER Journal of Orthopaedic Research 23 (2005) 520-525 Journal of Orthopaedic Research www.elsevier.com/locate/orthres Polyethylene wear rate and osteolysis: critical threshold versus continuous dose-response

More information

National Joint Replacement Registry. Lay Summary 2015 Annual Report Hip and Knee Replacement

National Joint Replacement Registry. Lay Summary 2015 Annual Report Hip and Knee Replacement National Joint Replacement Registry Lay Summary 2015 Annual Report Hip and Knee Replacement SUPPLEMENTARY REPORT 2015 TABLE OF CONTENTS Introduction... 1 A brief history of the Registry origins... 1 The

More information

RM Philosophy Innovation based on clinical evidence.

RM Philosophy Innovation based on clinical evidence. RM Philosophy Innovation based on clinical evidence RM Philosophy Innovation based on clinical evidence Design 1973 RM Classic Cup First uncoated RM Classic Cup out of Polyacetal was inserted by E. Morscher

More information

CHAPTER 1 INTRODUCTION

CHAPTER 1 INTRODUCTION 1 CHAPTER 1 INTRODUCTION 1.1 Development history of total hip replacement Owing to the crippling nature of arthritis, the medical profession has been trying for well over a century to successfully treat

More information

The Challenge of Durability in Hip Replacement

The Challenge of Durability in Hip Replacement The Challenge of Durability in Hip Replacement Gregory A. Tocks, D.O. Orthopedic Associates of Lancaster Ltd. INTRODUCTION Hip replacement is a very common operation, and the number of long-term postoperative

More information

Surface evaluation of orthopedic hip implants marketed in Brazil

Surface evaluation of orthopedic hip implants marketed in Brazil Surface evaluation of orthopedic hip implants marketed in Brazil M.M.Souza 1, R.M. Trommer 2, M.M.Maru 2, C.R.M.Roesler 3, W.S.Barros 1, M.S. Dutra 3 1 Dimensional Metrology Laboratory, Inmetro. Brazil;

More information

Instrumentation Options

Instrumentation Options Instrumentation Options With the introduction of Microplasty and Premier Instrumentation platforms and advancements such as our patented Slidex Technology, the Vanguard Complete Knee System is the surgeon

More information

TEN YEARS EXPERIENCE WITH NONCEMENTED REVISION SOCKETS. Lawrence D. Dorr, MD Zhinian Wan, MD

TEN YEARS EXPERIENCE WITH NONCEMENTED REVISION SOCKETS. Lawrence D. Dorr, MD Zhinian Wan, MD ""'".."c.",".~" ""ill""."""'""'" I TEN YEARS EXPERIENCE WITH NONCEMENTED REVISION SOCKETS. Lawrence D. Dorr, MD Zhinian Wan, MD f ABSTRACT A prospective study was completed on 167 noncemented revision

More information

Furlong H-A.C. THR System

Furlong H-A.C. THR System Important Information Please read prior to use in a clinical setting. The Surgeon should be familiar with the operative technique. Caution Federal (U.S.A) law restricts this device to sale by or on the

More information

Investigation performed at the Department of Orthopaedic Surgery, Wrightington Hospital, Wigan, United Kingdom

Investigation performed at the Department of Orthopaedic Surgery, Wrightington Hospital, Wigan, United Kingdom 71 COPYRIGHT 2006 BY THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED Long-Term Survival of the Acetabular Component After Total Hip Arthroplasty with Cement in Patients with Developmental Dysplasia

More information

Disclosure. Direct Anterior THA via Extension Table. History 11/3/2015

Disclosure. Direct Anterior THA via Extension Table. History 11/3/2015 Direct Anterior THA via Extension Table J. Masonis, M.D. OrthoCarolina Hip & Knee Center Chief Adult Hip & Knee reconstruction Carolinas Medical Center Disclosure Consultant & Design Smith & Nephew Zimmer

More information

Index. Gainslen prosthesis 125. Debonding 22-3 Degradation products 40-1 Diffusion theory of adhesive wear 137 Dual energy X-ray absorption 95-6, 98

Index. Gainslen prosthesis 125. Debonding 22-3 Degradation products 40-1 Diffusion theory of adhesive wear 137 Dual energy X-ray absorption 95-6, 98 Index Abrasion 40-1 mechanism of 137 Acetabular components 59-63 advantages 59-60 disadvantages 60-3 Harris Galante Porous cup 59, 60 Porous Coated Anatomic cup 60 Adhesins 155-6 Adhesive wear 137 Alloys

More information

Increase in early polyethylene wear after sterilization with ethylene oxideradiostereometric analyses of 201 total hips

Increase in early polyethylene wear after sterilization with ethylene oxideradiostereometric analyses of 201 total hips Acta Orthopaedica Scandinavica ISSN: 0001-6470 (Print) (Online) Journal homepage: https://www.tandfonline.com/loi/iort19 Increase in early polyethylene wear after sterilization with ethylene oxideradiostereometric

More information

Optimum implant geometry

Optimum implant geometry Surgical Technique Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Session 3: Controversies for Bearing Surfaces in THR

Session 3: Controversies for Bearing Surfaces in THR : Controversies for Bearing Surfaces in THR Learning Objectives Upon completion of this activity, participants should be able to: 1. Understand the indications and known methods for the use of each of

More information

PINNACLE Acetabular Cup System

PINNACLE Acetabular Cup System PINNACLE Acetabular Cup System Clinical Summary A Prospective, Randomized Study of Cross-Linked and Non-Cross-Linked Polyethylene for Total Hip Arthroplasty at 10-Year Follow-Up Engh CA Jr., Hopper RH

More information

Hedrocel trabecular metal monoblock acetabular cups : mid-term results

Hedrocel trabecular metal monoblock acetabular cups : mid-term results Acta Orthop. Belg., 2006, 72, 326-331 ORIGINAL STUDY Hedrocel trabecular metal monoblock acetabular cups : mid-term results Michiel MULIER, Bart RYS, Lieven MOKE From the University Hospital Pellenberg,

More information

Effect of Superior Placement of the Hip Center on Abductor Muscle Strength in Total Hip Arthroplasty

Effect of Superior Placement of the Hip Center on Abductor Muscle Strength in Total Hip Arthroplasty The Journal of Arthroplasty Vol. 24 No. 2 2009 Effect of Superior Placement of the Hip Center on Abductor Muscle Strength in Total Hip Arthroplasty Takahiko Kiyama, MD, Masatoshi Naito, MD, PhD, Hiroshi

More information

CAUTION: Ceramic liners are not approved for use in the United States.

CAUTION: Ceramic liners are not approved for use in the United States. Total Hip Prostheses, Self-Centering Hip Prostheses and Hemi-Hip Prostheses IMPORTANT: This essential product information sheet does not include all of the information necessary for selection and use of

More information

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS INTRODUCTION The Summit Tapered Hip System s comprehensive set of implants and instruments

More information

Exeter. Designed for Anatomic Restoration. Clinical Evidence Education

Exeter. Designed for Anatomic Restoration. Clinical Evidence Education Exeter Designed for Anatomic Restoration Clinical Evidence Education Anatomic Restoration Exeter re-creating biomechanics In hip replacement, the smallest adjustment can make a big difference. Stryker

More information

The information contained in this document is intended for healthcare professionals only.

The information contained in this document is intended for healthcare professionals only. The information contained in this document is intended for healthcare professionals only. Trident Poly Acetabular System Trident PSL HA Acetabular Shell with X3 Poly insert Trident Hemispherical Acetabular

More information

Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem

Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem Original Article Clinics in Orthopedic Surgery 2009;1:105-109 doi:10.4055/cios.2009.1.2.105 Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem Kyoung Ho Moon, MD, Joon Soon

More information

Cementless Acetabular Fixation With and Without Screws

Cementless Acetabular Fixation With and Without Screws Cementless Acetabular Fixation With and Without Screws Analysis of Stability and Migration The Journal of Arthroplasty Vol. 25 No. 2 2010 Richard Iorio, MD,* Brian Puskas, MD,y William L. Healy, MD,* John

More information

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter Design Philosophy Cementless Stems of 12/14 Taper Stem Length Neck shaft Angle STEM-N3 STEM-N4 STEM-N5 STEM-N6 STEM-N7 STEM-N8 STEM-N9 123 mm 125 mm 130 mm 135 mm 140 mm 145 mm 150 mm 132 Cement Polished

More information

Total Hip Replacement. Find out why the Anterior Approach may be right for you.

Total Hip Replacement. Find out why the Anterior Approach may be right for you. Total Hip Replacement Find out why the Anterior Approach may be right for you. UNDERSTANDING TOTAL HIP REPLACEMENT This brochure offers a brief overview of the Direct Anterior Approach to total hip arthroplasty.

More information

Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs: Report of Two Cases

Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs: Report of Two Cases SAGE-Hindawi Access to Research Advances in Orthopedics Volume 2011, Article ID 519254, 4 pages doi:10.4061/2011/519254 Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs:

More information

Understanding Hip Implant Options

Understanding Hip Implant Options Understanding Hip Implant Options Cup (Socket) Bearing (Liner) Head Stem (Femur) Modern Hip Implants (Ball) Metal Femoral Head on Marathon Cross-linked Polyethylene Liner Proven materials Polyethylene

More information

Ceramic Femoral Heads and Acetabular Cup Liners

Ceramic Femoral Heads and Acetabular Cup Liners Important Information: Please read prior to use in a clinical setting. The Surgeon should be familiar with the operative technique. Caution: Federal (U.S.A) law restricts this device to sale by or on the

More information

Femoral Revision in Hip Resurfacing Compared With Large-Bearing Metal-on-Metal Hip Arthroplasty

Femoral Revision in Hip Resurfacing Compared With Large-Bearing Metal-on-Metal Hip Arthroplasty The Journal of Arthroplasty Vol. 26 No. 8 2011 Femoral Revision in Hip Resurfacing Compared With Large-Bearing Metal-on-Metal Hip Arthroplasty Simon J.W. Garrett, BM, MRCS, FRCS(Tr&Orth),*y Ben J.R.F.

More information

Progeny Hip Stem. Surgical Protocol and Product Specifications

Progeny Hip Stem. Surgical Protocol and Product Specifications Progeny Hip Stem Surgical Protocol and Product Specifications Progeny Hip Stem Introduction With emphasis on maximum stability and ease of use, the StelKast ProgenyTM Hip System provides the surgeon with

More information

MicroSeal Total Hip Acetabular System. Like Nothing Else

MicroSeal Total Hip Acetabular System. Like Nothing Else MicroSeal Total Hip Acetabular System Like Nothing Else Like Nothing Else Unique Patented Sealing and Locking Design The MicroSeal design features: Locking/Sealing mechanisms work together to minimize

More information

Retrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report

Retrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report Journal of Long-Term Effects of Medical Implants, 18(1): 69 74 (2008) Retrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report W. Siebert, 1 S. Mai, 1 & S. Kurtz 2 1 Orthopädische Klinik

More information

6.1 Primary Total Knee Replacement with a Zirconia

6.1 Primary Total Knee Replacement with a Zirconia New Applications for Ceramics 183 6.1 Primary Total Knee Replacement with a Zirconia Ceramic Femoral Component B. S. Bal, D. D. Greenberg and T. J. Aleto Abstract This report presents the minimum two-year

More information

Ito Hiroshi, Tanino Hiromasa, Yamanaka Yasuhiro, Minami Akio, Matsuno Tateo

Ito Hiroshi, Tanino Hiromasa, Yamanaka Yasuhiro, Minami Akio, Matsuno Tateo Journal of arthroplasty (2013.Feb) 28 巻 2 号 :309~314. Intermediate to long-term results of periacetabular osteotomy in patients younger and older than forty years of age Ito Hiroshi, Tanino Hiromasa, Yamanaka

More information

Metal-on-Metal Bearings in Cementless Primary Total Hip Arthroplasty

Metal-on-Metal Bearings in Cementless Primary Total Hip Arthroplasty The Journal of Arthroplasty Vol. 19 No. 8 Suppl. 3 2004 Metal-on-Metal Bearings in Cementless Primary Total Hip Arthroplasty Christian P. Delaunay, MD Abstract: One hundred cementless titanium primary

More information

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing Journal of Orthopaedic Surgery 2001, 9(1): 45 50 Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing KY Chiu, TP Ng, WM Tang and P Lam Department of Orthopaedic Surgery, The University

More information

The Ringloc liner compared with the Hexloc liner in total hip arthroplasty

The Ringloc liner compared with the Hexloc liner in total hip arthroplasty The Ringloc liner compared with the Hexloc liner in total hip arthroplasty Olof Sköldenberg, Mats Salemyr, Olle Muren, Åke Johansson, Torbjörn Ahl, Henrik Bodén Karolinska Institutet, Department of Clinical

More information

Active Articulation E1

Active Articulation E1 Active Articulation E1 Dual Mobility Hip System Design Rationale Knees Hips Extremities Cement and Accessories PMI Trauma Technology Active Articulation E1 Dual Mobility Hip System Dual Mobility: Not a

More information

HYDROXYAPATITE IN REVISION OF TOTAL HIP REPLACEMENTS WITH MASSIVE ACETABULAR DEFECTS

HYDROXYAPATITE IN REVISION OF TOTAL HIP REPLACEMENTS WITH MASSIVE ACETABULAR DEFECTS HYDROXYAPATITE IN REVISION OF TOTAL HIP REPLACEMENTS WITH MASSIVE ACETABULAR DEFECTS 4- TO 10-YEAR CLINICAL RESULTS H. OONISHI, Y. IWAKI, N. KIN, S. KUSHITANI, N. MURATA, S. WAKITANI, K. IMOTO From Osaka-Minami

More information

Zimmer Longevity Highly Cross-linked Polyethylene. Clinical Value Dossier - APAC

Zimmer Longevity Highly Cross-linked Polyethylene. Clinical Value Dossier - APAC Zimmer Longevity Highly Cross-linked Polyethylene Clinical Value Dossier - APAC Table of Contents 1 Executive Summary...................................................... 1 2 Burden of Illness........................................................

More information

BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code

BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code Patrick Chaboche, MD patrick.chaboche@wanadoo.fr Former Head of Department of Orthopaedic Surgery. Centre

More information

TOTAL HIP REPLACEMENT:

TOTAL HIP REPLACEMENT: THR Prosthesis Design TOTAL HIP REPLACEMENT: PROSTHESIS DESIGN FEATURES JESS JOHNSTON & MELINDA ZIETH History of Hip Prosthesis Joint Replacement Registry Implant Design Technology & Future History and

More information

TriboFit Hip System. Issue 3 - August 2013 English

TriboFit Hip System. Issue 3 - August 2013 English Important Information: Please read prior to use in a clinical setting. The surgeon should be familiar with the operative technique and all the information in this insert. Description: The TriboFit Hip

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending the proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Effect of Femoral Head Size to Contact Stress at Pure Titanium Femoral Ball Head Outer Surface of Hip Joint Implant

Effect of Femoral Head Size to Contact Stress at Pure Titanium Femoral Ball Head Outer Surface of Hip Joint Implant Effect of Femoral Head Size to Contact Stress at Pure Titanium Femoral Ball Head Outer Surface of Hip Joint Implant Agung Setyo Darmawan, Waluyo Adi Siswanto, Haryanti Samekto Faculty of Mechanical and

More information

CONGENITAL HIP DISEASE IN YOUNG ADULTS CLASSIFICATION AND TREATMENT WITH THA. Th. KARACHALIOS, MD, DSc PROF IN ORTHOPAEDICS

CONGENITAL HIP DISEASE IN YOUNG ADULTS CLASSIFICATION AND TREATMENT WITH THA. Th. KARACHALIOS, MD, DSc PROF IN ORTHOPAEDICS CONGENITAL HIP DISEASE IN YOUNG ADULTS CLASSIFICATION AND TREATMENT WITH THA Th. KARACHALIOS, MD, DSc PROF IN ORTHOPAEDICS EDITOR IN CHIEF HIP INTERNATIONAL UNIVERSITY OF THESSALIA, LARISA HELLENIC REPUBLIC

More information

Wear Properties of UHMWPE in CHARITE Artificial Inter-vertebral Disc

Wear Properties of UHMWPE in CHARITE Artificial Inter-vertebral Disc Mechanics of Contact and Lubrication, MTM G230 Department of Mechanical & Industrial Enineering Northeastern University Spring 2006 Wear Properties of UHMWPE in CHARITE Artificial Inter-vertebral Disc

More information

Fracture of an Acetabular Component Inserted without Cement: A Case Report

Fracture of an Acetabular Component Inserted without Cement: A Case Report Syracuse University SURFACE Biomedical and Chemical Engineering College of Engineering and Computer Science 1997 Fracture of an Acetabular Component Inserted without Cement: A Case Report R T. Trousdale

More information