1 Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography Wilson Otto Gomes Batista Instituto Federal da Bahia Rua Emidio dos Santos s/n. Barbalho Salvador, Bahia, Brasil wilsonottobatista@gmail.com Abstract This study aimed to develop a geometry of irradiation applicable to the Software PCXMC and the consequent calculation of effective dose in applications of the Computed Tomography Cone Beam (CBCT). We evaluated two different teams of CBCT for dental applications: Carestream CS D tomograph; i-cat and GENDEX GXCB-500. Initially characterize each protocol measuring the surface kerma input and the product kerma air -area, P KA, with solid state detectors RADCAL and PTW transmission chamber. Then we introduce the technical parameters of each preset protocols and geometric conditions in the PCXMC software to obtain the values of effective dose. The calculated effective dose is within the range of 9.0 to 15.7 Sv for 3D computer 9000 CS; within the range 44.5 to 89 Sv for GXCB-500 equipment and in the range of Sv for equipment Classical i-cat. These values were compared with results obtained dosimetry using TLD implanted in anthropomorphic phantom and are considered consistent. Os effective dose results are very sensitive to the geometry of radiation (beam position in mathematical phantom). This factor translates to a factor of fragility software usage. But it is very useful to get quick answers to regarding process optimization tool conclusions protocols. We conclude that use software PCXMC Monte Carlo simulation is useful assessment protocols for na tests CBCT in dental applications. Keywords: Monte Carlo simulation, effective dose, dental tomography. 237

2 1. INTRODUCTION Technology Computed Tomography Cone Beam - CBCT was introduced in the late 1990s in dental radiology. Just as in other types of radiological image there is the need to review existing protocols and / or taken to obtain images of sufficient quality for a reliable diagnosis and also with the optimization purpose for proper application of techniques that use ionizing radiation. Usually tests protocols assessments are made, from the point of view of optimization of exposures with the use of magnitude effective dose (Batista et al., 2012; INTERNATIONAL ATOMIC ENERGY AGENCY, 2004). There are several techniques to evaluate the effective dose quantity. Among these various methods include those based experimental techniques in the use of detectors type TLD and OSL inserted in anthropomorphic phantoms (Soares et al., 2015b) and also, at the present time the use of radiochromic films (Al-Okshi et al., 2013). These methods are considered extensively laborious, require extensive time, several steps up to the data acquisition and also have significant associated uncertainties (Lee et al., 2016; Soares et al., 2015a). Another option is the use of conversion factors that may be derived from the dosimetry data with the aforementioned detectors or computer simulation data and associated with direct measurements of a physical quantity that characterizes the exposure conditions, such as kerma incident air; kerma input surface; product kerma air-area etc. obtained using ionisation chambers. The values of these quantities will be used as input parameters of the computational methods (Lee et al., 2016). The software using the Monte Carlo method for calculating effective dose in radiology are an alternative for this type of evaluation for numerous radiological imaging techniques. However, for technical panoramic radiography and computed tomography cone beam in dentistry using software prepared for use in medical radiology modalities not a simple task mainly due to the geometry and rotation of the beam around the patient (Koivisto et al., 2013; Lee et al., 2016). The software PCXMC is a Monte Carlo simulation application adapted for use in personal computer developed by Radiation and Nuclear Safety Authority in Finland (Tapiovaara M, 1997). Currently this software already incorporates the new data from ICRP 103 de 2007 (ICRP, 2007) and a module for rotational techniques. In this context the objective of this work is to implement a geometry and input parameters in the software based on Monte Carlo method to assess effective dose values in computed tomography cone beam for dental applications. The results will be compared with values obtained using dosimeters (TLD) inserted in anthropomorphic phantom in previous studies and has published (Batista et al., 2015; Soares et al., 2015a, 2015b). 238

3 2. Materials and methods To calculate the effective dose selected five test protocols in three different equipment we have effective dose data obtained from measurements with TLDs inserted into anthropomorphic phantom. We conducted a dosimetric characterization for these protocols in terms of kerma acquired with solid state sensors RADCAL Accu Gold + with AGMS-DM+ sensor and kerma product values in the air area (PKA) measured Diamentor E2 transmission ionization chamber Protocols The five protocols evaluated examinations were derived from three devices Carestream CS 9000; Gendex GXCB 500 and i-cat Classical. The protocols are identified by numbers and listed in Table 1. Table 5 Data of the five evaluated protocols. dosimetric information and geometric parameters and technical protocols. Protocol Equipment kv ma mas P KA FOV 1 FRD (mgy.cm 2 ) diameter x higher (cm) P1 P2 * P3 P4 P5 CS 9000 CS 9000 GXCB 500 i-cat Classical i-cat Classical cm x 3.7 cm 8.0 cm x 3.7 cm 8.5 cm x 8.5 cm 16.0 cm x 8.0 cm 16.0 cm x 6.0 cm FOV Field of View. *This FOV is a combination of 3 FOVs with 5.0 cm x 3.7 cm The protocols with combination FOVs are common in current equipment. Thus, to evaluate a geometry has been constructed and shown below in Figure 1, as a result of the combination of three FOVs 5.0 cm x 3.7 cm. Figure 1 shows a combination of three FOVs to form an image, example of equipment Carestream CS Figure 8 Representation of the combination of three FOVs to form an image. Adapted from 239

4 Zref, (cm) Zref, (cm) variable ISSSD Calculations with PCXMC For the simulation using the software PCXMC with the supplement for rotational techniques (Koivisto et al., 2012; Tapiovaara M, 1997). The necessary input data for simulation of execution are: patient data; beam parameters; irradiation geometry. Table 2 shows the input data to perform the actual dose calculations. Table 6 Input parameters for the software PCXMC. Protocol Filtration X-ray beam Input dose kv Xref Yref Zref number (mm Al) Width (cm) Height (cm) data P P KA value P2* variable P KA value P P KA value P P P KA value P P KA value * This FOV is a combination of 3 FOVs with 5.0 cm x 3.7 cm Figure 2 shows an example of geometry for generating the rotation and consequent calculation data. It is important to note that the reference point is located in the center of the FOV. That is, the focus distance reference (FRD) is the distance from the focal point to the center of the FOV Xref, (cm) Série1 FRD Yref, (cm) Série Figura 10 Geometria de irradiação e representação do FOV. Figura 9 Geometria de irradiação e representação do FOV. It used a -5 degree tilt in the flow direction to represent the configuration of the beams in CBCT that limits the divergence of the beam on the lower plane, thus avoiding high exposure of the thyroid gland. 240

5 The calculations were performed for full 360 degree rotation in increments of 2; 5:10 degrees to determine if there are influences the number of increments. Also, has been varied-zref, ie, the Z coordinate of the center of the FOV. 3. RESULTS The results obtained using rotation in increments of 1; 2; 5:10 degree influenced the results at 2.3% in the effective dose. Thus we chose 10 degree increment used to reduce the computation time. In the Table 3 presents the effective dose values for the three protocols evaluated. Table 7 - Effective dose values for the evaluated protocols. Protocols P2 P3 P4 P5 P6 Zref Effective Dose ( Sv) Individual and important contributions to effective dose assessment in dental radiology are presented in graphs, Figures 3, 4, 5, 6, 7 e 8. Figure 11 - Contributions to the effective dose of organ and tissue in Sv derived from P4 protocol. 241

6 Figure Contributions to the effective dose of organ and tissue in Sv derived from P3 protocol. Figure 13 - Contributions to the effective dose of organ and tissue in Sv derived from P2 protocol. 242

7 Figure 14 - Contributions to the effective dose of organ and tissue in Sv derived from P5 protocol. Figure 15 - Contributions to the effective dose of organ and tissue in Sv derived from P6 protocol. The effective dose values are dependent on the position Zref. This is because depending on this position some organs or tissues may or may not be included within FOV. Figure 9 shows the dependence of the effective dose to the Zref position of the center of the FOV. 243

8 Effective Dose, (microgray) ISSSD Zref position, (cm) dependence with the Zref position - P4 dependence with the Zref position - P3 dependence with the Zref position - P3 dependence with the Zref position- P5 dependence with the Zref position - P6 Figura 16 - dependence of effective dose vales with the Zref position. 4. DISCUSSION The contributions of individual organs and tissues, of course, depends on or not it is included within the field of vision, ie within the radiation beam. When the image purpose is the lower jaw must be considered gland thyroid. Because this is very sensitive to radiation and thus contribute to the effective dose more significantly. After publication of the ICRP # with your new factors and explicitness the salivary glands and the inclusion of the oral mucosa in other tissues changed the panorama of effective dose assessments in dentistry radiology (ICRP, 2007). This dependence can be visualized through the graph, Figure 9. Comparing the effective dose values presented in Table 4 with studies conducted with TLD for these same protocols (Batista et al., 2015; Soares et al., 2015a). It is observed that the maximum and minimum differences are presented in the Table 4: 244

9 Table 8 - Comparing between measure and calculated the effective dose values. Maximum and minimum differences Protocols P2 P3 P4 P5 P6 Zref Effective Dose ( Sv)/ Differences (%) % % % % % % % % % % % % % % % % % % % % % % % % It becomes clear dependence with Zref position. It is not easy to ensure real Zref position in studies with TLDs. However, other studies show differences of the same order of magnitude between assessments with TLDs and effective dose values calculated PCXMC (Lee et al., 2016) and (Koivisto et al., 2012). Such differences appear to be significant do not invalidate the use of simulation with PCXMC software. It is a practical and useful way in demonstrating the importance of positioning and thus contributing to the optimization of protocols. 5. CONCLUSION The results confirmed the usefulness and practicality in using PCXMC software for simulation and evaluation of effective doses in dental cone beam tomography. REERENCES Al-Okshi, A., Nilsson, M., Petersson, A., Wiese, M., Lindh, C., Using gafchromic film to estimate the effective dose from dental cone beam CT and panoramic radiography. Dentomaxillofacial Radiology 42. doi: /dmfr Batista, W.O., Soares, M.R., de Oliveira, M.V.L., Maia, A.F., Caldas, L.V.E., Assessment of protocols in cone-beam CT with symmetric and asymmetric beams usingeffective dose and air kerma-area product. Applied Radiation and Isotopes 100, doi: /j.apradiso Batista, W.O.G., Navarro, M.V.T., Maia, A.F., Effective doses in panoramic images from conventional and CBCT equipment. Radiation Protection Dosimetry 151, doi: /rpd/ncr

10 ICRP, ICRP 103: The 2007 Recommendations of the International Commission on Radiological Protection. Annals of the ICRP 37, 330. doi: /j.icrp INTERNATIONAL ATOMIC ENERGY AGENCY, Optimization of the Radiological Protection of Patients Undergoing Radiography, Fluoroscopy and Computed Tomography. INTERNATIONAL ATOMIC ENERGY AGENCY, Vienna. Koivisto, J., Kiljunen, T., Tapiovaara, M., Wolff, J., Kortesniemi, M., Assessment of radiation exposure in dental cone-beam computerized tomography with the use of metal-oxide semiconductor field-effect transistor (MOSFET) dosimeters and Monte Carlo simulations. Oral Surgery, Oral Medicine, Oral Pathology and Oral Radiology 114, doi: /j.oooo Koivisto, J., Kiljunen, T., Wolff, J., Kortesniemi, M., Assessment of effective radiation dose ofan extremity cbct, msctand conventional x ray for knee area using mosfet dosemeters. Radiation Protection Dosimetry 157, doi: /rpd/nct162 Lee, C., Lee, S.S., Kim, J.E., Huh, K.H., Yi, W.J., Heo, M.S., Choi, S.C., Comparison of dosimetry methods for panoramic radiography: Thermoluminescent dosimeter measurement versus personal computer-based Monte Carlo method calculation. Oral Surgery, Oral Medicine, Oral Pathology and Oral Radiology 121, doi: /j.oooo Soares, M.R., Batista, W.O., Antonio, P. de L., Caldas, L.V.E., Maia, A.F., 2015a. Study of effective dose of various protocols in equipment cone beam CT. Applied Radiation and Isotopes 100, doi: /j.apradiso Soares, M.R., Batista, W.O., de Lara Antonio, P., Caldas, L.V.E., Maia, A.F., 2015b. Effective dose comparison between stitched and single FOV in CBCT protocols for complete dental arcade. Radiation Physics and Chemistry 110, doi: /j.radphyschem Tapiovaara M, L.M.A.S.A., PCXMC: A PC-based Monte Carlo program for calculating patient doses in medical x-ray examinations. Helsinki. 246

Estimated radiation risk of cancer from dental cone-beam computed tomography imaging in orthodontics patients

Estimated radiation risk of cancer from dental cone-beam computed tomography imaging in orthodontics patients Yeh and Chen BMC Oral Health (2018) 18:131 https://doi.org/10.1186/s12903-018-0592-5 RESEARCH ARTICLE Estimated radiation risk of cancer from dental cone-beam computed tomography imaging in orthodontics

More information

Estimation of the effective dose of dental cone-beam computed tomography using personal computer-based Monte Carlo software

Estimation of the effective dose of dental cone-beam computed tomography using personal computer-based Monte Carlo software Imaging Science in Dentistry 2018; 48: 21-30 https://doi.org/10.5624/isd.2018.48.1.21 Estimation of the effective dose of dental cone-beam computed tomography using personal computer-based Monte Carlo

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Dosimetry analysis of panoramic-imaging devices in differentsized

Dosimetry analysis of panoramic-imaging devices in differentsized Received: 8 November 2016 Revised: 17 January 2017 Accepted: 19 January 2017 DOI: 10.1002/acm2.12059 RADIATION MEASUREMENTS Dosimetry analysis of panoramic-imaging devices in differentsized s Muizz A.

More information

Application of international standards to diagnostic radiology dosimetry

Application of international standards to diagnostic radiology dosimetry Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,

More information

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Investigation of the clinical performance of a novel solid-state diagnostic dosimeter Jason Tse, a Donald McLean Medical Physics and

More information

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of

More information

Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology

Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology Horst Aichinger Joachim Dierker Sigrid Joite-Barfuß Manfred Säbel Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology Physical Principles and Clinical Applications Second Edition Springer

More information

The impact of SID and collimation on backscatter in radiography

The impact of SID and collimation on backscatter in radiography The impact of SID and collimation on backscatter in radiography Poster No.: C-2931 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: M. Joyce 1, J. T. Ryan 2, M. F. Mc Entee

More information

RADIATION DOSES IN DENTAL PANORAMIC TOMOGRAPHY. J. Melgar, C. Martín, C. Montes, F. Sáez, P. Collado, P. Gómez

RADIATION DOSES IN DENTAL PANORAMIC TOMOGRAPHY. J. Melgar, C. Martín, C. Montes, F. Sáez, P. Collado, P. Gómez RADIATION DOSES IN DENTAL PANORAMIC TOMOGRAPHY J. Melgar, C. Martín, C. Montes, F. Sáez, P. Collado, P. Gómez Servicio de Radiofísica y Protección Radiológica. Hospital Universitario de Salamanca. Pº San

More information

Multislice CT versus cone beam CT: dosimetric and image quality comparision.

Multislice CT versus cone beam CT: dosimetric and image quality comparision. Multislice CT versus cone beam CT: dosimetric and image quality comparision. Poster No.: C-2589 Congress: ECR 2015 Type: Authors: Scientific Exhibit C. Dionisi 1, A. Crispo 2, E. Stefani 3, A. Lovaglio

More information

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development. Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte, MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 MODELLING A GAMMA IRRADIATION

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

Comparison of organ doses estimations in radiology with PCXMC application based on MIRD phantoms and CALDose-X application based on voxel phantoms

Comparison of organ doses estimations in radiology with PCXMC application based on MIRD phantoms and CALDose-X application based on voxel phantoms Comparison of organ doses estimations in radiology with PCXMC application based on MIRD phantoms and CALDose-X application based on voxel phantoms G. Gialousis 1, Z. Pappouli 2, A. Dimitriadis 2, E. Karavassilis

More information

User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT)

User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT) User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT) Poster No.: C-0756 Congress: ECR 2014 Type: Educational Exhibit Authors: J. Ukkonen, J. Asp; Helsinki/FI Keywords: Education

More information

Calibration of two 90 Sr+ 90 Y dermatological applicators

Calibration of two 90 Sr+ 90 Y dermatological applicators Calibration of two 90 Sr+ 90 Y dermatological applicators Patrícia L. Antonio *, Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares, IPEN-CNEN/SP Av. Prof. Lineu Prestes 2242, 05508-000,

More information

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS RADIOPROTECTION AND DOSIMETRY MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS DANIELA ADAM 1, ANA STOCHIOIU

More information

Chapter 10: Special Topics

Chapter 10: Special Topics Chapter 10: Special Topics Slide set of 71 slides based on the chapter authored by D. McLean and J. Shepherd of the publication (ISBN 978-92-0-131010-1): Diagnostic Radiology Physics: A Handbook for Teachers

More information

Conversion coefficients for the estimation of effective dose in cone-beam CT

Conversion coefficients for the estimation of effective dose in cone-beam CT Imaging Science in Dentistry 2014; 44: 21-9 http://dx.doi.org/10.5624/isd.2014.44.1.21 Conversion coefficients for the estimation of effective dose in cone-beam CT Dong-Soo Kim 1, Oyuntugs Rashsuren 1,

More information

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Clarita Saldarriaga Vargas, Corinne Amalberto, Lara Struelens,

More information

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: Rob Loader Oliver Gosling Introduction Approached by Dr Oliver Gosling (Research

More information

International Journal of Radiation Research, January 2017

International Journal of Radiation Research, January 2017 Downloaded from ijrr.com at : +0330 on Tuesday November th Volume 5, No International Journal of Radiation Research, January 7 Radiation dose to the thyroid, eyes and parotid glands of patients undergoing

More information

Performance of several active personal dosemeters in interventional radiology and cardiology

Performance of several active personal dosemeters in interventional radiology and cardiology Performance of several active personal dosemeters in interventional radiology and cardiology S. Chiriotti 1,*, M. Ginjaume 1, E. Vano 2,3, R. Sanchez 3, J.M. Fernandez 2,3, M.A. Duch 1, J. Sempau 1 1.

More information

Managing Cone Beam CT Dose in Paediatric Dental Imaging

Managing Cone Beam CT Dose in Paediatric Dental Imaging Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Managing Cone Beam CT Dose in Paediatric Dental Imaging Raija Seuri (HUS Medical Imaging Center, FI) Cristina Almeida (Centro Hospitalar

More information

Introduction and Background

Introduction and Background CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section

More information

Dosimetry of recently introduced CBCT Units for Oral and Maxillofacial Radiology

Dosimetry of recently introduced CBCT Units for Oral and Maxillofacial Radiology Dosimetry of recently introduced CBCT Units for Oral and Maxillofacial Radiology John B Ludlow, Laura E Davies-Ludlow, André Mol University of North Carolina, Chapel Hill, NC Background CBCT is seeing

More information

Mammography calibration qualities establishment in a Mo- Mo clinical system

Mammography calibration qualities establishment in a Mo- Mo clinical system Journal of Physics: Conference Series PAPER OPEN ACCESS Mammography calibration qualities establishment in a Mo- Mo clinical system To cite this article: E L Corrêa et al 2016 J. Phys.: Conf. Ser. 733

More information

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration Establishment of the new IEC 61267 mammography qualities in a clinical system used for instruments calibration 1 Corrêa, E.L.; 2 Vivolo, V.; 3 Potiens, M.P.A. Abstract Instituto de Pesquisas Energéticas

More information

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science PATIENT DOSIMETRY IN DIAGNOSTIC RADIOLOGY MODALITIES Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science Ankara University Institute of Nuclear Science USE OF RADIATION! INCREASING? Natural

More information

Agenda: Dental Cone Beam Imaging

Agenda: Dental Cone Beam Imaging Cone Beam Imaging Agenda: Dental Cone Beam Imaging *Definition and Functionality *Usage and diagnostics benefits *Comparative radiation information *Federal regulatory responsibilities: manufacturing *State

More information

Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters

Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters Patrícia de Lara Antonio 1*, Rogério M. V. Silva 2, Divanizia do Nascimento Souza 2 and Linda

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by: SRS PHANTOM SCAN PLAN LOCALIZE TREAT THE COMPREHENSIVE END-TO-END SRS PHANTOM distributed by: Tel: +33 (0) 42 88 68 41 info@orion-france.com www.orion-france.com 2, Avenue du General Balfourier 75016 Paris,

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY Room B1002, WIMR 1111 Highland Avenue Madison, WI 53705-2275 Larry A. DeWerd,

More information

ABSTRACT. Objectives: To investigate the radiation dose received by patients undergoing routine plain x-ray

ABSTRACT. Objectives: To investigate the radiation dose received by patients undergoing routine plain x-ray Radiation Dose Distribution for Patients Undergoing Routine Radiological Scans for Kidney Stone Diagnosis at the University Hospital of the West Indies DC Walker 1, WD Aiken 2, S Shah 3, MK Voutchkov 1,

More information

Diagnostic reference levels in intraoral dental radiography in Korea

Diagnostic reference levels in intraoral dental radiography in Korea Imaging Science in Dentistry 2012; 42 : 237-42 http://dx.doi.org/10.5624/isd.2012.42.4.237 Diagnostic reference levels in intraoral dental radiography in Korea Eun-Kyung Kim, Won-Jeong Han, Jin-Woo Choi,

More information

Kerma Area Product Method for Effective Dose Estimation During Lumbar Epidural Steroid Injection Procedures: Phantom Study

Kerma Area Product Method for Effective Dose Estimation During Lumbar Epidural Steroid Injection Procedures: Phantom Study Vascular and Interventional Radiology Original Research Kim et al. Dose for Lumbar Steroid Injection Vascular and Interventional Radiology Original Research Sangroh Kim 1 Greta Toncheva 2,3 Colin Anderson-Evans

More information

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report www.siemens.com/healthcare SOMATOM Drive System Owner Manual Dosimetry and imaging performance report Table of contents 1 Dosimetry and imaging performance report 5 1.1 Dose information 5 1.1.1 General

More information

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers Patient effective dose evaluation for chest X-ray examination in three digital radiography centers R. Paydar 1,3*, A. Takavar 1, M.R. Kardan 2,3, A. Babakhani 3,4, M.R. Deevband 3, S. Saber 5 1Medical

More information

EFFECTIVE DOSE AND IMAGE QUALITY OF CBCT DIAGNOSTIC IMAGING SCANS. Cameron G. Walker DDS, PhD. Chapel Hill 2012

EFFECTIVE DOSE AND IMAGE QUALITY OF CBCT DIAGNOSTIC IMAGING SCANS. Cameron G. Walker DDS, PhD. Chapel Hill 2012 EFFECTIVE DOSE AND IMAGE QUALITY OF CBCT DIAGNOSTIC IMAGING SCANS Cameron G. Walker DDS, PhD A thesis submitted to the faculty of the University of North Carolina at Chapel Hill in partial fulfillment

More information

Assessment of effective dose in paediatric CT examinations

Assessment of effective dose in paediatric CT examinations Assessment of effective dose in paediatric CT examinations E. Dougeni 1,2 CL. Chapple 1, J. Willis 1, G. Panayiotakis 2 1 Regional Medical Physics Department, Freeman Hospital, Freeman Road, Newcastle

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements Standards Revisions for Organizations Providing Fluoroscopy Services The Joint Commission has approved the following revisions for prepublication. While revised requirements

More information

Effective dose from direct and indirect digital panoramic units

Effective dose from direct and indirect digital panoramic units Imaging Science in Dentistry 2013; 43: 77-84 http://dx.doi.org/10.5624/isd.2013.43.2.77 Effective dose from direct and indirect digital panoramic units Gun-Sun Lee, Jin-Soo Kim, Yo-Seob Seo, Jae-Duk Kim

More information

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Radiation Safety Characteristics of the NOMAD Portable X-ray System Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 ARIBA X-Ray, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribaxray.com

More information

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

Manik Aima, Larry A. DeWerd, Wesley S. Culberson Manik Aima, Larry A. DeWerd, Wesley S. Culberson University of Wisconsin Medical Radiation Research Center, Madison, WI 25 th Annual Meeting of the Council of Ionizing Radiation Measurements and Standards,

More information

Bureau of Laboratory Quality Standards

Bureau of Laboratory Quality Standards 1. Condom -Lubricant ISO 4074:2002/Cor.1:2003 Annex C -Thickness ISO 4074:2002/Cor.1:2003 Annex F - Length ISO 4074:2002 / Cor1 : 2003 - Width ISO 4074 : 2002/Cor 1 : 2003 - Bursting pressure and Volume

More information

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center CT Dose Estimation John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology 1 University of California Davis Medical Center CT Dose Estimation Introduction The CTDI Family of Metrics

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

Unrivaled, End-to-End

Unrivaled, End-to-End PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by

More information

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS Etieli Camargo da Costa 1, Luiz Antonio Ribeiro da Rosa 2 Delano Valdivino Santos Batista 3 1 Instituto de Radioproteção

More information

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

A more accurate method to estimate patient dose during body CT examinations with tube current modulation A more accurate method to estimate patient dose during body CT examinations with tube current modulation Poster No.: C-0738 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Kawaguchi 1, Y. Matsunaga

More information

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou. Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution

More information

Characterization of MOSFET dosimeters for low-dose measurements in maxillofacial anthropomorphic phantoms

Characterization of MOSFET dosimeters for low-dose measurements in maxillofacial anthropomorphic phantoms JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Characterization of MOSFET dosimeters for low-dose measurements in maxillofacial anthropomorphic phantoms Juha H. Koivisto, 1a Jan

More information

Comparison of adsorbed skin dose received by patients in cone beam computed tomography, spiral and conventional computed tomography scanninng

Comparison of adsorbed skin dose received by patients in cone beam computed tomography, spiral and conventional computed tomography scanninng (1390 3 24 ) 3 3 2 - - - 1-1 -2-3 Comparison of adsorbed skin dose received by patients in cone beam computed tomography, spiral and conventional computed tomography scanninng Ghazi Khanlou Sani K 1, Eskandarlou

More information

Assessing the standard dose to standard patients for x-ray investigations

Assessing the standard dose to standard patients for x-ray investigations Assessing the standard dose to standard patients for x-ray investigations A. Almén and W. Leitz Swedish Radiation Protection Authority, SE-171 16 Stockholm, Sweden E-mail: anja.almen@ssi.se Abstract. The

More information

Evaluation of the doses delivered to premature babies in the Belgian Neonatal Intensive Care Units

Evaluation of the doses delivered to premature babies in the Belgian Neonatal Intensive Care Units Evaluation of the doses delivered to premature babies in the Belgian Neonatal Intensive Care Units PreDos Project Jérémie Dabin Lara Struelens Filip Vanhavere Table of Contents Executive summary... Introduction...

More information

Radiologic Units: What You Need to Know

Radiologic Units: What You Need to Know Radiologic Units: What You Need to Know TODD VAN AUKEN M.ED. RT (R)(MR) Agenda Greys, Sieverts, Coulombs per kg, & Becquerel's Conventional Units Other Concepts (LET, Q-Factor, Effective Dose, NCRP Report

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

Basic dosimetric data for treatment planning system

Basic dosimetric data for treatment planning system Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland IGRT1 technologies Paweł Kukołowicz Warsaw, Poland Minimal prerequisite for good, efficient radiotherapy ICTP 2015 Paweł Kukołowicz 2/29 Minimal prerequisite for good, efficient radiotherapy Well trained

More information

Planmeca Ultra Low Dose

Planmeca Ultra Low Dose OWN THE FUTURE Practice Ultra Low Dose 3D imaging with lower dose than panoramic imaging Ultra Low Dose Pioneering low dose 3D imaging 3D units offer a unique Ultra Low Dose imaging protocol that enables

More information

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals Iran. J. Radiat. Res., 2006; 4 (1): 29-33 Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals D. Shahbazi-Gahrouei * Department of Medical Physics and

More information

Quality and Safety of Patient Healthcare and Personalized Medicine Using Ionizing Radiation

Quality and Safety of Patient Healthcare and Personalized Medicine Using Ionizing Radiation Quality and Safety of Patient Healthcare and Personalized Medicine Using Ionizing Radiation Ana Catarina Antunes, Mariana Baptista, Ana Belchior, Jorge Borbinha, Salvatore di Maria, Célia Fernandes, Joana

More information

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith

More information

Asymmetric breast dose in coronary angiography

Asymmetric breast dose in coronary angiography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Asymmetric breast dose in coronary angiography Kam Shan Mong, 1,3a Anthony B. Wallace, 2 and Rick D. Franich 3 Medical Physics Department,

More information

Skyscan 1076 in vivo scanning: X-ray dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in

More information

Measurement of Absorbed Doses in Anatomical Phantoms

Measurement of Absorbed Doses in Anatomical Phantoms International Journal of Pure and Applied Physics. ISSN 0973-1776 Volume 8, Number 1 (2012), pp. 53-58 Research India Publications http://www.ripublication.com/ijpap.htm Measurement of Absorbed Doses in

More information

Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations

Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations

More information

Scientific Affairs. Scientific Affairs

Scientific Affairs. Scientific Affairs The use of cone-beam computed tomography in dentistry : An advisory statement from the American Dental Association Council on Scientific Affairs The American Dental Association Council on Scientific Affairs

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

Staff Exposure Monitoring in Interventional Radiology

Staff Exposure Monitoring in Interventional Radiology Conference on Physics in Medicine: From Diagnosis to Treatment "Enhancing safety and quality in radiation medicine". KFMC, Riyadh 7-9 November 2017 Staff Exposure Monitoring in Interventional Radiology

More information

Bwanga Phanzu. Degree of Master of Science in Dentistry by coursework and dissertation

Bwanga Phanzu. Degree of Master of Science in Dentistry by coursework and dissertation i EFFECTIVE DOSE OF RADIATION ON THE EYE, THYROID AND PELVIC REGION RESULTING FROM EXPOSURES TO THE GALILEOS COMFORT CONE BEAM COMPUTERIZED TOMOGRAPHIC SCANNER Bwanga Phanzu Degree of Master of Science

More information

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing

More information

ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED *

ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED * ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED * Olivera Ciraj-Bjelac 1**, Danijela Arandjic 1, Predrag Bozovic 1, Sandra Ceklic 2, Jelena Stankovic

More information

=::> (- j"' {_ }.; I {;, Effective Dose of Positioning Scans for Five CBCT Devices. Stona R. Jackson APPROVED:

=::> (- j' {_ }.; I {;, Effective Dose of Positioning Scans for Five CBCT Devices. Stona R. Jackson APPROVED: Effective Dose of Positioning Scans for Five CBCT Devices Stona R. Jackson APPROVED: Ryan L. Snyder, D.D.S., M.S., Supervising Professor ----=::> ------ - Brent J. Calle ari, D.D.S., M.S.D., Program Director

More information

Chu Wang. Graduate Program inmedical Physics Duke University. Date: Approved: Terry Yoshizumi, Supervisor. RathnayakaGunasingha.

Chu Wang. Graduate Program inmedical Physics Duke University. Date: Approved: Terry Yoshizumi, Supervisor. RathnayakaGunasingha. Evaluation of Patient Effective Dose of Neurovascular Imaging Protocols of a C- arm Cone-beam CT & Estimation of Current Source Radioactivity of a Cs-137 Irradiator by Chu Wang Graduate Program inmedical

More information

UvA-DARE (Digital Academic Repository)

UvA-DARE (Digital Academic Repository) UvA-DARE (Digital Academic Repository) Effective radiation dose of a MSCT, two CBCT and one conventional radiography device in the ankle region Koivisto, J.; Kiljunen, T.; Kadesjö, N.; Shi, X.Q.; Wolff,

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

organizations including the International Commission on Radiation Protection exposure dose management system to all the studies and

organizations including the International Commission on Radiation Protection exposure dose management system to all the studies and Background & Purpose Background Purpose As Through radiation this exposure research, in the we medical intend field to increases systematically worldwide, record efforts and are under manage way the to

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

Metrology Laboratory of Ionizing Radiation

Metrology Laboratory of Ionizing Radiation Metrology Laboratory of Ionizing Radiation Carlos Oliveira The Metrology Laboratory of Ionizing Radiation (LMRI) has been involved in activities related with scientific, technical and legal metrology.

More information

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology Iran. J. Radiat. Res., 2006; 4 (1): 21-27 Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology M.T. Bahreyni

More information

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

Influence of the exposure time in the area monitors at radiodiagnostic

Influence of the exposure time in the area monitors at radiodiagnostic Journal of Physics: Conference Series PAPER OPEN ACCESS Influence of the exposure time in the area monitors at radiodiagnostic To cite this article: F Leyton et al 2016 J. Phys.: Conf. Ser. 733 012087

More information

Downloaded from jdm.tums.ac.ir at 9:49 IRDT on Wednesday March 27th 2019 *** ****

Downloaded from jdm.tums.ac.ir at 9:49 IRDT on Wednesday March 27th 2019 *** **** CB **** -*** -** -* ** *** **** itle: Comparison of radiation absorbed dose in target organs in maxillofacial imaging with panoramic, conventional linear tomography, cone beam computed tomography and computed

More information

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic

More information

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute

More information

The need for standardization of dosimetry in experimental radiation biology

The need for standardization of dosimetry in experimental radiation biology The need for standardization of dosimetry in experimental radiation biology Kurt Pedersen University of Wisconsin Medial Radiation Research Center Under the direction of Larry DeWerd, PhD. NCCAAPM Fall

More information

Radiation Protection Services in R.Macedonia

Radiation Protection Services in R.Macedonia Radiation Protection Services in R.Macedonia Lidija Nikolovska Institute of Public Health Ionizing Radiation Laboratory Skopje, Republic of Macedonia Background - no NPP or research reactors, - limited

More information

SUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT

SUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT SUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT The use of dental CBCT equipment must comply with all the regulations (IRR99 and IR(ME)R2000)

More information

Patient dosimetry for total body irradiation using single-use MOSFET detectors

Patient dosimetry for total body irradiation using single-use MOSFET detectors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Patient dosimetry for total body irradiation using single-use MOSFET detectors Tina Marie Briere, 1 Ramesh Tailor, 1 Naresh Tolani,

More information

Low Dose Excellent Image Quality Rapid Reconstruction

Low Dose Excellent Image Quality Rapid Reconstruction Low Dose Excellent Image Quality Rapid Reconstruction Efficient 3 in 1 Dental X-ray System CBCT > Precise 3-D Anatomical structures - Accurate diagnosis for doctors - Safe implant for patients > Significant

More information