How accurately can we measure dose clinically? Dosimetric accuracy requirement. Accurate Clinical dosimetry. Part I: Fundamentals and Issues

Size: px
Start display at page:

Download "How accurately can we measure dose clinically? Dosimetric accuracy requirement. Accurate Clinical dosimetry. Part I: Fundamentals and Issues"

Transcription

1 Accurate Clinical dosimetry Part I: Fundamentals and Issues Jan Seuntjens McGill University Montreal, Canada, H3G 1A4 Dosimetric accuracy requirement effect of dose uncertainty on complication-free tumour control ( utility function, Schultheiss & Boyer, 1988) ICRU Report 24 (1976) Mijnheer et al (1987) Brahme (1988) 1% higher accuracy 2% increase of cure 5% 3.5% 3% The structures these numbers are referring to are targets volumes 1 Dosimetric and positional accuracy of TPA: acceptability criteria Homogeneous water slab - simple fields Stack of tissue slabs simple fields antropomorphic phantom and/or complex beams central axis 2% 3% high dose region low dose gradient 3% 3% 4% large dose gradient 4 mm 4 mm 4 mm low dose region low dose gradient 3% (50% loc.dose) 3% (50% loc.dose) 3% (50% loc.dose) Van Dyk (1993) How accurately can we measure dose clinically? Which accuracy do we expect for a given clinical situation? How should we measure dose accurately? when is a situation well-behaved? when do we have non-equilibrium conditions? What is a suitable detector? How do we handle complication areas? build-up region narrow stereotactic beams modulated fields and dynamic time-dependent measurement 2 3 1

2 Outline Principles of measurement dosimetry Detectors and phantoms Reference dosimetry in TG51 compliant beams Issues in non-equilibrium dosimetry Relative dosimetry in non-equilibrium situations Reference dosimetry in TG-51 non-compliant beams Classification of dosimeters Dosimeter Mechanism Gas-filled ionization chamber Ionization in gasses Liquid ionization chamber Semiconductors (diodes, diamond detectors, MOSFET) TLD Scintillation counters Film, gel, Fricke Calorimetry Ionization in liquids Ionization in solids Luminescence Fluorescence Chemical reactions Heat 4 5 Measurement dosimetry in medium where: D med (r) M(r) c det f med (r) D med (r) = c det M(r)f med (r) dose to medium at the point raw signal, corrected for environmental conditions as given by detector detector cavity dose calibration coefficient (coupling constant) dose conversion coefficient converts average detector dose into dose to medium at point 6 Dosimeter dependent coefficients & coupling constants Factor or coefficient Dosimetry technique Calorimetry Fricke dosimetry M T OD ρl c det C 1 ( 3 ε Fe + ) G f med Unity ( D) med Fricke Ion chamber dosimetry Q 1 Wgas mgas e s med,gas p Q 7 2

3 Interpretation of a dosimeter measurement (measurement at point r) D med (r) = c det M(r)s med,det (r)p Q (r) D det f med c det is the detector dose calibration factor and ties the clinical dose measurement to the chamber/detector calibration; for ionization chamber: cavity gas calibration coefficient M is the result of a measurement corrected for technical influence quantities to ensure that what we actually measure a quantity proportional to the dose to the detector material s med,det is the stopping power ratio medium to cavity material p Q is an overall perturbation correction factor that accounts for everything a cavity theory does not account for. f med is the ratio of dose to medium to dose to cavity; factorization of f med is questionable in regions of strong disequilibrium 8 Reference dosimetry: TG-51 calibration protocol Cavity gas calibration coefficient derived from an absorbed dose to water calibration at 60 N a.k.a c det = D,w N D,air = N gas ) s w,air p Q ( ) ( = N D w = D,w Ms ( s w,air p Q ) w,air p Q s = N D,wMkQ k Q = w,air p Q with s w,air p Q ( ) = f w 9 Q ( ) Calibration chain for ionization chambers calibration beam air kerma calibration coefficient absorbed dose calibration coefficient TG-51 or TRS 398 N Dw, TG-21 or TRS 277 NK or NX cavity gas calibration coefficient ' R * 50 ecal * gr or k k P ND, air or Ngas clinical beam absorbed dose calibration coefficient 10 k Q N D, w water L P air wallprepl ( / ρ) 11 3

4 Reference dosimetry: TG-51 calibration protocol (photons) Reference dosimetry: TG-51 calibration protocol (electrons) k Dw = MND,w ecal k'r50 Pgr k Dw = MND,w Q - Measurement at dref in water - 10 x 10 cm2 or 20 x 20 cm2 field - beam quality specifier: R50 - Measurement at 10 cm depth in water - 10 x 10 cm2 field - beam quality specifier: %ddx(10 cm) Almond et al, Reference Dosimetry New developments 13 Accuracy of clinical dosimetry A number of more recent chamber types are not listed in TG-51 Extensive experimental data on kq factors is available that allows to estimate uncertainties on kq as well as the overall reference dosimetry calibration process. Stability, linearity, reproducibility of the detector and the measurement setup Energy dependence - what is the accuracy with which one can convert detector dose to medium dose? CPE or TCPE and detector is small compared to the range of secondary electrons: fmed can be factorized as a stopping power ratio + correction factors (and the corrections are small or constant) If these conditions are not fulfilled: fmed cannot be factorized or very large correction and situation dependent correction factors are needed (unpractical) Monte Carlo calculations (assumed accurate within the constraints of the algorithm and basic cross-section datasets) Choose a more suitable detector (perturbation free, medium equivalent, etc) New AAPM Work Group (WGTG51, chaired by M. McEwen) to supply this information (addendum to TG-51)

5 Why do we worry about CPE or TCPE? For the measurement of a dose distribution or for measurement of output: In regions of CPE and TCPE: SPR corrections accurately represent detector response (and are small for air-filled chambers in photon beams) In regions of non-cpe: SPR corrections DO NOT accurately reflect changes in detector response and additional, sometimes large, corrections are needed build-up regions in any field, interface-proximal points in heterogeneous phantoms (build-up and build-down) narrow stereotactic fields intensity modulated fields 16 Different measurement issues in open photon fields in water in-field, CPE or TCPE standard measurement, SPR is nearly constant and represents detector response well; detector perturbations are small (point-ofmeasurement shift is 0.6*r inner upstream) in-field, non-cpe (build-up) non standard measurement, SPR varies modestly but does not represent variation in detector response well; detector perturbations are large, detector dependent. penumbra (non-cpe) detector size must be small relative to field-size and penumbra width, SPR does not represent detector response well, lateral fluence perturbations, detector dependent. out-of-field (CPE or TCPE) detector size must be sufficiently large to collect sufficiently large signal. SPR represents detector response well. 17 Stopping power ratios (SPR s) SPR s are depth, field size, and radiation quality dependent for reference dosimetry: values have been calculated with Monte Carlo techniques and are well documented in standard dosimetry protocols for relative dosimetry in TCPE: their variation relative to the reference point can be calculated using Monte Carlo techniques for relative dosimetry in non-equilibrium situations: their variation relative to the reference point can still be calculated but no longer represents an accurate dose conversion. s w,air (z) for plane parallel bremsstrahlung beams - no e - contamination, central axis MV 20 MV 30 MV 18 Andreo and Brahme, Phys. Med. Biol., 31, 839 (1986) 5

6 Narrow 1.5 mm field Ratio of dose to water to dose to air Small fields 1.70 averaged over cavity volume Dw/Dair llecting electrode diameter: 1.5 mm Separation: 1 mm 1.30 stopping power ratio w/air Paskalev, Seuntjens, Podgorsak (2002) AAPM Proc. Series 13, Med. Phys. Publishing, Madison, Wi, SanchezDoblado et al Phys. Med. Biol (2003) 20 Perturbation correction factors: traditional factorization for ionization chambers: PQ = Pwall Pgr Pfl Pcel Off-axis distance (mm) 8 21 Perturbation correction factors depth, field size, and radiation quality dependent for reference dosimetry using ionization chambers: based on Monte Carlo calculations, measurements and relatively well documented for relative dosimetry: their variation relative to the reference point is usually ignored but can be very significant in non-cpe conditions. Pwall: wall perturbation correction factor (pwall) Pgr: correction for gradient effect (effective p. of meas.) Pfl: fluence perturbation correction factor (pcav) Pcel: central electrode perturbation correction factor (pcel)

7 NACP chamber in 6 MeV and 20 MeV electrons NACP-02 Partial compensatio n of wall and fluence perturbation effects Buckley and Rogers, Med. Phys (2006) Verhaegen et al, Phys. Med. Biol (2006) Build-up dose measurements Discrepancies have been reported between measured and calculated build-up doses at high photon energies. Benchmark measurements are needed to ensure a treatment planning system accurately calculates dose in the build-up region. Benchmark measurements are prone to uncertainties ranging up to 50% of the local dose

8 NACP chamber response in buildup region Q W Dcav = mair e air Dcav ( d) = Q( d) Dcav ( dmax ) Q( dmax ) accurate DOSRZnrc cavity simulations NACP measurement Field size (cm 2 ) 10x10 40x40 Calculated relative surface cavity dose 45 ± 1 % 67 ± 2 % Measured relative surface ionization 47 ± 2 % 69.0 ± 0.3 % Perturbation correction factors 10x10 cm 2 water ( ) ( ) L( d) D water water d = Dcav d ρ Φ ( d) air air Depth (cm) Perturbation correction 0.78 ± ±

9 Effective point of measurement in build-up region is not the same as in CPE conditions Reconciling measurements and calculations in the build-up region by adjusting EPOM ->simple if it works; but is chamber dependent! 18 MV Kawrakow (2006) Med. Phys. 33, Reference dosimetry in TG- 51/TRS-398 non-compliant beams Many radiation treatment machines cannot produce reference field sizes or SSDs according to the recommendations of TG-51 or TRS-398 protocols (e.g., GammaKnife, CyberKnife, and TomoTherapy). IMRT/IMRS and SRS/SRT/SBRT treatment plans are designed to treat from multiple beam directions. Delivery of these plans use small fields that are not very similar to the reference conditions used for beam calibration. Reference conditions for beam calibration should be as close as possible to the patient irradiation conditions. 33 Mackie and Huq Possible Irradiation Protocol Reference Geometry Issues with non-compliant beams Calibration Lab Irradiation With a -60 Unit User Delivering Vendor s Protocol (Protocols will Vary from Vendor to Vendor) 34 Mackie and Huq There are no primary standards that measure dose directly in non-compliant beams Delivery protocols in non-compliant beams are not standardized (and not standardizable ) Fundamental studies are required to understand detector response in noncompliant beams - goal is to make an accurate connection between non-compliant beams and compliant beams 35 9

10 Proposed correction factor, C Q non-e How can this factor be determined? The calibration coefficient for any composed (possibly, non-equilibrium) field can be assessed as follows: non-e non-e Q N D,w = C Q ND,w Calibration coefficient applicable to composed field rrection factor Calibration coefficient in TG-51 reference conditions MLC photon transmission map Primary photon dose kernels Photon radial fluence Bouchard, H. and J. Seuntjens, Med. Phys. 31(9), 2004 Kernel acquisition 6 MV dose kernel to air cavity Chamber longitudinal kernel Longitudinal air-cavity kernels for the Exradin A12 Farmer chamber, z = 3 cm, 1e6 histories SSD = 100 cm 1.60E E E E cm 8.00E-26 on-axis (y = 0) off-axis (y = 0,14 cm) 6.00E-26 H 2 O Exradin A12 Farmer model (0.653 cm 3 ) 4.00E E Tantot et al E x/cm 10

11 mparison with measurements mparison with measurements Field name Static #1 Static #2 Static #3 Static #4 Static #5 Dynamic #1 Dynamic #2 Dynamic #3 Dynamic #4 Dynamic #5 Dynamic #6 Dynamic #7 Dynamic #8 C IMRT,6 MV measured ± 4.0% ± 6.1% ± 3.2% ± 6.0% ± 2.5% ± 3.0% ± 2.5% ± 3.9% ± 3.2% ± 5.4% ± 5.9% ± 5.7% ± 5.9% C IMRT,6 MV calculated Difference 6.8% 1.9% 2.7% 3.2% 2.7% -0.3% 0.7% 7.8% -2.4% 3.8% -1.4% 5.9% 6.9% Field name Static #1 Static #2 Static #3 Static #4 Static #5 Dynamic #1 Dynamic #2 Dynamic #3 Dynamic #4 Dynamic #5 Dynamic #6 Dynamic #7 Dynamic #8 C IMRT,6 MV measured ± 4.0% ± 6.1% ± 3.2% ± 6.0% ± 2.5% ± 3.0% ± 2.5% ± 3.9% ± 3.2% ± 5.4% ± 5.9% ± 5.7% ± 5.9% C IMRT,6 MV calculated Difference 6.8% 1.9% 2.7% 3.2% 2.7% -0.3% 0.7% 7.8% -2.4% 3.8% -1.4% 5.9% 6.9% Bouchard and Seuntjens, Med Phys 31, (2004) 40 Bouchard and Seuntjens, Med Phys 31, (2004) Summary Reviewed basics of dosimetry TG-51 and current extensions applicability of cavity theory to non-reference conditions Accurate measurements require accurate understanding of the detector response nventional cavity theory for detector response is not accurate in non-equilibrium conditions; this has, in some cases, significant clinical impact on a dose measurement suitable detectors or corrections are needed Calibration of TG51/TRS-398 non-compliant beams require more investigations 42 Acknowledgments Students: Emily Heath, Kristin Stewart, Khalid Al-Yahya, Keyvan Jabbari, Andrew Alexander, Adam Elliott, Arman Sarfehnia, Monica Margeanu, Edwin Sham Post-doc: Gabriela Stroian Staff: Frank Verhaegen, Wamied Abdel-Rahman, Slobodan Devic, Ervin Podgorsak, Michael Evans, William Parker Funding (operating): Canadian Inst. Health Research (CIHR), National Cancer Inst. Canada (NCIC), Natural Sciences and Engineering Research uncil (NSERC) 43 11

How accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement

How accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement Accurate clinical dosimetry Part I: Fundamentals and New developments in reference dosimetry Jan Seuntjens McGill University Montreal, Canada, H3G 1A4 jseuntjens@medphys.mcgill.ca Dosimetric accuracy requirement

More information

Much of what we will say

Much of what we will say New Dosimetry Measurements - New Developments J. Seuntjens J. Gibbons I. Das WA-A-211A Continuing Education Session Much o what we will say Clinical Dosimetry Measurements in Radiotherapy (AAPM 2009 Summer

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS} High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics

More information

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4 Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability

More information

Acknowledgments & Credits 8/2/2012

Acknowledgments & Credits 8/2/2012 Dosimetryfor Ir- HDR Brachytherapy Present Status and Future Direction TU-A-213AB L DeWerd 1 *, J Seuntjens 2 *, M Rivard 3 *, M McEwen 4 Approaches for the determination and dissemination of absorbed

More information

Applications of Monte Carlo simulations to radiation dosimetry

Applications of Monte Carlo simulations to radiation dosimetry Applications of Monte Carlo simulations to radiation dosimetry D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers

More information

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall

More information

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

Recent advances in dosimetry in reference conditions for proton and light-ion beams

Recent advances in dosimetry in reference conditions for proton and light-ion beams Recent advances in dosimetry in reference conditions for proton and light-ion beams S. Vatnitskiy a), P. Andreo b) and D.T.L. Jones c) a) MedAustron, Wiener Neustadt, Austria b) Medical Radiation Physics,

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams

AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams Peter R. Almond, Brown Cancer Center,Louisville, KY 40202 Peter J. Biggs, Department of Radiation Oncology,

More information

IAEA/AAPM Code of Practice for the Dosimetry of Static Small Photon Fields. Jan Seuntjens McGill University Montréal, Canada

IAEA/AAPM Code of Practice for the Dosimetry of Static Small Photon Fields. Jan Seuntjens McGill University Montréal, Canada IAEA/AAPM Code o Practice or the Dosimetry o Static Small Photon Fields Jan Seuntjens McGill University Montréal, Canada Acknowledgements IAEA/AAPM small and composite ield working group: Hugo Palmans

More information

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams Ernesto Mainegra-Hing a) and Bryan R. Muir Measurement Science and Standards, National Research Council of Canada,

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

Dosimetric Uncertainties in Reference and Relative Dosimetry of Small Fields

Dosimetric Uncertainties in Reference and Relative Dosimetry of Small Fields Dosimetric Uncertainties in Reerence and Relative Dosimetry o Small Fields Jan Seuntjens, Ph.D., FAAPM, FCCPM McGill University Health Centre Canada Outline o Presentation Uncertainty concepts and requirements

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA AAPM international Code of Practice for reference and relative dose determination Hugo Palmans a)

More information

Small Field Dosimetry

Small Field Dosimetry Small Field Dosimetry Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Emeritus Professor, Department of Medical Physics University of Wisconsin Madison WI Conflict

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the

More information

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons. Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom Iran. J. Radiat. Res., 2003; 1(3): 143-149 Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom A. Mostaar 1, M. Allahverdi 1,2, M. Shahriari 3 1 Medical Physics Department,

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning

Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Professor, Department of Medical Physics

More information

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article. Original Article PMP Progress in Medical Physics 29(1), March 218 https://doi.org/1.14316/pmp.218.29.1.16 pissn 258-4445, eissn 258-4453 Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers

More information

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013 Level of Accuracy Practically Achievable in Radiation Therapy David Followill and RPC staff August 6, 2013 Sources of Treatment Uncertainty Machine functioning Radiation dose determination Patient specific

More information

Dosimetry of Small Static Fields Used in External Beam Radiotherapy

Dosimetry of Small Static Fields Used in External Beam Radiotherapy Technical Reports SeriEs No.483 Dosimetry of Small Static Fields Used in External Beam Radiotherapy An International Code of Practice for Reference and Relative Dose Determination Sponsored by the IAEA

More information

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

PGY-1. Resident Review Session Schedule

PGY-1. Resident Review Session Schedule 1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,

More information

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of

More information

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe 3D printing of air slab caps for correction free small field dosimetry Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe Small fields? Defined as fields sizes too small to achieve

More information

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam Megan A. Hyun, M.S. Under the supervision of Larry A. DeWerd, Ph.D. NCCAAPM Fall Meeting October 30, 2015 Medical Radiation Research Center

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

Dosimetry and QA of proton and heavier ion beams

Dosimetry and QA of proton and heavier ion beams Dosimetry and QA of proton and heavier ion beams Stanislav Vatnitskiy EBG MedAustron GmbH Wiener Neustadt, Austria Content Introduction Reference dosimetry Methods, detectors, protocols Dosimetry in non-reference

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry C. Candela-Juan, J. Vijande, T. García-Martínez, Y. Niatsetski,

More information

Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition

Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of

More information

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions

More information

tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM

tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM A Thesis Presented to The Academic Faculty by Ryan J. Underwood In Partial Fulfillment of the Requirements for the Degree Master of Science in the School

More information

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation. Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation Samuel Chiquita 1 1 University of Porto, Porto, Portugal Abstract In

More information

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,

More information

IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy

IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK

More information

Transition to Heterogeneity Corrections. Why have accurate dose algorithms?

Transition to Heterogeneity Corrections. Why have accurate dose algorithms? Transition to Heterogeneity Corrections Eric E. Klein, M.S., Washington University, St. Louis, MO Craig Stevens, M.D., Ph.D., MD Anderson Cancer Center, Houston, TX Nikos Papinikolou, Ph.D., University

More information

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing

More information

Basic dosimetric data for treatment planning system

Basic dosimetric data for treatment planning system Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny

More information

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) IJDas (1) Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of

More information

Neutron dose evaluation in radiotherapy

Neutron dose evaluation in radiotherapy Neutron dose evaluation in radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Radiation therapy with a linear accelerator (LINAC) Photoneutron production in accelerator head

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Activity report from JCOG physics group

Activity report from JCOG physics group 2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer

More information

Application(s) of Alanine

Application(s) of Alanine Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

An Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements

An Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements An Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements Jean M. Moran, a Donald A. Roberts, Teamour S. Nurushev, Larry E. Antonuk, Youcef El-Mohri, and Benedick A. Fraass Department

More information

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline

More information

IMRT point dose measurements with a diamond detector

IMRT point dose measurements with a diamond detector IMRT point dose measurements with a diamond detector Erin Barnett, Marc MacKenzie, B. Gino Fallone Department of Physics, University of Alberta, and Department of Medical Physics, Cross Cancer Institute,

More information

IMRT QA: Point Dose Measurements or 2D Array?

IMRT QA: Point Dose Measurements or 2D Array? QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify

More information

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS Romanian Reports in Physics, Vol. 63, No. 3, P. 700 706, 2011 DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS ZOE GHITULESCU 1, ANA STOCHIOIU 2, MIHAI DUMITRACHE 3, 1 CNCAN- National

More information

Measuring radiation dose through the detection of radiation-induced acoustic waves

Measuring radiation dose through the detection of radiation-induced acoustic waves Measuring radiation dose through the detection of radiation-induced acoustic waves Susannah Hickling a, Hao Lei b, Maritza Hobson a, Issam El Naqa a,b a McGill University, Montreal, Canada b University

More information

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Dosimetric characterization of surface applicators for use with the Xoft ebx system Dosimetric characterization of surface applicators for use with the Xoft ebx system R.M. Kennedy University of Wisconsin Medical Research Center Madison, WI April 23, 2010 Introduction Squamous and basal

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti a*, M. Bucciolini a, L. Marrazzo a, D. Menichelli a. a) Department

More information

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA M. Pimpinella, A. S. Guerra, S. La Civita and R. F. Laitano Istituto

More information

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system Journal of Physics: Conference Series PAPER OPEN ACCESS Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system To cite this article: S Yani et al 2016 J.

More information

Dosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy

Dosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Dosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy Francis Kar-ho Lee, a Simon Kar-yiu

More information

SRS/SBRT Errors and Causes

SRS/SBRT Errors and Causes SRS/SBRT Errors and Causes Ryan Foster, Ph.D. Assistant Professor Director of Clinical Medical Physics Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX Disclosures I receive research

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Ion recombination corrections of ionization chambers in flattening filter-free photon radiation

Ion recombination corrections of ionization chambers in flattening filter-free photon radiation JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Ion recombination corrections of ionization chambers in flattening filter-free photon radiation Yuenan Wang, a Stephen B. Easterling,

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA Independent Dose Verification for IMRT Using Monte Carlo C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA Outline Why Monte Carlo for IMRT QA? Experimental verification

More information

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of

More information

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors G.Gambarini 1, C.Tenconi 1, N.Mantaut 1 M.Carrara 2, M.Borroni 2, E.Pignoli 2 D.Cutajar 3, M.Petasecca 3, I.Fuduli 3, M.Lerch

More information

This article was published in an Elsevier journal. The attached copy is furnished to the author for non-commercial research and education use, including for instruction at the author s institution, sharing

More information

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

EDITORIAL NOTE. factor, which has been used instead of the chamber factor N D

EDITORIAL NOTE. factor, which has been used instead of the chamber factor N D IAEA/WHO NETWORK OF SECONDARY STANDARD DOSIMETRY LABORATORIES Prepared by the Joint IAEA/WHO Secretariat of the SSDL Network Published semi-annually by the International Atomic Energy Agency Vienna No.

More information

Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams.

Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams. IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams. To cite this

More information

THE USE OF MONTE-CARLO CODES FOR TREATMENT PLANNING IN EXTERNAL- BEAM RADIOTHERAPY

THE USE OF MONTE-CARLO CODES FOR TREATMENT PLANNING IN EXTERNAL- BEAM RADIOTHERAPY THE USE OF MONTE-CARLO CODES FOR TREATMENT PLANNING IN EXTERNAL- BEAM RADIOTHERAPY Alan E. Nahum PhD Radiation Physics Dept. 3994, Copenhagen University Hospital, DK-2100 Copenhagen Ø, Denmark [alan.nahum@rh.dk]

More information

Implantable MOSFET dosimeter response to 192 Ir HDR radiation

Implantable MOSFET dosimeter response to 192 Ir HDR radiation Implantable MOSFET dosimeter response to 192 Ir HDR radiation Jessica Fagerstrom University of Wisconsin Madison, Department of Medical Physics Medical Radiation Research Center North Central Chapter AAPM

More information

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification Sathiyan Saminathan, a Ravikumar

More information

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom Gary Fisher, B.S. David Followill, Ph.D. Geoffrey Ibbott, Ph.D. This investigation

More information

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method Iran. J. Radiat. Res., 2007; 5 (3): 147-152 Dosimetric characteristics of Cs sources used in after loading Selectron system by Monte Carlo method M.B.Tavakoli, D. Shahbazi-Gahrouei *, M. Hosseinpour Department

More information

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers Rhys Carrington, Ysbyty Glan Clwyd, Bodelwyddan. MSc Medical Radiation Physics. Small field dosimetry

More information

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Clarita Saldarriaga Vargas, Corinne Amalberto, Lara Struelens,

More information

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information

Leila E. A. Nichol Royal Surrey County Hospital

Leila E. A. Nichol Royal Surrey County Hospital 2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry

More information

PRECISE DOSIMETRY OF SMALL PHOTON BEAMS COLLIMATED BY AN ADD-ON DYNAMIC MICRO-MULTILEAVES COLLIMATOR.

PRECISE DOSIMETRY OF SMALL PHOTON BEAMS COLLIMATED BY AN ADD-ON DYNAMIC MICRO-MULTILEAVES COLLIMATOR. IX Latin American IRPA Regional Congress on Radiation Protection and Safety - IRPA 2013 Rio de Janeiro, RJ, Brazil, April 15-19, 2013 SOCIEDADE BRASILEIRA DE PROTEÇÃO RADIOLÓGICA - SBPR PRECISE DOSIMETRY

More information

IORT with mobile linacs: the Italian experience

IORT with mobile linacs: the Italian experience IORT with mobile linacs: the Italian experience G. Tosi, M. Ciocca lntroduction At the beginning of 1999 a mobile Linac (Novac 7, manufactured by Hitesys, Aprilia, Italy) able to produce electron beams

More information

Radiochromic film dosimetry in water phantoms

Radiochromic film dosimetry in water phantoms INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) N27 N31 www.iop.org/journals/pb PII: S0031-9155(01)16858-2 NOTE Radiochromic film dosimetry in water phantoms

More information

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,

More information