1 Diversity in Medical Applications: The Linac Example Brian Bibb 19 January 2005 Outline Describe linac Describe some safety systems Arrive at some conclusions Diversity in Medical Applications: The Linac Example - 2

2 Treatment on a linac Diversity in Medical Applications: The Linac Example - 3 Layout of Bunker Diversity in Medical Applications: The Linac Example - 4

3 Linac Diversity in Medical Applications: The Linac Example - 5 Linear accelerator Diversity in Medical Applications: The Linac Example - 6

4 Treatment head Diversity in Medical Applications: The Linac Example - 7 Radiation profiles Diversity in Medical Applications: The Linac Example - 8

5 Risks Accidents Incorrect dose Incorrect volume Collision Cause Equipment error Use error Risk Control is often diversity Diversity in Medical Applications: The Linac Example - 9 Control room equipment 1Control cabinet 2 Keyboard 3 Keypad 4 Mouse 5 Monitor Diversity in Medical Applications: The Linac Example - 10

6 GUI Diversity in Medical Applications: The Linac Example - 11 Diverse Elements Prescribed and Backup MU Set and Actual Diaphragm numerical and visual display Gantry and head angle - mechanical and GUI display Segment timer Diversity in Medical Applications: The Linac Example - 12

7 Turn off radiation Emergency OFF Supplies Hardware and software control HT PRF Magnetron Linear accelerator Diversity in Medical Applications: The Linac Example - 13 Diverse elements No HT no radiation No PRF no radiation No mains power no radiation Diversity in Medical Applications: The Linac Example - 14

8 Start and Terminate Key START Button Hardware interlock Software TERMINATE Button Hardware interlock Software Diversity in Medical Applications: The Linac Example - 15 Diverse elements Hardwired START button Regulatory requirement Software control of START Hardwired TERMINATE button Software can always TERMINATE A reset key-switch must be operated in the event of an abnormal termination Diversity in Medical Applications: The Linac Example - 16

9 System level checks Tolerance checking Drive circuits Supplies Set value Actual value Supplies Monitor circuits Diversity in Medical Applications: The Linac Example - 17 Diverse elements Separate Demand and Monitor circuits Separate display of SET and ACTUAL values Visual and Automatic comparison of SET and actual values Separate supplies for Demand and Monitor circuits Diversity in Medical Applications: The Linac Example - 18

10 Course and fine pots +V Course -V +V -V Fine Position is also on mechanical scale Diversity in Medical Applications: The Linac Example - 19 Diverse elements Mechanical and electronic scales Consistent offset for the 2 wipers on Fine pot Defined relationship for Fine pot wipers and Course pot wiper Diversity in Medical Applications: The Linac Example - 20

11 Position A or B detection Position A Transition Position B The switches are: -part of drive and decode circuits -decoded in both hardware and software Indicates a fault Diversity in Medical Applications: The Linac Example - 21 Diverse elements Set position separate from monitor Software checks movement time FPLA and SW Monitors position: both check consistency with linac state Diversity in Medical Applications: The Linac Example - 22

12 Beam monitoring Radiation beam Ion chamber 1 Ion chamber 2 Diversity in Medical Applications: The Linac Example - 23 Hardware counter Ch1 Software: Counts Ch 1 Compares counters Checks dose rate Times beam Hardware counter Ch2 Terminate beam Terminate beam Display on VDU Display on LCD Terminate beam Display on VDU Diverse elements Ion chambers have different gains HW counters have different scaling factors Hardware and Software Counters Timer check regulatory requirement Dose rate check Separate display on VDU and LCD Diversity in Medical Applications: The Linac Example - 24

13 Dose Integration 1MU Accumulated dose Σ = 1MU -1MU Duplicated for 1MU and 1/64 MU Diversity in Medical Applications: The Linac Example - 25 Discharge time Ideal Actual Missed dose Capacitor discharge time Diversity in Medical Applications: The Linac Example - 26

14 Self test Self test signals Ion chamber signals Digital to Analogue Analogue to Digital Signal processing Diversity in Medical Applications: The Linac Example - 27 Single pulse check Typical pulse Detect if a single pulse is too large Abnormal pulse Diversity in Medical Applications: The Linac Example - 28

15 Overflow check Ion chamber signal Analogue to digital Signal processing 1MU Reset Overflow Pulse detected Count Counter Diversity in Medical Applications: The Linac Example - 29 Other checks Watchdog on particular events PSU monitoring Diversity in Medical Applications: The Linac Example - 30

16 Free checks Free means no extra hardware required Is the gain correct? Do the interlocks work? Is the noise level (base level) reasonable? Is the noise level consistent? Diagnostic checks Diversity in Medical Applications: The Linac Example - 31 Some conclusions More complex increased potential for design errors More complex, better functionality e.g. less analogue distortions Safety is a system property System level checks can still be used Potential for additional free checks It is worth looking for non complex checks Diversity in Medical Applications: The Linac Example - 32

Intensity Modulated RadioTherapy

Intensity Modulated RadioTherapy Intensity Modulated RadioTherapy A clinical application of a small accelerator University Medical Center Groningen A.A. van t Veld PhD I. Hoveijn PhD part 1 part2 CERN/KVI accelerator school, Zeegse, June

More information

Amendment No. 2. Item No. 2 (Rfx/ Event number )

Amendment No. 2. Item No. 2 (Rfx/ Event number ) Amendment 2 Sub: Amendment to the Document 06.12.2018 Ref.: Notice Inviting Bid ref. HITES/PCD/NCI-AIIMS/36/18-19 dated 26.09.2018 read with its Amendment no. 1 dated 19.11.18 The following changes have

More information

Teaching linear accelerator physics using simulation software. Marco Carlone, BC Cancer Agency Nicole Harnett, Princess Margaret Cancer Centre

Teaching linear accelerator physics using simulation software. Marco Carlone, BC Cancer Agency Nicole Harnett, Princess Margaret Cancer Centre Teaching linear accelerator physics using simulation software Marco Carlone, BC Cancer Agency Nicole Harnett, Princess Margaret Cancer Centre October, 27, 2017 About the presenter Linac Physics is Complicated

More information

Simulating Potential Layouts for a Proton Therapy Treatment Center

Simulating Potential Layouts for a Proton Therapy Treatment Center Simulating Potential Layouts for a Proton Therapy Treatment Center Stuart Price-University of Maryland Bruce Golden- University of Maryland Edward Wasil- American University Howard Zhang- University of

More information

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications Varian Treatment Streamlined Treatment Delivery Management Application Specifications Specifications Varian Treatment 1 Introduction Streamlined Treatment Delivery Management Varian Treatment verifies

More information

Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO

Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO Investigations and research Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO M. Kunze-Busch P. van Kollenburg Department of Radiation Oncology, Radboud University Nijmegen Medical

More information

Indiana University Health Proton Therapy Center. Chee-Wai Cheng, Ph.D.

Indiana University Health Proton Therapy Center. Chee-Wai Cheng, Ph.D. Indiana University Health Proton Therapy Center Chee-Wai Cheng, Ph.D. Machine configuration and layout Fixed beam line, double scattering and propeller Uniform scanning nozzle and snout Gantry room Range

More information

Dosimeter Setting Device

Dosimeter Setting Device Instruction Manual Dosimeter Setting Device For Electronic Personal Dosimeter Dose-i (Unit:Sv, Version:1.05 English) WTA529748 a 1 / 38 Foreword Thank you for purchasing the Dosimeter Setting Device; a

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

A new geometric and mechanical verification device for medical LINACs

A new geometric and mechanical verification device for medical LINACs JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 A new geometric and mechanical verification device for medical LINACs Keith T. Welsh,* Robert A. Wlodarczyk, and L. E. Reinstein

More information

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA CHAPTER 5 STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA 5.1 Introduction With the advent of new techniques like intensity modulated radiotherapy

More information

N-550 TROUBLESHOOTING TROUBLE SHOOTING (N-550 PULSE OXIMETER)

N-550 TROUBLESHOOTING TROUBLE SHOOTING (N-550 PULSE OXIMETER) TROUBLE SHOOTING (N-550 PULSE OXIMETER) 1. Problem Categories. This document is made for service training personnel, and based on Service manual. Problems with the N-550 are separated into the categories

More information

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Can you currently use direct documentation and measurement of delivered dose to ensure you did not underdose or

More information

Lessons learned from accidents in conventional external radiotherapy

Lessons learned from accidents in conventional external radiotherapy Lessons learned from accidents in conventional external radiotherapy Ola Holmberg 1,2, Renate Czarwinski 2 1 Radiation Protection of Patients Unit 2 Radiation Safety and Monitoring Section Division of

More information

Introductory Motor Learning and Development Lab

Introductory Motor Learning and Development Lab Introductory Motor Learning and Development Lab Laboratory Equipment & Test Procedures. Motor learning and control historically has built its discipline through laboratory research. This has led to the

More information

SNJB College of Engineering Department of Computer Engineering

SNJB College of Engineering Department of Computer Engineering 1. Intel s programmable device (8253) facilitates the generation of accurate time delays. a) counter b) timer c) both a & b d) none of these 2. The programmable timer device (8253) contains three independent

More information

Frequently Asked Questions

Frequently Asked Questions Frequently Asked Questions How does the SMRT-Y work? The SMRT-Y sensor accurately measures the moisture level in the soil. When the soil is dry, the SMRT-Y allows the irrigation controller to water as

More information

Radiotherapy. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiotherapy. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Brachytherapy Radioisotopes in contact with the tumor Overview

More information

ALTERNATIVE DESIGNS FOR MEGA VOLTAGE MACHINES FOR CANCER TREATMENT IN DEVELOPING COUNTRIES

ALTERNATIVE DESIGNS FOR MEGA VOLTAGE MACHINES FOR CANCER TREATMENT IN DEVELOPING COUNTRIES ALTERNATIVE DESIGNS FOR MEGA VOLTAGE MACHINES FOR CANCER TREATMENT IN DEVELOPING COUNTRIES C BORRAS Pan American Health Organization, Washington, DC, USA XA9642848 H. SVENSSON Imtemational Atomic Energy

More information

Cerus P-Series General Specifications Section 15172

Cerus P-Series General Specifications Section 15172 Cerus P-Series s Section 15172 Variable Frequency Drives PART 1-GENERAL 1.01 SECTION INCLUDES A. Variable Frequency Drive(VFD) 1.02 RELATED SECTIONS A. Section 16195 1.03 REFERENCES A. NEMA250 B. EN 50178

More information

Digital. hearing instruments have burst on the

Digital. hearing instruments have burst on the Testing Digital and Analog Hearing Instruments: Processing Time Delays and Phase Measurements A look at potential side effects and ways of measuring them by George J. Frye Digital. hearing instruments

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

Quality Assurance in Radiation Therapy

Quality Assurance in Radiation Therapy Quality Assurance in Radiation Therapy Purpose (from George) To make aware of the needs for quality assurance in radiation therapy To identify components of a quality assurance program in radiation therapy

More information

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis Acknowledgments Thanks to Saiful Huq for an illuminating conversation about the application of TG-100 Jennifer O Daniel, Ph.D. & Fang-Fang Yin, Ph.D. Duke University Medical Center Annual AAPM Meeting,

More information

IORT with mobile linacs: the Italian experience

IORT with mobile linacs: the Italian experience IORT with mobile linacs: the Italian experience G. Tosi, M. Ciocca lntroduction At the beginning of 1999 a mobile Linac (Novac 7, manufactured by Hitesys, Aprilia, Italy) able to produce electron beams

More information

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System Varian Edge Experience Jinkoo Kim, Ph.D Henry Ford Health System Disclosures I participate in research funded by Varian Medical Systems. Outline of Presentation Review advanced imaging in Varian Edge Linear

More information

RDP/DDP REDUNDANT/DUPLEX DRIVE PACKAGE

RDP/DDP REDUNDANT/DUPLEX DRIVE PACKAGE RDP/DDP REDUNDANT/DUPLEX DRIVE PACKAGE USER MANUAL Franklin Control Systems 11/11/2013 Table of Contents 1. Safety precautions... 1 2. Legends & Introduction... 2 3. Installation and wiring... 3 4. Setup...

More information

-SQA-SCOTTISH QUALIFICATIONS AUTHORITY. Hanover House 24 Douglas Street GLASGOW G2 7NG NATIONAL CERTIFICATE MODULE DESCRIPTOR

-SQA-SCOTTISH QUALIFICATIONS AUTHORITY. Hanover House 24 Douglas Street GLASGOW G2 7NG NATIONAL CERTIFICATE MODULE DESCRIPTOR -SQA-SCOTTISH QUALIFICATIONS AUTHORITY Hanover House 24 Douglas Street GLASGOW G2 7NG NATIONAL CERTIFICATE MODULE DESCRIPTOR -Module Number- 0069166 -Session-1986-87 -Superclass- PF -Title- DENTAL RADIOGRAPHY

More information

DIVISION 15 - MECHANICAL. Section Variable Frequency Drives. Introduction

DIVISION 15 - MECHANICAL. Section Variable Frequency Drives. Introduction DIVISION 15 - MECHANICAL Section 15870 Variable Frequency Drives Introduction This specification is to cover a complete variable frequency motor drive (VFD) consisting of a pulse width modulated (PWM)

More information

7/16/2009. Cost-benefit, QALYs and the Value of a Statistical Life

7/16/2009. Cost-benefit, QALYs and the Value of a Statistical Life 7/16/29 Cost-benefit, QALYs and the Value of a Statistical Life Alan McKenzie, Bristol The benefits of many developments in radiotherapy may be obvious for instance, replacing hand planning by computer

More information

For Electric Personal Dosimete Dose-i

For Electric Personal Dosimete Dose-i Instruction Manual Dosime eter Setting Device For Electric Personal Dosimete er Dose-i (Unit:rem, Version:1.05 English) Issued on March 2015 WTA529893 1 / 38 Foreword Thank you for purchasing the Dosimeter

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

The Heidelberg Ion Therapy (HIT) Accelerator Coming Into Operation. Presented at EPAC 2008, Genova D. Ondreka, GSI

The Heidelberg Ion Therapy (HIT) Accelerator Coming Into Operation. Presented at EPAC 2008, Genova D. Ondreka, GSI The Heidelberg Ion Therapy (HIT) Accelerator Coming Into Operation Presented at EPAC 2008, Genova D. Ondreka, GSI Introduction Heidelberg Ion Therapy Centre: Europe's first dedicated particle therapy facility

More information

Major accidents in radiotherapy

Major accidents in radiotherapy Major accidents in radiotherapy related to treatment units (b) IAEA International Atomic Energy Agency Hospital Clínico Zaragoza - Spain IAEA Prevention of accidental exposure in radiotherapy 2 Events:

More information

3D-CRT Breast Cancer Planning

3D-CRT Breast Cancer Planning 3D-CRT Breast Cancer Planning Tips and Tricks Bednář, V. 3D-CRT Obsolete or not? There are more advanced techniques than 3D-CRT, but 3D-CRT has some advantages: Availability and price Forward planning

More information

IMRT QA: Point Dose Measurements or 2D Array?

IMRT QA: Point Dose Measurements or 2D Array? QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify

More information

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of: Canadian Partnership for Quality Radiotherapy Technical Quality Control Guidelines for Gamma Knife Radiosurgery A guidance document on behalf of: Canadian Association of Radiation Oncology Canadian Organization

More information

Trane TR150 Variable Frequency Drive Guide Specification Section GENERAL II. PRODUCTS

Trane TR150 Variable Frequency Drive Guide Specification Section GENERAL II. PRODUCTS Trane TR150 Variable Frequency Drive Guide Specification Section 15172 I. GENERAL A. This specification covers all variable frequency drives (VFDs) designated on the drawing schedules. All standard and

More information

3100 & 7200 Programming Sheet V2.6

3100 & 7200 Programming Sheet V2.6 3100 & 7200 Programming Sheet V2.6 Main Board Version: V1.88B for 3100 & V1.8A PC for 7200 (voice) GSM Board Version: GSM V1.5A Enter Installer Programming Mode: [ Ж ] + [ MODE ] + [ 5 digits Installer

More information

PROGRESS IN HADRONTHERAPY

PROGRESS IN HADRONTHERAPY PROGRESS IN HADRONTHERAPY Saverio Braccini TERA Foundation for Oncological Hadrontherapy IPRD06 - Siena - 01.10.06 - SB 1 Outline Introduction Radiation therapy with X rays and hadrontherapy Hadrontherapy

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms 1 st Workshop: Radiotherapy Modelling - Luz Saint Sauveur, Sep 2016 A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms A R Barbeiro 1, A Ureba 2, J A Baeza 3,

More information

PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302)

PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302) PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302) Date of Interview/Examination (MM/DD/YYYY): A downloadable PDF

More information

510Ic Owner s Manual CONSOLE OVERVIEW & GENERAL NOTES CONSOLE KEYPAD FUNTION

510Ic Owner s Manual CONSOLE OVERVIEW & GENERAL NOTES CONSOLE KEYPAD FUNTION CONSOLE OVERVIEW & GENERAL NOTES A primary feature of the 510Ic is the ability to control and monitor the user s workout program via the brilliant LCD console. At any time, the user may view their speed,

More information

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331

More information

QUICK SETUP GUIDE SECULIFE DFBASE

QUICK SETUP GUIDE SECULIFE DFBASE QUICK SETUP GUIDE SECULIFE DFBASE SECULIFE DFBASE is a microprocessor-based instrument that is used in the testing of defibrillators. It measures the energy output and provides information about the defibrillator

More information

Development of an Electronic Glove with Voice Output for Finger Posture Recognition

Development of an Electronic Glove with Voice Output for Finger Posture Recognition Development of an Electronic Glove with Voice Output for Finger Posture Recognition F. Wong*, E. H. Loh, P. Y. Lim, R. R. Porle, R. Chin, K. Teo and K. A. Mohamad Faculty of Engineering, Universiti Malaysia

More information

Beam Loss and Collimation in the ESS Linac

Beam Loss and Collimation in the ESS Linac Beam Loss and Collimation in the ESS Linac Ryoichi Miyamoto (ESS) B. Cheymol, H. Danared, M. Eshraqi, A. Ponton, J. Stovall, L. Tchelidze (ESS) I. Bustinduy (ESS-Bilbao) H. D. Thomsen, A. I. S. Holm, S.

More information

Applications of Particle Accelerators

Applications of Particle Accelerators Applications of Particle Accelerators Prof. Rob Edgecock STFC Rutherford Appleton Laboratory EMMA Accelerator at Daresbury Laboratory Applications >30000 accelerators in use world-wide: 44% for radiotherapy

More information

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY

More information

Portable Equine Nebuliser System. User Manual

Portable Equine Nebuliser System. User Manual Portable Equine Nebuliser System User Manual Table of Contents INTENDED USE... 3 SAFETY INFORMATION... 3 TECHNICAL SPECIFICATION... 4 INSTRUCTIONS FOR USE... 6 MAINTENANCE... 12 TROUBLESHOOTING... 13 WARRANTY...

More information

Debate Motion. Theoretical benefits of VMAT: Real Costs of VMAT. "Arc Based Techniques Will Make Conventional IMRT Obsolete

Debate Motion. Theoretical benefits of VMAT: Real Costs of VMAT. Arc Based Techniques Will Make Conventional IMRT Obsolete Austin, TX July 21, 2014 Debate Motion "Arc Based Techniques Will Make Conventional IMRT Obsolete AGAINST the Motion: Peter Balter, Ph.D. Potential Conflicts of Interest: Dr. Balter has research support/collaborations

More information

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using

More information

Proton Treatment. Keith Brown, Ph.D., CHP. Associate Director, Radiation Safety University of Pennsylvania

Proton Treatment. Keith Brown, Ph.D., CHP. Associate Director, Radiation Safety University of Pennsylvania Proton Treatment Keith Brown, Ph.D., CHP Associate Director, Radiation Safety University of Pennsylvania Proton Dose vs. Depth Wilson,. R.R. Radiological use of fast protons. Radiology 47:487-491, 1946.

More information

Maintenance and Logistics Experience at Loma Linda Proton Treatment Facility

Maintenance and Logistics Experience at Loma Linda Proton Treatment Facility Maintenance and Logistics Experience at Loma Linda Proton Treatment Facility Baldev Patyal, Ph.D., Chief Medical Physicist Department of Radiation Medicine Outline Loma Linda Proton Treatment Facility

More information

11/18/13 ECG SIGNAL ACQUISITION HARDWARE DESIGN. Origin of Bioelectric Signals

11/18/13 ECG SIGNAL ACQUISITION HARDWARE DESIGN. Origin of Bioelectric Signals ECG SIGNAL ACQUISITION HARDWARE DESIGN Origin of Bioelectric Signals 1 Cell membrane, channel proteins Electrical and chemical gradients at the semi-permeable cell membrane As a result, we get a membrane

More information

PGY-1. Resident Review Session Schedule

PGY-1. Resident Review Session Schedule 1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT

More information

MCP PMT with high time response and linear output current for neutron time-of-flight detectors

MCP PMT with high time response and linear output current for neutron time-of-flight detectors Journal of Physics: Conference Series PAPER OPEN ACCESS MCP PMT with high time response and linear output current for neutron time-of-flight detectors To cite this article: A S Dolotov et al 216 J. Phys.:

More information

General Treatment Equipment

General Treatment Equipment General Treatment Equipment We have quality products in General Treatment Equipment. All our models are efficient and display quality in Panchakarma (Ayurvedic) products. IMI - 3111: EXAMINATION COUCH

More information

Voluntary Product Accessibility Template (VPAT)

Voluntary Product Accessibility Template (VPAT) Voluntary Product Accessibility Template (VPAT) Date: January 25 th, 2016 Name of Product: Mitel 6730i, 6731i, 6735i, 6737i, 6739i, 6753i, 6755i, 6757i, 6863i, 6865i, 6867i, 6869i, 6873i Contact for more

More information

Completion of Treatment Planning. Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA

Completion of Treatment Planning. Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA Completion of Treatment Planning Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA Outline Entering Prescription Plan Printout Print and Transfer DRR Segment BEV Export to R&V Physician approval

More information

An Electrically Actuated Pin-Puller for Space Application using Nickel- Titanium Memory Alloy

An Electrically Actuated Pin-Puller for Space Application using Nickel- Titanium Memory Alloy An Electrically Actuated Pin-Puller for Space Application using Nickel- Titanium Memory Alloy Peter M. Cipollo, Brendan S. Surrusco Advisor: Dr. Sven G. Bilén The Pennsylvania State University Frank J.

More information

Splitting of the characteristic K-alpha doublet radiation of molybdenum (Item No.: P )

Splitting of the characteristic K-alpha doublet radiation of molybdenum (Item No.: P ) Splitting of the characteristic K-alpha doublet radiation of molybdenum (Item No.: P2540701) Curricular Relevance Area of Expertise: ILIAS Education Level: Physik Topic: Hochschule Subtopic: Moderne Physik

More information

DC-6. Diagnostic Ultrasound System

DC-6. Diagnostic Ultrasound System DC-6 Diagnostic Ultrasound System MINDRAY is proud to introduce DC-6, a color Doppler ultrasound system for general applications. DC-6 incorporates the latest digital ultrasound image processing technology

More information

RISK ASSESSMENT IN RADIATION THERAPY OF PATIENTS USING PROBABILISTIC SAFETY ANALYSIS OF RADIOTHERAPY DEVICES

RISK ASSESSMENT IN RADIATION THERAPY OF PATIENTS USING PROBABILISTIC SAFETY ANALYSIS OF RADIOTHERAPY DEVICES RISK ASSESSMENT IN RADIATION THERAPY OF PATIENTS USING PROBABILISTIC SAFETY ANALYSIS OF RADIOTHERAPY DEVICES Natalia Spizhenko *, Sergii Luchkovskyi *,**, Igor Kadenko ** * Cyber Clinic of Spizhenko **

More information

MLAA Trade Test Outline

MLAA Trade Test Outline MLAA Trade Test Outline The Practical Trade Test includes the following twelve exercises: 1. Automotive 2. Master Keying* 3. Safe Combination Locks 4. Lock Picking 5. Lockfitting Timber** 6. Lockfitting

More information

No. Error/Event Description Possible causes Possible Solution Classif. Defective RAM memory. Defective RAM memory.

No. Error/Event Description Possible causes Possible Solution Classif. Defective RAM memory. Defective RAM memory. 1 2 Defective RM memory Defective RM memory Defective RM memory Perform a software download Perform a software download Replace the igem battery Defective Power Supply Replace the Power Supply 4 Power

More information

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D. Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation

More information

Dental Extraoral X-ray Systems

Dental Extraoral X-ray Systems Dental Extraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL EXTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability. A registrant s x-ray system used for dental

More information

Level 3 Diploma in Dental Nursing

Level 3 Diploma in Dental Nursing LearnerName CompanyName AssessorName Date Submitted Level 3 Diploma in Dental Nursing - 5234 Unit 314 Dental Radiography Below contains all areas that make up the 314 dental exam, please be aware of all

More information

Dosimeter Setting Device System NRZ

Dosimeter Setting Device System NRZ User s Manual Dosimeter Setting Device System NRZ For Dosimeter NRF series (Units: msv Version:0.26 English) TA5C0903 b 1 / 50 Preface Thank you for purchasing the Dosimeter Setting Device; a product by

More information

A UNIQUELY SMART RADIATION THERAPY SYSTEM

A UNIQUELY SMART RADIATION THERAPY SYSTEM A UNIQUELY SMART RADIATION THERAPY SYSTEM FULLY INTEGRATED TREATMENT PLANNING AND DELIVERY PLAN Accuray Precision Treatment Planning System As the radiation oncology field advances, interoperability is

More information

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah DEFIBRILLATORS Prof. Yasser Mostafa Kadah Basics Defibrillation is definitive treatment for life-threatening cardiac arrhythmias such as ventricular fibrillation Defibrillation consists of delivering therapeutic

More information

Deciding to design better user interfaces

Deciding to design better user interfaces 13th International Conference on Naturalistic Decision Making 2017, Bath, UK Deciding to design better user interfaces Daniel P. JENKINS a, Andrew WOLFENDEN b, David J. GILMORE b and Malcolm BOYD a a DCA

More information

Foot Rehabilitation Monitoring Device

Foot Rehabilitation Monitoring Device Foot Rehabilitation Monitoring Device M. Shukri Ahmad, A. Alabqari Ma Radzi Faculty of Electrical and Electronic Engineering Universiti Tun Hussein Onn Malaysia Parit Raja, Johor mshukri@uthm.edu.my M.

More information

Dental Intraoral X-ray Systems

Dental Intraoral X-ray Systems Dental Intraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 2.0 4732.#### DENTAL INTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Commented [JC(1]: Based on part 4732.0880. Subpart 1. Applicability.

More information

A UNIQUELY SMART RADIATION THERAPY SYSTEM

A UNIQUELY SMART RADIATION THERAPY SYSTEM A UNIQUELY SMART RADIATION THERAPY SYSTEM FULLY INTEGRATED TREATMENT PLANNING AND DELIVERY PLAN Accuray Precision Treatment Planning System As the radiation oncology field advances, interoperability is

More information

H. Röcken, M. Abdel-Bary, E. Akcöltekin, P. Budz, T. Stephani, J. Wittschen

H. Röcken, M. Abdel-Bary, E. Akcöltekin, P. Budz, T. Stephani, J. Wittschen : Medical Operation of the 2 nd Machine, Production and Commissioning Status of Machines No. 3 to 7 H. Röcken, M. Abdel-Bary, E. Akcöltekin, P. Budz, T. Stephani, J. Wittschen VARIAN Medical Systems Particle

More information

ATyS D10 & D20. Notice d utilisation - Operating instructions AFFICHEURS DÉPORTÉS - REMOTE INTERFACES MAKE YOUR BUSINESS SAFE

ATyS D10 & D20. Notice d utilisation - Operating instructions AFFICHEURS DÉPORTÉS - REMOTE INTERFACES MAKE YOUR BUSINESS SAFE AFFICHEURS DÉPORTÉS - REMOTE INTERFACES ATyS D10 & D20 Notice d utilisation - Operating instructions F GB MAKE YOUR BUSINESS SAFE SOCOMEC GROUP SWITCHING PROTECTION & UPS F Sommaire PRÉSENTATION GÉNÉRALE

More information

Quality Assurance of Helical Tomotherapy Machines

Quality Assurance of Helical Tomotherapy Machines Quality Assurance of Helical Tomotherapy Machines Wolfgang Tomé 1,2, Hazim Jaradat 2, David Westerly 1, John Fenwick 3, Nigel Orton 2, Rock Mackie 1,2, and Bhudatt Paliwal 1,2 1 Human Oncology and 2 Medical

More information

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,

More information

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT April 25 nd, 2015 Lung Cancer Lung cancer is the most lethal cancer: Over 224,000 new diagnoses in the U.S. predicted

More information

Application(s) of Alanine

Application(s) of Alanine Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for

More information

Bruker D8 Discover User Manual (Version: July 2017)

Bruker D8 Discover User Manual (Version: July 2017) University of Minnesota College of Science and Engineering Characterization Facility Eulerian cradle Reference 2θ (two-theta): Detector angle (GADDS angle # 1) Bruker D8 Discover User Manual (Version:

More information

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,

More information

Work Domain Analysis for Designing a Radiotherapy System Control Interface

Work Domain Analysis for Designing a Radiotherapy System Control Interface 224 Work Domain Analysis for Designing a Radiotherapy System Control Interface Chun Wu Institute of Biomaterials and Biomedical Engineering, University of Toronto Toronto, Ontario, Canada Jennifer Jeon,

More information

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer Tony Kin Ming Lam Radiation Planner Dr Patricia Lindsay, Radiation Physicist Dr John Kim, Radiation Oncologist Dr Kim Ann Ung,

More information

Disclosures. Objectives 8/4/2016. Troubleshooting issues with linear accelerators Steven Sutlief, PhD DABR FAAPM UC San Diego

Disclosures. Objectives 8/4/2016. Troubleshooting issues with linear accelerators Steven Sutlief, PhD DABR FAAPM UC San Diego Troubleshooting issues with linear accelerators Steven Sutlief, PhD DABR FAAPM UC San Diego 1 Disclosures I have nothing to disclose. 8/4/2016 2 Objectives Understand why unscheduled linac downtime should

More information

Graphical User Interface for Windows for MCM Module

Graphical User Interface for Windows for MCM Module Page 1 of 8 Installation:- Please ensure that you have the latest updated GUI and download the GUIv.x.x.x.x.ZIP from our website. Extract the folder to a temporary location (These files can be deleted

More information

STANDARD OPERATING PROCEDURE: RIGAKU RU 300 XRD POWDER SAMPLES

STANDARD OPERATING PROCEDURE: RIGAKU RU 300 XRD POWDER SAMPLES STANDARD OPERATING PROCEDURE: RIGAKU RU 300 XRD POWDER SAMPLES Purpose of this instrument: The Rigaku RU 300 XRD is used for phase analysis of polycrystalline films, characterizing single crystal, and

More information

ECG MACHINE KEC- Series

ECG MACHINE KEC- Series KIZLON ECG MACHINE KEC- Series 404 Crescent Royale, Off New Link Road Andheri West, Mumbai 40053, Phones: 022-6708747 Email: info@kizlon.com Website: www.kizlon.com ECG Machine KEC-A100 This ECG machine

More information

Global Radiation Therapy Market Report

Global Radiation Therapy Market Report Global Radiation Therapy Market Report ----------------------------------------- 2013 Executive Summary Radiotherapy is used to kill cancerous cells and is commonly used either alone or in combination

More information

NC DOUBLE OPPOSING HEAD DRILL

NC DOUBLE OPPOSING HEAD DRILL NC DOUBLE OPPOSING HEAD DRILL 720ATD 720ATD 1800 x 1100 mm 2600 x 1600 mm 3000 x 2200 mm 3000 x 2600 mm 1. GENERAL PERFORMANCE 720ATD/20 Maximum glass dimension 3000 x 2600 Total weight Maximum Glass weight

More information

Eric E. Klein, Ph.D. Chair of TG-142

Eric E. Klein, Ph.D. Chair of TG-142 Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing

More information

Online Care, Online Control, Online Confidence

Online Care, Online Control, Online Confidence Online Care, Online Control, Online Confidence * Online Treatment Monitoring and much more! Designed for Efficiency provides you with online control and online confidence of treating your patients exactly

More information

>Talko. Have your modular speak out the CV, the LPC, 8 bits, vintage way. Page 1

>Talko. Have your modular speak out the CV, the LPC, 8 bits, vintage way. Page 1 >Talko Have your modular speak out the CV, the LPC, 8 bits, vintage way. Page 1 Table of Contents 1Short description...3 2Technical details... 3 3Banks... 4 4Panel... 4 4.1Mode Switch...4 4.2Sound jack

More information

SRS Uncertainty: Linac and CyberKnife Uncertainties

SRS Uncertainty: Linac and CyberKnife Uncertainties SRS Uncertainty: Linac and CyberKnife Uncertainties Sonja Dieterich, PhD Linac/CyberKnife Technological Uncertainties 1 Linac Mechanical/Radiation Isocenters Depuydt, Tom, et al. "Computer aided analysis

More information

Multifunction device for estimating cardiorespiratory fitness in the Step Tests

Multifunction device for estimating cardiorespiratory fitness in the Step Tests Journal of Advanced Sport Technology 1(2):2-6 Technical Research Multifunction device for estimating cardiorespiratory fitness in the Step Tests Marefat Siahkouhian 1 1. Professor in Sport physiology,

More information

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques TROG 08.03 RAVES Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques Introduction Commissioning and quality assurance of planning systems and treatment delivery

More information