1 Translated English of Chinese Standard: YY ICS YY C 43 PHARMACEUTICAL INDUSTRY STANDARD OF THE PEOPLE S REPUBLIC OF CHINA YY Stereotactic radiotherapy system with gamma beam Part 1: Multi-source stereotactic radiotherapy system with gamma beam for head lesion γ 射束立体定向放射治疗系统第 1 部分 : 头部多源 γ 射束立体定向放射治疗系统 Issued on: December 31, 2011 Implemented on: June 1, 2013 Issued by: State Food and Drug Administration Page 1 of 16

2 Table of Contents Foreword Scope Normative references Terms and definitions Requirements Test methods... 8 Annex A (Informative) Materials, structure and size of special spherical die body and special focus measuring tools for test Page 2 of 16

3 Foreword All technical contents of this Part are mandatory. YY 0831 is consisted of 2 parts as follows: Part 1: Multi-source stereotactic radiotherapy system with gamma beam for head lesion; Part 2: Multi-source stereotactic radiotherapy system with gamma beam for body lesion. This Part is part 1 of YY This Part is drafted according to the rules specified in GB/T Please note that some contents in this Document may involve patents. The issuing authority of this Document does not undertake the responsibilities to identify these patents. The part was proposed by China Food and Drug Administration. The part shall be under the jurisdiction of National Standardization Technical Committee of Medical Appliances Radiotherapy and Technical Committee of Nuclear Medicine and Radiation Dosimetry Equipment (SAC/TC10/SC3). Drafting organization of this Part: Beijing Institute of Medical Device Testing, Shenzhen Haibo Technology Co., Ltd., Medical Science & Technology Development (Shenzhen) Co., Ltd. (MASEP), Shenzhen Shengai Medical Technology Development Co., Ltd., and Shenzhen Zunrui Technology Co., Ltd. The main drafters of the part: Zhang Zhaoyuan, Zhang Xin, Song Lianyou, Chen Jing, Zhu Weiqun, Xu Tao, Zheng Tie, and Liu Guangwu. Page 3 of 16

4 Stereotactic radiotherapy system with gamma beam Part 1: Multi-source stereotactic radiotherapy system with gamma beam for head lesion 1 Scope This Part of YY 0831 specifies the scope, terms, definitions, requirements, and test methods of the multi-source stereotactic radiotherapy system with gamma beam for head lesion. This Part applies to the multi-source stereotactic radiotherapy system with gamma beam for head lesion (hereinafter referred to as the system). The system simultaneously uses multiple 60 Co sealed radioactive sources (it may be dynamic or static) to conduct beaming irradiation to the head lesion. 2 Normative references The articles contained in the following documents have become part of this Document when they are quoted herein. For the dated documents so quoted, all the modifications (Including all corrections) or revisions made thereafter shall be applicable to this Document. GB Medical electrical equipment - Part 1: General requirements for safety GB Medical electrical equipment - Part 1: General requirements for safety - Collateral standard: Safety requirements for medical electrical systems GB Medical electrical equipment - Part 2: Particular requirements for the safety of gamma beam therapy equipment GB/T Medical radiology Terminology (Equipment for radiotherapy, nuclear medicine and radiation dosimetry) GB/T Radiotherapy equipment - Coordinates, movements and scales YY Medical electrical equipment - Requirements for the safety of radiotherapy therapeutic planning systems Page 4 of 16

5 c) For the focused field of which the nominal size is greater than 20 mm and less than or equal to 30 mm, the dose gradient of each focused field shall not exceed 10 mm. 4.6 Deviation of positioning reference point Deviation of the positioning reference point of the system shall meet the following requirements: a) For the minimum focused field, the deviation between the positioning reference point and the focus shall be less than or equal to 0.5 mm; b) The deviations of positioning reference point for each focused field shall be included in accompanying documents. 4.7 Integrated error of dose calculation For each focused field, the error between the absorbent dose value calculated by the therapeutic planning software AND the actual measured value shall not exceed ±5%. 4.8 Position error of therapeutic plan reference point Reconstruction position error of therapeutic planning software shall not exceed 1.5 mm. 4.9 Safety requirements The system shall meet the requirements of GB , GB , GB and YY Test methods 5.1 Coordinate system Check the accompanying documents. It shall comply with the requirements of Accompanying documents Check the accompanying documents. It shall comply with the requirements of Nominal absorbed dose rate of the focus The ionization chamber and the dosimeter used for measurement shall meet the requirements of the measurement conditions. Install the special spherical die body (hereinafter referred to as "Special Die body", see Appendix A) of which the diameter is 160 mm on the positioning bracket. The die body s center is located at the positioning reference point. Insert it into ionization chamber, so as to make its effective measurement point to coincide with the die body s center. Transfer the special die body along with the therapeutic bed into a predetermined irradiation position. Use the maximum focused field. Page 8 of 16

6 minimum focused field; Dmax Dosimeter readings at the focus in special spherical die body when using the maximum focused field. 5.4 Focused field size Grayscale-dose calibration Radiate a series of films in the allowed test area of grayscale-dose response curve of the film. Use a suitable mathematical model according to the radiating dose values and the corresponding film s grayscale value, Draw the grayscale-dose response curve for the radiation films used Focused field film radiation According to following steps to respectively measure the maximum deviations of the focused field size and the nominal value: a) Select a set of collimator for therapy. Measure at least on 2 focal planes; b) Put the film of which the size fits the measurement of focused field onto the central position of special die body. Let the film to be located on some focal plane; c) Transfer the special die body, together with the therapeutic bed, into focus position. Start to radiate. Radiation dose shall be within the appropriate response area of the grayscale-dose response curve of the film used. d) Repeat b). Let the film to be located on another focal plane. Repeat the radiation operation of c); e) Replace the collimator. Repeat tests b)-e). Complete the radiation of each group of focused field; f) Perform the therapy to the film Calculation of focused field size According to following steps to calculate the focused field size: a) Use a scanner to scan the film. The resolution shall not be less than 300 DPI. Respectively scan the focused field films of 5.4.2; b) Use the grayscale-dose response curve of to convert the film grayscale TO corresponding dose values. Draw a dose distribution curve on directions of 3 axes. c) Separately find out the dose of each film at positioning reference point. Define the dose value as the 100% dose value of the film; Page 10 of 16

7 planning software AND the actual measured absorbed dose value shall comply with the requirements of 4.7. Calculate the absorbed doses at 3 reference points. Calculate the relative deviation according to Formula (6); Where: ΔP Relative deviation between focus planned dose and measured dose, %; P0 Planned dose at die body center (i.e., focused plan dose), in unit of Gy; P1 Actual measured absorbed dose at die body center, in unit of Gy. 5.8 Position deviation of therapeutic planning reference point According to the image parameter s requirements that are given by accompanying documents and are required to achieve the reconstruction precision of three-dimensional image of therapeutic planning software, fix the imaged head die body that is installed with 3 known target volumes INTO the stereotaxic positioning apparatus. Mount together on an imaging system to scan, so as to obtain satisfactory image. Input the image data into therapeutic planning system. Calculate the coordinates of the 3 target volumes. Respectively calculate the position coordinates of the center of the 3 target volumes in the 3 directions of X, Y, Z. Calculate the reconstruction position deviation of the three-dimensional image of therapeutic planning software according to Formula (7): Where: Δ2 The distance between the center of the target volume and reconstruction target volume of therapeutic planning software, in unit of mm (mm); ΔX2 In X-axis direction, the distance between the center of the target volume and reconstruction target volume of therapeutic planning software, in unit of millimeter (mm); ΔY2 In Y-axis direction, the distance between the center of the target volume and reconstruction target volume of therapeutic planning software, in unit of millimeter (mm); ΔZ2 In Z-axis direction, the distance between the center of the target volume and reconstruction target volume of therapeutic planning software, in unit of Page 13 of 16

8 millimeter (mm). Check the accompanying documents. The maximum deviation s calculation results of reconstruction position of the center of the three target volumes shall comply with the requirements of 4.8. At the same time, it shall report ΔX2, ΔY2, ΔZ Safety requirements Implement it according to the methods specified in GB , GB , GB , and YY Page 14 of 16

YY/T Translated English of Chinese Standard: YY/T

YY/T Translated English of Chinese Standard: YY/T Translated English of Chinese Standard: YY/T1218-2013 www.chinesestandard.net Email: Sales@ChineseStandard.net YY ICS 11.100 C 44 PHARMACEUTICAL INDUSTRY STANDARD OF THE PEOPLE S REPUBLIC OF CHINA YY/T

More information

YY / ISO 11070:1998

YY / ISO 11070:1998 Translated English of Chinese Standard: YY0450.1-2003 www.chinesestandard.net Sales@ChineseStandard.net PHARMACEUTICAL INDUSTRY STANDARD YY OF THE PEOPLE S REPUBLIC OF CHINA ICS 11.120.30 C 48 YY 0450.1-2003

More information

YY/T PHARMACEUTICAL INDUSTRY STANDARD OF THE PEOPLE S REPUBLIC OF CHINA. Translated English of Chinese Standard: YY/T

YY/T PHARMACEUTICAL INDUSTRY STANDARD OF THE PEOPLE S REPUBLIC OF CHINA. Translated English of Chinese Standard: YY/T PHARMACEUTICAL INDUSTRY STANDARD OF THE PEOPLE S REPUBLIC OF CHINA Translated English of Chinese Standard: YY/T0520-2009 www.chinesestandard.net Sales@ChineseStandard.net YY ICS 11.060.10 C 33 YY/T 0520-2009

More information

YY/T Translated English of Chinese Standard: YY/T PHARMACEUTICAL INDUSTRY STANDARD

YY/T Translated English of Chinese Standard: YY/T PHARMACEUTICAL INDUSTRY STANDARD Translated English of Chinese Standard: YY/T1518-2017 www.chinesestandard.net Buy True-PDF Auto-delivery. Sales@ChineseStandard.net PHARMACEUTICAL INDUSTRY STANDARD YY OF THE PEOPLE S REPUBLIC OF CHINA

More information

YY Translated English of Chinese Standard: YY PHARMACEUTICAL INDUSTRY STANDARD

YY Translated English of Chinese Standard: YY PHARMACEUTICAL INDUSTRY STANDARD Translated English of Chinese Standard: YY0766-2009 www.chinesestandard.net Sales@ChineseStandard.net PHARMACEUTICAL INDUSTRY STANDARD YY OF THE PEOPLE S REPUBLIC OF CHINA ICS 11.040.70 C 41 YY 0766-2009

More information

YY Translated English of Chinese Standard: YY

YY Translated English of Chinese Standard: YY Translated English of Chinese Standard: YY0497-2005 www.chinesestandard.net Sales@ChineseStandard.net YY ICS 11.040.20 C 31 Pharmaceutical National Standard of the People s Republic of China YY 0497-2005

More information

YY/T Translated English of Chinese Standard: YYT PHARMACEUTICAL INDUSTRY STANDARD

YY/T Translated English of Chinese Standard: YYT PHARMACEUTICAL INDUSTRY STANDARD Translated English of Chinese Standard: YYT0325-2016 www.chinesestandard.net Buy True-PDF Auto-delivery. Sales@ChineseStandard.net PHARMACEUTICAL INDUSTRY STANDARD YY OF THE PEOPLE S REPUBLIC OF CHINA

More information

GB/T Translated English of Chinese Standard: GB/T

GB/T Translated English of Chinese Standard: GB/T Translated English of Chinese Standard: GB/T 19634-2005 www.chinesestandard.net Sales@ChineseStandard.net GB ICS 11.100 C 44 National Standard of the People s Republic of China GB/T 19634-2005 In vitro

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. YY/T Translated English of Chinese Standard: YY/T

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. YY/T Translated English of Chinese Standard: YY/T Translated English of Chinese Standard: YY/T0343-2002 www.chinesestandard.net Sales@ChineseStandard.net PHARMACEUTICAL INDUSTRY STANDARD YY OF THE PEOPLE S REPUBLIC OF CHINA ICS 11.040.40 C 30 Liquid penetrant

More information

YY Translated English of Chinese Standard: YY

YY Translated English of Chinese Standard: YY Translated English of Chinese Standard: YY0341-2009 www.chinesestandard.net Sales@ChineseStandard.net YY ICS 11.040.40 C 35 Pharmaceutical Industry Standard of the People s Republic of China YY/T 0341-2009

More information

JJF Translated English of Chinese Standard: JJF

JJF Translated English of Chinese Standard: JJF Translated English of Chinese Standard: JJF1070.2-2011 www.chinesestandard.net Sales@ChineseStandard.net JJF NATIONAL METROLOGICAL TECHNICAL SPECIFICATION OF THE PEOPLE S REPUBLIC OF CHINA JJF 1070.2-2011

More information

A Method for Statistical Process Control of Radiation Sterilization Facilities

A Method for Statistical Process Control of Radiation Sterilization Facilities A Method for Statistical Process Control of Radiation Sterilization Facilities 2006 A Method for Statistical Process Control (SPC) of Radiation Sterilization Facilities Scope This document describes a

More information

鱼类罐头卫生标准 Hygienic standard for canned fish

鱼类罐头卫生标准 Hygienic standard for canned fish Translated English of Chinese Standard: GB14939-2005 Translated by: www.chinesestandard.net Wayne Zheng et al. Email: Wayne@ChineseStandard.net ICS 67.040 C 53 GB National Standard of the People's Republic

More information

GB / ISO :2006

GB / ISO :2006 Translated English of Chinese Standard: GB18280.2-2015 www.chinesestandard.net Sales@ChineseStandard.net ICS 11.080.01 C 47 NATIONAL STANDARD OF THE PEOPLE'S REPUBLIC OF CHINA GB / ISO 11137-2:2006 Partly

More information

Sterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control

Sterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 11137-3 Second edition 2017-06 Sterilization of health care products Radiation Part 3: Guidance on dosimetric aspects of development, validation

More information

Hip Joint Prostheses

Hip Joint Prostheses Translated English of Chinese Standard: YY0118-2005 www.chinesestandard.net Sales@ChineseStandard.net PHARMACEUTICAL INDUSTRY STANDARD YY ICS 11.040.40 OF THE PEOPLE S REPUBLIC OF CHINA C 35 YY 0118-2005

More information

GB/T Translated English of Chinese Standard: GB/T NATIONAL STANDARD OF THE

GB/T Translated English of Chinese Standard: GB/T NATIONAL STANDARD OF THE Translated English of Chinese Standard: GB/T24253-2009 www.chinesestandard.net Sales@ChineseStandard.net GB ICS 59.080.30 W 04 NATIONAL STANDARD OF THE PEOPLE'S REPUBLIC OF CHINA GB/T 24253-2009 Textiles

More information

Procedure. Identify all possible radiation hazards.

Procedure. Identify all possible radiation hazards. Guidance Notes on Radiation Risk Assessment Prior to commencing any new work practice involving a source of ionising radiation it is important that a realistic assessment of the radiation risks is carried

More information

GB Translated English of Chinese Standard: GB NATIONAL STANDARD

GB Translated English of Chinese Standard: GB NATIONAL STANDARD Translated English of Chinese Standard: GB5009.5-2016 www.chinesestandard.net Sales@ChineseStandard.net GB NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA GB 5009.5-2016 National food safety standard

More information

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE Translated English of Chinese Standard: GB5009.268-2016 www.chinesestandard.net Buy True-PDF Auto-delivery. Sales@ChineseStandard.net GB NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA GB 5009.268-2016

More information

National Food Safety Standard Infant Formula

National Food Safety Standard Infant Formula Translated English of Chinese Standard: GB 10765-2010 www.chinesestandard.net Email: Sales@ChineseStandard.net NATIONAL STANDARD GB OF THE PEOPLE S REPUBLIC OF CHINA GB 10765-2010 National Food Safety

More information

Dosimetry of the Gamma Knife TM

Dosimetry of the Gamma Knife TM Dosimetry of the Gamma Knife TM Sheridan Griffin Meltsner, Ph.D. Aurora Health Care St. Luke s Medical Center NCCAAPM Spring Meeting Outline Introduction to stereotactic radiosurgery Elekta Gamma Knife

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard GB15980-1995 Translated by: www.chinesestandard.net Email: Sales@ChineseStandard.net ICS 11.120.20 C 59 National Standard of the People s Republic of China GB Hygienic

More information

Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects

Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects AAMI/FDS-1 11137-03 2006-01-12 (Revision of ANSI/AAMI/ISO 11137:1994; AAMI TIR27:2001; AAMI/ISO TIR13409:1996) Final draft American National Standard AAMI/ American National Standard NOTE - This document

More information

Activity report from JCOG physics group

Activity report from JCOG physics group 2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer

More information

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF Translated English of Chinese Standard: GB5009.4-2016 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF GB THE PEOPLE S REPUBLIC OF CHINA GB 5009.4-2016 National Food Safety Standard

More information

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV M. Mumot, G. V. Mytsin, Y. I. Luchin and A. G. Molokanov Medico-Technical

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. NB/T

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. NB/T Translated English of Chinese Standard: GB/T47013.5-2015 www.chinesestandard.net Sales@ChineseStandard.net ENERGY INDUSTRY STANDARD OF THE PEOPLE S REPUBLIC OF CHINA NB ICS 77.140.20 H 26 GB/T 47013.5-2015

More information

CT Quality Control Manual FAQs

CT Quality Control Manual FAQs CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Table of contents 1 Scope... 3

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Table of contents 1 Scope... 3 Translated English of Chinese Standard: GB28050-2011 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD GB OF THE PEOPLE S REPUBLIC OF CHINA National Standard on Food Safety National label

More information

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 12749-4 First edition 2015-08-15 Nuclear energy, nuclear technologies, and radiological protection Vocabulary Part 4: Dosimetry for radiation processing

More information

High-Dose Standardization Study For γ-rays radiation processing at NIM

High-Dose Standardization Study For γ-rays radiation processing at NIM High-Dose Standardization Study For γ-rays radiation processing at NIM ZHANG Yanli, GAO Junchen, YANG Yuandi. Division of Metrology in Ionizing Radiation and medicine, National Institute of Metrology,

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE Translated English of Chinese Standard: GB5009.85-2016 www.chinesestandard.net Buy True-PDF Auto-delivery. Sales@ChineseStandard.net GB NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA GB 5009.85-2016

More information

Dosimetry Standards and Dissemination Systems for Radiation Processing in China

Dosimetry Standards and Dissemination Systems for Radiation Processing in China Dosimetry Standards and Dissemination Systems for Radiation Processing in China Zhang Yanli National Institute of Metrology, China April 2017 in Vienna, Austria Organization Founded in 1955, National Institute

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB/T

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB/T Translated English of Chinese Standard: GB/T5009.197-2003 www.chinesestandard.net Sales@ChineseStandard.net ICS 67.040 C 53 NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA GB Determination of thiamine

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB Translated English of Chinese Standard: GB1987-2007 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA GB ICS 67.220.20 X 41 Replacing GB 1987-1986 Food

More information

Amendment No. 2. Item No. 2 (Rfx/ Event number )

Amendment No. 2. Item No. 2 (Rfx/ Event number ) Amendment 2 Sub: Amendment to the Document 06.12.2018 Ref.: Notice Inviting Bid ref. HITES/PCD/NCI-AIIMS/36/18-19 dated 26.09.2018 read with its Amendment no. 1 dated 19.11.18 The following changes have

More information

Agenda: Dental Cone Beam Imaging

Agenda: Dental Cone Beam Imaging Cone Beam Imaging Agenda: Dental Cone Beam Imaging *Definition and Functionality *Usage and diagnostics benefits *Comparative radiation information *Federal regulatory responsibilities: manufacturing *State

More information

Code of Practice for Radiation Protection in Dentistry. Code of Practice For Radiation Protection in Dentistry

Code of Practice for Radiation Protection in Dentistry. Code of Practice For Radiation Protection in Dentistry Code of Practice for Radiation Protection in Dentistry Code of Practice For Radiation Protection in Dentistry 10 OCTOBER 2017 CONTENTS 1. INTRODUCTION... 3 1.0 CITATION... 3 1.1 BACKGROUND... 3 1.2 PURPOSE

More information

Implementation Guide for the DoseControl Dosimetry System

Implementation Guide for the DoseControl Dosimetry System 1.0 SCOPE The GEX DoseControl System is intended for use as a single dosimetry system solution, specifically developed to satisfy the generic dosimetry needs of radiation sterilization application users

More information

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions)

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions) B3-20 Database System for GEX B3 Dosimetry (gamma and/or electron beam versions) and GEX combine to provide a Genesys 20 version of the flexible 21 CFR Part 11 certified database dosimetry system for use

More information

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF Translated English of Chinese Standard: GB13078-2001 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF GB THE PEOPLE S REPUBLIC OF CHINA ICS 65.120 B 46 GB 13078-2001 Replacing GB

More information

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics

More information

Evaluation of Several Dosimeters for Identification of Irradiated Foods Using a 5 MeV Electron Beam

Evaluation of Several Dosimeters for Identification of Irradiated Foods Using a 5 MeV Electron Beam Journal of Health Science, 48(1) 37 41 (2002) 37 Evaluation of Several Dosimeters for Identification of Irradiated Foods Using a 5 MeV Electron Beam Makoto Miyahara,*, a Hitoshi Ito, b Kouji Ueno, c Yutaka

More information

OIML R 131 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 131 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 131 Edition 2001 (E) Polymethylmethacrylate dosimetry systems for ionizing radiation processing of materials and products Systèmes de dosimétrie à polyméthylméthacrylate

More information

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte, MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 MODELLING A GAMMA IRRADIATION

More information

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development. Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY Room B1002, WIMR 1111 Highland Avenue Madison, WI 53705-2275 Larry A. DeWerd,

More information

Process Control Methods in Radiation Technologies

Process Control Methods in Radiation Technologies Process Control Methods in Radiation Technologies András Kovács Nuclear Security Department Centre for Energy Research Hungarian Academy of Sciences Radiation Processing Applications Radiation sterilization:

More information

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,

More information

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry Journal of Physics: Conference Series PAPER OPEN ACCESS Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry To cite this article: N A Costa et al 2016 J. Phys.: Conf. Ser. 733 012098 View

More information

Market Survey. Technical Description Irradiation Tests

Market Survey. Technical Description Irradiation Tests EDMS No. 1959184 The R2E Project Group Code: EN-STI Market Survey Technical Description Irradiation Tests Abstract This Technical Description concerns the performance of irradiation testing of CERN samples.

More information

Survey of patients CT radiation dose in Jiangsu Province

Survey of patients CT radiation dose in Jiangsu Province Original Article Page 1 of 6 Survey of patients CT radiation dose in Jiangsu Province Yuanyuan Zhou 1, Chunyong Yang 1, Xingjiang Cao 1, Xiang Du 1, Ningle Yu 1, Xianfeng Zhou 2, Baoli Zhu 1, Jin Wang

More information

FZ/T Translated English of Chinese Standard: FZ/T TEXTILE INDUSTRY STANDARD

FZ/T Translated English of Chinese Standard: FZ/T TEXTILE INDUSTRY STANDARD Translated English of hinese Standard: FZ/T01026-2009 www.hinesestandard.net Sales@hineseStandard.net TEXTILE INUSTRY STNR FZ OF THE PEOPLE S REPULI OF HIN IS 59.080.01 W 04 FZ/T 01026-2009 Replacing FZ/T

More information

Custom-Made Products / Scientific Research Instruments

Custom-Made Products / Scientific Research Instruments Synchrotron Radiation Instruments Double Crystal Monochromator A double crystal monochromator is an instrument that extracts light of a specific wavelength using crystal diffraction. Light of various wavelengths

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale OIML INTERNATIONAL RECOMMENDATION Equipment for speech audiometry Appareils pour l audiométrie vocale OIML R 122 Edition 1996 (E) CONTENTS Foreword... 3

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009. Translated English of Chinese Standard: GB5009.259-2016 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD GB OF THE PEOPLE S REPUBLIC OF CHINA National food safety standard Determination

More information

Radiologic Units: What You Need to Know

Radiologic Units: What You Need to Know Radiologic Units: What You Need to Know TODD VAN AUKEN M.ED. RT (R)(MR) Agenda Greys, Sieverts, Coulombs per kg, & Becquerel's Conventional Units Other Concepts (LET, Q-Factor, Effective Dose, NCRP Report

More information

FEE RULES RADIATION ONCOLOGY FEE SCHEDULE CONTENTS

FEE RULES RADIATION ONCOLOGY FEE SCHEDULE CONTENTS Tel: +27-21-9494060 Fax: +27-21-9494112 E-mail: leon.gouws@cancercare.co.za FEE RULES RADIATION ONCOLOGY FEE SCHEDULE CONTENTS 1. EXTERNAL BEAM RADIATION... 2 2. PLANNING OF TREATMENT... 2 3. DELIVERY

More information

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY E. BADITA 1, C.VANCEA 1,3, I. CALINA 1,3, D. STROE 2, M. DUMITRACHE 2,3,

More information

5th ADAMAS Workshop at GSI December 15-16, 2016, Darmstadt, Germany

5th ADAMAS Workshop at GSI December 15-16, 2016, Darmstadt, Germany Evaluation of 3D diamond detectors for application in medical radiation dosimetry K. Kanxheri (1,2), L. Servoli (2), C. Zucchetti (5), A. C. Dipilato (5), M. Iacco (5), S. Lagomarsino (3,4), A. Morozzi

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute

More information

Integrated Research Application System (IRAS)

Integrated Research Application System (IRAS) Integrated Research Application System (IRAS) Question specific guidance Part B Section 3 Exposure to ionising radiation The following document collates all guidance for the questions in Part B: Section

More information

Dosisverifikation mit Delta 4 Discover während der Behandlung

Dosisverifikation mit Delta 4 Discover während der Behandlung Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover

More information

RELIANT HOLDINGS LTD AND ITS AFFILIATES Safety Management System. Preparation: Safety Mgr Authority: CEO Issuing Dept: Safety Page: Page 1 of 5

RELIANT HOLDINGS LTD AND ITS AFFILIATES Safety Management System. Preparation: Safety Mgr Authority: CEO Issuing Dept: Safety Page: Page 1 of 5 Preparation: Safety Mgr Authority: CEO Issuing Dept: Safety Page: Page 1 of 5 Purpose The purpose of this program is to protect employees who may encounter ionizing radiation and its hazards while performing

More information

Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies

Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies J. Y. C. Cheung and K. N. Yu a Department of Physics and Materials Science, City University of Hong Kong, Kowloon Tong,

More information

ISO INTERNATIONAL STANDARD. Dentistry Implants Dynamic fatigue test for endosseous dental implants

ISO INTERNATIONAL STANDARD. Dentistry Implants Dynamic fatigue test for endosseous dental implants INTERNATIONAL STANDARD ISO 14801 Second edition 2007-11-15 Dentistry Implants Dynamic fatigue test for endosseous dental implants Art dentaire Implants Essai de fatigue dynamique pour implants dentaires

More information

Skyscan 1076 in vivo scanning: X-ray dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of

More information

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 6565 Fifth edition 2015-12-15 Tobacco and tobacco products Draw resistance of cigarettes and pressure drop of filter rods Standard conditions and measurement Tabac et produits

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

Draft for comments only Not to be cited as East African Standard

Draft for comments only Not to be cited as East African Standard EAST AFRICAN STANDARD Poultry egg fertility tester Specification EAST AFRICAN COMMUNITY CD/K/636:2010 ICS 67.120.20 EAC 2010 First Edition 2010 CD/K/636:2010 Foreword Development of the East African Standards

More information

Training Course on Small Field Dosimetry

Training Course on Small Field Dosimetry Training Course on Small Field Dosimetry Argonne, Illinois, United States of America 10 14 December 2015 Ref. No: E2-TR-51583 Prospectus Course Title: Place (Venue, City, Country): Training Course on Small

More information

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Gonzalo Walwyn Salas, Stefan Gutiérrez Lores Center for Radiation Protection and Hygiene,

More information

Slide 1. Slide 2. Slide 3. Introduction of INTRABEAM IORT. Disclosure. Contents. I have nothing to disclose.

Slide 1. Slide 2. Slide 3. Introduction of INTRABEAM IORT. Disclosure. Contents. I have nothing to disclose. Slide 1 AAPM 2017 Introduction of INTRABEAM IORT Hualin Zhang, Ph.D. Dept of Radiation Oncology, Northwestern University, Northwestern Memorial Hospital, Chicago, Illinois August 2, 2017 Slide 2 Disclosure

More information

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague Radiosurgery by Leksell gamma knife Josef Novotny, Na Homolce Hospital, Prague Radiosurgery - Definition Professor Lars Leksell The tools used by the surgeon must be adapted to the task and where the human

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60601-2-37 2001 AMENDMENT 1 2004-08 Amendment 1 Medical electrical equipment Part 2-37: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring

More information

Appendix I. List of stakeholders consulted with on the Patient Radiation Protection Manual and members of the Medical Exposure Radiation Unit

Appendix I. List of stakeholders consulted with on the Patient Radiation Protection Manual and members of the Medical Exposure Radiation Unit References References The accuracy, quality and relevance of these works are not guaranteed or uniform and more recent information may have superseded these works. This list is not exhaustive. It does

More information

Radiochromic film dosimetry in water phantoms

Radiochromic film dosimetry in water phantoms INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) N27 N31 www.iop.org/journals/pb PII: S0031-9155(01)16858-2 NOTE Radiochromic film dosimetry in water phantoms

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source

Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source Amanda Bravim, Leticia Lucente Campos Instituto de Pesquisas Energéticas e Nucleares (IPEN / CNEN - SP) Av. Professor

More information

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,

More information

IMRT QA: Point Dose Measurements or 2D Array?

IMRT QA: Point Dose Measurements or 2D Array? QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify

More information

Radiation Safety Guide. Analytical X-Ray Equipment

Radiation Safety Guide. Analytical X-Ray Equipment Radiation Safety Guide Analytical X-Ray Equipment Table of Content Page 1. Radiation 2 A. Radiation Quantities 2 B. Background Radiation 2 C. Biological Effect of Radiation 3 D. Radiation Injury To The

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 37-013 GARD Users Manual March 2005 Manual No. 37-013-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective

More information

45 Hr PET Registry Review Course

45 Hr PET Registry Review Course 45 HR PET/CT REGISTRY REVIEW COURSE Course Control Document Timothy K. Marshel, MBA, R.T. (R), (N)(CT)(MR)(NCT)(PET)(CNMT) The PET/CT Training Institute, Inc. SNMMI-TS 028600-028632 45hr CEH s Voice Credits

More information

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report www.siemens.com/healthcare SOMATOM Drive System Owner Manual Dosimetry and imaging performance report Table of contents 1 Dosimetry and imaging performance report 5 1.1 Dose information 5 1.1.1 General

More information

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331

More information

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

CAC/RCP (Rev ) 1

CAC/RCP (Rev ) 1 CAC/RCP 19-1979 (Rev.1-1983) Page 1 RECOMMENDED INTERNATIONAL CODE OF PRACTICE FOR THE OPERATION OF IRRADIATION FACILITIES USED FOR THE TREATMENT OF FOODS CAC/RCP 19-1979 (Rev. 1-1983) 1 1. INTRODUCTION

More information

Mitsubishi Heavy Industries Technical Review Vol. 51 No. 1 (March 2014)

Mitsubishi Heavy Industries Technical Review Vol. 51 No. 1 (March 2014) The Challenge of Innovative Cancer Treatments Enabled by Vero4DRT -Development of High-precision Dose Delivery Features for Reducing Radiation Exposure of Healthy Tissue- 76 YASUNOBU SUZUKI *1 KUNIO TAKAHASHI

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB1886.

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB1886. Translated English of Chinese Standard: GB1886.235-2016 www.chinesestandard.net Buy True-PDF Auto-delivery. Sales@ChineseStandard.net NATIONAL STANDARD OF THE GB PEOPLE S REPUBLIC OF CHINA National Food

More information