Advantages of Ion Mobility QTOF for Characterization of Large Molecules
|
|
- Anne Cummings
- 5 years ago
- Views:
Transcription
1 Advantages of Ion Mobility QTOF for Characterization of Large Molecules Add a New Dimension to your Research Capability with Agilent s New Drift Ion Mobility QTOF System Ning Tang, Ph.D. Application Scientist Anne Blackwell, Ph.D. Application Scientist 1
2 IM-QTOF Instrument Overview System sensitivity optimized using electrodynamic ion funnels to focus and transmit ions Ion Mobility resolution optimized while maintaining QTOF performance (mass resolution and accuracy) Ion Fragmentation can be selected using standard QTOF collision cell (CID) Bandwidth of QTOF data acquisition and processing channel was increased by 10 fold to match the ion mobility data rates 2
3 Ion Mobility System Design Ionization source: Ion generation (ESI, AJS, Nano ESI, ChipCube, APCI etc.) Front ion funnel: Efficient ion collection, desolvation and excess gas removal Trap funnel: Ion accumulation and introducing ion packets into drift cell Drift cell: Uniform low field ion mobility allows direct determination of accurate CCS (Ω) Rear funnel: Efficient ion refocusing and introduction into mass analyzer 3
4 Basic Operational Principle of Ion Mobility For Conventional DC Uniform Field IMS V H Ion Mobility Cell V L Analyte Ions Detector Gating Optics Electric Field Stacked ring ion guide gives linear field v = K E e E P T Ω 4
5 Resolution Is Important! Chromatographic Ion Mobility Mass ~seconds ~60 milli-seconds ~ 100 m seconds 5
6 Ion Mobility Resolution IMS Resolution R = t d t d = LEQ 16kTln
7 Ion Mobility Resolution - Continued Resolution = 84!
8 It s All About Separation Chromatography Ion Mobility Mass ~seconds ~60 milli-seconds ~ 100 m seconds 8
9 It s All About Separation Chromatography Ion Mobility Mass Peak Capacity = IM Resolution x Mass Resolution x Orthogonality Peak Capacity = 60 x 40,000 x 14% = 336,600 ~ 8-fold increase Dwivedi P, Schultz AJ, Hill HH, Metabolic profiling of human blood by high-resolution ion mobility mass spectrometry (IM-MS). Int J Mass Spectrom 298:
10 Separation of Isobaric Pesticides Theoretical Plot Aldicarb-sulfone (C 7 H 14 N 2 O 4 S) [M+Na] + = Acetamiprid (C 10 H 11 ClN 4 ) [M+Na] + = D mass is 0.2 mda requires ~ 2,000,000 resolution! IMS Drift Separation 4 x x Acetamiprid Aldicarb-sulfone Drift Time (ms) x10 4 +IMS DriftSpec (m/z: ) (rt: min) Aldicarbsulfone_A 5.5 * * Drift Time (ms) 10
11 Resolving Structural Sugar Isomers C 18 H 32 O 16 Raffinose Melezitose Resolving two isobaric tri-saccharides 11
12 Mobility Drift Time (ms) Crude Bacterial Extract Mass (Da) Mass (Da) siamycin II Ion Mobility Drift Time (ms) 12
13 IMS-MS for Proteomics: Transmembrane Spanning Peptides Drift Time (ms) vs. m/z Drift Time (ms) vs. m/z Characteristics: Large hydrophobic peptides X-ray: 7-helical domains Confirmation impacts function of membrane Membrane proteins are drug targets Working Hypothesis: α-helical membrane spanning domains stay intact, and differ in drift-time from denatured peptides Drift Time (ms) vs. m/z 13
14 IMS-MS for Proteomics: Transmembrane Spanning Peptides A B Drift Time (ms) vs. m/z Drift Time (ms) vs. m/z C D Drift Time (ms) vs. m/z Drift Time (ms) vs. m/z Figure Seq. Loc. Chg. State Helix A 1-40 [M+3H] 3+ A B 1-82 [M+6H] 6+ A-B C [M+5H] 5+ C-D-E D [M+4H] 4+ F-G 14
15 LC Ion Mobility QTOF Protein Isoform Analysis Cry34AB Drift Time (ms) vs. m/z
16 Resolving Isoforms of IgG2 16
17 Resolving Isoforms of IgG Drift Time (ms) vs. m/z x IMS Drift Spectrum ( ms) Counts vs. Drift Time (ms) Publication Paul Schnier, Synapt: 3uM direct infusion in 160 mm NH 4 Ac Drift Time (ms) vs. m/z 17
18 Why Ion Mobility? Separation! Protein Digest, Erin Baker, PNNL 18
19 Lipid Analysis: Mixture of L-α-phosphotidylethanolamine (PE) Lipids Ion Mobility Drift Time (ms) PE 92:N PE 90:N +1 lipids PE 19:N PE 48:N PE 46:N PE 44:N PE 40:N PE 38:N PE 37:N PE 36:N PE 34:N PE 33:N PE 29:N PE 63:N PE 61:N PE 59:N +2 lipids +3 lipids +4 lipids PE 38:8 PE 38:6 PE 38:4 PE 38:2 PE 38:0 PE 38:7 PE 38:5 PE 38:3 PE 38:1 PE 37:1 19
20 Mobility Drift Time (ms) Why Ion Mobility? Specificity! Integrated Mass Spectrum: Mobility-Filtered Mass Spectrum: Mass (Da) Mass (Da) Crude bacterial extract, John McLean, Vanderbilt Mass (Da) 20
21 single pulse 1 ms trapping single pulse 4 ms trapping multi-pulse 4x1 ms trapping single pulse 8 ms trap time multi-pulse 8x1 ms trapping Area Response Signal intensity (A.U.) Why Ion Mobility? Sensitivity! Linear Dynamic Range y = 59060x R² = E+06 5E+06 4E+06 3E+06 2E+06 1E Sample Amount (pg) 0E+00 Sensitivity: ~ 50 fg of Reserpine Dynamic range: ~ 3-4 orders Integrated signal intensity for tetrakis decyl ammonium bromide ion versus ion trapping time for single pulse and multi-pulse experiments. These data indicate that for the same amount of trapping time, multiplexing experiments result in at least 10X higher signal intensity possibly due to less space charge effects and detector saturation issues. 21
22 Why Ion Mobility? Selectivity! Data browser can be used to isolate a group of glycans from matrix Isolation of RNAseB native Glycans, Cathy Costello, Boston University 22
23 Why Ion Mobility? Selectivity! RNaseB native glycans Simplified view of RNaseB native glycan Glycan obscured by matrix is identified after cleaning up background chemical noise using ion mobility 23
24 Ion Mobility of Polymeric Ink Dispersants 24
25 Determining Cross Sectional Areas Charge state of the analyte ion Charge on an electron Electric field Ω = (18π)1/2 16 ze [ ] 1 (kbt) 1/2 2 t E d m i m B L 760 P T N Boltzmann constant Reduced mass of the ion and neutral Number density of the drift gas K 0 = 273 K T P 760 Torr L 2 Vt d Molecular size K 0 = Reduced Ion Mobility T = Temperature P = Pressure L = Drift length V = Voltage Drop across drift region t d = Drift time X-ray crystallography Ion mobility (using Helium) 25
26 Standard Procedure for Calculating CCS (Auto Charting, Curve Fitting and Calculating the CCS) Time segments with different drift fields 26
27 Collision Cross Section Benchmark --- Vanderbilt University Mobility Drift Time (ms) TAA-5 N-(CH 2 CH 2 CH 2 CH 2 CH 3 ) 4 Tetraalkylammonium salts (TAA) Proposed as an ideal ion mobility standard Wide CCS range (TAA-4 to TAA-18; 100 to 400 Å 2 ) TAA salts do not form clusters Literature CCS values exist N 2 drift gas TAA-8 TAA-7 TAA-6 TAA-5 TAA-4 TAA-10 TAA-12 TAA-16 TAA ions +2 ions +3 ions Mass-to-Charge (m/z)
28 Tetraalkylammonium Salts --- CCS Values Compared to Literature Analyte Measured Cross-Section [Å 2 ] TAA ± 0.5% TAA ± 0.6% TAA ± 0.3% TAA ± 0.2% TAA ± 0.1% TAA ± 0.1% TAA ± 0.2% TAA ± 0.2% TAA ± 0.3% Literature Cross-Section [Å 2 ] ± 0.3% ± 0.1% ± 0.3% ± 0.2% ± 0.4% Relative Standard Deviation [%] High experimental precision (< 0.5% relative deviation) Agreement with literature (most < 0.5% deviation) 28
29 Cross Section Calculation of Ubiquitin Charge States m/z 612.8, [M+14H] 14+ m/z 659.8, [M+13H] 13+ Automated collision cross section calculation without the use of calibration curves approximate region for compact structures approximate region for elongated structures m/z 714.7, [M+12H] 12+ m/z 779.6, [M+11H] 11+ m/z 857.5, [M+10H] 10+ m/z 952.6, [M+9H] 9+ m/z , [M+8H] 8+ m/z , [M+7H] 7+ m/z , [M+6H] 6+ m/z , [M+5H] (Å) Reference: Koeniger and Clemmer J Am Soc Mass Spectrom 2007, 18,
30 Cross Section Calculation of Ubiquitin Charge States Ion Charge State CCS experimental (Å 2 ) [M+5H] [M+6H] , 1658 CCS literature (Å 2 ) [M+7H] , [M+8H] [M+9H] [M+10H] , [M+11H] , [M+12H] , [M+13H] , [M+14H] , 2726 Automated collision cross section calculation without the use of calibration curves Reference: Bush et al., Anal Chem. 2010, 82,
31 Carbohydrates -- Great complexity by linkage Source: Blixt et al., PNAS, 2004 Current dominant strategies: MS n or Library searches 31
32 Carbohydrates IM-MS Ion Mobility Drift Time (ms) Lacto-N-difucohexaose I 60 Mixture of Lacto-N-difucohexaose I & II Fuc Fuc Gal 50 GlcNAc Gal Glc Lacto-N-difucohexaose 40 II Mass (Da) 30 Fuc Fuc Gal GlcNAc Gal Glc Lacto-N difucohexaose I Drift Time (ms) 10 Lacto-Ndifucohexaose II Mass (Da) Drift Time (ms) 32
33 IM Drift Spectrum on Cytochrome C at various RFs: S1: Native S2-S5: Denatured S3 Waters Synapt G2 IM data RF 90V RF 110V RF 130V RF 150V RF 180V S1? S2 S4 S5 Preserve protein native structures better! Due to the much lower ions heating. 33
34 IMS-MS for Proteomics: Transmembrane Spanning Peptides of HeLa digest All Ions Fragmentation Low dt High dt Low dt Drift Time (ms) vs. m/z Drift Time (ms) vs. m/z High dt Low dt 34
35 A Word About Instrument Design LC Drift IMS MS and MS/MS High Resolution Accurate Mass Feature Agilent Waters Drift Mobility Advantage Mobility Resolution Highest (can be > 80) 80cm drift tube (L) higher voltage (E) No RF fields, Uniform low DC field Generally around 30 25cm drift TriWave, Multi-section device RF fields Over 2X the IM resolution of T-wave Sensitivity High efficiency ion funnels - trapping and rear Step wave lens Pressure barrier between Q and TriWave 10X to 50X better than T-wave Collision Cross Section (CCS) measurement (Ω) Direct determination of Ω Low electric field and constant drift tube pressure Ω cannot be directly determined from drift time. Need calibration tables. 1-2% precision Much better than Synapt (5-10%) Molecular structures Lower RF fields, less ion heating. Higher RF fields, tendency for higher fragmentation and ion heating Lower RF allows preservation of molecular structures LC Q IMS MS High Resolution Accurate Mass 35
36 Summary Next generation of IM Q-TOF Technology Added dimension of separation based on size, charge and molecular conformation Resolve and characterize the complex samples -- Increased peak capacity Direct determination collision cross sections Preservation of molecular structures 36
37 Acknowledgments Ruwan Kurulugama Alex Mordehai Mark Werlich Chris Klein Tom Knotts Ed Darland Gregor Overney Bill Barry Nathan Sanders Lane Howard Mikhail Ugarov Gene Wong Yergeny Kaplun George Stafford Bill Frazier Bruce Wang Huy Bui Crystal Cody John Fjelsted Ken Imatani Robert Kincaid Robin Scheiderer Collaborators: PNNL NIH Texas A&M Vanderbilt University Boston University 37
38 38
39 Drift time (ms) Standard Procedure for Calculating CCS (Drift Time versus P/V Curves) Colchicine Reserpine Ondansetron t D = experimental drift time t 0 = time ions spend out side the drift cell t d = corrected drift time y = 8,019x R² = y = 6258x Ondansetron R² = y = 5,410x R² = Reserpine Colchicine P/V (Torr/V) Compound Drift time (ms) Slope Intercept t 0 (ms) Ondansetron Colchicine Reserpine
40 Standard Procedure for Calculating CCS - Auto Stepping the Low Field Gradient Time segments with different drift fields 40
41 Collision Cross Section (Å 2 ) Conformational Space Occupancy of Biomolecules Hypothetical Ordering of Biomolecular Classes lipids peptides carbohydrates oligonucleotides Mass (Da) 41
42 Ion Mobility System Design Uniform DC electric field (no RF voltage applied) Desolvation chamber at 0 V ~1700 V on chamber Ion source maintained at ground potential (no voltage offset) Uniform low static electric field across drift cell 80 cm long, approximately 20 V/cm with 4 Torr Nitrogen buffer gas Mobility resolution approaches theoretical limit Minimizes ion excitation or heating (helps to maintain ion structures and conformations) Uniform low field ion mobility allows direct determination of accurate CCS (Ω) 42
43 Ion Mobility Hardware Design Drift Cell Uniform field drift cell 80 cm Front funnel Trapping funnel Drift Cell (80 cm) Rear funnel Ion funnel technology improves sensitivity 43
Advantages of Ion Mobility Q-TOF for Characterization of Diverse Biological Molecules
Advantages of Ion Mobility Q-TOF for Characterization of Diverse Biological Molecules Add a New Dimension to your Research Capability with Agilent s New Drift Ion Mobility Q-TOF System Overview: 6560 IM
More informationTargeted and untargeted metabolic profiling by incorporating scanning FAIMS into LC-MS. Kayleigh Arthur
Targeted and untargeted metabolic profiling by incorporating scanning FAIMS into LC-MS Kayleigh Arthur K.Arthur@lboro.ac.uk Introduction LC-MS is a highly used technique for untargeted profiling analyses
More informationIntegrating omics for systems biology
Structural mass spectrometry for the analysis of complex biological samples: ion mobility-mass spectrometry for broad-scale systems and synthetic biology John A. McLean Laboratory for Structural Mass Spectrometry
More informationNew Developments in LC-IMS-MS Proteomic Measurements and Informatic Analyses
New Developments in LC-IMS-MS Proteomic Measurements and Informatic Analyses Erin Shammel Baker Kristin E. Burnum-Johnson, Xing Zhang, Cameron P. Casey, Yehia M. Ibrahim, Matthew E. Monroe, Tao Liu, Brendan
More informationSYNAPT G2-S High Definition MS (HDMS) System
SYNAPT G2-S High Definition MS (HDMS) System High performance, versatility, and workflow efficiency of your MS system all play a crucial role in your ability to successfully reach your scientific and business
More informationApplying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis
Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis Ying Qing Yu Waters Corporation, Milford, MA, USA APPLICATION BENEFITS
More informationProfiling Flavonoid Isomers in Highly Complex Citrus Juice Samples Using UPLC Ion Mobility Time-of-Flight Mass Spectrometry
Profiling Flavonoid Isomers in Highly Complex Citrus Juice Samples Using UPLC Ion Mobility Time-of-Flight Mass Spectrometry Michael McCullagh, 1 Kieran Neeson, 2 and Antonietta Gledhill 1 Waters Corporation,
More informationIncreasing Molecular Coverage in Complex Biological and Environmental Samples by Using IMS-MS
Increasing Molecular Coverage in Complex Biological and Environmental Samples by Using IMS-MS Erin Shammel Baker Kristin E. Burnum-Johnson, Jon M. Jacobs, Yehia M. Ibrahim, Daniel J. Orton, William F.
More informationFOURIER TRANSFORM MASS SPECTROMETRY
FOURIER TRANSFORM MASS SPECTROMETRY https://goo.gl/vx3ogw FT-ICR Theory Ion Cyclotron Motion Inward directed Lorentz force causes ions to move in circular orbits about the magnetic field axis Alan G. Marshall,
More informationChallenges in Separation Technologies
Challenges in Separation Technologies Erin Shammel Baker Xing Zhang, Yehia Ibrahim, Matthew Monroe, Dennis Mehinagic, Justin Teeguarden, Thomas Metz and Richard D. Smith Pacific Northwest National Laboratory
More informationMass Spectrometry at the Laboratory of Food Chemistry. Edwin Bakx Laboratory of Food Chemistry Wageningen University
Mass Spectrometry at the Wageningen University Mass Spectrometry at the 3 UPLC/CE - ESI - Ion trap MS systems UPLC Thermo Acella with a Velos or VelosPro CE Beckman PA800 with a Thermo VelosPro 1 UPLC-
More informationData Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section
Data Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section Emmanuelle Claude, Mark Towers, and Kieran Neeson Waters Corporation, Manchester,
More informationConsiderations of the use of Triple Quadrupoles or Ion Traps in Quantitative Applications
Considerations of the use of Triple Quadrupoles or Ion Traps in Quantitative Applications Triple Stage Quadrupole API MS / MS Full Scan Products / - IONS AND NEUTRALS FORMED IN API SOURCE Q0 LENS TRANSPORTS
More informationLECTURE 3. Ionization Techniques for Mass Spectrometry
LECTURE 3 Ionization Techniques for Mass Spectrometry Jack Henion, Ph.D. Emeritus Professor, Analytical Toxicology Cornell University Ithaca, NY 14850 Lecture 3, Page 1 Contents Electron ionization (EI)
More informationA NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.
A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS. Martin R Green*, Keith Richardson, John Chipperfield, Nick Tomczyk, Martin
More informationTime (min) Supplementary Figure 1: Gas decomposition products of irradiated DMC.
200000 C 2 CH 3 CH 3 DMC 180000 160000 140000 Intensity 120000 100000 80000 60000 40000 C 2 H 6 CH 3 CH 2 CH 3 CH 3 CCH 3 EMC DEC 20000 C 3 H 8 HCCH 3 5 10 15 20 25 Time (min) Supplementary Figure 1: Gas
More informationO O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions
Simplifying Qual/Quan Analysis in Discovery DMPK using UPLC and Xevo TQ MS Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION The determination of the drug metabolism
More informationAdvances in Hybrid Mass Spectrometry
The world leader in serving science Advances in Hybrid Mass Spectrometry ESAC 2008 Claire Dauly Field Marketing Specialist, Proteomics New hybrids instruments LTQ Orbitrap XL with ETD MALDI LTQ Orbitrap
More informationQuadrupole and Ion Trap Mass Analysers and an introduction to Resolution
Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution A simple definition of a Mass Spectrometer A Mass Spectrometer is an analytical instrument that can separate charged molecules according
More information[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS
Yun Wang Alelyunas, Henry Shion, Mark Wrona Waters Corporation, Milford, MA, USA APPLICATION BENEFITS mab LC-MS method which enables users to achieve highly sensitive bioanalysis of intact trastuzumab
More information1. Sample Introduction to MS Systems:
MS Overview: 9.10.08 1. Sample Introduction to MS Systems:...2 1.1. Chromatography Interfaces:...3 1.2. Electron impact: Used mainly in Protein MS hard ionization source...4 1.3. Electrospray Ioniztion:
More informationBiological Mass Spectrometry. April 30, 2014
Biological Mass Spectrometry April 30, 2014 Mass Spectrometry Has become the method of choice for precise protein and nucleic acid mass determination in a very wide mass range peptide and nucleotide sequencing
More information[ APPLICATION NOTE ] APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS
[ APPLICATI TE ] Ion Mobility-enabled Data-dependent Experiments Distinguishing Co-eluting Isomeric Metabolites Using an IMS-QTof Mass Spectrometer Jayne Kirk, 1 Russell Mortishire Smith, 1 Robert Beecher,
More informationNew Mass Spectrometry Tools to Transform Metabolomics and Lipidomics
New Mass Spectrometry Tools to Transform Metabolomics and Lipidomics July.3.13 Ken Miller Vice President of Marketing, Life Sciences Mass Spectrometry 1 The world leader in serving science Omics & the
More information2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry
Dr. Sanjeeva Srivastava 1. Fundamental of Mass Spectrometry Role of MS and basic concepts 2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry 2 1 MS basic concepts Mass spectrometry - technique
More informationThe New 6495 Triple Quadrupole LC/MS
The New 6495 Triple Quadrupole LC/MS Experience A New Level of Confidence Martin Haex Product Specialist MS and Automation 1 Overview of Topics The New 6495 QQQ LC/MS Technology Innovations Quantitation
More informationCharacterization of an Unknown Compound Using the LTQ Orbitrap
Characterization of an Unknown Compound Using the LTQ rbitrap Donald Daley, Russell Scammell, Argenta Discovery Limited, 8/9 Spire Green Centre, Flex Meadow, Harlow, Essex, CM19 5TR, UK bjectives unknown
More informationMASS SPECTROMETRY BASED METABOLOMICS. Pavel Aronov. ABRF2010 Metabolomics Research Group March 21, 2010
MASS SPECTROMETRY BASED METABOLOMICS Pavel Aronov ABRF2010 Metabolomics Research Group March 21, 2010 Types of Experiments in Metabolomics targeted non targeted Number of analyzed metabolites is limited
More informationAnalysis of N-Linked Glycans from Coagulation Factor IX, Recombinant and Plasma Derived, Using HILIC UPLC/FLR/QTof MS
Analysis of N-Linked Glycans from Coagulation Factor IX, Recombinant and Plasma Derived, Using HILIC UPLC/FLR/QTof MS Ying Qing Yu Waters Corporation, Milford, MA, U.S. A P P L I C AT ION B E N E F I T
More informationMass spectrometry Technologies in Lipid chemistry
Mass spectrometry Technologies in Lipid chemistry Rabah Soliymani University Of Helsinki Protein Chemistry Unit Biomedicum Helsinki Rabah.soliymani@helsinki.fi Complex_&_dynamic_mixtures (few copies to
More informationAgilent s LC/MS Portfolio and Applications Examples
Agilent s LC/MS Portfolio and Applications Examples Sean Orlowicz LC/MS Product Specialist Agilent Technologies 31August Agilent s > 40 year heritage in mass spectrometry 1971, 5930A GC/MS 1971-2011 2005,
More informationSCS Mass Spectrometry Laboratory
SCS Mass Spectrometry Laboratory Contact Information Staff 31 Noyes Laboratory (8:00-5:00 M-F) 217-333-2545 http://scs.illinois.edu/massspec/ Furong Sun (frs@illinois.edu) Furong Sun Director Training
More informationApplication of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices
Application ote #LCMS-2 esquire series Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices Introduction The simple monitoring
More informationFundamentals of Soft Ionization and MS Instrumentation
Fundamentals of Soft Ionization and MS Instrumentation Ana Varela Coelho varela@itqb.unl.pt Mass Spectrometry Lab Analytical Services Unit Index Mass spectrometers and its components Ionization methods:
More informationTowards High Resolution MS in Regulated Bioanalysis
Towards High Resolution MS in Regulated Bioanalysis Benno Ingelse MSD 3 rd EB focus meeting June 12 th 13 th Brussels, Belgium Contributors Gary Adamson Ken Anderson Kevin Bateman Cynthia Chavez-Eng Inhou
More informationIon Source. Mass Analyzer. Detector. intensity. mass/charge
Proteomics Informatics Overview of spectrometry (Week 2) Ion Source Analyzer Detector Peptide Fragmentation Ion Source Analyzer 1 Fragmentation Analyzer 2 Detector b y Liquid Chromatography (LC)-MS/MS
More informationExtended Mass Range Triple Quadrupole for Routine Analysis of High Mass-to-charge Peptide Ions
Extended Mass Range Triple Quadrupole for Routine Analysis of High Mass-to-charge Peptide Ions Application Note Targeted Proteomics Authors Linfeng Wu, Christine A. Miller, Jordy Hsiao, Te-wei Chu, Behrooz
More informationBiotherapeutics. Biopharmaceutical Sciences Group Waters Corporation Waters Corporation 1
Characterization of the Higher-Order Structure of Biotherapeutics Biopharmaceutical Sciences Group Waters Corporation 2011 Waters Corporation 1 FDA desire for specific innovations in biotherapeutic analysis
More informationMetabolomics: quantifying the phenotype
Metabolomics: quantifying the phenotype Metabolomics Promises Quantitative Phenotyping What can happen GENOME What appears to be happening Bioinformatics TRANSCRIPTOME What makes it happen PROTEOME Systems
More informationFlow-Through Electron Capture Dissociation in a novel Branched RF Ion Trap
Flow-Through Electron Capture Dissociation in a novel Branched RF Ion Trap Takashi Baba, J. Larry Campbell, Yves Le Blanc, Jim. W. Hager and Bruce A. Thomson ASMS, June 18 / 2014 1 2014 AB SCIEX Trapping
More informationION MOBILITY COUPLED TO HIGH RESOLUTION MASS SPECTROMETRY: THE POSSIBILITIES, THE LIMITATIONS
ION MOBILITY COUPLED TO HIGH RESOLUTION MASS SPECTROMETRY: THE POSSIBILITIES, THE LIMITATIONS Anton Kaufmann Official Food Control Authority of the Canton of Zurich (Kantonales Labor Zürich) Switzerland
More informationChemical Analysis Business Operations Waters Corporation Milford MA
The Detection and Identification of Unknown Contaminants During ToF Screening and Structural Elucidation for Pesticides in River Water Using an Integrated Software Approach Chemical Analysis Business Operations
More informationUltra High Definition Optimizing all Analytical Dimensions
Ultra High Definition Optimizing all Analytical Dimensions Sensitivity Dynamic Range Signal Response Linearity Separation Speed Peak Capacity Chromatogram Mass Spectrum Mass Accuracy Resolving Power Acquisition
More informationAB Sciex QStar XL. AIMS Instrumentation & Sample Report Documentation. chemistry
Mass Spectrometry Laboratory AIMS Instrumentation & Sample Report Documentation AB Sciex QStar XL chemistry UNIVERSITY OF TORONTO AIMS Mass Spectrometry Laboratory Department of Chemistry, University of
More informationThe use of mass spectrometry in lipidomics. Outlines
The use of mass spectrometry in lipidomics Jeevan Prasain jprasain@uab.edu 6-2612 utlines Brief introduction to lipidomics Analytical methodology: MS/MS structure elucidation of phospholipids Phospholipid
More informationIntroduction to LC/MS/MS
Trends in 2006 Introduction to LC/MS/MS By Crystal Holt, LC/MS Product Specialist, Varian Inc. Toxicology laboratories Increased use of LC/MS Excellent LD Cheaper (still expensive) Much more robust Solves
More informationMore structural information with MS n
PRODUCT SPECIFICATIONS The LTQ XL linear ion trap mass spectrometer More structural information with MS n The LTQ XL linear ion trap mass spectrometer delivers more structural information faster and with
More information2013 Waters Corporation 1
High Definition Mass Spectrometry (HDMS) application of ion mobility in oil and petroleum analysis Patrice Lemire Waters Corporation WCTOW Vancouver, May 13, 2013 2013 Waters Corporation 1 Introduction
More informationSUPPORTING INFORMATION. Lysine Carbonylation is a Previously Unrecognized Contributor. to Peroxidase Activation of Cytochrome c by Chloramine-T
Electronic Supplementary Material (ESI) for Chemical Science. This journal is The Royal Society of Chemistry 2019 SUPPORTING INFORMATION Lysine Carbonylation is a Previously Unrecognized Contributor to
More informationComprehensive Two-Dimensional HPLC and Informative Data Processing for Pharmaceuticals and Lipids
PO-CON1576E Comprehensive Two-Dimensional HPLC and Informative Data Processing for Pharmaceuticals and Lipids HPLC 2015 PSB-MULTI-06 Yoshiyuki WATABE, Tetsuo IIDA, Daisuke NAKAYAMA, Kanya TSUJII, Saki
More informationRapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer
Robert Plumb, Michael D. Jones, and Marian Twohig Waters Corporation, Milford, MA, USA INTRODUCTION The detection and characterization of impurities and degradation products of an active pharmaceutical
More informationAmadeo R. Fernández-Alba
% of compounds % of compounds % of compounds % of compounds Amadeo R. Fernández-Alba LC-Orbitrap QExactive Focus Instrumental LOQ 1% 9% 8% 7% 6% 5% 4% 3% 2% 1% %.1 mg/g.2 mg/g Tomato.5 mg/g ddms2 Target
More informationOMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (10) 86 2R QUALIFICATION OF EQUIPMENT ANNEX 7: QUALIFICATION OF MASS SPECTROMETERS Full document title and reference Document
More informationAutomating Mass Spectrometry-Based Quantitative Glycomics using Tandem Mass Tag (TMT) Reagents with SimGlycan
PREMIER Biosoft Automating Mass Spectrometry-Based Quantitative Glycomics using Tandem Mass Tag (TMT) Reagents with SimGlycan Ne uaca2-3galb1-4glc NAcb1 6 Gal NAca -Thr 3 Ne uaca2-3galb1 Ningombam Sanjib
More informationMS/MS Library Creation of Q-TOF LC/MS Data for MassHunter PCDL Manager
MS/MS Library Creation of Q-TOF LC/MS Data for MassHunter PCDL Manager Quick Start Guide Step 1. Calibrate the Q-TOF LC/MS for low m/z ratios 2 Step 2. Set up a Flow Injection Analysis (FIA) method for
More informationComparison of mass spectrometers performances
Comparison of mass spectrometers performances Instrument Mass Mass Sensitivity resolution accuracy Quadrupole 1 x 10 3 0.1 Da* 0.5-1.0 pmol DE-MALDI 2 x 10 4 20 ppm 1-10 fmol peptide 1-5 pmol protein Ion
More informationStructural Elucidation of N-glycans Originating From Ovarian Cancer Cells Using High-Vacuum MALDI Mass Spectrometry
PO-CON1347E Structural Elucidation of N-glycans Originating From Ovarian Cancer Cells Using High-Vacuum MALDI Mass Spectrometry ASMS 2013 TP-708 Matthew S. F. Choo 1,3 ; Roberto Castangia 2 ; Matthew E.
More informationSimple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID
Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID James Little, Eastman Chemical Company, Kingsport, TN Overview Phospholipids and
More informationACHIEVEMENTS ARE LIMITED ONLY BY ASPIRATIONS.
ACHIEVEMENTS ARE LIMITED ONLY BY ASPIRATIONS. BE FIRST TO DISCOVER, TO PUBLISH, TO DEVELOP, TO SUCCEED. ENTER A NEW DIMENSION OF PERFORMANCE. Your success depends on your ability to optimize laboratory
More informationLatest Innovations in LC/MS/MS from Waters for Metabolism and Bioanalytical Applications
Latest Innovations in LC/MS/MS from Waters for Metabolism and Bioanalytical Applications Ignatius J. Kass Senior Field Marketing Manager Pharmaceutical MS Challenges in Pharmaceutical Sample Analysis Quantitative
More information4th Multidimensional Chromatography Workshop Toronto (January, 2013) Herman C. Lam, Ph.D. Calibration & Validation Group
4th Multidimensional Chromatography Workshop Toronto (January, 2013) Herman C. Lam, Ph.D. Calibration & Validation Group MDLC for Shotgun Proteomics Introduction General concepts Advantages Challenges
More informationBioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS
Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Jenny Chen, Hongxia Wang, Zhiqi Hao, Patrick Bennett, and Greg Kilby Thermo Fisher
More informationAdvancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations
Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations For Research Use Only. Not for use in diagnostic procedures. 1 2015 AB Sciex RUO-MKT-11-1018-A For research
More informationMoving from targeted towards non-targeted approaches
Gesundheitsdirektion Moving from targeted towards non-targeted approaches Anton Kaufmann Official Food Control Authority of the Canton of Zurich () Switzerland 2 Overview I From single residue to multi
More informationMass Spectrometry Infrastructure
Mass Spectrometry Infrastructure Todd Williams, Ph.D. Director KU Mass Spectrometry and Analytical Proteomics Laboratory Mass Spectrometry Lab B025 Malott Hall Mission The Mass Spectrometry and analytical
More informationLOCALISATION, IDENTIFICATION AND SEPARATION OF MOLECULES. Gilles Frache Materials Characterization Day October 14 th 2016
LOCALISATION, IDENTIFICATION AND SEPARATION OF MOLECULES Gilles Frache Materials Characterization Day October 14 th 2016 1 MOLECULAR ANALYSES Which focus? LOCALIZATION of molecules by Mass Spectrometry
More informationChoosing the metabolomics platform
Choosing the metabolomics platform Stephen Barnes, PhD Department of Pharmacology & Toxicology University of Alabama at Birmingham sbarnes@uab.edu Challenges Unlike DNA, RNA and proteins, the metabolome
More informationAn Alternative Approach: Top-Down Bioanalysis of Intact Large Molecules Can this be part of the future? Lecture 8, Page 27
An Alternative Approach: Top-Down Bioanalysis of Intact Large Molecules Can this be part of the future? Lecture 8, Page 27 Top-down HRAM Bioanalysis of Native Proteins/Molecules Relative Abundance 100
More informationMethods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University
Methods in Mass Spectrometry Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University Sample Engineering Chemistry Biology Life Science Medicine Industry IDF / Police Sample
More informationREDOX PROTEOMICS. Roman Zubarev.
REDOX PROTEOMICS Roman Zubarev Roman.Zubarev@ki.se Physiological Chemistry I, Department for Medical Biochemistry & Biophysics, Karolinska Institutet, Stockholm What is (RedOx) Proteomics? Proteomics -
More informationAutomated Lipid Identification Using UPLC/HDMS E in Combination with SimLipid
Automated Lipid Identification Using UPLC/HDMS E in Combination with SimLipid Giorgis Isaac, Stephen McDonald, Giuseppe Astarita Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S
More informationSue D Antonio Application Chemist Cedar Creek, TX
Sue D Antonio Application Chemist Cedar Creek, TX What is Hemp Oil? CBD hemp oil is a natural botanical extract of the common hemp plant. CBD hemp oil is derived from the seeds and stem of the Cannabis
More informationAgilent 6410 Triple Quadrupole LC/MS. Sensitivity, Reliability, Value
Agilent 64 Triple Quadrupole LC/MS Sensitivity, Reliability, Value Sensitivity, Reliability, Value Whether you quantitate drug metabolites, measure herbicide levels in food, or determine contaminant levels
More informationRobust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS
TECHNICAL NOTE 21882 Robust extraction, separation, and quantitation of structural isomer steroids human plasma by SPE-UHPLC-MS/MS Authors Jon Bardsley 1, Kean Woodmansey 1, and Stacy Tremintin 2 1 Thermo
More informationApplication Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform
Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform Abstract Targeted proteomics for biomarker verification/validation
More informationSolving practical problems. Maria Kuhtinskaja
Solving practical problems Maria Kuhtinskaja What does a mass spectrometer do? It measures mass better than any other technique. It can give information about chemical structures. What are mass measurements
More informationSUPPORTING INFORMATION FOR: CONCENTRATIONS OF POLYBROMINATED DIPHENYL ETHERS, HEXABROMOCYCLODODECANES AND TETRABROMOBISPHENOL-A IN BREAST MILK FROM
SUPPORTING INFORMATION FOR: CONCENTRATIONS OF POLYBROMINATED DIPHENYL ETHERS, HEXABROMOCYCLODODECANES AND TETRABROMOBISPHENOL-A IN BREAST MILK FROM UNITED KINGDOM WOMEN DO NOT DECREASE OVER TWELVE MONTHS
More informationMatrix Assisted Laser Desorption Ionization Time-of-flight Mass Spectrometry
Matrix Assisted Laser Desorption Ionization Time-of-flight Mass Spectrometry Time-of-Flight Mass Spectrometry. Basic principles An attractive feature of the time-of-flight (TOF) mass spectrometer is its
More informationComparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D
Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D Julie A. Horner 1, Marta Kozak 1, Subodh Nimkar 1, and August A. Specht 1 1 Thermo Fisher Scientific, San Jose,
More informationQuantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System
Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System Marta Kozak Clinical Research Applications Group Thermo Fisher Scientific San Jose CA Clinical Research use only, Not for
More informationUPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes
UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes Catalin E. Doneanu, Weibin Chen, and Jeffrey R. Mazzeo Waters Corporation, Milford, MA, U.S. A P P L I C AT ION B E N E F
More informationDevelopment of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid
Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid Joanne ( 乔安妮 ) Mather Senior Scientist Waters Corporation Data courtesy of Erin Chambers and Mary
More informationDr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1
Development of an SPE-LC/MS/MS Assay for the Simultaneous Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid in Support of Alzheimer s Research Dr. Erin E. Chambers Waters Corporation Presented
More informationSmall Molecule Science: Experimental designs for achieving ultra trace analysis
Small Molecule Science: Experimental designs for achieving ultra trace analysis Michael P. Balogh Principal scientist Waters Corporation 211 Waters Corporation 1 www.cosmoscience.org The Society for Small
More informationIdentification of Ginsenosides Using the SCIEX X500R QTOF System
Identification of Ginsenosides Using the SCIEX X500R QTOF System Wang Sha, Cheng Haiyan, Liu Ting, Li Lijun, Jin Wenhai[Author] SCIEX, Pacific Applications Support Center (Beijing). China Background Ginseng
More informationBenefits and Characteristic Applications of High Resolution GC/MS and LC/MS. Frank David RIC and Ghent University
Benefits and Characteristic Applications of High Resolution GC/MS and LC/MS. Frank David RIC and Ghent University Mass Spectrometry Structure Elucidation Selective and Sensitive Detection Identification
More informationRapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS
Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS Mark Ritchie and Evelyn Goh Waters Pacific Pte Ltd., Singapore A P P L I C AT ION B E N E F I T S Delivers a rapid 10-min MRM method
More informationNIH Public Access Author Manuscript J Proteome Res. Author manuscript; available in PMC 2014 July 05.
NIH Public Access Author Manuscript Published in final edited form as: J Proteome Res. 2013 July 5; 12(7): 3071 3086. doi:10.1021/pr3011588. Evaluation and Optimization of Mass Spectrometric Settings during
More informationHigh-sensitivity Orbitrap mass analysis of intact macromolecular assemblies. R. J. Rose, E. Damoc, E. Denisov, A. Makarov, A. J. R.
High-sensitivity Orbitrap mass analysis of intact macromolecular assemblies R. J. Rose, E. Damoc, E. Denisov, A. Makarov, A. J. R. Heck SUPPLEMENTARY INFORMATION HCD multipole C -trap Transport octapole
More informationMALDI-TOF. Introduction. Schematic and Theory of MALDI
MALDI-TOF Proteins and peptides have been characterized by high pressure liquid chromatography (HPLC) or SDS PAGE by generating peptide maps. These peptide maps have been used as fingerprints of protein
More informationMetabolite identification in metabolomics: Database and interpretation of MSMS spectra
Metabolite identification in metabolomics: Database and interpretation of MSMS spectra Jeevan K. Prasain, PhD Department of Pharmacology and Toxicology, UAB jprasain@uab.edu utline Introduction Putative
More informationTechnical Note # TN-31 Redefining MALDI-TOF/TOF Performance
Bruker Daltonics Technical Note # TN-31 Redefining MALDI-TOF/TOF Performance The new ultraflextreme exceeds all current expectations of MALDI-TOF/TOF technology: A proprietary khz smartbeam-ii TM MALDI
More informationMetabolite identification in metabolomics: Metlin Database and interpretation of MSMS spectra
Metabolite identification in metabolomics: Metlin Database and interpretation of MSMS spectra Jeevan K. Prasain, PhD Department of Pharmacology and Toxicology, UAB jprasain@uab.edu Outline Introduction
More informationQuantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer
application Note Liquid Chromatography/ Mass Spectrometry Authors Sharanya Reddy Blas Cerda PerkinElmer, Inc. Shelton, CT USA Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate
More informationRapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition
Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition Evelyn Goh Waters Pacific, Singapore APPLICATION BENEFITS This method allows for the simultaneous analysis of 12 water-soluble
More informationThe J105 SIMS. A New Instrument for 3-Dimensional Imaging and Analysis. Paul Blenkinsopp, Ionoptika Ltd
The J105 SIMS A New Instrument for 3-Dimensional Imaging and Analysis Paul Blenkinsopp, Ionoptika Ltd The J105 SIMS Why a new ToF Mass Spectrometer? The J105 ToF has been designed to allow us to separate
More informationShotgun Proteomics MS/MS. Protein Mixture. proteolysis. Peptide Mixture. Time. Abundance. Abundance. m/z. Abundance. m/z 2. Abundance.
Abundance Abundance Abundance Abundance Abundance Shotgun Proteomics Protein Mixture 1 2 3 MS/MS proteolysis m/z 2 3 Time µlc m/z MS 1 m/z Peptide Mixture m/z Block Diagram of a Mass Spectrometer Sample
More informationLecture 3. Tandem MS & Protein Sequencing
Lecture 3 Tandem MS & Protein Sequencing Nancy Allbritton, M.D., Ph.D. Department of Physiology & Biophysics 824-9137 (office) nlallbri@uci.edu Office- Rm D349 Medical Science D Bldg. Tandem MS Steps:
More informationImpurity Identification using a Quadrupole - Time of Flight Mass Spectrometer QTOF
Impurity Identification using a Quadrupole - Time of Flight Mass Spectrometer QTOF PUSHER TOF DETECTOR ZSPRAY TM Ion Source SAMPLING CONE SKIMMER RF HEXAPOLE RF HEXAPOLE QUADRUPOLE IN NARROW BANDPASS MODE
More information